Norlake NSHI331WSG, NSHI331WSW, NSHI331WSG/0, NSHI331WSG/5, NSHI331WSW/0 Specifications

...
NOR-LAKE SCIENTIFIC
A Division of
NOR-LAKE, INCORPORATED
727 Second Street
P.O. Box 248
Hudson, Wisconsin 54016
800-477-5253 715-386-2323
715-386-4290 FAX
www.norlake.com
SCIENTIFIC
®
NOR-LAKE SCIENTIFIC
L
ARGE CAPACITY INCUBATORS
Nor-Lake, Inc.
Registered to ISO 9001:2000
File No. A3204
Designed to meet the demanding requirements of scientific and laboratory applications for dry heat or optional CO
2
environments.
Standard Product Features:
Programmable logic microprocessor controller
with LCD display control panel with message center
Programmable ramp soak feature
Continuous product temperature display
Adjustable control range from +5°C above
ambient to 70°C on solid door models and +5°C above ambient to 60°C on glass door models
Air probe with air temperature Hi/Lo alarms
Hi/Lo product temperature alarms, audible &
visual
2 levels of user password protection (set points
and parameters)
100 event alarm logging including date/time
stamp and product temperatures. Stored in non volatile flash memory
Integrated Hi/Lo alarm tests
Door ajar alarm with adjustable delay
time period
Power failure alarm
Over temperature safety limit switch
Remote alarm contacts
Exterior front, sides and back are painted white
Models are available in glass or solid door
Full size door with key lock
Stainless steel interior
3 solid stainless steel shelves
Streamlined design optimizes mechanical
convection air flow and temperature uniformity
Temperature uniformity to +
.5°C @ 37°C
Leveling legs
Complete cabinet is foamed-in-place with CFC
Free high density polyurethane foam insulation
Warranties: 18 month parts and labor
(US and Canada) 18 months parts (International)
UL, C-UL, CE Mark
Optional Features:
CO2control
(IR sensor)
RH display
Duplex interior
electrical outlet, 4 amp load
Extra shelves
Access port 2” sleeve
with cover
Humidity pan-stainless
steel
RS485 computer
interface
Field Reversible hinge
door (solid door only)
Temperature chart
recorder
Chart paper
Drawers
NSHI331WSW/ NSHI331WSG/
SPECIFICATIONS
Glass Door Models NSHI331WSG/0 NSHI331WSG/5 Solid Door Models NSHI331WSW/0 NSHI331WSW/5
Crated Height (in) (cm) 87-3/8 (221.8) Crated Width (in) (cm) 35-1/4 (89.5) Crated Depth (in) (cm) 39 (99) Crated Weight (lbs) (kg) 420 (190.9) Interior Height (in) (cm) 67 (170.1) Interior Width (in) (cm) 27-3/4 (70.4) Interior Depth (in) (cm) 31 (78.7) Overall Height (in) (cm) 82-3/4 (210.1) Overall Width (in) (cm) 31-3/4 (80.6) Overall Depth (in) (cm) 35-7/8 (91.1) Gross Cubage (ft.3) (m3)33.1 (.93) Shelf Area (sq. ft.) (m2)14.1 (1.3) Number of Shelves 3 Total Amp Draw 8.7 4.7 Electrical Characteristics 115/60/1 230/50/1
NSHI331
NSHI331
Revision Date: 4/06 © 2006 NOR-LAKE, INC. Printed in the U.S.A. Part Number: 132585
N
O
LAKE
SCIENTIFIC
®
Side
Front
Voltage Model
Suffix Code
Voltage
Description
NEMA
Plug
NEMA
Receptacle
/0 115V, 1PH, 60HZ 5-15P 5-15R /5 230V, 1PH, 50HZ
Power Inlet
(IEC 60320)
Module
Cord Supplied
Locally
Specifications subject to change without notice.
CO
2
range: 0-20%
Inlet pressure: - 4 PSI to 25 PSI
Sensor: - Infrared (IR)
CO
2
control - Integrated with primary controller PLC based
Humidity pan: - 9 quarts, 20.75” long x 12.75” wide x 2.5” deep
22 gauge stainless steel
CO2inputs (2)
CO2tanks / customer supplied (2) recommended
CO2Output (1)
CO
2
stability +.1%
CO2auto shutoff upon door opening
CO2 gas regulator not supplied
Automatic CO2tank switch over
Ambient operating conditions: 15°C (59°F) to 40°C 104°F Ambient Humidity: 80% RH maximum non condensing Voltage tolerance: +