• do not apply the pulse oximeter on the
same arm as a blood pressure cuff,
arterial catheter or infusion line(s) (IVs)
• excessive light, such as sunlight or
direct home lighting
• moisture in the device
• improperly applied device
• finger is outside recommended size
range
• poor pulse quality
• venous pulsations
• anemia or low hemoglobin
concentrations
• cardiogreen and other intravascular
dyes
• carboxyhemoglobin
• methemoglobin
• dysfunctional hemoglobin
• artificial nails or fingernail polish
0123
Onyx® Vantage 9590 Finger Pulse Oximeter
Indications for Use
The Nonin® Onyx Vantage 9590 Finger Pulse Oximeter is a small, lightweight, portable device indicated for use in
measuring and displaying functional oxygen saturation of arterial hemoglobin (%SpO
well or poorly perfused, during both motion and non-motion conditions. It is intended for spot-checking of adult and
pediatric patients on fingers, including the thumb, that are between 0.3 – 1.0 inch (0.8 – 2.5 cm) thick. Alternatively, the
Onyx Vantage 9590 may be placed on a toe of a patient during non-motion conditions. The device’s intended use
environments include hospitals, clinics, long-term care facilities, skilled nursing facilities, emergency medical services, and
home healthcare services.
CAUTION: Federal law (USA) restricts this device to sale by or on the order of a licensed practitioner.
Contraindications
• Do not use the device in an MR environment, in an explosive atmosphere, or on infant or neonatal patients.
• This device is not defibrillation proof per IEC 60601-1 clause 8.5.5.
Warnings
• This device is intended only as an adjunct in patient assessment. It must be used in conjunction with other methods of
assessing clinical signs and symptoms.
• The device must be able to measure the pulse properly to obtain an accurate SpO2 measurement. Verify that nothing is
hindering the pulse measurement before relying on the SpO2 measurement.
• Operation of this device below the minimum amplitude of 0.3% modulation may cause inaccurate results.
• General operation of the device may be affected by the use of an electrosurgical unit (ESU).
• This device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the device
should be observed carefully to verify normal operation.
• Keep the oximeter away from young children. Small items such as the battery door, battery, and lanyard are choking hazards.
• Certain activities may pose a risk of injury, including strangulation, if lanyard should become wrapped around your neck.
Cautions
• This device has no audible alarms and is intended only for spot-checking.
• This device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin. Factors that may
degrade pulse oximeter performance or affect the accuracy of the measurement include the following:
) and pulse rate of patients, who are
2
• The device may not work when circulation is reduced. Warm or rub the finger, or re-position the device.
• This device’s display will go blank after 30 seconds of no readings or poor readings.
• This device has motion tolerant software that minimizes the likelihood of motion artifact; however, the device may still interpret
motion as good pulse quality.
• Clean the device before applying it to a patient.
• Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or spray any liquids onto the device.
• Do not use caustic or abrasive cleaning agents, or any cleaning agent containing ammonium chloride or isopropyl alcohol.
• A flexible circuit connects the two halves. Do not twist or pull the flexible circuit or overextend the device’s spring. Do not hang
the lanyard from the device’s flexible circuit.
• A functional tester cannot be used to assess the accuracy of a pulse oximeter monitor.
• This equipment complies with IEC 60601-1-2 for electromagnetic compatibility for medical electrical equipment and/or
systems. This standard is designed to provide reasonable protection against harmful interference in a typical medical
installation. However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical
noise in healthcare and other environments, it is possible that high levels of such interference due to close proximity or strength
of a source might disrupt the performance of this device. Medical electrical equipment needs special precautions regarding
EMC, and all equipment must be installed and put into service according to the EMC information specified in this manual.
• Portable and mobile RF communications equipment can affect medical electrical equipment.
• Batteries may leak or explode if used or disposed of improperly. Remove batteries if the device will be stored for more than
30 days. Do not use different types of batteries at the same time. Do not mix fully charged and partially charged batteries at
the same time. These actions may cause the batteries to leak.
• Follow local, state, and national governing ordinances and recycling instructions regarding disposal or recycling of the device
and device components, including batteries.
• In compliance with the European Directive on Waste Electrical and Electronic Equipment (WEEE) 2002/96/EC, do not dispose
of this product as unsorted municipal waste. This device contains WEEE materials; please contact your distributor regarding
take-back or recycling of the device. If you are unsure how to reach your distributor, please call Nonin for your distributor ’s
contact information.
Nonin Medical, Inc.
13700 1st Avenue North
Plymouth, Minnesota 55441-5443, USA
+1 (763) 553-9968 (outside US and Canada)
+46 650 401500 (Europe)
800-356-8874 (US and Canada)
Authorized Representative in the European Community
CE Marking indicating conformance to EC Directive No. 93/42/EEC concerning medical devices.
Type BF Applied Part (patient isolation from electrical shock)
Not for Continuous Monitoring (no alarm for SpO2)
SN
IP32
Installing Batteries
Two 1.5 volt AAA-size batteries power the 9590 for about 6,000 spot checks or 36 hours of operation. Nonin recommends
using alkaline batteries (included with each new device). When batteries are low, the numeric displays flash once per
second. Remove batteries if the device will be stored for more than 30 days. Replace low batteries as soon as possible,
using the instructions below.
NOTE: Rechargeable batteries may be used; however, they require more frequent replacement.
1. Hold the 9590 as shown in figure A. To release the device’s battery tray, press upward and then pull outward slightly
with the thumb.
2. Remove the old batteries from the battery tray. Dispose of the batteries properly.
3. Insert two new 1.5 volt AAA-size batteries. Follow the polarity markings (+ and -) as illustrated in figure B. Proper
positioning of the batteries is essential for operation.
4. Carefully guide the battery tray back onto the device. Press downward and then push inward slightly to re-secure the
battery tray (figure C). Do not force it into place; it fits only when properly positioned.
5. Insert your finger into the device to verify operation. See the Activating the Onyx Vantage 9590 and Verifying
Operation section for more information.
Serial Number
Battery orientation
UL Mark for Canada and the United States with respect to electric shock, fire, and mechanical hazards only
in accordance with UL 60601-1 and CAN/CSA-C22.2 No. 601.1.
Indicates separate collection for electrical and electronic equipment (WEEE).
Protected against vertically falling water drops when enclosure is tilted up to 15 degrees and ingress of
solid foreign objects greater than or equal to 2.5 mm (0.1 in.) diameter per IEC 60529.
RoHS compliant (China)
Latex Free
2
Page 3
Activating the Onyx Vantage 9590 and Verifying Operation
!
%SpO
2
Pulse Quality
Indicator
The device contains numeric Light-Emitting Diodes (LEDs) that display oxygen saturation and pulse rate. A tricolor LED
display (pulse quality indicator, shown at left) provides a visual indication of the pulse signal quality, while blinking at the
corresponding pulse rate. This display changes colors to alert you to changes in pulse quality that may affect the readings:
• Green indicates a good pulse signal.
• Yellow indicates a marginal pulse signal.
• Red indicates an inadequate pulse signal.
Activate the 9590 by inserting the patient’s finger into the device. The device detects the inserted finger and automatically
illuminates the displays. Correct positioning of the device on the finger is critical for accurate measurements.
NOTE: While on the finger, do not press the device against any surface and do not squeeze or hold it together. The internal
spring provides the correct pressure; additional pressure may cause inaccurate readings.
1. Insert the patient’s finger, nail side up, into the 9590 until the fingertip touches the built-in stop guide.
2. Make sure the finger is lying flat (not on its side) and is centered within the device. For best results, keep the device at
the patient’s heart or chest level.
3. If the device does not turn on, remove the finger and wait a few seconds before reinserting it.
When a finger is inserted, the device performs a brief startup sequence. Verify that all LEDs illuminate during the startup
sequence. If any LED is not lit, do not use the 9590; contact Nonin Customer Support for repair or replacement. After the
startup sequence, the device begins sensing the pulse (indicated by the blinking pulse quality indicator). Allow the device
to stabilize and observe about 4 seconds of continuous green-colored pulse quality before relying on the displayed values.
It is common for the displayed values to fluctuate slightly over a period of several seconds. If the pulse quality indicator
blinks yellow or red, try another finger.
A minus sign (-) appears in the left-most digit of the %SpO
The last measured SpO
and pulse rate values display for 10 seconds while the device automatically turns off. The device
2
will automatically shut off (to conserve battery life) approximately 10 seconds after the finger is removed, or after a 2minute period of inadequate pulse signals.
display when the device senses the finger has been removed.
2
Using the Lanyard
WARNING: Certain activities may pose a risk of injury, including strangulation, if lanyard should become
wrapped around your neck.
A lanyard is provided for convenience. The device will function with or without the lanyard.
If lanyard use is desired, thread the lanyard as shown below.
CAUTION: A flexible circuit connects the two halves. Do not twist or pull the flexible circuit or overextend the
device’s spring. Do not hang the lanyard from the device’s flexible circuit.
3
Page 4
Onyx Vantage 9590 Care, Maintenance, and Cleaning
CAUTIONS:
!
The advanced digital circuitry within the device requires no calibration or periodic maintenance other than battery
replacement. Field repair of the 9590 circuitry is not possible. Do not attempt to open the case or repair the electronics.
Opening the case will damage the device and void the warranty. Do not open the 9590 more than 90°, and do not twist or
pull on the device when cleaning.
Cleaning the Onyx Vantage 9590
• Clean the device before applying it to a patient.
• Do not sterilize, autoclave, or immerse this device in liquid. Do not pour or spray any liquids onto the device.
• Do not use caustic or abrasive cleaning agents, or any cleaning agent containing ammonium chloride or isopropyl
alcohol.
1. To clean, wipe the surfaces with a soft cloth dampened with mild detergent or a 10% bleach solution (household
bleach [5.25% sodium hypochlorite]). Do not use undiluted bleach or any cleaning solution other than those
recommended here, as permanent damage could result.
2. Dry with a soft cloth, or allow to air dry. Ensure that all surfaces are completely dry.
Testing Summary
SpO2 accuracy, motion, and low perfusion testing was conducted by Nonin Medical, Inc. as described below.
SpO2 Accuracy Testing During Motion and Non-motion
SpO2 accuracy testing is conducted during induced hypoxia studies on healthy, non-smoking, light-to-dark-skinned
subjects in an independent research laboratory. The measured arterial hemoglobin saturation value (SpO
is compared to arterial hemoglobin oxygen (SaO
The accuracy of the device is in comparison to the co-oximeter samples measured over the SpO
Accuracy data is calculated using the root-mean-squared (A
Specification for Pulse Oximeters for Accuracy.
) value, determined from blood samples with a laboratory co-oximeter.
2
range of 70-100%.
value) for all subjects, per ISO 9919, Standard
rms
2
) of the device
2
Pulse Rate Motion Testing
This test measures pulse rate oximeter accuracy with motion artifact simulation introduced by a pulse oximeter tester. This
test determines whether the oximeter meets the criteria of ISO 9919 for pulse rate during simulated movement, tremor,
and spike motions.
Low Perfusion Testing
This test uses an SpO2 Simulator to provide a simulated pulse rate, with adjustable amplitude settings of various SpO2
levels. The device must maintain accuracy in accordance with ISO 9919 for pulse rate and SpO
pulse amplitude (0.3% modulation).
at the lowest obtainable
2
4
Page 5
Specifications
This graph shows plots of the error (SpO2 – SaO2)
by SaO
2
using the 9590 with a linear regression fit
and upper 95% and lower 95% limits of agreement.
Each sample data point is identified by subject
from a clinical study in non-motion conditions.
This graph shows plots of the error (SpO2 – SaO2)
by SaO
2
using the 9590 with a linear regression fit
and upper 95% and lower 95% limits of agreement.
Each sample data point is identified by subject
from a clinical study in motion conditions.
This graph shows plots of the error (SpO2 – SaO2)
by SaO
2
using the 9590 with a linear regression fit
and upper 95% and lower 95% limits of agreement.
Each sample data point is identified by subject
from a clinical study using toes in non-motion
conditions.
Oxygen Saturation Display Range:0% to 100% SpO
Pulse Rate Display Range:18 to 321 beats per minute (BPM)
Declared Accuracy:
The tables below show A
values measured using the Onyx Vantage 9590 in a clinical study in motion and non-motion conditions.
rms
Accuracy Summary by Decade – Finger and Thumb
Decade
Oxygen Saturation
(A
)
rms
Low Perfusion
Oxygen Saturation (A
70 – 80%±2±2±3±3
80 – 90%±2±2±3±4
90 – 100%±2±2±2±3
70 – 100%±2±2±3±3
Accuracy Summary by Decade – Toe
DecadeOxygen Saturation (A
70 – 80%±2±2
80 – 90%±3±2
90 – 100%±3±2
70 – 100%±3±2
2
Motion Oxygen Saturation
)
rms
)
rms
Oxygen Saturation (A
(A
) (Finger)
rms
Low Perfusion
Motion Oxygen Saturation
(A
) (Thumb)
rms
)
rms
5
Page 6
Pulse Rate Declared Accuracy Range (A
Low Perfusion Pulse Rate Declared Accuracy Range (A
Motion Pulse Rate Declared Accuracy Range (A
*):20 to 250 BPM ±3 digits
rms
*):40 to 240 BPM ±3 digits
rms
*):40 to 240 BPM ±3 digits
rms
Measurement Wavelengths and Output Power**:
Red:
Infrared:
660 nanometers @ 0.8 mW maximum average
910 nanometers @ 1.2 mW maximum average
Temperature:
Operating:
Storage/Transportation:
-5 °C to 40 °C (23 °F to 104 °F)
-40 °C to 70 °C (-40 °F to 158 °F)
Humidity:
Operating:
Storage/Transportation:
10% to 90% non-condensing
10% to 95% non-condensing
Altitude:
Operating:
Hyperbaric Pressure:
Up to 12,192 meters (40,000 feet)
Up to 4 atmospheres
Battery Life:
Operating:
Approximately 6,000 spot checks or 36 hours of continuous
operation using new alkaline batteries.
Storage:
12 months
Classifications per IEC 60601-1 / CAN/CSA-C22.2 No. 601.1/ UL 60601-1:
Degree of Protection:
Enclosure Degree of Ingress Protection:
Mode of Operation:
Type BF-Applied Part
IP32
Continuous
This product complies with ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation and testing.
*± 1 A
represents approximately 68% of measurements.
rms
**This information is especially useful for clinicians performing photodynamic therapy.
6
Page 7
Warranty
NONIN MEDICAL, INCORPORATED, (Nonin) warrants to the purchaser, for a period of 4 years from the date
of purchase, each Onyx Vantage 9590 exclusive of the batteries, spring, lanyard, and lanyard lock.
Nonin shall repair or replace any Onyx Vantage 9590 found to be defective in accordance with this warranty,
free of charge, for which Nonin has been notified by the purchaser by serial number that there is a defect,
provided notification occurs within the applicable warranty period. This warranty shall be the sole and exclusive
remedy by the purchaser hereunder for any Onyx Vantage 9590 delivered to the purchaser which is found to be
defective in any manner whether such remedies be in contract, tort or by law.
This warranty excludes cost of delivery to and from Nonin. All repaired units shall be received by the purchaser
at Nonin’s place of business. Nonin reserves the right to charge a fee for a warranty repair request on any Onyx
Vantage 9590 found to be within specifications.
Onyx Vantage 9590 is a precision electronic instrument and must be repaired by trained Nonin personnel only.
Any sign or evidence of opening the Onyx Vantage 9590, field service by non-Nonin personnel, tampering, or
any kind of misuse of the Onyx Vantage 9590, shall void the warranty. All non-warranty work shall be done at
Nonin’s standard rates and charges in effect at the time of delivery to Nonin.
Nonin Medical, Inc.
13700 1st Avenue North
Plymouth, Minnesota 55441-5443 USA
(800) 356-8874 (USA and Canada)
+1 (763) 553-9968 (outside USA and Canada)
Fax: +1 (763) 553-7807
E-mail: technicalservice@nonin.com
www.nonin.com
Nonin Medical AB
Fibervägen 2
82450 Hudiksvall, Sweden
+46 650 401500 (Europe)
Fax: +46 650 401514
E-mail: serviceintl@nonin.se
www.nonin.com
Manufacturer’s Declaration
Refer to the following tables for specific information regarding this device’s compliance to IEC 60601-1-2.
This device is intended for use in the electromagnetic environment specified below.
The customer and/or user of this device should ensure that it is used in such an environment.
RF Emissions
CISPR 11
RF Emissions
CISPR 11
Harmonic Emissions
IEC 61000-3-2
Voltage Fluctuations/Flicker
Emissions
IEC 61000-3-3
Group 1This device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
Class BThis device is suitable for use in all establishments, including
domestic and those directly connected to the public low-
N/A
N/A
voltage power supply network that supplies buildings used for
domestic purposes.
7
Page 8
Table 2: Electromagnetic Immunity
Immunity TestIEC 60601 Test Level
This device is intended for use in the electromagnetic environment specified below.
The customer and/or user of this device should ensure that it is used in such an environment.
Electrostatic Discharge
(ESD)
±6 kV contact
±8 kV air
IEC 61000-4-2
Electrical Fast Transient/
Burst
±2 kV for power supply lines
±1 kV for input/output lines
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions, and voltage
variations on power supply
input lines
±1 kV differential mode
±2 kV common mode
±5% U
(>95% dip in UT) for 0.5 cycle
T
±40% U
±70% U
<5% U
(60% dip in UT) for 5 cycles
T
(30% dip in UT) for 25 cycles
T
(>95% dip in UT) for 5 sec.
T
IEC 61000-4-11
Power Frequency
3 A/m3 A/mPower frequency magnetic fields
(50/60 Hz) Magnetic Field
IEC 61000-4-8
NOTE: UT is the AC mains voltage before application of the test level.
Compliance
Level
±6 kV contact
±8 kV air
Electromagnetic Environment—
Guidance
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, relative humidity
should be at least 30%.
N/AMains power quality should be that of a
typical commercial or hospital
environment.
N/AMains power quality should be that of a
typical commercial or hospital
environment.
N/AMains power quality should be that of a
typical commercial or hospital
environment.
should be at levels characteristic of a
typical location in a typical commercial
or hospital environment.
8
Page 9
Table 3: Guidance and Manufacturer’s Declaration—Electromagnetic Immunity
d1.17 P=
d1.17 P=
80 MHz to 800 MHz
d2.33 P=800 MHz to 2.5 GHz
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment—Guidance
This device is intended for use in the electromagnetic environment specified below.
The customer and/or user of this device should ensure that it is used in such an environment.
Portable and mobile RF communications equipment should be used no closer to any part of the device, including cables, than
the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended Separation Distance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Radiated RF per
ISO 9919 clause 36
and ISO 80601-2-61
clause 202.6.2.3
3 V
rms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
20 V/m
80 MHz to 2.5 GHz
N/A
[3] V/m
[20] V/m
where P is the maximum output power rating of the transmitter in
watts (W) according to the transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey
level in each frequency range.
a
, should be less than the compliance
b
Interference may occur in the vicinity of equipment marked with the
following symbol:
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the device.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than [3] V/m.
NOTES:
• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.
9
Page 10
The following table details the recommended separation distances between portable and mobile RF communications equipment and this
d1.17 P=
d1.17 P=
d2.33 P=
device.
Table 4: Recommended Separation Distances
This device is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
Users of this device can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communication equipment (transmitters) and the device as recommended below, according to
maximum output power of the communications equipment.
Separation Distance According to Frequency of Transmitter
Rated Maximum Output
Power of Transmitter
W
0.010.120.120.23
0.10.370.370.74
11.21.22.3
103.73.77.4
100121223
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer.
NOTES:
• At 80 MHz and 800 MHz, the higher frequency range applies.
• These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects, and people.
150 kHz to 80 MHz
80 MHz to 800 MHz
800 MHz to 2.5 GHz
10
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.