Nonin GO2 Achieve User Manual

User Guide
7077-000-02 Cover.indd 1 11/18/2009 2:27:52 PM
i ii
ENGLISH
Introduction ....................................................................................................1
Contents of Package .......................................................................................1
Symbols ..........................................................................................................2
Intended Use ...................................................................................................4
Installing the Battery.......................................................................................8
Applying the GO2 Achieve to Your Finger ......................................................10
Activating the Device and Verifying Operation ..............................................12
Reading Your Results ....................................................................................13
Care and Maintenance ..................................................................................15
Specifications ................................................................................................20
Equipment Response Time ............................................................................23
Testing Summary ..........................................................................................24
SpO2 Accuracy Testing ...........................................................................24
Pulse Rate Motion Testing ......................................................................24
Low Perfusion Testing .............................................................................24
Warranty .......................................................................................................25
Table of Contents
7077-000-02.indd 1-2 11/18/2009 2:20:13 PM
1 2
ENGLISH
Introduction
Thank you for buying the GO2 Achieve Fingertip Pulse Oximeter. This small, portable device will give you important information about your oxygen saturation (the amount of oxygen in your blood) and pulse rate at your fingertip! The GO2 Achieve is easy to use and needs no routine maintenance except battery replacement. This User Guide explains how to use and care for your GO2 Achieve.
Contents of Package
 • GO2 Achieve
 • One AAA Alkaline Battery
 • User Guide
 • Quick Guide
Symbols
The following symbols are associated with your GO2 Achieve.
Symbol Definition of Symbol
Consult Instructions for Use
!
Caution!
0123
CE Marking: conformance to EC Directive No. 93/42/EEC for medical devices
Type BF Applied Part (patient isolation from electrical shock).
Not for Continuous Monitoring (no alarm for SpO2)
UL Mark for Canada and the United States with respect to electric shock, fire, and mechanical hazards only in accordance with IEC 60601-1, UL 60601-1 and CAN/CSA-C22.2 No. 601.1.
7077-000-02.indd 1-2 11/18/2009 2:20:14 PM
3 4
ENGLISH
Battery
Indicates separate collection for electrical and electronic equipment (WEEE).
IP33
Protected against spraying water and against access to hazardous parts with a tool, per IEC 60529.
SN
Serial Number
Sensor disconnect; the pulse signal is not detected or there is excessive motion.
EC
REP
Authorized Representative in the European Community
+70°C
-30°C
Temperature Limitation for storage/shipping
Intended Use
GO2 Achieve is intended to measure blood oxygen saturation (%SpO2) (the amount of oxygen in your blood) and pulse rate of both adults and children. It is designed for fingers (not the thumb) between 0.3 and 1.0 inch (0.8 – 2.5cm) thick. The index finger (pointer finger) is most recommended.
Warnings
• Keep the oximeter away from young children. Small items such as the battery door,
battery, and lanyard are choking hazards.
• Certain activities may pose a risk of injury, including strangulation, if the lanyard
should become wrapped around your neck. Use the lanyard with caution.
7077-000-02.indd 3-4 11/18/2009 2:20:16 PM
5 6
ENGLISH
Cautions
• The GO2 Achieve might misinterpret excessive movement as good pulse strength. Limit
finger movement as much as possible when using the device.
• The GO2 Achieve must be able to measure your pulse properly to give you an accurate
reading. Do not put the device on the same hand/arm when using a blood pressure cuff or monitor.
• The GO2 Achieve has no alarms. It will not sound if the amount of oxygen in your blood
is low or if your pulse rate is too high or too low.
• Do not place the GO2 Achieve in liquid or clean it with agents containing ammonium
chloride, isopropyl alcohol, or products that are not listed in this User’s Guide.
• Any of the following conditions may reduce the performance of the GO2 Achieve:
– flickering or very bright light; – weak pulse quality (low perfusion); – low hemoglobin; – arterial catheters; – nail polish, and/or artificial nails; and – any tests recently performed on you that required an injection of intravascular dyes.
• The GO2 Achieve may not work if you have poor circulation. Rub your finger to
increase circulation, or place the device on another finger.
• The GO2 Achieve measures oxygen saturation of functional hemoglobin. High levels
of dysfunctional hemoglobin (caused by sickle cell anemia, carbon monoxide, etc.) could affect the accuracy of the measurements.
• Batteries can leak or explode if used or disposed of improperly. Remove the battery if
the GO2 Achieve will be stored for more than 30 days.
• Do not use the GO2 Achieve in a combustible environment (oxygen enriched
environment).
• Do not use the GO2 Achieve outside the specified operating and storage temperature
ranges.
• Do not use the GO2 Achieve for more than 30 minutes without relocating the device
to another finger.
• The GO2 Achieve needs to be used according to information provided in the User
Guide.
• Do not tamper with, or hang lanyard from the flexible circuit.
!
7077-000-02.indd 5-6 11/18/2009 2:20:16 PM
7 8
ENGLISH
• Radios and cell phones or similar devices may affect the GO2 Achieve and should be kept
at least 7 feet away from the device.
• Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast towers and TV broadcast towers may affect accuracy.
• Use in emergency vehicles with communication systems may affect accuracy.
• Functional tester cannot be used to assess the accuracy of this pulse oximeter.
• Follow local disposal and recycling laws for the GO2 Achieve and its components,
including the battery.
• The GO2 Achieve is a precision electronic instrument and must be repaired by Nonin
Technical Service.
Installing the Battery
One 1.5 volt AAA-size (LR03) battery powers the GO2 Achieve for approximately 2400 measurements. Nonin recommends using alkaline batteries (one is included with each new GO2 Achieve). NOTE: You may use rechargeable batteries; however, they may require more frequent replacement.
1. Remove the battery door located on the left side of the GO2 Achieve by sliding it towards you.
2. Insert one new 1.5 volt AAA-size battery. Follow the plus (+) and minus (-) markings for battery direction (as shown inside of the battery compartment).
GO
2
7077-000-02.indd 7-8 11/18/2009 2:20:17 PM