Nonin 3150 Operator's Manual

Page 1
Operator’s Manual
0123
WristOx
Wrist-Worn Pulse Oximeter
Model 3150
Pulse Oximeter
2
®
Wireless Technology
English
Page 2
CAUTION: Federal law (USA) restricts this device to sale by or on the order of a
EC
REP
licensed practitioner.
Consult Instructions for Use.
Nonin® reserves the right to make changes and improvements to this manual and the products it describes at any time, without notice or obligation.
Nonin Medical, Inc.
13700 1st Avenue North
Plymouth, MN 55441-5443
USA
+ 1 (763) 553-9968
800-356-8874 (USA and Canada)
Fax: + 1 (763) 553-7807 e-mail: mail@nonin.com
www.nonin.com
0123
MPS, Medical Product Service GmbH
Borngasse 20
D-35619 Braunfels, Germany
References to “Nonin” in this man Nonin, WristOx
Medical, Inc. The Bluetooth of such marks by Nonin Medical, Inc. is under license. Other trademarks and trade names are those of their respective owners.
, PureLight, and nVISION are registered trademarks or trademarks of Nonin
2
word mark and logo are owned by the Bluetooth SIG, Inc. and any use
ual imply Nonin Medical, Inc.
© 2010 Nonin Medical, Inc.
7622-001-01
Page 3
Contents
Indications for Use.................................................................................................... 1
Contraindications....................................................................................................1
Warnings ................................................................................................................1
Cautions .................................................................................................................2
Declaration of Conformity with FCC and Canadian Ministry
of Health Rules for Electromagnetic Compatibility..............................................3
Federal Communications Commission (FCC) Notice............................................. 3
Guide to Symbols...................................................................................................... 4
Displays, Controls, and Indicators.......................................................................... 5
Introduction ...............................................................................................................8
Unpacking the WristOx
Standard Kit....................................................................................................... 8
Starter Kit .......................................................................................................... 8
Batteries ................................................................................................................ 9
Bluetooth Technology............................................................................................ 9
Operation Modes.....................................................................................................10
Cable ................................................................................................................... 10
Standby .............................................................................................................. 10
On........................................................................................................................ 10
Spot Check Mode ............................................................................................ 10
Sensor Activation Mode................................................................................... 11
Programmed Mode.......................................................................................... 11
Using the WristOx
, Model 3150............................................................................ 12
2
Installing Batteries............................................................................................... 12
Wrist Straps......................................................................................................... 13
Ring Strap Description..................................................................................... 13
Attach Ring Strap to Device............................................................................. 14
Adjustable Strap Description............................................................................ 15
Attach Adjustable Strap to Device ................................................................... 15
Attaching the Sensor........................................................................................... 18
Patient Application............................................................................................... 19
Two-Piece Wrist Band Application................................................................... 19
Single Strap Application................................................................................... 21
Verifying Operation.............................................................................................. 24
Startup Sequence and Self-Test.......................................................................24
Activation Switch ................................................................................................. 25
Activate Bluetooth Radio.................................................................................. 25
Activate Device ........................................................................................ ........ 25
Error Codes......................................................................................................... 25
Troubleshooting..................................................................................................... 26
Care and Maintenance............................................................................................ 28
Cleaning the Device ............................................................................................ 28
Cleaning the Sensor............................................................................................ 28
Cleaning the Wrist Band...................................................................................... 28
Storing................................................................................................................. 28
, Model 3150 ....................................................................8
2
i
Page 4
Contents (Continued)
Memory and Data.....................................................................................................29
nVISION Software ....................................................................................................30
nVISION Settings................................................................................................. 30
Accessing nVISION Settings............................................................................ 30
Cable Connection................................................................................................. 31
USB Driver Installation (XP) ............................................................................. 32
USB Driver Installation (Vista).......................................................................... 32
USB Driver Installation (Windows 7)................................................................. 33
Bluetooth Connection........................................................................................... 33
Parts and Accessories ........................................................................................... 35
Service, Support, and Warranty ............................................................................ 36
Service and Support............................................................................................. 36
Warranty............................................................................................................... 36
Technical Information............................................................................................. 37
Manufacturer’s Declaration................................................................................ 37
Equipment Response Time................................................................................ 40
Testing Summary............................................................................................... 41
SpO
Accuracy Testing .................................................................................. 41
2
Pulse Rate Motion Testing.............................................................................. 41
Low Perfusion Testing.................................................................................... 41
Specifications..................................................................................................... 42
Oximeter Specifications.................................................................................. 42
System Specifications..................................................................................... 43
Transmitter ..................................................................................................... 44
ii
Page 5
Figures
Figure 1. Front Display (Startup Screen).............................................................. 5
Figure 2. Comparison of Full and Partial Display................................................11
Figure 3. Remove Battery Door.......................................................................... 12
Figure 4. Insert Batteries.................................................................................... 12
Figure 5. Ring Strap........................................................................................... 13
Figure 6. Thread Ring Strap............................................................................... 14
Figure 7. Attach Ring Strap................................................................................ 14
Figure 8. Adjustable Strap.................................................................................. 15
Figure 9. Thread Adjustable Wrist Strap............................................................ 16
Figure 10. Attach Adjustable Wrist Strap ........................................................... 16
Figure 11. Device with Wrist Straps Attached (Front and Back Views).............. 17
Figure 12. Attach Sensor.................................................................................... 18
Figure 13. Verify Two-Piece Wrist Band Attachment......................................... 19
Figure 14. Thread and Tighten Two-Piece Wrist Band...................................... 20
Figure 15. Fasten Two-Piece Wrist Band........................................................... 20
Figure 16. Apply Sensor to Patient .................................................................... 21
Figure 17. Verify Single Strap Attachment......................................................... 21
Figure 18. Thread Single Strap ......................................................................... 22
Figure 19. Tighten Single Strap.......................................................................... 22
Figure 20. Fasten Single Strap........................................................................... 23
Figure 21. Apply Sensor to Patient..................................................................... 23
Figure 22. nVISION Settings Window................................................................. 31
Tables
Table 1. Labeling Symbols.................................................................................. 4
Table 2. Error Codes......................................................................................... 25
Table 3. Electromagnetic Emissions................................................................. 37
Table 4. Electromagnetic Immunity................................................................... 38
Table 5. Guidance and Manufacturer’s Declaration—
Electromagnetic Immunity ................................................................39
Table 6. Recommended Separation Distances................................................. 40
iii
Page 6
Indications for Use
Indications for Use
The Nonin Model 3150, WristOx
Pulse Oximeter is a small, wrist-worn device indicated for
2
use in measuring, displaying, and storing functional oxygen saturation of arterial hemoglobin (%SpO
) and pulse rate of adult and pediatric patients. It is intended for spot-checking and/or
2
data collection and recording of patients during both no motion and motion conditions, and for patients who are well or poorly perfused. The intended use environments are sleep and pulmonary rehab labs, surgical recovery, critical care, emergency room, long-term care, home use, and mobile units.
Contraindications
Do not use this device in a Magnetic Resonance (MR) environment or in the presence of flammable anesthetics or gases.
This device is not defibrillation proof per IEC 60601-1:1988/A2:1995 clause 17h.
Warnings
This device is intended only as an adjunct device in patient assessment. It must be used in conjunction with other methods of assessing clinical signs and symptoms.
Carefully route patient cables and connections to reduce the possibility of patient entanglement or strangulation.
®
Use only Nonin-branded PureLight accuracy specifications for Nonin pulse oximeters. Using other manufacturers’ sensors can result in improper pulse oximeter performance.
pulse oximeter sensors. These sensors are manufactured to meet the
This device is a precision electronic instrument and must be repaired by qualified technical professionals. Field repair of this device is not possible. Except to replace batteries, do not attempt to open the case or repair the electronics. Opening the case may damage the device and void the warranty.
The USB cable must be unplugged from the device before replacing batteries. This device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary,
the device should be observed carefully to verify normal operation. The use of accessories, sensors, and cables other than those listed in this manual may result in increased
electromagnetic emission and/or decreased immunity of this device. Do not use the device when alarms are required.
Do not use a damaged sensor. This equipment complies with International IEC 60601-1-2:2007 for electromagnetic compatibility for medical
electrical equipment and/or systems. This standard is designed to provide reasonable protection against harmful interference in a typical medical installation. However, because of the proliferation of radio-frequency transmitting equipment and other sources of electrical noise in healthcare and other environments, it is possible that high levels of such interference due to close proximity or strength of a source might disrupt the performance of this device. Medical electrical equipment needs special precautions regarding EMC, and all equipment must be installed and put into service according to the EMC information specified in this manual.
Only use Nonin-branded sensors with a length of 1 meter or less. Accuracy may degrade if sensor cable is over 1 meter in length. Using the sensor cable adapter does not affect accuracy.
1
Page 7
Indications for Use
Cautions
If this device fails to respond as described, refer to “Troubleshooting” or discontinue use until the situation has been corrected. Contact Nonin Technical Service.
This device has motion tolerant software that minimizes the likelihood of motion artifact being misinterpreted as good pulse quality. In some circumstances, however, the device may still interpret motion as good pulse quality.
Check the pulse oximeter sensor application site every 6-8 hours to determine the positioning of the sensor and the circulation and skin sensitivity of the patient. Patient sensitivity varies depending on medical status or skin condition.
Do not place liquids on top of this device. Do not place the WristOx
alcohol. Refer to the “Care and Maintenance” section of this operator’s manual. Use a detergent that is safe for skin and washable surfaces. Most detergents can be high sudsing, so use sparingly.
Wipe with a damp, detergent free cloth to remove residue. Follow local, state, and national governing ordinances and recycling instructions regarding disposal or recycling of
the device and device components, including batteries. In compliance with the European Directive on Waste Electrical and Electronic Equipment (WEEE) 2002/96/EC, do
not dispose of this product as unsorted municipal waste. This device contains WEEE materials; please contact your distributor regarding take-back or recycling of the device. If you are unsure how to reach your distributor, please call Nonin for your distributor’s contact information.
This device is designed to determine the percentage of arterial oxygen saturation of functional hemoglobin. Factors that may degrade pulse oximeter performance or affect the accuracy of the measurement include the following:
• excessive ambient light
• excessive motion
• electrosurgical interference
• blood flow restrictors (arterial catheters, blood pressure cuffs, infusion lines, etc.)
• moisture in the sensor
• improperly applied sensor
• incorrect sensor type
Do not perform any testing or maintenance on this device while it is being used to monitor a patient. Verify all visible indicators appear during the start-up (initialization) sequence. If any indicator does not appear, do
not use the device. Contact Nonin Technical Service for assistance.
, Model 3150, in liquid or clean it with agents containing ammonium chloride or isopropyl
2
• poor pulse quality
• venous pulsations
• anemia or low hemoglobin concentrations
• cardiogreen and other intravascular dyes
• carboxyhemoglobin
• methemoglobin
• dysfunctional hemoglobin
• artificial nails or fingernail polish
Portable and mobile RF communications equipment can affect medical electrical equipment. Batteries may leak or explode if used or disposed of improperly. Remove batteries if the device will be stored for
more than 30 days. Do not use different types of batteries at the same time. Do not mix fully charged and partially charged batteries at the same time. These actions may cause the batterie s to le ak.
To avoid the risk of confusing or misinterpreting patient data when transmitting data via Bluetooth, verify the device is paired with the correct display unit.
The pulse oximeter may not work when circulation is reduced. Warm or rub the finger or reposition the sensor. A functional tester cannot be used to assess the accuracy of the oximeter or sensor. Do not fasten the device too tightly around the patient’s wrist. Inaccurate readings and patient discomfort could
result.
2
Page 8
Indications for Use
Declaration of Conformity with FCC and Canadian Ministry of Health Rules for Electromagnetic Compatibility
• Nonin Medical, Inc., of 13700 1st Avenue North, Plymouth, Minnesota, 55441, declares under its sole responsibility that Model 3150, WristOx declaration relates, complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
• Ministry of Health (Canada), Safety Code 6: standards include a substantial safety margin designed to ensure the safety of all persons, regardless of age and health. The exposure standard for wireless mobile phones employs a unit of measurement known as the Specific Absorption Rate, or SAR. The SAR limit set by the FCC is 1.6 W/kg.
Pulse Oximeter, to which this
2
Federal Communications Commission (FCC) Notice
This device has been tested and found to comply with the limits for a class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This device generates, uses, and can radiate radio frequency energy. If not installed and used in accordance with the instructions, it may cause harmful interference to radio or television reception, which can be determined by turning the device off and on. The user is encouraged to try to correct the interference by one or more of the following measures:
• Reorient or relocate the receiving antenna.
• Increase the distance between the device and the receiver.
• Connect the device to an outlet on a circuit different from the outlet where the receiver is connected
• Consult the dealer or an experienced radio/TV technician for assistance.
• RF Exposure: For body worn operation, to maintain compliance with FCC RF exposu re guidelines, use only accessories that contain no metallic components. Use of other accessories may violate FCC RF exposure guidelines and should be avoided.
• The WristOx2, Model 3150, is designed and manufactured not to exceed the emission limits for exposure to radio frequency (RF) energy set by the United States FCC. These limits are part of comprehensive guidelines and establish permitted levels of RF energy for the general population. The guidelines are based on the safety standards previously set by both U.S. and international standards bodies. This device has been shown to be compliant for localized specific absorption rate (SAR) for uncontrolled environment/ general population exposure limits specified in ANSI/IEEE Std. C95.1-2005.
• The FCC requires the user to be notified that any changes or modifications to this device that are not expressly approved by Nonin Medical, Inc. may void the user’s authority to operate the device.
3
Page 9
Guide to Symbols
!
EC
REP
0123
Guide to Symbols
This chapter describes the symbols that are found in this manual and on the WristOx2, Model 3150. Detailed information about display symbols can be found in “Displays, Controls, and Indicators.”
Table 1: Labeling Symbols
Symbol Description
Caution!
Consult Instructions for Use.
Authorized Representative in the European Commun ity.
CE Marking indicating conformance to EC dir concerning medical devices.
Type BF-Applied Part (patient isolation from electrical shock)
No alarms
Indicates separate collection for electrical and electronic equipment
Continua Certified™ sign be interoperable with other products that carry the Continua Ce rtified symbol.
®
Bluetooth
Non-ionizing electromagnetic radiation. Equipment includes RF transmitters. In
terference may occur in the vicinity of equipment marked with this symbol.
UL Mark for Canada and the United S tates w and mechanical hazards only in accordance with UL 60601-1 and CAN/CSA C22.2 No. 601.1.
®
figure mark
ifies that this product has been tested and proven to
ective No. 93/42/EEC
ith respect to electric shock, fire,
(WEEE).
IP33
4
Protected against spraying water and against access to hazardous p arts with a tool, per IEC 60529.
Page 10
Activation Switch
Sensor Port
Displays, Controls, and Indicators
Displays, Controls, and Indicators
Figure 1: Front Display (Startup Screen)
%SpO2 Display
This 3-digit display, located in the upper left corner of the LCD, shows percent blood oxygen saturation (%SpO from 0 to 100 %.
This display also shows the month, year, and hour (24-hour clock format) during startup.
). The range is
2
Pulse Rate Display
This 3-digit display, located below the %SpO2 display, shows the pulse rate in beats per minute (BPM). The range is from 18 to 32 1 BPM.
This display also shows the day and minute during startup.
5
Page 11
Full
Half
Critical
Low
Displays, Controls, and Indicators
Activation Switch
This switch is located next to the sensor port. Pressing this switch activates the Bluetooth radio for 3 minutes. It can also be used to turn the device on when it is in Standby mode.
See “Activation Switch” section for more information.
Sensor Fault Indicator
This indicator displays if the device determines a sensor fault exists (e.g., sensor disconnect, misalignment, or incompatibility with the device). It also displays when the finger is removed from the sensor.
Pulse Strength Indicator
This heart-shaped indicator is followed by up to nine curved bars. The heart shape is always visible, and the number of curved bars on the display depends on the pulse strength as determin ed by the oximeter.
Poor Pulse Signal Indicator
This indicator displays when the pulse signal is inadequate or the device does not sense a pulse. It may also display if there is excessive motion at the sensor site.
Battery Indicator
This indicator shows remaining battery life as eithe r full, half, low, and critical (as shown at left).
Replace the batteries when device reaches low state. When the battery reaches critical state, all indicators clear from the
display except for the blinking critical battery indicator, the current session closes, and the Bluetooth radio shuts down.
6
Page 12
Bluetooth
indicator
Bluetooth indicator with animated bars
Displays, Controls, and Indicators
Bluetooth Indicator
This indicator displays when the Bluetooth radio is on. It appears as either the Bluetooth logo or the Bluetooth logo with animated bars.
This indicator displays for the first 2 minutes the device is on. If a master device does not connect to the device in those 2 minutes, the Bluetooth radio shuts down and the icon no longer displays. When the device is connected to a master device, the indicator displays with animated bars.
If the Bluetooth radio is on when the device enters Standby mode or connects to the USB interface cable, the Bluetooth indicator appears on the LCD while the Bluetooth radio shuts down. It will be the only indicator on the LCD and will display for up to 10 seconds.
SmartPoint Indicator
This indicator displays during the startup sequence.
7
Page 13
Introduction
Introduction
The Bluetooth-enabled WristOx2, Model 3150, is a small, wrist-worn device that displays, measures, and stores patient SpO radio with a range (spherical radius) of approximately 100 meters (328 feet).
The device ships ready to use in Spot Check turn on mode. In Spot Check turn on mode, inserting a finger in the sensor automatically turns the device on. Approximately 10 seconds after the finger is removed, the device enters Standby mode.
Advanced memory and programming features are available with Nonin’s nVISION® software (version 6.3 or greater). See the “nVISION Software” section to learn more about using the device with nVISION.
Unpacking the WristOx2, Model 3150
The WristOx2, Model 3150, standard or starter kit includes the items listed below . Once the shipping carton is unpacked, verify these items were received. Contact the carrier immediately if the shipping carton is damaged.
and pulse rate data. The device includes a Bluetooth
2
Standard Kit
• Model 3150, WristOx2 Pulse Oximeter
• Model 8000SM-WO2, reusable soft sensor
• 3 two-piece wrist bands, 1 each of the following:
• 6 in. (15 cm)
• 8 in. (20 cm)
• 10 in. (25 cm)
• 2 AAA (1.5 volt) alkaline batteries
• Operator’s manual (CD)
• USB driver software (on operator’s manual CD) – required to use the PC USB interface cable
Starter Kit
A starter kit is required to configure the device and download data to a PC. The starter kit consists of the standard kit, plus:
• 9 pack of two-piece wrist bands, 3 each of the following:
• 6 in. (15 cm)
• 8 in. (20 cm)
• 10 in. (25 cm)
• nVISION SpO2 data management software (CD)
• PC USB interface cable
8
Page 14
Introduction
Batteries
The device uses 2 AAA alkaline batteries. With new alkaline batteries, battery life is approximately 48 hours (minimum) when not
connected to a Bluetooth device. When connected to a Bluetooth device, battery life will vary depending on class of operation. See “Specifications” for detailed battery life information.
The battery indicator shows one of four states: full, half, low, and critical. Replace the batteries when device reaches low state. A low battery has a minimum of 10 minutes before it reaches critical state. Actual battery life depends on Bluetooth radio use. In critical battery mode, the battery indicator blinks and the device no longer monitors or records patient data.
When batteries are removed, the device maintains the time and da te for up to 30 seconds. If replacing batteries takes more than 30 seconds, or if the battery level is at or below the critical level, clock settings are lost and the device reverts to Spot Check mode. Use nVISION software to reset the clock and change the operation mode.
Remove the batteries and disconnect the sensor if the device is to be stored for more than 1 month. In storage, battery life is approximately 9 months.
NOTES:
• This device contains non-volatile memory. Removing or replacing batteries does not affect the data stored in memory. Stored data remains in memory until overwritten by newer data or cleared from memory with nVISION software (version 6.3 or greater).
• If batteries are replaced while recording data, the session will terminate and some data from the session may not be saved. The terminated session will be time stamped with the current date/time the next time the device turns on.
• If clock settings are lost, the date and time restarts at 01:01:10:00:00.
Bluetooth Technology
Bluetooth technology allows wireless connections between electronic communications and computing devices. The technology is based on a radio link that offers fast and reliable data transmissions. Bluetooth uses a license-free, globally available frequency range in the ISM band—intended to ensure communication compatibility worldwide.
Nonin’s use of Bluetooth wireless technology allows SpO2 and pulse rate data to be transmitted through a Bluetooth radio to a compatible Bluetooth-enabled device. Nonin’s wireless system removes the cable connection from the device, giving patients increased ability to move freely.
To make efficient use of battery life, Nonin’s WristOx2, Model 3150, uses an automatically switchable Class 1/Class 2 Bluetooth radio with a maximum range (spherical radius) of about 100 meters (328 feet). Obstacles and other conditions may affect range, and class of operation and connection mode will impact battery life. See “S pecifications” fo r detailed battery life information.
9
Page 15
Operation Modes
Operation Modes
The WristOx2, Model 3150, has three states: Cable, Standby, and On.
Cable
The device is in Cable mode when it is connected to a PC using the USB interface cable. While in Cable mode, the device does not collect or save data and the Bluetooth radio is off.
Standby
When the device is in S tandby mode, the screen is blank and the device appears to be of f. In Standby, it is ready for a signal that will turn the device on (e.g., pressing activation switch, inserting finger in sensor [S pot Check mode], connecting sensor [Sensor Activation mode], or programmed start time [Programmed mode]). While in Standby mode, the device does not collect or save data and the Bluetooth radio is off.
On
When the device is on, it can collect and save data. The device features three turn on modes:
• Spot Check mode
• Sensor Activation mode
• Programmed mode The device is delivered in Spot Check mode . nV ISION software (version 6.3 or higher) is
needed to access the device settings and change Spo t Check mode to Sensor Activa tion or Programmed mode (see “nVISION Software”).
The device recalls the active settings when the device is shut off and turned on again.
Spot Check Mode
Spot Check mode is the default turn on operation mode. The device automatically turns on when a finger is inserted into the sensor. It enters
Standby mode 10 seconds after the finger is removed. If the sensor is disconnected, the device enters Standby mode immediately.
In this mode, the sensor can be left connected to the device.
NOTE: If the device determines that a sensor fault exists (a sensor failure, misalignment, or incompatibility with the device) or if a pulse oximeter sensor signal cannot be detected, the device enters Standby mode after 3 minutes.
10
Page 16
Operation Modes
Full Display
Partial Display
Sensor Activation Mode
Sensor Activation mode may be selected through nVISION software. In this mode, the device turns on when the activation switch is pressed or when the sensor is disconnected and reconnected. This mode is useful when using a sensor that is not easily remove d from the sensor site (e.g., disposable or wrap sensor).
NOTE: The sensor doe
If the sensor is not used for at least 10 minutes or if an
s not need to be applied to a finger to turn the device on.
inadequate pulse signal is detected, the device automatically enters Standby mode. To turn the device on again, press the activation switch or disconnect and reconnect the sensor.
This mode allows for Full or Partial display (see figure 2 for display comparison). When using Partial d
isplay, the SpO
and pulse rate readings do not display. The user will only
2
see the battery indicator and the animated pulse strength indicator.
Figure 2: Comparison of Full and Partial Display
Programmed Mode
Programmed mode may be selected and setup through nVISION software. With the software, the user can program the device to start and stop for up to three sessions. Once programmed, the next start time displays on the LCD every 30 seconds in HH:MM format.
A sensor must be connected for Programmed mode to function. If the programmed device is in St andby mode and the a
ctivation switch is pressed, the user activates the Bluetooth radio and the device for 3 minutes. During this time, the user is able to take and store measurements. After 3 minutes, the device returns to Standby mode.
This mode allows for Full or Partial display (see figure 2 above for display comparison). When
using Partial display , the SpO
and pulse rate readings do not display. The user will
2
only see the battery indicator and the animated pulse strength indicator.
NOTE: A programmed device reverts to Spot Check mode if the clock is not set or if the clock settings are lost when replacing the batteries.
11
Page 17
Using the WristOx2, Model 3150
Using the WristOx2, Model 3150
WARNING: Do not use the device when alarms are required.
W
ARNING: The USB cable must be unplugged from the device before replacing
batteries.
Installing Batteries
1. Open the battery compartment by sliding the battery door off the back of the device (figure 3).
Figure 3: Remove Battery Door
2. Insert 2 new AAA alkaline batteries (figure 4). Battery orient inside the battery compartment.
3. Replace battery door by sliding it back into place.
4. Inserting batteries does not turn the device on. In Spot Check mode, the device turns on when a finger is inserted in the sensor.
NOTE: When batteries are removed, the device maintains the time and date for up to 30 seconds. If replacing batteries takes more than 30 seconds, or if the battery level is at or below the critical level, the operation mode reverts to Spot Check mode. Use nVISION software to reset the clock and change the operation mode.
Figure 4: Insert Batteries
ation is shown
12
Page 18
Using the WristOx2, Model 3150
Attach this end to strap bar by sensor port
Fastener
Hook and Loop Side
Woven Side
Wrist Straps
The WristOx2, Model 3150, is designed to be applied to the patient’s wrist using a two­piece wrist band. The two-piece wrist band consists of a ring strap and an adjustable strap.
For pediatrics and petite adults, the device may b
e applied using a single adjustable strap. The single adjustable strap is the longer of the two straps in the two-piece wrist band package.
Straps have hook and loop fasteners on one side. The other side has a smooth, woven fabric. Wh
en the straps are attached to the device, the woven side should touch the
patient’s skin.
NOTE: A wrist strap has a maximum of 10 uses before replacement is required.
This section contains descriptions of the straps and instructions for atta ching the straps to the de
vice. See the “Patient Application” section for instructions on how to apply the device
to the patient.
Ring Strap Description
One end of the ring strap has a plastic ring. The square end has a hook and loop faste ner (figure 5).
The square end attaches to the strap bar at the top of the device, by the sensor port (see “Attach Ring Strap to Device”).
Figure 5: Ring Strap
13
Page 19
Using the WristOx2, Model 3150
Strap bar at top of device (by sensor port)
Attach Ring Strap to Device
1. Place device face down.
2. Locate strap bar at top of device (by sensor port).
3. Position ring strap so hook and loop side of strap faces up.
4. Thread square end fastener through the strap bar by the sensor port. Start from the back of the device and thread it towards the front of the device (figure 6).
.
5. Fold strap so fastener adheres to the strap (figure 7).
Figure 6: Thread Ring Strap
Figure 7: Attach Ring Strap
6. Continue to “Attach Adjustable Strap to Device.”
14
Page 20
Using the WristOx2, Model 3150
Hook and Loop Side
Woven Side
Attach square end to strap bar by battery door
Rounded End Fastener
Square End Fastener
Adjustable Strap Description
This strap is part of the two-piece wrist band and may also be used to apply the device using a single strap.
The adjustable strap is available in various length
s (see “Parts and Accessories”). It has
hook and loop fasteners at both ends (figure 8). The square end attaches to the strap bar at the bottom of the device, by the battery door
“Attach Adjustable Strap to Device” section).
(see
Attach Adjustable Strap to Device
1. Place device face down.
2. Locate strap bar at bottom of device (by battery door).
3. Position adjustable strap so hook and loop side of strap faces up.
Figure 8: Adjustable Strap
15
Page 21
Using the WristOx2, Model 3150
Strap bar by battery door
4. Thread square end fastener through the strap bar by the battery door. Start from the back of the device and thread it towards the front of the device (figure 9).
Figure 9: Thread Adjustable Wrist Strap
5. Fold square end of strap so square end fastener adheres to the strap (figure 10).
Figure 10: Attach Adjustable Wrist Strap
16
Page 22
Using the WristOx2, Model 3150
Front View
Back View
6. Verify straps are attached properly (figure 11). The smooth, woven side of the strap should contact the patient.
Figure 11: Device with Wrist Straps Attached (Front and Back Views)
17
Page 23
Using the WristOx2, Model 3150
!
Attaching the Sensor
When applying the device using a single strap, connect the sensor to the device after the device has been applied to the patient (see “Single Strap Application”).
When using a two-piece wrist band, the sensor can be connected to the device before or af
ter applying the device to the patient (see “Two-Piece Wrist Band Application”).
The following steps apply to these Nonin sensors:
• 8000SS-WO2, 8000SM-WO2, 8000SL-WO2
• 8000AA-WO2
• 8000J-WO2 If using another Nonin-branded sensor, use sensor adap
ter cable 3150I (see “Parts and
Accessories”).
CAUTION: Only use Nonin-branded sensors with a length of 1 meter or less. Accuracy may degrade if sensor cable is over 1 meter in length. Using the sensor cable adapter does not affect accuracy.
1. Insert the sensor connector into the sensor port at the top of the device (figure 12). The Non
in logo on the sensor connector should face the front of the device.
2. Push the connector until it clicks into place.
3. The device is ready to use.
18
Figure 12: Attach Sensor
Page 24
Using the WristOx2, Model 3150
Back View
Front View
Patient Application
The WristOx2, Model 3150, is usually worn on the back of a patient’s wrist. This section describes how to secure the device to a wrist using either the two-piece wrist band or a single strap.
NOTE: Straps can be used to secure the device to an alternate location (e.g., the upper arm or a bed rail). See “Part and Accessories” for additional strap lengths.
Two-Piece Wrist Band Application
NOTE: Ensure the wrist band fits comfortably on the patient’s arm. Do not over-tighten
the wrist band.
1. Verify the ring strap and the adjustable strap have been attached properly to the device (figure 13). If straps have not been attached to the device, see “Attach Ring Strap to Device” and “Attach Adjustable Strap to Device.”
Figure 13: Verify Two-Piece Wrist Band Attachment
19
Page 25
Using the WristOx2, Model 3150
Fastener
Plastic
Ring
2. Place the device on the back of the patient’s arm.
3. Thread the rounded end of the adjustable strap through the plastic ring on the ring stra
p. Pull the adjustable strap through the plastic ring until the device fit s comforta bly
on the wrist (figure 14).
Figure 14: Thread and Tighten Two-Piece Wrist Band
4. Fold the adjustable strap back over the plastic ring and attach the fastener to the
able strap (figure 15).
adjust
Figure 15: Fasten Two-Piece Wrist Band
5. Attach the sensor if it is not already connected (see “Att
6. Apply the sensor to the patient (figure 16). Refer proper sensor application sites and sensor application cautions and warnings.
20
aching the Sensor”).
to the sensor Instructions for Use for
Page 26
Using the WristOx2, Model 3150
Front View
Back View
Figure 16: Apply Sensor to Patient
7. When in Spot Check mode, inserting a finger in the sensor automatically turns the device
on. When the finger is removed, the device enters Standby mode in
approximately 10 seconds.
NOTE: Depending on the sensor and ambient light conditions, it may take up to 3 minutes for the device to enter Standby mode.
8. If the device does not turn on, verify battery orientation, operation mode, and sensor
Single Strap Application
NOTE: Patients using a single strap may need assistance applying the device.
NOTE: Ensure the wrist band fits comfortably on the patient’s arm. Do not over-tighten
the wrist band.
1. Verify the adjustable strap has been attached to the device by the battery door
con
nection. Refer to “Troubleshooting” for additional information.
(figure 17). If strap has not been attached, see “Attach Adjustable Strap to Device.”
Figure 17: Verify Single Strap Attachment
21
Page 27
Using the WristOx2, Model 3150
2. Place the device on the back of the patient’s arm and wrap the strap around the patient’s arm.
3. Thread the rounded end of the strap through the strap bar by the sensor port (figure 18). S
tart from the back of the device and pull it towards the front of the device.
Figure 18: Thread Single Strap
4. Pull the strap through the strap bar until the device fits comfortably on the patient’s wrist (figure
19).
Figure 19: Tighten Single Strap
22
Page 28
Using the WristOx2, Model 3150
5. Fold the strap over the strap bar so the fastener attaches to the strap (figure 20).
Figure 20: Fasten Single Strap
6. Attach the sensor by inserting the sensor connector into the sensor port at the top of evice. The Nonin logo on the sensor connector should face the front of the
the d device. (See “Attaching the Sensor” for more information.)
7. Push the connector until it clicks into place.
8. Apply the sensor to the patient (figure 21). Refer t o the sensor Instructions for Use for
prope
r sensor application sites and sensor application cautions and warnings.
Figure 21: Apply Sensor to Patient
9. When in Spot Check mode, inserting a finger in the sensor automatically turns the
device
on. When the finger is removed, the device goes into Standby mode in
approximately 10 seconds.
NOTE: Depe
nding on the sensor and ambient light conditions, it may take up to 3 minutes
for the device to enter Standby mode.
10. If the device does not turn on, verify battery orientation, operation mode, and sensor
nection. Refer to “Troubleshooting” for additional information.
con
23
Page 29
Using the WristOx2, Model 3150
Verifying Operation
When the WristOx2, Model 3150, first turns on, it performs a startup sequence and self­test. It occurs:
• When a sensor is applied to a patient (Spot Check mode).
• When a sensor is attached to the device (Sensor Activation mode).
• At a programmed start time when a sensor is attached to the device (Programmed mode).
• After the activation switch is pressed while the device is in Standby mode.
• After the device disconnects from nVISION (Bluetooth connection only).
Verify all indicators display d uring the startup sequence. Indicators ap pear in the following order for 1 second each.
Startup Sequence and Self-Test
1. r and the software revision level:
2. All display icons:
3. Date/time using 24-hour clock format (MM:DD:YY:HH:MM) (example shows 23 April 2010 at 5:57 p.m.):
Month and Day Year (MM:DD) (YY)
Hour and Minutes (HH:MM)
If the time is not set, the device displays 01:01:10:00:00. If any indicator does not display, do not use the device. Contact Nonin Technical Service
for assistance.
24
Page 30
Using the WristOx2, Model 3150
Activation Switch
The activation switch is located next to the sensor port at the top of the WristOx2, Model
3150. It is primarily used to:
• Activate the Bluetooth radio when the device is either on or in Standby.
• Activate the device when it is in Sensor Activation mode so the user does not need to disconnect and reconnect the sensor.
It will also activate the device when it is in Spot Check and Programmed modes.
Activate Bluetooth Radio
When the device’s Bluetooth radio is on, a master device can connect to it. If a connection is not made, the Bluetooth radio shuts down.
Pressing the activation switch turns the Bluetooth radio on for 3 minutes. The device will remain on until the Bluetooth radio shuts down. For example, if in Sensor Activatio n mode, unplugging the sensor will not put the device in Standby.
Activate Device
When in Sensor Activation mode, the device enters S tandby mode after 10 minutes without a signal. Pressing the activation switch allows the user to turn the device on without disconnecting and reconnecting the sensor.
Error Codes
This device includes error codes that indicate problems with the unit. When an error occurs, the device displays the letters “Er” and a two-digit code (table 2).
Table 2: Error Codes
Error Code Description
01 Configuration sector error 02 Patient data pointer error
03
04 Data format 13 stored packet pointer error
05
Main memory pointer error (Device memory is intact; however, the most recent session may be missing from the device.)
Main data format 13 pointer error (Device memory is inta ct; however, the most recent stored measurement may be missing from the device.)
Some error codes may be corrected by the user. See “Troubleshooting” for more information.
25
Page 31
Troubleshooting
Problem Possible Cause Possible Solution
Troubleshooting
Device will not activate.
%SpO do not display.
Poor pulse signal indicator displays.
Poor pulse signal indicator displays and pulse strength indicator shows two bars or less.
and pulse rate
2
Batteries inserted wrong. Check batteries. Batteries are depleted. Replace batteries. Sensor is disconnected. Reconnect sensor. Device is in Sensor Activation
mode and has timed out.
Device is in Programmed mode. Use nVISION software to select Spot
Device set in Partial Display mode.
Excessive patient motion. Reduce patient motion.
Inadequate pulse signal. Reposition or replace sensor , or place
Hands are cold. Warm sensor application site.
Press the activation switch. Disconnect and then reconnect the
sensor.
Check or Sensor Activation mode. Use nVISION software to select Full
Display mode. Reconnect sensor.
sensor on a different finger. Remove and reconnect sensor .
No pulse display on pulse strength bar graph indicator.
26
Sensor applied incorrectly. Refer to sensor Instructions for Use
for proper sensor application. Device needs repair. Contact Nonin Technical Service. Possible interference from blood
flow restrictors (arterial catheters, blood pressure cuffs, infusion lines, etc.).
Reduced circulation due to excess pressure from sensor.
Excessive ambient light. Shield sensor from light source.
Reduce or eliminate restriction.
Check sensor alignment, reposition
sensor, verify correct sensor size.
Check sensor alignment.
Page 32
Troubleshooting
Problem Possible Cause Possible Solution
No pulse display on pulse strength indicator (continued).
Er 01 displays on LCD.
Er 02 or 04 displays on LCD.
Er 03 or 05 displays on LCD.
Dashes continually display on LCD.
Device does not record in Programmed mode.
Sensor applied to polished or artificial nail.
Sensor Light-Emitting Diode (LED) is not lit.
Device configuration memory failure.
Device memory failure. Contact Nonin Technical Service.
Device failure. Device memory intact, but device may have lost most recent session or stored data.
Sensor malfunction. Replace sensor with a Nonin-branded
Data collection start and stop times are set incorrectly.
Clock settings are lost after replacing batteries.
Remove fingernail polish or an
artificial nail.
Contact Nonin Technical Service.
Device reverts to default settings
(Spot-Check mode, 4-second sample
rate). Use nVISION software to
change settings. If error code
continues, contact Nonin Technical
Service.
If error code continues, contact Nonin
Technical Service.
sensor.
Use nVISION software to program
correct start and stop times.
Use nVISION software to reset clock.
Devices will not pair. Device is out of range. Verify device is in range while being
paired (approximately 100 meters
[328 feet] spherical radius). Bluetooth radio has timed out. Press activation switch to turn on
Bluetooth radio.
If these solutions do not correct the problem, please contact Nonin Technical Service at (800)
356-8874 (USA and Canada) or + 1 (763) 553-9968.
27
Page 33
Care and Maintenance
!
Care and Maintenance
The device requires no calibration or maintenance other than battery replacement.
Cleaning the Device
Wipe the device with a soft cloth dampened with a 10% bleach solution. Do not use undiluted bleach or any cleaning solution other than those recommended here, as permanent damage could result. Dry with a soft cloth, or allow to air dry.
Clean once per week or more frequently if handled by multiple users.
CAUTION: Do not place the WristOx2, Model 3150, in liquid or clean it with agents containing ammonium chloride or isopropyl alcohol.
Cleaning the Sensor
Refer to the sensor Instructions for Use for cleaning information.
Cleaning the Wrist Band
The wrist band is designed for single-patient use and may be used up to 10 times. If it needs to be cleaned, hand wash with a mild detergent (see note) in cool water (30 °C/ 86 °F). Allow to a
Do not machine wash or dry. The wrist band will shrink if placed in a dryer.
NOTE: Mild and grease. To clean washable surfaces, use in a solution of warm water.
CAUTION: Use a can be high sudsing, so use sparingly. Wipe with a damp, detergent free cloth to remove residue.
ir dry.
detergents, such as hand and dish washing liquid detergents, dissolve dirt
detergent that is safe for skin and washable surfaces. Most detergents
Storing
Store the device within the stated environmental specifications. See “Specifications” for additional information.
Remove the batteries and disconnect the sensor if it is to be stored for more than 1 month.
28
Page 34
Memory and Data
Memory and Data
The WristOx2, Model 3150 measures, collects, and stores up to 1,080 hours of SpO2 and pulse rate data with a 4-second data collection rate. Data collected at a 1 or 2-second rate reduces memory capacity to 270 or 540 hours, respectively.
When the memory is full, the device overwrites the oldest existing data with the new data. Each time the device is turned on, data are automatically stored in memory . Dat a collection of less than 1 minute is not retained in memory.
Each time the device turns on, the current oximeter time and date (if the clock is set properly) are stored in memory to allow quick differentiation of recording sessions. Patient SpO
and pulse rate are stored every 4 seconds (default), or every 1 or 2 seconds if
2
programmed using nVISION software. The oxygen saturation values are stored in 1% increments in the range of 0 to 100%.
This device contains non-volatile memory . Removing or replacing batteries does not affect the data stored in memory . S tored dat a remains in memory until overwritten by newer data or cleared from memory with nVISION software.
NOTE: Downloading data in memory does not clear memory. To clear memory, see “nVISION Settings.”
29
Page 35
nVISION Software
nVISION Software
Nonin’s nVISION software (version 6.3 or greater) works with Microsoft Windows® 2000/ XP/Vista/7 operating systems. It allows users to transfer recorded patient data from the device to a PC and then analyze, report, and archive the data. T he software is required t o access the device’s additional modes of operation and advanced features.
nVISION Settings
The following WristOx2, Model 3150, settings are programmed using nVISION:
• Date and time – 24-hour clock format
• Display options – allows clinicians to choose the best display option for each patient:
• Full display shows %SpO2 and pulse rate data
• Partial display shows pulse strength indicator, but not %SpO
• Patient data storage (sample) rate – 1, 2, or 4 seconds
and pulse rate data
2
• Operation Modes – Sensor Activation, Spot Checking, and Programmed (see “Activation Options”)
• Patient ID – up to 50 alphanumeric characters
• Bluetooth Radio – disable at startup
• Synchronize device time/date to the PC time/date
• Download and save patient data to a PC
• Clear device memory
T o access nVISION settings, connect the device to a PC using either the PC USB interface cable or a Bluetooth connection.
Accessing nVISION Settings
1. Connect the device to a PC using the USB interface cable (see “Cable Connection”)
or Bluetooth (see “Bluetooth Connection”).
2. Open nVISION.
3. Click the Data Capture icon, or select New Data Capture from the File drop down
menu.
4. Select 3150 from the list of oximeters.
5. Click Settings.
6. “Enter Wrist Oximeter Settings” window opens (figure 22). Update or cha nge settings
as needed.
7. Click OK.
8. For more information, see nViSION Help.
30
Page 36
nVISION Software
.
Figure 22: nVISION Settings Window
Cable Connection
To connect the device to a PC, use the PC USB interface cable found in the starter kit. Once connected to a PC, the device settings may be accessed and data can be downloaded using nVISION software.
The USB driver software for the cable needs to be installed before the device can connect to the
PC. The software is located in the USB Driver folder on the Operator’s Manual CD.
1. Install USB driver if needed. See appropriate “USB Drive more information.
2. Connect the cable to the USB port on the PC.
3. Connect the cable to the device’s sensor port.
r Installation” section for
31
Page 37
nVISION Software
4. When the device is ready to use with nVISION, these indicators display on the LCD:
CP
• Battery indicator
5. For more information about nVISION, refer to nVISION Help. NOTE: Disconnect the USB interface cable from the device when the data transfer or
device configuration is complete. Leaving the cable connected will reduce battery life.
USB Driver Installation (XP)
1. The USB driver software is on the Operator’s Manual CD. Insert the CD into the PC’s
CD-DVD drive.
2. Connect the USB cable to the sensor port on the device and a USB port on the PC.
3. The Found New Hardware wizard opens and asks if Windows should connect to
Windows Update to search for software. Select No, not this time and click Next.
4. Select Install from a list or specific location (Advanced) and click Next.
5. When asked to choose search and installation options:
a. Select Search for the best driver in these locations. b. Unselect Search removable media c. Select Include this location in the search: d. Browse to the USB Driver folder on the Operator’s Manual CD and click OK. e. Click Next.
6. If the Hardware Installation/Windows logo testing window appears, click Continue
Anyway to continue the installation.
7. When the wizard is done with the software installation, click Finish.
8. Look up the communications (comm or COM) port for the device:
a. Click Start / Settings / Control Panel. b. Select System. System Properties window opens. c. On the Hardware tab, select Device Manager. d. Expand Ports (COM & LPT). One port should say “Nonin Model 3150 (COM#).”
Make a note of the COM#. It is needed to set up the device with nVISION.
32
Page 38
nVISION Software
USB Driver Installation (Vista)
1. Connect the USB cable to the sensor port on the device and a USB port on the PC.
2. In the Found New Hardware window, select Locate and install driver software (recommended).
3. Windows needs your permission to continue pop up appears. Continue with installation.
4. The USB driver software is on the Operator’s Manual CD. Insert the CD into the PC’s CD-DVD drive and click Next.
5. Update Driver Software - Model 3150 window opens. Choose Browse my computer for driver software.
6. Browse to the USB Driver folder on the Operator’s Manual CD and click OK.
7. Click Next.
8. In the Windows Security pop-up window, select Install this driver software anyway.
9. Driver software installs. When Windows has successfully update d the driver software, click Close.
10. Open the Device Manager by clicking Start / Control Panel / Device Manager.
11. In the Device Manager window, look up the communications (comm or COM) port for the device. Expand Ports (COM & LPT). One port should say “Nonin Model 3150 (COM#).” Make a note of the COM#. It is needed to set up the device with nVISION.
USB Driver Installation (Windows 7)
1. The USB driver software is on the Operator’s Manual CD. Insert the CD into the PC’s CD-DVD drive.
2. Connect the USB cable to the sensor port on the device and a USB port on the PC.
3. Open the Device Manager by clicking Start / Control Panel / System and then selecting Device Manager.
4. Expand Other devices.
5. Right click Model 3150 and select Update Driver Software...
6. Update Driver Software - Model 3150 window opens. Choose Browse my computer for driver software.
7. Browse to the USB Driver folder on the Operator’s Manual CD and click OK.
8. Click Next.
9. In the Windows Security pop-up window, select Install this driver software anyway.
10. Driver software installs. When Windows has successfully updated the driver software , click Close.
11. In the Device Manager window, look up the communications (comm or COM) port for the device. Expand Ports (COM & LPT). One port should say “Nonin Model 3150 (COM#).” Make a note of the COM#. It is needed to set up the device with nVISION.
33
Page 39
nVISION Software
Bluetooth Connection
NOTE: Etched onto the device is the word “pin” followed by a 6-digit number. This is the device’s unique identification number, also known as the Bluetooth Passkey or PIN Code. This number is used when pairing the device to the host system. Refer to the host system’s operator’s manual for additional information.
Before a Bluetooth master device can connect with the WristOx2, Model 3150 (slave device), the devices must be paired. Once paired, the WristOx automatically connect with the last paired master device when turned on or activated.
1. To connect the WristOx2, Model 3150, to a PC or another device using Bluetooth, see Nonin’s online Bluetooth Connection Tutorial:
http://www.nonin.com/training/products/3150/bluetooth_connection_tutorial/
2. When nVISION connects to the WristOx2, Model 3150, the device stops recording patient data and the following indicators display on the LCD:
, Model 3150, will
2
CP
• Battery indicator
• Bluetooth icon with animated bars
3. For more information about nVISION, refer to nVISION Help.
34
Page 40
Parts and Accessories
Parts and Accessories
For more information about Nonin parts, acce ssories and sensors, contact your distributor , or contact Nonin at (800) 356-8874 (USA and Canada) or +1 (763) 553-9968. This information is also available on Nonin’s website: www.nonin.com.
Model Number Description
3100CC Carrying Case 3150 Manual CD with Operator’s Manual and USB Driver Software 3150SC PC USB Interface Cable nVISION nVISION Software (version 6.3 or greater). Used with Microsoft
Windows 2000/XP/Vista/7 operating systems.
3150I Sensor Adapter Cable. Used to connect 1-meter, 9-pin connector
sensors to the WristOx sensors, contact Nonin, your distributor, or visit www.nonin.com.
, Model 3150. For compatible 1-meter
2
3150WB Two-piece Wrist Band. Available lengths:
6 in. (15 cm) 8 in. (20 cm) 10 in. (25 cm) 13 in. (33 cm)
3150WBE Wrist Band Extender, 5 in. (13 cm)
Reusable Pulse Oximeter Sensors
8000AA-WO2 Adult Articulated 8000J-WO2 Adult Flex Sensor 8000SS-WO2 Soft Sensor Small 8000SM-WO2 Soft Sensor Medium 8000SL-WO2 Soft Sensor Large
Finger Clip Sensor
35
Page 41
Service, Support, and Warranty
Service, Support, and Warranty
Service and Support
For information about the device and accessories, contact your local sales represent ative or distributor. For the sales representative or distributor in your area, contact Nonin.
A return authorization number is required before returning any product to Nonin. To obtain this return authorization number, contact Nonin’s Technical Service Department at:
Nonin Medical, Inc.
13700 1st Avenue North
Plymouth, Minnesota 55441-5443 USA
(800) 356-8874 (USA and Canada)
+ 1 (763) 553-9968
Fax: + 1 (763) 553-7807
E-mail: info@nonin.com
www.nonin.com
Warranty
NONIN MEDICAL, INCORPORATED, (NONIN) warrants to the purchaser the Model 3150, WristOx WristOx which NONIN has been notified by the purchaser by serial number that there is a defect, provided said notification occurs within the applicable warranty period. This warranty shall be the sole and exclusive remedy by the purchaser hereunder for any WristOx purchaser that is found to be defective in any manner , whether such remedies be in contract, tort, or by law.
This warranty excludes cost of delivery to and from NONIN. All repaired units shall be received by the purchaser at NONIN’s place of business. NONIN reserves the right to charge a fee for a warranty repair request on any unit found to be within specifications.
The WristOx2, Model 3150, is a precision electronic instrument and must be repaired by knowledgeable and specially trained NONIN personnel only. Accordingly, any sign or evidence of opening the WristOx of misuse of the WristOx
All non-warranty work shall be performed according to NONIN standard rates and charges in effect at the time of delivery to NONIN.
DISCLAIMER/EXCLUSIVITY OF WARRANTY:
Pulse Oximeter for 3 years from the date of purchase. NONIN shall rep air or replace any
2
, Model 3150, found to be defective in accordance with this warranty, free of charge, for
2
, Model 3150, delivered to the
2
, Model 3150, field service by non-NONIN personnel, t ampering, or any kind
2
, Model 3150, shall void the warranty.
2
THE WARRANTIES IN THIS MANUAL ARE EXCLUSIVE, AND NO OTHER WARRANTIES OF ANY KIND, WHETHER STATUTORY, WRITTEN, ORAL, OR IMPLIED, SHALL APPLY.
36
Page 42
Technical Information
Technical Information
Manufacturer’s Declaration
Refer to the following table for specific information regarding this device’s compliance to IEC 60601-1-2.
Table 3: Electromagnetic Emissions
Emissions Test Compliance Electromagnetic Environment—Guidance
This device is intended for use in the electromagnetic environment specified below
The customer and/or user of this device should ensure that it is used in such an environment.
RF Emissions CISPR 11
RF Emissions CISPR 11
Harmonic Emissions IEC 61000-3-2
Voltage Fluctuatio Flicker Emissions IEC 61000-3-3
ns/
Group 2 This device must emit electro
to perform its intended function. Nearby electronic equipment may be affected.
Class B This device is suitable for use in all establishments,
includ
ing domestic and those directly connected to the
N/A
N/A
public low-voltage power supply network that supplies buildings used for domestic purposes.
magnetic energy in order
.
37
Page 43
Technical Information
Table 4: Electromagnetic Immunity
Immunity
Test
This device is intended for use in the electromagnetic environment specified below
The customer and/or user of this device should ensure that it is used in such an environment.
Electrostatic Disch
arge (ESD)
IEC 61000-4-2
Electrical Fast
ransient/Burst
T IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short in
terruptions, and voltage variations on power supply input lines IEC 61000-4-11
Power Frequency (50/60 Hz) Magnetic
ld
Fie IEC 61000-4-8
NOTE: U
is the AC mains voltage before application of the test level.
T
IEC 60601 Test Level
Compliance
Level
Electromagnetic
Environment—Guidance
.
±6 kV contact ±8 kV air
±6 kV contact ±8 kV air
Floors should be wood, concrete, or ceramic tile. If floors are covered with synthetic material, relative humidity should be at least 30%.
±2 kV for power supply lines ±1 kV for input/output lines
±1 kV differential mode ±2 kV common mode
N/A Mains power quality should be
that of a typical commercial or hospital environment.
N/A Mains power quality should be
tha
t of a typical commercial or
hospital environment.
±5% U
(>95% dip in UT)
T
for 0.5 cycle ±40% U
(60% dip in UT)
T
N/A Mains power quality should be
tha
t of a typical commercial or
hospital environment. for 5 cycles ±70% U
(30% dip in UT)
T
for 25 cycles <5% U
(>95% dip in UT)
T
for 5 sec. 3 A/m 3 A/m Power frequency magnetic fields
should be at levels c
haracteristic of a typical location in a typical commercial or hospital environment.
38
Page 44
d 1.17 P=
d 1.17 P=
d 2.33 P=
Technical Information
Table 5: Guidance and Manufacturer’s Declaration—
Electromagnetic Immunity
Immunity Test IEC 60601 Test
Level
This device is intended for use in the electromagnetic en
Compliance
Level
Electromagnetic Environment—
Guidance
vironment specified below.
The customer and/or user of this device should ensure that it is used in such an environment.
Portable and mobile RF communications equipm ent sh ou ld be use d no close r to any part of the devic e, includ
ing cables, than the recommended separation distance calculated from the equation applicable to
the frequency of the transmitter.
Recommended Separation Distance
Conducted RF IEC 61000-4-6
Radiated RF IEC 61000-4-3
3 Vrms 150 kHz to 80 MH
z
3 V/m 80 MHz to 2.5
z
GH
3 Vrms
3 V/m
3 V/m
80 kHz to 800 MHz 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
termined by an electromagnetic site survey
de should be less than the compliance level in each frequency range.
b
a
,
Interference may occur in the vicinity of equipment marked with the symbol:
NOTES:
1. At 80 MHz and 800 MHz, the higher frequency ra nge ap p lies.
2. These guidelines may not apply in all situation
s. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
a. Field strengths from fixed transmitters, such as base st ations f or radio (cellular/ cordless) telephon es and land mobile radios, amateur radio, AM
and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation. If abn ormal per formance is observed, additional measures may be necessary, such a
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
s reorienting or relocating the device.
39
Page 45
Technical Information
d 1.17 P=
d 1.17 P=
d 2.33 P=
Table 6: Recommended Separation Distances
This table details the recommended separation distances between portable and mobile RF communications equipment and this device
This device is intended for use in an electromagnetic e
of this device can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communication equipment (transmitters) and the device as recommended below, according to maximum output
power of the communications equipment.
Separation Distance According to Frequency of Transmitter
Rated Maximum Output Power
of
150 kHz to 80 MHz
Transmitter W
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.73 1 1.2 1.2 2.3 10 3.7 3.7 7.3 100 12 12 23 For transmitters rated at a maximum output power n
distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTES:
1. At 80 MHz and 800 MHz, the higher frequency range applies.
2. These guidelines may not apply in all situations. Electro absorption and reflection from structures, objects, and people.
nvironment in which radiated RF disturbances are controlled. Users
80 MHz to 800 MHz
800 MHz to 2.5 GHz
ot listed above, the recommended separation
magnetic propagation is affected by
Equipment Response Time
SpO2 Values
Standard/Fast Averaged SpO
2
Pulse Rate Values Response Latency
Standard/Fast Averaged Pulse Rate 4 beat exponential 2 beats
Example - SpO2 Exponential Averaging
decreases 0.75% per second (7.5% over 10 seconds)
SpO
2
Pulse Rate = 75 BPM
40
Average Latency
4 beat exponential 2 beats
Page 46
60
65
70
75
80
85
90
95
100
0.0
8.0
16.0
24.0
32.0
40.0
48.0
56.0
64.0
72.0
80.0
Time in seconds
SpO2
SaO2 Reference 4 B eat Average
Technical Information
Specific to this example:
• The response of the 4-beat average is 1.5 seconds.
Testing Summary
SpO2 accuracy and low perfusion testing was conducted by Nonin Medical, Inc., as described below.
SpO2 Accuracy Testing
SpO2 accuracy testing is conducted during induced hypoxia studies on healthy, non­smoking, light- to dark-skinned subjects during motion and no-motion conditions in an independent research laboratory. The measured arterial hemoglobin saturation value
) of the sensors is compared to arterial hemoglobin oxygen (SaO2) value,
(SpO
2
determined from blood samples with a laboratory co-oximeter . The accuracy of the sensors in comparison to the co-oximeter samples measured over the SpO 70 - 100%. Accuracy data is calculated using the root-mean-squared (A subjects, per ISO 9919:2005, Medical Electrical Equipment—Particular requireme the basic safety and essential performance of pulse oximeter equipment for medical use.
Pulse Rate Motion Testing
This test measures pulse rate oximeter accuracy with motion artifact simulation introduced by a pulse oximeter tester . This test determines whether the oximeter meets the criteria of ISO 9919:2005 for pulse rate during simulated movement, tremor, and spike motions.
Low Perfusion Testing
This test uses an SpO2 Simulator to provide a simulated pulse rate, with adjustable amplitude settings at various SpO maintain accuracy in accordance with ISO 9919:2005 for heart rate and S pO obtainable pulse amplitude (0.3% modulation).
levels for the oximeter to read. The oximeter must
2
range of
2
rms
value) for all
nts for
at the lowest
2
41
Page 47
Specifications
Oximeter Specifications
Technical Information
Oxygen Saturation Display Range: 0 to 100 % SpO
2
Pulse Rate Display Range: 18 to 321 beats per minute (BPM) Displays:
Numeric: 3-digit LCD
Pulse Strength: Pulse Strength Bar Graph
Saturation Accuracy A
%SpO
: No Motion Motion
2
rms
a
:
70 % to 100 %
b
Low Perfusion 8000AA-WO2: ±2 units ±2 units ±2 units 8000J-WO2: ±3 units NA ±2 units 8000SM-WO2: ±2 units NA ±2 units
Pulse Rate Accuracy: ±3 units ±3 units ±3 units
c
Measurement Wavelengths and Output Power
:
Red: 660 nanometers @ 0.8 mW max. avg.
Infrared: 910 nanometers @ 1.2 mW max. avg.
a. ±1 A b. SpO2 accuracy in motion specified for 8000AA-W sensor. c. This information is especially useful for clinicians performing photodynamic therapy.
represents approximately 68% of measurements.
rms
42
Page 48
Technical Information
System Specifications
Temperature:
Operating: -5 °C to 40 °C (23 °F to 104 °F)
Storage/Transportation: -40 °C to 70 °C (-40 °F to 158 °F)
Humidity:
Operating: 10 % to 95 % noncondensing
Storage/Transportation: 10 % to 95 % noncondensing
Operating Altitude: Up to 12,192 meters (40,000 feet) Operating Hyperbaric Pressure: Up to 4 atmospheres Power Requirements: Two AAA (1.5V) alkaline batteries Battery Life:
Storage: Approx. 9 months
Operating without Bluetooth: Approx. 48 hours (minimum) of continuous
operation
a
Approx. 8 hours minimum
Operating with Bluetooth radio – Class 1
:
Operating with Bluetooth radio – Class 2: Approx. 24 hours minimum
Dimensions (without sensor or strap): 51 mm x 73 mm x 19 mm (H x W x D)
(2.0 in. x 2.9 in. x 0.75 in.)
Weight (with batteries and wrist strap): 64.0 g (2.3 oz) Memory:
Type: Non-volatile
Capacity: up to 1,080 hours (4 sec. data storage rate)
up to 540 hours (2 sec. data storage rate) up to 270 hours (1 sec. data storage rate)
Classification per IEC 60601-1 / CAN/CSA-C22.2 No. 601.1 / UL60601-1:
Type of Protection: Internally powered (battery power) Degree of Protection: Type BF-Applied Part Mode of Operation: Continuous Enclosure Degree of Ingress Protection: IP33
a. When operating with Bluetooth, typical battery life may vary depending on proximity to host connection
and configuration of host-to-device communications. Times provided are minimum times for common configurations.
43
Page 49
Transmitter
Bluetooth Compliance: Version 2.0 Operating Frequency: 2.4 to 2.4835 GHz Output Power: < 20 dBm Operating Range: 100-meter (328-foot) radius indoors Network Topology: Point-to-Point Operation: Slave Antenna Type: Internal Modulation Type: Frequency Shift Keying
Frequency Hopping Spread Spectrum
Technical Information
Band Width: 1 MHz
44
Loading...