Niigata seiki LHT-400 Instruction Manual

PROFESSIONAL
Instruction Manual
Digital Hardness Tester for Metals (Rebound Type)
Model No.
LHT-400
LEEB HARDNESS TESTER
Thank you for purchasing the Niigata Seiki Leeb-Type Hardness Tester. This digital instrument measures the hardness of (non-magnetized) metals: steel, cast steel, tool steel, stainless steel, heat resistant steel, standard (gray) cast iron, ductile (nodular cast) iron, cast aluminum alloys, brass, bronze, and copper. Hardness is measured by impacting the surface of the sample with a test probe and measuring the rebound velocity.
For safe and proper use of this product, please read this
instruction manual before use and follow the procedures described. Please keep manual where it is accessible to user for future reference.
Keep this manual with the instrument if transferred or leased to
a third party.
For inquiries about this product, please contact dealer or
Niigata Seiki at the address listed on the following page.
PART IDENTIFICATION & FUNCTION
[Main Unit]
LCD Display
Measuring
Ring
Probe
Cover
Measure Button
USB Port
Handle
Up Button
Down Button
Power Button
In this manual, indicates RISK OF PERSONAL INJURY OR PROPERTY DAMAGE if not followed. The symbol indicates something which is PROHIBITED, and the symbol indicates REQUIRED step or necessary condition.
SAFETY NOTIFICATIONS
Probe ·························· Impactor which strikes surface
during measurement.
Measuring Ring ·········· Pressed to surface during
measurement.
LCD Display ··············· For display of measurements
and settings.
USB Port ···················· Connects to included USB cable
for charging and data output.
Cover ·························· Cover for USB Port. Please
keep closed when not in use.
Measure Button ··········· Press to release Probe and
trigger a measurement.
Handle ························· Move down to position Probe
for measurement.
Power Button ·············· Power ON/OFF, and menu
select button.
Up Button ···················· Button to select a setting, and
to control display size.
Down Button ················ Button to select a setting, and
to delete last measurement.
[Accessories]
⑪USB Cable
Cleaning
Brush
Measuring Ring (sml.)
Case ● Manual
[Measurement Screen]
Ⓑ Ⓚ  Ⓒ
H
HL
AUTO
Standard
Display
D
ST&CAST ST
ⒾⒽ
HLD: MAX: MIN: AVG:
419 411 414
USB Cable ·················· Connects for charging Main
Unit and transferring data.
Cleaning Brush ··········· For cleaning inside of Main
Measuring Ring (sml
⑭AC Adapter
Calibration
Block
Ⓓ Ⓔ Ⓕ Ⓖ
%
Large Readout
Ⓑ ⓀⒸ
AUTO
419
Display
HLD:HHL
AC Adapter ················· Charges Main Unit via the
Calibration Block ········ Test Block of specified hardness
Storage Case ·············· Storage case for Main Unit
Instruction Manual ······ This user guide.
Scaled t / Average
Scale Units
Measurement
Maximum
Minimum
Average
Record Number
Press the Up Button to toggle between Standard and Large readout
Unit.
.) ··· Replaceable tip for use in
tight places.
USB Cable.
checking, and calibrating
for
unit before each test.
and accessories.
Battery Charge Level
No Measurements / Total
Material Setting
HighLow Limit Exceeded
H) (L)
Memory Auto / Off
Ⓛ ⓂProbe Type
SAFETY PRECAUTIONS
Please Observe
Always follow the procedures specified below in order to prevent harm to yourself or others, and to prevent damage to property.
Content marked as follows indicates risk of injury or
damage if not followed.
WARNING
Indicates risk of personal injury or property damage if not followed.
These symbols mark content that must be observed.
Denotes a prohibition – You MUST NOT do
Denotes a requirement – You MUST do
CAUTION
Read the manual and follow all instructions.
Use of product other than as described in the manual may
cause accident.
Use only for measuring hardness.
Use for any purpose other than measuring may damage or
wear the instrument. Improper use may also cause accident.
Use in an environment which meets the following conditions:
● Keep away from direct sunlight
● Keep in dry location protected from rain and water.
● Protected from use by children and unauthorized people.
Use in location contrary to the above may cause poor accuracy,
damage to the product, or may result in accident or injury.
Handle With Care.
Do not drop or subject to shock, do not place under heavy
objects. Damage may cause failure or poor accuracy.
Preparation of Test Surface
The workpiece and measured surface must meet the following conditions.
① Mass: 5 kg or more
For workpiece weighing between 2 and 5 kg, or for thin and easily
deformed parts, place on a surface plate or stable surface of more than 5kg.
For workpiece weighing less than 2kg, secure it to a surface plate
of more than 5kg using a coupling agent, such as petroleum jelly, with a tight fit to stabilize.
② Surface Roughness: Ra 2μm or less
If surface is rougher than Ra 2μm, please polish to meet
specification.
③ Radius of Curvature: greater than 30mm ④ Surface Temperature: 80°C or less ⑤ For testing surface hardness, a hardened surface thickness:
0.8mm or greater
⑥ Not Magnetized
Be sure to demagnetize if there is any residual magnetism.
⑦ Bare (unpainted) Surface
If sample is painted, remove any paint or coating before test.
⑧ Not easily fractured, or damaged by the surface indents
resulting from the test
This instrument measures hardness by impacting a probe, with a
hardness of HLD 1600, against the test surface. Please note: after testing, some small dents will remain on workpiece.
Charging
The Main Unit contains a rechargeable battery. When power level drops below 20%, please charge as follows.
① Connect with the provided USB Cable
Connect to the AC Adapter using the included USB cable.
Plug in the AC adapter
Plug the AC adapter into the power mains (100V AC).
Not for use on on materials which are easily fractured, or which will be damaged by the dents created during tests.
This instrument measures hardness by impacting a probe, with
a hardness of HLD 1600, against the test surface. Please note: after testing, some small dents will remain on workpiece.
Do not use on magnetized material or in presence of magnetic fields.
Hardness is measured based on rebound speed of the Impact
Probe. If magnetic field is present, the rebound speed may be affected and a precise measurement is not possible.
Do not dryfire instrument.
If test is performed without test surface under probe, the
instrument may be damaged and affect accuracy.
Do not Modify or disassemble beyond the
procedures described in this manual.
It may damage product or affect accuracy. For service,
please contact distributor or place of purchase.
Accuracy Confirmation ・ Calibration
Instrument accuracy will be affected by shock or Probe wear. Please check accuracy and calibrate before each use using the supplied Calibration Block.
① Confirm the hardness written on Calibration Block
The hardness of the calibration master is marked on the block surface.
779HLD
Calibration Block
② Measure the Calibration Block hardness
Use the "Hardness Test Procedure" on the following page to test the hardness.
Material = Steel  ● Unit = HL ● No. of Measurements = 5
Make sure the Calibration Block surface is clean before measurements.
Take the measurements in a spot free of marks from previous
measurements.
③ Error Correction
If the measured hardness from step is within ±17HL of the
hardness from step , then the accuracy is confirmed within the guaranteed range and the instrument is ready to use.
If the measured hardness from step is NOT within ±17HL of the
hardness from step , then the calibration must be adjusted. Enter the measurement error In the "Setup Menu" "6. Calibration" as shown in the following page.
Connect
USB Port
Main Unit
USB
Cable
Block hardness
AC Adapter
USB Port
Connect
Power ONOFF
Power ON ··········· Press the Power Button. The power will turn on and the Measurement Screen will display on the LCD.
Power OFF ········ From the Measurement Screen, press and hold the Power Button for 3 seconds. Power and LCD will turn off.
Hardness Test Procedure
Follow the steps below for a basic hardness test.
① Turn the power ON.
Press the Power Button to turn on the power.
② Set the material and measurement units.
Specify the workpiece material and desired scale units in the "Setup Menu"
"3. Material & Scale" After entering the settings, select "1.
Measurement" to return to the Measurement Screen.
③ Press the Measuring Ring against the surface.
Make sure the workpiece and test spot meet the conditions listed in the section "Preparation of Test Surface", and press the Measuring Ring onto the surface.
Make sure there are no scratches or marks from previous measurements.
④ Push down on the Handle, and then pull it back up. (Figure 1)
Push down on the Handle to charge the internal probe for the test, then lift the Handle to bring it into position for the test.
Make sure you hold the Handle secure while bringing it up to the
top. If the Handle slips, it may return with a shock and cause damage.
⑤ Press the Measure Button. (Figure 2)
With the Measuring Ring held firm to the surface, press the Measure Button. The Probe will release and impact the surface. Hardness will be calculated from the rebound velocity. Be careful not to push down on the Handle when pressing the
Measure Button, it will cause inaccurate readings.
Repeat the measurement in steps ④~⑤ .
Repeat the required number of times specified in the "Setup Menu" "2. Average" for calculating an averaged reading.
⑦ The measurements are displayed. (Figure 3)
The average value, maximum reading, and minimum reading are displayed on the LCD display.
Because workpiece hardness my not be uniform, measurements should be
taken from multiple locations on the surface of the workpiece for averaging.
Press
Push down
Average Value
Automatic Angle Co
The instrument automatically senses the measurement angle and applies a correction factor the reading. Measurement can be made freely over 360° range. Please make sure the Measuring Ring is always pressed firm against the surface when taking a measurement.
Return to top
Handle
Workpiece
(Fig.1)
AUTO
D
mpensation
Measure Button
HL
ST&CAST ST
Workpiece
to
HLD: MAX: MIN: AVG:
easuring
M Ring
00003 3: 075
419 411 414
(Fig.2)
Maximum
Minimum
Average
%
(Fig.3)
Measure
Setup Menu
When power is ON, pressing the Power Button will display the Setup Menu screen. Within the Setup Menu, the Up and Down buttons are used to select items and to change values, and the power button is used to enter the selection.
1. Measurement ·····
2. Average ·······
3. Material & Scale ···
4. Tolerance
········
5. Memory ·········
6. Calibration ·······
7. Auto Shutdown ····
8. Contrast ·········
9. Impact Type ······
10. Language ·······
11. Default ·········
Return to the Measurement Screen.
Change the number of measurements averaged for each reading. You can select 2-8 measurements.
Specify the workpiece material, and set the measurement units. When specifying the material and units, options are shown, so select as required.
To set up tolerance function. Press Up Button to change the value, and Down Button to move to next digit. "H" or "L" will appear on the display for readings above or below the set limits.
Control the storage of measured data.
1. Memory on/off ······ ON - Measurements saved sequentially up to number specified below. OFF - Measurement data is not saved
2. File Setting ··········· Specify the storage destination for the group of measurements. Ten groups, A to J, can be stored with 200 Measurement averages each.
3. Delete Single ······· Delete data from the specified group.
4. Delete All ············· Clear the Main Unit stored measurement data.
Correct the measurement by entering the error value determined in the accuracy check from the "Calibration" section on the previous page.
Automatic power shut off can be selected for 1, 2, 5 minutes; or disabled.
Adjust the contrast of the display.
Set parameters to match the type of probe
Select from Japanese / English.
Restore all settings to factory default.
Back Page: Installing DL-Type Probe (sold separately)
Back Page: Meas. Range Listing for Each Material, Units
Measurement Data Output
Niigata seiki Co., Ltd.
5-3-14, Tsukanome, Sanjo, Niigata, Japan, 955-0055 Tel. : +81-256-33-5522 Fax. : +81-256-33-5518
MAIL intl.sales@niigataseiki.co.jp URL http://www.niigataseiki.co.jp
① Connect the Main Unit to the computer
Please connect the USB Port of the Main Unit to the computer's USB Port using the USB Cable and turn on the power. The Display will show "Connected between instrument and PC"and the computer will show a CD drive icon.
② Open the CD drive icon on the computer.
The data is stored as text (.txt) files labeled AJ, and you can open the desired file to view.
The file contents are listed below
Num Impact Material Scale AVGDATA DATA1~8
Files can not be deleted or created through the computer, you must delete at the Main Unit.
File Number Probe type: D/DL The material type set in the "Material & Scale" setting The scale set in the "Material & Scale" setting Average of Measurements Measurement data. The number will correspond to the " Average" setting
Changing The Measuring Ring, Cleaning
The Measuring Ring can be exchanged with a small one if required to fit in tight locations. Also, the ring can be removed to for periodic cleaning inside of the mechanism.
① Press the Measure Button.
Release the probe from charged position.
Turn the Measuring Ring to remove it from the Main Unit (Figure 4) and take out the Probe.
③ Cleaning.
Using the supplied brush, clean the inside of the main unit as well as the Probe to remove any dust or dirt.
④ Replace the Probe and reattach the Measuring Ring to the Main Unit.
Insert the Probe in the direction shown in the drawing. (Figure 5)Screw the Measuring Ring back on to the main body.
Computer USB Port
Connect
USB Cable
Connect
USB Port
Main Unit
Connected between instrument and PC
Before cleaning, or replacing the Measuring Ring,
make sure the power is turned OFF.
Loosen
Measuring
Ring
Threads
(Fig.4)
Tighten
Measuring Ring
Installing DL-Type Probe (sold separately)
Measurement can be made in narrow spaces, such as the bottom groove of a large gear, by replacing the Probe with the DL-type Probe available for the Hardness Tester.
Press the Measure Button.
Release the Probe from charged position.
② Turn the Measuring Ring to remove it from the Main Unit (Figure 4)
and take out the (Dtype) Probe.
③ Insert the DLtype Probe as shown in the drawing and attach the
measurement guide. (Figure 6)
④ Change "Setup Menu → "9. Impact Type" from "D" to "DL".
(Fig.5)
Probe
Measurement guide
DL-Type Probe
(Fig.6)
Insert
Insert
Measurement Range Listing by Material, Units / Specifications
Material
ST& CAST ST.
1.
CWT STEEL
2.
STAINLESS ST.
3.
G CAST IRON
4.
N CAST IRON
5.
CAST ALUM
6.
BRASS
7.
BRONZES
8.
COPPER
9.
Accuracy: within ±17HL
Operating Temperature: 0+40°C
Unit Scale
HL HRC HRB HB HS HV
300 〜 900 300 〜 640 300 〜 800 360 〜 650 400 〜 660 174 〜 560 200 〜 550 300 〜 700 200 〜 690
20.0 〜 68.0
20.4 〜 67.1
19.6 〜 62.4
Weight: Main unit … 97g
Data Storage Capacity: Max. 2,000 avg. readings
38.4 〜 99.5
46.5 〜 100.7
13.5 〜 95.3
80 〜 647
− 85 〜 656 90 〜 334
131 〜 367
20 〜 190 40 〜 173 60 〜 290 45 〜 315
32.5 〜 99.5
Auto Power Off: 1, 2, 5 minutes, or disabled
Charge time: approx. 3 hours
80 〜 940 80 〜 898 85 〜 802
σb
373 〜 2008
Maintenance and Storage
Remove any dust or grime with a dry cloth.
Any contamination entering inside the Main Unit will prevent smooth operation of the Probe.
Store in a cool, dry, and dark location in the provided case.
Keep out of direct sunlight
and moisture, and please keep secure from unauthorized personnel.