Niigata seiki DI-1058, DI-10KD, WDI-10, WDI-0342, WDI-0542 User Manual

...
……………
………
………………
………………………
……………………
…………
………………………
…………
…………
DIAL GAUGE
Dial Indicator / Comparator
PROFESSIONAL
Model No.
User Manual
APPLICATIONS
PART IDENTIFICATION, FUNCTION
Model WDI-0558SP shown
Comparing parts to a master part during inspectionMeasuring machine tools positioning accuracy
Measuring runout for rotary shaftsChecking vise parallelism on milling machinesMeasuring flatness of surfaces and assembliesConfirming machine tool feed distance
Thank you for purchasing the Niigata Seiki Dial Gauge. Used with a Magnetic Base or Indicator Stand, this gauge will show the difference in height or position relative to a zero point set at a reference position.
For safe and proper use of this product, please read this
instruction manual before use and follow the procedures described. Please keep manual where it is accessible to user for future reference.
Keep this manual with the instrument if transferred or
leased to a third party.
For inquiries about this product, please contact dealer or
Niigata Seiki at the address listed on the following page.
Contact Point (Probe
)
Spindle
Stem
Pointer
(Needle
)
Bezel ClampScrew
Bezel
(
φ8mm
)
Flat Rear Plate
Limit Markers
Main Scale
(Indicator Face
)
Rev. Counter Scale
Revolution Counter
Lug
Mounting
Hole
(
φ6. 5mm
)
Rear Plate with Lug
To Use the Rear Plate with
Lug, unscrew the Standard Flat Rear Plate screws, replace, and use screws to secure.
[
Accessories
]
SAFETY NOTIFICATIONS
Throughout this manual, symbol indicates RISK OF PERSONAL INJURY OR PROPERTY DAMAGE if not followed. The symbol indicates something which is PROHIBITED, and the symbol Indicates REQUIRED step or necessary condition.
DI-10KD WDI-10 DI-1058
WDI-0342 WDI-0542
WDI-0558SP WDI-1058SP WDI-0158SP
Limit Markers For marking acceptance range for Pass/Fail testing
Pointer (Needle) Shows measurement on Main Scale
Main Scale Main Scale, rotate Bezel (⑩) to turn
Bezel Clamp Screw Clamps Scale to prevent rotation
(DI-10KD/WDI-0558SP/
WDI-1058SP/WDI-0158SP only
)
Revolution Counter Pointer indicates Main Scale ×100
Rev. Counter Scale Each division is Main Scale ×100
Stem For holding and mounting Gauge
Spindle Shaft moves up and down with measurement
Contact Point Probe which contacts workpiece
Bezel Rotate to turn Main Scale (③) ⑪Flat Rear Plate Standard back coverLug Alternate mounting point Mounting Hole For mounting GaugeRear Plate with Lug Accessory back cover for
mounting Gauge using Lug
SAFETY PRECAUTIONS
PREPARATION - Mounting
WARNING
Indicates risk of personal injury or property damage if not followed.
Please Observe
Always follow the procedures specified below in order to prevent harm to yourself or others, and to prevent damage to property.
Dial gauge must be securely mounted such as on a comparator stand or magnetic base. Please follow these guidelines.
Content marked as follows indicates risk of injury or damage if not followed
CAUTION
Denotes a prohibition – You MUST NOT do Denotes a requirement – You MUST do
These symbols mark content that must be observed.
Read the manual and follow all instructions.
Use of product other than as described in the
manual may cause accident.
Use only as indicator Gauge.
Use for any purpose other than measuring
may damage or wear the instrument. Improper use may also cause accident.
Use in an environment which meets the following conditions:
Temperature within range of 0~40°C
humidity 30~70% (non-condensing.
)
Location with minimal dust, oil, oil mist,
and protected from direct sunlight.
Location protected from use by children
and unauthorized people.
Use in location contrary to the above may
cause poor accuracy, damage to the product, or may result in accident or injury.
Handle With Care.
Do not drop or subject to shock, do not place
under heavy objects. Damage may cause failure or poor accuracy.
Make sure Gauge holder is rigid.
Holder must be sufficiently secure to prevent deflection
from the weight of the Gauge.
Holder must be rigid enough to hold Gauge and not to
lift from measurement force.
Holder support arm should be as short as possible to
prevent deflection.
Deflection or lifting will cause measurement error such
as origin position error and inaccuracies in measured reading.
Dial Gauge must only be attached by Stem or Rear Lug.
Mounting of gauge by other than Stem or Lug will cause
inaccuracy and product damage.
During installation, do not over-tighten the Stem.
Excessive force on the Stem may cause Spindle to bind.
Do not disassemble or modify.
It may damage Gauge and cause poor
accuracy
If Bezel Clamp Screw is removed, internal
components may come loose and become misaligned causing product failure.
Do not shock Spindle.
Rapid motion, or lateral force may damage
Gauge and cause poor accuracy.
Gauge Holder
Support Arm
Stem
Mounting Example
Minimize Length
No Rapid Motion No Lateral Force
HOW TO USE - Preparation
Confirm that the Contact Point and Rear Plate are tightly fastened.
If loose, tighten Contact Point and Rear Plate Screws.
Attach to the Gauge Holder using the Stem or Rear Lug.
Mounting Gauge by other than Stem or Lug will cause inaccuracy and product damage.
Confirm that Pointer and Revolution Counter movement is smooth.
Using fingertip, gently press on Contact Point to move the Spindle up and down. Motion of Pointers should be smooth. If it is not smooth make sure Stem is not clamped too tight, and adjust. Also make sure Pointer is stable at set position.
Make sure Spindle axis is perpendicular to measured surface.
If Spindle (Contact Point) is not at a right angle to surface, Gauge will not operate properly and measurement will be inaccurate. Always keep the Spindle axis perpendicular.
 ※When used to check parallelism of Milling
Machine vise, use a Magnetic Base to mount the Gauge perpendicular to surface, and move it out of the way during operation to prevent interference.
HOW TO USE - Comparison Measurements
Set Up Reference Part.
Carefully lift Spindle with fingertip, and, taking care not to hit Spindle from the side, insert the Reference Part or Master under Contact Point.
Set the Origin.
Adjust the gauge mount or rotate the Bezel to set the Gauge to "0".
Remove Reference, and begin measurements.
Remove Reference or Master, careful not to shock Spindle. Insert part to be measured and read the measurement off the Scale.
 ※Setting the Limit Markers
Limit Markers can be moved to show acceptance range
for measurements.
HOW TO USE - Parallelism, Flatness, Runout, etc.
Position Contact Point on surface.
Carefully lifting Spindle with fingertip, and taking care not to hit Spindle from the side, position the surface to be measured under the Contact Point.
Set the Origin.
Adjust the gauge mount, or rotate the Bezel to set the Gauge to "0".
Read the scale as the measured part is moved.
Slowly move the part while monitoring the Pointer and reading the measurement.
Right Angle
Pointer
Limit Markers
Ex.) Pass Range :
0.05 to 0.15mm
Measured Object Measured Object
TROUBLESHOOTING
SPECIFICATIONS
Origin position shifts during measurement.
Temperature changes during measurement can cause repeatability error. Please try the following solutions:
Use in location with constant temperature.
When taking measurements, periodically adjust zero point using a Master reference to correct for temperature
induced drift.
Measurement is not stable, or measurement accuracy is poor.
Contact Point may be worn.
Worn Contact Point will affect accuracy. Periodically check for wear, and replace if wear is affecting measurement accuracy.
Remove any dust or dirt after use. ※Do not lubricate.
Wipe any contamination from Spindle sliding surface using a dry cloth, or cloth moistened with alcohol.To clean other surfaces, wipe with a soft dry cloth, or a cloth moistened with a mild cleaner.
Check for wear of Contact Point.
Measurement accuracy will be affected by worn Contact Point. Regularly check for wear and replace
Contact Point if worn.
Store in provided case in a cool, dark, and dry location.
During storage, make sure there is no force on the Spindle (such as pushed in, or lateral force.
)
Keep away from moisture and direct sunlight, and secure from unauthorized personnel.
I3 60 -K1 1808
Niigata seiki Co., Ltd.
6-15-22, Tsukanome, Sanjo, Niigata, Japan, 955-0055 Tel. : +81-256-31-5670 Fax. : +81-256-39-7730
MAIL intl.sales@niigataseiki.co.jp URL http://www.niigataseiki.co.jp
CALIBRATION
Replacement Contact Point Part No.DI-10KDDI-CPKWDI-10DI-1058WDI-0342WDI-0542DI-CP
         WDI-0558SPWDI-1058SPWDI-0158SPWDI-W
AFTER USE CARE, STORAGE
To maintain measurement accuracy, periodic calibration is recommended.
(
For reference, we recommend a calibration interval of 3~4 months when used in a factory.
)
Outside Japan, Please contact distributor or place of purchase to inquire about calibrations services.
DI-10KD
WDI-10
DI-1058
WDI-0342
WDI-0542
WDI-0558SP
WDI-1058SP
WDI-0158SP
0.01
0.01
0.01
0.01
0.01
0.01
0.01
0.001
5
5
5
6
6
3
3
2
1.5N
1.5N
1.5N
1.5N
1.5N
1.5N
1.5N
2N
1
1
1
0.5
0.5
1
1
0.2
010 010 010 0 3 0 5 0 5 010 0 1
±0-100 ±0-100 ±0-100 ±0-50 ±0-50 ±0-100 ±0-100
0-100-0
Model No.
Scale
Contact
Point
125
180
180
100
100
190
190
190
Weight
(g)
Graduation
(mm)
Measurement
Range (mm)
Retrace
Error
(μm)
Measuring
Force
(N)
Range
(
1 rev.
)
(mm)
8
8
8
8
9
8
10
3
±20 ±20 ±20 ±14 ±16 ±16 ±20
± 5
5
5
5
5
5
3
3
0.3
Adjacent
Error
(
1/10rev.):
(μm)
Full Range
Accuracy
(μm)
Repeat
Accuracy
(μm)
Carbide
Ball
Steel
Ball