■MEASUREMENT
For accurate measurements, please follow the procedure below.
①In cross section perpendicular to the axis of the bore, position the gauge along diameter ⓐ. (Figure: 1
)
This is the maximum distance, and at this point the Indicator will read the maximum value.
The Guide will automatically position the Measurement Head along diameter ⓐ.
※ CDI-10D Does not have a guide, manually position the probe for maximum reading in this direction.
②In cross section parallel to the axis of the bore through diameter ⓐ, position the indicator to minimize the distance ⓑ. (Figure: 2
)
In this position the Indicator will read the minimum value. This position must be manually achieved by pivoting the gauge and noting
the point where the Indicator shows the minimum value. (Figure: 3
)
①Before use, please make sure the Measurement Head components are securely tightened. Tighten as necessary.
②Clean the gauge head and the ID of a Ring Gauge Standard using parts cleaner or cleaning solution.
③Press the [ZERO/ON] Button to turn on power.※Power is turned off with a long press to the [ZERO/ON] Button.
■OPERATION
(
Figure: 1
) (
Figure: 3
)(
Figure: 2
)
①Turn on power. If the “ ”, “ ” symbols are displayed on the LCD, press the [MIN] Button to turn off.
②Make the “SET” value of the Bore Gauge match the ID of the Ring Gauge Standard. Press and hold the [ABS/SET] Button until
“SET” blinks on the display.
③Continue holding the [ABS/SET] Button and the display character to the left will blink. Use a short press of the [ABS/SET] Button to
change the value, and a long press to change the character being modified.
Repeat this process to enter the dimension of the Ring Gauge Standard on the LCD display. (Figure: 4
)
④When the value is entered, the flashing position will return to “SET”.
A short press of the [ABS/SET] Button to turn off the flashing.
⑤Insert the Measurement Head into the Ring Gauge Standard and
press the [MIN] Button to put the Indicator into minimum-mode. Slowly
pivot the gauge (Figure: 3) to display the minimum value.
⑥Remove the gauge from the Ring Gauge and press the [CAL] Button
until “OK” is momentarily displayed.
Preparation for measurement is now complete.
⑦Put the gauge into the workpiece to be measured, and pivot the gauge
slowly. (Figure: 3) The LCD will display the measured value. (Figure: 5
)
※Press the [ON/ZERO] Button momentarily to reset display in
preparation to measure the next workpiece.
ABSOLUTE MODE (ABS
)
※Displays absolute measurement of workpiece.
①Turn on power. If “ ”, or “ ” symbols are displayed on the LCD,
press the [MIN] Button to turn off. If “ABS” is displayed press the [ABS]
Button to turn off.
②Insert the measuring head into the Ring Gauge Standard and press
[
MIN] Button to put the Indicator into minimum-mode. Slowly pivot the
gauge (Figure: 3) to display the minimum value.
③Remove the gauge from the Ring Gauge and press the [CAL] Button
until “OK” is momentarily displayed.
Preparation for measurement is now complete.
④Put the gauge into the workpiece to be measured, and pivot the gauge
slowly. (Figure: 3) The LCD will display the difference between the
measured workpiece and the reference standard. (Figure: 6)
※Press the [ON/ZERO] Button momentarily to reset display in
preparation to measure the next workpiece.
※Display dif ference between workpiece and
reference standard.
COMPARATIVE MEASUREMENT
■TROUBLESHOOTING
ERROR CONDITION CORRECTIVE ACTION
LCD does not display
Displayed value is unstable
Display is not clear
Cycle power by removing and
replacing battery.
Replace battery with new one.
●
●
●
●
●
※If problem persists, or if you have any questions,
please contact distributor or place of purchase.
※Please note, manufacturer is unable to respond
to inquires or provide service directly. Please
contact distributor or place of purchase.
Niigata seiki Co., Ltd.
6-15-22, Tsukanome, Sanjo, Niigata, Japan, 955-0055
Tel. : +81-256-31-5660 Fax. : +81-256-39-7730
MAIL:intl.sales@niigataseiki.co.jp
URL
http://www.niigataseiki.co.jp
I240-K 1506
(
Figure: 4
)
(
Figure: 5
)
(
Figure: 6
)
Ring Gauge
Standard
Adjust Reading
Workpiece
Workpiece Dimension
Workpiece
Difference between
Workpiece and
Reference Standard