15 ELECTROMAGNETIC COMPATIBILITY
GUIDANCE AND MANUFACTURER’S DECLARATION ELECTROMAGNETIC EMISSIONS
The NICE1 is intended for use in the electromagnetic environment specied below. The customer
or the user of the NICE1 should assure that it is used in such an environment.
EMISSIONS TEST COMPLIANCE ELECTROMAGNETIC ENVIRONMENT GUIDANCE
RF emissions
CISPR 11
Group 1 The NICE1 uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
RF emissions
CISPR 11
Class B The NICE1 is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage uctuations/
icker emissions
IEC 61000-3-3
Complies
GUIDANCE AND MANUFACTURER’S DECLARATION ELECTROMAGNETIC IMMUNITY
The NICE1 is intended for use in the electromagnetic environment specied below. The customer
or the user of the NICE1 should assure that it is used in such an environment.
IMMUNITY TEST IEC 60601
TEST LEVEL
COMPLIANCE LEVEL ELECTROMAGNETIC
ENVIRONMENT GUID
ANCE
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete, or ceramic tile.
If oors are covered with
synthetic material,
the relative humidity
should be
at least 30 %.
Electrical fast
transient/burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/output lines
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV dierential mode Mains power quality
should be that of a
typical commercial or
hospital environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT) for 5 s
<5 % UT
(>95 % dip in UT)
for 0,5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT) for 5 s
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the NICE1
requires continued
operation during power
mains interruptions, it is
recommended that the
NICE1 be powered from
an uninterruptible power
supply or a battery.
Power frequency (50/60
Hz) magnetic eld
IEC 61000-4-8
N/A N/A NICE1 does not contain
magnetically sensitive
components
NOTE UT is the a.c. mains voltage prior to application of the test level.