NI FP-AI-100, FP-AI-110, FP-AI-111, FP-RTD-122, FP-TC-120 User Manual

C
ALIBRATINGFIELDPOINT
C
WITH
ALIBRATIONEXECUTIVE

Contents

Introduction.............................................................................................2
What Is Calibration? ........................................................................ 2
Why Should You Calibrate? ............................................................ 2
How Often Should You Calibrate?.................................................. 2
Equipment and Other Test Requirements ...............................................2
Test Equipment ................................................................................ 2
Documentation.................................................................................3
Software ........................................................................................... 4
Test Conditions ................................................................................ 5
Calibration Procedures............................................................................ 5
Configuring Your FieldPoint AI Module for Calibration................ 6
Connecting the Calibrator to an FP-TC-120 Module ...................... 10
Connecting the Calibrator to an FP-AI-1xx Module ...................... 11
Connecting the Calibrator to an FP-RTD-122 Module ................... 12
Running the Calibration Executive Procedure................................. 13
Troubleshooting and Common Questions............................................... 14
Technical Support Resources.................................................................. 15
NI Web Support ............................................................................... 15
Worldwide Support.......................................................................... 15
A
NALOGINPUT
M
ODULES
FieldPoint™, National Instruments™,NI herein are trademarks or trade names of their respective companies.
323124A-01 Copyright © 2001 National Instruments Corp. All rights reserved. July 2001
, and ni.com
are trademarks of National Instruments Corporation. Product and company names mentioned

Introduction

This document contains step-by-step instructions for loading and running National Instruments Calibration Executive to calibrate FieldPoint analog input (AI) modules, including the FP-AI-100, FP-AI-110, FP-AI-111, FP-RTD-122, and FP-TC-120.

What Is Calibration?

Calibration consists of verifying the measurement accuracy of a device
and adjusting for any measurement error. Verification is measuring performance of a device and comparing the results to the factory specifications. Calibration Executive automates the verification and adjustment of National Instruments (NI) measurement devices.

Why Should You Calibrate?

The accuracy of electronic components drifts with time and temperature, which can affect measurement accuracy as the device ages. Calibration restores the device to its specified accuracy and ensures that it still meets National Instruments standards.

How Often Should You Calibrate?

The measurement accuracy requirements of your application determine how often you should calibrate a FieldPoint AI module. NI recommends that you perform a complete calibration at least once every year. You can shorten this interval to 90 days or six months based on the demands of your application.

Equipment and Other Test Requirements

This section describes the equipment, documentation, software, and test conditions you need to calibrate a FieldPoint module using Calibration Executive.

Test Equipment

Calibration requires using a high-precision calibrator with the voltage accuracy in parts per million (ppm) listed in Table 1.
Calibrating FieldPoint AI Modules with Calibration Executive 2 ni.com
NI recommends that you use the following instruments for calibration:

Table 1. Test Equipment Specifications and Recommended Models

FieldPoint Module Voltage Accuracy (ppm) Recommended Calibrator
FP-AI-100 500 ppm Fluke 5500A or better
FP-AI-110 100 ppm Fluke 5500A or better
FP-AI-111 100 ppm Fluke 5500A or better
FP-RTD-122 20 ppm Fluke 5700A
FP-TC-120 30 ppm Fluke 5520A or better
FieldPoint FP-1000 network module
FP-TB-CAL FieldPoint Calibration Terminal Base (part
number 777519-90), a modified FieldPoint base that allows you to
write new calibration constants to the module EEPROM
11-24 V power supply—NI PS-2 power supply
RS-232 serial cable
If a Fluke 5700A is not available, use the accuracy specifications to choose a substitute calibration standard.

Documentation

This section describes the documentation you need to calibrate a FieldPoint AI module. In addition to this calibration procedure, you may need to refer to the following documentation:
Calibration Executive Software User Manual
FP-1000/1001 User Manual
FP-1000/1001 Quick Start Guide
Operating instructions for the AI module
You can download these documents from the NI Web site at
ni.com/manuals
© National Instruments Corporation 3 Calibrating FieldPoint AI Modules with Calibration Executive
.

Software

To calibrate the FieldPoint AI modules with Calibration Executive, you need to install the FieldPoint calibration procedure for Calibration Executive and FieldPoint Explorer on your system.
Complete the following steps to install the FieldPoint calibration procedure for Calibration Executive:
1. Make sure that your computer and monitor are powered on and that you have installed Windows NT/Me/98/95.
2. Close all open applications.
3. Insert the installation CD into the CD-ROM drive.
4. Choose the Run option from the Start menu on the desktop task bar.
5. In the command line box, enter
x:\setup.exe
(wherexis the letter
of the CD-ROM drive you are using), then click OK.
6. Follow the onscreen instructions to complete the installation.
The setup program installs Calibration Executive as well as the associated files listed in Table 2.

Table 2. Calibration Executive Directories and Supporting Files

Directory Name Contents
Calibration Executive \Procedures\FP-XX-XXX\Limits
Calibration Executive\Procedures\FP-XX-XXX
Calibration Executive\ Procedures\FP-XX-XXX\Support Files
Complete the following steps to install FieldPoint Explorer:
1. Insert the FieldPoint Explorer CD into the CD-ROM drive on your computer.
2. Choose the Run option from the Start menu on the desktop task bar.
3. In the command line box, enter of the CD-ROM drive you are using), then click OK.
4. Follow the onscreen instructions to complete the installation.
Microsoft Access database that stores the calibration limits for FieldPoint AI modules (excluding FP-TC-120)
Directory structure that contains the calibration procedures
Required support files
x:\setup.exe
(wherexis the letter
Calibrating FieldPoint AI Modules with Calibration Executive 4 ni.com

Test Conditions

Follow these guidelines to optimize connections and test conditions during calibration:
Keep connections to the module short. Long cables and wires act as antennae, picking up extra noise that can affect measurements.
Use shielded copper wire for all cable connections to the device. Use twisted-pair wire to eliminate noise and thermal offsets.
Maintain a temperature between 18 and 28 °C. To operate the module at a specific temperature outside this range, calibrate the module at that temperature.
Keep relative humidity below 80%.
Allow a warm-up time of at least 15 minutes to ensure that the
measurement circuitry is at a stable operating temperature.

Calibration Procedures

This section explains how to configure and set up the module for calibration, and how to run the Calibration Executive procedure. Refer to Table 3 for the approximate times to complete the procedure in both automated and manual modes.

Table 3. Approximate Time to Complete Calibration Executive Procedure

Module Automated Mode Manual Mode
FP-TC-120 10 minutes 15 minutes
FP-AI-100 70 minutes 120 minutes
FP-AI-110 20 minutes 30 minutes
FP-AI-111 30 minutes 45 minutes
FP-RTD-122 10 minutes 15 minutes
Note
For an explanation of automated versus manual calibration, refer to the Automated Versus Manual Calibration section in Chapter 2, Calibration Executive System Overview, of the Calibration Executive Software User Manual.
© National Instruments Corporation 5 Calibrating FieldPoint AI Modules with Calibration Executive
Calibration consists of two main steps:
1. Configuring the module for calibration, which is described in the
Configuring the FieldPoint AI Module for Calibration section of this
document.
2. Running the procedure to calibrate the module, which is described in the Running the Calibration Executive Procedure section of this document.

Configuring the FieldPoint AI Module for Calibration

To verify the operation of the test fixture, build a FieldPoint bank with a network module, calibration terminal base, power supply, and the FieldPoint AI module you are calibrating. Build the FieldPoint bank as follows:
1. Attach the network module to the calibration terminal base as shown in Figure 1.
Local Bus Connectors
Firmly Mated
Protective
Cover

Figure 1. Connecting FieldPoint Terminal Bases

2. Place the FieldPoint AI module on top of the calibration terminal base.
3. Connect the bank to the host computer using an RS-232 cable.
4. Select the appropriate baud rate that your computer supports on the network interface module. Table 4 shows the available baud rates. The default baud rate for the FP-1000 is 115,200, which provides the fastest performance. You do not need to change this setting unless you experience communications problems.
Calibrating FieldPoint AI Modules with Calibration Executive 6 ni.com

Table 4. Baud Rate Switch Settings for the FP-1000/1001

Switch Positions 6–8 Network Module Baud Rate
123 45678
123 45678
123 45678
123 45678
123 45678
123 45678
300
1,200
2,400
9,600
19,200
38,400
123 45678
115,200
123 45678
© National Instruments Corporation 7 Calibrating FieldPoint AI Modules with Calibration Executive
57,600
5. Connect the positive wire of the power supply into the first V terminal of the network module. Connect the negative wire of the power supply into the first C terminal of the FP-1000 network module (see Figure 2).
+
To DC V Power Supply

Figure 2. FP-1000/1001 Power Connector Pinout

VV
CC
6. Launch FieldPoint Explorer.
7. Expand the + directory tree, which is on the left side of the FieldPoint Explorer window (see Figure 3).

Figure 3. FieldPoint Explorer

8. Right-click the FieldPoint icon and select Add a comm resource to this server.
9. In the Comm Resource Configuration dialog box, which is shown in Figure 4, select the port to which the FieldPoint system is connected. Make sure the baud rate matches whatthe network module is set to use, then click Find Devices.
Calibrating FieldPoint AI Modules with Calibration Executive 8 ni.com

Figure 4. Comm Resource Configuration

10. Expand the view to show the devices found on the network, as shown in Figure 5. If you do not see the module that you want to calibrate, the network module is not communicating correctly with either the host or the FieldPoint module.

Figure 5. Expanding the View of the + Directory Tree

11. Close FieldPoint Explorer before you go on to the next section. You do not need to save the new comm resource.
© National Instruments Corporation 9 Calibrating FieldPoint AI Modules with Calibration Executive
Caution
Yo u must close FieldPoint Explorer and keep it closed during the remainder of
this procedure to prevent a system crash.

Connecting the Calibrator to an FP-TC-120 Module

If you are calibrating the FP-TC-120, make the following connections as showninFigure6:
1. Connect the positive output (voltage) of the calibrator to terminal 1 of the calibration terminal base
2. Connect the negative output (voltage) of the calibrator to terminal 2 of the calibration terminal base.
1
7
VV
CC
6
5
1 FP-1000 Network Module 2 TB-Cal Calibration Terminal Base 3 Connect to positive output (voltage) of calibrator 4 Connect to negative output (voltage) of calibrator
3 4
5 Connect to DC V power supply 6 Connect RS-232 serial cable to COM port 7 Baud Rate Switches
67892345 141516V10 11 12 13V1
2
22 23 24 2518 19 20 21 30 31 32 C26 27 28 29C17

Figure 6. FP-TC-120 Connections to Calibrator

Calibrating FieldPoint AI Modules with Calibration Executive 10 ni.com

Connecting the Calibrator to an FP-AI-1xx Module

If you are calibrating an FP-AI-100, FP-AI-110, or FP-AI-111 module, make the connections shown in Figure 7.
1. Connect the positive output (current) of the calibrator to terminal 2 of the calibration terminal base.
2. Connect the negative output (current) of thecalibrator to terminal 18 of the calibration terminal base.
1
7
VV
CC
6
5
1 FP-1000 Network Module 2 FP-TB-CAL Calibration Terminal Base 3 Connect to positive output (current) of calibrator 4 Connect to negative output (current) of calibrator
Figure 7.
FP-AI-1xx Connections to Calibrator for Current Offset Testing
67892345 141516V10 11 12 13V1
22 23 24 2518 19 20 21 30 31 32 C26 27 28 29C17
3
4
5 Connect to DC V power supply 6 Connect RS-232 serial cable to COM port 7 Baud Rate Switches
2
© National Instruments Corporation 11 Calibrating FieldPoint AI Modules with Calibration Executive

Connecting the Calibrator to an FP-RTD-122 Module

If you are calibrating the FP-RTD-122 module, make the following connections as shown in Figure 8.
1. Connect the positive output (ohms) and positive sense (ohms) of the calibrator to terminal 1 of the calibration base.
2. Connect the negative output (ohms) and the negative sense (ohms) of the calibrator to terminal 18 of the calibration base.
3. On the calibration terminal base, connect terminal 2 to terminal 18.
While the Calibration Executive procedure runs, you are prompted to make additional connections between the calibrator and the module you are calibrating. Follow the onscreen instructions at each prompt.
1
10
9
1 FP-1000 Network Module 2 TB-Cal Calibration Terminal Base 3 Connect terminal 2 to terminal 18 4 Connect to positive output of calibrator 5 Connect to positive sense of calibrator
VV
CC
5
4
8
6 Connect to negative output of calibrator 7 Connect to negative sense of calibrator 8 11-24 DC V Power Supply PS-2 9 RS-232 Serial Cable 10 Baud Rate Switches
3
67892345 141516V10 11 12 13V1
22 23 24 2518 19 20 21 30 31 32 C26 27 28 29C17
7
6
2

Figure 8. FP-RTD-122 Connections to Calibrator

Calibrating FieldPoint AI Modules with Calibration Executive 12 ni.com

Running the Calibration Executive Procedure

To run the Calibration Executive procedure, complete the following steps:
1. Launch Calibration Executive and follow the instructions in the Calibration Configuration Wizard to load the appropriate FieldPoint calibration procedure for the module. Refer to Chapter 1, Introduction
to Calibration Executive,intheCalibration Executive Software User Manual if you need more information on configuring and loading a
calibration procedure.
2. Enter the following information about the installed hardware when prompted by Calibration Executive:
Selected Calibrator InformationThe calibrator you are using
for this calibration procedure (for example, the Fluke 5700A).
Available AddressSelect
mode; otherwise, select the GPIB address of the calibrator.
3. When the procedure is loaded, click Run Procedure to begin.
4. Enter the following information in the first dialog box, which is titled FieldPoint Bank Configuration:
Baud RateThe correct rate from Table 4 to communicate with
the network module.
COM Port NumberThe serial port used to connect to the
FP-1000 network module, for example, COM1, COM2, and so on.
Device NumberThe address of the I/O module you want to
verify or calibrate. The first module to the right of a network module has an address of 1.
If you enter a value incorrectly, Calibration Executive prompts you to re-enter the information.
5. Follow any instructions you receive from Calibration Executive.
if you are using the manual
None
After the procedure finishes, complete the following steps to see the calibration report:
1. Click View»Reports.
2. Select View to launch your browser and view the report. The calibration report appears as a printable HTML file.
You have completed calibrating your FieldPoint AI module with Calibration Executive.
Note
If the FieldPoint AI module fails after calibration, return it to NI for repair or
replacement.
© National Instruments Corporation 13 Calibrating FieldPoint AI Modules with Calibration Executive

Troubleshooting and Common Questions

When I change the baud rate of the FP-1000 and run the Calibration Executive procedure, a dialog box prompts me to re-enter the correct baud rate. What steps do I need to take to do this?
Power down the FP-1000, then power back up again to enable the new settings that you entered.
Some test points fail even after I run the Calibration Executive procedure. What can I do to correct this problem?
Before running the Calibration Executive procedure, allow the FieldPoint bank to warm up at least 15 minutes to ensure accurate readings. Also, make sure you are using shielded copper wire for all cable connections to the module.
In the FP-AI-100 current offset test, my test points are significantly out of specification. What causes this problem?
The voltage terminals (1, 3, 5, 7, 9, 11, 13, and 15) on the calibration base might still be connected together from the voltage offset test. Disconnect all the wires that connect these terminals before attempting the current offset test.
Communication with the FieldPoint AI module times out. What causes this to happen?
The watchdog timer on the FP-1000 might be enabled. Refer to the FP-1000/1001 User Manual to find out how to disable the watchdog timer.
Calibrating FieldPoint AI Modules with Calibration Executive 14 ni.com

Technical Support Resources

NI Web Support

National Instruments Web support is your first stop for help in solving installation, configuration, and application problems and questions. Online problem-solving and diagnostic resources include frequently asked questions, knowledge bases, product-specific troubleshooting wizards, manuals, drivers, software updates, and more. Web support is available through the Technical Support section of

Worldwide Support

National Instruments has offices located around the world to help address your support needs. You can access our branch office Web sites from the Worldwide Offices section of up-to-date contact information, support phone numbers, e-mail addresses, and current events.
If you have searched the technical support resources on our Web site and still cannot find the answers you need, contact your local office or National Instruments corporate. For telephone support in the United States, dial 512 795 8248. For telephone support outside the United States, contact your local branch office:
ni.com
ni.com.
. Branch office Web sites provide
Australia 03 9879 5166, Austria 0662 45 79 90 0, Belgium 02 757 00 20, Brazil 011 284 5011, Canada (Calgary) 403 274 9391, Canada (Montreal) 514 288 5722, Canada (Ottawa) 613 233 5949, Canada (Québec) 514 694 8521, Canada (Toronto) 905 785 0085, China (Shanghai) 021 6555 7838, China (ShenZhen) 0755 3904939, Denmark 45 76 26 00, Finland 09 725 725 11, France 01 48 14 24 24, Germany 089 741 31 30, Greece 30 1 42 96 427, Hong Kong 2645 3186, India 91805275406, Israel 03 6120092, Italy 02 413091, Japan 03 5472 2970, Korea 02 596 7456, Malaysia 603 9596711, Mexico 5 280 7625, Netherlands 0348 433466, New Zealand 09 914 0488, Norway 32 27 73 00, Poland 0 22 528 94 06, Portugal 351 1 726 9011, Singapore 2265886, Spain 91 640 0085, Sweden 08 587 895 00, Switzerland 056 200 51 51, Taiwan 02 2528 7227, United Kingdom 01635 523545
© National Instruments Corporation 15 Calibrating FieldPoint AI Modules with Calibration Executive
Loading...