Accuracy Under Calibration Conditions .................................................................................. 15
Worldwide Support and Services ............................................................................................. 16
ni.com/calibration.
Software
Calibrating the NI 9234 requires the installation of NI-DAQmx 9.2 or later on the calibration
system. You can download NI-DAQmx from ni.com/downloads. NI-DAQmx supports
LabVIEW, LabWindows
NI-DAQmx you only need to install support for the ADE that you intend to use.
TM
/CVI, C/C++, C#, and Visual Basic .NET. When you install
Documentation
Consult the following documents for information about the NI 9234, NI-DAQmx, and your
application software. All documents are available on
software.
NI cDAQ-9174/9178 USB Chassis Quick Start
NI-DAQmx installation and hardware setup
NI 9234 Operating Instructions and Specifications
NI 9234 specific information, specifications, and calibration interval
NI-DAQmx Readme
Operating system and application software support in NI-DAQmx
LabVIEW Help
LabVIEW programming concepts and reference information about NI-DAQmx
VIs and functions
ni.com and help files install with the
NI-DAQmx C Reference Help
Reference information for NI-DAQmx C functions and NI-DAQmx C properties
NI-DAQmx .NET Help Support for Visual Studio
Reference information for NI-DAQmx .NET methods and NI-DAQmx .NET
properties, key concepts, and a C enum to .NET enum mapping table
2 | ni.com | NI 9234 Calibration Procedure
Test Equipment
Table 1 lists the equipment recommended for the performance verification and adjustment
procedures. If the recommended equipment is not available, select a substitute using the
requirements listed in Table 1.
Table 1. Recommended Equipment
Recommended
Equipment
Model
Where UsedRequirements
CalibratorFluke 5700AAccuracy,
Adjustment
Function
Generator
NI 4461Gain
Matching
Phase
Matching
CMRRUse a function generator capable of
DMMNI 4070IEPE
Current
Use a high-precision voltage source
with an accuracy ≤100 ppm and an
output that is able to be not grounded
(floating).
Use a function generator capable of
sourcing 1 kHz and 10 kHz ±4.5 V
pk
AC sinusoidal signals with amplitude
uncertainty of ±10% or less.
Use a function generator capable of
sourcing both 1 kHz and 10 kHz
±4.5 Vpk AC sinusoidal signals with
amplitude uncertainty of ±10% or less.
sourcing 1 kHz ±4.5 Vpk AC sinusoidal
signals with amplitude uncertainty of
±0.04dB or less.
Use DMM in current mode in the
smallest range to measure 2 mA DC
with a measurement uncertainty
≤0.5%.
Source
Measure
Keithley Model
2400
Unit (SMU)
BNC Cable—AllUse a length appropriate for your
ChassisNI cDAQ-9178All—
IEPE
Compliance
Voltage
Use an SMU capable of sourcing 20 V
DC with an accuracy ≤0.1% while
sinking >2 mA DC.
The following setup and environmental conditions are required to ensure the NI 9234meets
published specifications.
•Keep connections to the device as short as possible. Long cables and wires act as antennas,
picking up extra noise that can affect measurements.
•Verify that all connections to the device are secure.
•Use shielded copper wire for all cable connections to the device. Use twisted-pairs wire to
eliminate noise and thermal offsets.
•Maintain an ambient temperature of 23 ±5 °C. The device temperature will be greater than
the ambient temperature.
•Keep relative humidity below 80%.
•Allow a warm-up time of at least 10 minutes to ensure that the NI 9234 measurement
circuitry is at a stable operating temperature.
Initial Setup
Complete the following steps to set up the NI 9234.
1.Install NI-DAQmx.
2.Make sure the NI cDAQ-9178 power source is not connected.
3.Connect the NI cDAQ-9178 to the system safety ground.
a.Attach a ring lug to a 14 AWG (1.6 mm) wire.
b.Connect the ring lug to the ground terminal on the side of the NI cDAQ-9178 using
the ground screw.
c.Attach the other end of the wire to the system safety ground.
4.Install the module in slot 8 of the NI cDAQ-9178 chassis. Leave slots 1 through 7 of the
NI cDAQ-9178 chassis empty.
5.Connect the NI cDAQ-9178 chassis to your host computer.
6.Connect the power source to the NI cDAQ-9178 chassis.
7.Launch Measurement & Automation Explorer (MAX).
8.Right-click the device name and select Self-Test to ensure that the module is working
properly.
4 | ni.com | NI 9234 Calibration Procedure
Verification
The following performance verification procedures describe the sequence of operation and test
points required to verify the NI 9234 and assumes that adequate traceable uncertainties are
available for the calibration references.
Accuracy Verification
Complete the following procedure to determine the As-Found status of the NI 9234.
1.Set the calibrator to Standby mode (STBY).
2.Connect the NI 9234 to the calibrator as shown in Figure 1.
Figure 1. Accuracy Verification Connections to the NI 9234
HI
LO
–
+
AIx
NI 9234Calibrator
3.Verify that the calibrator output is not grounded (floating).
NotesThe analog inputs on the NI 9234 are grounded, so the calibrator must be
not grounded (floating) to prevent ground loop measurement errors.
To float the output of the Fluke 5700A, enable External Guard mode (EX GRD).
4.Set the calibrator voltage to a Test Point value indicated in Table 4.
5.Set the calibrator to Operate mode (OPR).
6.Acquire and average samples.
a.Create an AI voltage task on the NI 9234 according to Table 2.
Table 2. NI 9234 Configuration for Accuracy Verification
Min (V)Max (V)Scaled UnitsChannel Configuration
-55Vo l t sDC Coupled
b.Configure the AI voltage task timing according to Table 3.
8.Repeat steps 3 through 7 for each test point in Table 4.
Table 4. NI 9234 Verification Test Limits for Accuracy
Test Point1-Year Limits
LocationValue (V)Lower Limit (V)Upper Limit (V)
Max4.03.99524.0048
Mid0.0-0.00120.0012
Min-4.0-4.0048-3.9952
NoteThe test limits listed in Table 4 are derived using the values in Table 21.
9.Disconnect the calibrator from the device.
10. Repeat steps 1 through 9 for each channel on the NI 9234.
Gain Matching Verification
Complete the following procedure to determine the As-Found status of the NI 9234.
1.Set the calibrator to Standby mode (STBY).
2.Connect the NI 9234 to the calibrator. Refer to Figure 2 for a connection diagram.
Figure 2. Gain Matching Verification Connections to the NI 9234
HI
LO
3.Verify that the calibrator output is not grounded (floating).
NotesThe analog inputs on the NI 9234 are grounded, so the calibrator must be
not grounded (floating) to prevent ground loop measurement errors.
–
+
AIx
NI 9234Calibrator
To float the output of the Fluke 5700A, enable External Guard mode (EX GRD).
4.Set the calibrator voltage to 4.0 V.
5.Set the calibrator to Operate mode (OPR).
6 | ni.com | NI 9234 Calibration Procedure
6.Acquire and average samples.
Gain dB() 2010log×
V
1V2
–
8
------------------
=
a.Create an AI voltage task on the NI 9234 according to Table 5.
Table 5. NI 9234 Configuration for Gain Matching Verification
Min (V)Max (V)Scaled UnitsChannel Configuration
-55Vo l t sDC Coupled
b.Configure the AI voltage task timing according to Table 6.
Table 6. Timing Configuration
Sample ModeSamples Per ChannelRate (kHz)
Finite Samples1024010.24
c.Start the task.
d.Average the readings and record the average as V
.
1
e.Clear the task.
f.Set the calibrator to Standby mode (STBY).
g.Set the calibrator voltage to -4.0 V.
h.Set the calibrator to Operate mode (OPR).
i.Start the task.
j.Average the readings and record the average as V
.
2
k.Clear the task.
7.Set the calibrator to Standby mode (STBY).
8.Disconnect the calibrator from the device.
9.Perform the following calculation using the recorded V
and V2 values.
1
10. Repeat steps 1 through 9 for each channel.
11. Compare the channel-to-channel gain difference to the maximum limit of ±40 mdB.
If the Gain Matching verification procedure determines that the NI 9234 is outside of the limits,
refer to Worldwide Support and Services for assistance in returning the device to NI.
Complete the following procedure to determine the As-Found status of the NI 9234.
1.Connect the NI 9234 to the function generator as shown in Figure 3.
Figure 3. Phase Matching Verification Connections to the NI 9234
HI
LO
Function
Generator
–
+
AI0
–
+
AI1
–
+
AI2
–
+
AI3
NI 9234
2.Configure function generator according to Table 7.
Table 7. Function Generator Configuration for Phase Matching Verification
Signal TypeAmplitude (Vpk)
Sinewave±4.5
3.Set the function generator frequency to a value indicated in Table 10.
4.Acquire and average samples.
a.Create an AI voltage task on the NI 9234 according to Table 8.
Table 8. NI 9234 Configuration for Phase Matching Verification
Min (V)Max (V)Scaled UnitsChannel Configuration
-55Vo l t sAC Coupled
NoteWait 20 s for the AC coupling filter on the NI 9234 to settle.
b.Configure the AI voltage task timing according to Table 9.
Table 9. NI 9234 Timing Configuration
Sample ModeSamples Per ChannelRate (kHz)
Finite Samples1638451.2
c.Enable the function generator.
d.Start the task.
8 | ni.com | NI 9234 Calibration Procedure
e.Disable the function generator.
f.Clear the task.
5.Calculate the phase for each channel on the NI 9234. NI recommends using the Extract
Single Tone Information VI to calculate phase.
6.Compare the channel-to-channel phase difference to the limits in Table 10.
7.Repeat steps 2 through 6 for each test point in Table 10.
Table 10. NI 9234 Verification Test Limits for Phase Matching
Generator Frequency (kHz)Phase Matching Channel-to-Channel Limit
1±0.085°
10±0.49°
If the Phase Matching verification procedure determines that the NI 9234 is outside of the limits,
refer to Worldwide Support and Services for assistance in returning the device to NI.
Common-Mode Rejection Ratio Verification
Complete the following procedure to determine the As-Found status of the NI 9234.
1.Connect the NI 9234 to the function generator and short the positive pin and negative shell
of the NI 9234 channel.
2.Connect the function generator to the chassis ground as shown in Figure 4.
Figure 4. CMRR Verification Connections to the NI 9234
HI
LO
Function
Generator
–
+
AIx
NI 9234
3.Configure the function generator according to Table 11.
Table 11. Function Generator Configuration for CMRR Verification
a.Create and configure an AI voltage task on the NI 9234 according to Table 12.
Table 12. NI 9234 Configuration for CMRR Verification
Min (V)Max (V)Scaled UnitsChannel Configuration
-55Vo l t sDC Coupled
b.Configure the AI voltage task timing according to Table 13.
Table 13. NI 9234 Timing Configuration
Sample ModeSamples Per ChannelRate (kHz)
Finite Samples1638451.2
c.Enable the function generator.
d.Start the task.
e.Record the measured amplitude as Amplitude
Acq
.
f.Disable the function generator.
g.Clear the task.
5.Perform the following calculation using the recorded Amplitude
Acq
.
h.Compare the calculation result to the minimum CMRR limit of 40 dB.
6.Repeat steps 1 through h for each channel.
If the Common-Mode Rejection Ratio verification procedure determines that the NI 9234 is
outside of the limits, refer to Worldwide Support and Services for assistance in returning the
device to NI.
10 | ni.com | NI 9234 Calibration Procedure
IEPE Current Verification
Complete the following procedure to determine the As-Found status of the NI 9234.
1.Connect the NI 9234 to the DMM as shown in Figure 5.
Figure 5. IEPE Current Verification Connections to the NI 9234
HI
LO
DMMNI 9234
–
+
AIx
2.Configure the DMM for a DC current measurement in the lowest appropriate range to
measure 2 mA.
3.Measure IEPE Source short circuit current.
a.Create an AI voltage task on the NI 9234 according to Table 14.
Table 14. NI 9234 Configuration for IEPE Current Verification
Min (V)Max (V)
Scaled
Units
Channel
Configuration
IEPE Excitation
-55Vo l t sAC CoupledEnabled
b.Configure the AI voltage task timing according to Table 15.
Table 15. NI 9234 Timing Configuration
Sample ModeRate (kHz)
Continuous Samples51.2
c.Start the task.
d.Record the DMM measurement.
e.Clear the task.
4.Compare the DMM measurement to the minimum current limit of 2.0 mA.
5.Repeat steps 1 through 4 for each channel.
If the IEPE Current verification procedure determines that the NI 9234 is outside of the limits,
refer to Worldwide Support and Services for assistance in returning the device to NI.
b.Configure the AI voltage task timing according to Table 19.
Table 19. NI 9234 Timing Configuration
Sample ModeRate (kHz)
Continuous Samples51.2
c.Set the source current on the SMU to the recorded SMU
SourceCurrent
value.
d.Start the task.
e.Enable the SMU.
f.Record the SMU measurement SMU
Voltage
.
g.Disable the SMU.
h.Clear the task.
8.Compare the voltage measurement from the SMU to the minimum voltage limit of 19 V.
9.Repeat steps 1 through 8 for each channel.
If the IEPE Compliance Voltage verification procedure determines that the NI 9234 is outside of
the limits, refer to Worldwide Support and Services for assistance in returning the device to NI.
The following performance adjustment procedure describes the sequence of operations required
to adjust the NI 9234.
Accuracy Adjustment
Complete the following procedure to adjust the accuracy performance of the NI 9234.
1.Set the calibrator to Standby mode (STBY).
2.Connect the NI 9234 to the calibrator as shown in Figure 1.
3.Adjust the NI 9234.
a.Initialize a calibration session on the NI 9234. The default password is
b.Input the external temperature in degrees Celsius.
c.Call the NI 9234 get C Series adjustment points function to obtain an array of
recommended calibration voltages for the NI 9234.
d.Set the calibrator to a reference value determined by the array of recommended
calibration voltages.
e.Set the calibrator to Operate mode (OPR).
f.Call and configure the NI 9234 gain instance of the DAQmx adjustment function
according to Table 20.
Table 20. Adjustment Configuration
Physical ChannelReference Value
cDAQMod8/aixThe reference value from the array of adjustment points
g.Set the calibrator to Standby mode (STBY).
h.Repeat steps d through g for each calibration voltage in the array.
i.Disconnect the NI 9234 from the calibrator.
NI.
j.Short AI+ and AI- on the BNC connector together for the channel under adjustment
on the NI 9234.
k.Take a measurement using the NI 9234 offset instance of the DACQmx adjust function
for the channel under adjustment.
l.Close the calibration session.
4.Repeat steps 1 through 3 for each channel on the NI 9234.
EEPROM Update
When an adjustment procedure is completed, the NI 9234 internal calibration memory
(EEPROM) is immediately updated.
If you do not want to perform an adjustment, you can update the calibration date and onboard
calibration temperature without making any adjustments by initializing an external calibration,
setting the C Series calibration temperature, and closing the external calibration.
14 | ni.com | NI 9234 Calibration Procedure
Reverification
Repeat the Verification section to determine the As-Left status of the device.
NoteIf any test fails Reverification after performing an adjustment, verify that you
have met the Test Conditions before returning your device to NI. Refer to Worldwide
Support and Services for assistance in returning the device to NI.
Accuracy Under Calibration Conditions
The following calibration specifications are valid under the following conditions:
•Ambient temperature 23 ±5 °C
•NI 9234 installed in slot 8 of an NI cDAQ-9178 chassis
•Slots 1 through 7 of the NI cDAQ-9178 chassis are empty
NoteThe test limits listed in Table 4 are derived using the values in Table 21.
Table 21. NI 9234 Accuracy Under Calibration Conditions
Percent of Reading (Gain Error)Percent of Range* (Offset Error)
±0.09%±0.024% (±1.2 mV)
*
Range equals 5.1 V
NoteFor operational specifications, refer to the most recent NI 9234 Operating
Instructions and Specifications online at ni.com/manuals.
The National Instruments website is your complete resource for technical support. At ni.com/
support
self-help resources to email and phone assistance from NI Application Engineers.
Visit ni.com/services for NI Factory Installation Services, repairs, extended warranty, and
other services.
Visit
facilitates technical support and ensures that you receive important information updates from NI.
A Declaration of Conformity (DoC) is our claim of compliance with the Council of the European
Communities using the manufacturer’s declaration of conformity. This system affords the user
protection for electromagnetic compatibility (EMC) and product safety. You can obtain the DoC
for your product by visiting
you can obtain the calibration certificate for your product at ni.com/calibration.
National Instruments corporate headquarters is located at 11500 North Mopac Expressway,
Austin, Texas, 78759-3504. National Instruments also has offices located around the world. For
telephone support in the United States, create your service request at
1 866 ASK MYNI (275 6964). For telephone support outside the United States, visit the
Worldwide Offices section of
provide up-to-date contact information, support phone numbers, email addresses, and current
events.
you have access to everything from troubleshooting and application development
ni.com/register to register your National Instruments product. Product registration
ni.com/certification. If your product supports calibration,
ni.com/support or dial
ni.com/niglobal to access the branch office websites, which
Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for more information on National Instruments trademarks. Other
product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National
Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the
National Instruments Patents Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party
legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance
for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES
NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR
ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable
limited rights and restricted data rights as set forth in FAR 52.227-14s, DFAR 252.227-7014, and DFAR 252.227-7015.