Page 8 of 10
If error continues after replacing the lamps, the bal-
last must be replaced by a qualied person.
Lamp 1/2 = Ballast 1 Lamp 3/4 = Ballast 2
Lamp 5/6 = Ballast 3 Lamp 7/8 = Ballast 4
Lamp 9/10 = Ballast 5 Lamp 11/12 = Ballast 6
71 Color Lamp: 1, 3, 5, 7, 9, 11
78 Color Lamp: 2, 4, 6, 8,10,12
Always change 2 lamps 71/78 together
ERROR 5: will appear if the
the UV Lamps have burned
for 750 hours. (Lamp time).
Between the 740 and 750
hours burned the change alll
lamps message will appear
in the display. Indicating that
all lamps must be replaced
before reaching 750h.
When the 750h burning time of the lamps is reached
the machine will stop and beep, showing ERR5
The change all lamps message will appear.
Replace all lamps with new ones according to the
schedule above.
After changing the lamps reset the timer by pressing
the decreasing button for 5 seconds.
The lamp time will now read 000h.
Machine will be ready for use.
When replacing the lamps, always disconnect the power cable, or switch off the machine to
prevent burns. Let the lamps cool down for several
minutes before replacing them.
Maintenance
General instructions
-
-
-
-
-
Repair or maintenance of the machine, must be
carried out by qualied persons only.
Any repair or maintenance must be carried out with
ocial replacement parts supplied from the manufacturer/distributor.
Before cleaning, please remove the power plug from
the power supply.
For cleaning, please use a soft cloth with mild cleaning agent.
Operation of this machine must be done according
to this instructions for use.
Error message by automatic self-checking
system.
The message “OPEN” will appear on the display when a
user opens the cover during
operation. At the same time,
the machine beeps three
times.
ERROR 1: In case the
temperature around ballast
reaches over 60°C, the
machine stops automatically
and Err1 will appear on the
display.
ERROR 2: When there is a
problem with the temperature sensors, the machine
stops automatically and Err2
will appear on the display.
(contact distributor)
ERROR 3: If there is a
defective lamp, numbers
belonging to the defective
lamp or lamps will change to
red on the display. ERR3 &
Check IFU message will also
appear on the display. When
operation time elapses,
change the indicated lamps.
Always replace both indicat-
ed lamps, to insure constant
UV output. This error will also occur if the voltage is
not set to match the country supplied voltage.
Working Progress
11/12
5/6
9/10
7/8
1/2 3/4
OPEN
10:31
Lamp Time: 004h
Lamp Condition
Check IFU
11/12
5/6
9/10
7/8
1/2 3/4
ERR1
00:31
Lamp Time: 004h
Lamp Condition
Check IFU
11/12
5/6
9/10
7/8
1/2 3/4
ERR2
00:31
Lamp Time: 004h
Lamp Condition
Check IFU
11/12
5/6
9/10
7/8
1/2 3/4
ERR3
00:31
Lamp Time: 004h
Lamp Condition
Check IFU
11/12
5/6
9/10
7/8
1/2 3/4
ERR5
00:31
Lamp Time: 750h
Lamp Condition
Please note that electrical/electronic units must be disposed off separately. Do not dispose with household waste. The black bar on the “garbage bin” symbol indicates that the unit was put into circulation
after August 13, 2005. Please note that the unit is subject to regulation 2002/96/EC (WEEE) and applicable following national laws and must be disposed off accordingly. Please contact your dealer if the unit
needs to be disposed of.