Newtech 6300A Owner's Manual

Page 1
OPERATOR’S MANUAL
VENTILATOR
6300A
NewTech Medical Limited
8400 Normandale Lake Boulevard, Suite 920, Bloomington 55437, Minnesota, U.S.A Tel: 1-952-837-2688 Fax: 1-952-400-8947 http://www.newtech-medical.com E-mail: [email protected]
Page 2
CONTENTS
1 GENERAL ..........................................................................................................................................1
1.1 Main Technical Specification.......................................................................................................1
1.2 Operational environment requirements.......................................................................................2
1.3 Volume and weight......................................................................................................................2
1.4 Storage........................................................................................................................................2
1.5 Symbols.......................................................................................................................................2
2 SYSTEM OPERATION PRINCIPLE...................................................................................................3
2.1 System-Block-diagram................................................................................................................3
2.2 Basic Operation Principle............................................................................................................4
2.3 The rear panel.............................................................................................................................7
3 INSTALLATION AND CHECK............................................................................................................8
3.1 Ventilator Component..................................................................................................................8
3.2 Ventilator Installation...................................................................................................................9
3.3 Check ..........................................................................................................................................9
4 OPERATION.....................................................................................................................................11
4.1 Parameters’ setting....................................................................................................................11
4.2 Operation...................................................................................................................................12
5 OPERATION METHODS OF VARIOUS VENTILATION MODES ...................................................13
5.1 C: also called control respiration...............................................................................................13
5.2 A: Namely assist respiration......................................................................................................13
5.3 SIMV: Synchronization Intermission Mandatory Ventilation......................................................14
5.4 A/C+SIGH..................................................................................................................................14
5.5 SPONT......................................................................................................................................15
6 CLEANING AND MAINTENANCE...................................................................................................16
6.1 General principle.......................................................................................................................16
6.2 Maintenance..............................................................................................................................17
7 TROUBLESHOOTING .....................................................................................................................19
WARRANTY
Page 3
WARNINGS
The following warnings must be read and understood before performing any of the procedures described in this section
z Under no circumstances should this medical device be operated in the presence of flammable
anesthetics or other volatile materials due to a possible explosion hazard.
z Lipid spilled or dripped into the unit may cause damage to the unit or result in an electrical
shock hazard.
z Oxygen vigorously accelerates combustion. Do not use any gauges, valves or other
equipment that have been exposed to oil or grease contamination to avoid violent ignition.
z When verifying the operation of this medical device, do not breathe directly from the machine.
Always use a fresh bacterial filter and test circuit. A hazard to the health of the service person may result.
z Hazardous voltages exist within this medical device. Always observes appropriate safety
precaution when working on the unit while the machine is connected to an electrical power source to prevent possible accidental injurious electrical shock.
z This medical device should never be serviced or the covers removed when connected to a
patient. Intern maintenance may affect the operation of the unit, jeopardizing patient care.
z Disconnect all electrical power, air and oxygen sources before attempting any disassembly.
Failure to do so could result in injury to the service technician and /or damage to the equipment
z Removed and/or replaced. Failure to do so could result in malfunction of this medical device
and/or injury to the patient.
Page 4
IMPORTANCE
You are welcome to purchase and utilize our equipment. The careful and comprehensive service to you is our commitment. Please read thoroughly and carefully the Operation Manual before putting the machine into
operation and strictly follow all the procedures of operation. Our company guarantees the quality of the product. Any time and in any case when you need our
help we are always ready to serve you. One who is not authorized by our company shall not be allowed to open and dismantle the
machine. Please directly contact Service Department in case of your need.
Page 5
6300A Operator’s Manual
- 1 - NewTech Medical Limited
1 GENERAL
6300A microprocessor ventilator is pneumatic driven, electronic-controlled, pressure-limit and time­cycled multifunctional ventilator also with perfect multi-ventilating mode and monitoring system. Its panel configuration refers to the figure.
1.1 Main Technical Specification
1.1.1 The performance of Airway System
a) Safety pressure of the airway system: no more than 6kPa; b) Inspiration Fi - oxygen: no more than 45%
1.1.2 Ventilation mode
a) A/C b) A/C+ SIGH c) SIMV (f / 2, f / 4) d) SPONT
1.1.3 Ventilation Performance
a) Respiration frequency f 6~60 times/min b) I: E Ratio: 1:4, 1:3, 1:2, 1:1.5, 1:1,2:1, ±15% c) Tidal volume 0, 80~1200mL, ±15%
d) Ventilating volume: >18L/min
1.1.4 Monitor Performance
a) Tidal volume digital display with error of ±20% b) Ventilation frequency digital display c) Airway pressure analogue display
1.1.5 Alarm Performance
a) Airway pressure high limit alarm 2~ 6kPa ±15% b) Airway pressure low limit alarm 0~0.5kPa ±100Pa; 0.5 ~ 2kPa ±15% c) Power-off alarm >120s d) Silence <120s
1.1.6 Performance of the whole set
a) Power <50W b) Noise of whole set <65dB(A) c) Electrical safety Meet requirements for I type B mode equipment specified in IEC60601-1 <<Medical Electrical equipment: Part one: General requirement for safety>>
Page 6
6300A Operator’s Manual
NewTech Medical Limited - 2 -
1.2 Operational environment requirements
a) Ambient temperature +5 °C ~ +40°C b) Relative humidity not more than 80% c) Atmospheric pressure 86kPa ~104kPa d) Requirements for Gas Source 0.3MPa ~0.5MPa, Oxygen e) Power supply ~220V, 50Hz f) Warming up no less than 5min
1.3 Volume and weight
a) Volume 376mm×246mm×196mm (L×W×H) b) Weight 9kg
1.4 Storage
The product should be stored in the room with temperature of 0 ~ 40°C, relative humidity of no more than 80%, good ventilation and without corrosion gas.
Note: When the storage conditions are beyond the requirements of operational environment, and the storage state is transferred into operation state, the product only can be used after being stored in environment for over 8 hours.
1.5 Symbols
ON (Power)
Type B equipment
OFF (Power)
Warning or Caution, ISO 7000-0434
Alternating Current
NOTE: refer to the manual, IEC601-1
Direct Current
Dangerous Voltage
Protectively earth
Battery
Equipotential
Buzzer
Page 7
6300A Operator’s Manual
- 3 - NewTech Medical Limited
2 SYSTEM OPERATION PRINCIPLE
2.1 System-Block-diagram
Figure 2-1 6300A Diagram
Page 8
6300A Operator’s Manual
NewTech Medical Limited - 4 -
2.2 Basic Operation Principle
The high-pressure oxygen enters the pressure reduction valve I, and the output pressure from the pressure reduction valve I is stabilized at 0.25MPa (already adjusted before ex-factory). The output gas from the pressure reduction valve I enters the electromagnetic valve. In inspiring, the electromagnetic valve is open; the two-way outputs enter the flow valve and the pressure reduction valve II respectively. The one entering the pressure reduction valve II presses on the diaphragm in the expiration valve by means of the gas pressure to control the opening and closing of the expiration valve, and the output pressure is stabilized at 0.05MPa (already adjusted before ex­factory). The other one enters the flow valve. The tidal volume value could be changed through adjusting the flow valve. The pure oxygen entering the gas line is mixed in accordance with a certain proportion in the gas room. The air enters from the venturi-type valve, and a unidirectional valve is installed in the venturi-type valve to prevent the gas overflowing. The sponge gas entrance is adopted for the unidirectional valve to reduce noise and filtrate the gas to be breathed in. A safety valve is installed on the side face of the gas room to prevent that the air pressure is too high and will make hurt to a patient. When the air pressure exceeds the set value of the safety valve (5.1kPa~6kPa), the safety valve can open automatically and the gas will be discharged from the safety valve. The electromagnetic valve will be closed in expiration, the pressure exerted on the diaphragm of the expiration valve will disappear, and the diaphragm will loose. Thus the gas room connects with the atmosphere directly. The gas compressed in the bellows by the ventilator will discharge to the atmosphere through the expiration valve. The process above will be repeated along with the breath rhythm.
During the above procedures, it is the electromagnetic valve that controls the gas flowing direction, and is controlled by the mainframe board. In inspiration the electromagnetic valve will be opened while in expiration it will be closed. In the Fig. 2, the display plate is used for the digital display for the respiration rate, breath frequency, the tidal volume and the oxygen concentration, the display of the corresponding display lamps and the display of the luminous line of the pressure. The mainframe board is the central processing unit of the ventilator circuit part. It not only controls the opening and closing of the electromagnetic valve, but also receives and processes the pressure signal, the flow signal. It receives the inputs from the panel and sends the signal that will be outputted to display to the display plate. The voltage stabilizing power supply provides the circuit needed voltages (+5V, +12V, −12V). The functions of the subsystems above are set and adjusted by the keys on the panel.
The front and back panels drawings of the ventilator are shown in Figure 2-2, Figure 2-3.
Page 9
6300A Operator’s Manual
- 5 - NewTech Medical Limited
The front panel
Figure 2-2
2.2.1 Display screen
See Fig. 4. Table1
Figure 2-3
Table 2-1
Display symbol’s implication Function
TIDAL VOL Display the tidal volume value.
AIRWAY PRESSURE
Airway pressure display: Reflect the airway pressure change when patients respiration.
MV Display minute ventilation
TOTAL Total frequency display.
TRIG PRESS Trigger display.
Page 10
6300A Operator’s Manual
NewTech Medical Limited - 6 -
2.2.2 Parameter setting key
Figure 2-4
Table 2-2
Parameter Function
RATE Respiratory frequency setting key ranging from six to sixty bpm.
I: E
Inspiration and respiratory ratio setting key, totally six positions: 1:4, 1:3, 1:2, 1:1.5, 1:1, 2:1.
TRIG PRESS Trigger pressure setting key.
Δ
Increase each parameter’s value on screen. (Increasing key)
∇
Decrease each parameter’s value on screen. (Decreasing key)
HI PRESS High airway pressure limit setting key.
LO PRESS Low airway pressure limit setting key.
POWER
Power Indicator.
2.2.3 Function key
Figure 2-5
Table 2-3
Symbol Function
A/C
Intermittent positive pressure ventilating key. Synchronous intermittent positive pressure ventilating key.
A/C+ SIGH Sigh key.
SIMV f/2
Synchronous intermittent positive pressure mandatory ventilating key, at this moment, the respiratory frequency is one-second of the A/C ventilating frequency.
SIMV f/4
Synchronous intermittent positive pressure mandatory ventilating key, at this moment, the respiratory frequency is one-fourth of the A/C ventilating frequency.
SPONT Spontaneous ventilation key.
Page 11
6300A Operator’s Manual
- 7 - NewTech Medical Limited
2.2.4 Alarm Information Display
Figure 2-6
Table 2-4
Symbol Function
Oxygen Failure
Supply Oxygen Failure Alarm Display.
HI PRESS
Airway High Pressure Alarm Display.
LO PRESS Airway Low Pressure Alarm Display.
Apnea Apnea Alarm
Tidal Volume Tidal volume regulating knob. Regulate ranging :0, 80ml~1200ml
2min Eliminate alarm. When alarming, press the key to eliminate alarm.
2.2.5 Others
Figure 2-7
2.3 The rear panel
Oxygen
Do Not Block
Air Inlet
Pressure Relief Valve
Power Switch250V0.5A 5X20(T)
Beep
名称:呼吸机
出厂日期:
地址:北京市石景山区八大处高科技园区双园路9号
北京谊安世纪医疗器械有限公司
出厂编号:
型号:
产品注册号:
功率:
频率:
电压:
Figure 2-8
Page 12
6300A Operator’s Manual
NewTech Medical Limited - 8 -
3 INSTALLATION AND CHECK
3.1 Ventilator Component
Figure 3-1
1 Ventilator 2 Machine Arm 3 Power Cable For Ventilator 4 Check Valve 5 Oxygen Pipeline 6 Cart 7 Wheels 8 Bag 9 Water Trap 10 Lines For Measure Temperature 11 Pressure Sampling Tube And Joint Of Y Pipe 12 Humidifier 13 Threading Pipe 14 Respiratory Valve
Page 13
6300A Operator’s Manual
- 9 - NewTech Medical Limited
3.2 Ventilator Installation
3.2.1 Take out all parts of ventilator from the box
3.2.2 Main-machine connection
See Figure 3-1
3.2.3 Power supply connection
There is a power socket box behind the cart of the ventilator, whose input cable plug of 3m long can be connected to the power (~220V, 50Hz) socket of the ward, while the ventilator power plug, humidifier plug, monitor (optional) power plug is plugged in the power socket. After it is turned on, the indicating lamp will light up, the voltmeter will indicate the voltage value. The 6300A ventilator is used under the ~220V ±10% power supply. If the power supply voltage in the ward varies greatly, a voltage-stabilizing power supply should be used.
The fuse of the 6300A ventilator is at the power socket position of the rear panel. The fuse, with the value of 2A, can be replaced after the power cable is unplugged and the fuse box rear cover is opened with a small dash-line screwdriver.
3.2.4 Gas source connection
The 6300A ventilator need compressed oxygen for operation. For the hospital with centralized oxygen supply, the pipe can be connected to the air and oxygen socket on the wall directly. For the hospital without centralized oxygen supply, generally the compressed oxygen is obtained from the oxygen bottle, with a pressure reducer connected to the output of the oxygen bottle.
3.3 Check
3.3.1 Ventilator and relevant parts’ check
Operators should check the machine before operating the ventilator to ensure the perfect.
3.3.2 Power-off alarming check
Switching on the ventilator’s power, then off, should cause sound alarm.
Page 14
6300A Operator’s Manual
NewTech Medical Limited - 10 -
3.3.3 Switch on the ventilator’s power
Set the ventilator standard state Upper pressure limit at 6kPa Lower pressure limit at 0.5kPa The respiratory frequency at 20bpm The ventilation mode as A/C I: E as 1:2 The tidal volume 700ml
3.3.4 Airway upper pressure limit alarming function check
Adjust the upper pressure setting key, the display value is a little lower than the airway pressure peak value, set out the sound-light alarm, then the machine turns into respiration immediately with decrease of the airway pressure.
3.3.5 Airway lower pressure limit alarming function check
Take off the inspiration channel hose. After ten to fifteen seconds sound-light alarm occurs.
3.3.6 Trigger pressure check
Set the trigger pressure at 0.1cm H2O. Pinch the simulated lung with both hands. When the airway pressure is a little lower than the preset value, trigger inspiration and the indicating light to flash.
3.3.7 SIMV (f/2, f/4) function check
Set the respiratory frequency at twenty times/min. and the ventilating mode as A/C. Choose the respiratory function to be SIMV f/2 or SIMV f/4. After one minute, observe TOTAL reading ten times/min. or five times/min.
3.3.8 Spontaneous function check
Set the function at Spontaneous and the trigger voltage at -3cmH2O. Pinch the simulated lung with hand to form negative pressure; at this time the ventilator should deliver gas. When airway pressure reaches 3cmH
2
O, start ventilation.
3.3.9 A/C+SIGH function check
First set the ventilating mode as A/C, turn the flow regulating knob to the minimum, then change the ventilating mode to SIGH, following that is one sigh with 1.5 times of the tidal volume.
Page 15
6300A Operator’s Manual
- 11 - NewTech Medical Limited
4 OPERATION
This chapter will mainly introduce the detailed operating method of 6300A ventilator.
4.1 Parameters’ setting
Table 4-1
Parameter key Operating method
Tidal Volume
Turn the knob to change the tidal volume value. Increase by clockwise and decrease by counter-clockwise.
I: E
Inspiration and respiratory ratio setting: When pressing I: E key, the light at the left upper corner of the key flashes and the revised part flashes. At this moment press Δ or ∇ key, and set the I: E mode.
RATE
Respiratory frequency setting: When pressing RATE key, the light at the left upper corner of the key flashes and the revised part flashes. At this moment press Δ or ∇ key, and set the respiratory frequency.
HI PRESS
Airway upper pressure limit setting: When pressing HI PRESS key, the light at the left upper corner of the key flashes and the revised part flashes. At this moment press Δ or ∇ key, and set the airway upper pressure limit.
LO PRESS
Airway lower pressure limit setting: When pressing LO PRESS key, the light at the left upper corner of the key flashes and the revised part flashes. At this moment press Δ or ∇ key, and set the airway lower pressure limit.
TRIG
Trigger pressure setting: When pressing TRIG key, the light at the left upper corner of the key flashes and the revised part flashes. At this moment press Δ or ∇ key, and change the trigger pressure value.
2min
Eliminate alarm When alarming, press the key to eliminate alarm.
Page 16
6300A Operator’s Manual
NewTech Medical Limited - 12 -
4.2 Operation
Table 4-2
Parameter key Operating method
A/C
Control ventilation: When pressing the key, the light at the left upper corner of the key flashes. At this mode, in case of the patients’ respiratory condition, ventilate for the patients according to the pre-adjusted ventilating parameters.
Assist ventilation: When pressing the key, the light at the left upper corner of the key flashes. At this mode, inspiration depends on patients. Other parameters ventilate for the patients according to the pre-adjusted ventilating parameters.
A/C+ SIGH
Sigh: When pressing the key, the light at the left upper corner of the key flashes. During A/C, every certain time provide double tidal volume once.
SIMV f/2
Synchronous intermittent mandatory ventilation: When pressing the key, the light at the left upper corner of the key flashes. Patient controls the spontaneous respiration frequency, but the value is one-second of the preset A/C frequency.
SIMVf/4
Synchronous intermittent mandatory ventilation: When pressing the key, the light at the left upper corner of the key flashes. Patient controls the spontaneous respiration frequency, but the value is one-fourth of the preset A/C frequency.
SPONT
When pressing the key, the light at the left upper corner of the key flashes. The patient spontaneous respiration via continuous positive pressure gas flow system.
Page 17
6300A Operator’s Manual
- 13 - NewTech Medical Limited
5 Operation Methods of Various Ventilation
Modes
5.1 C: also called control respiration
The characteristic is that patient cannot control releasing airflow, and all respiratory movement all must be carried out according to the conditions set by ventilator. Operation methods is as following: a) Set the breath frequency f. b) Set the ventilation volume c) Upper limit setting for airway pressure: it can be set higher when turn on the ventilator, read the airway pressure peak value from the airway manometer after ventilator runs normally in A/C mode, then set the upper pressure limit 1kPa higher than the airway pressure peak value.
Figure 5-1 Control Ventilation Waveform
5.2 A: Namely assist respiration
It is featured by that the patient can control the respiration frequency, while the tidal volume of the respiration, the time of the inspiration and expiration are still controlled by ventilator. To the patients who remain fully conscious and have spontaneous respiration ability, but can’t conduct enough breath function, the assist respiration of the ventilator is needed. a) Control the ventilation frequency f, the set value should be lower several times every minute than f
total,
b) Set the burst electrical level: adjust around –0.3kPa. To A, several above-mentioned parameters must be adjusted properly and must assist with vigor analysis. Other parameters adjustments are all same to C.
Figure 5-2 Assist Respiration Waveform
Page 18
6300A Operator’s Manual
NewTech Medical Limited - 14 -
5.3 SIMV: Synchronization Intermission Mandatory Ventilation
This is a kind of method combined by patient independence breath and machine mandatory ventilation. The mandatory ventilation is touching off synchronous. It is used in the procedure from mandatory ventilation to independence breath before carrying down machine. a) SIMV f/2: mandatory ventilation frequency is one half of A/C ventilation frequency. The mandatory ventilation of SIMV and independence breath is both touch off synchronous with patient, so the touch off electrical level should be set. Normal it is adjusted around –0.3kPa. The independence breathe when Other parameter settings are same as that of A/C. b) SIMV f/4: The command ventilation frequency is one fourth of A/C ventilation frequency. Others are same as that of f/2.
Figure 5-3 SIMV Waveform
5.4 A/C+SIGH
SIGH is that on the basis of A/C make a deep inspiration of double tidal volume every 100 times, which is adapted to the long term mechanically ventilation patient and can be used in “expanding lung” of thorax operation. When expanding lung, because requires continuous several sighs it is needed to convert ventilation method conversion switch several times between A/C and SIGH. When signing because tidal volume is doubled and the peak value of airway press increases, the upper limit set value of airway press will increase. That is the peak value is higher 1kPa than the value of sighing, the set of other parameters are same to A/C.
Page 19
6300A Operator’s Manual
- 15 - NewTech Medical Limited
5.5 SPONT
In some hospitals, after patient have spontaneous respiration, select SPONT should be provided for the patient through facemask before offline so as to train the spontaneous respiration ability of the patient. As a routine, the methods for realizing persistent gas flow are as follows:
Figure 5-4 Spontaneous Waveform
Page 20
6300A Operator’s Manual
- 16 - NewTech Medical Limited
6 CLEANING AND MAINTENANCE
6.1 General principle
Generally, the gas in the inspiration circuit (the part before the humidifier) is dry and clean, and flows to the patient with a certain velocity, so it is unsuitable for the bacteria to grow. While the gas in the expiration circuit is from patient, it is humid, and some patients have infectious disease, which is easy for the bacteria to increase, so the cleaning and disinfectant operations are specified in principle as follows.
6.1.1 Non-infective patient
Just conduct disinfectant for the connectors such as patient circuit (silicon rubber threaded pipe), water trap and Y joint out of the airway for each patient. The disinfectant method can be fumigating, infection liquid soaking or high-temperature sterilization, decided in accordance with the hospital's conditions.
6.1.2 Infective patient
Besides conducting disinfectant for the patient circuit, the gas delivery element (in the airway tank) of the expiration circuit should also be disinfected so as to avoid cross infestation between patients and prevent the ward from being polluted so that the health of staffs is affected. Refer to the above for the disinfectant method. But the flow sensor should be paid special attention, as it is precision and damageable part. For disinfectant, it is better to soak it in alcohol for an hour, and then dry it. While soaking, the plug part of the sensor should be upward and exposed in the air, it will be OK if the alcohol height can drown the ventilation hole. When positioning the disinfected sensor, pay attention to positioning the side marked with arrowhead upward (the arrow head means the flow direction).
6.1.3 Surface cleaning and sterilization
Clean the machine's panel and all surfaces with soft cloth soaked with the soluble sterilizing agent. The formation and preparation of the agent must be done in accordance with the direction given by the manufacturer. One must prevent the sterilizing agent drops from entering the anesthetic machine and the misusage of organic agent for cleaning the machine.
6.1.4 Expiratory valves cleaning
Dismount the cover of the expiratory valves by rotating it counter clockwise, then clean all parts of them with the gauze socked with water soluble sterilizing agent, after all parts cleaned and dried recover it in original integration. Then one must check the leakage and the movement of the expiratory valves in accordance with the required regulation and checking procedure. Please handle all parts with care preventing any damage.
Page 21
6300A Operator’s Manual
- 17 - NewTech Medical Limited
6.2 Maintenance
6.2.1 Maintainability
The 6300A ventilator is designed to place reliability at the principal position, and it is required that the mean time of break free (MTBF) of the whole machine be greater than 2000 hours. But human operates the machine; the operating life of the machine in a great extent is dependent on the human factors. Correct operation and careful maintenance will greatly prolong the life of the machine, and reduce the appearance of faults. The repairing operation is divided into maintenance and repair. By maintenance it is meant the fault location and removing after fault appears. The repair is under the charge of the repairing personnel from the factory or specified maintenance point; the user should not carry out repairing by him. For the doctor or nurse who operates the ventilator, stress should be laid on the regular maintenance of the machine.
6.2.2 Disinfect, cleaning, drying and correctly install of the easy damageable part.
The pipeline for connecting the machine and patient is easy to be polluted, so disinfectant is needed, and the method is shown in the operation manual. The airway channel、connector and sensor, etc. After disinfectant should be correctly installed in position without gas leakage. After installation, trial running should be conducted, and the machine can be connected to patient and used only after it is operated normally. It should be specially specified that the arrowed head on the flow sensor means the flow direction, and it should not be installed in reverse. Also the sensor of the inspiration circuit and that of the expiration circuit can’t be mistaken. Generally, the expiration channel is easy to be polluted, thus just disinfecting the expiratory flow sensor is OK. After disinfectant and cleaning, it must be dried, otherwise, if there is water on the wire mesh, the resistance will increase, the expiration flow will become inaccurate, thus the reading on the expired minute volume meter will be affected. Those parts that directly touch the patient's body are easily polluted, the sterilization must be done after every use as explained before. It must be here emphasized is that: One must do strictly the correct installation after every cleaning and sterilization. The recovered machine must be tested and ensured in good condition before putting in reuse for patients.
6.2.3 Maintaining good contact for connectors
The connector comprises airway connector and circuit connector. For all the joints, rubber connectors and delivering pipes of the airway, good contact should be ensured so as to guarantee the gas circulating without gas leakage. For all the connectors in the circuit, check should be conducted before operation, especially connectors of inspiration valve and expiration valve, as they have no locking device, main check should be carried out to prevent them from loosening.
6.2.4 Correctly operation of panel knob
The panel knob should be operated carefully with slight motion. In special, the potentiometer knobs should not be screwed with force when the terminal position is attained as this may lead to dislocation. After dislocation, the set value is not accurate, which even may cause abnormal operation of the machine.
Page 22
6300A Operator’s Manual
- 18 - NewTech Medical Limited
6.2.5 Maintenance during operation and transportation
The location of the machine at the operation place (such as operating room, ward, monitor room) should be appropriate, so that the doctor or nurse may not touch the machine during operation, therapy or nursing of the patient. Special the airway pipeline and power supply cable on the ground should not affect walking of person so as to avoid them being contacted and causing gas and electricity failure. The knob on the panel should not be misplaced due to involuntary contact to make the set value abnormal. In transporting the machine, especially going up and down the stairs, it should be protected to avoid the cart from dumping and damaging the valuable machine.
Page 23
6300A Operator’s Manual
- 19 - NewTech Medical Limited
7 TROUBLESHOOTING
Table 7
Troubles Causes Methods
Corrugated hose damaged o
r
connector loose
Replace or reinstall properly
Sensor connector loose Reinstall Patients’ respiratory loop leaks Check as the above method
Patients breathing system leaks
Flow regulating switch closes Turn on the flow regulating
switch
Pressure is too high in pneumatic system under “ Manual breathe”
Semi-tight valve regulated improperly
Regulate again properly
Ventilator doesn't work normally
Source pressure is too low Regulate source pressure Power off Switch on the power source
Digital number tube doesn’t flash, Ventilator doesn’t work
Fuse melts Replace
During operation, ventilator stops suddenly, indicating light goes out with sound-alarm
Power off
Shift to be artificial ventilating mod
Airway upper pressure limi
t
setting too low
Reset
Upper airway pressure limit alarms
Airway block Remove Airway lower pressure limi
t
setting too high
Reset
Airway leaks or hoses and channels loose
Check or reconnect
Lower airway pressure limit alarms
Patients’ compatibility changes Pressure sensor connecting
hose loose
Reconnect
Pressure displays zero or no readings
Pressure sampling hole blocks Remove Flow sensor damaged Contact our company
Tidal volume displays lower or doesn’t display
Ventilator in trouble Contact our company
Page 24
6300A Operator’s Manual
NewTech Medical Limited - 20 -
VENTILATOR
WARRANTY
NewTech Medical Limited (NewTech) guarantees to the customers
that NewTech VENTILATOR machines to be free from defects for a
period of eighteen (18) months and accessories for a period of three
(3) months. The warranty period starts at the date of purchase.
The warranty is void if the ventilator machines have not been used
in accordance with the operation manual or improperly
disassembled by anyone other than NewTech.
Vinson Lau-President
NEWTECH MEDICAL LIMITED
Http://www.newtech-medical.com
Loading...