Newport PR50, PR50PP, PR50CC User Manual

Page 1
Compact High-Speed
Rotation Stages
PR50 Series
USERS MANUAL
G
U
R
A
N
T
E
E
D
S
P
E
C
I
F
I
C
A
T
I
O
N
S
Page 2
EDH0176En1051 — 12/18 ii
PR50 Series Compact High-Speed Rotation Stages
Warranty
Newport Corporation warrants this product to be free from defects in material and workmanship for a period of 1 year from the date of shipment. If found to be defective during the warranty period, the product will either be repaired or replaced at Newport’s discretion.
To exercise this warranty, write or call your local Newport representative, or contact Newport headquarters in Irvine, California. You will be given prompt assistance and return instructions. Send the instrument, transportation prepaid, to the indicated service facility. Repairs will be made and the instrument returned, transportation prepaid. Repaired products are warranted for the balance of the original warranty period, or at least 90 days.
Limitation of Warranty
This warranty does not apply to defects resulting from modification or misuse of any product or part.
This warranty is in lieu of all other warranties, expressed or implied, including any implied warranty of merchantability or fitness for a particular use. Newport Corporation shall not be liable for any indirect, special, or consequential damages.
by Newport Corporation, Irvine, CA. All rights reserved.
Original instructions.
No part of this document may be reproduced or copied without the prior written approval of Newport Corporation. This document is provided for informa ti on only, and pro du ct specifications are subject to change without notice. Any change will be reflected in future publishings.
CAUTION
Warranty does not apply to damages resulting from:
• Incorrect usage:
– Load on the stage greater than maximum specified load.
– Carriage speed higher than specified speed.
–Improper grounding.
¬ Connectors must be properly secured.
¬ When the load on the stage represents an electrical risk, it must
be connected to ground.
– Excessive or improper cantilever loads.
• Modification of the stage or any part thereof.
CAUTION
Please return equipment in the original (or equivalent) packing.
You will be responsible for damage incurred from inadequate packaging if the original packaging is not used.
© 2018
Page 3
iii EDH0176En1051 — 12/18
PR50 Series Compact High-Speed Rotation Stages
Table of Contents
Warranty .................................................................................................................ii
EC Declaration of Conformity...............................................................................v
Definitions and Symbols.......................................................................................vi
Warnings ...............................................................................................................vii
Caution .................................................................................................................viii
1.0 — Introduction.................................................................................1
2.0 — Description ...................................................................................2
2.1 Design Details ............................................................................................2
3.0 — Characteristics............................................................................3
3.1 Definitions ..................................................................................................3
3.2 Mechanical Specifications .......................................................................4
3.3 Load Specification Definitions.................................................................4
3.4 Load Characteristics and Stiffness .........................................................4
3.5 Stage Weights ............................................................................................5
4.0 — Drives and Motors ....................................................................5
4.1 Stepper Drive Version ..............................................................................5
4.2 DC-Motor Drive Version ...........................................................................6
5.5 Mechanical Zero ........................................................................................7
5.6 Feedback Signal Position (PR50CC Only) ..............................................7
4.5 Pinouts........................................................................................................8
5.0 — Connection to Newport Controllers ................................9
5.1 Warnings on Controllers ..........................................................................9
5.2 Connection ...............................................................................................10
5.3 Cables .......................................................................................................10
6.0 — Connection to Non-Newport Electronics ....................12
6.1 Connections .............................................................................................12
7.0 — Dimensions.................................................................................13
8.0 — Accessory: EQ45 Bracket ....................................................13
Page 4
EDH0176En1051 — 12/18 iv
PR50 Series Compact High-Speed Rotation Stages
9.0 — Maintenance ..............................................................................14
9.1 Maintenance ............................................................................................14
9.2 Repair .......................................................................................................14
9.3 Calibration ...............................................................................................14
Service Form .........................................................................................................15
Page 5
v EDH0176En1051 — 12/18
PR50 Series Compact High-Speed Rotation Stages
PR50 Series
EU Declaration of Conformity
following Annex II-1A of Directive 2006/42/EC on machinery
The manufacturer:
MICRO-CONTROLE Spectra-Physics, 9, rue du bois sauvage F-91055 Evry FRANCE
Hereby declares that the machinery:
y Description: " PR50 " y Function: Compact High-Speed Rotation Stages y Models: PR50CC/PP
– the technical file of which was compiled by:
Mr Hervé LE COINTE , Quality Director, MICRO-CONTROLE Spectra-Physics, Zone Industrielle - B.P.29 F-45340 Beaune La Rolande France
– complies with all the relevant provisions of the Directive 2006/42/EC on machinery. – complies with all the relevant provisions of the Directive 2014/30/EU relating to electro­magnetic compatibility. – complies with all the relevant provisions of the Directive 2011/65/EU relating to RoHS2.
– was designed and built in accordance with the following harmonised standards:
y NF EN 61326-1:2013 « Electrical equipment for measurement, control and laboratory use – EMC requirements – Part 1: General requirements » y NF EN 55011:2010/A1:2011 Class A
– was designed and built in accordance with the following other standards:
y NF EN 61000-4-2 y NF EN 61000-4-3 y NF EN 61000-4-4 y NF EN 61000-4-5 y NF EN 61000-4-6
ORIGINAL DECLARATION
Done in Beaune La Rolande on 16 May 2017 Hervé LE COINTE Quality Director
DC1-EN rev:A
EC Declaration of Conformity
Page 6
EDH0176En1051 — 12/18 vi
PR50 Series Compact High-Speed Rotation Stages
Definitions and Symbols
The following terms and symbols are used in this documentation and also appear on the product where safety-related issues occur.
General Warning or Caution
The exclamation symbol may appear in warning and caution tables in this document. This symbol designates an area where personal injury or damage to the equipment is possible.
The following are definitions of the Warnings, Cautions and Notes that may be used in this manual to call attention to important information regarding personal safety, safety and preservation of the equipment, or important tips.
WA RN IN G
Warning indicates a potentially dangerous situation which can result in bodily harm or death.
CAUTION
Caution indicates a potentially hazardous situation which can result in damage to product or equipment.
NOTE
Note indicates additional information that must be considered by the user or operator.
European Union CE Mark
The presence of the CE Mark on Newport Corporation equipment means that it has been designed, tested and certified as complying with all applicable European Union (CE) regulations and recommendations.
Warnings and Cautions
ATTENTIO N
This stage is a Class A device. In a residential environment, this device can cause electromagnetic interference. In this case, suitable measures must be taken by the user.
Page 7
vii EDH0176En1051 — 12/18
PR50 Series Compact High-Speed Rotation Stages
Warnings
W
ARNING
The motion of objects of all types carries potential risks for operators. Ensure the protection of operators by prohibiting access to the dangerous area and by informing the personnel of the potential risks involved.
WA RN IN G
Do not use this stage when its motor is emitting smoke or is unusually hot to the touch or is emitting any unusual odor or noise or is in any other abnormal state.
Stop using the stage immediately, switch off the motor power and then disconnect the electronics power supply.
After checking that smoke is no longer being emitted contact your Newport service facility and request repairs. Never attempt to repair the stage yourself as this can be dangerous.
WA RN IN G
Make sure that this stage is not exposed to moisture and that liquid does not get into the stage.
Nevertheless, if any liquid has entered the stage, switch off the motor power and then disconnect the electronics from power supply.
Contact your Newport service facility and request repairs.
WA RN IN G
Do not insert or drop objects into this stage, this may cause an electric shock, or lock the drive.
Do not use this stage if any foreign objects have entered the stage. Switch off the motor power and then disconnect the electronics power supply.
Contact your Newport service facility for repairs.
WA RN IN G
Do not place this stage in unstable locations such as on a wobbly table or sloping surface, where it may fall or tip over and cause injury.
If this stage has been dropped or the case has been damaged, switch off the motor power and then disconnect the electronics power supply.
Contact your Newport service facility and request repairs.
WA RN IN G
Do not attempt to modify this stage; this may cause an electric shock or downgrade its performance.
WA RN IN G
Do not exceed the usable depth indicated on the mounting holes (see section “Dimensions”). Longer screws can damage the mechanics or cause a short-circuit.
Page 8
EDH0176En1051 — 12/18 viii
PR50 Series Compact High-Speed Rotation Stages
Caution
CAUTION
Due to its very low profile, the PR50 stage contains pressure sensitive components. For usage at a pressure level below 0.25 bar, please contact our sales department.
CAUTION
Do not place this stage in a hostile environment such as X-Rays, hard UV,… or in any vacuum environment.
CAUTION
Do not place this stage in a location affected by dust, oil fumes, steam or high humidity. This may cause an electric shock.
CAUTION
Do not leave this stage in places subject to extremely high temperatures or low temperatures. This may cause an electric shock.
• Operating temperature: +10 to +35 °C
• Storage temperature: -10 to +40 °C (in its original packaging)
CAUTION
Do not move this stage if its motor power is on.
Make sure that the cable to the electronics is disconnected before moving the stage. Failure to do so may damage the cable and cause an electrical shock.
CAUTION
Be careful that the stage is not bumped when it is being carried. This may cause it to malfunction.
CAUTION
When handling this stage, always unplug the equipment from the power source for safety.
CAUTION
When the carriage is in its end-of-run position, it is strongly recommended not to go beyond this point as this may damage the stage mechanism.
CAUTION
Contact your Newport service facility to request cleaning and specification control every year.
Page 9
1 EDH0176En1051 — 12/18
PR50 Series Compact High-Speed Rotation Stages
Compact High-Speed Rotation Stages PR50 Series
1.0 —Introduction
This manual provides operating instructions for the rotation stage that you have purchased in the PR50 Series:
REMARK
Due to its very low profile, the PR50 stage contains pressure sensitive components. For usage at a pressure level below 0.25 bar.
PR50 rotation stage.
RECOMMENDATION
We recommend you read carefully the chapter “Connection to electronics” before using the PR50 rotation stage.
PR50 rotation stage and its 2 retaining rings.
• PR50PP • PR50CC
Page 10
EDH0176En1051 — 12/18 2
PR50 Series Compact High-Speed Rotation Stages
2.0 —Description
The PR50 Series low profile rotation stages are designed to rotate up to 1
in. diameter optical components, such as polarizers, wave plates, or wedges. The small size (only 60 mm width x 110 mm length x 21 mm height) allows use in confined spaces such as laser cavities, disk texturing machines, or manufacturing areas for optical components.
Similar to our higher-resolution rotation stages SR50, the PR50 stages fea­t
ure a low-profile, lightweight design based on an aluminum body construc­tion with preloaded ball bearings. The motor is folded inside the outer body providing a small footprint.
The PR50 stages are a cost-efficient alternative to the SR50 for applications that do not require the high-resolution and precision of the SR50, or for applications that could benefit from higher speeds up to 20 °/s. The stages feature angular resolution of 0.01°, which may be achieved with either DC­Servo or open-loop stepper motors in half-step mode. The stages are equipped with a hardware origin and allow for continuous rotation of 360°.
For optimal performance, we recommend the use of our motion con­trollers.
The PR50 rotation stages are equipped with a 3-meter cable for connection to our motion controllers.
2.1 Design Details
Base Material Aluminum Bearings Ball bearings Drive Mechanism Ground worm gear Worm Gear Ratio 1:63 Reduction Gear 11.8642:1 Feedback PR50CC: Motor mounted rotary encoder, 48 pts/rev
PR50PP: Open-loop, no encoder Origin Optical Motor PR50CC: DC servo motor UE17CC
PR50PP: 2-phase stepper motor UE16PPSC, 1 Full step = 0.002° Cable 3 m
Page 11
3 EDH0176En1051 — 12/18
PR50 Series Compact High-Speed Rotation Stages
3.0 —Characteristics
3.1 Definitions
Specifications of our products are established in reference to ISO 230 standard part II “Determination of accuracy and repeatability of positioning numerically controlled axes”.
T
his standard gives the definition of position uncertainty which depends
on the 3 following parameters:
Absolute Accuracy
Difference between ideal position and real position.
Accuracy
Difference between ideal position and real position after the compensation of linear errors.
Linear errors include: cosine errors, inaccuracy of screw or linear scale pitch, angular deviation at the measuring point (Abbe error) and thermal expansion effects. All Newport motion electronics can compensate for linear errors.
The relation between absolute accuracy and on-axis accuracy is as follows:
Absolute Accuracy = Accuracy + Correction Factor
x Travel
Repeatability
Ability of a system to achieve a commanded position over many attempts.
Reversal Value (Hysteresis)
Difference between actual position values obtained for a given target position when approached from opposite directions.
Minimum Incremental Motion (MIM or Sensitivity)
The smallest increment of motion a device is capable of delivering consistently and reliably.
Resolution
The smallest increment that a motion device can theoretically move and/or detect. Resolution is not achievable, whereas MIM, is the real output of a motion system.
Eccentricity
Displacement of the geometric center of a rotation stage from the rotation axis in the plane defined by bearings.
Wobble
Tilt of rotation axis during rotation of a stage, measured on a reference surface.
The testing of accuracy, repeatability, and reversal error are made systematically with test equipment in controlled environment (20±1°C).
A linear cycle with 21 data points on the travel and 4 cycles in each direction gives a total of 168 points.
Page 12
EDH0176En1051 — 12/18 4
PR50 Series Compact High-Speed Rotation Stages
3
.2 Mechanical Specifications
CAUTION
To reach specifications stated, stages must be fixed on a plane surface with a flatness of 5 µm.
3.3 Load Specification Definitions
Normal Load Capacity (Cz)
Maximum load a rotation stage can move while maintaining specifications.
This value is given with speed and acceleration specified for each rotation stage, and with a load perpendicular to bearings.
Off-Centered Load (Q)
Maximum cantilever-load a rotation stage can move: Q Cz ÷ (1 + D/a)
D: Cantilever distance.
a: Construction parameter
3.4 Load Characteristics and Stiffness
G
U
A
R
A
N
T
E
E
D
S
P
E
C
I
F
I
C
A
T
I
O
N
S
Travel Range (°) 360 continuous Minimum Incremental Motion (°) 0.02 Uni-directional Repeatability, Typical (°) ±0.005 Bidirectional Repeatability, Typical
(1)
(°) ±0.030
Absolute, Typical
(
1)
(°) ±0.025
Maximum Speed
(2)
(°/s) 20
Wobble
(1)
(µrad) 100 or ±50
MTBF (h) 10,000
1
)
For the definition of Typical and Guaranteed specifications see “Motion Basics Terminology & Standards” Tutorial at www.newport.com
2)
Max. speed with SMC100PP: 8 °/s for PR50PP.
G
uaranteed and Typical Specifications
Guaranteed maximum performance values are verified per Newport's A167 metrology test procedure. For more information, please consult the m
etrology tutorial section in the Newport catalog or at www.newport.com
PR50CC PR50PP
Cz, Normal centered load capacity (N) 10 Ka, Transversal Stiffness (µrad/Nm) 100 Mz, Nominal Torque (Nm) 0.1 Q, Off-center load (N) Q Cz ÷ (1 + D/22) Where D = Cantilever distance (mm)
k
α
Z
Q
D
Cz
X
Mz
Page 13
5 EDH0176En1051 — 12/18
PR50 Series Compact High-Speed Rotation Stages
3.5 Stage Weights
The weights indicated into the below table is the value for the rotation s
tage with its cable installed.
T
he weight variation between drive units is not very significant.
4.0 —Drives and Motors
4.1 Stepper Drive Version
Stepper-motor-driven stages are offered in one half-step drive version: PR50PP.
Half-Step Drive
Is used for stepper motors, when 1 pulse emitted by electronic corresponds to theoretical physical motion of 1/2 of a full step of the motor.
Stepper Motor Performance Specifications and Characteristics
NOTES
• The stepper motor does not use an encoder.
• The real resolution for PR50PP rotation stages is 0.0100341°.
Weight [lb (kg)] PR50 0.88 (0.4)
Resolution Speed Angle by Step RMS Current Resistance Inductance
(°) (°/s) (°) per Phase (A) (Ω) (mH)
PR50PP 0.01 20 15
0.25
(1)
12.5 5.5
0.18
(2)
1)
1 phase at once: 0.25 A.
2)
2 phases at once: 0.18 A by phase.
Page 14
EDH0176En1051 — 12/18 6
PR50 Series Compact High-Speed Rotation Stages
Command Signals for the Stepper Motor
4.2 DC-Motor Drive Version
One DC-motor-driven configuration is available: PR50CC.
DC-Motor Performance Specifications and Characteristics
NOTE
The real resolution for PR50CC rotation stages is 0.0100341°.
Command Signals for the DC-Motors
In the above drawings, + Motor signal is referred to – Motor signal.
When the stage moves in + Direction, the + Motor voltage is higher than – Motor voltage.
When the stage moves in – Direction, the + Motor voltage is lower than – Motor voltage.
P
hase 1a
1
Phase 2a
Phase 1b
Phase 2b
D
irection – MOTION
2345 6 7 8
D
irection +
D
irection +
Resolution Speed Nominal Max RMS Max. Peak Resistance Inductance Tachometer (°) (°/s) Voltage (V) Current (A) Current (A) (Ω) (mH) Const. (V/krpm) PR50CC 0.01 20 12 0.15 0.3 24 0.75 –
+V
+
Motor
– Motor
–V
MOTION Direction +
MOTIONDirection –
Direction +
Page 15
7 EDH0176En1051 — 12/18
PR50 Series Compact High-Speed Rotation Stages
5.5 Mechanical Zero
Mechanical Zero is a 5 V open collector type.
Mechanical Zero repeatability is <0.01°.
CAUTION
“Mechanical Zero” is an active signal and should not be connected to any other source.
5.6 Feedback Signal Position (PR50CC Only)
The incremental sensor consists of an optical scale and an encoder head. When the carriage moves, the encoder head generates square signals in quadrature and sends to pins #19, #20, #23 and #24 of the SUB-D25 connector.
“Encoder” are “differential pair” (type RS-422) type output signals. Using these signals permits a high immunity to noise. Emission circuits generally used by Newport are 26LS31 or MC3487. Reception circuits to use are 26LS32 or MC3486.
Direction +
0
1
M
echanical
Z
ero
1
80 ±5
1
80 ±5
M
OTION Direction +
Pin #21
Pin #22
NEWPORT
STAGE
USER
Output Signals
Encoder Phase B
Encoder Phase B
Encoders & Sensors Power Supply
+5 V 5% 150 mA max.
0 V
Pin #24
Pin #20
Encoder Phase A
Encoder Phase A
Pin #23
Pin #19
Pin #25
Pin #15
}
0
1
Encoder Phase A
0
1
Encoder Phase A
1234
0
1
E
ncoder
P
hase B
0
1
Encoder Phase B
Direction – MOTION Direction +
Direction +
Page 16
EDH0176En1051 — 12/18 8
PR50 Series Compact High-Speed Rotation Stages
4.5 Pinouts
The pinout diagrams for the PR50 stages SUB-D25M connector are shown b
elow.
PR50CC
1 N.C. 2 N.C. 3 N.C. 4 N.C. 5 + Motor 6 + Motor 7 – Motor 8 – Motor 9 N.C. 10 N.C. 11 N.C. 12 N.C. 13 Mechanical Zero 14 Ground 15 N.C. 16 0 V 17 Do not connect 18 Do not connect 19 Encoder Phase A 20 Encoder Phase B 21 +5 V 22 0 V 23 Encoder Phase /A 24 Encoder Phase /B 25 N.C.
PR50PP
1 Phase 1a 2 N.C. 3 Phase 1b 4 N.C. 5 Phase 2a 6 N.C. 7 Phase 2b 8 N.C. 9 N.C. 10 N.C. 11 N.C. 12 N.C. 13 Mechanical Zero 14 Ground 15 N.C. 16 0 V 17 Do not connect 18 Do not connect 19 N.C. 20 N.C. 21 +5 V 22 0 V 23 N.C. 24 N.C. 25 N.C.
13
2
5
1
4
1
Page 17
9 EDH0176En1051 — 12/18
PR50 Series Compact High-Speed Rotation Stages
5.0 —Connection to Newport Controllers
5.1 Warnings on Controllers
Controllers are intended for use by qualified personnel who recognize shock hazards and are familiar with safety precautions required to avoid p
ossible injury. Read the controller user’s manual carefully before operating the instrument and pay attention to all written warnings and cautions.
WA RN IN G
Disconnect the power plug under the following circumstances:
• If the power cord or any attached cables are frayed or damaged in
any way.
• If the power plug is damaged in any way.
• If the unit is exposed to rain, excessive moisture, or liquids are spilled
on the unit.
• If the unit has been dropped or the case is damaged.
• If you suspect service or repair is required.
• Whenever you clean the electronics unit.
CAUTION
To protect the unit from damage, be sure to:
• Keep all air vents free of dir t and dust.
• Keep all liquids away from the unit.
• Do not expose the unit to excessive moisture (85% humidity).
• Read this manual before using the unit for the first time.
WA RN IN G
All attachment plug receptacles in the vicinity of this unit are to be of the grounding type and properly polarized.
Contact your electrician to check your receptacles.
WA RN IN G
This product is equipped with a 3-wire grounding type plug.
Any interruption of the grounding connection can create an electric shock hazard.
If you are unable to insert the plug into your wall plug receptacle, contact your electrician to perform the necessary alterations to ensure that the green (green-yellow) wire is attached to ear th ground.
WA RN IN G
This product operates with voltages that can be lethal.
Pushing objects of any kind into cabinet slots or holes, or spilling any liquid on the product, may touch hazardous voltage points or short out parts.
Page 18
EDH0176En1051 — 12/18 10
PR50 Series Compact High-Speed Rotation Stages
5.2 Connection
There is a label on every stage indicating its part and serial numbers.
W
ARNING
Always turn the controller's power OFF before connecting to a stage.
NOTE
These stages are ESP compatible. Enhanced System Performance is Newport's exclusive technology that enables Newport ESP motion controllers to recognize the connected Newport ESP stage and upload the stage parameters. This ensures that the user can operate the motion system quickly and safely.
5.3 Cables
The PR50 stages are delivered with a 3-meter cable. This cable is equipped with a SUB-D25M connector for direct connection to Newport Controllers.
BENDING
STATIC CABLE: .94 (24) CABLE IN MOTION: 2.76 (70)
ø
.24
(
6)
L
OCKING
K
NOBS
2
.28 (58)
D
ISCONNECTED
C
ABLE
C
ONNECTOR
STAGE
SUB-D25 CONNECTOR SHOWN
DIMENSIONS IN INCHES (AND MILLIMETERS)
WA RN IN G
This cable is shielded correctly. For a correct operation, make sure to lock connectors (ground continuity provided by the cable).
For applications where the cable attached to your stage is not adequate, Newport offers longer length cables designed to ensure the integrity of your positioning application.
These cables are specially shielded and terminated with Newport’s standard SUB-D25 connectors. They are available in 5-m (MMCABLE-5)or 7-m (MMCABLE-7) lengths.
Page 19
11 EDH0176En1051 — 12/18
PR50 Series Compact High-Speed Rotation Stages
CABLE
CONNECTOR
S
TAGE
L
OCKING
K
NOBS
B
ENDING
S
TATIC CABLE: 1.97 (50)
C
ABLE IN MOTION: 4.72 (120)
2
.60 (66)
D
ISCONNECTED
ø.39
(10.6)
SUB-D25 CONNECTOR SHOWN
DIMENSIONS IN INCHES (AND MILLIMETERS)
WA RN IN G
Keep the motor cables at a safe distance from other electrical cables in your environment to avoid potential cross talk.
For cable lengths in excess of 3 meters, we recommend the MMCABLE-REG to ensure a high quality, regulated 5 V supply to the stages.
This regulator is available as an option. Please note that for best efficiency, this regulator should be attached to the stage to re-adjust the 5 V coming from the controller through the long cable.
2.1
(54)
2.4 (62)
25-PIN SUB-D
MALE
2
5-PIN SUB-D
F
EMALE
DIMENSIONS IN INCHES
(AND MILLIMETERS)
Page 20
EDH0176En1051 — 12/18 12
PR50 Series Compact High-Speed Rotation Stages
6.0 —Connection to Non-Newport Electronics
6.1 Connections
WARNING
N
ewport is not responsible for malfunction or damage of XXXX stages
when used with non-Newport controllers.
WARNING
Newport guarantees
“”compliance of PR50 stages only if used with
Newport cables and controllers.
It is the customer’s responsibility to modify the cable and take care of sensor signal connections, when using the stage with non-Newport controllers.
End-of-Runs and Mechanical Zero are open collector type with a 5.6 V protective Zener diode.
I
i
n
max.: 16 mA
V max.: 5.25 V
NEWPORT
STAGE
5
.6 V
Page 21
13 EDH0176En1051 — 12/18
PR50 Series Compact High-Speed Rotation Stages
7.0 —Dimensions
Interface
The central aperture is threaded 1.063-20 in. PR50 stages are delivered with two retaining rings with the same diameter to allow mounting of any 1” (25.4 mm) diameter optics within the aperture. 4 M3 holes allow for fixation of custom top plates.
8.0 —Accessory: EQ45 Bracket
The EQ45 bracket (to order separately) can be used for vertical mounting of an PR50 rotation stage.
1.14 (29)
1
.14
(
29)
NOTE: TOP VIEW SHOWN WITH RETAINING RINGS.
4
.65
(
118)
1
.97
(
50)
.39 (10)
.063 (1.6)
.
39
(
10)
.
83
(
21)
2
HOLES M3 THD,
D
EPTH: .24 (6)
H
OLE 8-32 THD,
D
EPTH: .18 (4.5)
ø.83 (21)
2 RETAINING RINGS
THICKNESS: .26 (6.5)
2
.28
(
58)
1.97 (50)
4 HOLES M3 THD ON SQR .98 (25),
DEPTH: .12 (3)
S
UB-D25 CONNECTOR,
C
ABLE LENGTH 9.8 FT (3 M)
PP
.
25 (6.5)
RETAINING RING
.
82
(
Ø 21)
HOLE 8-32 THD, DEPTH: .18 (4.5)
2 HOLES M3 THD,
DEPTH: .24 (6)
ø1.0625"-20 UN 2B THD
ø1.01 (25.7)
4 HOLES Ø 3.5 ON 1.98
x .99 (50.4 x 25.2),
C'BORED ø.24 (6), DEPTH: .13 (3.2)
NOTE: BOTTOM VIEW SHOWN WITHOUT RETAINING RINGS.
MODEL SHOWN: PR50
DIMENSIONS IN INCHES (AND MILLIMETERS)
.63
(16)
2.28 (58)
.31 (8)
2.28 (58)
1.69 (43)
6°
2.83 (72)
4 HOLES ø.13 (3.4) ON SQR .98 (25)
4 HOLES M3 THD ON 1.98
x .99 (50.4 x 25.2)
DEPTH: 8
1.14 (29)
.71 (18)
ø.83 (21)
ø.83 (21)
MODEL SHOWN: EQ45
DIMENSIONS IN INCHES (AND MILLIMETERS)
Page 22
EDH0176En1051 — 12/18 14
PR50 Series Compact High-Speed Rotation Stages
9.0 —Maintenance
R
ECOMMENDATION
Please contact Technical Sales Support team for recommendations on application specific maintenance.
9.1 Maintenance
The PR50 rotation stage requires no particular maintenance. Nevertheless, this is a precision mechanical device that must be kept and operated with caution.
PRECAUTIONS
The PR50 rotation stage must be used or stocked in a clean environment, without dust, humidity, solvents or other substances.
RECOMMENDATION
It is recommended to return the stage to Newport for re-lubrication after 2000 hours of use.
If the PR50 rotation stage is mounted on a workstation and cannot be easily removed, please contact Newport's After Sales Service for further instructions.
9.2 Repair
CAUTION
Never attempt to disassemble a component of the rotation stage that has not been covered in this manual.
To disassemble a non specified component can cause a malfunction of the stage.
If you observe a malfunction in your rotation stage, please contact us immediately to arrange for a repair.
CAUTION
Any attempt to disassemble or repair a rotation stage without prior authorization will void your warranty.
9.3 Calibration
CAUTION
It is recommended to return your PR50 rotation stage to Newport once a year for recalibration to its original specifications.
Page 23
15 EDH0176En1051 — 12/18
Your Lo ca l Rep re se nt at iv e
Tel .: F
ax:
N
ame:
Company:
Address:
Country:
P. O . N u m b e r :
Item(s) Being Returned:
Model #:
Description:
Reasons of return of goods (please list any specific problems):
R
eturn authorization #:
(Please obtain prior to return of item)
Date:
Phone Number:
Fax Number:
Serial #:
Service Form
Page 24
Visit Newport Online at:
www.newport.com
North America & Asia
Newport Corporation 1791 Deere Ave.
Irvine, CA 92606, USA
Sales
Tel.: (800) 222-6440 e-mail: sales@newport.com
Technical Support
Tel.: (800) 222-6440 e-mail: tech@newport.com
Service, RMAs & Returns
Tel.: (800) 222-6440 e-mail: service@newport.com
Europe
MICRO-CONTROLE Spectra-Physics S.A.S 9, rue du Bois Sauvage
91055 Évry CEDEX France
Sales & Technical Support
Tel.: +33 (0)1.60.91.68.68 e-mail: france@newport.com
Service & Returns
Tel.: +33 (0)2.38.40.51.55
Loading...