Newport Corporation warrants this product to be free from defects in material and workmanship for a period of 1 year from the date of shipment. If
found to be defective during the warranty period, the product will either be
repaired or replaced at Newport’s option.
To exercise this warranty, write or call your local Newport representative, or
contact Newport headquarters in Irvine, California. You will be given prompt
assistance and return instructions. Send the instrument, transportation
prepaid, to the indicated service facility. Repairs will be made and the
instrument returned, transportation prepaid. Repaired products are warranted for the balance of the original warranty period, or at least 90 days.
Limitation of Warranty
This warranty does not apply to defects resulting from modification or misuse
of any product or part. This warranty also does not apply to fuses, batteries,
or damage from battery leakage.
This warranty is in lieu of all other warranties, expressed or implied, including
any implied warranty of merchantability or fitness for a particular use.
Newport Corporation shall not be liable for any indirect, special, or consequential damages.
Statement of Calibration
This instrument has been inspected and tested in accordance with specifications published by Newport Corporation.
The accuracy and calibration of this instrument and photodetector (where
applicable) is traceable to the National Institute for Standards and Technology through equipment which is calibrated at planned intervals by comparison to the certified standards maintained at Newport Corporation.
Analog Bandwidth35 Hz35 Hz1.5 kHz1.5 kHz5 kHz15 kHz10 kHz20 kHz
Full-Scale Accuracy
4
0.2 %0.2 %0.2 %0.2 %0.2 %0.2 %0.2 %0.2 %
(Typical)
Full-Scale Accuracy
4
0.4 %0.4 %0.4 %0.4 %0.4 %0.4 %0.4 %0.4 %
(Worst Case)
1
Listed signal ranges specify meter capability. Available signal ranges are detector dependent.
2
Maximum measurable signal is detector dependent. See description of detector saturation message “SA” in Table 2.
3
Full scale current may vary due to the Auto-Calibration compensation of amplifier DC offsets.
4
After 60 minute warm-up, followed by execution of an Auto-Calibration. See Section 2.5 and Section 6.2 (“O” Command).
Analog Output
Full-Scale Voltage:2V into 1MΩ
Full-Scale Accuracy: ±1%
Maximum AC Noise: Range 1:<4 mV
(open input)Ranges 2-8:<1 mV
p-p
p-p
x
DisplayDisplay
CalculationUnitComment
I/RWZERO Disabled
(I-Iz)/RWZERO Enabled
101log
/IR
mW
dBmZERO Disabled
(
101log
10 log
I
STOREF
10 log
II
STOREFZ
I
I
STOREF
II
−
Z
II
−
STOREFZ
−
Z
mW
I
−
II
/II R
)
Z
−
dBmZERO Enabled
dBZERO Disabled
dBZERO Enabled
RELZERO Disabled
RELZERO Enabled
WhereI= detector current
Iz= detector background current defined when the
ZERO key was pressed
R = responsivity of the detector (A/W)
I
STOREF
= referenced detector current defined when the
STOREF key was pressed
Table 1. Measurement Modes
xi
xii
Section 1
General Information
1.1System Overview
The 1830-C Optical Power Meter is a ±20,000 count A/D resolution, autoranging picoammeter. Measurements are displayed on a large 41/2 digit
liquid crystal display (LCD) and can also be taken remotely via the RS-232C
port or the IEEE 488 GPIB bus. The 1830-C is designed to take continuous
wave optical power measurements and is compatible with all of Newport’s
Low-Power Semiconductor photodetectors. A family tree of the 1830-C
compatible detectors and accessories is shown in Figure 1 below.
FP3-FH1
818-FA2
818-UV/CM
818-SL/CM
818-IR/CM
818-FA3-SMA
818-FA3-ST
818-FA3-FC
818-F-SL
818-F-IR
■
Optical Power Meter ■
Model 1830-C
(LOCAL)
R/S
AVG
STOREF
UNITSZERO
BEEPBKLT
ATTN
AUTO
INPUT
RANGEλλRANGE
818-ST/CM
Figure 1. 1830-C Power Meter and Compatible Detectors
Detectors connect to the 1830-C through a calibration module, in which the
responsivities and other information unique to the detectors is stored.
Calibration modules are ordered with the detectors at the time of purchase
and are labeled with the detectors’ model number and serial number.
Detectors with calibration modules have a “/CM” appended to their model
number.
EXAMPLE:818-SL(no calibration module)
818-SL/CM(with calibration module)
1
1.2Scope of this Manual
1.3Unpacking and Inspection
Please carefully read this instruction manual before using the 1830-C. Be
especially careful to observe the warnings and cautions throughout this
manual (see Safety Symbols and Terms). If any operating instructions are
not clear, contact Newport Corporation.
This instruction manual contains the necessary information for operation
and maintenance of the 1830-C, as well as information for troubleshooting
and obtaining service if necessary. This information is divided into the
following sections:
Section 1General Information
Section 2Initial Setup
Section 3System Operation
Section 4Performing Measurements
Section 5Computer Interfacing
Section 6Remote Command Set
Section 7Troubleshooting and Maintenance
Section 8Factory Service
1.4Preparation for Use
All 1830-C meters are carefully assembled, tested and inspected before
shipment. Upon receiving this instrument, check for any obvious signs of
physical damage that might have occurred during shipment. Report any
such damage to the shipping agent immediately.
NOTE
Retain the original packing materials in case reshipment becomes necessary.
The 1830-C should have some basic operations performed before measurements are made. These include:
Setting the Line Voltage Selector Switch ( Section 2.2)
This section contains information on how to configure the 1830-C to your
local line voltage and how to connect the detector and calibration module
to the meter. It also includes a short discussion about the remote interface
and the 1830-C’s power-up and auto-calibration.
2.2Setting the Line Voltage Selector Switch
The 1830-C can be configured to operate with line voltages of 100-120 VAC
and 220-240 VAC, and line frequencies between 50-60 Hz. Before turning the
meter on, configure it to the local voltage by setting the voltage selection
switches on the 1830-C back panel to match the nominal local voltage. See
Figure 2. Plug the AC line power cord to the rear of the 1830-C and then
connect the cord to AC power.
Figure 2. Line Voltage Selection Switch
WARNING
This product is equipped with a 3 wire grounding type plug. Any
interruption of the grounding connection can create an electric shock
hazard. If you are unable to insert the plug into your wall plug receptacle, contact your electrician to perform the necessary alterations to
assure that the green (green-yellow) wire is attached to earth ground.
3
2.3Detector Connection and Setup
Connect the detector to its calibration module as shown in Figure 3. The
detector’s model and serial numbers must match the model and serial
numbers found on the calibration module. Insert the calibration module,
with its groove facing left, into the connector port on the front panel of the
1830-C, as shown in Figure 4. An alignment pin assures the proper orientation of the calibration module.
Detector Cable BNCCalibration Module
Figure 3. Connecting a Detector with its Calibration Module
Plug calibration
module into this port.
INPUT
Always power off the meter before removing or inserting the calibration
module.
2.4Configuring the Computer Interface
The RS-232C baud rate and the IEEE 488 GPIB address must be properly set
via the back panel system switches if either interface is to be used. Please
refer to Section 6 for detailed instructions.
Figure 4. Calibration Module Connector Port
NOTE
4
2.5Power Up / Auto Calibration
Turn on the 1830-C by depressing the red power switch, located in the lower
left-hand corner of the front panel, until it clicks and remains in its depressed
position. At power-up, the 1830-C will perform the following sequence:
1) Momentarily display all the segments on the display
2) Display the software version number
3) Perform an Auto-Calibration, designated by displaying CAL on the
4) Display the serial number of the detector/calibration module
5) Display the wavelength to which the meter is set (If no calibration
The auto-calibration process involves measuring amplifier offset voltages
which arise from aging and temperature effects. These offsets are then
appropriately corrected for during normal operation. The 1830-C will
automatically disconnect the input signal from the amplifier during the
auto-calibration mode. To achieve stable readings at the specified accuracy, auto-calibration should be executed after a minimum 60 minute warmup period. This can be done either by resetting the 1830-C with the RESET
button on the back of the meter, or by sending the auto-calibration remote
command, O.
display
(If no calibration module is installed, the meter will display 000 as the
serial number.)
module is installed, the meter will display 0257nm as the wavelength.)
NOTE
The 1830-C must be powered-up with the calibration module connected to
the meter in order for the calibration data to be downloaded into the 1830-C.
Always power off the meter before removing or inserting the calibration
module.
5
Section 3
System Operation
3.1Introduction
3.2Digital Display
This section contains the information needed to operate the 1830-C power
meter, via the front panel key pad. The meanings of the display annunciators and the operation of all keys are discussed later in this section.
The 1830-C has a large 4-1/2 digit, back-lit liquid crystal display (LCD)
which can be seen at large angles of view. Figure 5 illustrates the layout of
the 1830-C’s display. Table 2, on the following page, explains the meaning of
the various annunciators and messages on the display.
Figure 5. The 1830-C Liquid Crystal Display
6
Annunciator/MessageComment
SNThis annunciator denotes serial number and is displayed at turn-on when
the detector’s serial number is displayed.
dBmThis annunciator specifies that either dB or dBm measurements are being
displayed.
RELThis annunciator specifies that relative measurements are being displayed.
k µ mOnly one of the characters of this annunciator will light at any time to specify
engineering prefixes of kilo, micro, nano, or milli. The “m” in this annunciator
serves as both “m” and “n”.
WThis annunciator specifies that measurements in units of watts are being
displayed.
nmThis annunciator indicates nanometer whenever the wavelength is displayed.
ATTNThe responsivity in use includes the effect of the detector’s OD3 attenuator.
AUTOAutomatic signal ranging is activated.
ZEROBackground signal subtraction (zeroing) is activated.
HOLDNo new readings will be displayed, nor be available on the remote interface bus.
REMThe meter has received a command/query either through the IEEE 488 or
the RS-232C bus.
LLOThe meter has been set to local-lockout from the remote interface bus and
will not respond to any front panel keys.
AVG: S M FEither Slow, Medium, or Fast numerical averaging for the displayed
measurement is activated.
OLThis message indicates that the input signal level exceeds the capability of
the present signal range. Either use auto-ranging or increment the signal
range until the OL message goes away.
SAThis message indicates that the input signal level exceeds the specified
saturation current of the detector being used with the meter. This level is
detector dependent.
CALThis message indicates that the meter is currently performing an auto-
calibration. The auto-calibration process involves measuring amplifier offset
voltages. See Section 2.5 for more details.
Table 2. 1830-C Display Annunciators/Messages
7
3.3Front Panel Key Functions
The front panel key pad of the 1830-C, Figure 6, provides access to all the
basic measurement functions. Table 3 and Sections 3.3.1 through 3.3.14 list
and describe in detail each key function. For power-up default conditions,
please see Table 4 at the end of this section.
(LOCAL)
R/S
ZERO
ATTN
UNITS
BKLT
STOREF
BEEP
AVG
AUTO
RANGERANGE
λλ
Figure 6. Front Panel Key Pad
KeypadRemote CommandDescription
(LOCAL)L0Enables local mode.
R/SG0, G1Run/Stop signal acquisition.
ZEROZ0, Z1Zero the display by subtracting the present reading from all
subsequent readings.
UNITSU1-U4Cycles between the four available measurement units (Watt, dB,
dBm, Relative).
STOREFSStores last measurement for future dB or relative measure-
ments.
AVGF1-F3Cycles between Slow, Medium, or Fast numerical averaging of
readings that are within a certain percentage of each other.
λ 䉱WnnnnIncrements the calibration wavelength in use.
λ 䉲WnnnnDecrements the calibration wavelength in use.
NoneTurns the meter on/off.
ATTNA0, A1Sets the responsivity value for either:
BKLTK0, K1, K2Cycles the display backlight brightness between Off, Medium,
BEEPB0, B1Turns the variable frequency beeper on/off.
AUTOR0, R1-R8Turns the automatic signal ranging on/off.
RANGE 䉱RxIncrements the signal range and disables the automatic signal
RANGE 䉲RxDecrements the signal range and disables the automatic signal
Table 3. 1830-C Key Functions and Associated Remote Commands
1) detector alone, or
2) detector and OD3 optical attenuator.
or High.
ranging.
ranging.
8
3.3.1
(LOCAL)
R/S
Local Mode
Run / Stop
This key serves two purposes:
1) When the meter receives a command through either the RS-232C or the
IEEE 488 GPIB interface, the meter is automatically put into remote
mode and the REM annunciator is displayed. Now all front panel keys
are disabled except this (LOCAL) key which, when pressed, puts the
meter back into the local mode. The meter can also be put in the local
mode by sending the remote command, L0.
2) When the meter is in the local mode, pressing this key will toggle
between run/stop acquisition modes. The HOLD annunciator is displayed when the meter is in the stop acquisition mode.
3.3.2
ZERO
Offset Subtraction
This key turns the offset subtraction on and off. When turned on, the ZERO
annunciator is displayed and the last measurement is saved as Pz and
subtracted from all subsequent measurements P. This causes subsequent
measurement calculations shown on the display and available on the
remote interface to use the value ( P− Pz) instead of P.
Offset subtraction allows one to remove the effects of ambient DC signals,
by zeroing the display before making a measurement. A second ZERO key
press turns off the ZERO annunciator and stops offset subtraction.
3.3.3
UNITS
Display Units
Measurements can be displayed in units of watts W, decibels dB or dBm,
and relative REL. Pressing the UNITS key repeatedly cycles the display
through these four units.
The W mode is a straightforward measurement which converts the current
from the detector into a power reading via the responsivity of the detector
at the set wavelength.
The dB mode uses the relationship dB = 10 * log( P / P
most recent measurement and P
is the reference measurement. At power-
ref
), where P is the
ref
up the reference for the dB mode is a level equivalent to 1mW. This can be
changed by pressing STOREF, which makes the most recent measurement
the new P
.
ref
The dBm mode uses the relationship dBm = 10 * log( P / 1mW ), where P is
the most recent measurement.
The REL mode uses the relationship ( P / P
measurement and P
is the reference measurement. At power-up the
ref
), where P is the most recent
ref
reference for the REL mode is a level equivalent to 1mW. This can be
changed by pressing STOREF, which makes the most recent measurement
the new P
.
ref
When the engineering prefixes n, µ, m and k are displayed in the REL
mode, they multiply the reading on the display by their appropriate
weights. For example, a display of 1.8300µ
REL
means ( P/ P
) * 10−6.
ref
9
3.3.4
STOREF
Store Reference Value
Pressing STOREF causes the most recent measurement P to be stored as P
ref
for subsequent use in relative measurement calculations. When the units
are dB and STOREF has been pressed, the displayed value is the function
10 * log( P/P
displayed value is the ratio P/P
P
is always a power reading stored in the units of Watts. Pressing STOREF
ref
causes a new P
3.3.5
). When the units are REL and STOREF has been pressed, the
ref
to overwrite the existing P
ref
AVG
Numerical Averaging
.
ref
value.
ref
Pressing AVG causes the numerical averaging feature to cycle through
slow S, medium M, and fast F modes. The slow mode takes the last 16
measurements that are within ±9 counts of the oldest measurement and
averages them for the displayed reading. The medium mode averages the
last 4 measurements, and the fast mode does no averaging.
The averaging buffer is a sliding buffer that always maintains 16 or 4
measurements, depending on the mode. Initially the buffer is completely
filled with the same measurement, and thereafter subsequent measurements replace the older measurements in the buffer. If, at any time, a single
measurement is acquired that is not within ±9 counts of the oldest measurement, the averaging algorithm starts over, not retaining any of the
previous measurements.
3.3.6
λ
Wavelength Increment
Pressing this key once causes the wavelength to be displayed. If the key is
pressed again or held down, the wavelength will increment or scroll,
respectively.
The detector calibration module contains responsivity data at discrete
wavelengths for its associated detector. By entering the wavelength which
is being measured, the correct responsivity value is used by the 1830-C in
calculating the measured power. When a wavelength falls between two
calibration points, linear interpolation is used to approximate the true
responsivity value.
Upon the initial power-up on a new calibration module, the meter defaults
to the shortest wavelength available for that specific detector. Whenever
the calibration wavelength is changed, this new wavelength is written to the
PROM in the calibration module and is used as the default wavelength at
subsequent power-ups.
NOTE:
If the detector and calibration module are changed to a different set, the
power meter must be powered-up again to allow for the downloading of
the new calibration data into the power meter. Always remember to
power off the meter before removing or inserting the calibration module.
10
Loading...
+ 51 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.