Doc: 1200-02 Copyright New Era Pump Systems Inc. Rev. 14 V3.9 08/27/14
Page 2
New Era Pump Systems Inc.
www.SyringePump.com
Model NE-300 Syringe Pump
Quick Start Instructions
• Plug in the pump.
• Press the power switch to turn on power.
• Press any key to stop the display from blinking.
Setup Pumping Parameters
To Change Numbers:
• Use the arrow keys to increment individual digits.
•
To set or clear the decimal point:
decimal point position. Or, press and hold the left-most up-arrow key for at least 1 second. When the digit
increments from 9 to 0, the decimal point will begin to shift. Release the key when the decimal point is correct.
•Press any non-arrow key, or wait 2 seconds, to enter the new setting. The display will blink when a new value is
entered and stored in memory.
Set the Syringe Inside Diameter:
• Momentarily press the ‘Diameter’ key. Set the inside diameter of the syringe in millimeters (mm).
Set the Pumping Rate.
• Momentarily press the ‘Rate’ key.
• To change the
− Momentarily press the ‘Rate’ key again. The display will show:
− While the units are blinking, press any arrow key to select the next available rate units.
− Press any non-arrow key, or wait 2 seconds, to set the rate units.
• Set the pumping rate. If the pumping rate is out of range, the display will show:
pumping rate units
Simultaneously press the 2 up-arrow keys under the 2 digits next to the
:
Load the Syringe
• Press in the white drive-nut button to move the pusher block.
• Lift and turn the syringe clamp away from the syringe holder block
• Position the syringe on the pump with the flange to the left of the syringe holder block.
• Lift and turn the syringe clamp onto the syringe body.
• Move the pusher block next to the syringe plunger.
Purge:
Press and hold the ‘Start/Stop’ key for one second. Release to stop.
Press and hold the direction button to reverse the pump and release the pusher block if
jammed.
Start the Pump:
Press and release the ‘Start/Stop’ key to start or stop the pump.
When Pumping
• The pumping rate can be changed.
PUMP RESET:
PUMP RESET:
PUMP RESET:PUMP RESET:
Press and hold the right-most up-arrow key while turning on power to the pump.
Need more pumping features?
Visit www.SyringePump.com to see our full-featured syringe pumps.
Copyright: New Era Pump Systems Inc. i08/27/14
Page 3
New Era Pump Systems Inc.
Table of Con
tents
1.
GENERAL INFORMATION AND PACKING LIST .................................................... 1
Thank you for purchasing the Model NE-300 “Just Infusion”™ Syringe Pump. With the NE-300 syringe
pump you will be able to digitally set an infusion rate for continuous pumping.
Please familiarize yourself with the NE-300’s operation by reading this user's manual. For future reference,
record the serial number, located on the rear of the pump, and the date of purchase.
New Era Pump Systems Inc., located in Farmingdale, NY USA, can be contacted at:
This Operating Manual, and the NE-300’s hardware, electronics and firmware are copyrighted.
Copyright 1999-2014, all rights reserved.
www.SyringePump.com
WWW.SYRINGEPUMP.COM
Model NE-300 Syringe Pump
Packing List
Included with the NE-300 Syringe Pump are the following items:
•One of the following external unregulated power supply adapters:
Input: One of: 120V AC 60 Hz, 220V AC 50 Hz, 240V AC 50 HZ
Output: 12V DC @ 800 mA (or compatible regulated power supply)
•This Operating Manual
1.1
Warnings! and Cautions
!
Read the user’s manual
!
No user serviceable parts are inside.
!
Disconnect power from the pump when
connecting or disconnecting cables.
!
Do not immerse the pump in liquid
!
Install on a stable surface.
!
Keep hands and loose clothing away from the
pump's moving parts.
!
The pump can automatically start when the
Pumping Parameters is operating or when
attached to an external control device.
!
Prevent liquids from entering openings in the
rear of the pump.
Use only with the supplied power supply
connected to a power source as specified on the
power supply label.
!
Do not push objects of any kind into the chassis
openings, except for appropriate cables and
connectors.
!
If the pump becomes damaged, do not use unless
certified safe by a qualified technician. Damage
includes, but is not excluded to, frayed cords and
deterioration in performance.
Discharge static from control cables before
connecting by touching the cable to ground.
Before touching the pump, discharge static by
touching ground.
1.2
Disclaimer and Warranty
Disclaimer
New Era Pump Systems Inc. makes no representations or warranties, expressed, statutory or implied, regarding the fitness
or merchantability of this product for any particular purpose. Further, New Era Pump Systems Inc. is not liable for any
damages, including but not limited to, lost profits, lost savings, or other incidental or consequential damages arising from
ownership or use of this product, or for any delay in the performance of its obligations under the warranty due to causes
beyond its control. New Era Pump Systems Inc. also reserves the right to make any improvements or modifications to the
product described in this manual at any time, without notice of these changes.
New Era Pump Systems Inc. products are not designed, intended, or authorized for use in applications or as system
components intended to support or sustain human life, as a clinical medical device for humans, or for any application in
which the failure of the product could create a situation where personal injury or death may occur.
All brand and product names used in this manual are the trademarks of their respective owners.
Copyright: New Era Pump Systems Inc. 108/27/14
Page 5
New Era Pump Systems Inc.
2
3
www.SyringePump.com
Model NE-300 Syringe Pump
Warranty
New Era Pump Systems Inc. warranties this product and accessories for a period of two year, parts and labor, from the date
of purchase. The repaired unit will be covered for the period of the remainder of the original warranty or 90 days,
whichever is greater.
A return authorization number must be obtained from New Era Pump Systems Inc. before returning a unit for repair.
Warranty covered repairs will not be performed without a return authorization number. At the option of New Era Pump
Systems Inc., a defective unit will be either repaired or replaced.
This warranty does not cover damage by any cause including, but not limited to, any malfunction, defect or failure caused
by or resulting from unauthorized service or parts, improper maintenance, operation contrary to furnished instructions,
shipping or transit accidents, modifications or repair by the user, harsh environments, misuse, neglect, abuse, accident,
incorrect line voltage, fire, flood, other natural disasters, or normal wear and tear. Changes or modifications not approved
by New Era Pump Systems Inc. could void the warranty. Wearable parts, such as drive nuts, are not covered by the
warranty.
The foregoing is in lieu of all other expressed warranties and New Era Pump Systems Inc. does not assume or authorize
any party to assume for it any other obligation or liability.
2.
Overview
The NE-300 is a general purpose single syringe pump capable of infusing at digitally set pumping rates. It is
controlled from a microcontroller based system which drives a step motor, allowing a wide range of pumping
rates configured to the inside diameter of the loaded syringe. The syringe is driven from a drive-screw and
drive-nut mechanism.
Features:
♦ Infusion pumping of syringes up to 140 mL.
♦ Pumping rates from 0.73 µL/hr with a 1 mL to
1500 mL/hr with a 60 mL.
♦ Display of infused volume.
♦ Non-volatile memory of all operating parameters.
♦ Power Failure Mode: Re-starts pumping after a
power interruption.
2.1
Glossary of Terminology and Concepts
4
5
6
7
1
8
10
When a device has as many features as the NE-1000 series, understanding its operation could be a daunting
task at first. By understanding the key concepts and terminology used in this manual, the operation of the
NE-300 will become quite intuitive. Every effort has been made to design the NE-300 with a consistent and
intuitive user interface.
9
Copyright: New Era Pump Systems Inc. 2 08/27/14
Page 6
New Era Pump Systems Inc.
Momentary Press:
Display Blink:
Plunger
Barrel
Barrel Flange
To facilitate and enhance your understanding of the NE-300’s operation, please take the time to familiarize
yourself with the basic concepts below:
www.SyringePump.com
Model NE-300 Syringe Pump
Parts of the Pump
1) Pusher Block
2) End Plate
3) Power On/Off Switch
4) Drive-Nut Button
5) Drive-Screw
6) Syringe Holder Block
7) ‘V’ Slot (on Syringe Holder Block)
8) Syringe Clamp
9) Keypad / User Interface
10) Guide Rod (2 guide rods)
Parts of a Syringe
Flange
Plunger
Terminology
3.
Setup
♦ Place the pump on a stable surface.
♦ Plug the round connector end of the supplied power supply adapter into the power plug located on the
lower right of the pump's rear. See section 8, Rear of Pump, for a diagram of the rear of the pump. Plug
the other end of the power supply adapter into an appropriate electrical outlet. The pump will be powered
when the bottom of the power switch, located on the upper right of the rear of the pump, labeled ‘1’, is
pressed. The red indicator on the switch is visible when the power switch is in the ‘on’ position. After
power is applied to the pump, the pump’s display will flash.
♦ Next, the Pumping Parameters can be entered. Before the pump can be started, the pump needs the
measurement of the inside diameter, in millimeters (mm), of the syringe that will be loaded. The syringe
diameter is entered using the keypad on the front panel of the pump.
♦ Finally, the syringe can be loaded and the pump started.
4.
Loading Syringes
The syringe is loaded by securing the barrel and the pusher flange as follows:
1: Press in fully the white drive-nut button on the pusher block, releasing the block. Taking care not to
drag the drive-nut on the drive-screw, slide the block away from the syringe holder, providing sufficient
space for the loaded syringe. Then release the white button.
A quick press, less than 1 second, and then release of a key on the keypad.
A momentary blanking of the LCD display. This indicates that the new data
entered by the user is valid and has taken effect.
2: Lift the syringe clamp above the syringe holder block. Turn it 1/4 turn and then lower it onto the
3: Load the syringe with the barrel over the syringe holder and the syringe plunger towards the middle of
4: Lift the syringe clamp to slightly above the height of the syringe barrel and then turn the syringe clamp 1/4
5: Press the white drive-nut button to slide the pusher block against the syringe plunger. Release the white
⇒ To unload the syringe, reverse the instructions for syringe loading.
Copyright: New Era Pump Systems Inc. 308/27/14
syringe holder block. The syringe clamp should be out of the ‘V’ slot.
the pump. Place the barrel on the syringe holder, in the ‘V’ slot, with the barrel flange to the left of the
syringe holder block.
turn back to its original position and lower it onto the syringe barrel.
drive-nut button.
Page 7
New Era Pump Systems Inc.
Dispensed
Phase #
Function
µ
Volume Dispensed
Indicator
5.
User Interface
www.SyringePump.com
Units
Indicators
Model NE-300 Syringe Pump
Reverse Purge
Indicator
Motor
Operating
Volume
min
hr
Volume
Withdraw
Diameter
Setup
Pumping
Start
Stop
Reverse Purge Key
Decimal Point: Press 2 keys
Figure 1: Front Panel
5.1
Entering Values
Syringe ID
mm
mL
L
Rate
Pumping Program
Unused Functions
When applicable, values can be changed by displaying the current value, then using the arrow keys. The new
value will be stored in the pump’s non-volatile memory, meaning that the new value will not be lost the next
time that power is applied to the pump.
A displayed value can be changed by pressing the up-arrow keys below each digit. To set or clear the decimal
point, press the 2 up-arrow keys below the 2 digits next to the decimal point position. If the value to be
changed is not currently displayed, when applicable, press the key associated with the required value. The
display will show the setting’s current value and its units, if any.
While the current value is being changed, the units LEDs associated with the value, if any, will blink. Except
where noted, the new value is stored, and/or the selected operation takes effect, when either
1) A non-arrow key is pressed or
2) After a 2 second delay since the last arrow key was pressed.
If the new value is valid and different from the original value, the display will blink, indicating that the new
value was stored. Otherwise, if the value was invalid, an error message will be displayed. Pressing any key
clears the error message and restores the original value.
In general, if a parameter has 2 values, ‘off’ and ‘on’, they are represented by the numbers ‘0’ and ‘1’,
respectfully.
5.2
LCD Display
The display consists of a 4 digit reflective LCD display. This is the general purpose user display device for
displaying floating point values, functions and parameters. The colon (:) is used for separating function
abbreviations from their parameter values.
Publication #1200-01 408/27/14
Page 8
New Era Pump Systems Inc.
µ
Pumping rate units are expressed using 2 LEDs:
mL
min
hr
µL
mL
min
hr
µL
mL
min
hr
µL
mL
min
hr
µL
5.3
LEDs
To the right of the LCD are 8 red, round, LED indicators. The first 2 columns display the units of the
displayed values. Units are expressed using 1 or 2 LEDs. For instance, ‘mL / hr’ is expressed by lighting the
‘mL’ and the ‘hr’ LEDs.
LEDDescription
mm Millimeters
mL Milliliters
min Minutes
L
hr Hours
Withdraw Pumping Direction:
Dispensed Displayed volume is dispensed volume
Pumping Lit: Motor is operating
5.4
Each of the four digits in the display is associated with the up arrow key directly below it. When applicable,
the arrow key is used to increment the value of that digit, or advance to the next setting.
Microliters
Lit: Withdraw Purge Pumping
Not lit: The Pump is stopped
Up Arrow and Decimal Point
www.SyringePump.com
Model NE-300 Syringe Pump
‘mL/min’ =
‘mL/hr’ =
‘µL/min’ =
‘µL/hr’ =
Each press of an up arrow key will increase the digit by 1, up to 9, and then back to 0. The arrow keys may
also be held down for continuous incrementing of numbers. Some parameters have a fixed range of values,
such as some setup parameters that are either turned on or off. In these cases, the arrow key will only scroll up
to the maximum value for that parameter, then back to the minimum value.
When changing a value’s units, each press of any arrow key will change the unit’s LEDs to the next unit
selection.
When the display blinks, the new value is stored and takes effect. This will occur when a non-arrow key is
pressed or after a 2 second delay since the last key press.
Decimal Point
There are 4 decimal point positions on the LCD display. Each decimal point position is to the right of a digit
in the display. The last decimal point position, to the right of the right-most digit is not displayed, indicating
whole numbers with no decimal point.
To set or clear the decimal point position, simultaneously press the 2 up-arrow keys under the 2 digits next to
the decimal point position.
Alternatively, use the left-most up-arrow key / decimal point key (/). Press and hold this key for at least 1
second and wait until the left-most digit scrolls from ‘9’ to ‘0’. While continuing to hold this key, the decimal
point will shift 1 position to the right. After the right-most decimal point position, the decimal point will shift
to the first decimal point position. Release the key when the decimal point is in the required position.
5.5
‘Diameter’ and ‘Setup’ Key
The ‘Diameter’ key allows the syringe inside diameter to be viewed and set. While displaying the diameter,
the ‘mm’ LED is lit. When pump is stopped, pressing the arrow keys will change the current diameter (see
sec. 5.4, Up Arrow and Decimal Point). The ‘mm’ LED will blink while the diameter is being changed.
If the ‘Diameter’ key is pressed and held, ‘Setup’ mode will be entered. (see sec. 5.10, ‘Setup’ ).
5.6
‘Rate’ Key
The ‘Rate’ key allows the pumping rate to be viewed or changed while pumping or stopped. While the pump
is stopped, momentary presses of this key will switch between the ‘Rate’ display and the select rate units
mode.
To change the pumping rate displayed, while pumping or stopped, use the up arrow keys (see sec. 5.4, Up
Arrow and Decimal Point). The rate units will blink while the rate is being changed. The new pumping rate
Publication #1200-01 508/27/14
Page 9
New Era Pump Systems Inc.
n
is entered and takes effect immediately when the display blinks after a 2 second delay or when a non-arrow
key is pressed. The new pumping rate is stored in memory.
See section 9.5, “Syringe Diameters and Rate Limits”, for a list of minimum and maximum pumping rates. A
pumping rate of 0.0 will stop the pump. When the pumping rate is changed, if it is out of range of the pumping
rate limits, the display will show . Pressing any key clears the message and returns to the previous
pumping rate.
www.SyringePump.com
Model NE-300 Syringe Pump
5.6.1 Pumping Rate Units
The pumping rate units can only be changed when the Pump is stopped. A momentary press of the ‘Rate’ key
will enter Rate Units Change mode. The 2 LEDs representing the units will blink and the display will show:
.
Each press of any up arrow key selects the next rate units, as indicated by the blinking units LEDs. When the
required rate units are blinking, press any non-arrow key or wait 2 seconds. The display will blink, indicating
the rate units are stored in memory.
5.7
‘Volume’ Key
Pressing this key will display the “Volume Dispensed”, as indicted by the ‘Dispensed’ LED. The units of the
volume are set according to the syringe diameter, but can be changed.
5.7.1 Clearing “Volume Dispensed”
With the pump stopped and displaying the “Volume Dispensed”, pressing and holding any up-arrow key for
one second will reset the dispensed volume to 0.000.
5.7.2 Changing Volume Dispensed Units
With the “Volume Dispensed” 0.000 displayed, press any up-arrow key to enter “Set Volume Units” mode.
The display will show: and the Volume units will blink. Press an up-arrow key to toggle the
volume units between ‘mL’ and ‘µL’ while the units are blinking.
5.8
Reverse Purge Key
The direction key, ‘’, is used to un-jam the pusher block after an over infusion, making it difficult to
release the nut block. Press and hold this button for at least 1 second to reverse purge the pump until the
pusher block is released. Release this key to stop the pump. While this key is held, the display will show:
. The Withdraw and Pumping LED’s will be lit during the reverse purge.
5.9
‘Start’/’Stop’ Key
The ‘Start/Stop’ key starts or stops the Pump’s operation. Pressing this key switches between Pumping and the
Pump stopped. The ‘Pumping’ LED will indicate that the Pump is pumping.
Pressing and holding this key while starting the Pump will start Purge Mode. Purge will begin after the key is
held for one second, and continues until the key is released. The pump stops after the key is released.
5.10
The secondary function of the ‘Diameter’ key is ‘Setup’. While the Pump is stopped, press and hold the
‘Diameter’ key until the setup configuration parameter, “Power Failure Mode”, is displayed:
The display will consecutively display, for about 2 seconds, each Setup Configuration parameter and its
current setting. Pressing any non-arrow key will immediately advance to the next Setup Configuration
parameter, if any.
‘Setup’ Key
.
To change a Setup Configuration parameter, press an arrow key under the parameter’s value. To store the new
value, press any non-arrow key or wait 2 seconds. If the parameter value differs from its previous value, the
display will blink. The new parameter value will be stored and the next parameter will be displayed, if any.
See section 7, Setup Configuration for a complete description of the Setup Configurations.
After the last configuration parameter is displayed, the display reverts back to displaying the syringe diameter.
Any new parameter value will take effect immediately upon being stored.
Publication #1200-01 608/27/14
Page 10
New Era Pump Systems Inc.
www.SyringePump.com
Model NE-300 Syringe Pump
5.11
The following special functions are accessed by pressing the relevant key, while turning on power to the
pump.
Special Power-Up Functions
5.11.1 Firmware Version Display
To display the pump’s firmware version, press the left-most up arrow key (/) while turning on power to
the pump. The display will show:
, or similar. Pressing any key will clear the display.
5.11.2 Reset Pumping Parameters
To clear out the current Pump setups, press the right-most up arrow key () while turning on power to the
pump. The display will show . Pressing any key will clear the display.
With pumps having as many complex features as the NE-1000 series, it is easy for a novice user experimenting
with the pump's setup to get the pump into a 'weird' state. Performing this reset function will bring the pump
out of a 'weird' state.
5.12
Error and Status Messages
Value entered is ‘Out Of Range’ of the pump’s operational limits. Check the pumping
rate and/or the syringe inside diameter setting is correct.
Indicates pumping rate or volume units change mode. The units LED's will also be
blinking. Press any up arrow key to change the units.
Key pressed is not currently applicable.
Indicates that the pump is purging. Displayed while holding down the 'Start/Stop' key.
Displayed while holding down the direction key and performing a reverse purge.
6.
Operation
Before the pump can be operated, the pumping data must be setup. The syringe inside
diameter and a non-zero pumping rate needs to be set. The operation of the pump can
then be started by pressing the ‘Start / Stop’ key.
6.1
Syringe Inside Diameter
The syringe inside diameter can only be set while the Pump is stopped. Use the up arrow keys to set the
diameter value. While the diameter value is being set, the ‘mm’ LED will blink. The new diameter value is
stored after pressing any non-arrow key, or after a 2 second delay.
Valid syringe diameters are from 0.1 mm to 50.0 mm. If the diameter is out of this range, the display will
show ‘oor’. Pressing any key restores the diameter display to its previous value. Changing the syringe
diameter will not zero the pumping rate. Section 9.5, “Syringe Diameters and Rate Limits”, is a
representative list, for reference, of syringe diameters for various syringe manufacturers and syringe sizes.
6.1.1 Volume Units: Default and How to Change
The units of the accumulated infusion volume is set according to the syringe diameter setting. If the default
volume units are changed (see next section), the selected volume units will remain in effect until a reset
function is performed.
From 0.1 to 14.0 mm Syringes smaller than 10 mL: Volume units are ‘µL’
From 14.01 to 50.0 mm Syringes greater than or equal to 10 mL: Volume units are ‘mL’
Changing Volume Units
The volume units used for the accumulated volume can be changed to either ‘mL’ or ‘µL’. Volume units can
only be changed when the pump is stopped.
Display the “Volume Dispensed” by pressing the “Volume” key. The current volume units and the
“Dispensed” LED will be lit.
Publication #1200-01 708/27/14
Page 11
New Era Pump Systems Inc.
Set the Volume Dispensed to 0.000 if it is not zero: Press and hold any up arrow key until the Volume
Dispensed is set to 0.000.
Then, pressing any up-arrow key will change the display to and the current volume units will
blink.
Now press any up-arrow key to switch the volume units between ‘mL’ and ‘µL’. Press any non-arrow key or
wait 2 seconds to enter the new volume units. The display will blink when entered. The selected volume units
will remain in effect and override the default volume units. Changing the diameter will no longer change the
volume units. Performming a system reset will cancel the override and allow the volume units to change to the
default volume units when setting the syringe diameter.
6.2
Operating the Pump
When the “Start/Stop” key is pressed, the Pump will begin pumping, and the ‘Pumping’ LED will be lit. The
pump will pump continuously at the set pumping rate until stopped.
Either the pumping rate, “Volume Dispensed”, or syringe diameter can be displayed while pumping. Press the
relevant key to change the display. While displaying the pumping rate, press the up arrow keys to change the
pumping rate.
Warning: The NE-300 does not have stall detection. The pump must be stopped manually when the
syringe is empty or the motor force is exceeded. The white drive nut block can be damaged when the
pump is operating while the motor cannot turn. This will violate the warranty and replacement of the
drive nut block will not be covered.
6.3
Purging
www.SyringePump.com
Model NE-300 Syringe Pump
To purge the syringe, with the Pump stopped, press and hold the 'Start/Stop' key. The Pump will start
pumping, then, after one second, purge will begin. The pump will pump at its top speed. Purging will
continue until the 'Start/Stop' key is released, and then the pump will stop. While purging the display will
show: .
Press and hold the direction button, to perfrom a reverse purge. The display will show:
6.4
Changing the Pumping Rate While Pumping
The pumping rate can be changed while the pump is operating. To change the pumping rate, display the
current pumping rate by pressing the ‘Rate’ key, if needed. With the pumping rate displayed, press the up
arrow keys to change the rate. The rate units will blink while the rate is being changed. Rate units cannot be
changed while pumping.
The new rate is stored after a 2 second delay or by pressing a non-arrow key. If the new rate is within the
operating range of the pump, the display will blink and the new rate will be stored in memory and the pump
will begin pumping at the new rate. If the new rate is out of the operating range of the pump, the display will
show . Pressing any key clears the message and restores the previous rate.
6.5
Volume Dispensed
The accumulated Volume Dispensed can be displayed at any time by pressing the Volume key. The display
will show the total accumulated volume pumped with the ‘mL’ or ‘µL’ LED lit and the ‘Dispensed’ LED lit.
Volume is computed based upon the syringe inside diameter setting.
The “Volume Dispensed” accumulation is reset to 0 when:
A) With the pump stopped, pressing and holding any up arrow key while displaying the volume.
B) The syringe diameter is changed.
C) The accumulated Volume Dispensed rolls over from 9999 to 0.
D) The pump is powered on.
Publication #1200-01 808/27/14
Page 12
New Era Pump Systems Inc.
n
n
NE-300
7.
Setup Configuration
To change or view the setup configuration, the Pump must be stopped. Press and hold the ‘Diameter’/‘Setup’
key until the parameter, ‘PF’ is displayed. After 2 seconds, or when any non-arrow key is pressed, the next
parameter will be displayed. (See sec. 5.10, ‘Setup’ Key). Pressing an arrow key under a value will change the
parameter. The following will be displayed:
Power Failure Mode. Where ‘n’ is the current setting: ‘0’ = Disabled, ‘1’ = Enabled.
When enabled, if the Pump was pumping when power to the pump was disrupted, the Pump will automatically
start pumping when power is reconnected to the pump. The accumulated volume will be reset to 0. Pressing
any key on the keypad while powering up the pump will stop the Pump from starting in Power Failure Mode.
Low Noise Mode, Where ‘n’ is the current setting: ‘0’ = Disabled, ‘1’ = Enabled.
A side effect of the NE-300’s high precision micro-stepped motor driver is a high frequency resonance sound
at very low pumping speeds. This mode minimizes this sound by reducing the micro-stepping, increasing
pulsations.
8.
Rear of Pump
www.SyringePump.com
Model NE-300 Syringe Pump
NEW ERA PUMP SYSTEM S, Inc.
WWW.SYRINGEPUM P.COM
(631) 249-1392
Figure 2: Rear of Pump
9.
Appendix
9.1
Accessories
Farmingdale, NY info@syringepump.com
9.1.1 Syringe Heater
See www.SyringeHeater.com for details.
Flexible heating pads that wraps around the syringe. Thermo-Kinetic Heat Clamping controller will heat a
syringe to a set temperature up to 100 C.
9.1.2 Firmware and Hardware Upgrade
Contact your dealer about upgrading your pump with other features from the NE-1000 Programmable Syringe
Pump Series. See www.SyringePump.com
Publication #1200-01 908/27/14
Page 13
New Era Pump Systems Inc.
9.2
Troubleshooting and Maintenance
Maintenance: Periodically, a small amount of all-purpose oil should be applied to the guide rods.
The mechanism should be kept clean to prevent impeded operation.
No other special maintenance or calibrations are needed
Pusher block makes a snap or click sound when the pump is started: This is a normal condition. When
the pusher block is manually moved, the drive-nut may not have been fully engaged on the drive screw. The
sound heard is the drive-nut engaging on the drive screw.
9.3
Specifications
www.SyringePump.com
Model NE-300 Syringe Pump
Mechanical & Electrical
Syringe sizes: Up to 60 mL. 140 mL partially filled.
Number of syringes: 1
Motor type: Step motor
Motor steps per revolution: 400
Micro-stepping: 1/8 to 1/2 depending on motor speed
Advance per step: 0.2126 µm to 0.8504 µm depending on motor speed
Motor to drive screw ratio: 15/28
Drive screw pitch: 20 revolutions/”
Power connector: 2.1 mm, center positive, DC
Voltage at power connector: 12V DC at full load
Amperage: 750 mA at full load
Power supply type: External wall adapter, country and power source specific
Power supply output rating: 12V DC @ 800 mA (or greater if regulated power supply)
Dimensions: 8 3/4” x 5 3/4” x 4 1/2” (L x W x H)
(22.86 cm x 14.605 cm x 11.43 cm)
Weight: 3.6 lbs. (1.63 kg)
Operational
Accuracy Within +/- 1% over length of syringe,
exclusive of syringe variations.
Reproducibility +/- 0.1%
Maximum speed: 3.7742 cm/min
Minimum speed: 0.004205 cm/hr
Maximum pumping rate: 1257 mL/hr with a B-D 60 mL syringe
Minimum pumping rate: 0.73 µL /hr with a B-D 1 mL syringe
Maximum force: 35 lbs. at minimum speed, 18 lbs. at maximum speed
Syringe inside diameter range: 0.100 to 50.00 mm
9.4
Custom Applications
For specialized and OEM applications, contact your dealer
or New Era Pump Systems Inc. Custom modifications can
be made to the mechanics or the firmware.
Publication #1200-01 1008/27/14
Page 14
New Era Pump Systems Inc.
Syringe
Syringe
Inside
Maximum
Minimum
Maximum
1
4.699
3
8.585
5
11.99
10
14.43
20
19.05
30
21.59
60
26.59
1
4.69
3
9.65
5
12.45
10
15.9
20
20.05
30
22.9
50
29.2
1
5.74
3
8.941
6
12.7
12
15.72
20
20.12
35
23.52
60
26.64
140
38
1
4.7
3
8.95
5
13
10
15.8
20
20.15
30
23.1
60
29.7
1
6.7
2
8.91
3
9.06
5
11.75
10
14.67
20
19.62
30
22.69
50
26.96
1
9.538
3
9.538
5
12.7
8
9.538
20
19.13
50
28.6
Syringe
Diameter
Maximum
Minimum
Syringe
Diameter
Maximum
Minimum
5
0.343
2.303
0.176
10
0.485
57
0.351
25
0.728
1
4.606
0.701
50
1.03
7.284
1.752
100
1.457
5
10.3
3.504
0.5
0.103
10
14.57
7.01
1
0.146
25
23.03
17.52
2
0.206
50
27.5
24.98
0.326
34.99
40.43
9.5
Syringe Diameters and Rate Limits
www.SyringePump.com
Model NE-300 Syringe Pump
Manufacturer
(all names ™)
B-D
HSW
Norm-Ject
Monoject
Terumo
Poulten &
Graf
(Glass)
Steel
Syringes
SGE
(gas tight)
Hamilton
Microliter
(mL)
(µL)
Diameter
Inside
5
(mm)
(mm)
Rate
(mL/hr)
39.27
255.6 4.748
370.3 6.876
645.4 11.99
1257 23.35
39.11 0.727
165.6 3.076
275.6 5.119
449.6 8.349
714.9 13.28
932.6 17.32
1516 28.16
58.59 1.088
142.1
286.8 5.326
439.4 8.161
719.9 13.37
983.8 18.27
1262 23.44
2568 47.69
39.28
142.4 2.646
300.5 5.581
443.9 8.244
722.1 13.41
1568 29.13
79.83 1.483
141.1 2.622
145.9 2.711
245.5 4.559
382.7 7.107
684.6 12.72
915.6 17.01
1292 24.01
161.7 3.005
161.7 3.005
286.8 5.326
161.7 3.005
650.8 12.09
1454 27.01
Rate
(µL/hr)
209.2
418.3
942.5
1886
3775
18.86
37.91
75.47
(µL/hr)
131 2.434
829
949 17.63
(µL/hr)
189
Rate
0.73
15.4
2.64
0.73
Rate
0.004
0.008
0.018
0.036
0.071
0.001
0.001
0.002
0.004
Rate
(mL/min)
0.654
2.184
4.261
6.172
10.75
13.81
20.95
0.651
2.76
4.594
7.493
11.91
15.54
25.27
0.976
2.369
4.78
7.324
11.99
16.39
21.03
42.8
0.654
2.374
5.009
7.399
12.03
15.81
26.14
1.33
2.353
2.433
4.092
6.379
11.41
15.26
21.54
2.696
2.696
4.78
2.696
10.84
24.24
(mL)
0.25
0.5
2.5
100
Inside
(mm)
3.2
Rate
(mL/hr)
9.432
18.86
37.73
94.36
188.6
377.5
943.2
1344
2177
Rate
(µL/hr)
Publication #1200-01 1108/27/14
Loading...
+ hidden pages