8 UNPACKING AND ACCEPTANCE TESTING.................................................8-1
9 EC DECLARATION OF CONFORMITY ..........................................................9-1
Document Number 111/00 Operators Manual August 2007
NT-500 Lesion Generator
NT-500 RF Lesion Generator ™
1-1
OPERATORS MANUAL
1 GENERAL INTRODUCTION WITH WARNINGS AND CAUTIONS
The NT-500 has been designed to offer the full range of features required to perform
Radiofrequency denervation.
The front panel has been ergonomically designed and allows the clinician direct
manipulation of the controls. Each functional effect of the device is defined within its
own discrete area.
It is designed for safe use in a doctor’s office, surgery centre or hospital environment.
Linear Bar Displays are used to facilitate immediate visual interpretation of
temperature, RF current and RF voltage, and digital displays are used for
impedance, stimulate voltage and lesion time.
The NT-500 has full electronic interlocking to prevent accidental switching to lesion
power and stimulation voltage.
The internal settings of the machine have been factory set and should not be
adjusted except by approved technicians authorised by the company.
The machine is designed for use with NT-500 Thermocouple Probes only. The use of
probes from other manufacturers could give serious errors in the temperature
reading and may compromise the safety of the patient, and would negate the
warranty.
Regularly inspect the accessories of the NT-500, in particular electrode cables
should be checked for possible damage to the insulation.
The accessories are not appropriate for endoscopic use.
Warnings
A warning indicates a potentially harmful situation to yourself or others.
Ensure you read this Operators Manual before operating the NT-500.
HAZARDOUS ELECTRICAL OUTPUT:- The equipment is for use ONLY by qualified
medical personnel.
DO NOT under any circumstances perform any Testing or Maintenance on the
equipment while it is being used on a patient.
DO NOT use extension cords or adapters of any type. The power cord and plug must
be intact and undamaged.
Document Number 111/00 Operators Manual August 2007
NT-500 Lesion Generator
NT-500 RF Lesion Generator ™
1-2
OPERATORS MANUAL
Should the power cord or plug become cracked, frayed, broken or otherwise
damaged, it must be replaced immediately.
If the equipment has in any way suffered mechanical damage, it should be returned
to the Supplier for Inspection and Test before further use.
The power cord should always be unplugged before cleaning or service.
The operator should not perform any servicing of the equipment. Any servicing
should only be performed by qualified personnel.
EXPLOSION HAZARD: - This equipment is not suitable for use in the presence of a
flammable anaesthetic mixture with air, oxygen or nitrous oxide.
EXPLOSION SHOCK HAZARD:-Always turn the equipment off before cleaning and
DO NOT allow ANY fluid to enter the ventilation holes or sockets.
ELECTRIC SHOCK HAZARD:- DO NOT touch any exposed wiring or conductive
surface while cover is off and the equipment is energised. The voltage present when
the electric power is connected to the equipment can cause injury or death. NEVER
wear a grounding wrist strap when working on energised equipment.
FUSE REPLACEMENT:- For continued protection against fire hazard, replace only
with same type and rating of fuse as displayed on the rear Serial Number Plate.
When carrying out treatment take care to avoid the following risks:-
RISK OF RF BURNS TO PATIENT:- Ensure the patient does not come into contact
with metal parts of the table and its accessories – antistatic sheeting recommended.
RISK OF RF BURNS TO PATIENT: - Avoid skin to skin contact between different
parts of patient’s body (for example between the arms and the body of the patient, or
between legs) – use dry gauze if necessary.
RISK OF RF BURNS TO PATIENT: - Avoid using physiological monitoring
equipment during a procedure – if monitoring is required, monitoring electrodes
should be placed as far as possible from the NT-500 cannula. Monitoring devices
which use needle electrodes are not recommended.
RISK OF RF BURNS TO PATIENT: - Position all cables to the NT-500 cannula and
dispersive plate in such a way to avoid contact with the patient or other leads.
INTERFERENCE WITH ACTIVE INPLANTS: - Check whether the patient has a
cardiac pacemaker or any other active implant. A possible hazard exists because
interference with the action of the pacemaker may occur or the pacemaker may be
damaged. In case of doubt, obtain qualified advice.
Document Number 111/00 Operators Manual August 2007
NT-500 Lesion Generator
NT-500 RF Lesion Generator ™
1-3
OPERATORS MANUAL
INTERFERENCE WITH OTHER EQUIPMENT: - During RF Treatment procedures
the radiated electrical fields may interfere with other electrical medical equipment.
(See Section 2.5 to Minimise Electromagnetic Interference)
USE OF FLUIDS: - Ensure that if fluids (saline etc.) are being used during a
procedure they should be positioned away from the NT-500.
RISK OF RF BURNS TO PATIENT; - In Lesion Mode select the lowest possible
power for intended purpose.
RISK OF RF BURNS TO PATIENT: - Check the Dispersive (Neutral) Lead and the
Dispersive Pad before applying power to the patient.
PROBES: - Use only “NT-500 “probes.
Cautions
A CAUTION indicates a condition that may lead to equipment damage or
malfunction.
DO NOT Activate the output of the NT-500 until the probe is properly positioned in
the patient.
DO NOT Remove the top cover of the NT-500, as it will expose voltage which can
cause injury or death.
Servicing of the equipment in accordance with the appropriate service manual. This
process should never be undertaken in the absence of proper tools, test equipment
and the most recent revision of the service manual which is clearly and thoroughly
understood.
To reduce risk of electrical shock DO NOT remove the back panel of the NT-500.
Refer servicing to qualified personnel.
When cleaning the outer casing of front panel of the equipment DO NOT use
abrasive agents or solvents.
If erratic readings of voltage, current or impedance or temperature are
observed, the procedure should be halted until a determination of the source
is identified.
If at any time the device is behaving erratically, turn the RF Power off.
Document Number 111/00 Operators Manual August 2007
USA 110 volt 60Hz Fused 2 Ampere on live and neutral
UK 230 volt 50Hz Fused 1 Ampere on live and neutral
Rated Power Consumption 100 Watts
The power supply is built to Class II standard. The instrument is not
connected to mains earth. (Class II BF).
Medical With respect to electrical shock fire and mechanical
Equipment hazards only in accordance with UL60601-1, IEC60601-1,
CAN/CSA C22.2 No 601.1 and IEC60601-2-2
0473 CE mark
Attention: Consult accompanying
documents
!
Document Number 111/00 Operators Manual August 2007
Digital Display Reading Digital display reads biological impedance
STIMULATION
Voltage Amplitude Continuously variable between 0 - 3 volts ±10%.
The voltage supplied is displayed on the Digital
Meter in 0.01 Volt steps.
Accuracy ± 10 %
Pulse Rates 2 (Motor) (pulses per second)
50 or 100 (Sensory)
(Default 110v units 50Hz and 100 Hz for 230 V units)
Pulse Rate Accuracy ± 5 %
Pulse Width 1mSec ± 10 %
Waveform Biphasic pulses. Negative pulse leading
Lamp Indicator Green LED flashes at pulse rate according to the
switch which has been selected (Motor or Sens)
Indicator only lights when output is present.
Safety Interlock When stimulate Mode is selected Stimulation
Voltage cannot be delivered unless the Stimulate
Output Voltage Control is first set to the “Off”
position (fully anticlockwise). A buzzer will sound if
this is attempted.
RF LESION POWER RF frequency 480 KHz (±10%) sine wave
Document Number 111/00 Operators Manual August 2007
NT-500 Lesion Generator
NT-500 RF Lesion Generator ™
2-3
OPERATORS MANUAL
Power Output Continuously variable. Maximum power
output 25 Watts into 200 Ohms.
Voltage Indication Bar Graph 0 -80 Volts RF
Current Indication Bar Graph 0-800mA RF
Lamp Indicator Amber Lesion Power LED Flashes when
Lesion Power is being delivered
Audible Tone When Lesion is complete an audible tone
will sound until the RF Output control is
turned off (fully anticlockwise)
Pulsed / Pulse Dose (Optional When ordering unit)
RF bursts of 20mS at a frequency of 2.0 Hz
Pulse Temp limit 42º C (option of no limit when ordering)
Pulse Dose Limit 42º C
Pulsed RF Option (optional when ordering unit)
If the unit has the option to perform a procedure using Pulsed RF the
concept is the same as conventional RF lesions, except that the Physician
selects the pulsed option by pressing the “Pulsed” button.
In Pulsed RF Mode the unit will deliver a Burst of RF 20mS long at a
frequency of 2 pulses per second. The default temperature limit is set to 42ºC
(or no limit as ordered with unit)
And the default time is set to 120 seconds. The physician can select other
time options from the “Time / Count” button these are 120, 240, 30, 60, and
90 seconds
The sensory and motor stimulation should be carried out to ensure the
correct placement of the electrodes. When the physician is satisfied that the
electrode is in the correct position. Select the RF Output button and rotate the
Control anticlockwise until the RF Voltage indicator reads 45 volts
It should be noted that this is not performing a Thermal Lesion in pulsed RF
Mode, and it is quite normal for the Temperature to not reach the 42ºC setting.
The perfect treatment delivers 45 volt pulses for the full duration of the
treatment. If the 42ºC limit is reached the NT-500 will reduce the width of the
RF Burst until the temperature is below 42ºC. Below 42ºC the pulse will return
to 20 mS full width pulses.
Pulse Dose Option (optional when ordering unit)
Document Number 111/00 Operators Manual August 2007
NT-500 Lesion Generator
NT-500 RF Lesion Generator ™
2-4
OPERATORS MANUAL
Pulse Dose delivers a pre determined number of full 20mS bursts of
RF. Set the RF output to 45 volts on the RF Voltage indicator.
If the unit temperature reaches 42ºC then the counter will stop and no
further Pulses will be delivered, until the temperature falls below 42ºC
at this point the count will continue until the counter reaches the set
limit. As only full pulses are being delivered the physician can set the
number of pulses from 240 (Default) 360, 480, or 120. This
removes the variables of other Pulse applications and ensures repeatable results.
To Use Pulse Dose (optional when ordering unit)
Set up the NT-500 to enable the electrode to be placed as for a conventional
lesion. Carry out the sensory and motor stimulation tests until the Physician is
satisfied with the placement of the electrode. Make an X-Ray record for the
patient file and prepare the patient for the procedure.
Press the Pulsed buttonand if required change the number of pulses by
pressing the Time/Count button.
The count will cycle through the option of count numbers release the button
when the count number is reached. Select RF Output and rotate the RF
Output control anticlockwise until the RF volt indicator reads 45 volts and
press the Start button. The counter will count down to zero, if the temperature
goes above 42ºC, the count will stop until the temperature is below 42ºC. The
count will resume at this point until the process is complete.
At the end of the procedure the power will be switched off and a tone will
sound. To stop the tone set the RF Output to the off position
LESION TIMER
Time Start The timer will start after the Timer start button is pressed, At the
end of the procedure an alarm tone will sound and RF power is
removed from the output. The timer resets to the default setting.
The Alarm tone can be stopped by turning the RF output Control
off. (Fully anticlockwise)
Time Indicator Clock indicates the amount of time
remaining for lesioning.
Lesion Time Defalt is 60 Seconds
Timer/Count Time set button cycles 30, 60, 90, 120 and
240 seconds (Pulsed RF Version) default
120 Seconds
Count Pulse Dose Count set cycles 120, 240, 360, and 480
counts (Pulse dose Version) default 240
Seconds
Audible Indicator A one second “tick” indicates the clock is
operating.
TEMPERATURE MONITOR
Document Number 111/00 Operators Manual August 2007
NT-500 Lesion Generator
NT-500 RF Lesion Generator ™
2-5
OPERATORS MANUAL
Meter Range 30ºC to 100º C Bar Display
The Bar Display has 3 colour segments
30ºC - 42ºC Amber
42ºC - 95ºC Green
95ºC - 100ºC Red
Temperature set Modes The lesion Temperature can be selected
from the front panel buttons 70, 80, and
90ºC (Default setting 80ºC)
Probes Use only NT-500 probes - no individual
adjustment necessary.
Thermocouple Indicator LED shows Green when probe is
connectedor Red when disconnected or
faulty.
SAFETY FEATURES
Safety Cut Out RF Power is automatically reduced if
temperature reaches set limit as
determined by Maximum Temperature
Control, (either 70ºC, 80ºC or 90ºC). RF
Power will automatically proceed again once the
temperature of the tip of the probe
drops1ºC.
Safety Interlock a) RF Power cannot be delivered unless
the RF Power Control is first set to
the OFF position, this prevents any
accidental application of the RF
power. If the lesion power is selected
and power control is on, a warning
tone is given out and no RF Power
is provided to the probe.
b) The Lesion Power will also be cut off
if any of the conditions listed below
occur.
a) Measured temperature goes
above 95ºC
b) RF Power Oscillator detects a
short circuit or low patient
impedance.
2.2 EARTH LEAKAGE DATA
Tests carried out as per IEC 601-1
Document Number 111/00 Operators Manual August 2007
NT-500 Lesion Generator
NT-500 RF Lesion Generator ™
2-6
OPERATORS MANUAL
Typical Max Allowable
1 Enclosure leakage current
Normal
Reverse
Single Fault condition
Normal
Reverse
2 Patient Leakage Current
Normal (AC)
Reverse (AC)
Single Fault condition
Normal (AC)
Reverse (AC)
3 Patient Leakage Current
Normal (DC)
Reverse (DC)
Single Fault Condition
Normal (DC)
Reverse (DC)
4 Patient Auxiliary Leakage
Normal (AC)
Reverse (AC)
Single Fault condition
Normal (AC)
Reverse (AC)
5 Patient Auxiliary Leakage
Normal (DC)
Reverse (DC)
Single Fault condition
Normal (DC)
Reverse (DC)
6 Patient Leakage Floating Type
Normal
Reverse
Single Fault condition
Normal
Reverse
40 microamps
40 microamps
40 microamps
40 microamps
4 microamps
4 microamps
4 microamps
4 microamps
4 microamps
4 microamps
4 microamps
4 microamps
4 microamps
4 microamps
4 microamps
4 microamps
4 microamps
4 microamps
4 microamps
4 microamps
25 microamps
25 microamps
25 microamps
25 microamps
100 microamps
40 microamps
500 microamps
500 microamps
100 microamps
100 microamps
500 microamps
500 microamps
10 microamps
10 microamps
50 microamps
50 microamps
100 microamps
100 microamps
500 microamps
500 microamps
10 microamps
10 microamps
50 microamps
50 microamps
5000 microamps
5000 microamps
5000 microamps
5000 microamps
Document Number 111/00 Operators Manual August 2007
NT-500 Lesion Generator
Loading...
+ 22 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.