The NeurOptics® PLR®-3000 Pupillometer offers researchers quantitative infrared technology to objectively
and accurately measure pupil size and dynamics in an advanced design. The system acquires images using
self- contained infrared and visible illumination sources and a digital camera. It analyzes the captured image
data and displays a summary of the measurement in the LCD window. The NeurOptics PLR-3000 Pupillometer
uses a menu driven graphical user interface (GUI), with a color touchscreen LCD screen for data display.
A keypad completes the user interface and enables manual entry of individual patient identification (ID).
Indications for Use
The PLR-3000 Pupillometer is a hand held optical scanner which measures pupil size and dynamics under
different light and experimental protocols. The results obtained from the Pupillometer scans are used for
information only and are not to be used for clinical diagnostic purposes. It should only be operated by properly
trained clinical personnel, under the direction of a qualified physician.
Contraindications
Avoid use when the orbit structure is damaged, or surrounding soft tissue is edematous or has an open lesion.
Warnings and Cautions appear throughout this manual where
they are relevant. The Warnings and Cautions listed here apply
generally anytime you operate the device.
• The Pupillometer is intended for use by trained personnel.
• If a problem is recognized while operating the device, the
device must be removed from use and referred to qualified
personnel for servicing. Using an inoperative device may
result in inaccurate readings.
• Electric shock hazard - Do not open the device or the charging
station. There are no user serviceable parts.
• The battery in the PLR®-3000 Pupillometer is only replaceable
by a qualified ser vice technician. Contact NeurOptics if you
suspect an inoperable battery.
• Use only the NeurOptics PLR®-3000 Charging Station for
charging the Pupillometer.
• Risk of fire or chemical burn: This device and its components
may present a risk of fire or chemical burn if mistreated. Do not
disassemble, expose to heat above 100°C, incinerate, or
dispose of in fire.
Cautions
The following cautions apply when cleaning the device or when
sterilizing device accessories.
• The internal components of the Pupillometer are not compatible
with sterilization techniques, such as ETO, Steam Sterilization,
Heat Sterilization and Gamma.
• DO NOT submerge the device or pour cleaning liquids over or
into the device.
• DO NOT use acetone to clean any surface of the Pupillometer
or Charging Station.
Electromechanical Compatibility (EMC) Notice
This device generates, uses, and can radiate radio frequency
energy. If not set up and used in accordance with the instructions
in this manual, electromagnetic interference may result. The
equipment has been tested and found to comply with the
limits set forth in EN60601-1-2 for Medical Products. These
limits provide reasonable protection against electromagnetic
interference when operated in the intended use environments
(e.g. hospitals, research laboratories).
Magnetic Resonance Imaging (MRI) Notice
This device contains components whose operation can be
affected by intense electromagnetic fields. Do not operate the
device in an MRI environment or in the vicinity of high-frequency
surgical diathermy equipment, defibrillators, or short-wave therapy
equipment. Electromagnetic interference could disrupt the
operation of the device.
Bluetooth® Notice
Do not attempt to pair the PLR-3000 Pupillometer with the
NeurOptics® Antimicrobial Barcode Scanner by Socket®
while simultaneously using another barcode scanner in
close proximity.
Classification
Type of Equipment: Medical Equipment, Class 1 886.1700
Trade name: NeurOptics® PLR®-3000 Pupillometer
Manufactured by:
NeurOptics, Inc.
18101 Von Karman Avenue, Suite 1940
Irvine, CA 92612, USA
p: 949.250.9792
Toll Free North America: 866.99.PUPIL
info@NeurOptics.com
This work is protected under Title 17 of the U.S. Code and is
the sole property of NeurOptics, Inc. (the Company). No part
of this document may be copied or otherwise reproduced, or
stored in any electronic information retrieval system, except
as specif ically permitted under U.S. Copyright law, without the
prior written consent of the Company.
This device complies with Part 15 of the Federal
Communications Commission (FCC) Rules. Operation is
subject to the following two conditions: (1) this device may
not cause harmful interference, and (2) this device must
accept any interference received, including interference
which may cause undesired operation.
•Please review the following safety information prior to operating the device.
•Please read the Operating Instructions fully before attempting to use the Pupillometer. Attempting to operate the
device without fully understanding its features and functions may result in unsafe operating conditions and/or
inaccurate results.
•If you have a question regarding the installation, setup, operation,
or maintenance of the device, please contact NeurOptics.
Unpacking the Pupillometer
The NeurOptics PLR®-3000 Pupillometer is packaged
with the following components (Ex. 1):
•PLR-3000 Pupillometer
•PLR®-3000 Charging Station
•Eye Cups (2)
•PLR®-3000 Power Supply Adaptor
•PLR®-3000 Instructions for Use
Ex. 1
Power Up
Initial Set-up
Connect the PLR-3000 Pupillometer Power Supply to the PLR-3000 Charging Station
and plug into a power outlet. The green light at the base of the Charging Station will
indicate power has been established (Ex. 2).
Place the PLR-3000 into its Charging Station. After powering on, the touchscreen
will display a blue battery icon indicating the PLR-3000 is charging. The battery
icon will turn green when fully charged (Ex. 3).
To modify the date and time, from the main screen, select the
Settings icon and then select Set Date and Set Time
(Ex. 4 and Ex. 5). Follow the prompts to input the proper date
and time using 24 hour time configuration and select Accept.
Turning On the PLR-3000
Ex. 2
Ex. 3
Ex. 4Ex. 5
When not in use, the PLR-3000 should be kept in the Charging Station. If the PLR-3000
is not in the Charging Station, to conserve battery life the Pupillometer will:
•Go into sleep mode after 5 minutes. Touch the screen to turn on.
•Power down after 30 minutes. Press and hold the UP arrow to turn on (red circle, Ex. 6).
Patient ID is reported in the main screen (Ex. 3) and it enables recall of patient data.
To assign a new Patient ID, from the main screen select icon and then select either
Barcode Scanner or Manual ID to indicate the patient ID entry method to be used (Ex. 7).
In case other ID numbers are included in the PLR®-3000 database stored in memory,
these will be listed in the same window (e.g. ID=123ABC, Ex. 7) and they can be
immediately re-entered.
Pairing the PLR-3000 to the NeurOptics Antimicrobial Barcode Scanner
Connect the NeurOptics Antimicrobial Barcode Scanner and Charging Cradle to the
power supply and plug into a power outlet (Ex. 8). Turn on the Barcode Scanner until
an audible beep is heard and a blue light on the device flashes. Position the
Barcode Scanner next to the PLR-3000.
On the PLR-3000, select Barcode Scanner. The PLR-3000 will
display
successfully paired, the touchscreen will prompt when the
device is ready to scan the patient ID barcode (Ex. 10).
“Connecting...”
on the touchscreen (Ex. 9). Once
Ex. 7
Ex. 8
The patient ID will now appear on the PLR-3000 touchscreen. Confirm the patient
information is correct and select Accept (Ex. 11). Use Customize ID function
in the Settings menu (icon from the main screen) to customize the barcode
reading (for example if only a part of the barcode has to be reported).
The PLR-3000 will display the patient ID number and read
“Ready to scan”
(Ex. 12).
Manual Entry of the Patient ID
Press Manual ID Entry. Using the touchscreen, press the Patient ID.
Select Shift to toggle from alpha to numeric as required. When
the patient ID number has been manually entered, check for
accuracy and press Enter (Ex. 13 & 14).
From the main screen (Ex. 12), select the Settings icon
and then the top right icon (Define Protocol, Ex. 15) to
navigate to the Change Protocol Parameters menu (Ex. 16).
Each parameter listed in the two pages of this menu (Ex. 16)
can be changed by toggling the values reported using the
left “<” and right “>” keys in the directional keyboard.
Ex. 15
Five different experimental protocols can be saved in
memory and retrieved later when a specific protocol is needed. Press Select to active a protocol. Use the
left “<” and right “>” keys in the directional keyboard to change the protocol number. Use the list of parameters
below to change the settings of the selected protocol.
Use this setting to change the type of the experiment:
Positive Pulse Stimulus: standard pupil light reflex, a bright pulse over a dimmer background triggers a pupil constriction.
Negative Pulse Stimulus: dim pulse over a brighter background triggers a dark reflex and pupil dilation.
Static Stimulus: no light stimulation and no pupil reflex, just a continuous and static light background.
Extended Mode: pupil is recorded at a lower frequency for an extended period of time (maximum duration 10 minutes).
Pulse Intensity. Use this setting to change the intensity of the light stimulus. Unit of light is radiometric and provided
in micro Watts, uW. Five different intensities are provided: 0uW, 1uW, 10uW, 50uW, 121uW, and 180uW.
Background Intensity. Use this setting to change the background light intensity. Note that in case of a Positive Pulse
Stimulus protocol, Background Intensity must be less than Pulse Intensity whereas, for a Negative Pulse Stimulus,
protocol it is the opposite -- Background must be higher than the Pulse Intensity.
Ex. 16
Measurement Duration. This setting is for setting the duration of the measurement, a minimum of 3 seconds up
to a maximum of 41 seconds.
Pulse Duration. This is the duration of the light pulse stimulation, a minimum of 0.03 sec for the Positive Pulse
and 1 sec for the Negative Pulse.
Pulse Onset. This is the start of the Pulse stimulation.
In case there is a conflict in the settings, for example the Background Intensity is set brighter than the Pulse Intensity
in a Positive Pulse Stimulus protocol, then the corresponding icons are shown in red and the Define Protocol menu
cannot be exited until the conflict is resolved.
Measure Pupils
The PLR-3000 should not be used without the eye cup positioned correctly (Ex. 17). It is
very important that the eye cup be correctly fitted. A snug fit helps reduce the possibility
of stray light entering the eye while the scans are taking place. The eye cup has a tab in
the rim, which fits into the indentation in the lens shield of the Pupillometer. Position the
tab in the eye cup rim into the indent in the lens shield and press into place. The tabs on
either side of the lens shield should also snap into the holes on either side of the eye cup.
Position the PLR-3000 with the eye cup at a right angle
to the patient’s axis of vision, minimizing any tilting of the
device (Ex. 18).
Press and hold either the RIGHT or LEFT until the eye is centered on the touch screen
and the display shows a circle around the pupil (Ex.19). Once the circle appears (green
for RIGHT and yellow for LEFT), release the button, holding the PLR-3000 in place for
the entire duration of the measurement. When the pupil measurement is complete, pupil
data are analyzed and then results will be displayed (Ex.20).
Results page for Static Stimulus
The results page for the Static Stimulus (Ex. 20) shows the diameter of the pupil in bold
and the standard deviation of pupil diameter measured (in parenthesis and smaller font)
during the scan. It also includes the ID number of the subject, the date and time of the
measurement and, finally, which eye, (Right or Left) was measured.
Results page for Positive or Negative Stimulus
The results page for the Positive Stimulus (Ex. 21) shows different information. The pupil
diameter waveform is plotted as a function of time. The two vertical yellow lines show
where the stimulus started and ended. The green vertical line shows the latency and the
blue line the T75. Latency and T75 are two of the variables calculated by the analysis and
they are explained in Appendix A. If a variable could not be calculated
(for example, because of a blink) it is reported with dashes or in red in the table.
Ex. 19
Ex. 20
Ex. 21
The characteristics of the light stimulus are summarized in the top right corner: P/DC reports the intensity of the
stimulus and the background in uW (here, 50uW for the pulse intensity over a 0uW background). S/D reports the
Pulse Onset (2.01 sec in the example) and Pulse Duration (0.80 seconds in the example).
Results page for Extended Mode
The results page for the Extended Mode shows the whole pupil function as a function
of time (Ex. 22). The colored vertical lines correspond to the five different keys of the
directional keypad. The user can press any of those keys during the recording and the
time(s) of the press (or presses) are reported in the plot and saved with the record.
Note that an extended pupil recording is ended by a press of the RIGHT or LEFT
key - the duration of the measurement is not defined.
From the results page, you may select the Video icon to view the video playback of
the scan. Records can also be deleted using Delete icon or printed using a Bluetooth
printer using the Print icon . Contact NeurOptics® for information about the printer.
To browse, retrieve and print data from previous measurements select icon from the
main screen. In the browse records menu (Ex. 23) select All Records to browse all records in
memory, or Specify Patient ID if only one specific patient needs to be retrieved. All
the most recent patient IDs are reported in the browse records catalog so that it is possible to
select directly from the catalog without having to re-enter the Patient ID using option Specify
Patient ID (e.g. ID=123ABC , Ex. 24).
Download Records
Data can be downloaded to an external computer via Bluetooth. Verify the computer is in
“Discoverable” mode (i.e., in the Bluetooth Settings, the box for “Allow Bluetooth devices to find this computer” should be checked.) Select the Bluetooth icon in the Settings menu
of the Pupillometer and then follow the instructions. The device might need to be paired
the first time. If so, a PIN is provided in the instructions. Initiate the Receive Files protocol
in the receiving external computer. You can either right click on the Bluetooth icon in the
notification area near the task bar and then click on “Receive a File” or:
• In Windows 7, click on the Start button and type fsquirt in the box that says Search programs and files.
Ex. 24
Ex. 25
• In Windows 8 and 8.1, open the Start screen and type fsquirt.
• In Windows 10, right-click Start and select Run, type fsquirt into the box and press Enter.
This will open the Bluetooth File Transfer window. Select the Receive Files option then press “Dismiss” on the Pupillometer
and go to the next page.
Select what type of data should be downloaded (Ex. 25).
• Select Records to download a file (Excel compatible) with all the records saved in memory.
• Select Video to download the video of the latest collected measurement.
The video is not in standard format. NeurOptics can provide a Matlab® script for reading and interpreting the video file (contact
NeurOptics® Customer Service at Toll Free North America: 866.99.PUPIL or +1 949.250.9792.)
The Pupillometer will display a list of available Bluetooth devices. Select the computer name set to Receive Files. Be aware
that the search for all available Bluetooth devices around the Pupillometer could take several minutes. The list is updated
continuously on the screen of Pupillometer. Follow the instructions shown on the Computer and Pupillometer then wait until
the download is complete. Press the Close link button on the Pupillometer to complete the download.
Power Down
To turn the PLR-3000 off, select the from the main screen and confirm Yes.
ALWAYS handle the PLR-3000 Pupillometer and PLR-3000 Charging Station with care because sensitive metal, glass,
plastic and electronic components are contained inside. The PLR-3000 Pupillometer and PLR-3000 Charging Station can
be damaged if dropped, or if they come in contact with liquid.
The PLR-3000 Pupillometer and PLR-3000 Charging Station do not require any regularly scheduled maintenance. If
the PLR-3000 Pupillometer and PLR-3000 Charging Station are not working properly, or are believed to have been
damaged, immediately contact NeurOptics Customer Service at Toll Free North America: 866.99.PUPIL (866-997-8745),
international: +1-949-250-9792, or email: Support@NeurOptics.com.
Cleaning the PLR-3000 Pupillometer and PLR-3000 Charging Station
Isopropyl alcohol (IPA)-based cleaning solutions, in formula concentrations up to 70% IPA (70% IPA), are recommended
for use in cleaning the PLR-3000 Pupillometer and PLR-3000 Charging Station. Do not use chemicals that can damage
the pupillometer and charging station surface. Some chemicals can weaken or damage plastic parts and may cause
instruments to not operate as intended. Use all cleaning products per manufacturer’s instructions, being careful to
squeeze out excess liquid prior to wiping the pupillometer and charging station and do not use an oversaturated cloth.
Wipe all exposed surfaces. Follow the cleaner’s manufacturer instructions as to the time required to leave the solution
on the device surface.
• DO NOT allow any other cleaner other than 70% IPA to contact the gold connector blades located on the bottom of the PLR-3000 Pupillometer handle, and the gold connector pins located in the base of the PLR-3000
Charging Station.
• DO NOT use an oversaturated cloth. Be sure to squeeze out excess liquid prior to wiping the PLR-3000 Pupillometer or the PLR-3000 Charging Station.
• DO NOT allow the cleaner to collect on the instrument.
• DO NOTuse any hard, abrasive or pointed objects to clean any part of the PLR-3000 Pupillometer or PLR-3000
Charging Station.
• DO NOT immerse the PLR-3000 Pupillometer or the PLR-3000 Charging Station in liquid, or attempt to sterilize the product, as damage to the electronic and optical componentry could occur.
Drying and Inspection Following Cleaning
Confirm the PLR-3000 Pupillometer is thoroughly dry before placing in the PLR-3000 Charging Station to charge.
Once thoroughly dry, place the PLR-3000 Pupillometer into the PLR-3000 Charging Station and plug in the PLR-3000
Power Supply to the back of the charging station to power ON.
• DO NOT place the PLR-3000 Pupillometer into the PLR-3000 Charging Station until all components are completely dry.
• DO NOT reconnect the PLR-3000 Power Supply to the PLR-3000 Charging Station until all components are completely dry.
Cleaning Considerations: Gold Connector Pins and Blades
In instances where there is concern of exposure to highly resistant bacteria or viruses (ie: Clostridium difficile, or “C.
diff”), we understand that hospital protocols may require use of cleaning solutions containing sodium hypochlorite
(bleach) when cleaning equipment. Please be aware solutions containing sodium hypochlorite (bleach) will corrode
the gold connector blades located on the bottom of the PLR-3000 Pupillometer handle (Figure 26), and the gold
connector pins located in the base of the PLR-3000 Charging Station (Figure 27.)
PLR-3000 Pupillometer
Figure 26- Gold Connector Blades
• DO NOTuse products containing sodium hypochlorite (bleach) to clean the gold connector blades located
on the bottom of the PLR-3000 Pupillometer handle, and the gold connector pins located in the base of the
PLR-3000 Charging Station
Gold
Connector
Blades
PLR-3000 Pupillometer
Gold
Connector
Pins
Figure 27- Gold Connector Pins
If products containing sodium hypochlorite (bleach) are used to clean the gold connector blades located on the
bottom of the PLR-3000 Pupillometer and the gold connector pins located in the base of the PLR-3000 Charging
Station, the cleaning process should be followed by a second cleaning using 70% IPA solution to ensure that all
residue is completely removed from the device in order to minimize damage to the gold connector pins and blades.
For best protection of the liquid crystal display (LCD), use a clean, soft, lint-free cloth and 70% IPA cleaning solution to
clean the pupillometer optics.
In instances where there is concern of exposure to highly resistant bacteria or viruses (ie: Clostridium difficile, or “C.
diff”), we understand that hospital protocols may require use of cleaning solutions containing sodium hypochlorite
(bleach) when cleaning equipment. If products containing sodium hypochlorite (bleach) are used to clean the LCD of
the PLR-3000 Pupillometer, the cleaning process should be followed by a second cleaning solution with a 70% IPA
solution to ensure that all bleach residue is completely removed from the LCD using a clean, soft, lint-free cloth.
BCS-CC-01 NeurOptics® Antimicrobial Barcode Scanner by Socket®
NEUR-PRTS445-BTSeiko printer
Customer Service
For technical support, or if you have a question about your order, please contact NeurOptics Customer Service.
Toll Free North America: 866.99.PUPIL(866 - 997-8745) |
email: Support@neuroptics.com
p: 949.250.9792
Returned Goods Policy
Products must be returned in unopened packages, with manufacturer’s seals intact, to be accepted for replacement
or credit, unless returned due to a complaint of product defect or mislabeling. Determination of a product defect or
mislabeling will be made by NeurOptics, which determination will be final. Products will not be accepted for replacement
or credit if they have been in the possession of the customer for more than 30 days.