LS-901AJ/AK/AG Photo Control Unit ......................................................................... 8.3
JE-913A/AG Mini Junction Box..................................................................................8.6
C.4Service Manual EEG-9100/9200
GENERAL HANDLING PRECAUTIONS
This device is intended for use only by qualified medical personnel.
Use only Nihon Kohden approved products with this device. Use of non-approved products or
in a non-approved manner may affect the performance specifications of the device. This
includes, but is not limited to, batteries, recording paper, pens, extension cables, electrode
leads, input boxes and AC power.
Please read these precautions thoroughly before attempting to operate the instrument.
1. To safely and effectively use the instrument, its operation must be fully understood.
2. When installing or storing the instrument, take the following precautions:
(1) Avoid moisture or contact with water, dust, extreme atmospheric pressure, excessive humidity and temperatures,
poorly ventilated areas, and saline or sulphuric air.
(2) Place the instrument on an even, level floor. Avoid vibration and mechanical shock, even during transport.
(3) Avoid placing in an area where chemicals are stored or where there is danger of gas leakage.
(4) The power line source to be applied to the instrument must correspond in frequency and voltage to product
specifications, and have sufficient current capacity.
(5) Choose a room where a proper grounding facility is available.
3. Before Operation
(1) Check that the instrument is in perfect operating order.
(2) Check that the instrument is grounded properly.
(3) Check that all cords are connected properly.
(4) Pay extra attention when the instrument is in combination with other instruments to avoid misdiagnosis or other
problems.
(5) All circuitry used for direct patient connection must be doubly checked.
(6) Check that battery level is acceptable and battery condition is good when using battery-operated models.
4. During Operation
(1) Both the instrument and the patient must receive continual, careful attention.
(2) Turn power off or remove electrodes and/or transducers when necessary to assure the patient’s safety.
(3) Avoid direct contact between the instrument housing and the patient.
5. To Shutdown After Use
(1) Turn power off with all controls returned to their original positions.
(2) Remove the cords gently; do not use force to remove them.
(3) Clean the instrument together with all accessories for their next use.
6. The instrument must receive expert, professional attention for maintenance and repairs. When the instrument is
not functioning properly, it should be clearly marked to avoid operation while it is out of order.
7. The instrument must not be altered or modified in any way.
8. Maintenance and Inspection:
(1) The instrument and parts must undergo regular maintenance inspection at least every 6 months.
(2) If stored for extended periods without being used, make sure prior to operation that the instrument is in perfect
operating condition.
Service Manual EEG-9100/9200i
(3) Technical information such as parts list, descriptions, calibration instructions or other information is available for
qualified user technical personnel upon request from your Nihon Kohden distributor.
9. When the instrument is used with an electrosurgical instrument, pay careful attention to the application and/or
location of electrodes and/or transducers to avoid possible burn to the patient.
10. When the instrument is used with a defibrillator, make sure that the instrument is protected against defibrillator
discharge. If not, remove patient cables and/or transducers from the instrument to avoid possible damage.
WARRANTY POLICY
Nihon Kohden Corporation (NKC) shall warrant its products against all defects in materials and workmanship for one year
from the date of delivery. However, consumable materials such as recording paper, ink, stylus and battery are excluded from
the warranty.
NKC or its authorized agents will repair or replace any products which prove to be defective during the warranty period,
provided these products are used as prescribed by the operating instructions given in the operator’s and service manuals.
No other party is authorized to make any warranty or assume liability for NKC’s products. NKC will not recognize any other
warranty, either implied or in writing. In addition, service, technical modification or any other product change performed by
someone other than NKC or its authorized agents without prior consent of NKC may be cause for voiding this warranty.
Defective products or parts must be returned to NKC or its authorized agents, along with an explanation of the failure.
Shipping costs must be pre-paid.
This warranty does not apply to products that have been modified, disassembled, reinstalled or repaired without Nihon
Kohden approval or which have been subjected to neglect or accident, damage due to accident, fire, lightning, vandalism,
water or other casualty, improper installation or application, or on which the original identification marks have been
removed.
In the USA and Canada other warranty policies may apply.
CAUTION
United States law restricts this device to sale by or on the order of a physician.
iiService Manual EEG-9100/9200
EMC RELATED CAUTION
This equipment and/or system complies with the International Standard IEC60601-1-2 for electromagnetic
compatibility for medical electrical equipment and/or system. However, an electromagnetic environment
that exceeds the limits or levels stipulated in the IEC60601-1-2, can cause harmful interference to the
equipment and/or system or cause the equipment and/or system to fail to perform its intended function or
degrade its intended performance. Therefore, during the operation of the equipment and/or system, if
there is any undesired deviation from its intended operational performance, you must avoid, identify and
resolve the adverse electromagnetic effect before continuing to use the equipment and/or system.
The following describes some common interference sources and remedial actions:
1.Strong electromagnetic interference from a nearby emitter source such as an authorized radio station or
cellular phone:
Install the equipment and/or system at another location if it is interfered with by an emitter source such
as an authorized radio station. Keep the emitter source such as cellular phone away from the equipment
and/or system.
2.Radio-frequency interference from other equipment through the AC power supply of the equipment and/
or system:
Identify the cause of this interference and if possible remove this interference source. If this is not
possible, use a different power supply.
3.Effect of direct or indirect electrostatic discharge:
Make sure all users and patients in contact with the equipment and/or system are free from direct or
indirect electrostatic energy before using it. A humid room can help lessen this problem.
4.Electromagnetic interference with any radio wave receiver such as radio or television:
If the equipment and/or system interferes with any radio wave receiver, locate the equipment and/or
system as far as possible from the radio wave receiver.
If the above suggested remedial actions do not solve the problem, consult your Nihon Kohden
Corporation subsidiary or distributor for additional suggestions.
The CE mark is a protected conformity mark of the European Community. The products herewith comply
with the requirements of the Medical Device Directive 93/42/EEC.
The CE mark is only applied to the EEG-9100K/G and EEG-9200K/G Electroencephalograph.
This equipment complies with EUROPEAN STANDARD EN-60601-1-2 (1993) which requires EN-55011,
class B.
Service Manual EEG-9100/9200iii
Conventions Used in this Manual and Instrument
Warnings, Cautions and Notes
Warnings, cautions and notes are used in this manual to alert or signal the reader to specific information.
WARNING
A warning alerts the user to the possible injury or death associated with the use or misuse of the
instrument.
CAUTION
A caution alerts the user to possible injury or problems with the instrument associated with its use or
misuse such as instrument malfunction, instrument failure, damage to the instrument, or damage to other
property.
NOTE
A note provides specific information, in the form of recommendations, prerequirements, alternative
methods or supplemental information.
Explanations of the Symbols in this Manual and Instrument
The following symbols found in this manual/instrument bear the respective descriptions as given.
Power supply unit, SC-901A/AK/AG (for EEG-9100A/J/K/G)
Isolation unit, SM-930AA/AJ/AK (for EEG-9200A/J/K/G)
SymbolDescriptionSymbolDescription
Alternative current
Equipotential ground terminalSerial number
Attention, consult operator’s
manual
Protective groundDate of manufacture
The CE Mark is a protected
conformity mark of the
European Community. The
products herewith comply with
the requirements of the Medical
Device Directive 93/42/EEC.
ivService Manual EEG-9100/9200
EMC
The product complies with
IEC60601-1-2 (1993)
(for sales in Japan only)
PC unit, CC-901AK (for EEG-9100A/J/K/G)
SymbolDescriptionSymbolDescription
Power/suspend indicatorRS-232C connector
9
A
Symbols on the PC unit differ according to model. Refer to the Operator’s manual of the PC unit.
For the symbols of the following equipments, refer to each Operator’s manual.
• Magneto-optical disk drive
• Printer
Num Lock indicatorMouse connector
Caps Lock indicatorUSB connector
Scroll Lock indicatorPRT connector
Hard disk access lamp indicatorUnlock icon
Battery charging indicatorProtective ground
Power socket
Attention, consult operator’s
manual
Service Manual EEG-9100/9200v
PC unit, CC-902AK (for EEG-9200A/J/K/G)
SymbolDescriptionSymbolDescription
Standby (power on/off)USB connector
Hard disk access lampVideo connector
Headphone connectorNetwork connector
Printer portAudio jack (Microphone)
Serial portAudio jack (LINE OUT)
Mouse connectorAudio jack (LINE IN)
Keyboard connector
Protective ground
Attention, consult operator’s
manual
Symbols on the PC unit differ according to model. Refer to the Operator’s manual of the PC unit.
For the symbols of the following equipments, refer to each Operator’s manual.
• Display
• Magneto-optical disk drive
• Laser printer
viService Manual EEG-9100/9200
Options
Electrode junction box, JE-910A/AG, JE-911A/AG
SymbolDescriptionSymbolDescription
Type BF applied part
Flash lamp assembly, LS-703A/LS-706A
SymbolDescriptionSymbolDescription
Hot surface
Photo control unit, LS-901AJ/AK/AG
SymbolDescriptionSymbolDescription
Attention, consult operator’s
manual
Attention, consult operator’s
manual
Attention, consult operator’s
manual
Serial number
Photo control unit (inside)
SymbolDescriptionSymbolDescription
Alternative currentDate of manufacture
Equipotential ground terminal
High voltageGround
Protective ground
Service Manual EEG-9100/9200vii
On screen
SymbolDescriptionSymbolDescription
Warning query that displays a
Display of list box
warning or caution for
operation.
Warning message that displays a
Scrolling of data, list and others
warning or caution for operation
you to do something.
Check box
Maximize
Restore
Minimize
Window maximize/resize
minimize button
Option buttonClose button
viiiService Manual EEG-9100/9200
Precautions for Input Jack Use
NOTE
Do not perform EEG measurement without the Z, C3, C4, A1 and A2 electrodes.
Use of input jack Z
Connect the lead from the electrode (Z electrode) attached on the patient’s nasion to the input jack Z on the electrode
junction box. The purpose of this input jack is to eliminate AC interference positively.
NOTE
The input jack Z is also used for checking electrode impedance.
Use of input jacks C3 and C4
Connect the leads from the electrodes attached on the positions C3 and C4 to the input jacks C3 and C4 respectively.
NOTE
• The C3 and C4 electrodes are the system reference electrodes for EEG measurement.
• The input jacks C3 and C4 must be attached for EEG measurement even when the C3 and C4 are not
programmed in any montage.
Use of input jacks A1 and A2 (or FP1 and FP2), C3 and C4 during skin-electrode impedance check
When checking each electrode impedance, connect the leads from the electrode attached on the positions A1, A2, C3 and C4
to the input jacks A1, A2, C3 and C4 respectively.
NOTE
• The A1 and A2 (or FP1 and FP2) electrodes are the reference electrodes for skin-electrode impedance
check.
• The input jacks A1 and A2 (or FP1 and FP2) in addition to the Z, C3 and C4 must be attached for the
electrode impedance check.
PG1
22
1
T1
25
F7
13
3
Fz11T315C35Cz24C46T416A2
T5
7
17
9
FP1
F3
P3
OI
Z
Fz
19
Pz
20
PG2
23
FP2
2
T2
F8
F4
4
P4
8
O2
10
26
14
12
T6
18
Checking original electrode potentials for all active electrodes
Check the original electrode potential for all active electrodes by programming a montage with the system reference (Select
the 0 V button for reference electrode on the Montage dialog box). Refer to “Programming Patterns” in Section 4.
The digital EEG displays the EEG waveform in each channel by subtracting two electrode potentials selected to a montage.
The subtracted result will be incorrect, if the electrode attachment is not correct, the original electrode potential is flat or
unstable, or artifact is superimposed on the original electrode potential. Omit the measurement result if the displayed EEG
In the E11CFG.ini configuration file, you can change the system settings which
cannot be changed in the System Program. This configuration file manages
advanced settings in the Acquisition and Review program
CAUTION
Only use the specified personal computer . Otherwise the EEG-9000
system program does not operate correctly.
NOTE
• When changing the settings in the Configuration file, close the all
EEG-9000 application programs.
• Before changing the contents of the E11CFG.ini configuration file,
back up the original E11CFG.ini configuration file, just in case.
Opening the Configuration Settings File Editor
Window
1. Right-click the My Computer icon on the desktop. The
pop-up menu opens.
2. Select Explore. The My Computer window opens.
3. Double- click the C:\nfx11\INIEditor.exe file. The File
Open dialog box opens.
Service Manual EEG-9100/92002.1
2. CHANGING SETTINGS
4. Double-click the E11CFG.ini file. The Configuration Settings
File Editor window opens.
5. Double-click the item that you want to change. The Change
dialog box opens.
6. Type the new setting in the Value text box and click the
OK button.
To cancel changing click the Cancel button.
You can change the configuration settings by opening the E11CFG.ini
configuration file. To open the configuration file, type C:/NFX11/E11CFG.INI in
the open text box (Start menu → Run→ Reun dialog box).
2.2Service Manual EEG-9100/9200
2. CHANGING SETTINGS
Configuration File List and Settings
The following describes the default settings of the configuration file.
///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////
// //
// Set various advanced settings in EEG-1000 Acquisition program and Review program //
// //
///////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////////
[Acquisition]
/////////////// Advanced settings in Acquisition /////////////////////////////
// Up to version 02-10
// Amp bar sizeWHS800:=48,18,101024:=72,22,141280:=90,28,171600:=112,34,21
AmpBarFontSize=14
// Auto record timer mode Increase/Decrease
AutoRecordTimer=Decrease
// Automatic file close interval (msec)
Backup_msec_Interval=2000
// Main patient database drive
MainDataBaseDrive=C
// Main patient database maximum size KB
MainDataBaseSpace=61440
// Mark event name
MarkerStartStatus=MARK ON
MarkerStopStatus=MARK OFF
// Show patient infomation dialog bo x at start. Clear items or not OFF/Clear/NotClear
PatientDlgonNew=NotClear
// Resource DLL for acquisition program
ResourceFileName=c:\nfx11\A11ResEng.dll
// Pop up message bo xStorageMsgMode 0:filing 1:not filing 2:record time
//StorageMsgInterval 1000:1 sec, ...5000:5 sec, 0:Do not display, 1:Do not blink
StorageMsgMode=1
StorageMsgInterval=2000
StorageMsgColor=0,0,128
StorageMessage=Filing
StorageStopMessage=Not Filing
StorageMsgFont=Arial Bold
StorageMsgSize=48
// CAL voltage in a EEG file 0-9 CODE5=50uV
StorageCalVolt=5
// Local patient database maximum size KB
// Tool bar sizeWH800:=23,231024:=27,271280:=35,351600:=45,45
ToolBarSwitchHeight=27
//Change display resolution automatically
WndResolution=0
// V0301
Service Manual EEG-9100/92002.3
2. CHANGING SETTINGS
// Link HV unit and HV timer 0:Not link, 1:Link
HvTimerLinkMode=0
// Size for the LTM bar
LtmBarFontSize=16
LtmBarSwitchHeight=23
LtmBarSwitchWidth=77
// Size for the Photo and Timer bar.
PhotoAndTimerFontSize=14
PhotoAndTimerSwitchWidth=26
// V0310
// Adjust PC clock time to acquisition program clock time when file close 0:OFF, 1:ON 0:OFF, 1:ON)
AdjustClockTime=0
// If file must be saved after the setting time past (ON) ON or OFF
//=File must be saved after the setting time past. Unit min
// The automatic sa ve function at the time of patient information change. (ON or OFF)
AutoP atientChange=OFF
// Time of operation when using the automatic sa v e function. (Unit:Minutes)
AutoP atientChangeTime=20
// Impedance check time length in auto record (10) Unit:Second
AutoRecordImpchkTime=10
// Application name started at the time of WIF interface operation.
CreateProcess=
// Deterrence of a disk full message. (ON or OFF)
DisableDiskMsg=OFF
// Event String for Storage disk full.
DiskFull=Disk Full
// EEG-9000: Weight time of data reception processing of bulk-in transmission. (Unit:millisecond)
InputBoxDeviceBulkInSleep=3
// EEG-9000: Stole state recognition time. (Unit:millisecond)
InputBoxDeviceStallInterval=500
// EEG-9000: Automatic pipe reset at the time of stole detection. (ON or OFF)
InputBoxDeviceStallSw=OFF
// Event string for disconnecting Junction Box.
InputBoxDisconnect=JBox Disconnected
// EEG-9000: Time to deter ring buffer overflo w from stole state detection. (Unit:Second)
InputBoxOverrunDisableTime=60
// File automatic review function after Acquisition. (ON or OFF)
QuickReview=OFF
// V0401
// Acquisition Priority (-3,-2,-1,0,1, 0:Default)
ChangePriority=1
// Dummy bar hight
DummyBarHeight=10
// Dummy bar width
DummyBarWidth=456
// Use dummy bar ON/OFF (Used to display EEG wave half of the desktop area when using camera window)
DummyBarSW=OFF
// D V region check application name
LTMDVIndicatorPath=
2.4Service Manual EEG-9100/9200
// Open/close Timer bar link ed to Photo/HV bar . 0:Independent(Not link) 1:link
OpenCloseLinkPhotoAndTimer=0
// Pen recorder record timer. 0:Link to filing 1:Link to pen recorder
RecordTimerMode=0
// Show time in Camera window 0:Not show 1:show
ShowTimeInCameraWnd=1
// Show topographical montage top most mode
TopMostMontageMap=OFF
WriteBlockSize=1024 * 8
AmpBarSwitchWidth=72
AmpBarSwitchHeight=20
ToolBarSwitchWidth=29
PhotoAndTimerSwitchHeight=23
AutoFileSave=ON
AutoFileSaveTime=120
LTMExecDvtrJpn=c:\dvc\dvrec.exe /m:r /s:2
LTMExecDvtrUsa=c:\dvc\dvrec.exe /m:r /s:2
SubDataBaseSpace=10240
DelayTimeP ercentage=50
FileFlushCloseSw=ON
FileFlushCycleWriteInterval=0
2. CHANGING SETTINGS
[Common]
///////////////////////////////// Common settings /////////////////////////////
// Up to version 02-10
// ACC dialog auto close ON/OFF
AccAutoClose=ON
// Separator between year and month in patient infomation
AgeSeparator=M
// Channel number font size800:=111024:=121280:=131600:=14
ChannelFontSize=14
// Use same amp condition in all pattern ON/OFF
ConditionFreeMode=OFF
// Wav e width0:One line1:2:
CrtWaveWidth=0
// Create EEG-1000 registry when EEG registry not found
DummyRegistryCreate=1
//Color of EEG scale
EEGScaleColor=255,0,0
// Event dialog auto close ON/OFF
EventAutoClose=ON
// Put event before current time (sec)
EventDelay=0
// Event character font size800:=141024:=141280:=181600:=22
EventFontSize=14
// Horizontal cursor width
HorizontalCursor=0
// Mark channel wave color R,G,B
Service Manual EEG-9100/92002.5
2. CHANGING SETTINGS
MarkChColor=128,128,128
// Montage character font size800:=121024:=121280:=141600:=18
MontageFontSize=14
// A name of era next to Heisei
NewEra=NO_D ATA
// Patient search keyID/NAME/DATE/EEG
PatientSearchKey=ID
// Patient condition items (used f or Acquisition, Launcher, Fileutility, Database Maneger. not for Review)
//PatientTitleID=
PatientTitleName=Name
//PatientTitleSex=
//PatientTitleDOB=
//PatientTitleAge=
//PatientTitleHandedness=
//PatientTitleHeight=
//PatientTitleW eight=
//PatientTitleDate=
//PatientTitleEEGNo=
//PatientTitleInOut=
//PatientTitleRef erDept=
//PatientTitlePhysician=
//PatientTitleOperator=
//PatientTitleComment=
//PatientTitlePatientCondition=
//PatientTitleSeizureType=
//PatientTitleSleepDisorder=
//PatientTitleMedicalHistory=
//PatientTitleMedication=
// Channel number in Pattern Table
PatternCh=24
// Right side montage position
RMontPosition=100
// Time jump mode CLOCK/ELAPSED
TimeJumpMode=CLOCK
// Time mark color R,G,B
TimerMarkColor=255,255,0
// 30 seconds time mark in 5 min/page ON/OFF
TimerMarkof5min=ON
// Title in about dialog box
// Title descriptionID/NAME/DATE/EEG/VOL
TitleDescription=NAME,DATE,VOL
// Cursor width800:=01024:=01280:=11600:=1
VerticalCursor=0
// Wave color list for pattern setting
// W av eColor1-8 could not be changed because of EEG-2100 color
//Wa v eColor1=0,0,0
//WaveColor2=80,0,0
//WaveColor3=0,80,0
//WaveColor4=80,80,0
2.6Service Manual EEG-9100/9200
2. CHANGING SETTINGS
//WaveColor5=0,0,80
//WaveColor6=80,0,80
//WaveColor7=80,80,0
//WaveColor8=255,255,255
//Wa v eColor9=80,80,80
//Wa v eColor10=255,0,0
//Wa v eColor11=0,255,0
//WaveColor12=255,255,0
//Wa v eColor13=0,0,255
//WaveColor14=255,0,255
//WaveColor15=0,255,255
//WaveColor16=192,192,192
// Do not change following settings
// V0301
// Display type for the amp bar (0:Button , 1:Combobox)
AccSelectType=0
// Arrange the camera window and the Acquisition dialog automatically
// when opening the camera window . (0 or 1)
AutoWndArrangeWithCamera=0
// Font size of the time imformation displayed on the camera window.
CameraWndFontSize=24
// Width of the DC channel bar .
ChannelBarWidth=110
// Display Scale on the DC channel bar . (ON or OFF)
ChannelBarScaleDisp=ON
// Display limitation for the DC channel bar.
DcChannelLimit=10
// Width limitation f or the DC channel bar. (ON or OFF)
DCWaveNoLimit=ON
// Select OS when using camera IF card 0:win98 1:win2k or later
// V0310
// Event color of the manual event. (R.G.B)
DefaultEventColor=128,128,128
// EEG filter is made to continue at the time of pattern change. (ON or OFF)
EcgFilterAutoOff=OFF
// Drawing method when an event line overlaps with the time scale. (ON:Overwrite, OFF:XOR)
EventLinePriorityOver=ON
// Style of the event line. (0:Solid, 1:Dash, 2:Dot, 3:Dash_Dot, 4,Dash_Dot_Dot)
EventLineStyle=0
// Thickness of the event line. (from 1 to 3)
EventLineWidth=1
// The 15Hz high cut filter is transposed to 1Hz. (ON or OFF)
ExHf15Mode=OFF
// String of the 15Hz when a setup of ExHf15Mode is set as ON.
ExHf15String=1
// Display the Event line at the time of Mark-ON event. (ON or OFF)
MarkOnEventLine=ON
// Line color of the Page Comennt. (R,G,B)
PComEventColor=255,255,0
Service Manual EEG-9100/92002.7
2. CHANGING SETTINGS
// Width of a page comment pre vie w screen.
PCommentPreviewWidth=50
// The application name started at the time of page comment editing.
PComLaunchApp=mspaint
//EEG-9000: Size of the USB storage buffer (Unit:Second)
// Wav eform drawing margin. (Unit:%)
Wav eDr a wMargin=0
// V0401
// Montage font name and font size of impedance check dialog and average delete dialog
AmpListFont=Microsoft Sans Serif
AmpListSize=16
// Camera window time font color R,G,B
CameraWndTimeColor=0,0,0
// Camera window time area back color R,G,B (If no value sets, time area becomes transparent)
CameraWndTimeBkColor=192,192,192
// Minimum memory size when using BG FFT (Do not change)
MemoryCheck=120
// Montage focus move mode 0:G1->G1-> 1:G1->G2->G1->
MontageFocusMode=1
// Electrode font name and font size of pattern table and average delete dialog
MontButtonFont=Microsoft Sans Serif
MontButtonSize=14
// Show Fp1 and Fp2 electrode name in montage map
MontTopoFp1Fp2=ON
// Font name in montage map window
MontTopoFontName=MS Sans Serif
// Font size in montage map window (unit 0.01mm)
MontTopoFontSize=420
// Line width of the Topographical montage(unit 0.01 mm)
MontTopoLineWidth=10
// Show pattern name in Topographical montage window
MontTopoPattern=ON
// Font name and size of the montage setting dialog bo x in Pattern table
PattListF ont=Microsoft Sans Serif
PattListSize=16
// Motage description shift (-1000-1000) 0:No shift(over lap on waveform)
ShiftMontageString=-20
UsbStorageBuffSize=20
StorageBuffSize=30
WaveBuffSize=30
UseAltCamera=1
WDMVideoDevice=0
WDMAudioDevice=0
WDMVideoInputSource=0
WDMMasterStream=-1
WDMUseFrameRate=1
WDMVideoSignalType=1
WDMVideoBrightness=128
WDMVideoContrast=128
[Review]
////////////// Advanced settings in Review ////////////////////////////////////
// Up to version 02-10
// Amp bar sizeWHS800:=50,16,101024:=64,20,121280:=80,25,151600:=100,32,19
// Epoch time
EpochTime=10
// Beep at a event ON/OFF
EventBeep=OFF
// Page bar sizeWH800:=1024:=1280:=1600:=
PageControlBarSwitchHeight=25
// Page n umberON:absolute OFF:relative
PageSelect=OFF
// Font and font size in printing
PrintFontName=Microsoft Sans Serif
// Resource DLL for review program
ResourceFileName=c:\nfx11\R11ResEng.dll
// Selected wave background color
// Tool bar sizeWH800:=21,211024:=25,251280:=33,331600:=42,42
ToolBarSwitchHeight=25
//Change display resolution automatically
WndResolution=0
// V0301
// Wa v e position f or the Time information of FileMapping. (0-300)
FmapCurPos=0
// Font size of the Page comment.
PageCommentF ontSiz e=15
// Background color of the Page comment. (R,G,B)
PageCommentBgColor=255,255,155
//V0310
// Application name started at the time of WIF interface operation.
CreateProcess=
// Coefficient for DC inputs exporting as ASCII file.
DcConvCoef=1.00
// Height of the Event jump bar.
EventBarHeight=90
// Width of the Ev ent jump bar.
// Size of the Event jump bar. (ON:Half size OFF:Full Size)
EventBarHalfMode=OFF
// Hight of the button in the Event jump bar.
EventJumpBarButtonHeight=24
// Width of the button in the Event jump bar.
Service Manual EEG-9100/92002.9
2. CHANGING SETTINGS
EventJumpBarButtonWidth=90
// Font size of the Event jump bar.
EventJumpBarFontSize=12
// Add events when printing waveforms. (ON or OFF)
PrintedEvent=OFF
// Width of the event line at the time of printing (1/100mm unit)
PrintEventLineWidth=15
// Font size at the time of printing (1/100mm unit)
PrintFontSizeEx=300
// V0401
// DSA redraw interval in Jump bar(ms)
DrawIntervalDSABar=10
// DSA Dlg redraw interbal(ms)
DrawIntervalDSADlg=50
// DSA drawing volume per a timer event
DsaCalcIterationCount=2
// Spectral edge line color
DsaColorEdgeLine=255,255,0
// Spectral peek line color
DsaColorPeakLine=0,255,255
// Rec off area color in DSA dlg
DsaColorRecOff=192,192,192
// No (stage ) data area color in DSA dlg
DsaColorStageLost=0,255,255
// No data area color in DSA dlg
DsaColorStageOut=128,128,128
// Maximum DSA line in DSA Dlg(max=5)
DsaDlgBarSetCount=3
// Temporary sav ed DSA line bitmap image in bac kg round
DsaDlgBitimapStoreCount=24
// DSA voltage compress method when over lapped (1:peek 0:average)
DsaDrawType=1
// DSA height in DSA dialog
DsaEventHeight=25
// DSA height in jump bar
DsaHeight=50
// Time font size in DSA dialog Initial=13
DsaTimeScaleFontSize=13
// DSA window heignt in DSA dialog
DsaWndDlgHeight=50
// Digital video application name (This Item is set by Digital Video Installer)
DVLaunchPath=c:\dvc\dvplay.exe /s:2 /m:p /t:d
// Selected waveforms when stage changed in LIF review 0:Show dialog if save or not 1:Save 2:Not save (0 fixed)
ExportModeWhenChangingFile=0
// Set focus to the review after the review launched. Seconds after launched. 0:Not set focus
ForegroundTime=1
// Over write file when stage changed in LIF review. 0:Sho w dialog if overwrite or not 1:overwrite 2:Not Save (0 fixed)
LifAutoSave=0
// Maximum number of sample reference waveforms Initial value:100
2.10Service Manual EEG-9100/9200
2. CHANGING SETTINGS
MaxNumOfSampleData=100
// Maximum number of reference waveforms saved in drive C Initial:1000
MaxNumOfSnapData=1000
// Use waveform calculation Initial OFF
MeasureFunc=OFF
// Which mouse wheel direction is linked to page f ow ard. 0:fold forefinger=page f orwad 1:e xtension forefinger = page
forward default 0
MouseWheelPageMode=0
// Printed out font size of montage map (unit 0.01mm)
MontTopoPrintFontSize=200
// Transparent montage map when pringing ( ON / OFF(transparent))
MontTopoPrintPenetrate=OFF
// Printed out montage map position 0:center 1:right side down 2:left side down 3:right side up 4:left side up, (1)
MontTopoPrintPosition=1
// Printed out montage map size (unit 0.01mm)
MontTopoPrintSize=3500
// Selected area back color (R, G, B) (Initial 220,255,220)
SelectedAreaBKColor=220,255,220
// Show the time in camera window 0:do not show the time, 1:show the time (1)
ShowTimeInCameraWnd=1
// Show center line of the note wave ON/OFF (ON)
SnapMarkerLine=ON
// Note wave time length (unit second) (20)
SnapSampleLength=20
// Zoom window frame color (cannot use in V04-01)
Wav eMeasureFrameColor=0,0,255
SelectColor=180,250,180
EventBarWidth=180
AmpBarSwitchWidth=58
AmpBarSwitchHeight=20
AmpBarFontSize=12
ToolBarSwitchWidth=24
PageControlBarSwitchWidth=22
[PhotoUnitComm]
// COM port configuration for the Photo control unit.(LS-901A)
PhotoUnitCommPortNo=1
PhotoUnitCommPortSts=19200,n,8,1
// DVHS(JlIP) settings
DvhsControlCommPortNum=1
DvhsControlCommPort=9600,o,8,1
DvhsID1=1
DvhsID2=2
DvhsID3=3
DvhsID4=4
DvhsID5=5
// ALL TAPE WILL BE EJECTED
DvhsInitProcess1=08,41,60,00,00,00,00
// Time difference to the jump end point
JlipJumpFinishSec=40
// Time diff erence changing FF to Play FF
JlipJumpSlowSec=600
DvhsJump=ON
[EEG-Scope]
// Size for the Amp bar.
AmpBarFontSize=14
// Interval for the Auto refresh.
AutoRefreshInterval=5000
// Basic port number for the socket interface.
BaseSocketPort=2100
// Wait time for closing file
CloseWaitSec=10
// Intervals for connection
ConnectInterval=20
// Size for the Page control bar.
PageControlBarSwitchHeight=25
// Regtry count for the connection of socket interface.
PatientEditRetryCount=3
// Size for the Tool bar.
ToolBarSwitchHeight=25
//V0310
// Height of the Event jump bar.
EventBarHeight=90
2.12Service Manual EEG-9100/9200
// Width of the Ev ent jump bar.
// Size of the Event jump bar. (ON:Half size OFF:Full Size)
EventBarHalfMode=OFF
// Font size of the Remote annotation.
EventDisplayMsgSize=30
// Font color of the Remote annotation. (R,G,B)
EventDisplayMsgColor=0,0,255
// Font of the Remote annotation.
EventDisplayMsgFont=MS Sans Serif
// Display time of the Remote annotation. (Unit:Second)
EventDisplayMsgTime=10
// Hight of the button in the Event jump bar.
EventJumpBarButtonHeight=24
// Width of the button in the Event jump bar.
EventJumpBarButtonWidth=90
// Font size of the Event jump bar.
EventJumpBarFontSize=12
// V0401
// FFT data number per a dot in DSA window of the jump bar
DsaAver ageNum=1
// File access delay time (sec) after EEG-Scope getting a file name from acquisition
FileReadWaitTime=3
AmpBarSwitchWidth=68
AmpBarSwitchHeight=20
PageControlBarSwitchWidth=25
ToolBarSwitchWidth=25
EventBarWidth=180
[SCOPEMENUINFO]
///////////////////////////// SCOPEMENUINFO /////////////////////////
MenuTitle=”Option(&X)”
count=1
// QP/QV -110AK Digital Video remote re vie w mode (Ver.03-01)
MENUITEM1=”Digital Video Pla y”, 30001, 0000, “c:\dvc\dvplay.exe /S:3 /M:P /T:D”,SW_SHOWNORMAL
[AlphaBlending]
//////////////////////////// Transparent Dialog Box ////////////////////////
// Dialog name=***(T r ansparent ratio (0:Transparent - 255:Not Transparent)
// Montage map dialog in Aquisition
AcqMontageMapDlg=255
// Aquisition Patient Dialog
AcqPatientDlg=255
// Auto record dialog
AUT ODLG=255
// Pattern Table
PatternTableDlg=255
// Dsa Dialog in Review
RevDsaDlg=180
// Montage map dialog in Review
RevMontageMapDlg=255
// Review Patient Dialog
RevP atientDlg=255
// Not storaged (Blink) dialog
StorageDlg=180
[BGFFT]
////////////////////// Background (online) FFT /////////////////////
// Cal Code using in FFT =CAL Code Number (5=50uV)
CAL=5
// HF Code using in FFT = HF Code Number(2=30Hz)
HF=2
// Maximum FFT channels
MaxNumOfBGFftCh=32
// TC Code using in FFT = TC Code Number(5=0.3sec)
TC=5
2.14Service Manual EEG-9100/9200
2. CHANGING SETTINGS
Changing the MO User Label When Installing Two or More
Instruments in an Area or Connecting the Instrument to a Network
General
Procedure
When installing two or more EEG instruments in an area or connecting the
instrument to a network with other NK digital EEGs, assign the different MO disk
volume number for each instrument. The default “MO User Label” in the MO disk
volume number is “A”. Refer to “Writing Down the Data in the Volume Set Dialog
Box Before PC Unit or Hard Disk Replacement” in Section 6.
1. Click the Start button on the taskbar. The Satrt menu opens.
2. Click Run. The Run dialog box opens.
3. Type in C:\NFX11\VINST.EXE in the Open text box and click the OK button.
The Volume Set dialog box opens.
4. Type in the MO disk volume number with a capital letter from “A” to “Z” in
the MO USER LABEL text box.
5. Click the OK button.
To cancel the change, click the Cancel button.
Service Manual EEG-9100/92002.15
2. CHANGING SETTINGS
This page is intentionally left blank.
2.16Service Manual EEG-9100/9200
Section 3Troubleshooting and
Error Messages
How to Troubleshoot .......................................................................................................... 3.1
Closing the Program and Shutting Down Windows ....................................... 3.2
Acquisition Program/Review Program ................................................................... 3.14
File Utility Program ................................................................................................ 3.18
System Program .................................................................................................... 3.19
Open File Dialog Box ............................................................................................. 3.20
Service Manual EEG-9100/92003C.1
This page is intentionally left blank.
3C.2Service Manual EEG-9100/9200
How to Troubleshoot
3. TROUBLESHOOTING AND ERROR MESSAGES
Use this section to locate, identify and solve a problem in the instrument or an error
message displayed on the screen. The troubleshooting tables in this section are
divided into general problems and displayed error messages.
1. Determine which troubleshooting table to use.
2. In the “Problem” or “Error Message” column, find the trouble item that
matches the problem or error message.
3. Do the action recommended in the “Action” column. (Do the first action
recommended in the “Action” column).
4. If the problem or error message is not solved, do the next action recommended
in the “Action” column. (If this does not solve the problem, do the next
recommended sections.)
5. If none of the actions solve the problem, contact your Nihon Kohden
distributor or representative.
WARNING
When checking a cable connection, close the EEG-9000 application
program, turn off the power of the PC unit and all components, and
unplug the AC power cord from the AC outlet (For the procedure,
refer to the next page). Failure to follow this warning may cause
electrical shock.
After checking the cable connection, turn on the power of all components, then
restart the PC unit.
NOTE
Before contacting your NK distributor or representative for technical
support, please complete a copy of the Maintenance Check Sheet
(the original copy is provided at the end of the Section 6
“Maintenance”), and if possible, provide additional detailed
information on the problem. Send the complete copy of the
Maintenance Check Sheet to your NK distributor or representative.
This will allow your NK distributor or representative to provide you
with the best support.
Service Manual EEG-9100/92003.1
3. TROUBLESHOOTING AND ERROR MESSAGES
Closing the Program and Shutting Down Windows
1. Open the Windows Task Manager.
When the mouse does not operate:
1) Press the Ctrl + Alt + Del key. The Windows Security dialog box opens.
2) Select Task Manager to open the Windows Task Manager dialog box.
When the keyboard does not operate:
1) Right-click the task bar. The pop-up menu opens.
2) Select Task Manager to open the Windows Task Manager dialog box.
2. Select the program to close.
3. Select the End Task button.
4. Shut down Windows.
1) Select Shut Down from the Start menu. Or, press the Ctrl + Esc key,
then press the U key. The Shut Down Windows dialog box opens.
2) Select “Shut Down” in the “What do you want the computer to do ?”
list box.
3) EEG-9100:
Click the OK button. The PC unit is automatically turned off.
EEG-9200:
Click the OK button. When the “It is now safe to turn off your
computer” message appears, press the powr switch of the PC unit to
turn off the PC unit.
3.2Service Manual EEG-9100/9200
3. TROUBLESHOOTING AND ERROR MESSAGES
Troubleshooting
Waveform AcquisitionNOTE
Before measurement, confirm the following:
• The examination room is free from artifact-causing sources.
• The electrodes are firmly attached to the patient.
• The electrodes and electrode leads are not dirty, damaged or
frayed.
If not, pulse noise caused by static electricity or generated by a
display may be superimposed on the EEG waveform. Refer to
“Instrument Location” in Section 2 of the EEG-9100/9200 Operator’s
manual.
ProblemPossible causeAction
Noise or artifact is superimposed on the
waveforms.
The electrode lead is faulty.Check the continuity of the electrode
lead with a multimeter. If the
electrode lead is faulty, replace it with a
new one.
One or more of the leads from the Z,
C3 and C4 input jacks are not attached
to the patient.
Attach these leads to the patient
because the Z electrode and C3 and C4
electrodes are necessary for EEG
measurement.
The bed is not grounded.If the bed is metal, ground it.
The instrument is not grounded.If the AC outlet on the wall does not
have a ground terminal, ground the
instrument with the provided ground
lead.
Several medical electronic instruments
are used together.
There is an AC outlet or table tap near
the patient or bed.
The PC unit or printer is placed near
the patient or electrode junction box.
A desk lamp or fluorescent light is
turned on.
The patient touched some metal part.Prevent the patient from touching metal
The patient is using an electric blanket.Turn the electric blanket off and unplug
There is a cellular phone near the
patient.
Perform equipotential grounding for
each instrument.
Arrange the measurement environment
so that there is no influence from an AC
power line.
Arrange the measurement environment
so that unwanted radio frequency does
not affect the measurement.
Turn the desk lamp or fluorescent light
off.
parts.
the AC power cord, then use another
warming method.
Turn the cellular phone off.
Service Manual EEG-9100/92003.3
3. TROUBLESHOOTING AND ERROR MESSAGES
ProblemPossible causeAction
The waveform is not stable.One or more of the leads from the Z,
C3 and C4 input jacks are not attached
to the patient.
New and old electrodes or different
types of electrodes are used together.
The waveform sometimes becomes flat.The skin electrode contact impedance
of the C3 or C4 is high.
During waveform acquisition the
following message appears.
“The disk is full. Close the current file
to exit the acquisition program. Insert a
new disk.
[CAUTION]
Do not turn off the power of the main
unit or connected instruments (Photo
Drive unit or MO drive).
This can cause loss of EEG data and
damage to the hard disk.”
The waveform is not displayed.The color of the waveform and
The waveforms do not sweep smoothly.Many channels are displayed.
The AC filter does not function.The AC filter setting is not correct.Select the correct AC filter setting (50
Noise in AV derivation.An unused electrode for AV derivation
The electrode name on the screen is
indicated in red.
The storage drive (the MO disk or hard
disk which saves the EEG data file) is
almost full.
background is the same.
The Display setting in the Pattern table
is set to "OFF".
The FFT analysis function is set to on.
Noise is not caused by AC line
influence.
is selected in the AV Delete dialog box.
The electrode that is used for
measurement is not selected for the
storage electrode.
Attach these leads to the patient
because the Z electrode and C3 and C4
electrodes are necessary for EEG
measurement.
Do not use new and old electrodes or
different types of electrodes together.
This may cause high polarization
voltage.
Clean the electrode attachment to
reduce the impedance, and reattach the
electrode.
Click the OK button on the message
dialog box, then end the measurement
and save the file. Refer to "Starting
and Ending EEG Measurement -
Ending the Measurement and Saving
the File" in Section 5 of the Operator’s
manual. After saving the file, prepare
a new MO disk, or delete unnecessary
files in the hard disk.
Use a different color for waveform and
background.
Set the Display setting to "ON" for
necessary channels.
The PC unit cannot process all running
programs. Reduce the channels to
display or set the FFT analysis function
to off.
or 60 Hz) in the System program.
Use the proper filter according to the
artifact.
Delete unnecessary electrodes for AV
derivation in the AV Delete dialog box.
Select the electrode for the storage
electrode in the Electrodes to be Saved
dialog box of the System program.
The electrode is selected for the AV
derivation but not selected for the
storage electrode.
Select all electrodes which are selected
for the AV derivation for the storage
electrode in the Electrodes to be Saved
dialog box of the System program.
3.4Service Manual EEG-9100/9200
ProblemPossible causeAction
The amplifier setting (sensitivity, time
constant or high-cut filter) does not
change with the Amp bar.
The Acquisition program does not
open.
No calibration waveform appears on
the screen.
Artifact is superimposed on all
calibration waveforms.
Only baselines appear on the screen.
No waveform appears on the screen.
Artifact is superimposed on all signals
input from the electrode jacks.
A specific electrode signal does not
appear.
Artifact is superimposed on a specific
electrode signal.
No DC input signal appears on the
screen.
Artifact is superimposed on all DC
input signals.
A specific DC input signal does not
appear on the screen.
Artifact is superimposed on a specific
DC input signal.
3. TROUBLESHOOTING AND ERROR MESSAGES
The amplifier setting is not set to
"ACC".
---When an error appears, follow the
Faulty USB cable connection.Turn off the power of the PC unit, then
Faulty electrode junction box.Replace the electrode junction box.
Faulty EEG MOTHER board.Replace the EEG MOTHER board or
Faulty EEG MOTHER board if no
SELECTOR (Impedance threshold
display) LED on the electrode junction
box lights.
Faulty EEG AMP board.Replace the EEG AMP board.
Faulty electrode lead.Replace the electrode lead.
Faulty EEG AMP board. An amplifier
corresponding to the signal is faulty.
Faulty electrode lead.Replace the electrode lead.
Faulty electrode jack on the electrode
junction box.
Faulty EEG AMP board. An amplifier
corresponding to the signal is faulty.
Faulty EEG AMP board.Replace the EEG AMP board or
Faulty EEG AMP board. An amplifier
corresponding to the signal is faulty.
Set the amplifier setting to ACC with
the pattern table in the System
program. You can temporarily change
the amplifier setting in the Acquisition
program and Review program.
instructions on the dialog box.
check that the USB cable from the
electrode junction box is correctly
connected to the PC unit.
electrode junction box.
Replace the EEG MOTHER board or
electrode junction box.
Replace the EEG AMP board or
electrode junction box.
Replace the EEG INPUT board or
electrode junction box.
Replace the EEG AMP board or
electrode junction box.
electrode junction box.
Replace the EEG AMP board.
Service Manual EEG-9100/92003.5
3. TROUBLESHOOTING AND ERROR MESSAGES
Skin-electrode Impedance Check
ProblemPossible causeAction
The skin-electrode impedance check
result does not appear.
The skin-electrode impedance check
result is faulty.
A specific impedance display LED on
the electrode junction box does not
light.
Faulty IMPEDANCE CHECK keyReplace the EEG INPUT board or
Faulty EEG MOTHER board if no
SELECTOR (Impedance threshold
display) LED on the electrode junction
box lights.
One or more of the following
electrodes that are used for the
impedance check are not attached to the
patient: Z, A1 and A2 (or Fp1 and Fp2),
C3 and C4.
Faulty EEG INPUT board. An
electrode jack corresponding to the
faulty impedance check result is faulty.
Faulty EEG AMP board. An amplifier
corresponding to the signal is faulty.
Faulty EEG MOTHER board.Replace the EEG MOTHER board or
Faulty LED.Replace the EEG INPUT board.
electrode junction box.
Replace the EEG MOTHER board or
electrode junction box.
Attach these electrodes to the patients
firmly.
Replace the EEG INPUT board or
electrode junction box.
Replace the EEG AMP board or
electrode junction box.
electrode junction box.
3.6Service Manual EEG-9100/9200
3. TROUBLESHOOTING AND ERROR MESSAGES
Control
ProblemPossible causeAction
When the power is turned on, the
instrument does not start but the BIOS
setup screen is displayed on the screen.
When the power is turned on, nothing
is displayed on the screen. (EEG-9100)
When the power is turned on, nothing is
displayed on the screen (EEG-9200)
When the power is turned on,
Windows does not start.
The screen is dark.The brightness or contrast is not
The settings in the BIOS setup screen
are not correct.
The AC adapter is not correctly
connected to the power supply unit and
PC unit, and the rechargeable battery of
the PC unit is discharged.
Faulty PC unit.Replace the PC unit.
The power of the display is not turned
on.
The brightness or contrast of the display
is not appropriate.
The display cable is not connected to the
PC unit correctly.
The input line (BNC/D-Sub) is not
correct.
Dell Optiplex GX240 SMT:
The display cable from the CRT display
is not connected to the video connector
on the QI-111A Camera Interface Board
or display cable of the A/V input cable
from the Camera Interface Board is not
connected to the 15 pin video connector
on the PC unit.
A floppy disk is inserted into the floppy
disk drive.
Faulty electrode junction box.1. Turn the power of the PC unit off.
The settings in the BIOS setup screen
are not correct.
Problem with the Windows operating
system.
Faulty PC unit.Replace the PC unit.
appropriate.
Check and correct the settings in the
BIOS setup screen according to “BIOS
Default Setting” in Section 6.
Connect the AC adapter to the power
unit and PC unit correctly.
Press the power switch of the display to
turn the display on.
Adjust the brightness or contrast.
Refer to the Operator's manual of the
display and PC unit.
Connect the display cable to the video
connector on the PC unit correctly.
Select the correct input line. Refer to
the Operator's manual of the display.
Connect the display cable from the CRT
display and the display cable of the A/V
cable correctly.
Remove the floppy disk.
2. Remove the USB cable from the PC
unit.
3. Restart the PC unit.
If Windows starts correctly, the
electrode junction box is faulty.
Replace the electrode junction box.
Check and correct the settings in the
BIOS setup screen according to “BIOS
Default Setting” in Section 6.
Reinstall Windows 2000 and EEG-
9000 system program.
Adjust the brightness or contrast.
Refer to the Operator’s manual of the
display or PC unit.
Service Manual EEG-9100/92003.7
3. TROUBLESHOOTING AND ERROR MESSAGES
ProblemPossible causeAction
The EEG-9000 application program
does not work correctly.
The mouse does not function (EEG-
9100.
The keyboard does not function (EEG-
9100).
The mouse or keyboard does not
function (EEG-9200).
A screen saver program is active.Close the screen saver program.
Another windows application program
is active.
The USB cable from the electrode
junction box is not connected to the PC
unit.
Problem with the EEG-9000 application
program.
Faulty PC unit.Replace the PC unit.
The mouse is not connected to the PC
unit.
Program malfunction.Use the Task Manager to close the
Faulty mouse.Replace the mouse.
Temporary error.Turn off the power of the PC unit, then
The active window is behind an inactive
window. The title bar of the active
window is highlighted.
Program malfunction.Use the Task Manager to close the
Faulty PC unit.Replace the PC unit.
The mouse is not connected to the PC
unit.
The mouse is connected to the keyboard
connector.
The keyboard is not connected to the PC
unit.
The keyboard is connected to the mouse
connector.
The displayed window is inactive. The
title bar of the window is dimmed.
Faulty mouse.
Faulty keyboard.
Program malfunction.
Close all Windows application
programs. Or, delete the application
program if it conflicts with the EEG-
9000 system program.
Turn off the power of the PC unit, then
connect the USB cable to the PC unit
correctly.
Reinstall the EEG-9000 system
program.
Turn off the power of the PC unit, then
connect the mouse cable to mouse
connector on the PC unit correctly.
program, then shut down Windows.
Refer to “Closing the Program and
Shutting Down Windows” in Page 3.2
on again.
When the AC power cord of the PC
unit is disconnected and the PC unit
operates on battery power, this trouble
occurs. Click the title bar of the
window so that the window is active.
program or shut down Windows.
Refer to “Closing the Program and
Shutting Down Windows” in Page 3.2
Turn the power off, then connect the
mouse cable to mouse connector on the
PC unit correctly.
Turn the power off, then connect the
keyboard cable to keyboard connector
on the PC unit correctly.
Click the title bar of the window so that
the window is active.
Use the Task Manager to close the
program or shut down Windows. Refer
to “Closing the Program and Shutting
Down Windows” in Page 3.2
3.8Service Manual EEG-9100/9200
3. TROUBLESHOOTING AND ERROR MESSAGES
ProblemPossible causeAction
The manual mark (MARK ON,
MARK OFF) does not appear
The layout of the window is changed.The size of the window is changed.
The date and time is not correct.The date and time setting is not correct.Set the correct date and time. Refer to
Faulty mark cord connection.Connect the mark cord to the electrode
junction box correctly
Faulty mark cord.Replace the mark cord.
Faulty EEG MOTHER board.Replace the EEG MOTHER board.
Change the size of the window
When the size of the window is
changed, the position of the button
changes.
The backup battery of the PC unit is
discharged.
properly.
the Windows online help.
Contact your NK distributor or
representative.
Service Manual EEG-9100/92003.9
3. TROUBLESHOOTING AND ERROR MESSAGES
Activation
ProblemPossible causeAction
The flash lamp does not light.The AC power cord is not connected to
the photo control unit correctly.
Faulty flash lamp assembly cable
connection.
Faulty RS-232C cable connectionCheck that the RS-232C cable is
The power of the photo control unit is
turned on after the Acquisition program
opens.
The “Use photic stimulation" check
box on the Photic Stimulation dialog
box (System program) is not checked.
The photic stimulation mode is set to
“Single”.
The flash lamp is faulty.Replace the flash lamp.
The photic stimulation mark or HV
mark does not appear.
The photo mark connection cable is not
connected to the electrode junction box
and the photo control unit correctly.
Faulty photo mark connection cable.Replace the photo mark connection
The flash lamp is faulty.Replace the flash lamp.
Faulty EEG MOTHER boardReplace the EEG MOTHER board.
Faulty PHOTO STIM board.Replace the PHOTO STIM board.
Connect the AC power cord correctly.
Check that the flash lamp assembly
cable is correctly connected to the
PHOTIC LAMP connector on the
photo control unit.
correctly connected to the photo control
unit and PC unit.
1. Close the Acquisition program.
2. Turn on the photo control unit
power.
3. Open the Acquisition program.
Check the “Use stimulation” check
box.
Select the correct photic stimulation
mode.
Turn off the power of the PC unit and
photo control unit, then check that the
photo mark connection cable is
connected to the electrode junction box
and the photo control unit correctly.
cable.
3.10Service Manual EEG-9100/9200
3. TROUBLESHOOTING AND ERROR MESSAGES
Power
ProblemPossible causeAction
The PC unit operates on battery power
(EEG-9100 Only).
The MO disk drive does not operate.
The printer does not operate.
The flash lamp does not light.
The photo control unit is not
recognized (The Display Activation
Control button on the tool bar is
dimmed).
No pacing sound (SPEAKER
OUTPUT) occurs.
The photo control unit does not
operate.
The flash lamp does not light.Faulty +512 V on the PHOTO STIM
Faulty isolation AC power on the
power supply unit.
Faulty +5 VD and/or +12 VA on the
PHOTO STIM board.
Faulty –12 VA on the PHOTO STIM
board.
Faulty +5 VD and/or +12 VA on the
PHOTO STIM board.
Faulty isolation AC power on the photo
control unit.
board.
Check that the power fuse is not blown.
If the fuse is not blown, replace the
power supply unit.
Check that the F012 fuse is not blown.
If the fuse is not blown, replace the
PHOTO STIM board.
Check that the F013 fuse is not blown.
If the fuse is not blown, replace the
PHOTO STIM board.
Check that the F012 fuse is not blown.
If the fuse is not blown, replace the
PHOTO STIM board.
Check that the power fuse is not blown.
If the fuse is not blown, replace the
photo control unit.
Check that the F011 and/or F014 is not
blown. If the fuse is not blown, replace
the PHOTO STIM board.
* The +5 VD power is generated from the +12 VA.
PrinterThe following table explains general troubleshooting used with the
instrument. For details, refer to the printer Operator’s manual.
ProblemPossible CauseAction
The printer does not operate.The printer power cord is not connected
to the power supply unit.
The printer cable is not connected to
the PC unit or printer.
The printer does not print correctly.The printer driver is not installed or an
incorrect printer driver is installed.
An error message appears on the
printer.
The size of the recording paper does
not match the size of the printing data.
Malfunction occurs during printingThe printer power cord is connected to
the 3 prong outlet on the rear panel of
the main unit.
Both the instrument and isolation
transformer for the printer power cord
are connected to the same AC outlet.
Connect the AC power cord correctly
Connect the printer cable correctly.
Install the correct printer driver.
Refer to the Operator's manual of the
printer.
Match the size of the recording paper
and printing data.
Supply the printer power from a
medical isolation transformer.
Use a different outlet for the instrument
and transformer because the laser
printer consumes a lot of AC power.
Service Manual EEG-9100/92003.11
3. TROUBLESHOOTING AND ERROR MESSAGES
MO Disk DriveThe following table explains the general troubleshooting in combination with
the instrument. For details, refer to the MO disk drive Operator’s manual.
ProblemPossible CauseAction
Cannot assign a volume number to the
MO disk.
Cannot read/write a file from/to the
MO disk.
The MO disk cannot be removed by
pressing the eject switch on the MO
disk drive.
The MO disk is not a type specified by
Nihon Kohden.
The SCSI device ID number of the MO
disk drive is not set to “4”.
The MO disk is write-protected.Release the write-protect and try again.
The MO disk is not formatted.Format the MO disk and try again.
The MO disk is not specified by Nihon
Kohden.
The SCSI device ID number of the MO
disk drive is not set to “4”.
The MO disk drive is not recognized by
the instrument because the MO disk
drive is not turned on or it was turned
on after the PC unit was turned on.
The MO disk drive is not specified by
Nihon Kohden.
The same SCSI device ID number is
used for more than one SCSI devices.
A terminator is not connected to the
end of the SCSI device.
The MO disk is write-protected.Release the write-protect and try again.
The MO disk is not formatted.Format the MO disk and assign the
The MO drive is not selected as the
storage drive.
The SCSI cable is not connected to the
PC unit or MO disk drive.
The MO disk is faulty.
The volume number is not assigned to
the MO disk.
The Enable library check box
(Computer Management window
Optional Device Properties → General
page) is checked.
→
Use only the specified MO disk type
(512 bytes/sector).
Set the SCSI device ID number to “4”
and turn the instrument on again.
Use only the specified MO disk (512
bytes/sector).
Set the SCSI device ID number to “4”
and turn the instrument on again.
Check that the MO disk drive is
recognized as a Removable Drive in the
My Computer. If not, turn the power
switch of the MO disk drive on and
turn the PC unit on again.
Use only the specified MO disk drive.
Set a different SCSI device ID number
for each SCSI device. Refer to the
Operator's manual of each device.
Connect the terminator to the end of the
SCSI device.
volume number.
Select the MO disk drive as the storage
drive in the System Setting dialog box
of the System program.
Connect the SCSI cable correctly then
turn the power on again.
Use the Check Disk function to recover
the MO disk.
Clean the MO disk. Refer to the
Operator's manual of the MO disk
drive.
Assign the MO disk. Refer to
“Formatting a MO Disk” in Section 3
of the Operator's manual.
Uncheck the Enable library check box.
Refer to “Setting the Properties for 5
inch Magneto-optical Disk” in Section
of the Operator's manual.
3.12Service Manual EEG-9100/9200
3. TROUBLESHOOTING AND ERROR MESSAGES
CD-RW Drive (EEG-9200 Only)
ProblemPossible CauseAction
Cannot assign a volume number to the
CD-R/CD-RW disk.
Cannot read/write a file from/to the
CD-R/CD-RW disk.
The following table explains the general troubleshooting. For details, refer to
PC unit Operator’s manual and CD-R/CD-RW recorder software online help..
The CD-R/CD-RW disk is not
formatted as a Direct CD.
The CD-R/CD-RW disk is not
formatted.
The volume number is not assigned to
the CD-R/CD-RW disk.
The CD-R/CD-RW disk is faulty.
The CD-RW drive is not recognized by
the Windows 2000 operating system.
Format the CD-R/CD-RW disk with the
Direct CD format utility and try again.
Refer to “Formatting an Magneto-
optical Disk, CD-R Disk and CD-RW
Disk” in Section 3 of the Operator’s
manual.
Format the CD-R/CD-RW disk with the
Direct CD format utility and assign the
volume number.
Assign the volume number to the CD-
R/CD-RW disk.
Use the Scan Disk function of the
Direct CD utility to recover the CD-
R/CD-RW disk.
Clean the CD-R/CD-RW disk.
Check that the CD-RW drive is
recognized as a Compact Disk in the
My Computer window. If not, the PC
unit is faulty. Replace the PC unit.
Service Manual EEG-9100/92003.13
3. TROUBLESHOOTING AND ERROR MESSAGES
Error Messages
Following is a list of error messages during operation or the power on self check
for the PC unit. To solve the problem if an error message appears, find the
displayed error message from the table, then do the action recommended in the
Action column.
Acquisition Program/Review Program
Error MessagePossible CauseAction
A communication error in the Photo
Drive unit.
Please restart the unit.
Acquisition or Review program is
open.
Please close it and start again.
An error was detected by the diagnostic
program of the Photo Drive unit.
Please restart the unit.
An error was detected during operation
of the Photo Drive unit.
Confirm that the unit is connected and
the power is on.
Another file with the same name was
found in this disk.
A new file name has been set.
There was a possibility of abnormal
exit last time.
To be on the safe side, please check
your disk and system with
SCANDISK.EXE.
Another file with the same name was
found in this disk.
Overwrite the previous file?
Another software of this EEG system is
working.
Please close it and start again.
Can open the file no more.You tried to open an EEG data file
Cannot access the patient register.The database file cannot be opened.Open the Acquisition program to
Cannot access the selected file.The drive which saves EEG data file of
Faulty communication between the
photo control unit and PC unit
You tried to open another EEG-9000
application program when the
Acquisition or Review program was
open. Only one EEG-9000 application
program can be open at a time.
Faulty RS-232C cable connection.Check that the RS-232C cable is
Faulty PHOTO STIM boardReplace the PHOTO STIM board.
Faulty communication between the
photo control unit and PC unit
The Acquisition program did not close
correctly and the registry settings are
not correct.
You tried to save the ASCII file with a
file name that already exists in the
folder.
You tried to open an EEG-9000
application program when another
EEG-9000 application program was
open. Only one EEG-9000
application program can be open at a
time.
when four EEG data files were open.
the selected patient has a malfunction.
Check that the RS-232C cable is
correctly connected to the photo control
unit and PC unit, then open the
Acquisition program. If the same
error message appears again, the
PHOTO STIM board is faulty.
Replace the PHOTO STIM board.
Close the Acquisition or Review
program, then open the required
program.
correctly connected to the photo control
unit and PC unit, then open the
Acquisition program.
Check that the RS-232C cable is
correctly connected to the photo control
unit and PC unit, then open the
Acquisition program.
Check for disk error with the Check
Disk function. Refer to “Checking for
Disk Damage Using Check Disk” in
Section 3.
Use a different file name.
Close the EEG-9000 application
program, then open the required
program.
No more than four EEG data files can
be open at the same time.
recreate the database file.
Check the drive with the Windows
Explore or Check Disk function.
3.14Service Manual EEG-9100/9200
3. TROUBLESHOOTING AND ERROR MESSAGES
Error MessagePossible CauseAction
Cannot read the storage drive.
Please verify that the drive door is
closed and that the disk is formatted
and free of errors.
Cannot revise the multiple events at the
same time.
Could not open the Communication
port (#%d) for the Photo Drive unit.
Confirm the COM port and the
E11CFG.INI file setting.
Date or Date of Birth was not entered
correctly.
Please re-enter it using the date format
in the Windows Control Panel.
<<DSP017>> Error found while
checking waveform data buffer.
Error found while saving data.
Check the drive or media.
Event was not created correctly.The system resources are low.Restart the PC unit. If necessary add
The %d channel is used for ECG
trigger.
Cannot set this channel to Off.
The IMPEDANCE CHECK key on the
electrode junction box has been
pressed.
The Photo Drive unit is not connected
or the power is off.
The SELECTOR key on the electrode
junction box was pressed during
startup.
The MO disk is write-protected.Find the trouble source and correct it,
referring to “Possible Cause” and check
that a file can be saved in the MO disk
with Windows Explore.
The SCSI cable is not correctly
connected to the PC unit and MO disk
drive.
The SCSI ID of the MO disk is not
correct.
The terminator setting of the MO disk
drive is not correct.
Two or more events are selected when
changing the event name.
The RS-232C port of the photo control
unit is not recognized.
You tried to enter the “Date” or “Date
of Birth” in an incorrect format.
The system resources are low.Restart the PC unit. If necessary add
Faulty storage drive (the MO disk or
hard disk which saves the EEG data
file).
You tried to set the “Display” setting of
the channel that is set to the trigger
channel for the ECG filer to “OFF”.
When the power is turned on, the
IMPEDANCE CHECK key on the
electrode junction box is pressed, or the
IMPEDANCE CHECK key makes a
short-circuit.
Faulty RS-232C cable connection.Check that the RS-232C cable is
The photo control unit is not turned on.
Faulty RS-232C cableReplace the RS-232C cable.
Faulty PHOTO STIM boardReplace the PHOTO STIM board.
Faulty photo control unitReplace the photo control unit.
When the power is turned on, the
SELECTOR key on the electrode
junction box is pressed, or the
SELECTOR key makes a short-circuit.
Change the event name one by one.
Close the Acquisition program, then
check the description of the
[PhotoControlComm] section in the
E11CFG.INI file
Check the format of the Date and Date
of Birth in the Windows Control Panel.
system memory.
Check for disk error with the Check
Disk function. Refer to “Checking for
Disk Damage Using Check Disk” in
Section 3.
system memory.
Do not change the “Display” setting to
“OFF” when the channel is set to the
trigger channel for the ECG filter.
Do not press the IMPEDANCE
CHECK key when the power is turned
on. If the IMPEDANCE CHECK key
is faulty, replace the EEG INPUT
board.
correctly connected to the photo control
unit and PC unit, then open the
Acquisition program.
Close the Acquisition program.
1.
Turn the photo control unit on.
2.
Open the Acquisition program.
3.
If the same error message appears
again, check the fuses on the photo
control unit.
Do not press the SELECTOR key when
the power is turned on. If the
SELECTOR key is faulty, replace the
EEG INPUT board.
Service Manual EEG-9100/92003.15
3. TROUBLESHOOTING AND ERROR MESSAGES
Error MessagePossible CauseAction
The disk is full.
Close the current file to exit the
acquisition program. Insert a new disk.
[CAUTION]
Do not turn off the power of the main
unit or connected instruments (Photo
Drive unit or MO drive).
This can cause loss of EEG data and
damage to the hard disk.
The file has not been saved yet.
Save the file and open a new file?
If you select [No], the data will be lost.
The file has not been saved yet.
Save the file?
The data will be lost when you select
[NO].
The file is being saved.
Please wait for moment.
The following character cannot be
entered.
| '
The instrument is running on battery
power.
Connect AC power.
The instrument temporarily running on
battery power.
Connect AC power.
The maximum number of events have
been registered.
Cannot add more.
The number of information in the
patient register is over the maximum
number.
Save the current register to a file and
delete the register by using the
Database Manager.
The printer driver is not installed.
Please install the printer driver.
The specified Time has no EEG data.
Type a new time "hhhh:mm:ss".
The specified epoch number is not
correct.
Please input the epoch number again.
The storage drive (the MO disk or hard
disk which saves the EEG data file) is
almost full.
From the File menu, New is selected
after the Starts/Stop Filing button is
clicked.
You tried to close the Acquisition
program without saving the EEG data
file.
The EEG data file is being saved.No operation can be done until this
You tried to enter the character “ ’ ” or
“|”.
The AC adapter is disconnected from
the PC unit.
The AC power cord of the AC adapter
is disconnected.
The number of saved events has
reached 5000.
The maximum number of registered
patient information in the system
database has reached.
The printer driver is not installed.Install the correct printer driver.
There is no waveform data
corresponding to the specified time to
jump to.
There is no epoch data corresponding
to the specified epoch number.
Click the OK button on the message
dialog box, then end the measurement
and save the file. Refer to "Starting
and Ending EEG Measurement Ending the Measurement and Saving
the File" in Section 5 of the Operator’s
manual. After saving the file, prepare
the new MO disk, or delete
unnecessary files in the hard disk.
Select the proper button on the dialog
box. Refer to “Start and Ending EEG
Measurement – Closing the File
without Saving” in Section 5 of the
Operator’s manual.
Click the appropriate button on the
message dialog box. Refer to "Starting
and Ending EEG Measurement –
Closing the File without saving" in
Section 5 of the Operator’s manual.
message disappears.
The characters “ ’ ” and “|” cannot be
entered.
1. Close the EEG-9000 application
program.
2. Turn off the power of the PC unit
and all components.
3. Connect the AC adapter to the PC
unit correctly.
4. Turn on the all components and
restart the PC unit.
1. Close the EEG-9000 application
program.
2. Turn off the power of the PC unit
and all components.
3. Connect the AC adapter to the PC
unit correctly.
4. Turn on all components and restart
the PC unit.
No more events can be added. Delete
unnecessary events.
No more patient information can be
added. Delete unnecessary patient
information.
Enter the correct time corresponding to
the waveform data.
Enter the correct epoch number
corresponding to the epoch data.
3.16Service Manual EEG-9100/9200
Error MessagePossible CauseAction
The starting option parameter is not
correct.
Please check the setting of the shortcut.
The value for Time is invalid.
Type a new time "hhhh:mm:ss".
There was a possibility of abnormal
exit last time.
To be on the safe side, check your disk
and system with SCANDISK.EXE.
<< USB 901 >> The electrode junction
box has been disconnected.
The Acquisition program will be closed
after saving the acquired data.
Please connect the electrode junction
box, and restart the program.
<< USB 902 >> No electrode junction
box was detected.
Confirm the electrode junction box is
connected and restart the Acquisition
program.
<< USB 903 >> The electrode junction
box is disconnected or there is a data
transmission error.
After the Acquisition program closes,
turn off the power,
confirm the electrode junction box is
connected and restart the Acquisition.
<< USB 904 >> An overflow occurred
during data transmission from the
electrode junction box.
Restart the Acquisition program
Volume name was not found on the
storage Drive.
New volume name %s was created.
Windows data format must be set to 4
digit year.
3. TROUBLESHOOTING AND ERROR MESSAGES
Invalid shortcut target or the target file
does not exist.
You tried to enter the time in an
incorrect format.
The EEG-9000 application program did
not close correctly.
Faulty USB cable connection.Check that the USB cable from the
The electrode junction box is not
recognized.
Faulty data communication between the
electrode junction box and PC unit.
Overflow occurs during data transfer.1. Close the Acquisition program.
No volume number is assigned to the
MO disk.
This message appears when a two digit
year format will be used from the year
2009.
Correct the shortcut target description
in the Properties of the shortcut icon, or
make a correct shortcut icon.
Enter the time in the “hhhh:mm:ss”
format.
Check the disk damage with the Check
Disk function. Refer to “Checking for
Disk Damage Using Check Disk” in
Section 3.
electrode junction box is correctly
connected to the PC unit, then open the
Acquisition program.
Check that the USB cable from the
electrode junction box is correctly
connected to the PC unit, then open the
Acquisition program. If the same error
message appears again, reinstall the
electrode junction box driver, or replace
the electrode junction box.
1. Close the Acquisition program.
2. Turn the PC unit off.
3. Check that the USB cable is
correctly connected to the PC unit.
4. Restart the PC unit.
5. Open the Acquisition program.
2. Turn the PC unit off.
3. Check that the USB cable is
correctly connected to the PC unit.
4. Restart the PC unit.
5. Open the Acquisition program.
The volume number is automatically
assigned.
Change the year format to 4 digits
(Control panel → Regional Options
Data page → Short date area → Short
date format).
Service Manual EEG-9100/92003.17
3. TROUBLESHOOTING AND ERROR MESSAGES
File Utility Program
Error MessagePossible CauseAction
Another program of this EEG system is
working.
Please close it and start again.
Cannot change file name.You tried to change the file name when
Cannot open database file.The database file is faulty.Reinstall the EEG-9000 system
Local Drive [***] has not been
initialized.
Please execute initialize.
Not enough disk space on the
destination drive to copy your file.
Not enough disk space on the
destination drive to move your file.
Select a different drive.You tried to copy the file to the same
The disk is write protected.The MO disk is write-protected.Release the write-protect.
The selected drive has not been
formatted.
This drive has already been initialized.
Initialize canceled.
***.ini was not found.
Close File Utility.
You tried to open the File Utility
program when another EEG-9000
application program was open.
copying the file to a folder.
No volume number is assigned.Check the disk, then assign the volume
There is not enough free disk space to
copy the file.
There is not enough free disk space to
move the file.
drive.
You tried to move the file to the same
drive.
The MO disk drive that you want to
copy a file to is write-protected.
The MO disk that you want to move a
file to is write-protected.
The disk is not formatted.Format the disk. Refer to “Formatting
You tried to assign a volume number to
a disk or folder that already has a
volume number.
The initial file for the File Utility
program is lost.
Close the other EEG-9000 application
program, then open the File Utility
program.
The file name cannot be changed when
copying the file to a folder.
program.
number if necessary. Refer to
“Formatting a Magneto-optical Disk Assigning the Volume Number to a
Magneto-optical Disk” in Section 3 of
the Operator’s manual.
Prepare a new MO disk, or delete
unnecessary files in the hard disk
Prepare a new MO disk, or delete
unnecessary files in the hard disk
The file can only be copied to another
drive.
The file can only be moved to another
drive.
a Magneto-optical Disk - Assigning the
Volume Number to a Magneto-optical
Disk” in Section 3 of the Operator’s
manual.
Check the volume number, then assign
a new volume number if necessary.
Refer to “Formatting a Magneto-optical
Disk - Assigning the Volume Number
to a Magneto-optical Disk” in Section 3
of the Operator’s manual.
Reinstall the EEG-9000 system
program.
3.18Service Manual EEG-9100/9200
3. TROUBLESHOOTING AND ERROR MESSAGES
System Program
Error MessagePossible CauseAction
At least one electrode must be selected.No electrode is selected and saved for
waveform acquisition.
Drive "D" is CD-ROM drive.
Storage drive is set to "C".
Drive "D" not found.
Storage drive is set to "C".
Please close Acquisition program
before you start the System program.
Please close Database Manager
program before you start the System
program.
Please close File Utility program before
you start the System program.
Please close Review program before
you start the System program.
The storage folder and the automatic
copy folder
cannot be the same folder.
Time-out error (over 30 seconds).
The settings file may be damaged.
The MO disk drive is not recognized.Check that the SCSI cable is correctly
The MO disk drive is not recognized.Check that the SCSI cable is correctly
You tried to open the System program
when the Acquisition program was
open.
You tried to open the System program
when the Database Manager program
was open.
You tried to open the System program
when the File Utility program was
open.
You tried to open the System program
when the Review program was open.
The same directory is selected for the
“Storage drive” option and “Automatic
copy after acquisition” option.
The system setting file is faulty.Reinstall the EEG-9000 system
Select and save the proper electrodes.
Refer to “Selecting and Saving the
Electrodes for Waveform Acquisition”
in Section 4 of the Operator’s manual.
connected to the MO disk drive and PC
unit.
connected to the MO disk drive and PC
unit.
Close the Acquisition program, then
open the System program.
Close the Database Manager program,
then open the System program.
Close the File Utility program, then
open the System program.
Close the Review program, then open
the System program.
Change the directory for the
“Automatic copy after acquisition”
option, or uncheck the “Automatic
copy after acquisition” option.
program.
Service Manual EEG-9100/92003.19
3. TROUBLESHOOTING AND ERROR MESSAGES
Open File Dialog Box
Error MessagePossible CauseAction
429ActiveX Component can’t create
Object.
Cannot access the selected file.
Confirm the selected drive.
Could not create new database.
Exit application.
Device not ready.The selected drive is not recognized.Insert the correct MO disk into the MO
Drive is write-protected.
Open in read-only mode.
Dskvol21.vol not found in selected
drive.
File is write-protected.
Open in read-only mode.
File not found.The MO disk is not inserted into the
Initialize Error: could not write the
E11Launcher.ini file.
Exit application.
No Data.There is no EEG data file in the
No EEG folder in selected DriveThere is no folder to save EEG data
No volume number for Dskvol21.vol.No volume number is assigned.Check the disk, then assign a volume
Selected file is not on selected drive.Insert the correct MO disk into the MO
This file is not entered.Add the EEG data file to the system
Too many files selected.Four EEG data files have been opened.Close unnecessary EEG data files.
Faulty “Data Access Object (DAO)”
installation.
The MO disk that has the selected EEG
data file is not inserted into the MO
disk drive.
The disk is write-protected.Release the write-protect, if necessary.
No volume number is assigned.Check the disk, then assign a volume
MO disk drive.
The E11Launcer.ini file is not found.Check that the E11Launch.ini file exists
selected drive.
files
Install the DAO*, or install the EEG9000 system program.
Insert the correct MO disk into the MO
disk drive.
Reinstall the EEG-9000 system
program.
disk drive, or select the correct drive or
folder in the hard disk.
number if necessary. Refer to
“Formatting a Magneto-optical Disk Assigning the Volume Number to a
Magneto-optical Disk” in Section 3 of
the Operator’s manual.
If necessary, change the attribute of the
file to “Archive” (Tools menu
Change File Attributes).
Insert the correct MO disk, or select the
correct drive or folder in the hard disk.
in C:/NFX11. If it does not exist,
reinstall the EEG-9000 system
program.
Insert the correct MO disk into the MO
disk drive, or select the correct drive or
folder in the hard disk.
Assign a volume number. Refer to
“Formatting a Magneto-optical Disk Assigning the Volume Number to a
Magneto-optical Disk” in Section 3 of
the Operator’s manual.
number if necessary. Refer to
“Formatting a Magneto-optical Disk Assigning the Volume Number to a
Magneto-optical Disk” in Section 3 of
the Operator’s manual.
disk drive, or select the correct drive or
folder in the hard disk.
database. Refer to “Adding Files to the
System Database” in Section 8 of the
Operator’s manual.
→
* To install the DAO:
Double-clicking C:\nfx11\VXXXX\Dao35\Setup.exe.
The subdirectory name VXXXX differs depending on the software version.