NeuMoDx Molecular, Inc. CONFIDENTIAL P/N 40600195_Rev B
March 2019
Pg. 3 of 4 For Ex-US Distribution Only
2. If a valid calibration does not exist, the NeuMoDx System will prompt the user to process external calibrators (and external controls)
before sample results can be reported.
3. If calibrators are required, process the NeuMoDx HCV calibrators with the following barcodes (1 high calibrator and 1 low calibrator per
reagent lot):
4. Retrieve a set of NeuMoDx HCV external calibrators from freezer and allow the vials to set at room temperature (15-30°C) until
completely thawed. If using an already thawed set of calibrators, ensure that the thawed calibrators were stored at 4°C and are not
more than 24 hours old.
5. Vortex gently to ensure homogeneity.
6. Load the calibrator vials into a standard 32-Tube Carrier, and ensure caps are removed from all tubes.
7. Place the Tube Carrier on the Autoloader shelf and use the touchscreen to load carrier into the NeuMoDx System.
8. The NeuMoDx System will recognize the barcode and start processing the specimen tubes unless reagents or consumables required for
testing are not available.
9. To generate valid results, at least 2 out of the 3 replicates must yield results within predefined parameters. The low calibrator nominal
target is 3.0 Log10 IU/mL and the high calibrator nominal target is 5.0 Log10 IU/mL.
10. Discrepant result handling for external calibrators should be performed as follows:
a. If one or both the calibrators fails the validity check, repeat processing of the failed calibrator(s) using a new vial. In the event
one calibrator fails validity, it is possible to only repeat the failed calibrator as system does not require the user to run both
calibrators.
b. If problem persists, contact NeuMoDx Molecular, Inc.
11. External controls [REF 900201] must be processed after calibrator validity has been established, prior to obtaining test results from
human clinical samples.
LIMITATIONS
1. The NeuMoDx HCV Calibrators can only be used in conjunction with the NeuMoDx HCV Quant Test Strips on the NeuMoDx System.
2. A valid calibration of the NeuMoDx HCV Quant Test Strip using the NeuMoDx HCV Calibrators [REF 800200] is required before the NeuMoDx
HCV External Controls [REF 900201] can be processed.
3. Erroneous results could occur from improper handling, storage, or other technical error.
4. Operation of the NeuMoDx System is limited to use by personnel trained on the use of the NeuMoDx System.
REFERENCES
1. Biosafety in Microbiological and Biomedical Laboratories, 5th edition. HHS Publication No. (CDC) 21-1112, Revised December 2009
2. Clinical And Laboratory Standards Institute (CLSI). Protection of Laboratory Workers From Occupationally Acquired Infections; Approved
Guideline – Fourth Edition. CLSI document M29-A4; May 2014
TRADEMARKS
NeuMoDx™ is a trademark of NeuMoDx Molecular, Inc.
TaqMan® is a registered trademark of Roche Molecular Systems, Inc.