The NeuLog relative humidity sensor can be used for any science experiment or
activity which requires an accurate relative humidity reading such as in the fields of
Biology, Physiology, Exercise Science, Chemistry, Environmental Science, Earth
Science, Weather Science, Botany, Entomology, etc.
The sensor comes pre-calibrated so you can start experimentation right out of the box
using any of the following guides.
Just a few of the thousands of possible experiments that can be done with the NUL207 sensors are: perspiration, photosynthesis rates, insect behavior, water vapor
emitting chemical reactions, ecological studies, animal behavior, weather patterns,
and many more.
Relative humidity is a measurement which examines the ratio of water vapor present
in the air compared with the maximum amount of water vapor that the air can hold for
a specific temperature.
Data is presented as a percentage which is a ratio of the amount of water
vapor in the air versus the maximum amount that could be allowed at a
given temperature.
Quick start procedure:
PC or Mac Computer
Materials needed:
NUL-207 Relative Humidity Sensor
USB-200 USB Module
A USB to mini USB cable (which comes with the USB-200)
Your relative humidity sensor needs to be connected to a USB-200 module. The
USB-200 module then connects to a computer via a USB to mini-USB cable. Please
note that you cannot plug the relative humidity sensor directly into the computer.
Resident PC software and browser based software can be downloaded for free at
www.NeuLog.com/dowload as well as a full software user guide. Note: Make sure not
to download and install both types of software they will conflict on the computer.
Procedure:
1. Install the NeuLog software
2. Connect the USB-200 module to the PC or Mac
3. Connect the relative humidity sensor to the USB-200 module (they directly
plug together). Please note there is no calibration required for this sensor.
4. Open the NeuLog software.
5. Once a relative humidity sensor logo appears on the left side of the screen
the probe has been automatically identified and you can begin
experimentation.
6. If the relative humidity sensor is not automatically identified then click the
“Search for sensors” icon to find the sensor.
7. Select the “On-line experiment” button; this will open a graph below.
8. Click on the “Module setup” button located on the relative humidity sensor
icon in the module window to change the sensor settings if need be.
9. Click on the experiment set up button to change the experiment settings if
need be (experiment duration for example).
10. The relative humidity sensor will give a live reading in the box to the left of
the screen while plugged in.
11. To run an experiment and collect data click “Run experiment”.
12. To end data collection early, click “Stop experiment”.
Your relative humidity sensor needs to be connected to a WIFI-201 module. The WIFI201 module will create a closed NeuLog wifi network which will stream the NeuLog
data to a device of your choosing. Once your device is wirelessly connected to the
NeuLog network you can run experiments and collect data through a browser of your
choosing.
NEULOG RELATIVE HUMIDITY SENSOR GUIDE
Procedure:
1. Connect the relative humidity sensor directly to the left side of a WIFI-201
module (no wires required).
2. Connect a BAT-200 module to the right side of the WIFI-201 module.
3. Although not required, we recommend plugging the BAT-200 to an outlet
using a USB to micro USB charger (such as a typical cell phone charger).
The WIFI-201 module will run for 15-75 minutes (depending on the sensor)
without being plugged in.
4. For further WIFI-201 instructions or the WIFI-201 quick start guide please
visit www.NeuLog.com/download.
5. The WIFI-201 indicator lights will flash; take no action until the LED to the
far left turns blue.
6. Take your tablet or smart phone and go to the Wi-Fi settings and select the
NeuLog network which matches the WIFI module ID found on the back of
the WIFI-201 device. (NeuLogXXXX for example)
7. Give your device 1-2 minutes to connect to the WIFI-201 network.
8. Once the device is connected go to your browser and type in the website
wifi201.com into the URL bar, then wait for 30-60 seconds.
9. You will see a “Control mode” icon in the browser, click on this icon.
10. The browser will then load a new screen and begin to auto detect the
sensors, this can take a few minutes.
11. If the browser does not auto detect, select “Search for sensors”
12. Once the sensor is found you will see an icon on the left side of the screen
for the relative humidity sensor. The icon will display data in real time.
13. To set the relative humidity sensor’s settings click on “Module setup” button
located on the relative humidity sensor icon on the left side of your screen.
14. To change the experimental settings click the “Experiment setup” button
(experiment duration for example).
15. To run an experiment and collect data click “Run experiment”.
Materials needed to configure your offline experiment:
USB-200 USB Module or WIFI-201 WIFI module or VIEW-101 Viewing Module
A USB to mini USB cable (which comes with the USB-200)
Procedure:
1. Connect the relative humidity sensor directly to the left side of a charged BAT-200 module.
2. When ready to collect data press the “Start/Stop” button on the relative humidity sensor, a red light will turn on for the duration of the experiment.
NEULOG RELATIVE HUMIDITY SENSOR GUIDE
Sensor specifications
Relative humidity level
Range and operation modes
0 to 95% RH
ADC resolution
16 bit digital
Accuracy
± 5% RH
Resolution
0.1%
Max sample rate (S/sec)
100
3. To change the experiment settings first connect to your chosen device (PC,
Mac, tablet, smart device) and select the “Module setup” button. (For more
information on how to change the experiment settings view quick start
procedure section for your chosen device in this document.)
4. After the experiment has concluded (when the “Start/Stop” button is pressed
again or the data collection period ends) connect the sensor to your chosen
device. (full instructions in each device’s quick start procedure section)
5. Open NeuLog software.
6. Click the “Off-line experiment” button.
7. Click the “Load data from sensors” button.
8. Select which experimental data to upload. (5 experimental runs can be stored at one time on the NUL-207 relative humidity sensor)
Restoring sensor’s factory default settings:
Procedure:
1. Connect the USB-200 to a PC, Mac, tablet, or smart device.
2. Connect the relative humidity sensor to a USB-200 module (they directly
plug together). Please note there is no calibration required for this sensor.
Fully digital data.
Rugged plastic ergonomic case.
Push button switch for Start/Stop experiments in off line mode.
LED indicator of experiment status (blinks while collecting data).
Pre-calibrated sensing equipment.
Note: NeuLog products are intended for educational use.
Videos and experiment examples:
Videos, literature and other probes can be found at www.NeuLog.com
http://www.neulog.com/Portals/3/Users/Experiment B-9 Thermoregulation
and Perspiration Ver 3.2.4.pdf
Technical background:
The philosophy behind NeuLog’splug and play technology is based on each sensor’s
ability to store its own data due to an internal flash memory chip and micro-controller
in each plastic NeuLog body. This technology allows the sensor to collect and then
store the digital data in the correct scientific units (oC, oF, Lux, %, ppm, for example).
NEULOG RELATIVE HUMIDITY SENSOR GUIDE
The sensor is pre-calibrated at the factory. The built-in software in the logger can be
upgraded for free at any time using the provided firmware update.
The relative humidity sensor’s probe houses a capacitive polymer which reacts with
water vapor present in the air. The reaction with the polymer produces a voltage
which is proportional to the concentration of water vapor in the air. An internal
voltmeter reads the voltage and using the internal software a relative humidity is
calculated from the voltage.
Maintenance and storage:
Never submerge the NeuLog plastic body in any liquid.
Do not allow liquid into the relative humidity sensor’s body.
After use, gently wipe away any foreign material from the relative humidity
sensor.
Store in a box at room temperature out of direct sunlight.
Warranty:
We promise to deliver our sensor free of defects in materials and workmanship for a
period of 3 years from the date of purchase. Our warranty does not cover damage of
the product caused by improper use, abuse, or incorrect storage. Sensors with a shelf
life such as ion selective proves have a warranty of 1 year. Should you need to act
upon the warranty please contact your distributor. Your sensor will be repaired or
replaced.
Thank you for using NeuLog!
Flexible, simple, fast, forward thinking.
W: www.neulog.com
E: info@neulog.com
A: 850 St Paul Street, Suite 15, Rochester, NY 14605
P: 1.866.553.8536 V04302013
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.