INSTALLING THE NETRONIX™ THIAMIS
™
GSM/GPS COMMUNICATION MODEM INTO
DUSTTRAK
This guide will help you quickly install the Netronix™
Thiamis™ GSM/GPS Communication Modem onto the
various DustTrak™ Environmental Monitors.
Unpacking the Kit
1. Unpack the Netronix Thiamis GSM/GPS
Communication Modem (801905) and verify that all
items listed in the following table are present.
2. Contact TSI immediately if items are missing or
broken.
NOTE: Power should be disconnected from the
instrument prior to starting this procedure.
™
ENVIRONMENTAL MONITORS
Tools Needed for Installation
Phillips Screw Driver
_______________
TSI and TSI logo are registered trademarks of TSI Incorporated. DustTrak is a trademark of TSI Incorporated.
Environet, Netronix and Thiamis are trademarks of Netronix Inc.

1. Remove slip nut from
GSM/GPS antenna (nut
can be discarded).
2. Screw the GSM/GPS
antenna onto the post.
3. Feed the
GSM/GPS
antenna cables
through the fitting
on the top of the
enclosure and
push the antenna
onto the enclosure
bulkhead.
4. Remove the mounting
plate from the back of
the modem. To
separate the two
pieces, push the
button on the top of
the modem, rotate the
plate slightly away,
and slide the mounting
plate downwards.
Be careful not to rotate
the plate too far or the
tabs on the bottom
may break.
Installing the Modem
5. Fasten the mounting plate to the adapter plate using
the 6-32 x 3/16” pan head screws (qty. 2).
6. Snap the modem into the mounting plate.
–2–

7. Connect the GSM and GPS antenna cables to
the modem.
9. Connect the other end power cable and the RS232
to USB cables to Thiamis 1000 as shown.
8. Connect the power cable and the RS232 to USB
converter cable to the side of the DustTrak
photometer as shown.
10. Plug the AC/DC power supply into an AC power
source. Instrument and modem should power on.
11. The LED on the modem will blink while the modem
searches for a suitable network. If the LED stops
blinking and is solid blue, a network has been found
and proper communication with Environet™ should
be possible. If the LED changes to solid yellow, the
modem has not found a network. A power cycle may
resolve the issue or the unit may need to be moved
to a different location where signal strength is
improved.
–3–

USA Tel: +1 800 680 1220
UK Tel: +44 149 4 459200
France Tel: +33 1 41 19 21 99
Germany Tel: +49 241 523030
India Tel: +91 80 67877200
China Tel: +86 10 8219 7688
Singapore Tel: +65 6595 6388
TSI Incorporated – Visit our website www.tsi.com for more information.
P/N 6008413 Rev C ©2018 TSI Incorporated Printed in U.S.A.