NEOTECHA SAPRO SAMPLING DEVICES
MAINTENANCE AND REPAIR INSTRUCTIONS
Before installation these instructions must be fully read and understood
1 SAPRO SAMPLING DEVICES
Wafer body bayonet adaptor
Wafer body bottle adaptor
Neotecha Sapro sampling devices the following instructions should be read and understood thoroughly. Any questions about these instructions should be directed to Neotecha or any authorized distributor.
2 OPERATION
Details of the installation and operating procedures will be found in a separate publication: Maintenance and Installation Instructions VCIOM-01977 and VCIOM-01978.
Flanged body bayonet adaptor
Flanged body bottle adaptor
3 SAFETY
Prior to any disassembly, ensure that the unit has been thoroughly cleaned. In every case, protective equipment such as gloves, eye protection etc., should be used when operating, cleaning or maintaining the system. Attention must be given to any national or company regulations regarding dangerous media. Environmental protection regulations may vary according to the country or industry involved and must be strictly observed.
Sapro piston syringe
Emerson.com/FinalControl © 2017 Emerson. All Rights Reserved. VCIOM-01984-EN 21/04
NEOTECHA SAPRO SAMPLING DEVICES
MAINTENANCE AND REPAIR INSTRUCTIONS
4 ROUTINE INSPECTION
Sapro sampling valves and syringes are precision engineered devices and need to be kept clean and well maintained to ensure perfect operation. The Sapro units should be inspected periodically to ensure that seals are not leaking, as follows:
a. Valve seat
Excessive moisture on the head of the safety plug may indicate a leaking valve seat. (Refertoreplacement instruction further on).
b. Spindle seal
Remove the yellow screwed plastic plug at theleak detection port on the Sapro valve body. Excessive moisture at this point may indicate a leaking spindle seal. (Refer to replacement instructions further on).
c. Syringe nozzle seal
When a sample has been drawn into the syringe, examine the nozzle seal area and the O-ring and piston seals on the cylinder. Any leakage indicates the need to replace the relevant seals. (Refer to replacement instruction further on).
5 CLEANING THE SYRINGE
After each use of the Sapro syringe, the unit should be cleaned by thoroughly rinsing with an appropriate cleaning fluid as follows:
1. Press the operating lever (8) against thecylinder cage (9) and lock into position by moving with the locking lever (6) from position B and to position C.
2. Immerse the tip of the nozzle into a jar containing the cleaning fluid.
3. Pull and push the piston rod repeatedly until the syringe cylinder is clean.
9
8
C
B
Cleaning the syringe
6
2
NEOTECHA SAPRO SAMPLING DEVICES
MAINTENANCE AND REPAIR INSTRUCTIONS
Figure 1
a
b
11
10
9
5
6
2/3
7
4a 4
Figure 2
6 SAPRO SYRINGE: MAINTENANCE
6.1 Syringe disassembly
Disassembly of the syringe unit for cleaning, sterilization or replacement of components, isundertaken as follows (see Fig. 1).
1. Unscrew the cage end cap (13) and remove the cylinder assembly (10, 11, 12, 13 and 14) from the syringe cage (9).
2. Unscrew the grip (14) and remove the cage end cap (13).
3. Push the piston assembly (11) out of the cylinder (12) in the direction of arrow (a), thus pushing off the cylinder base (10) from the cylinder. The direction of removal is important to prevent damage to the piston seals.
4. Clean all disassembled parts and replace asnecessary.
13
14
12
Figure 3
6.4 Seal replacement
1. Place the nozzle (5) in tool No. 4, asshown (see Fig. 4).
2. Apply a little silicone oil to the replacement nozzle seal assembly (2 and 3) and insert it into the nozzle cavity.
3. Insert tool No. 3 into the nozzle and place tool No. 2 over tool No. 3, as shown.
4. Tap tool No. 2 with a mallet until the nozzle seal assembly is correctly located within thenozzle.
5. Replace the spacer (4) and washer (4a) over the syringe spindle as shown in Fig. 2.
6. Replace the nozzle assembly over the syringe spindle and screw into position.
7. Secure the nozzle by tightening set screws(6).
Nozzle tool No. 1
Figure 4
Nozzle tool No. 2
Nozzle tool No. 3
Nozzle tool No. 4
CAUTION
Reassemble the syringe by reversing the above procedure, ensuring that the piston is inserted into the cylinder in the direction of arrow (b).
6.2 Replacement of syringe nozzle seal
If the nozzle seal is leaking, it should be replaced by implementing the following procedure.
6.3 Seal removal (see Fig. 2)
1. With the syringe cap removed, loosen thetwo set screws (6).
2. Unscrew the nozzle (5) from the syringe head (7).
3. Ensure that the spacer (4) and washer (4a) are removed from the nozzle.
4. Place the nozzle on a firm surface. Place tool No. 1 on top of the nozzle and tap with a mallet until the nozzle seal assembly (2 and 3) falls out of the nozzle (see Fig. 3).