NEC MD215MG, MD242C2, MD211G5-N1, MD211G5-A1, MDG5-BNDN1 Miscellaneous Information

...
MultiSync MD Series: Quality assurance made easy
MultiSync MD Series displays are designed for impeccable image performance over their entire lifetime. Each unit is factory calibrated to the DICOM GSDF standard, and most models feature built-in sensors to ensure that this calibration is maintained.
GammaCompMD QA
Every MultiSync MD Series display comes with a copy of NEC’s full-feature quality assurance software, GammaCompMD QA Client. This software is recommended for calibration and quality assurance testing of the displays.
GammaCompMD QA can calibrate all MD Series displays listed on the next page to DICOM GSDF and can perform conformance testing to the AAPM TG-18 standard, providing a report of the displays’ performance. These tests can also be scheduled to execute
at times when the workstation is not in use. The software keeps a record of past calibrations and conformance tests, which is useful for Quality Assurance purposes and can automatically back up data to network locations.
GammaCompMD QA Server enhances the functionality of the Client software by providing a centralized interface for management of all Client installations on a network. Display status is available remotely as well as the ability to perform remote conformance tests and calibrations for most displays1.
For more information on the features and the latest version of the GammaCompMD QA software, visit www.necdisplay.com/GammaCompMD
Technology Paper
MultiSync MD Series Calibration Options
PRODUCT
MegaPixels
Front Sensor
Standalone Calibration Sensor Option
Quick QA Feature
PRODUCT
MegaPixels
Front Sensor
Standalone Calibration Sensor Option
Quick QA Feature
1
Scheduled tests and remote calibration and conformance are only available on displays with a front calibration sensor or the sentrydx remote calibration
sensor (MD-N2M5B).
MD211C2 MD212MC MD211C3 MD213MC MD211G3
2 2 3 3 3
Yes
MDSVSENSOR2 or i1 Display 2
Yes No Yes No No
MD213MG MD301C4 MD304MC MD215MG
3 4 4 5
Yes sentrydx option Yes
MDSVSENSOR2 or i1 Display 2 No
No
Standalone/Self Calibration
Every MultiSync MD Series display includes a standalone or self-calibration feature. This feature uses an external sensor (NEC MDSVSENSOR2 or X-Rite i1 Display2) plugged into a USB port on the display to perform a DICOM calibration without any software.
This calibration option is useful for situations where a computer is not available to run the GammaCompMD QA software. It does not provide any information on the parameters measured during calibration, so standalone calibration may not be ideal for ongoing quality assurance.
Quick QA
MultiSync MD211C2 and MD211C3 models include a Quick screen QA feature, available at the push of a button. This feature uses the built-in front calibration sensor on the display and is designed to allow the user to perform a simple visual test, while providing a measurement of the luminance and DICOM GSDF error percentage.
This QA test is ideal for times when a simple visual test is needed and there is a method for documenting the results, as the results of this test are not stored with the display.
This technology paper was published in and based on information as of January 2013. Technical information is subject to change.
©2013 NEC Display Solutions of America, Inc. All rights reserved.