21.3” grayscale medical-grade LCD displays ideal for diagnostic imaging
www.necdisplay.com
Modern medical technology enables complex examinations. But in the end, it is the radiologist who is responsible for fast and accurate diagnostic findings. To support these findings in a soft copy read­ing environment, flat-panel displays need to comply with the highest visual requirements. Their image quality has a decisive influence on diagnostic accuracy.
MultiSync MD Series grayscale displays – in combination with the appropriate graphics controller – are designed for diagnostic imaging and for use in Picture Archiving and Communication System (PACS) applications. Furthermore, diagnostic imaging displays from NEC Display Solutions are based on technologies and functions which actively contribute to a reduction of cost of ownership.
Figure 1.
9
1
2
3
4 5
X-Light technology - a view behind the screen
1 LCD panel 2 Polariser diffusing plate, etc. 3 Light guide plate 4 Color sensor 5 Temperature sensor 6 Current data of color sensor and temperature sensor 7 Control of 3 inverters 8 Calibration of whitepoint via pre-calibration in factory, software (GammaComp MD) or reference value of external sensor
9 Light sources, white light using a mixture of different phosphor colors
Factory calibration and consistency. Using state-
as it enables a more flexible use of display equipment. Since all X-Light-equipped monitors are set for the same whitepoint dur­ing production, they can be exchanged at any time within one location and swapped in existing dual- or multi-screen configu-
rations without calibration. monitor is calibrated to the DICOM display function for luminance and to a
8
desired whitepoint (color of the white). NEC Display
7
Solutions’ patent-pending X-LightTM technology (Fig.
1) has the unique ability to control and adjust the luminance and whitepoint
6
via an integrated fast-feed­back internal backlight sensor, which continuously
monitors and realigns these settings to maintain the factory calibration throughout the life of the monitor. X-Light technol­ogy also alleviates the color shift of the backlight to the yellow spectrum.
Standalone calibration and monitor matching. NEC MultiSync
MD Series monitors have been designed to maintain their fac-
tory calibration. However, if calibration or matching (Fig. 2) to
a replacement is required, the monitors can be configured using
the built-in, standalone algorithm programmed into the monitor’s
firmware. This feature is useful for monitors that are not used in
a setup where calibration can easily be performed with a PC and
external software or non-supported
operating systems.
GammaComp™ MD calibration and
conformance software. Quality control
and calibration/matching can also
be accomplished using GammaComp
Figure 2.
USB
RESULT OF
MEASUREMENT
Target: Luminance 400 cd/m
x cord 0.265 y cord 0.320
MD (shipped with display), ensuring consistent image quality on a single display system. The software main-
Source Display
Target Display
tains the monitor’s conformance to the DICOM standard, while providing an easy-to-use QA environment for medi-
2
USB
cal imaging. Optionally, GammaComp
Monitor matching and re­duced downtime. X-Light
helps reduce service costs
MD Administrator provides centralized control and management of multiple display systems.