National Instruments VXI-SC-1102 User Manual

VXI-SC-1102/B/C
User Manual
32-Channel Multiplexer Amplifier Submodules
August 1997 Edition
Part Number 321385B-01
© Copyright 1997 National Instruments Corporation. All rights reserved.
Internet Support
support@natinst.com
E-mail: info@natinst.com FTP Site: ftp.natinst.com Web Address: http://www.natinst.com
Bulletin Board Support
BBS United States: (512) 794-5422 BBS United Kingdom: 01635 551422 BBS France: 01 48 65 15 59
Fax-on-Demand Support
(512) 418-1111
Telephone Support (U.S.)
Tel: (512) 795-8248 Fax: (512) 794-5678
International Offices
Australia 03 9879 5166, Austria 0662 45 79 90 0, Belgium 02 757 00 20, Canada (Ontario) 905 785 0085, Canada (Québec) 514 694 8521, Denmark 45 76 26 00, Finland0972572511, France0148142424, Germany0897413130, Hong Kong 2645 3186, Israel 03 5734815, Italy 02 413091, Japan 03 5472 2970, Korea 02 596 7456, Mexico 5 520 2635, Netherlands 0348 433466, Norway 32 84 84 00, Singapore 2265886, Spain 91 640 0085, Sweden 08 730 49 70, Switzerland 056 200 51 51, Taiwan 02 377 1200, United Kingdom 01635 523545
National Instruments Corporate Headquarters
6504 Bridge Point Parkway Austin, TX 78730-5039 Tel: (512) 794-0100

Important Information

Warranty

The VXI-SC-1102,VXI-SC-1102B, and VXI-SC-1102C submodules are warranted against defects in materials and workmanship for a period of one year from the date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that proves to be defective during the warranty period. This warranty includes parts and labor.
The media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects in materials an d work manship, fo r a pe riod of 90 days from date of shipment , as evi denced by receipts or other documentation. National Instruments will, at its option, repair or replace software media that do not execute programming instructions if National Instruments receives notice of such defects during the warranty period. National Instruments does not warrant that the operation of the software shall be uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are covered by warranty.
National Instruments believes that the information in this manual is accurate. The document has been carefully reviewed for technical accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subseq uent editio ns of th is do cum ent wi thout prio r not ice to ho lders of this edit ion. Th e reader should consult National Instruments if errors are suspected. In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.
XCEPT AS SPECIFIED HEREIN
E
SPECIFICALLY DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE
USTOMER’S RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART OF NATIONAL
C
NSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE CUSTOMER
I
WILL NOT BE LIABLE FOR DAMAGES RESULTING FROM LOSS OF DATA, PROFITS, USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY THEREOF
Instruments will apply regardless of the form of action, whether in contract or tort, including negligence. Any action against National Instruments must be brought within one year after the cause of action accrues. National Instruments shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not cover damages, defects, malfunctions, or service failures caused by owner’s failure to follow the National Instruments installation, operation, or maintenance instructions; owner’s modification of the product; owner’s abuse, misuse, or negligent acts; and power failure or surges, fire, flood, accident, actions of third parties, or other events outside reasonable control.
ATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR IMPLIED, AND
, N
.
ATIONAL INSTRUMENTS
. N
. This limitation of the liability of National

Copyright

Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical, including photocopying, recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National Instruments Corporation.

Trademarks

LabVIEW®, NI-DAQ®, RTSI®, ComponentWorks™, CVI™, Measure™, SCXI™, and VirtualBench™ are trademarks of National Instruments Corporation.
Product and company names listed are trademarks or trade names of their respective companies.
WARNING REGARDING MEDICAL AND CLINICAL USE OF NATIONAL INSTRUMENTS PRODUCTS
National Instruments products are not designed with components and testing intended to ensure a level of reliability suitable for use in treatment and diagnosis of humans. Applications of National Instruments products involving medical or clinical treatment can create a potential for accidental injury caused by product failure, or by errors on the part of the user or application designer. Any use or application of National Instruments products for or involving medical or clinical treatment must be performed by properly trained and qualified medical perso nnel, and all traditi onal medical safeguards, equipment, and procedures that are appropriate in the particular situation to prevent serious injury or death should always continue to be used when National Instruments products are being used. National Instruments products are NOT intended to be a substitute for any form of established process, procedure, or equipment used to monitor or safeguard human health and safety in medical or clinical treatment.
About This Manual
Organization of This Manual........................................................................................ix
Conventions Used in This Manual................................................................................x
National Instruments Documentation...........................................................................xi
Related Documentation.......................................... .......................................................xii
Customer Communication............................................................................................ xii
Chapter 1 Introduction
About the VXI-SC-1102/B/C .......................................................................................1-2
What You Need to Get Started .....................................................................................1-3
Software Programming Choices...................................................................................1-4
National Instruments Application Software................................................... 1-4
NI-DAQ Driver Software...............................................................................1-5
VXIplug&play Instrument Driver..................................................................1-6
Optional Equipment......................................................................................................1-6
Unpacking.....................................................................................................................1-7
Table
of
Contents
Chapter 2 Installation
Current-Loop Receivers................................................................................................2-3
Installing Your Hardware .............................................................................................2-4
Chapter 3 Signal Connections
Front Connector.............................................................. ..................................... .........3-1
Signal Descriptions........................................................................................3-3
Channel Input Signal Connections.................................................................3-3
Ground-Referencing Your Signals...................................................3-4
Cold-Junction Sensor Connection..................................................................3-6
©
National Instruments Corporation v VXI-SC-1102/B/C User Manual
Table of Contents
Chapter 4 Theory of Operation
Functional Overview....................................................................................................4-1
VXIbus Interface..........................................................................................................4-2
Digital Control Circuitry .............................................................................................. 4-2
Analog Circuitry........................................................................................................... 4-3
Analog Input Channels................................................................................... 4-3
Analog Output ................................................................................................4-4
Chapter 5 Calibration
Calibration Equipment Requirements .......................................................................... 5-1
Gain and Offset Calibration..........................................................................................5-2
Appendix A Specifications
Appendix B Calibration Sample Program
Sample Program for Calibration................................................................................... B-1
Appendix C Customer Communication
Glossary
Index

Figures

Figure 1-1. VXI Signal Conditioning System ..........................................................1-1
Figure 1-2. The Relationship between the Programming Environment,
Your Instrument Driver, and Your VXI-DAQ Hardware..................... 1-6
Figure 2-1. VXI-SC-1102/B/C Parts Locator Diagram ........................................... 2-2
Figure 2-2. Bent and Trimmed Resistor................................................................... 2-3
VXI-SC-1102/B/C User Manual vi
©
National Instruments Corporation
Tables
Table of Contents
Figure 3-1. VXI-SC-1102/B/C Front Connector Pin Assignments..........................3-2
Figure 3-2. Ground-Referenced Signal Connection.................................................3-5
Figure 3-3. Floating Signal Connection Referenced to Chassis Ground..................3-6
Figure 4-1. VXI-SC-1102/B/C Block Diagram........................................................4-1
Figure 5-1. Analog Bus Receptacle Pin Assignment ...............................................5-2
Table 3-1. Front Connector Signal Descriptions.....................................................3-3
©
National Instruments Corporation vii VXI-SC-1102/B/C User Manual
About
This
Manual
This manual describes the electrical and mechanical aspects of the VXI-SC-1102 family and contains information concerning their installation and operation. The VXI-SC-1102 family is a group of signal conditioning submodules in the VXI-data acquisition (VXI-DAQ) line of National Instruments products.
The VXI-SC-1102 family consists of the following submodules:
VXI-SC-1102
VXI-SC-1102B
VXI-SC-1102C Unless otherwise noted, VXI-SC-1102/B/C will hereafter refer to all
three submodules in the VXI-SC-1102 family. The VXI-SC-1102/B/C submodules are designed for signal
conditioning of thermocouples, volt and millivolt sources, and 4–20 mA sources or 0–20 mA process-current sources. The VXI-SC-1102/B/C submodules have 32 differential analog input channels and one cold-junction sensor channel.

Organization of This Manual

The VXI-SC-1102/B/C User Manual is organized as follows:
Chapter 1, Introduction, describes the VXI-SC-1102 family of submodules, lists what you need to get started, optional software and optional equipment, and explains how to unpack yo ur VXI-SC-1102/B/C submodule.
Chapter 2, Installation, describes how to install current-loop receivers on your VXI-SC-1102/B/C and gives information about installing your VXI-SC-1102/B/C into the VXI-SC-1000 carrier module.
Chapter 3, Signal Connections, describes the analog input signal connections to the VXI-SC-1102/B/C submodule via the front connector.
©
National Instruments Corporation ix VXI-SC-1102/B/C User Manual
About This Manual
Chapter 4, Theory of Operation , contains a functional overview of the VXI-SC-1102/B/C submodules and explains the operation of each functional unit.
Chapter 5, Calibration, discusses the calibration procedures for the VXI-SC-1102/B/C submodules.
Appendix A, Specifications, lists the specifications for the VXI-SC-1102/B/C submodules.
Appendix B, Calibration Sample Program, contains a sample program to help you calibrate your submodule.
Appendix C, Customer Communication, contains forms you can use to request help from National Instruments or to comment on our products.
•The Glossary contains an alphabetical list and description of terms used in this manual, including abbreviations, acronyms, metric prefixes, mnemonics, symbols, and general data acquisition terms.
•The Index contains an alphabetical list of key terms and topics in this manual, including the page where you can find each one.

Conventions Used in This Manual

The following conventions are used in this manual.
< > Angle brackets containing numbers separated by an ellipsis represent a
range of values associated with a port, b it, or signal name (f or example, ACH<0..7> stands for the signals ACH0 through ACH7).
The ♦ symbol indicates that the text following it applies only to specific
VXI-SC-1102 modules. This icon to the left of bold italicized text denotes a note, which alerts
you to important information.
!
bold italic Bold italic text denotes a note, caution, or warning.
italic Italic text denotes emphasis, a cross reference, or an introduction to a
monospace Denotes text or characters that are to be literally input from the
VXI-SC-1102/B/C User Manual x
This icon to the left of bold italicized text denotes a caution, which advises you of precautions to take to avoid injury, data loss, or a system crash.
This icon to the left of bold italicized text denotes a warning, which advises you of precautions to take to avoid being electrically shocked.
key concept.
keyboard, sections of code, programming examples, and syntax
©
National Instruments Corporation
About This Manual
examples. This font is also used for the proper names of disk drives, paths, directories, programs, subprograms, subroutines, device names, functions, variables, file names, and extensions, and for statements and comments taken from program code.
carrier module Refers to the VXI-SC-1000, which plugs directly into a VXI slot and on
which VXI-SC submodules are installed. PC Refers to the IBM PC/XT, the IBM PC AT, and compatible computers. SCMP Refers to the signal conditioning minipods in the VXI signal
conditioning system. VXI-MIO module Refers to any of the National Instruments VXI-MIO series of plug-in
data acquisition devices. VXI-SC submodule Refers to VXI signal conditioning hardware that installs onto the carrier
module. VXI-SC-1102/B/C Refers to all submodules in the VXI-SC-1102 family, unless otherwise
submodules noted.
The Glossary lists abbreviations, acronyms, metric prefixes,
mnemonics, symbols, and terms.

National Instruments Documentation

The VXI-SC-1102/B/C User Manual is one piece of the documentation
set for your VXI-DAQ system. You could have any of several types of
documents, depending on the hardware and software in your system.
Use the different types of documents you have as follows:
Your VXI-DAQ hardware documentation—These documents have detailed information about the V XI-DAQ hardware that plugs into or is connected to your VXIbus chassis. Use these documents for hardware installation and configuration instructions, specification information about your VXI-DAQ hardware, and application hi nts.
Software documentation—You may have both application software and driver software documentation. National Instruments application software includes ComponentWorks, LabVIEW, LabWindows Instruments driver software includes NI-DAQ and VXIplug&play instrument drivers. After you set up your hardware system, use either your application or driver software documentation to help you write your application. If you have a large, complicated system, it is worthwhile to look through the software documentation before you configure your hardware.
©
National Instruments Corporation xi VXI-SC-1102/B/C User Manual
®
/CVI, Measure, and VirtualBench. National
About This Manual
Accessory installation guides or manuals—If you are using accessory products, read the terminal block and cable assembly installation guides or accessory device user manuals. They explain how to physically connect the relevant pieces of the system. Consult these guides when you are making your connections.

Related Documentation

The following document contains information you may find helpful:
VXI-SC-1000 Carrier Module Installation Guide

Customer Communication

National Instruments wants to receive your comments on our products and manuals. We are interested in the applications you develop with our products, and we want to help if you have problems with them. To make it easy for you to contact us, this manual contains comment and configuration forms for you to complete. These forms are in Appendix C, Customer Communication, at the end of this manual.
VXI-SC-1102/B/C User Manual xii
©
National Instruments Corporation
Chapter
Introduction
This chapter describes the VXI-SC-1102 family of submodules; lists what you need to get started, optional soft ware, and optional equipment; and explains how to unpack your VXI-SC-1102/B/C submodule.
The VXI-SC-1102 family is part of the VXI signal conditioning system, which consists of a carrier module and one or more VXI signal conditioning submodules or filler panels as shown in Figure 1-1.
2
1
4
3
5
To VXIbus
Chassis
6
8
1
9
1 Analog Bus Receptacle 2 SCMPs 3 SCMP Socket
©
National Instruments Corporation 1-1 VXI-SC-1102/B/C User Manual
4 Address Switch 5 VXI-SC Subslot 1 6 VXI-SC Subslot 2
Figure 1-1.
7 VXI-SC Carrier Module 8 VXI-SC-1150 Submodule 9 VXI-SC Submodule
VXI Signal Conditioning System
7
Chapter 1 Introduction
The VXI-SC-1000 is a carrier module that installs into a slot in a VXIbus chassis and is populated with one or two submodules. The VXI-SC-1102/B/C is a submodule that installs into VXI-SC subslots on a VXI-SC-1000 carrier module.
Your VXI-SC-1000 carrier module is shipped with a filler panel installed in one of the submodule sites. This filler panel is necessary for the attachment of the front panel, so you should leave it in place until you are ready to install a VXI-SC submodule.
About the VXI-SC-1102/B/C
The VXI-SC-1102 is a submodule for the signal conditioning of
thermocouples, low-bandwidth volt and millivolt sources, 4-20 mA current sources, and 0-20 mA process-current sources. The VXI-SC-1102 has 32 differential analog input channels and one cold-junction sensor channel. On each channel, the VXI-SC-1102 has a three-pole lowpass filter with a 2 Hz cutoff frequency to reject 60 Hz noise. Each channel also has an amplifier with a selectable gain of 1 or 100. You can multiplex the VXI-SC-1102 inputs to a single output, which drives a single VXI-MIO module channel.
The VXI-SC-1102B is a submodule for the signal conditioning of
thermocouples, medium-bandwidth volt and millivolt sources, 4-20 mA current sources, and 0-20 mA process-current sources. The VXI-SC-1102B has 32 differential analog input channels and one cold-junction sensor channel. On each channel, the VXI-SC-1102B has a three-pole lowpass filter with a 200 Hz cutoff frequency. Each channel also has an amplifier with a selectable gain of 1 or 100. You can multiplex the VXI-SC-1102B inputs to a single output, which drives a single VXI-MIO module channel.
The VXI-SC-1102C is a submodule for the signal conditioning of
thermocouples, high-bandwidth volt and millivolt sources, 4-20 mA current sources, and 0-20 mA process-current sources. The VXI-SC-1102C has 32 differential analog input channels and one cold-junction sensor channel. On each channel, the VXI-SC-1102C has a three-pole lowpass filter with a 10 kHz cutoff frequency. Each channel also has an amplifier with a selectable gain of 1 o r 100. You can multiplex the VXI-SC-1102C inputs to a single output, which drives a single VXI-MIO module channel.
VXI-SC-1102/B/C User Manual 1-2
©
National Instruments Corporation
The VXI-SC-1102/B/C submodules operate with full functionality with National Instruments VXI-MIO modules. You can multiplex several VXI-SC-1102/B/C submodules and other VXI signal conditioning submodules into a single channel on the VXI-MIO module, greatly increasing the number of analog input signals that you can digitize.
You can also use National Instruments terminal blocks, which have screw terminals to which you attach the input signals for the VXI-SC-1102/B/C submodules. In addition, some of these terminal blocks have a temperature sensor for thermocouple cold-junction compensation (CJC). The VXI-SC-1102/B/C submodules multiplex this cold-junction sensor with the 32 input channels during a hardware scan. A National Instruments isothermal terminal block is recommended for thermocouple applications.
Detailed specifications of the VXI-SC-1102/B/C submodules are listed in Appendix A, Specifications.

What You Need to Get Started

To set up and use your VXI-SC-1102/B/C, you will need the following items:
Chapter 1 Introduction
One of the following submodules:
VXI-SC-1102 VXI-SC-1102B VXI-SC-1102C
VXI-SC-1102/B/C User ManualVXI-SC-1000 carrier module and documentationVXI-MIO module and documentation One of the following software packages and documentation:
ComponentWorks LabVIEW for Windows LabWindows/CVI for Windows Measure NI-DAQ for PC compatibles VirtualBench VXIplug&play instrument driver
Your VXIbus system
©
National Instruments Corporation 1-3 VXI-SC-1102/B/C User Manual
Chapter 1 Introduction

Software Programming Choices

There are several options to choose from when programming your National Instruments VXI-DAQ hardware. You can use LabVIEW, LabWindows/CVI, ComponentWorks, VirtualBench, or other application development environments with either NI-DAQ or the VXIplug&play instrument driver. Both NI-DAQ and the VXIplug&play instrument driver access the VXI-DAQ hardware through the VISA driver software.

National Instruments Application Software

LabVIEW features interactive graphics, a state-of-the-art user interface, and a powerful graphical programming language. The LabVIEW Data Acquisition VI Library, a series of VIs for using LabVIEW with National Instruments DAQ hardware, is included with LabVIEW. The LabVIEW Data Acquisition VI Library is functionally equivalent to the NI-DAQ software.
LabWindows/CVI features interactive graphics, a state-of-the-art user interface, and uses the ANSI standard C programming language. The LabWindows/CVI Data Acquisition Library, a series of functions for using LabWindows/CVI with National Instruments DAQ hardware, is included with the NI-DAQ software kit. The LabWindows/CVI Data Acquisition Library is functionally equivalent to the NI-DAQ so ftware.
ComponentWorks contains tools for data acquisition and instrument control built on NI-DAQ driver software. Compo nentWorks provi des a higher-level programming interface for building virtual instruments through standard OLE controls and DLLs. With ComponentWork s, you can use all of the configuration tools, resource management utilities, and interactive control utilities included with NI-DAQ.
VirtualBench features VIs that combine DAQ products, software, and your computer to create a stand-alone instrument with the added benefit of the processing, display, and storage capabilities of your computer. VirtualBench instruments load and save waveform data to disk in the same forms that can be used in popular spreadsheet programs and word processors.
Using LabVIEW, LabWindows/CVI, ComponentWorks, or VirtualBench software will greatly reduce the development time for your data acquisition and control application.
VXI-SC-1102/B/C User Manual 1-4
©
National Instruments Corporation
NI-DAQ Driver Software
The NI-DAQ driver software is included at no charge with all National Instruments DAQ hardware. NI-DAQ is not packaged with accessory products. NI-DAQ has an extensive library of functions that you can call from your application programming environment. These functions include routines for analog input (A/D conversion), buffered data acquisition (high-speed A/D conversion), analog output (D/A conversion), waveform generation, digital I/O, counter/timer operations, SCXI, RTSI, self-calibration, messaging, and acquiring data to extended memory .
NI-DAQ has both high-level DAQ I/O functions for maximum ease of use and low-level DAQ I/O functions for maximum flexibility and performance. Examples of high-level functions are streaming data to disk or acquiring a certain number of data points. An example of a low-level function is writing directly to registers on the DAQ device. NI-DAQ does not sacrifice the performance of National Instruments DAQ devices because it lets multiple devices operate at their peak performance.
NI-DAQ maintains a consistent software interface among its different versions so that you can change platforms with minimal modifications to your code.
Chapter 1 Introduction
©
National Instruments Corporation 1-5 VXI-SC-1102/B/C User Manual
Chapter 1 Introduction
VXI
plug&play
Instrument Driver
National Instruments distributes VXIplug&play instrument drivers free of charge. VXIplug&play instrument drivers are one level above the NI-DAQ device driver and contain high-level software functions whose architecture is specified by the VXIplug&play Systems Alliance. The VXIplug&play standards increase interoperability with other vendors, and ensure that drivers are designed and presented in a consistent fashion that facilitates ease of use. Refer to Figure 1-2 to see the relationship between your VXI-DAQ hardware and your software.
Figure 1-2.

Optional Equipment

National Instruments offers a variety of products to use with your VXI-SC-1102/B/C submodule, as follows:
Terminal blocks and 96-pin cables that allow you to attach input signals to your submodule
SCXI Process Current Resistor Kit
ComponentWorks,
LabVIEW, LabWindows/CVI,
or VirtualBench
plug&play
VXI
Instrument Driver
NI-DAQ Driver Software
VISA
VXI-DAQ Hardware
The Relationship between the Programming Environment,
Your Instrument Driver, and Your VXI-DAQ Hardware
Other Application
Development Environments
For more specific information about these products, refer to your National Instruments catalogue or call the office nearest you.
VXI-SC-1102/B/C User Manual 1-6
©
National Instruments Corporation

Unpacking

Chapter 1 Introduction
Your VXI-SC-1102/B/C submodule is shipped in an antistatic package to prevent electrostatic damage to the submodule. Electrostatic discharge can damage several components on the submodule. To avoid such damage in handling the submodule, take the following precautions:
Ground yourself via a grounding strap or by holding a grounded object.
Touch the antistatic package to a metal part of your VXIbus chassis before removing the submodule from the package.
Remove the submodule from the package and inspect the submodule for loose components or any other sign of damage. Notify National Instruments if the module appears damaged in any way. Do not install a damaged submodule into your VXIbus chassis.
Never touch the exposed pins of connectors.
©
National Instruments Corporation 1-7 VXI-SC-1102/B/C User Manual
Chapter
Installation
2
This chapter describes how to install current-loop receivers on your VXI-SC-1102/B/C and gives information about installing your VXI-SC-1102/B/C into the VXI-SC-1000 carrier module.
Refer to Figure 2-1, VXI-SC-1102/B/C Parts Locator Diagram, for information as you read the instructions in this chapter.
©
National Instruments Corporation 2-1 VXI-SC-1102/B/C User Manual
Chapter 2 Installation
2
3
1
4
1 Serial Number 2 Product Name 3 Assembly Number 4 RCLx Pads
Figure 2-1. VXI-SC-1102/B/C Parts Locator Diagram
VXI-SC-1102/B/C User Manual 2-2
©
National Instruments Corporation
Current-Loop Receivers
The VXI-SC-1102/B/C submodules have pads for transforming individual channels to current-to-voltage converters. National Instruments offers an SCXI Process Current Resistor Kit of four 249 ,
0.1%, 5 ppm, 1/4 W resistors. The reference designators for the current
loop resistors have the following format: for input channel x, the resistor is RCLx. For example, the resistor for channel 28 is RCL28.
Caution: Before installing the resistors in your submodule, make sure that there are
!
no signals connected to your submodule front connector.
Install the resistors by performing the following steps:
1. Follow the instructions in the VXI-SC-1000 Carrier Module Installation Guide to remove your VXI-SC-1102/B/C submodule from the carrier assembly.
2. Bend and trim the resistor lead as shown in Figure 2-2. Be sure that the resistor does not extend more than 0.65 in. above the surface of the circuit board.
Chapter 2 Installation
Figure 2-2.
3. Insert the resistor into the appropriate pa d, labeled RCLx.
4. Solder the leads to the pad on the bottom side of the module.
5. Follow the instructions in the VXI-SC-1000 Carrier Module Installation Guide to replace the VXI-SC-1102/B/C submodule in the carrier assembly.
©
National Instruments Corporation 2-3 VXI-SC-1102/B/C User Manual
Bent and Trimmed Resistor
Chapter 2 Installation

Installing Your Hardware

Follow the instructions in the VXI-SC-1000 Carrier Module Installation Guide to install the assembled VXI-SC carrier module in the VXIbus
chassis.
VXI-SC-1102/B/C User Manual 2-4
©
National Instruments Corporation
Chapter
Signal Connections
This chapter describes the analog input signal connections to the VXI-SC-1102/B/C submodule via the front connector.
Caution:
!

Front Connector

Static electricity is a major cause of component failure. To prevent damage to the electrical components in the module, observe antistatic techniques whenever removing a module from the chassis or whenever working on a module.
Figure 3-1 shows the pin assignments for the VXI-SC-1102/B/C submodule front connector.
3
©
National Instruments Corporation 3-1 VXI-SC-1102/B/C User Manual
Chapter 3 Signal Connections
CGND
NC NC NC NC NC NC NC
CGND
NC NC NC NC NC NC NC
CGND
NC NC NC NC NC NC NC NC NC NC
CGND
CJSENSOR CJSENSOR
CGND
+5 V
32 31 30 29 28 27 26 25 24 23 22 21 20 19 18 17 16 15 14 13 12 11 10
A
CH0­CH1­CH2­CH3­CH4­CH5­CH6­CH7­CH8-
CH9­CH10­CH11­CH12­CH13­CH14­CH15­CH16­CH17­CH18­CH19­CH20­CH21­CH22-
9 8 7 6 5 4 3 2 1
CH23­CH24­CH25­CH26­CH27­CH28­CH29­CH30­CH31-
32 31 30 29 28 27 26 25 24 23 22 21 20 19 18 17 16 15 14 13 12 11 10
B
CH0+ CH1+ CH2+ CH3+ CH4+ CH5+ CH6+ CH7+ CH8+
CH9+ CH10+ CH11+ CH12+ CH13+ CH14+ CH15+ CH16+ CH17+ CH18+ CH19+ CH20+ CH21+ CH22+
9 8 7 6 5 4 3 2 1
CH23+ CH24+ CH25+ CH26+ CH27+ CH28+ CH29+ CH30+ CH31+
C
32 31 30 29 28 27 26 25 24 23 22 21 20 19 18 17 16 15 14 13 12 11 10
9 8 7 6 5 4 3 2 1
Figure 3-1. VXI-SC-1102/B/C Front Connector Pin Assignments
VXI-SC-1102/B/C User Manual 3-2
©
National Instruments Corporation

Signal Descriptions

Chapter 3 Signal Connections
Table 3-1.
Pin Signal Name Description
A1 +5 V +5 VDC Source—Powers the
A2, A5, A16, A24, A32
A3, A4 CJSENSOR Cold-Junction Temperature
B1–B32 CH31-
C1–C32 CH31+
All other pins are not connected.
Front Connector Signal Descriptions
temperature sensor on the terminal block. 0.2 mA of source not protected.
CGND Chassis Ground—Tied to the
VXIbus chassis.
Sensor Input—Connects to the temperature sensor of the terminal block.
Negative Input Channels 31–0 through CH0-
through CH0+
—Negative side of differential
input channels.
Positive Input Channels 31–0
—Positive side of differential
input channels.

Channel Input Signal Connections

The signal terminals for the positive input channels are located in column C of the connector. The signal terminal for each corresponding negative input channel is located in column B of the connector. Each input goes to a separate filter and amplifier that is multiplexed to the module output buffer. The terminal block temperature sensor output— connected to pins A3 and A4 (CJSENSOR)—is also filtered and multiplexed to the module output buffer.
The differential input signal range of a VXI-SC-1102/B/C input channel is ±10 V/G, where G is the gain selected on the VXI-SC-1102/B/C input channel. This differential input range is the maximum measurable voltage difference between the positive and negative channel inputs. The common-mode input signal range of a VXI-SC-1102/B/C input
©
National Instruments Corporation 3-3 VXI-SC-1102/B/C User Manual
Chapter 3 Signal Connections
Warning: Exceeding the input damage level (±42 VAC peak or VDC between input
channels and chassis ground) can damage the VXI-SC-1102/B/C submodule, the VXIbus, and the VXI-MIO module. National Instruments is
NOT
Applying a voltage greater than ±42 VAC peak or VDC to the VXI-SC-1102/B/C is an electrical shock hazard. National Instruments is
NOT
application.
Note: Exceeding the differential or common-mode input channel ranges results
in a distorted signal measurement.
channel is ±11 V. This common-mode input range for either positive or negative channel input is the maximum input voltage (relative to CGND) that will result in a valid measurement. Each channel includes input protection circuitry to withstand the accidental application of voltages up to ±42 VAC peak or VDC.
liable for any injuries resulting from such signal connections.
liable for any damages or injuries resulting from such voltage
Ground-Referencing Your Signals
Your input signals can be either ground-referenced, as shown in Figure 3-2, or floating, as shown in Figure 3-3. Before you connect your thermocouple or any other signal, determine whether it is fl oatin g or ground-referenced. If it is a floating signal, you must ground­reference the signal in one of two ways. You can connect the negative channel input to chassis ground as shown in Figure 3-3 or you can use the clamping resistors on a VXI-TB Series terminal block. The VXI-TB Series terminal block may also have a resistor pack for pulling up the positive inputs for open-thermocouple detection. Consult the terminal block installation guide for details.
Do not ground signals that are already ground -referenced; doing so will result in a ground loop, which adversely affects your measurement accuracy.
VXI-SC-1102/B/C User Manual 3-4
©
National Instruments Corporation
Chapter 3 Signal Connections
CH0 + CH0 -
+
V
diff
-
+
V
cm
-
CH1 +
CH1 -
Connector
Front Signal
+
-
+
-
VXI-SC-1102/B/C

Figure 3-2. Ground-Referenced Signal Connection

©
National Instruments Corporation 3-5 VXI-SC-1102/B/C User Manual
Chapter 3 Signal Connections
CH0 + CH0 -
Floating
Signal
Make this connection to
ground reference the signal.
+
V
s
-
Reference to
Chassis Ground
CH1 + CH1 -
Front Signal
CGND
Figure 3-3. Floating Signal Connection Referenced to Chassis Ground

Cold-Junction Sensor Connection

Pins A3 and A4 (CJSENSOR) connect the temperature sensor located on the National Instruments VXI-TB Series terminal block to the VXI-SC-1102/B/C. The CJSENSOR signal is measured relative to CGND. Pins A3 and A4 are connected together within the VXI-SC-1102/B/C so the position of the MTEMP/DTEMP switch on the VXI-TB Series terminal block does not matter. The input is overvoltage-protected to 15 VDC with the power on and off.
+
-
+
-
Connector
VXI-SC-1102/B/C
Caution: Exceeding the overvoltage protection on the CJSENSOR input can damage
!
the VXI-SC-1102/B/C submodule, the VXIbus, and the VXI-MIO board. National Instruments is
NOT
from such signal connections.
VXI-SC-1102/B/C User Manual 3-6
liable for any damages or injuries resulting
©
National Instruments Corporation
Chapter
Theory of Operation
This chapter contains a functional overview of the VXI-SC-1102/B/C submodules and explains the operation of each functional unit.

Functional Overview

The block diagram in Figure 4-1 illustrates the key functional components of the VXI-SC-1102/B/C submodules.
CH 0 +
CH 0 –
+
Inst. Amp
Input Protection
and Lowpass Filter
Gain 0
Lowpass
Filter
Buffer
4
Buffer
Mux
32-to-1 Mux
CH 31+
CH 31 –
Front Signal Connector
CJSENSOR
©
National Instruments Corporation 4-1 VXI-SC-1102/B/C User Manual
+
Inst. Amp
Input Protection
and Lowpass Filter
Input Protection
and Lowpass Filter
Gain 31
Lowpass
Filter
Lowpass
Filter
Buffer
Buffer
Figure 4-1.
Switch Switch
Gain
Register
Calibration EEPROM
Digital
Control
VXI-SC-1102/B/C Block Diagram
Analog Bus + Analog Bus -
Connector to Carrier Module
Chapter 4 Theory of Operation
The major components of the VXI-SC-1102/B/C submodules are as follows:
VXIbus interface
Digital control circuitry
Analog circuitry The VXI-SC-1102/B/C submodules consist of 32 multiplexed input
channels, each with a software-programmable gain of 1 or 100. Each input channel has its own lowpass filter. The VXI-SC-1102/B/C submodules also have digital control circuitry for automatic control of channel scanning, temperature sensor selection, and gain selection.

VXIbus Interface

The VXI-DAQ module controls the VXI-SC-1102/B/C over the VXIbus. The VXI-SC-1000 carrier module serves as an interface from the VXI-SC-1102/B/C submodules to the VXIbus.

Digital Control Circuitry

The digital control circuitry consists of the address handler and the following registers: Module ID, Configuration, Status, EEPROM, Gain, and Channel. The address handler controls which re gister is being addressed. The Module ID Register contains a code unique to each type of VXI-SC-1102/B/C submodule:
VXI-SC-1102—The Module ID is 42 decimal. VXI-SC-1102B—The Module ID is 43 decimal. VXI-SC-1102C—The Module ID is 63 decimal.
The Configuration Register configures the VXI-SC-1102/B/C submodules for the desired scanning mode and connection to the rear signal connector. The Status Register indicates whether the input channels have settled after a change in the gains. The EEPROM Register is the address for interfacing with the submodule’s EEPROM, which contains calibration information. The Gain Register selects between gains of 1 or 100 for each of the 32 channels. The Channel Register selects a channel for a single measurement or a start channel for a scan. Refer to Software Programming Choices in Chapter 1,
VXI-SC-1102/B/C User Manual 4-2
©
National Instruments Corporation
Introduction, of this manual to learn about options for programming the control circuitry.

Analog Circuitry

The analog circuitry per channel consists of a lowpass filter and an amplifier with a software-selectable gain of 1 or 100. The CJSENSOR channel also has a buffered lowpass filter but has no amplifier. The channels and CJSENSOR are multiplexed to a single output buffer.

Analog Input Channels

Each of the 32 analog input channels drives a separate amplifier with a programmable gain of 1 or 100. Then the signal passes through a three-pole lowpass filter.
Note: Because of the 2 Hz bandwidth of the VXI-SC-1102 input channels, you
must wait approximately 3 s after changing the gains before the channels settle and you can take an accurate measurement. NI-DAQ automatically reads the Status Register to determine when the module output has settled. This time is approximately 100 ms and 1 ms for the VXI-SC-1102B and VXI-SC-1102C, respectively.
Chapter 4 Theory of Operation
The temperature sensor consists of a thermistor located on a National Instruments VXI terminal block. The temperature sensor is for cold-junction compensation of thermocouples. The CJSENSOR channel also passes through a 2 Hz lowpass filter to reject unwanted noise. Along with the other 32 input channels, the CJSENSOR is multiplexed to the output buffer, wh ere it can be read by the VXI-MIO module.
For a measurement accuracy of 0.012% of full scale, the minimum scan interval is 3 µs. This is the smallest interval in which you can switch between analog channels on the submodule and still measure accurate voltages. The 3 µs scan interval gives you a maximum sampling rate o f 333 kHz. Because each VXI-SC-1102/B/C channel contains a lowpass filter, the 333 kHz sample rate allows you to sample multiple channels without undersampling any channel.
©
National Instruments Corporation 4-3 VXI-SC-1102/B/C User Manual
Chapter 4 Theory of Operation

Analog Output

The 32 input channels and the CJSENSOR are multiplexed into one output. This output is passed to the VXI-SC-1000 carrier module, where it is connected to the analog bus for measurement by a VXI-MIO module.
VXI-SC-1102/B/C User Manual 4-4
©
National Instruments Corporation
Chapter
Calibration
This chapter discusses the calibration procedures for the VXI-SC-1102/B/C submodules.
Using the procedure described in this chapter, you will be able to calculate the gain error and voltage offset on a per channel per gain basis. You can store these constants in the onboard EEPROM for future use and for automatic calibration when you are using National Instruments software. The VXI-SC-1102/B/C submodules come from the factory with factory-determined calibration constants in the EEPROM. However, National Instruments recommends that you recalibrate your VXI-SC-1102/B/C at least once per year or when you operate the submodule outside the 20° to 30° C temperature range.
In order to calibrate the VXI-SC-1102/B/C, you will need to apply precision voltages to the channel inputs and/or ground the channel inputs.

Calibration Equipment Requirements

5
According to standard practice, the equipment you use to calibrate the VXI-SC-1102/B/C should be 10 times as accurate as the VXI-SC-1102/B/C submodule itself. Calibration equipment with four times the accuracy of the VXI-SC-1102/B/C is generally considered acceptable. To calibrate the VXI-SC-1102/B/C submodules, you need a voltmeter with the following specifications:
Accuracy ±6 ppm standard reading ±15 ppm sufficient reading
Range ±10 V
Resolution 8.5 digits
A multiranging 8.5-digit digital multimeter (DMM) can perform the necessary calibrations.
To make sure that the DMM does not introdu ce an additional offset, you can determine the offset error of the DMM by shorting its leads together
©
National Instruments Corporation 5-1 VXI-SC-1102/B/C User Manual
Chapter 5 Calibration
and reading the measured value. This value, the DMM offset, must be subtracted from all subsequent measurements.
Measure the submodule’s output at the analog bus receptacle (see Figure 1-1 for the location of this receptacle). Refer to Figure 5-1 to connect the positive DMM input to MUX+ and the negative DMM input to MUX-.
MUX+ MUX- CAL- CAL+
MUX- MUX+ CAL+ CAL-

Figure 5-1. Analog Bus Receptacle Pin Assignment

Gain and Offset Calibration

To determine the offset and gain calibration factors of the VXI-SC-1102/B/C submodules for a given gain, perform the following steps for a two-point calibration. For two-point calibration, it is best to use input voltages that correspond to the signal range of interest. For example, to measure bipolar voltages over the submodule’s full input range, choose –9.9 V/G (negative full-scale) and +9.9 V/G (positive full-scale) as your two input voltages.
1. Select the desired channel. Set the channel gain to the desired gain.
2. Apply the input for the first calibration point. To select negative full scale as the calibration point, apply –9.9 V/G for an input voltage.
a. Apply the input voltage to the channel selected in Step 1.
VXI-SC-1102/B/C User Manual 5-2
©
National Instruments Corporation
Chapter 5 Calibration
Note: To make one of yo ur calibrat ion po int s the zero point, connect the positive
and negative channel leads to one of the chassis ground pins on the front connector or terminal block.
b. Measure the input voltage with the DMM. Call the measured
voltage
input1.
Note: If you are using a calibrator that supplies accurate volt ages, you can simply
use the known applied voltage for
input1 instead of measuring.
c. Measure the module output at the analog bus receptacle with
the DMM. Call the measured voltage
output1
.
3. Repeat Step 2, applying the input for the second calibration point. Call the measured voltages
input2 and output2. To select
positive full scale as the calibration point, repeat Step 2 and apply +9.9 V/G.
4. You now have two pairs of voltages (
(
input2, output2
). Each pair consists of an input voltage and an
input1, output1
) and
output voltage.
5. Convert the output voltages from volt units to your VXI-MIO module binary unit. You must take into consideration the polarity of your VXI-MIO module, its resolution (12 bits or 16 bits), and gain. For example, if you are using the 12-bit VXI-MIO-64E-1 in bipolar mode with the gain set to G
, your output voltages for the
MIO
autozeroing option will be represented in binary units as given by the following formula:
Binary
=
Output
----------------- ­5 V
12
2
G
MIO
For other VXI-MIO modules, refer to the device user manual to determine the appropriate formula.
6. You now have a new set of pairs referred to as voltage binary pairs
input1, bin_output1) and (input2, bin_output2). Pass these
( pairs to the
SCXI_Cal_Constants function or VI as described in
your software user documentation.
Note: When you use 0 V and positive full-scale for your two calibration points,
you eliminate the error at 0 V and at positive full-scale voltage. However, because of nonlinearity, the error at the negative full-scale voltage will be two times the nonlinearity error. This is also true for the positive full-scale voltage if you use the negative full-scale voltage and 0 V as your two calibration points.
©
National Instruments Corporation 5-3 VXI-SC-1102/B/C User Manual
Chapter 5 Calibration
When you make a measurement using LabVIEW, ComponentWorks, or VirtualBench, the driver automatically uses the calibration constants to correct the measured voltages.
When you use application development environments other than LabVIEW, ComponentWorks, or VirtualBench, you have the option to apply the calibration constants to measured voltages through the NI-DAQ API.
Refer to Appendix B, Calibration Sample Program, for an example of a calibration program for your submodule.
VXI-SC-1102/B/C User Manual 5-4
©
National Instruments Corporation
Appendix
Specifications
This appendix lists the specifications for the VXI-SC-1102/B/C submodules. These specifications are typical at 25° C unless otherwise noted.
VXI-SC-1102/B/C
Analog Input
Input Characteristics
Number of channels............................32 differential
Input signal ranges................................±100 mV (gain = 100) or
Max working voltage
(signal + common mode) ....................Each input should remain
A
±10 V (gain = 1)
within ±10 V of CGND
Input damage level ...............................±42 VAC peak or VDC
Inputs protected............................CH<0..31>, CJSENSOR
Transfer Characteristics
Nonlinearity........................................0.005% FSR
Offset error
Gain = 1
After calibration .................... 300 µV max
Before calibration..................600 µV
Gain = 100
After calibration .................... 15 µV max
Before calibration..................100 µV
©
National Instruments Corporation A-1 VXI-SC-1102/B/C User Manual
Appendix A Specifications
Gain error
Gain = 1
After calibration.................... 0.015% of reading max
Before calibration ................. 0.04% of reading
Gain = 100
After calibration.................... 0.02% of reading max
Before calibration ................. 0.1% of reading
Amplifier Characteristics
Input impedance
Normal powered on ..................... >1 G
Powered off................................. 10 k
Overload ...................................... 10 k
Input bias current...................................±0.5 nA
Input offset current................................±1.0 nA
CMRR
50 to 60 Hz, either gain 110 dB 90 dB 90 dB 0 Hz, gain 1 75 dB min 75 dB min 75 dB min 0 Hz, gain 100 100 dB min 100 dB min 100 dB min
Output range ..........................................±10 V
Output impedance ..............................0.1
Dynamic Characteristics
Bandwidth (-3 dB) ............................. 2 Hz (1102)
Scan interval (per channel, any gain)
0.012% ........................................ 3 µs
0.0061% ......................................10 µs
VXI-SC-1102/B/C User Manual A-2
1102 1102B 1102C
200 Hz (1102B) 10 kHz (1102C)
©
National Instruments Corporation
Filters
Appendix A Specifications
System noise (related to input)
1102 1102B 1102C
Gain = 1 50 µVrms 50 µVrms 70 µVrms Gain = 100 5 µVrms 5 µVrms 10 µVrms
Cutoff frequency (–3 dB)....................2 Hz (1102)
200 Hz (1102B) 10 kHz (1102C)
NMR (60 Hz)......................................40 dB
Step response (either gain)
1102 1102B 1102C
To 0.1% 1 s 10 ms 200 µs To 0.01% 10 s 100 ms 1 ms
Stability
Recommended warm-up time..............20 min.
Offset temperature coefficient
Gain = 1 .......................................20 µV/°C
Gain = 100 ...................................1 µV/°C
Gain temperature coefficient ............... 10 ppm/ °C
Physical
Dimensions.........................................115 by 273 mm
(4.54 by 10.75 in.)
I/O connector......................................50-pin male ribbon cable
rear connector 96-pin male DIN C front connector
©
National Instruments Corporation A-3 VXI-SC-1102/B/C User Manual
Appendix A Specifications
Environment
Operating temperature ........................ 0° to 50° C
Storage temperature ........................... –55° to 150° C
Relative humidity............................... 5% to 90% noncondensing
Power Requirements
5 V supply .......................................... 15 mA max
±15 V supply (regulated
from ±24 V supply)............................ 150 mA max
VXI-SC-1102/B/C User Manual A-4
©
National Instruments Corporation
Appendix
Calibration Sample Program
This chapter contains a sample program to help you calibrate your submodule.
B
Sample Program for Calibration
The following sample C program will help you calibrate the VXI-SC-1102/B/C submodules. The calibration constants are stored only in NI-DAQ memory. You must enter the (input, output) voltage pairs read by a DMM by hand. Before running the sample program below, you must run the NI-DAQ Configuration Utility to configure the VXI-SC chassis and to configure the DAQ device that will communicate with the VXI-SC-1102/B/C.
The following sample program is written to work for a DAQ device, such as the VXI-MIO-64E-1, with a 5 V full-scale input:
#include <stdio.h> #include <math.h> #include <dataacq.h>
#define MIO_FULL_SCALE 5.0 /* valid for VXI-MIO-64E-1 */ #define N_VXI_SC_SLOTS 24 #define NIDAQMEM 0 #define ALL -1
int enterInt (char *prompt) {
short
ret;
int
value;
©
National Instruments Corporation B-1 VXI-SC-1102/B/C User Manual
Appendix B Calibration Sample Program
do {
fputs(prompt, stdout); fflush(stdin);
ret = scanf("%d", &value); } while (!ret); return (value);
}
double enterFloat (char *prompt) {
short
ret; double
value;
do {
fputs(prompt, stdout);
fflush(stdin);
ret = scanf("%lf", &value); } while (!ret); return (value);
}
void main() {
int
vxiChassisID, commPath, moduleSlot, /* slot of module to calibrate */ moduleChan;
short
dummyRangeCode = 0, dummyDAQboard, dummyDAQchan = 0, dummyDAQgain = 1, dummyDAQrange; /* These dummy variables would be
VXI-SC-1102/B/C User Manual B-2
used if the measurement actually came from a DAQ board and not an external DMM. */
©
National Instruments Corporation
Appendix B Calibration Sample Program
double
scale, gain, vInput1, vOutput1, vInput2, vOutput2, binOutput1, binOutput2, binOffset, gainerr, offset;
do
{
vxiChassisID = enterInt ("\nEnter chassis ID of the V XI-SC chassis:"); } while (vxiChassisID < 1); do {
moduleSlot = enterInt ("Enter slot of the VXI-SC-1102: "); } while (moduleSlot < 1 || N_VXI_SC_SLOTS < moduleSlot); do {
commPath = enterInt ("Enter device number of the DAQ board"
"controlling the chassis: "); } while (commPath < 0); dummyDAQboard = commPath;
dummyDAQrange = MIO_FULL_SCALE / dummyDAQgain; scale = pow(2.0, 12.0) * dummyDAQgain / 10;
/* factor for converting to format
of 12-bit bipolar DAQ board */
SCXI_Reset (vxiChassisID, moduleSlot);
SCXI_Single_Chan_Setup (vxiChassisID, moduleSlot, 0, dummyDAQboard);
/* This is necessary only so that
SCXI_Change_Chan won't return
an error. */ for (gain = 1; gain <= 100; gain = gain + 99) {
SCXI_Set_Gain (vxiChassisID, moduleSlot, ALL, gain);
/* Set the gain on all channels. */
for (moduleChan = 0; moduleChan < 32; moduleChan++)
©
National Instruments Corporation B-3 VXI-SC-1102/B/C User Manual
Appendix B Calibration Sample Program
{
SCXI_Change_Chan (vxiChassisID, moduleSlot, moduleChan);
printf("Apply input voltage for point 1, channel %d, gain %.0f.\n"
moduleChan, gain);,
vInput1 = enterFloat ("Enter VXI-SC-1102 input voltage: ");
vOutput1 = enterFloat ("Enter VXI-SC-1102 output voltage: ");
printf("Apply input voltage for point 2, channel %d, gain
%.0f.\n",moduleChan, gain);
vInput2 = enterFloat ("Enter VXI-SC-1102 input voltage: ");
vOutput2 = enterFloat ("Enter VXI-SC-1102 output voltage: ");
binOutput1= vOutput1*scale; binOutput2= vOutput2*scale; /* Convert to DAQ board's format. */
SCXI_Cal_Constants (vxiChassisID, moduleSlot, moduleChan,
offset = binOffset/scale; /* Convert from DAQ board format. */
printf("Calculated offset %f V, gain ratio (actual/ideal) %f"
"and stored in NI-DAQ memory.\n\n", offset, gainerr);
}
}
}
/* Select the channel. */
/* User enters vInput1. */
/* User enters vOutput1. */
/* User enters vInput2. */
/* User enters vOutput2. */
2, NIDAQMEM, dummyRangeCode, gain, dummyDAQboard, dummyDAQchan, dummyDAQgain, 1, vInput1, binOutput1, vInput2, binOutput2, &binOffset, &gainerr);
/* Calculate offset & gain error. */
VXI-SC-1102/B/C User Manual B-4
©
National Instruments Corporation
Appendix
Customer Communication
For your convenience, this appendix contains forms to help you gather the information necessary to help us solve your technical problems and a form you can use to comment on the product documentation. When you contact us, we need the information on the Technical Support Form and the configuration form, if your manual contains one, about your system configuration to answer your questions as quickly as possible.
National Instruments has technical assistance through electronic, fax, and telephone systems to quickly provide the information you need. Our electronic services include a bulletin board service, an FTP site, a fax-on-demand system, and e-mail support. If you have a hardware or software problem, first try the electronic support systems. If the information available on these systems does not answer your questions, we offer fax and telephon e support through our technical sup port centers, which are staffed by applications engineers.
C
Electronic Services
Bulletin Board Support
National Instruments has BBS and FTP sites dedicated for 24-hour support with a collection of files and documents to answer most common customer questions. From these sites, you can also download the latest instrument drivers, updates, and example programs. For recorded instructions on how to use the bulletin board and FTP services and for BBS automated information, call (512) 795-6990. You can access these services at:
United States: (512) 794-5422
Up to 14,400 baud, 8 data bits, 1 stop bit, no parity
United Kingdom: 01635 551422
Up to 9,600 baud, 8 data bits, 1 stop bit, no parity
France: 01 48 65 15 59
Up to 9,600 baud, 8 data bits, 1 stop bit, no parity
FTP Support
To access our FTP site, log on to our Internet host, Internet address, such as documents are located in the
©
National Instruments Corporation C-1 VXI-SC-1102/B/C User Manual
joesmith@anywhere.com, as your password. The support files and
/support directories.
ftp.natinst.com, as anonymous and use your
Fax-on-Demand Support
Fax-on-Demand is a 24-hour information retrieval system containing a library of documents on a wide range of technical information. You can access Fax-on-Demand from a touch-tone telephone at (512) 418-1111.
E-Mail Support (currently U.S. only
You can submit technical support questions to the applications engineering team through e-mail at the Internet address listed below. Remember to include your name, address, and phone number so we can contact you with solutions and suggestions.
support@natinst.com
)
Telephone and Fax Support
National Instruments has branch offices all over the world. Use the list below to find the technical support number for your country. If there is no National Instruments office in your country, contact the source from which you purchased your software to obtain support.
Telephone Fax
Australia 03 9879 5166 03 9879 6277 Austria 0662 45 79 90 0 0662 45 79 90 19 Belgium 02 757 00 20 02 757 03 11 Canada (Ontario) 905 785 0085 905 785 0086 Canada (Quebec) 514 694 8521 514 694 4399 Denmark 45 76 26 00 45 76 26 02 Finland 09 725 725 11 09 725 725 55 France 01 48 14 24 24 01 48 14 24 14 Germany 089 741 31 30 089 714 60 35 Hong Kong 2645 3186 2686 8505 Israel 03 5734815 03 5734816 Italy 02 413091 02 41309215 Japan 03 5472 2970 03 5472 2977 Korea 0 2 596 7456 02 596 7455 Mexico 5 520 2635 5 520 3282 Netherlands 0348 433466 0348 430673 Norway 32 84 84 00 32 84 86 00 Singapore 2265886 2265887 Spain 91 640 0085 91 640 0533 Sweden 08 730 49 70 08 730 43 70 Switzerland 056 200 51 51 056 200 51 55 Taiwan 02 377 1200 02 737 4644 United Kingdom 01635 523545 01635 523154 United States 512 795 8248 512 794 5678
Technical Support Form
Photocopy this form and update it each time you make changes to your software or hardware, and use the completed copy of this form as a reference for your current configuration. Completing this form accurately before contacting National Instruments for technical support helps our applications engineers answer your questions more efficiently.
If you are using any National Instruments hardware or software products related to this problem, include the configuration forms from their user manuals. Include additional pages if necessary.
Name ____________________________________________________________ ______________ Company _______________________________________________________________________ Address ________________________________________________________________________
_______________________________________________________________________________ Fax ( ___ )___________________ Phone ( ___ ) _______________________________________ Computer brand ________________ Model ________________ Processor___________________ Operating system (include version number) ____________________________________________ Clock speed ______MHz RAM _____MB Display adapter __________________________ Mouse ___yes ___no Other adapters installed _________________________ ___ ___________ Hard disk capacity _____MB Brand _____________________________________________ Instruments used _________________________________________________________________
_______________________________________________________________________________ National Instruments hardware product model __________ Revision ______________________ Configuration _______________________________________________________________ ____ National Instruments software product ____________________________ Version ____________ Configuration ___________________________________________________________________ The problem is: __________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________ List any error messages: ___________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________ The following steps reproduce the problem:____________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
VXI-SC-1102/B/C Submodule Hardware and Software Configuration Form
Record the settings and revisions of your hardware and software on the line to the right of each item. Complete a new copy of this form each time you revise your software or hardware configuration, and use this form as a reference for your current configuration. Completing this form accurately before contacting National Instruments for technical support helps our applications engineers answer your questions more efficiently.
National Instruments Products
Application Software and Version _____________________________________ _______________
Other Products
Computer make and model _________________________________________________________ Microprocessor ___________________________________ ________________________________ Clock frequency or speed ___________________________________________________________ Type of video board installed ________________________________________________ ________ Operating system version ___________________________________________________________ Operating system mode ____________________________________________________________ Programming language ____________________________________________________________ Programming language version ______________________________________________________ Other boards in system ____________________________________ _________________________ Base I/O address of other boards _____________________________________________________ DMA channels of other boards ______________________ ________________________________ Interrupt level of other boards _______________________________________________________
Documentation Comment Form
National Instruments encourages you to comment on the documentation supplied with our products. This information helps us provide quality products to meet your needs.
Title: VXI-SC-1102/B/C User Manual Edition Date: August 1997 Part Number: 321385B-01
Please comment on the completeness, clarity, and organization of the manual.
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
If you find errors in the manual, please record the page numbers and describe the errors.
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
_______________________________________________________________________________
Thank you for your help. Name _________________________________________________________________________ Title __________________________________________________________________________ Company _______________________________________________________________________ Address _________________________________________ _______________________________
_________________________________________ ______________________________________ Phone ( ___ )__________________________ Fax ( ___ ) ________ ________________________
Mail to: Technical Publications Fax to: Technical Publications
National Instruments Corporation National Instruments Corporation 6504 Bridge Point Parkway (512) 794-5678 Austin, TX 78730-5039
Prefix Meaning Value

Glossary

p- pico- 10 n- nano- 10 µ-micro-10
m- milli- 10
k- kilo- 10
M- mega- 10
G- giga- 10
Numbers/Symbols
+5 V +5 volt signal
° degrees ohms
% percent ± plus or minus
–12
–9
–6
–3
3
6
9
A
AC alternating current A/D analog-to-digital ADC analog-to-digital converter–an electronic device, often an integrated
circuit, that converts an analog voltage to a digital number
©
National Instruments Corporation G-1 VXI-SC-1102/B/C User Manual
Glossary
amplification a type of signal conditioning that improves accuracy in the resulting
digitized signal and reduces noise
API application programming interface
B
bus the group of conductors that interconnect individual circuitry in a
computer. Typically, a bus is the expansion vehicle to which I/O or other devices are connected. Examples of PC buses are the AT bus, NuBus, Micro Channel, and EISA bus.
C
C Celsius CGND chassis ground signal channel pin or wire lead to which you apply or from which you read the analog
or digital signal. Analog signals can be single-ended or differential. For digital signals, you group channels to form ports. Ports usually consist
of either four or eight digital channels. CH<0..31> - negative input channels 0 through 31 CH<0..31>+ positive input channels 0 through 31 CJC cold-junction compensation—a method of compensating for
inaccuracies in thermocouple circuits CJSENSOR cold-junction sensor
D
DAQ data acquisition—(1) collecting and measuring electrical signals from
sensors, transducers, and test probes or fixtures and inputting them to a
computer for processing; (2) collecting and measuring the same kinds
of electrical signals with A/D and/or DIO boards plugged into a
computer, and possibly generating control signals with D/ A and/or DIO
boards in the same computer DC direct current
©
VXI-SC-1102/B/C User Manual G-2
National Instruments Corporation
Glossary
DIFF differential configuration differential input an analog input consisting of two terminals, both of which are isolated
from computer ground, whose difference is measured
drivers/driver software software that controls a specific hardware device such as a DAQ board
F
filtering a type of signal conditioning that allows you to filter unwanted signals
from the signal you are trying to measure
FSR full-scale range
G
G gain—the factor by which a signal is amplified, sometimes expressed in
decibels
H
hex hexadecimal Hz hertz
I
in. inches input impedance the measured resistance and capacitance between the input terminals of
a circuit
instrument driver a set of high-level software functions that controls a specific VXI or
RS-232 programmable instrument or a specific plug-in DAQ board. Instrument drivers are available in several forms, ranging from a function callable language to a virtual instrument (VI) in LabVIEW.
©
National Instruments Corporation G-3 VXI-SC-1102/B/C User Manual
Glossary
L
logical address An 8-bit number that uniquely identifies each VXIbus device in a
system. It defines the A16 register addresses of a device, and indicates
Commander and Servant relationships. LSB least significant bit
M
mainframe The chassis of a VXIbus system that mechanically contains VXI
modules inserted into the backplane, ensuring that connectors fit
properly and that adjacent modules do not contact each other. It also
provides cooling airflow, and ensures that modules do not disengage
from the backplane due to vibration or shock. MB megabytes of memory MIO multifunction I/O MSB most significant bit mux multiplexer—a switching device with multiple inputs that sequentially
connects each of its inputs to its output, typically at high speeds, in
order to measure several signals with a single analog input channel
N
NC not connected (signal) noise an undesirable electrical signal—Noise comes from external sources
such as the AC power line, motors, generators, transformers,
fluorescent lights, soldering irons, CRT displays, computers, electrical
storms, welders, radio transmitters, and internal sources such as
semiconductors, resistors, and capacitors. Noise corrupts signals you
are trying to send or receive.
©
VXI-SC-1102/B/C User Manual G-4
National Instruments Corporation
Glossary
R
RMA Return Material Authorization rms root mean square
S
SCMP signal conditioning minipod signal conditioning the manipulation of signals to prepare them for digitizing slot A position where a module can be inserted into a VXIbus backplane.
Each slot provides the 96-pin J connectors to interface with the board P connectors. A slot can have one, two, or three connectors.
S/s samples per second
T
thermocouple a temperature sensor created by joining two dissimilar metals. The
junction produces a small voltage as a function of the temperature.
transducer excitation a type of signal conditioning that uses external voltages and currents
to excite the circuitry of a signal conditioning system into measuring physical phenomena
V
Vvolts VAC volts alternating current V
em
VDC volts, direct current V
diff
©
National Instruments Corporation G-5 VXI-SC-1102/B/C User Manual
common-mode voltage
differential voltage
Glossary
VI virtual instrument—(1) a combination of hardware and/or software
elements, typically used with a PC, that has the functionality of a classic
stand-alone instrument (2) a LabVIEW software module (VI), which
consists of a front panel user interface and a block diagram program V
IH
V
IL
V
in
volts, input high
volts, input low
volts in VISA a new driver software architecture developed by National Instruments
to unify instrumentation softwareGPIB, DAQ, and VXI. It has been
accepted as a standard for VXI by the VXIplug&play Systems Alliance. V
OH
V
OL
volts, output high
volts, output low VXIbus VMEbus eXtensions for Instrumentation VXIplug&play A group of VXI developers dedicated to making VXI devices as easy to
Systems Alliance use as possible, primarily by simplifying software development
W
waveform a time-varying physical phenomenon, often measured in voltage
VXI-SC-1102/B/C User Manual G-6
©
National Instruments Corporation

Index

Numbers
+5 V signal (table), 3-3
A
address handler, 4-2 analog circuitry, 4-3 to 4-4
analog input channels, 4-3 analog output, 4-4
analog input specifications, A-1 to A-4
amplifier characteristics, A-2 dynamic characteristics, A-2 to A-3 input characteristics, A-1 transfer characteristics, A-1 to A-2
analog output, 4-4
B
bulletin board support, C-1
C
calibration, 5-1 to 5-4
equipment requirements, 5-1 to 5-2 gain and offset calibration, 5-2 to 5-4 overview, 5-1
sample program, B-1 to B-4 CGND signal (table), 3-3 CH31- through CH0- signals (table), 3-3 CH31+ through CH0+ signals (table), 3-3
channel input signal connections, 3-3 to 3-4
common-mode input signal range,
3-3 to 3-4 differential input signal range, 3-3 exceeding input levels and ranges
(warning), 3-4 signal terminals, 3-3
Channel Register, 4-2 CJSENSOR signal
cold-junction sensor connection, 3-6 description (table), 3-3 exceeding overvoltage protection
(warning), 3-6
cold-junction sensor connection, 3-6 common-mode input signal range, 3-3 to 3-4 ComponentWorks application software, 1-4 Configuration Register, 4-2 current-loop receivers, installing, 2-3 customer communication, xii, C-1 to C-2
D
differential input signal range, 3-3 digital control circuitry, 4-2 to 4-3 documentation
conventions used in manual, x to xi National Instruments documentation,
xi to xii organization of manual, ix to x related documentation, xii
©
National Instruments Corporation I-1 VXI-SC-1102/B/C User Manual
Index
E
e-mail support, C-2 EEPROM Register, 4-2 electronic support services, customer
communication environment specifications, A-4 equipment, optional, 1-6
F
FaxBack support, C-2 filler panel, 1-2 filter specifications, A-3 floating signal connection
referenced to chassis ground (figure), 3-6 front connector pin assignments (figure), 3-2 FTP support, C-1
G
gain and offset calibration, 5-2 to 5-4 Gain Register, 4-2 ground-referencing signals, 3-4
floating signal connection
referenced to chassis ground (figure), 3-6
ground-referenced signal connection
(figure), 3-5
H
hardware installation. See installation.
I
input damage level, exceeding (warning), 3-4 installation, 2-1 to 2-4
current-loop receivers, 2-3 hardware installation, 2-4
parts locator diagram, 2-2 unpacking the VXI-SC-1102/B/C, 1-7
L
LabVIEW application software, 1-4 LabWindows/CVI application software, 1-4
M
manual. See documentation. Module ID Register, 4-2
N
NI-DAQ driver software, 1-5
P
parts locator diagram, 2-2 physical specifications, A-3 pins
front connector pin assignments
(figure), 3-2
front connector signal descriptions
(table), 3-3
power requirements, A-4
R
requirements for getting started, 1-3 resistors
bent and trimmed resistor (figure), 2-3 disconnecting signals before installing
(warning), 2-3
installing current-loop receivers, 2-3
VXI-SC-1102/B/C User Manual I-2
©
National Instruments Corporation
Index
S
scan interval, minimum, 4-3
settling time (note), 4-3
signal connections, 3-1 to 3-6
channel input signal connections,
3-3to3-4 cold-junction sensor connection, 3-6 front connector pin assignments
(figure), 3-2 ground-referencing signals, 3-4
floating signal connection
referenced to chassis ground (figure), 3-6
ground-referenced signal connection
(figure), 3-5 signal descriptions (table), 3-3 static electricity damage (caution), 3-1
signal terminals for input channels, 3-3 software programming choices, 1-4 to 1-6
National Instruments application
software, 1-4 to 1-6 NI-DAQ driver software, 1-5 VXIplug&play instrument driver, 1-6
specifications, A-1 to A-4
amplifier characteristics, A-2 dynamic characteristics, A-2 to A-3 environment, A-4 filters, A-3 input characteristics, A-1 physical, A-3 power requirements, A-4 stability, A-3 transfer characteristics, A-1 to A-2
stability specifications, A-3 Status Register, 4-2
T
technical support, customer communication telephone and fax support, C-2 temperature sensor, 4-3 theory of operation, 4-1 to 4-4
analog circuitry, 4-3 to 4-4
analog input channels, 4-3
analog output, 4-4 block diagram of VXI-SC-1102, 4-1 digital control circuitry, 4-2 to 4-3 functional overview, 4-1 to 4-2 VXIbus interface, 4-2
U
unpacking the VXI-SC-1102, 1-7
V
VirtualBench application software, 1-4 VXI-SC-1000 carrier module, 1-2 VXI-SC-1102/B/C. See also theory
of operation.
features, 1-2 to 1-3 optional equipment, 1-6 parts locator diagram, 2-2 requirements for getting started, 1-3 software programming choices, 1-4 to 1-6
National Instruments application
software, 1-4 NI-DAQ driver software, 1-5 VXIplug&play instrument
driver, 1-6
unpacking, 1-7 VXI signal conditioning system
(figure), 1-1 VXI signal conditioning system (figure), 1-1 VXIbus interface, 4-2 VXIplug&play instrument driver, 1-6
©
National Instruments Corporation I-3 VXI-SC-1102/B/C User Manual
Loading...