National Instruments SLSC-12201 User Manual

GETTING STARTED GUIDE
Digital Input/Output
32 Ch, 5V to 33V
SLSC-12201
01
(Ports 0 and 1)
(Ports 2 and 3)
1
16
31
1
16
31
SLSC-12201
32 Channel, 5 V to 33 V, Digital Input/Output
This document explains how to connect the SLSC-12201.
Note Before you begin, complete the software and hardware installation
procedures applicable to your application.
Note The guidelines in this document are specific to the SLSC-12201. The other
components in the system might not meet the same safety ratings. Refer to the documentation of each component in the system to determine the safety and EMC ratings for the entire system.

Safety Guidelines

Caution Observe all instructions and cautions in the user documentation. Using
the model in a manner not specified can damage the model and compromise the built-in safety protection. Return damaged models to NI for repair.
Attention Suivez toutes les instructions et respectez toutes les mises en garde de la
documentation utilisateur. L'utilisation d'un modèle de toute autre façon que celle spécifiée risque de l'endommager et de compromettre la protection de sécurité intégrée. Renvoyez les modèles endommagés à NI pour réparation.

Safety Voltages

Measurement category O
Isolation
Channel-to-channel None
Channel-to-earth ground None
Caution Do not connect the SLSC-12201 to signals or use for measurements
within Measurement Categories II, III, or IV.
Attention Ne connectez pas le SLSC-12201 à des signaux et ne l'utilisez pas pour
effectuer des mesures dans les catégories de mesure II, III ou IV.
Measurement Category I is for measurements performed on circuits not directly connected to the electrical distribution system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system that powers equipment. This category is for measurements of voltages from specially protected secondary circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of equipment, circuits powered by regulated low­voltage sources, and electronics.
Note Measurement Categories CAT I and CAT O are equivalent. These test and
measurement circuits are for other circuits not intended for direct connection to the MAINS building installations of Measurement Categories CAT II, CAT III, or CAT IV.

EMC Guidelines

This product was tested and complies with the regulatory requirements and limits for electromagnetic compatibility (EMC) stated in the product specifications. These requirements and limits provide reasonable protection against harmful interference when the product is operated in the intended operational electromagnetic environment.
2 | ni.com | SLSC-12201 Getting Started Guide
This product is intended for use in industrial locations. However, harmful interference may occur in some installations, when the product is connected to a peripheral device or test object, or if the product is used in residential or commercial areas. To minimize interference with radio and television reception and prevent unacceptable performance degradation, install and use this product in strict accordance with the instructions in the product documentation.
Furthermore, any changes or modifications to the product not expressly approved by National Instruments could void your authority to operate it under your local regulatory rules.
Notice To ensure the specified EMC performance, operate this product only with
shielded cables and accessories.
Notice To ensure the specified EMC performance, the length of all I/O cables must
be no longer than 3 m (10 ft).

Preparing the Environment

Ensure that the environment in which you are using the SLSC-12201 meets the following specifications.
Module operating temperature
0 °C to 85 °C
(IEC 60068-2-1, IEC 60068-2-2)
Operating humidity (IEC 60068-2-78) 10% RH to 90% RH, noncondensing
Pollution Degree 2
Maximum altitude 2,000 m
1
Indoor use only.
Note Refer to the device specifications on ni.com/manuals for complete
specifications.
1
The chassis internal ambient temperature may reach 85 °C with all slots at the maximum allowed power dissipation.
SLSC-12201 Getting Started Guide | © National Instruments | 3

Circuitry

ChannelEnable ChannelDirection
Control
Logic
InputThreshold
Px.y
GND
DIO(z)
GND
Rear I/O
Front I/O
SLSC-12201
Vsup_0
Vsup_1
VsupSelectPullupEnable
RangeEnable5V
Hardware Software Signal
Note Diagram only shows one channel.
All voltages are relative to GND unless otherwise noted.
Note You can configure the power-on configuration in the software. The factory
default power-on configuration sets the front I/O channels to sinking input and rear I/O channels to input.
4 | ni.com | SLSC-12201 Getting Started Guide

SLSC-12201 Pinout

GND
P0.3
NC
NC
GND
P0.7
NC
NC
GND
P1.3
NC
NC
GND
P1.7
P0.0
GND
P0.1
GND
GND
P0.2
NC
NC
GND
P0.4
GND
P0.5
GND
P0.6
NC
NC
GND
P1.0
GND
P1.1
GND
P1.2
NC
NC
GND
P1.4
GND
P1.5
GND
P1.6
15
30
14
29
13
28
12
27
11
26
10
25
9
24
8
23
7
22
6
21
5
20
4
19
3
18
2
17
1
16
44
43
42
41
40
39
38
37
36
35
34
33
32
31
J1
GND
P2.3
NC
NC
GND
P2.7
NC
NC
GND
P3.3
NC
NC
GND
P3.7
P2.0
GND
P2.1
GND
GND
P2.2
NC
NC
GND
P2.4
GND
P2.5
GND
P2.6
NC
NC
GND
P3.0
GND
P3.1
GND
P3.2
NC
NC
GND
P3.4
GND
P3.5
GND
P3.6
15
30
14
29
13
28
12
27
11
26
10
25
9
24
8
23
7
22
6
21
5
20
4
19
3
18
2
17
1
16
44
43
42
41
40
39
38
37
36
35
34
33
32
31
J2
Front Panel Pinout
Px.y Line y in Portx
GND Ground connection
NC No connection
Signal Description
Table 1. Front Panel Signal Descriptions
SLSC-12201 Getting Started Guide | © National Instruments | 5
XJ2 Connector Pinout
1525 11 1
edcba
Table 2. XJ2 Connector Pin Assignments
Row e d c b a
1 DIO3 DIO2 NC DIO1 DIO0
2 DIO7 DIO6 NC DIO5 DIO4
3 GND GND GND GND GND
4 DIO11 DIO10 NC DIO9 DIO8
5 DIO15 DIO14 NC DIO13 DIO12
6 GND GND GND GND GND
7 DIO19 DIO18 NC DIO17 DIO16
8 DIO23 DIO22 NC DIO21 DIO20
9 GND GND GND GND GND
10 DIO27 DIO26 NC DIO25 DIO24
6 | ni.com | SLSC-12201 Getting Started Guide
Table 2. XJ2 Connector Pin Assignments (Continued)
NC
NC
NC
NC
GND
Vsup_1
Vsup_0 GND
Row e d c b a
11 DIO31 DIO30 NC DIO29 DIO28
15 NC NC NC NC NC
16 NC NC NC NC NC
17 NC NC NC NC NC
18 NC NC NC NC NC
19 NC NC NC NC NC
20 NC NC NC NC NC
21 NC NC NC NC NC
22 NC NC NC NC NC
23 NC NC NC NC NC
24 NC NC NC NC NC
25 NC NC NC NC NC
Table 3. XJ2 Connector Signal Descriptions
Signal Description
DIO Digital input/output signal connection
GND Ground connection
NC No connection
XJ3 Connector Pinout
SLSC-12201 Getting Started Guide | © National Instruments | 7
Table 4. XJ3 Connector Signal Descriptions
DC Power Supply
DC Power Supply
FPGA
PXIe-7821R
Measurement System
RTI-12301
SLSC-12201
Device Under Test
P3.7 as Sourcing Output
Vsup
Vsup
Vsup
Power
DIO30
DIO31
DIO2
DIO1
DIO0
JR2 DIO30
JR2 DIO31
JR2 DIO2
JR2 DIO1
JR2 DIO0
JR3 Vsup_1
JR3 Vsup_0
Load
P3.6 as Sinking Output
P0.2 as Push-pull Output
P0.1 as Sourcing Input
P0.0 as Sinking Input
Signal Description
Vsup_<0, 1> Voltage supply connection for Bank 0 and Bank 1. You can select either for a
given bank.
GND Ground connection
NC No connection

Connecting Digital Devices to the SLSC-12201

The SLSC-12201 has 32 digital input/output channels that allow the direction to be configured for each channel. The 32 channels are organized into two banks of 16 channels each. Each bank can be powered from either of the Vsup connections. The banks are further organized into two ports of eight channels each. In the example shown below, the SLSC-12201 connects to the PXIe-7821R through the RTI-12301.

Short-Circuit Protection

Digital output channels are protected against short-circuit faults.
8 | ni.com | SLSC-12201 Getting Started Guide
Note The SLSC-12201 supports up to 10 channels in simultaneous short-circuit
fault.

Worldwide Support and Services

The NI website is your complete resource for technical support. At ni.com/support, you have access to everything from troubleshooting and application development self-help resources to email and phone assistance from NI Application Engineers.
Visit ni.com/services for information about the services NI offers.
Visit ni.com/register to register your NI product. Product registration facilitates technical support and ensures that you receive important information updates from NI.
NI corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504. NI also has offices located around the world. For support in the United States, create your service request at ni.com/support or dial 1 866 ASK MYNI (275 6964). For support outside the United States, visit the Worldwide Offices section of ni.com/niglobal to access the branch office websites, which provide up-to-date contact information.
SLSC-12201 Getting Started Guide | © National Instruments | 9
Information is subject to change without notice. Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.
© 2017—2019 National Instruments. All rights reserved.
377035C-01 July 25, 2019
Loading...