National Instruments SCXI-1520 User Manual

TM
SCXI

SCXI-1520 User Manual

SCXI-1520 User Manual
May 2009 372583E-01

Support

Worldwide Technical Support and Product Information

National Instruments Corporate Headquarters

11500 North Mopac Expressway Austin, Texas 78759-3504 USA Tel: 512 683 0100

Worldwide Offices

Australia 1800 300 800, Austria 43 662 457990-0, Belgium 32 (0) 2 757 0020, Brazil 55 11 3262 3599, Canada 800 433 3488, China 86 21 5050 9800, Czech Republic 420 224 235 774, Denmark 45 45 76 26 00, Finland 358 (0) 9 725 72511, France 01 57 66 24 24, Germany 49 89 7413130, India 91 80 41190000, Israel 972 3 6393737, Italy 39 02 41309277, Japan 0120-527196, Korea 82 02 3451 3400, Lebanon 961 (0) 1 33 28 28, Malaysia 1800 887710, Mexico 01 800 010 0793, Netherlands 31 (0) 348 433 466, New Zealand 0800 553 322, Norway 47 (0) 66 90 76 60, Poland 48 22 328 90 10, Portugal 351 210 311 210, Russia 7 495 783 6851, Singapore 1800 226 5886, Slovenia 386 3 425 42 00, South Africa 27 0 11 805 8197, Spain 34 91 640 0085, Sweden 46 (0) 8 587 895 00, Switzerland 41 56 2005151, Taiwan 886 02 2377 2222, Thailand 662 278 6777, Turkey 90 212 279 3031, United Kingdom 44 (0) 1635 523545
For further support information, refer to the Technical Support Information Document. To comment on National Instruments documentation, refer to the National Instruments Web site at
feedback.
ni.com/info and enter the info code
© 2001–2009 National Instruments Corporation. All rights reserved.

Important Information

Warranty

The SCXI-1520 is warranted against defects in materials and workmanship for a period of one year from the date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that proves to be defective during the warranty period. This warranty includes parts and labor.
The media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects in materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace software media that do not execute programming instruc tions if National Instruments receives notice of such defects during the warranty period. National Instruments does not warrant that the operation of the software shall be uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are covered by warranty.
National Instruments believes that the information in this document is accurate. The document has been carefully reviewed for technical accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent editions of this document without prior notice to holders of this edition. The reader should consult National Instruments if errors are suspected. In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.
XCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF
E
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. CUSTOMERS RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART OF NATIONAL
NSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE CUSTOMER. NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR DAMAGES RESULTING FROM
I
LOSS OF DATA, PROFITS, USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY THEREOF. This limitation of the
liability of National Instruments will apply regardless of the form of action, whether in contract or tort, including negligence. Any action against National Instruments must be brought within one year after the cause of action accrues. National Instruments shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not cover damages, defects, malfunctions, or service failures caused by owner’s failure to follow the National Instruments installation, operation, or maintenance instructions; owner’s modification of the product; owner’s abuse, misuse, or negligent acts; and power failure or surges, fire, flood, accident, actions of third parties, or other events outside reasonable control.

Copyright

Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical, including photocopying, recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National Instruments Corporation.
National Instruments respects the intellectual property of others, and we ask our users to do the same. NI software is protected by copyright and other intellectual property laws. Where NI software may be used to reproduce software or other materials belonging to others, you may use NI software only to reproduce materials that you may reproduce in accordance with the terms of any applicable license or other legal restriction.

Trademarks

National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation. Refer to the Terms of Use section
ni.com/legal for more information about National Instruments trademarks.
on
Other product and company names mentioned herein are trademarks or trade names of their respective companies.
Members of the National Instruments Alliance Partner Program are business entities independent from National Instruments and have no agency, partnership, or joint-venture relationship with National Instruments.

Patents

For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents.

WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS

(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING FOR A LEVEL OF RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL IMPLANTS OR AS CRITICAL COMPONENTS IN ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE TO PERFORM CAN REASONABLY BE EXPECTED TO CAUSE SIGNIFICANT INJURY TO A HUMAN.
(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE SOFTWARE PRODUCTS CAN BE IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT LIMITED TO FLUCTUATIONS IN ELECTRICAL POWER SUPPLY, COMPUTER HARDWARE MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE FITNESS, FITNESS OF COMPILERS AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION, INSTALLATION ERRORS, SOFTWARE AND HARDWARE COMPATIBILITY PROBLEMS, MALFUNCTIONS OR FAILURES OF ELECTRONIC MONITORING OR CONTROL DEVICES, TRANSIENT FAILURES OF ELECTRONIC SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR MISUSES, OR ERRORS ON THE PART OF THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS THESE ARE HEREAFTER COLLECTIVELY TERMED “SYSTEM FAILURES”). ANY APPLICATION WHERE A SYSTEM FAILURE WOULD CREATE A RISK OF HARM TO PROPERTY OR PERSONS (INCLUDING THE RISK OF BODILY INJURY AND DEATH) SHOULD NOT BE RELIANT SOLELY UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM FAILURE. TO AVOID DAMAGE, INJURY, OR DEATH, THE USER OR APPLICATION DESIGNER MUST TAKE REASONABLY PRUDENT STEPS TO PROTECT AGAINST SYSTEM FAILURES, INCLUDING BUT NOT LIMITED TO BACK-UP OR SHUT DOWN MECHANISMS. BECAUSE EACH END-USER SYSTEM IS CUSTOMIZED AND DIFFERS FROM NATIONAL INSTRUMENTS' TESTING PLATFORMS AND BECAUSE A USER OR APPLICATION DESIGNER MAY USE NATIONAL INSTRUMENTS PRODUCTS IN COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT EVALUATED OR CONTEMPLATED BY NATIONAL INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS ULTIMATELY RESPONSIBLE FOR VERIFYING AND VALIDATING THE SUITABILITY OF NATIONAL INSTRUMENTS PRODUCTS WHENEVER NATIONAL INSTRUMENTS PRODUCTS ARE INCORPORATED IN A SYSTEM OR APPLICATION, INCLUDING, WITHOUT LIMITATION, THE APPROPRIATE DESIGN, PROCESS AND SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.

Conventions

The following conventions are used in this manual:
<> Angle brackets that contain numbers separated by an ellipsis represent a
range of values associated with a bit or signal name—for example, AO <3. .0>.
» The » symbol leads you through nested menu items and dialog box options
to a final action. The sequence File»Page Setup»Options directs you to pull down the File menu, select the Page Setup item, and select Options from the last dialog box.
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to avoid injury, data loss, or a system crash. When this symbol is marked on a product, refer to the Read Me First: Safety and Radio-Frequency
Interference for information about precautions to take.
bold Bold text denotes items that you must select or click in the software, such
as menu items and dialog box options. Bold text also denotes parameter names.
italic Italic text denotes variables, emphasis, a cross-reference, or an introduction
to a key concept. Italic text also denotes text that is a placeholder for a word or value that you must supply.
monospace Text in this font denotes text or characters that you should enter from the
keyboard, sections of code, programming examples, and syntax examples. This font is also used for the proper names of disk drives, paths, directories, programs, subprograms, subroutines, device names, functions, operations, variables, filenames, and extensions.
monospace bold Bold text in this font denotes the messages and responses that the computer
automatically prints to the screen. This font also emphasizes lines of code that are different from the other examples.
monospace italic
Italic text in this font denotes text that is a placeholder for a word or value that you must supply.

Contents

Chapter 1 About the SCXI-1520
What You Need to Get Started ......................................................................................1-1
National Instruments Documentation ............................................................................1-3
Installing Application Software, NI-DAQ, and the E/M Series DAQ Device .............. 1-5
Installing the SCXI-1520 Module into the SCXI Chassis...............................1-5
Connecting the SCXI-1520 in an SCXI Chassis to an E/M Series
DAQ Device for Multiplexed Scanning .......................................................1-5
Connecting the SCXI-1520 in a PXI/SCXI Combination Chassis to an
E/M Series DAQ Device for Multiplexed Scanning ....................................1-5
Verifying the SCXI-1520 Installation in Software ........................................................1-6
Installing SCXI Using NI-DAQmx in Software .............................................1-6
Manually Adding Modules in NI-DAQmx .....................................................1-6
Installing SCXI Using Traditional NI-DAQ (Legacy) in Software ................ 1-6
Manually Adding Modules in Traditional NI-DAQ (Legacy) ........................1-6
Verifying and Self-Testing the Installation .....................................................1-6
Troubleshooting the Self-Test Verification ...................................................................1-7
Troubleshooting in NI-DAQmx ......................................................................1-7
Troubleshooting in Traditional NI-DAQ (Legacy) ......................................... 1-8
Chapter 2 Connecting Signals
Connecting Bridge Sensor Signals ................................................................................ 2-1
Quarter-Bridge Type I .....................................................................................2-1
Quarter-Bridge Type II....................................................................................2-2
Half-Bridge Type I ..........................................................................................2-4
Half-Bridge Type II.........................................................................................2-5
Full-Bridge Type I...........................................................................................2-6
Full-Bridge Type II.......................................................................................... 2-7
Full-Bridge Type III ........................................................................................2-8
Remote Sense ..................................................................................................2-9
Pin Assignments ............................................................................................................2-10
© National Instruments Corporation v SCXI-1520 User Manual
Contents
Chapter 3 Configuring and Testing
SCXI-1520 Software-Configurable Settings .................................................................3-1
Common Software-Configurable Settings ...................................................... 3-1
Bridge Configuration........................................................................ 3-1
Excitation Level................................................................................ 3-2
Filter Bandwidth ............................................................................... 3-2
Gain/Input Range.............................................................................. 3-3
Null Potentiometers .......................................................................... 3-4
Shunt Calibration Switches .............................................................. 3-4
Modes of Operation .......................................................................... 3-5
Simultaneous Sample and Hold........................................................ 3-5
Configurable Settings in MAX...................................................................................... 3-5
NI-DAQmx ..................................................................................................... 3-6
Creating a Strain Global Channel or Task........................................ 3-7
Creating a Custom Voltage with Excitation Global Channel
or Task ........................................................................................... 3-8
Traditional NI-DAQ (Legacy) ........................................................................ 3-9
Configuring Module Property Pages in Traditional NI-DAQ
(Legacy)......................................................................................... 3-10
Creating a Strain Virtual Channel .................................................... 3-11
Calibrating a Strain Virtual Channel ................................................ 3-11
Verifying the Signal ...................................................................................................... 3-12
Verifying the Signal in NI-DAQmx Using a Task or Global Channel........... 3-12
Verifying the Signal in Traditional NI-DAQ (Legacy) .................................. 3-13
Verifying the Signal Using Channel Strings .................................... 3-13
Verifying the Signal Using Strain Virtual Channel .......................... 3-14
Using the Strain Calibration Wizard in NI-DAQmx..................................................... 3-14
Chapter 4 Theory of Operation
Strain-Gauge Theory ..................................................................................................... 4-1
Wheatstone Bridges ........................................................................................ 4-1
Strain Gauges .................................................................................................. 4-2
Acronyms, Formulas, and Variable Definitions ............................................. 4-3
Software Scaling and Equations ..................................................................... 4-4
Quarter-Bridge Type I..................................................................................... 4-4
Quarter-Bridge Type II ................................................................................... 4-6
Half-Bridge Type I.......................................................................................... 4-9
Half-Bridge Type II ........................................................................................ 4-11
Full-Bridge Type I .......................................................................................... 4-13
SCXI-1520 User Manual vi ni.com
Full-Bridge Type II.......................................................................................... 4-14
Full-Bridge Type III ........................................................................................4-16
SCXI-1520 Theory of Operation ...................................................................................4-18
Bridge Configuration and Completion ............................................................4-21
Excitation.........................................................................................................4-22
Remote Sense....................................................................................4-23
Gain ................................................................................................................. 4-24
Filter Bandwidth and Cutoff Frequency..........................................................4-25
Offset Null Compensation...............................................................................4-25
Shunt Calibration.............................................................................................4-27
Simultaneous Sample and Hold.......................................................................4-28
Maximum Simultaneous Sample and Hold Sample Rate Using
Maximum Simultaneous Sample and Hold Using Traditional
Modes of Operation......................................................................................... 4-34
Theory of Multiplexed Mode Operation ...........................................4-34
Theory of Parallel Mode Operation ..................................................4-35
Chapter 5 Using the SCXI-1520
Developing Your Application in NI-DAQmx ...............................................................5-1
Typical Program Flowchart.............................................................................5-2
General Discussion of Typical Flowchart .......................................................5-4
Creating a Task Using DAQ Assistant or Programmatically ...........5-4
Adjusting Timing and Triggering ..................................................... 5-4
Configuring Channel Properties........................................................5-5
Performing Offset Null Compensation .............................................5-7
Performing Shunt Calibration ...........................................................5-8
Acquiring, Analyzing, and Presenting ..............................................5-8
Completing the Application .............................................................. 5-9
Developing an Application Using LabVIEW..................................................5-9
Using a DAQmx Channel Property Node in LabVIEW ...................5-11
Specifying Channel Strings in NI-DAQmx.....................................................5-12
Text Based ADEs..............................................................................5-14
Measurement Studio (Visual Basic, .NET, and C#)........................................5-14
Programmable NI-DAQmx Properties..............................................5-16
Developing Your Application in Traditional NI-DAQ (Legacy) ..................................5-18
Traditional NI-DAQ (Legacy) in LabVIEW................................................... 5-19
Typical Program Flow ......................................................................5-20
Configuring the SCXI-1520 Settings Using Traditional NI-DAQ
(Legacy) in LabVIEW ..................................................................................5-22
Contents
NI-DAQmx ....................................................................................4-30
NI-DAQ (Legacy) ..........................................................................4-32
© National Instruments Corporation vii SCXI-1520 User Manual
Contents
Performing Offset Null Compensation Using Traditional NI-DAQ
(Legacy) in LabVIEW.................................................................................. 5-24
Performing Shunt Calibration Using Traditional NI-DAQ (Legacy)
in LabVIEW ................................................................................................. 5-26
Configure, Start Acquisition, and Take Readings Using Traditional
NI-DAQ (Legacy) in LabVIEW .................................................................. 5-27
Converting Scaling Using Traditional NI-DAQ (Legacy) in LabVIEW ........ 5-27
Analyze and Display Using Traditional NI-DAQ (Legacy) in LabVIEW ..... 5-28
Traditional NI-DAQ (Legacy) in Text-Based ADEs ...................................... 5-28
Low-Level DAQ Functions ............................................................................ 5-29
Configuring System Settings Using Traditional NI-DAQ (Legacy) C API ... 5-31 Configuring Module Settings Using Traditional NI-DAQ (Legacy) C API... 5-32 Performing Offset Null Compensation Using Traditional NI-DAQ
(Legacy) C API ............................................................................................ 5-34
Performing Shunt Calibration Using Traditional NI-DAQ
(Legacy) C API ............................................................................................ 5-34
Performing Acquisition Using Traditional NI-DAQ (Legacy) C API............ 5-35
Performing Scaling, Analysis, and Display .................................................... 5-35
Other Application Documentation and Material ........................................................... 5-36
Traditional NI-DAQ (Legacy) CVI Examples................................................ 5-36
Traditional NI-DAQ (Legacy) Measurement Studio Examples ..................... 5-36
Calibrating the Strain System........................................................................................ 5-37
Calibrating the SCXI-1520 ............................................................................. 5-37
Internal Calibration Procedure.......................................................... 5-37
Internal Calibration Using LabVIEW............................................... 5-37
Internal Calibration Using a C-Based ADE ..................................... 5-38
External Calibration.......................................................................... 5-38
Calibrating the System .................................................................................... 5-39
Offset Null Compensation ................................................................ 5-39
Shunt Calibration.............................................................................. 5-39
Appendix A Specifications
Appendix B Using SCXI Channel Strings with Traditional NI-DAQ (Legacy) 7.0 or Later
Appendix C Removing the SCXI-1520
SCXI-1520 User Manual viii ni.com
Appendix D Common Questions
Glossary
Index
Figures
Figure 2-1. Quarter-Bridge I Circuit Diagram .........................................................2-2
Figure 2-2. Quarter-Bridge II Circuit Diagram........................................................2-3
Figure 2-3. Half-Bridge Type I Circuit Diagram .....................................................2-4
Figure 2-4. Half-Bridge Type II Circuit Diagram....................................................2-5
Figure 2-5. Full-Bridge Type I Circuit Diagram......................................................2-6
Figure 2-6. Full-Bridge Type II Circuit Diagram ....................................................2-7
Figure 2-7. Full-Bridge Type III Circuit Diagram ...................................................2-8
Figure 2-8. Remote-Sense Circuit Diagram.............................................................2-9
Figure 4-1. Basic Wheatstone Bridge Circuit Diagram ...........................................4-1
Figure 4-2. Quarter-Bridge Type I Measuring Axial and Bending Strain ...............4-4
Figure 4-3. Quarter-Bridge I Circuit Diagram .........................................................4-5
Figure 4-4. Quarter-Bridge Type II Measuring Axial and Bending Strain..............4-7
Figure 4-5. Quarter-Bridge II Circuit Diagram........................................................4-7
Figure 4-6. Half-Bridge Type I Measuring Axial and Bending Strain ....................4-9
Figure 4-7. Half-Bridge Type I Circuit Diagram .....................................................4-9
Figure 4-8. Half-Bridge Type II Rejecting Axial and
Measuring Bending Strain.....................................................................4-11
Figure 4-9. Half-Bridge Type II Circuit Diagram....................................................4-11
Figure 4-10. Full-Bridge Type I Rejecting Axial and Measuring Bending Strain..... 4-13
Figure 4-11. Full-Bridge Type I Circuit Diagram......................................................4-13
Figure 4-12. Full-Bridge Type II Rejecting Axial and Measuring Bending Strain ...4-15
Figure 4-13. Full-Bridge Type II Circuit Diagram ....................................................4-15
Figure 4-14. Full-Bridge Type III Measuring Axial and
Rejecting Bending Strain.......................................................................4-17
Figure 4-15. Full-Bridge Type III Circuit Diagram ...................................................4-17
Figure 4-16. Block Diagram of SCXI-1314/SCXI-1520 Combination .....................4-19
Figure 4-17. Signal During Simultaneous Sample-and-Hold Sampling ....................4-29
Contents
Figure 5-1. Typical Program Flowchart...................................................................5-3
Figure 5-2. LabVIEW Channel Property Node with Filtering Enabled
at 10 kHz and SS/H Disabled................................................................5-12
Figure 5-3. Typical SCXI-1520 Program Flow with
Traditional NI-DAQ (Legacy)...............................................................5-21
© National Instruments Corporation ix SCXI-1520 User Manual
Contents
Tables
Figure 5-4. Using the AI Parameter VI to Set Up the SCXI-1520 .......................... 5-24
Figure 5-5. Offset Null and Shunt Calibration Flowchart ....................................... 5-30
Figure C-1. Removing the SCXI-1520..................................................................... C-2
Table 1-1. Accessories Available for the SCXI-1520............................................ 1-2
Table 2-1. Front Signal Pin Assignments ..............................................................2-11
Table 2-2. Rear Signal Pin Assignments................................................................ 2-13
Table 2-3. SCXI-1520 Communication Signals..................................................... 2-14
Table 3-1. Excitation Voltage for Configuration and Gauge Resistances ............. 3-2
Table 4-1. Strain-Gauge Configurations ................................................................4-2
Table 4-2. Control Codes for Coarse and Fine Offset Null Potentiometers........... 4-26
Table 4-3. NI-DAQmx Values Used to Determine Maximum Sample Rate
in Multiplexed Mode............................................................................. 4-31
Table 4-4. NI-DAQmx Values Used to Determine Maximum Sample Rate
in Parallel Mode.................................................................................... 4-32
Table 4-5. Traditional NI-DAQ (Legacy) Values Used to Determine
Maximum Sample Rate in Multiplexed Mode...................................... 4-33
Table 5-1. NI-DAQmx Properties .......................................................................... 5-5
Table 5-2. Programming a Task in LabVIEW ....................................................... 5-9
Table 5-3. NI-DAQmx Properties .......................................................................... 5-16
Table 5-4. Settings for Configuring the SCXI-1520 Through the AI Parameter ...5-22
Table 5-5. Configuration Functions ....................................................................... 5-31
Table 5-6. NI-DAQ Functions Used to Configure SCXI-1520.............................. 5-33
Table D-1. Digital Signals on the SCXI-1520 ....................................................... D-2
SCXI-1520 User Manual x ni.com
About the SCXI-1520
The SCXI-1520 module is an eight-channel module for interfacing to strain-gauge bridges and other Wheatstone-bridge based sensors.
Note Descriptions in this chapter explicitly refer to the first channel (channel 0); however,
the same descriptions are applicable to channels <1..7>.
You can configure all settings on a per channel basis in software. The SCXI-1520 is configured using Measurement & Automation Explorer (MAX) or through NI-DAQmx property nodes.
With the SCXI-1520 and the accessory SCXI-1314 terminal block, you can do the following:
Connect sensors of all bridge configurations, including quarter-, half-, and full-bridge
Set the DC voltage excitation between 0 and 10 V (increments dependent upon the driver software)
Programmatically offset null bridge circuits connected to the SCXI-1520
Set the analog input lowpass filter cut-off frequency to 10 Hz, 100 Hz, 1 kHz, 10 kHz, or bypass
Set the analog input gain between 1 and 1000 at any one of 49 settings
Implement shunt calibration using two independent circuits
Connect the bridge for remote-sense voltage excitation
1

What You Need to Get Started

To set up and use the SCXI-1520, you need the following:
Hardware
SCXI-1520 module
One of the following terminal blocks:
SCXI-1314 terminal block
SCXI-1314T terminal block
© National Instruments Corporation 1-1 SCXI-1520 User Manual
Chapter 1 About the SCXI-1520
SCXI or PXI/SCXI combo chassis
E/M Series DAQ device
Computer, if using an SCXI chassis
Cabling, cable adapter, and sensors as required for your
application
Software
NI-DAQ 7.0 or later
Application software, such as LabVIEW, LabWindows
Measurement Studio, or other programming environments
Documentation
Read Me First: Safety and Radio-Frequency Interference
DAQ Getting Started Guide
SCXI Quick Start Guide
SCXI-1520 User Manual
Documentation for your hardware
Documentation for your software
/CVI™,
The optional accessories listed in Table 1-1 are available for the SCXI-1520.

Table 1-1. Accessories Available for the SCXI-1520

Accessory Description
SCXI-1314
Screw terminal block—Mounts on the front of the SCXI-1520 module. It includes connections and sockets for two shunt calibration resistors and a quarter-bridge completion resistor per channel.
SCXI-1314T
††
RJ-50 terminal block—Mounts on the front of the SCXI-1520 module. It features eight RJ-50 10-position/10-conductor (10p10c) modular plugs for connection to hardware TEDS smart sensors.
SCXI-1520 User Manual 1-2 ni.com
Chapter 1 About the SCXI-1520
Table 1-1. Accessories Available for the SCXI-1520 (Continued)
Accessory Description
SCXI-1310
Connector and shell assembly—The SCXI-1310 provides 96 eyelet-type terminals for easy hook-and-solder signal connection and custom mass termination connectivity.
TBX-96
DIN-rail mounted terminal block with 96 generic screw terminals.
One of the following cables is required to connect the TBX-96 to an SCXI module:
SH96-96 shielded cable, 1 m
R96-96 unshielded ribbon cable, 1 m
You must wire a shunt resistor between the pins that correspond to the SCA on the SCXI-1314 and the appropriate legs of
the bridge sensor. Refer to Chapter 4, Theory of Operation, for more information.
††
The SCXI-1314T only supports SCA. It does not support SCB. There are no quarter-bridge completion resistors in the SCXI-1314T. If you are connecting a quarter-bridge sensor to an SCXI-1520 using an SCXI-1314T, you must place an external resistor between the pins that correspond to PX– and SX+ on the terminals. Refer to Chapter 2, Connecting Signals, for more information.
There are no quarter-bridge completion resistors in the SCXI-1310 or TBX-96. If you are connecting a quarter-bridge sensor to an SCXI-1520 using an SCXI-1310 or TBX-96, you must place an external resistor between the pins that correspond to PX– and SX+ on the terminals. Refer to Chapter 2, Connecting Signals, for more information.

National Instruments Documentation

The SCXI-1520 User Manual is one piece of the documentation set for data acquisition (DAQ) systems. You could have any of several types of manuals depending on the hardware and software in the system. Use the manuals you have as follows:
Getting Started with SCXI—This is the first manual you should read. It gives an overview of the SCXI system and contains the most commonly needed information for the modules, chassis, and software.
SCXI or PXI/SCXI chassis manual—Read this manual for maintenance information on the chassis and for installation instructions.
The DAQ Getting Started Guide—This document has information on installing NI-DAQ and the E/M Series DAQ device. Install these before you install the SCXI module.
The SCXI Quick Start Guide—This document contains a quick overview for setting up an SCXI chassis, installing SCXI modules and terminal blocks, and attaching sensors. It also describes setting up the SCXI system in MAX.
© National Instruments Corporation 1-3 SCXI-1520 User Manual
Chapter 1 About the SCXI-1520
The SCXI hardware user manuals—Read these manuals for detailed information about signal connections and module configuration. They also explain, in greater detail, how the module works and contain application hints.
Accessory installation guides or manuals—Read the terminal block and cable assembly installation guides. They explain how to physically connect the relevant pieces of the system. Consult these guides when you are making the connections.
The E/M Series DAQ device documentation—This documentation has detailed information about the DAQ device that plugs into or is connected to the computer. Use this documentation for hardware installation and configuration instructions, specification information about the DAQ device, and application hints.
Software documentation—You may have both application software and NI-DAQ software documentation. National Instruments (NI) application software includes LabVIEW, LabWindows/CVI, and Measurement Studio. After you set up the hardware system, use either your application software documentation or the NI-DAQ documentation to help you write your application. If you have a large, complex system, it is worthwhile to look through the software documentation before you configure the hardware.
One or more of the following help files for software information:
Start»Programs»National Instruments»NI-DAQ»
NI-DAQmx Help
Start»Programs»National Instruments»NI-DAQ»
Traditional NI-DAQ User Manual
Start»Programs»National Instruments»NI-DAQ»
Traditional NI-DAQ Function Reference Help
NI strain-gauge application notes or tutorials—NI has additional material about strain gauges and strain measurements available at
ni.com/support.
You can download NI documents from the latest version of NI-DAQ, click Download Software at
SCXI-1520 User Manual 1-4 ni.com
ni.com/manuals. To download
ni.com.
Chapter 1 About the SCXI-1520

Installing Application Software, NI-DAQ, and the E/M Series DAQ Device

Refer to the DAQ Getting Started Guide packaged with the NI-DAQ software to install your application software, NI-DAQ driver software, and the DAQ device to which you will connect the SCXI-1520. NI-DAQ 7.0 or later is required to configure and program the SCXI-1520 module. If you do not have NI-DAQ 7.0 or later, you can either contact a NI sales representative to request it on a CD or download the latest NI-DAQ version
ni.com.
from
Note Refer to the Read Me First: Radio-Frequency Interference document before
removing equipment covers or connecting or disconnecting any signal wires.

Installing the SCXI-1520 Module into the SCXI Chassis

Refer to the SCXI Quick Start Guide to install your SCXI-1520 module.

Connecting the SCXI-1520 in an SCXI Chassis to an E/M Series DAQ Device for Multiplexed Scanning

Refer to the SCXI Quick Start Guide to install the cable adapter and connect the SCXI modules to the DAQ device.
If you have already installed the appropriate software, refer to Chapter 3,
Configuring and Testing, to configure the SCXI-1520 module(s).

Connecting the SCXI-1520 in a PXI/SCXI Combination Chassis to an E/M Series DAQ Device for Multiplexed Scanning

Refer to the SCXI Quick Start Guide to connect the SCXI modules to the DAQ device.
If you have already installed the appropriate software, refer to Chapter 3,
Configuring and Testing, to configure the SCXI-1520 module(s).
© National Instruments Corporation 1-5 SCXI-1520 User Manual
Chapter 1 About the SCXI-1520

Verifying the SCXI-1520 Installation in Software

Refer to the SCXI Quick Start Guide for information on verifying the SCXI installation.

Installing SCXI Using NI-DAQmx in Software

Refer to the SCXI Quick Start Guide for information on installing modules using NI-DAQmx in software.

Manually Adding Modules in NI-DAQmx

If you did not auto-detect the SCXI modules, you must manually add each of the modules. Refer to the SCXI Quick Start Guide to manually add modules.
Note NI recommends auto-detecting modules for the first time configuration of the
chassis.

Installing SCXI Using Traditional NI-DAQ (Legacy) in Software

Refer to the SCXI Quick Start Guide for information on installing modules using Traditional NI-DAQ (Legacy) in software.

Manually Adding Modules in Traditional NI-DAQ (Legacy)

If you did not auto-detect the SCXI modules, you must manually add each of the modules. Refer to the SCXI Quick Start Guide to manually add modules.
Note NI recommends auto-detecting modules for the first time configuration of the
chassis.

Verifying and Self-Testing the Installation

The verification procedure for the SCXI chassis is the same for both NI-DAQmx and Traditional NI-DAQ (Legacy). To test the successful installation for the SCXI chassis, refer to the SCXI Quick Start Guide. Verify that the chassis is powered on and correctly connected to an E/M Series DAQ device.
SCXI-1520 User Manual 1-6 ni.com
Chapter 1 About the SCXI-1520
After verifying and self-testing the installation, the SCXI system should operate properly with your ADE software. If the test did not complete successfully, refer to Chapter 3, Configuring and Testing, for troubleshooting steps.

Troubleshooting the Self-Test Verification

If the Self-Test Verification did not verify the chassis configuration, complete the steps in this section to troubleshoot the SCXI configuration.

Troubleshooting in NI-DAQmx

If you get a Verify SCXI Chassis message box showing the SCXI
chassis model number, Chassis ID: x, and one or more messages stating Slot Number: x Configuration has module: SCXI-XXXX or 1520, hardware in chassis is: Empty, take the following troubleshooting actions:
Make sure the SCXI chassis is powered on.
Make sure all SCXI modules are properly installed in the chassis.
Refer to the SCXI Quick Start Guide for proper installation instructions.
Make sure the cable between the SCXI chassis and E/M Series
DAQ device is properly connected.
Inspect the cable connectors for bent pins.
Make sure you are using the correct NI cable assembly.
Test the DAQ device to verify it is working properly. Refer to the
DAQ device help file for more information.
If you get a Verify SCXI Chassis message box showing the SCXI
chassis model number,
Number:
hardware in chassis is: SCXI-
complete the following troubleshooting steps to correct the error.
1. Expand the list of NI-DAQmx devices by clicking the + next to
2. Right-click the SCXI chassis and click Properties to load the
3. Under the Modules tab, ensure that the cabled module is listed in
x
Configuration has module: SCXI-
NI-DAQmx Devices.
chassis configurator.
the correct slot.
Chassis ID:
x
, and the message Slot
YYYY
XXXX
or 1520,
, 1520, or Empty,
© National Instruments Corporation 1-7 SCXI-1520 User Manual
Chapter 1 About the SCXI-1520
4. If the cabled module is not listed in the correct slot, complete the
following troubleshooting steps:
a. If the cabled module is not listed in the correct slot and the
slot is empty, click the drop-down listbox next to the correct slot and select the cabled module. Configure the cabled module following the steps listed in the SCXI Quick Start Guide. Click OK.
b. If another module appears where the cabled module should
be, click the drop-down listbox next to the correct slot and select the cabled module. A message box appears asking you to confirm the module replacement. Click OK. Configure the cabled module following the steps listed in the SCXI Quick Start Guide. Click OK.
Ensure that you have the highest priority SCXI module cabled to the E/M Series DAQ device. Refer to the SCXI Quick Start Guide to find out which SCXI module in the chassis should be cabled to the DAQ device.
After checking the preceding items, return to the Troubleshooting the
Self-Test Verification section and retest the SCXI chassis.
If these measures do not successfully configure the SCXI system, contact NI. Refer to the Technical Support Information document for contact information.

Troubleshooting in Traditional NI-DAQ (Legacy)

If you get the message Unable to test chassis at this time, you have not designated at least one module as connected to a E Series DAQ device. Refer to the Traditional NI-DAQ (Legacy) section of Chapter 3, Configuring and Testing, the cabled module in the system from Connected to: None to Connected to: Device x.
If you get the message codes and the message
Failed to find followed by the module Unable to communicate with chassis,
take the following troubleshooting actions:
Make sure the SCXI chassis is powered on.
Make sure the cable between the SCXI chassis and E Series DAQ
device is properly connected.
Inspect the cable connectors for bent pins.
Make sure you are using the correct NI cable assembly.
Test the DAQ device to verify it is working properly. Refer to the
DAQ device help file for more information.
SCXI-1520 User Manual 1-8 ni.com
and change the configuration of
Chapter 1 About the SCXI-1520
If you get the message Failed to find, followed by module codes and the message
Instead found: module with ID 0Xxx, refer
to the Traditional NI-DAQ (Legacy) section of Chapter 3, Configuring
and Testing, and make sure the correct module is in the specified slot.
Delete the incorrect module as described in Appendix C, Removing the
SCXI-1520, and add the correct module as described in the Traditional NI-DAQ (Legacy) section of Chapter 3, Configuring and Testing.
If you get the message and the message
Failed to find, followed by a module code
Slot x is empty, make sure the configured module
is installed in the specified slot. If not, install the module by following the instructions in the Installing the SCXI-1520 Module into the SCXI
Chassis section. If the module is installed in the correct slot, power off
the chassis, remove the module as specified in Appendix C, Removing
the SCXI-1520, and verify that no connector pins are bent on the rear
signal connector. Reinstall the module as described in the Installing the
SCXI-1520 Module into the SCXI Chassis section, ensuring
the module is fully inserted and properly aligned in the slot.
After checking the preceding items, return to the Troubleshooting the
Self-Test Verification section and retest the SCXI chassis.
If these measures do not successfully configure the SCXI system, contact NI. Refer to the Technical Support Information document for contact information.
© National Instruments Corporation 1-9 SCXI-1520 User Manual
Connecting Signals
This chapter describes how to connect Wheatstone bridge sensors to the SCXI-1520 in quarter-, half-, and full-bridge configurations and for remote sensing. It also provides the front and rear signal pin assignments of the module.

Connecting Bridge Sensor Signals

This section discusses how to connect the signals of supported strain-gauge configuration types as well as full-bridge sensors such as load, force, torque, and pressure sensors. It also discusses connecting leads for remote sensing and shunt calibration. Refer to Chapter 4, Theory of Operation, for a discussion of strain-gauge concepts. Refer to the SCXI-1314 Installation Guide for more signal connection information.
Notes The circuits in this section illustrate circuits using the SCXI-1314 terminal block.
If you are using the SCXI-1314T terminal block, refer to the SCXI-1314T TEDS Bridge Sensor Terminal Block Installation Guide for the permitted circuit configuration diagrams.
2
Refer to Figure 2-5 and use its wiring diagram for full-bridge sensors such as load, force, torque, and pressure sensors.

Quarter-Bridge Type I

This section provides information for connecting the quarter-bridge strain-gauge configuration type I. Figure 2-1 shows the quarter-bridge type I circuit wiring diagram.
Note S– is left unwired.
© National Instruments Corporation 2-1 SCXI-1520 User Manual
Chapter 2 Connecting Signals
SCXI-1520 Set Bridge
Configuration to
Quarter Bridge
+
+ –
V
CH
S+
P+
R
L
R
L
R
L
QTR
SCA
SCA
Shunt Cal A
P
SCXI-1314
R
1
R
2
R
S
R
3
R
4
(gauge)
V
EX
+
You must connect
the shunt calibration
wires externally using
the terminal block
screw connections.
Transducer
Figure 2-1. Quarter-Bridge I Circuit Diagram

Quarter-Bridge Type II

SCXI-1520 User Manual 2-2 ni.com
The following symbols apply to the circuit diagram and equations:
•R
and R2 are half-bridge completion resistors.
1
•R
is the quarter-bridge completion resistor.
3
•R
is the active element measuring tensile strain (+ε).
4
V
R
V
Note The value of the quarter-bridge completion resistor, R
resistance of the strain gauge. NI recommends using a 0.1% precision resistor.
This section provides information for connecting the quarter-bridge strain-gauge configuration type II. Figure 2-2 shows the quarter-bridge type II circuit wiring diagram.
is the excitation voltage.
EX
is the lead resistance.
L
is the measured voltage.
CH
, must equal the nominal
3
Chapter 2 Connecting Signals
+
+
S+
P+
P
R
L
R
L
R
L
SCA
SCA
Shunt Cal A
SCXI-1314Transducer
V
CH
R
1
R
2
R
S
R
4
(gauge)
V
EX
R
3
(dummy)
R
L
R
L
SCXI-1520 Set Bridge
Configuration to Quarter Bridge
Note The quarter-bridge type II configuration is often confused with the more commonly
used half-bridge type I configuration. In the half-bridge type I configuration, the R
3
element is active and bonded to the strain specimen to measure Poisson's ratio, while in the quarter-bridge type II configuration, the R
element does not actively measure strain, but
3
is in close thermal contact with the strain specimen. In quarter-bridge type II configuration,
element is not bonded to the specimen. Typically it is either physically close to the
the R
3
specimen or mounted on the same type material at the same temperature, but is not under strain.
Figure 2-2. Quarter-Bridge II Circuit Diagram
The following symbols apply to the circuit diagram and equations:
© National Instruments Corporation 2-3 SCXI-1520 User Manual
Note S– and QTR are left unwired.
•R
and R2 are half-bridge completion resistors.
1
•R
is the quarter-bridge temperature-sensing element (dummy gauge).
3
•R
is the active element measuring tensile strain (+ε).
4
V
R
V
is the excitation voltage.
EX
is the lead resistance.
L
is the measured voltage.
CH
Chapter 2 Connecting Signals
+
+
S+
P+
P
R
L
R
L
R
L
SCA
SCA
Shunt Cal A
+
v
SCXI-1520 Set Bridge
Configuration to Half Bridge
SCXI-1314
Transducer
V
CH
R
1
R
2
R
S
R
4
(gauge)
V
EX
R
3
(gauge)
R
L
R
L

Half-Bridge Type I

Note S– is left unwired.
This section provides information for connecting the half-bridge strain-gauge configuration type I. Figure 2-3 shows the half-bridge type I circuit wiring diagram.
Figure 2-3. Half-Bridge Type I Circuit Diagram
The following symbols apply to the circuit diagram and equations:
•R
and R2 are half-bridge completion resistors.
1
•R
is the active element measuring compression from Poisson
3
effect (–νε).
Note As shown in Figure 2-4, for greatest calibration accuracy, use separate wires
•R
V
R
V
between the bridge and the SCA terminals. Do not directly connect S+ or P– to the
SCXI-1520 User Manual 2-4 ni.com
SCA terminals inside the SCXI-1314 terminal block unless the strain-gauge cable length is very short.
is the active element measuring tensile strain (+ε).
4
is the excitation voltage.
EX
is the lead resistance.
L
is the measured voltage.
CH

Half-Bridge Type II

+
+
S+
P+
P
R
L
R
L
R
L
SCA
SCA
Shunt Cal A
+
SCXI-1314Transducer
V
CH
R
1
R
2
R
S
R
4
(gauge)
V
EX
R
3
(gauge)
R
L
R
L
V
out
+
V
out
SCXI-1520 Set Bridge
Configuration to Half Bridge
Note S– is left unwired.
Chapter 2 Connecting Signals
This section provides information for connecting the half-bridge strain-gauge configuration type II. Figure 2-4 shows the half-bridge type II circuit wiring diagram.
Figure 2-4. Half-Bridge Type II Circuit Diagram
The following symbols apply to the circuit diagram and equations:
•R
and R2 are half-bridge completion resistors.
1
•R
is the active element measuring compressive strain (–ε).
3
•R
© National Instruments Corporation 2-5 SCXI-1520 User Manual
Note As shown in Figure 2-3, for greatest calibration accuracy, use separate wires
between the bridge and the SCA terminals. Do not directly connect S+ or P– to the SCA terminals inside the SCXI-1314 terminal block unless the strain-gauge cable length is very short.
is the active element measuring tensile strain (+ε).
4
V
R
V
is the excitation voltage.
EX
is the lead resistance.
L
is the measured voltage.
CH
Chapter 2 Connecting Signals
+
V
EX
+
V
EX
+
S+
S
P+
P
R
1
R
2
R
4
R
3
SCA
SCA
Shunt Cal A
+
+
R
L
R
L
SCXI-1520 Set Bridge
Configuration to Full BridgeSCXI-1314
Transducer
V
CH
R
S
CH+
CH
R
L
R
L

Full-Bridge Type I

This section provides information for connecting the full-bridge strain-gauge configuration type I. Figure 2-5 shows the full-bridge type I circuit wiring diagram.
Figure 2-5. Full-Bridge Type I Circuit Diagram
The following symbols apply to the circuit diagram and equations:
•R
is an active element measuring compressive strain (–ε).
1
•R
is an active element measuring tensile strain (+ε).
2
•R
is an active element measuring compressive strain (–ε).
3
•R
is an active element measuring tensile strain (+ε).
4
SCXI-1520 User Manual 2-6 ni.com
Note As shown in Figure 2-5, for greatest calibration accuracy, use separate wires
between the bridge and the SCA terminals. Do not directly connect S+ or P– to the SCA terminals inside the SCXI-1314 terminal block unless the strain-gauge cable length is very short.
V
R
V
is the excitation voltage.
EX
is the lead resistance.
L
is the measured voltage.
CH

Full-Bridge Type II

+
+
S+
S
P+
P
SCA
SCA
Shunt Cal A
+
–v
+v
R
L
R
L
SCXI-1520 Set Bridge
Configuration to Full BridgeSCXI-1314
Transducer
R
1
R
2
R
4
R
3
V
CH
R
S
V
EX
+
V
EX–
R
L
R
L
Chapter 2 Connecting Signals
This section provides information for connecting the full-bridge strain-gauge configuration type II. Figure 2-6 shows the full-bridge type II circuit wiring diagram.
Figure 2-6. Full-Bridge Type II Circuit Diagram
The following symbols apply to the circuit diagram and equations:
•R
is an active element measuring compressive Poisson effect (–νε).
1
•R
is an active element measuring tensile Poisson effect (+νε).
2
•R
is an active element measuring compressive strain (–ε).
3
© National Instruments Corporation 2-7 SCXI-1520 User Manual
Note As shown in Figure 2-6, for greatest calibration accuracy, use separate wires
between the bridge and the SCA terminals. Do not directly connect S+ or P– to the SCA terminals inside the SCXI-1314 terminal block unless the strain-gauge cable length is very short.
V
R
V
•R
is an active element measuring tensile strain (+ε).
4
is the excitation voltage.
EX
is the lead resistance.
L
is the measured voltage.
CH
Chapter 2 Connecting Signals
+
+
S+
S
P+
P
SCA
SCA
Shunt Cal A
+
–v
–v
+
SCXI-1520 Set Bridge
Configuration to Full BridgeSCXI-1314
Transducer
R
1
R
2
R
4
R
3
V
CH
R
S
V
EX
+
V
EX–
R
L
R
L

Full-Bridge Type III

This section provides information for connecting the full-bridge strain-gauge configuration type I. The full-bridge type III only measures axial strain. Figure 2-7 shows the full-bridge type III circuit wiring diagram.
Figure 2-7. Full-Bridge Type III Circuit Diagram
The following symbols apply to the circuit diagram and equations:
•R
is an active element measuring compressive Poisson effect (–νε).
1
•R
is an active element measuring tensile strain (+ε).
2
•R
is an active element measuring compressive Poisson effect (–νε).
3
•R
is an active element measuring the tensile strain (+ε).
4
is the excitation voltage.
EX
is the lead resistance.
L
is the measured voltage.
CH
Note As shown in Figure 2-7, for greatest calibration accuracy, use separate wires
between the bridge and the SCA terminals. Do not directly connect S+ or P– to the SCA terminals inside the SCXI-1314 terminal block unless the strain-gauge cable
V
R
V
length is very short.
SCXI-1520 User Manual 2-8 ni.com

Remote Sense

Note NI recommends using remote sense if your application requires the improved
accuracy. Refer to Chapter 4, Theory of Operation, for more information about using remote sense.
Run Separate Wires Between Remote-Sense Terminals and Bridge
R
1
Chapter 2 Connecting Signals
Wire the SCXI-1520 for remote sense as shown in Figure 2-8.
SCXI-1520SCXI-1314Transducer
RS+
V
EX
+
R
4
P+
+
Feedback
R
2
V
EX–
R
3
P
RS
Figure 2-8. Remote-Sense Circuit Diagram
Note
If you use remote sense, set RL to zero in the equations for measured strain (ε).
© National Instruments Corporation 2-9 SCXI-1520 User Manual
Chapter 2 Connecting Signals

Pin Assignments

The pin assignments for the SCXI-1520 front signal connector are shown in Table 2-1. The front signal connector is a special 96-pin DIN C male connector through which you make all signal connections. The terminal assignments are as follows:
•SX+ and SX– are for analog input
•RSX+ and RSX– are for remote sense
•PX+ and PX– are for excitation output
•SCAX are for shunt calibration circuit A
•SCBX are for shunt calibration circuit B
where X is the channel number.
The negative terminals are listed in Column B and the positive terminals are listed in Column C. The pins labeled RSVD are reserved. Do not make any connections to the RSVD pins.
SCXI-1520 User Manual 2-10 ni.com
Column
A B C
32
31
30
29
28
27
26
25
24
23
22
21
20
19
18
17
16
15
14
13
12
11
10
9
8
7
6
5
4
3
2
1
© National Instruments Corporation 2-11 SCXI-1520 User Manual
Chapter 2 Connecting Signals

Table 2-1. Front Signal Pin Assignments

Front Connector Diagram Pin Number Column A Column B Column C
32 SCB0 S0– S0+
31 SCB0 RS0– RS0+
30 SCB1 P0– P0+
29 SCB1 SCA0 SCA0
28 RSVD S1– S1+
27 RSVD RS1– RS1+
26 RSVD P1– P1+
25 RSVD SCA1 SCA1
24 SCB2 S2– S2+
23 SCB2 RS2– RS2+
22 SCB3 P2– P2+
21 SCB3 SCA2 SCA2
20 RSVD S3– S3+
19 RSVD RS3– RS3+
18 RSVD P3– P3+
17 RSVD SCA3 SCA3
16 SCB4 S4– S4+
15 SCB4 RS4– RS4+
14 SCB5 P4– P4+
13 SCB5 SCA4 SCA4
12 RSVD S5– S5+
11 RSVD RS5– RS5+
10 RSVD P5– P5+
9 RSVD SCA5 SCA5
8 SCB6 S6– S6+
7 SCB6 RS6– RS6+
6 SCB7 P6– P6+
5 SCB7 SCA6 SCA6
4 RSVD S7– S7+
3 RSVD RS7– RS7+
2 RSVD P7– P7+
1 RSVD SCA7 SCA7
Loading...
+ 133 hidden pages