National Instruments SCXI -1314 Installation Manual

Page 1
INSTALLATION GUIDE
SCXI -1314 U
NIVERSAL STRAIN TERMINALBLOCK
This guide describes how to install and use the SCXI-1314 terminal block with your SCXI-1520 module.

Introduction

The SCXI-1314 terminal block is used with the SCXI-1520 universal strain/bridge module enabling you to conveniently connect strain gauges through screw terminals. There are 88 screw terminals arranged in eight groups of 11. Each group corresponds to one of the eight channels available on the SCXI-1520. (Refer to Figure 2 for a diagram of terminal locations.) The signal names of the terminals for each channel are listed in Table 1.
Table 1.
Signal Name Condition
P+ Positive excitation signal
P– Negative excitation signal
S+ Positive input signal
S– Negative input signal
QTR Quarter-bridge completion resistor connection
RS+ Positive remote sense signal
RS– Negative remote sense signal
SCA
(2 terminals)
SCB
(2 terminals)
Each channel input contains screw terminals that connect to the inputs of the SCXI-1520, two precision 100 k shunt calibration resistors, and a precision quarter-bridge completion resistor. The SCXI-1520 contains
Shunt calibration A signal
Shunt calibration B signal
Signal Names
National Instruments™, ni.com™, and SCXI™ are trademarks of National Instruments Corporation. Product and company names mentioned herein are trademarks or trade names of their respective companies.
322859A-01
Copyright 2000 National Instruments Corp. All rights reserved. August 2000
Page 2
relays that switch the resistors in and out of the input circuit to provide shunt calibration and quarter-bridge completion.
The quarter-bridge completion resistor has a factory default value of 350 for use with a 350 strain gauge. These resistors are placed in sockets on the terminal block so you can easily replace them with resistors suitable to your specific strain gauge. If your application requires a 120 Ω strain gauge, use the eight 120 resistors included in the kit.
For a complete description of the use and operation of the SCXI-1520 module, refer to the SCXI-1520 User Manual.

What You Need to Get Started

To install and use your SCXI-1314 terminal block, you need the following items:
SCXI-1314 terminal block
SCXI-1314 Universal Strain Terminal Block Installation Guide
Eight precision 120 resistors for use in quarter-bridge completion
circuits (if using 120 Ω strain gauges)
SCXI chassis
SCXI-1520 module
SCXI-1520 User Manual
Number 1 and 2 Phillips-head screwdrivers
1/8 in. flathead screwdriver
Long-nose pliers
Wire cutter
Wire insulation stripper
SCXI-1314 Universal Strain Terminal Block 2 ni.com
Page 3

Connecting the Signals

To connect the signal to the terminal block, perform the following steps, referring to Figures 1 and 2 as necessary:
6
5
1 Top Cover Screws 2 Top Cover
7
4
3 Thumbscrew 4 Mating Connector
3
5 Strain-Relief Bar 6 Strain-Relief Screws
Figure 1.
SCXI-1314 Parts Locator Diagram
1
2
7 Safety-Ground
Lug
1. Unscrew the top cover screws and remove the top cover.
2. Loosen the strain-relief screws and remove the strain-relief bar.
3. Run the signal wires through the strain-relief opening. You can add insulation or padding if necessary.
4. Prepare your signal wire by stripping the insulation no more than 7 mm.
5. Connect the signal wires to the screw terminals by inserting the stripped end of the wire fully into the terminal. No bare wire should extend past the screw terminal. Exposed wire increases the risk of a short circuit that can cause circuit failure.
When connecting your signals to the SCXI-1314, follow the labeling on the SCXI-1314, as indicated in Figure 2.
National Instruments Corporation 3 SCXI-1314 Universal Strain Terminal Block
Page 4
R1
R3
R2
SCB
SCA
QTR
RPJ
J2 J3 J4 J5
CH0 CH1 CH2 CH3 CH4 CH5 CH6 CH7
SCB
SCB
SCB
SCB
SCB
SCB
SCA
SCA
SCA
SCA
P–
P–
P+
P+
SCA
SCA
QTR
QTR
S+
S+
S–
S–
RS+
RS+
RS–
RS–
QTR
QTR
UJ
SCB
SCB
SCB
SCA
SCA
P–
P+
SCA
QTR
S+
S–
RS+
RS–
QTR
UNIVERSAL STRAIN
SCB
SCB
SCB
SCB
SCA
SCA
SCA
SCA
P–
P–
P+
P+
QTR
QTR
S+
S+
S–
S–
RS+
RS+
RS–
RS–
SCXI-1314
J6 J7 J8 J9
SCB
SCB
SCB
SCB
SCA
SCA
QTR
RS+
RS–
SCBSCAQTR
SCA
SCA
P–
P–
P+
P+
QTR
S+
S+
S–
S–
RS+
RS–
J1
R4
SCB
SCB
SCB
SCB
SCBSCAQTR
SCA
SCA
SCA
SCA
QTR
RS+
RS–
P–
P–
P+
P+
QTR
S+
S+
S–
S–
RS+
RS–
1 2 3 4
5
C2
SCB
SCB
SCBSCAQTR
SCA
SCA
P–
P+
QTR
S+
S–
RS+
RS–
S/N
ASSY187212A-01
C3
Q1
SCBSCAQTR
R1
R3
R2
SCB
6
SCA
7
8
QTR
RPJ
J2 J3 J4 J5
SCB
SCB
SCB
SCB
SCA
SCA
P–
P+
SCA
QTR
S+
S–
RS+
RS–
QTR
CH0 CH1 CH2 CH3 CH4 CH5 CH6 CH7
UJ
SCB
SCB
SCB
SCB
SCA
SCA
SCA
SCA
P–
P–
P+
P+
SCA
QTR
QTR
S+
S+
S–
S–
RS+
RS+
RS–
RS–
QTR
1 Safety Ground Lug 2 Cable Strain Relief 3 Serial Number 4 Assembly Number 5 Screw Terminal and Associated Resistors
for <CH0...CH7>
6. Tighten the terminal screws to a torque of 5 to 7 in.-lb.
7. Reinstall the strain-relief bar and tighten the strain-relief screws.
8. Reinstall the top cover and tighten the top cover screws.
9. Connect the terminal block to the module front connector as explained in the Installing the Terminal Block section.
SCB
SCB
SCB
SCB
SCB
SCA
SCA
SCA
SCA
P–
P–
P+
P+
SCA
QTR
QTR
S+
S+
S–
S–
RS+
RS+
RS–
RS–
QTR
SCXI-1314
UNIVERSAL STRAIN
6 Shunt Calibration B Resistors, 8 places
7 Shunt Calibration A Resistors, 8 places
8 Quarter-Bridge Completion Resistor, 8 places
Figure 2.
J1
J6 J7 J8 J9
SCB
SCB
SCB
SCB
SCA
SCA
P–
P+
QTR
S+
S–
RS+
RS–
SCB
SCB
SCBSCAQTR
SCA
SCA
SCA
SCA
QTR
RS+
RS–
P–
P–
P+
P+
QTR
S+
S+
S–
S–
RS+
RS–
(Factory Default = 100 k
(Factory Default = 100 k
(Factory Default = 350
R4
SCB
SCB
SCBSCAQTR
SCA
SCA
P–
P+
QTR
S+
S–
RS+
RS–
,
SCXI-1314 Signal Connections
C2
SCB
SCB
SCBSCAQTR
SCA
SCA
P–
P+
QTR
S+
S–
RS+
RS–
S/N
ASSY187212A-01
,
0.1%)
,
0.1%)
0.1%)
C3
Q1
SCBSCAQTR
SCXI-1314 Universal Strain Terminal Block 4 ni.com
Page 5

Installing the Terminal Block

To connect the terminal block to the SCXI module front connector, perform the following steps:
1. Connect the module front connector to its mating connector on the terminal block.
2. Tighten the top and bottom thumbscrews on the back of the terminal block to hold it securely in place.

Cleaning the Terminal Block

Clean the terminal block by brushing off light dust with a soft, nonmetallic brush. Remove other contaminants with deionized water and a stiff nonmetallic brush. The unit must be completely dry and free from contaminants before returning it to service.

Calibrating the Terminal Block

For information on calibrating the SCXI-1314 see the SCXI-1314 Universal Strain Terminal Block Calibration Procedure located online at
ni.com/calibration
Procedures.
and follow the link for Manual Calibration

Specifications

All specifications are typical at 25 °C unless otherwise specified.

Electrical

Accuracy of resistors
120 completion resistor............... ±0.1%
350 completion resistor............... ±0.1%
100 k shunt calibration resistor.... ±0.1%
Temperature coefficient of resistors
120 completion resistor............... ±10 ppm/°C
350 completion resistor............... ±10 ppm/°C
100 k shunt calibration resistor.... ±10 ppm/°C
National Instruments Corporation 5 SCXI-1314 Universal Strain Terminal Block
Page 6

Mechanical

Environment

Resistor sockets
Mating lead size...............................0.023 to 0.026 in.
Mating lead length...........................0.110 to 0.175 in.
Lead spacing....................................0.500 in.
Operating temperature ............................0 to 50 °C
Storage temperature................................–20 to 70 °C
Relative humidity ...................................10 to 90%
Indoor use only
SCXI-1314 Universal Strain Terminal Block 6 ni.com