6504 Bridge Point Parkway
Austin, TX 78730-5039
(512) 794-0100
Technical support fax: (800) 328-2203
(512) 794-5678
Branch Offices:
Australia (03) 879 9422, Austria (0662) 435986, Belgium 02/757.00.20, Canada (Ontario) (519) 622-9310,
Canada (Québec) (514) 694-8521, Denmark 45 76 26 00, Finland (90) 527 2321, France (1) 48 14 24 24,
Germany 089/741 31 30, Italy 02/48301892, Japan (03) 3788-1921, Netherlands 03480-33466, Norway 32-848400,
Spain (91) 640 0085, Sweden 08-730 49 70, Switzerland 056/20 51 51, U.K. 0635 523545
Limited Warranty
The SCXI-1163 is warranted against defects in materials and workmanship for a period of one year from the date of
shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace
equipment that proves to be defective during the warranty period. This warranty includes parts and labor.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the
outside of the package before any equipment will be accepted for warranty work. National Instruments will pay the
shipping costs of returning to the owner parts which are covered by warranty.
National Instruments believes that the information in this manual is accurate. The document has been carefully
reviewed for technical accuracy. In the event that technical or typographical errors exist, National Instruments
reserves the right to make changes to subsequent editions of this document without prior notice to holders of this
edition. The reader should consult National Instruments if errors are suspected. In no event shall National
Instruments be liable for any damages arising out of or related to this document or the information contained in it.
EXCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTSMAKES NO WARRANTIES, EXPRESS OR IMPLIED,
AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE
OF
NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR DAMAGES RESULTING FROM LOSS OF DATA, PROFITS,
USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY
THEREOF
whether in contract or tort, including negligence. Any action against National Instruments must be brought within
one year after the cause of action accrues. National Instruments shall not be liable for any delay in performance due
to causes beyond its reasonable control. The warranty provided herein does not cover damages, defects,
malfunctions, or service failures caused by owner's failure to follow the National Instruments installation, operation,
or maintenance instructions; owner's modification of the product; owner's abuse, misuse, or negligent acts; and
power failure or surges, fire, flood, accident, actions of third parties, or other events outside reasonable control.
. CUSTOMER'S RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART
NATIONAL INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE CUSTOMER.
. This limitation of the liability of National Instruments will apply regardless of the form of action,
Copyright
Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or
mechanical, including photocopying, recording, storing in an information retrieval system, or translating, in whole or
in part, without the prior written consent of National Instruments Corporation.
Trademarks
®
LabVIEW®, NI-DAQ®, and RTSI
Product and company names listed are trademarks or trade names of their respective companies.
are trademarks of National Instruments Corporation.
WARNING REGARDING MEDICAL AND CLINICAL USE
OF NATIONAL INSTRUMENTS PRODUCTS
National Instruments products are not designed with components and testing intended to ensure a level of reliability
suitable for use in treatment and diagnosis of humans. Applications of National Instruments products involving
medical or clinical treatment can create a potential for accidental injury caused by product failure, or by errors on the
part of the user or application designer. Any use or application of National Instruments products for or involving
medical or clinical treatment must be performed by properly trained and qualified medical personnel, and all
traditional medical safeguards, equipment, and procedures that are appropriate in the particular situation to prevent
serious injury or death should always continue to be used when National Instruments products are being used.
National Instruments products are NOT intended to be a substitute for any form of established process, procedure, or
equipment used to monitor or safeguard human health and safety in medical or clinical treatment.
Contents
About This Manual............................................................................................................. xi
Organization of This Manual ......................................................................................... xi
Conventions Used in This Manual................................................................................. xii
Related Documentation.................................................................................................. xiii
Customer Communication ............................................................................................. xiv
Chapter 1
Introduction
What You Need To Get Started ..................................................................................... 1-2
This manual describes the electrical and mechanical aspects of the SCXI-1163 and contains
information concerning its operation and programming. The SCXI-1163 is a member of the
National Instruments Signal Conditioning eXtensions for Instrumentation (SCXI) Series modules
for the National Instruments data acquisition plug-in boards. This module provides 32 isolated
digital outputs configured in ports of four channels. Each port is isolated from other ports and
from the internal circuitry of the SCXI-1163.
This manual describes the installation, basic programming considerations, and theory of
operation for the SCXI-1163.
Organization of This Manual
The SCXI-1163 User Manual is organized as follows:
• Chapter 1, Introduction, describes the SCXI-1163; lists the contents of your SCXI-1163 kit;
describes the optional software, optional equipment, and custom cables; and explains how to
unpack the SCXI-1163 kit.
• Chapter 2, Configuration and Installation, describes the SCXI-1163 jumper configurations,
installation of the SCXI-1163 into the SCXI chassis, signal connections to the SCXI-1163,
and cable wiring.
• Chapter 3, Theory of Operation, contains a functional overview of the SCXI-1163 module
and explains the operation of each functional unit making up the SCXI-1163.
• Chapter 4, Register Descriptions, describes in detail the SCXI-1163 Address Handler, the
Module ID Register, the Data Register, the Status registers, and the SCXI-1000/1001 Slot 0
registers.
• Chapter 5, Programming, contains a functional programming description of the SCXI-1163
and Slot 0.
• Appendix A, Specifications, lists the specifications for the SCXI-1163.
• Appendix B, Rear Signal Connector, describes the pinout and signal names for the
SCXI-1163 50-pin rear signal connector, including a description of each connection.
• Appendix C, SCXIbus Connector, describes the pinout and signal names for the SCXI-1163
SCXIbus connector, including a description of each connection.
• Appendix D, SCXI-1163 Front Connector, describes the pinout and signal names for the
SCXI-1163 front connector, including a description of each connection.
• Appendix E, SCXI-1163 Cabling, describes how to use and install the hardware accessories
for the SCXI-1163.
• Appendix F, Customer Communication, contains forms you can use to request help from
National Instruments or to comment on our products.
• The Glossary contains an alphabetical list and description of terms used in this manual,
including abbreviations, acronyms, metric prefixes, mnemonics, and symbols.
• The Index contains an alphabetical list of key terms and topics in this manual, including the
page where you can find each one.
Conventions Used in This Manual
The following conventions are used in this manual:
DIO boardDIO board refers to the National Instruments AT-DIO-32F, MC-DIO-24,
MC-DIO-32F, NB-DIO-24, NB-DIO-32F, NB-DIO-96, PC-DIO-24, and
PC-DIO-96 digital I/O data acquisition boards unless otherwise noted.
DIO-type boardDIO-type board refers to National Instruments data acquisition boards that
have only digital inputs and outputs. These boards include the DIO-24,
DIO-32F, and DIO-96 boards unless otherwise noted.
italicItalic text denotes emphasis, a cross reference, or an introduction to a key
concept.
Lab boardLab board refers to the National Instruments Lab-LC, Lab-NB, Lab-PC,
and Lab-PC+ boards unless otherwise noted.
MCMC refers to the Micro Channel series computers.
MIO boardMIO board refers to the National Instruments AT-MIO-16, AT-MIO-16D,
names, functions, variables, filenames, and extensions, and for statements
and comments taken from program code.
NBNB refers to the NuBus series computers.
PCPC refers to the IBM PC/XT, the IBM PC AT, and compatible computers.
SCXIbusSCXIbus refers to the backplane in the chassis. A signal on the backplane
is referred to as the SCXIbus <signal name> line (or signal).
The SCXIbus descriptor may be omitted when the meaning is clear.
Descriptions of all SCXIbus signals are given in Appendix C, SCXIbus
Connector.
Slot 0Slot 0 refers to the power supply and control circuitry in the SCXI chassis.
National Instruments wants to receive your comments on our products and manuals. We are
interested in the applications you develop with our products, and we want to help if you have
problems with them. To make it easy for you to contact us, this manual contains comment and
configuration forms for you to complete. These forms are in Appendix F, Customer
This chapter describes the SCXI-1163; lists the contents of your SCXI-1163 kit; describes the
optional software, optional equipment, and custom cables; and explains how to unpack the
SCXI-1163 kit.
Figure 1-1 shows the SCXI-1163 module.
Figure 1-1. SCXI-1163 Digital Output Module
The SCXI-1163 is an SCXI module consisting of 32 optically isolated digital outputs. The
SCXI-1163 module provides digital signals where common-mode voltages may be present.
The SCXI-1163 operates with full functionality with National Instruments MIO boards; Lab-NB,
Lab-PC, Lab-PC+, Lab-LC, and PC-LPM-16 boards; and with the DIO-24, DIO-32F, and
DIO-96 boards. You can also use the SCXI-1163 with other systems that comply with the
specifications given in Chapter 2, Configuration and Installation. You can control several
SCXI-1163s in a single chassis with one data acquisition board.
An additional shielded terminal block, the SCXI-1326, has screw terminals for easy signal
attachment to the SCXI-1163.
With the SCXI-1163, you can use the SCXI chassis as a controller in laboratory testing,
production testing, and industrial-process monitoring.
What You Need To Get Started
SCXI-1163 module
SCXI-1163 User Manual
SCXI chassis
Your computer
Optional Software
This manual contains complete instructions for directly programming the SCXI-1163. You can
order separate software packages for controlling the SCXI-1163 from National Instruments.
When you combine the PC, AT, and MC data acquisition boards with the SCXI-1163, you can
use LabVIEW for Windows or LabWindows for DOS. LabVIEW and LabWindows are
innovative program development software packages for data acquisition and control applications.
LabVIEW uses graphical programming, whereas LabWindows enhances Microsoft C and
QuickBASIC. Both packages include extensive libraries for data acquisition, instrument control,
data analysis, and graphical data presentation.
Your National Instruments data acquisition board is shipped with the NI-DAQ software.
NI-DAQ has a library of functions that can be called from your application programming
environment. These functions include routines for analog input (A/D conversion), buffered data
acquisition (high speed A/D conversion), analog output (D/A conversion), waveform generation,
digital I/O, counter/timer, SCXI, RTSI, and self-calibration. NI-DAQ maintains a consistent
software interface among its different versions so you can switch between platforms with
minimal modifications to your code.
You can also use the SCXI-1163, together with the PC, AT, and MC data acquisition boards,
with NI-DAQ software for DOS/Windows/LabWindows. NI-DAQ software for
DOS/Windows/LabWindows comes with language interfaces for Professional BASIC, Turbo
Pascal, Turbo C, Turbo C++, Borland C++, and Microsoft C for DOS; and Visual Basic, Turbo
Pascal, Microsoft C with SDK, and Borland C++ for Windows. NI-DAQ software for
DOS/Windows/LabWindows is on high-density 5.25 in. and 3.5 in. diskettes.
You can use the SCXI-1163, together with the NB Series data acquisition boards, with
LabVIEW for Macintosh, a software system that features interactive graphics, a state-of-the-art
user interface, and a powerful graphical programming language. The LabVIEW Data
Acquisition VI Library, a series of VIs for using LabVIEW with National Instruments boards, is
included with LabVIEW. The LabVIEW Data Acquisition VI Library is functionally equivalent
to the NI-DAQ software for Macintosh.
You can also use the SCXI-1163, combined with the NB Series data acquisition boards, with
NI-DAQ software for Macintosh. NI-DAQ software for Macintosh, which is shipped with all
National Instruments Macintosh data acquisition boards, comes with language interfaces for
MPW C, THINK C, Pascal, and Microsoft QuickBASIC. Any language that uses Device
Manager Toolbox calls can access NI-DAQ software for Macintosh.
Optional Equipment
• SCXI-1326 front terminal block
• SCXI-1340 cable assembly
• SCXI-1341 Lab-NB/Lab-PC/Lab-PC+ cable assembly
• SCXI-1342 PC-LPM-16 cable assembly
• SCXI-1343 rear screw terminal adapter
• SCXI-1344 Lab-LC cable assembly
• SCXI-1348 DIO-32F cable assembly
• SCXI-1350 multichassis adapter
• SCXI-1351 one-slot cable extender
• Standard ribbon cable 0.5 m
1.0 m
• NB5 cable 0.5 m
1.0 m
Refer to the Signal Connections section in Chapter 2, Configuration and Installation, and to
Appendix E, SCXI-1163 Cabling, for additional information on cabling, connectors, and
adapters.
Custom Cables
The SCXI-1163 rear signal connector is a 50-pin male ribbon-cable header. The manufacturer
part number National Instruments uses for this header is as follows:
•AMP Inc. (part number 1-103310-0)
The mating connector for the SCXI-1163 rear signal connector is a 50-position polarized
ribbon-socket connector with strain relief. National Instruments uses a polarized or keyed
connector to prevent inadvertent upside-down connection to the SCXI-1163. Recommended
manufacturer part numbers for this mating connector are as follows:
•Electronic Products Division/3M (part number 3425-7650)
•T&B/Ansley Corporation (part number 609-5041CE)
Standard 50-conductor 28 AWG stranded ribbon cables that work with these connectors are as
follows:
•Electronic Products Division/3M (part number 3365/50)
•T&B/Ansley Corporation (part number 171-50)
The SCXI-1163 front connector is a special 48-pin DIN C male connector. The manufacturer
part number National Instruments uses for this connector is as follows:
•ERNI Components, Inc. (part number 033-273)
The mating connector for the SCXI-1163 front connector is a special 48-pin reversed DIN C
female connector. National Instruments uses a polarized and keyed connector to prevent
inadvertent upside-down connection to the SCXI-1163. The manufacturer part number National
Instruments uses for this mating connector is as follows:
•ERNI Components, Inc. (part number 913-495)
National Instruments selected these connectors to meet UL 1244 for 450 Vrms working isolation.
Unpacking
Your SCXI-1163 module is shipped in an antistatic package to prevent electrostatic damage to
the module. Several components on the module can be damaged by electrostatic discharge. To
avoid such damage in handling the module, take the following precautions:
•Touch the antistatic package to a metal part of your SCXI chassis before removing the
module from the package.
•Remove the module from the package and inspect the module for loose components or any
other sign of damage. Notify National Instruments if the module appears damaged in any
way. Do not install a damaged module into your SCXI chassis.
This chapter describes the SCXI-1163 jumper configurations, installation of the SCXI-1163 into
the SCXI chassis, signal connections to the SCXI-1163, and cable wiring.
Module Configuration
The SCXI-1163 module consists of five user-configurable jumpers and one reserved jumper as
shown in the parts locator diagram, Figure 2-1.
The SCXI-1163 can accept data in two forms–serial and parallel. Serial data is transferred across
the SCXIbus using the Serial Peripheral Interface (SPI) protocol. Using serial communication,
one data acquisition board that is cabled to one SCXI module in an SCXI system can control the
digital communication for all the boards in the system. Parallel data is read from the rear
connector and transferred directly to the outputs at the front connector. No extra communication
through the rear connector is needed or permitted with parallel communication; thus you can
transfer parallel data to the outputs more rapidly than serial data. However, each module that is
utilizing parallel data transfers requires a separate data source, such as a data acquisition board.
You can configure the SCXI-1163 to primarily accept either parallel or serial data. In one
configuration, the rear connector accepts parallel data; in the other configuration, the rear
connector creates a serial communication link with the SCXIbus. Configure the module for
parallel data transfers when you want to send data directly from the rear connector to the outputs.
Configure the module for serial communication whenever you want to send only serial data to
the module. You must configure the module for serial communication if you want a data
acquisition board that is cabled to the module rear connector to control the SCXIbus.
When you cable the SCXI-1163 to a data acquisition board, you must also configure the
SCXI-1163 to recognize the board either as a DIO-type or an MIO-type board. DIO-type boards
are National Instruments boards that have only digital inputs and outputs. These boards include
the DIO-24, DIO-32F, and DIO-96. MIO-type boards are National Instruments boards that have
both analog and digital inputs and digital outputs. These boards consist of MIO boards; Lab-NB,
Lab-PC, Lab-PC+, and Lab-LC boards; and PC-LPM-16 boards. You can use either type for
serial communication; for parallel communication, you can use only the DIO-type boards.
However, when you use the DIO-24 for parallel communication, only channels 0 through 23 can
be controlled on the SCXI-1163.
To configure the SCXI-1163 module, use the five user-configurable jumpers shown in the parts
locator diagram, Figure 2-1. There is also one reserved jumper that should remain empty. To
change the configuration of the module, use the following procedure:
1. Remove the grounding screw of the top cover.
2. Snap out the top cover of the shield by placing a screwdriver in the groove at the bottom of
the module.
3. Remove the jumpers to be changed and replace them on the appropriate pins.
4. Record the new jumper settings on the SCXI-1163 Hardware Configuration Formin
Appendix F, Customer Communication.
5. Snap the top cover back in place.
6. Replace the grounding screw to ensure proper shielding.
The following warnings contain important safety information concerning hazardous voltages.
Warnings:Keep away from live circuits. Do not remove equipment covers or shields unless
you are trained to do so. If signal wires are connected to the module or terminal
block, dangerous voltages may exist even when the equipment is turned off. To
avoid dangerous electrical shock, do not perform procedures involving cover or
shield removal unless you are qualified to do so.
Do not operate damaged equipment. The safety protection features built into this
module can become impaired if the module becomes damaged in any way. If it is
damaged, turn the module off and do not use until service-trained personnel can
check its safety. If necessary, return the module to National Instruments for
service and repair to ensure that its safety is not compromised.
Do not substitute parts or modify equipment. Because of the danger of
introducing additional hazards, do not install unauthorized parts or modify the
module. Return the module to National Instruments for service and repair to
ensure that its safety features are not compromised.
When using the terminal block with high common-mode voltages, you must
insulate your signal wires appropriately. National Instruments is not liable for any
damages or injuries resulting from inadequate signal wire insulation.
When connecting or disconnecting signal lines to the SCXI-1326 terminal block
screw terminals, make sure the lines are powered off to prevent shock hazard.
Connections, including power signals to ground and vice versa, that exceed any of
the maximum signal ratings on the SCXI-1163 can damage any or all of the
boards connected to the SCXI chassis, the host computer, and the SCXI-1163
module. National Instruments is not liable for any damages or injuries resulting
from incorrect signal connections.
If high voltages (≥42 Vrms) are present, you must connect the safety earth groundto the strain relief tab. This complies with UL 1244 and fully protects against
electric shock when the terminal block is not connected to the chassis. To connect
the safety earth ground to the strain relief tab, run an earth ground wire in the
cable from the signal source to the terminal block. National Instruments is not
liable for any damages or injuries resulting from inadequate safety earth ground
connections.
Jumper Use
Use the jumpers as follows:
•Reserved jumper
-Jumper W1 is reserved; do not connect this jumper.
•User-configurable jumpers
-Jumpers W2 and W3 select whether the SCXI-1163 is to be connected to a DIO-type
board or to an MIO-type board. If you are controlling the SCXI-1163 module via the
SCXI backplane, the position of these jumpers is irrelevant.
-Jumper W4 connects a pullup resistor to the SERDATOUT signal on the rear signal
connector.
-Jumper W5 carries the SCXIbus MISO line, after buffering, to the SERDATOUT signal
on the rear signal connector.
-Jumper W6 selects the primary operating mode, either serial or parallel, and configures
the rear connector accordingly.
Further configuration of the board is software controlled and is discussed later in this chapter.
Jumper Configuration
All five user-configurable jumpers (W2 through W6) on the SCXI-1163 are for digital
communication between the data acquisition board and the SCXI-1163 module.
The SCXI-1163 has two jumpers, W2 and W3, that indicate to the module what type of data
acquisition board is to be connected to the module rear signal connector. The SCXI-1163 also
has two jumpers, W4 and W5, for communication between the data acquisition board and the
SCXIbus. There is also one jumper, W6, that you use to set the primary mode of operation. See
Tables 2-1 and 2-2 for the description and configuration of the jumper settings.
Jumpers W2 and W3
You can use two types of data acquisition boards to communicate with the SCXI-1163–DIO-type
boards and MIO-type boards. Jumpers W2 and W3 configure the rear connector to accept
signals from either a DIO-type or an MIO-type data acquisition board. Both jumpers have labels
indicating the DIO and MIO configurations. If you want to use the SCXI-1163 with a DIO-type
board, place the jumpers in the D position. This position is the factory setting. If you want to
use the SCXI-1163 with an MIO-type board, place the jumpers in the M position. If you are
controlling the SCXI-1163 via the SCXI backplane and are not using the rear connector, the
position of these jumpers is irrelevant.
Jumper W4
Position A connects a 2.2 kΩ pullup resistor to the SERDATOUT line. This is the factorydefault setting. Position B does not connect the pullup resistor to the SERDATOUT line.
Jumper W5
Position DIO connects the SCXIbus MISO line, after buffering, to the SERDATOUT pin of the
rear signal connector when used with a DIO-type board. Position MIO connects the SCXIbus
MISO line, after buffering, to the SERDATOUT pin of the rear signal connector when used with
an MIO-type board. In these settings, when you have set jumper W4 properly, the data
acquisition board can read the Module ID Register or the Status Registers of the SCXI-1163.
See the Timing Requirements and Communication Protocolsection later in this chapter, and
Chapter 4, Register Descriptions, for information on reading the Module ID Register and the
Status registers. See Appendix E, SCXI-1163 Cabling, for the pin equivalences of the
SCXI-1163 rear signal connector and the data acquisition board I/O connector. The factorydefault setting is position DIO.
Position PAR disconnects SERDATOUT from the SCXIbus MISO line. Use this position when
you will connect the module in parallel to a data acquisition board.
Using Jumpers W4 and W5
If a module is not connected to a data acquisition board, the positions of jumpers W4 and W5 are
irrelevant. Jumpers W4 and W5 give the data acquisition board access to the MISO line on the
backplane. You use the MISO line to read the Module ID Register of modules. National
Instruments software does not use this ability. You must indicate to the software which module
is in which slot. In addition to reading the Module ID Register, you can use the MISO line on
the SCXI-1163 to read the Status Registers, which indicate the primary and output modes of
operation. National Instruments software uses this feature.
The SERDATOUT line is driven with an open-collector driver, which is a driver that actively
drives low or goes to a high-impedance state, relying on a pullup resistor to make the signal line
go high. When using a single chassis, set jumper W4 to position A on the SCXI-1163 that is
connected to the data acquisition board and set jumper W5 to DIO or MIO. In this setting, the
module drives MISO to SERDATOUT and connects the necessary pullup resistor to the
SERDATOUT line. When using multiple chassis, set jumper W5 to position DIO or MIO on all
of the SCXI-1163s that are cabled to the data acquisition board. Only one of the SCXI-1163s
that are cabled to the data acquisition board should have jumper W4 in position A. It does not
matter which of the SCXI-1163s that are cabled to the data acquisition board has the pullup
connected. You should set jumper W4 to position B on all of the other SCXI-1163 modules that
are cabled to the data acquisition board. If you attach too many pullup resistors to the
SERDATOUT line, the drivers cannot drive the line low.
Jumper W6
Jumper W6 sets the primary mode of operation to either serial mode or parallel mode and
configures the rear connector accordingly. With the primary mode of operation set to parallel,
the rear connector is configured to accept parallel data from a DIO board. With the primary
mode of operation set to serial, the rear connector is configured to allow serial communication
between a data acquisition board and the SCXIbus.
To set the primary mode of operation to parallel, set jumper W6 to position P. In this mode,
jumper W5 should be set to PAR to prevent the MISO line from inadvertently driving one of the
outputs. To set the primary mode of operation to serial, set jumper W6 to position S.
You can install the SCXI-1163 in any available SCXI chassis. After you have made any
necessary changes and have verified and recorded the jumper settings on the form in Appendix F,
Customer Communication, you are ready to install the SCXI-1163. The following are general
installation instructions, but consult the user manual or technical reference manual of your SCXI
chassis for specific instructions and warnings.
1. Turn off the computer that contains the data acquisition board or disconnect the computer
from your SCXI chassis.
2. Turn off the SCXI chassis. Do not insert the SCXI-1163 into a chassis that is turned on.
3. Insert the SCXI-1163 into the board guides. Gently guide the module into the back of the
slot until the connectors make good contact. If you have already installed a cable assembly
in the rear of the chassis, you must firmly engage the module and cable assembly; however,
do not force the module into place.
4. Screw the front mounting panel of the SCXI-1163 to the top and bottom threaded strips of
your SCXI chassis.
5. If this module is to be connected to an MIO-16 or to a DIO-24 data acquisition board, attach
the connector at the metal end of the SCXI-1340 cable assembly to the rear signal connector
on the SCXI-1163 module. Screw the rear panel to the rear threaded strip. Attach the loose
end of the cable to the MIO-16 or DIO-24 board.
Note: For installation procedures with other SCXI accessories and data acquisition boards,
consult Appendix E, SCXI-1163 Cabling.
6. Check the installation.
7. Turn on the SCXI chassis.
8. Turn on the computer or reconnect the computer to your chassis.
The SCXI-1163 board is installed and ready for operation.
Signal Connections
This section describes the signal connections to the SCXI-1163 board via the SCXI-1163 front
connector and rear signal connector, and includes specifications and connection instructions for
the signals given on the SCXI-1163 connectors.
Warning:Connections that exceed any of the maximum ratings of input signals on the
SCXI-1163 can damage the SCXI-1163 board and the SCXIbus. Maximum input
ratings for each signal are given in this chapter under the discussion of that signal.
National Instruments is not liable for any damages resulting from signal
connections that exceed these ratings.
The SCXI-1163 consists of eight ports of four optically isolated digital outputs. Each port of
outputs has a separate connection for a +5 V power supply (Vcc) and for a ground (GND) for
that port. Digital logic powered by the port Vccs and referenced to the port grounds drives the
outputs of each port; therefore, for any port of outputs to work, you must connect a Vcc and a
ground to that particular port. The outputs work on a +5 V supply (in other words, Vcc must be
5 V above ground for any given port), and supply voltages of 7 V or more may damage the
output circuitry. It is recommended you keep the Vcc between 4.5 and 5.5 V above ground. The
Vcc, ground, and four outputs of one port are isolated from the Vcc, ground, and four outputs of
any other port, as well as from the internal circuitry of the module.
All signals have a working isolation voltage of 450 VAC or VDC referenced to chassis earth
ground and to signals of other ports. There is no isolation barrier between signals of the sameport. The signal-to-ground and port-to-port breakdown voltage is 1,900 Vrms.
Terminal Block
To connect the signals to the SCXI-1163 outputs, you can use a National Instruments terminal
block with screw terminals for easy connection.
The terminal block kit is listed in the Optional Equipment section in Chapter 1, Introduction.
SCXI-1326 Terminal Block
The SCXI-1326 front terminal block consists of a shielded board with supports to connect it to
the SCXI-1163 front connector. Forty-eight screw terminals are available for easy signal
connection.
When connecting your signals to the SCXI-1326 for use with the SCXI-1163, you should follow
the labeling as indicated in Figure 2-3, SCXI-1326 Parts Locator Diagram.
The following warnings contain important safety information concerning hazardous voltages.
Warnings:If high voltages (≥42 Vrms) are present, you must connect the safety earth ground
to the strain relief tab. This complies with UL 1244 and fully protects against
electric shock when the terminal block is not connected to the chassis. To connect
the safety earth ground to the strain relief tab, run an earth ground wire in the
cable from the signal source to the terminal block. National Instruments is not
liable for any damages or injuries resulting from inadequate safety earth ground
connections.
When using the terminal block with high common-mode voltages, you must
insulate your signal wires appropriately. National Instruments is not liable for any
damages or injuries resulting from inadequate signal wire insulation.
When connecting or disconnecting signal lines to the SCXI-1326 terminal block
screw terminals, make sure the lines are powered off to prevent shock hazard.
Connections, including power signals to ground and vice versa, that exceed any of
the maximum signal ratings on the SCXI-1163 can damage any or all of the
boards connected to the SCXI chassis, the host computer, and the SCXI-1163
module. National Instruments is not liable for any damages or injuries resulting
from incorrect signal connections.
Signal Connection
To connect the signal to the terminal block, use the following procedure:
1. Remove the grounding screw of the top cover.
2. Snap out the top cover of the shield by placing a screwdriver in the groove at the bottom of
the terminal block.
3. Slide the signal wires one at a time through the front panel strain relief. You can add
insulation or padding if necessary.
4. Connect the wires to the screw terminals, making sure you connect a Vcc and a ground to
every port you use.
5. Tighten the strain relief by tightening the larger screws.
6. Snap the top cover back in place.
7. Replace the grounding screw to ensure proper shielding.
9. Connect the terminal block to the SCXI-1163 front connector as explained in the SCXI-1326
Installation section, the next section in this chapter.
Figure 2-3 shows a parts locator diagram for the SCXI-1326 terminal block.
Warning:If high voltages (≥42 Vrms) are present, you must connect the safety earth ground
to the strain relief tab. This complies with UL 1244 and fully protects against
electric shock when the terminal block is not connected to the chassis. To connect
the safety earth ground to the strain relief tab, run an earth ground wire in the
cable from the signal source to the terminal block. National Instruments is not
liable for any damages or injuries resulting from inadequate safety earth ground
connections.
SCXI-1326 Installation
To connect the terminal block to the SCXI-1163 front connector, follow these steps:
1. Connect the SCXI-1163 front connector to its mating connector on the terminal block.
2. Make sure that the SCXI-1163 top and bottom thumbscrews do not obstruct the rear panel of
the terminal block.
3. Tighten the top and bottom screws on the back of the terminal block to hold the terminal
block securely in place.
Rear Signal Connector
Note: If you will be using the SCXI-1163 with a National Instruments data acquisition board
and cable assembly, you do not need to read the remainder of this chapter. If you will
also be using the SCXI-1180 feedthrough panel or the SCXI-1343 rear screw terminal
adapter with the SCXI-1163, you should read this section.
The rear signal connector is configured based on the jumper settings described earlier in this
chapter. Jumper W6 determines whether the rear connector will be used for serial or parallel
communication. If you set jumper W6 to serial, jumpers W2, W3, and W5 determine whether
the rear connector is configured for an MIO-type connection or a DIO-type connection.
Figure 2-4A shows the pin assignments for the SCXI-1163 rear signal connector configured for
serial communication. Figure 2-4B shows the pin assignments for the rear signal connector
configured for parallel communication.