National Instruments SCXI-1162HV User Manual

Page 1
SCXI-1162HV
User Manual
32-Channel Isolated AC/DC Digital Input Module
August 1994 Edition
Part Number 320790A-01
© Copyright 1994 National Instruments Corporation.
All Rights Reserved.
Page 2
6504 Bridge Point Parkway Austin, TX 78730-5039 (512) 794-0100 Technical support fax: (800) 328-2203
(512) 794-5678
Branch Offices:
Australia (03) 879 9422, Austria (0662) 435986, Belgium 02/757.00.20, Canada (Ontario) (519) 622-9310, Canada (Québec) (514) 694-8521, Denmark 45 76 26 00, Finland (90) 527 2321, France (1) 48 14 24 24, Germany 089/741 31 30, Italy 02/48301892, Japan (03) 3788-1921, Netherlands 03480-33466, Norway 32-848400, Spain (91) 640 0085, Sweden 08-730 49 70, Switzerland 056/20 51 51, U.K. 0635 523545
Page 3
Limited Warranty
The SCXI-1162HV is warranted against defects in materials and workmanship for a period of one year from the date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that proves to be defective during the warranty period. This warranty includes parts and labor.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are covered by warranty.
National Instruments believes that the information in this manual is accurate. The document has been carefully reviewed for technical accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent editions of this document without prior notice to holders of this edition. The reader should consult National Instruments if errors are suspected. In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.
EXCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR IMPLIED,
AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE OF
NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR DAMAGES RESULTING FROM LOSS OF DATA, PROFITS,
USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY THEREOF
whether in contract or tort, including negligence. Any action against National Instruments must be brought within one year after the cause of action accrues. National Instruments shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not cover damages, defects, malfunctions, or service failures caused by owner's failure to follow the National Instruments installation, operation, or maintenance instructions; owner's modification of the product; owner's abuse, misuse, or negligent acts; and power failure or surges, fire, flood, accident, actions of third parties, or other events outside reasonable control.
. CUSTOMER'S RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART
NATIONAL INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE CUSTOMER.
. This limitation of the liability of National Instruments will apply regardless of the form of action,
Copyright
Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical, including photocopying, recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National Instruments Corporation.
Trademarks
LabVIEW®, NI-DAQ®, and RTSI® are trademarks of National Instruments Corporation. Product and company names listed are trademarks or trade names of their respective companies.
Page 4
Warning Regarding Medical and Clinical Use
of National Instruments Products
National Instruments products are not designed with components and testing intended to ensure a level of reliability suitable for use in treatment and diagnosis of humans. Applications of National Instruments products involving medical or clinical treatment can create a potential for accidental injury caused by product failure, or by errors on the part of the user or application designer. Any use or application of National Instruments products for or involving medical or clinical treatment must be performed by properly trained and qualified medical personnel, and all traditional medical safeguards, equipment, and procedures that are appropriate in the particular situation to prevent serious injury or death should always continue to be used when National Instruments products are being used. National Instruments products are NOT intended to be a substitute for any form of established process, procedure, or equipment used to monitor or safeguard human health and safety in medical or clinical treatment.
Page 5
Contents
About This Manual............................................................................................................vii
Organization of This Manual........................................................................................vi i
Conventions Used in This Manual................................................................................vii
The National Instruments Documentation Set..............................................................ix
Related Documentation.................................................................................................ix
Customer Communication ............................................................................................x
Chapter 1 Introduction
About the SCXI-1162HV..............................................................................................1-1
What You Need to Get Started .....................................................................................1-1
Software Programming Choices ...................................................................................1-1
Optional Equipment......................................................................................................1-4
Unpacking.....................................................................................................................1-6
.........................................................................................................................1-1
LabVIEW and LabWindows Application Software .........................................1-2
NI-DAQ Driver Software..................................................................................1-2
Register-Level Programming............................................................................1-4
Custom Cables ..................................................................................................1-5
Chapter 2 Configuration and Installation
Module Configuration...................................................................................................2-1
Configuration Options.......................................................................................2-1
Configuring the Rear Connector for Parallel Mode..............................2-2
Configuring the Rear Connector for Serial Mode.................................2-2
Jumper Use........................................................................................................2-3
Supplementary Jumper W3 Configuration Installation ........................2-6
Hardware Installation....................................................................................................2-6
Software Installation.....................................................................................................2-7
Chapter 3 Signal Connections
Front Connector ................................................................................................3-3
Front Connector Signal Descriptions................................................................3-4
Terminal Block .................................................................................................3-6
Rear Signal Connector......................................................................................3-6
Rear Signal Connector Signal Descriptions, Serial Configuration...................3-8
Rear Signal Connector Signal Descriptions, Parallel Configuration................3-9
............................................................................................................3-1
Input Channels ......................................................................................3-4
Sensing DC Voltages............................................................................3-4
Sensing AC Voltages............................................................................3-5
Signal Isolation .....................................................................................3-5
Signal Connection Example..................................................................3-5
......................................................................................2-1
© National Instruments Corporation v SCXI-1162HV User Manual
Page 6
Contents
Appendix A Specifications
.......................................................................................................................A-1
Appendix B Customer Communication
..............................................................................................B-1
Glossary........................................................................................................................Glossary-1
Index..................................................................................................................................Index-1
Figures
Figure 1-1. The Relationship between the Programming Environment,
NI-DAQ, and Your Hardware...........................................................................1-3
Figure 2-1. Removing the SCXI Module Cover..................................................................2-3
Figure 2-2. SCXI-1162HV Parts Locator Diagram.............................................................2-4
Figure 3-1. SCXI-1162HV Front Connector Pin Assignment.............................................3-3
Figure 3-2. Signal Connection Example..............................................................................3-6
Figure 3-3. SCXI-1162HV Rear Signal Connector Pin Assignments.................................3-7
Tables
Table 2-1. Digital Signal Connections, Jumper Settings....................................................2-5
SCXI-1162HV User Manual vi © National Instruments Corporation
Page 7
About This Manual
This manual describes the electrical and mechanical aspects of the SCXI-1162HV and contains information concerning its operation. The SCXI-1162HV is a member of the National Instruments Signal Conditioning eXtensions for Instrumentation (SCXI) Series modules for the National Instruments plug-in DAQ boards. This module switches and controls low-level and power signals. The SCXI-1162HV operates as 32 digital input channels, arranged as eight banks of four channels. Each bank is optically isolated.
Organization of This Manual
The SCXI-1162HV User Manual is organized as follows:
Chapter 1, Introduction, describes the SCXI-1162HV; lists what you need to get started with your SCXI-1162HV; describes the software programming choices and optional equipment; and explains how to unpack the SCXI-1162HV kit.
Chapter 2, Configuration and Installation, describes the SCXI-1162HV jumper configurations and installation of the SCXI-1162HV into the SCXI chassis.
Chapter 3, Signal Connections, describes the SCXI-1162HV signal connections via the front and rear connectors, and has specifications and connection instructions for the SCXI­1162HV connector signals.
Appendix A, Specifications, lists the specifications for the SCXI-1162HV.
Appendix B, Customer Communication, contains forms you can use to request help from National Instruments or to comment on our products.
The Glossary contains an alphabetical list and description of terms used in this manual, including abbreviations, acronyms, metric prefixes, mnemonics, and symbols.
The Index contains an alphabetical list of key terms and topics in this manual, including the page where you can find each one.
Conventions Used in This Manual
The following conventions are used in this manual: bold italic Bold italic text denotes a note, caution, or warning. DIO board DIO board refers to the National Instruments AT-DIO-32F, MC-DIO-24,
MC-DIO-32F, NB-DIO-24, NB-DIO-96, NB-DIO-32F, PC-DIO-24, and PC-DIO-96 digital I/O DAQ boards unless otherwise noted.
© National Instruments Corporation vii SCXI-1162HV User Manual
Page 8
About This Manual
DIO-24 board DIO-24 board refers to the National Instruments MC-DIO-24,
NB-DIO-24, and PC-DIO-24 digital I/O DAQ boards.
DIO-type board DIO-type board refers to National Instruments DAQ boards that have only
digital inputs and outputs. These boards include the DIO-24, DIO-32F, and DIO-96 boards unless otherwise noted.
italic Italic text denotes emphasis, a cross reference, or an introduction to a key
concept.
Lab board Lab board refers to the National Instruments Lab-LC, Lab-NB, Lab-PC,
and Lab-PC+ boards unless otherwise noted. MC MC refers to the Micro Channel Series computers. MIO board MIO board refers to the National Instruments AT-MIO-16, AT-MIO-16D,
AT-MIO-16F-5, AT-MIO-16X, AT-MIO-64F-5, MC-MIO-16,
NB-MIO-16, and NB-MIO-16X multichannel I/O DAQ boards unless
otherwise noted. MIO-type board MIO-type board refers to National Instruments DAQ boards that have at
least analog and digital inputs and outputs. These boards include the MIO
boards, the Lab boards, and the PC-LPM-16 board unless otherwise noted. monospace Lowercase text in this font denotes text or characters that are to be literally
input from the keyboard, sections of code, programming examples, and
syntax examples. This font is also used for the proper names of disk
drives, paths, directories, programs, subprograms, subroutines, device
names, functions, variables, filenames, and extensions, and for statements
and comments taken from program code. NB NB refers to the NuBus Series computers. PC PC refers to the IBM PC/XT, the IBM PC AT, and compatible computers. SCXIbus SCXIbus refers to the backplane in the chassis. A signal on the backplane
is referred to as the SCXIbus <signal name> line (or signal). The
SCXIbus descriptor may be omitted when the meaning is clear. Slot 0 Slot 0 refers to the power supply and control circuitry in the SCXI chassis. Abbreviations, acronyms, metric prefixes, mnemonics, symbols, and terms are listed in the
Glossary.
SCXI-1162HV User Manual viii © National Instruments Corporation
Page 9
About This Manual
The National Instruments Documentation Set
The SCXI-1162HV User Manual is one piece of the documentation set for your DAQ or SCXI system. You could have any of several types of manuals, depending on the hardware and software in your system. Use these different types of manuals as follows:
Getting Started with SCXI–If you are using SCXI, this is the first manual you should read. It
gives an overview of the SCXI system and contains the most commonly needed information for the modules, chassis, and software.
Your SCXI hardware user manuals–If you are using SCXI, read these manuals next for
detailed information about signal connections and module configuration. They also explain in greater detail how the module works and contain application hints.
Your DAQ hardware user manuals–These manuals have detailed information about the DAQ
hardware that plugs into or is connected to your computer. Use these manuals for hardware installation and configuration instructions, specification information about your DAQ hardware, and application hints.
Software manuals–Examples of software manuals you may have are the LabVIEW and
LabWindows system, use either the application software (LabVIEW or LabWindows) manuals or the NI-DAQ manuals to help you write your application. If you have a large and complicated system, it is worthwhile to look through the software manuals before you configure your hardware.
®
manual sets and the NI-DAQ manuals. After you have set up your hardware
Accessory installation guides or manuals–If you are using accessory products, read the
terminal block and cable assembly installation guides or accessory board user manuals. They explain how to physically connect the relevant pieces of the system. Consult these guides when you are making your connections.
SCXI chassis manuals–If you are using SCXI, read these manuals for maintenance
information on the chassis and installation instructions.
Related Documentation
In addition, the following National Instruments manual contains detailed information for the register-level programmer.
SCXI-1162HV Register-Level Programmer Manual This manual is available from National Instruments by request. If you are using NI-DAQ,
LabVIEW, or LabWindows, you should not need the register-level programmer manual. Using NI-DAQ, LabVIEW, or LabWindows is quicker and easier than and as flexible as using the low­level programming described in the register-level programmer manual. Refer to Software Programming Choices in Chapter 1, Introduction, of this manual to learn about your programming options.
© National Instruments Corporation ix SCXI-1162HV User Manual
Page 10
About This Manual
Customer Communication
National Instruments wants to receive your comments on our products and manuals. We are interested in the applications you develop with our products, and we want to help if you have problems with them. To make it easy for you to contact us, this manual contains comment and configuration forms for you to complete. These forms are in Appendix B, Customer
Communication, at the end of this manual.
SCXI-1162HV User Manual x © National Instruments Corporation
Page 11
Chapter 1 Introduction
This chapter describes the SCXI-1162HV; lists what you need to get started with your SCXI-1162HV; describes the software programming choices and optional equipment; and explains how to unpack the SCXI-1162HV kit.
About the SCXI-1162HV
The SCXI-1162HV module consists of 32 optically isolated, wide range, AC or DC digital inputs. The SCXI-1162HV is a module for sensing the presence of AC or DC voltages where high common-mode voltages may be present.
The SCXI-1162HV works with National Instruments MIO Series boards; the SCXI-1200 module; Lab series boards; and with the DIO-24, DIO-32F, and DIO-96 boards. You can read 24 inputs in parallel mode with a DIO-24 board, or all 32 inputs with a DIO-32F or a DIO-96 board. You can also use the SCXI-1162HV in parallel mode with other digital logic inputs that comply with the specifications in Chapter 3, Signal Connections. You can control several SCXI-1162HVs in series mode in a single chassis with one DAQ board.
An additional shielded terminal block, the SCXI-1326, has screw terminals for easy signal attachment to the SCXI-1162HV.
With the SCXI-1162HV, the SCXI chassis can serve as a digital signal monitor in laboratory testing, production testing, and industrial process monitoring.
What You Need to Get Started
To set up and use your SCXI-1162HV, you will need the following:
SCXI-1162HV module
SCXI-1162HV User Manual
SCXI chassis
Software Programming Choices
There are four options to choose from when programming your National Instruments DAQ and SCXI hardware. You can use LabVIEW, LabWindows, NI-DAQ, or register-level programming software.
The SCXI-1162HV works with LabVIEW for Windows, LabVIEW for Macintosh, LabWindows for DOS, LabWindows/CVI for Windows, NI-DAQ software for PC compatibles.
© National Instruments Corporation 1-1 SCXI-1162HV User Manual
Page 12
Introduction Chapter 1
LabVIEW and LabWindows Application Software
LabVIEW and LabWindows are innovative program development software packages for data acquisition and control applications. LabVIEW uses graphical programming, whereas LabWindows enhances traditional programming languages. Both packages include extensive libraries for data acquisition, instrument control, data analysis, and graphical data presentation.
LabVIEW currently runs on three different platforms—AT/MC/EISA computers running Microsoft Windows, the Macintosh platform, and the Sun SPARCstation platform. LabVIEW features interactive graphics, a state-of-the-art user interface, and a powerful graphical programming language. The LabVIEW Data Acquisition VI Library, a series of VIs for using LabVIEW with National Instruments DAQ hardware, is included with LabVIEW. The LabVIEW Data Acquisition VI Libraries are functionally equivalent to the NI-DAQ software, except that the SCXI functions are not included in the LabVIEW software for Sun.
LabWindows has two versions—LabWindows for DOS is for use on PCs running DOS, and LabWindows/CVI is for use on PCs running Windows and for Sun SPARCstations. LabWindows/CVI features interactive graphics, a state-of-the-art user interface, and uses the ANSI standard C programming language. The LabWindows Data Acquisition Library, a series of functions for using LabWindows with National Instruments DAQ hardware, is included with the NI-DAQ software kit. The LabWindows Data Acquisition libraries are functionally equivalent to the NI-DAQ software, except that the SCXI functions are not included in the LabWindows/CVI software for Sun.
Using LabVIEW or LabWindows software will greatly reduce the development time for your data acquisition and control application.
NI-DAQ Driver Software
The NI-DAQ driver software is included at no charge with all National Instruments DAQ hardware. NI-DAQ is not packaged with SCXI or accessory products except for the SCXI-1200. NI-DAQ has an extensive library of functions that you can call from your application programming environment. These functions include routines for analog input (A/D conversion), buffered data acquisition (high-speed A/D conversion), analog output (D/A conversion), waveform generation, digital I/O, counter/timer operations, SCXI, RTSI, self-calibration, messaging, and acquiring data to extended memory.
NI-DAQ also internally addresses many of the complex issues between the computer and the DAQ hardware such as programming interrupts and DMA controllers. NI-DAQ maintains a consistent software interface among its different versions so that you can change platforms with minimal modifications to your code. Figure 1-1 illustrates the relationship between NI-DAQ and LabVIEW and LabWindows. You can see that the data acquisition parts of LabVIEW and LabWindows are functionally equivalent to the NI-DAQ software.
SCXI-1162HV User Manual 1-2 © National Instruments Corporation
Page 13
Chapter 1 Introduction
Conventional
Programming
Environment
(PC, Macintosh, or
Sun SPARCstation)
LabVIEW
(PC, Macintosh, or
Sun SPARCstation)
NI-DAQ
Driver Software
LabWindows
(PC or
Sun SPARCstation)
DAQ or
SCXI Hardware
Personal
Computer
or
Workstation
Figure 1-1. The Relationship between the Programming Environment,
NI-DAQ, and Your Hardware
The National Instruments PC, AT, MC, EISA, DAQCard, and DAQPad Series DAQ hardware are packaged with NI-DAQ software for PC compatibles. NI-DAQ software for PC compatibles comes with language interfaces for Professional BASIC, Turbo C++, Borland C++, and Microsoft C for DOS; and Visual Basic, Turbo Pascal, Microsoft C with SDK, and Borland C++ for Windows. You can use your SCXI-1163R, together with other PC, AT, MC, DAQCard, and SCXI hardware, with NI-DAQ software for PC compatibles.
The National Instruments NB Series DAQ boards are packaged with NI-DAQ software for Macintosh. NI-DAQ software for Macintosh comes with language interfaces for MPW C, THINK C, Pascal, and Microsoft QuickBASIC. Any language that uses Device Manager Toolbox calls can access NI-DAQ software for Macintosh. You can use NB Series DAQ boards and SCXI hardware with NI-DAQ software for Macintosh.
The National Instruments SB Series DAQ boards are packaged with NI-DAQ software for Sun, which comes with a language interface for ANSI C.
© National Instruments Corporation 1-3 SCXI-1162HV User Manual
Page 14
Introduction Chapter 1
Register-Level Programming
The final option for programming any National Instruments DAQ hardware is to write register­level software. Writing register-level programming software can be very time-consuming and inefficient, and is not recommended for most users. The only users who should consider writing register-level software should meet at least one of the following criteria:
National Instruments does not support your operating system or programming language.
You are an experienced register-level programmer who is more comfortable writing your
own register-level software.
Even if you are an experienced register-level programmer, always consider using NI-DAQ, LabVIEW, or LabWindows to program your National Instruments DAQ hardware. Using the NI-DAQ, LabVIEW, or LabWindows software is easier than, and as flexible as, register-level programming, and can save you weeks of development time.
The SCXI-1162HV User Manual and your software manuals contain complete instructions for programming your SCXI-1162HV with NI-DAQ, LabVIEW, or LabWindows. If you are using NI-DAQ, LabVIEW, or LabWindows to control your board, you should not need the register­level programmer manual.
The SCXI-1162HV Register-Level Programmer Manual contains low-level programming details, such as register maps, bit descriptions, and register programming hints, that you will need only for register-level programming. If you want to obtain the register-level programmer manual, please fill out the Register-Level Programmer Manual Request Form in Appendix B, Customer Communication, at the end of this manual and send it to National Instruments.
Optional Equipment
Contact National Instruments to order any of the following optional equipment:
SCXI-1326 front terminal block SCXI-1340 cable assembly
SCXI-1341 Lab-NB/Lab-PC/Lab-PC+ cable assembly
SCXI-1342 PC-LPM-16 cable assembly
SCXI-1343 rear screw terminal adapter
SCXI-1344 Lab-LC cable assembly
SCXI-1350 multichassis adapter
SCXI-1351 one slot extender cable
SCXI-1162HV User Manual 1-4 © National Instruments Corporation
Page 15
Chapter 1 Introduction
SCXI-1348 DIO-32F cable assembly
Standard ribbon cable in lengths of 0.5 or 1.0 m
NB5 cable in lengths of 0.5 or 1.0 m
Refer to Chapter 3, Signal Connections, and to your cable or accessory installation guide for additional information on cabling, connectors, and adapters.
Custom Cables
The SCXI-1162HV rear signal connector is a 50-pin male ribbon-cable header. The manufacturer part number for the header National Instruments uses is as follows:
AMP Inc. (part number 1-103310-0) The mating connector for the SCXI-1162HV rear signal connector is a 50-position polarized
ribbon-socket connector with strain relief. National Instruments uses a polarized or keyed connector to prevent inadvertent upside-down connection to the SCXI-1162HV. Recommended manufacturer part numbers for this mating connector are as follows:
Electronic Products Division/3M (part number 3425-7650)
T&B/Ansley Corporation (part number 609-5041CE) Standard 50-conductor 28 AWG stranded ribbon cables that work with these connectors are as
follows:
Electronic Products Division/3M (part number 3365/50)
T&B/Ansley Corporation (part number 171-50) The SCXI-1162HV front connector is a special 48-pin DIN C male connector. The manufacturer
part number of the connector National Instruments uses is as follows:
ERNI Components, Inc. (part number 033-273) The mating connector for the SCXI-1162HV front connector is a special 48-pin reversed DIN C
female connector. National Instruments uses a polarized and keyed connector to prevent inadvertent upside-down connection to the SCXI-1162HV. The manufacturer part number of the mating connector National Instruments uses is as follows:
ERNI Components, Inc. (part number 913-271) These connectors were selected to meet UL 1244 for 450 Vrms working isolation.
© National Instruments Corporation 1-5 SCXI-1162HV User Manual
Page 16
Introduction Chapter 1
Unpacking
Your SCXI-1162HV module is shipped in an antistatic package to prevent electrostatic damage to the module. Electrostatic discharge can damage several components on the module. To avoid such damage in handling the module, take the following precautions:
Ground yourself via a grounding strap or by holding a grounded chassis such as an
SCXI chassis.
Touch the antistatic package to a metal part of your SCXI chassis before removing the
module from the package.
Remove the module from the package and inspect the module for loose components or any
other sign of damage. Notify National Instruments if the module appears damaged in any way. Do not install a damaged module into your SCXI chassis.
Never touch the exposed pins of connectors.
SCXI-1162HV User Manual 1-6 © National Instruments Corporation
Page 17
Chapter 2 Configuration and Installation
This chapter describes the SCXI-1162HV jumper configurations and installation of the SCXI-1162HV into the SCXI chassis.
Module Configuration
The SCXI-1162HV has six user-configurable jumpers and one reserved jumper that are shown in the parts locator diagram in Figure 2-1. You use these jumpers to configure the SCXI-1162HV rear connector.
Notes: When you are controlling the SCXI-1162HV through the SCXIbus and are not using
the rear connector (not cabled directly to a plug-in DAQ board), the positions of the configuration jumpers are irrelevant.
If you want to use a single DAQ board to control a chassis which contains analog modules in addition to your SCXI-1162HV, then you must cable your DAQ board to one of the analog modules rather than to the SCXI-1162HV.
You can configure the SCXI-1162HV rear connector to send data directly from the inputs in parallel, or to form a serial communication link with the SCXIbus. In the latter case, you can also configure the rear connector to connect to a DIO-type or an MIO-type DAQ board.
You must configure the module in the chassis that is cabled to the DAQ board to allow reads from module registers. (Specifically, the module must be configured to send data from the SCXIbus MISO line to the SERDATOUT pin of the module rear connector. Refer to the user manual for the SCXI module you are cabling to a DAQ board for information on how to accomplish this.)
Configuration Options
The SCXI-1162HV is shipped configured for serial communication through a DIO-type board or the SCXI chassis. Change the jumper configuration from the factory default settings only if you plan to cable the SCXI-1162HV rear connector to an MIO-type board, or to use the SCXI-1162HV in parallel mode, or if the SCXI-1162HV is a controller module for a multichassis system.
© National Instruments Corporation 2-1 SCXI-1162HV User Manual
Page 18
Configuration and Installation Chapter 2
Configuring the Rear Connector for Parallel Mode
When the rear connector is configured for parallel mode, the data from each input is sent directly to a rear connector pin. You do not need serial communication with the SCXI chassis or modules to read the data when the module is in parallel mode. This mode may be useful when you want to send data from the SCXI-1162HV directly to equipment other than a DAQ board.
When the SCXI-1162HV is configured for parallel mode, a DAQ board cabled to the module can only read from that module; communication with the SCXI chassis or other modules in that chassis must be through another DAQ board. Only DIO DAQ boards can read the data from the SCXI-1162HV in parallel mode; MIO boards can only read from the SCXI-1162HV in serial mode.
When you configure the SCXI-1162HV for parallel mode, the module can drive either all 32 digital lines of the rear connector or only the first 24 lines. When you cable a DIO-24 or the DIO portion of an AT-MIO-16D to an SCXI-1162HV in parallel mode, the SCXI-1162HV must be configured to drive only 24 of the rear connector lines, because the DIO-24 and the AT-MIO-16D only have 24 digital inputs. Do not attempt to connect the SCXI-1162HV to a DIO-24 or an AT-MIO-16D with the module configured to drive all 32 lines in parallel, as this may cause driver contention.
Note: When the rear connector is configured for parallel mode, the data can still be read in
serial mode through the SCXIbus connector. The jumper set configuration only affects the rear signal connector; no other module functionality is affected by the jumper positions.
Configuring the Rear Connector for Serial Mode
When the rear connector is configured for serial mode, there is a serial data link between the rear connector and the SCXIbus. This lets a DAQ board cabled to the rear connector communicate with the SCXI-1162HV and with all other modules in the chassis. This is the recommended mode of operation because you only need one DAQ board for an entire SCXI system.
When the SCXI-1162HV is configured for serial mode, only digital data is transferred between the SCXIbus and the rear signal connector. Thus, in a mixed system of analog modules and digital modules, the DAQ board must be cabled to an analog module. If the DAQ board were cabled to a digital module such as the SCXI-1162HV, the analog signals from analog modules could not reach the DAQ board.
SCXI-1162HV User Manual 2-2 © National Instruments Corporation
Page 19
Chapter 2 Configuration and Installation
Jumper Use
To configure the SCXI-1162HV module, use the six user-configurable jumpers (W2–W7) shown in the parts locator diagram, Figure 2-2. (Jumper W1 is a reserved jumper and should remain unconnected.) If you are controlling the SCXI-1162HV via the SCXIbus and are not using the rear connector, then the positions of these jumpers are irrelevant. Tables 2-1 and 2-2 list the description and configuration of the user-configurable jumpers. To change the configuration of the module, refer to Figures 2-1 and 2-2 as you perform the following steps:
1. Remove the grounding screw of the top cover.
2. Snap out the top cover of the shield by placing a screwdriver in the groove at the bottom of
the module.
3. Remove the jumpers to be changed and replace them on the appropriate pins.
4. Record the new jumper settings on the SCXI-1162HV Hardware Configuration Form in
Appendix B, Customer Communication.
5. Snap the top cover back in place.
6. Replace the grounding screw to ensure proper shielding.
Removable Cover
Top of Module
450 V MAX BANK-TO-BANK
SCXI
1162HV
450 V MAX BANK-TO-EARTH
32-CHANNEL
DIGITAL INPUT
HIGH VOLTAGE
Grounding screw
(on rear panel)
Front Connector
Figure 2-1. Removing the SCXI Module Cover
© National Instruments Corporation 2-3 SCXI-1162HV User Manual
Page 20
Configuration and Installation Chapter 2
Figure 2-2. SCXI-1162HV Parts Locator Diagram
SCXI-1162HV User Manual 2-4 © National Instruments Corporation
Page 21
Chapter 2 Configuration and Installation
Table 2-1. Digital Signal Connections, Jumper Settings
Jumper Description Configuration
24
32
32
P
P
DIO
24
S
S
MIO
W6
Factory setting—Drives only 24 lines when W2 is
W5
set to PAR. Used with DIO-24 or MIO-16D.
Drives all 32 rear connector data lines when W2 is set to PAR. Used with DIO-32F or DIO-96.
Factory setting—Configures rear connector for serial communication.
Configures rear connector for parallel communication. DIO boards only.
Factory setting—Configures rear connector for a DIO board.
W4
W7
Configures rear connector for an MIO board. Includes MIO Series and Lab Series boards, and PC-LPM-16.
Factory setting—Configures rear connector for a DIO board.
Configures rear connector for an MIO board. Includes MIO series, Lab series, and PC-LPM-16.
MIO
DIO
DIO
MIO
DIO
MIO
(continues)
© National Instruments Corporation 2-5 SCXI-1162HV User Manual
Page 22
Configuration and Installation Chapter 2
Table 2-1. Digital Signal Connections, Jumper Settings (Continued)
Jumper Description Configuration
P
W3
Position P (pull-up)—Use this setting for a single-
NP
chassis system or for the first chassis in a
NP
P
multichassis system. Connects a pull-up resistor to SERDATOUT. (Factory setting)
Position NP (no pull-up)—Use this setting for additional chassis in a multichassis system. No pull­up resistor is connected to SERDATOUT.
W2
Parking position—Disconnects SERDATOUT line. Use for parallel mode.
Configures SERDATOUT for MIO-type board. Includes MIO series, Lab series, and PC-LPM-16.
DIO
DIO
PAR
PAR
MIO
MIO
PAR
Factory setting—Configures SERDATOUT for DIO-
DIO
type board.
MIO
Note: The shaded area indicates the position of the jumper. Supplementary Jumper W3 Configuration Information
Jumper W3, when set to position P, connects a 2.2 k pull-up resistor to the SERDATOUT line. An open-collector driver actively drives low or goes to a high-impedance state, relying on a pull­up resistor to make the signal line go high. If too many pull-up resistors are attached to the SERDATOUT line, the drivers cannot drive the line low. To prevent this, set jumper W3 to position P on only one of the SCXI-1162HVs that are cabled to the DAQ board in a multichassis system. It does not matter which of the SCXI-1162HVs cabled to the DAQ board has the pull-up connected.
Hardware Installation
You can install the SCXI-1162HV in any available SCXI chassis. After you have made any necessary changes and have verified and recorded the jumper settings on the form in Appendix B, Customer Communication, you are ready to install the SCXI-1162HV. The following are general installation instructions; consult the user manual or technical reference manual of your SCXI chassis for specific instructions and warnings.
1. Turn off the computer that contains the DAQ board or disconnect it from your
SCXI chassis.
2. Turn off the SCXI chassis. Do not insert the SCXI-1162HV into a chassis that is turned on.
SCXI-1162HV User Manual 2-6 © National Instruments Corporation
Page 23
Chapter 2 Configuration and Installation
3. Insert the SCXI-1162HV into the board guides. Gently guide the module into the back of the
slot until the connectors make good contact. If you have already installed a cable assembly in the rear of the chassis, you must firmly engage the module and cable assembly; however, do not force the module into place.
4. Screw the front mounting panel of the SCXI-1162HV to the top and bottom threaded strips
of your SCXI chassis.
5. If this module is to be connected to an MIO-16 or a DIO-24 DAQ board, attach the connector
at the metal end of the SCXI-1340 cable assembly to the rear signal connector on the SCXI-1162HV module. Screw the rear panel to the rear threaded strip. Attach the loose end of the cable to the DAQ board.
Note: For installation procedures with other SCXI accessories, consult the appropriate
installation guides.
6. Check the installation.
7. Turn on the SCXI chassis.
8. Turn on the computer or reconnect it to your chassis. The SCXI-1162HV board is installed. You are now ready to install and configure your software.
Software Installation
If you are using NI-DAQ, refer to the NI-DAQ User Manual for PC Compatibles. The software installation and configuration instructions are in Chapter 1, Introduction to NI-DAQ. Find the installation and system configuration section for your operating system and follow the instructions given there.
If you are using LabVIEW, the software installation instructions are in your LabVIEW release notes. After you have installed LabVIEW, refer to the Configuring LabVIEW section of Chapter 1 of your LabVIEW user manual for software configuration instructions.
If you are using LabWindows, the software installation instructions are in Part 1, Introduction to LabWindows, of the Getting Started with LabWindows manual. After you have installed LabWindows, refer to Chapter 1, Configuring LabWindows, of the LabWindows User Manual for software configuration instructions.
If you are a register-level programmer, refer to the SCXI-1162HV Register-Level Programmer
Manual.
© National Instruments Corporation 2-7 SCXI-1162HV User Manual
Page 24
Chapter 3 Signal Connections
This chapter describes the SCXI-1162HV signal connections via the front and rear connectors, and has specifications and connection instructions for the SCXI-1162HV connector signals.
The following warnings contain important safety information concerning hazardous voltages.
Warning: Connections that exceed any of the maximum ratings of input signals on the
SCXI-1162HV can damage the SCXI-1162HV board and the SCXIbus. Maximum input ratings for each signal are given in this chapter under the discussion of that signal. National Instruments is not liable for any damages resulting from signal connections that exceed these ratings.
K
EEP AWAY FROM LIVE CIRCUITS. Do not remove equipment covers or shields
unless you are trained to do so. If signal wires are connected to the module or terminal block, dangerous voltages may exist even when the equipment is turned off. To avoid dangerous electrical shock, do not perform procedures involving cover or shield removal unless you are qualified to do so.
D
O NOT OPERATE DAMAGED EQUIPMENT. The safety protection features built
into this module can become impaired if the module becomes damaged in any way. If it is damaged, turn the module off and do not use until service-trained personnel can check its safety. If necessary, return the module to National Instruments for service and repair to ensure that its safety is not compromised.
D
O NOT SUBSTITUTE PARTS OR MODIFY EQUIPMENT. Because of the danger of
introducing additional hazards, do not install unauthorized parts or modify the module. Return the module to National Instruments for service and repair to ensure that its safety features are not compromised.
When using the terminal block with high common-mode voltages, you insulate your signal wires appropriately. National Instruments is NOT liable for any damages or injuries resulting from inadequate signal wire insulation.
When connecting or disconnecting signal lines to the SCXI-1326 terminal block screw terminals, make sure the lines are powered off to prevent shock hazard.
MUST
© National Instruments Corporation 3-1 SCXI-1162HV User Manual
Page 25
Signal Connections Chapter 3
Connections, including power signals to ground and vice versa, that exceed any of the maximum signal ratings on the SCXI-1162HV can damage any or all of the boards connected to the SCXI chassis, the host computer, and the SCXI-1162HV module. National Instruments i
DAMAGES OR INJURIES resulting from incorrect signal connections.
If high voltages (
SAFETY EARTH GROUND TO THE STRAIN-RELIEF TAB. This complies with
42 Vrms or 60 VDC) are present, YOU MUST CONNECT THE
S NOT LIABLE FOR ANY
UL 1244 and protects against electric shock when the terminal block is not connected to the chassis. To connect the safety earth ground to the strain-relief tab, run an earth ground wire in the cable from the signal source to the terminal block. National Instruments is
NOT liable for any damages or injuries
resulting from inadequate safety earth ground connections.
SCXI-1162HV User Manual 3-2 © National Instruments Corporation
Page 26
Chapter 3 Signal Connections
Front Connector
Figure 3-1 shows the pin assignments for the SCXI-1162HV front connector.
Pin
Number
30
29
26
25
22
21
Signal Name
In 10 In 11
In 0 In 1 In 2 In 3
In 4 In 5 In 6 In 7
In 8 In 9
Column
A B C
Signal Name
NC
GND
NC
GND
NC
GND
18
17
14
13
10
In 12 In 13 In 14 In 15
In 19 In 18 In 17 In 16
In 23 In 22 In 21
9
6
5
2
1
In 20
In 27 In 26 In 25 In 24
In 31 In 30 In 29 In 28
NC
GND
GND
NC
GND
NC
GND
NC
GND
NC
Figure 3-1. SCXI-1162HV Front Connector Pin Assignment
© National Instruments Corporation 3-3 SCXI-1162HV User Manual
Page 27
Signal Connections Chapter 3
Front Connector Signal Descriptions
Pin Signal Description
B30, A30, B29, A29, C29
B26, A26, B25, A25, C25
B22, A22, B21, A21, C21
B18, A18, B17, A17, C17
A13, B13, A14, B14, C14
A9, B9, A10, B10, C10
A5, B5, A6, B6, C6
A1, B1, A2, B2, C2
C1, C5, C9, C13, C18, C22, C26, C30
In <0..3>, GND Bank 0 input channels and ground reference.
In <4..7>, GND Bank 1 input channels and ground reference.
In <8..11>, GND Bank 2 input channels and ground reference.
In <12..15>, GND Bank 3 input channels and ground reference.
In <16..19>, GND Bank 4 input channels and ground reference.
In <20..23>, GND Bank 5 input channels and ground reference.
In <24..27>, GND Bank 6 input channels and ground reference.
In <28..32>, GND Bank 7 input channels and ground reference.
NC (No Connect) Unused pins.
Input Channels
The optically isolated inputs of the SCXI-1162HV consist of bidirectional light-emitting diodes and current-limiting circuitry. The SCXI-1162HV has eight banks of four of these optically isolated inputs. Each bank of inputs has a separate connection for the ground reference (GND) for that bank. The four inputs of each bank are referenced to the GND of that bank, so for any bank of inputs to work, a GND must be connected to that bank. When a voltage is applied between an input and its ground, signal current flows through the input and the diode emits light that is sensed by a photosensitive circuit. When this circuit senses the light, it returns a logic­high digital signal.
Sensing DC Voltages
When a positive or negative DC voltage with a magnitude of at least 2 V is referenced to the GND of a bank and is applied to an input of that bank, the SCXI-1162HV registers a logic high for that input. Current-limiting circuitry limits the input current to 1.5 mA for input voltages up to 240 VDC referenced to the GND signal. If there is no such voltage, the SCXI-1162HV will register a logic low for that input. Thus, you can use the SCXI-1162HV to sense the presence of a wide range of DC signals; from digital logic levels to DC power supply levels up to 240 V.
There is a delay of approximately 1 ms between the time a voltage is removed and the time the SCXI-1162HV recognizes this change.
SCXI-1162HV User Manual 3-4 © National Instruments Corporation
Page 28
Chapter 3 Signal Connections
Sensing AC Voltages
The SCXI-1162HV senses a wide range of AC signals by registering a constant logic high while an AC voltage (referenced to a GND) is present at an input. Signals with low amplitude and low frequency appear as signals that are alternately turned on and off; therefore the module alternately registers logic highs and logic lows for that signal. For sinusoidal signals, a 50 or 60 Hz signal with a voltage of at least 20 Vrms returns a constant logic high level. For 1 kHz and higher frequency signals, a 2 Vrms signal returns a constant logic high level.
The current-limiting circuitry limits the current to 1.5 mA for input voltages up to 240 Vrms for a sinusoidal signal.
Signal Isolation
The GND and inputs of each bank are isolated from the GND and inputs of any other bank. In addition, all signals are isolated from the chassis earth ground and the internal circuitry of the module. These barriers provide isolation for voltages up to 450 VAC or VDC and protect the user and the SCXI system. There is no isolation between signals of the same bank. You must not exceed the voltage limits of the inputs referenced to their respective GND signals.
Signal Connection Example
In Figure 3-2, one channel of the SCXI-1162HV is being used to sense that a load is being powered. The load is connected to a power supply by means of a switch. This power supply can be AC or DC and can be any voltage within the SCXI-1162HV range. Common supplies may include +5 VDC, -48 VDC, or 120 VAC 60 Hz. When the switch is open, no current flows through the load and no voltage is applied to the load or to the SCXI-1162HV input. The digital logic of the SCXI-1162HV then registers a logic low or 0 for the channel. When the switch is closed, the power supply voltage is applied across both the load and the input, and the SCXI-1162HV registers a logic high or 1 for the channel.
© National Instruments Corporation 3-5 SCXI-1162HV User Manual
Page 29
Signal Connections Chapter 3
High common-mode voltage
Power Supply
+
-
Up to 450 VAC or VDC
Load
In
GND
SCXI-1162 HV
1.5 mA Current Limiter
Bidirectional
V
in
LED
Optical isolation barrier
Figure 3-2. Signal Connection Example
Photosensitive digital logic
Logic high when
> 2 V
V
in
Logic low when
V
= 0
in
Terminal Block
The SCXI-1326 terminal block has screw terminals for easy signal connection to the SCXI-1162HV inputs. This kit is listed in the Optional Equipment section in Chapter 1, Introduction. See the SCXI-1326 Terminal Block Installation Guide for more information on connecting it to your SCXI-1162HV.
Rear Signal Connector
Note: If you are using the SCXI-1162HV with a National Instruments DAQ board and cable
assembly, you do not need to read the remainder of this chapter. If you are using the SCXI-1180 feedthrough panel, the SCXI-1343 rear screw terminal adapter, or the SCXI-1351 one-slot cable extender with the SCXI-1162HV, read this section.
The rear signal connector is configured based on the jumper settings described earlier in this chapter. Figure 3-3A shows the pin assignments for the SCXI-1162HV rear signal connector configured for serial communication. Figure 3-3B shows the pin assignments for the SCXI-1162HV rear signal connector configured for parallel communication.
SCXI-1162HV User Manual 3-6 © National Instruments Corporation
Page 30
Chapter 3 Signal Connections
1 2 3 4
Out 23 Out 22
1 2 3 4
SERDATIN
DAQD*/A
SLOT0SEL*
(DIO) SERCLK
(MIO) SERCLK
5 6 7 8
9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26
27 28 29 30
31 32 33 34 35 36 37 38
DIGGND (MIO) SERDATOUT (MIO)
Out 21 Out 20
Out 19
Out 18
Out 17
Out 16
Out 15
Out 14
Out 13
Out 12 Out 11
Out 10
Out 9 Out 8 Out 7 Out 6 Out 5
5 6 7 8
9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26
27 28 29 30
31 32 33 34 35 36 37 38
Out Out
Out
(DIO) SERDATOUT
A. Jumper W2 in SER Position
Figure 3-3. SCXI-1162HV Rear Signal Connector Pin Assignments
39 40 41 42 43 44 45 46 47 48 49 50
DIGGND (DIO)
Out 4 Out 3 Out 2 Out 1 Out 0
39 40 41 42 43 44
45 46 47 48
49 50
Out Out Out Out
Out
GND
B. Jumper W2 in PAR Position
With W1
set to 32
© National Instruments Corporation 3-7 SCXI-1162HV User Manual
Page 31
Signal Connections Chapter 3
Rear Signal Connector Signal Descriptions, Serial Configuration
Pin Signal Description
24 or 50 DIGGND Digital Ground—Supplies the reference for DAQ
board digital signals and is tied to the module digital ground. Pin 50 is for DIO boards. Pin 24 is for MIO boards and jumper W4 selects it.
25 SERDATIN Serial Data In—Taps into the SCXIbus MOSI line
to provide serial input data to a module or Slot 0.
26 or 47 SERDATOUT Serial Data Out—Taps into the SCXIbus MISO
line to accept serial output data from a module. Pin 47 is for DIO boards. Pin 26 is for MIO boards and jumper W6 selects it.
27 DAQD*/A DAQ Board Data/Address Line—Taps into the
SCXIbus D*/A line to indicate to the module whether the incoming serial stream is data or address information.
29 SLOT0SEL* Slot 0 Select—Taps into the SCXIbus INTR* line
to indicate whether the information on MOSI is sent to a module or to Slot 0.
31 or 37 SERCLK Serial Clock—Taps into the SCXIbus SPICLK
line to clock the data on the MOSI and MISO lines. Pin 31 is for DIO boards. Pin 37 is for MIO boards and jumper W3 selects it.
* Indicates active low. All other pins are not connected. See Chapter 2, Timing, of the SCXI-1162HV Register-Level Programmer Manual for
information on serial communication and programming.
SCXI-1162HV User Manual 3-8 © National Instruments Corporation
Page 32
Chapter 3 Signal Connections
Rear Signal Connector Signal Descriptions, Parallel Configuration
Pin Signal Description
1, 3, 5, 7, 9, 11, 13, 15,
Out <23..0> Parallel outputs driven to the logic levels seen at
inputs 23 through 0. 17, 19, 21, 23, 25, 27, 29, 31, 33, 35, 37, 39, 41, 43, 45, 47
34, 36, 38, 40, 42, 44,
Out <31..24> Parallel outputs driven to the logic levels seen at
inputs 31 through 24 when jumper W1 is set to 32. 46, 48
50 GND Ground—Supplies ground reference for the
parallel outputs.
All other pins are not connected. The following specifications apply to the digital output lines (referenced to DIGGND):
•V
•V
output logic high voltage 3.7 V min at 4 mA max
OH
output logic low voltage 0.4 V max at 4 mA max
OL
© National Instruments Corporation 3-9 SCXI-1162HV User Manual
Page 33
Appendix A Specifications
This appendix lists the specifications of the SCXI-1162HV. These specifications are typical at 25° C unless otherwise noted.
Number of channels .....................................................32 organized into 8 optically isolated banks of 4 inputs each
Digital logic levels ...................................................
Input low voltage
(DC or Peak AC)
Input high voltage
DC ±2 VDC ±240 VDC 50–60 Hz AC 10 VAC 240 VAC 1 kHz AC 2 VAC 240 VAC
Input current limiting ..............................................................1 mA
Common-mode isolation,
bank-to-bank and bank-to-ground
Working ..........................................................................450 Vrms max
Common-mode rejection.........................................................500 V/µs typ
Maximum speed
Propagation delay, parallel mode
(Vcc = 5 V) ......................................................................1.5 ms
Serial data rate .................................................................1,000 words/s
Level Min Max
±1 V
(VAC assumes sinusoidal waveforms)
1
Physical
Dimensions...................................................................1.2 by 6.8 in. by 8.0 in. (33.78 by 9.91 cm)
I/O connector................................................................50-pin male ribbon-cable rear connector
48-pin DIN C male front connector (SCXI-1326 screw terminal adapter available)
Environment
Operating temperature..................................................0˚ to 50˚ C
Storage temperature .....................................................-40˚ to 125˚ C
Relative humidity .........................................................5% to 90% noncondensing
1
Transfer rate depends on the computer and software. These tests were made using a 50 MHz 80486 computer
running NI-DAQ software.
© National Instruments Corporation A-1 SCXI-1162HV User Manual
Page 34
Appendix B Customer Communication
For your convenience, this appendix contains forms to help you gather the information necessary to help us solve technical problems you might have as well as a form you can use to comment on the product documentation. Filling out a copy of the Technical Support Form before contacting National Instruments helps us help you better and faster.
National Instruments provides comprehensive technical assistance around the world. In the U.S. and Canada, applications engineers are available Monday through Friday from 8:00 a.m. to 6:00 p.m. (central time). In other countries, contact the nearest branch office. You may fax questions to us at any time.
Corporate Headquarters
(512) 795-8248 Technical support fax: (800) 328-2203
(512) 794-5678
Branch Offices Phone Number Fax Number
Australia (03) 879 9422 (03) 879 9179 Austria (0662) 435986 (0662) 437010-19 Belgium 02/757.00.20 02/757.03.11 Denmark 45 76 26 00 45 76 71 11 Finland (90) 527 2321 (90) 502 2930 France (1) 48 14 24 00 (1) 48 14 24 14 Germany 089/741 31 30 089/714 60 35 Italy 02/48301892 02/48301915 Japan (03) 3788-1921 (03) 3788-1923 Netherlands 03480-33466 03480-30673 Norway 32-848400 32-848600 Spain (91) 640 0085 (91) 640 0533 Sweden 08-730 49 70 08-730 43 70 Switzerland 056/20 51 51 056/20 51 55 U.K. 0635 523545 0635 523154
© National Instruments Corporation B-1 SCXI-1162HV User Manual
Page 35
Technical Support Form
___________________________________________________
Photocopy this form and update it each time you make changes to your software or hardware, and use the completed copy of this form as a reference for your current configuration. Completing this form accurately before contacting National Instruments for technical support helps our applications engineers answer your questions more efficiently.
If you are using any National Instruments hardware or software products related to this problem, include the configuration forms from their user manuals. Include additional pages if necessary.
Name Company Address
Fax ( ) Phone ( ) Computer brand Model Processor
Operating system Speed MHz RAM MB Display adapter Mouse yes no Other adapters installed Hard disk capacity MB Brand Instruments used
National Instruments hardware product model Revision
Configuration
National Instruments software product Version
Configuration
The problem is
List any error messages
The following steps will reproduce the problem
Page 36
SCXI-1162HV Hardware Configuration Form
Record the settings and revisions of your hardware and software on the line to the right of each item. Complete a new copy of this form each time you revise your software or hardware configuration, and use this form as a reference for your current configuration. Completing this form accurately before contacting National Instruments for technical support helps our applications engineers answer your questions more efficiently.
SCXI-1162HV Revision
Chassis Slot
Chassis Type
Jumper W1
Jumper W2
Jumper W3
Jumper W4
Jumper W5
Jumper W6
My Setting
24
32
S
P
DIO
MIO
DIO
MIO
P
NP
PAR
DIO
MIO
Factory Setting
24
32
S
P
DIO
MIO
DIO
MIO
P
NP
PAR
MIO
DIO
Note: Mark your jumper positions on the jumper diagrams in the left column.
Other Modules and Chassis in System
DAQ Boards Installed
Page 37
Register-Level Programmer Manual Request Form
National Instruments offers a register-level programmer manual at no charge to customers who are not using National Instruments software.
Title: SCXI-1162HV Register-Level Programmer Manual Part Number: 340735A-01 Please indicate your reasons for obtaining the register-level programmer manual. Check all that apply.
National Instruments does not support your operating system or programming language.
You are an experienced register-level programmer who is more comfortable writing your own register-level software.
Other. Please explain.
Thank you for your help. Name Title Company Shipping Address (Street Address, Not P.O. Box) Phone ( )
Mail to: Data Entry Department Fax to: Data Entry Department
National Instruments Corporation National Instruments Corporation 6504 Bridge Point Parkway (512) 794-8411 Austin, TX 78730-5039
Page 38
Documentation Comment Form
National Instruments encourages you to comment on the documentation supplied with our products. This information helps us provide quality products to meet your needs.
Title: SCXI-1162HV User Manual Edition Date: August 1994 Part Number: 320790A-01 Please comment on the completeness, clarity, and organization of the manual.
If you find errors in the manual, please record the page numbers and describe the errors.
Thank you for your help. Name Title Company Address
Phone ( ) Mail to: Technical Publications Fax to: Technical Publications
National Instruments Corporation National Instruments Corporation 6504 Bridge Point Parkway, MS 53-02 MS 53-02 Austin, TX 78730-5039 (512) 794-5678
Page 39
Glossary
___________________________________________________
Prefix Meaning Value
p - pico­n- nano­µ- micro-
m- milli-
k - kilo-
M- mega-
+5 V +5 VDC Source signal ˚ degrees ohms A amperes ADC A/D converter Arms amperes, root mean square AWG American Wire Gauge BCD binary-coded decimal C Celsius CHS Chassis bit CMRR common-mode rejection ratio dB decibels DIN Deutsche Industrie Norme DMM digital multimeter F farads FIFO first-in-first-out ft feet GND ground signal hex hexadecimal H z hertz I
I
I
in
I/O input/output I
out
in. inches IN Input bit INTR* Interrupt signal ksamples 1,000 samples LSB least significant bit m meters MB megabytes of memory MSB most significant bit
input current leakage input current
output current
10
10 10 10
10 10
-12
-9
-6
-3
3 6
© National Instruments Corporation Glossary-1 SCXI-1162HV User Manual
Page 40
Glossary
Out <23..0> Parallel outputs driven to the logic levels seen at inputs 23 through 0. Out <31..24> Parallel outputs driven to the logic levels seen at inputs 31 through 24
when jumper W1 is set to 32. ppm parts per million RAM random-access memory rms root mean square s seconds SPDT single-pole double throw SPI serial peripheral interface SS* Slot Select signal UL Underwriters Laboratory V volts VAC volts alternating current Vcc positive supply voltage VDC volts direct current V
V V V V V
IH IL
in
OH OL
out
volts input high
volts input low
volts in
volts output high
volts output low
volts out Vrms volts, root mean square
W watts
SCXI-1162HV User Manual Glossary-2 © National Instruments Corporation
Page 41
Index
A
AC voltages, 3-5
C
cables, custom, 1-5 to 1-6 configuration. See also installation.
connection to MIO-type or DIO-type
DAQ boards, 2-1 factory configuration, 2-1 jumpers
parts locator diagram, 2-4 procedure for changing jumpers, 2-3 rear signal connector, 3-6 supplementary W3
configuration, 2-6
W2 and W3 jumper configuration
for DIO- and MIO-type boards (table), 2-6
W4, W5, W6, and W7 jumper
configuration for DIO- and
MIO-type boards (table), 2-5 options, 2-1 parallel mode operation, 2-2 serial mode operation, 2-2
custom cables, 1-5 to 1-6 customer communication, x, B-1
E
electrical hazards (warning), 3-1 environment specifications, A-1 equipment, optional, 1-4 to 1-5
F
front connector
optically isolated solid-state relay
channels, 3-4 pin assignments (illustration), 3-3 signal descriptions (table), 3-4
input channels, 3-4 sensing AC voltages, 3-5 sensing DC voltages, 3-4 signal connection example, 3-5
to 3-6
signal isolation, 3-5
G
GND signal (table), 3-9
H
D
DAQD*/A signal (table), 3-8 DC voltages, 3-4 DIGGND signal (table), 3-8 documentation
conventions used in manual, vii customer communication, x National Instruments documentation
set, ix organization of this manual, vii related documentation, ix
© National Instruments Corporation Index-1 SCXI-1162HV User Manual
hardware installation, 2-6 to 2-7
I
input channels, 3-4 installation. See also configuration.
hardware installation, 2-6 to 2-7 SCXI-1326 terminal block, 3-6 to 3-7 software installation, 2-7 unpacking the SCXI-1162HV, 1-6
Page 42
Index
J
jumpers
parts locator diagram, 2-4 procedure for changing jumpers, 2-3 rear signal connector, 3-7 supplementary W3 configuration, 2-6 use, 2-3 W2 and W3 jumper configuration for
DIO- and MIO-type boards (table), 2-6 W4, W5, W6, and W7 jumper
configuration for DIO- and MIO-type
boards (table), 2-5
L
LabVIEW and LabWindows software, 1-2
M
manual. See documentation. module, configuration, 2-1
N
NI-DAQ driver software, 1-2 to 1-3
P
parallel mode operation, 2-1 to 2-2 parts locator diagram, 2-4 physical specifications, A-1 pin assignments
front connector (illustration), 3-3 rear signal connector (illustration), 3-7
R
rear signal connector, 3-6 to 3-9
configuration, 3-6
parallel mode, 2-2 serial mode, 2-2
pin assignments (illustration), 3-7
signal descriptions
parallel configuration (table), 3-9 serial configuration (table), 3-8
register-level programming, 1-4
S
SCXI-1162HV
capabilities, 1-1 custom cables, 1-5 to 1-6 kit contents, 1-1 optional equipment, 1-4 to 1-5 software programming choices, 1-1
to 1-3
LabVIEW and LabWindows
software, 1-2 NI-DAQ driver software, 1-2 to 1-3 register-level programming, 1-4
unpacking, 1-6 SCXI-1326 terminal block, 3-5 SERCLK signal (table), 3-8 SERDATIN signal (table), 3-8 SERDATOUT signal (table), 3-8 serial mode operation, 2-2 signal connections
electrical hazards (warning), 3-1
front connector
pin assignments (illustration), 3-3 signal descriptions (table), 3-4
input channels, 3-4 sensing AC voltages, 3-5 sensing DC voltages, 3-4 signal connection example, 3-5
to 3-6
signal isolation, 3-5
rear signal connector, 3-6 to 3-9
configuration, 3-6 pin assignments (illustration), 3-7 signal descriptions
parallel configuration (table), 3-9 serial configuration (table), 3-8
terminal block, 3-6 SLOT0SEL* signal (table), 3-8 software programming choices, 1-1 to 1-3
LabVIEW and LabWindows software,
1-2 to 1-3 NI-DAQ driver software, 1-2 to 1-3 register-level programming, 1-4
SCXI-1162HV User Manual Index-2 © National Instruments Corporation
Page 43
specifications
environment, A-1 physical, A-1
T
technical support, B-1 terminal block, 3-6
U
unpacking the SCXI-1162HV, 1-6
Index
© National Instruments Corporation Index-3 SCXI-1162HV User Manual
Loading...