6504 Bridge Point Parkway
Austin, TX 78730-5039
(512) 794-0100
Technical support fax: (800) 328-2203
(512) 794-5678
Branch Offices:
Australia (03) 879 9422, Austria (0662) 435986, Belgium 02/757.00.20, Canada (Ontario) (519) 622-9310,
Canada (Québec) (514) 694-8521, Denmark 45 76 26 00, Finland (90) 527 2321, France (1) 48 14 24 24,
Germany 089/741 31 30, Italy 02/48301892, Japan (03) 3788-1921, Netherlands 03480-33466, Norway 32-848400,
Spain (91) 640 0085, Sweden 08-730 49 70, Switzerland 056/20 51 51, U.K. 0635 523545
Limited Warranty
The SCXI-1120 is warranted against defects in materials and workmanship for a period of one year from the date of
shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace
equipment that proves to be defective during the warranty period. This warranty includes parts and labor.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the
outside of the package before any equipment will be accepted for warranty work. National Instruments will pay the
shipping costs of returning to the owner parts which are covered by warranty.
National Instruments believes that the information in this manual is accurate. The document has been carefully
reviewed for technical accuracy. In the event that technical or typographical errors exist, National Instruments
reserves the right to make changes to subsequent editions of this document without prior notice to holders of this
edition. The reader should consult National Instruments if errors are suspected. In no event shall National
Instruments be liable for any damages arising out of or related to this document or the information contained in it.
EXCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTSMAKES NO WARRANTIES, EXPRESS OR IMPLIED,
AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE
OF
NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR DAMAGES RESULTING FROM LOSS OF DATA, PROFITS,
USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY
THEREOF
whether in contract or tort, including negligence. Any action against National Instruments must be brought within
one year after the cause of action accrues. National Instruments shall not be liable for any delay in performance due
to causes beyond its reasonable control. The warranty provided herein does not cover damages, defects,
malfunctions, or service failures caused by owner's failure to follow the National Instruments installation, operation,
or maintenance instructions; owner's modification of the product; owner's abuse, misuse, or negligent acts; and
power failure or surges, fire, flood, accident, actions of third parties, or other events outside reasonable control.
. CUSTOMER'S RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART
NATIONAL INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE CUSTOMER.
. This limitation of the liability of National Instruments will apply regardless of the form of action,
Copyright
Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or
mechanical, including photocopying, recording, storing in an information retrieval system, or translating, in whole or
in part, without the prior written consent of National Instruments Corporation.
Trademarks
LabVIEW®, NI-DAQ®, and RTSI® are trademarks of National Instruments Corporation.
Product and company names listed are trademarks or trade names of their respective companies.
Warning Regarding Medical and Clinical Use
of National Instruments Products
National Instruments products are not designed with components and testing intended to ensure a level of reliability
suitable for use in treatment and diagnosis of humans. Applications of National Instruments products involving
medical or clinical treatment can create a potential for accidental injury caused by product failure, or by errors on the
part of the user or application designer. Any use or application of National Instruments products for or involving
medical or clinical treatment must be performed by properly trained and qualified medical personnel, and all
traditional medical safeguards, equipment, and procedures that are appropriate in the particular situation to prevent
serious injury or death should always continue to be used when National Instruments products are being used.
National Instruments products are NOT intended to be a substitute for any form of established process, procedure, or
equipment used to monitor or safeguard human health and safety in medical or clinical treatment.
Contents
About This Manual............................................................................................................. xi
Organization of This Manual ......................................................................................... xi
Conventions Used in This Manual................................................................................. xii
Related Documentation.................................................................................................. xiii
Customer Communication ............................................................................................. xiii
Chapter 1
Introduction
What Your Kit Should Contain...................................................................................... 1-2
This manual describes the electrical and mechanical aspects of the SCXI-1120 and contains
information concerning its operation and programming. The SCXI-1120 is a class I module that
operates as eight isolated input channels. Refer to your chassis manual for a description of the
different module classes. Each channel is isolated and independently configurable via jumpers.
Refer to your chassis manual for a description of the different module classes. The SCXI-1120 is
a member of the National Instruments Signal Conditioning eXtensions for Instrumentation
(SCXI) Series for the National Instruments data acquisition plug-in boards. This board is
designed for low-cost signal conditioning of thermocouples, volt sources, millivolt sources, and
4 to 20 mA sources or 0 to 20 mA process-current sources where high common-mode voltages
exist.
This manual describes the installation, theory of operation, and basic programming
considerations for the SCXI-1120.
Organization of This Manual
The SCXI-1120 User Manual is organized as follows:
•Chapter 1, Introduction, describes the SCXI-1120; lists the contents of your SCXI-1120 kit;
describes the optional software, optional equipment, and custom cables; and explains how to
unpack the SCXI-1120 kit.
•Chapter 2, Configuration and Installation, describes the SCXI-1120 jumper configurations,
installation of the SCXI-1120 into the SCXI chassis, signal connections to the SCXI-1120,
and cable wiring.
•Chapter 3, Theory of Operation, contains a functional overview of the SCXI-1120 module
and explains the operation of each functional unit making up the SCXI-1120.
•Chapter 4, Register Descriptions, describes in detail the SCXI-1120 Module ID Register, the
Configuration Register, the Slot 0 registers, and multiplexer addressing.
•Chapter 5, Programming, contains a functional programming description of the SCXI-1120
and Slot 0.
•Appendix A, Specifications, lists the specifications for the SCXI-1120.
•Appendix B, Rear Signal Connector, describes the pinout and signal names for the
SCXI-1120 50-pin rear signal connector, including a description of each connection.
•Appendix C, SCXIbus Connector, describes the pinout and signal names for the SCXI-1120
96-pin SCXIbus connector, including a description of each connection.
•Appendix D, SCXI-1120 Front Connector, describes the pinout and signal names for the
SCXI-1120 front connector, including a description of each connection.
•Appendix E, SCXI-1120 Cabling, describes how to use and install the hardware accessories
for the SCXI-1120.
•Appendix F, Revision A and B Photo and Parts Locator Diagrams, contains a photograph of
the Revision A and B SCXI-1120 signal conditioning module and the general and detailed
parts locator diagrams.
•Appendix G, Customer Communication, contains forms you can use to request help from
National Instruments or to comment on our products and manuals.
•The Glossary contains an alphabetical list and description of terms used in this manual,
including abbreviations, acronyms, metric prefixes, mnemonics, symbols, and terms.
•The Index contains an alphabetical list of key terms and topics in this manual, including the
page where you can find each one.
Conventions Used in This Manual
The following conventions are used in this manual.
italicItalic text denotes emphasis, a cross reference, or an introduction to a key
concept.
Lab boardLab board refers to the Lab-LC, Lab-NB, Lab-PC, and Lab-PC+ boards,
unless otherwise noted.
MCMC refers to the Micro Channel series computers.
MIO boardMIO board refers to the AT-MIO-16, AT-MIO-16D, AT-MIO-16F-5,
AT-MIO-16X, AT-MIO-64F-5, MC-MIO-16, NB-MIO-16, and
NB-MIO-16X series of multichannel I/O data acquisition boards.
monospaceLowercase text in this font denotes text or characters that are to be literally
input from the keyboard, sections of code, programming examples, and
syntax examples. This font is also used for the proper names of disk
names, functions, variables, filenames, and extensions, and for statements
and comments taken from program code.
NBNB refers to the NuBus series computers.
PCPC refers to the IBM PC/XT, the IBM PC AT, and compatible computers.
SCXIbusSCXIbus refers to the backplane in the chassis. A signal on the backplane
is referred to as the SCXIbus <signal name> line (or signal). The
SCXIbus descriptor may be omitted when the meaning is clear.
Descriptions of all SCXIbus signals are given in Appendix C, SCXIbus
Connector.
Slot 0Slot 0 refers to the power supply and control circuitry in the SCXI chassis.
Abbreviations, acronyms, metric prefixes, mnemonics, symbols, and terms are listed in the
Glossary.
Related Documentation
The following documents contain information that you may find helpful as you read this manual:
•Your DAQ board user manual
•Your SCXI chassis user manual
Customer Communication
National Instruments wants to receive your comments on our products and manuals. We are
interested in the applications you develop with our products, and we want to help if you have
problems with them. To make it easy for you to contact us, this manual contains comment and
configuration forms for you to complete. These forms are in Appendix G, Customer
This chapter describes the SCXI-1120; lists the contents of your SCXI-1120 kit; describes the
optional software, optional equipment, and custom cables; and explains how to unpack the
SCXI-1120 kit.
The SCXI-1120 is a class I module consisting of eight isolated input channels. The SCXI-1120
is a module for signal conditioning of thermocouples, volt sources, millivolt sources, 4 to 20 mA
current sources, and 0 to 20 mA process-current sources. Refer to your chassis manual for a
description of the different module classes. If external excitation is provided, thermistors, RTDs,
and strain gauges can also be measured. The SCXI-1120 can operate in two output modes–in the
Parallel-Output mode with all eight input channels connected in parallel to eight data acquisition
board channels, or in the Multiplexed-Output mode with all eight channels multiplexed into a
single data acquisition board channel.
The SCXI-1120 operates with full functionality with the National Instruments MIO-16 boards.
You can use the Lab-NB, the Lab-PC, the Lab-PC+, the Lab-LC, and the PC-LPM-16 boards
with the SCXI-1120, but these boards cannot scan the module when it is configured in the
Multiplexed-Output mode. These boards can perform only single-channel reads in this mode.
You can also use the SCXI-1120 with other systems that comply with the specifications given in
Chapter 2, Configuration and Installation. You can multiplex several SCXI-1120s into a single
channel, thus greatly increasing the number of analog input signals that can be digitized.
The addition of a shielded terminal block provides screw terminals for easy signal attachment to
the SCXI-1120. In addition, a temperature sensor for cold-junction compensation of
thermocouples is included on the terminal block. This cold-junction reference (CJR) is either
multiplexed along with the eight channels or connected by jumpers to a different channel of the
data acquisition board.
With the SCXI-1120, the SCXI chassis can serve as a fast-scanning signal conditioner for
laboratory testing, production testing, and industrial process monitoring.
The contents of the SCXI-1120 kit (part number 776572-20) are listed as follows.
Kit ComponentPart Number
SCXI-1120 module181695-01
SCXI-1120 User Manual320425-01
If your kit is missing any of the components, contact National Instruments.
Optional Software
This manual contains complete instructions for directly programming the SCXI-1120. You can
order separate software packages for controlling the SCXI-1120 from National Instruments.
When you combine the PC, AT, and MC data acquisition boards with the SCXI-1120, you can
use LabVIEW for Windows or LabWindows for DOS. LabVIEW and LabWindows are
innovative program development software packages for data acquisition and control applications.
LabVIEW uses graphical programming, whereas LabWindows enhances Microsoft C and
QuickBASIC. Both packages include extensive libraries for data acquisition, instrument control,
data analysis, and graphical data presentation.
Your National Instruments data acquisition board is shipped with the NI-DAQ software.
NI-DAQ has a library of functions that can be called from your application programming
environment. These functions include routines for analog input (A/D conversion), buffered data
acquisition (high speed A/D conversion), analog output (D/A conversion), waveform generation,
digital I/O, counter/timer, SCXI, RTSI, and self-calibration. NI-DAQ maintains a consistent
software interface among its different versions so you can switch between platforms with
minimal modifications to your code.
You can also use the SCXI-1120, together with the PC, AT, and MC data acquisition boards,
with NI-DAQ software for DOS/Windows/LabWindows. NI-DAQ software for
DOS/Windows/LabWindows comes with language interfaces for Professional BASIC, Turbo
Pascal, Turbo C, Turbo C++, Borland C++, and Microsoft C for DOS; and Visual Basic, Turbo
Pascal, Microsoft C with SDK, and Borland C++ for Windows. NI-DAQ software for
DOS/Windows/LabWindows is on high-density 5.25 in. and 3.5 in. diskettes.
You can use the SCXI-1120, together with the NB Series data acquisition boards, with
LabVIEW, a software system that features interactive graphics, a state-of-the-art user interface,
and a powerful graphical programming language. The LabVIEW Data Acquisition VI Library, a
series of VIs for using LabVIEW with National Instruments boards, is included with LabVIEW.
The LabVIEW Data Acquisition VI Library is functionally equivalent to the NI-DAQ software
for Macintosh.
You can also use the SCXI-1120, combined with the NB Series data acquisition boards, with
NI-DAQ software for Macintosh. NI-DAQ software for Macintosh, which is shipped with all
National Instruments Macintosh data acquisition boards, comes with language interfaces for
MPW C, THINK C, Pascal, and Microsoft QuickBASIC. Any language that uses Device
Manager Toolbox calls can access NI-DAQ software for Macintosh.
with 1 m cable
with 2 m cable
with 5 m cable
with 10 m cable
SCXI-1349 SCXI shielded cable assembly
with 1 m cable
with 2 m cable
with 5 m cable
with 10 m cable
SCXI-1350 multichassis adapter776575-50
SCXI process-current resistor kit776582-01
Standard ribbon cable0.5 m
1.0 m
NB6 cable
0.5 m
1.0 m
776574-471
776574-472
776574-475
776574-470
776574-491
776574-492
776574-495
776574-490
180524-05
180524-10
181305-01
181305-10
Refer to the Signal Connections section in Chapter 2, Configuration and Installation, and to
Appendix E, SCXI-1120 Cabling, for additional information on cabling, connectors, and
adapters.
The SCXI-1120 rear signal connector is a 50-pin male ribbon-cable header. The manufacturer
part number used by National Instruments for this header is as follows:
•AMP Inc. (part number 1-103310-0)
The mating connector for the SCXI-1120 rear signal connector is a 50-position polarized
ribbon-socket connector with strain relief. National Instruments uses a polarized or keyed
connector to prevent inadvertent upside-down connection to the SCXI-1120. Recommended
manufacturer part numbers for this mating connector are as follows:
•Electronic Products Division/3M (part number 3425-7650)
•T&B/Ansley Corporation (part number 609-5041CE)
Standard 50-conductor, 28 AWG, stranded ribbon cables that can be used with these connectors
are as follows:
•Electronic Products Division/3M (part number 3365/50)
•T&B/Ansley Corporation (part number 171-50)
The SCXI-1120 front connector is a 32-pin DIN C male connector with columns A and C even
pins only. The manufacturer part number used by National Instruments for this connector is as
follows:
•Panduit Corporation (part number 100-932-023)
The mating connector for the SCXI-1120 front connector is a 32-pin DIN C female connector.
National Instruments uses a polarized connector to prevent inadvertent upside-down connection
to the SCXI-1120. Recommended manufacturer part numbers for this mating connector are as
follows:
•Panduit Corporation (part number 100-932-434; straight-solder eyelet pins)
•Panduit Corporation (part number 100-932-633; right-angle pins)
These connectors were selected to meet UL 1950 and UL 1244 for 1,500 Vrms isolation.
Unpacking
Your SCXI-1120 module is shipped in an antistatic package to prevent electrostatic damage to
the module. Several components on the module can be damaged by electrostatic discharge. To
avoid such damage in handling the module, take the following precautions:
•Touch the antistatic package to a metal part of your SCXI chassis before removing the
module from the package.
•Remove the module from the package and inspect the module for loose components or any
other sign of damage. Notify National Instruments if the module appears damaged in any
way. Do not install a damaged module into your SCXI chassis.
This chapter describes the SCXI-1120 jumper configurations, installation of the SCXI-1120 into
the SCXI chassis, signal connections to the SCXI-1120, and cable wiring.
Module Configuration
The SCXI-1120 includes 46 jumpers that are shown in the parts locator diagrams in Figures 2-1
and 2-2.
-On Revision A and B modules, jumper W42 is unused and should not be connected.
-Jumper W45 is reserved and should not be reconfigured.
-On Revision A and B modules, jumper W44 carries the SLOT0SEL* signal from the rear
signal connector, after buffering, to the SCXIbus INTR* line and should be left in the
factory-default position (position 1). On Revision C and later modules, jumper W44 does
not exist.
•User-configurable jumpers
-Jumper W43 carries the SCXIbus MISO line, after buffering, to the SERDATOUT signal
on the rear signal connector.
-On Revision C and later modules, jumper 42 connects a pullup resistor to the
SERDATOUT signal on the rear signal connector.
-Jumper W46 configures the guard, the analog output ground, and enables the
Pseudodifferential Reference mode.
-Jumpers W1 through W8 configure the first-stage gain of channels 0 through 7,
respectively.
-Jumpers W9 through W16 configure the second-stage gain of channels 0 through 7,
respectively.
-Jumpers W17 through W24 configure the first-stage filtering of input channels 0 through
7, respectively.
-Jumpers W25 through W40 configure the second-stage filtering of input channels 0
through 7, respectively.
-Jumper W41 directly connects the temperature sensor to the rear signal connector. To do
so, set jumper W46 in the AB-R2 or AB-R0 position first.
Further configuration of the board is software controlled and will be discussed later in this
chapter.
Digital Signal Connections
The four digital signal connection jumpers have position 1 marked on the board. Position 3 is
not explicitly marked on the board.
The SCXI-1120 has three jumpers dedicated for communication between the data acquisition
board and the SCXIbus. These jumpers are W42, W43, and W44.
On Revision A and B modules, position 1 connects, after buffering, SLOT0SEL* to the SCXIbus
INTR* line. This is the factory-default setting and should not be changed. In this setting, the
data acquisition board controls the SCXIbus INTR* line. See the Timing Requirements andCommunication Protocol section later in this chapter, and Chapter 5, Programming, for
information on the use of the INTR* line. See Appendix E, SCXI-1120 Cabling, for the pin
equivalences of the SCXI-1120 rear signal connector and the data acquisition board I/O
connector.
Position 3 is reserved and should not be used.
On Revision C and later modules, jumper W44 is not loaded. SLOT0SEL* is always buffered to
the INTR* line.
Jumper W43
Position 1 connects, after buffering, the SCXIbus MISO line to the SERDATOUT pin of the rear
signal connector. In this setting, along with the proper setting of jumper W42, the data
acquisition board can read the Module ID Register of the SCXI-1120. This is the factory-default
setting. See the Timing Requirements and Communication Protocol section later in this chapter,
and Chapter 4, Register Descriptions, for information on reading the Module ID Register. See
Appendix E, SCXI-1120 Cabling, for the pin equivalences of the SCXI-1120 rear signal
connector and the data acquisition board I/O connector.
Position 3 disconnects SERDATOUT from the SCXIbus MISO line.
Jumper W42
On Revision A and B modules, jumper W42 should not be connected. On Revision C and later
modules, position 1 connects a 2.2 kΩ pullup resistor to the SERDATOUT line. Position 3 does
not connect the pullup resistor to the SERDATOUT line.
Using Jumpers W42 and W43
If the SCXI-1120 is not cabled to a data acquisition board, the positions of these jumpers do not
matter, so leave them in their factory-default positions (both in position 1).
If the SCXI-1120 is cabled to a data acquisition board, and the SCXI chassis that the SCXI-1120
is in, is the only SCXI chassis cabled to that data acquisition board, leave the jumpers in their
factory-default positions (both in position 1).
If the SCXI-1120 is cabled to a data acquisition board, and there are multiple SCXI chassis
cabled to that data acquisition board with shielded cables (you are using SCXI-1346 shielded
cable multichassis adapters), leave the jumpers in their factory-default positions (both in
position 1).
If the SCXI-1120 is cabled to a data acquisition board, and there are multiple SCXI chassis
cabled to that data acquisition board with ribbon cables (you are using SCXI-1350 multichassis
adapters), leave jumper W43 in its factory-default position (position 1). On all but one of the
SCXI-1120s that are cabled to the data acquisition board, move jumper W42 to position 3. It
does not matter which of the SCXI-1120 modules that are cabled to the data acquisition board
has jumper W42 set to position 1. If you have different types of modules cabled to the data
acquisition board, those different modules will have jumpers similar to W43 and W42 of the
SCXI-1120. Set those jumpers on the different modules using the same method described here
for the SCXI-1120.
On Revision A and B SCXI-1120s, jumper W42 is not used. You set jumper W43 as explained
in the cases above, except in the case of a multichassis ribbon cable system. In a multichassis
ribbon cable system with Revision A and B SCXI-1120s cabled to the data acquisition board,
you can access the MISO line in only one chassis. Pick one of the chassis that has the
SCXI-1120 cabled to the DAQ board. Set jumper W43 on the SCXI-1120 to position 1. On the
SCXI-1120s that are in the other chassis and cabled to the data acquisition board, set jumper
W43 to position 3. Notice that you will only be able to access digital information from the
chassis that has the SCXI-1120 with jumper W43 set to position 1.
On Revision C and later modules, the SERDATOUT line is driven with an open-collector driver
(a driver that actively drives low or goes to a high-impedance state, relying on a pullup resistor to
make the signal line go high). When using a single chassis, leave W42 and W43 in position 1,
the factory default, on the SCXI-1120 that is connected to the data acquisition board. In this
setting, the module drives MISO to SERDATOUT and connects the necessary pullup resistor to
the SERDATOUT line. When using multiple chassis, leave jumper W43 in position 1 on all of
the SCXI-1120s that are cabled to the data acquisition board. Only one of the SCXI-1120s that
are cabled to the data acquisition board should have jumper W42 in position 1. It does not matter
which of the SCXI-1120s that are cabled to the data acquisition board has the pullup connected.
All of the other SCXI-1120 modules that are cabled to the data acquisition board should have
jumper W42 in position 3. The reason for this is that if too many pullup resistors are attached to
the SERDATOUT line, the drivers cannot drive the line low. See Table 2-1 for the description
and configuration of the jumper settings.
Table 2-1. Digital Signal Connections, Jumper Settings
JumperDescriptionConfiguration
W42
W42
W43
W43
Connects pullup to SERDATOUT
(Revision C and later) factorydefault setting.
Parking position (not connected on
Revision A and B)
Connects MISO to SERDATOUT
Factory-default setting.
Table 2-1. Digital Signal Connections, Jumper Settings (Continued)
JumperDescriptionConfiguration
W45
W44
Factory default
Factory default
(Revision A and B modules only)
3
•
2
•
1
•
3
•
2
•
1
•
Analog Configuration
The SCXI-1120 has 42 analog configuration jumpers.
Before starting, notice that the jumper configurations for each channel are similar; only the
jumper reference designator number changes. When you learn how to configure one channel,
you can configure the other channels as well.
Grounding, Shielding, and Reference-Mode Selection
Jumper W46
Position B-R0R1 is the parked position and the factory-default setting.
Position AB-R0 connects the analog reference to the analog output ground (pins 1 and 2 on the
rear signal connector). Select this configuration when using an RSE data acquisition board. You
should not use differential input data acquisition boards when jumper W46 is in the AB-R0
position.
Position AB-R1 connects the analog reference to the SCXIbus guard.
Position AB-R2 enables the Pseudodifferential Reference mode and connects the analog
reference to the OUTREF pin on the rear signal connector. Select this mode when the
SCXI-1120 has to operate with data acquisition boards that have a nonreferenced single-ended
input (NRSE). Do not use differential-input data acquisition boards when jumper W46 is in the
AB-R2 position.
Note: The SCXI-1120 will drive pins 4, 6, 8, 10, 12, 14, 16, and 18 on the rear signal connector,
although the SCXI-1120 is in pseudodifferential mode.
AOGND (pins 1 and 2 of the rear
signal connector). Use this connection
with RSE data acquisition boards.
W46Connects SCXIbus guard to the analog
reference
W46Enables the Pseudodifferential
Reference mode (pin 19 of the rear
signal connector is connected to the
analog reference)
•••••
R
2 R1 R0
•••••
R
2 R1 R0
•••••
R
2 R1 R0
•••••
R
2 R1 R0
B
•
A
B
•
A
B
•
A
B
•
A
Direct Temperature Connection
Jumper W41
When your SCXI-1120 is operating with a data acquisition board configured in a differential
mode, such as the MIO board, you can access the temperature sensor only in the MultiplexedOutput mode. Direct temperature connection is not permissible to the data acquisition board.
This is the factory setting and is position 1 of the jumper block. Position 3 is not explicitly
marked.
If the temperature sensor needs to be accessed in parallel with the other eight outputs, or
scanned–without software interference–along with the other channels, you need to place W41 in
position 3 (this position is not explicitly marked on the module) and to configure your reference
to pseudodifferential or single-ended operation (jumper W46) with your data acquisition board
configured for the same reference scheme. In this position, the direct temperature sensor
connection is made on pin 18 of the rear signal connector and corresponds to ACH15 of the MIO
boards or the PC-LPM-16.
Each input channel has two gain stages. The first gain stage provides gains of 1, 10, 50, and 100,
and the second stage provides gains of 1, 2, 5, 10, and 20. Tables 2-4 and 2-5 show how to set
up the gain for each channel.
Table 2-4. Gain Jumper Allocation
Input Channel NumberFirst Gain JumperSecond Gain Jumper
The board is shipped with the first-stage gain set to 100 (position A), and a second-stage gain set
to 10 (position D). To change the gain of your module, move the appropriate jumper on your
module to the position indicated in Tables 2-3 and 2-4. Refer to Figure 2-1, SCXI-1120 GeneralParts Locator Diagram, and Figure 2-2, Detailed Parts Locator Diagram, for jumper locations
on your module.
To determine the overall gain of a given channel use the following formula:
Overall gain = First-stage gain x second-stage gain.
Two-stage filtering is also available on your SCXI-1120 module. The first stage is located in the
isolated section of the input channel, whereas the second stage is located in the nonisolated
section of your input channel. Two-stage filtering eliminates the noise generated by the isolation
amplifier, producing a higher signal-to-noise ratio. Furthermore, two filter bandwidths are
available, 10 kHz and 4 Hz.
Your SCXI-1120 is shipped in the 4 Hz position. Remember to make sure that both stages are
set to the same bandwidth to ensure that the required bandwidth is achieved. Notice that one
jumper block is available for each filter stage.
You can install the SCXI-1120 in any available SCXI chassis. After you have made any
necessary changes and have verified and recorded the jumper settings on the form in
Appendix G, Customer Communication, you are ready to install the SCXI-1120. The following
are general installation instructions, but consult the user manual or technical reference manual of
your SCXI chassis for specific instructions and warnings.
1. Turn off the computer that contains the data acquisition board or disconnect it from your
SCXI chassis.
2. Turn off the SCXI chassis. Do not insert the SCXI-1120 into a chassis that is turned on.
3. Insert the SCXI-1120 into the module guides. Gently guide the module into the back of the
slot until the connectors make good contact. If a cable assembly has already been installed in
the rear of the chassis, the module and cable assembly must be firmly engaged; however, do
not force the module into place.
4. Screw the front mounting panel of the SCXI-1120 to the top and bottom threaded strips of
your SCXI chassis.
5. If this module is to be connected to an MIO-16 data acquisition board, attach the connector at
the metal end of the SCXI-1340 cable assembly to the rear signal connector on the
SCXI-1120 module. Screw the rear panel to the rear threaded strip. Attach the loose end of
the cable to the MIO-16 board.
Note: For installation procedures with other SCXI accessories and data acquisition boards,
consult Appendix E, SCXI-1120 Cabling.
6. Check the installation.
7. Turn on the SCXI chassis.
8. Turn on the computer or reconnect it to your chassis.
The SCXI-1120 board is installed and ready for operation.
Signal Connections
This section describes the input and output signal connections to the SCXI-1120 board via the
SCXI-1120 front connector and rear signal connector, and includes specifications and connection
instructions for the signals given on the SCXI-1120 connectors.
Warning:Connections that exceed any of the maximum ratings of input or output signals on
the SCXI-1120 can result in damage to the SCXI-1120 board and to the SCXIbus.
Maximum input ratings for each signal are given in this chapter under the
discussion of that signal. National Instruments is not liable for any damages
resulting from such signal connections.
A2 CHSGNDChassis Ground – Tied to the SCXI chassis.
C2DTEMPDirect Temperature Sensor – Connects the
temperature sensor to pin 18 of the rear signal
connector MCH7- when the terminal block is
configured for direct temperature connection and
jumper W41 is in position 3.
A4+5 V+5 VDC Source – Used to power the temperature
sensor on the terminal block. 0.2 mA of source not
protected.
C4MTEMPMultiplexed Temperature Sensor – Connects the
temperature sensor to the output multiplexer.
A6, C6, C8RSVDReserved – Reserved for future use. Do not connect
any signals to these pins. TTL/CMOS output. They
are not protected.
A8, A10, C10,No ConnectDo not connect any signals to these pins.
A16, C16,A22,
C22, A28, C28
A12, A14, A18, CH7+ through CH0+Positive Input Channels – The positive inputs to
A26, A20, A24, channels 7 through 0, respectively.
A30, A32
C12, C14, C18, CH7- through CH0-Negative Input Channels – The negative inputs to
C20, C24, C26, channels 7 through 0, respectively.
C30, C32
The signals on the front connector are all analog with the exceptions of pins A6, C6, and C8.
The analog signals can be divided into two groups–the analog input channels, and the
temperature sensor.
Analog Input Channels
The positive input channels are located in column A. Their corresponding negative input
channels are located in column C. Each input corresponds to a separate amplifier and is fully
isolated from the other channels and from earth ground. The inputs are designed in a floating
single-ended configuration, thus the measured signal can be referenced to a ground level with
common-mode voltage up to 250 Vrms. For better noise immunity, connect the negative input
channel to the signal reference. If the measured signals are floating, connect the negative input
channel to chassis ground on the terminal block. Figure 2-4 shows how to connect a groundreferenced signal. Figure 2-5 shows how to connect a floating signal. Figures 2-6 and 2-7 show
how to connect AC-coupled signals.
Figure 2-7. AC-Coupled Signal Connection with High Common-Mode Voltage
For AC-coupled signals, an external resistor from the positive input channel to the signal
reference should be connected. This is needed to provide the DC path for the positive input bias
current. Typical resistor values range from 100 kΩ to 1 MΩ. This solution, although necessary
in this case, lowers the input impedance of the input channel amplifier and introduces an
additional offset voltage proportional to the input bias current and to the resistor value used. The
typical input bias current of the amplifier consists of ±80 pA and a negligible offset drift current.
When a 100 kΩ resistor is used, this will result into ±8 µV of offset, which is insignificant in
most applications. However, if larger valued bias resistors are used, significant input offset may
result. To determine the maximum offset introduced by the biasing resistor, use the following
equation:
V
The input signal range of an SCXI-1120 input channel is ±5 V/ G
input, where G
ofsbias
= I
x R
bias
total
bias
referenced to its negative
total
is equal to the product of the first-stage and second-stage gains. In addition,
the input channels are overvoltage protected to 240 Vrms with power on or off at a maximum of
4.5 mArms sink or source.
Warning: Exceeding the input signal range and the common-mode input range results in
distorted signals. Exceeding the maximum input voltage rating (250 Vrms between
terminals, and between any terminal and ground) can result in damage to the
SCXI-1120, the SCXIbus, and the data acquisition board. National Instruments is
not liable for any damages resulting from such signal connections.
Temperature Sensor Connection
Pins C2 and C4 are dedicated for connecting the temperature sensor to the SCXI-1120. The
temperature sensor is not isolated and is referenced to the chassis ground. The connection is
overvoltage protected to ±25 VDC with power on and ±15 VDC with power off.
Warning: Exceeding the overvoltage protection on the temperature connections can result in
damage to the SCXI-1120, the SCXIbus, and the data acquisition board. National
Instruments is not liable for any damages resulting from such signal connections.
Two types of signal connectors are available to connect the signals to the SCXI-1120 inputs.
The first, the SCXI-1330 32-pin DIN C female connector-and-shell assembly, is available in a kit
listed in the Optional Equipment section in Chapter 1, Introduction. The connector has eyelet
ends for easy hook-and-solder wire connection. With this kit, you can build your own signal
cable to connect to the SCXI-1120 inputs. After you have built the cable, the shell covers and
protects the connector. Perform the following steps to assemble and mount the connector-andshell assembly to your SCXI module:
1.Refer to Figure 2-8, Assembling and Mounting the SCXI-1330 Connector-and-ShellAssembly, and the diagram included with your SCXI-1330 kit to build the connector-andshell assembly.
2.Turn off the computer that contains your data acquisition board or disconnect the board
from your SCXI chassis.
3.Turn off your SCXI chassis.
4.Slide the selected module out of the SCXI chassis.
5.Remove the module cover.
6.Place one jack screw as indicated in Figure 2-8.
7.While holding the jack screw in place, insert the lock washer and then the nut. Notice that
you might need long-nose pliers to insert the washer and nut.
8.Tighten the nut by holding it firmly and rotating the jack screw.
9.Repeat steps 6 through 8 for the second jack screw.
10.Replace the module cover and tighten the grounding screw.
11.Slide the module back in place.
12.Connect the SCXI-1330 to your module connector and secure it by tightening both
mounting screws.
Figure 2-8. Assembling and Mounting the SCXI-1330 Connector-and-Shell Assembly
SCXI-1320 and SCXI-1328 Terminal Blocks
Grounding Screw
The second type of connector available to connect the signals to the SCXI-1120 inputs is a
terminal block with an onboard temperature sensor and screw terminals for easy connection.
One terminal block, the SCXI-1328 isothermal terminal block, has a high-accuracy onboard
temperature sensor. The terminal block kits are listed in the Optional Equipment section in
Chapter 1, Introduction.
The SCXI-1320 terminal block and SCXI-1328 high-accuracy isothermal terminal block consist
of a shielded board with supports for connection to the SCXI-1120 input connector. The
terminal blocks have 18 screw terminals for easy connection. Eight pairs of screw terminals are
for signal connection to the eight inputs of the SCXI-1120, and one pair of screw terminals
connects to the chassis ground.
The following warnings contain important safety information concerning hazardous voltages and
terminal blocks.
Warnings:When using the terminal block with high common-mode voltages, you must
insulate your signal wires appropriately. National Instruments is not liable for any
damages or injuries resulting from inadequate signal wire insulation.
If high voltages (≥42 Vrms) are present, you must connect the safety earth groundto the strain-relief tab. This complies with UL 1244 and protects against electric
shock when the terminal block is not connected to the chassis. To connect the
safety earth ground to the strain-relief tab, run an earth ground wire in the cable
from the signal source to the terminal block. National Instruments is not liable for
any damages or injuries resulting from inadequate safety earth ground
connections.
When connecting your signals to the SCXI-1320 terminal block for use with the SCXI-1120,
follow the labeling on the SCXI-1320 indicated under the module type column for the
SCXI-1120 as indicated in Figure 2-9.
When connecting your signals to the SCXI-1328 high-accuracy isothermal terminal block for use
with the SCXI-1120, follow the labeling on the SCXI-1328 indicated along the module type row
for the SCXI-1120 as indicated in Figure 2-10.
Terminal Block Temperature Sensor
To accommodate thermocouples with the SCXI-1120, the terminal block has a temperature
sensor for cold-junction compensation. You can connect the temperature sensor in two ways:
•You can connect the temperature sensor to the MTEMP pin (C4) on the module front
connector and multiplex the sensor at the output multiplexer along with amplifier outputs.
This is the Multiplexed Temperature Sensor (MTS) mode. Refer to the ConfigurationRegister section in Chapter 4, Register Descriptions, for further details.
•You can connect the temperature sensor to a separate data acquisition channel via pin 18 on
the module rear signal connector when you set SCXI-1120 jumper W41 to position 3. This is
the Direct Temperature Sensor (DTS) mode.
Note:Use an average of a large number of samples to obtain the most accurate reading.
Noisy environments require more samples for greater accuracy.
The SCXI-1320 temperature sensor outputs 10 mV/°C and has an accuracy of ±1° C over the
0° to 55° C temperature range. To determine the temperature, use the following formulas:
T (°C) = 100(V
T (° C)
T (°F) =
[]
TEMPOUT
9
)
+32
5
where V
TEMPOUT
is the temperature sensor output and T (°F) and T (°C) are the temperature
readings in degrees Fahrenheit and degrees Celsius, respectively.
The SCXI-1328 temperature sensor outputs 0.62 to 0.07 V from 0° to 55° C and has an accuracy
of ±0.35° C over the 15° to 35° C range and ±0.65° C over the 0° to 15° and 35° to 55° C ranges.
To determine the temperature, use the following formulas:
T (°C) = T = T
where T
K
TK=
a = 1.288 x 10
b = 2.356 x 10
c = 9.556 x 10
is the temperature in kelvin
a + b lnR
[]
- 273.15
K
1
-3
-4
-8
+c lnR
()
T
T
()
3
RT = resistance of the thermistor in Ω
R
= 50,000
T
V
TEMPOUT
T (°F) =
V
TEMPOUT
2.5− V
TEMPOUT
= output voltage of the temperature sensor
T (° C)
[]
9
+32
5
where T (°F) and T (°C) are the temperature readings in degrees Fahrenheit and degrees Celsius,
respectively.
Terminal Block Jumper Configuration
In addition to the screw terminals, the terminal block has one jumper for configuring the onboard
temperature sensor. When you set jumper W1 on the terminal block to the MTEMP position, the
jumper connects the temperature sensor to the SCXI-1120 output multiplexer. This is the factory
setting. The DTEMP position connects the temperature sensor to SCXI-1120 jumper W41.
In both MTS and DTS modes, the reference to the temperature sensor signal is the SCXI-1120
analog ground that is connected to MCHO- in the MTS mode, or to OUTREF or AOGND (with
SCXI-1120 jumper W46 set in positions AB-R2 and AB-R0, respectively) in the DTS mode.
One jumper block comprises both positions; thus, you can use only one type of configuration at a
time. The parking position for the jumper block is the MTEMP position (the temperature sensor
is disabled until the RTEMP bit in the Configuration Register selects the sensor).
Tables 2-7 and 2-8 show the jumper settings on the SCXI-1320 and SCXI-1328 terminal blocks.
Table 2-7. Jumper Settings on the SCXI-1320 Terminal Block
JumperPositionDescription
W1
W1
•
•
MTEMP
•
•
DTEMP
•
•
MTEMP
•
•
DTEMP
MTS mode selected; factory
setting; parking position
Table 2-8. Jumper Settings on the SCXI-1328 Terminal Block
JumperPositionDescription
W1
W1
DTEMPMTEMP
•••
•
•••
DTEMPMTEMP
•
MTS mode selected; factory
setting; parking position
DTS mode selected
Terminal Block Signal Connection
Warnings:The chassis GND terminals on your terminal block are for grounding high
impedance sources such as a floating source (1 mA maximum). Do not use these
terminals as safety earth grounds.
If high voltages (≥42 Vrms) are present, you must connect the safety earth groundto the strain-relief tab. This complies with UL 1244 and protects against electric
shock when the terminal block is not connected to the chassis. To connect the
safety earth ground to the strain-relief tab, run an earth ground wire in the cable
from the signal source to the terminal block. National Instruments is not liable for
any damages or injuries resulting from inadequate safety earth ground
connections.
Shock Hazard: This unit should only be opened by qualified personnel aware of
!
the dangers involved. Disconnect all power before removing cover. Always
install grounding screw.
To connect the signal to the terminal block, use the following procedure:
1. Remove the grounding screw of the top cover.
2. Snap out the top cover of the shield by placing a screwdriver in the groove at the bottom of
the terminal block.
3. Slide the signal wires, one at a time, through the front panel strain-relief opening. You can
add insulation or padding if necessary.
4. Connect the wires to the screw terminals by wrapping the wires around the screws and
tightening the screws without letting the wires slip out.
5. Tighten the larger strain-relief screws.
6. Snap the top cover back in place.
7. Reinsert the grounding screw to ensure proper shielding.
8. Connect the terminal block to the SCXI-1120 front connector as explained in the TerminalBlock Installation section later in this chapter.
Figure 2-9 shows a parts locator diagram for the SCXI-1320 terminal block. Figure 2-10 shows
a parts locator diagram for the SCXI-1328 terminal block.
Terminal Block Installation
To connect the terminal block to the SCXI-1120 front connector, perform the following steps:
1. Connect the SCXI-1120 front connector to its mating connector on the terminal block.
2. Make sure that the SCXI-1120 top and bottom thumbscrews do not obstruct the rear panel of
the terminal block.
3. Tighten the top and bottom screws on the back of the terminal block to hold it securely in
place.
Rear Signal Connector
Note: If you will be using the SCXI-1120 with a National Instruments data acquisition board
and cable assembly, you do not need to read the remainder of this chapter. If you will
also be using the SCXI-1180 feedthrough panel, the SCXI-1343 rear screw terminal
adapter, or the SCXI-1351 one-slot cable extender with the SCXI-1120, you should read
this section.
Figure 2-11 shows the pin assignments for the SCXI-1120 rear signal connector.
1-2AOGNDAnalog Output Ground – Connected to the analog
reference when jumper W46 is in position AB-R0.
3-18MCH0± through MCH7± Analog Output Channels 0 through 7 – Connects to
the data acquisition board differential analog input
channels.
19OUTREFOutput Reference – Serves as the reference node for
the analog output channels and the temperature
sensor–in the DTS mode–in the Pseudodifferential
Reference mode. It should be connected to the
analog input sense of the NRSE data acquisition
board.
24, 33DIG GNDDigital Ground – Supply the reference for data
acquisition board digital signals and are tied to the
module digital ground.
25SERDATINSerial Data In – Taps into the SCXIbus MOSI line
to provide serial input data to a module or Slot 0.
26SERDATOUTSerial Data Out – Taps into the SCXIbus MISO line
to accept serial output data from a module.
27DAQD*/AData Acquisition Board Data/Address Line – Taps
into the SCXIbus D*/A line to indicate to the
module whether the incoming serial stream is data
or address information.
29SLOT0SEL*Slot 0 Select – Taps into the SCXIbus INTR* line
to indicate whether the information on MOSI is
being sent to a module or Slot 0.
36SCANCLKScan Clock – Indicates to the SCXI-1120 that a
sample has been taken by the data acquisition board
and causes the SCXI-1120 to change channels.
37SERCLKSerial Clock – This signal taps into the SCXIbus
SPICLK line to clock the data on the MOSI and
MISO lines.
43RSVDReserved.
All other pins are not connected.
See the Timing Requirements and Communication Protocol section later in this chapter for more
The signals on the rear signal connector can be classified as analog output signals, digital I/O
signals, or timing I/O signals. Signal connection guidelines for each of these groups are given in
the following section.
Analog Output Signal Connections
Pins 1 through 19 of the rear signal connector are analog output signal pins. Pins 1 and 2 are
AOGND signal pins. AOGND is an analog output common signal that is routed through jumper
W46 to the analog reference on the SCXI-1120. You can use these pins for a general analog
power ground tie point to the SCXI-1120 if necessary. In particular, when using differential
input data acquisition boards, such as the MIO-16 board, it is preferable to leave jumper W46 in
its factory setting or in position AB-R1 to avoid ground loops. With data acquisition boards that
are configured for referenced single-ended (RSE) measurements, jumper W46 should be in
position AB-R0 to connect the SCXI-1120 analog ground reference to the data acquisition analog
ground. Pin 19 is the OUTREF pin, and is connected internally to the analog reference when
jumper W46 is in position AB-R2. Pins 3 through 18 are the analog output channels of the
SCXI-1120. Pins 3 and 4, or MCH0±, are a multiplexed output of all eight channels and the
temperature sensor output. Pins 5 through 18, or MCH1± through MCH7±, are a parallel
connection of channels 1 through 7 to the rear signal connector. Pin 18 is a direct connection to
the temperature sensor when jumper W41 is in position 3. Notice that the temperature sensor is
located on the terminal block. For further details on configuring the temperature sensor output,
refer to the SCXI-1320 and SCXI-1328 Terminal Blocks section earlier in this chapter.
Warning: The SCXI-1120 analog outputs are not overvoltage protected. Applying external
voltage to these outputs can result in damage to the SCXI-1120. National
Instruments is not liable for any damages resulting from such signal connections.
Note: The SCXI-1120 analog outputs are short-circuit protected.
Digital I/O Signal Connections
Pins 24 through 27, 29, 33, 36, 37, and 43 constitute the digital I/O lines of the rear signal
connector. They are divided into three categories–the digital input signals, the digital output
signals, and the digital timing signals.
The digital input signals are pins 24, 25, 27, 29, 33, and 37. The data acquisition board uses
these pins to configure the SCXI module that is under data acquisition board control. Each
digital line emulates the SCXIbus communication signals as follows:
•Pin 25 is SERDATIN and is equivalent to the SCXIbus MOSI serial data input line.
•Pin 27 is DAQD*/A and is equivalent to the SCXIbus D*/A line. It indicates to the module
whether the incoming serial stream on SERDATIN is data (DAQD*/A = 0) or address
(DAQD*/A = 1) information.
• Pin 29 is SLOT0SEL* and is equivalent to the SCXIbus INTR* line. It indicates whether the
data on the SERDATIN line is being sent to Slot 0 (SLOT0SEL* = 0) or to a module
(SLOT0SEL* = 1).
• Pins 24 and 33 are the digital ground references for the data acquisition board digital signals
and are tied to the module digital ground.
• Pin 37 is SERCLK and is equivalent to the SCXIbus SPICLK line and is used to clock the
serial data on the SERDATIN line into the module registers.
•Pin 26 is SERDATOUT and is equivalent to SCXIbus MISO when jumper W43 is in
position 1.
The digital I/O signals of the SCXI-1120 match the digital I/O lines of the MIO-16 board. When
used with an SCXI-1341, SCXI-1342, or SCXI-1344 cable assembly, the SCXI-1120 signals
match the digital lines of the Lab-NB/Lab-PC+/Lab-LC boards and the PC-LPM-16 board,
respectively. Table 2-9 lists the equivalences. For more information, consult Appendix E,
SCXI-1120 Cabling.
Table 2-9. SCXIbus to SCXI-1120 Rear Signal Connector to
•Pin 36 is used as a clock by the SCXI-1120 to increment the MUXCOUNTER after each
conversion by the data acquisition board during scanning. This signal is referred to as
SCANCLK. See Chapter 3, Theory of Operation, for a description of MUXCOUNTER.
•Pin 43 is a reserved digital input.
The following specifications and ratings apply to the digital I/O lines.
Absolute maximum voltage
input rating5.5 V with respect to DIG GND
Digital input specifications (referenced to DIG GND):
V
input logic high voltage2 V minimum
IH
input logic low voltage0.8 V maximum
V
IL
I
input current leakage±1 µA maximum
I
Digital output specifications (referenced to DIG GND):
V
output logic high voltage3.7 V minimum at 4 mA maximum
OH
output logic low voltage0.4 V maximum at 4 mA maximum
Timing Signal
The data acquisition timing signal is SCANCLK.
SCANCLK is used to increment MUXCOUNTER on its rising edge. Figure 2-12 shows the
timing requirements of the SCANCLK signal. These requirements will ensure that SCANCLK
is properly transmitted over TRIG0.
T
high
SCANCLK
T
low
T
T
low
high
Time low before rising edge400 nsec minimum
Time high before falling edge250 nsec minimum
Figure 2-12. SCANCLK Timing Requirements
For output selection time specifications, refer to Appendix A, Specifications.
Communication Signals
This section describes the methods for communicating on the Serial Peripheral Interface (SPI)
bus and their timing requirements. The communication signals are SERDATIN, DAQD*/A,
SLOT0SEL*, SERDATOUT, and SERCLK. Furthermore, SS* is produced by Slot 0 according
to data acquisition board programming, and SS* timing relationships will also be discussed. For
information on the Slot 0 Slot-Select Register, consult Chapter 4, Register Descriptions.
The data acquisition board determines to which slot it will talk by writing a slot-select number to
Slot 0. In the case of an SCXI-1001 chassis, this write also determines to which chassis the data
acquisition board will talk. Writing a slot-select number is also used in programming the Slot 0
hardscan circuitry. See Chapter 5, Programming, for information on programming the Slot 0
hardscan circuitry.
The following sections detail the procedure for selecting a slot in a particular chassis.
Figure 2-13 illustrates the timing of this procedure with the example case of selecting Slot 11 in
Chassis 9. Notice that the factory-default chassis address for the SCXI-1000 is address 0. For
information on changing the address of your chassis, consult the SCXI-1000/1001 User Manual.
An SCXI-1000 chassis will respond to any chassis number.
2. Clear SLOT0SEL* to 0. This will deassert all SS* lines to all modules in all chassis.
3. For each bit, starting with the MSB, perform the following action:
a. SERDATIN = bit to be sent. These bits are the data that are being written to the
Slot-Select Register.
b. SERCLK = 0.
c. SERCLK = 1. This rising edge clocks the data.
4. Set SLOT0SEL* to 1. This will assert the SS* line of the module whose slot number was
written to Slot 0. If multiple chassis are being used, only the appropriate slot in the chassis
whose address corresponds to the written chassis number will be selected. When no
communication is taking place between the data acquisition board and any modules, you
should write zero to the Slot-Select Register to ensure that no accidental writes occur.
Figure 2-14 shows the timing requirements on the SERCLK and SERDATIN signals. You must
observe these timing requirements for all communications. T
Minimum low time65 nsec minimum
Minimum high time400 nsec minimum
SERDATIN setup time 200 nsec minimum
SERDATIN hold time200 nsec minimum
SERDATOUT delay350 nsec maximum
T
hold
Figure 2-14. Serial Data Timing Diagram
After the Slot-Select line to an SCXI-1120 has been asserted, you can write to its Configuration
Register and read from its Module ID Register using the following protocols. The contents of
the Module ID Register are reinitialized by deasserting Slot-Select. After the 32 bits of data are
read from the Module ID Register, further data will be zeros until reinitialization occurs.
To write to the Configuration Register, follow these steps:
1. Initial conditions:
SS* asserted low.
SERDATIN = don't care.
DAQD*/A = 0 (indicates data will be written to Configuration Register).
SLOT0SEL* = 1.
SERCLK = 1 (and has not transitioned since SS* went low).
2. For each bit to be written:
Establish the desired SERDATIN level corresponding to this bit.
SERCLK = 0.
SERCLK = 1. Clock the data.
3. Pull SLOT0SEL* low to deassert the SS* line and establish conditions for writing a new
slot-select number to the Slot 0 Slot-Select Register.
4. If you are not selecting another slot, you should write zero to the Slot 0 Slot-Select Register.
Figure 2-15 illustrates a write to the SCXI-1120 Configuration Register of the binary pattern:
To read from the Module ID Register, follow these steps:
1. Initial conditions:
SS* asserted low.
SERDATIN = don't care.
DAQD*/A = 1. Make sure DAQD*/A does not go low or erroneous data will be written
to the Configuration Register.
SLOT0SEL* = 1.
SERCLK = 1 (and has not changed since SS* went low).
2. For each bit to be read:
SERCLK = 0.
SERCLK = 1. Clock the data.
Read the level of the SERDATOUT line.
3. Pull SLOT0SEL* low to deassert the SS* line and establish conditions for writing a new
slot-select number to the Slot 0 Slot-Select Register.
4. If you are not selecting another slot, you should write zero to the Slot 0 Slot-Select Register.
Figure 2-16 illustrates a read of the SCXI-1120 Module ID Register.
SLOT0SEL*
SS*
SERCLK
T
SERDATOUT
T
delay
01000000000000000000000000000000
byte 0 = 4byte 3 = 0byte 2 = 0byte 1 = 0
SS* high to SERDATOUT high350 nsec maximum
delay
Figure 2-16. SCXI-1120 Module ID Register Timing Diagram
For further details on programming these signals, refer to Chapter 5, Programming.
The major components of the SCXI-1120 are as follows:
•SCXIbus connector
•Digital interface
•Digital control circuitry
•Timing and analog circuitry
The SCXI-1120 consists of eight isolated amplifier channels with gains of 1, 2, 5, 10, 20, 50,
100, 200, 500, 1,000, and 2,000. The SCXI-1120 also has a digital section for automatic control
of channel scanning, temperature selection, and MUXCOUNTER clock selection.
The theory of operation of each of these components is explained in the rest of this chapter.
SCXIbus Connector
Figure 3-2 shows the pinout of the SCXIbus connector.
A24TRIG0TRIG0 – General-purpose trigger line used by the
SCXI-1120 to send SCANCLK to other modules or
receive SCANCLK from other modules. Open
collector. I/O.
B24SS*Slot Select – When low, enables module
communications over the SCXIbus. Totem pole.
Input.
C24SCANCONScanning Control – Combination output enable and
reload signal for scanning operations. Totem pole.
Input.
All other pins are not connected.
MOSI, MISO, SPICLK, and SS* form a synchronous communication link that conforms with
SPI using an idle-high clock and second-edge data latching. D*/A, INTR*, and RESET* are
additional control signals.
When the module is being used in an SCXI-1000 or SCXI-1001 chassis, the data acquisition
board, via the module rear signal connector, must tap into the open-collector backplane signal
lines as a master to write to the module. The signal connections from the rear signal connector to
the backplane are shown in Table 3-1.
Table 3-1. SCXIbus Equivalents for the Rear Signal Connector
Rear Signal Connector
SCXIbus Equivalent
Signal
SERDATINMOSI
DAQD*/AD*/A
SLOT0SEL*INTR*Jumper W44 must be set to position 1
(Revision A and B modules only)
SERCLKSPICLK
SERDATOUTMISOJumper 43 must be set to position 1
The SCXI-1120 module converts the data acquisition board signals to open-collector signals on
the backplane of the SCXI chassis. In order for the data acquisition board to talk to a slot, the
board must first assert the SS* for that slot. This is done by asserting INTR* low, writing a
16-bit number over MOSI corresponding to the desired slot (and chassis if an SCXI-1001 chassis
is being used), and then releasing INTR* high. At this point, the SS* of the desired slot is
asserted low and the data acquisition board can communicate with the module in that slot
according to the SPI protocol.
Figure 3-3 shows a diagram of the SCXI-1120 and SCXIbus digital interface circuitry.
SCXIbus
SS*
MOSI
D*/A
MISO
INTR*
SPICLK
SERDATIN
Buffered Serial
Data
Digital
Interface
Buffered Digital
Signal Controls
DAQD*/A
SLOT0SEL*
SERCLK
SERDATOUT
Rear Signal Connector
Figure 3-3. Digital Interface Circuitry Block Diagram
The digital interface circuitry is divided into a data acquisition section and an SCXIbus section.
The SCXI-1120 connects to the SCXIbus via a 4x24 metral receptacle and to the data acquisition
board via a 50-pin ribbon-cable header. The digital interface circuitry buffers the digital signals
from the data acquisition board and the SCXIbus and sends signals back and forth between the
data acquisition board and the SCXIbus.
Figure 3-4 diagrams the SCXI-1120 digital control.
Serial Data Out
Input Channel
Select
Module ID Register
Configuration
Register
Output
Stage
Control
Hardware
Control
SCANCLK
Path
Control
Scan
Figure 3-4. SCXI-1120 Digital Control
Buffered
Serial Data In
Buffered Digital
Control Signals
The digital control section consists of the Configuration Register and the Module ID Register.
The Configuration Register is a two-byte, serial-in, parallel-out shift register. Data is received on
the MOSI line from either Slot 0 or the data acquisition board when SS* is enabled and D*/A
indicates data transfer (D*/A low). The Configuration Register provides channel selection and
configures the SCXI-1120 for scanning options. All the control bits are fed into a latch before
being routed to the rest of the module. The channel-select bits are taken directly from the shift
register. Complete descriptions of the register bits are given in Chapter 4, Register Descriptions.
Writes to the Configuration Register require the following steps:
1. SS* goes low, enabling communication with the board.
2. D*/A goes low, indicating that the information sent on the MOSI line is data.
3. The serial data is available on MOSI and SPICLK clocks it into the register.
4. SS* goes high and D*/A goes high, indicating an end of communication. This action latches
the Configuration Register bits.
When the SCXIbus is reset, all bits in the Configuration Register are cleared.
The Module ID Register connects to MISO on the SCXIbus. The Module ID Register is an 8-bit,
parallel/serial-in, serial-out shift register and an SPI communication adapter. The contents of the
Module ID Register are written onto MISO during the first four bytes of transfer after SS* has
been asserted low. Zeros are written to MISO thereafter until SS* is released and reasserted.
The SCXI-1120 module ID is hex 00000004.
Analog and Timing Circuitry
The SCXIbus provides analog power (±18.5 VDC) that is regulated on the SCXI-1120 to
±15 VDC, a guard, an analog bus (AB0±), and a chassis ground (CHSGND). AB0± buses the
SCXI-1120 output to other modules or receives outputs from other modules via the SCXIbus.
The guard guards the analog bus and can be connected via jumper W46 to the analog ground
reference or can be left floating (a connection can be made by another board).
The data acquisition board analog input and timing is the interface between the SCXI-1120
output and the data acquisition board. This is described in the following section.
Analog Input Channels
Figure 3-5 is a diagram of the analog input block.
The analog input consists of eight isolated single-ended noninverting amplifiers. In addition,
lowpass filtering is available at the inputs. You can jumper select one of two bandwidths,
10 kHz or 4 Hz. The amplifier gain is divided into two stages, a first stage providing gains of 1,
10, 50, and 100, and a second stage providing gains of 1, 2, 5, 10, and 20. Each channel is
configurable to a different bandwidth and gain.
Use the following formula to determine the overall gain of a given amplifier input channel:
G
total
= G
1st
x G
2nd
where G
is the overall gain and G
total
and G
1st
are the first- and second-stage gains. Here it is
2nd
important to note that the choice of gain in each stage will affect the amplifier bandwidth. To
determine the bandwidth of a given gain stage use the following formula:
BW = GBWP/G
where BW is a given amplifier stage bandwidth, GBWP is the gain bandwidth product (typically
800 kHz), and G is the gain at this stage. This BW might be of concern at high first-stage gains
such as 50 and 100. In this case, the first-stage amplifier has a BW equal to 16 kHz and 8 kHz,
respectively. Due to this decrease in the amplifier bandwidth, you will notice a decrease in the
channel overall bandwidth, but a better noise immunity. If this bandwidth limitation is
unacceptable, you should spread the gains over both stages, thus increasing the BW of each
amplifier stage. This will introduce, in most cases, a negligible effect on the channel bandwidth.
For example, to achieve a gain of 100, use G
G
= 50 and G
1st
2nd
= 20.
= 10 and G
1st
= 10; for a gain of 1,000, use
2nd
All the amplifier input channels are overvoltage protected to 240 Vrms with power on or off.
The isolated amplifiers fulfill two purposes on the SCXI-1120 module. They convert a small
signal riding on a high common-mode voltage into a single-ended signal with respect to the
SCXI-1120 chassis ground. With this conversion, the input analog signal can be extracted from
a high common-mode voltage or noise before being sampled and converted by the data
acquisition board. The isolated amplifier also amplifies and conditions an input signal, which
results in an increase in measurement resolution and accuracy.
After isolation, further filtering is available to increase the noise immunity of the amplifier
channel. It is important to note that the overall amplifier bandwidth is determined by both
filtering stages, so to achieve the required bandwidth, both filtering sections should be set the
same, as indicated in Chapter 2, Configuration and Installation.
Calibration
Calibration Equipment Requirements
For best measurement results, calibrate the SCXI-1120 so that its offset is adjusted to 0 ± 3 mV
RTO and 0 ± 6 µV RTI. No special equipment is needed other than a regular voltmeter with the
following specifications.
A multiranging 3 1/2-digit digital multimeter can provide you with the necessary function as
described previously. We will refer to the measuring instrument as a digital multimeter (DMM).
Each channel on the SCXI-1120 has two potentiometers dedicated for calibration. One
potentiometer is used to null the output offset; the other is used to null the input offset.
Offset Null Adjust
To null the offset of the amplifier channels, complete the following steps:
1.Set the DMM range to the smallest range that can measure ±3 mV.
2.Short the inputs of the DMM together and then to chassis ground.
3.Record the measurement indicated by the DMM display. This is the DMM inherent offset
and it should be subtracted from subsequent measurements.
4.Short the desired channel inputs together and then to chassis ground.
5.Set the amplifier gain to 1.
6.Connect the amplifier output to the DMM. Make sure that the DMM can achieve the
accuracy and resolution you need.
7.Adjust the output potentiometer of the desired channel until the output is 0 ± 3 mV.
8.Set the DMM range to the smallest range that can measure ±6 mV.
9.Set the amplifier gain to 1,000.
10.Adjust the input potentiometer of the desired channel until the output is 0 ± 6 mV.
11.Go to the next channel.
To avoid erroneous results when nulling the amplifier, follow these steps in the order indicated.
You may seal the potentiometers after calibration with antisabotage lacquer to avoid tampering
with the calibration.
Table 3-2 lists the potentiometer reference designators that correspond to each channel.
The SCXI-1120 output circuitry consists of a buffered-output multiplexer and channel-select
hardware. The channel-select hardware consists of a three-bit counter, MUXCOUNTER. This
counter is needed when the board is operating in the Multiplexed-Output mode. The counter
output is sent to the output multiplexer address pins to determine which of the eight channels is
to be connected to MCH0. In the Single-Channel Read mode, the MUXCOUNTER is loaded
with the desired channel number. In the Scanning mode, the counter is loaded with the first
channel to be read. During the scan, the counter is clocked by SCANCLK from the data
acquisition board, or TRIG0 from the SCXIbus, depending on the state of the CLKSELECT bit
in the Configuration Register. During scanning operations, the MUXCOUNTER is reloaded
with the channel value stored in the Configuration Register when SCANCON is high (inactive),
and will count upwards on each rising clock edge when SCANCON is low (active). In the
Parallel-Output mode, the MUXCOUNTER is disabled and its output indicates binary 00; thus,
amplifier Channel 0 is selected at the output multiplexer and is connected to MCH0. The seven
other channels are hardwired to MCH1 through MCH7 on the rear signal connector.
The output multiplexer multiplexes all eight amplifier outputs and the temperature-sensor
reading provided on the MTEMP line. To read the temperature sensor when it is multiplexed
with the other input channels, the RTEMP bit of the Configuration Register must be set high.
This measurement is only software controlled. For hardware control of the temperature sensor
reading, connect the temperature sensor to pin 18 on the rear signal connector as described in
Chapter 2, Configuration and Installation. The multiplexer output connects to the MCH0± and
is connected to the data acquisition board analog channel input. In the case of the MIO data
acquisition boards, MCH0± on the rear signal connector corresponds to ACH0 and ACH8.
Furthermore, the multiplexed output of the SCXI-1120 can be bused, via switches to AB0± on
the SCXIbus, to other modules. When you use multiple modules, it is possible to bus the module
output via AB0 to the module that connects to the data acquisition board. In this case, the AB0
switches of all the modules are closed, whereas the output multiplexer of all the modules, except
the one being read, are disabled. Refer to chapters 2 and 5 for further details on how to configure
and program multiple modules.
In addition to the Multiplexed-Output mode described in the previous paragraph, it is possible to
operate the SCXI-1120 in the Parallel-Output mode. In this mode, you need no software–other
than the software used with your data acquisition board–to control the scanning of the eight
channels or to perform a single read. To access the temperature sensor in this mode, configure
the temperature sensor in the DTS mode. At power up or reset, amplifier Channel 0 is selected
on the output multiplexer, and thus connects to MCH0. The other seven amplifier channels are
hardwired to the rear signal connector. Notice that when the Multiplexed-Output mode is
selected, pins 5 through 18 on the rear signal connector are still driven by the SCXI-1120. The
SCXI-1120 outputs on the rear signal connector are short-circuit protected.
Refer to the following Scanning Modes section for further details on how to scan the SCXI-1120
channels.
Scanning Modes
The SCXI-1120 has four basic types of scanning modes–single-module parallel scanning,
single-module multiplexed scanning, multiple-module multiplexed scanning, and
multiple-chassis scanning, which is possible only with the SCXI-1001 chassis. For additional
information, consult Chapter 2, Configuration and Installation, Chapter 5, Programming, your
data acquisition board user manual, or your SCXI chassis user manual. If you need more
information, contact National Instruments.
Single-Module Parallel Scanning
Single-module parallel scanning is the simplest scanning mode. Directly cable the SCXI-1120 to
the data acquisition board as shown in Figure 3-7. In this configuration, each analog signal has
its own channel. Timing signals are not necessary for this type of scanning because the module
provides all channels to the data acquisition board at all times. You can implement
single-module parallel scanning with any data acquisition board that is appropriately cabled to
the SCXI-1120.
Eight Isolated
Floating
Single-Ended
Inputs
CH0
CH1
CH2
CH3
CH4
CH5
CH6
CH7
SCXI-1120
Cable Assembly
MCH0
MCH1
MCH2
MCH3
MCH4
MCH5
MCH6
MCH7
Data Acquisition Board
Analog Input 0
Analog Input 1
Analog Input 2
Analog Input 3
Analog Input 4
Analog Input 5
Analog Input 6
Analog Input 7
Figure 3-7. Single-Module Parallel Scanning
Multiplexed Scanning
Only the MIO-16 data acquisition boards support multiplexed scanning on the SCXI-1120.
During multiplexed scanning, a module sends the SCANCLK signal to Slot 0 over the TRIG0
backplane line, and Slot 0 sends SCANCON signals to each module. Each module uses its
SCANCON signal to reload MUXCOUNTER and to determine when the SCXI-1120 output is
enabled. Slot 0 contains a module scan list first-in-first-out (FIFO) memory chip, similar to the
Channel/Gain FIFO on an MIO-16 board, except that instead of having a channel number and
gain setting for each entry, the Slot 0 FIFO contains a slot number and a sample count for each
entry. The list in Slot 0 will determine which module is being accessed and for how many
samples. It is important that you make sure that the lists on the data acquisition board and Slot 0
are compatible so that the samples are acquired as intended. See your SCXI chassis manual for
more information.
Single-Module Multiplexed Scanning
Single-Module Multiplexed Scanning (Direct)
This is the simplest multiplexed-scanning mode. Directly cable the SCXI-1120 to the data
acquisition board as shown in Figure 3-8. The module sends SCANCLK onto TRIG0, and Slot 0
sends SCANCON back to the module. SCANCON will be low at all times during the scan
except during changes from one Slot 0 scan list entry to the next, when SCANCON pulses high
to make the MUXCOUNTER reload its starting channel. Notice that although you are using
only a single module, you can put many entries with different counts in the Slot 0 FIFO, so that
some channels are read more often than others. You cannot change the start channel in the
module Configuration Register during a scan.
Single-Module Multiplexed Scanning (Indirect)
In this mode, the SCXI-1120 is not directly cabled to the data acquisition board. Instead, you
connect another module to the data acquisition board, and the analog output of the SCXI-1120 is
sent over Analog Bus 0, through the intermediate module, and then to the data acquisition board.
The SCXI-1120 receives its MUXCOUNTER clock from TRIG0, which is sent by the
intermediate module, as shown in Figure 3-9. Slot 0 operation is the same for direct connection
scanning.
In this mode, all the modules tie into Analog Bus 0 and SCANCON enables the output of their
amplifiers. The module that is directly cabled to the data acquisition board sends SCANCLK
onto TRIG0 for the other modules and Slot 0, as shown in Figure 3-10. The scan list in Slot 0 is
programmed with the sequence of modules and the number of samples per entry.
SCXI-1000 or SCXI-1001 Chassis
SCANCON X
Data Acquisition Board
Timing
Output
Analog
Input
Cable Assembly
SCANCON A
SLOT 0
SCANCON B
SLOT A
TRIG0
SCXI Module
SLOT B
SCANCLK
MCH0
SCXI ModuleSCXI Module
SLOT X
Analog Bus 0
Figure 3-10. Multiple-Module Multiplexed Scanning
Multiple-Chassis Scanning
In this mode, you attach each SCXI-1001 chassis to a daisy chain of cable assemblies and
multichassis adapter boards, as shown in Figure 3-11. You program each chassis separately, and
each chassis occupies a dedicated channel of the data acquisition board. Within each chassis,
scanning operations act as if the other chassis are not being used, with one exception–you must
program the Slot 0 scan list in each chassis with dummy entries of Slot 13 to fill the samples
when the data acquisition board will be sampling another chassis or data acquisition board
channel. This will keep the chassis synchronized. Notice that you can only perform
multiple-chassis scanning with the SCXI-1001 chassis and MIO-16 data acquisition boards. See
Chapter 5, Programming, for more information on multiple-chassis scanning. See Appendix E,
SCXI-1120 Cabling, for more information on the necessary cable accessories for multichassis
scanning.
This chapter describes in detail the SCXI-1120 Module ID Register, the Configuration Register,
the Slot 0 registers, and multiplexer addressing.
Note: If you plan to use a programming software package such as NI-DAQ, LabWindows, or
LabVIEW with your SCXI-1120 board, you do not need to read this chapter.
Register Description
Register Description Format
This register description chapter discusses each of the SCXI-1120 registers and the Slot 0
registers. A detailed bit description of each register is given. The individual register description
gives the type, word size, and bit map of the register, followed by a description of each bit.
The register bit map shows a diagram of the register with the MSB shown on the left (bit 15 for a
16-bit register, bit 7 for an 8-bit register), and the LSB shown on the right (bit 0). A rectangle is
used to represent each bit. Each bit is labeled with a name inside its rectangle. An asterisk (*)
after the bit name indicates that the bit is inverted (negative logic). The Module ID register has a
unique format and is described in the Module ID Register section.
In many of the registers, several bits are labeled with an X, indicating don't care bits. When you
write to a register, you may set or clear these bits without effect.
SCXI-1120 Registers
The SCXI-1120 has two registers. The Module ID Register is a four-byte, read-only register that
contains the Module ID number of the SCXI-1120. The Configuration Register is a 16-bit,
write-only register that controls the functions and characteristics of the SCXI-1120.
The Module ID Register contains the 4-byte module ID code for the SCXI-1120. This code
number will be read as the first four bytes on the MISO line whenever the module is accessed.
The bytes will appear least significant byte first. Within each byte, data is sent out MSB first.
Additional data transfers will result in all zeros being sent on the MISO line. The Module ID
Register is reinitialized to its original value each time the SCXI-1120 is deselected by the SS*
signal on the backplane.
The Configuration Register contains 16 bits that control the functions of the SCXI-1120. When
SS* is asserted (low) and D*/A indicates data (low), the register will shift in the data present on
the MOSI line, bit 15 first, and then latch it when the SCXI-1120 is deselected by the SS* signal
on the backplane. The Configuration Register initializes to all zeros when the SCXI chassis is
reset or first turned on.
Type:Write-only
Word Size:16-bit
Bit Map:
15141312111098
CLKOUTENCLKSELECTXXXCHAN2CHAN1CHAN0
76543210
XXRTEMPRSVDSCANCLKENSCANCONENAB0ENFOUTEN*
BitNameDescription
15CLKOUTENScan Clock Output Enable – This bit determines whether
the SCANCLK signal from the rear signal connector is sent
out, in inverted form, to the TRIG0 backplane signal. If
CLKOUTEN is set to 1, SCANCLK* is transmitted on
TRIG0. If CLKOUTEN is cleared to 0, SCANCLK* is not
transmitted on TRIG0.
14CLKSELECTScan Clock Select – This bit determines whether the
SCXI-1120 uses SCANCLK or the inverted form of TRIG0
to clock the MUXCOUNTER for the purposes of scanning
through the analog channels. If CLKSELECT is cleared
to 0, SCANCLK is used to clock MUXCOUNTER. If
CLKSELECT is set to 1, TRIG0* is used as the source to
clock MUXCOUNTER.
13-11, 7-6XDon't care bits – Unused.
10-8CHAN<2..0>Channel Select – These bits determine the channel number
(zero to seven) that is loaded into the MUXCOUNTER to
determine the analog channel to be read during a single
read, or the starting channel on the module for a scanned
Slot 0 has three registers. The Slot-Select Register is a 16-bit, write-only register that determines
with which slot the data acquisition board will speak when SLOT0SEL* is released high. In the
case of the SCXI-1001 chassis, the Slot-Select Register also determines in which chassis the
desired slot is. The FIFO Register is a 16-bit, write-only register used for storing the Slot 0 scan
list that determines the chassis scan sequence. The Hardscan Control Register (HSCR) is an
8-bit, write-only register used for setting up the timing circuitry in Slot 0. The Slot-Select
Register is written to by using the SLOT0SEL* line. The HSCR and the FIFO Register are
written to as if they were registers located on modules in Slots 13 and 14. You should maintain
software copies of the Slot-Select Register, HSCRs, and all the Slot 0 scan lists that correspond
to the writes to FIFO Registers.
If you are using multiple chassis, it is important to understand the architectural differences of the
Slot-Select Register as compared to the HSCR and the FIFO Register. Although each chassis
has its own physical Slot-Select Register, all are written to at the same time. The jumper settings
in Slot 0 of a chassis determine with which chassis number Slot 0 is identified. From the
software perspective, only one Slot-Select Register exists in a system composed of multiple
chassis. The HSCR and FIFO Register, on the other hand, are unique to each chassis and you
must program them separately.
The Slot-Select Register contains 16 bits that determine which module in which chassis will be
enabled for communication when the SLOT0SEL* line is high. An SCXI-1000 chassis will
select the appropriate module in its chassis, regardless of the chassis number written. The
Slot-Select Register will shift in the data present on the MOSI line, bit 16 first, when
SLOT0SEL* is low.
Type:Write-only
Word Size:16-bit
Bit Map:
15141312111098
XXXXXXXCHS4
76543210
CHS3CHS2CHS1CHS0SL3SL2SL1SL0
BitNameDescription
15-9XDon't care bits – Unused.
8-4CHS<4..0>Chassis Bit 4 through 0 – These bits determine
which chassis is selected. On the SCXI-1000
chassis, these are don't care bits.
3-0SL<3..0>Slot Bit 3 through 0 – These bits determine which
The HSCR contains eight bits that control the setup and operation of the hardscan timing
circuitry of Slot 0. To write to the HSCR, follow the procedure given in the Register Writes
section in Chapter 5, Programming, using 13 as the slot number and writing eight bits to the
HSCR. The register will shift in the data present on the MOSI line, bit 7 first, when Slot 13 is
selected by the Slot-Select Register.
Type:Write-only
Word Size:8-bit
Bit Map:
76543210
RSVDFRTRDONCEHSRS*LOAD*SCANCONENCLKEN
BitNameDescription
7RSVDReserved.
6FRTForced Retransmit – This bit, when clear, causes the scan
list in the FIFO to be reinitialized to the first entry, thus
allowing the scan list to be reprogrammed in two steps
instead of having to rewrite the entire list. When this bit is
set, it has no effect.
5RDRead – This bit, when clear, prevents the FIFO from being
read. When set, the FIFO is being read except at the end
of a scan list entry during scanning, when reading is briefly
disabled to advance to the next scan list entry.
4ONCEOnce – When set, this bit will cause the Hardscan circuitry
to shut down at the end of the scan list circuitry during a
data acquisition. When clear, the circuitry will wrap
around and continue seamlessly with the first scan list entry
after the entry is finished.
3HSRS*Hardscan Reset – When clear, this bit causes all the
hardware scanning circuitry, including the FIFO, to be reset
to the power up state. When set, this bit has no effect.
2LOAD*Load – This bit, when clear, forces a loading of the Slot 0
sample counter with the output of the FIFO. When set,
this bit has no effect.
1SCANCONENScan Control Enable – When set, this bit enables the
SCANCON lines. When clear, all SCANCON lines are
disabled (high).
0CLKENClock Enable – When set, this bit enables TRIG0 as a
clock for the hardscan circuitry. When clear, TRIG0 is
The FIFO Register is used to add entries to the Slot 0 FIFO. The FIFO contains the Slot 0 scan
list. Each entry contains a slot number to be accessed, and a count number to determine the
number of samples to be taken from that slot. To write to the FIFO Register, follow the
procedure given in the Register Writes section in Chapter 5, Programming, using 14 as the slot
number, and writing 16 bits to the FIFO Register. The register will shift in the data present on
the MOSI line, bit 7 first, when Slot 14 is selected by the Slot-Select Register. The Slot 0 scan
list is created by consecutive writes to the FIFO Register. Each write creates a new entry at the
end of the scan list. The maximum number of entries is 256. To clear the FIFO of all entries,
clear the HSRS* bit in the HSCR.
Type:Write-only
Word Size:16-bit
Bit Map:
15141312111098
XXXXXMOD3MOD2MOD1
76543210
MOD0CNT6CNT5CNT4CNT3CNT2CNT1CNT0
BitNameDescription
15-11XDon't care bits – Unused.
10-7MOD<3..0>Module Number – The value of these bits plus one
determines the number of the slot to be accessed for
this scan entry. For example, to access Slot 6,
MOD<3..0> would be 0101.
6-0CNT<6..0>Count – The value of these bits plus one determines
how many samples will be taken before the next
scan list entry becomes active. A value of zero
corresponds to one sample and a value of 127
corresponds to 128 samples.
This chapter contains a functional programming description of the SCXI-1120 and Slot 0.
Note: If you plan to use a programming software package such as NI-DAQ, LabWindows, or
LabVIEW with your SCXI-1120 board, you do not need to read this chapter.
Programming Considerations
Programming the SCXI-1120 involves writing to the Configuration Register. Programming
Slot 0 involves writing to the HSCR and FIFO Register. Programming the data acquisition
boards involves writes to their registers. See your data acquisition board user manual for more
information. The programming instructions list the sequence of steps to take. The instructions
are language independent; that is, they instruct you to write a value to a given register without
presenting the actual code.
Notation
For the bit patterns to be written, the following symbols are used:
0binary zero
1binary one
Xdon't care; either zero or one may be written
Cone of three bits used to specify the channel to be loaded into the MUXCOUNTER. This
value will either be the channel to be read for single reads, or a starting channel for
scanned measurements.
The 16-bit patterns are presented MSB first, left to right.
Register Writes
This section describes how to write to the Configuration Register, HSCR, and FIFO Register,
including the procedure for writing to the Slot-Select Register to select the appropriate slot. For
timing specifics, refer to the Timing Requirements and Communication Protocol section in
Chapter 2, Configuration and Installation. The rear signal connector pin equivalences to the
different National Instruments data acquisition boards are given in Table 5-1. Also see
Appendix E, SCXI-1120 Cabling. The Configuration Register, the HSCR, and the FIFO Register
are write-only registers.
The different bits in these registers often control independent pieces of circuitry. There are times
when you may want to set or clear a specific bit or bits without affecting the remaining bits.
However, a write to one of these registers will affect all bits simultaneously. You cannot read
the registers to determine which bits have been set or cleared in the past; therefore, you should
maintain a software copy of these registers. You can then read the software copy to determine
the status of the register. To change the state of a single bit without disturbing the remaining
bits, set or clear the bit in the software copy and write the software copy to the register.
2. Clear SLOT0SEL* to 0. This will deassert all SS* lines to all modules in all chassis.
3. For each bit, starting with the MSB (bit 15):
a. SERDATIN = bit to be sent. These bits are the data that is being written to the
Slot-Select Register.
b. Clear SERCLK to 0.
c. Set SERCLK to 1. This rising edge clocks the data. (If you are using an MIO-16 board,
writing to the EXTSTROBE* register will pulse EXTSTROBE* low and then high,
accomplishing steps 3b and 3c.)
4. Set SLOT0SEL* to 1. This will assert the SS* line of the module whose slot number was
written to Slot 0. If you are using multiple chassis, only the appropriate slot in the chassis
whose address corresponds to the written chassis number will be selected automatically.
When no communications are taking place between the data acquisition board and any
modules, write zero to the Slot-Select Register to ensure that no accidental writes occur.
5. If you are writing to a Configuration Register, clear DAQD*/A to 0 (this indicates data will
be written to Configuration Register). If you are writing to the HSCR or FIFO Register,
leave DAQD*/A high.
6. For each bit to be written to the Configuration Register:
a. Establish desired SERDATIN level corresponding to this bit.
b. Clear SERCLK to 0.
c. Set SERCLK to 1 (clock the data). (If you are using an MIO-16 board, writing to the
EXTSTROBE* register will pulse EXTSTROBE* low and then high, accomplishing
steps 6b and 6c.)
7. Pull SLOT0SEL* low to deassert the SS* line, latch the data into the Configuration Register
and establish conditions for writing a new slot-select number to the Slot 0 Slot-Select
Register.
8. If you are not selecting another slot, write zero to the Slot 0 Slot-Select Register. If you are
selecting another slot, start at step 3.
For a timing illustration of a Configuration Register write, see Figure 2-15, ConfigurationRegister Write Timing Diagram, which shows the proper write to configure an SCXI-1120 that is
directly cabled to an MIO-16 board for multiple-module multiplexed scanning with a start
channel of 3.
Initialization
The SCXI-1120 powers up with its Configuration register cleared to all zeros. You can force this
state by sending an active low signal on the RESET* pin of the SCXIbus connector. In the reset
state, CH0 through CH3 are routed to MCH0 through MCH3 on the rear signal connector. The
module is disconnected from Analog Bus 0 and disabled from scanning.
Single-Channel Measurements
This section describes how to program the SCXI-1120, either alone or in conjunction with other
modules, to make single-channel, or nonscanned, measurements.
Direct Measurements
Parallel Output
To perform a parallel output measurement, you must cable the SCXI-1120 rear signal connector
to a data acquisition board with each output connected to a different data acquisition board
channel. See Chapter 2, Configuration and Installation, for more information. For information
on how to make the voltage measurement with your data acquisition board, consult your data
acquisition board user manual. Remember to account for the gains of both the SCXI-1120 and
the data acquisition board when calculating the actual voltage present at the input of the
SCXI-1120.
To measure one of the eight differential input channels to the SCXI-1120, or the DTEMP line if
the module has been configured appropriately, perform the following steps:
1. Write the binary pattern 00XXX000 XX000000 to the SCXI-1120 Configuration Register.
Notice that this can be the RESET state.
2. Measure the voltage with the data acquisition board.
Multiplexed Output
To perform a direct multiplexed output measurement, you must cable the SCXI-1120 rear signal
connector to a data acquisition board. See Chapter 2, Configuration and Installation, for more
information. For information on how to make the voltage measurement with your data
acquisition board, consult your data acquisition board user manual. Remember to account for the
gains of both the SCXI-1120 and the data acquisition board when calculating the actual voltage
present at the input of the SCXI-1120.
To measure one of the eight differential input channels to the SCXI-1120, perform the following
steps:
1. Write the binary pattern 00XXXXCC XX000000 to the SCXI-1120 Configuration Register.
2. Measure the voltage with the data acquisition board.
To measure the voltage on the MTEMP line, perform the following steps:
1. Write the binary pattern 00XXXXXX XX100000 to the SCXI-1120 Configuration Register.
2. Measure the voltage with the data acquisition board.
Indirect Measurements
Indirect measurements involve one module sending a signal to Analog Bus 0, where it is picked
up by another module and transmitted to the data acquisition board.
Measurements from Other Modules
To perform measurements from other modules, you must cable the SCXI-1120 rear signal
connector to a data acquisition board. See Chapter 2, Configuration and Installation, for more
information. To make a measurement from another module, perform the following steps:
1. Perform any necessary programming to ensure that no modules are driving Analog Bus 0.
For an SCXI-1120, clearing AB0EN in the Configuration Register will ensure that its output
is not driving AB0.
2. Write the binary pattern 00XXXXXX XX100011 to the SCXI-1120 Configuration Register.
This step disables the SCXI-1120 from driving Analog Bus 0 and allows Analog Bus 0 to
drive MCH0 through the output buffer.
3. Program the other module to drive Analog Bus 0 with the signal to be measured.
4. Measure the voltage with the data acquisition board.
Measurements from the SCXI-1120 via Another Module
To perform measurements via another module, you must cable the other module rear signal
connector to a data acquisition board. The other module must be able to transfer Analog Bus 0 to
the data acquisition board. See Chapter 2, Configuration and Installation, for more information.
To measure one of the eight differential input channels to the SCXI-1120, perform the following
steps:
1. Perform any necessary programming to ensure that no modules are driving Analog Bus 0.
For an SCXI-1120, clearing AB0EN in the Configuration Register will ensure that its output
is not driving AB0.
2. Program the other module not to drive Analog Bus 0, but to send Analog Bus 0 to the data
acquisition board.
3. Write the binary pattern 00XXXCCC XX000010 to the SCXI-1120 Configuration Register.
4. Measure the voltage with the data acquisition board.
To measure the voltage on the MTEMP line, perform the following steps:
1. Perform any necessary programming to ensure that no modules are driving Analog Bus 0.
For an SCXI-1120, clearing AB0EN in the Configuration Register will ensure that its output
is not driving AB0.
2. Program the other module not to drive Analog Bus 0, but to send Analog Bus 0 to the data
acquisition board.
3. Write the binary pattern 00XXXXXX XX100010 to the SCXI-1120 Configuration Register.
4. Measure the voltage with the data acquisition board.
Scanning Measurements
Programming for scanned data acquisition involves programming your data acquisition board,
modules, and Slot 0. In general, the steps to be taken are as follows:
1. Perform all data acquisition board programming to the point of enabling the data
acquisition.
2. Perform all module programming.
3. Program the Slot 0 hardscan circuitry.
4. Enable the data acquisition, trigger it either through software or hardware, and service the
data acquisition.
The MIO boards can do all types of scanning. Lab-NB, Lab-PC, Lab-PC+, Lab-LC, and
PC-LPM-16 boards support only single-module parallel scanning, and do not support any of the
multiplexed scanning modes. Notice that single-module parallel scanning is typically done
without any module or Slot 0 programming; only programming the data acquisition board is
necessary.
The programming steps for your data acquisition board are given in your data acquisition board
user manual. You should follow the instructions in the following sections:
•AT-MIO-16 User Manual
-Multiple A/D Conversions with Continuous Channel Scanning (Round Robin)
-Multiple A/D Conversions with Interval Channel Scanning (Pseudosimultaneous)
•AT-MIO-16D User Manual
-Multiple A/D Conversions with Continuous Channel Scanning (Round Robin)
-Multiple A/D Conversions with Interval Channel Scanning (Pseudosimultaneous)
•AT-MIO-16F-5 User Manual
-Posttrigger Data Acquisition with Continuous Channel Scanning
-Posttrigger Data Acquisition with Interval Channel Scanning
•AT-MIO-16X User Manual
-Continuous Channel Scanning Data Acquisition
-Interval Channel Scanning Data Acquisition
•AT-MIO-64F-5 User Manual
-Continuous Channel Scanning Data Acquisition
-Interval Channel Scanning Data Acquisition
•Lab-LC User Manual
-Programming Multiple A/D Conversions with Channel Scanning
•Lab-NB User Manual
-Programming Multiple A/D Conversions with Channel Scanning
•Lab-PC User Manual
-Programming Multiple A/D Conversions with Channel Scanning
•Lab-PC+ User Manual
-Programming Multiple A/D Conversions with Channel Scanning
-Programming Multiple A/D Conversions with Interval Scanning
-Programming Multiple A/D Conversions in Single-Channel Interval Acquisition Mode
-Multiple A/D Conversions with Continuous Channel Scanning (Round Robin)
-Multiple A/D Conversions with Interval Channel Scanning (Pseudosimultaneous)
•NB-MIO-16 User Manual
-Programming Multiple A/D Conversions with Channel Scanning
•NB-MIO-16X User Manual
-Multiple A/D Conversions with Continuous Channel Scanning (Round Robin)
-Multiple A/D Conversions with Interval Channel Scanning (Pseudosimultaneous)
•PC-LPM-16 User Manual
-Programming Multiple A/D Conversions with Channel Scanning
Follow the instructions in these sections through the part labeled as follows:
• Clear the A/D Circuitry and Reset the Mux Counter in the MIO board user manual (except for
the AT-MIO-16X and AT-MIO-64F-5). Do not continue to the part called Enable theScanning Data Acquisition Operation. You will do this after you have programmed the
modules and Slot 0.
• Program the Sample Counter (if you are doing continuous channel scanning) or Program the
Scan-Interval Counter (if you are doing interval channel scanning) in the AT-MIO-16X and
AT-MIO-64F-5 user manuals. Do not continue to the part labeled Enable a Scanning DataAcquisition Operation or Enable an Interval Scanning Data Acquisition Operation. You will
do this after you have programmed the modules and Slot 0.
Note: For multiplexed scanning with an MIO board, it is important that you follow the
instructions in the channel scanning sections, not the single-channel sections. Although
you may be using only one MIO board channel, the channel scanning programming will
ensure that the MIO board outputs SCANCLK, which is needed by the SCXI-1120 and
Slot 0.
• Clear the A/D Circuitry in the Lab-LC User Manual. Do not continue to the part called
Program the Sample-Interval Counter. You will do this after you have programmed the
modules and Slot 0.
• Clear the A/D Circuitry in the Lab-PCUser Manual, the Lab PC+ User Manual, and the
PC-LPM-16 User Manual. Do not continue to the part called Start and Service the Data
Acquisition Operation. You will do this after you have programmed the modules and Slot 0.
• Clear the A/D Circuitry in the Lab-NB User Manual. Do not continue to the part called
Program the Sample-Interval Counter (Counter A0). You will do this after you have
Counter 1 and SCANDIV
All MIO boards can operate their data acquisition board scan lists in two ways–they can acquire
one sample per data acquisition board scan list entry; or they can acquire N samples per data
acquisition board scan list entry, where N is a number from 2 to 65,535 that is programmed in
Counter 1. This second method of operation is especially useful when the data acquisition board
scan list length is limited to 16 entries, as it is on all MIO boards except the AT-MIO-16F-5,
which can have up to 512 entries. Because you can multiplex many SCXI-1120s in one chassis
to one MIO board channel, often the simplest way to program the MIO board is to use only one
data acquisition board scan list entry, and make N the total number of samples to be taken on all
modules in one scan. Check your MIO board user manual for limitations in the data acquisition
board scan list format.
To program the MIO board to take N samples per data acquisition board scan list entry, perform
the following additional programming steps at the end of the Enable the Scanning DataAcquisition Operation section in the appropriate data acquisition board user manual:
1. Write FF01 to the Am9513 Command Register to select Counter 1 Mode Register.
2. Write 0325 (hex) to the Am9513 Data Register to store Counter 1 Mode Value for most MIO
boards. For the AT-MIO-16F-5, and the AT-MIO-16X, and the AT-MIO-64F-5, write 1325
(hex).
3. Write FF09 to the Am9513 Command Register to select Counter 1 Load Register.
4. Write the number of samples to be taken per scan list entry (2 to 65,535) to the Am9513
Data Register to load Counter 1.
5. Write FF41 to the Am9513 Command Register to load Counter 1.
6. Write FFF1 to the Am9513 Command Register to step Counter 1.
7. Write FF21 to the Am9513 Command Register to arm Counter 1.
8. Set the SCANDIV bit in Command Register 1.
2. Module Programming
This section describes the programming steps for various scanning possibilities.
Single-Module Parallel Scanning
To perform single-module parallel scanning, you must cable the SCXI-1120 rear signal
connector to a data acquisition board with each output connected to a different data acquisition
board channel. See Chapter 2, Configuration and Installation, for more information.
To program the SCXI-1120 for single-module parallel scanning, write the binary pattern
00XXXX00 XX000000 to the SCXI-1120 Configuration Register. Notice that this can be the
RESET state.
Single-Module Multiplexed Scanning (Direct)
To perform simple channel scanning, you must cable the SCXI-1120 to a data acquisition board.
See Chapter 2, Configuration and Installation, for more information.
To program the module for scanned-channel measurements, write the binary pattern 10XXXCCC
XX001101 to the SCXI-1120 Configuration Register. CCC represents the starting channel
number.
Single-Module Multiplexed Scanning (Indirect)
To indirectly scan a module, send the output of the scanned module onto Analog Bus 0, where it
is picked up by another module and transmitted to the data acquisition board.
Channel Scanning from Other Modules. To scan measurements from other modules, you must
cable the SCXI-1120 to a data acquisition board. See Chapter 2, Configuration and Installation,
for more information. The module programming steps are as follows:
1. Perform any necessary programming to ensure that no modules are driving Analog Bus 0.
For an SCXI-1120, clearing AB0EN in the Configuration Register will ensure that its output
is not driving AB0.
2. Write the binary pattern 10XXXXXX XX100011 to the SCXI-1120 Configuration Register.
This step disables the SCXI-1120 from driving Analog Bus 0 and allows Analog Bus 0 to
drive MCH0 through the output buffer.
3. Program the other module to be scanned.
Channel Scanning from the SCXI-1120 via Another Module. To scan the SCXI-1120 via other
modules, you must cable the other module to a data acquisition board, and the other module must
be able to transfer Analog Bus 0 to the data acquisition board. The other module must also be
able to send a SCANCLK*-compatible signal on TRIG0. See Chapter 2, Configuration andInstallation, for more information. The module programming steps are as follows:
1. Perform any necessary programming to ensure that no modules are driving Analog Bus 0.
For an SCXI-1120, clearing AB0EN in the Configuration Register will ensure that its output
is not driving AB0.
2. Program the other module not to drive Analog Bus 0, but to send Analog Bus 0 to the data
acquisition board. Also program the other module to send a SCANCLK*-compatible signal
to TRIG0.
3. Write the binary pattern 01XXXCCC XX001111 to the SCXI-1120 Configuration Register,
where CCC is the starting channel number.
Multiple-Module Multiplexed Scanning
To scan multiple modules, you must connect one module to the data acquisition board, and the
module must be able to transfer Analog Bus 0 to the data acquisition board. This module must
also be able to send a SCANCLK*-compatible signal on TRIG0. See Chapter 2, Configuration
and Installation, for more information. The module programming steps are as follows:
1. Perform any necessary programming to ensure that no modules are driving Analog Bus 0.
For an SCXI-1120, clearing AB0EN in the Configuration Register will ensure that its output
is not driving AB0.
2. Program the module that is connected to the data acquisition board to connect Analog Bus 0
to the data acquisition board but not drive Analog Bus 0 unless it is receiving an active low
signal on SCANCON. Also program the module to send a SCANCLK*-compatible signal
onto TRIG0. If this module is an SCXI-1120, this programming is accomplished by writing
the binary pattern 10XXXCCC XX001111 to its Configuration Register.
Note: If this module is an SCXI-1120 and is not going to be scanned (it is just being used as
an interface), write a 0 to bit 2 (SCANCONEN) in the Configuration Register. The
start channel bits become don't care bits.
3. Program the other modules to be used in the scan to connect their outputs to Analog Bus 0
but not drive Analog Bus 0 unless receiving an active low signal on SCANCON. Also
program the modules to use TRIG0 as their clock source. For SCXI-1120 modules, this
programming is accomplished by writing the binary pattern 01XXXCCC XX001111 to
their Configuration Registers.
Multiple-Chassis Scanning
To scan modules on multiple chassis, you must use the SCXI-1001 chassis. The cable from the
data acquisition board must bus the digital lines to one module on each chassis. Additionally, the
cable must provide each chassis with its own analog channel. The data acquisition board must be
able to take several readings at a time on a given channel before accessing a new channel. See
the Counter 1 and SCANDIV subsection of the 1. Data Acquisition Board Setup Programming
section earlier in this chapter. You can use the MIO-16 boards, in conjunction with the
SCXI-1350 multichassis adapter, for multichassis scanning.
For each chassis, program the modules according to the appropriate mode of operation,
disregarding the fact that other chassis will be involved.
For example, you want to scan thirteen modules. Twelve modules are in one chassis, and the
thirteenth is in the second chassis and is to be scanned through a fourteenth module that is cabled
to the data acquisition board but is not involved in the scan. Program the twelve modules in the
first chassis according to the steps in the previous Multiple-Module Multiplexed Scanning
section, and program the thirteenth and fourteenth modules according to Channel Scanning fromthe SCXI-1120 via Another Module earlier in this chapter.
3. Programming the Slot 0 Hardscan Circuitry
The following section describes how to program the Slot 0 circuitry for scanning operations. For
a more detailed description of the Slot 0 scanning circuitry, consult the SCXI-1000/1001 UserManual. Descriptions of the Slot 0 registers are in the Slot 0 Register Descriptions section of
Chapter 4, Register Descriptions. It is not necessary to read this section if you are performing
single-module parallel scanning.
To program the hardscan circuitry, perform the following steps:
1. Write binary 0000 0000 to the HSCR.
2. Write binary 0000 1000 to the HSCR.
3. Write the Slot 0 scan list to the FIFO.
4. Write binary 0010 1100 to the HSCR.
5. Write binary 101S 1100 to the HSCR.
6. Write binary 101S 1110 to the HSCR.
7. Write binary 101S 1111 to the HSCR.
To program the hardscan circuitry to use the current scan list, perform the following steps:
1. Write binary 0000 1000 to the HSCR.
2. Write binary 0100 1000 to the HSCR.
3. Write binary 0000 1000 to the HSCR.
4. Write binary 0010 1100 to the HSCR.
5. Write binary 101S 1100 to the HSCR.
6. Write binary 101S 1110 to the HSCR.
7. Write binary 101S 1111 to the HSCR.
In the preceding steps:
S = 0 if you want the scanning to repeat when the end of the list is reached.
S = 1 if you want the circuitry to shut down after a single scan.
When you are writing multiple entries to the same register, for example, repetitive writes to the
HSCR or several FIFO entries, it is important that SS*13 or SS*14 go inactive (high) between
each entry. Select another slot or toggle the SLOT0SEL* line to temporarily deassert the
appropriate SS* line.
If consecutive scan list entries access an SCXI-1120, the module will reload the
MUXCOUNTER with the starting channel after each entry. Thus, two entries with counts of
four for one module will yield different behavior than one entry with a count of eight.
For multiple-chassis scanning, program each Slot 0 with dummy entries to fill the sample counts
when the data acquisition board is accessing other chassis. Use Slot 13 as the dummy entry slot.
At this point, you should now continue from where you left off in the 1. Data Acquisition Board
Setup Programming section of this chapter. Perform the following steps given in your data
acquisition board user manual.
•MIO board user manual
-Enable the scanning data acquisition operation.
-Apply a trigger.
-Service the data acquisition operation.
•Lab-PC User Manual, Lab-PC+ User Manual, and PC-LPM-16 User Manual
-Start and service the data acquisition operation.
•Lab-LC User Manual
-Program the sample-interval counter.
-Service the data acquisition operation.
•Lab-NB User Manual
-Program the sample-interval counter (Counter A0).
-Service the data acquisition operation.
Scanning Examples
The following examples are intended to aid your understanding of module and Slot 0
programming. It will be helpful to refer to the bit descriptions for the Configuration Register and
the FIFO Register in Chapter 4, Register Descriptions.
Example 1
You want to scan, in Multiplexed mode, channels 1 through 4 on an SCXI-1120 in Slot 1 of an
SCXI-1000 chassis. The SCXI-1120 is directly cabled to a data acquisition board.
The programming steps are as follows:
1. Program your data acquisition board as described in the 1. Data Acquisition Board SetupProgramming section of this chapter.
2. Following the procedure given in the Register Writes section, write 10000100 00001101
to the Configuration Register of the SCXI-1120 in Slot 1.
3. Follow the steps outlined in the section earlier in this chapter, 3. Programming the Slot 0
Hardscan Circuitry, where step 3, Write the Slot 0 scan list to the FIFO, consists of the
following:
Write 00000000 00000011 to the FIFO Register. This corresponds to Slot 1 for four
samples.
4. Follow the procedure given in the 4. Acquisition Enable, Triggering, and Servicing section
earlier in this chapter.
Example 2
An SCXI-1000 chassis has four SCXI-1120 modules in Slots 1, 2, 3, and 4. The SCXI-1120 in
Slot 4 is cabled to the data acquisition board. You want to scan channels 3 through 7 on the
SCXI-1120 in Slot 1, channels 0 through 6 on the SCXI-1120 in Slot 4, and channels 7 through 3
on the SCXI-1120 in Slot 3.
The programming steps are as follows:
1. Program your data acquisition board as described in the 1. Data Acquisition Board Setup Programming section.
2. Following the procedure given in the Register Writes section, write 00000000 00000000
to the Configuration Register of the SCXI-1120 in Slot 2. This step resets the module,
including the clearing of the AB0EN bit (bit 0). Notice that a complete reset of this module
is not necessary, but is used for simplicity.
3. Following the procedure given in the Register Writes section, write 10XXX000 00001111
to the Configuration Register of the SCXI-1120 in Slot 4.
4. Following the procedure given in the Register Writes section, write 01XXX011 00001111
to the Configuration Register of the SCXI-1120 in Slot 1.
5. Following the procedure given in the Register Writes section, write 01XXX111 00001111
to the Configuration Register of the SCXI-1120 in Slot 3. Notice that after Channel 7, the
SCXI-1120 will wrap around to Channel 0.
6. Follow the steps given in the section earlier in this chapter, 3. Programming the Slot 0Hardscan Circuitry, where step 3, Write the Slot 0 scan list to the FIFO, consists of the
following:
a. Write 00000000 00000100 to the FIFO Register. This corresponds to Slot 1 for five
samples.
b. Write 00000001 10000110 to the FIFO Register. This corresponds to Slot 4 for
seven samples.
c. Write 00000001 00000100 to the FIFO Register. This corresponds to Slot 3 for five
samples.
Make sure to toggle SLOT0SEL* or reselect the FIFO Register from scratch between steps
6a, 6b, and 6c.
7. Follow the procedure given in the 4. Acquisition Enable, Triggering, and Servicing section
earlier in this chapter.
Example 3
You want to scan five channels on an SCXI-1120 in Slot 4 of Chassis 1, then seven channels of
an SCXI-1120 in Slot 11 of Chassis 2, three channels of an SCXI-1120 in Slot 3 in Chassis 3,
and one channel of an SCXI-1120 in Slot 8 of Chassis 3.
Assuming that the modules are correctly cabled and programmed, the Slot 0 scan lists should be
as follows:
Chassis 1Chassis 2Chassis 3
EntrySlot
Number
CountEntrySlot
Number
CountEntrySlot
Number
Count
1 4 51135 11312
213112117 2 3 3
3134 3 8 1
Other solutions are possible.
In the section earlier in this chapter, 3. Programming the Slot 0 Hardscan Circuitry, step 3, Write
the Slot 0 scan list to the FIFO, consists of the following steps:
This appendix lists the specifications for the SCXI-1120. These are typical at 25° C unless
otherwise stated. The operating temperature range is 0° to 50° C.
Analog Input
Gain (jumper-selectable)1, 2, 5, 10, 20, 50, 100, 200, 500, 1,000, 2,000
Output range±5 V
Number of channels8
Gain accuracy0.15% of full scale
Offset voltage
1-2AOGNDAnalog Output Ground – Connected to the analog
reference when jumper W46 is in position AB-R0.
3-18MCH0± through MCH7± Analog Output Channels 0 through 7 – Connects to
the data acquisition board differential analog
inputchannels.
19OUTREFOutput Reference – Serves as the reference node for
the analog output channels and the temperature
sensor–in the DTS mode–in the Pseudodifferential
Reference mode. It should be connected to the
analog input sense of the NRSE data acquisition
board.
24, 33DIG GNDDigital Ground – Supply the reference for data
acquisition board digital signals and are tied to the
module digital ground.
25SERDATINSerial Data In – Taps into the SCXIbus MOSI line
to provide serial input data to a module or Slot 0.
26SERDATOUTSerial Data Out – Taps into the SCXIbus MISO line
to accept serial output data from a module.
27DAQD*/AData Acquisition Board Data/Address Line – Taps
into the SCXIbus D*/A line to indicate to the
module whether the incoming serial stream is data
or address information.
29SLOT0SEL*Slot 0 Select – Taps into the SCXIbus INTR* line
to indicate whether the information on MOSI is
being sent to a module or Slot 0.
36SCANCLKScan Clock – Indicates to the SCXI-1120 that a
sample has been taken by the data acquisition board
and causes the SCXI-1120 to change channels.
37SERCLKSerial Clock – This signal taps into the SCXIbus
SPICLK line to clock the data on the MOSI and
MISO lines.
43RSVDReserved.
All other pins are not connected.
See the Timing Requirements and Communication Protocol section in Chapter 2, Configuration
and Installation, for more detailed information on timing. Detailed signal specifications are also
This appendix describes how to use and install the hardware accessories for the SCXI-1120:
•SCXI-1340 cable assembly
•SCXI-1341 Lab-NB, Lab-PC, and Lab-PC+ cable assembly
•SCXI-1342 PC-LPM-16 cable assembly
•SCXI-1344 Lab-LC cable assembly
•SCXI-1180 feedthrough panel
•SCXI-1302 50-pin terminal block
•SCXI-1351 one-slot cable extender
•SCXI-1350 multichassis adapter
•SCXI-1343 screw terminal adapter
SCXI-1340 Cable Assembly
The SCXI-1340 cable assembly connects an MIO-16 board to an SCXI-1120 module. The
SCXI-1340 consists of a 50-conductor ribbon cable that has mounting bracket at one end and a
50-pin female connector at the other end. The female connector connects to the I/O connector of
the MIO-16 board. Attached to the mounting bracket is a 50-pin female mounting-bracket
connector that connects to the module rear signal connector. To extend the signals of the
MIO-16 board to an SCXI-1180 feedthrough panel or an SCXI-1181 breadboard module, you
can use the male breakout connector that is near the mounting bracket on the ribbon cable. All
50 pins from the MIO-16 board go straight through to the rear signal connector. You can use a
standard 50-pin ribbon cable instead of the SCXI-1340 cable assembly. The SCXI-1340 has the
following advantages over the ribbon cable:
•The SCXI-1340 has strain relief so that you cannot accidentally disconnect the cable.
•The SCXI-1340 includes a mounting bracket that mounts to the chassis so that you can
remove and reinsert the module without explicitly removing the cable from the back of the
chassis. This is especially useful when the SCXI chassis is rack mounted, making rear access
difficult.
•The SCXI-1340 has an extra male breakout connector for use with the SCXI-1180
feedthrough panel or additional modules or breadboards that need a direct connection to the
MIO-16 board.
•The SCXI-1340 rear panel gives the module and the chassis both mechanical and electrical