National Instruments PXIe-5665 Calibration Manual

Page 1
CALIBRATION PROCEDURE
PXIe-5665
3.6 GHz and 14 GHz RF Vector Signal Analyzer
This document contains the verification procedures for the National Instruments PXIe-5665 RF vector signal analyzer (VSA). Refer to calibration solutions.
When not otherwise specified, the procedures in this document refer to both the PXIe-5665
Note PXIe-5665 tuned frequencies greater than 3.6 GHz and procedures with the
preselector enabled for frequencies greater than 3.6 GHz apply only to the PXIe-5665 14 GHz VSA.
NI warrants the PXIe-5665 to meet its published specifications if the individual modules are calibrated and operating within specifications. For more information about RF system calibration, visit
ni.com/manuals and search for Letter of Conformance.

Contents

Software.................................................................................................................................... 2
Documentation.......................................................................................................................... 2
Test Equipment......................................................................................................................... 3
Test Conditions......................................................................................................................... 10
System Options......................................................................................................................... 11
Initial Setup............................................................................................................................... 11
Characterizing the Test System ................................................................................................ 11
Zeroing and Calibrating the Power Sensor ....................................................................... 11
Characterizing Power Splitter Reference Output ............................................................. 11
Characterizing Power Splitter Difference......................................................................... 12
Characterizing RF Source Power (Direct) ........................................................................ 13
Characterizing RF Source Power (with Lowpass Filter).................................................. 15
Characterizing RF Source Power (Combined) for Frequencies ≤700 MHz..................... 17
Characterizing RF Source Power (Combined) for Frequencies >700 MHz..................... 19
Characterizing RF Source Power (Through Splitter) ....................................................... 21
Characterizing RF Source Power (with Splitter and Attenuator) ..................................... 23
Characterizing Spectrum Analyzer Response .................................................................. 25
As-Found and As-Left Limits................................................................................................... 27
Verification ............................................................................................................................... 27
Verifying Reference Accuracy ......................................................................................... 27
Verifying Phase Noise ...................................................................................................... 27
ni.com/calibration for more information about
Page 2
Verifying Frequency Response and Absolute Amplitude Accuracy ................................ 28
Verifying Frequency Response and Absolute Amplitude Accuracy for
Frequencies >10 MHz ........................................................................................... 28
Verifying Average Noise Level ........................................................................................ 35
Verifying Non-Input-Related Spurs (Residual Spurs)......................................................40
Verifying LO-Related Spurs (Sideband Spurs) ................................................................43
Verifying Image Rejection................................................................................................ 48
Verifying Third-Order Intermodulation Distortion ..........................................................53
Frequencies 700 MHz.............................................................................................53
Frequencies >700 MHz............................................................................................. 57
Verifying Second Harmonic Intercept (SHI).................................................................... 62
Verifying Gain Compression ............................................................................................67
Frequencies 700 MHz............................................................................................. 68
Frequencies >700 MHz.............................................................................................73
Verifying LO Output Power ............................................................................................. 79
Reverification............................................................................................................................ 83
Appendix A: Anti-Distortion Test Fixture................................................................................ 83
Appendix B: Power Sensor Calibration Factor Uncertainty .....................................................84
Worldwide Support and Services ............................................................................................. 85

Software

Calibrating the PXIe-5665 requires you to install the following software on the calibration system:
NI-RFSA 2.5 or later
NI Spectral Measurements Toolkit 2.5 or later
You can download all required software from
The software supports programming the calibration procedures in the LabVIEW, C, and LabWindows™/CVI™ application development environments (ADEs). When you install the software, you need to install support only for the ADE that you intend to use.
ni.com/downloads.

Documentation

You might find the following documents helpful as you perform the calibration procedure:
NI 5665 RF Vector Signal Analyzer Getting Started Guide
NI RF Vector Signal Analyzers Help
PXIe-5665 Specifications
The latest versions of these documents are available at ni.com/manuals.
This calibration procedure calibrates the PXIe-5665 as a single system. To calibrate the PXIe-5653 or PXIe-5622 individually, refer to their calibration procedures, available at
ni.com/manuals.
2 | ni.com | PXIe-5665 Calibration Procedure
Page 3

Test Equipment

Table 1 lists the equipment NI recommends for the performance verification procedures. If the recommended equipment is not available, select a substitute using the minimum requirements listed in the table.
PXIe-5665 Calibration Procedure | © National Instruments | 3
Page 4
4 | ni.com | PXIe-5665 Calibration Procedure
Equipment Recommended Models Where Used Minimum Requirements

Table 1. Recommended Equipment for PXIe-5665 Calibration

Power meter Anritsu ML2438A Verifying frequency response,
Verifying absolute amplitude accuracy,
Verifying LO output power
Display resolution: ≤0.01 dB
Settling: ±0.1%
Instrumentation accuracy: <±0.5%
Noise, zero set, and drift: ±0.5% full-scale (lowest range)
Reference power uncertainty: ≤±0.9%
Reference output VSWR: <1.04 : 1
Power sensors (2x) Anritsu SC7400 Verifying frequency response,
Verifying absolute amplitude accuracy,
Verifying LO output power
Power range: -55 dBm to 20 dBm
Frequency range: 10 MHz to 16 GHz
Input VSWR:
10 MHz to 150 MHz .......... 1.17 : 1
150MHz to 2GHz............. ≤1.08 : 1
2 GHz to 12.4 GHz............. 1.16 : 1
12.4 GHz to 18 GHz........... 1.21 : 1
Linearity:
-55 dBm to 20 dBm............ 1.8%
Calibration factor uncertainty: Refer to Appendix B:
Power Sensor Calibration Factor Uncertainty
Page 5
PXIe-5665 Calibration Procedure | © National Instruments | 5
Table 1. Recommended Equipment for PXIe-5665 Calibration (Continued)
Equipment Recommended Models Where Used Minimum Requirements
Signal generator (RF source 1)
Anritsu MG3692C
Options 2A, 3, 4, 15A, and 22
Verifying third-order intermodulation distortion,
Verifying gain compression
Frequency range: 8 MHz to 16 GHz
Leveled power: -115 dBm to 18 dBm
Power accuracy: ±1.5 dB
Harmonics:
0.1 MHz to ≤10 MHz ......... <-30 dBc
>10 MHz to ≤100 MHz ...... <-40 dBc
>100 MHz to ≤2.2 GHz ..... <-50 dBc
>2.2 GHz to ≤16 GHz ........ <-30 dBc
Nonharmonic spurious:
0.1 MHz to ≤10 MHz ......... <-30 dBc
>10 MHz to ≤2.2 GHz ....... <-60 dBc
>2.2 GHz to ≤16 GHz ........ <-60 dBc
Output VSWR: <2.0 : 1
Signal generator (RF source 2)
Anritsu MG3692C
Options 2A, 3, 4, 15A, and 22
Verifying frequency response,
Verifying absolute amplitude accuracy,
Verifying LO-related spurs,
Verifying image rejection,
Verifying third-order intermodulation distortion,
Verifying second harmonic intercept,
Verifying gain compression,
Verifying LO output power
Frequency range: 0.1 Hz to 16 GHz
Leveled power: -115 dBm to 18 dBm
Power accuracy: ±1.5 dB
Harmonics:
0.1 MHz to ≤10 MHz ......... <-30 dBc
>10 MHz to ≤100 MHz ...... <-40 dBc
>100 MHz to ≤2.2 GHz ..... <-50 dBc
>2.2 GHz to ≤16 GHz ........ <-30 dBc
Nonharmonic spurious:
0.1 MHz to 10 MHz ........... <-30 dBc
10 MHz to ≤2.2 MH z......... <-60 dBc
>2.2 GHz to ≤16 GHz ........ <-60 dBc
Output VSWR: < 2.0 : 1
Page 6
6 | ni.com | PXIe-5665 Calibration Procedure
Equipment Recommended Models Where Used Minimum Requirements
Spectrum analyzer Rohde & Schwarz FSUP26
Options B60 and B61
Table 1. Recommended Equipment for PXIe-5665 Calibration (Continued)
Verifying frequency response,
Verifying phase noise,
Verifying LO-related spurs,
Verifying LO output power
Frequency range: 10 MHz to 18 GHz
Noise floor: <-152 dBm/Hz
Phase noise measurement using cross-correlation
Frequency counter marker feature
Spectrum analysis capabilities
50 termination (three are included in the PXIe-5665 kit)
NI 778353-01 Verifying average noise level,
Verifying non-input-related (residual) spurs
PXI Express chassis PXIe-1065 or PXIe-1075
PXI Express controller
SMA (m)-to­SMA (m) semi-rigid cables (3x)
*
SMA (m)-to­SMA (m) semi-rigid
*
cable
SMA (m)-to­SMA (m) semi-flexible cable
SMA (m)-to­SMA (m) flexible
*
cable
PXIe-8105 or PXIe-8133
NI 151611A-01
NI 151612A-01
NI 190412B-04
*
NI 151890A-1R25
Page 7
SMA (m)-to­SMA (m) cables (36 in.) (3x)
3.5 mm (m)-to-
3.5 mm (m) adaptor
PXIe-5665 Calibration Procedure | © National Instruments | 7
3.5 mm (f)-to-
3.5 mm (f) adaptor
SMA (m)-to­SMA (f) 10 dB attenuator
Table 1. Recommended Equipment for PXIe-5665 Calibration (Continued)
Equipment Recommended Models Where Used Minimum Requirements
MegaPhase G916-SISI-36
Frequency range: DC to 18 GHz
Insertion loss: 2 dB at 18 GHz
Impedance: 50
VSWR: 1.35 : 1 at 18 GHz
Huber+Suhner 32_PC35-50-0-2/199_NE
Frequency range: DC to 18 GHz
Impedance: 50
Return loss:
DC to 1.5 GHz.................... ≥35 dB
1.5 GHz to 6.0 GHz............ ≥30 dB
6.0 GHz to 18.0 GHz.......... ≥20 dB
Huber+Suhner 31_PC35-50-0-2/199_N
Frequency range: DC to 18 GHz
Impedance: 50
Return loss:
DC to 1.5 GHz.................... ≥35 dB
1.5 GHz to 6.0 GHz............ ≥30 dB
6.0 GHz to 18.0 GHz.......... ≥20 dB
Huber+Suhner 6610_SMA-50-1/199N
Frequency range: DC to 18 GHz
Attenuation: 10 dB (nominal)
Power rating: 2 W average
Impedance: 50
VSWR:
DC to 4 GHz....................... 1.15 : 1
4 GHz to 8 GHz.................. 1.20 : 1
8 GHz to 12.4 GHz............. 1.25 : 1
12.4 GHz to 18 GHz........... 1.35 : 1
Page 8
8 | ni.com | PXIe-5665 Calibration Procedure
Table 1. Recommended Equipment for PXIe-5665 Calibration (Continued)
Equipment Recommended Models Where Used Minimum Requirements
SMA (m)-to­SMA (f) 20 dB attenuator
Huber+Suhner 6620_SMA-50-1/199N
Frequency range: DC to 18 GHz
Attenuation: 20 dB (nominal)
Power rating: 2 W average
Impedance: 50
VSWR:
DC to 4 GHz....................... 1.15 : 1
4 GHz to 8 GHz.................. 1.20 : 1
8 GHz to 12. GHz............... 1.25 : 1
12.4 GHz to 18 GHz........... 1.35 : 1
Power splitter (two-resistor type)
Anti-distortion test fixture
Aeroflex/Weinschel 1593 Frequency range: DC to 18 GHz
Amplitude tracking: <0.25 dB
Phase tracking: <4°
Insertion loss: 8.5 dB (6 dB, nominal)
Power rating: 1 W
Impedance: 50
VSWR:
DC to 18 GHz..................... 1.25 : 1
Equivalent output VSWR:
DC to 18 GHz..................... 1.25 : 1
Connectors: 3.5 mm (f)
NI 166375A-01 Verifying third-order intermodulation
distortion,
Verifying second harmonic intercept,
Verifying gain compression
Refer to Appendix A: Anti-Distortion Test Fixture for specifications.
Note: The anti-distortion test fixture requires an external +12 VDC supply and USB interface. Contact NI for programming details.
Page 9
Table 1. Recommended Equipment for PXIe-5665 Calibration (Continued)
Equipment Recommended Models Where Used Minimum Requirements
Frequency reference source
Symmetricom 8040 rubidium frequency standard
Frequency: 10 MHz
Frequency accuracy: ±1 × 10
-9
Torque wrench Refer to Test Conditions for torque wrench specifications.
*
Included in the PXIe-5665 cable accessory kit.
PXIe-5665 Calibration Procedure | © National Instruments | 9
Page 10

Test Conditions

The following setup and environmental conditions are required to ensure the PXIe-5665 meets published specifications.
Keep cabling as short as possible. Long cables and wires act as antennae, picking up extra noise that can affect measurements.
Verify that all connections, including front panel connections and screws, are secure.
Maintain an ambient temperature of 23 °C ±5 °C.
Keep the relative humidity between 10% and 90%, noncondensing.
Allow a warm-up time of at least 30 minutes after NI-RFSA is loaded and recognizes the PXIe-5665. The warm-up time ensures that the measurement circuitry of the PXIe-5665 is at a stable operating temperature.
Perform self-calibration on the PXIe-5665.
Use a torque wrench appropriate for the type of RF connector that you are using. NI recommends a 0.565 N · m (5 lb · in.) wrench for SMA connectors and a 0.90 N · m (8 lb · in.) wrench for 3.5 mm connectors.
Connect the frequency reference source to the REF IN connector on the back of the PXI Express chassis with a standard BNC (m)-to-BNC (m) cable. This connection replaces the connection from the PXIe-5653 REF OUT (10 MHz) connector to the PXI Express chassis REF IN connector, if present.
Lock all test equipment to the REF OUT signal on the back of the PXI Express chassis. Refer to the NI 5665 Timing Configurations topic in the NI RF Vector Signal Analyzers Help for more information about configuring clocking sources.
Ensure that the PXI/PXI Express chassis fan speed is set to HIGH, that the fan filters, if present, are clean, and that the empty slots contain filler panels. For more information, refer to the Maintain Forced-Air Cooling Note to Users document available at
ni.com/manuals.
10 | ni.com | PXIe-5665 Calibration Procedure
Page 11

System Options

The PXIe-5665 is available with options without certain features. The calibration procedure must be modified to handle the following options:
No preamplifier: Skip steps in the verification procedures that call for the preamplifier to be enabled.
25 MHz instantaneous bandwidth: Verification procedures include any changes necessary to accommodate this option. There are no changes to the PXIe-5665 specifications for this digitizer option.

Initial Setup

Refer to the NI PXIe-5665 RF Vector Signal Analyzer Getting Started Guide for information about how to install the software and hardware and how to configure the device in Measurement & Automation Explorer (MAX).

Characterizing the Test System

You use the measured response of the test system during verification tests.
The power splitter and attenuation response is measured at the RF input frequencies used in the verification tests.
Caution The connectors on the device under test (DUT) and test equipment are
fragile. Perform the steps in these procedures with great care to prevent damaging any DUTs or test equipment.

Zeroing and Calibrating the Power Sensor

1. Connect channel A of the power meter to power sensor A.
2. Connect channel B of the power meter to power sensor B.
3. Zero and calibrate the power sensors using the built-in functions in the power meter.

Characterizing Power Splitter Reference Output

Designate either of the two outputs of the power splitter as the reference output. This output must be consistently used as the reference output throughout the characterization and verification procedures.
PXIe-5665 Calibration Procedure | © National Instruments | 11
Page 12

Characterizing Power Splitter Difference

4
6
1
5
3
2
1. Connect the RF source 2 output to the power splitter input through the SMA (m)-to­SMA (m) cable.
2. Connect power sensor A to the reference output of the power splitter through the SMA (m)-to-SMA (m) cable.
3. Connect power sensor B to the other output of the power splitter through the SMA (m)-to-SMA (m) cable. The completed setup is shown in Figure 1.

Figure 1. Power Splitter Difference Characterization Equipment Setup

1 RF Source 2 2 SMA (m)-to-SMA (m) cable
3 Power Splitter 4 Power Sensor A
5 Power Sensor B 6 Power Meter
4. Set the RF source 2 frequency to 4 GHz.
5. Set the RF source 2 power to 0 dBm.
6. Measure the signal at 4 GHz with both power sensors.
7. Set the power sensors calibration factor for the measurement frequency at 4 GHz.
8. Calculate the Power Splitter at 4 GHz Correction Factor using the following formula:
Power Splitter at 4 GHz Correction Factor = Corrected Power Sensor B Power - Corrected
Power Sensor A Power
12 | ni.com | PXIe-5665 Calibration Procedure
Page 13

Characterizing RF Source Power (Direct)

Note Prior to starting this procedure, zero and calibrate the power sensor according
to the Zeroing and Calibrating the Power Sensor section of this document.
1. Connect the RF source 2 output to the power sensor B input using the SMA (m)-to­SMA (m) cable and 3.5 mm (f)-to-3.5 mm (f) adaptor. The completed equipment setup is shown in Figure 2.

Figure 2. RF Source 2 Output Power Characterization Equipment Setup

5
4
3
1
2
1 RF Source 2 2 SMA (m)-to-SMA (m) Cable
3 3.5 mm (f)-to-3.5 mm (f) Adaptor 4 Power Sensor B
5 Power Meter
2. Set the RF source 2 frequency according to the first row in Table 2 or Table 3 as appropriate.

Table 2. PXIe-5665 3.6 GHz VSA RF Source 2 Characterization Frequencies

Start Frequency Stop Frequency Step Size
100.033325 MHz 2.000033325 GHz 100 MHz
2.099966675 GHz 3.599966675 GHz 100 MHz
9.325033325 GHz 11.225033325 GHz 100 MHz
11.324966675 GHz 12.824966675 GHz 100 MHz
1.325033325 GHz 3.225033325 GHz 100 MHz
3.324966675 GHz 4.824966675 GHz 100 MHz
74.966675 MHz 274.966675 MHz 100 MHz
PXIe-5665 Calibration Procedure | © National Instruments | 13
Page 14
Table 2. PXIe-5665 3.6 GHz VSA RF Source 2 Characterization Frequencies (Continued)
Start Frequency Stop Frequency Step Size
25.033325 MHz 1.625033325 GHz 100 MHz
1.724966675 GHz 3.224966675 GHz 100 MHz
3.3875 GHz
987.5 MHz

Table 3. PXIe-5665 14GHz VSA RF Source 2 Characterization Frequencies

Start Frequency Stop Frequency Step Size
100.033325 MHz 1.975033325 GHz 125 MHz
2.099966675 GHz 3.599966675 GHz 125 MHz
3.724966675 GHz 13.974966675 GHz 125 MHz
13.999966675 GHz
9.325033325 GHz 11.200033325 GHz 125 MHz
11.324966675 GHz 12.824966675 GHz 125 MHz
4.949966675 GHz 15.199966675 GHz 125 MHz
15.224966675 GHz
1.325033325 GHz 3.200033325 GHz 125 MHz
3.324966675 GHz 4.824966675 GHz 125 MHz
3.349966675 GHz 13.599966675 GHz 125 MHz
13.624966675 GHz
24.966675 MHz 274.966675 MHz 125 MHz
1.724966675 GHz 3.224966675 GHz 125 MHz
3.3875 GHz
987.5 MHz
3. Set the RF source 2 power to 0 dBm.
4. Measure the Channel B Power using the appropriate calibration factor for the power sensor frequency. Calculate the RF Source Power Direct using the following equation:
RF Source Power Direct = -Channel B Power
5. Repeat steps 2 to 4 for all remaining frequencies in Table 2 or Table 3.
14 | ni.com | PXIe-5665 Calibration Procedure
Page 15

Characterizing RF Source Power (with Lowpass Filter)

Note Prior to starting this procedure, zero and calibrate the power sensor according
to the Zeroing and Calibrating the Power Sensor section of this document.
1. Connect the RF source 2 output to power sensor B through the anti-distortion test fixture. The completed equipment setup is shown in Figure 3.

Figure 3. RF Source Power (with Lowpass Filter) Characterization Equipment Setup

5
6
1
3
2
1 RF Source 1 2 RF Source 2 3 SMA (m)-to-SMA (m) Cable 4 SMA (m)-to-SMA (m) Cable 5 Anti-Distortion Test Fixture
2. Disable the RF source 1 output.
NI-5665 ANTI-DISTORTION ET
RF
INPUT 1
166375A
RF
INPUT 2
6 SMA (m)-to-SMA (m) Cable 7 3.5 mm (f)-to-3.5 mm (f) Adaptor 8 Power Sensor B
RF
OUTPUT
4
9
9 Power Meter
7
8
PXIe-5665 Calibration Procedure | © National Instruments | 15
Page 16
3. Set the RF source 2 frequency and power according to the first row in Table 4.

Table 4. RF Source 2 Settings

Start Frequency Stop Frequency Step Size Power (dBm)
300 MHz 1.8 GHz 10 MHz 0
1.9 GHz 7.0 GHz 100 MHz 0
300.0333125 MHz 1790.0333125 MHz 10 MHz -30
1.8 GHz -30
4. Configure the anti-distortion test fixture to use the appropriate lowpass filter path according to Table 5.

Table 5. Lowpass Filter Frequency Ranges

Lowpass Filter Path (MHz) Frequency Range
470 1 MHz to 470 MHz
735 >470 MHz to 735 MHz
1,150 >735 MHz to 1.150 GHz
1,800 >1.150 GHz to 1.800 GHz
5. Enable the RF source 2 output.
6. Measure the channel B power using the appropriate calibration factor for the power sensor frequency. Adjust the RF source 2 power until this measured power is within ±0.1 dB of the value listed in Table 4. The final RF source 2 power setting is the RF Source Power LPF for that frequency.
7. Repeat steps 3 to 6 for all the frequencies and powers in Table 4.
16 | ni.com | PXIe-5665 Calibration Procedure
Page 17

Characterizing RF Source Power (Combined) for Frequencies 700 MHz

Note Prior to starting this procedure, zero and calibrate the power sensor according
to the Zeroing and Calibrating the Power Sensor section of this document.
1. Connect the RF source 1 and RF source 2 outputs to power sensor B through the anti-distortion test fixture. The completed equipment setup is shown in Figure 4.
Figure 4. RF Source Power (Combined) Characterization Equipment Setup
(Frequencies 700 MHz)
5
6
NI-5665 ANTI-DISTORTION ET
INPUT 1
166375A
RF
RF
INPUT 2
OUTPUT
RF
1
3
4
9
2
1 RF Source 1 2 RF Source 2 3 SMA (m)-to-SMA (m) Cable 4 SMA (m)-to-SMA (m) Cable 5 Anti-Distortion Test Fixture
6 SMA (m)-to-SMA (m) Cable 7 3.5 mm (f)-to-3.5 mm (f) Adaptor 8 Power Sensor B 9 Power Meter
2. Set the anti-distortion test fixture to the 700 MHz combiner path.
7
8
PXIe-5665 Calibration Procedure | © National Instruments | 17
Page 18
3. Set the RF source 1 and RF source 2 frequency and power values according to the first row in Table 6.

Table 6. RF Source Power (Combined) Test Settings (Frequencies 700 MHz)

Start
Frequency
(MHz)
12.05 92.05 10 +350 kHz -10 -350 kHz -10
100 700 100 +350 kHz -10 -350 kHz -10
12.05 92.05 10 +350 kHz -30 -350 kHz -30
100 700 100 +350 kHz -30 -350 kHz -30
10 100 10 +1 MHz +9.1 0Hz -24
10 100 10 +1 MHz -16 0Hz -54
200 700 100 +1 MHz +9.1 0Hz -24
200 700 100 +1 MHz -13 0Hz -54
10 100 10 +1 MHz +8 0Hz -24
200 700 100 +1 MHz +8 0Hz -24
10 100 10 +1 MHz -18 0Hz -54
200 700 100 +1 MHz -15 0Hz -54
Stop
Frequency
(MHz)
Step Size
(MHz)
RF Source 1 RF Source 2
Frequency
Offset
Power (dBm)
Frequency
Offset
4. Disable the RF source 1 output, and enable the RF source 2 output.
Note When disabled, the RF source 1 output signal should be less than -60 dBm.
Power (dBm)
5. Measure the channel B power using the appropriate sensor calibration factor for the RF source 2 frequency.
6. Repeat step 5, adjusting the RF source 2 power until the channel B reading is within ±0.1 dB of the power listed in Table 6. The RF source 2 power is the RF Source 2 Programmed Power.
7. Disable the RF source 2 output, and enable the RF source 1 output.
Note When disabled, the RF source 2 output signal should be less than - 60 dBm.
8. Measure the channel B power using the appropriate sensor calibration factor for the RF source 1 frequency.
9. Repeat step 8, adjusting the RF source 1 power until the channel B power is within ±0.1 dB of the power listed in Table 6. The RF source 1 power is the RF Source 1 Programmed Power.
10. Repeat steps 3 to 9 for all remaining frequency and power values in Table 6.
18 | ni.com | PXIe-5665 Calibration Procedure
Page 19

Characterizing RF Source Power (Combined) for Frequencies >700 MHz

Note Prior to starting this procedure, zero and calibrate the power sensor according
to the Zeroing and Calibrating the Power Sensor section of this document.
1. Connect the RF source 1 and RF source 2 outputs to power sensor B through the anti-distortion test fixture. The completed equipment setup is shown in Figure 5.
Figure 5. RF Source Power (Combined) Characterization Equipment Setup
(Frequencies >700 MHz)
5
6
NI-5665 ANTI-DISTORTION ET
INPUT 1
166375A
RF
RF
INPUT 2
OUTPUT
RF
1
3
4
9
2
1 RF Source 1 2 RF Source 2 3 SMA (m)-to-SMA (m) cable 4 SMA (m)-to-SMA (m) cable
6 SMA (m)-to-SMA (m) cable 7 3.5 mm (f)-to-3.5 mm (f) Adaptor 8 Power Sensor B 9 Power Meter
5 Anti-Distortion Test Fixture
2. Set the anti-distortion test fixture to the >700 MHz combiner path.
7
8
PXIe-5665 Calibration Procedure | © National Instruments | 19
Page 20
3. Set the RF source 1 and RF source 2 frequency and power values according to the first row in Table 7.
Table 7. PXIe-5665 RF Source Power (Combined) Test Settings (Frequencies
>700 MHz)
RF Source 1 RF Source 2
Start
Frequency
800 MHz 3.500 GHz 100 +350 kHz -10 -350 kHz -10
3598.95 MHz +350 kHz -10 -350 kHz -10
3.6 GHz 13.8 GHz 200 +350 kHz -10 -350 kHz -10
13.99895 GHz +350 kHz -10 -350 kHz -10
800 MHz 3.500 GHz 100 +350 kHz -30 -350 kHz -30
3598.95 MHz +350 kHz -30 -350 kHz -30
3.6 GHz +350 kHz -30 -350 kHz -30
800 MHz 1.700 GHz 100 +1 MHz +9.1 0Hz -24
800 MHz 1.700 GHz 100 +1 MHz +8 0Hz -24
800 MHz 1.700 GHz 100 +1 MHz -13 0Hz -54
800 MHz 1.700 GHz 100 +1 MHz -15 0Hz -54
Stop
Frequency
Step Size
(MHz)
Frequency
Offset
Power (dBm)
Frequency
Offset
Powe r (dBm)
1.800 GHz 14.000 GHz 100 +1 MHz +7.1 0Hz -24
1.800 GHz 14.000 GHz 100 +1 MHz +6 0Hz -24
1.800 GHz 3.600 GHz 100 +1 MHz -16 0Hz -54
1.800 GHz 3.600 GHz 100 +1 MHz -18 0Hz -54
20 | ni.com | PXIe-5665 Calibration Procedure
Page 21
4. Disable the RF source 1 output, and enable the RF source 2 output.
Note When disabled, the RF source 1 output signal should be less than -60 dBm.
5. Measure the channel B power using the appropriate sensor calibration factor for the RF source 2 frequency.
6. Repeat step 5, adjusting the RF source 2 power until the channel B reading is within ±0.1 dB of the power in Table 7. The RF source 2 power is the RF Source 2 Programmed Power.
7. Disable the RF source 2 output, and enable the RF source 1 output.
Note When disabled, the RF source 2 output signal should be less than -60 dBm.
8. Measure the channel B power using the appropriate sensor calibration factor for the RF source 1 frequency.
9. Repeat step 8, adjusting the RF source 1 power until the channel B power is within ±0.1 dB of the power in Table 7. The RF source 1 power is the RF Source 1 Programmed Power.
10. Repeat steps 3 to 9 for all remaining frequency and power values listed in Table 7.

Characterizing RF Source Power (Through Splitter)

Note Prior to starting this procedure, zero and calibrate the power sensor and define
the power splitter reference output using the procedures in the Zeroing and
Calibrating the Power Sensor and Characterizing Power Splitter Reference Output
sections of this document.
1. Connect the RF source 2 output to the power splitter input through the SMA (m)-to­SMA (m) cable.
2. Connect a 50 termination load to the reference output of the power splitter.
PXIe-5665 Calibration Procedure | © National Instruments | 21
Page 22
3. Connect power sensor B to the other output of the power splitter through the
1
3
5 6 7
8
2
4
3.5 mm (m)-to-3.5 mm (m) adaptors and the 3.5 mm (f)-to-3.5 mm (f) adaptor. The completed equipment setup is shown in Figure 6.

Figure 6. RF Source Power (Through Splitter) Characterization Equipment Setup

1 RF Source 2 2 SMA (m)-to-SMA (m) Cable 350Ω Termination Load 4 Power Splitter
22 | ni.com | PXIe-5665 Calibration Procedure
5 3.5 mm (m)-to-3.5 mm (m) Adaptor 6 3.5 mm (f)-to-3.5 mm (f) Adaptor 7 Power Sensor B 8Power Meter
Page 23
4. Set the RF source 2 power to -10 dBm.
5. Set the RF source 2 frequency according to the first row in Table 8.

Table 8. RF Source 2 Frequencies

Start Frequency Stop Frequency Step Size
100 MHz 3.5 GHz 200 MHz
3.6 GHz
3.7 GHz 14 GHz 100 MHz
6. Measure the Channel B Power using the appropriate calibration factor for the RF source 2 frequency. Adjust the RF source 2 power until this measured power is within ±0.1 dB of
-10 dBm.
7. Store the RF source 2 set power as the RF Source Power through Splitter.
8. Repeat steps 5 to 7 for all remaining frequencies in Table 8.

Characterizing RF Source Power (with Splitter and Attenuator)

Note Prior to starting this procedure, zero and calibrate the power sensor and define
the power splitter reference output using the procedures in the Zeroing and
Calibrating the Power Sensor and Characterizing Power Splitter Reference Output
sections of this document.
1. Connect the RF source 2 output to the power splitter input through the SMA (m)-to­SMA (m) cable.
2. Connect power sensor A to the reference output of the power splitter.
PXIe-5665 Calibration Procedure | © National Instruments | 23
Page 24
3. Connect power sensor B to the other output of the power splitter through the 3.5 mm (m)-
3
5
7
6
8
4
1
9
2
to-3.5 mm (m) adaptor, the 3.5 mm (f)-to-3.5 mm (f) adaptor, and the 20 dB attenuator. The completed equipment setup is shown in Figure 7.
Figure 7. RF Source Power (with Splitter and Attenuator) Characterization
Equipment Setup
1 RF Source 2 2 SMA (m)-to-SMA (m) Cable 3 Power Sensor A 4 Power Splitter 5 3.5 mm (m)-to-3.5 mm (m) Adaptor
4. Set the RF source 2 power to 0 dBm.
24 | ni.com | PXIe-5665 Calibration Procedure
6 20 dB Attenuator 7 3.5 mm (f)-to-3.5 mm (f) Adaptor 8 Power Sensor B 9 Power Meter
Page 25
5. Set the RF source 2 frequency according to the first row in Table 9.

Table 9. RF Source 2 Frequencies

Start Frequency Stop Frequency Step Size
10 MHz 100 MHz 10 MHz
200 MHz 600 MHz 100 MHz
612.5 MHz
700 MHz 3.6 GHz 100 MHz
3.8 GHz 14 GHz 200 MHz
6. Measure the Channel A Power and the Channel B Power using each sensor and the appropriate calibration factor for the RF source 2 frequency.
7. Calculate the DUT to Power Meter through Attenuator Path Loss and RF Source to
DUT through Attenuator Path Loss for that frequency using the following formulas:
DUT to Power Meter through Attenuator Path Loss =
Channel A Power - Channel B Power
RF Source to DUT through Attenuator Path Loss = -Channel B Power
8. Repeat steps 5 to 7 for all remaining frequencies in Table 9.

Characterizing Spectrum Analyzer Response

1. Connect the RF source 2 output to the power splitter input through the SMA (m)-to­SMA (m) cable.
2. Connect power sensor A to the power splitter reference output through the SMA (m)-to­SMA (m) cable.
PXIe-5665 Calibration Procedure | © National Instruments | 25
Page 26
3. Connect the spectrum analyzer to the other power splitter output through the SMA (m)-to-SMA (m) cable. The completed equipment setup is shown in Figure 8.

Figure 8. Spectrum Analyzer Response Characterization Equipment Setup

7
8
1
1 RF Source 2 2 SMA (m)-to-SMA (m) cable 3 Power Splitter
4
4 Power Sensor A 5 10 dB Attenuator 6 SMA (m)-to-SMA (m) cable
3 65
2
7 Spectrum Analyzer 8 Power Meter
4. Set the RF source 2 power to 0 dBm.
5. Set the RF source 2 frequency to 4 GHz.
6. Configure the spectrum analyzer according to the following settings:
Resolution bandwidth: 1 kHz
Center frequency: 4 GHz
Frequency span: 0 Hz
Reference level: 0 dBm
7. Measure the power sensor A power.
8. Correct the power sensor A measured power for the measurement frequency of 4 GHz.
9. Correct the power sensor A measured power for the Power Splitter at 4 GHz Correction Factor from the Characterizing Power Splitter Difference equipment characterization section of this document using the following equation:
Corrected Power Sensor Power = Power Sensor A reading + Power Splitter at 4 GHz
Correction Factor
26 | ni.com | PXIe-5665 Calibration Procedure
Page 27
10. Measure the spectrum analyzer peak power.
11. Calculate the Spectrum Analyzer Correction Factor using the following formula:
Spectrum Analyzer Correction Factor =
Corrected Power Sensor Power - Spectrum Analyzer Power

As-Found and As-Left Limits

The as-found limits are the published specifications for the PXIe-5665. NI uses these limits to determine whether the PXIe-5665 meets the device specifications when it is received for calibration.
The as-left limits are equal to the published NI specifications for the PXIe-5665, less guard bands for manufacturing test measurement uncertainty, temperature drift, and drift over time. NI uses these limits to determine whether the PXIe-5665 meets the device specifications over its calibration interval.

Verification

The performance verification procedures assume that adequate traceable uncertainties are available for the calibration references.
In the event of a failure during the verification of the PXIe-5665, perform a calibration of the individual modules. Return the PXIe-5603 or PXIe-5605 module to NI for calibration and adjustment, if needed.

Verifying Reference Accuracy

The PXIe-5665 reference accuracy specification is derived from the PXIe-5653 reference accuracy. For more information about PXIe-5653 calibration values, refer to the NI PXIe-5653 Calibration Procedure available at
ni.com/manuals.

Verifying Phase Noise

The phase noise for the PXIe-5665 system is determined by the phase noise of the PXIe-5653 module. To verify the PXIe-5653 phase noise specification, you must use the phase noise verification test for the PXIe-5653. For more information about calibration and verification of the PXIe-5653, refer to the NI PXIe-5653 Calibration Procedure, available at
ni.com/manuals.
PXIe-5665 Calibration Procedure | © National Instruments | 27
Page 28
Verifying Frequency Response and Absolute Amplitude
OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD SENSITIVE
OUT
ALL PORTS
50 Ω
ACCESS
ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p NOM
+ 20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
ESD SENSITIVE
ACCESS
ACTIVE
OUT
IN
LO3
LO2
LO1
ALL PORTS
50 Ω
NI PXIe-5603
Downconverter 20 Hz - 3.6 GHz
RF IN
0 V DC, DC COUPLED
25 V DC MAX, AC COUPLED
+20 dBm MAX > 10 MHz
0 dBm MAX < 10 MHz
+
15 dBm MAX 25 V DC MAX
+
15 dBm MAX
0 V DC
IF OUT
+
22 dBm MAX
0 V DC
K
K
J
J
J
J
J
J
1
765
4
2
3
Accuracy

Verifying Frequency Response and Absolute Amplitude Accuracy for Frequencies >10 MHz

Note Prior to starting this procedure, zero and calibrate the power sensor according
to the Zeroing and Calibrating the Power Sensor section of this document.
1. Connect the RF source 2 to the power splitter input using the SMA (m)-to-SMA (m) cable.
2. Connect the power splitter reference output to power sensor A.
3. Connect the other power splitter output to the PXIe-5665 RF IN connector using the
3.5 mm (m)-to-3.5 mm (m) adaptor and 20 dB attenuator. The completed equipment setup is shown in Figure 9 and Figure 10.
Figure 9. PXIe-5665 3.6 GHz VSA Frequency Response and Absolute Amplitude Accuracy
Verification (Frequencies >10 MHz) Equipment Setup
1 RF Source 2 2 Power Meter 3 SMA (m)-to-SMA (m) cable 4 Power Sensor A
28 | ni.com | PXIe-5665 Calibration Procedure
5 Power Splitter 6 3.5 mm (m)-to-3.5 mm (m) Adaptor 7 20 dB Attenuator
Page 29
Figure 10. PXIe-5665 14 GHz VSA Frequency Response and Absolute Amplitude
OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD SENSITIVE
OUT
ALL PORTS
50 Ω
NI PXIe-5605
Downconverter 20 Hz - 14 GHz
ACCESS ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p
NOM
+
20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
IF OUT
+ 22 dBm MAX
0 V DC
LO1
LO2
LO3
IN
ESD SENSITIVE
OUT
LO3
LO2
LO1
ALL PORTS
50 Ω
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
ACCESS
ACTIVE
RF IN
0 V DC +20 dBm MAX > 10 MHz +10 dBm MAX < 10 MHz
K
K
J
J
J
J
J
J
1
4
2
3
7
6
5
Accuracy Verification (Frequencies >10 MHz) Equipment Setup
1 RF Source 2 2 Power Meter 3 SMA (m)-to-SMA (m) cable 4 Power Sensor A
5 Power Splitter 6 3.5 mm (m)-to-3.5 mm (m) Adaptor 7 20 dB Attenuator
4. Create a new session for the PXIe-5665.
5. Configure the PXIe-5665 according to the following fixed property settings. These settings remain unchanged during the test.
Acquisition Type: Spectrum
Averaging Mode: RMS Averaging
Number of Averages: 10
Digital IF Equalization Enabled: TRUE
Digitizer Dither Enabled: Enabled
Ref Clock Source: PXI_Clk
Channel Coupling: AC Coupled
Span: 100 kHz
Resolution Bandwidth: 10 kHz
6. Configure the PXIe-5665 according to the following variable property settings:
Preamp Enabled: Disabled
Device Instantaneous Bandwidth: 300 kHz
(PXIe-5665 14 GHz VSA) Preselector Enabled: Disabled
PXIe-5665 Calibration Procedure | © National Instruments | 29
Page 30
7. Set the PXIe-5665 reference level using the following values:
Start power: -50 dBm
Stop power: -10 dBm
8. Set the PXIe-5665 center frequency and the RF source 2 frequency according to the first row in Table 10.
Table 10. Frequency Response and Absolute Amplitude Accuracy Verification
(Frequencies >10 MHz) Test Frequencies
Start Frequency Stop Frequency Step Size
*
10 MHz
100 MHz 30 MHz
200 MHz 600 MHz 200 MHz
612.5 MHz
800 MHz 3.6 GHz 400 MHz
4 GHz 14 GHz 500 MHz
*
The 10 MHz to 100 MHz frequency range applied only to the 300 kHz device instantaneous
bandwidth value.
9. Set the RF source 2 amplitude for that RF source 2 frequency using the following equation:
RF Source 2 Amplitude = PXIe-5665 Reference Level +
RF Source to DUT through Attenuator Path Loss
10. Commit the PXIe-5665 settings to hardware.
11. Take five readings of the RF source 2 amplitude using power sensor A. Calculate the average power.
12. Repeat step 11 until any two readings measured differ no more than 0.02 dB and no two averages differ by more than 0.005 dB. This average is the Measured Average Power.
Note More than 30 averages total indicate a test failure that you must diagnose.
13. Calculate the Corrected Input Power using the following formula:
Corrected Input Power = Measured Average Power -
DUT to Power Meter through Attenuator Path Loss
Note The RF Source to DUT through Attenuator Path and DUT to Power Meter
through Attenuator Path Loss values were measured in the Characterizing RF
Source Power (with Splitter and Attenuator) equipment characterization procedure.
Use the values that correspond to the frequency range you are testing.
14. Read the PXIe-5665 power spectrum. The PXIe-5665 Power is the peak value of that spectrum.
30 | ni.com | PXIe-5665 Calibration Procedure
Page 31
15. Calculate the Absolute Amplitude Accuracy at each RF frequency using the following formula:
Absolute Amplitude Accuracy = PXIe-5665 Power - Corrected Input Power
16. Repeat steps 8 to 15 for all remaining frequencies in Table 10.
17. Calculate the Frequency Response at each RF frequency using the following formula:
Frequency Response = Absolute Amplitude Accuracy -
Absolute Amplitude Accuracy at 612.5 MHz
Note Calculate Frequency Response only for PXIe-5665 center frequencies less
than or equal to 3.6 GHz.
18. Repeat steps 9 to 17 for all reference levels listed in step 7.
19. Repeat steps 8 to 18 for an instantaneous bandwidth of 50 MHz (25 MHz for the PXIe-5665 with 25 MHz bandwidth).
20. Repeat steps 8 to 19 with the PXIe-5665 preamplifier enabled.
21.
(PXIe-5665 14 GHz VSA) Repeat steps 8 to 19 for test frequencies greater than 3.6 GHz
with the PXIe-5665 preselector enabled and preamplifier disabled.
Note The PXIe-5665 preselector has a maximum instantaneous bandwidth of
47 MHz. For steps that call for a 50 MHz instantaneous bandwidth, use a 40 MHz instantaneous bandwidth when using the preselector.
PXIe-5665 Calibration Procedure | © National Instruments | 31
Page 32
22. Compare the Frequency Response to the verification test limits in Table 11 or Table 12 as appropriate.

Table 11. Frequency Response Verification Test Limits (Preamplifier Disabled)

Frequency Device
As-Found Limit
(dB)
As-Left Limit
*
(dB)
10 MHz to 100 MHz PXIe-5665 ±0.60 ±0.35
>100 MHz to 1.7 GHz ±0.35 ±0.25
>1.7 GHz to 2.8 GHz PXIe-5665
±0.40 ±0.30
3.6 GHz VSA
PXIe-5665
±0.42
14 GHz VSA
>2.8 GHz to 3.6 GHz PXIe-5665
±0.45 ±0.35
3.6 GHz VSA
PXIe-5665
±0.62
14 GHz VSA
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
Table 12. Frequency Response Verification Test Limits
(Preamplifier Enabled)
Frequency Device
10 MHz to
PXIe-5665 ±0.75 ±0.50
As-Found Limit
(dB)
As-Left Limit
*
(dB)
100 MHz
>100 MHz to
±0.45 ±0.30
2.8 GHz
>2.8 GHz to
3.6 GHz
PXIe-5665
3.6 GHz VSA
PXIe-5665
±0.45 ±0.30
±0.50 ±0.45
14 GHz VSA
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
32 | ni.com | PXIe-5665 Calibration Procedure
Page 33
23. Compare the Absolute Amplitude Accuracy to the verification test limits in Table 13, Table 14, or Table 15 as appropriate.
Table 13. Absolute Amplitude Accuracy Verification Test Limits
(Preamplifier Disabled, PXIe-5665 14 GHz VSA Preselector Disabled)
Frequency Device
612.5 MHz PXIe-5665
3.6 GHz VSA
PXIe-5665
14 GHz VSA
10 MHz to
100 MHz
PXIe-5665
3.6 GHz VSA
PXIe-5665
14 GHz VSA
>100 MHz to
1.7 GHz
PXIe-5665
3.6 GHz VSA
PXIe-5665
14 GHz VSA
>1.7 GHz to
2.8 GHz
PXIe-5665
3.6 GHz VSA
PXIe-5665
14 GHz VSA
>2.8 GHz to
3.6 GHz
PXIe-5665
3.6 GHz VSA
PXIe-5665
14 GHz VSA
As-Found Limit
(dB)
±0.35 ±0.25
±0.46 ±0.38
±0.95 ±0.60
±1.06 ±0.73
±0.70 ±0.50
±0.81 ±0.63
±0.75 ±0.55
±0.88 ±0.68
±0.80 ±0.60
±1.08 ±0.73
As-Left Limit
(dB)
*
>3.6 GHz to
7.5 GHz
>7.5 GHz to
PXIe-5665
14 GHz VSA
±0.70 ±0.50
±0.80 ±0.50
8.5 GHz
>8.5 GHz to
±1.25 ±0.75
14 GHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
PXIe-5665 Calibration Procedure | © National Instruments | 33
Page 34

Table 14. Absolute Amplitude Accuracy Verification Test Limits (Preamplifier Enabled)

Frequency Device
612.5 MHz PXIe-5665
As-Found Limit
(dB)
±0.35 ±0.30
As-Left Limit
(dB)
*
3.6 GHz VSA
PXIe-5665
±0.70 ±0.60
14 GHz VSA
10 MHz to
100 MHz
PXIe-5665
3.6 GHz VSA
PXIe-5665
±1.10 ±0.80
±1.45
14 GHz VSA
>100 MHz to
2.8 GHz
PXIe-5665
3.6 GHz VSA
PXIe-5665
±0.80 ±0.60
±1.15
14 GHz VSA
>2.8 GHz to
3.6 GHz
PXIe-5665
3.6 GHz VSA
PXIe-5665
±0.80 ±0.60
±1.20 ±0.75
14 GHz VSA
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
Table 15. PXIe-5665 14 GHz VSA Absolute Amplitude Accuracy Verification Test Limits
(Preamplifier Disabled, PXIe-5665 14 GHz VSA Preselector Enabled)
Frequency As-Found Limit (dB) As-Left Limit (dB)
*
>3.6 GHz to 7.5 GHz ±4.0 ±1.5
>7.5 GHz to 8.5 GHz ±4.0 ±1.5
>8.5 GHz to 14 GHz ±4.0 ±1.5
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
24. Close the PXIe-5665 session.
If the frequency response and absolute amplitude accuracy verification procedures determines that the PXIe-5665 is outside its limits, refer to Worldwide Support and Services for information about support resources or service requests.
34 | ni.com | PXIe-5665 Calibration Procedure
Page 35

Verifying Average Noise Level

1. Connect a 50 termination to the PXIe-5665 RF IN connector. The completed equipment setup is shown in Figure 11 and Figure 12.

Figure 11. PXIe-5665 3.6 GHz VSA Average Noise Level Verification Equipment Setup

1
150Ω termination
NI PXIe-562 2
16-Bit IF Digitizer
ACCESS ACTIVE
IF IN
K
+
20 dBm MAX
50 Ω
PFI 1
CLK IN
6.3 Vp-p
CLK OUT
ESD SENSITIVE
ACCESS
IF OUT
+
22 dBm MAX
0 V DC
LO3
TTL
MAX
50 Ω
LO2
2 Vp-p NOM
50 Ω
LO1
NI PXIe-5603
Downconverter 20 Hz - 3.6 GHz
ACTIVE
RF IN
0 V DC, DC COUPLED
25 V DC MAX, AC COUPLED
+20 dBm MAX > 10 MHz
0 dBm MAX < 10 MHz
K
OUT
+
+
15 dBm MAX
0 V DC
15 dBm MAX
25 V DC MAX
ESD SENSITIVE
ALL PORTS
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
IN
J
J
J
50 Ω
800 MHz
3.2 GHz - 8.3 GHz
OUT
LO3
J
LO2
4 GHz
J
LO1
J
OUT
ESD SENSITIVE
ALL PORTS
50 Ω
PXIe-5665 Calibration Procedure | © National Instruments | 35
Page 36
OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD SENSITIVE
OUT
ALL PORTS
50 Ω
NI PXIe-5605
Downconverter 20 Hz - 14 GHz
ACCESS
ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p
NOM
+
20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-562 2
16-Bit IF Digitizer
IF IN
IF OUT
+
22 dBm MAX
0 V DC
LO1
LO2
LO3
IN
ESD SENSITIVE
OUT
LO3
LO2
LO1
ALL PORTS
50 Ω
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
ACCESS
ACTIVE
RF IN
0 V DC +20 dBm MAX > 10 MHz +10 dBm MAX < 10 MHz
K
K
1
J
J
J
J
J
J

Figure 12. PXIe-5665 14 GHz VSA Average Noise Level Verification Equipment Setup

150Ω termination
2. Create a new session for the PXIe-5665.
3. Configure the PXIe-5665 according to the following property settings:
Acquisition Type: Spectrum
Averaging Mode: RMS Averages
Channel Coupling: AC Coupled
Digital IF Equalization Enabled: TRUE
Digitizer Dither Enabled: Enabled
Number of Averages: 20
Ref Clock Source: PXI_Clk
Power Spectrum Units: Volts Squared
Preamp Enabled: Disabled
Resolution Bandwidth Type: ENBW
Reference Level: -50 dBm
Resolution Bandwidth: 300 Hz
RF Attenuation: 0 dB
FFT Window Type: Blackman-Harris
36 | ni.com | PXIe-5665 Calibration Procedure
(PXIe-5665 14 GHz VSA) Preselector Enabled: Disabled
Page 37
4. Set the PXIe-5665 center frequency according to Table 16.

Table 16. Average Noise Level Verification Test Frequencies

Start Frequency Stop Frequency Step Size (MHz)
10 MHz 90 MHz 20
100 MHz 3.6 GHz 500
4 GHz 14 GHz 500
5. Set the PXIe-5665 channel coupling, RF attenuation, device instantaneous bandwidth, span, resolution bandwidth, sampling ratio, and number of averages according to Table 17.

Table 17. PXIe-5665 Settings for Average Noise Level Verification

Test Frequency
Device Instantaneous
Bandwidth
Span
OSP Sampling
Ratio
10 MHz to <100 MHz 300 kHz 100 kHz 32
100 MHz to ≤3.6 GHz 300 kHz 100 kHz 32
*
50 MHz
3.8 GHz to 14 GHz 50 MHz
*
Set the PXIe-5665 Device Instantaneous Bandwidth to 25 MHz for the PXIe-5665 with
*
5MHz 1
5 MHz 1
25 MHz bandwidth.
6. Commit the PXIe-5665 settings to hardware.
7. Read the power spectrum from the PXIe-5665. Convert the power spectrum to a power spectral density (dBm/Hz) value.
8. Remove five points from around the center of the power spectral density from step 7.
9. Convert the power spectral density with the points removed to watts/Hz, take the mean, and then convert the result back to dBm/Hz. This value is the PXIe-5665 Average Noise (dBm/Hz).
10. Repeat steps 4 to 9 for all frequencies in Table 16.
11. Repeat steps 4 to 10 for frequencies less than or equal to 3.6 GHz with the PXIe-5665 preamplifier enabled.
12.
(PXIe-5665 14 GHz VSA) Repeat steps 4 to 10 for frequencies greater than 3.6 GHz with
the PXIe-5665 preselector enabled and preamplifier disabled.
Note The PXIe-5665 preselector has a maximum instantaneous bandwidth of
47 MHz. For steps that call for a 50 MHz instantaneous bandwidth, use a 40 MHz instantaneous bandwidth when using the preselector.
PXIe-5665 Calibration Procedure | © National Instruments | 37
Page 38
13. Compare the PXIe-5665 Average Noise (dBm/Hz) to the verification test limits in Table 18, Table 19, or Table 20 as appropriate.
Table 18. Average Noise Verification Test Limits (Preamplifier Disabled, PXIe-5665
14 GHz VSA Preselector Disabled)
*
Frequency
As-Found Limit
(dBm/Hz)
As-Left Limit
(dBm/Hz)
>10 MHz to 100 MHz -149 -151
>100 MHz to 300 MHz -152 -154
>300 MHz to 1.7 GHz -151 -153
>1.7 GHz to 2.8 GHz -149 -151
>2.8 GHz to 3.6 GHz -148 -150
>3.6 GHz to 7.5 GHz
>7.5 GHz to 8.5 GHz
>8.5 GHz to 12 GHz
>12 GHz to 14 GHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
-148 -150
-146 -147
-147 -148
-145 -146
about as-left limits.
Frequencies >3.6 GHz apply only to the PXIe-5665 14 GHz VSA.

Table 19. Average Noise Verification Test Limits (Preamplifier Enabled)

*
Frequency
As-Found Limit
(dBm/Hz)
As-Left Limit
(dBm/Hz)
>10 MHz to 100 MHz -161 -163
>100 MHz to 300 MHz -162 -165
>300 MHz to 1.7 GHz -162 -164
>1.7 GHz to 2.8 GHz -161 -163
>2.8 GHz to 3.6 GHz -160 -163
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
38 | ni.com | PXIe-5665 Calibration Procedure
Page 39
Table 20. Average Noise Verification Test Limits (Preamplifier Disabled, PXIe-5665
14 GHz VSA Preselector Enabled)
*
Frequency
As-Found Limit
(dBm/Hz)
As-Left Limit
(dBm/Hz)
>3.6 GHz to 7.5 GHz -144 -145
>7.5 GHz to 8.5 GHz -140 -141
>8.5 GHz to 12 GHz -141 -142
>12 GHz to 14 GHz -140 -141
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
14. Close the PXIe-5665 session.
If the average noise level verification procedure determines that the PXIe-5665 is outside its limits, refer to Worldwide Support and Services for information about support resources or service requests.
PXIe-5665 Calibration Procedure | © National Instruments | 39
Page 40

Verifying Non-Input-Related Spurs (Residual Spurs)

OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD SENSITIVE
OUT
ALL PORTS
50 Ω
ACCESS ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p
NOM
+
20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-562 2
16-Bit IF Digitizer
IF IN
ESD SENSITIVE
ACCESS
ACTIVE
OUT
IN
LO3
LO2
LO1
ALL PORTS
50 Ω
NI PXIe-5603
Downconverter 20 Hz - 3.6 GHz
RF IN
0 V DC, DC COUPLED
25 V DC MAX, AC COUPLED
+20 dBm MAX > 10 MHz
0 dBm MAX < 10 MHz
+
15 dBm MAX 25 V DC MAX
+
15 dBm MAX
0 V DC
IF OUT
+
22 dBm MAX
0 V DC
K
K
J
J
J
J
J
J
1
1. Connect a 50 termination to the PXIe-5665 RF IN connector. The completed equipment setup is shown in Figure 13 and Figure 14.
Figure 13. PXIe-5665 3.6 GHz VSA Non-Input-Related Spurs Verification
Equipment Setup
150Ω Termination
40 | ni.com | PXIe-5665 Calibration Procedure
Page 41
OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD SENSITIVE
OUT
ALL PORTS
50 Ω
NI PXIe-5605
Downconverter 20 Hz - 14 GHz
ACCESS
ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p
NOM
+
20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-562 2
16-Bit IF Digitizer
IF IN
IF OUT
+
22 dBm MAX
0 V DC
LO1
LO2
LO3
IN
ESD SENSITIVE
OUT
LO3
LO2
LO1
ALL PORTS
50 Ω
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
ACCESS
ACTIVE
RF IN
0 V DC +20 dBm MAX > 10 MHz +10 dBm MAX < 10 MHz
K
K
J
J
J
J
J
J
1
Figure 14. PXIe-5665 14 GHz VSA Non-Input-Related Spurs Verification
Equipment Setup
150Ω Termination
2. Create a new session for the PXIe-5665.
3. Configure the PXIe-5665 according to the following property settings:
Acquisition Type: Spectrum
Averaging Mode: RMS Averaging
Number of Averages: 4
Ref Clock Source: PXI_Clk
Ref Clock Rate: 10 MHz
PXIe-5665 Calibration Procedure | © National Instruments | 41
Channel Coupling: AC Coupled
Preamp Enabled: Disabled
Span: 50 MHz (25 MHz for the PXIe-5665 with 25 MHz bandwidth)
Device Instantaneous Bandwidth: 50 MHz (25 MHz for the PXIe-5665 with 25 MHz bandwidth)
Resolution Bandwidth: 2 kHz
RF Attenuation: 0 dB
Reference Level: -60 dBm
FFT Window Type: Flat Top
(PXIe-5665 14 GHz VSA) Preselector Enabled: Disabled
Page 42
4. Set the PXIe-5665 center frequency using the following values:
Start frequency: 100 MHz
Stop frequency:
(PXIe-5665 3.6 GHz VSA) 3.6 GHz
(PXIe-5665 14 GHz VSA) 14 GHz
Step size: 9.997331 MHz
5. Commit the PXIe-5665 settings to hardware.
6. Read the power spectrum from the PXIe-5665.
7. Measure the highest power in the spectrum returned from the PXIe-5665. This value is the PXIe-5665 Non-Input-Related Spurious Level.
8. Repeat steps 4 to 7 for all frequencies in step 4.
9. Compare the PXIe-5665 Non-Input-Related Spurious Level to the verification test limits in Table 21.

Table 21. Non-Input-Related Spurs Verification Test Limits

Center Frequency As-Found Limit (dBm) As-Left Limit (dBm)
*
100 MHz to <1.650 GHz -95 -96
1.650 GHz to 1.750 GHz -85 -88
>1.750 GHz to 3.6 GHz -95 -96
>3.6 GHz to 7.5 GHz
>7.5 GHz to 8.5 GHz
>8.5 GHz to 14 GHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
-92 -94
-90 -92
-90 -92
about as-left limits.
Center frequencies >3.6 GHz apply only to the PXIe-5665 14 GHz VSA.
10. Close the PXIe-5665 session.
If the non-input-related spurs (residual spurs) verification procedure determines that the PXIe-5665 is outside its limits, refer to Worldwide Support and Services for information about support resources or service requests.
42 | ni.com | PXIe-5665 Calibration Procedure
Page 43

Verifying LO-Related Spurs (Sideband Spurs)

1. Connect RF source 2 to the power splitter input through the SMA (m)-to-SMA (m) cable.
2. Connect the power splitter reference output to the spectrum analyzer through the SMA (m)-to-SMA (m) cable.
3. Connect the other power splitter output to the PXIe-5665 RF IN connector using the
3.5 mm (m)-to-3.5 mm (m) adaptor. The completed equipment setup is shown in Figure 15 and Figure 16.

Figure 15. PXIe-5665 3.6 GHz VSA LO-Related Spurs Verification Equipment Setup

2
4
5
6
1
1 RF Source 2 2 Spectrum Analyzer 3 SMA (m)-to-SMA (m) cable
IF IN
50 Ω
PFI 1
K
ACCESS
ACTIVE
+
TTL
6.3 Vp-p MAX
50 Ω
2 Vp-p
NOM
50 Ω
NI PXIe-5603
Downconverter 20 Hz - 3.6 GHz
ACTIVE
0 V DC, DC COUPLED
25 V DC MAX, AC COUPLED
+20 dBm MAX > 10 MHz
0 dBm MAX < 10 MHz
K
IF OUT
22 dBm MAX
0 V DC
OUT
LO3
LO2
LO1
+
15 dBm MAX
0 V DC
+
ESD SENSITIVE
ALL PORTS
15 dBm MAX 25 V DC MAX
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
IN
J
J
J
50 Ω
RF IN
NI PXIe-5622
16-Bit IF Digitizer
ACCESS
+
20 dBm MAX
3
CLK OUT
ESD SENSITIVE
CLK IN
4 SMA (m)-to-SMA (m) cable 5 Power Splitter 6 3.5 mm (m)-to-3.5 mm (m) Adaptor
NI PXIe-5653
Synthesizer
3.2 GHz - 8.3 GHz
OUT
LO3
800 MHz
J
LO2
4 GHz
J
LO1
J
OUT
ESD SENSITIVE
ALL PORTS
50 Ω
PXIe-5665 Calibration Procedure | © National Instruments | 43
Page 44

Figure 16. PXIe-5665 14 GHz VSA LO-Related Spurs Verification Equipment Setup

K
K
J
J
J
J
J
J
2
4
5
6
1
1 RF Source 2 2 Spectrum Analyzer 3 SMA (m)-to-SMA (m) cable
4. Create a new session for the PXIe-5665.
NI PXIe-5622
16-Bit IF Digitizer
ACCESS ACTIVE
ACCESS
IF IN
+ 20 dBm MAX
50 Ω
+
PFI 1
TTL
CLK IN
6.3 Vp-p
3
4 SMA (m)-to-SMA (m) cable 5 Power Splitter 6 3.5 mm (m)-to-3.5 mm (m) Adaptor
CLK OUT
MAX
50 Ω
2 Vp-p NOM
50 Ω
ESD SENSITIVE
NI PXIe-5605
Downconverter 20 Hz - 14 GHz
ACTIVE
RF IN
0 V DC +20 dBm MAX > 10 MHz +10 dBm MAX < 10 MHz
IF OUT
22 dBm MAX
0 V DC
OUT
LO3
LO2
LO1
+
15 dBm MAX
0 V DC
ESD SENSITIVE
ALL PORTS
50 Ω
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
IN
LO3
LO2
LO1
+
15 dBm MAX
25 V DC MAX
3.2 GHz - 8.3 GHz
OUT
LO3
800 MHz
LO2
4 GHz
LO1
OUT
ESD SENSITIVE
ALL PORTS
50 Ω
44 | ni.com | PXIe-5665 Calibration Procedure
Page 45
5. Configure the PXIe-5665 according to the following fixed property settings. These settings remain unchanged during the test.
Acquisition Type: Spectrum
Averaging Mode: RMS Averaging
Number of Averages: 10
Ref Clock Source: PXI_Clk
Ref Clock Rate: 10 MHz
Preamp Enabled: Disabled
Reference Level: -10 dBm
FFT Window Type: Flat Top
Resolution Bandwidth Type: Bin Width
Channel Coupling: AC Coupled
Device Instantaneous Bandwidth: 50 MHz (25 MHz for the PXIe-5665 with the
25 MHz bandwidth)
(PXIe-5665 14 GHz VSA) Preselector Enabled: Disabled
6. Set the RF source 2 mode to Single Frequency.
7. Connect the frequency reference source to the spectrum analyzer reference frequency input.
8. Set the spectrum analyzer reference source to External.
Note If the RF source you use for this test has consistent, known offset spurs, these
spurs can be characterized independently of the LO-related spurs. You can ignore these offset frequencies to save time during testing.
9. Set the PXIe-5665 center frequency according to the first row in either Table 22 or Table 23.

Table 22. PXIe-5665 3.6 GHz VSA Test Frequencies

Start Frequency Stop Frequency Step Size
100 MHz 3.5 GHz 200 MHz
3.6 GHz

Table 23. PXIe-5665 14 GHz VSA Test Frequencies

Start Frequency Stop Frequency Step Size
100 MHz 13.9 GHz 200 MHz
14 GHz
PXIe-5665 Calibration Procedure | © National Instruments | 45
Page 46
10. Configure the RF Source 2 according to the following settings:
Center frequency: PXIe-5665 center frequency
Output power: RF Source Power through Splitter for that frequency
11. Set the spectrum analyzer center frequency to the RF source 2 frequency.
12. Set the PXIe-5665 span and resolution bandwidth according to the first row in Table 24.

Table 24. LO-Related Spurs Verification Bandwidth and Span Settings

PXIe-5665
Span
200 kHz 10 Hz 20 kHz (1999) 600 Hz 50 Hz 6
2MHz 40 Hz 200 kHz (4999) 300 Hz 50 Hz 3
20 MHz 60 Hz 2 MHz (33333) 300 Hz 50 Hz 3
PXIe-5665
RBW
PXIe-5665 Notch BW
(Number of Bins at
Stated RBW)
Spectrum
Analyzer
Span
Spectrum
Analyzer
RBW
Spectrum
Analyzer
Averages
13. Set the spectrum analyzer span, resolution bandwidth, and number of averages according to the values in Table 24.
14. Commit the PXIe-5665 settings to hardware.
15. Wait 250 ms before making the first measurement and wait 100 ms before making a subsequent measurements to allow the PXIe-5665 and RF source 2 amplitudes to settle.
Note Settling times are a characteristic of the RF source device. Refer to the Test
Equipment section of this document for a list of NI-recommended source devices and
device specifications. If you use an RF source device with different device specifications, your settling times may differ from those listed in this procedure.
16. Measure the RF source 2 spectrum with the spectrum analyzer.
17. Perform a peak search on the spectrum from step 16. The value of the peak is the Spectrum Analyzer Tone Power. This value is used in step 19(c).
46 | ni.com | PXIe-5665 Calibration Procedure
Page 47
18. Measure the spectrum on the PXIe-5665.
a. Find the peak value of the spectrum.
b. Confirm that the peak value is between -16 dBm and -4 dBm. The peak value is the
PXIe-5665 Tone Power. This value is used in step 18(d).
c. Notch filter the spectrum by replacing the appropriate number of bins from the center
of the spectrum according to Table 24 with bins of -130 dB in value. The resulting spectrum is the Notched Spectrum.
d. Calculate the Notched Spectrum Response using the following equation:
Notched Spectrum Response (dB) = Notched Spectrum (dBm) - PXIe-5665 Tone Power (dBm)
e. Create a list of target spurs from the Notched Spectrum Response (dB) that have a
level greater than -90 dB.
f. If any two spurs are within ±500 Hz of each other, retain only the spur with the greater
power magnitude. This reduced list is used in step 19(a).
Note The RF source 2 indicated for use in this procedure exhibits spurious
responses that have a frequency spread of 500 Hz. Any target spur found within ±500 Hz of an RF source 2 spurious response is considered to be from the RF source 2 because of this frequency spread.
19. Eliminate any RF source 2 spurious responses from the list of target spurs.
a. Tune the spectrum analyzer to the frequency of each target spur in the reduced list
from step 18(f).
b. Measure the spectrum and perform a peak search on the measured spectrum. This
value is the Spectrum Analyzer Peak Value.
c. Convert the Spectrum Analyzer Peak Value units to dB using the following equation:
Spectrum Analyzer Peak (dB) = Spectrum Analyzer Peak Value (dBm) -
Spectrum Analyzer Tone Power
d. If the value of the Spectrum Analyzer Peak (dB) is 5 dB or more below the spur level
in the reduced list from step 18(f), retain the spur level, otherwise discard the value.
20. Compare the spur values from the sorted list in step 19(d) to the verification test limits in Table 25.
Table 25. LO-Related Spurious Responses (Sideband Spurs)
Verification Test Limits
Center Frequency As-Found Limit (dBc) As-Left Limit (dBc)
*
100 MHz to 3.6 GHz -73 -75
3.6 GHz to 14 GHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
-73 -75
about as-left limits.
Center frequencies >3.6 GHz apply only to the PXIe-5665 14 GHz VSA.
PXIe-5665 Calibration Procedure | © National Instruments | 47
Page 48
21. Repeat steps 11 to 19 for all spans in Table 24.
22. Repeat steps 9 to 20 for all center frequencies in Table 22.
(PXIe-5665 14 GHz VSA) Repeat steps 9 to 20 for center frequencies from 3.8 GHz to
23. 14 GHz in 200 MHz steps with the PXIe-5665 preselector enabled.
Note The PXIe-5665 preselector has a maximum instantaneous bandwidth of
47 MHz. For steps that call for a 50 MHz instantaneous bandwidth, use a 40 MHz instantaneous bandwidth when using the preselector.
24. Close the PXIe-5665 session.
If the LO-related spurs (sideband spurs) verification procedure determines that the PXIe-5665 is outside its limits, refer to Worldwide Support and Services for information about support resources or service requests.

Verifying Image Rejection

This procedure verifies the following image responses of the PXIe-5665:
For test frequencies <3.6 GHz:
IF1 Image = Test Frequency + 9.225 GHz
IF2 Image = 3.3875 GHz
IF3 Image = 987.5 MHz
IF2 Image at RF = Test Frequency + 1.225 GHz
IF3 Image at RF = |Test Frequency - 375 MHz | (Evaluated only for images greater than 10 MHz)
For test frequencies >3.6 GHz,
IF1 Image = Test Frequency + 1.225 GHz
IF2 Image = 987.5 GHz
IF2 Image at RF = Test Frequency - 375 MHz
where Test Frequency is the PXIe-5665 center frequency.
48 | ni.com | PXIe-5665 Calibration Procedure
Page 49
1. Connect RF source 2 to the PXIe-5665 RF IN using the SMA (m)-to-SMA (m) cable. The
OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD SENSITIVE
OUT
ALL PORTS
50 Ω
ACCESS
ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p NOM
+
20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
ESD SENSITIVE
ACCESS
ACTIVE
OUT
IN
LO3
LO2
LO1
ALL PORTS
50 Ω
NI PXIe-5603
Downconverter 20 Hz - 3.6 GHz
RF IN
0 V DC, DC COUPLED
25 V DC MAX, AC COUPLED
+20 dBm MAX > 10 MHz
0 dBm MAX < 10 MHz
+
15 dBm MAX 25 V DC MAX
+
15 dBm MAX
0 V DC
IF OUT
+ 22 dBm MAX
0 V DC
K
K
J
J
J
J
J
J
1
2
completed equipment setup is shown in Figure 17 and Figure 18.

Figure 17. PXIe-5665 3.6 GHz VSA Image Rejection Verification Equipment Setup

1 RF Source 2 2 SMA (m)-to-SMA (m) cable
PXIe-5665 Calibration Procedure | © National Instruments | 49
Page 50

Figure 18. PXIe-5665 14 GHz VSA Image Rejection Verification Equipment Setup

OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD SENSITIVE
OUT
ALL PORTS
50 Ω
NI PXIe-5605
Downconverter 20 Hz - 14 GHz
ACCESS
ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p
NOM
+
20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
IF OUT
+
22 dBm MAX
0 V DC
LO1
LO2
LO3
IN
ESD SENSITIVE
OUT
LO3
LO2
LO1
ALL PORTS
50 Ω
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
ACCESS
ACTIVE
RF IN
0 V DC +20 dBm MAX > 10 MHz +10 dBm MAX < 10 MHz
K
K
J
J
J
J
J
J
1
2
1 RF Source 2 2 SMA (m)-to-SMA (m) cable
2. Create a new session for the PXIe-5665.
3. Configure the RF source 2 using the following values:
Mode: Single frequency
Power level: 0 dBm
4. Configure the PXIe-5665 according to following property settings:
Acquisition Type: Spectrum
Averaging Mode: RMS Averaging
Number of Averages: 20
Ref Clock Source: PXI_Clk
Channel Coupling: AC Coupled
Preamp Enabled: Disabled
Span: 20 kHz
50 | ni.com | PXIe-5665 Calibration Procedure
Resolution Bandwidth: 100 Hz
Device Instantaneous Bandwidth: 40 MHz (25 MHz for the PXIe-5665 with 25 MHz bandwidth)
RF Attenuation: 10 dB
Reference Level: 0 dBm
(PXIe-5665 14 GHz VSA) Preselector Enabled: Enabled
Page 51
5. Set the PXIe-5665 center frequency according to Table 26 or Table 27 as appropriate.

Table 26. PXIe-5665 3.6 GHz VSA Image Rejection Verification Test Frequencies

Start Frequency Stop Frequency (GHz) Step Size (MHz)
100.033325 MHz 2.000033325 400
2.399966675 GHz 3.599966675 400

Table 27. PXIe-5665 14 GHz VSA Image Rejection Verification Test Frequencies

Start Frequency Stop Frequency (GHz) Step Size (MHz)
100.033325 MHz 2.000033325 400
2.399966675 GHz 13.999966675 400
6. Commit the PXIe-5665 settings to hardware.
7. For each test frequency from Table 26 or Table 27, calculate PXIe-5665 image frequencies using the following formulas:
For test frequencies <3.6 GHz,
IF1 Image = Test Frequency + 9.225 GHz
IF2 Image = 3.3875 GHz
IF3 Image = 987.5 MHz
IF2 Image at RF = Test Frequency + 1.225 GHz
IF3 Image at RF = | Test Frequency - 375 MHz | (Evaluated only for images greater
than 10 MHz)
For test frequencies > 3.6 GHz,
IF1 Image = Test Frequency + 1.225 GHz
IF2 Image = 987.5 MHz
IF2 Image at RF = Test Frequency - 375 MHz
where Test Frequency is the PXIe-5665 center frequency.
8. For each image and test frequency in the list from step 7 and the PXIe-5665 center frequency from step 5, set the RF source 2 to the frequency with a Power Setting of 0 dB. The RF Source Power Direct for this frequency is used in step 11.
Note RF Source Power Direct value was measured in the Characterizing RF
Source Power (Direct) equipment characterization procedure.
PXIe-5665 Calibration Procedure | © National Instruments | 51
Page 52
9. Wait for 200 ms to allow the PXIe-5665 and RF source 2 amplitudes to settle.
Note Settling times are a characteristic of the RF source device. Refer to the Test
Equipment section of this document for a list of NI-recommended source devices and
device specifications. If you use an RF source device with different device specifications, your settling times may differ from those listed in this procedure.
10. Acquire the spectrum and determine the peak power for each frequency. The peak power value is the Measured Level.
11. For each Measured Level, calculate the Response Correction using the following equation:
Response Correction = Measured Level + RF Source Power Direct
12. Calculate the Image Rejection at each of the image frequencies using the following equation:
Image Rejection = Response Correction (at test frequency) - Response Correction (at each
image frequency)
13. Repeat steps 5 to 12 for all center frequencies in Table 26 or Table 27.
14. Compare the Image Rejection results to the verification test limits in Table 28.

Table 28. Image Rejection Verification Test Limits

Center Frequency As-Found Limit As-Left Limit
*
100 MHz to 2.2 GHz -80 dBc -83 dBc
>2.2 GHz to 3.6 GHz -77 dBc -80 dBc
>3.6 GHz to 14 GHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
-80 dBc -82 dBc
about as-left limits.
Center frequencies >3.6 GHz apply only to the PXIe-5665 14 GHz VSA.
15. Close the PXIe-5665 session.
If the image rejection verification procedure determines that the PXIe-5665 is outside its limits, refer to Worldwide Support and Services for information about support resources or service requests.
52 | ni.com | PXIe-5665 Calibration Procedure
Page 53

Verifying Third-Order Intermodulation Distortion

OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD SENSITIVE
OUT
ALL PORTS
50 Ω
ACCESS ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p NOM
+ 20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
ESD SENSITIVE
ACCESS
ACTIVE
OUT
IN
LO3
LO2
LO1
ALL PORTS
50 Ω
NI PXIe-5603
Downconverter 20 Hz - 3.6 GHz
RF IN
0 V DC, DC COUPLED
25 V DC MAX, AC COUPLED
+20 dBm MAX > 10 MHz
0 dBm MAX < 10 MHz
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
IF OUT
+
22 dBm MAX
0 V DC
J
J
J
J
J
J
K
K
RF
OUTPUT
RF
INPUT 1
NI-5665 ANTI-DISTORTION ET
166375A
RF
INPUT 2
6
1
5
2
4
3

Frequencies 700 MHz

1. Connect RF source 1 and RF source 2 to the PXIe-5665 RF IN connector through the anti-distortion test fixture as shown in Figure 19 and Figure 20.
Figure 19. PXIe-5665 3.6 GHz VSA Third-Order Intermodulation Verification
(Frequencies 700 MHz) Equipment Setup
1 RF Source 1 2 RF Source 2 3 SMA (m)-to-SMA (m) cable
4 SMA (m)-to-SMA (m) cable 5 Anti-Distortion Test Fixture 6 SMA (m)-to-SMA (m) cable
PXIe-5665 Calibration Procedure | © National Instruments | 53
Page 54
Figure 20. PXIe-5665 14 GHz VSA Third-Order Intermodulation Verification
OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD SENSITIVE
OUT
ALL PORTS
50 Ω
NI PXIe-5605
Downconverter 20 Hz - 14 GHz
ACCESS
ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p
NOM
+
20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
IF OUT
+ 22 dBm MAX
0 V DC
LO1
LO2
LO3
IN
ESD SENSITIVE
OUT
LO3
LO2
LO1
ALL PORTS
50 Ω
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
ACCESS
ACTIVE
RF IN
0 V DC +20 dBm MAX > 10 MHz +10 dBm MAX < 10 MHz
K
K
J
J
J
J
J
J
RF
OUTPUT
RF
INPUT 1
NI-5665 ANTI-DISTORTION ET 166375A
RF
INPUT 2
1
5
2
4
3
6
(Frequencies 700 MHz) Equipment Setup
1 RF Source 1 2 RF Source 2 3 SMA (m)-to-SMA (m) cable
4 SMA (m)-to-SMA (m) cable 5 Anti-Distortion Test Fixture 6 SMA (m)-to-SMA (m) cable
2. Create a new session for the PXIe-5665.
3. Configure the PXIe-5665 according to the following property settings:
Acquisition Type: Spectrum
Averaging Mode: RMS Averaging
Number of Averages: 10
Digital IF Equalization Enabled: TRUE
Digitizer Dither Enabled: Enabled
Ref Clock Source: PXI_Clk
Channel Coupling: AC Coupled
RF Attenuation: 0 dB
Device Instantaneous Bandwidth: 300 kHz
IF Output Power Level: 0 dB
Resolution Bandwidth: 750 Hz
Span: 150 kHz
4. Configure the anti-distortion test fixture to use the 700 MHz combiner path.
54 | ni.com | PXIe-5665 Calibration Procedure
Page 55
5. Disable the RF source 1 and RF source 2 outputs.
Note When disabled, the RF source 1 and RF source 2 output signals should be less
than -60 dBm.
6. Set the RF source 1 power to the RF Source 1 Programmed Power
1
corresponding to one
of the following values:
Preamplifier disabled: -10 dBm
Preamplifier enabled: -30 dBm
7. Set the RF source 2 power to the RF Source 2 Programmed Power
2
corresponding to one
of the following values:
Preamplifier disabled: -10 dBm
Preamplifier enabled: -30 dBm
8. Configure the PXIe-5665 reference level to one of the following values:
Preamplifier disabled: -10 dBm
Preamplifier enabled: -30 dBm
9. For each test frequency from Table 29, calculate the following four frequencies:
IMD Low Frequency = Test Frequency - 1.05 MHz
P1 Frequency = Test Frequency + 350 kHz
P2 Frequency = Test Frequency - 350 kHz
IMD High Frequency = Test Frequency + 1.05 MHz
where IMD is the intermodulation distortion.

Table 29. TOI (Frequencies 700 MHz) Verification Test Frequencies

Start Frequency Stop Frequency (MHz) Step Size (MHz)
12.05 MHz 92.05 40
100 MHz 700 300
10. Set the RF source 1 frequency to P1 Frequency.
11. Set the RF source 2 frequency to P2 Frequency.
12. Set the PXIe-5665 center frequency to IMD Low Frequency.
13. Commit the PXIe-5665 settings to hardware.
14. Enable the RF source 1 and the RF source 2 outputs.
1
The RF Source 1 Programmed Power value was measured in the RF Source Power (Combined) for
Frequencies
2
The RF Source 2 Programmed Power value was measured in the RF Source Power (Combined) for
Frequencies
700 MHz equipment characterization procedure.
700 MHz equipment characterization procedure.
PXIe-5665 Calibration Procedure | © National Instruments | 55
Page 56
15. Wait 2.5 s before making the first measurement and wait 100 ms before making subsequent measurements to allow the PXIe-5665, RF source 1, and RF source 2 amplitudes to settle.
Note Settling times are a characteristic of the RF source device. Refer to the Test
Equipment section of this document for a list of NI-recommended source devices and
device specifications. If you use an RF source device with different device specifications, your settling times may differ from those listed in this procedure.
16. Measure the power spectrum with the PXIe-5665.
17. Find the highest peak within the measurement bandwidth. This value is the IMD Low Power.
18. Repeat steps 11 to 16 for the PXIe-5665 center frequencies of P1 Frequency,
P2 Frequency, and IMD High Frequency. The values from step 16 are IMD Low Power, P1 Power, P2 Power, and IMD High Power, respectively.
19. Calculate the upper and lower third-order intercept (TOI) points using the following equations:
TOI Upper = P1 Power + (P2 Power - IMD High Power)/2
TOI Lower = P2 Power + (P1 Power - IMD Low Power)/2
20. The smaller of the TOI Lower and TOI Upper values is the TOI Minimum.
21. Repeat steps 8 to 19 for all test frequencies in Table 29.
22. Repeat steps 5 to 20 with the PXIe-5665 preamplifier enabled.
23. Compare the TOI Minimum to the verification test limits in Table 30 or Table 31 as appropriate.

Table 30. TOI (Frequencies 700 MHz) Verification Test Limits (Preamplifier Disabled)

Center Frequency As-Found Limit (dBm) As-Left Limit (dBm)
10 MHz to <100 MHz +16 +17
100 MHz to 700 MHz +19 +20
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.

Table 31. TOI (Frequencies 700 MHz) Verification Test Limits (Preamplifier Enabled)

Center Frequency As-Found Limit (dBm) As-Left Limit (dBm)
10 MHz to <100 MHz -3 -2
100 MHz to 700 MHz +2 +3
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
56 | ni.com | PXIe-5665 Calibration Procedure
*
*
Page 57
24. Disable the output of the RF source 1 and RF source 2.
OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD SENSITIVE
OUT
ALL PORTS
50 Ω
ACCESS ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p NOM
+ 20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
ESD SENSITIVE
ACCESS
ACTIVE
OUT
IN
LO3
LO2
LO1
ALL PORTS
50 Ω
NI PXIe-5603
Downconverter 20 Hz - 3.6 GHz
RF IN
0 V DC, DC COUPLED
25 V DC MAX, AC COUPLED
+20 dBm MAX > 10 MHz
0 dBm MAX < 10 MHz
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
IF OUT
+
22 dBm MAX
0 V DC
J
J
J
J
J
J
K
K
RF
OUTPUT
RF
INPUT 1
NI-5665 ANTI-DISTORTION ET
166375A
RF
INPUT 2
6
1
5
2
4
3
Note When disabled, the RF source 1 and RF source 2 output signals should be less
than -60 dBm.
25. Close the PXIe-5665 session.
If the third-order intermodulation distortion for frequencies 700 MHz verification procedure determines that the PXIe-5665 is outside its limits, refer to Worldwide Support and Services for information about support resources or service requests.

Frequencies >700 MHz

1. Connect RF source 1 and RF source 2 to the PXIe-5665 RF IN connector through the anti-distortion test fixture as shown in Figure 21 and Figure 22.
Figure 21. PXIe-5665 3.6 GHz VSA Third-Order Intermodulation (Frequencies >700 MHz)
Verification Equipment Setup
1 RF Source 1 2 RF Source 2 3 SMA (m)-to-SMA (m) cable
4 SMA (m)-to-SMA (m) cable 5 Anti-Distortion Test Fixture 6 SMA (m)-to-SMA (m) cable
PXIe-5665 Calibration Procedure | © National Instruments | 57
Page 58
Figure 22. PXIe-5665 14 GHz VSA Third-Order Intermodulation (Frequencies >700 MHz)
OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD SENSITIVE
OUT
ALL PORTS
50 Ω
NI PXIe-5605
Downconverter 20 Hz - 14 GHz
ACCESS
ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p
NOM
+
20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
IF OUT
+ 22 dBm MAX
0 V DC
LO1
LO2
LO3
IN
ESD SENSITIVE
OUT
LO3
LO2
LO1
ALL PORTS
50 Ω
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
ACCESS
ACTIVE
RF IN
0 V DC +20 dBm MAX > 10 MHz +10 dBm MAX < 10 MHz
K
K
J
J
J
J
J
J
RF
OUTPUT
RF
INPUT 1
NI-5665 ANTI-DISTORTION ET 166375A
RF
INPUT 2
1
5
2
4
3
6
Verification Equipment Setup
1 RF Source 1 2 RF Source 2 3 SMA (m)-to-SMA (m) cable
4 SMA (m)-to-SMA (m) cable 5 Anti-Distortion Test Fixture 6 SMA (m)-to-SMA (m) cable
2. Create a new session for the PXIe-5665.
3. Configure the PXIe-5665 according to the following property settings:
Acquisition Type: Spectrum
Averaging Mode: RMS Averaging
Number of Averages: 10
Digital IF Equalization Enabled: TRUE
Ref Clock Source: PXI_Clk
Channel Coupling: AC Coupled
RF Attenuation: 0 dB
Device Instantaneous Bandwidth: 300 kHz
IF Output Power Level: 0 dB
Resolution Bandwidth: 750 Hz
Span: 150 kHz
(PXIe-5665 14 GHz VSA) Preselector Enabled: Disabled
4. Configure the anti-distortion test fixture to use the >700 MHz combiner path.
58 | ni.com | PXIe-5665 Calibration Procedure
Page 59
5. Disable the RF source 1 and RF source 2 outputs.
Note When disabled, the RF source 1 and RF source 2 output signals should be less
than -60 dBm.
6. Set the RF source 1 power to the RF Source 1 Programmed Power
1
corresponding to one
of the following values:
Preamplifier disabled: -10 dBm
Preamplifier enabled: -30 dBm
7. Set the RF source 2 power to the RF Source 2 Programmed Power
2
corresponding to one
of the following values:
Preamplifier disabled: -10 dBm
Preamplifier enabled: -30 dBm
8. Configure the PXIe-5665 Reference Level to one the following values:
Preamplifier disabled: -10 dBm
Preamplifier enabled: -30 dBm
9. For each test frequency from Table 32 or Table 33, calculate the following four frequencies:
IMD Low Frequency = Test Frequency - 1.05 MHz
P1 Frequency = Test Frequency + 350 kHz
P2 Frequency = Test Frequency - 350 kHz
IMD High Frequency = Test Frequency + 1.05 MHz
Table 32. PXIe-5665 3.6 GHz VSA TOI (Frequencies >700 MHz)
Verification Test Frequencies
Start Frequency Stop Frequency Step Size
800 MHz 3.3 GHz 500 MHz
3.59895 GHz
1
The RF Source 1 Programmed Power value was measured in the Characterizing RF Source Power
(Combined) for Frequencies >700 MHz equipment characterization procedure.
2
The RF Source 2 Programmed Power value was measured in the Characterizing RF Source Power
(Combined) for Frequencies >700 MHz equipment characterization procedure.
PXIe-5665 Calibration Procedure | © National Instruments | 59
Page 60
Table 33. PXIe-5665 14 GHz VSA TOI (Frequencies >700 MHz)
Verification Test Frequencies
Stop Frequency
Preamp Enabled Start Frequency
(GHz)
Step Size (MHz)
Enabled 800 MHz 3.3 500
Disabled 800 MHz 3.3 500
3.6 GHz 13.6 500
13.99895 GHz
10. Set the RF source 1 frequency to P1 Frequency.
11. Set the RF source 2 frequency to P2 Frequency.
12. Set the PXIe-5665 center frequency to IMD Low Frequency.
13. Commit the PXIe-5665 settings to hardware.
14. Enable the RF source 1 and RF source 2 outputs.
15. Wait 2.5 s before making the first measurement and wait 100 ms before making subsequent measurements to allow the PXIe-5665, RF source 1, and RF source 1 amplitudes to settle.
Note Settling times are a characteristic of the RF source device. Refer to the Test
Equipment section of this document for a list of NI-recommended source devices and
device specifications. If you use an RF source device with different device specifications, your settling times may differ from those listed in this procedure.
16. Measure the power spectrum with the PXIe-5665.
17. Find the highest peak within the measurement bandwidth. This value is the IMD Low Power.
18. Repeat steps 11 to step 16 for the PXIe-5665 center frequencies of P1 Frequency,
P2 Frequency, and IMD High Frequency. The values from step 16 are IMD Low Power, P1 Power, P2 Power, and IMD High Power, respectively.
19. Calculate the upper and lower TOI points using the following equations:
TOI Upper= P1 Power + (P2 Power - IMD High Power)/2
TOI Lower= P2 Power + (P1 Power - IMD Low Power)/2
The smaller value of the TOI Lower and TOI Upper is the TOI Minimum.
20. Repeat steps 8 to 18 for all test frequencies in Table 32.
21. Repeat steps 5 to 19 with the PXIe-5665 preamplifier enabled for frequencies greater than or equal to 3.6 GHz.
60 | ni.com | PXIe-5665 Calibration Procedure
Page 61
22. Compare the TOI Minimum to the verification test limits in Table 34 or Table 35 as appropriate.
Table 34. TOI (Frequencies >700 MHz) Verification Test Limits (Preamplifier Disabled,
PXIe-5665 14 GHz VSA Preselector Disabled)
Center Frequency As-Found Limit (dBm) As-Left Limit (dBm)
*
>700 MHz to 3.6 GHz +20 +21
>3.6 GHz to 14 GHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
+20 +21
about as-left limits.
Center frequencies >3.6 GHz apply only to the PXIe-5665 14 GHz VSA.

Table 35. TOI (Frequencies >700 MHz) Verification Test Limits (Preamplifier Enabled)

Center Frequency As-Found Limit As-Left Limit
*
>700 MHz to 3.6 GHz +2.5 dBm +3.5 dBm
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
23. Disable the output of the RF source 1 and RF source 2.
Note When disabled, the RF source 1 and RF source 2 output signals should be less
than -60 dBm.
24. Close the PXIe-5665 session.
If the third-order intermodulation distortion for frequencies >700 MHz verification procedure determines that the PXIe-5665 is outside its limits, refer to Worldwide Support and Services for information about support resources or service requests.
PXIe-5665 Calibration Procedure | © National Instruments | 61
Page 62

Verifying Second Harmonic Intercept (SHI)

OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD SENSITIVE
OUT
ALL PORTS
50 Ω
ACCESS ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p NOM
+ 20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
ESD SENSITIVE
ACCESS
ACTIVE
OUT
IN
LO3
LO2
LO1
ALL PORTS
50 Ω
NI PXIe-5603
Downconverter 20 Hz - 3.6 GHz
RF IN
0 V DC, DC COUPLED
25 V DC MAX, AC COUPLED
+20 dBm MAX > 10 MHz
0 dBm MAX < 10 MHz
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
IF OUT
+
22 dBm MAX
0 V DC
J
J
J
J
J
J
K
K
RF
OUTPUT
RF
INPUT 1
NI-5665 ANTI-DISTORTION ET
166375A
RF
INPUT 2
6
1
5
2
4
3
1. Connect RF source 1 and RF source 2 to the PXIe-5665 RF IN connector through the anti-distortion test fixture as shown in Figure 23 and Figure 24.

Figure 23. PXIe-5665 3.6 GHz VSA SHI Verification Equipment Setup

1 RF Source 1 2 RF Source 2 3 SMA (m)-to-SMA (m) Cable
62 | ni.com | PXIe-5665 Calibration Procedure
4 SMA (m)-to-SMA (m) Cable 5 Anti-Distortion Test Fixture 6 SMA (m)-to-SMA (m) Cable
Page 63

Figure 24. PXIe-5665 14 GHz VSA SHI Verification Equipment Setup

OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD SENSITIVE
OUT
ALL PORTS
50 Ω
NI PXIe-5605
Downconverter 20 Hz - 14 GHz
ACCESS
ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p
NOM
+
20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
IF OUT
+ 22 dBm MAX
0 V DC
LO1
LO2
LO3
IN
ESD SENSITIVE
OUT
LO3
LO2
LO1
ALL PORTS
50 Ω
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
ACCESS
ACTIVE
RF IN
0 V DC +20 dBm MAX > 10 MHz +10 dBm MAX < 10 MHz
K
K
J
J
J
J
J
J
RF
OUTPUT
RF
INPUT 1
NI-5665 ANTI-DISTORTION ET 166375A
RF
INPUT 2
1
5
2
4
3
6
1 RF Source 1 2 RF Source 2 3 SMA (m)-to-SMA (m) Cable
4 SMA (m)-to-SMA (m) Cable 5 Anti-Distortion Test Fixture 6 SMA (m)-to-SMA (m) Cable
2. Create a new session for the PXIe-5665.
3. Configure the PXIe-5665 according to the following property settings:
Acquisition Type: Spectrum
Digital IF Equalization Enabled: TRUE
Digitizer Dither Enabled: Enabled
Ref Clock Source: PXI_Clk
Channel Coupling: AC Coupled
Span: 150 kHz
Device Instantaneous Bandwidth: 300 kHz
Resolution Bandwidth: 750 Hz
RF Attenuation: 10 dB
Power Spectrum Units: Volts Squared
(PXIe-5665 14 GHz VSA) Preselector Enabled: Disabled
4. Configure the anti-distortion test fixture to use the 470 MHz Lowpass Filter path.
PXIe-5665 Calibration Procedure | © National Instruments | 63
Page 64
5. Disable the RF source 1 and RF source 2 outputs.
Note When disabled, the RF source 1 and RF source 2 output signals should be less
than -60 dBm.
6. Disable the PXIe-5665 preamplifier.
7. Set the PXIe-5665 reference level to one of the following values:
Preamplifier disabled: 0 dBm
Preamplifier enabled: -30 dBm
8. Set the RF source 2 frequency and the PXIe-5665 center frequency to the values shown in the first row of Table 36.

Table 36. SHI Verification Test Frequencies

Start Frequency Stop Frequency Step Size (MHz)
300 MHz 700 MHz 200
800 MHz 1.8 GHz 500
2 GHz 3.5 GHz 500
3.8 GHz 6.8 GHz 500
7 GHz
9. Configure the anti-distortion test fixture to use the appropriate Lowpass Filter Path as shown in Table 37.

Table 37. Lowpass Filter Specifications

Lowpass Filter Path (MHz) Frequency Range
470 1 MHz to 470 MHz
735 >470 MHz to 735 MHz
1,150 >735 MHz to 1.150 GHz
1,800 >1.150 GHz to 1.800 GHz
2,530 >1.800 GHz to 2.530 GHz
3,550 >2.530 GHz to 3.550 GHz
4,985 >3.550 GHz to 4.985 GHz
7,000 >4.985 GHz to 7.000 GHz
10. Commit the PXIe-5665 settings to hardware.
64 | ni.com | PXIe-5665 Calibration Procedure
Page 65
11. Set the RF source 2 power to the RF Source Power LPF to one of the following values for
Noise Measurement BW
Span
2
------------
10 kHz
2
------------------=
Noise CF Center Frequency
Span
4
------------
10 kHz
4
------------------++=
Average Noise Power Power in Band 10
Noise Measurement BW
Resolution Bandwidth
----------------------------------------------------------


10
log=
the RF source 2 frequency:
Preamplifier disabled: 0 dBm
Preamplifier enabled: -30 dBm
12. Enable the RF source 2 output.
13. Wait 500 ms before making the first measurement and wait 100 ms before making subsequent measurements to allow the PXIe-5665 and RF source 2 amplitudes to settle.
Note Settling times are a characteristic of the RF source device. Refer to the Test
Equipment section of this document for a list of NI-recommended source devices and
device specifications. If you use an RF source device with different device specifications, your settling times may differ from those listed in this procedure.
14. Measure the power spectrum with the PXIe-5665.
15. Find the highest peak within the measurement bandwidth (BW). This value is the Peak Power.
16. Calculate the Noise Measurement BW and the Noise Center Frequency (CF) using the following equations:
17. Calculate the Power in Band for the spectrum from step 13 using the Noise Measurement BW and Noise CF values from step 16.
Note Power in Band is a built-in function in the NI Spectral Measurements Toolkit.
18. Calculate the Average Noise Power using the following equation:
19. Record either Peak Power (from step 14) or Average Noise Power (from step 17). According to the frequency tuned to with the PXIe-5665, record whichever value is greater as the Fundamental Power or the Harmonic Power.
20. Repeat steps 7 to 19 for the second harmonic, which is twice the test frequency.
21. Calculate the SHI using the following equation:
SHI = 2(Fundamental Power) - Harmonic Power - RF Attenuation
22. Repeat steps 7 to 20 for all test frequencies listed in Table 36.
PXIe-5665 Calibration Procedure | © National Instruments | 65
Page 66
23. (PXIe-5665 14 GHz VSA) Repeat steps 6 to 21 for test frequencies between 3.8 GHz and 7 GHz with the PXIe-5665 preselector enabled.
24. Repeat steps 6 to 22 for the frequencies in Table 38 with the PXIe-5665 preamplifier enabled.

Table 38. SHI Verification Test Frequencies

Start Frequency (MHz) Stop Frequency Step Size
300.0333125 1.3000333125 GHz 200 MHz
1,800
25. Compare the SHI to the verification test limits in Table 39, Table 40, or Table 41 as appropriate.
Table 39. SHI Verification Test Limits (Preamplifier Disabled,
PXIe-5665 14 GHz VSA Preselector Enabled)
Source
Frequency
300 MHz to
Device
PXIe-5665 +42 +43
As-Found Limit
(dBm)
As-Left Limit
*
(dBm)
<700 MHz
700 MHz to
1.8 GHz
PXIe-5665
3.6 GHz VSA
PXIe-5665
+50 +51
+44 +45
14 GHz VSA
>1.8 GHz to
3.75 GHz
>3.75 GHz to
4.25 GHz
>4.25 GHz to
7 GHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
PXIe-5665
14 GHz VSA
+54 +56
+54 +56
+54 +56
about as-left limits.
Source frequencies >1.8 GHz apply only to the PXIe-5665 14 GHz VSA.
66 | ni.com | PXIe-5665 Calibration Procedure
Page 67

Table 40. SHI Verification Test Limits (Preamplifier Enabled)

Source
Frequency
300 MHz to
1.8 GHz
Device
PXIe-5665
3.6 GHz VSA
PXIe-5665
As-Found Limit
(dBm)
+15 +18
+20 +25
As-Left Limit
*
(dBm)
14 GHz VSA
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.

Table 41. SHI Verification Test Limits (PXIe-5665 14 GHz VSA Preselector Disabled)

Frequency (GHz) As-Found Limit (dBm) As-Left Limit (dBm)
>1.80 to 4.25 +28 +30
>4.25 to 7.00 +18 +20
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
26. Disable the output of RF source 2.
Note When disabled, the RF source 2 output signal should be less than -60 dBm.
27. Close the PXIe-5665 session.
*
If the second harmonic intercept verification procedure determines that the PXIe-5665 is outside its limits, refer to Worldwide Support and Services for information about support resources or service requests.

Verifying Gain Compression

This procedure verifies that <1 dB of gain compression occurs when the specified power level is applied at the RF IN connector of the PXIe-5665.
Note The power level you apply differs whether you are verifying the as-found or
as-left limit.
PXIe-5665 Calibration Procedure | © National Instruments | 67
Page 68

Frequencies 700 MHz

OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD SENSITIVE
OUT
ALL PORTS
50 Ω
ACCESS ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p NOM
+ 20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
ESD SENSITIVE
ACCESS
ACTIVE
OUT
IN
LO3
LO2
LO1
ALL PORTS
50 Ω
NI PXIe-5603
Downconverter 20 Hz - 3.6 GHz
RF IN
0 V DC, DC COUPLED
25 V DC MAX, AC COUPLED
+20 dBm MAX > 10 MHz
0 dBm MAX < 10 MHz
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
IF OUT
+
22 dBm MAX
0 V DC
J
J
J
J
J
J
K
K
RF
OUTPUT
RF
INPUT 1
NI-5665 ANTI-DISTORTION ET
166375A
RF
INPUT 2
6
1
5
2
4
3
1. Connect RF source 1 and RF source 2 to the PXIe-5665 RF IN connector through the anti-distortion test fixture, as shown in Figure 25 and Figure 26.
Figure 25. PXIe-5665 3.6 GHz VSA Gain Compression Verification
(Frequencies 700 MHz) Equipment Setup
1 RF Source 1 2 RF Source 2 3 SMA (m)-to-SMA (m) Cable
68 | ni.com | PXIe-5665 Calibration Procedure
4 SMA (m)-to-SMA (m) Cable 5 Anti-Distortion Test Fixture 6 SMA (m)-to-SMA (m) Cable
Page 69
Figure 26. PXIe-5665 14 GHz VSA Gain Compression Verification
OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD SENSITIVE
OUT
ALL PORTS
50 Ω
NI PXIe-5605
Downconverter 20 Hz - 14 GHz
ACCESS
ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p
NOM
+
20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
IF OUT
+ 22 dBm MAX
0 V DC
LO1
LO2
LO3
IN
ESD SENSITIVE
OUT
LO3
LO2
LO1
ALL PORTS
50 Ω
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
ACCESS
ACTIVE
RF IN
0 V DC +20 dBm MAX > 10 MHz +10 dBm MAX < 10 MHz
K
K
J
J
J
J
J
J
RF
OUTPUT
RF
INPUT 1
NI-5665 ANTI-DISTORTION ET 166375A
RF
INPUT 2
1
5
2
4
3
6
(Frequencies 700 MHz) Equipment Setup
1 RF Source 1 2 RF Source 2 3 SMA (m)-to-SMA (m) Cable
4 SMA (m)-to-SMA (m) Cable 5 Anti-Distortion Test Fixture 6 SMA (m)-to-SMA (m) Cable
2. Create a new session for the PXIe-5665.
3. Configure the PXIe-5665 according to the following property settings:
Acquisition Type: Spectrum
Averaging Mode: No Averaging
Ref Clock Source: PXI_Clk
Channel Coupling: AC Coupled
Span: 100 kHz
Device Instantaneous Bandwidth: 300 kHz
Resolution Bandwidth: 500 Hz
RF Attenuation: 0 dB
FFT Window Type: Flat Top
4. Configure the anti-distortion test fixture to use the 700 MHz combiner path.
PXIe-5665 Calibration Procedure | © National Instruments | 69
Page 70
5. Disable the RF source 1 and RF source 2 outputs.
Note When disabled, the RF source 1 and RF source 2 output signals should be less
than -60 dBm.
6. Disable the PXIe-5665 preamplifier.
7. Set the RF source 2 frequency according to Table 42.
8. Set the RF source 1 frequency 1 MHz above the RF source 2 frequency.
9. Set the PXIe-5665 center frequency according to the first row in Table 42.

Table 42. Gain Compression Verification Test Frequencies (Frequencies 700 MHz)

Start Frequency (MHz) Stop Frequency (MHz) Step Size (MHz)
10 90 40
100 700 200
10. Set the RF source 2 power to the RF Source 2 Programmed Power in Table 43 or Table 44 as appropriate.
11. Set the RF source 1 power to the RF Source 1 Programmed Power in Table 43 or Table 44 as appropriate.
12. Set the PXIe-5665 reference level according to Table 43 or Table 44 as appropriate.

Table 43. Gain Compression Verification Test Settings (Preamplifier Disabled)

Setting Frequency Range (MHz) Limit Value ( dBm )
RF Source 1
Programmed Power
RF Source 2
11 to 701 As-Found +8.0
*
As-Left
+9.1
10 to 700 Both -24
Programmed Power
PXIe-5665 Reference
10 to 700 Both 0
Level
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
70 | ni.com | PXIe-5665 Calibration Procedure
Page 71

Table 44. Gain Compression Test Settings (Preamplifier Enabled)

Setting Frequency Range (MHz) Limit Valu e (dBm)
RF Source 1
Programmed Power
11 to 101 As-Found -18
*
As-Left
-16
>101 to 701 As-Found -15
*
RF Source 2
As-Left
10 to 700 Both -54
-13
Programmed Power
PXIe-5665 Reference
10 to 700 Both -30
Level
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
13. Commit the PXIe-5665 settings to hardware.
14. Enable the RF source 2 output and wait 250 ms, or wait 100 ms if the output is already enabled. This wait time allows the PXIe-5665 and RF source 2 amplitudes to settle.
Note Settling times are a characteristic of the RF source device. Refer to the Test
Equipment section of this document for a list of NI-recommended source devices and
device specifications. If you use an RF source device with different device specifications, your settling times may differ from those listed in this procedure.
15. Read the power spectrum from the PXIe-5665.
16. Depending on your equipment configuration, using the spectrum returned from the PXIe-5665, measure the power at the center frequency and record this value as the Small
Tone First Measurement (RF source 1 output disabled) or as the Small Tone Second Measurement (RF source 1 output enabled).
17. Repeat steps 7 to 16 for all test frequencies in Table 42.
18. Repeat steps 7 to 17 with the PXIe-5665 preamplifier enabled.
19. Enable the RF source 1 output.
20. Repeat steps 6 to 18 with the RF source 1 output enabled.
21. Calculate the PXIe-5665 Gain Compression using the following equation:
Gain Compression = Small Tone First Measurement - Small Tone Second Measurement
PXIe-5665 Calibration Procedure | © National Instruments | 71
Page 72
22. Compare the Gain Compression to the verification test limits in Table 45 or Table 46 as appropriate.
Table 45. Gain Compression Verification Test Limits
(Frequencies
700 MHz, Preamplifier Disabled)
Compression
Frequency Limit
10 MHz to
700 MHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
As-Found 1 +8
*
As-Left
(dB)
1 +9.1
Applied Power
(dBm)
about as-left limits.
The device specification calls for less than or equal to 1 dB gain compression for a minimum applied power. The applied power includes the measurement uncertainty to guarantee the required minimum applied power.
Table 46. Gain Compression Verification Test Limits
(Frequencies
Frequency Limit
10 MHz to
100 MHz
>100 MHz to
700 MHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
700 MHz, Preamplifier Enabled)
Compression
(dB)
Applied Power
(dBm)
As-Found 1 -18
*
As-Left
1 -16
As-Found 1 -15
*
As-Left
1 -13
about as-left limits.
The device specification calls for less than or equal to 1 dB gain compression for a minimum applied power. The applied power includes the measurement uncertainty to guarantee the required minimum applied power.
23. Disable the output of RF source 1 and RF source 2.
Note When disabled, the RF source 1 and RF source 2 output signals should be less
than -60 dBm.
24. Close the PXIe-5665 session.
If the gain compression for frequencies ≤700 MHz verification procedure determines that the PXIe-5665 is outside its limits, refer to Worldwide Support and Services for information about support resources or service requests.
72 | ni.com | PXIe-5665 Calibration Procedure
Page 73

Frequencies >700 MHz

OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD SENSITIVE
OUT
ALL PORTS
50 Ω
ACCESS ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p NOM
+ 20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
ESD SENSITIVE
ACCESS
ACTIVE
OUT
IN
LO3
LO2
LO1
ALL PORTS
50 Ω
NI PXIe-5603
Downconverter 20 Hz - 3.6 GHz
RF IN
0 V DC, DC COUPLED
25 V DC MAX, AC COUPLED
+20 dBm MAX > 10 MHz
0 dBm MAX < 10 MHz
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
IF OUT
+
22 dBm MAX
0 V DC
J
J
J
J
J
J
K
K
RF
OUTPUT
RF
INPUT 1
NI-5665 ANTI-DISTORTION ET
166375A
RF
INPUT 2
6
1
5
2
4
3
1. Connect RF source 1 and RF source 2 to the PXIe-5665 RF IN connector through the anti-distortion test fixture, as shown in Figure 27 and Figure 28.
Figure 27. PXIe-5665 3.6 GHz VSA Gain Compression Verification
(Frequencies >700 MHz) Equipment Setup
1 RF Source 1 2 RF Source 2 3 SMA (m)-to-SMA (m) Cable
4 SMA (m)-to-SMA (m) Cable 5 Anti-Distortion Test Fixture 6 SMA (m)-to-SMA (m) Cable
PXIe-5665 Calibration Procedure | © National Instruments | 73
Page 74
Figure 28. PXIe-5665 14 GHz VSA Gain Compression Verification
OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD SENSITIVE
OUT
ALL PORTS
50 Ω
NI PXIe-5605
Downconverter 20 Hz - 14 GHz
ACCESS
ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p
NOM
+
20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
IF OUT
+ 22 dBm MAX
0 V DC
LO1
LO2
LO3
IN
ESD SENSITIVE
OUT
LO3
LO2
LO1
ALL PORTS
50 Ω
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
ACCESS
ACTIVE
RF IN
0 V DC +20 dBm MAX > 10 MHz +10 dBm MAX < 10 MHz
K
K
J
J
J
J
J
J
RF
OUTPUT
RF
INPUT 1
NI-5665 ANTI-DISTORTION ET 166375A
RF
INPUT 2
1
5
2
4
3
6
(Frequencies >700 MHz) Equipment Setup
1 RF Source 1 2 RF Source 2 3 SMA (m)-to-SMA (m) Cable
4 SMA (m)-to-SMA (m) Cable 5 Anti-Distortion Test Fixture 6 SMA (m)-to-SMA (m) Cable
2. Create a new session for the PXIe-5665.
3. Configure the PXIe-5665 according to the following property settings:
Acquisition Type: Spectrum
Averaging Mode: No Averaging
Ref Clock Source: PXI_Clk
Channel Coupling: AC Coupled
Span: 100 kHz
Device Instantaneous Bandwidth: 300 kHz
Resolution Bandwidth: 500 Hz
RF Attenuation: 0 dB
FFT Window Type: Flat Top
4. Configure the anti-distortion test fixture to use the >700 MHz combiner path.
74 | ni.com | PXIe-5665 Calibration Procedure
Page 75
5. Disable the RF source 1 and RF source 2 outputs.
Note When disabled, the RF source 1 and RF source 2 output signals should be less
than -60 dBm.
6. Disable the PXIe-5665 preamplifier.
7.
(PXIe-5665 14 GHz VSA) Disable the PXIe-5665 preselector.
8. Set the RF source 2 frequency according to the first row in Table 47 or Table 48 as appropriate.
Table 47. PXIe-5665 3.6 GHz Gain Compression Test Frequencies
(Frequencies >700 MHz)
Start Frequency Stop Frequency Step Size
800 MHz 3.3 GHz 500 MHz
3.6 GHz
Table 48. PXIe-5665 14 GHz Gain Compression Test Frequencies
(Frequencies >700 MHz)
Start Frequency Stop Frequency Step Size
800 MHz 3.3 GHz 500 MHz
3.6 GHz
3.8 GHz 13.8 GHz 500 MHz
14 GHz
9. Set the RF source 1 frequency 1 MHz above the RF source 2 frequency.
10. Set the PXIe-5665 center frequency according to the values listed in Table 47 or Table 48.
11. Set the RF source 2 power to the RF Source 2 Programmed Power from Table 49 or Table 50 as appropriate.
12. Set the RF source 1 power to the RF Source 1 Programmed Power from Table 49 or Table 50 as appropriate.
PXIe-5665 Calibration Procedure | © National Instruments | 75
Page 76
13. Set the PXIe-5665 reference level according to Table 49 or Table 50 as appropriate.
Table 49. Gain Compression Verification Test Settings
(Preamplifier Disabled)
Frequency
Device Setting
Range
Limit Value (dB m)
PXIe-5665
3.6 GHz VSA
RF Source 1
Programmed
Power
RF Source 2
Programmed
801 MHz to
1.701 GHz
>1.701 GHz to
3.601 GHz
800 MHz to
3.600 GHz
As-Found +8.0
*
As-Left
+9.1
As-Found +6.0
*
As-Left
+7.1
Both -24
Power
PXIe-5665
Reference
800 MHz to
3.600 GHz
Both 0
Level
PXIe-5665
14 GHz VSA
RF Source 1
Programmed
Power
RF Source 2
Programmed
801 MHz to
1.701 GHz
>1.701 GHz to
14.001 GHz
800 MHz to
14.00 GHz
As-Found +8.0
*
As-Left
+9.1
As-Found +6.0
*
As-Left
+7.1
Both -24
Power
PXIe-5665
Reference
800 MHz to
14.00 GHz
Both 0
Level
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
76 | ni.com | PXIe-5665 Calibration Procedure
Page 77
Table 50. Gain Compression Verification Test Settings
Device Setting
(Preamplifier Enabled)
Frequency
Range
Limit Value (dB m)
PXIe-5665
3.6 GHz VSA
RF Source 1 Programmed
Power
RF Source 2 Programmed
801 MHz to
1.701 GHz
>1.701 GHz to
3.601 GHz
800 MHz to
3.600 GHz
As-Found -15
*
As-Left
-13
As-Found -18
*
As-Left
-16
Both -54
Power
PXIe-5665
Reference
800 MHz to
3.600 GHz
Both -30
Level
PXIe-5665
14 GHz VSA
RF Source 1 Programmed
Power
RF Source 2 Programmed
801 MHz to
1.701 GHz
>1.701 GHz to
14.001 GHz
800 MHz to
14.00 GHz
As-Found -15
*
As-Left
-13
As-Found -18
*
As-Left
-16
Both -54
Power
PXIe-5665
Reference
80 MHz to
14.00 GHz
Both -30
Level
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
14. Commit the PXIe-5665 settings to hardware.
15. Enable the RF source 2 output and wait 250 ms, or wait 100 ms if the output is already enabled. This wait time allows the PXIe-5665 and RF source 2 amplitudes to settle.
Note Settling times are a characteristic of the RF source device. Refer to the Test
Equipment section of this document for a list of NI-recommended source devices and
device specifications. If you use an RF source device with different device specifications, your settling times may differ from those listed in this procedure.
16. Read the power spectrum from the PXIe-5665.
PXIe-5665 Calibration Procedure | © National Instruments | 77
Page 78
17. Depending on your equipment configuration, using the spectrum returned from the PXIe-5665, measure the power at the center frequency and record this value as the Small
Tone First Measurement (RF source 1 output disabled) or as the Small Tone Second Measurement (RF source 1 output enabled).
18. Repeat steps 8 to 17 for all test frequencies in step 8.
19. (PXIe-5665 14 GHz VSA) Repeat steps 8 to 18 for frequencies greater than 3.6 GHz with the PXIe-5665 preselector enabled.
20. Repeat steps 7 to 18 with the PXIe-5665 preamplifier enabled for frequencies less than
3.6 GHz.
21. Enable the RF source 1 output.
22. Repeat steps 6 to 20 with the RF source 1 output enabled.
23. Calculate the PXIe-5665 Gain Compression using the following equation:
Gain Compression = Small Tone First Measurement - Small Tone Second Measurement
24. Compare the Gain Compression to the verification test limits in Table 51 or Table 52.

Table 51. Gain Compression Verification Test Limits (>700 MHz, Preamplifier Disabled)

Compression
Frequency Limit
>700 MHz to
1.70 GHz
>1.70 GHz to
3.6 GHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
As-Found 1 +8.0
*
As-Left
As-Found 1 +6.0
*
As-Left
(dB)
1 +9.1
1 +7.1
Applied Power
(dBm)
about as-left limits.
The device specification calls for less than or equal to 1 dB gain compression for a minimum applied power. The applied power includes the measurement uncertainty to guarantee the required minimum applied power.
78 | ni.com | PXIe-5665 Calibration Procedure
Page 79

Table 52. Gain Compression Verification Test Limits (>700 MHz, Preamplifier Enabled)

Compression
Frequency Limit
>700 MHz to
1.70 GHz
>1.70 GHz to
3.6 GHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
As-Found 1 -15
*
As-Left
As-Found 1 -18
*
As-Left
(dB)
1 -13
1 -16
Applied Power
(dBm)
about as-left limits.
The device specification calls less than or equal to 1 dB gain compression for a minimum applied power. The applied power includes the measurement uncertainty to guarantee the required minimum applied power.
25. Disable the output of RF source 1 and RF source 2.
Note When disabled, the RF source 1 and RF source 2 output signals should be less
than -60 dBm.
26. Close the PXIe-5665 session.
If the gain compression for frequencies >700 MHz verification procedure determines that the PXIe-5665 is outside its limits, refer to Worldwide Support and Services for information about support resources or service requests.

Verifying LO Output Power

Note Prior to starting this procedure, zero and calibrate the power sensor according
to the Zeroing and Calibrating the Power Sensor section of this document.
1. Connect the RF source 2 output to the PXIe-5605 RF IN connector using the SMA (m)-to­SMA (m) cable.
2. Connect power sensor A to the PXIe-5605 LO3 OUT connector using the SMA (m)-to­SMA (m) cable.
3. Connect the spectrum analyzer to the PXIe-5605 LO2 OUT connector.
PXIe-5665 Calibration Procedure | © National Instruments | 79
Page 80
4. Connect power sensor B to the PXIe-5605 LO1 OUT connector. The completed setup is
1
2
3
5
6
7
4
ACCESS
ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p
NOM
+
20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-562 2
16-Bit IF Digitizer
IF IN
ESD SENSITIVE
ACCESS
ACTIVE
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
OUT
ALL PORTS
50 Ω
NI PXIe-5653
Synthesizer
REF IN
10 MHz
2 V p-p MAX
REF OUT
10 MHz
0.5 V rms
REF OUT
100 MHz
0.5 V rms
ESD SENSITIVE
ACCESS
ACTIVE
OUT
IN
LO3
LO2
LO1
ALL PORTS
50 Ω
NI PXIe-5603
Downconverter 20 Hz - 3.6 GHz
RF IN
0 V DC, DC COUPLED
25 V DC MAX, AC COUPLED
+20 dBm MAX > 10 MHz
0 dBm MAX < 10 MHz
IF OUT
+
22 dBm MAX
0 V DC
+
15 dBm MAX 25 V DC MAX
+
15 dBm MAX
0 V DC
K
K
J
J
J
J
J
J
8
shown in Figure 29 and Figure 30.

Figure 29. PXIe-5665 3.6 GHz VSA LO Output Power Verification Equipment Setup

1 RF Source 2 2 Spectrum Analyzer 3 Power Meter
Note Attach the 10 dB attenuator directly to the PXIe-5606 LO2 OUT connector.
4 SMA (m)-to-SMA (m) Cable 5 SMA (m)-to-SMA (m) Cable 6 Power Sensor A
80 | ni.com | PXIe-5665 Calibration Procedure
7 Power Sensor B 8 10 dB Attenuator
Page 81

Figure 30. PXIe-5665 14 GHz VSA LO Output Power Verification Equipment Setup

OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD SENSITIVE
OUT
ALL PORTS
50
NI PXIe-5605
Downconverter 20 Hz - 14 GHz
ACCESS
ACTIVE
CLK IN
CLK OUT
PFI 1
ESD SENSITIVE
TTL
6.3 Vp-p MAX
2 Vp-p NOM
+ 20 dBm MAX
50 Ω
50
50
NI PXIe-5622
16-Bit IF Digitizer
IF IN
LO1
LO2
LO3
IN
ESD SENSITIVE
OUT
LO3
LO2
LO1
ALL PORTS
50
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
ACCESS
ACTIVE
RF IN
0 V DC +20 dBm MAX > 10 MHz +10 dBm MAX < 10 MHz
IF OUT
+
22 dBm MAX
0 V DC
K
K
J
J
J
J
J
J
1
2
3
4
7
8
6
5
1 RF Source 2 2 Spectrum Analyzer 3 Power Meter
4 SMA (m)-to-SMA (m) Cable 5 SMA (m)-to-SMA (m) Cable 6 Power Sensor A
7 Power Sensor B 8 10 dB Attenuator
Note Attach the 10 dB attenuator directly to the PXIe-5606 LO2 OUT connector.
5. Configure the spectrum analyzer according to the following settings:
Resolution Bandwidth: 1 kHz
Span: 0 Hz
Reference Level: +20 dBm
Center Frequency: 4 GHz
6. Configure the RF source 2 according to the following settings:
•Power: -20dBm
Frequency: 1 GHz
Note The RF source 2 settings do not affect this test. It is programmed to a known
state for convenience only.
7. Create a new session for the PXIe-5665.
PXIe-5665 Calibration Procedure | © National Instruments | 81
Page 82
8. Configure the PXIe-5665 according to the following property setting:
LO Export Enabled: True
9. Initiate the PXIe-5665.
10. Close the PXIe-5665 session.
11. Create a new session for the PXIe-5653.
12. Configure the PXIe-5653 according to the following property settings:
Ref Clock Source: PXI -Clk
Frequency: 4.05625 GHz
13. Read the power meter channel B power and correct for the PXIe-5653 frequency.
14. Repeat step 13 and record the frequency where the minimum measured power occurs for the following PXIe-5653 frequencies:
Minimum frequency: 4.05625 GHz
Maximum frequency: 8.2125 GHz
Frequency step: 50 MHz
15. Tune the PXIe-5653 to the frequency found in step 14.
16. Measure the LO1 power with power meter channel B. Correct the reading for the LO1 frequency.
17. Measure the LO2 power with the spectrum analyzer. Correct the reading for the 4 GHz LO2 frequency using the following equation:
LO2 Power = Spectrum Analyzer Reading + Spectrum Analyzer Correction Factor
1
18. Measure the LO3 power with power meter channel A. Correct the reading for the LO3 frequency of 800 MHz.
19. Compare the readings from steps 16 to 18 to the following verification test limits in Table 53.

Table 53. LO Output Power Verification Test Limits

LO As-Found (dBm) As-Left (dBm)
LO1 >7.0 7.0 to 8.0
LO2 >7.0 9.0 to 10.0
LO3 >7.0 9.0 to 10.0
20. Close the PXIe-5653 session.
If the LO output power verification procedure determines that the PXIe-5665 is outside its limits, refer to Worldwide Support and Services for information about support resources or service requests.
1
The Spectrum Analyzer Correction Factor value was calculated in the Characterizing Spectrum Analyzer
Response equipment characterization procedure.
82 | ni.com | PXIe-5665 Calibration Procedure
Page 83

Reverification

Repeat the Verification section to determine the as-left status of the device.

Appendix A: Anti-Distortion Test Fixture

Note The anti-distortion test fixture specifications are for the entire fixture and
include the performance of all components internal to the fixture.

Table 54. Lowpass Filter Passband Paths

Filter Path (MHz) Start (MHz) Stop (MHz)
470 10 470 2.5 1.75
735 >470 735 2.5 1.75
1,150 >735 1,150 2.5 1.75
1,800 >1,150 1,800 2.5 1.75
2,530 >1,800 2,530 2.5 1.75
3,550 >2,530 3,550 2.5 1.75
4,985 >3,550 4,985 2.5 1.75
7,000 >4,985 7,000 2.5 1.75
*
Input or output ports.
Insertion Loss
(dB, max.)
VSWR* (max.)

Table 55. Lowpass Filter Stopband Paths

Filter Path (MHz) Start (MHz) Stop (MHz) Rejection
470 600 940 50
735 >940 1,470 50
1,150 >1,470 2,300 50
1,800 >2,300 3,600 50
*
2,530 >3,600 5,060 70
3,550 >5,060 7,100 70
4,985 >7,100 9,970 70
7,000 >9,970 14,000 70
*
Rejection equals Fundamental Power - Second Harmonic Power
PXIe-5665 Calibration Procedure | © National Instruments | 83
Page 84

Table 56. Combiner Paths

Tot al
*
Combiner Path Start Stop
2 1 >700 MHz 10 GHz 8.0 20 1.75
2 18.0
2 1 >10 GHz 14 GHz 8.0 20 2.00
2 18.0
*
Total Loss includes the splitter loss at either input or output ports. The unused input port is terminated
in 50 Ω.
Isolation is the loss between input ports. The output port is terminated in 50 Ω.
VSWR includes any port. All other ports are terminated in 50 Ω.
Loss
(dB, max.)
Isolation (dB, min.)
VSWR
(max.)

Appendix B: Power Sensor Calibration Factor Uncertainty

Table 57. Maximum Power Sensor Calibration Factor Uncertainty

Frequency Calibration Factor Uncertainty (%)
10 MHz 1.15
20 MHz 0.80
30 MHz 0.65
40 MHz to 200 MHz ≤0.55
300 MHz to 400 MHz ≤0.60
500 MHz to 1 GHz ≤0.50
2GHz to 4GHz ≤0.60
5 GHz 0.75
6 GHz 0.80
7 GHz 0.85
8 GHz 0.90
9 GHz 0.95
10 GHz 1.00
84 | ni.com | PXIe-5665 Calibration Procedure
Page 85
Table 57. Maximum Power Sensor Calibration Factor Uncertainty (Continued)
Frequency Calibration Factor Uncertainty (%)
11 GHz 1.05
12 GHz 1.10
13 GHz 1.15
14 GHz 1.20
15 GHz 1.25

Worldwide Support and Services

The National Instruments website is your complete resource for technical support. At ni.com/
support
self-help resources to email and phone assistance from NI Application Engineers.
Vis it other services.
Vis it facilitates technical support and ensures that you receive important information updates from NI.
A Declaration of Conformity (DoC) is our claim of compliance with the Council of the European Communities using the manufacturer’s declaration of conformity. This system affords the user protection for electromagnetic compatibility (EMC) and product safety. You can obtain the DoC for your product by visiting you can obtain the calibration certificate for your product at
you have access to everything from troubleshooting and application development
ni.com/services for NI Factory Installation Services, repairs, extended warranty, and
ni.com/register to register your National Instruments product. Product registration
ni.com/certification. If your product supports calibration,
ni.com/calibration.
National Instruments corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504. National Instruments also has offices located around the world. For telephone support in the United States, create your service request at
ni.com/support or dial
512 795 8248. For telephone support outside the United States, visit the Worldwide Offices section of
ni.com/niglobal to access the branch office websites, which provide up-to-date
contact information, support phone numbers, email addresses, and current events.
Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for more information on National Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patents Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and how to obtain re levant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.
© 2011–2017 National Instruments. All rights reserved.
373205G-01 Nov17
Loading...