This document contains the verification procedures for the National Instruments PXIe-5665
RF vector signal analyzer (VSA). Refer to
calibration solutions.
When not otherwise specified, the procedures in this document refer to both the PXIe-5665
3.6 GHz VSA and the PXIe-5665 14 GHz VSA products. In places where the procedures differ
between the two products, the appropriate device settings are specified.
Note PXIe-5665 tuned frequencies greater than 3.6 GHz and procedures with the
preselector enabled for frequencies greater than 3.6 GHz apply only to the PXIe-5665
14 GHz VSA.
NI warrants the PXIe-5665 to meet its published specifications if the individual modules are
calibrated and operating within specifications. For more information about RF system
calibration, visit
ni.com/manuals and search for Letter of Conformance.
Appendix A: Anti-Distortion Test Fixture................................................................................ 83
Appendix B: Power Sensor Calibration Factor Uncertainty .....................................................84
Worldwide Support and Services ............................................................................................. 85
Software
Calibrating the PXIe-5665 requires you to install the following software on the calibration
system:
•NI-RFSA 2.5 or later
•NI Spectral Measurements Toolkit 2.5 or later
You can download all required software from
The software supports programming the calibration procedures in the LabVIEW, C, and
LabWindows™/CVI™ application development environments (ADEs). When you install the
software, you need to install support only for the ADE that you intend to use.
ni.com/downloads.
Documentation
You might find the following documents helpful as you perform the calibration procedure:
•NI 5665 RF Vector Signal Analyzer Getting Started Guide
•NI RF Vector Signal Analyzers Help
•PXIe-5665 Specifications
The latest versions of these documents are available at ni.com/manuals.
This calibration procedure calibrates the PXIe-5665 as a single system. To calibrate the
PXIe-5653 or PXIe-5622 individually, refer to their calibration procedures, available at
ni.com/manuals.
2 | ni.com | PXIe-5665 Calibration Procedure
Page 3
Test Equipment
Table 1 lists the equipment NI recommends for the performance verification procedures. If the
recommended equipment is not available, select a substitute using the minimum requirements
listed in the table.
The following setup and environmental conditions are required to ensure the PXIe-5665 meets
published specifications.
•Keep cabling as short as possible. Long cables and wires act as antennae, picking up extra
noise that can affect measurements.
•Verify that all connections, including front panel connections and screws, are secure.
•Maintain an ambient temperature of 23 °C ±5 °C.
•Keep the relative humidity between 10% and 90%, noncondensing.
•Allow a warm-up time of at least 30 minutes after NI-RFSA is loaded and recognizes the
PXIe-5665. The warm-up time ensures that the measurement circuitry of the PXIe-5665 is
at a stable operating temperature.
•Perform self-calibration on the PXIe-5665.
•Use a torque wrench appropriate for the type of RF connector that you are using.
NI recommends a 0.565 N · m (5 lb · in.) wrench for SMA connectors and a 0.90 N · m
(8 lb · in.) wrench for 3.5 mm connectors.
•Connect the frequency reference source to the REF IN connector on the back of the
PXI Express chassis with a standard BNC (m)-to-BNC (m) cable. This connection replaces
the connection from the PXIe-5653 REF OUT (10 MHz) connector to the PXI Express
chassis REF IN connector, if present.
•Lock all test equipment to the REF OUT signal on the back of the PXI Express chassis.
Refer to the NI 5665 Timing Configurations topic in the NI RF Vector Signal Analyzers Help for more information about configuring clocking sources.
•Ensure that the PXI/PXI Express chassis fan speed is set to HIGH, that the fan filters, if present,
are clean, and that the empty slots contain filler panels. For more information, refer to the
Maintain Forced-Air Cooling Note to Users document available at
ni.com/manuals.
10 | ni.com | PXIe-5665 Calibration Procedure
Page 11
System Options
The PXIe-5665 is available with options without certain features. The calibration procedure
must be modified to handle the following options:
•No preamplifier: Skip steps in the verification procedures that call for the preamplifier to
be enabled.
•25 MHz instantaneous bandwidth: Verification procedures include any changes necessary
to accommodate this option. There are no changes to the PXIe-5665 specifications for this
digitizer option.
Initial Setup
Refer to the NI PXIe-5665 RF Vector Signal Analyzer Getting Started Guide for information
about how to install the software and hardware and how to configure the device in Measurement
& Automation Explorer (MAX).
Characterizing the Test System
You use the measured response of the test system during verification tests.
The power splitter and attenuation response is measured at the RF input frequencies used in the
verification tests.
Caution The connectors on the device under test (DUT) and test equipment are
fragile. Perform the steps in these procedures with great care to prevent damaging any
DUTs or test equipment.
Zeroing and Calibrating the Power Sensor
1.Connect channel A of the power meter to power sensor A.
2.Connect channel B of the power meter to power sensor B.
3.Zero and calibrate the power sensors using the built-in functions in the power meter.
Characterizing Power Splitter Reference Output
Designate either of the two outputs of the power splitter as the reference output. This output must
be consistently used as the reference output throughout the characterization and verification
procedures.
1.Connect the RF source 2 output to the power splitter input through the SMA (m)-toSMA (m) cable.
2.Connect power sensor A to the reference output of the power splitter through the
SMA (m)-to-SMA (m) cable.
3.Connect power sensor B to the other output of the power splitter through the
SMA (m)-to-SMA (m) cable. The completed setup is shown in Figure 1.
Figure 1. Power Splitter Difference Characterization Equipment Setup
1 RF Source 2
2 SMA (m)-to-SMA (m) cable
3 Power Splitter
4 Power Sensor A
5 Power Sensor B
6 Power Meter
4.Set the RF source 2 frequency to 4 GHz.
5.Set the RF source 2 power to 0 dBm.
6.Measure the signal at 4 GHz with both power sensors.
7.Set the power sensors calibration factor for the measurement frequency at 4 GHz.
8.Calculate the Power Splitter at 4 GHz Correction Factor using the following formula:
Power Splitter at 4 GHz Correction Factor = Corrected Power Sensor B Power - Corrected
Power Sensor A Power
12 | ni.com | PXIe-5665 Calibration Procedure
Page 13
Characterizing RF Source Power (Direct)
Note Prior to starting this procedure, zero and calibrate the power sensor according
to the Zeroing and Calibrating the Power Sensor section of this document.
1.Connect the RF source 2 output to the power sensor B input using the SMA (m)-toSMA (m) cable and 3.5 mm (f)-to-3.5 mm (f) adaptor. The completed equipment setup is
shown in Figure 2.
Figure 2. RF Source 2 Output Power Characterization Equipment Setup
5
4
3
1
2
1 RF Source 2
2 SMA (m)-to-SMA (m) Cable
3 3.5 mm (f)-to-3.5 mm (f) Adaptor
4 Power Sensor B
5 Power Meter
2.Set the RF source 2 frequency according to the first row in Table 2 or Table 3 as
appropriate.
Table 3. PXIe-5665 14GHz VSA RF Source 2 Characterization Frequencies
Start FrequencyStop FrequencyStep Size
100.033325 MHz1.975033325 GHz125 MHz
2.099966675 GHz3.599966675 GHz125 MHz
3.724966675 GHz13.974966675 GHz125 MHz
13.999966675 GHz——
9.325033325 GHz11.200033325 GHz125 MHz
11.324966675 GHz12.824966675 GHz125 MHz
4.949966675 GHz15.199966675 GHz125 MHz
15.224966675 GHz——
1.325033325 GHz3.200033325 GHz125 MHz
3.324966675 GHz4.824966675 GHz125 MHz
3.349966675 GHz13.599966675 GHz125 MHz
13.624966675 GHz——
24.966675 MHz274.966675 MHz125 MHz
1.724966675 GHz3.224966675 GHz125 MHz
3.3875 GHz——
987.5 MHz——
3.Set the RF source 2 power to 0 dBm.
4.Measure the Channel B Power using the appropriate calibration factor for the power sensor
frequency. Calculate the RF Source Power Direct using the following equation:
RF Source Power Direct = -Channel B Power
5.Repeat steps 2 to 4 for all remaining frequencies in Table 2 or Table 3.
14 | ni.com | PXIe-5665 Calibration Procedure
Page 15
Characterizing RF Source Power (with Lowpass Filter)
Note Prior to starting this procedure, zero and calibrate the power sensor according
to the Zeroing and Calibrating the Power Sensor section of this document.
1.Connect the RF source 2 output to power sensor B through the anti-distortion test fixture.
The completed equipment setup is shown in Figure 3.
3.Set the RF source 2 frequency and power according to the first row in Table 4.
Table 4. RF Source 2 Settings
Start FrequencyStop FrequencyStep SizePower (dBm)
300 MHz1.8 GHz10 MHz0
1.9 GHz7.0 GHz100 MHz0
300.0333125 MHz1790.0333125 MHz10 MHz-30
1.8 GHz——-30
4.Configure the anti-distortion test fixture to use the appropriate lowpass filter path according
to Table 5.
Table 5. Lowpass Filter Frequency Ranges
Lowpass Filter Path (MHz)Frequency Range
4701 MHz to 470 MHz
735>470 MHz to 735 MHz
1,150>735 MHz to 1.150 GHz
1,800>1.150 GHz to 1.800 GHz
5.Enable the RF source 2 output.
6.Measure the channel B power using the appropriate calibration factor for the power sensor
frequency. Adjust the RF source 2 power until this measured power is within ±0.1 dB of the
value listed in Table 4. The final RF source 2 power setting is the RF Source Power LPF
for that frequency.
7.Repeat steps 3 to 6 for all the frequencies and powers in Table 4.
16 | ni.com | PXIe-5665 Calibration Procedure
Page 17
Characterizing RF Source Power (Combined) for
Frequencies ≤700 MHz
Note Prior to starting this procedure, zero and calibrate the power sensor according
to the Zeroing and Calibrating the Power Sensor section of this document.
1.Connect the RF source 1 and RF source 2 outputs to power sensor B through the
anti-distortion test fixture. The completed equipment setup is shown in Figure 4.
Figure 4. RF Source Power (Combined) Characterization Equipment Setup
(Frequencies ≤700 MHz)
5
6
NI-5665 ANTI-DISTORTION ET
INPUT 1
166375A
RF
RF
INPUT 2
OUTPUT
RF
1
3
4
9
2
1 RF Source 1
2 RF Source 2
3 SMA (m)-to-SMA (m) Cable
4 SMA (m)-to-SMA (m) Cable
5 Anti-Distortion Test Fixture
6 SMA (m)-to-SMA (m) Cable
7 3.5 mm (f)-to-3.5 mm (f) Adaptor
8 Power Sensor B
9 Power Meter
2.Set the anti-distortion test fixture to the ≤700 MHz combiner path.
3.Set the RF source 1 and RF source 2 frequency and power values according to the first row
in Table 6.
Table 6. RF Source Power (Combined) Test Settings (Frequencies ≤700 MHz)
Start
Frequency
(MHz)
12.0592.0510+350 kHz-10-350 kHz-10
100700100+350 kHz-10-350 kHz-10
12.0592.0510+350 kHz-30-350 kHz-30
100700100+350 kHz-30-350 kHz-30
1010010+1 MHz+9.10Hz-24
1010010+1 MHz-160Hz-54
200700100+1 MHz+9.10Hz-24
200700100+1 MHz-130Hz-54
1010010+1 MHz+80Hz-24
200700100+1 MHz+80Hz-24
1010010+1 MHz-180Hz-54
200700100+1 MHz-150Hz-54
Stop
Frequency
(MHz)
Step Size
(MHz)
RF Source 1RF Source 2
Frequency
Offset
Power
(dBm)
Frequency
Offset
4.Disable the RF source 1 output, and enable the RF source 2 output.
Note When disabled, the RF source 1 output signal should be less than -60 dBm.
Power
(dBm)
5.Measure the channel B power using the appropriate sensor calibration factor for the
RF source 2 frequency.
6.Repeat step 5, adjusting the RF source 2 power until the channel B reading is within
±0.1 dB of the power listed in Table 6. The RF source 2 power is the
RF Source 2 Programmed Power.
7.Disable the RF source 2 output, and enable the RF source 1 output.
Note When disabled, the RF source 2 output signal should be less than - 60 dBm.
8.Measure the channel B power using the appropriate sensor calibration factor for the
RF source 1 frequency.
9.Repeat step 8, adjusting the RF source 1 power until the channel B power is within ±0.1 dB
of the power listed in Table 6. The RF source 1 power is the RF Source 1 Programmed Power.
10. Repeat steps 3 to 9 for all remaining frequency and power values in Table 6.
18 | ni.com | PXIe-5665 Calibration Procedure
Page 19
Characterizing RF Source Power (Combined) for
Frequencies >700 MHz
Note Prior to starting this procedure, zero and calibrate the power sensor according
to the Zeroing and Calibrating the Power Sensor section of this document.
1.Connect the RF source 1 and RF source 2 outputs to power sensor B through the
anti-distortion test fixture. The completed equipment setup is shown in Figure 5.
Figure 5. RF Source Power (Combined) Characterization Equipment Setup
(Frequencies >700 MHz)
5
6
NI-5665 ANTI-DISTORTION ET
INPUT 1
166375A
RF
RF
INPUT 2
OUTPUT
RF
1
3
4
9
2
1 RF Source 1
2 RF Source 2
3 SMA (m)-to-SMA (m) cable
4 SMA (m)-to-SMA (m) cable
6 SMA (m)-to-SMA (m) cable
7 3.5 mm (f)-to-3.5 mm (f) Adaptor
8 Power Sensor B
9 Power Meter
5 Anti-Distortion Test Fixture
2.Set the anti-distortion test fixture to the >700 MHz combiner path.
3.Set the RF source 1 and RF source 2 frequency and power values according to the first row
in Table 7.
Table 7. PXIe-5665 RF Source Power (Combined) Test Settings (Frequencies
>700 MHz)
RF Source 1RF Source 2
Start
Frequency
800 MHz3.500 GHz100+350 kHz-10-350 kHz-10
3598.95 MHz——+350 kHz-10-350 kHz-10
3.6 GHz13.8 GHz200+350 kHz-10-350 kHz-10
13.99895 GHz——+350 kHz-10-350 kHz-10
800 MHz3.500 GHz100+350 kHz-30-350 kHz-30
3598.95 MHz——+350 kHz-30-350 kHz-30
3.6 GHz——+350 kHz-30-350 kHz-30
800 MHz1.700 GHz100+1 MHz+9.10Hz-24
800 MHz1.700 GHz100+1 MHz+80Hz-24
800 MHz1.700 GHz100+1 MHz-130Hz-54
800 MHz1.700 GHz100+1 MHz-150Hz-54
Stop
Frequency
Step Size
(MHz)
Frequency
Offset
Power
(dBm)
Frequency
Offset
Powe r
(dBm)
1.800 GHz14.000 GHz100+1 MHz+7.10Hz-24
1.800 GHz14.000 GHz100+1 MHz+60Hz-24
1.800 GHz3.600 GHz100+1 MHz-160Hz-54
1.800 GHz3.600 GHz100+1 MHz-180Hz-54
20 | ni.com | PXIe-5665 Calibration Procedure
Page 21
4.Disable the RF source 1 output, and enable the RF source 2 output.
Note When disabled, the RF source 1 output signal should be less than -60 dBm.
5.Measure the channel B power using the appropriate sensor calibration factor for the
RF source 2 frequency.
6.Repeat step 5, adjusting the RF source 2 power until the channel B reading is within
±0.1 dB of the power in Table 7. The RF source 2 power is the RF Source 2 Programmed Power.
7.Disable the RF source 2 output, and enable the RF source 1 output.
Note When disabled, the RF source 2 output signal should be less than -60 dBm.
8.Measure the channel B power using the appropriate sensor calibration factor for the
RF source 1 frequency.
9.Repeat step 8, adjusting the RF source 1 power until the channel B power is within ±0.1 dB
of the power in Table 7. The RF source 1 power is the RF Source 1 Programmed Power.
10. Repeat steps 3 to 9 for all remaining frequency and power values listed in Table 7.
Characterizing RF Source Power (Through Splitter)
Note Prior to starting this procedure, zero and calibrate the power sensor and define
the power splitter reference output using the procedures in the Zeroing and
Calibrating the Power Sensor and Characterizing Power Splitter Reference Output
sections of this document.
1.Connect the RF source 2 output to the power splitter input through the SMA (m)-toSMA (m) cable.
2.Connect a 50 Ω termination load to the reference output of the power splitter.
3.Connect power sensor B to the other output of the power splitter through the
1
3
567
8
2
4
3.5 mm (m)-to-3.5 mm (m) adaptors and the 3.5 mm (f)-to-3.5 mm (f) adaptor. The
completed equipment setup is shown in Figure 6.
Figure 6. RF Source Power (Through Splitter) Characterization Equipment Setup
1 RF Source 2
2 SMA (m)-to-SMA (m) Cable
350Ω Termination Load
4 Power Splitter
22 | ni.com | PXIe-5665 Calibration Procedure
5 3.5 mm (m)-to-3.5 mm (m) Adaptor
6 3.5 mm (f)-to-3.5 mm (f) Adaptor
7 Power Sensor B
8Power Meter
Page 23
4.Set the RF source 2 power to -10 dBm.
5.Set the RF source 2 frequency according to the first row in Table 8.
Table 8. RF Source 2 Frequencies
Start FrequencyStop FrequencyStep Size
100 MHz3.5 GHz200 MHz
3.6 GHz——
3.7 GHz14 GHz100 MHz
6.Measure the Channel B Power using the appropriate calibration factor for the RF source 2
frequency. Adjust the RF source 2 power until this measured power is within ±0.1 dB of
-10 dBm.
7.Store the RF source 2 set power as the RF Source Power through Splitter.
8.Repeat steps 5 to 7 for all remaining frequencies in Table 8.
Characterizing RF Source Power (with Splitter and
Attenuator)
Note Prior to starting this procedure, zero and calibrate the power sensor and define
the power splitter reference output using the procedures in the Zeroing and
Calibrating the Power Sensor and Characterizing Power Splitter Reference Output
sections of this document.
1.Connect the RF source 2 output to the power splitter input through the SMA (m)-toSMA (m) cable.
2.Connect power sensor A to the reference output of the power splitter.
3.Connect the spectrum analyzer to the other power splitter output through the
SMA (m)-to-SMA (m) cable. The completed equipment setup is shown in Figure 8.
1 RF Source 2
2 SMA (m)-to-SMA (m) cable
3 Power Splitter
4
4 Power Sensor A
5 10 dB Attenuator
6 SMA (m)-to-SMA (m) cable
365
2
7 Spectrum Analyzer
8 Power Meter
4.Set the RF source 2 power to 0 dBm.
5.Set the RF source 2 frequency to 4 GHz.
6.Configure the spectrum analyzer according to the following settings:
•Resolution bandwidth: 1 kHz
•Center frequency: 4 GHz
•Frequency span: 0 Hz
•Reference level: 0 dBm
7.Measure the power sensor A power.
8.Correct the power sensor A measured power for the measurement frequency of 4 GHz.
9.Correct the power sensor A measured power for the Power Splitter at 4 GHz Correction Factor from the Characterizing Power Splitter Difference equipment characterization
section of this document using the following equation:
Corrected Power Sensor Power = Power Sensor A reading + Power Splitter at 4 GHz
Correction Factor
26 | ni.com | PXIe-5665 Calibration Procedure
Page 27
10. Measure the spectrum analyzer peak power.
11. Calculate the Spectrum Analyzer Correction Factor using the following formula:
Spectrum Analyzer Correction Factor =
Corrected Power Sensor Power - Spectrum Analyzer Power
As-Found and As-Left Limits
The as-found limits are the published specifications for the PXIe-5665. NI uses these limits to
determine whether the PXIe-5665 meets the device specifications when it is received for
calibration.
The as-left limits are equal to the published NI specifications for the PXIe-5665, less guard
bands for manufacturing test measurement uncertainty, temperature drift, and drift over time. NI
uses these limits to determine whether the PXIe-5665 meets the device specifications over its
calibration interval.
Verification
The performance verification procedures assume that adequate traceable uncertainties are
available for the calibration references.
In the event of a failure during the verification of the PXIe-5665, perform a calibration of the
individual modules. Return the PXIe-5603 or PXIe-5605 module to NI for calibration and
adjustment, if needed.
Verifying Reference Accuracy
The PXIe-5665 reference accuracy specification is derived from the PXIe-5653 reference
accuracy. For more information about PXIe-5653 calibration values, refer to the NI PXIe-5653 Calibration Procedure available at
ni.com/manuals.
Verifying Phase Noise
The phase noise for the PXIe-5665 system is determined by the phase noise of the PXIe-5653
module. To verify the PXIe-5653 phase noise specification, you must use the phase noise
verification test for the PXIe-5653. For more information about calibration and verification of
the PXIe-5653, refer to the NI PXIe-5653 Calibration Procedure, available at
11. Take five readings of the RF source 2 amplitude using power sensor A. Calculate the
average power.
12. Repeat step 11 until any two readings measured differ no more than 0.02 dB and no
two averages differ by more than 0.005 dB. This average is the Measured Average Power.
Note More than 30 averages total indicate a test failure that you must diagnose.
13. Calculate the Corrected Input Power using the following formula:
Corrected Input Power = Measured Average Power -
DUT to Power Meter through Attenuator Path Loss
Note The RF Source to DUT through Attenuator Path and DUT to Power Meter
through Attenuator Path Loss values were measured in the Characterizing RF
Source Power (with Splitter and Attenuator) equipment characterization procedure.
Use the values that correspond to the frequency range you are testing.
14. Read the PXIe-5665 power spectrum. The PXIe-5665 Power is the peak value of that
spectrum.
30 | ni.com | PXIe-5665 Calibration Procedure
Page 31
15. Calculate the Absolute Amplitude Accuracy at each RF frequency using the following
formula:
Absolute Amplitude Accuracy = PXIe-5665 Power - Corrected Input Power
16. Repeat steps 8 to 15 for all remaining frequencies in Table 10.
17. Calculate the Frequency Response at each RF frequency using the following formula:
Frequency Response = Absolute Amplitude Accuracy -
Absolute Amplitude Accuracy at 612.5 MHz
Note Calculate Frequency Response only for PXIe-5665 center frequencies less
than or equal to 3.6 GHz.
18. Repeat steps 9 to 17 for all reference levels listed in step 7.
19. Repeat steps 8 to 18 for an instantaneous bandwidth of 50 MHz (25 MHz for the
PXIe-5665 with 25 MHz bandwidth).
20. Repeat steps 8 to 19 with the PXIe-5665 preamplifier enabled.
21.
(PXIe-5665 14 GHz VSA) Repeat steps 8 to 19 for test frequencies greater than 3.6 GHz
with the PXIe-5665 preselector enabled and preamplifier disabled.
Note The PXIe-5665 preselector has a maximum instantaneous bandwidth of
47 MHz. For steps that call for a 50 MHz instantaneous bandwidth, use a 40 MHz
instantaneous bandwidth when using the preselector.
Table 14. Absolute Amplitude Accuracy Verification Test Limits (Preamplifier Enabled)
FrequencyDevice
612.5 MHzPXIe-5665
As-Found Limit
(dB)
±0.35±0.30
As-Left Limit
(dB)
*
3.6 GHz VSA
PXIe-5665
±0.70±0.60
14 GHz VSA
10 MHz to
100 MHz
PXIe-5665
3.6 GHz VSA
PXIe-5665
±1.10±0.80
±1.45
14 GHz VSA
>100 MHz to
2.8 GHz
PXIe-5665
3.6 GHz VSA
PXIe-5665
±0.80±0.60
±1.15
14 GHz VSA
>2.8 GHz to
3.6 GHz
PXIe-5665
3.6 GHz VSA
PXIe-5665
±0.80±0.60
±1.20±0.75
14 GHz VSA
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
Table 15. PXIe-5665 14 GHz VSA Absolute Amplitude Accuracy Verification Test Limits
(Preamplifier Disabled, PXIe-5665 14 GHz VSA Preselector Enabled)
FrequencyAs-Found Limit (dB)As-Left Limit (dB)
*
>3.6 GHz to 7.5 GHz±4.0±1.5
>7.5 GHz to 8.5 GHz±4.0±1.5
>8.5 GHz to 14 GHz±4.0±1.5
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
24. Close the PXIe-5665 session.
If the frequency response and absolute amplitude accuracy verification procedures determines
that the PXIe-5665 is outside its limits, refer to Worldwide Support and Services for information
about support resources or service requests.
34 | ni.com | PXIe-5665 Calibration Procedure
Page 35
Verifying Average Noise Level
1.Connect a 50 Ω termination to the PXIe-5665 RF IN connector. The completed equipment
setup is shown in Figure 11 and Figure 12.
Figure 11. PXIe-5665 3.6 GHz VSA Average Noise Level Verification Equipment Setup
4.Set the PXIe-5665 center frequency according to Table 16.
Table 16. Average Noise Level Verification Test Frequencies
Start FrequencyStop FrequencyStep Size (MHz)
10 MHz90 MHz20
100 MHz3.6 GHz500
4 GHz14 GHz500
5.Set the PXIe-5665 channel coupling, RF attenuation, device instantaneous bandwidth,
span, resolution bandwidth, sampling ratio, and number of averages according to Table 17.
Table 17. PXIe-5665 Settings for Average Noise Level Verification
Test Frequency
Device Instantaneous
Bandwidth
Span
OSP Sampling
Ratio
10 MHz to <100 MHz300 kHz100 kHz32
100 MHz to ≤3.6 GHz300 kHz100 kHz32
*
50 MHz
3.8 GHz to 14 GHz50 MHz
*
Set the PXIe-5665 Device Instantaneous Bandwidth to 25 MHz for the PXIe-5665 with
*
5MHz1
5 MHz1
25 MHz bandwidth.
6.Commit the PXIe-5665 settings to hardware.
7.Read the power spectrum from the PXIe-5665. Convert the power spectrum to a power
spectral density (dBm/Hz) value.
8.Remove five points from around the center of the power spectral density from step 7.
9.Convert the power spectral density with the points removed to watts/Hz, take the mean, and
then convert the result back to dBm/Hz. This value is the PXIe-5665 Average Noise (dBm/Hz).
10. Repeat steps 4 to 9 for all frequencies in Table 16.
11. Repeat steps 4 to 10 for frequencies less than or equal to 3.6 GHz with the PXIe-5665
preamplifier enabled.
12.
(PXIe-5665 14 GHz VSA) Repeat steps 4 to 10 for frequencies greater than 3.6 GHz with
the PXIe-5665 preselector enabled and preamplifier disabled.
Note The PXIe-5665 preselector has a maximum instantaneous bandwidth of
47 MHz. For steps that call for a 50 MHz instantaneous bandwidth, use a 40 MHz
instantaneous bandwidth when using the preselector.
13. Compare the PXIe-5665 Average Noise (dBm/Hz) to the verification test limits in Table 18,
Table 19, or Table 20 as appropriate.
Table 18. Average Noise Verification Test Limits (Preamplifier Disabled, PXIe-5665
14 GHz VSA Preselector Disabled)
*
Frequency
As-Found Limit
(dBm/Hz)
As-Left Limit
(dBm/Hz)
>10 MHz to 100 MHz-149-151
>100 MHz to 300 MHz-152-154
>300 MHz to 1.7 GHz-151-153
>1.7 GHz to 2.8 GHz-149-151
>2.8 GHz to 3.6 GHz-148-150
>3.6 GHz to 7.5 GHz
>7.5 GHz to 8.5 GHz
>8.5 GHz to 12 GHz
>12 GHz to 14 GHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
†
†
†
†
-148-150
-146-147
-147-148
-145-146
about as-left limits.
†
Frequencies >3.6 GHz apply only to the PXIe-5665 14 GHz VSA.
Table 19. Average Noise Verification Test Limits (Preamplifier Enabled)
*
Frequency
As-Found Limit
(dBm/Hz)
As-Left Limit
(dBm/Hz)
>10 MHz to 100 MHz-161-163
>100 MHz to 300 MHz-162-165
>300 MHz to 1.7 GHz-162-164
>1.7 GHz to 2.8 GHz-161-163
>2.8 GHz to 3.6 GHz-160-163
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
38 | ni.com | PXIe-5665 Calibration Procedure
Page 39
Table 20. Average Noise Verification Test Limits (Preamplifier Disabled, PXIe-5665
14 GHz VSA Preselector Enabled)
*
Frequency
As-Found Limit
(dBm/Hz)
As-Left Limit
(dBm/Hz)
>3.6 GHz to 7.5 GHz-144-145
>7.5 GHz to 8.5 GHz-140-141
>8.5 GHz to 12 GHz-141-142
>12 GHz to 14 GHz-140-141
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
14. Close the PXIe-5665 session.
If the average noise level verification procedure determines that the PXIe-5665 is outside its
limits, refer to Worldwide Support and Services for information about support resources or
service requests.
4.Set the PXIe-5665 center frequency using the following values:
•Start frequency: 100 MHz
•Stop frequency:
–
(PXIe-5665 3.6 GHz VSA) 3.6 GHz
–(PXIe-5665 14 GHz VSA) 14 GHz
•Step size: 9.997331 MHz
5.Commit the PXIe-5665 settings to hardware.
6.Read the power spectrum from the PXIe-5665.
7.Measure the highest power in the spectrum returned from the PXIe-5665. This value is the
PXIe-5665 Non-Input-Related Spurious Level.
8.Repeat steps 4 to 7 for all frequencies in step 4.
9.Compare the PXIe-5665 Non-Input-Related Spurious Level to the verification test limits in
Table 21.
Table 21. Non-Input-Related Spurs Verification Test Limits
Center FrequencyAs-Found Limit (dBm)As-Left Limit (dBm)
*
100 MHz to <1.650 GHz-95-96
1.650 GHz to 1.750 GHz-85-88
>1.750 GHz to 3.6 GHz-95-96
>3.6 GHz to 7.5 GHz
>7.5 GHz to 8.5 GHz
>8.5 GHz to 14 GHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
†
†
†
-92-94
-90-92
-90-92
about as-left limits.
†
Center frequencies >3.6 GHz apply only to the PXIe-5665 14 GHz VSA.
10. Close the PXIe-5665 session.
If the non-input-related spurs (residual spurs) verification procedure determines that the
PXIe-5665 is outside its limits, refer to Worldwide Support and Services for information about
support resources or service requests.
42 | ni.com | PXIe-5665 Calibration Procedure
Page 43
Verifying LO-Related Spurs (Sideband Spurs)
1.Connect RF source 2 to the power splitter input through the SMA (m)-to-SMA (m) cable.
2.Connect the power splitter reference output to the spectrum analyzer through the
SMA (m)-to-SMA (m) cable.
3.Connect the other power splitter output to the PXIe-5665 RF IN connector using the
3.5 mm (m)-to-3.5 mm (m) adaptor. The completed equipment setup is shown in Figure 15
and Figure 16.
10. Configure the RF Source 2 according to the following settings:
•Center frequency: PXIe-5665 center frequency
•Output power: RF Source Power through Splitter for that frequency
11. Set the spectrum analyzer center frequency to the RF source 2 frequency.
12. Set the PXIe-5665 span and resolution bandwidth according to the first row in Table 24.
Table 24. LO-Related Spurs Verification Bandwidth and Span Settings
PXIe-5665
Span
200 kHz10 Hz20 kHz (1999)600 Hz50 Hz6
2MHz40 Hz200 kHz (4999)300 Hz50 Hz3
20 MHz60 Hz2 MHz (33333)300 Hz50 Hz3
PXIe-5665
RBW
PXIe-5665 Notch BW
(Number of Bins at
Stated RBW)
Spectrum
Analyzer
Span
Spectrum
Analyzer
RBW
Spectrum
Analyzer
Averages
13. Set the spectrum analyzer span, resolution bandwidth, and number of averages according
to the values in Table 24.
14. Commit the PXIe-5665 settings to hardware.
15. Wait 250 ms before making the first measurement and wait 100 ms before making a
subsequent measurements to allow the PXIe-5665 and RF source 2 amplitudes to settle.
Note Settling times are a characteristic of the RF source device. Refer to the Test
Equipment section of this document for a list of NI-recommended source devices and
device specifications. If you use an RF source device with different device
specifications, your settling times may differ from those listed in this procedure.
16. Measure the RF source 2 spectrum with the spectrum analyzer.
17. Perform a peak search on the spectrum from step 16. The value of the peak is the Spectrum Analyzer Tone Power. This value is used in step 19(c).
46 | ni.com | PXIe-5665 Calibration Procedure
Page 47
18. Measure the spectrum on the PXIe-5665.
a.Find the peak value of the spectrum.
b.Confirm that the peak value is between -16 dBm and -4 dBm. The peak value is the
PXIe-5665 Tone Power. This value is used in step 18(d).
c.Notch filter the spectrum by replacing the appropriate number of bins from the center
of the spectrum according to Table 24 with bins of -130 dB in value. The resulting
spectrum is the Notched Spectrum.
d.Calculate the Notched Spectrum Response using the following equation:
Notched Spectrum Response (dB) = Notched Spectrum (dBm) - PXIe-5665 Tone Power (dBm)
e.Create a list of target spurs from the Notched Spectrum Response (dB) that have a
level greater than -90 dB.
f.If any two spurs are within ±500 Hz of each other, retain only the spur with the greater
power magnitude. This reduced list is used in step 19(a).
Note The RF source 2 indicated for use in this procedure exhibits spurious
responses that have a frequency spread of 500 Hz. Any target spur found within
±500 Hz of an RF source 2 spurious response is considered to be from the
RF source 2 because of this frequency spread.
19. Eliminate any RF source 2 spurious responses from the list of target spurs.
a.Tune the spectrum analyzer to the frequency of each target spur in the reduced list
from step 18(f).
b.Measure the spectrum and perform a peak search on the measured spectrum. This
value is the Spectrum Analyzer Peak Value.
c.Convert the Spectrum Analyzer Peak Value units to dB using the following equation:
21. Repeat steps 11 to 19 for all spans in Table 24.
22. Repeat steps 9 to 20 for all center frequencies in Table 22.
(PXIe-5665 14 GHz VSA) Repeat steps 9 to 20 for center frequencies from 3.8 GHz to
23.
14 GHz in 200 MHz steps with the PXIe-5665 preselector enabled.
Note The PXIe-5665 preselector has a maximum instantaneous bandwidth of
47 MHz. For steps that call for a 50 MHz instantaneous bandwidth, use a 40 MHz
instantaneous bandwidth when using the preselector.
24. Close the PXIe-5665 session.
If the LO-related spurs (sideband spurs) verification procedure determines that the PXIe-5665 is
outside its limits, refer to Worldwide Support and Services for information about support
resources or service requests.
Verifying Image Rejection
This procedure verifies the following image responses of the PXIe-5665:
For test frequencies <3.6 GHz:
•IF1 Image = Test Frequency + 9.225 GHz
•IF2 Image = 3.3875 GHz
•IF3 Image = 987.5 MHz
•IF2 Image at RF = Test Frequency + 1.225 GHz
•IF3 Image at RF = |Test Frequency - 375 MHz | (Evaluated only for images greater than
10 MHz)
For test frequencies >3.6 GHz,
•IF1 Image = Test Frequency + 1.225 GHz
•IF2 Image = 987.5 GHz
•IF2 Image at RF = Test Frequency - 375 MHz
where Test Frequency is the PXIe-5665 center frequency.
48 | ni.com | PXIe-5665 Calibration Procedure
Page 49
1.Connect RF source 2 to the PXIe-5665 RF IN using the SMA (m)-to-SMA (m) cable. The
OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD
SENSITIVE
OUT
ALL PORTS
50 Ω
ACCESS
ACTIVE
CLK IN
CLK OUT
PFI 1
ESD
SENSITIVE
TTL
6.3 Vp-p
MAX
2 Vp-p
NOM
+
20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
ESD
SENSITIVE
ACCESS
ACTIVE
OUT
IN
LO3
LO2
LO1
ALL PORTS
50 Ω
NI PXIe-5603
Downconverter 20 Hz - 3.6 GHz
RF IN
0 V DC, DC COUPLED
25 V DC MAX, AC COUPLED
+20 dBm MAX > 10 MHz
0 dBm MAX < 10 MHz
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
IF OUT
+
22 dBm MAX
0 V DC
K
K
J
J
J
J
J
J
1
2
completed equipment setup is shown in Figure 17 and Figure 18.
5.Set the PXIe-5665 center frequency according to Table 26 or Table 27 as appropriate.
Table 26. PXIe-5665 3.6 GHz VSA Image Rejection Verification Test Frequencies
Start FrequencyStop Frequency (GHz)Step Size (MHz)
100.033325 MHz2.000033325400
2.399966675 GHz3.599966675400
Table 27. PXIe-5665 14 GHz VSA Image Rejection Verification Test Frequencies
Start FrequencyStop Frequency (GHz)Step Size (MHz)
100.033325 MHz2.000033325400
2.399966675 GHz13.999966675400
6.Commit the PXIe-5665 settings to hardware.
7.For each test frequency from Table 26 or Table 27, calculate PXIe-5665 image frequencies
using the following formulas:
For test frequencies <3.6 GHz,
•IF1 Image = Test Frequency + 9.225 GHz
•IF2 Image = 3.3875 GHz
•IF3 Image = 987.5 MHz
•IF2 Image at RF = Test Frequency + 1.225 GHz
•IF3 Image at RF = | Test Frequency - 375 MHz | (Evaluated only for images greater
than 10 MHz)
For test frequencies > 3.6 GHz,
•IF1 Image = Test Frequency + 1.225 GHz
•IF2 Image = 987.5 MHz
•IF2 Image at RF = Test Frequency - 375 MHz
where Test Frequency is the PXIe-5665 center frequency.
8.For each image and test frequency in the list from step 7 and the PXIe-5665 center
frequency from step 5, set the RF source 2 to the frequency with a Power Setting of 0 dB.
The RF Source Power Direct for this frequency is used in step 11.
Note RF Source Power Direct value was measured in the Characterizing RF
Source Power (Direct) equipment characterization procedure.
9.Wait for 200 ms to allow the PXIe-5665 and RF source 2 amplitudes to settle.
Note Settling times are a characteristic of the RF source device. Refer to the Test
Equipment section of this document for a list of NI-recommended source devices and
device specifications. If you use an RF source device with different device
specifications, your settling times may differ from those listed in this procedure.
10. Acquire the spectrum and determine the peak power for each frequency. The peak power
value is the Measured Level.
11. For each Measured Level, calculate the Response Correction using the following equation:
Response Correction = Measured Level + RF Source Power Direct
12. Calculate the Image Rejection at each of the image frequencies using the following
equation:
Image Rejection = Response Correction (at test frequency) - Response Correction (at each
image frequency)
13. Repeat steps 5 to 12 for all center frequencies in Table 26 or Table 27.
14. Compare the Image Rejection results to the verification test limits in Table 28.
Table 28. Image Rejection Verification Test Limits
Center FrequencyAs-Found LimitAs-Left Limit
*
100 MHz to 2.2 GHz-80 dBc-83 dBc
>2.2 GHz to 3.6 GHz-77 dBc-80 dBc
>3.6 GHz to 14 GHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
†
-80 dBc-82 dBc
about as-left limits.
†
Center frequencies >3.6 GHz apply only to the PXIe-5665 14 GHz VSA.
15. Close the PXIe-5665 session.
If the image rejection verification procedure determines that the PXIe-5665 is outside its limits,
refer to Worldwide Support and Services for information about support resources or service
requests.
52 | ni.com | PXIe-5665 Calibration Procedure
Page 53
Verifying Third-Order Intermodulation Distortion
OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD
SENSITIVE
OUT
ALL PORTS
50 Ω
ACCESS ACTIVE
CLK IN
CLK OUT
PFI 1
ESD
SENSITIVE
TTL
6.3 Vp-pMAX
2 Vp-p
NOM
+
20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
ESD
SENSITIVE
ACCESS
ACTIVE
OUT
IN
LO3
LO2
LO1
ALL PORTS
50 Ω
NI PXIe-5603
Downconverter 20 Hz - 3.6 GHz
RF IN
0 V DC, DC COUPLED
25 V DC MAX, AC COUPLED
+20 dBm MAX > 10 MHz
0 dBm MAX < 10 MHz
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
IF OUT
+
22 dBm MAX
0 V DC
J
J
J
J
J
J
K
K
RF
OUTPUT
RF
INPUT 1
NI-5665 ANTI-DISTORTION ET
166375A
RF
INPUT 2
6
1
5
2
4
3
Frequencies ≤700 MHz
1.Connect RF source 1 and RF source 2 to the PXIe-5665 RF IN connector through the
anti-distortion test fixture as shown in Figure 19 and Figure 20.
Figure 19. PXIe-5665 3.6 GHz VSA Third-Order Intermodulation Verification
15. Wait 2.5 s before making the first measurement and wait 100 ms before making subsequent
measurements to allow the PXIe-5665, RF source 1, and RF source 2 amplitudes to settle.
Note Settling times are a characteristic of the RF source device. Refer to the Test
Equipment section of this document for a list of NI-recommended source devices and
device specifications. If you use an RF source device with different device
specifications, your settling times may differ from those listed in this procedure.
16. Measure the power spectrum with the PXIe-5665.
17. Find the highest peak within the measurement bandwidth. This value is the IMD Low Power.
18. Repeat steps 11 to 16 for the PXIe-5665 center frequencies of P1 Frequency,
P2 Frequency, and IMD High Frequency. The values from step 16 are IMD Low Power, P1
Power, P2 Power, and IMD High Power, respectively.
19. Calculate the upper and lower third-order intercept (TOI) points using the following
equations:
TOI Upper = P1 Power + (P2 Power - IMD High Power)/2
TOI Lower = P2 Power + (P1 Power - IMD Low Power)/2
20. The smaller of the TOI Lower and TOI Upper values is the TOI Minimum.
21. Repeat steps 8 to 19 for all test frequencies in Table 29.
22. Repeat steps 5 to 20 with the PXIe-5665 preamplifier enabled.
23. Compare the TOI Minimum to the verification test limits in Table 30 or Table 31 as
appropriate.
Table 30. TOI (Frequencies ≤700 MHz) Verification Test Limits (Preamplifier Disabled)
Center FrequencyAs-Found Limit (dBm)As-Left Limit (dBm)
10 MHz to <100 MHz+16+17
100 MHz to 700 MHz+19+20
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
Table 31. TOI (Frequencies ≤ 700 MHz) Verification Test Limits (Preamplifier Enabled)
Center FrequencyAs-Found Limit (dBm)As-Left Limit (dBm)
10 MHz to <100 MHz-3-2
100 MHz to 700 MHz+2+3
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
56 | ni.com | PXIe-5665 Calibration Procedure
*
*
Page 57
24. Disable the output of the RF source 1 and RF source 2.
OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD
SENSITIVE
OUT
ALL PORTS
50 Ω
ACCESS ACTIVE
CLK IN
CLK OUT
PFI 1
ESD
SENSITIVE
TTL
6.3 Vp-pMAX
2 Vp-p
NOM
+
20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
ESD
SENSITIVE
ACCESS
ACTIVE
OUT
IN
LO3
LO2
LO1
ALL PORTS
50 Ω
NI PXIe-5603
Downconverter 20 Hz - 3.6 GHz
RF IN
0 V DC, DC COUPLED
25 V DC MAX, AC COUPLED
+20 dBm MAX > 10 MHz
0 dBm MAX < 10 MHz
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
IF OUT
+
22 dBm MAX
0 V DC
J
J
J
J
J
J
K
K
RF
OUTPUT
RF
INPUT 1
NI-5665 ANTI-DISTORTION ET
166375A
RF
INPUT 2
6
1
5
2
4
3
Note When disabled, the RF source 1 and RF source 2 output signals should be less
than -60 dBm.
25. Close the PXIe-5665 session.
If the third-order intermodulation distortion for frequencies ≤700 MHz verification procedure
determines that the PXIe-5665 is outside its limits, refer to Worldwide Support and Services for
information about support resources or service requests.
Frequencies >700 MHz
1.Connect RF source 1 and RF source 2 to the PXIe-5665 RF IN connector through the
anti-distortion test fixture as shown in Figure 21 and Figure 22.
Table 33. PXIe-5665 14 GHz VSA TOI (Frequencies >700 MHz)
Verification Test Frequencies
Stop Frequency
Preamp EnabledStart Frequency
(GHz)
Step Size (MHz)
Enabled800 MHz3.3500
Disabled800 MHz3.3500
3.6 GHz13.6500
13.99895 GHz——
10. Set the RF source 1 frequency to P1 Frequency.
11. Set the RF source 2 frequency to P2 Frequency.
12. Set the PXIe-5665 center frequency to IMD Low Frequency.
13. Commit the PXIe-5665 settings to hardware.
14. Enable the RF source 1 and RF source 2 outputs.
15. Wait 2.5 s before making the first measurement and wait 100 ms before making subsequent
measurements to allow the PXIe-5665, RF source 1, and RF source 1 amplitudes to settle.
Note Settling times are a characteristic of the RF source device. Refer to the Test
Equipment section of this document for a list of NI-recommended source devices and
device specifications. If you use an RF source device with different device
specifications, your settling times may differ from those listed in this procedure.
16. Measure the power spectrum with the PXIe-5665.
17. Find the highest peak within the measurement bandwidth. This value is the IMD Low Power.
18. Repeat steps 11 to step 16 for the PXIe-5665 center frequencies of P1 Frequency,
P2 Frequency, and IMD High Frequency. The values from step 16 are IMD Low Power,
P1 Power, P2 Power, and IMD High Power, respectively.
19. Calculate the upper and lower TOI points using the following equations:
TOI Upper= P1 Power + (P2 Power - IMD High Power)/2
TOI Lower= P2 Power + (P1 Power - IMD Low Power)/2
The smaller value of the TOI Lower and TOI Upper is the TOI Minimum.
20. Repeat steps 8 to 18 for all test frequencies in Table 32.
21. Repeat steps 5 to 19 with the PXIe-5665 preamplifier enabled for frequencies greater than
or equal to 3.6 GHz.
60 | ni.com | PXIe-5665 Calibration Procedure
Page 61
22. Compare the TOI Minimum to the verification test limits in Table 34 or Table 35 as
appropriate.
Table 34. TOI (Frequencies >700 MHz) Verification Test Limits (Preamplifier Disabled,
PXIe-5665 14 GHz VSA Preselector Disabled)
Center FrequencyAs-Found Limit (dBm)As-Left Limit (dBm)
*
>700 MHz to 3.6 GHz+20+21
>3.6 GHz to 14 GHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
†
+20+21
about as-left limits.
†
Center frequencies >3.6 GHz apply only to the PXIe-5665 14 GHz VSA.
Table 35. TOI (Frequencies >700 MHz) Verification Test Limits (Preamplifier Enabled)
Center FrequencyAs-Found LimitAs-Left Limit
*
>700 MHz to 3.6 GHz+2.5 dBm+3.5 dBm
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
23. Disable the output of the RF source 1 and RF source 2.
Note When disabled, the RF source 1 and RF source 2 output signals should be less
than -60 dBm.
24. Close the PXIe-5665 session.
If the third-order intermodulation distortion for frequencies >700 MHz verification procedure
determines that the PXIe-5665 is outside its limits, refer to Worldwide Support and Services for
information about support resources or service requests.
13. Wait 500 ms before making the first measurement and wait 100 ms before making
subsequent measurements to allow the PXIe-5665 and RF source 2 amplitudes to settle.
Note Settling times are a characteristic of the RF source device. Refer to the Test
Equipment section of this document for a list of NI-recommended source devices and
device specifications. If you use an RF source device with different device
specifications, your settling times may differ from those listed in this procedure.
14. Measure the power spectrum with the PXIe-5665.
15. Find the highest peak within the measurement bandwidth (BW). This value is the Peak Power.
16. Calculate the Noise Measurement BW and the Noise Center Frequency (CF) using the
following equations:
17. Calculate the Power in Band for the spectrum from step 13 using the Noise Measurement BW and Noise CF values from step 16.
Note Power in Band is a built-in function in the NI Spectral Measurements Toolkit.
18. Calculate the Average Noise Power using the following equation:
19. Record either Peak Power (from step 14) or Average Noise Power (from step 17).
According to the frequency tuned to with the PXIe-5665, record whichever value is greater
as the Fundamental Power or the Harmonic Power.
20. Repeat steps 7 to 19 for the second harmonic, which is twice the test frequency.
21. Calculate the SHI using the following equation:
SHI = 2(Fundamental Power) - Harmonic Power - RF Attenuation
22. Repeat steps 7 to 20 for all test frequencies listed in Table 36.
23. (PXIe-5665 14 GHz VSA) Repeat steps 6 to 21 for test frequencies between 3.8 GHz and
7 GHz with the PXIe-5665 preselector enabled.
24. Repeat steps 6 to 22 for the frequencies in Table 38 with the PXIe-5665 preamplifier
enabled.
Table 38. SHI Verification Test Frequencies
Start Frequency (MHz)Stop FrequencyStep Size
300.03331251.3000333125 GHz200 MHz
1,800——
25. Compare the SHI to the verification test limits in Table 39, Table 40, or Table 41 as
appropriate.
Table 39. SHI Verification Test Limits (Preamplifier Disabled,
PXIe-5665 14 GHz VSA Preselector Enabled)
Source
Frequency
300 MHz to
Device
PXIe-5665+42+43
As-Found Limit
(dBm)
As-Left Limit
*
(dBm)
<700 MHz
700 MHz to
1.8 GHz
PXIe-5665
3.6 GHz VSA
PXIe-5665
+50+51
+44+45
14 GHz VSA
>1.8 GHz to
3.75 GHz
†
>3.75 GHz to
4.25 GHz
†
>4.25 GHz to
†
7 GHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
PXIe-5665
14 GHz VSA
+54+56
+54+56
+54+56
about as-left limits.
†
Source frequencies >1.8 GHz apply only to the PXIe-5665 14 GHz VSA.
66 | ni.com | PXIe-5665 Calibration Procedure
Page 67
Table 40. SHI Verification Test Limits (Preamplifier Enabled)
Source
Frequency
300 MHz to
1.8 GHz
Device
PXIe-5665
3.6 GHz VSA
PXIe-5665
As-Found Limit
(dBm)
+15+18
+20+25
As-Left Limit
*
(dBm)
14 GHz VSA
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
Table 41. SHI Verification Test Limits (PXIe-5665 14 GHz VSA Preselector Disabled)
Frequency (GHz)As-Found Limit (dBm)As-Left Limit (dBm)
>1.80 to 4.25+28+30
>4.25 to 7.00+18+20
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
26. Disable the output of RF source 2.
Note When disabled, the RF source 2 output signal should be less than -60 dBm.
27. Close the PXIe-5665 session.
*
If the second harmonic intercept verification procedure determines that the PXIe-5665 is outside
its limits, refer to Worldwide Support and Services for information about support resources or
service requests.
Verifying Gain Compression
This procedure verifies that <1 dB of gain compression occurs when the specified power level
is applied at the RF IN connector of the PXIe-5665.
Note The power level you apply differs whether you are verifying the as-found or
5.Disable the RF source 1 and RF source 2 outputs.
Note When disabled, the RF source 1 and RF source 2 output signals should be less
than -60 dBm.
6.Disable the PXIe-5665 preamplifier.
7.Set the RF source 2 frequency according to Table 42.
8.Set the RF source 1 frequency 1 MHz above the RF source 2 frequency.
9.Set the PXIe-5665 center frequency according to the first row in Table 42.
Table 42. Gain Compression Verification Test Frequencies (Frequencies ≤700 MHz)
Start Frequency (MHz)Stop Frequency (MHz)Step Size (MHz)
109040
100700200
10. Set the RF source 2 power to the RF Source 2 Programmed Power in Table 43 or Table 44
as appropriate.
11. Set the RF source 1 power to the RF Source 1 Programmed Power in Table 43 or Table 44
as appropriate.
12. Set the PXIe-5665 reference level according to Table 43 or Table 44 as appropriate.
Table 43. Gain Compression Verification Test Settings (Preamplifier Disabled)
SettingFrequency Range (MHz)LimitValue ( dBm )
RF Source 1
Programmed Power
RF Source 2
11 to 701As-Found+8.0
*
As-Left
+9.1
10 to 700Both-24
Programmed Power
PXIe-5665 Reference
10 to 700Both0
Level
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
70 | ni.com | PXIe-5665 Calibration Procedure
Page 71
Table 44. Gain Compression Test Settings (Preamplifier Enabled)
SettingFrequency Range (MHz)LimitValu e (dBm)
RF Source 1
Programmed Power
11 to 101As-Found-18
*
As-Left
-16
>101 to 701As-Found-15
*
RF Source 2
As-Left
10 to 700Both-54
-13
Programmed Power
PXIe-5665 Reference
10 to 700Both-30
Level
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
13. Commit the PXIe-5665 settings to hardware.
14. Enable the RF source 2 output and wait 250 ms, or wait 100 ms if the output is already
enabled. This wait time allows the PXIe-5665 and RF source 2 amplitudes to settle.
Note Settling times are a characteristic of the RF source device. Refer to the Test
Equipment section of this document for a list of NI-recommended source devices and
device specifications. If you use an RF source device with different device
specifications, your settling times may differ from those listed in this procedure.
15. Read the power spectrum from the PXIe-5665.
16. Depending on your equipment configuration, using the spectrum returned from the
PXIe-5665, measure the power at the center frequency and record this value as the Small
Tone First Measurement (RF source 1 output disabled) or as the Small Tone Second
Measurement (RF source 1 output enabled).
17. Repeat steps 7 to 16 for all test frequencies in Table 42.
18. Repeat steps 7 to 17 with the PXIe-5665 preamplifier enabled.
19. Enable the RF source 1 output.
20. Repeat steps 6 to 18 with the RF source 1 output enabled.
21. Calculate the PXIe-5665 Gain Compression using the following equation:
Gain Compression = Small Tone First Measurement - Small Tone Second Measurement
22. Compare the Gain Compression to the verification test limits in Table 45 or Table 46 as
appropriate.
Table 45. Gain Compression Verification Test Limits
(Frequencies
≤700 MHz, Preamplifier Disabled)
Compression
FrequencyLimit
10 MHz to
700 MHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
As-Found1+8
*
As-Left
(dB)
1+9.1
Applied Power
(dBm)
about as-left limits.
The device specification calls for less than or equal to 1 dB gain compression for a minimum
applied power. The applied power includes the measurement uncertainty to guarantee the
required minimum applied power.
Table 46. Gain Compression Verification Test Limits
(Frequencies
FrequencyLimit
10 MHz to
100 MHz
>100 MHz to
700 MHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
≤700 MHz, Preamplifier Enabled)
Compression
(dB)
Applied Power
(dBm)
As-Found1-18
*
As-Left
1-16
As-Found1-15
*
As-Left
1-13
about as-left limits.
The device specification calls for less than or equal to 1 dB gain compression for a minimum
applied power. The applied power includes the measurement uncertainty to guarantee the
required minimum applied power.
23. Disable the output of RF source 1 and RF source 2.
Note When disabled, the RF source 1 and RF source 2 output signals should be less
than -60 dBm.
24. Close the PXIe-5665 session.
If the gain compression for frequencies ≤700 MHz verification procedure determines that the
PXIe-5665 is outside its limits, refer to Worldwide Support and Services for information about
support resources or service requests.
72 | ni.com | PXIe-5665 Calibration Procedure
Page 73
Frequencies >700 MHz
OUT
NI PXIe-5653
Synthesizer
REF IN
10 MHz
5 V p-p MAX
REF OUT
10 MHz
1.5 V p-p MAX
REF OUT
100 MHz
1.5 V p-p MAX
LO3
800 MHz
LO2
4 GHz
LO1
3.2 GHz - 8.3 GHz
ESD
SENSITIVE
OUT
ALL PORTS
50 Ω
ACCESS ACTIVE
CLK IN
CLK OUT
PFI 1
ESD
SENSITIVE
TTL
6.3 Vp-pMAX
2 Vp-p
NOM
+
20 dBm MAX
50 Ω
50 Ω
50 Ω
NI PXIe-5622
16-Bit IF Digitizer
IF IN
ESD
SENSITIVE
ACCESS
ACTIVE
OUT
IN
LO3
LO2
LO1
ALL PORTS
50 Ω
NI PXIe-5603
Downconverter 20 Hz - 3.6 GHz
RF IN
0 V DC, DC COUPLED
25 V DC MAX, AC COUPLED
+20 dBm MAX > 10 MHz
0 dBm MAX < 10 MHz
+
15 dBm MAX
25 V DC MAX
+
15 dBm MAX
0 V DC
IF OUT
+
22 dBm MAX
0 V DC
J
J
J
J
J
J
K
K
RF
OUTPUT
RF
INPUT 1
NI-5665 ANTI-DISTORTION ET
166375A
RF
INPUT 2
6
1
5
2
4
3
1.Connect RF source 1 and RF source 2 to the PXIe-5665 RF IN connector through the
anti-distortion test fixture, as shown in Figure 27 and Figure 28.
Figure 27. PXIe-5665 3.6 GHz VSA Gain Compression Verification
13. Set the PXIe-5665 reference level according to Table 49 or Table 50 as appropriate.
Table 49. Gain Compression Verification Test Settings
(Preamplifier Disabled)
Frequency
DeviceSetting
Range
LimitValue (dB m)
PXIe-5665
3.6 GHz VSA
RF Source 1
Programmed
Power
RF Source 2
Programmed
801 MHz to
1.701 GHz
>1.701 GHz to
3.601 GHz
800 MHz to
3.600 GHz
As-Found+8.0
*
As-Left
+9.1
As-Found+6.0
*
As-Left
+7.1
Both-24
Power
PXIe-5665
Reference
800 MHz to
3.600 GHz
Both0
Level
PXIe-5665
14 GHz VSA
RF Source 1
Programmed
Power
RF Source 2
Programmed
801 MHz to
1.701 GHz
>1.701 GHz to
14.001 GHz
800 MHz to
14.00 GHz
As-Found+8.0
*
As-Left
+9.1
As-Found+6.0
*
As-Left
+7.1
Both-24
Power
PXIe-5665
Reference
800 MHz to
14.00 GHz
Both0
Level
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
76 | ni.com | PXIe-5665 Calibration Procedure
Page 77
Table 50. Gain Compression Verification Test Settings
DeviceSetting
(Preamplifier Enabled)
Frequency
Range
LimitValue (dB m)
PXIe-5665
3.6 GHz VSA
RF Source 1
Programmed
Power
RF Source 2
Programmed
801 MHz to
1.701 GHz
>1.701 GHz to
3.601 GHz
800 MHz to
3.600 GHz
As-Found-15
*
As-Left
-13
As-Found-18
*
As-Left
-16
Both-54
Power
PXIe-5665
Reference
800 MHz to
3.600 GHz
Both-30
Level
PXIe-5665
14 GHz VSA
RF Source 1
Programmed
Power
RF Source 2
Programmed
801 MHz to
1.701 GHz
>1.701 GHz to
14.001 GHz
800 MHz to
14.00 GHz
As-Found-15
*
As-Left
-13
As-Found-18
*
As-Left
-16
Both-54
Power
PXIe-5665
Reference
80 MHz to
14.00 GHz
Both-30
Level
*
Refer to the As-Found and As-Left Limits section of this document for more information
about as-left limits.
14. Commit the PXIe-5665 settings to hardware.
15. Enable the RF source 2 output and wait 250 ms, or wait 100 ms if the output is already
enabled. This wait time allows the PXIe-5665 and RF source 2 amplitudes to settle.
Note Settling times are a characteristic of the RF source device. Refer to the Test
Equipment section of this document for a list of NI-recommended source devices and
device specifications. If you use an RF source device with different device
specifications, your settling times may differ from those listed in this procedure.
17. Depending on your equipment configuration, using the spectrum returned from the
PXIe-5665, measure the power at the center frequency and record this value as the Small
Tone First Measurement (RF source 1 output disabled) or as the Small Tone Second
Measurement (RF source 1 output enabled).
18. Repeat steps 8 to 17 for all test frequencies in step 8.
19. (PXIe-5665 14 GHz VSA) Repeat steps 8 to 18 for frequencies greater than 3.6 GHz with
the PXIe-5665 preselector enabled.
20. Repeat steps 7 to 18 with the PXIe-5665 preamplifier enabled for frequencies less than
3.6 GHz.
21. Enable the RF source 1 output.
22. Repeat steps 6 to 20 with the RF source 1 output enabled.
23. Calculate the PXIe-5665 Gain Compression using the following equation:
Gain Compression = Small Tone First Measurement - Small Tone Second Measurement
24. Compare the Gain Compression to the verification test limits in Table 51 or Table 52.
Table 51. Gain Compression Verification Test Limits (>700 MHz, Preamplifier Disabled)
Compression
FrequencyLimit
>700 MHz to
1.70 GHz
>1.70 GHz to
3.6 GHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
As-Found1+8.0
*
As-Left
As-Found1+6.0
*
As-Left
(dB)
1+9.1
1+7.1
Applied Power
(dBm)
about as-left limits.
The device specification calls for less than or equal to 1 dB gain compression for a minimum
applied power. The applied power includes the measurement uncertainty to guarantee the
required minimum applied power.
78 | ni.com | PXIe-5665 Calibration Procedure
Page 79
Table 52. Gain Compression Verification Test Limits (>700 MHz, Preamplifier Enabled)
Compression
FrequencyLimit
>700 MHz to
1.70 GHz
>1.70 GHz to
3.6 GHz
*
Refer to the As-Found and As-Left Limits section of this document for more information
As-Found1-15
*
As-Left
As-Found1-18
*
As-Left
(dB)
1-13
1-16
Applied Power
(dBm)
about as-left limits.
The device specification calls less than or equal to 1 dB gain compression for a minimum
applied power. The applied power includes the measurement uncertainty to guarantee the
required minimum applied power.
25. Disable the output of RF source 1 and RF source 2.
Note When disabled, the RF source 1 and RF source 2 output signals should be less
than -60 dBm.
26. Close the PXIe-5665 session.
If the gain compression for frequencies >700 MHz verification procedure determines that the
PXIe-5665 is outside its limits, refer to Worldwide Support and Services for information about
support resources or service requests.
Verifying LO Output Power
Note Prior to starting this procedure, zero and calibrate the power sensor according
to the Zeroing and Calibrating the Power Sensor section of this document.
1.Connect the RF source 2 output to the PXIe-5605 RF IN connector using the SMA (m)-toSMA (m) cable.
2.Connect power sensor A to the PXIe-5605 LO3 OUT connector using the SMA (m)-toSMA (m) cable.
3.Connect the spectrum analyzer to the PXIe-5605 LO2 OUT connector.
18. Measure the LO3 power with power meter channel A. Correct the reading for the LO3
frequency of 800 MHz.
19. Compare the readings from steps 16 to 18 to the following verification test limits in
Table 53.
Table 53. LO Output Power Verification Test Limits
LOAs-Found (dBm)As-Left (dBm)
LO1>7.07.0 to 8.0
LO2>7.09.0 to 10.0
LO3>7.09.0 to 10.0
20. Close the PXIe-5653 session.
If the LO output power verification procedure determines that the PXIe-5665 is outside its limits,
refer to Worldwide Support and Services for information about support resources or service
requests.
1
The Spectrum Analyzer Correction Factor value was calculated in the Characterizing Spectrum Analyzer
Response equipment characterization procedure.
82 | ni.com | PXIe-5665 Calibration Procedure
Page 83
Reverification
Repeat the Verification section to determine the as-left status of the device.
Appendix A: Anti-Distortion Test Fixture
Note The anti-distortion test fixture specifications are for the entire fixture and
include the performance of all components internal to the fixture.
Table 54. Lowpass Filter Passband Paths
Filter Path (MHz)Start (MHz)Stop (MHz)
47010≤4702.51.75
735>470≤7352.51.75
1,150>735≤1,1502.51.75
1,800>1,150≤1,8002.51.75
2,530>1,800≤2,5302.51.75
3,550>2,530≤3,5502.51.75
4,985>3,550≤4,9852.51.75
7,000>4,985≤7,0002.51.75
*
Input or output ports.
Insertion Loss
(dB, max.)
VSWR* (max.)
Table 55. Lowpass Filter Stopband Paths
Filter Path (MHz)Start (MHz)Stop (MHz)Rejection
470600≤94050
735>940≤1,47050
1,150>1,470≤2,30050
1,800>2,300≤3,60050
*
2,530>3,600≤5,06070
3,550>5,060≤7,10070
4,985>7,100≤9,97070
7,000>9,970≤14,00070
*
Rejection equals Fundamental Power - Second Harmonic Power
Total Loss includes the splitter loss at either input or output ports. The unused input port is terminated
in 50 Ω.
†
Isolation is the loss between input ports. The output port is terminated in 50 Ω.
‡
VSWR includes any port. All other ports are terminated in 50 Ω.
Loss
(dB, max.)
Isolation
(dB, min.)
†
VSWR
(max.)
Appendix B: Power Sensor Calibration Factor
Uncertainty
Table 57. Maximum Power Sensor Calibration Factor Uncertainty
FrequencyCalibration Factor Uncertainty (%)
10 MHz≤1.15
‡
20 MHz≤0.80
30 MHz≤0.65
40 MHz to 200 MHz≤0.55
300 MHz to 400 MHz≤0.60
500 MHz to 1 GHz≤0.50
2GHz to 4GHz≤0.60
5 GHz≤0.75
6 GHz≤0.80
7 GHz≤0.85
8 GHz≤0.90
9 GHz≤0.95
10 GHz≤1.00
84 | ni.com | PXIe-5665 Calibration Procedure
Page 85
Table 57. Maximum Power Sensor Calibration Factor Uncertainty (Continued)
FrequencyCalibration Factor Uncertainty (%)
11 GHz≤1.05
12 GHz≤1.10
13 GHz≤1.15
14 GHz≤1.20
15 GHz≤1.25
Worldwide Support and Services
The National Instruments website is your complete resource for technical support. At ni.com/
support
self-help resources to email and phone assistance from NI Application Engineers.
Vis it
other services.
Vis it
facilitates technical support and ensures that you receive important information updates from NI.
A Declaration of Conformity (DoC) is our claim of compliance with the Council of the European
Communities using the manufacturer’s declaration of conformity. This system affords the user
protection for electromagnetic compatibility (EMC) and product safety. You can obtain the DoC
for your product by visiting
you can obtain the calibration certificate for your product at
you have access to everything from troubleshooting and application development
ni.com/services for NI Factory Installation Services, repairs, extended warranty, and
ni.com/register to register your National Instruments product. Product registration
ni.com/certification. If your product supports calibration,
ni.com/calibration.
National Instruments corporate headquarters is located at 11500 North Mopac Expressway,
Austin, Texas, 78759-3504. National Instruments also has offices located around the world. For
telephone support in the United States, create your service request at
ni.com/support or dial
512 795 8248. For telephone support outside the United States, visit the Worldwide Offices
section of
ni.com/niglobal to access the branch office websites, which provide up-to-date
contact information, support phone numbers, email addresses, and current events.
Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for more information on National Instruments trademarks. Other
product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National
Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the
National Instruments Patents Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party
legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance
for the National Instruments global trade compliance policy and how to obtain re levant HTS codes, ECCNs, and other import/export data. NI MAKES
NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR
ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable
limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.