250 kHz to 20 GHz RF Analog Signal Generator with Amplitude
Extender
This document contains the verification and adjustment procedures for the PXIe-5654 RF
Analog Signal Generator when used with the PXIe-5696 Amplitude Extender.
Refer to ni.com/calibration for more information about calibration solutions.
Updating Calibration Date and Time...................................................................................... 37
Worldwide Support and Services............................................................................................ 37
Page 2
Required Software
Calibrating the PXIe-5654 requires you to install the following package on the calibration
system:
•NI-RFSG 14.5 or later
You can download all required software from ni.com/downloads.
The software supports programming the verification procedures in the LabVIEW, C, and
LabWindows™/CVI™ application development environments (ADE) and programming the
adjustment procedures in LabVIEW. When you install the software, you need to install support
only for the ADE that you intend to use.
Related Documentation
For additional information, refer to the following documents as you perform the calibration
procedure:
•PXIe-5654 Getting Started Guide
•PXIe-5654 Calibration Procedure
•PXIe-5696 Getting Started Guide
•PXIe-5654 Specifications (includes specifications for the PXIe-5696)
•NI RF Signal Generators Help
Visit ni.com/manuals for the latest versions of these documents.
Test Equipment
The following table lists the equipment NI recommends for the performance verification and
adjustment procedures. If the recommended equipment is not available, select a substitute
using the minimum requirements listed in the following table.
2 | ni.com | PXIe-5654 with PXIe-5696 Calibration Procedure
Page 3
Table 1. Recommended Equipment for PXIe-5654 Calibration
EquipmentRecommended
Model
Where UsedMinimum
Requirements
Signal source
analyzer
Frequency
reference
Rhode & Schwarz
(R&S) FSUP26
Signal Source
Analyzer
Table 1. Recommended Equipment for PXIe-5654 Calibration (Continued)
EquipmentRecommended
Model
Where UsedMinimum
Requirements
Power meterAnritsu ML2438A
with a SC7413A
Power Sensor
K(m)-to-K(m)
cable, 36 in.
(x2)
Florida RF Labs
KMS-160-36.0KMS
Verifications:
•RF OUT maximum
power
•RF OUT amplitude
accuracy
Adjustments:
•RF OUT power
•RF attenuator
accuracy
•Amplitude accuracy
(low harmonic paths)
•Amplitude accuracy
(high power paths)
•Automatic Level
Control (ALC)
Verifications:
•RF OUT frequency
accuracy
•RF OUT phase noise
•RF OUT amplitude
accuracy
Adjustments:
•Frequency reference
accuracy
Range: -40 dBm to
20 dBm
Frequency range:
250 kHz to 20 GHz
Accuracy: <±4.0%
Length: 36 in.
Loss: <0.7 dB/ft.
(typical) at 20 GHz
Impedance: 50 Ω
K(m)-to-K(m)
cable, 12 in.
(x2)
Florida RF Labs
KMS-160-12.0KMS
Verifications:
•RF OUT amplitude
accuracy
•RF OUT maximum
power
Adjustments:
•RF attenuator path
4 | ni.com | PXIe-5654 with PXIe-5696 Calibration Procedure
Length: 12 in.
Loss: <0.5 dB/ft.
(typical) at 20 GHz
Impedance: 50 Ω
Page 5
Table 1. Recommended Equipment for PXIe-5654 Calibration (Continued)
EquipmentRecommended
Model
Where UsedMinimum
Requirements
K(m)-to-K(f)
Anritsu 41KB-20Adjustments:
20 dB
attenuator
K(m)-to-K(f)
Anritsu 41KB-6Test system
6 dB attenuator
(x2)
Power splitterAeroflex/Weinschel
1593
•RF attenuator path
characterization
Verifications:
•RF OUT amplitude
accuracy
•RF OUT maximum
power
Test system
characterization
Verifications:
•RF OUT amplitude
accuracy
•RF OUT maximum
power
Frequency range: DC to
20 GHz
Voltage standing wave
ratio (VSWR): <1.18
Accuracy: <±0.5 dB
Impedance: 50 Ω
Frequency range: DC to
20 GHz
VSWR: <1.18
Accuracy: <±0.5 dB
Impedance: 50 Ω
Frequency range: DC to
20 GHz
VSWR: <1.35
Insertion loss: <8.5 dB
Impedance: 50 Ω
K(m)-to-K(f)
1
adapter
Anritsu K224BTest system
characterization
Frequency range: DC to
20 GHz
VSWR: ≤1.12
Impedance: 50 Ω
The following table lists the equipment NI recommends for optional performance verification
procedures for non-warranted specifications. If the recommended equipment is not available,
select a substitute using the minimum requirements listed in the table.
1
The connector on the power sensor listed above may be difficult to access with a torque wrench
when there is adjacent connected hardware. Use this adapter if you need to extend the capabilities
of the power sensor to make it accessible by a torque wrench.
Table 2. Recommended Equipment for Optional PXIe-5654 Calibration
EquipmentRecommended ModelWhere UsedMinimum
Requirements
USB digital outputUSB-6501 Digital I/O
Device
BNC(m)-to-SMB(f)
cable
BNC(m)-to-screw
terminal block
adapter
BNC(m)-toBNC(m) cable
BNC Tee adapter
(f-f-f)
Radiall
R284C0351028
Clever Little Box
CLB-JL73
Pasternack PE3087Verifications:
Pasternack PE9003Verifications:
Verifications:
•PULSE IN
operation
Verifications:
•PULSE IN
operation
Verifications:
•PULSE IN
operation
•PULSE IN
operation
•PULSE IN
operation
Active drive
capability
Voltage:
<3.3 V CMOS
Impedance: 50 Ω
Impedance: 50 Ω
Impedance: 50 Ω
Impedance: 50 Ω
Test Conditions
The following setup and environmental conditions are required to ensure the PXIe-5654 and
PXIe-5696 meet published specifications.
•Keep cabling as short as possible. Long cables act as antennas, picking up extra noise that
can affect measurements.
•Verify that all connections to the PXIe-5654 and PXIe-5696, including front panel
connections and screws, are secure.
•Maintain an ambient temperature of 23 °C ± 5 °C.
•Keep relative humidity between 10% and 90%, noncondensing.
•Allow a warm-up time of at least 30 minutes after the chassis is powered on. The warmup time ensures that the PXIe-5654 and PXIe-5696 are at a stable operating temperature.
•In each verification procedure, insert a delay between configuring all instruments and
acquiring the measurement. This delay may need to be adjusted depending on the
instruments used but should always be at least 1,000 ms for the first iteration, 1,000 ms
when the power level changes, and 100 ms for each other iteration.
6 | ni.com | PXIe-5654 with PXIe-5696 Calibration Procedure
Page 7
•Plug the chassis and the calibration instrument(s) into the same power strip to avoid
ground loops.
•Use an appropriate torque wrench to tighten all module RF connectors (SMA, 3.5 mm, or
K). NI recommends a 0.565 N · m (5 lb · in) wrench for SMA connectors and a
0.90 N · m (8 lb · in) wrench for 3.5 mm or K connectors.
•Connect the frequency reference source to the signal source analyzer REF IN back panel
connector using a BNC(m)-to-BNC(m) cable, and connect the signal source analyzer
REF OUT connector to the PXIe-5654 REF IN connector using a BNC(m)-to-SMA(m)
cable.
•Ensure that the PXI chassis fan speed is set to HIGH, that the fan filters, if present, are
clean, and that the empty slots contain filler panels. For more information about cooling,
refer to the Maintain Forced-Air Cooling Note to Users document available at ni.com/
manuals.
•Frequencies less than and equal to 10 GHz apply to the 10 GHz and 20 GHz PXIe-5654;
frequencies above 10 GHz apply only to the 20 GHz PXIe-5654.
•The PXIe-5654 must be verified as a stand-alone device using the PXIe-5654 CalibrationProcedure before performing the procedures in this document.
Initial Setup
Refer to the PXIe-5654 Getting Started Guide and PXIe-5696 Getting Started Guide for
information about how to install the software and the hardware and how to configure the
devices in MAX.
Test System Characterization
The following procedures characterize the test equipment used during verification.
Note Before starting characterization, unassociate the PXIe-5654 and PXIe-5696
in Measurement & Automation Explorer (MAX) by setting the PXIe-5696
connected to the PXIe-5654 to None.
Zeroing and Calibrating the Power Sensor
1.Connect channel A of the power meter to the power sensor.
2.Zero and calibrate the power sensor using the built-in functions in the power meter.
Characterizing the Power Splitter
Several procedures in this document require using a splitter that has been characterized to
remove error from future measurements. Complete the following steps to characterize a
splitter using a PXIe-5654, power meter, and signal source analyzer, as shown in the following
figure.
1.Connect the frequency reference source to the signal source analyzer REF IN back panel
connector using a BNC(m)-to-BNC(m) cable, and connect the signal source analyzer
REF OUT connector to the PXIe-5654 REF IN connector using a BNC(m)-to-SMA(m)
cable.
2.Connect the power meter power sensor to the PXIe-5654 RF OUT front panel connector.
3.Create a new device session for the PXIe-5654.
4.Create a list of test frequencies using the values in the following table.
Table 3. Power Splitter Characterization Test Frequencies
Start Frequency (MHz)Stop Frequency (MHz)Frequency Step (MHz)
0.25——
25250112.5
2751,000145
1,25020,000250
5.Configure the PXIe-5654 using the following settings:
•Frequency: Frequency from the list in step 4.
•Power Level: -10 dBm
6.Commit the settings to hardware if this is the first iteration of running this procedure;
otherwise skip to step 9.
7.Initiate signal generation.
8.Enable the RF output.
9.Check the signal generation status and verify that there are no reported errors or
warnings.
10. Use the power meter to measure the output power. Store this value as the Power MeterReading.
11. Repeat steps 5 through 10 for each remaining test frequency in the list created in step 4.
12. Disconnect the power meter from the PXIe-5654 RF OUT front panel connector.
13. Connect the PXIe-5654 RF OUT front panel connector to the 6 dB attenuator.
14. Connect the open end of the attenuator to the input of the power splitter using a K(m)-toK(m) 12 in. cable.
15. Connect an available output port of the power splitter to the power sensor.
16. Connect a 6 dB attenuator to the signal source analyzer RF input.
17. Connect the other available output port of the power splitter to the attenuator connected to
the signal source analyzer RF input connector using a K(m)-to-K(m) 36 in. cable.
The hardware setup is shown in the following image, with the dotted line surrounding the
equipment fixture.
Note After assembling the hardware and performing characterization, do not
disassemble or alter the torque applied to the connectors of the fixture. If any
hardware within the fixture is altered, you must characterize again.
8 | ni.com | PXIe-5654 with PXIe-5696 Calibration Procedure
Page 9
Figure 1. Power Splitter Characterization Equipment Setup
250 kHz–20 GHz Signal Generator
ACCESS ACTIVE
REF IN
1–20 MHz
±10 dBm NOM
17 dBm MAX
PULSE
IN
–0.5V MIN
+5.5V MAX
FM IN
±1 VDC NOM
±2 VDC MAX
AM IN
±1 VDC NOM
±2 VDC MAX
ALC IN
±15 VDC MAX
RF OUT
250 kHz –20 GHz
REF OUT
10 MHz
5 dBm ±2 dBm NOM
REF OUT 2
100 MHz
5 dBm ±2 dBm NOM
TRIG
IN/OUT
3.3V CMOS
IN/OUT
–0.5V MIN IN
5.5V MAX IN
ESD
SENSITIVE
ALL PORTS
50 Ω
23
8
5
2
4
7
6
1
1. PXIe-5654 RF Analog Signal Generator
2. K(m)-to-K(f) 6 dB Attenuator
3. K(m)-to-K(m) Cable, 12 in.
4. Power Sensor
5. Power Meter
6. Power Splitter
7. K(m)-to-K(m) Cable, 36 in.
8. Signal Source Analyzer
18. Configure the reference level of the spectrum analyzer to -30 dBm.
The spectrum analyzer is used only for termination.
19. Configure the PXIe-5654 using the following settings:
•Frequency: Frequency from the list in step 4.
•Power Level: -10 dBm
20. Check the signal generation status and verify that there are no reported errors or
warnings.
21. Use the power meter to measure the output power. Store this value as Splitter.
22. Calculate the splitter loss using the following formula:
Splitter Loss = Power Meter Reading - Splitter
23. Store the calculated values.
24. Repeat steps 19 through 23 for each remaining test frequency in the list created in step 4.
25. Close the device session.
As-Found and As-Left Limits
The as-found limits are the published specifications for the PXIe-5654 with PXIe-5696.
NI uses these limits to determine whether the PXIe-5654 with PXIe-5696 meets the
specifications when it is received for calibration. Use the as-found limits during initial
verification.
The as-left calibration limits are equal to the published NI specifications for the
PXIe-5654 with PXIe-5696, less guard bands for measurement uncertainty, temperature drift,
and drift over time. NI uses these limits to reduce the probability that the instrument will be
outside the published specification limits at the end of the calibration cycle. Use the as-left
limits when performing verification after adjustment.
Verification
The performance verification procedures assume that adequate traceable uncertainties are
available for the calibration references.
Before starting verification, complete the following steps.
1.Complete the verification procedures for the PXIe-5654 as a stand-alone device.
2.Associate the PXIe-5654 and PXIe-5696 in MAX.
3.Make the following reference connections.
a)Connect the frequency reference source to the signal source analyzer REF IN back
panel connector using a BNC(m)-to-BNC(m) cable.
b)Connect the signal source analyzer REF OUT connector to the PXIe-5654 REF IN
connector using a BNC(m)-to-SMA(m) cable.
Related Information
For instructions on verifying the PXIe-5654 as a stand-alone device, refer to the PXIe-5654
Calibration Procedure.
Verifying RF OUT Frequency Accuracy
This procedure verifies that the internal signal generation circuitry is adjusted for correct
frequency accuracy.
1.Connect the PXIe-5654 to the PXIe-5696.
2.Connect the PXIe-5696 RF OUT front panel connector to the signal source analyzer RF
input connector using the K(m)-to-K(m) cable, as shown in the following figure.
10 | ni.com | PXIe-5654 with PXIe-5696 Calibration Procedure
Page 11
Figure 2. RF OUT Frequency Accuracy Verification Equipment Setup (Reference
250 kHz–20 GHz Signal Generator
ACCESS ACTIVE
REF IN
1–20 MHz
±10 dBm NOM
17 dBm MAX
PULSE
IN
–0.5V MIN
+5.5V MAX
FM IN
±1 VDC NOM
±2 VDC MAX
AM IN
±1 VDC NOM
±2 VDC MAX
ALC IN
±15 VDC MAX
RF OUT
250 kHz –
20 GHz
REF OUT
10 MHz
5 dBm ±2 dBm NOM
REF OUT 2
100 MHz
5 dBm ±2 dBm NOM
TRIG
IN/OUT
3.3V CMOS
IN/OUT
–0.5V MIN IN
5.5V MAX IN
ESD
SENSITIVE
ALL PORTS
50 Ω
250 kHz–20 GHz Amplitude Extender
ACCESS ACTIVE
RF AMP
OUT
ATTN IN
+30 dBm MAX
ALC OUT
±15 VDC MAX
RF OUT
Reverse Power
±30 dBm MAX
RF IN
±15 dBm
MAX
ESD
SENSITIVE
ALL PORTS
50 Ω
T
T
3
4
12
Connections Not Shown)
1. PXIe-5654 RF Analog Signal Generator
2. PXIe-5696 Amplitude Extender
3. Signal Source Analyzer
4. K(m)-to-K(m) Cable
3.Create a new device session for the PXIe-5654 with PXIe-5696.
4.Create a list of test frequencies, including endpoints, containing 100 MHz and 1 GHz,
according to the frequency range of your PXIe-5654 model, with 1 GHz increments.
5.Configure the PXIe-5654 with PXIe-5696 using the following settings:
•Frequency: Frequency from the list in step 4
6.Commit the settings to hardware if this is the first iteration of running this procedure;
7.Initiate signal generation.
8.Enable the RF output.
9.Check the signal generation status and verify that there are no reported errors or
10. Configure the signal source analyzer using the following settings:
•Center frequency: Frequency from the list in step 4
•Reference level: 20 dBm
•Frequency span: 1 MHz
•Reference Clock source: External
•Frequency counter resolution: 0.1 Hz
•Sweep: Manual
•Number of sweeps: 1
•Auto Sweep Time
11. Measure the frequency of the peak that is returned by the signal source analyzer at
approximately the corresponding point in the frequency list you created in step 4.
12. Calculate the deviation using the following formula:
Δf =
f
Measured
MHz − f
f
Expected
Expected
MHz
MHz
ppm
13. Ensure that the deviation found in the previous step is less than the result of the following
equations:
As Left Calculation: Initial Accuracy + Temperature Stability
As Found Calculation: Initial Accuracy + Aging + Temperature Stability
where
Initial Accuracy = ±0.1 ppm
Temperature Stability (15 °C to 35 °C) = ±0.2 ppm
Ten Year Aging = 1.25 ppm
14. Repeat steps 5 through 13 for the remaining frequencies in the list created in step 4.
15. Close the device session.
Verifying RF OUT Phase Noise
This procedure verifies that the internal frequency generation circuitry is adjusted for correct
phase noise accuracy.
1.Connect the PXIe-5654 to the PXIe-5696.
2.Connect the PXIe-5696 RF OUT front panel connector to the signal source analyzer RF
input connector using the K(m)-to-K(m) cable.
The hardware setup is shown in the following figure.
12 | ni.com | PXIe-5654 with PXIe-5696 Calibration Procedure
Carrier Frequency OffsetResolution BandwidthCross Correlation
100 Hz to 300 Hz10 Hz100
300 Hz to 1 kHz30 Hz1,000
1 kHz to 3 kHz100 Hz10,000
3 kHz to 10 kHz300 Hz10,000
10 kHz to 30 kHz1 kHz10,000
30 kHz to 100 kHz1 kHz5,000
100 kHz to 300 kHz3 kHz1,000
300 kHz to 1 MHz10 kHz1,000
6.Configure the PXIe-5654 with PXIe-5696 using the following settings:
•Frequency: 500 MHz
•Power Level: +8 dBm
7.Commit the settings to hardware if this is the first iteration of running the procedure;
otherwise skip to step 10.
8.Initiate signal generation.
9.Enable the RF output.
10. Set the signal source analyzer center frequency to 500 MHz if the signal source analyzer
does not automatically center for phase noise measurements.
11. Check the signal generation status and verify that there are no reported errors or
warnings.
12. Measure the phase noise using the signal source analyzer for every offset value in the
following table. Record the measurements.
13. Ensure that the recorded measurements are within the limits set in the following tables.
Table 5. RF OUT Phase Noise (dBc/Hz), As-Found Limits
OffsetCarrier Frequency
500 MHz1 GHz5 GHz10 GHz20 GHz
100 Hz≤-107≤-101≤-87≤-81≤-75
1 kHz≤-126≤-121≤-109≤-103≤-97
10 kHz≤-135≤-130≤-120≤-114≤-108
100 kHz≤-137≤-131≤-122≤-117≤-111
1 MHz≤-138≤-132≤-125≤-119≤-113
14 | ni.com | PXIe-5654 with PXIe-5696 Calibration Procedure
Page 15
Table 6. RF OUT Phase Noise (dBc/Hz), As-Left Limits
OffsetCarrier Frequency
500 MHz1 GHz5 GHz10 GHz20 GHz
100 Hz≤-108.5≤-102.5≤-88.5≤-82.5≤-76.5
1 kHz≤-127.5≤-122.5≤-110.5≤-104.5≤-98.5
10 kHz≤-135.5≤-130.5≤-120.5≤-114.5≤-108.5
100 kHz≤-138≤-132≤-123≤-118≤-112
1 MHz≤-138.5≤-132.5≤-125.5≤-119.5≤-113.5
14. Repeat steps 6 through 13 for 1 GHz, 5 GHz, 10 GHz, and 20 GHz carrier frequencies.
15. Close the device session.
Verifying RF OUT Amplitude Accuracy
This procedure verifies that the PXIe-5654 with PXIe-5696 is adjusted for correct amplitude
accuracy.
1.Connect the PXIe-5654 to the PXIe-5696.
2.Using the fixture that was previously characterized in Characterizing the Power Splitter,
connect the input of the fixture to the PXIe-5696 RF OUT front panel connector.
3.Connect the outputs of the fixture to the power sensor and to the signal source analyzer
RF input connector, as shown in the following image.
All other connections should not have changed since characterization.
36. Verify that the Power Deviation (dB) results in the preceding step are within the limits
listed in Tables 8 and 9.
37. Repeat steps 30 through 36 for each test frequency in the frequency list created in step 29.
The following steps apply to using an output power of <-40 dBm.
38. Create a list of test frequencies from Table 7 that apply to a power level of -80 dBm
39. Configure the PXIe-5654 with PXIe-5696 using the following settings:
•Frequency: Frequency from the list in step 38
•Power Level: -80 dBm, which is the Expected Power (dBm)
40. Commit the settings to hardware if this is the first iteration of running this procedure;
otherwise skip to step 41.
41. Check the signal generation status and verify that there are no reported errors or
warnings.
42. Configure the signal source analyzer with the following settings:
•Center frequency: Frequency in step 38
•RF Input coupling: DC
•Reference level: -30 dBm
•Frequency span: 0 Hz
•Resolution bandwidth: 500 Hz
•Sweep time: 200 ms
•Reference Clock source: External
•RF attenuation: 5 dB
•Preamplifier: On
43. Measure the power level with the signal source analyzer. Record the value as the SignalSource Analyzer Reading.
44. Correct the reading with previously stored characteristic values.
Power Deviation (dB) = Signal Source Analyzer Reading + Corrections stored in step 26
45. Verify that the Power Deviation (dB) results in the preceding step are within the limits
listed in Tables 8 and 9.
46. Repeat steps 39 through 45 for each test frequency in the frequency list created in step 38.
47. Stop signal generation.
48. Close the device session.
Verifying RF OUT Maximum Power
This procedure verifies that the PXIe-5654 with PXIe-5696 produces the correct maximum
output power.
1.Connect the PXIe-5654 to the PXIe-5696.
2.Using the fixture that was previously characterized in Characterizing the Power Splitter,
connect the input of the fixture to the PXIe-5696 RF OUT front panel connector.
3.Connect the outputs of the fixture to the power sensor and to the signal source analyzer
RF input connector, as shown in the following image.
All other connections should not have changed since characterization.
22 | ni.com | PXIe-5654 with PXIe-5696 Calibration Procedure
Page 23
Figure 5. RF OUT Maximum Power Verification Equipment Setup
250 kHz–20 GHz Signal Generator
ACCESS ACTIVE
REF IN
1–20 MHz
±10 dBm NOM
17 dBm MAX
PULSE
IN
–0.5V MIN
+5.5V MAX
FM IN
±1 VDC NOM
±2 VDC MAX
AM IN
±1 VDC NOM
±2 VDC MAX
ALC IN
±15 VDC MAX
RF OUT
250 kHz –
20 GHz
REF OUT
10 MHz
5 dBm ±2 dBm NOM
REF OUT 2
100 MHz
5 dBm ±2 dBm NOM
TRIG
IN/OUT
3.3V CMOS
IN/OUT
–0.5V MIN IN
5.5V MAX IN
ESD
SENSITIVE
ALL PORTS
50 Ω
250 kHz–20 GHz Amplitude Extender
ACCESS ACTIVE
RF AMP
OUT
ATTN IN
+30 dBm MAX
ALC OUT
±15 VDC MAX
RF OUT
Reverse Power
±30 dBm MAX
RF IN
±15 dBm
MAX
ESD
SENSITIVE
ALL PORTS
50 Ω
T
T
4
9
6
3
5
8
7
2
1
3
1. PXIe-5654 RF Analog Signal Generator
2. PXIe-5696 Amplitude Extender
3. K(m)-to-K(f) 6 dB Attenuator
4. K(m)-to-K(m) Cable, 12 in.
6. Power Meter
7. Power Splitter
8. K(m)-to-K(m) Cable, 36 in.
9. Signal Source Analyzer
5. Power Sensor
4.Configure the reference level of the spectrum analyzer to -30 dBm.
The spectrum analyzer is used only for termination.
5.Create a new device session for the PXIe-5654 with PXIe-5696.
6.Create a list of test frequencies using the values in the following table.
Table 10. RF OUT Maximum and Minimum Power Test Frequencies
7.Configure the PXIe-5654 with PXIe-5696 using the following settings:
•Frequency: Start frequency from the list in step 6
•Power Level: Power level from the list in step 6
8.Initiate signal generation if this is the first iteration of running the procedure; otherwise
skip to step 11.
9.Enable the RF output.
10. Measure the PXIe-5696 RF OUT front panel connector power using the power meter and
the fixture created in Characterizing the Power Splitter. Correct the power meter reading
for the RF frequency and previously characterized power splitter loss.
11. Verify that the power output measured in step 10 meets the limits in the following table.
Table 11. Maximum Power Limits
Frequency (MHz)As-Found Limit (dBm)As-Left Limit (dBm)
0.25 to ≤250≥10≥11
>250 to ≤1,000≥20≥21
>1,000 to ≤3,000≥24≥25
>3,000 to ≤6,000>23>24
>6,000 to ≤15,000>20>21
>15,000 to ≤20,000>18>19
12. Repeat steps 7 through 11 for each frequency from the list created in step 6.
13. Close the device session.
Optional Verification
Use the following procedures to verify nonwarranted specifications for the PXIe-5654 with
PXIe-5696.
Verifying PULSE IN Operation
This procedure verifies that the PXIe-5654 PULSE IN connection is functioning.
1.Connect the PXIe-5654 to the PXIe-5696.
2.Connect the signal source analyzer RF input to the PXIe-5696 RF OUT front panel
connector using the K(m)-to-K(m) cable.
3.Connect the Port 0, Line 0 (P0.0) output of the USB digital output device to two separate
connectors using a BNC Tee connector and a BNC(m)-to-terminal block adapter.
a)Connect the P0.0 output of the USB digital output device to the signal source
analyzer EXT TRIG/GATE IN back panel connector using a BNC(m)-to-BNC(m)
cable. (Not pictured.)
b)Connect the P0.0 output of the USB digital output device to the PXIe-5654 front
panel PULSE IN connector using BNC(m)-to-SMB(f) cable.
24 | ni.com | PXIe-5654 with PXIe-5696 Calibration Procedure
Page 25
The hardware setup is shown in the following figure.
250 kHz–20 GHz Signal Generator
ACCESS ACTIVE
REF IN
1–20 MHz
±10 dBm NOM
17 dBm MAX
PULSE
IN
–0.5V MIN
+5.5V MAX
FM IN
±1 VDC NOM
±2 VDC MAX
AM IN
±1 VDC NOM
±2 VDC MAX
ALC IN
±15 VDC MAX
RF OUT
250 kHz –
20 GHz
REF OUT
10 MHz
5 dBm ±2 dBm NOM
REF OUT 2
100 MHz
5 dBm ±2 dBm NOM
TRIG
IN/OUT
3.3V CMOS
IN/OUT
–0.5V MIN IN
5.5V MAX IN
ESD
SENSITIVE
ALL PORTS
50 Ω
250 kHz–20 GHz Amplitude Extender
ACCESS ACTIVE
RF AMP
OUT
ATTN IN
+30 dBm MAX
ALC OUT
±15 VDC MAX
RF OUT
Reverse Power
±30 dBm MAX
RF IN
±15 dBm
MAX
ESD
SENSITIVE
ALL PORTS
50 Ω
T
T
4
6
5
3
2
1
NI USB-6501
24-line Digital I/O
Figure 6. PULSE IN Operation Verification Equipment Setup
1. USB-6501 Digital I/O Device
2. BNC(m)-to-SMB(f) Cable
3. PXIe-5654 RF Analog Signal Generator
4. PXIe-5696 Amplitude Extender
5. K(m)-to-K(m) Cable
6. Signal Source Analyzer
4.Create a new device session for the PXIe-5654 with PXIe-5696.
5.Configure the signal source analyzer using the following settings:
•Reference level: 5 dBm
•Start frequency: 0.25 MHz
•Stop frequency: 10,000 MHz
•Span (kHz): 2 KHz
•Resolution bandwidth: 30 Hz
•Video bandwidth: 100 Hz
•RF attenuation: 30 dB
6.Configure the PXIe-5654 with PXIe-5696 using the following settings:
10. Create a USB digital output channel for the Port 0, Line 0 of the USB digital output
device.
11. Configure the channel to be Active Drive.
12. Enable the output on the channel created in step 10.
13. Start the USB digital output task for the channel.
14. Sweep the signal source analyzer and read the marker. This value is the OnPower.
15. Stop the USB digital output task.
16. Clear the USB digital output task.
17. Repeat steps 10 and 11, disabling the output of the USB digital output channel.
18. Sweep the signal source analyzer and read the marker. This value is the OffPower.
19. Verify that the following equation is true.
OnPower - OffPower ≥ 80 dB
20. Close the device sessions.
Adjustment
This section describes the steps needed to adjust the PXIe-5654 with PXIe-5696 to meet
published specifications. The PXIe-5696 is adjusted while connected to the PXIe-5654.
Note Adjusting the PXIe-5696 requires a calibrated 20 GHz PXIe-5654. If you
performed verification using a 10 GHz PXIe-5654, power off the system and replace
the 10 GHz PXIe-5654 with a 20 GHz PXIe-5654 prior to starting adjustment.
Note Before starting this adjustment on the PXIe-5654 with PXIe-5696,
unassociate the PXIe-5654 and PXIe-5696 in MAX, then complete the adjustment
procedures for the PXIe-5654 as a stand-alone device.
Related Information
For instructions on adjusting the PXIe-5654 as a stand-alone device, refer to the PXIe-5654
Calibration Procedure.
Measuring RF Attenuator Path Input Power
This procedure measures the input power used for PXIe-5696 attenuator path adjustments.
Using the measured input power yields a more accurate adjustment for the PXIe-5696
attenuator paths.
1.Connect the PXIe-5654 RF OUT front panel connector to the power meter through the
20 dB attenuator as shown in the following figure.
26 | ni.com | PXIe-5654 with PXIe-5696 Calibration Procedure
3.Initialize a new external calibration session for the PXIe-5696.
4.Call the niRFSG Initialize External Calibration VI using the appropriate password.
Note The default password for password-protected operations is NI.
5.Call the niRFSG 5696 Attenuator Path Cal Initialize VI with the following inputs:
•NI 5654 resource name
•measurement port: Input Power
•input power file: Specify a fully qualified path to the Atten_Zero_Loss.bin file
6.Call the niRFSG 5696 Attenuator Path Cal Configure VI.
7.Measure the PXIe-5654 output power through the 20 dB attenuator with the power meter.
Ensure that the power meter is settled within 0.1% before taking a measurement.
8.Correct the power meter reading for the RF frequency using the frequency to measure
output of the niRFSG 5696 Attenuator Path Cal Configure VI.
Note The VI creates this file. Specify a path to a writeable location for
9.Use the corrected reading from step 8 as the value of the measured power input of the
250 kHz–20 GHz Signal Generator
ACCESS ACTIVE
REF IN
1–20 MHz
±10 dBm NOM
17 dBm MAX
PULSE
IN
–0.5V MIN
+5.5V MAX
FM IN
±1 VDC NOM
±2 VDC MAX
AM IN
±1 VDC NOM
±2 VDC MAX
ALC IN
±15 VDC MAX
RF OUT
250 kHz –
20 GHz
REF OUT
10 MHz
5 dBm ±2 dBm NOM
REF OUT 2
100 MHz
5 dBm ±2 dBm NOM
TRIG
IN/OUT
3.3V CMOS
IN/OUT
–0.5V MIN IN
5.5V MAX IN
ESD
SENSITIVE
ALL PORTS
50 Ω
250 kHz–20 GHz Amplitude Extender
ACCESS ACTIVE
RF AMP
OUT
ATTN IN
+30 dBm MAX
ALC OUT
±15 VDC MAX
RF OUT
Reverse Power
±30 dBm MAX
RF IN
±15 dBm
MAX
ESD
SENSITIVE
ALL PORTS
50 Ω
21
3
5
4
niRFSG 5696 Attenuator Path Cal Adjust VI.
10. Repeat steps 6 through 9 until the attenuator calibration complete output of the
niRFSG 5696 Attenuator Path Cal Adjust VI returns a value of TRUE.
11. Call the niRFSG Close External Calibration VI to close the session. Set the writecalibration to hardware? parameter to TRUE.
Adjusting RF Attenuator Path
This procedure characterizes the RF attenuator path in the PXIe-5696 and updates the values
stored in the PXIe-5696 EE PROM. This adjustment yields a more accurate output power for
the PXIe-5696 RF OUT.
1.Connect the PXIe-5654 RF OUT to the PXIe-5696 ATTN IN front panel connector
through the 20 dB attenuator.
2.Connect the PXIe-5696 ATTN OUT front panel connector to the power meter as shown
in the following figure.
3.Initialize a new external calibration session for the PXIe-5696.
28 | ni.com | PXIe-5654 with PXIe-5696 Calibration Procedure
4. Power Sensor
5. 20 dB Attenuator
Page 29
4.Call the niRFSG Initialize External Calibration VI using the appropriate password.
Note The default password for password-protected operations is NI.
5.Call the niRFSG 5696 Attenuator Path Cal Initialize VI with the following inputs:
•NI 5654 resource name
•measurement port: Output Power
6.Select the file created during Measuring RF Attenuator Path Input Power on page 26 as
the input power file of the niRFSG Attenuator Path Cal Initialize VI.
7.Call the niRFSG 5696 Attenuator Path Cal Configure VI.
8.Measure the PXIe-5696 output power with the power meter. Correct the measurement
using the frequency to measure output of the niRFSG 5696 Attenuator Path Cal
Configure VI. Ensure that the power meter is settled within 0.1% before taking a
measurement.
9.Use the corrected measurement from step 8 as the value of the measured power input of
the niRFSG 5696 Attenuator Path Cal Adjust VI.
10. Repeat steps 7 through 9 until the attenuator calibration complete output of the
niRFSG 5696 Attenuator Path Cal Adjust VI returns a value of TRUE.
11. Call the niRFSG Close External Calibration VI to close the session. Set the writecalibration to hardware? parameter to TRUE to store the results to the EEPROM on the
PXIe-5696.
Adjusting RF Attenuator Accuracy
This procedure characterizes the RF attenuator in the PXIe-5696 and updates the values stored
in the PXIe-5696 EEPROM. This adjustment yields a more accurate output power for the
PXIe-5696 RF OUT.
1.Connect the power sensor directly to the PXIe-5696 RF OUT front panel connector as
shown in the following figure.
2.Initialize a new external calibration session for the PXIe-5696.
3.Call the niRFSG Initialize External Calibration VI using the appropriate password.
Note The default password for password-protected operations is NI.
4.Call the niRFSG 5696 Attenuator Cal Initialize VI with the NI 5654 resource name as
input.
5.Call the niRFSG 5696 Attenuator Cal Configure VI.
6.Measure the PXIe-5696 output power with the power meter. Correct the measurement
7.Use the reading from step 6 as the value of the measured power input of the to niRFSG
8.Repeat steps 6 and 7 until the attenuator calibration complete output of the niRFSG
9.Call the niRFSG Close External Calibration VI to close the session. Set the write
30 | ni.com | PXIe-5654 with PXIe-5696 Calibration Procedure
using the frequency to measure output of the niRFSG 5696 Attenuator Cal Configure
VI. Ensure that the power meter is settled within 0.1% before taking a reading.
5696 Attenuator Cal Adjust VI.
5696 Attenuator Cal Adjust VI returns a value of TRUE.
calibration to hardware? parameter to TRUE to store the results to the EEPROM on the
PXIe-5696.
Page 31
Measuring Amplifier Path Input Power
250 kHz–20 GHz Signal Generator
ACCESS ACTIVE
REF IN
1–20 MHz
±10 dBm NOM
17 dBm MAX
PULSE
IN
–0.5V MIN
+5.5V MAX
FM IN
±1 VDC NOM
±2 VDC MAX
AM IN
±1 VDC NOM
±2 VDC MAX
ALC IN
±15 VDC MAX
RF OUT
250 kHz –20 GHz
REF OUT
10 MHz
5 dBm ±2 dBm NOM
REF OUT 2
100 MHz
5 dBm ±2 dBm NOM
TRIG
IN/OUT
3.3V CMOS
IN/OUT
–0.5V MIN IN
5.5V MAX IN
ESD
SENSITIVE
ALL PORTS
50 Ω
2
1
43
This procedure measures the input power used for PXIe-5696 RF amplifier path adjustments.
Using the measured input power yields a more accurate adjustment.
1.Connect the PXIe-5654 RF OUT front panel connector directly to the power meter as
shown in the following figure.
Figure 10. PXIe-5696 Amplifier Path Input Power Measurement Equipment Setup
1. PXIe-5654 RF Analog Signal Generator
2. Power Meter
3. Power Sensor
2.Initialize the power meter.
3.Initialize a new external calibration session for the PXIe-5696.
4.Call the niRFSG Initialize External Calibration VI using the appropriate password.
Note The default password for password-protected operations is NI.
5.Call the niRFSG 5696 Amplifier Cal Initialize VI with the following inputs:
•input power file: Specify a fully qualified path to the
High_Power_Path_Input_Power.bin file
Note The VI creates this file. Specify a path to a writeable location for
the file.
6.Call the niRFSG 5696 Amplifier Cal Configure VI.
7.Measure the PXIe-5654 output power with the power meter. Ensure that the power meter
is settled within 0.1% before taking a measurement.
8.Correct the measurement from step 7 using the frequency to measure output of the
niRFSG 5696 Amplifier Cal Configure VI.
9.Use the corrected reading from step 8 as the value of the measured power input of the
niRFSG 5696 Amplifier Cal Adjust VI.
10. Repeat steps 6 through 9 until the amp calibration complete output of the niRFSG 5696
Amplifier Cal Adjust VI returns a value of TRUE.
11. Call the niRFSG Close External Calibration VI to close the session. Set the writecalibration to hardware? parameter to TRUE.
12. Repeat steps 5 through 10 with the following inputs:
•path to calibrate: Low Harmonic
•measurement port: Input Power
•input power file: Specify a fully qualified path to the
Low_Harmonic_Path_Input_Power.bin file
Note The VI creates this file. Specify a path to a writeable location for
the file.
13. Call the niRFSG Close External Calibration VI to close the session. Set the writecalibration to hardware? parameter to TRUE.
Adjusting Amplitude Accuracy (Low Harmonic Paths)
This procedure characterizes the low harmonic paths in the PXIe-5696 and updates the values
stored in the PXIe-5696 EEPROM. This adjustment yields a more accurate output power for
the PXIe-5696 RF OUT.
1.Connect the power sensor directly to the PXIe-5696 RF OUT front panel connector as
shown in the following figure.
32 | ni.com | PXIe-5654 with PXIe-5696 Calibration Procedure
2.Initialize a new external calibration session for the PXIe-5696.
3.Call the niRFSG Initialize External Calibration VI using the appropriate password.
Note The default password for password-protected operations is NI.
4.Call the niRFSG 5696 Amplifier Cal Initialize VI with the following inputs:
•path to calibrate: Low Harmonic
•measurement port: Output Power
5.Select the file created during Measuring Amplifier Path Input Power on page 31 for the
Low Harmonic path as the input power file of the niRFSG Amplifier Cal Initialize VI.
6.Call the niRFSG 5696 Amplifier Cal Configure VI.
7.Measure the PXIe-5696 output power with the power meter. Correct the measurement
using the frequency to measure output of the niRFSG 5696 Amplifier Cal Configure VI.
Ensure that the power meter is settled within 0.1% before taking a reading.
8.Use the reading from step 7 as the value of the measured power input of the niRFSG
5696 Amplifier Cal Adjust VI.
9.Repeat steps 7 and 8 until the amp calibration complete output from the niRFSG 5696
250 kHz–20 GHz Signal Generator
ACCESS ACTIVE
REF IN
1–20 MHz
±10 dBm NOM
17 dBm MAX
PULSE
IN
–0.5V MIN
+5.5V MAX
FM IN
±1 VDC NOM
±2 VDC MAX
AM IN
±1 VDC NOM
±2 VDC MAX
ALC IN
±15 VDC MAX
RF OUT
250 kHz –
20 GHz
REF OUT
10 MHz
5 dBm ±2 dBm NOM
REF OUT 2
100 MHz
5 dBm ±2 dBm NOM
TRIG
IN/OUT
3.3V CMOS
IN/OUT
–0.5V MIN IN
5.5V MAX IN
ESD
SENSITIVE
ALL PORTS
50 Ω
250 kHz–20 GHz Amplitude Extender
ACCESS ACTIVE
RF AMP
OUT
ATTN IN
+30 dBm MAX
ALC OUT
±15 VDC MAX
RF OUT
Reverse Power
±30 dBm MAX
RF IN
±15 dBm
MAX
ESD
SENSITIVE
ALL PORTS
50 Ω
T
T
3
21
44
Amplifier Cal Adjust VI returns a value of TRUE.
10. Call the niRFSG Close External Calibration VI to close the session. Set the writecalibration to hardware? parameter to TRUE to store the results to the EEPROM on the
PXIe-5696.
Related Information
Measuring Amplifier Path Input Power on page 31
Adjusting Amplitude Accuracy (High Power Paths)
This procedure characterizes the high power paths in the PXIe-5696 and updates the values
stored in the PXIe-5696 EEPROM. This adjustment yields a more accurate output power for
the PXIe-5696 RF OUT.
1.Connect the power sensor directly to the PXIe-5696 RF OUT front panel connector as
shown in the following figure.
Figure 12. PXIe-5696 Amplitude Accuracy (High Power Paths) Adjustment Equipment
Setup
1. PXIe-5654 RF Analog Signal Generator
2. PXIe-5696 Amplitude Extender
34 | ni.com | PXIe-5654 with PXIe-5696 Calibration Procedure
3. Power Meter
4. Power Sensor
Page 35
2.Initialize a new external calibration session for the PXIe-5696.
3.Call the niRFSG Initialize External Calibration VI using the appropriate password.
Note The default password for password-protected operations is NI.
4.Call the niRFSG 5696 Amplifier Cal Initialize VI with the following input parameters:
•path to calibrate: High Power
•measurement port: Output Power
5.Select the file created during Measuring Amplifier Path Input Power on page 31 for the
High Power path as the input power file of the niRFSG Amplifier Cal Initialize VI.
6.Call the niRFSG 5696 Amplifier Cal Configure VI.
7.Measure the PXIe-5696 output power with the power meter. Correct the measurement
using the frequency to measure output of the niRFSG 5696 Amplifier Cal Configure VI.
Ensure that the power meter is settled within 0.1% before taking a reading.
8.Use the reading from step 7 as the value of the measured power input of the niRFSG
5696 Amplifier Cal Adjust VI.
9.Repeat steps 7 and 8 until the amp calibration complete output from the niRFSG 5696
Amplifier Cal Adjust VI returns a value of TRUE.
10. Call the niRFSG Close External Calibration VI to close the session. Set the writecalibration to hardware? parameter to TRUE to store the results to the EEPROM on the
PXIe-5696.
Related Information
Measuring Amplifier Path Input Power on page 31
Adjusting ALC
This procedure characterizes the ALC path in the PXIe-5654 with PXIe-5696 and updates the
values stored in the PXIe-5696 EEPROM. This adjustment yields a more accurate output
power for the PXIe-5696 RF OUT.
1.Connect the power sensor directly to the PXIe-5696 RF OUT front panel connector as
shown in the following figure.
2.Initialize a new external calibration session for the PXIe-5696.
3.Call the niRFSG Initialize External Calibration VI using the appropriate password.
Note The default password for password-protected operations is NI.
4.Call the niRFSG 5696 ALC Cal Initialize VI.
5.Call the niRFSG 5696 ALC Cal Configure VI
6.Measure the PXIe-5696 output power with the power meter. Correct the measurement
using the frequency to measure output of the niRFSG 5696 ALC Cal Configure VI.
Ensure that the power meter is settled within 0.1% before taking a reading.
7.Use the reading from step 6 as the measured power input of the niRFSG 5696 ALC Cal
Adjust VI.
8.Repeat steps 6 and 7 until ALC calibration complete output of the niRFSG 5696 ALC
Cal Adjust VI returns a value of TRUE.
9.Call the niRFSG Close External Calibration VI to close the session. Set the writecalibration to hardware? parameter to TRUE to store the results to the EEPROM on the
PXIe-5696.
36 | ni.com | PXIe-5654 with PXIe-5696 Calibration Procedure
Page 37
Reverification
Repeat the Verification section to determine the as-left status of the PXIe-5654 or the
PXIe-5696.
Note If any test fails reverification after performing an adjustment, verify that you
have met the test conditions before returning your PXIe-5654 or PXIe-5696 to NI.
Refer to the Worldwide Support and Services section for information about support
resources or service requests.
Updating Calibration Date and Time
This procedure updates the date and time of the last calibration of the PXIe-5696.
1.Call the niRFSG Initialize External Calibration VI.
2.Call the niRFSG Update External Calibration Date and Time VI.
3.Call the niRFSG Close External Calibration VI to close the session. Set the writecalibration to hardware? parameter to TRUE to store the results to the EEPROM on the
PXIe-5696.
Worldwide Support and Services
The NI website is your complete resource for technical support. At ni.com/support, you have
access to everything from troubleshooting and application development self-help resources to
email and phone assistance from NI Application Engineers.
Visit ni.com/services for information about the services NI offers.
Visit ni.com/register to register your NI product. Product registration facilitates technical
support and ensures that you receive important information updates from NI.
NI corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas,
78759-3504. NI also has offices located around the world. For support in the United States,
create your service request at ni.com/support or dial 1 866 ASK MYNI (275 6964). For
support outside the United States, visit the Worldwide Offices section of ni.com/niglobal to
access the branch office websites, which provide up-to-date contact information.
Information is subject to change without notice. Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for
information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their
respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your
software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find
information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer
to the Export Compliance Information at ni.com/legal/export-compliance for the NI global trade compliance policy and how
to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS
TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S.
Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable
limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.