National Instruments PXIe-5413 Calibration Manual

CALIBRATION PROCEDURE
PXIe-5413
PXIe, 20 MHz Bandwidth, 16-Bit PXI Waveform Generator
This document contains the verification and adjustment procedures for the PXIe-5413. Refer to ni.com/calibration for more information about calibration solutions.
Contents
Required Software.....................................................................................................................1
Related Documentation.............................................................................................................2
Test Equipment..........................................................................................................................2
Test Conditions..........................................................................................................................4
Password................................................................................................................................... 5
Calibration Interval................................................................................................................... 5
As-Found and As-Left Limits................................................................................................... 5
Calibration Overview................................................................................................................5
Verifying the PXIe-5413........................................................................................................... 6
Adjusting the PXIe-5413........................................................................................................ 14
Reverification..........................................................................................................................18
Setting the Calibration Date and Due Date.............................................................................18
Worldwide Support and Services............................................................................................ 18
Required Software
Calibrating the PXIe-5413 requires you to install the following software on the calibration system:
NI-FGEN 17.1 or later
The PXIe-5413 was first supported in NI-FGEN 17.1.
LabVIEW 2016 or later, Full Development System or Professional Development System
You can download all required software from ni.com/downloads.
Related Documentation
For additional information, refer to the following documents as you perform the calibration procedure:
PXIe-5413 Getting Started Guide
PXIe-5413 Specifications
NI Signal Generators Help
LabVIEW Help
Visit ni.com/manuals for the latest versions of these documents.
Test Equipment
Refer to the following table for a list of necessary test equipment, minimum requirements for the test equipment, and model recommendations for calibration of the PXIe-5413.
If you do not have the recommended equipment, select a substitute calibration standard using the minimum requirements specifications listed in the table.
Table 1. PXIe-5413 Test Equipment
Equipment Recommended
Model
Where Used Minimum Requirements
Oscilloscope PXIe-5160 or
PXIe-5171
Verifications:
Timebase accuracy
Adjustments:
Timebase
Frequency resolution:
0.1 Hz
The ability to measure sine waves ≥10 MHz with amplitudes of 2 V
pk-pk
to
5 V
pk-pk
with 50 Ω input
impedance.
SMA (m)-to­SMA (m) cable
Verifications:
Timebase accuracy
Adjustments:
Timebase
Frequency range: DC to 100 MHz
Impedance: 50 Ω
Length: <1 m
2 | ni.com | PXIe-5413 Calibration Procedure
Table 1. PXIe-5413 Test Equipment (Continued)
Equipment Recommended
Model
Where Used Minimum Requirements
Precision frequency reference
PXIe-6674T or Rubidium reference
Verifications:
Timebase accuracy
Adjustments:
Timebase
Frequency accuracy: 100 ppb
The ability to PLL lock to the oscilloscope.
DMM PXI-4071 Verifications:
DC voltage accuracy
AC amplitude accuracy
Adjustments:
DC voltage
DC voltage range: ±5 V
DC voltage accuracy: <40 ppm
AC voltage accuracy: ±0.1% of reading at 50 kHz
AC Input Range: 0.05 V
RMS
to 5 V
RMS
AC Input Impedance: ≥10 MΩ
Bandwidth: ≥100 kHz
Double banana plug-to-BNC (f)
Pasternak PE9008 Verifications:
DC voltage accuracy
AC amplitude accuracy
Adjustments:
DC voltage
Impedance: 50 Ω
SMA (m)-to­BNC (m) cable
Verifications:
DC voltage accuracy
AC amplitude accuracy
Adjustments:
DC voltage
Frequency range: DC to 1 MHz
Impedance: 50 Ω
Length: <1 m
PXIe-5413 Calibration Procedure | © National Instruments | 3
Table 1. PXIe-5413 Test Equipment (Continued)
Equipment Recommended
Model
Where Used Minimum Requirements
Power sensor Rohde & Schwarz
(R&S) NRP-Z91
Verifications:
Passband flatness accuracy
Adjustments:
Passband flatness
Range: -30 dBm to 23 dBm
Frequency range: 50 kHz to 100 MHz
Absolute Power Accuracy: <0.048 dB for <100 MHz
Relative Power Accuracy: <0.022 dB for <100 MHz
VSWR: <1.11
SMA (m)-to-N (f) adapter
Verifications:
Passband flatness accuracy
Adjustments:
Passband flatness
Frequency range: DC to 100 MHz
Impedance: 50 Ω
Test Conditions
The following setup and environmental conditions are required to ensure the PXIe-5413 meets published specifications:
The PXIe-5413 is warmed up for 15 minutes at ambient temperature. Warm up begins after the chassis is powered and the PXIe-5413 is recognized by the host and configured using NI-FGEN.
All test instruments are warmed up for at least the period of time stated in the documentation for each instrument and self-calibration is performed. The warm-up time ensures that the test instruments are at a stable operating temperature.
Keep cabling as short as possible. Long cables act as antennas, picking up extra noise that can affect measurements.
Verify that all connections to the PXIe-5413, including front panel connections and screws, are secure.
Use shielded copper wire for all cable connections to the device. Use twisted-pair wire to eliminate noise and thermal offsets.
Maintain an ambient temperature of 23 °C ± 3 °C.
Keep relative humidity between 10% and 90%, noncondensing.
4 | ni.com | PXIe-5413 Calibration Procedure
Ensure that the PXI chassis fan speed is set to HIGH, that the fan filters (if present) are clean, and that the empty slots contain slot blockers and filler panels. For more information about cooling, refer to the Maintain Forced-Air Cooling Note to Users document available at ni.com/manuals.
Plug the PXI chassis and any other instrument standards into the same power strip to avoid ground loops.
Password
The default password for password-protected operations is NI.
Related Information
Adjusting the PXIe-5413 on page 14 Setting the Calibration Date and Due Date on page 18
Calibration Interval
Recommended calibration interval 2 years
As-Found and As-Left Limits
The as-found limits are the published specifications for the PXIe-5413. NI uses these limits to determine whether the PXIe-5413 meets the specifications when it is received for calibration. Use the as-found limits during initial verification.
The as-left calibration limits are equal to the published NI specifications for the PXIe-5413, less guard bands for measurement uncertainty, temperature drift, and drift over time. NI uses these limits to reduce the probability that the instrument will be outside the published specification limits at the end of the calibration cycle. Use the as-left limits when performing verification after adjustment.
Calibration Overview
Install the PXIe-5413 and configure it in NI Measurement & Automation Explorer (MAX) before calibrating.
Calibration includes the following steps:
1. Self-calibration—Let the PXIe-5413 warm up for 15 minutes before self-calibrating the PXIe-5413 by clicking the Self-Calibrate button in MAX or calling niFgen Self Cal.
2. Verification—Verify the existing operation of the device. This step confirms whether the device is operating within the published specification prior to adjustment.
3. Adjustment—Perform an external adjustment of the calibration constants of the device. The adjustment procedure automatically stores the calibration date and temperature on the EEPROM to allow traceability.
PXIe-5413 Calibration Procedure | © National Instruments | 5
4. Self-calibration—Let the PXIe-5413 warm up for 15 minutes before self-calibrating the PXIe-5413 by clicking the Self-Calibrate button in MAX or calling niFgen Self Cal.
5. Reverification—Repeat the Verification procedure to ensure that the device is operating within the published specifications after adjustment.
Refer to the following sections to complete each procedure.
Related Information
Verifying the PXIe-5413 on page 6
Adjusting the PXIe-5413 on page 14
Verifying the PXIe-5413
The performance verification procedures assume that adequate traceable uncertainties are available for the calibration references.
Self-calibrate the PXIe-5413 and all test instruments before beginning the verification procedures.
Note Verification of the PXIe-5413 is complete only after you have successfully
completed all verification tests in this section using the as-found limits.
Related Information
Calibration Overview on page 5
Verifying Timebase Accuracy
Follow this procedure to verify the frequency accuracy of the PXIe-5413 onboard timebase (TCXO) using an oscilloscope.
Refer to the following table as you complete the following steps.
Table 3. Timebase Accuracy Verification
As-Found Limit As-Left Limit
±(1.5 ppm + 1 ppm per year1) ±0.28 ppm
1. Connect one end of the SMA (m)-to-SMA (m) cable to a channel of the PXIe-5413 and the other end to a channel of the oscilloscope.
Note Timebase accuracy verification can be performed on either channel of
the two-channel PXIe-5413.
2. Lock the oscilloscope to the precision frequency reference using the PXI Clock.
1
Where time drift starts at the latest external calibration date.
6 | ni.com | PXIe-5413 Calibration Procedure
Loading...
+ 14 hidden pages