National Instruments PXIe-5413 Calibration Manual

Page 1
CALIBRATION PROCEDURE
PXIe-5413
PXIe, 20 MHz Bandwidth, 16-Bit PXI Waveform Generator
This document contains the verification and adjustment procedures for the PXIe-5413. Refer to ni.com/calibration for more information about calibration solutions.
Contents
Required Software.....................................................................................................................1
Related Documentation.............................................................................................................2
Test Equipment..........................................................................................................................2
Test Conditions..........................................................................................................................4
Password................................................................................................................................... 5
Calibration Interval................................................................................................................... 5
As-Found and As-Left Limits................................................................................................... 5
Calibration Overview................................................................................................................5
Verifying the PXIe-5413........................................................................................................... 6
Adjusting the PXIe-5413........................................................................................................ 14
Reverification..........................................................................................................................18
Setting the Calibration Date and Due Date.............................................................................18
Worldwide Support and Services............................................................................................ 18
Required Software
Calibrating the PXIe-5413 requires you to install the following software on the calibration system:
NI-FGEN 17.1 or later
The PXIe-5413 was first supported in NI-FGEN 17.1.
LabVIEW 2016 or later, Full Development System or Professional Development System
You can download all required software from ni.com/downloads.
Page 2
Related Documentation
For additional information, refer to the following documents as you perform the calibration procedure:
PXIe-5413 Getting Started Guide
PXIe-5413 Specifications
NI Signal Generators Help
LabVIEW Help
Visit ni.com/manuals for the latest versions of these documents.
Test Equipment
Refer to the following table for a list of necessary test equipment, minimum requirements for the test equipment, and model recommendations for calibration of the PXIe-5413.
If you do not have the recommended equipment, select a substitute calibration standard using the minimum requirements specifications listed in the table.
Table 1. PXIe-5413 Test Equipment
Equipment Recommended
Model
Where Used Minimum Requirements
Oscilloscope PXIe-5160 or
PXIe-5171
Verifications:
Timebase accuracy
Adjustments:
Timebase
Frequency resolution:
0.1 Hz
The ability to measure sine waves ≥10 MHz with amplitudes of 2 V
pk-pk
to
5 V
pk-pk
with 50 Ω input
impedance.
SMA (m)-to­SMA (m) cable
Verifications:
Timebase accuracy
Adjustments:
Timebase
Frequency range: DC to 100 MHz
Impedance: 50 Ω
Length: <1 m
2 | ni.com | PXIe-5413 Calibration Procedure
Page 3
Table 1. PXIe-5413 Test Equipment (Continued)
Equipment Recommended
Model
Where Used Minimum Requirements
Precision frequency reference
PXIe-6674T or Rubidium reference
Verifications:
Timebase accuracy
Adjustments:
Timebase
Frequency accuracy: 100 ppb
The ability to PLL lock to the oscilloscope.
DMM PXI-4071 Verifications:
DC voltage accuracy
AC amplitude accuracy
Adjustments:
DC voltage
DC voltage range: ±5 V
DC voltage accuracy: <40 ppm
AC voltage accuracy: ±0.1% of reading at 50 kHz
AC Input Range: 0.05 V
RMS
to 5 V
RMS
AC Input Impedance: ≥10 MΩ
Bandwidth: ≥100 kHz
Double banana plug-to-BNC (f)
Pasternak PE9008 Verifications:
DC voltage accuracy
AC amplitude accuracy
Adjustments:
DC voltage
Impedance: 50 Ω
SMA (m)-to­BNC (m) cable
Verifications:
DC voltage accuracy
AC amplitude accuracy
Adjustments:
DC voltage
Frequency range: DC to 1 MHz
Impedance: 50 Ω
Length: <1 m
PXIe-5413 Calibration Procedure | © National Instruments | 3
Page 4
Table 1. PXIe-5413 Test Equipment (Continued)
Equipment Recommended
Model
Where Used Minimum Requirements
Power sensor Rohde & Schwarz
(R&S) NRP-Z91
Verifications:
Passband flatness accuracy
Adjustments:
Passband flatness
Range: -30 dBm to 23 dBm
Frequency range: 50 kHz to 100 MHz
Absolute Power Accuracy: <0.048 dB for <100 MHz
Relative Power Accuracy: <0.022 dB for <100 MHz
VSWR: <1.11
SMA (m)-to-N (f) adapter
Verifications:
Passband flatness accuracy
Adjustments:
Passband flatness
Frequency range: DC to 100 MHz
Impedance: 50 Ω
Test Conditions
The following setup and environmental conditions are required to ensure the PXIe-5413 meets published specifications:
The PXIe-5413 is warmed up for 15 minutes at ambient temperature. Warm up begins after the chassis is powered and the PXIe-5413 is recognized by the host and configured using NI-FGEN.
All test instruments are warmed up for at least the period of time stated in the documentation for each instrument and self-calibration is performed. The warm-up time ensures that the test instruments are at a stable operating temperature.
Keep cabling as short as possible. Long cables act as antennas, picking up extra noise that can affect measurements.
Verify that all connections to the PXIe-5413, including front panel connections and screws, are secure.
Use shielded copper wire for all cable connections to the device. Use twisted-pair wire to eliminate noise and thermal offsets.
Maintain an ambient temperature of 23 °C ± 3 °C.
Keep relative humidity between 10% and 90%, noncondensing.
4 | ni.com | PXIe-5413 Calibration Procedure
Page 5
Ensure that the PXI chassis fan speed is set to HIGH, that the fan filters (if present) are clean, and that the empty slots contain slot blockers and filler panels. For more information about cooling, refer to the Maintain Forced-Air Cooling Note to Users document available at ni.com/manuals.
Plug the PXI chassis and any other instrument standards into the same power strip to avoid ground loops.
Password
The default password for password-protected operations is NI.
Related Information
Adjusting the PXIe-5413 on page 14 Setting the Calibration Date and Due Date on page 18
Calibration Interval
Recommended calibration interval 2 years
As-Found and As-Left Limits
The as-found limits are the published specifications for the PXIe-5413. NI uses these limits to determine whether the PXIe-5413 meets the specifications when it is received for calibration. Use the as-found limits during initial verification.
The as-left calibration limits are equal to the published NI specifications for the PXIe-5413, less guard bands for measurement uncertainty, temperature drift, and drift over time. NI uses these limits to reduce the probability that the instrument will be outside the published specification limits at the end of the calibration cycle. Use the as-left limits when performing verification after adjustment.
Calibration Overview
Install the PXIe-5413 and configure it in NI Measurement & Automation Explorer (MAX) before calibrating.
Calibration includes the following steps:
1. Self-calibration—Let the PXIe-5413 warm up for 15 minutes before self-calibrating the PXIe-5413 by clicking the Self-Calibrate button in MAX or calling niFgen Self Cal.
2. Verification—Verify the existing operation of the device. This step confirms whether the device is operating within the published specification prior to adjustment.
3. Adjustment—Perform an external adjustment of the calibration constants of the device. The adjustment procedure automatically stores the calibration date and temperature on the EEPROM to allow traceability.
PXIe-5413 Calibration Procedure | © National Instruments | 5
Page 6
4. Self-calibration—Let the PXIe-5413 warm up for 15 minutes before self-calibrating the PXIe-5413 by clicking the Self-Calibrate button in MAX or calling niFgen Self Cal.
5. Reverification—Repeat the Verification procedure to ensure that the device is operating within the published specifications after adjustment.
Refer to the following sections to complete each procedure.
Related Information
Verifying the PXIe-5413 on page 6
Adjusting the PXIe-5413 on page 14
Verifying the PXIe-5413
The performance verification procedures assume that adequate traceable uncertainties are available for the calibration references.
Self-calibrate the PXIe-5413 and all test instruments before beginning the verification procedures.
Note Verification of the PXIe-5413 is complete only after you have successfully
completed all verification tests in this section using the as-found limits.
Related Information
Calibration Overview on page 5
Verifying Timebase Accuracy
Follow this procedure to verify the frequency accuracy of the PXIe-5413 onboard timebase (TCXO) using an oscilloscope.
Refer to the following table as you complete the following steps.
Table 3. Timebase Accuracy Verification
As-Found Limit As-Left Limit
±(1.5 ppm + 1 ppm per year1) ±0.28 ppm
1. Connect one end of the SMA (m)-to-SMA (m) cable to a channel of the PXIe-5413 and the other end to a channel of the oscilloscope.
Note Timebase accuracy verification can be performed on either channel of
the two-channel PXIe-5413.
2. Lock the oscilloscope to the precision frequency reference using the PXI Clock.
1
Where time drift starts at the latest external calibration date.
6 | ni.com | PXIe-5413 Calibration Procedure
Page 7
3. Configure the oscilloscope to acquire a 10 MHz sine wave with the following settings:
Vertical range: 5 V
pk-pk
Sample rate: 250 MS/s
Input impedance: 50 Ω
Minimum number of points: 1,048,576 samples
4. Configure the PXIe-5413 with the following settings:
Channel: Channel of the PXIe-5413 connected to the oscilloscope
Output impedance: 50 Ω
Output mode: Standard Function
Waveform type: Sine
Frequency: 10 MHz
Amplitude: 2.5 V
pk-pk
Offset: 0 V
Reference Clock source: Onboard Reference Clock
5. Wait 1 second for settling.
6. Measure the peak frequency (Hz) using the oscilloscope and the Extract Single Tone Information VI and record this as F
Measured
.
7. Calculate the timebase error using the following formula:
Timebase Error (ppm) =
Measured
10
7
10
8. Compare the frequency error to the appropriate limit from the Timebase Accuracy
Verification table.
Note Timebase accuracy verification is only required on one channel of the
two-channel PXIe-5413.
Verifying DC Voltage Accuracy
Follow this procedure to verify the DC voltage accuracy of the PXIe-5413 by comparing the voltage sourced by the PXIe-5413 to the value measured by the DMM.
Refer to the following table as you complete the following steps.
PXIe-5413 Calibration Procedure | © National Instruments | 7
Page 8
Table 4. DC Voltage Accuracy Verification
DMM
Range (V)
Amplitude
Range (V
pk-pk
)
Offset (V) Expected
Amplitude (V)
As-Found
Test Limit
(mV)
As-Left
Test Limit
(mV)
1 10 7.340 3.670 3.670 ±39.04 ±11.17
2 2.268 1.134 1.134 ±12.41 ±3.45
3 2.185 1.092 1.092 ±11.97 ±3.33
4 2.102 1.051 1.051 ±11.54 ±3.20
5 2.019 1.009 1.009 ±11.10 ±3.07
6 1 1.936 0.968 0.968 ±10.66 ±2.95
7 1.854 0.927 0.927 ±10.23 ±2.82
8 1.770 0.885 0.885 ±9.79 ±2.69
9 1.686 0.843 0.843 ±9.35 ±2.57
10 0.696 0.348 0.348 ±4.15 ±1.06
11 0.214 0.107 0.107 ±1.62 ±0.68
12 0.1 0.066 0.033 0.033 ±0.85 ±0.53
13 10 7.340 -3.670 -3.670 ±39.04 ±11.17
14 2.268 -1.134 -1.134 ±12.41 ±3.45
15 2.185 -1.092 -1.092 ±11.97 ±3.33
16 2.102 -1.051 -1.051 ±11.54 ±3.20
17 2.019 -1.009 -1.009 ±11.10 ±3.07
18 1 1.936 -0.968 -0.968 ±10.66 ±2.95
19 1.854 -0.927 -0.927 ±10.23 ±2.82
20 1.770 -0.885 -0.885 ±9.79 ±2.69
21 1.686 -0.843 -0.843 ±9.35 ±2.57
22 0.696 -0.348 -0.348 ±4.15 ±1.06
23 0.214 -0.107 -0.107 ±1.62 ±0.68
24 0.1 0.066 -0.033 -0.033 ±0.85 ±0.53
8 | ni.com | PXIe-5413 Calibration Procedure
Page 9
1. Connect one end of the SMA (m)-to-BNC (m) cable to CH 0 of the PXIe-5413. Connect the other end to the double banana plug-to-BNC (f) adapter, and connect the double banana plug-to-BNC (f) adapter to the HI and LO connectors of the DMM.
2. Configure the DMM with the following settings:
Function: DC Voltage
Resolution: 7.5 digits
Aperture time: 100 ms
Number of averages: 4
Auto zero: Enabled
ADC calibration: Enabled
Vertical range: DMM range listed in the DC Voltage Accuracy Verification table
3. Configure the PXIe-5413 with the following settings:
Channel: Channel of the PXIe-5413 connected to the DMM
Load impedance: 1 MΩ
Output mode: Standard Function
Waveform type: User
Amplitude: Amplitude range listed in the DC Voltage Accuracy Verification table
Offset: Offset listed in the DC Voltage Accuracy Verification table
Waveform data: Array of zeros
Waveform size: 8,192 samples
4. Wait 1 second for settling.
5. Measure the DC voltage (V) using the DMM and record this as DC Voltage
Measured
.
6. Calculate the DC voltage error using the following formula:
DC Voltage Error (mV) = (DC Voltage
Measured
- Expected Amplitude) * 10
3
where Expected Amplitude is the expected amplitude (V) listed in the DC Voltage
Accuracy Verification table.
7. Compare the DC voltage error to the appropriate limit from the DC Voltage Accuracy
Verification table.
8. Repeat steps 2 through 7 for each configuration listed in the DC Voltage Accuracy
Verification table.
9. If you are verifying a two-channel PXIe-5413, complete the following steps to verify CH 1:
a) Abort waveform generation on the PXIe-5413. b) Disconnect the SMA (m)-to-BNC (m) cable from CH 0 and connect it to CH 1 of the
PXIe-5413.
c) Repeat steps 2 through 8.
Verifying AC Amplitude Accuracy
Follow this procedure to verify the AC amplitude accuracy of the PXIe-5413 by comparing the amplitude sourced by the PXIe-5413 to the amplitude measured by the DMM.
Refer to the following table as you complete the following steps.
PXIe-5413 Calibration Procedure | © National Instruments | 9
Page 10
Table 5. AC Amplitude Accuracy Verification
DMM Range
(V
RMS
)
Amplitude Range
(V
pk-pk
)
Expected
Amplitude
(V
pk-pk
)
As-Found
Test Limit
(mV)
As-Left Test
Limit (mV)
1 5 7.34 7.34 ±74.40 ±32.66
2 2.27 2.27 ±23.70 ±10.10
3 2.18 2.18 ±22.80 ±9.70
4 2.1 2.1 ±22.00 ±9.35
5 2.02 2.02 ±21.20 ±8.99
6 1.94 1.94 ±20.40 ±8.63
7 1.85 1.85 ±19.50 ±8.23
8 1.77 1.77 ±18.70 ±7.88
9 1.68 1.68 ±17.80 ±7.48
10 0.5 0.69 0.69 ±7.90 ±3.07
11 0.21 0.21 ±3.10 ±1.50
12 0.05 0.065 0.065 ±1.65 ±0.81
1. Connect one end of the SMA (m)-to-BNC (m) cable to CH 0 of the PXIe-5413. Connect the other end to the double banana plug-to-BNC (f) adapter, and connect the double banana plug-to-BNC (f) adapter to the HI and LO connectors of the DMM.
2. Configure the DMM with the following settings:
Function: AC Voltage
Resolution: 6.5 digits
Aperture time: 100 ms
Number of averages: 4
Minimum frequency: 49 kHz
Auto zero: Enabled
ADC calibration: Enabled
Vertical range: DMM range listed in the AC Amplitude Accuracy Verification table
3. Configure the PXIe-5413 with the following settings:
Channel: Channel of the PXIe-5413 connected to the DMM
Load impedance: 1 MΩ
Output mode: Standard Function
Waveform type: Sine
Frequency: 50 kHz
10 | ni.com | PXIe-5413 Calibration Procedure
Page 11
Amplitude: Amplitude range listed in the AC Amplitude Accuracy Verification table
Offset: 0 V
4. Wait 1 second for settling.
5. Measure the AC amplitude (V
RMS
) using the DMM and record this as
AC Amplitude
Measured
.
6. Convert the recorded AC Amplitude
Measured
voltage from V
RMS
to V
pk-pk
using the
following formula:
pk pk
=
RMS
* 2 2
7. Calculate the AC amplitude error using the following formula:
AC Amplitude Error (mV) = (AC Amplitude
Measured
- Expected Amplitude) * 10
3
where Expected Amplitude is the expected amplitude (V
pk-pk
) listed in the AC Amplitude
Accuracy Verification table.
8. Compare the AC amplitude error to the appropriate limit from the AC Amplitude
Accuracy Verification table.
9. Repeat steps 2 through 8 for each configuration listed in the AC Amplitude Accuracy
Verification table.
10. If you are verifying a two-channel PXIe-5413, complete the following steps to verify CH 1:
a) Abort waveform generation on the PXIe-5413. b) Disconnect the SMA (m)-to-BNC (m) cable from CH 0 and connect it to CH 1 of the
PXIe-5413.
c) Repeat steps 2 through 9.
Verifying Passband Flatness Accuracy
Follow this procedure to verify the passband flatness accuracy of the PXIe-5413 by comparing the flatness sourced by the PXIe-5413 to the value measured by the power sensor.
Refer to the following table as you complete the following steps.
Table 6. Passband Flatness Accuracy Verification
Amplitude Range
(V
pk-pk
)
Frequency (MHz) As-Found Test
Limit (dB)
As-Left Test Limit
(dB)
1 3.500 0.05
2 1 ±0.40 ±0.25
3 20 ±0.60 ±0.14
PXIe-5413 Calibration Procedure | © National Instruments | 11
Page 12
Table 6. Passband Flatness Accuracy Verification (Continued)
Amplitude Range
(V
pk-pk
)
Frequency (MHz) As-Found Test
Limit (dB)
As-Left Test Limit
(dB)
5 1.040 0.05
6 1 ±0.40 ±0.25
7 20 ±0.40 ±0.11
11 0.956 0.05
12 1 ±0.40 ±0.25
13 20 ±0.40 ±0.11
17 0.867 0.05
18 1 ±0.40 ±0.25
19 20 ±0.40 ±0.11
23 0.780 0.05
24 1 ±0.40 ±0.25
25 20 ±0.40 ±0.11
29 0.321 0.05
30 1 ±0.40 ±0.25
31 20 ±0.40 ±0.11
35 0.099 0.05
36 1 ±0.40 ±0.25
37 20 ±0.40 ±0.11
1. Connect the SMA (m)-to-N (f) adapter to the power sensor.
2. Zero the power sensor. a) Ensure that the power sensor is not connected to any signals. b) Zero the power sensor using the built-in function, according to the power sensor
documentation.
3. Connect the power sensor with the SMA (m)-to-N (f) adapter attached to CH 0 of the PXIe-5413.
Note Ensure the power sensor is supported to the extent that excessive strain
is not applied to the connected channel of the PXIe-5413 or to the power sensor connector.
12 | ni.com | PXIe-5413 Calibration Procedure
Page 13
4. Configure the power sensor with the following settings:
Maximum noise ratio: 0.005 dB
Auto range: Enabled
Power measurement: Continuous average
Path selection: Automatic
Averaging: Automatic
Aperture: 20 ms
Timeout: 10 s
Frequency: Frequency listed in the Passband Flatness Accuracy Verification table
5. Configure the PXIe-5413 with the following settings:
Load impedance: 50 Ω
Output mode: Standard Function
Waveform type: Sine
Amplitude: Amplitude range listed in the Passband Flatness Accuracy Verification table
Frequency: Frequency listed in the Passband Flatness Accuracy Verification table
Offset: 0 V
6. Wait 1 second for settling.
7. Measure and record the power (dBm) using the power sensor.
When the power is measured at 0.05 MHz, record this as P
Measured at 0.05 MHz
.
When the power is measured at all frequencies other than 0.05 MHz, record this as
P
Measured at Current Test Frequency
.
8. Skip this step if the PXIe-5413 frequency is configured to 0.05 MHz. Calculate the passband flatness error using the following formula:
Passband Flatness Error (dB) = P
Measured at 0.05 MHz
- P
Measured at Current Test Frequency
where P
Measured at 0.05 MHz
(dBm) and P
Measured at Current Test Frequency
(dBm) are values for
power measured at the same amplitude range listed in the Passband Flatness Accuracy
Verification table.
9. Repeat steps 4 through 8 for each configuration listed in the Passband Flatness Accuracy
Verification table.
10. For each configuration, compare the passband flatness error to the appropriate limit from the Passband Flatness Accuracy Verification table.
11. If you are verifying a two-channel PXIe-5413, complete the following steps to verify CH 1:
a) Abort waveform generation on the PXIe-5413. b) Disconnect the SMA (m)-to-N (f) adapter and power sensor assembly from CH 0
and connect the assembly to CH 1 of the PXIe-5413.
c) Repeat steps 4 through 10.
PXIe-5413 Calibration Procedure | © National Instruments | 13
Page 14
Adjusting the PXIe-5413
Follow this procedure to adjust the PXIe-5413 to meet published specifications.
Note All adjustments must be completed in the same external calibration session.
1. Call niFgen ni5433 Open Ext Cal Session to obtain an external calibration session.
2. Complete the Adjusting Timebase procedure.
3. Complete the Adjusting DC Voltage procedure.
4. Complete the Adjusting Passband Flatness procedure.
5. Call niFgen ni5433 Close Ext Cal Session with the following settings:
Action: If the external adjustment procedure completed without any errors, set this control to Commit to store the new calibration constants, adjustment time, adjustment date, and adjustment temperature to the onboard memory. If any errors occurred during the external adjustment procedure, or if you want to abort the operation, set the control to Abort to discard the new calibration constants without changing any of the calibration data stored in the onboard memory.
Self-calibrate the PXIe-5413 after completing adjustment and before reverification.
Related Information
Password on page 5
Calibration Overview on page 5
Adjusting Timebase
Follow this procedure to adjust the internal timebase reference (TCXO) of the PXIe-5413.
1. Connect one end of the SMA (m)-to-SMA (m) cable to a channel of the PXIe-5413 and the other end to a channel of the oscilloscope.
Note Timebase adjustment can be performed on either channel of the two-
channel PXIe-5413.
2. Lock the oscilloscope to the precision frequency reference using the PXI Clock.
3. Configure the oscilloscope with the following settings:
Vertical range: 5 V
pk-pk
Sample rate: 250 MS/s
Input impedance: 50 Ω
Minimum number of points: 1,048,576 samples
4. Call niFgen ni5433 Timebase Cal Initialize with the following settings:
Channel: Channel of the PXIe-5413 connected to the oscilloscope
5. Call niFgen ni5433 Timebase Cal Configure.
6. Wait 1 second for settling.
7. Measure the frequency of the signal using the oscilloscope and the Extract Single Tone Information VI.
14 | ni.com | PXIe-5413 Calibration Procedure
Page 15
8. Call niFgen ni5433 Timebase Cal Adjust with the following settings:
Measured Frequency: Frequency of the waveform present on the connected channel of the PXIe-5413 and measured by the oscilloscope in the previous step
9. If the Timebase Cal Complete output of niFgen ni5433 Timebase Cal Adjust returns
FALSE, repeat steps 5 through 8 until Timebase Cal Complete returns TRUE.
Note Timebase adjustment is only required on one channel of the two-channel
PXIe-5413.
Adjusting DC Voltage
Follow this procedure to adjust the DC voltage of the PXIe-5413.
1. Connect one end of the SMA (m)-to-BNC (m) cable to a channel of the PXIe-5413. Connect the other end to the double banana plug-to-BNC (f) adapter, and connect the double banana plug-to-BNC (f) adapter to the HI and LO connectors of the DMM.
Note DC voltage adjustment can be performed on either channel of the two-
channel PXIe-5413.
2. Call niFgen ni5433 ADC Cal Initialize with the following settings:
Channel: Channel of the PXIe-5413 connected to the DMM
3. Call niFgen ni5433 ADC Cal Configure to obtain the Approximate Voltage Generated (V
pk-pk
) output.
4. Wait 1 second for settling.
5. Configure the DMM with the following settings:
Function: DC Voltage
Resolution: 7.5 digits
Aperture time: 100 ms
Number of averages: 4
Auto zero: Enabled
ADC calibration: Enabled
Range: Range returned by the Approximate Voltage Generated (V
pk-pk
) output of
niFgen ni5433 ADC Cal Configure in step 3
6. Measure the voltage of the signal using the DMM.
7. Call niFgen ni5433 ADC Cal Adjust with the following settings:
Measured Voltage: Voltage of the waveform present on the connected channel of
the PXIe-5413 and measured by the DMM in the previous step
8. If the ADC Cal Complete output of niFgen ni5433 Timebase Cal Adjust returns FALSE, repeat steps 3 through 7 until ADC Cal Complete returns TRUE.
Note DC voltage adjustment is only required on one channel of the two-channel
PXIe-5413.
PXIe-5413 Calibration Procedure | © National Instruments | 15
Page 16
Adjusting Passband Flatness
Complete one of the following procedures depending on the channel count of the PXIe-5413 you are adjusting:
Adjusting Passband Flatness of a One-Channel PXIe-5413
Adjusting Passband Flatness of a Two-Channel PXIe-5413
Adjusting Passband Flatness of a One-Channel PXIe-5413
Follow this procedure to adjust the passband flatness of a one-channel PXIe-5413.
Note Ensure the power sensor, with the SMA (m)-to-N (f) adapter attached, has
been zeroed within the past hour.
1. Call niFgen ni5433 Flatness Cal Initialize.
2. Call niFgen ni5433 Flatness Cal Configure.
3. Call niFgen ni5433 Flatness Cal Adjust with the following settings:
Measured Power (dBm): 0 dBm
4. Connect the power sensor with the SMA (m)-to-N (f) adapter attached to CH 0 of the PXIe-5413.
5. Call niFgen ni5433 Flatness Cal Configure to obtain the Current Configuration output.
6. Configure the power sensor with the following settings:
Auto range: Enabled
Maximum noise ratio: 0.005 dB
Power measurement: Continuous average
Path selection: Automatic
Averaging: Automatic
Aperture: 20 ms
Timeout: 10 s
Frequency: Frequency returned by the Current Configuration output of niFgen
ni5433 Flatness Cal Configure in the previous step
7. Wait 1 second for settling.
8. Measure the power of the signal using the power sensor.
9. Call niFgen ni5433 Flatness Cal Adjust with the following settings:
Measured Power (dBm): Power of the waveform present on CH 0 of the
PXIe-5413 and measured by the power sensor in the previous step in dBm
10. If the Flatness Cal Complete output of niFgen ni5433 Timebase Cal Adjust returns
FALSE, repeat steps 5 through 9 until Flatness Cal Complete returns TRUE.
Adjusting Passband Flatness of a Two-Channel PXIe-5413
Follow this procedure to adjust the passband flatness of a two-channel PXIe-5413.
Note Ensure the power sensor, with the SMA (m)-to-N (f) adapter attached, has
been zeroed within the past hour.
1. Call niFgen ni5433 Flatness Cal Initialize.
16 | ni.com | PXIe-5413 Calibration Procedure
Page 17
2. Call niFgen ni5433 Flatness Cal Configure.
3. Call niFgen ni5433 Flatness Cal Adjust with the following settings:
Measured Power (dBm): 0 dBm
4. Connect the power sensor with the SMA (m)-to-N (f) adapter attached to CH 0 of the PXIe-5413.
5. Call niFgen ni5433 Flatness Cal Configure to obtain the Current Configuration and Next Configuration outputs.
6. Configure the power sensor with the following settings:
Auto range: Enabled
Maximum noise ratio: 0.005 dB
Power measurement: Continuous average
Path selection: Automatic
Averaging: Automatic
Aperture: 20 ms
Timeout: 10 s
Frequency: Frequency returned by the Current Configuration output of niFgen
ni5433 Flatness Cal Configure in the previous step
7. Wait 1 second for settling.
8. Measure the power of the signal using the power sensor.
9. Call niFgen ni5433 Flatness Cal Adjust with the following settings:
Measured Power (dBm): Power of the waveform present on CH 0 of the
PXIe-5413 and measured by the power sensor in the previous step in dBm
10. If the Next Configuration output of niFgen ni5433 Flatness Cal Configure obtained in step 5 returns CH 0, repeat steps 5 through 9. If the Next Configuration output of niFgen ni5433 Flatness Cal Configure obtained in step 5 returns CH 1, continue to step 11.
11. Connect the SMA (m)-to-N (f) adapter and power sensor assembly to CH 1 of the PXIe-5413.
12. Call niFgen ni5433 Flatness Cal Configure to obtain the Current Configuration output.
13. Configure the power sensor with the following settings:
Auto range: Enabled
Maximum noise ratio: 0.005 dB
Power measurement: Continuous average
Path selection: Automatic
Averaging: Automatic
Aperture: 20 ms
Timeout: 10 s
Frequency: Frequency returned by the Current Configuration output of niFgen
ni5433 Flatness Cal Configure in the previous step
14. Wait 1 second for settling.
15. Measure the power of the signal using the power sensor.
PXIe-5413 Calibration Procedure | © National Instruments | 17
Page 18
16. Call niFgen ni5433 Flatness Cal Adjust with the following settings:
Measured Power (dBm): Power of the waveform present on CH 1 of the PXIe-5413 and measured by the power sensor in the previous step in dBm
17. If the Flatness Cal Complete output of niFgen ni5433 Timebase Cal Adjust returns
FALSE, repeat steps 12 through 16 until Flatness Cal Complete returns TRUE.
Reverification
Repeat the Verification section to determine the as-left status of the PXIe-5413.
Note If any test fails reverification after performing an adjustment, verify that you
have met the test conditions before returning your PXIe-5413 to NI. Refer to the
Worldwide Support and Services section for information about support resources or
service requests.
Setting the Calibration Date and Due Date
After completing all verifications or reverifications, set the calibration date and a calibration due date for the PXIe-5413 using either Measurement Automation Explorer (MAX) or the LabVIEW NI System Configuration API.
NI recommends a calibration due date of the date of external calibration plus the external calibration interval for the PXIe-5413.
MAX—Navigate to the External Calibration section of the Settings tab and update the Calibration Date and Calibration Due Date entries. Click Save.
LabVIEW NI System Configuration API—Use the Update Calibration VI to set the calibration date and calibration due date.
The calibration due date can be either a specific date or an interval in months.
Related Information
Password on page 5
Worldwide Support and Services
The NI website is your complete resource for technical support. At ni.com/support, you have access to everything from troubleshooting and application development self-help resources to email and phone assistance from NI Application Engineers.
Visit ni.com/services for NI Factory Installation Services, repairs, extended warranty, and other services.
Visit ni.com/register to register your NI product. Product registration facilitates technical support and ensures that you receive important information updates from NI.
18 | ni.com | PXIe-5413 Calibration Procedure
Page 19
A Declaration of Conformity (DoC) is our claim of compliance with the Council of the European Communities using the manufacturer’s declaration of conformity. This system affords the user protection for electromagnetic compatibility (EMC) and product safety. You can obtain the DoC for your product by visiting ni.com/certification. If your product supports calibration, you can obtain the calibration certificate for your product at ni.com/calibration.
NI corporate headquarters is located at 11500 North Mopac Expressway, Austin, Texas, 78759-3504. NI also has offices located around the world. For telephone support in the United States, create your service request at ni.com/support or dial 1 866 ASK MYNI (275 6964). For telephone support outside the United States, visit the Worldwide Offices section of ni.com/
niglobal to access the branch office websites, which provide up-to-date contact information,
support phone numbers, email addresses, and current events.
PXIe-5413 Calibration Procedure | © National Instruments | 19
Page 20
Information is subject to change without notice. Refer to the NI Trademarks and Logo Guidelines at ni.com/trademarks for information on NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents. You can find information about end-user license agreements (EULAs) and third-party legal notices in the readme file for your NI product. Refer to the Export Compliance Information at ni.com/legal/export-compliance for the NI global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data. NI MAKES NO EXPRESS OR IMPLIED WARRANTIES AS TO THE ACCURACY OF THE INFORMATION CONTAINED HEREIN AND SHALL NOT BE LIABLE FOR ANY ERRORS. U.S. Government Customers: The data contained in this manual was developed at private expense and is subject to the applicable limited rights and restricted data rights as set forth in FAR 52.227-14, DFAR 252.227-7014, and DFAR 252.227-7015.
© 2017—2018 National Instruments. All rights reserved.
376945B-01 April 24, 2018
Loading...