National Instruments PXI-1006 User Manual

PXI

PXI-1006 User Manual

PXI-1006 User Manual
February 2001 Edition
Part Number 323006A-01

Support

Worldwide Technical Support and Product Information

National Instruments Corporate Headquarters

11500 North Mopac Expressway Austin, Texas 78759-3504 USA Tel: 512 794 0100

Worldwide Offices

Australia 03 9879 5166, Austria 0662 45 79 90 0, Belgium 02 757 00 20, Brazil 011 284 5011, Canada (Calgary) 403 274 9391, Canada (Ottawa) 613 233 5949, Canada (Québec) 514 694 8521, China (Shanghai) 021 6555 7838, China (ShenZhen) 0755 3904939, Denmark 45 76 26 00, Finland 09 725 725 11, France 01 48 14 24 24, Germany 089 741 31 30, Greece 30 1 42 96 427, Hong Kong 2645 3186, India 91805275406, Israel 03 6120092, Italy 02 413091, Japan 03 5472 2970, Korea 02 596 7456, Mexico 5 280 7625, Netherlands 0348 433466, New Zealand 09 914 0488, Norway 32 27 73 00, Poland 0 22 528 94 06, Portugal 351 1 726 9011, Singapore 2265886, Spain 91 640 0085, Sweden 08 587 895 00, Switzerland 056 200 51 51, Taiwan 02 2528 7227, United Kingdom 01635 523545
For further support information, see the Technical Support Resources appendix. To comment on the documentation, send e-mail to techpubs@ni.com
© Copyright 2001 National Instruments Corporation. All rights reserved.

Important Information

Warranty

The PXI-1006 is warranted against defects in materials and workmanship for a period of one year from the date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that proves to be defective during the warranty period. This warranty includes parts and labor.
The media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects in materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace software media that do not execute programming instructions if National Instruments receives notice of such defects during the warranty period. National Instruments does not warrant that the operation of the software shall be uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are covered by warranty.
National Instruments believes that the information in this document is accurate. The document has been carefully reviewed for technical accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent editions of this document without prior notice to holders of this edition. The reader should consult National Instruments if errors are suspected. In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.
XCEPT AS SPECIFIED HEREIN,NATIONAL INSTRUMENTS MAKES NO W ARRANTIES, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF
E
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE
NATIONAL INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID B Y THE CUSTOMER.NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR
DAMAGES RESULTING FROM LOSS OF DATA
. This limitation of the liability of National Instruments will apply regardless ofthe form of action,whether in contract ortort, including
THEREOF
negligence. Any action against National Instruments must be brought within one year after the cause of action accrues. National Instruments shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not cover damages, defects, malfunctions, or service failures caused by owner’s failure to follow the National Instruments installation, operation, or maintenance instructions; owners modification of the product; owners abuse, misuse, or negligent acts; and power failure or surges, fire, flood, accident, actions of third parties, or other events outside reasonable control.
, PROFITS, USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY

Copyright

Under the copyright laws, this publication may not be reproduced ortransmitted in any form, electronic or mechanical, including photocopying, recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National Instruments Corporation.

Trademarks

National Instruments™,ni.com™,andPXI™are trademarks of National Instruments Corporation.
Product and company names mentioned herein are trademarks or trade names of their respective companies.
.CUSTOMERS RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART OF

WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS

(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING FOR A LEVEL OF RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL IMPLANTS OR AS CRITICAL COMPONENTS IN ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE TO PERFORM CAN REASONABLY BE EXPECTED TO CAUSE SIGNIFICANT INJURY TO A HUMAN.
(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE SOFTWARE PRODUCTS CAN BE IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT LIMITED TO FLUCTUATIONS IN ELECTRICAL POWER SUPPLY, COMPUTER HARDWARE MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE FITNESS, FITNESS OF COMPILERS AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION, INSTALLATION ERRORS, SOFTWARE AND HARDWARE COMPATIBILITY PROBLEMS, MALFUNCTIONS OR FAILURES OF ELECTRONIC MONITORING OR CONTROL DEVICES, TRANSIENT FAILURES OF ELECTRONIC SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR MISUSES, OR ERRORS ON THE PART OF THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS THESE ARE HEREAFTER COLLECTIVELY TERMED SYSTEM FAILURES). ANY APPLICATION WHERE A SYSTEM FAILURE WOULD CREATE A RISK OF HARM TO PROPERTY OR PERSONS (INCLUDING THE RISK OF BODILY INJURY AND DEATH) SHOULD NOT BE RELIANT SOLELY UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM FAILURE. TO AVOID DAMAGE, INJURY, OR DEATH, THE USER OR APPLICATION DESIGNER MUST TAKE REASONABLY PRUDENT STEPS TO PROTECT AGAINST SYSTEM FAILURES, INCLUDING BUT NOT LIMITED TO BACK-UP OR SHUT DOWN MECHANISMS. BECAUSE EACH END-USER SYSTEM IS CUSTOMIZED AND DIFFERS FROM NATIONAL INSTRUMENTS' TESTING PLATFORMS AND BECAUSE A USER OR APPLICATION DESIGNER MAY USE NATIONAL INSTRUMENTS PRODUCTS IN COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT EVALUATED OR CONTEMPLATED BY NATIONAL INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS ULTIMATELY RESPONSIBLE FOR VERIFYING AND VALIDATING THE SUITABILITY OF NATIONAL INSTRUMENTS PRODUCTS WHENEVER NATIONAL INSTRUMENTS PRODUCTS ARE INCORPORATED IN A SYSTEM OR APPLICATION, INCLUDING, WITHOUT LIMITATION, THE APPROPRIATE DESIGN, PROCESS AND SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.

Compliance

FCC/Canada Radio Frequency Interference Compliance*
Determining FCC Class
The Federal Communications Commission (FCC) has rules to protect wireless communications from interference. The FCC places digital electronics into two classes. These classes are known as Class A (for use in industrial-commercial locations only) or Class B (for use in residential or commercial locations). Depending on where it is operated, this product could be subject to restrictions in the FCC rules. (In Canada, the Department of Communications (DOC), of Industry Canada, regulates wireless interference in much the same way.)
Digital electronics emit weak signals during normal operation that can affect radio, television, or other wireless products. By examining the product you purchased, you can determine the FCC Class and therefore which of the two FCC/DOC Warnings apply in the following sections. (Some products may not be labeled at all for FCC; if so, the reader should then assume these are Class A devices.)
FCC Class A products only display a simple warning statement of one paragraph in length regarding interference and undesired operation. Most of our products are FCC Class A. The FCC rules have restrictions regarding the locations where FCC Class A products can be operated.
FCC Class B products display either a FCC ID code, starting with the letters EXN, or the FCC Class B compliance mark that appears as shown here on the right.
Consult the FCC web site
http://www.fcc.gov
FCC/DOC Warnings
This equipment generates and uses radio frequency energyand, if not installed and used in strict accordance with the instructions in this manual and the CE Mark Declaration of Conformity**, may cause interference to radio and television reception. Classification requirements are the same for the Federal Communications Commission (FCC) and the Canadian Department of Communications (DOC).
Changes or modifications not expressly approved by National Instruments could void the users authority to operate the equipment under the FCC Rules.
for more information.
Class A
Federal Communications Commission
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Canadian Department of Communications
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.
Class B
Federal Communications Commission
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
Canadian Department of Communications
This Class B digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations. Cet appareil numérique de la classe B respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.
European Union - Compliance to EEC Directives
Readers in the EU/EEC/EEA must refer to the Manufacturer's Declaration of Conformity (DoC) for information** pertaining to the CE Mark compliance scheme.The Manufacturer includes a DoC for most every hardware product except for those bought for OEMs, if also available from an original manufacturer that also markets in the EU, or where compliance is not required as for electrically benign apparatus or cables.
* Certain exemptions may apply in the USA, see FCC Rules §15.103 Exempted devices,and§15.105(c). Also available in
sections of CFR 47.
** The CE Mark Declaration of Conformity will contain important supplementary information and instructions for the user or
installer.

Contents

About This Manual
Conventions Used in This Manual.................................................................................ix
Related Documentation..................................................................................................ix
For Your Safety
Connecting Safety Ground............................................................................................. xi
Chapter 1 Installing and Verifying Operation
Unpacking the PXI-1006 and Installing a Controller ....................................................1-1
Chapter 2 PXI-1006 Features
Key Features ..................................................................................................................2-1
PXI-1006 Backplane Overview .....................................................................................2-2
Interoperability with CompactPCI ..................................................................2-2
System Controller Slot .................................................................................... 2-4
Star Trigger Slot ..............................................................................................2-4
Peripheral Slots................................................................................................2-4
Local Bus.........................................................................................................2-4
Trigger Bus......................................................................................................2-6
System Reference Clock..................................................................................2-6
Rack Mounting ..............................................................................................................2-6
Remote Power Monitoring and Inhibiting Interface......................................................2-7
Using the Inhibiting Feature............................................................................ 2-7
Using the Power Monitoring Feature .............................................................. 2-8
Installing Filler Panels ...................................................................................................2-9
Using the Chassis Initialization File ..............................................................................2-9
Cleaning .........................................................................................................................2-9
Interior Cleaning..............................................................................................2-9
Exterior Cleaning............................................................................................. 2-9
Cleaning the Fan Filters................................................................................... 2-10
Chapter 3 Troubleshooting
Troubleshooting the PXI-1006 ......................................................................................3-1
Resetting the AC Mains Circuit Breaker .......................................................................3-1
© National Instruments Corporation vii PXI-1006 User Manual
Contents
Appendix A Specifications
Appendix B Pinouts
Appendix C Technical Support Resources
Glossary
Index

Figures

Figure 1-1. Installing Controller and PXI Modules ................................................. 1-2
Figure 2-1. Front View of the PXI-1006 Chassis.................................................... 2-3
Figure 2-2. Rear View of the PXI-1006 Chassis ..................................................... 2-3
Figure 2-3. PXI Local Bus and Star Trigger Routing..............................................2-5
Figure A-1. PXI-1006 Dimensions........................................................................... A-6

Tables

Table 2-1. DB-9 Connector Pinout ........................................................................ 2-7
Table 2-2. Power Supply Voltages at Power Monitoring Connector (DB-9) ........ 2-8
Table 3-1. Troubleshooting .................................................................................... 3-1
Table B-1. P1 (J1) Connector Pinout for the System Controller Slot..................... B-2
Table B-2. P2 (J2) Connector Pinout for the System Controller Slot..................... B-3
Table B-3. P1 (J1) Connector Pinout for the Star Trigger Slot...............................B-4
Table B-4. P2 (J2) Connector Pinout for the Star Trigger Slot...............................B-5
Table B-5. P1 (J1) Connector Pinout for the Peripheral Slot.................................. B-6
Table B-6. P2 (J2) Connector Pinout for the Peripheral Slot.................................. B-7
PXI-1006 User Manual viii ni.com

About This Manual

The PXI-1006 User Manual describes the features of the PXI-1006 chassis and contains information about configuring the chassis, installing the modules, and operating and using the PXI-1006.

Conventions Used in This Manual

The following conventions appear in this manual:
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to avoid injury, data loss, or a system crash.
This icon denotes a warning, which advises you of precautions to take to avoid being electrically shocked.
bold Bold text denotes items that you must select or click on in the software,
such as menu items and dialog box options. Bold text also denotes parameter names.
italic Italic text denotes variables, emphasis, a cross reference, or an introduction
to a key concept. This font also denotes text that is a placeholder for a word or value that you must supply.
monospace
Text in this font denotes text or characters that you should enter from the keyboard, sections of code, programming examples, and syntax examples. This font is also used for the proper names of disk drives, paths, directories, programs, subprograms, subroutines, device names, functions, operations, variables, filenames and extensions, and code excerpts.

Related Documentation

The following documents contain information that you may find helpful as you read this manual:
Compact PCI Specification PICMG 2.0 R3.0
PXI Specification Revision 2.0
IEEE 1101.1-1991, IEEE Standard for Mechanical Core
Specifications for Microcomputers Using IEC 603-2 Connectors
© National Instruments Corporation ix PXI-1006 User Manual
About This Manual
IEEE 1101.10 and P1101.11, IEEE Standard for Additional
Mechanical Specifications for Microcomputers Using IEEE 1101.1 Equipment Practice
PXI-1006 User Manual x ni.com

For Your Safety

Caution
carefully read the warning and caution notices.
Before undertaking any troubleshooting, maintenance, or exploratory procedure,
This equipment contains voltage hazardous to human life and safety, and can inflict personal injury.
Chassis GroundingThe PXI-1006 requires a connection from the
premise wire safety ground to the PXI-1006 chassis ground. The earth safety ground must be connected during use of this equipment to minimize shock hazards. See the Connecting Safety Ground section below.
Live CircuitsOperating personnel and service personnel must not
remove protective covers when operating or servicing the PXI-1006.
Explosive AtmosphereDo not operate the chassis in conditions
where flammable gases are present. Under such conditions this equipment is unsafe and may ignite the gases or gas fumes.
ModificationDo not modify any part of the chassis from its original
condition. Unsuitable modifications may result in safety hazards.

Connecting Safety Ground

Warning
that connects the ground line to the chassis ground. To minimize shock hazard, make sure your electrical power outlet has an appropriate earth safety ground that is connected whenever you power up the chassis.
The PXI-1006 chassis is designed with a three-position NEMA 15-5 style plug
If your power outlet does not have an appropriate ground connection, you must connect the premise wire safety ground to the chassis grounding screw located on the rear panel. To connect the safety ground, complete the following steps:
1. Connect a 16 AWG (1.3 mm) wire to the chassis grounding screw on the rear of the chassis using a toothed grounding lug. The wire must have green insulation with a yellow stripe or must be noninsulated (bare).
2. Attach the opposite end of the wire to permanent earth ground using toothed washers or a toothed lug.
© National Instruments Corporation xi PXI-1006 User Manual
1
Installing and Verifying Operation
This chapter describes how to unpack the PXI-1006, install a controller, and verify that the chassis is operating correctly.

Unpacking the PXI-1006 and Installing a Controller

This installation example assumes you are using an embedded PXI controller.
1. Unpack your PXI-1006 chassis.
2. Make sure the power switch is in the Standby (Off) position.
3. Plug in the AC power cord.
Power cords are available from National Instruments, or use a power cord that meets the requirements of ANSI/UL817 for use in the United States, CSA C22.2 21 and 49 for use in Canada, and IEC 227 or 245 for use in the European Union. AC mains power supply cords used with the PXI-1006 in other countries must be approved by the authority having jurisdiction in that country.
© National Instruments Corporation 1-1 PXI-1006 User Manual
Chapter 1 Installing and Verifying Operation
4. Install your controller in Slot 1 of your PXI-1006. Install peripheral cards, if any, in slots 2 through 18.
1
4
2
3
1 PXI-1006 Chassis 2 Peripheral Card
3 Ejector Handle in Down Position 4 PXI Controller

Figure 1-1. Installing Controller and PXI Modules

5. Set the fan speed selector switch (located on the rear of the chassis) to either HI for maximum cooling or AUTO to employ the temperature sensing module that controls the fan speed.
6. Power on the chassis. Verify that all fans become operational.
7. Verify that your controller boots the operating system.
Note
If the chassis does not power on, see Chapter 3, Troubleshooting.
PXI-1006 User Manual 1-2 ni.com
PXI-1006 Features
This chapter describes the key features of the PXI-1006 chassis and lists optional equipment you can order from National Instruments.

Key Features

The PXI-1006 combines a high-performance 18-slot PXI backplane with a high-output power supply and a structural design that has been optimized for maximum usability in a wide range of applications. The chassis modular design ensures the highest level of maintainability resulting in a very low mean time to repair (MTTR). The PXI-1006 fully complies with the PXI Specification, Revision 2.0, offering advanced timing and synchronization features.
The key features of the PXI-1006 include:
PXI and CompactPCI (PICMG 2.0 R 3.0) module compatibility
3U-sized, 18-slot chassis
600 W of usable power
Universal AC input: auto-voltage and auto-frequency ranging
Over-current protection via push-reset circuit breaker (no loose fuses
to replace)
Removable modular power supply
Remote power status and inhibit via a rear panel connector
On/Off (Standby) switch located on the front panel for easy access
Selectable fan speed for maximum cooling or quiet operation
Carrying handle for portability
Tilt feet for bench-top applications
Rack mount compatibility
2
© National Instruments Corporation 2-1 PXI-1006 User Manual
Chapter 2 PXI-1006 Features

PXI-1006 Backplane Overview

Interoperability with CompactPCI

The PXI-1006 backplane is interoperable with PXI-compatible products and standard CompactPCI products. This is an important feature, as many PXI-compatible systems may require components that do not implement PXI-specific features. For example, you may want to use a standard CompactPCI network interface card in a PXI chassis.
The signals on the P1 connector of the backplane meet the requirements of the CompactPCI specification for both the peripheral and system modules.
The PXI-specific signals are located on P2 and are found on only the signals that are reserved or not used in the CompactPCI 64-bit specification. Therefore, all modules that meet the requirements of the CompactPCI 64-bit specification will function in the PXI-1006.
The PXI-1006 has three PCI bus segments linked via two 64-bit PCI-to-PCI bridges. The PCI bus segment divisions are represented on the front of the chassis by vertical bars. See Figure 2-1 for the PCI segment divider locations.
Figures 2-1 and 2-2 show some of the key features and components of the PXI-1006 chassis. Figure 2-1 shows the front view of the PXI-1006. Figure 2-2 shows the rear view.
PXI-1006 User Manual 2-2 ni.com
Chapter 2 PXI-1006 Features
1
2
3
1 Remote On/Off (Standby) Switch 2 Optional Mounting Brackets 3 Removable Feet
4
6
5
4 Three Controller Expansion Slots 5 System Controller Slot 6 Star Trigger/Peripheral Slot
5
4
3
1 Fan Speed Selector 2 Remote Power Status and Inhibit Connector
7
Figure 2-1.
8
Front View of the PXI-1006 Chassis
3 Chassis Ground Screw 4 Universal AC Input
9
7 Segment Break 1 8 17 Peripheral Slots 9 Segment Break 2
1
AUTO
5 Circuit Breake
2
3
2
r

Figure 2-2. Rear View of the PXI-1006 Chassis

© National Instruments Corporation 2-3 PXI-1006 User Manual
Chapter 2 PXI-1006 Features

System Controller Slot

The System Controller slot is Slot 1 of the chassis as defined by the PXI specification. It has three controller expansion slots, which are used for system controller modules that are wider than one slot. As defined in the PXI specification, these slots allow the controller to expand to the left to prevent the controller from using up peripheral slots.

Star Trigger Slot

The Star Trigger (ST) slot is Slot 2. This slot has a dedicated trigger line between itself and slots 3-15. Slots 16-18 do not use the Star Trigger functionality because there are 13 Star Trigger lines, as specified in the PXI specification (see Figure 2-3). This slot is intended for modules with ST functionality that can provide individual triggers to the peripherals installed in slots 3-15. However, if you do not require advanced trigger functionality, you can install any standard peripheral module into this slot.

Peripheral Slots

There are 17 peripheral slots including the Star Trigger controller slot.

Local Bus

The PXI backplanes local bus is a daisy-chained bus that connects each peripheral slot with its adjacent peripheral slots to the left and right, as shown in Figure 2-3.
For example, a given peripheral slots right local bus connects to the adjacent slot’s left local bus, and so on. Each local bus is 13 lines wide and can pass analog signals between cards or provide a high-speed side-band communication path that does not affect the PXI bandwidth.
The PXI-1006 has local bus connections between all slots except slots 1 and 2, in accordance with the PXI specification. See Figure 2-3 for details.
PXI-1006 User Manual 2-4 ni.com
Clock 10 Buffer and Detection Circuitry
PXI-1006 Backplane Configuration
Chapter 2 PXI-1006 Features
18
Bus
Local
17
Bus
Local
16
Bus
Local
15
Bus
Local
14
Bus
Local
PXI Trigger Bus Segment 3 and PCI Bus Segment 3
13
Bus
Local
12
Bus
Local
11
Bus
Local
10
Bus
Local
9
Bus
Local
8
Bus
Local
PXI Trigger Bus Segment 2 and PCI Bus Segment 2
7
Bus
Local
6
Bus
Local
5
Bus
Local
4
Bus
Local
3
Bus
Local
Star Trigger Controller
2
2
PXI Trigger Bus Segment 1 and PCI Bus Segment 1
PXI Star Triggers
System Controller
1

Figure 2-3. PXI Local Bus and Star Trigger Routing

© National Instruments Corporation 2-5 PXI-1006 User Manual
Chapter 2 PXI-1006 Features

Trigger Bus

The PXI-1006 has three PXI segments. Vertical bars represent the segment divisions on the front of the chassis. (See Figure 2-1 for the segment divider locations.) The PXI trigger bus is divided into three separate trigger buses at these locations. You can use the trigger lines in a variety of ways. For example, you can use triggers to synchronize the operation of several different PXI peripheral modules. In other applications, one module can control carefully timed sequences of operations performed on other modules in the system. Modules can pass triggers to one another, allowing precisely timed responses to asynchronous external events the system is monitoring or controlling.

System Reference Clock

The PXI-1006 supplies the PXI 10 MHz system clock signal (PXI_CLK10) independently to every peripheral slot. An independent buffer (having a source impedance matched to the backplane and a skew of less than 1 ns between slots) drives the clock signal to each peripheral slot. You can use this common reference clock signal to synchronize multiple modules in a measurement or control system. You can drive PXI_CLK10 from an external source through the PXI_CLK10_IN pin on the P2 connector of the star trigger slot. (See Table B-1, P1 (J1) Connector Pinout for the System
Controller Slot.) Sourcing an external clock on this pin automatically
disables the backplanes 10 MHz source.

Rack Mounting

Rack-mount applications require the optional rack-mount kit available from National Instruments. Refer to the instructions supplied with the rack-mount kit to install your PXI-1006 in a standard 19 in. instrument rack.
PXI-1006 User Manual 2-6 ni.com
Chapter 2 PXI-1006 Features

Remote Power Monitoring and Inhibiting Interface

The PXI-1006 chassis supports remote power monitoring and inhibiting via a 9-pin D-sub connector located on the rear panel. Table 2-1 shows the pinout of the DB-9 connector.

Table 2-1. DB-9 Connector Pinout

DB-9 Pin Signal
1 Logic Ground
2 +5 V
3 Reserved
4 +3.3 V
5 Inhibit*
6 +12 V
7 Reserved
8 –12 V
9 Logic Ground
7
6
1
2
89
3
4
5

Using the Inhibiting Feature

You can use the Inhibit signal (active low) to turn off the power supply outputs. To use this feature, connect the Inhibit pin (pin 5) to a Logic Ground pin (pin 1 or 9). Make sure the front (standby) switch is in the ON position. As long as the connection is made, the power supply inhibits its DC outputs. DC output resumes when Inhibit is no longer connected. Note that the power (standby) switch, located on the front of the chassis, uses this inhibiting feature. For remote reset, connect a momentary switch between pin5andpin1(orpin9).
© National Instruments Corporation 2-7 PXI-1006 User Manual
Chapter 2 PXI-1006 Features

Using the Power Monitoring Feature

You can use a digital voltmeter to ensure all voltage levels in your PXI-1006 are within the allowable limits.
Caution
When connecting digital voltmeter probes to the rear D-sub connector, be careful
not to short the probe leads together. Doing so could damage the power supply.
Referring to Table 2-2, connect one lead of the voltmeter to a supply pin on the remote power monitoring connector (9-pin D-sub) on the rear panel. Refer to Table 2-1 for a pinout diagram of the remote power monitoring connector. Connect the reference lead of the voltmeter to one of the ground pins. Compare each voltage reading to the values listed in Table 2-2.
Note
Use the rear-panel D-sub connector to check voltages only. Do not use the connector
to supply power to external devices.

Table 2-2. Power Supply Voltages at Power Monitoring Connector (DB-9)

Pin Supply Acceptable Voltage Range
2 +5 V 4.75 to 5.25 V
4 +3.3 V 3.135 to 3.465 V
6 +12 V 11.4 to 12.6 V
8 –12 V –12.6 to –11.4 V
1, 9 Logic Ground N/A
If the voltages fall within the specified ranges, the chassis complies with the CompactPCI voltage limit specifications. Notice that the rear-panel D-sub connector is to be used to check voltages only. Do not use these voltages to supply power to external devices.
Note
If the fans or power unit fail to function properly, refer to Chapter 3,
Troubleshooting.
PXI-1006 User Manual 2-8 ni.com

Installing Filler Panels

To optimize module cooling performance, install filler panels into unused or empty slots. Secure with the captive mounting screws.

Using the Chassis Initialization File

To assist system integrators, the PXI specification requires manufacturers of PXI chassis and system modules to document the capabilities of their products. The minimum documentation requirements are contained in read the use this file. The PXI-1006 chassis initialization file, is included on the diskette for your PXI-1006.
files, which consist of ASCII text. The system integrator can
.ini
file, and configuration utilities and device drivers can also
.ini

Cleaning

Cleaning procedures consist of exterior and interior cleaning of the chassis and cleaning the fan filters. Refer to your module user documentation for information on cleaning the individual CompactPCI or PXI modules.
Chapter 2 PXI-1006 Features
chassis.ini
,
Caution
servicing the chassis.
Mise en garde
de nettoyer ou dintervenir sur le châssis.
Always power-off the chassis and disconnect the power cord before cleaning or
Veillez àéteindrelechâssis et déconnecter le cordon d’alimentation avant

Interior Cleaning

Use a dry, low-velocity stream of air to clean the interior of the chassis. Use a soft-bristle brush for cleaning around components. If you must use a liquid for minor interior cleaning, use a 75% isopropyl alcohol solution and rinse with deionized water.

Exterior Cleaning

Clean the exterior surfaces of the chassis with a dry lint-free cloth or a soft-bristle brush. If any dirt remains, wipe with a cloth moistened in a mild soap solution. Remove any soap residue by wiping with a cloth moistened with clear water. Do not use abrasive compounds on any part of the chassis.
© National Instruments Corporation 2-9 PXI-1006 User Manual
Chapter 2 PXI-1006 Features
Caution
Avoid getting moisture inside the chassis during exterior cleaning. Use just enough moisture to dampen the cloth. Do not wash the front- or rear-panel connectors or switches. Cover these components while cleaning the chassis. Do not use chemical cleaning agents; they may damage the chassis. Avoid chemicals that contain benzene, toluene, xylene, acetone, or similar solvents.

Cleaning the Fan Filters

You can easily remove the chassis cooling filters from the rear of the chassis by removing the plastic housing attached to each fan.
Clean the fan filters by washing them in a mild soap solution and then vacuuming or blowing air through them. Rinse the filters with water and allow them to dry before reinstalling them on the chassis.
PXI-1006 User Manual 2-10 ni.com
Troubleshooting
This chapter describes basic troubleshooting procedures you can perform on the PXI-1006 chassis.

Troubleshooting the PXI-1006

Refer to Table 3-1 to troubleshoot the PXI-1006 chassis. The table lists possible causes for power failure and recommends ways to correct the problem.
3
Table 3-1.
Possible Cause What to Do
PXI-1006 chassis is not connected to power source.
Power switch is not switched on. Set the power switch to the On position.
Remote inhibiting input on the rear panel of the chassis is active.
Circuit breaker is tripped. Reset the circuit breaker. Refer to the Resetting the AC
Power supply has failed. Contact National Instruments.
Troubleshooting
Make sure that the PXI-1006 is connected to a live electrical outlet. Try operating another piece of equipment from this outlet.
Deactivate your system’s remote inhibiting control.
Mains Circuit Breaker section in this chapter.

Resetting the AC Mains Circuit Breaker

If your PXI-1006 is connected to an AC source and encounters an over-current condition, the circuit breaker located on the rear panel will trip to prevent damage to the chassis. Complete the following steps to reset the circuit breaker:
1. Turn the power switch to the Standby position.
2. Disconnect the AC line cord.
© National Instruments Corporation 3-1 PXI-1006 User Manual
Chapter 3 Troubleshooting
3. Depress the circuit breaker to reset it.
4. Reconnect the AC line cord.
5. Turn the power switch to the On position.
If the circuit breaker trips again, complete the following steps:
1. Turn the front panel power switch to the Standby position.
2. Disconnect the chassis from the AC mains power source.
3. Remove all modules from the chassis.
4. Complete the test procedure described in the Using the Power
Monitoring Feature section in Chapter 2, PXI-1006 Features.
5. If any voltages are outside the acceptable limits, contact National Instruments.
6. If all voltages are within the acceptable limits, verify that your PXI-1006 can meet the power requirements of your CompactPCI or PXI modules. Overloading the chassis can cause the breaker to trip. Refer to Appendix A, Specifications.
7. The over-current condition that caused the circuit breaker to trip may be due to a faulty CompactPCI or PXI module. Refer to the documentation supplied with the modules for troubleshooting your modules.
PXI-1006 User Manual 3-2 ni.com
Specifications
This appendix contains specifications for the PXI-1006 chassis.
Electrical
AC Input
Input voltage range.................................90–264 VAC
Input frequency range ............................ 47 to 63 Hz
Over-current protection.......................... 10 A circuit breaker
Maximum steady state
operating current .................................... 10 A
Line regulation ....................................... ± 0.1% over operating line range
Efficiency ............................................... 70-80% typical
Power disconnect ................................... The (standby) power switch
A
causes the power module to supply DC power to the CompactPCI/PXI backplane. The rear-panel D-sub connector facilitates remote inhibiting operation. The (standby) switch must be in the On position prior to use of remote inhibit. The power cord provides main power disconnect.
© National Instruments Corporation A-1 PXI-1006 User Manual
Appendix A Specifications
DC Output
Maximum usable power .........................600 W
DC current capacity (I
Vo l t ag e IMP(Steady-State Current)
+3.3 V 60 A
+5 V 60 A
+12 V 9A
–12 V 1.8 A
Load regulation
Vo l t ag e Regulation
+3.3 V 0.4% or 20 mV max
+12 V 0.4% or 20 mV max
+5 V 0.4% or 20 mV max
–12 V 0.4% or 20 mV max
Maximum ripple and noise .....................1% ripple, 1% noise
Over-current protection...........................105–140% of rated output
MP
)
20 MHz bandwidth
current; automatic recovery
Over-voltage protection ..........................3.3 V, 5 V clamp at 122–134%
of output voltage +12 V and –12 V clamp at 110–120% of output voltage
Power supply/fan unit MTTR.................Replacement in under 5 minutes
Cooling
Module cooling system...........................Forced air circulation
(positive pressurization) via three 140 cfm fans with HI/AUTO speed selector
PXI-1006 User Manual A-2 ni.com
Slot airflow direction.............................. P1 to P2, bottom of module to top
Module/power supply cooling intake..... Three rear fans
Module cooling exhaust ......................... Along both sides and top of
Power supply cooling system................. Forced air circulation via
Power supply cooling exhaust................ Along upper left side
Power supply/fan unit ............................ Replacement in under 5 minutes
Electromagnetic Compatibility
EMC/EMI............................................... CE, C-Tick and FCC Part 15
Electrical emissions................................ EN 55011 Class A @ 10 m.
Appendix A Specifications
of module
chassis
integrated fan
(Class A) compliant
FCC Part 15A above 1 GHz
Note
This device should be operated only with shielded cabling for full EMC and EMI compliance. See the Declaration of Conformity for this product for any additional regulatory compliance information.
Safety
Environmental
Electrical immunity................................ Evaluated to EN 61326:1998,
Tabl e 1
Designed in accordance with IEC 61010-1, EN 61010-1, UL 3111-1, and CAN/CSA C22.2 No. 1010.1 for electrical measuring and test equipment Installation Category II Pollution Degree 2
Operating temperature............................ 0 to 50 °C
Storage temperature ............................... –20 to 70 °C
Operating relative humidity ................... Maximum 80% for temperatures
up to 31 °C, decreasing linearly to 50% at 40 °C
© National Instruments Corporation A-3 PXI-1006 User Manual
Appendix A Specifications
Backplane
Functional shock (operating) ..................MIL-T-28800E CLASS 3,
Half-Sine Shock Pulse, 11 ms duration, 30 g peak
Operating location ..................................Indoor use only
Random vibration
Operational
Non-operational1.............................10 to 500 Hz, 2.46 g
1
.....................................5 to 500 Hz, 0.31 g
RMS
RMS
EMC emissions.......................................FCC Class A compliant and
EN 55011 Group 1 Class A Compliant
EMC immunity .......................................Refer to DOC supplied with
chassis for compliance to relevant directives.
Altitude ...................................................2 km (1.24 mi) max
Size .........................................................3U-sized; one system slot
(with three system expansion slots)and17peripheralslots. Compliant with IEEE 1101.10 mechanical packaging. PXI Specification Revision 2.0 compliant. Accepts both PXI and CompactPCI (PICMG 2.0 R3.0) 3U modules.
Backplane bare-board material ...............UL 94 V-0 recognized
Backplane connectors .............................Conform to IEC 917 and
IEC 1076-4-101, and are UL 94 V-0 rated
1
Random vibration profiles were developed in accordance with MIL-T-28800E CLASS 3 and MIL-STD-810E Method 514 Test levels exceed those recommended in MIL-STD-810E for Category 1 (Basic Transportation), Figures 514.4-1 through
514.4-3.
PXI-1006 User Manual A-4 ni.com
Mechanical
Appendix A Specifications
Weight .................................................... 13.2 kg (29.2 lb.)
Maximum module weight ...................... 1.8 kg (4 lb.)
Materials................................................. Sheet Aluminum (5052-H32 and
3003-H14) and Cold Rolled Steel
Finish
Unpainted aluminum....................... Conductive clear conversion
coating
Cold rolled steel .............................. Clear chromate zinc plating
Paint ................................................ Polyurethane enamel
© National Instruments Corporation A-5 PXI-1006 User Manual
Appendix A Specifications
0.78 in.
(19.8 mm)
Figure A-1 shows the PXI-1006 dimensions. The holes shown are for the installation of the optional rack-mount kit. You can install this kit on the front or rear of the chassis, depending on which end of the chassis you want to face toward the front of the instrument cabinet. Note that the front and rear chassis mounting holes (size M4) are symmetrical.
6.97 in.
(177.0 mm)
17.54 in. (445.5 mm)
16.89 in. (429.0 mm)
1.82 in. (46.3 mm)
1.50 in.
(38.0 mm)
3.54 in.
(90.0 mm)
1.88 in. (47.8 mm)
1.84 in.
(46.8 mm)
0.57 in. (14.5 mm)

Figure A-1. PXI-1006 Dimensions

PXI-1006 User Manual A-6 ni.com
Pinouts
Note
B
This appendix describes the P1 and P2 connector pinouts for the PXI-1006 backplane.
Table B-1 shows the P1 (J1) connector pinout for the System Controller slot.
Table B-2 shows the P2 (J2) connector pinout for the System Controller slot.
Table B-3 shows the P1 (J1) connector pinout for the Star Trigger slot.
Table B-4 shows the P2 (J2) connector pinout for the Star Trigger slot.
Table B-5 shows the P1 (J1) connector pinout for the peripheral slots.
Table B-6 shows the P2 (J2) connector pinout for the peripheral slots.
PXI signals are shown in bold.
© National Instruments Corporation B-1 PXI-1006 User Manual
Appendix B Pinouts

Table B-1. P1 (J1) Connector Pinout for the System Controller Slot

Pin Z A B C D E F
25 GND 5V REQ64# ENUM# 3.3V 5V GND
24 GND AD[1] 5V V(I/O) AD[0] ACK64# GND
23 GND 3.3V AD[4] AD[3] 5V AD[2] GND
22 GND AD[7] GND 3.3V AD[6] AD[5] GND
21 GND 3.3V AD[9] AD[8] M66EN C/BE[0]# GND
20 GND AD[12] GND V(I/O) AD[11] AD[10] GND
19 GND 3.3V AD[15] AD[14] GND AD[13] GND
18 GND SERR# GND 3.3V PA R C/BE[1]# GND
17 GND 3.3V IPMB_SCL IPMB_SDA GND PERR# GND
16 GND DEVSEL# GND V(I/O) STOP# LOCK# GND
15 GND 3.3V FRAME# IRDY# GND TRDY# GND
12–14 Key Area
11 GND AD[18] AD[17] AD[16] GND C/BE[2]# GND
10 GND AD[21] GND 3.3V AD[20] AD[19] GND
9 GND C/BE[3]# GND AD[23] GND AD[22] GND
8 GND AD[26] GND V(I/O) AD[25] AD[24] GND
7 GND AD[30] AD[29] AD[28] GND AD[27] GND
6 GND REQ0# GND 3.3V CLK0 AD[31] GND
5 GND BRSVP1A5 BRSVP1B5 RST# GND GNT0# GND
4 GND IPMB_PWR HEALTHY# V(I/O) INTP INTS GND
3 GND INTA# INTB# INTC# 5V INTD# GND
2 GND TCK 5V TMS TDO TDI GND
1 GND 5V –12V TRST# +12V 5V GND
PXI-1006 User Manual B-2 ni.com
Appendix B Pinouts
Table B-2.
Pin Z A B C D E F
22 GND GA4 GA3 GA2 GA1 GA0 GND
21 GND CLK6 GND RSV RSV RSV GND
20 GND CLK5 GND RSV GND RSV GND
19 GND GND GND SMB_SDA SMB_SCL SMB_ALERT# GND
18 GND PXI_TRIG3 PXI_TRIG4 PXI_TRIG5 GND PXI_TRIG6 GND
17 GND
16 GND PXI_TRIG1 PXI_TRIG0 DEG# GND PXI_TRIG7 GND
15 GND
14 GND AD[35] AD[34] AD[33] GND AD[32] GND
13 GND AD[38] GND V(I/O) AD[37] AD[36] GND
12 GND AD[42] AD[41] AD[40] GND AD[39] GND
11 GND AD[45] GND V(I/O) AD[44] AD[43] GND
10 GND AD[49] AD[48] AD[47] GND AD[46] GND
9 GND AD[52] GND V(I/O) AD[51] AD[50] GND
8 GND AD[56] AD[55] AD[54] GND AD[53] GND
7 GND AD[59] GND V(I/O) AD[58] AD[57] GND
6 GND AD[63] AD[62] AD[61] GND AD[60] GND
PXI_TRIG2
PXI_BRSVA15
P2 (J2) Connector Pinout for the System Controller Slot
GND PRST# REQ6# GNT6# GND
GND FA L# REQ5# GNT5# GND
5 GND C/BE[5]# GND V(I/O) C/BE[4]# PA R 64 GND
4 GND V(I/O) PXI_BRSVB4 C/BE[7]# GND C/BE[6]# GND
3 GND CLK4 GND GNT3# REQ4# GNT4# GND
2 GND CLK2 CLK3 SYSEN# GNT2# REQ3# GND
1 GND CLK1 GND REQ1# GNT1# REQ2# GND
© National Instruments Corporation B-3 PXI-1006 User Manual
Appendix B Pinouts

Table B-3. P1 (J1) Connector Pinout for the Star Trigger Slot

Pin Z A B C D E F
25 GND 5V REQ64# ENUM# 3.3V 5V GND
24 GND AD[1] 5V V(I/O) AD[0] ACK 64# GND
23 GND 3.3V AD[4] AD[3] 5V AD[2] GND
22 GND AD[7] GND 3.3V AD[6] AD[5] GND
21 GND 3.3V AD[9] AD[8] M66EN C/BE[0]# GND
20 GND AD[12] GND V(I/O) AD[11] AD[10] GND
19 GND 3.3V AD[15] AD[14] GND AD[13] GND
18 GND SERR# GND 3.3V PA R C/BE[1]# GND
17 GND 3.3V IPMB_SCL IPMB_SDA GND PERR# GND
16 GND DEVSEL# GND V(I/O) STOP# LOCK# GND
15 GND 3.3V FRAME# IRDY# BD_SEL# TRDY# GND
12–14 Key Area
11 GND AD[18] AD[17] AD[16] GND C/BE[2]# GND
10 GND AD[21] GND 3.3V AD[20] AD[19] GND
9 GND C/BE[3]# IDSEL AD[23] GND AD[22] GND
8 GND AD[26] GND V(I/O) AD[25] AD[24] GND
7 GND AD[30] AD[29] AD[28] GND AD[27] GND
6 GND REQ# GND 3.3V CLK AD[31] GND
5 GND BRSVP1A5 BRSVP1B5 RST# GND GNT# GND
4 GND IPMB_PWR HEALTHY# V(I/O) INTP INTS GND
3 GND INTA# INTB# INTC# 5V INTD# GND
2 GND TCK 5V TMS TDO TDI GND
1 GND 5V –12V TRST# +12V 5V GND
PXI-1006 User Manual B-4 ni.com
Appendix B Pinouts
Table B-4.
Pin Z A B C D E F
22 GND GA4 GA3 GA2 GA1 GA0 GND
21 GND PXI_LBR0 GND PXI_LBR1 PXI_LBR2 PXI_LBR3 GND
20 GND PXI_LBR4 PXI_LBR5 PXI_STAR0 GND PXI_STAR1 GND
19 GND PXI_STAR2 GND PXI_STAR3 PXI_STAR4 PXI_STAR5 GND
18 GND PXI_TRIG3 PXI_TRIG4 PXI_TRIG5 GND PXI_TRIG6 GND
17 GND
16 GND PXI_TRIG1 PXI_TRIG0 RSV GND PXI_TRIG7 GND
15 GND
14 GND AD[35] AD[34] AD[33] GND AD[32] GND
13 GND AD[38] GND V(I/O) AD[37] AD[36] GND
12 GND AD[42] AD[41] AD[40] GND AD[39] GND
11 GND AD[45] GND V(I/O) AD[44] AD[43] GND
10 GND AD[49] AD[48] AD[47] GND AD[46] GND
9 GND AD[52] GND V(I/O) AD[51] AD[50] GND
8 GND AD[56] AD[55] AD[54] GND AD[53] GND
7 GND AD[59] GND V(I/O) AD[58] AD[57] GND
6 GND AD[63] AD[62] AD[61] GND AD[60] GND
PXI_TRIG2
PXI_BRSVA15
P2 (J2) Connector Pinout for the Star Trigger Slot
GND RSV
GND RSV
PXI_CLK10_IN PXI_CLK10
PXI_STAR6 PXI_LBR6
GND
GND
5 GND C/BE[5]# GND V(I/O) C/BE[4]# PA R 64 GND
4 GND V(I/O) PXI_BRSVB4 C/BE[7]# GND C/BE[6]# GND
3 GND
2 GND
1 GND PXI_STAR9 GND PXI_STAR10 PXI_STAR11 PXI_STAR12 GND
© National Instruments Corporation B-5 PXI-1006 User Manual
PXI_LBR7
PXI_LBR11 PXI_LBR12
GND
PXI_LBR8 PXI_LBR9 PXI_LBR10
UNC
PXI_STAR7 PXI_STAR8
GND
GND
Appendix B Pinouts

Table B-5. P1 (J1) Connector Pinout for the Peripheral Slot

Pin Z A B C D E F
25 GND 5V REQ64# ENUM# 3.3V 5V GND
24 GND AD[1] 5V V(I/O) AD[0] ACK 64# GND
23 GND 3.3V AD[4] AD[3] 5V AD[2] GND
22 GND AD[7] GND 3.3V AD[6] AD[5] GND
21 GND 3.3V AD[9] AD[8] M66EN C/BE[0]# GND
20 GND AD[12] GND V(I/O) AD[11] AD[10] GND
19 GND 3.3V AD[15] AD[14] GND AD[13] GND
18 GND SERR# GND 3.3V PA R C/BE[1]# GND
17 GND 3.3V IPMB_SCL IPMB_SDA GND PERR# GND
16 GND DEVSEL# GND V(I/O) STOP# LOCK# GND
15 GND 3.3V FRAME# IRDY# BD_SEL# TRDY# GND
12–14 Key Area
11 GND AD[18] AD[17] AD[16] GND C/BE[2]# GND
10 GND AD[21] GND 3.3V AD[20] AD[19] GND
9 GND C/BE[3]# IDSEL AD[23] GND AD[22] GND
8 GND AD[26] GND V(I/O) AD[25] AD[24] GND
7 GND AD[30] AD[29] AD[28] GND AD[27] GND
6 GND REQ# GND 3.3V CLK AD[31] GND
5 GND BRSVP1A5 BRSVP1B5 RST# GND GNT# GND
4 GND IPMB_PWR HEALTHY# V(I/O) INTP INTS GND
3 GND INTA# INTB# INTC# 5V INTD# GND
2 GND TCK 5V TMS TDO TDI GND
1 GND 5V –12V TRST# +12V 5V GND
PXI-1006 User Manual B-6 ni.com
Appendix B Pinouts
Table B-6.
Pin Z A B C D E F
22 GND GA4 GA3 GA2 GA1 GA0 GND
21 GND PXI_LBR0 GND PXI_LBR1 PXI_LBR2 PXI_LBR3 GND
20 GND PXI_LBR4 PXI_LBR5 PXI_LBL0 GND PXI_LBL1 GND
19 GND PXI_LBL2 GND PXI_LBL3 PXI_LBL4 PXI_LBL5 GND
18 GND PXI_TRIG3 PXI_TRIG4 PXI_TRIG5 GND PXI_TRIG6 GND
17 GND
16 GND PXI_TRIG1 PXI_TRIG0 RSV GND PXI_TRIG7 GND
15 GND
14 GND AD[35] AD[34] AD[33] GND AD[32] GND
13 GND AD[38] GND V(I/O) AD[37] AD[36] GND
12 GND AD[42] AD[41] AD[40] GND AD[39] GND
11 GND AD[45] GND V(I/O) AD[44] AD[43] GND
10 GND AD[49] AD[48] AD[47] GND AD[46] GND
9 GND AD[52] GND V(I/O) AD[51] AD[50] GND
8 GND AD[56] AD[55] AD[54] GND AD[53] GND
7 GND AD[59] GND V(I/O) AD[58] AD[57] GND
6 GND AD[63] AD[62] AD[61] GND AD[60] GND
PXI_TRIG2
PXI_BRSVA15
P2 (J2) Connector Pinout for the Peripheral Slot
GND RSV
GND RSV
PXI_STAR PXI_CLK10
PXI_LBL6 PXI_LBR6
GND
GND
5 GND C/BE[5]# GND V(I/O) C/BE[4]# PA R 64 GND
4 GND V(I/O) PXI_BRSVB4 C/BE[7]# GND C/BE[6]# GND
3 GND
2 GND
1 GND PXI_LBL9 GND PXI_LBL10 PXI_LBL11 PXI_LBL12 GND
© National Instruments Corporation B-7 PXI-1006 User Manual
PXI_LBR7
PXI_LBR11 PXI_LBR12
GND
PXI_LBR8 PXI_LBR9 PXI_LBR10
UNC
PXI_LBL7 PXI_LBL8
GND
GND
Technical Support Resources
Web Support
National Instruments Web support is your first stop for help in solving installation, configuration, and application problems and questions. Online problem-solving and diagnostic resources include frequently asked questions, knowledge bases, product-specific troubleshooting wizards, manuals, drivers, software updates, and more. Web support is available through the Technical Support section of
NI Developer Zone
C
The NI Developer Zone at building measurement and automation systems. At the NI Developer Zone, you can easily access the latest example programs, system configurators, tutorials, technical news, as well as a community of developers ready to share their own techniques.
Customer Education
National Instruments provides a number of alternatives to satisfy your training needs, from self-paced tutorials, videos, and interactive CDs to instructor-led hands-on courses at locations around the world. Visit the Customer Education section of syllabi, training centers, and class registration.
System Integration
If you have time constraints, limited in-house technical resources, or other dilemmas, you may prefer to employ consulting or system integration services. You can rely on the expertise available through our worldwide network of Alliance Program members. To find out more about our Alliance system integration solutions, visit the System Integration section of
ni.com
ni.com/zone
ni.com
is the essential resource for
for online course schedules,
© National Instruments Corporation C-1 PXI-1006 User Manual
Appendix C Technical Support Resources
Worldwide Support
National Instruments has offices located around the world to help address your support needs. You can access our branch office Web sites from the Worldwide Offices section of up-to-date contact information, support phone numbers, e-mail addresses, and current events.
If you have searched the technical support resources on our Web site and still cannot find the answers you need, contact your local office or National Instruments corporate. Phone numbers for our worldwide offices are listed at the front of this manual.
. Branch office Web sites provide
PXI-1006 User Manual C-2 ni.com

Glossary

Prefix Meanings Value
n- nano- 10
µ- micro- 10
m- milli- 10
c- centi- 10
k- kilo- 10
M- mega- 10

Symbols

° Degrees
Equal or greater than
Equal or less than
9
6
3
2
3
6
% Percent
A
A Amperes
AC Alternating current
Ah Ampere hours
ANSI American National Standards Institute
AWG American Wire Gauge
© National Instruments Corporation G-1 PXI-1006 User Manual
Glossary
B
backplane An assembly, typically a printed circuit board, with connectors and signal
paths that bus the connector pins
C
CCelsius
cfm Cubic feet per minute
CFR Cooperative Fuel Research
CSA Canadian Standards Association
D
daisy-chain A method of propagating signals along a bus, in which the devices are
prioritized on the basis of their position on the bus
DC Direct current
E
ECL Emitter-coupled logic
EIA Electronic Industries Association
EMC Electromagnetic Compatibility
F
FCC Federal Communications Commission
G
g 1) grams 2) A measure of acceleration equal to 9.8 m/s
GPIB General Purpose Interface Bus (IEEE 488)
2
PXI-1006 User Manual G-2 ni.com
Glossary
g
RMS
A measure of random vibration. The root mean square of acceleration levels in a random vibration test profile.
H
Hz Hertz; cycles per second
I
IEC International Electrotechnical Commission; an organization that sets
international electrical and electronics standards
IEEE Institute of Electrical and Electronics Engineers
I
MP
in. Inches
Mainframe peak current
L
lb Pounds
M
m Meters
MTBF Mean time between failure
MTTR Mean time to repair
N
NEMA National Electrical Manufacturers Association
P
PXI PCI eXtensions for Instrumentation
© National Instruments Corporation G-3 PXI-1006 User Manual
Glossary
R
RH Relative humidity
RMS Root mean square. A method used to measure electrical output in volts and
watts
S
s Seconds
ST Star Trigger
Star Trigger slot This slot is located at slot 2 and has a dedicated trigger line between each
peripheral slot. Use this slot for a module with ST functionality that can provide individual triggers to all other peripherals.
System controller A module configured for installation in Slot 0 of a VXIbus mainframe. This
device is unique in the VXIbus system in that it performs the VMEbus system controller functions, including clock sourcing and arbitration for data transfers across the backplane. Installing such a device into any other slot can damage the device, the VXIbus backplane, or both.
U
UL Underwriter’s Laboratories
V
VVolts
VAC Volts alternating current
V
PP
Peak to peak voltage
W
WWatts
PXI-1006 User Manual G-4 ni.com

Index

B
backplane, 2-2
interoperability with CompactPCI, 2-2 local bus, 2-4 overview, 2-2 peripheral slots, 2-4 specifications, A-4 Star Trigger (ST) slot, 2-4 system reference clock, 2-6 trigger bus, 2-6
C
chassis initialization file, 2-9 chassis segment dividers, 2-6 CompactPCI
interoperability with PXI-1006
backplane, 2-2 connector pinouts. See pinouts. cooling
filler panel installation, 2-9 cooling specifications, A-2 customer education, C-1
E
electrical specifications, A-1 electromagnetic compatibility specifications
(table), A-3
F
filler panel installation, 2-9
G
ground, connecting, xi
I
IEC 320 inlet, 2-3 installation, configuration, and operation
chassis initialization file, 2-9 connecting safety ground, xi filler panel installation, 2-9 rack mounting, 2-6 remote power monitoring and inhibiting
interface, 2-7
interoperability with CompactPCI, 2-2
D
DB-9 connector
pinout (table), 2-7
power supply voltages (table), 2-8 dimensions (figure), A-6 documentation
conventions used in manual, ix
related documentation, ix
© National Instruments Corporation I-1 PXI-1006 User Manual
K
key features, 2-1
L
local bus
routing (figure), 2-4, 2-5
Index
M
maintenance of PXI-1006, 3-1
cleaning
exterior cleaning, 2-9 fan filters, 2-10 interior cleaning, 2-9
resetting the AC mains circuit
breaker, 3-1
troubleshooting
causes, what to do (table), 3-1
mechanical specifications, A-5
N
National Instruments Web support, C-1 NI Developer Zone, C-1
P
P1 (J1) connector
peripheral slot (table), B-6 Star Trigger slot (table), B-4 system controller slot (table), B-2
P2 (J2) connector
peripheral slot (table), B-7 Star Trigger slot (table), B-5 system controller slot (table), B-3
peripheral slots
overview, 2-4 P1 (J1) connector pinouts (table), B-6 P2 (J2) connector pinouts (table), B-7
pinouts, B-1
DB-9 connector (table), 2-7 P1 (J1) connector
peripheral slot (table), B-6 Star Trigger slot (table), B-4 system controller slot (table), B-2
P2 (J2) connector
peripheral slot (table), B-7 Star Trigger slot (table), B-5 system controller slot (table), B-3
power monitoring connector. See DB-9
connector. power monitoring feature, 2-8 power problems, troubleshooting, 3-1 power supply
remote power monitoring and inhibiting
interface, 2-7
voltages at power monitoring connector
(DB-9) (table), 2-7 PXI_CLK10, 2-6 PXI_CLK10_IN pin, 2-6 PXI-1006
frontview (figure), 2-3 installation
See installation, configuration, and
operation. key features, 2-1 maintenance
See maintenance of PXI-1006. rack mounting, 2-6 rearview of AC-only (figure), 2-3 safety ground, connecting, xi
PXI-1006 backplane, 2-2
interoperability with CompactPCI, 2-2 local bus, 2-4 overview, 2-2 peripheral slots, 2-4 specifications, A-4 Star Trigger (ST) slot, 2-4 system reference clock, 2-6 trigger bus, 2-6
PXI-1006 User Manual I-2 ni.com
Index
R
rack mounting, 2-6 remote power monitoring and inhibiting
interface, 2-7
S
safety ground, connecting, xi safety specifications (table), A-3 safety, warning and caution notices, xi specifications, A-1
backplane, A-4 cooling, A-2 dimensions (figure), A-6 electrical, A-1 electromagnetic compatibility, A-3 environmental, A-3 mechanical, A-5 safety, A-3
StarTrigger(ST)slot
description, 2-4 P1 (J1) connector pinouts (table), B-4 P2 (J2) connector pinouts (table), B-5
system controller slot
description, 2-4 P1 (J1) connector pinouts (table), B-2 P2 (J2) connector pinouts (table), B-3
system integration, by National Instruments,
C-1
system reference clock, 2-6
T
technical support resources, C-1 trigger bus, 2-6 troubleshooting the PXI-1006 (table), 3-1
V
voltages at power monitoring connector
(DB-9) (table), 2-8
W
Web support from National Instruments, C-1 worldwide technical support, C-2
© National Instruments Corporation I-3 PXI-1006 User Manual
Loading...