The PXI-1006 is warranted against defects in materials and workmanship for a period of one year from the date of shipment, as evidenced by
receipts or other documentation. National Instruments will, at its option, repair or replace equipment that proves to be defective during the
warranty period. This warranty includes parts and labor.
The media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects
in materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other documentation. National
Instruments will, at its option, repair or replace software media that do not execute programming instructions if National Instruments receives
notice of such defects during the warranty period. National Instruments does not warrant that the operation of the software shall be
uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before
any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are
covered by warranty.
National Instruments believes that the information in this document is accurate. The document has been carefully reviewed for technical
accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent
editions of this document without prior notice to holders of this edition. The reader should consult National Instruments if errors are suspected.
In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.
XCEPT AS SPECIFIED HEREIN,NATIONAL INSTRUMENTS MAKES NO W ARRANTIES, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF
E
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE
NATIONAL INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID B Y THE CUSTOMER.NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR
DAMAGES RESULTING FROM LOSS OF DATA
. This limitation of the liability of National Instruments will apply regardless ofthe form of action,whether in contract ortort, including
THEREOF
negligence. Any action against National Instruments must be brought within one year after the cause of action accrues. National Instruments
shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not cover
damages, defects, malfunctions, or service failures caused by owner’s failure to follow the National Instruments installation, operation, or
maintenance instructions; owner’s modification of the product; owner’s abuse, misuse, or negligent acts; and power failure or surges, fire,
flood, accident, actions of third parties, or other events outside reasonable control.
, PROFITS, USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY
Copyright
Under the copyright laws, this publication may not be reproduced ortransmitted in any form, electronic or mechanical, including photocopying,
recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National
Instruments Corporation.
Trademarks
National Instruments™,ni.com™,andPXI™are trademarks of National Instruments Corporation.
Product and company names mentioned herein are trademarks or trade names of their respective companies.
.CUSTOMER’S RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART OF
WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS
(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING FOR A LEVEL OF
RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL IMPLANTS OR AS CRITICAL COMPONENTS IN
ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE TO PERFORM CAN REASONABLY BE EXPECTED TO CAUSE SIGNIFICANT
INJURY TO A HUMAN.
(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE SOFTWARE PRODUCTS CAN BE
IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT LIMITED TO FLUCTUATIONS IN ELECTRICAL POWER SUPPLY,
COMPUTER HARDWARE MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE FITNESS, FITNESS OF COMPILERS
AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION, INSTALLATION ERRORS, SOFTWARE AND
HARDWARE COMPATIBILITY PROBLEMS, MALFUNCTIONS OR FAILURES OF ELECTRONIC MONITORING OR CONTROL
DEVICES, TRANSIENT FAILURES OF ELECTRONIC SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR
MISUSES, OR ERRORS ON THE PART OF THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS THESE ARE
HEREAFTER COLLECTIVELY TERMED “SYSTEM FAILURES”). ANY APPLICATION WHERE A SYSTEM FAILURE WOULD
CREATE A RISK OF HARM TO PROPERTY OR PERSONS (INCLUDING THE RISK OF BODILY INJURY AND DEATH) SHOULD
NOT BE RELIANT SOLELY UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM FAILURE. TO AVOID
DAMAGE, INJURY, OR DEATH, THE USER OR APPLICATION DESIGNER MUST TAKE REASONABLY PRUDENT STEPS TO
PROTECT AGAINST SYSTEM FAILURES, INCLUDING BUT NOT LIMITED TO BACK-UP OR SHUT DOWN MECHANISMS.
BECAUSE EACH END-USER SYSTEM IS CUSTOMIZED AND DIFFERS FROM NATIONAL INSTRUMENTS' TESTING
PLATFORMS AND BECAUSE A USER OR APPLICATION DESIGNER MAY USE NATIONAL INSTRUMENTS PRODUCTS IN
COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT EVALUATED OR CONTEMPLATED BY NATIONAL
INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS ULTIMATELY RESPONSIBLE FOR VERIFYING AND VALIDATING
THE SUITABILITY OF NATIONAL INSTRUMENTS PRODUCTS WHENEVER NATIONAL INSTRUMENTS PRODUCTS ARE
INCORPORATED IN A SYSTEM OR APPLICATION, INCLUDING, WITHOUT LIMITATION, THE APPROPRIATE DESIGN,
PROCESS AND SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.
Compliance
FCC/Canada Radio Frequency Interference Compliance*
Determining FCC Class
The Federal Communications Commission (FCC) has rules to protect wireless communications from interference. The FCC
places digital electronics into two classes. These classes are known as Class A (for use in industrial-commercial locations only)
or Class B (for use in residential or commercial locations). Depending on where it is operated, this product could be subject to
restrictions in the FCC rules. (In Canada, the Department of Communications (DOC), of Industry Canada, regulates wireless
interference in much the same way.)
Digital electronics emit weak signals during normal operation that can affect radio, television, or other wireless products. By
examining the product you purchased, you can determine the FCC Class and therefore which of the two FCC/DOC Warnings
apply in the following sections. (Some products may not be labeled at all for FCC; if so, the reader should then assume these are
Class A devices.)
FCC Class A products only display a simple warning statement of one paragraph in length regarding interference and undesired
operation. Most of our products are FCC Class A. The FCC rules have restrictions regarding the locations where FCC Class A
products can be operated.
FCC Class B products display either a FCC ID code, starting with the letters EXN,
or the FCC Class B compliance mark that appears as shown here on the right.
Consult the FCC web site
http://www.fcc.gov
FCC/DOC Warnings
This equipment generates and uses radio frequency energyand, if not installed and used in strict accordance with the instructions
in this manual and the CE Mark Declaration of Conformity**, may cause interference to radio and television reception.
Classification requirements are the same for the Federal Communications Commission (FCC) and the Canadian Department
of Communications (DOC).
Changes or modifications not expressly approved by National Instruments could void the user’s authority to operate the
equipment under the FCC Rules.
for more information.
Class A
Federal Communications Commission
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC
Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated
in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and
used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct
the interference at his own expense.
Canadian Department of Communications
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.
Class B
Federal Communications Commission
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the
FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can
be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of
the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
Canadian Department of Communications
This Class B digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe B respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.
European Union - Compliance to EEC Directives
Readers in the EU/EEC/EEA must refer to the Manufacturer's Declaration of Conformity (DoC) for information** pertaining
to the CE Mark compliance scheme.The Manufacturer includes a DoC for most every hardware product except for those bought
for OEMs, if also available from an original manufacturer that also markets in the EU, or where compliance is not required as
for electrically benign apparatus or cables.
* Certain exemptions may apply in the USA, see FCC Rules §15.103 Exempted devices,and§15.105(c). Also available in
sections of CFR 47.
** The CE Mark Declaration of Conformity will contain important supplementary information and instructions for the user or
installer.
Contents
About This Manual
Conventions Used in This Manual.................................................................................ix
Related Documentation..................................................................................................ix
For Your Safety
Connecting Safety Ground............................................................................................. xi
Chapter 1
Installing and Verifying Operation
Unpacking the PXI-1006 and Installing a Controller ....................................................1-1
Chapter 2
PXI-1006 Features
Key Features ..................................................................................................................2-1
Table B-1.P1 (J1) Connector Pinout for the System Controller Slot..................... B-2
Table B-2.P2 (J2) Connector Pinout for the System Controller Slot..................... B-3
Table B-3.P1 (J1) Connector Pinout for the Star Trigger Slot...............................B-4
Table B-4.P2 (J2) Connector Pinout for the Star Trigger Slot...............................B-5
Table B-5.P1 (J1) Connector Pinout for the Peripheral Slot.................................. B-6
Table B-6.P2 (J2) Connector Pinout for the Peripheral Slot.................................. B-7
PXI-1006 User Manualviiini.com
About This Manual
The PXI-1006 User Manual describes the features of the PXI-1006 chassis
and contains information about configuring the chassis, installing the
modules, and operating and using the PXI-1006.
Conventions Used in This Manual
The following conventions appear in this manual:
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to
avoid injury, data loss, or a system crash.
This icon denotes a warning, which advises you of precautions to take to
avoid being electrically shocked.
boldBold text denotes items that you must select or click on in the software,
such as menu items and dialog box options. Bold text also denotes
parameter names.
italicItalic text denotes variables, emphasis, a cross reference, or an introduction
to a key concept. This font also denotes text that is a placeholder for a word
or value that you must supply.
monospace
Text in this font denotes text or characters that you should enter from the
keyboard, sections of code, programming examples, and syntax examples.
This font is also used for the proper names of disk drives, paths, directories,
programs, subprograms, subroutines, device names, functions, operations,
variables, filenames and extensions, and code excerpts.
Related Documentation
The following documents contain information that you may find helpful as
you read this manual:
•Compact PCI Specification PICMG 2.0 R3.0
•PXI Specification Revision 2.0
•IEEE 1101.1-1991, IEEE Standard for Mechanical Core
Specifications for Microcomputers Using IEC 603-2 Connectors
•IEEE 1101.10 and P1101.11, IEEE Standard for Additional
Mechanical Specifications for Microcomputers Using IEEE 1101.1
Equipment Practice
PXI-1006 User Manualxni.com
For Your Safety
Caution
carefully read the warning and caution notices.
Before undertaking any troubleshooting, maintenance, or exploratory procedure,
This equipment contains voltage hazardous to human life and safety, and
can inflict personal injury.
•Chassis Grounding—The PXI-1006 requires a connection from the
premise wire safety ground to the PXI-1006 chassis ground. The earth
safety ground must be connected during use of this equipment to
minimize shock hazards. See the Connecting Safety Ground section
below.
•Live Circuits—Operating personnel and service personnel must not
remove protective covers when operating or servicing the PXI-1006.
•Explosive Atmosphere—Do not operate the chassis in conditions
where flammable gases are present. Under such conditions this
equipment is unsafe and may ignite the gases or gas fumes.
•Modification—Do not modify any part of the chassis from its original
condition. Unsuitable modifications may result in safety hazards.
Connecting Safety Ground
Warning
that connects the ground line to the chassis ground. To minimize shock hazard, make sure
your electrical power outlet has an appropriate earth safety ground that is connected
whenever you power up the chassis.
The PXI-1006 chassis is designed with a three-position NEMA 15-5 style plug
If your power outlet does not have an appropriate ground connection, you
must connect the premise wire safety ground to the chassis grounding
screw located on the rear panel. To connect the safety ground, complete the
following steps:
1.Connect a 16 AWG (1.3 mm) wire to the chassis grounding screw on
the rear of the chassis using a toothed grounding lug. The wire must
have green insulation with a yellow stripe or must be noninsulated
(bare).
2.Attach the opposite end of the wire to permanent earth ground using
toothed washers or a toothed lug.
This chapter describes how to unpack the PXI-1006, install a controller,
and verify that the chassis is operating correctly.
Unpacking the PXI-1006 and Installing a Controller
This installation example assumes you are using an embedded PXI
controller.
1.Unpack your PXI-1006 chassis.
2.Make sure the power switch is in the Standby (Off) position.
3.Plug in the AC power cord.
Power cords are available from National Instruments, or use a power
cord that meets the requirements of ANSI/UL817 for use in the United
States, CSA C22.2 21 and 49 for use in Canada, and IEC 227 or 245
for use in the European Union. AC mains power supply cords used
with the PXI-1006 in other countries must be approved by the authority
having jurisdiction in that country.
4.Install your controller in Slot 1 of your PXI-1006. Install peripheral
cards, if any, in slots 2 through 18.
1
4
2
3
1 PXI-1006 Chassis
2 Peripheral Card
3 Ejector Handle in Down Position
4 PXI Controller
Figure 1-1. Installing Controller and PXI Modules
5.Set the fan speed selector switch (located on the rear of the chassis) to
either HI for maximum cooling or AUTO to employ the temperature
sensing module that controls the fan speed.
6.Power on the chassis. Verify that all fans become operational.
7.Verify that your controller boots the operating system.
Note
If the chassis does not power on, see Chapter 3, Troubleshooting.
PXI-1006 User Manual1-2ni.com
PXI-1006 Features
This chapter describes the key features of the PXI-1006 chassis and lists
optional equipment you can order from National Instruments.
Key Features
The PXI-1006 combines a high-performance 18-slot PXI backplane with a
high-output power supply and a structural design that has been optimized
for maximum usability in a wide range of applications. The chassis’
modular design ensures the highest level of maintainability resulting in a
very low mean time to repair (MTTR). The PXI-1006 fully complies with
the PXI Specification, Revision 2.0, offering advanced timing and
synchronization features.
The key features of the PXI-1006 include:
•PXI and CompactPCI (PICMG 2.0 R 3.0) module compatibility
•3U-sized, 18-slot chassis
•600 W of usable power
•Universal AC input: auto-voltage and auto-frequency ranging
•Over-current protection via push-reset circuit breaker (no loose fuses
to replace)
•Removable modular power supply
•Remote power status and inhibit via a rear panel connector
•On/Off (Standby) switch located on the front panel for easy access
•Selectable fan speed for maximum cooling or quiet operation
The PXI-1006 backplane is interoperable with PXI-compatible products
and standard CompactPCI products. This is an important feature, as many
PXI-compatible systems may require components that do not implement
PXI-specific features. For example, you may want to use a standard
CompactPCI network interface card in a PXI chassis.
The signals on the P1 connector of the backplane meet the requirements of
the CompactPCI specification for both the peripheral and system modules.
The PXI-specific signals are located on P2 and are found on only the
signals that are reserved or not used in the CompactPCI 64-bit
specification. Therefore, all modules that meet the requirements of the
CompactPCI 64-bit specification will function in the PXI-1006.
The PXI-1006 has three PCI bus segments linked via two 64-bit PCI-to-PCI
bridges. The PCI bus segment divisions are represented on the front of the
chassis by vertical bars. See Figure 2-1 for the PCI segment divider
locations.
Figures 2-1 and 2-2 show some of the key features and components of the
PXI-1006 chassis. Figure 2-1 shows the front view of the PXI-1006.
Figure 2-2 shows the rear view.
The System Controller slot is Slot 1 of the chassis as defined by the PXI
specification. It has three controller expansion slots, which are used for
system controller modules that are wider than one slot. As defined in the
PXI specification, these slots allow the controller to expand to the left to
prevent the controller from using up peripheral slots.
Star Trigger Slot
The Star Trigger (ST) slot is Slot 2. This slot has a dedicated trigger line
between itself and slots 3-15. Slots 16-18 do not use the Star Trigger
functionality because there are 13 Star Trigger lines, as specified in the PXI
specification (see Figure 2-3). This slot is intended for modules with ST
functionality that can provide individual triggers to the peripherals installed
in slots 3-15. However, if you do not require advanced trigger functionality,
you can install any standard peripheral module into this slot.
Peripheral Slots
There are 17 peripheral slots including the Star Trigger controller slot.
Local Bus
The PXI backplane’s local bus is a daisy-chained bus that connects each
peripheral slot with its adjacent peripheral slots to the left and right, as
shown in Figure 2-3.
For example, a given peripheral slot’s right local bus connects to the
adjacent slot’s left local bus, and so on. Each local bus is 13 lines wide and
can pass analog signals between cards or provide a high-speed side-band
communication path that does not affect the PXI bandwidth.
The PXI-1006 has local bus connections between all slots except slots 1
and 2, in accordance with the PXI specification. See Figure 2-3 for details.
PXI-1006 User Manual2-4ni.com
Clock 10 Buffer and Detection Circuitry
PXI-1006 Backplane Configuration
Chapter 2PXI-1006 Features
18
Bus
Local
17
Bus
Local
16
Bus
Local
15
Bus
Local
14
Bus
Local
PXI Trigger Bus Segment 3 and PCI Bus Segment 3
13
Bus
Local
12
Bus
Local
11
Bus
Local
10
Bus
Local
9
Bus
Local
8
Bus
Local
PXI Trigger Bus Segment 2 and PCI Bus Segment 2
7
Bus
Local
6
Bus
Local
5
Bus
Local
4
Bus
Local
3
Bus
Local
Star Trigger Controller
2
2
PXI Trigger Bus Segment 1 and PCI Bus Segment 1
PXI Star Triggers
System Controller
1
Figure 2-3. PXI Local Bus and Star Trigger Routing
The PXI-1006 has three PXI segments. Vertical bars represent the segment
divisions on the front of the chassis. (See Figure 2-1 for the segment divider
locations.) The PXI trigger bus is divided into three separate trigger buses
at these locations. You can use the trigger lines in a variety of ways. For
example, you can use triggers to synchronize the operation of several
different PXI peripheral modules. In other applications, one module can
control carefully timed sequences of operations performed on other
modules in the system. Modules can pass triggers to one another, allowing
precisely timed responses to asynchronous external events the system is
monitoring or controlling.
System Reference Clock
The PXI-1006 supplies the PXI 10 MHz system clock signal (PXI_CLK10)
independently to every peripheral slot. An independent buffer (having a
source impedance matched to the backplane and a skew of less than 1 ns
between slots) drives the clock signal to each peripheral slot. You can use
this common reference clock signal to synchronize multiple modules in a
measurement or control system. You can drive PXI_CLK10 from an
external source through the PXI_CLK10_IN pin on the P2 connector of the
star trigger slot. (See Table B-1, P1 (J1) Connector Pinout for the System
Controller Slot.) Sourcing an external clock on this pin automatically
disables the backplane’s 10 MHz source.
Rack Mounting
Rack-mount applications require the optional rack-mount kit available
from National Instruments. Refer to the instructions supplied with the
rack-mount kit to install your PXI-1006 in a standard 19 in. instrument
rack.
PXI-1006 User Manual2-6ni.com
Chapter 2PXI-1006 Features
Remote Power Monitoring and Inhibiting Interface
The PXI-1006 chassis supports remote power monitoring and inhibiting via
a 9-pin D-sub connector located on the rear panel. Table 2-1 shows the
pinout of the DB-9 connector.
Table 2-1. DB-9 Connector Pinout
DB-9 PinSignal
1Logic Ground
2+5 V
3Reserved
4+3.3 V
5Inhibit*
6+12 V
7Reserved
8–12 V
9Logic Ground
7
6
1
2
89
3
4
5
Using the Inhibiting Feature
You can use the Inhibit signal (active low) to turn off the power supply
outputs. To use this feature, connect the Inhibit pin (pin 5) to a Logic
Ground pin (pin 1 or 9). Make sure the front (standby) switch is in the ON
position. As long as the connection is made, the power supply inhibits its
DC outputs. DC output resumes when Inhibit is no longer connected. Note
that the power (standby) switch, located on the front of the chassis, uses this
inhibiting feature. For remote reset, connect a momentary switch between
pin5andpin1(orpin9).
You can use a digital voltmeter to ensure all voltage levels in your
PXI-1006 are within the allowable limits.
Caution
When connecting digital voltmeter probes to the rear D-sub connector, be careful
not to short the probe leads together. Doing so could damage the power supply.
Referring to Table 2-2, connect one lead of the voltmeter to a supply pin on
the remote power monitoring connector (9-pin D-sub) on the rear panel.
Refer to Table 2-1 for a pinout diagram of the remote power monitoring
connector. Connect the reference lead of the voltmeter to one of the ground
pins. Compare each voltage reading to the values listed in Table 2-2.
Note
Use the rear-panel D-sub connector to check voltages only. Do not use the connector
to supply power to external devices.
Table 2-2. Power Supply Voltages at Power Monitoring Connector (DB-9)
PinSupplyAcceptable Voltage Range
2+5 V4.75 to 5.25 V
4+3.3 V3.135 to 3.465 V
6+12 V11.4 to 12.6 V
8–12 V–12.6 to –11.4 V
1, 9Logic GroundN/A
If the voltages fall within the specified ranges, the chassis complies with the
CompactPCI voltage limit specifications. Notice that the rear-panel D-sub
connector is to be used to check voltages only. Do not use these voltages to
supply power to external devices.
Note
If the fans or power unit fail to function properly, refer to Chapter 3,
Troubleshooting.
PXI-1006 User Manual2-8ni.com
Installing Filler Panels
To optimize module cooling performance, install filler panels into unused
or empty slots. Secure with the captive mounting screws.
Using the Chassis Initialization File
To assist system integrators, the PXI specification requires manufacturers
of PXI chassis and system modules to document the capabilities of their
products. The minimum documentation requirements are contained
in
read the
use this file. The PXI-1006 chassis initialization file,
is included on the diskette for your PXI-1006.
files, which consist of ASCII text. The system integrator can
.ini
file, and configuration utilities and device drivers can also
.ini
Cleaning
Cleaning procedures consist of exterior and interior cleaning of the chassis
and cleaning the fan filters. Refer to your module user documentation for
information on cleaning the individual CompactPCI or PXI modules.
Chapter 2PXI-1006 Features
chassis.ini
,
Caution
servicing the chassis.
Mise en garde
de nettoyer ou d’intervenir sur le châssis.
Always power-off the chassis and disconnect the power cord before cleaning or
Veillez àéteindrelechâssis et déconnecter le cordon d’alimentation avant
Interior Cleaning
Use a dry, low-velocity stream of air to clean the interior of the chassis. Use
a soft-bristle brush for cleaning around components. If you must use a
liquid for minor interior cleaning, use a 75% isopropyl alcohol solution and
rinse with deionized water.
Exterior Cleaning
Clean the exterior surfaces of the chassis with a dry lint-free cloth or a
soft-bristle brush. If any dirt remains, wipe with a cloth moistened in a mild
soap solution. Remove any soap residue by wiping with a cloth moistened
with clear water. Do not use abrasive compounds on any part of the chassis.
Avoid getting moisture inside the chassis during exterior cleaning. Use just
enough moisture to dampen the cloth. Do not wash the front- or rear-panel connectors or
switches. Cover these components while cleaning the chassis. Do not use chemical
cleaning agents; they may damage the chassis. Avoid chemicals that contain benzene,
toluene, xylene, acetone, or similar solvents.
Cleaning the Fan Filters
You can easily remove the chassis cooling filters from the rear of the
chassis by removing the plastic housing attached to each fan.
Clean the fan filters by washing them in a mild soap solution and then
vacuuming or blowing air through them. Rinse the filters with water and
allow them to dry before reinstalling them on the chassis.
PXI-1006 User Manual2-10ni.com
Troubleshooting
This chapter describes basic troubleshooting procedures you can perform
on the PXI-1006 chassis.
Troubleshooting the PXI-1006
Refer to Table 3-1 to troubleshoot the PXI-1006 chassis. The table lists
possible causes for power failure and recommends ways to correct the
problem.
3
Table 3-1.
Possible CauseWhat to Do
PXI-1006 chassis is not connected to
power source.
Power switch is not switched on.Set the power switch to the On position.
Remote inhibiting input on the rear
panel of the chassis is active.
Circuit breaker is tripped.Reset the circuit breaker. Refer to the Resetting the AC
Power supply has failed.Contact National Instruments.
Troubleshooting
Make sure that the PXI-1006 is connected to a live
electrical outlet. Try operating another piece of
equipment from this outlet.
Deactivate your system’s remote inhibiting control.
Mains Circuit Breaker section in this chapter.
Resetting the AC Mains Circuit Breaker
If your PXI-1006 is connected to an AC source and encounters an
over-current condition, the circuit breaker located on the rear panel will trip
to prevent damage to the chassis. Complete the following steps to reset the
circuit breaker:
If the circuit breaker trips again, complete the following steps:
1.Turn the front panel power switch to the Standby position.
2.Disconnect the chassis from the AC mains power source.
3.Remove all modules from the chassis.
4.Complete the test procedure described in the Using the Power
Monitoring Feature section in Chapter 2, PXI-1006 Features.
5.If any voltages are outside the acceptable limits, contact National
Instruments.
6.If all voltages are within the acceptable limits, verify that your
PXI-1006 can meet the power requirements of your CompactPCI or
PXI modules. Overloading the chassis can cause the breaker to trip.
Refer to Appendix A, Specifications.
7.The over-current condition that caused the circuit breaker to trip
may be due to a faulty CompactPCI or PXI module. Refer to the
documentation supplied with the modules for troubleshooting your
modules.
PXI-1006 User Manual3-2ni.com
Specifications
This appendix contains specifications for the PXI-1006 chassis.
Electrical
AC Input
Input voltage range.................................90–264 VAC
Input frequency range ............................ 47 to 63 Hz
Over-current protection.......................... 10 A circuit breaker
Maximum steady state
operating current .................................... 10 A
Line regulation ....................................... ± 0.1% over operating line range
Power disconnect ................................... The (standby) power switch
A
causes the power module to
supply DC power to
the CompactPCI/PXI backplane.
The rear-panel D-sub connector
facilitates remote inhibiting
operation. The (standby) switch
must be in the On position prior to
use of remote inhibit. The power
cord provides main power
disconnect.
Maximum usable power .........................600 W
DC current capacity (I
Vo l t ag eIMP(Steady-State Current)
+3.3 V60 A
+5 V60 A
+12 V9A
–12 V1.8 A
Load regulation
Vo l t ag eRegulation
+3.3 V0.4% or 20 mV max
+12 V0.4% or 20 mV max
+5 V0.4% or 20 mV max
–12 V0.4% or 20 mV max
Maximum ripple and noise .....................1% ripple, 1% noise
Over-current protection...........................105–140% of rated output
MP
)
20 MHz bandwidth
current; automatic recovery
Over-voltage protection ..........................3.3 V, 5 V clamp at 122–134%
of output voltage
+12 V and –12 V clamp at
110–120% of output voltage
Power supply/fan unit MTTR.................Replacement in under 5 minutes
Cooling
Module cooling system...........................Forced air circulation
(positive pressurization) via
three 140 cfm fans with
HI/AUTO speed selector
PXI-1006 User ManualA-2ni.com
Slot airflow direction.............................. P1 to P2, bottom of module to top
Module/power supply cooling intake..... Three rear fans
Module cooling exhaust ......................... Along both sides and top of
Power supply cooling system................. Forced air circulation via
Power supply cooling exhaust................ Along upper left side
Power supply/fan unit ............................ Replacement in under 5 minutes
Electromagnetic Compatibility
EMC/EMI............................................... CE, C-Tick and FCC Part 15
Electrical emissions................................ EN 55011 Class A @ 10 m.
Appendix ASpecifications
of module
chassis
integrated fan
(Class A) compliant
FCC Part 15A above 1 GHz
Note
This device should be operated only with shielded cabling for full EMC and
EMI compliance. See the Declaration of Conformity for this product for any additional
regulatory compliance information.
Safety
Environmental
Electrical immunity................................ Evaluated to EN 61326:1998,
Tabl e 1
Designed in accordance with IEC 61010-1, EN 61010-1, UL 3111-1, and
CAN/CSA C22.2 No. 1010.1 for electrical measuring and test equipment
Installation Category II
Pollution Degree 2
Operating temperature............................ 0 to 50 °C
Storage temperature ............................... –20 to 70 °C
Operating relative humidity ................... Maximum 80% for temperatures
Functional shock (operating) ..................MIL-T-28800E CLASS 3,
Half-Sine Shock Pulse, 11 ms
duration, 30 g peak
Operating location ..................................Indoor use only
Random vibration
Operational
Non-operational1.............................10 to 500 Hz, 2.46 g
1
.....................................5 to 500 Hz, 0.31 g
RMS
RMS
EMC emissions.......................................FCC Class A compliant and
EN 55011 Group 1 Class A
Compliant
EMC immunity .......................................Refer to DOC supplied with
chassis for compliance to
relevant directives.
Altitude ...................................................2 km (1.24 mi) max
Size .........................................................3U-sized; one system slot
(with three system expansion
slots)and17peripheralslots.
Compliant with IEEE 1101.10
mechanical packaging.
PXI Specification Revision 2.0
compliant.
Accepts both PXI and
CompactPCI (PICMG 2.0 R3.0)
3U modules.
Backplane bare-board material ...............UL 94 V-0 recognized
Backplane connectors .............................Conform to IEC 917 and
IEC 1076-4-101, and are
UL 94 V-0 rated
1
Random vibration profiles were developed in accordance with MIL-T-28800E CLASS 3 and MIL-STD-810E Method 514
Test levels exceed those recommended in MIL-STD-810E for Category 1 (Basic Transportation), Figures 514.4-1 through
514.4-3.
PXI-1006 User ManualA-4ni.com
Mechanical
Appendix ASpecifications
Weight .................................................... 13.2 kg (29.2 lb.)
Maximum module weight ...................... 1.8 kg (4 lb.)
Materials................................................. Sheet Aluminum (5052-H32 and
Figure A-1 shows the PXI-1006 dimensions. The holes shown are for the
installation of the optional rack-mount kit. You can install this kit on the
front or rear of the chassis, depending on which end of the chassis you want
to face toward the front of the instrument cabinet. Note that the front and
rear chassis mounting holes (size M4) are symmetrical.
6.97 in.
(177.0 mm)
17.54 in. (445.5 mm)
16.89 in. (429.0 mm)
1.82 in. (46.3 mm)
1.50 in.
(38.0 mm)
3.54 in.
(90.0 mm)
1.88 in. (47.8 mm)
1.84 in.
(46.8 mm)
0.57 in. (14.5 mm)
Figure A-1. PXI-1006 Dimensions
PXI-1006 User ManualA-6ni.com
Pinouts
Note
B
This appendix describes the P1 and P2 connector pinouts for the PXI-1006
backplane.
Table B-1 shows the P1 (J1) connector pinout for the System
Controller slot.
Table B-2 shows the P2 (J2) connector pinout for the System
Controller slot.
Table B-3 shows the P1 (J1) connector pinout for the Star Trigger slot.
Table B-4 shows the P2 (J2) connector pinout for the Star Trigger slot.
Table B-5 shows the P1 (J1) connector pinout for the peripheral slots.
Table B-6 shows the P2 (J2) connector pinout for the peripheral slots.
National Instruments Web support is your first stop for help in solving
installation, configuration, and application problems and questions. Online
problem-solving and diagnostic resources include frequently asked
questions, knowledge bases, product-specific troubleshooting wizards,
manuals, drivers, software updates, and more. Web support is available
through the Technical Support section of
NI Developer Zone
ni.com
C
The NI Developer Zone at
building measurement and automation systems. At the NI Developer Zone,
you can easily access the latest example programs, system configurators,
tutorials, technical news, as well as a community of developers ready to
share their own techniques.
Customer Education
National Instruments provides a number of alternatives to satisfy your
training needs, from self-paced tutorials, videos, and interactive CDs to
instructor-led hands-on courses at locations around the world. Visit the
Customer Education section of
syllabi, training centers, and class registration.
System Integration
If you have time constraints, limited in-house technical resources, or other
dilemmas, you may prefer to employ consulting or system integration
services. You can rely on the expertise available through our worldwide
network of Alliance Program members. To find out more about our
Alliance system integration solutions, visit the System Integration section
of
National Instruments has offices located around the world to help address
your support needs. You can access our branch office Web sites from the
Worldwide Offices section of
up-to-date contact information, support phone numbers, e-mail addresses,
and current events.
If you have searched the technical support resources on our Web site and
still cannot find the answers you need, contact your local office or National
Instruments corporate. Phone numbers for our worldwide offices are listed
at the front of this manual.
RMSRoot mean square. A method used to measure electrical output in volts and
watts
S
sSeconds
STStar Trigger
Star Trigger slotThis slot is located at slot 2 and has a dedicated trigger line between each
peripheral slot. Use this slot for a module with ST functionality that can
provide individual triggers to all other peripherals.
System controllerA module configured for installation in Slot 0 of a VXIbus mainframe. This
device is unique in the VXIbus system in that it performs the VMEbus
system controller functions, including clock sourcing and arbitration for
data transfers across the backplane. Installing such a device into any other
slot can damage the device, the VXIbus backplane, or both.
U
ULUnderwriter’s Laboratories
V
VVolts
VACVolts alternating current
V
PP
Peak to peak voltage
W
WWatts
PXI-1006 User ManualG-4ni.com
Index
B
backplane, 2-2
interoperability with CompactPCI, 2-2
local bus, 2-4
overview, 2-2
peripheral slots, 2-4
specifications, A-4
Star Trigger (ST) slot, 2-4
system reference clock, 2-6
trigger bus, 2-6