National Instruments NI PS-15, PS-15 User Manual

Page 1
Power Supply
NI PS-15 Power Supply User Manual
NI PS-15 Power Supply User Manual
June 2011 372911B-01
Page 2
Worldwide Technical Support and Product Information
ni.com
Visit ni.com/niglobal to access the branch office Web sites, which provide up-to-date contact information, support phone numbers, email addresses, and current events.
National Instruments Corporate Headquarters
11500 North Mopac Expressway Austin, Texas 78759-3504 USA Tel: 512 683 0100
For further support information, refer to the Technical Support and Professional Services appendix. To comment on National Instruments documentation, refer to the National Instruments Web site at the Info Code
feedback.
© 2009–2011 National Instruments Corporation. All rights reserved.
ni.com/info and enter
Page 3
Important Information
Warranty
The NI PS-15 is warranted against defects in materials and workmanship for a period of one year from the date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that proves to be defective during the warranty period. This warranty includes parts and labor.
The media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects in materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace software media that do not execute programming instructions if National Instruments receives notice of such defects during the warranty period. National Instruments does not warrant that the operation of the software shall be uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are covered by warranty.
National Instruments believes that the information in this document is accurate. The document has been carefully reviewed for technical accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent editions of this document without prior notice to holders of this edition. The reader should consult National Instruments if errors are suspected. In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.
XCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF
E
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. CUSTOMERS RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART OF NATIONAL
NSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE CUSTOMER. NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR DAMAGES RESULTING
I
FROM LOSS OF DATA, PROFITS, USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY THEREOF. This limitation of
the liability of National Instruments will apply regardless of the form of action, whether in contract or tort, including negligence. Any action against National Instruments must be brought within one year after the cause of action accrues. National Instruments shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not cover damages, defects, malfunctions, or service failures caused by owner’s failure to follow the National Instruments installation, operation, or maintenance instructions; owner’s modification of the product; owner’s abuse, misuse, or negligent acts; and power failure or surges, fire, flood, accident, actions of third parties, or other events outside reasonable control.
Copyright
Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical, including photocopying, recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National Instruments Corporation.
National Instruments respects the intellectual property of others, and we ask our users to do the same. NI software is protected by copyright and other intellectual property laws. Where NI software may be used to reproduce software or other materials belonging to others, you may use NI software only to reproduce materials that you may reproduce in accordance with the terms of any applicable license or other legal restriction.
Trademarks
LabVIEW, National Instruments, NI, ni.com, the National Instruments corporate logo, and the Eagle logo are trademarks of National Instruments Corporation. Refer to the Trademark Information at
Other product and company names mentioned herein are trademarks or trade names of their respective companies.
Members of the National Instruments Alliance Partner Program are business entities independent from National Instruments and have no agency, partnership, or joint-venture relationship with National Instruments.
ni.com/trademarks for other National Instruments trademarks.
Patents
For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents.
Export Compliance Information
Refer to the Export Compliance Information at ni.com/legal/export-compliance for the National Instruments global trade compliance policy and how to obtain relevant HTS codes, ECCNs, and other import/export data.
WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS
(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING FOR A LEVEL OF RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL IMPLANTS OR AS CRITICAL COMPONENTS IN ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE TO PERFORM CAN REASONABLY BE EXPECTED TO CAUSE SIGNIFICANT INJURY TO A HUMAN.
(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE SOFTWARE PRODUCTS CAN BE IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT LIMITED TO FLUCTUATIONS IN ELECTRICAL POWER SUPPLY, COMPUTER HARDWARE MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE FITNESS, FITNESS OF COMPILERS AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION, INSTALLATION ERRORS, SOFTWARE AND HARDWARE COMPATIBILITY PROBLEMS, MALFUNCTIONS OR FAILURES OF ELECTRONIC MONITORING OR CONTROL DEVICES, TRANSIENT FAILURES OF ELECTRONIC SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR MISUSES, OR ERRORS ON THE PART OF THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS THESE ARE HEREAFTER COLLECTIVELY TERMED “SYSTEM FAILURES”). ANY APPLICATION WHERE A SYSTEM FAILURE WOULD CREATE A RISK OF HARM TO PROPERTY OR PERSONS (INCLUDING THE RISK OF BODILY INJURY AND DEATH) SHOULD NOT BE RELIANT SOLELY UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM FAILURE. TO AVOID DAMAGE, INJURY, OR DEATH, THE USER OR APPLICATION DESIGNER MUST TAKE REASONABLY PRUDENT STEPS TO PROTECT AGAINST SYSTEM FAILURES, INCLUDING BUT NOT LIMITED TO BACK-UP OR SHUT DOWN MECHANISMS. BECAUSE EACH END-USER SYSTEM IS CUSTOMIZED AND DIFFERS FROM NATIONAL INSTRUMENTS' TESTING PLATFORMS AND BECAUSE A USER OR APPLICATION DESIGNER MAY USE NATIONAL INSTRUMENTS PRODUCTS IN COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT
Page 4
EVALUATED OR CONTEMPLATED BY NATIONAL INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS ULTIMATELY RESPONSIBLE FOR VERIFYING AND VALIDATING THE SUITABILITY OF NATIONAL INSTRUMENTS PRODUCTS WHENEVER NATIONAL INSTRUMENTS PRODUCTS ARE INCORPORATED IN A SYSTEM OR APPLICATION, INCLUDING, WITHOUT LIMITATION, THE APPROPRIATE DESIGN, PROCESS AND SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.
Page 5
Compliance
Electromagnetic Compatibility Information
This hardware has been tested and found to comply with the applicable regulatory requirements and limits for electromagnetic compatibility (EMC) as indicated in the hardware’s Declaration of Conformity (DoC) designed to provide reasonable protection against harmful interference when the hardware is operated in the intended electromagnetic environment. In special cases, for example when either highly sensitive or noisy hardware is being used in close proximity, additional mitigation measures may have to be employed to minimize the potential for electromagnetic interference.
While this hardware is compliant with the applicable regulatory EMC requirements, there is no guarantee that interference will not occur in a particular installation. To minimize the potential for the hardware to cause interference to radio and television reception or to experience unacceptable performance degradation, install and use this hardware in strict accordance with the instructions in the hardware documentation and the DoC
If this hardware does cause interference with licensed radio communications services or other nearby electronics, which can be determined by turning the hardware off and on, you are encouraged to try to correct the interference by one or more of the following measures:
Reorient the antenna of the receiver (the device suffering interference).
Relocate the transmitter (the device generating interference) with respect to the receiver.
Plug the transmitter into a different outlet so that the transmitter and the receiver are on different branch circuits.
Some hardware may require the use of a metal, shielded enclosure (windowless version) to meet the EMC requirements for special EMC environments such as, for marine use or in heavy industrial areas. Refer to the hardware’s user documentation and the DoC
When the hardware is connected to a test object or to test leads, the system may become more sensitive to disturbances or may cause interference in the local electromagnetic environment.
Operation of this hardware in a residential area is likely to cause harmful interference. Users are required to correct the interference at their own expense or cease operation of the hardware.
Changes or modifications not expressly approved by National Instruments could void the user’s right to operate the hardware under the local regulatory rules.
1
for product installation requirements.
1
.
1
. These requirements and limits are
1
The Declaration of Conformity (DoC) contains important EMC compliance information and instructions for the user or installer. To obtain the DoC for this product, visit and click the appropriate link in the Certification column.
ni.com/certification, search by model number or product line,
Page 6
Contents
About This Manual
Conventions ...................................................................................................................viii
Related Documentation..................................................................................................ix
Chapter 1 Getting Started
Unpacking......................................................................................................................1-1
What You Need to Get Started ......................................................................................1-1
Key Features ..................................................................................................................1-1
Power Supply Description .............................................................................................1-2
Output Terminals.............................................................................................1-3
Output Voltage Potentiometer.........................................................................1-4
DC OK LED....................................................................................................1-4
Input Terminals ...............................................................................................1-4
Mounting Equipment .....................................................................................................1-5
Side Mounting Kit ...........................................................................................1-5
Panel Mounting Kit .........................................................................................1-5
Chapter 2 Installation and Configuration
Mounting Orientation and Installation...........................................................................2-1
Wiring The Terminals....................................................................................................2-3
Operating the NI PS-15..................................................................................................2-5
Serial Operation...............................................................................................2-5
Parallel Operation............................................................................................2-6
Two-Phase Power Operation...........................................................................2-6
External Input Protection.................................................................................2-7
Operation in a Sealed Enclosure......................................................................2-7
Cooling ............................................................................................................2-8
Hazardous Risks ..............................................................................................2-8
Service Parts ....................................................................................................2-8
Peak Current Capability ..................................................................................2-9
Charging Batteries ...........................................................................................2-9
Back Feeding Loads ........................................................................................2-10
Output Circuit Breakers...................................................................................2-10
Inductive and Capacitive Loads ......................................................................2-12
© National Instruments Corporation vi NI PS-15 Power Supply User Manual
Page 7
Contents
Appendix A Specifications
Dimensions and Weight ................................................................................................ A-1
AC Input ........................................................................................................................A-3
Input Current Inrush Surge............................................................................................ A-5
Hold-up Time ................................................................................................................ A-6
Output ............................................................................................................................ A-7
Peak Current Capability .................................................................................. A-8
Efficiency and Power Losses......................................................................................... A-8
Reliability ...................................................................................................................... A-9
Dielectric Strength......................................................................................................... A-10
Used Substances ............................................................................................................ A-11
Environment .................................................................................................................. A-12
Protection Features ........................................................................................................ A-13
Appendix B Technical Support and Professional Services
Index
NI PS-15 Power Supply User Manual vii ni.com
Page 8
About This Manual
The NI PS-15 Power Supply User Manual describes the features and specifications of the NI PS-15 power supply and contains information about installing the power supply.
Conventions
The following conventions are used in this manual:
» The » symbol leads you through nested menu items and dialog box options
to a final action. The sequence Options»Settings»General directs you to pull down the Options menu, select the Settings item, and select General from the last dialog box.
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to avoid injury, data loss, or a system crash.
When this symbol is marked on a product, it denotes a warning advising you to take precautions to avoid electrical shock.
bold Bold text denotes items that you must select or click in the software, such
as menu items and dialog box options. Bold text also denotes parameter names.
italic Italic text denotes variables, emphasis, a cross-reference, or an introduction
to a key concept. Italic text also denotes text that is a placeholder for a word or value that you must supply.
monospace Text in this font denotes text or characters that you should enter from the
keyboard, sections of code, programming examples, and syntax examples. This font is also used for the proper names of disk drives, paths, directories, programs, subprograms, subroutines, device names, functions, operations, variables, filenames, and extensions.
© National Instruments Corporation viii NI PS-15 Power Supply User Manual
Page 9
Related Documentation
The following documents contain information that you might find helpful as you read this manual:
NI PS-15/16/17 Side Mount Brackets Installation Guide
NI PS-15/16/17 Panel Mount Brackets Installation Guide
NI PS-15/16/17 Instruction Manual
Page 10
Getting Started
This chapter describes the key features of the NI PS-15 power supply and lists the kit contents and mounting equipment you can order from National Instruments.
Unpacking
Carefully inspect the shipping container and the power supply for damage. Check for visible damage to the metal work. If damage appears to have been caused during shipment, file a claim with the carrier. Retain the packing material for possible inspection and/or reshipment.
What You Need to Get Started
1
Key Features
The NI PS-15 power supply kit contains the following items:
NI PS-15 power supply
Printed NI PS-15/16/17 Instruction Manual
The NI PS-15 includes a power reserve of 20%. This extra current may be used continuously at temperatures up to +45 °C. Additionally, the AC 100–120 V/200–240 V Auto-select input makes worldwide installation and usage very simple.
The key features of the NI PS-15 power supply include the following:
Small size—power supply width of only 32 mm
High efficiency—efficiency up to 90.2%
Wide temperature range—full output power between –25 °C and +60 °C
AC 100–120/200–240 V Auto-select input
20% output power reserves
Minimal inrush current surge
© National Instruments Corporation 1-1 NI PS-15 Power Supply User Manual
Page 11
Chapter 1 Getting Started
+
+
– –
Power
Converter
Output
Filter
DC OK
L
N
Input Fuse
And
Input Filter
Input
Rectifier
And Inrush
Limiter
115/230 V
Auto Select
Output
Over-Voltage
Protection
Output
Voltage
Regulator
V
OUT
Hazardous Voltages Must be mounted in an enclosure by qualified personnel.
This power supply is designed for installation in an enclosure and is intended for general use, such as in industrial control, office, communication, and instrumentation equipment. Do not use this device in aircraft, trains and nuclear equipment, where malfunctioning of the power supply may cause severe personal injury or threaten human life. For more information, refer to Appendix A, Specifications.
Power Supply Description
Figures 1-1 and 1-2 show the functional description and front panel features of the NI PS-15 power supply.
Figure 1-1 provides the functional diagram for the NI PS-15.
NI PS-15 Power Supply User Manual 1-2 ni.com
Figure 1-1. Functional Diagram for the NI PS-15 Power Supply
Page 12
Chapter 1 Getting Started
DC 24 V 10 A
DC OK
NL
AC 100-120 / 200-240 V
24 - 28V
Power Supply
NI PS-16
2
3
4
1
Figure 1-2 provides the front panel features of the NI PS-15.
1Output Terminals 2Output Voltage Potentiometer
3 DC OK LED 4Input Terminals
Figure 1-2. Front View of the NI PS-15 Power Supply
Output Terminals
The NI PS-15 has a total of four output terminals, providing two positive (+) output terminals and two negative (–) output terminals. Both positive terminals are wired together internally, and both negative terminals are wired together internally, as shown in Figure 1-1. The output
© National Instruments Corporation 1-3 NI PS-15 Power Supply User Manual
terminals provide 24 V
with 5 A of current.
DC
Page 13
Chapter 1 Getting Started
Output Voltage Potentiometer
Note You must open the protective flap to turn the potentiometer.
Output voltage from the NI PS-15 is set by the output voltage potentiometer, shown in Figure 1-2. The factory setting output voltage is 24.1 V ±0.2% (at full load when the power supply is cold), and the potentiometer allows the output voltage to be adjusted from 24–28 V on any unit. The output voltage may be adjusted above 28 V by the potentiometer, but voltages beyond 28 V are not guaranteed.
Note Output voltages greater than 28 V are not supported on an NI PS-15 unit.
DC OK LED
This green LED indicates the status of available DC power through the output terminals. If the LED is lit, DC output of greater than 21 V is available for use through the output terminals. If the LED is not lit, DC is not currently available. The DC OK LED is wired internally to the power conversion circuitry prior to the output filtering stage, as shown in Figure 1-1.
Input Terminals
Caution National Instruments recommends that you wire all three input terminals for
proper operation of the NI PS-15.
If the DC OK LED does not light when power is provided through the input terminals, it may indicate a problem with the power supply. Contact National Instruments for more details.
The NI PS-15 power supply derives power through the input terminals on the front panel, shown in Figure 1-2. There are three terminals corresponding to the Neutral input, the Line (or hot) input, and the Protective Earth (PE) input. The NI PS-15 rectifies both single-phase and two-phase AC input. The Neutral input terminal provides a MAINS return path for the input circuitry. The Line input is the primary power input for the supply. The PE input corresponds to an earth ground. As shown in Figure 1-1, the power supply case itself is grounded to the PE input.
NI PS-15 Power Supply User Manual 1-4 ni.com
Page 14
Mounting Equipment
Contact National Instruments to order the following mounting options for the NI PS-15 power supply. Refer to Table 1-1 for part numbers.
Part Number Mounting Kit
199429-01 SIDE MOUNTING KIT FOR NI PS-15
199432-01 PANEL MOUNTING KIT FOR NI PS-15/16/17
Side Mounting Kit
The Side Mounting Kit (199429-01) allows you to mount the NI PS-15 on its side to a wall, panel surface, or a DIN-Rail for reduced installation depth. Refer to the NI PS-15/16/17 Side Mount Brackets Installation Guide
ni.com for more information.
at
Chapter 1 Getting Started
Table 1-1. Mounting Equipment
Panel Mounting Kit
The Panel Mounting Kit (199432-01) allows you to mount the NI PS-15 to a wall or panel surface without using a DIN-Rail. Refer to the NI PS-15/16/17 Panel Mount Brackets Installation Guide at for more information.
ni.com
© National Instruments Corporation 1-5 NI PS-15 Power Supply User Manual
Page 15
Installation and Configuration
This chapter describes how to prepare and operate the NI PS-15 power supply.
Mounting Orientation and Installation
This section describes the different mounting orientations, and the effect that mounting orientation has on power supply performance.
Mounting orientations other than input terminals on the bottom and output on the top require a reduction in continuous output power or a limitation in the maximum allowed ambient temperature. The amount of reduction influences the lifetime expectancy of the power supply. Therefore, two different derating curves for continuous operation are referenced in Figure 2-1:
2
Curve A1—Recommended output current.
Curve A2—Maximum allowed output current (which results in
approximately half the lifetime expectancy for the power supply when following curve A1).
Note National Instruments recommends that the power supply be oriented such that the
output terminals are located on top and the input terminals located on bottom. Figure 2-1 refers to this as the Standard Orientation.
© National Instruments Corporation 2-1 NI PS-15 Power Supply User Manual
Page 16
Chapter 2 Installation and Configuration
Mounting Orientation A (Standard Orientation)
OUTPUT
Power
Supply
INPUT
Output Current
12 A
9
6
3
0
Ambient Temperature
10 20 30 40
50
A1
60°C
Mounting Orientation B (Upside Down)
Mounting Orientation C (Table-top Mounting)
Mounting Orientation D (Horizontal ccw)
INPUT
INPUT
Supply
Power
Output Current
12 A
9
6
A2
A1
3
OUTPUT
0
10 20 30 40
Output Current
Ambient Temperature
50
60°C
12 A
9
6
A2 A1
3
0
10 20 30 40
Output Current
Ambient Temperature
50
60°C
12 A
Supply
Power
OUTPUT
9
6
A2 A1
3
0
10 20 30 40
Ambient Temperature
50
60°C
Mounting Orientation E (Horizontal ccw)
OUTPUT
Mount the NI PS-15 power supply according to the installation instructions included with your mounting kit. For details on the mounting options available, refer to the Mounting Equipment section of Chapter 1, Getting Started.
NI PS-15 Power Supply User Manual 2-2 ni.com
Output Current
12 A
9
6
Power
INPUT
Supply
3
0
Ambient Temperature
10 20 30 40
A2 A1
50
60°C
Figure 2-1. NI PS-15 Mounting Orientations
Page 17
Wiring The Terminals
This section describes wiring for the NI PS-15 power supply. Table 2-1 provides a list of basic requirements for wiring.
Solid wire 0.5–6 mm
Stranded wire 0.5–4 mm
American wire gauge 20–10 AWG
Ferrules Allowed, but not required
Wire stripping length 7 mm (0.275 in.)
Chapter 2 Installation and Configuration
Table 2-1. Wiring Requirements
Type Spring-Clamp Terminals
Consider the following when wiring the NI PS-15.
Use appropriate copper cables that are designed for an operating temperature of:
60 °C for ambient up to 45 °C.
75 °C for ambient up to 60 °C minimum.
Follow national installation codes and installation regulations.
Up to two stranded wires with the same cross section are permitted in one connection point (except PE wire).
Do not use the unit without the PE connection being wired.
Complete the following steps to connect wires to the input and output terminals.
1. Ensure that none of the wires are connected to live power.
2. Strip the ends of the wires according to the recommendations in Tab le 2-1 .
© National Instruments Corporation 2-3 NI PS-15 Power Supply User Manual
Page 18
Chapter 2 Installation and Configuration
3. Ensure that the terminal lever is in an open position as shown in Figure 2-2.
4. Insert the end of the wire into the terminal until the exposed portion of the wire is completely inside of the terminal connection as shown in Figure 2-2. If you are using stranded wire, ensure that all strands of the wire enter the terminal connection.
Inserting The Wire Closing The Lever
Figure 2-2. Connecting a Wire
5. Move the lever until the lever snaps into the closed position, as shown in Figure 2-2.
6. Repeat steps 4–5 for each of the other terminals.
7. Ensure that all wires are properly seated and not loose.
8. Ensure that the rest of your equipment is ready to be powered without creating a hazard.
9. Apply MAINS voltage to the NI PS-15 power supply.
NI PS-15 Power Supply User Manual 2-4 ni.com
Page 19
Operating the NI PS-15
This section provides general information on the operation of the NI PS-15 power supply.
Serial Operation
The NI PS-15 can operate in series to increase the output voltage. Figure 2-3 shows the NI PS-15 in a serial configuration.
Unit A
AC
Unit B
AC
DC
Chapter 2 Installation and Configuration
+
+
Load
+
DC
Figure 2-3. NI PS-15 in Serial Operation
Earth
Before operating the NI PS-15 in a serial configuration, consider the following:
It is possible to connect as many units in series as needed, providing the sum of the output voltage does not exceed 150 V
Voltages with a potential above 60 V
are not SELV-compliant any
DC
DC
.
more and can be dangerous. Such voltages must be installed with a protection against touching.
Use power supplies of the same type for serial operation.
Earthing of the output is required when the sum of the output voltage is above 60 V
DC
.
Keep an installation clearance of 15 mm (left/right) between two power supplies and avoid installing the power supplies on top of each other.
Caution Avoid return voltage (for example, from a decelerating motor or battery) which is
applied to the output terminals.
© National Instruments Corporation 2-5 NI PS-15 Power Supply User Manual
Page 20
Chapter 2 Installation and Configuration
Parallel Operation
Caution The power supply must not be used in parallel for the purpose of increasing the
output power.
Power supplies can be paralleled for 1+1 redundancy to gain a higher system availability. Redundant systems require a certain amount of extra power to support the load in case one power supply unit fails. The simplest way is to put two NI PS-15 power supplies in parallel. If one power supply unit fails, the other one is automatically able to support the load current without any interruption. When using this method of building a redundant system consider the following:
The faulty power supply can not be recognized. The green LED will still be on since it is reverse-powered from the other power supply.
It does not cover failures such as an internal short circuit in the secondary side of the power supply. In such a case (nearly impossible), the defective unit becomes a load for the other power supplies and the output voltage can not be maintained.
Recommendations for building redundant power systems:
Use separate input fuses for each power supply.
When possible, connect each power supply to different phases or circuits.
Two-Phase Power Operation
The NI PS-15 power supply can operate with two-phase power, as shown in Figure 2-4.
L3
L1
L2
Power Supply
AC
internal
fused
Fuse
L
N
PE
DC
max.
+10%
240 V
Before operating the NI PS-15 in this configuration, consider the following:
NI PS-15 Power Supply User Manual 2-6 ni.com
Figure 2-4. NI PS-15 in Two-Phase Operation
Page 21
A phase-to-phase connection is allowed as long as the supplying voltage is below 240 V +10%.
Use a fuse or a circuit breaker to protect the N (Neutral) input. The N input is not protected internally and in two-phase configuration would be connected to a hot wire.
Appropriate fuses and circuit breakers are specified in the External Input Protection section.
External Input Protection
The NI PS-15 power supply is tested and approved for branch circuits up to 20 A. External protection is only required if the supplying branch has an ampacity greater than 20 A. In some countries local regulations might apply, so check local codes and local requirements.
If an external fuse is utilized, a minimum value is required to avoid undesired tripping of the fuse, shown in Table 2-2.
Chapter 2 Installation and Configuration
Table 2-2. Maximum and Minimum Ampacities for External Fuses
Ampacity B-Characteristic C-Characteristic
Minimum 10 A 6 A
Maximum 20 A 20 A
Operation in a Sealed Enclosure
When the power supply is installed in a tightly sealed enclosure, the temperature inside the enclosure will be higher than outside. The inside temperature defines the ambient temperature for the power supply.
The following is the result of such an installation, where the NI PS-15 power supply was placed in the middle of a sealed enclosure, and no other heat producer was present:
Enclosure................................................ Rittal Type IP66 Box PK 9516
100, plastic, 110 mm × 180 mm × 165 mm
Load ....................................................... 24 V, 4 A; (= 80%) load is placed
Input ....................................................... 230 V
© National Instruments Corporation 2-7 NI PS-15 Power Supply User Manual
outside the box
AC
Page 22
Chapter 2 Installation and Configuration
Temperature inside the box ....................44.3 °C (in the middle of the right
Temperature outside the box ..................23.3 °C
Temperature rise .....................................21 °C
Cooling
The NI PS-15 is convection cooled, and direct cooling is not required. However, you must not cover the ventilation grid (for example, with cable conduits) by more than 30%.
Proper installation clearance for the NI PS-15 is 40 mm on top, 20 mm on the bottom, 5 mm on the left and right side when loaded permanently with full power. If the adjacent device is a heat source, 15 mm clearance is recommended between the NI PS-15 and the adjacent device.
side of the power supply with a distance of 2 cm)
Hazardous Risks
Cautions Do not use the unit without the proper earth connection (Protective Earth).
Use the PE pin on the front panel terminal block for earth connection instead of one of the screws on the housing.
Turn the power off before working on the power supply. Protect against inadvertent re-powering.
Make sure the wiring is correct by following all local and national codes.
Do not open, modify, or repair the unit.
Use caution to prevent any foreign objects from entering into the housing.
Do not use in wet locations or in areas where moisture or condensation can be expected.
Service Parts
The NI PS-15 power supply does not contain any serviceable parts. If an internal fuse trips, it is caused by an internal defect. If damage or malfunction occurs during operation, immediately turn the power off and send the NI PS-15 to National Instruments for inspection.
Note Attempting to repair or modify the NI PS-15 power supply will void your warranty.
NI PS-15 Power Supply User Manual 2-8 ni.com
Page 23
Peak Current Capability
Solenoids, contactors and pneumatic modules often have a steady state coil and a pick-up coil. The inrush current demand of the pick-up coil is several times higher than the steady state current and usually exceeds the nominal output current (including the PowerBoost). The same situation applies when starting a capacitive load.
Branch circuits are often protected with circuit breakers or fuses. In case of a short or an overload in the branch circuit, the fuse needs a certain amount of over-current to trip or to blow. The peak current capability ensures the safe operation of subsequent circuit breakers.
Assuming the input voltage is turned on before such an event, the built-in large sized output capacitors inside the power supply can deliver extra current. Discharging this capacitor causes a voltage dip on the output. Figure 2-5 illustrates two typical voltage dips.
Chapter 2 Installation and Configuration
24 V
T
10 A
0A
10 ms/DIV
Peak load 10 A (resistive) for 50 ms
Output voltage dips from 24 V to 18.5 V.
Charging Batteries
Output
Voltage
18.5 V
Output
Current
24 V
T
25 A
0 A
1 ms/DIV
Peak load 25 A (resistive) for 5 ms
Output voltage dips from 24 V to 12 V.
Output
Voltage
20 V
Output
Current
Figure 2-5. Peak Current Capacity Output Voltage Dips
The NI PS-15 power supply should not be used to charge batteries.
© National Instruments Corporation 2-9 NI PS-15 Power Supply User Manual
Page 24
Chapter 2 Installation and Configuration
AC
DC
+
Load
+
S1
Circuit
Breaker
I
Power
Supply
Back Feeding Loads
Loads such as decelerating motors and inductors can feed voltage back to the power supply. This feature is also called return voltage immunity or resistance against back-EMF (Electro Magnetic Force). The NI PS-15 power supply is resistant to this and does not malfunction when a load feeds back voltage to the power supply, regardless of whether the power supply itself is on or off.
The maximum allowed feed back voltage is 35 V can be calculated according to the built-in large sized output capacitor, which is specified in the Output section of Appendix A, Specifications.
Output Circuit Breakers
Standard miniature circuit breakers (MCBs) can be used for branch protection. Ensure that the MCB is also rated for DC voltage. The following tests show which circuit breakers the power supply typically trips.
Note Circuit breakers have huge tolerances in their tripping behavior. Therefore, these
typical tests can only be used as a recommendation or for comparing two different power supplies. Furthermore, the loop impedance has a major influence on whether a breaker trips or not.
Test 1: Short circuit with S1 on the power supply end of the cable (loop impedance approximately 20 mΩ). The input voltage was 230 V load current was 0 A.
. The absorbing energy
DC
and the
AC
NI PS-15 Power Supply User Manual 2-10 ni.com
Figure 2-6. Breaker Trip Test 1
Page 25
Chapter 2 Installation and Configuration
Circuit
Breaker
I
Power
Supply
AC
DC
+
Load
+
S1
R
The following circuit breaker tripped during the test:
A or Z-Characteristic ............................. ≤ 8 A
B-Characteristic ..................................... No tripping ≥ 6 A
No breaker available <6 A
C-Characteristic ..................................... ≤ 4 A
Test 2: Short circuit with S1 on the load end (additional impedance is included in the form of longer load wire length). The input voltage was 230 V
and the load current was 0 A.
AC
Figure 2-7. Breaker Trip Test 2
The following circuit breaker tripped during the test:
A or Z-Characteristic ............................. ≤ 6 A and R = 180 mΩ
B-Characteristic ..................................... No tripping ≥ 6 A
No breaker available <6 A
C-Characteristic ..................................... ≤ 3 A and R = 270 mΩ
Table 2-3 provides a comparison of resistances in terms of wire gauge and length.
Table 2-3. Resistances for Wire Gauges and Lengths
Resistance 0.5 mm
2
0.7 mm
2
1.0 mm
2
1.5 mm
2
2.5 mm
2
4.0 mm
180 mΩ 5.0 m 7.0 m 10 m 15 m 25 m 40 m
270 mΩ 7.5 m 10.5 m 15 m 23 m 38 m 60 m
2
© National Instruments Corporation 2-11 NI PS-15 Power Supply User Manual
Page 26
Chapter 2 Installation and Configuration
For example: Which wire gauge must be used to trip a C-Characteristic circuit breaker with a rating of 3 A? The load wire length is 21 m.
Answer: A 3 A C-Characteristic circuit breaker requires a loop impedance of less than 270 mΩ (test results). Table 2-3 shows that up to 23 m of wire with a cross section of 1.5 mm wire smaller than 1.5 mm
Inductive and Capacitive Loads
The NI PS-15 is designed to supply any kind of load, including unlimited capacitive and inductive loads.
2
is below 270 mΩ. You should not use a
2
.
NI PS-15 Power Supply User Manual 2-12 ni.com
Page 27
Specifications
This appendix contains specifications for the NI PS-15 power supply.
Note Specifications are subject to change without notice.
Hazardous Voltages Must be mounted in an enclosure by qualified personnel. Refer to
Figure A-1 for more information.
This power supply is designed for installation in an enclosure and is intended for general use, such as in industrial control, office, communication, and instrumentation equipment. Do not use this device in aircraft, trains and nuclear equipment, where malfunctioning of the power supply may cause severe personal injury or threaten human life.
Dimensions and Weight
A
Dimensions
Width .............................................. 32 mm
Height.............................................. 124 mm
Depth............................................... 117 mm
Weight.................................................... 500 g (1.1 lb)
© National Instruments Corporation A-1 NI PS-15 Power Supply User Manual
Page 28
Appendix A Specifications
DIN-Rail Depth
Depth: 117 mm (4.61 in.)
Height: 124 mm (4.88 in.)
19.14.7
25.43
24–28 V
DC OK
+ + – –
N L
Width: 32 mm
(1.26 in.)
AC 100-120 / 200-240 V
DC 24 V 5 A
1
2
1 Suitably rated NEMA or IP enclosure that requires tool access 2NI PS-15 Power Supply
Figure A-1. NI PS-15 Power Supply
Figure A-2. NI PS-15 Dimensions
NI PS-15 Power Supply User Manual A-2 ni.com
Page 29
Appendix A Specifications
AC Input
AC input Nominal AC 100 –120 V/200 –240 V Auto-select input, TN-, TT-, IT-Mains, refer to Figure A-3
AC input range 90–132 V
180–264 V
85–90 V
264–300 V
AC
AC
AC
AC
100–120 V range, continuous operation
200–240 V range, continuous operation
Short term or with output derating
<0.5 s
Input frequency Nominal 50–60 Hz ±6%
DC input Not allowed
Typ ic al/
Maximum
AC
100 V
AC
120 V
AC
230 V
Input current Typical 2.34 A 2.05 A 1.23 A At 24 V, 5 A refer to Figure A-3
Power factor
Crest factor
*
**
Typical 0.58 0.56 0.47 At 24 V, 5 A refer to Figure A-3
Typical 2.9 3.1 3.7 At 24 V, 5 A
Start-up delay Typical 740 ms 900 ms 720 ms Refer to Figure A-4
Rise time Typical
Typical
8 ms 25 ms
8 ms 25 ms
8 ms 25 ms
0 mF, 24 V, 5 A, refer to Figure A-4 5 mF, 24 V, 5 A, refer to Figure A-4
Turn-on
Maximum 400 mV 400 mV 400 mV Refer to Figure A-4
overshoot
Turn-on voltage Typical 75 V
Shut-down
Typical 55 V
AC
AC
75 V
55 V
AC
AC
Steady-state value, refer to Figure A-3
Steady-state value, refer to Figure A-3
voltage
*
The power factor is the ratio of the true (or real) power to the apparent power in an AC circuit.
**
The crest factor is the mathematical ratio of the peak value to the RMS value of the input current waveform.
© National Instruments Corporation A-3 NI PS-15 Power Supply User Manual
Page 30
Appendix A Specifications
Turn-on
90 V
Rated Input Ranges
V
IN
P
OUT
264 Vac180 V
Shut-down
132 V
No
Harm
From
132 to
180 V
Range
100-
120 V
Range
200-
240 V
Start-up
Delay
Rise
Time
Overshoot
–5%
Output
Voltage
Intput
Voltage
Figure A-3. Input Voltage Range
NI PS-15 Power Supply User Manual A-4 ni.com
Figure A-4. Turn On Behavior, Definitions
Input Current, Typ.
3A
2.5
2
1.5
1
0.5
0
100 Vac
120 Vac
230 Vac
Output Current
Figure A-5. Input Current Vs. Output Load
6A13452.5 3.5 4.5 5.51.5 2
Page 31
Figure A-6. Power Factor Vs. Output Load
0.35
0.4
Output Current
0.45
0.5
0.55
0.6
0.65
Power Factor, Typ.
230 Vac
120 Vac
100 Vac
6 A13452.5 3.5 4.5 5.51.5 2
Input Current
Input Voltage
Output Voltage
A
A:
Input:
Output:
Ambient:
Upper curve:
Medium curve:
Lower curve:
Time scale:
Start-up delay = Inrush delay
230 Vac
24 V, 10 A
25 °C
Input current 10 A / DIV
Input voltage 500 V / DIV
Output voltage 20 V / DIV
100 ms / DIV
Input Current Inrush Surge
Appendix A Specifications
An active inrush limitation circuit limits the input inrush current after input voltage is applied. The charging current into EMI suppression capacitors is disregarded in the first milliseconds after power up.
Typical/
Maximum
Inrush current Maximum
Typica l
Inrush energy Ty pica l 1 A2s 1 A2s 1 A2s –25 °C to +70 °C
AC 100 V AC 120 V AC 230 V
10 A 3 A
peak
peak
10 A 3 A
peak
peak
10 A 3 A
peak
peak
–25 °C to +70 °C –25 °C to +70 °C
© National Instruments Corporation A-5 NI PS-15 Power Supply User Manual
Figure A-7. Input Inrush Current, Typical Behavior
Page 32
Appendix A Specifications
0
25
50
150 ms
85
Hold-up Time
100
75
125
Input Voltage
a) 24 V 2,5 A typ. b) 24 V 2,5 A min. c) 24 V 5 A typ. d) 24 V 5 A min. e) 24 V 6 A min.
a
a
b
c
c d
d e
e
b
180 230 Vac100 120
Hold-up Time
Typical/
Maximum
AC
100V
AC
120V
AC
230V
Hold-up Time Typic al 109 ms 165 ms 161 ms 2.5 A, 24 V, refer to Figure A-8
Typic al 50 ms 80 ms 78 ms 5 A, 24 V, refer to Figure A-8
Typic al 37 ms 62 ms 63 ms 6 A, 24 V, refer to Figure A-8
NI PS-15 Power Supply User Manual A-6 ni.com
Note
At no load, the hold-up time can be up to several seconds. The green DC OK LED
is lit during this time.
Figure A-8. Hold-Up Time Vs. Input Voltage
Intput Voltage
Output Voltage
Figure A-9. Shutdown Behavior, Definitions
Zero Transition
– 5%
Hold-up Time
Page 33
Output
Output Voltage
0
8
4
8
12
28 V
16
20
24
12 A1062
Adjustment
Range
Output Current
04
Output voltage Nominal 24 V
Appendix A Specifications
Adjustment range Minimum
Maximum
24–28 V 30 V
Guaranteed At clockwise end position of potentiometer
Factory setting 24.1 V ±0.2%, at full load, cold unit
Line regulation Maximum 70 mV 90 VAC to 132 VAC or 180 VAC to 264 V
AC
Load regulation Maximum 100 mV Static value, 0A 5A → 0A
Ripple and noise voltage Maximum 50 mVpp 20 Hz to 20 MHz, 50 Ω
Output capacitance Typic al 1,800 µF
Output current Nominal
Nominal Nominal Nominal
Output power Nominal
Nominal
Short-circuit current Minimum
Maximum
*
The NI PS-15 may respond with a thermal shut-down when continuously loaded with more than 120 W and operated with
*
6 A 5 A
5.1 A
4.3 A
144 W 120 W
10 A 14 A
At 24 V, ambient < 45 °C, refer to Figure A-10 At 24 V, ambient < 60 °C, refer to Figure A-10
*
At 28 V, ambient < 45 °C, refer to Figure A-10 At 28 V, ambient < 60 °C, refer to Figure A-10
*
Ambient < 45 °C Ambient < 60 °C
Load impedance 200 mΩ, refer to Figure A-10 Load impedance 200 mΩ, refer to Figure A-10
a MAINS voltage of 100 V or below.
© National Instruments Corporation A-7 NI PS-15 Power Supply User Manual
Figure A-10. Output Voltage Vs. Output Current, Typical
Page 34
Appendix A Specifications
Peak Current Capability
The power supply can deliver a peak current which is higher than the specified short term current. This helps to start current-demanding loads or to safely operate subsequent circuit breakers.
The extra current is supplied by the output capacitors inside the power supply. During this event, the capacitors will be discharged and cause a voltage dip on the output. Detailed curves can be found in the Peak Current
Capability section of Chapter 2, Installation and Configuration.
Peak current voltage dips Typical From 24 V to 18.5 V At 10 A for 50 ms, resistive load
Typical From 24 V to 22 V At 25 A for 2 ms, resistive load
Typical From 24 V to 20 V At 25 A for 5 ms, resistive load
Efficiency and Power Losses
AC
100V
Efficiency Typical 88.8% 89.4% 90.2% 5 A, 24 V
Power losses Typical 1.9 W 2.0 W 1.7 W 0 A
Typical 9.1 W 8.8 W 8.2 W 2.5 A, 24 V
Typical 15.3 W 14.5 W 13.2 W 5 A, 24 V
Typical 19.4 W 18.2 W 16.1 W 6 A, 24 V
AC
120V
AC
230V
NI PS-15 Power Supply User Manual A-8 ni.com
Page 35
Appendix A Specifications
Efficiency vs. output current at 24 V Losses vs. output current at 24 V
Efficiency
1
86
88
90
85
91%
100 Vac
120 Vac
230 Vac
6 A2 3
87
89
Power Losses
0
3
9
15
18 W
120 Vac
100 Vac
230 Vac
Output CurrentOutput Current
6
3012 456
12
Efficiency vs. input voltage, 24 V, 5 A Losses vs. input voltage, 24 V, 5 A
Efficiency
260
Vac
Input Voltage Input Voltage
91%
88.5
89.5
90.5
90.0
89.0
88.0
Power Losses
0
3
6
9
12
15
18 W
260
Vac
54
85 120 155 190 225
85 120 155 190 225
Reliability
The lifetime expectancy shown in Table A-1 indicates the service life of the NI PS-15, and is determined by the lifetime expectancy of the built-in electrolytic capacitors. Lifetime expectancy is specified in operational hours. Lifetime expectancy is calculated according to the capacitor’s manufacturer specification. The prediction model allows a calculation of up to 15 years from the date of shipment.
MTBF (Mean Time Between Failure) is calculated according to statistical device failures and indicates reliability of a device. It is the statistical representation of the likelihood of a unit to fail and does not necessarily represent the life of a product.
© National Instruments Corporation A-9 NI PS-15 Power Supply User Manual
Figure A-11. NI PS-15 Efficiency and Losses
Page 36
Appendix A Specifications
Table A-1. Reliability Specifications
Minimum/ Maximum
Lifetime expectancy Minimum 52,000 hours 58,000 hours 72,000 hours 40 °C, 24 V, 5 A
Minimum 27,000 hours 34,000 hours 42,000 hours 40 °C, 24 V, 6 A
Minimum 135,000 hours 128,000 hours 144,000 hours 40 °C, 24 V, 2.5A
Minimum 142,000 hours 15 years 15 years 25 °C, 24 V, 5 A
MTBF SN 29500, IEC 61709
MTBF MIL HDBK 217F 552,000 hours 546,000 hours 574,000 hours 40 °C, 24 V, 5 A,
638,000 hours 661,000 hours 869,000 hours 40 °C, 24 V, 5 A
AC
100V
542,000 hours 562,000 hours 739,000 hours 40 °C, 24 V, 6 A
1,077,000 hours 1,111,000 hours 1,495,000 hours 25 °C, 24 V, 5 A
497,000 hours 491,000 hours 517,000 hours 40 °C, 24 V, 6 A,
788,000 hours 775,000 hours 800,000 hours 25 °C, 24 V, 5 A,
AC
120V
AC
230V
Ground Benign GB40
Ground Benign GB40
Ground Benign GB25
Dielectric Strength
Notes The output voltage is floating and has no ohmic connection to ground.
To fulfill the PELV requirements according to EN60204-1 § 6.4.1, we recommend that either the + pole, the – pole or any other part of the output circuit should be connected to the protective earth system. This helps to avoid situations in which a load starts unexpectedly or can not be switched off when unnoticed earth faults occur.
NI PS-15 Power Supply User Manual A-10 ni.com
Page 37
Appendix A Specifications
A
N
L
Input
Earth, PE
Output
+
C
B
Figure A-12. Dielectric Strength
Table A-2 lists the tests that have been run to determine the NI PS-15 dielectric strength, and the results of each test.
Table A-2. Dielectric Strength Test Results
Te st Duration A B C
Type tes t 60 s 2500 V
Factory test 5 s 2500 V
Field test 5 s 2000 V
Type tests and factory tests
Conducted by the manufacturer. Do not repeat test in field.
Rules for field test
Use appropriate test equipment which applies the voltage with a slow ramp. Connect L and N together as well as all output poles.
Used Substances
The unit does not release any silicone and is suitable for the use in paint shops.
Electrolytic capacitors included in this unit do not use electrolytes such as Quaternary Ammonium Salt Systems.
Plastic housings and other molded plastic materials are free of halogens, wires and cables are not PVC insulated.
AC
AC
AC
3000 V
2500 V
2000 V
AC
AC
AC
500 V
500 V
500 V
AC
AC
AC
© National Instruments Corporation A-11 NI PS-15 Power Supply User Manual
Page 38
Appendix A Specifications
Environment
The production material within our production does not include following toxic chemicals: Polychlorized Biphenyl (PCB), Polychlorized Terphenyl (PCT), Pentachlorophenol (PCP), Polychlorinated naphthalene (PCN), Polybrom Biphenyl (PBB), Polybrom Bipheny-oxyd (PBO), Polybrominated Diphenylether (PBDE), Polychlorinated Diphenylether (PCDE), Polydibromphenyl Oxyd (PBDO), Cadmium, Asbest, Mercury, Silicia.
Operational temperature –25 °C to +70 °C
(–13 °F to 158 °F)
Output de-rating 1.6 W/°C 45–60 °C (113 °F to 140 °F)
3 W/°C 60–70 °C (140 °F to 158 °F)
Storage temperature –40 to +85 °C (–40 °F to 185 °F) Storage and transportation
Humidity 5 to 95% r.H. IEC 60068-2-30
Vibration sinusoidal 2–17.8 Hz: ±1.6 mm;
17.8–500 Hz: 2 g 2 hours / axis
Shock 30 g 6 ms, 20 g 11 ms
3 bumps/direction, 18 bumps total
Altitude 0 to 6000 m (0 to 20,000 ft) Reduce output power or ambient
Output de-rating (for altitude) 7.5 W/1000 m or 5 °C/1000 m Above 2000 m (6500 ft), refer to
Over-voltage category III EN 50178, altitudes up to 2000 m
Reduce output power according to Figure A-13
Do not energize while condensation is present
IEC 60068-2-6
IEC 60068-2-27
temperature above 2000 m sea level
Figure A-14
Degree of pollution 2 EN 50178, not conductive
NI PS-15 Power Supply User Manual A-12 ni.com
II Altitudes from 2000 m to 6000 m
Page 39
Appendix A Specifications
0
1
2
3
4
5
6 A
–25 0 20 40 60
70 °C
Short Term
Allowable Output Current at 24a V
Ambient Temperature
Continuous
Figure A-13. Output Current Vs. Ambient Temperature
Allowable Output Current at 24 V
6 A
5
4
3
2
A... Tamb < 60 °C
1
B... Tamb < 50 C... Tamb < 40
0
0 2000 4000 6000 m
B
° °
Short Term
C
A
C C
Figure A-14. Output Current Vs. Altitude
Note
The ambient temperature is defined as the temperature 2 cm below the NI PS-15.
Protection Features
Output protection Electronically protected against overload, no-load and short-circuits
Altitude
Output over-voltage protection
Output over-current protection
Degree of protection IP 20 EN/IEC 60529
Penetration protection >3.5 mm From screws, small parts, and so on
© National Instruments Corporation A-13 NI PS-15 Power Supply User Manual
Typical 3 5 V Maximum 39 V
Electronically limited Refer to Figure A-10.
DC
DC
In case of an internal power supply defect, a redundant circuitry limits the maximum output voltage. The output shuts down and automatically attempts to restart.
Page 40
Appendix A Specifications
Over-temperature protection
Input transient protection MOV Metal Oxide Varistor
Internal input fuse T4A H.B.C. Not user replaceable
Ye s Output shut-down with automatic restart
Note In case of a protection event, audible noise may occur.
Safety
This product is designed to meet the requirements of the following standards of safety for industrial control and information technology equipment:
IEC/EN 60950-1, UL 508
Note For UL and other safety certifications, refer to the product label or the Online
Product Certification section.
Electromagnetic Compatibility
This product is designed to meet the requirements of the following standards of EMC for industrial control and information technology equipment:
EMC Immunity EN 61000-6-2
EN 61000-6-1
Electrostatic discharge EN 61000-4-2 Contact discharge
Electromagnetic RF field EN 61000-4-3 80 MHz–1 GHz 10 V/m Criterion A
Fast transients (Burst) EN 61000-4-4 Input lines
Surge voltage on input EN 61000-4-5 L N
Surge voltage on output EN 61000-4-5 +
Conducted disturbance EN 61000-4-6 0.15–80 MHz 10 V Criterion A
Generic standards
Air discharge
Output lines
PE
N/L
PE
+/–
8 kV 15 kV
4 kV 2 kV
2 kV 4 kV
500 V 500 V
Criterion A Criterion A
Criterion A Criterion A
Criterion A Criterion A
Criterion A Criterion A
NI PS-15 Power Supply User Manual A-14 ni.com
Page 41
Appendix A Specifications
Mains voltage dips EN 61000-4-11 0% of 100 V
40% of 100 V
70% of 100 V
0% of 200 V
40% of 200 V
70% of 200 V
Voltage interruptions EN 61000-4-11 0 VAC, 5000 ms Criterion C
Powerful transients VDE 0160 Over entire load range 750 V, 1.3 ms Criterion A
Criterions:
A: Power supply shows normal operation behavior within the defined limits.
C: Temporary loss of function is possible. Power supply might shut-down and restarts by itself. No damages or hazards for
the power supply occur.
AC
AC
AC
AC
AC
AC
0 VAC, 20 ms Criterion A
40 VAC, 200 ms Criterion C
70 VAC, 500 ms Criterion A
0 VAC, 20 ms Criterion A
80 VAC, 200 ms Criterion C
140 VAC, 500 ms Criterion A
Switching Frequency
175 kHz to 225 kHz ........................ input voltage dependent 24 V,
2.5 A
100 kHz to 130 kHz ........................ input voltage dependent 24 V, 5 A
EMC Emission EN 61000-6-4 Generic standards
Conducted emission EN 55011, EN 55022, FCC Part 15,
CISPR 11, CISPR 22
EN 55022 Class A, output lines
Radiated emission EN 55011, EN 55022 Class B
Harmonic input current EN 61000-3-2 >2.7 A output current not fulfilled
Voltage fluctuations, flicker EN 61000-3-3 fulfilled
Class B, input lines
Notes This device complies with FCC Part 15 rules.
Operation is subjected to following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Above an average output current of 2.7 A, the harmonic current standard EN 61000-3-2 is not fulfilled.
© National Instruments Corporation A-15 NI PS-15 Power Supply User Manual
Page 42
Appendix A Specifications
A power supply has to comply with EN 61000-3-2 (Standard for harmonic input current) when:
the end-device is used within the European Union and
the end-device is connected to a public mains supply with a nominal voltage greater than or equal to 220 V
AC
and
the power supply is:
fitted in an end-device with an average input power
in excess of 75 W
or
fitted in an end-device with a continuous input power
in excess of 75 W
or
part of a lighting system.
Exceptions
End-devices for professional applications with an input power > 1000 W do not need to fulfill EN 61000-3-2.
Notes For the standards applied to assess the EMC of this product, refer to the Online
Product Certification section.
For EMC compliance, operate this device with shielded cabling.
CE Compliance
Comments
The average input power must be determined in accordance with EN 61000-3-2.
Industrial MAINS supplies with their own transformer are considered to be non-public.
Where individual self-contained items of equipment are installed in a rack or case (for example, devices connected in parallel), they are regarded as being individually connected to the MAINS supply. The rack or case need not be tested as a whole. Alternatively, it is also permitted to assess the whole rack or case. This is recommended for devices used in professional applications with an input power greater than 1000 W.
This product meets the essential requirements of applicable European Directives as follows:
2006/95/EC; Low-Voltage Directive (safety)
2004/108/EC; Electromagnetic Compatibility Directive (EMC)
NI PS-15 Power Supply User Manual A-16 ni.com
Page 43
Certifications
⬉ᄤֵᙃѻક∵ᶧ᥻ࠊㅵ⧚ࡲ⊩ ˄Ё೑
RoHS
˅
Ё೑ᅶ᠋
National Instruments
ヺড়Ё೑⬉ᄤֵᙃѻકЁ䰤ࠊՓ⫼ᶤѯ᳝ᆇ⠽䋼ᣛҸ
(RoHS)
DŽ
݇Ѣ
National Instruments
Ё೑
RoHS
ড়㾘ᗻֵᙃˈ䇋ⱏᔩ
ni.com/environment/rohs_china
DŽ
(For information about China RoHS compliance, go to
ni.com/environment/rohs_china
.)
18WM
LISTED
IND. CONT. EQ.
Appendix A Specifications
LISTED as Industrial Control Equipment (UL 508)
RECOGNIZED as Information Technology Equipment (UL 60950-1)
GL
Online Product Certification
Refer to the product Declaration of Conformity (DoC) for additional regulatory compliance information. To obtain product certifications and the DoC for this product, visit number or product line, and click the appropriate link in the Certification column.
Environmental Management
NI is committed to designing and manufacturing products in an environmentally responsible manner. NI recognizes that eliminating certain hazardous substances from our products is beneficial to the environment and to NI customers.
For additional environmental information, refer to the NI and the Environment Web page at environmental regulations and directives with which NI complies, as well as other environmental information not included in this document.
GL (Germanischer Lloyd) classified for marine and offshore applications. Environmental category: C, EMC2
ni.com/certification, search by model
ni.com/environment. This page contains the
Waste Electrical and Electronic Equipment (WEEE)
EU Customers At the end of the product life cycle, all products must be sent to a WEEE
recycling center. For more information about WEEE recycling centers, National Instruments WEEE initiatives, and compliance with WEEE Directive 2002/96/EC on Waste and Electronic Equipment, visit
© National Instruments Corporation A-17 NI PS-15 Power Supply User Manual
ni.com/environment/weee.
Page 44
Technical Support and Professional Services
Visit the following sections of the award-winning National Instruments Web site at
ni.com for technical support and professional services:
B
Support—Technical support at
following resources:
Self-Help Technical Resources—For answers and solutions,
visit
ni.com/support for software drivers and updates,
a searchable KnowledgeBase, product manuals, step-by-step troubleshooting wizards, thousands of example programs, tutorials, application notes, instrument drivers, and so on. Registered users also receive access to the NI Discussion Forums
ni.com/forums. NI Applications Engineers make sure every
at question submitted online receives an answer.
Standard Service Program Membership—This program
entitles members to direct access to NI Applications Engineers via phone and email for one-to-one technical support as well as exclusive access to on demand training modules via the Services Resource Center. NI offers complementary membership for a full year after purchase, after which you may renew to continue your benefits.
For information about other technical support options in your area, visit
ni.com/contact.
ni.com/services, or contact your local office at
ni.com/support includes the
Training and Certification—Visit
System Integration—If you have time constraints, limited in-house
© National Instruments Corporation B-1 NI PS-15 Power Supply User Manual
ni.com/training for
self-paced training, eLearning virtual classrooms, interactive CDs, and Certification program information. You also can register for instructor-led, hands-on courses at locations around the world.
technical resources, or other project challenges, National Instruments Alliance Partner members can help. To learn more, call your local NI office or visit
ni.com/alliance.
Page 45
Appendix B Technical Support and Professional Services
Declaration of Conformity (DoC)—A DoC is our claim of compliance with the Council of the European Communities using the manufacturer’s declaration of conformity. This system affords the user protection for electromagnetic compatibility (EMC) and product safety. You can obtain the DoC for your product by visiting
ni.com/certification.
Calibration Certificate—If your product supports calibration, you can obtain the calibration certificate for your product at
ni.com/calibration.
You also can visit the Worldwide Offices section of
ni.com/niglobal
to access the branch office Web sites, which provide up-to-date contact information, support phone numbers, email addresses, and current events.
NI PS-15 Power Supply User Manual B-2 ni.com
Page 46
Index
A
AC input specifications, A-3
B
back feeding loads, 2-10
C
calibration certificate (NI resources), B-2 capacitive loads, 2-12 CE compliance, specifications, A-16 certification specifications, A-17 charging batteries, use, 2-9 circuit breakers, output, 2-10 conventions used in the manual, viii cooling the NI PS-15, 2-8 current capability, peak, 2-9
figure, 2-9
D
environment specifications, A-12 environmental management,
specifications, A-17 examples (NI resources), B-1 external input protection, 2-7
H
hazardous risks, 2-8 help, technical support, B-1 hold-up time specifications, A-6
I
inductive loads, 2-12 input current inrush surge, A-5 input protection, 2-7 installation, configuration, and operation
mounting the NI PS-15, 2-1 unpacking the NI PS-15, 1-1 wiring the NI PS-15, 2-3
instrument drivers (NI resources), B-1
Declaration of Conformity (NI resources), B-2 diagnostic tools (NI resources), B-1 dielectric strength specifications, A-10 dimension specifications, A-1 documentation
conventions used in manual, viii NI resources, B-1 related documentation, ix
drivers (NI resources), B-1
E
efficiency specifications, A-8 electromagnetic compatibility, A-14 enclosure, sealed operation, 2-7
© National Instruments Corporation I-1 NI PS-15 Power Supply User Manual
K
key features, 1-1 kit contents, 1-1 KnowledgeBase, B-1
L
loads
back feeding, 2-10 inductive, capacitive, 2-12
Page 47
Index
M
mounting equipment, 1-5
panel kit, 1-5 side brackets, 1-5
mounting orientation, 2-1
N
National Instruments support and services,
B-1
NI PS-15
cooling, 2-8 external input protection, 2-7 hazardous risks, 2-8 key features, 1-1 mounting equipment, 1-5
panel brackets, 1-5 side brackets, 1-5
operation, 2-5
parallel, 2-6 serial, 2-5
figure, 2-5
two-phase, 2-6 peak current capability, 2-9 service parts, 2-8 specifications, A-1 unpacking, 1-1 wiring terminals, 2-3
NI support and services, B-1
O
operating in a sealed enclosure, 2-7 operation
parallel, 2-6 serial, 2-5
figure, 2-5 two phase, 2-6
figure, 2-6
output circuit breakers, 2-10 output specifications, A-7
P
panel mounting kit, 1-5 parallel operation, 2-6 parts, service, 2-8 peak current capability, 2-9
figure, 2-9
specifications, A-8 power loss specifications, A-8 programming examples (NI resources), B-1 protection features specifications, A-13 PS-15, orientation, 2-1
figure, 2-2
R
related documentation, ix reliability specifications, A-9 risks, 2-8
S
safety, specifications, A-14 sealed enclosure, operation, 2-7 serial operation, 2-5 service parts, 2-8 side mounting kit, 1-5 software (NI resources), B-1 specifications
AC input, A-3
CE compliance, A-16
certifications, A-17
current inrush surge, A-5
dielectric strength, A-10
dimensions, A-1
efficiency, A-8
electromagnetic compatibility, A-14
environment, A-12
environmental management, A-17
hold-up time, A-6
online product certification, A-17
NI PS-15 Power Supply User Manual I-2 ni.com
Page 48
Index
output, A-7 peak current capability, A-8 power loss, A-8 protection features, A-13 reliability, A-9 safety, A-14 substances used, A-11
weight, A-1 support, technical, B-1 surge, input current inrush, A-5
T
technical support, B-1 terminal wiring, 2-3
requirements (table), 2-3 training and certification (NI resources), B-1 troubleshooting (NI resources), B-1 two-phase operation, 2-6
figure, 2-6
U
unpacking the NI PS-15 power supply, 1-1 use for charging batteries, 2-9 used substances specifications, A-11
W
Web resources, B-1 weight specifications, A-1 wiring the terminals, 2-3
requirements (table), 2-3
© National Instruments Corporation I-3 NI PS-15 Power Supply User Manual
Loading...