National Instruments NI SMB-2163 User Manual

USER GUIDE
NI SMB-2163
Single-Ended Digital I/O Accessory
Ρ΅νιϋΠȂ඾ུࢊβȜΐ܄ȃ
The NI SMB-2163 is a breakout box for National Instruments single-ended digital waveform generator/analyzer modules. The NI SMB-2163 provides an easy way to connect to other devices for testing and debugging.
This guide explains how to set up and use the NI SMB-2163 single-ended digital I/O (DIO) accessory with NI single-ended digital waveform generator/analyzer modules.

Contents

Conventions ............................................................................................2
What You Need to Get Started ...............................................................2
Related Documentation........................................................................... 3
Parts Locator ........................................................................................... 3
Installing Cables...................................................................................... 4
Connecting Signals ................................................................................. 6
Cleaning the Accessory...........................................................................8
Specifications .......................................................................................... 9
Technical Support Resources.................................................................. 10

Conventions

The following conventions are used in this manual:
<> Angle brackets that contain numbers separated by an ellipsis represent a
range of values associated with a bit or signal name—for example, DIO <0..3>.
» The » symbol leads you through nested menu items and dialog box options
to a final action. The sequence File»Page Setup»Options directs you to pull down the File menu, select the Page Setup item, and select Options from the last dialog box.
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to avoid injury, data loss, or a system crash. When this symbol is marked on a product, refer to the Specifications section for information about precautions to take.
bold Bold text denotes items that you must select or click in the software, such
as menu items and dialog box options. Bold text also denotes parameter names.
italic Italic text denotes variables, emphasis, a cross reference, or an introduction
to a key concept. This font also denotes text that is a placeholder for a word or value that you must supply.
monospace Text in this font denotes text or characters that you should enter from the
keyboard, sections of code, programming examples, and syntax examples. This font is also used for the proper names of disk drives, paths, directories, programs, subprograms, subroutines, device names, functions, operations, variables, filenames, and extensions.

What You Need to Get Started

To set up and use the NI SMB-2163, you need the following items:
SHC68-C68-D2 cable assembly
NI digital waveform generator/analyzer, installed in a PXI
or CompactPCI chassis
50 Ω cables with SMB connectors
NI SMB-2163 User Guide 2 ni.com
The documentation included with the digital waveform
SMB-2163
High-Speed DIO Accessory
DIO 20
DIO 24
DIO 25 DIO 26 DIO 27 DIO 28
DIO 29 DIO 30
DIO 31
PFI 1 PFI 2
PFI 3
DDC CLK OUT
STROBE DIO 1 DIO 2
DIO 3 DIO 4 DIO 5
DIO 6
DIO 7
DIO 8 DIO 9
DIO 10
DIO 11 DIO 12 DIO 13 DIO 14 DIO 15
DIO 0
DIO 16
DIO 17
DIO 18
DIO 19
DIO 21 DIO 22
DIO 23
generator/analyzer and driver software
(Optional) Rack Mount Kit for CA-1000 (part number 777665-01),
available from National Instruments

Related Documentation

The NI digital waveform generator/analyzer ships with several documents designed to familiarize you with different aspects of the module. The documentation set includes the following pieces:
NI Digital Waveform Generator/Analyzer Getting Started Guide—Read this printed document to set up the digital waveform generator/analyzer and configure it to complete your first acquisition or generation.
NI Digital Waveform Generator/Analyzer Help—This online document provides more in-depth information about the hardware capabilities of the module, theory of operation, programming flow, and software reference.
NI digital waveform generator/analyzer specifications—This printed document provides specifications for your module.

Parts Locator

Refer to Figure 1 to locate connectors on the NI SMB-2163.
3
3
1
2
4
1STROBE 2 DDC CLK OUT 3 DIO <0..31>
4 68-Pin Digital Data & Control (DDC) Connector 5 Programmable Function Interface (PFI) <1..3> 6 Shield Ground Lug

Figure 1. NI SMB-2163 Parts Locator Diagram

© National Instruments Corporation 3 NI SMB-2163 User Guide
5
6

Installing Cables

Complete the following steps to install the NI SHC68-C68-D2 cable and any 50 Ω SMB cables.
Caution Disconnect power from the device, accessory, and any other connected hardware
before connecting the cable to prevent damage to the hardware and personal injury. NI is not liable for damage resulting from improper connections.
1. Install the driver software and digital waveform generator/analyzer
Note Always install the module in a computer or PXI/CompactPCI chassis before
attaching any cables or accessories.
2. Attach either end of an NI SHC68-C68-D2 cable assembly to the DDC
Notes Do not use cables other than the NI SHC68-C68-D2 cable with this accessory.
Do not use the NI SMB-2163 accessory with differential digital signals, such as the digital pattern output of the NI PXI-5421 signal generator.
by following the installation procedure in the NI Digital Waveform Generator/Analyzer Getting Started Guide.
connector of the digital waveform generator/analyzer and secure the cable with the captive screws on the cable connector.
The NI SMB-2163 is not compatible with the NI 6533/6534 products.
3. Attach and secure the other end of the NI SHC68-C68-D2 cable assembly to the DDC connector of the NI SMB-2163 and secure them together with the captive screws on the cable connector.
NI SMB-2163 User Guide 4 ni.com
Figure 2 shows how the NI SHC68-C68-D2 cable connects the NI SMB-2163 and the NI digital waveform generator/analyzer.
NI PXI-1042
NI PXI-6551
50 MHz Digital I/O
ACCESS ACTIVE
CLK IN
PFI 0
CLK OUT
DIGITAL DATA & CONTROL
1
2
3
1 PXI Chassis with NI Digital Waveform Generator/Analyzer 2 NI SMB-2163 3 NI SHC68-C68-D2 Cable Assembly
Figure 2. Connecting a Single-Ended Digital Waveform Generator/Analyzer
to the NI SMB-2163
4. Make signal connections by connecting SMB cables to NI SMB-2163 signal terminals.
Note To ensure a solid ground connection, tighten the SMB connectors by gently
snapping them into place.
© National Instruments Corporation 5 NI SMB-2163 User Guide

Connecting Signals

The NI SMB-2163 provides connectivity to up to 32 single-ended DIO channels, three programmable function interface (PFI) channels, and two clock channels of an NI digital waveform generator/analyzer. You can connect to these signals from the labeled SMB connectors on the NI SMB-2163.
Caution Connections that exceed any of the maximum ratings for the NI SMB-2163
or the NI digital waveform generator/analyzer can damage the device and the computer. Maximum input ratings are provided in the specifications document that shipped with the NI digital waveform generator/analyzer. NI is not liable for any damages resulting from such signal connections.
Figure 3 shows the pinout of the DDC connector of the accessory. Each DIO, PFI, and clock channel is connected to a corresponding pin on the NI SMB-2163.
Table 1 describes the DDC signals in the pinout.
NI SMB-2163 User Guide 6 ni.com
DIO 31
GND
DIO 29
GND
DIO 27
GND
DIO 25
RESERVED
DIO 23
GND
DIO 21
GND
DIO 19
GND
DIO 17
GND
DIO 15
GND
DIO 13
GND
DIO 11
GND
DIO 9
GND
DIO 7
PFI 1
DIO 5
GND
DIO 3
PFI 3
DIO 1
GND
DDC CLK OUT
GND
10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32 33 34
35
1 2 3 4 5 6 7 8 9
36 37 38 39 40 41 42 43 44 45 46 47 48 49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64 65 66 67 68
DIO 30 GND DIO 28 GND DIO 26 GND DIO 24 GND DIO 22 GND DIO 20 GND DIO 18 GND DIO 16 GND DIO 14 RESERVED DIO 12 GND DIO 10 GND DIO 8 GND DIO 6 RESERVED DIO 4 GND DIO 2 PFI 2 DIO 0 GND STROBE GND

Figure 3. DDC Connector Pinout

Note
Refer to the NI Digital Waveform Generator/Analyzer Help at Program Files»National Instruments»NI-HSDIO»Documentation»NI Digital Waveform Generator/Analyzer Help for information about the number of available DIO channels
on your device. DIO <20..31> may not apply to your device.
© National Instruments Corporation 7 NI SMB-2163 User Guide
DDC Pin Signal Description
DIO <0..31> Bidirectional digital data channels 0 though 31.
STROBE External Sample clock source for pattern
DDC CLK OUT Exported Sample clock signal.
PFI <1..3> Programmable function interface channels
GND Ground reference for signals.
RESERVED These channels are reserved for system use.
Note For additional shielding, you can connect the shield ground connector on the
NI SMB-2163 to earth/hard ground. This terminal is connected to the shielded enclosure ground. The shield ground lug is shown in Figure 1.

Cleaning the Accessory

Disconnect all cables to the NI SMB-2163 before cleaning. To remove light dust, use a soft, nonmetallic brush. To remove other contaminants, use alcohol wipes. The unit must be completely dry and free from contaminants before returning to service.

Table 1. DDC Connector Pinout Descriptions

acquisition.
1 through 3.
Do not connect signals to these channels.
NI SMB-2163 User Guide 8 ni.com