The NI SMB-2163 is a breakout box for National Instruments single-ended
digital waveform generator/analyzer modules. The NI SMB-2163 provides
an easy way to connect to other devices for testing and debugging.
This guide explains how to set up and use the NI SMB-2163 single-ended
digital I/O (DIO) accessory with NI single-ended digital waveform
generator/analyzer modules.
Technical Support Resources.................................................................. 10
Conventions
The following conventions are used in this manual:
<>Angle brackets that contain numbers separated by an ellipsis represent a
range of values associated with a bit or signal name—for example,
DIO <0..3>.
»The » symbol leads you through nested menu items and dialog box options
to a final action. The sequence File»Page Setup»Options directs you to
pull down the File menu, select the Page Setup item, and select Options
from the last dialog box.
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to
avoid injury, data loss, or a system crash.When this symbol is marked on a
product, refer to the Specifications section for information about
precautions to take.
boldBold text denotes items that you must select or click in the software, such
as menu items and dialog box options. Bold text also denotes parameter
names.
italicItalic text denotes variables, emphasis, a cross reference, or an introduction
to a key concept. This font also denotes text that is a placeholder for a word
or value that you must supply.
monospaceText in this font denotes text or characters that you should enter from the
keyboard, sections of code, programming examples, and syntax examples.
This font is also used for the proper names of disk drives, paths, directories,
programs, subprograms, subroutines, device names, functions, operations,
variables, filenames, and extensions.
What You Need to Get Started
To set up and use the NI SMB-2163, you need the following items:
❑ SHC68-C68-D2 cable assembly
❑ NI digital waveform generator/analyzer, installed in a PXI
or CompactPCI chassis
❑ 50 Ω cables with SMB connectors
NI SMB-2163 User Guide2ni.com
❑ The documentation included with the digital waveform
SMB-2163
High-Speed DIO Accessory
DIO 20
DIO 24
DIO 25 DIO 26 DIO 27 DIO 28
DIO 29 DIO 30
DIO 31
PFI 1 PFI 2
PFI 3
DDC CLK OUT
STROBEDIO 1 DIO 2
DIO 3 DIO 4 DIO 5
DIO 6
DIO 7
DIO 8 DIO 9
DIO 10
DIO 11 DIO 12 DIO 13 DIO 14 DIO 15
DIO 0
DIO 16
DIO 17
DIO 18
DIO 19
DIO 21 DIO 22
DIO 23
generator/analyzer and driver software
❑ (Optional) Rack Mount Kit for CA-1000 (part number 777665-01),
available from National Instruments
Related Documentation
The NI digital waveform generator/analyzer ships with several documents
designed to familiarize you with different aspects of the module. The
documentation set includes the following pieces:
•NI Digital Waveform Generator/Analyzer Getting Started Guide—Read this printed document to set up the digital waveform
generator/analyzer and configure it to complete your first acquisition
or generation.
•NI Digital Waveform Generator/Analyzer Help—This online
document provides more in-depth information about the hardware
capabilities of the module, theory of operation, programming flow,
and software reference.
•NI digital waveform generator/analyzer specifications—This printed
document provides specifications for your module.
Parts Locator
Refer to Figure 1 to locate connectors on the NI SMB-2163.
3
3
1
2
4
1STROBE
2 DDC CLK OUT
3 DIO <0..31>
4 68-Pin Digital Data & Control (DDC) Connector
5 Programmable Function Interface (PFI) <1..3>
6 Shield Ground Lug
Complete the following steps to install the NI SHC68-C68-D2 cable and
any 50 Ω SMB cables.
Caution Disconnect power from the device, accessory, and any other connected hardware
before connecting the cable to prevent damage to the hardware and personal injury. NI is
not liable for damage resulting from improper connections.
1.Install the driver software and digital waveform generator/analyzer
Note Always install the module in a computer or PXI/CompactPCI chassis before
attaching any cables or accessories.
2.Attach either end of an NI SHC68-C68-D2 cable assembly to the DDC
Notes Do not use cables other than the NI SHC68-C68-D2 cable with this accessory.
Do not use the NI SMB-2163 accessory with differential digital signals, such as the digital
pattern output of the NI PXI-5421 signal generator.
by following the installation procedure in the NI Digital Waveform Generator/Analyzer Getting Started Guide.
connector of the digital waveform generator/analyzer and secure the
cable with the captive screws on the cable connector.
The NI SMB-2163 is not compatible with the NI 6533/6534 products.
3.Attach and secure the other end of the NI SHC68-C68-D2 cable
assembly to the DDC connector of the NI SMB-2163 and secure them
together with the captive screws on the cable connector.
NI SMB-2163 User Guide4ni.com
Figure 2 shows how the NI SHC68-C68-D2 cable connects the
NI SMB-2163 and the NI digital waveform generator/analyzer.
NI PXI-1042
NI PXI-6551
50 MHz Digital I/O
ACCESS ACTIVE
CLK
IN
PFI 0
CLK
OUT
DIGITAL DATA & CONTROL
1
2
3
1 PXI Chassis with NI Digital Waveform Generator/Analyzer
2 NI SMB-2163
3 NI SHC68-C68-D2 Cable Assembly
Figure 2. Connecting a Single-Ended Digital Waveform Generator/Analyzer
to the NI SMB-2163
4.Make signal connections by connecting SMB cables to NI SMB-2163
signal terminals.
Note To ensure a solid ground connection, tighten the SMB connectors by gently
The NI SMB-2163 provides connectivity to up to 32 single-ended DIO
channels, three programmable function interface (PFI) channels, and
two clock channels of an NI digital waveform generator/analyzer. You
can connect to these signals from the labeled SMB connectors on the
NI SMB-2163.
Caution Connections that exceed any of the maximum ratings for the NI SMB-2163
or the NI digital waveform generator/analyzer can damage the device and the computer.
Maximum input ratings are provided in the specifications document that shipped with the
NI digital waveform generator/analyzer. NI is not liable for any damages resulting from
such signal connections.
Figure 3 shows the pinout of the DDC connector of the accessory. Each
DIO, PFI, and clock channel is connected to a corresponding pin on the
NI SMB-2163.
DIO 30
GND
DIO 28
GND
DIO 26
GND
DIO 24
GND
DIO 22
GND
DIO 20
GND
DIO 18
GND
DIO 16
GND
DIO 14
RESERVED
DIO 12
GND
DIO 10
GND
DIO 8
GND
DIO 6
RESERVED
DIO 4
GND
DIO 2
PFI 2
DIO 0
GND
STROBE
GND
Figure 3. DDC Connector Pinout
Note
Refer to the NI Digital Waveform Generator/Analyzer Help at Program
Files»National Instruments»NI-HSDIO»Documentation»NI Digital Waveform
Generator/Analyzer Help for information about the number of available DIO channels
on your device. DIO <20..31> may not apply to your device.
DIO <0..31>Bidirectional digital data channels 0 though 31.
STROBEExternal Sample clock source for pattern
DDC CLK OUTExported Sample clock signal.
PFI <1..3>Programmable function interface channels
GNDGround reference for signals.
RESERVEDThese channels are reserved for system use.
Note For additional shielding, you can connect the shield ground connector on the
NI SMB-2163 to earth/hard ground. This terminal is connected to the shielded enclosure
ground. The shield ground lug is shown in Figure 1.
Cleaning the Accessory
Disconnect all cables to the NI SMB-2163 before cleaning. To remove
light dust, use a soft, nonmetallic brush. To remove other contaminants, use
alcohol wipes. The unit must be completely dry and free from contaminants
before returning to service.
Table 1. DDC Connector Pinout Descriptions
acquisition.
1 through 3.
Do not connect signals to these channels.
NI SMB-2163 User Guide8ni.com