National Instruments NI PXI-4224 User Manual

PXI

NI PXI-4224 User Manual

NI PXI-4224 User Manual
August 2008 373752G-01

Support

Worldwide Technical Support and Product Information

ni.com
11500 North Mopac Expressway Austin, Texas 78759-3504 USA Tel: 512 683 0100

Worldwide Offices

Australia 1800 300 800, Austria 43 662 457990-0, Belgium 32 (0) 2 757 0020, Brazil 55 11 3262 3599, Canada 800 433 3488, China 86 21 5050 9800, Czech Republic 420 224 235 774, Denmark 45 45 76 26 00, Finland 358 (0) 9 725 72511, France 01 57 66 24 24, Germany 49 89 7413130, India 91 80 41190000, Israel 972 3 6393737, Italy 39 02 41309277, Japan 0120-527196, Korea 82 02 3451 3400, Lebanon 961 (0) 1 33 28 28, Malaysia 1800 887710, Mexico 01 800 010 0793, Netherlands 31 (0) 348 433 466, New Zealand 0800 553 322, Norway 47 (0) 66 90 76 60, Poland 48 22 3390150, Portugal 351 210 311 210, Russia 7 495 783 6851, Singapore 1800 226 5886, Slovenia 386 3 425 42 00, South Africa 27 0 11 805 8197, Spain 34 91 640 0085, Sweden 46 (0) 8 587 895 00, Switzerland 41 56 2005151, Taiwan 886 02 2377 2222, Thailand 662 278 6777, Turkey 90 212 279 3031, United Kingdom 44 (0) 1635 523545
For further support information, refer to the Signal Conditioning Technical Support Information document. To comment on National Instruments documentation, refer to the National Instruments Web site at ni.com/info and enter the info code feedback.
© 2004–2008 National Instruments Corporation. All rights reserved.

Important Information

Warranty

The NI PXI-4224 is warranted against defects in materials and workmanship for a period of one year from the date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that proves to be defective during the warranty period. This warranty includes parts and labor.
The media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects in materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace software media that do not execute programming instruc tions if National Instruments receives notice of such defects during the warranty period. National Instruments does not warrant that the operation of the software shall be uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are covered by warranty.
National Instruments believes that the information in this document is accurate. The document has been carefully reviewed for technical accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent editions of this document without prior notice to holders of this edition. The reader should consult National Instruments if errors are suspected. In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.
E
XCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE. CUSTOMERS RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART OF NATIONAL
I
NSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE CUSTOMER. NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR DAMAGES RESULTING
FROM LOSS OF DATA, PROFITS, USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY THEREOF. This limitation of
the liability of National Instruments will apply regardless of the form of action, whether in contract or tort, including negligence. Any action against National Instruments must be brought within one year after the cause of action accrues. National Instruments shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not cover damages, defects, malfunctions, or service failures caused by owner’s failure to follow the National Instruments installation, operation, or maintenance instructions; owner’s modification of the product; owner’s abuse, misuse, or negligent acts; and power failure or surges, fire, flood, accident, actions of third parties, or other events outside reasonable control.

Copyright

Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical, including photocopying, recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National Instruments Corporation.
National Instruments respects the intellectual property of others, and we ask our users to do the same. NI software is protected by copyright and other intellectual property laws. Where NI software may be used to reproduce software or other materials belonging to others, you may use NI software only to reproduce materials that you may reproduce in accordance with the terms of any applicable license or other legal restriction.

Trademarks

National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation. Refer to the Terms of Use section on
ni.com/legal for more information about National Instruments trademarks.
Other product and company names mentioned herein are trademarks or trade names of their respective companies.

Patents

For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software, the
patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents.

WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS

(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING FOR A LEVEL OF RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL IMPLANTS OR AS CRITICAL COMPONENTS IN ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE TO PERFORM CAN REASONABLY BE EXPECTED TO CAUSE SIGNIFICANT INJURY TO A HUMAN.
(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE SOFTWARE PRODUCTS CAN BE IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT LIMITED TO FLUCTUATIONS IN ELECTRICAL POWER SUPPLY, COMPUTER HARDWARE MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE FITNESS, FITNESS OF COMPILERS AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION, INSTALLATION ERRORS, SOFTWARE AND HARDWARE COMPATIBILITY PROBLEMS, MALFUNCTIONS OR FAILURES OF ELECTRONIC MONITORING OR CONTROL DEVICES, TRANSIENT FAILURES OF ELECTRONIC SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR MISUSES, OR ERRORS ON THE PART OF THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS THESE ARE HEREAFTER COLLECTIVELY TERMED “SYSTEM FAILURES”). ANY APPLICATION WHERE A SYSTEM FAILURE WOULD CREATE A RISK OF HARM TO PROPERTY OR PERSONS (INCLUDING THE RISK OF BODILY INJURY AND DEATH) SHOULD NOT BE RELIANT SOLELY UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM FAILURE. TO AVOID DAMAGE, INJURY, OR DEATH, THE USER OR APPLICATION DESIGNER MUST TAKE REASONABLY PRUDENT STEPS TO PROTECT AGAINST SYSTEM FAILURES, INCLUDING BUT NOT LIMITED TO BACK-UP OR SHUT DOWN MECHANISMS. BECAUSE EACH END-USER SYSTEM IS CUSTOMIZED AND DIFFERS FROM NATIONAL INSTRUMENTS' TESTING PLATFORMS AND BECAUSE A USER OR APPLICATION DESIGNER MAY USE NATIONAL INSTRUMENTS PRODUCTS IN COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT EVALUATED OR CONTEMPLATED BY NATIONAL INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS ULTIMATELY RESPONSIBLE FOR VERIFYING AND VALIDATING THE SUITABILITY OF NATIONAL INSTRUMENTS PRODUCTS WHENEVER NATIONAL INSTRUMENTS PRODUCTS ARE INCORPORATED IN A SYSTEM OR APPLICATION, INCLUDING, WITHOUT LIMITATION, THE APPROPRIATE DESIGN, PROCESS AND SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.

Conventions

The following conventions are used in this manual:
<> Angle brackets that contain numbers separated by an ellipsis represent a
range of values associated with a bit or signal name—for example, AO <3. .0>.
» The » symbol leads you through nested menu items and dialog box options
to a final action. The sequence File»Page Setup»Options directs you to pull down the File menu, select the Page Setup item, and select Options from the last dialog box.
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to avoid injury, data loss, or a system crash. When this symbol is marked on the product, refer to the Read Me First: Safety and Radio-Frequency Interference document, shipped with the product, for precautions to take.
When symbol is marked on a product it denotes a warning advising you to take precautions to avoid electrical shock.
When symbol is marked on a product it denotes a component that may be hot. Touching this component may result in bodily injury.
bold Bold text denotes items that you must select or click in the software, such
as menu items and dialog box options. Bold text also denotes parameter names.
italic Italic text denotes variables, emphasis, a cross-reference, hardware labels,
or an introduction to a key concept. Italic text also denotes text that is a placeholder for a word or value that you must supply.
monospace Text in this font denotes text or characters that you should enter from the
keyboard, sections of code, programming examples, and syntax examples. This font is also used for the proper names of disk drives, paths, directories, programs, subprograms, subroutines, device names, functions, operations, variables, filenames and extensions, and code excerpts.

Contents

Chapter 1 About the NI PXI-4224
What You Need to Get Started ......................................................................................1-2
National Instruments Documentation ............................................................................1-3
Installing the Application Software, NI-DAQ, and the DAQ Device ...........................1-3
Installing the NI PXI-4224 ............................................................................................1-4
LED Pattern Descriptions ..............................................................................................1-4
Chapter 2 Connecting Signals
Connecting Signals to the NI PXI-4224 ........................................................................2-1
Front Signal Connector....................................................................................2-1
Analog Input Connections...............................................................................2-3
Floating Signal Source Connection...................................................2-11
Ground-Referenced Signal Connection ............................................2-12
Shielded Ground-Referenced Signal Connection
(Recommended) .............................................................................2-12
Chapter 3 Configuring and Testing
Verifying and Self-Testing the Signals Using Test Panels ............................................3-1
Configuring the NI PXI-4224 in MAX..........................................................................3-2
Creating a Voltage Task or Global Channel Using NI-DAQmx.....................3-2
Verifying and Self-Testing an NI-DAQmx Task or Global Channel.............. 3-3
Chapter 4 Theory of Operation
Theory of Operation.......................................................................................................4-1
Signal Conditioning Functional Overview ......................................................4-3
Measurement Considerations ..........................................................................4-3
Input Impedance................................................................................4-3
Common-Mode Rejection Ratio .......................................................4-4
Effective CMR ..................................................................................4-5
Timing and Control Functional Overview ......................................................4-5
Programmable Function Inputs .......................................................................4-6
Device and PXI Clocks ...................................................................................4-7
© National Instruments Corporation v NI PXI-4224 User Manual
Contents
Chapter 5 Using the NI PXI-4224
Developing Your Application ....................................................................................... 5-1
Typical Program Flow Chart........................................................................... 5-1
Overview of Typical Flow Chart .................................................................... 5-3
Creating a Task Using DAQ Assistant or Programmatically ........... 5-3
Adjusting Timing and Triggering..................................................... 5-3
Configuring Channel Properties ....................................................... 5-4
Acquiring, Analyzing, and Presenting.............................................. 5-5
Completing the Application.............................................................. 5-5
Developing an Application Using LabVIEW ................................................. 5-5
Using a DAQmx Channel Property Node in LabVIEW................... 5-7
Synchronization and Triggering...................................................................... 5-8
Synchronizing the NI PXI-4224 ..................................................................... 5-8
Synchronizing the NI PXI-4224 Using LabVIEW ........................... 5-10
Other Application Documentation and Material ........................................................... 5-11
Calibrating the NI PXI-4224 ......................................................................................... 5-12
Loading Calibration Constants........................................................................ 5-12
Self-Calibration............................................................................................... 5-12
External Calibration ........................................................................................ 5-13
Appendix A Specifications
Appendix B Timing Signal Information
Appendix C Removing the NI PXI-4224
Appendix D Common Questions
Glossary
Index
NI PXI-4224 User Manual vi ni.com
Figures
Contents
Figure 2-1. NI PXI-4224 Front Label ......................................................................2-3
Figure 2-2. Unshielded Floating Signal Source Connection
Using a D-SUB Connector ....................................................................2-4
Figure 2-3. Unshielded Grounded Signal Source Connection
Using a D-SUB Connector ....................................................................2-5
Figure 2-4. Shielded Floating Signal Source Connection
Using a D-SUB Connector ....................................................................2-6
Figure 2-5. Shielded Grounded Signal Source Connection
Using a D-SUB Connector ....................................................................2-7
Figure 2-6. Unshielded Floating Signal Source Connection
Using a Terminal Block ........................................................................2-8
Figure 2-7. Unshielded Grounded Signal Source Connection
Using a Terminal Block ........................................................................2-9
Figure 2-8. Shielded Floating Signal Source Connection
Using a Terminal Block ........................................................................2-10
Figure 2-9. Shielded Grounded Signal Source Connection
Using a Terminal Block ........................................................................2-11
Figure 4-1. Block Diagram of NI PXI-4224 ............................................................4-2
Figure 4-2. Effect of Input Impedance on Signal Measurements ............................4-4
Figure 4-3. AI CONV CLK Signal Routing ............................................................4-6
Figure 4-4. NI PXI-4224 PXI Trigger Bus Signal Connection................................4-8
Figure 5-1. Typical Program Flowchart...................................................................5-2
Figure 5-2. General Synchronizing Flowchart.........................................................5-9
Figure A-1. PXI-4224 Dimensions ...........................................................................A-4
Figure B-1. Typical Posttriggered Sequence ............................................................B-2
Figure B-2. Typical Pretriggered Sequence..............................................................B-2
Figure B-3. AI START TRIG Input Signal Timing .................................................B-3
Figure B-4. AI START TRIG Output Signal Timing...............................................B-3
Figure B-5. AI REF TRIG Input Signal Timing.......................................................B-4
Figure B-6. AI REF TRIG Output Signal Timing ....................................................B-5
Figure B-7. AI SAMP CLK Input Signal Timing ....................................................B-6
Figure B-8. AI SAMP CLK Output Signal Timing..................................................B-6
Figure B-9. AI CONV CLK Input Signal Timing ....................................................B-7
Figure B-10. AI CONV CLK Output Signal Timing .................................................B-8
Figure B-11. AI SAMPLE CLK TIMEBASE Signal Timing....................................B-9
Figure B-12. AI HOLD COMPLETE Signal Timing.................................................B-10
Figure C-1. Injector/Ejector Handle Position Before Device Removal....................C-2
© National Instruments Corporation vii NI PXI-4224 User Manual
Contents
Tables
Table 2-1. NI PXI-4224 25-Pin D-SUB Terminal Pin Assignments ..................... 2-2
Table 4-1. Signal Conditioning Functional Blocks ................................................ 4-3
Table 4-2. PXI Trigger Bus Timing Signals .......................................................... 4-9
Table 5-1. NI-DAQmx Properties .......................................................................... 5-4
Table 5-2. Programming a Task in LabVIEW ....................................................... 5-6
Table 5-3. Synchronizing the NI PXI-4224 Using LabVIEW ............................... 5-10
Table A-1. Maximum Sampling Rates.................................................................... A-1
NI PXI-4224 User Manual viii ni.com
About the NI PXI-4224
This chapter provides an introduction to the NI PXI-4224 device and its installation.
The NI PXI-4224 is part of the NI PXI-4200 series of data acquisition (DAQ) devices with integrated signal conditioning. The PXI-4200 series reduces measurement setup and configuration complexity by integrating signal conditioning and DAQ on the same product.
The NI PXI-4224 is an 8-channel isolated analog input device with a ±10 V input range. It allows isolated analog measurements directly on the PXI platform.
The NI PXI-4224 has the following characteristics:
Each channel has a gain of either 1 or 10.
An isolation rating of 42.4 V
The front connector is a 25-pin D-SUB connector, with 16 pins for analog input.
or 60 VDC, Category I.
peak
1
Signal connections are made through a TB-2725 terminal block that provides connections for all eight analog input channels. You can optionally connect a standard 25-pin D-SUB cable to the device and cable it as needed for your application.
Note Go to ni.com/products to determine if newly developed terminal blocks are
available.
You can configure most settings on a per-channel basis through software. The NI PXI-4224 is configured using Measurement & Automation Explorer (MAX) or through function calls to NI-DAQmx.
Note The NI PXI-4224 is supported in NI-DAQmx only.
© National Instruments Corporation 1-1 NI PXI-4224 User Manual
Chapter 1 About the NI PXI-4224

What You Need to Get Started

To set up and use the NI PXI-4224, you need the following:
Hardware
–NIPXI-4224
One of the following:
TB-2725 terminal block
25-pin D-SUB female connector
PXI or PXI/SCXI combination chassis
Software
NI-DAQ 7.3.1 or later
One of the following:
•LabVIEW
Measurement Studio
LabWindows
Documentation
NI PXI-4224 User Manual
Read Me First: Safety and Radio-Frequency Interference
DAQ Getting Started Guide
PXI or PXI/SCXI combination chassis user manual
Documentation for your software
/CVI
Tools
1/8 in. flathead screwdriver
You can download NI documents from
NI PXI-4224 User Manual 1-2 ni.com
ni.com/manuals.

National Instruments Documentation

The NI PXI-4224 User Manual is one piece of the documentation set for your DAQ system. You could have any of several types of manuals depending on the hardware and software in your system. Use the manuals you have as follows:
DAQ Getting Started Guide—This document describes how to install NI-DAQ devices and NI-DAQ. Install NI-DAQmx before you install the SCXI module.
SCXI Quick Start Guide—This document describes how to set up an SCXI chassis, install SCXI modules and terminal blocks, and configure the SCXI system in MAX.
PXI or PXI/SCXI combination chassis manual—Read this manual for maintenance information about the chassis and for installation instructions.
Accessory installation guides or manuals—If you are using accessory products, read the terminal block installation guides. They explain how to physically connect the relevant pieces of the system.
Software documentation—You may have both application software and NI-DAQmx software documentation. NI application software includes LabVIEW, Measurement Studio, and LabWindows/CVI. After you set up the hardware system, use either your application software documentation or the NI-DAQmx documentation to help you write your application. If you have a large, complicated system, it is worthwhile to look through the software documentation before you configure the hardware.
Chapter 1 About the NI PXI-4224

Installing the Application Software, NI-DAQ, and the DAQ Device

Refer to the DAQ Getting Started Guide, packaged with the NI-DAQ software, for instructions for installing your application software, NI-DAQ driver software, and the DAQ device to which you will connect the NI PXI-4224.
NI-DAQ 7.3.1 or later is required to configure and program the NI PXI-4224 device. If you do not have NI-DAQ 7.3.1 or later, you can either contact an NI sales representative to request it on a CD or download
ni.com.
it from
© National Instruments Corporation 1-3 NI PXI-4224 User Manual
Chapter 1 About the NI PXI-4224

Installing the NI PXI-4224

Note Refer to the Read Me First: Radio-Frequency Interference document before
removing equipment covers or connecting or disconnecting any signal wires.
Refer to the DAQ Getting Started Guide to unpack, install, and configure the NI PXI-4224 in a PXI chassis, and then to the SCXI Quick Start Guide if you are using a PXI/SCXI combination chassis.

LED Pattern Descriptions

The following LEDs on the NI PXI-4224 front panel confirm the system is functioning properly:
The ACCESS LED is normally green and blinks yellow for a minimum of 100 ms during the NI PXI-4224 configuration.
The ACTIVE LED is normally green and blinks yellow for a minimum of 100 ms during data acquisition.
NI PXI-4224 User Manual 1-4 ni.com
Connecting Signals
This chapter provides details about the front signal connector of the NI PXI-4224 and how to connect signals to the NI PXI-4224.

Connecting Signals to the NI PXI-4224

After you have verified that the NI PXI-4224 is installed correctly and self-tested the device, refer to the following sections to connect signals to the device.
Caution Refer to the Read Me First: Safety and Radio-Frequency Interference document
before removing equipment covers, or connecting or disconnecting any signal wires.

Front Signal Connector

The NI PXI-4224 connection interface consists of a 25-pin D-SUB connector and one SMB connector. You can program SMB connector as a PFI 0 line or for external calibration. Table 2-1 shows the signal assignments of the D-SUB connector for the NI PXI-4224. Figure 2-1 shows the front label, with each set of screw terminals labeled according to the corresponding differential input signal for the NI PXI-4224.
2
To connect a signal to the NI PXI-4224, use a TB-2725 terminal block designed specifically for the NI-PXI-4224, or use a 25-pin D-SUB to build a connector to suit your application. Refer to the TB-2725 Terminal Block Installation Guide if you are using the TB-2725 terminal block. Use Table 2-1 to make the signal connections if you are constructing a connector using a 25-pin D-SUB connector.
Connect a timing or triggering signal to the PFI 0/CAL SMB connector using a cable with an SMB signal connector.
© National Instruments Corporation 2-1 NI PXI-4224 User Manual
Chapter 2 Connecting Signals
Caution The PFI 0/CAL SMB connector is for low-voltage timing and calibration signals
only. Voltages greater than ±15 V can damage the device.
If you are building a 25-pin D-SUB connector for your application, make sure you use a connector and wires that are safety rated for the voltage and category of the signals in your application.
Table 2-1. NI PXI-4224 25-Pin D-SUB Terminal Pin Assignments
Front Connector
Diagram
14 15 16 17 18 19 20 21 22 23 24 25
12345678910111213
Pin Number Signal Names Pin Number Signal Names
14 AI 0 – 1 AI 0 +
15 AI 1 – 2 AI 1 +
16 AI 2 – 3 AI 2 +
17 AI 3 – 4 AI 3 +
18 AI 4 – 5 AI 4 +
19 AI 5 – 6 AI 5 +
*
NC—No Connection
20 AI 6 – 7 AI 6 +
21 AI 7 – 8 AI 7 +
22 No Pin 9 No Pin
23 NC
*
10 D GND
24 SPI CLK 11 MISO
25 SELECT 12 MOSI
N/A N/A 13 +5 V
NI PXI-4224 User Manual 2-2 ni.com
Chapter 2 Connecting Signals
NI PXI-4224
8 Chan Isolation Amp
ACCESS ACTIVE
2
PFI 0/
CAL
12345678910111213
14 15 16 17 18 19 20 21 22 23 24 25
1
3
1 ACCESS and ACTIVE LEDs 2 SMB PFI 0/CAL Connector
3 25-Pin D-SUB or TB-2725 Terminal
Block Connector
Figure 2-1. NI PXI-4224 Front Label

Analog Input Connections

The following sections provide a definition of the signal source characteristics, descriptions of various ways to connect signals to the NI PXI-4224, and electrical diagrams showing the signal source and connections. Whenever possible, use shielded twisted-pair field wiring and grounding to reduce the effects of unwanted noise sources.
© National Instruments Corporation 2-3 NI PXI-4224 User Manual
Chapter 2 Connecting Signals
Caution If you are building a 25-pin D-SUB connector for your application, make sure
you use a connector and signal wires that are safety rated for the voltage and category of the signals in your application.
In the electrical diagrams, two different ground symbols are used. These symbols indicate that you cannot assume that the indicated grounds are at the same potential. Refer to Appendix A, Specifications, for maximum working voltage specifications.
You can make signal connections to the NI PXI-4224 through either an NI terminal block, such as the TB-2725, or you can build a connector using a 25-pin D-SUB.
Figures 2-2 through 2-5 illustrate connecting signals using a D-SUB connector.
Signal Source
+
V
SIG
Tw isted-Pair
Wiring
AI 0 +
AI 0 –
AI 7 +
AI 7 –
CH 0
CH 7
Figure 2-2. Unshielded Floating Signal Source Connection Using a D-SUB Connector
NI PXI-4224 User Manual 2-4 ni.com
Chapter 2 Connecting Signals
Signal Source
+
V
SIG
V
Ground
SIG
Reference
Tw isted-Pair
Wiring
AI 0 +
CH 0
AI 0 –
CH 7
AI 7 +
AI 7 –
Figure 2-3. Unshielded Grounded Signal Source Connection Using a D-SUB Connector
© National Instruments Corporation 2-5 NI PXI-4224 User Manual
Chapter 2 Connecting Signals
Signal Source
+
V
SIG
Shielding
Wiring
CH 0
AI 0 +
AI 0 –
CH 7
AI 7 +
AI 7 –
Figure 2-4. Shielded Floating Signal Source Connection Using a D-SUB Connector
Tw isted-Pair
NI PXI-4224 User Manual 2-6 ni.com
Chapter 2 Connecting Signals
Signal Source
+
V
SIG
V
Ground
SIG
Reference
Shielding
Wiring
CH 0
AI 0 +
AI 0 –
CH 7
AI 7 +
AI 7 –
Figure 2-5. Shielded Grounded Signal Source Connection Using a D-SUB Connector
Tw isted-Pair
© National Instruments Corporation 2-7 NI PXI-4224 User Manual
Chapter 2 Connecting Signals
Figures 2-6 through 2-9 illustrate connecting signals using a terminal block.
Signal Source
+
V
SIG
Tw isted-Pair
Wiring
Terminal Block
AI 0 +
AI 0 –
AI 7 +
AI 7 –
CH 0
CH 7
Figure 2-6. Unshielded Floating Signal Source Connection Using a Terminal Block
NI PXI-4224 User Manual 2-8 ni.com
Chapter 2 Connecting Signals
Signal Source
+
V
SIG
V
Ground
SIG
Reference
Tw isted-Pair
Wiring
Terminal Block
AI 0 +
CH 0
AI 0 –
CH 7
AI 7 +
AI 7 –
Figure 2-7. Unshielded Grounded Signal Source Connection Using a Terminal Block
© National Instruments Corporation 2-9 NI PXI-4224 User Manual
Chapter 2 Connecting Signals
Signal Source
+
V
SIG
Tw isted-Pair
Wiring
Shielding
Terminal Block
CH 0
AI 0 +
AI 0 –
CH 7
AI 7 +
AI 7 –
Figure 2-8. Shielded Floating Signal Source Connection Using a Terminal Block
NI PXI-4224 User Manual 2-10 ni.com
Chapter 2 Connecting Signals
Signal Source
+
V
SIG
V
Ground
SIG
Reference
Tw isted-Pair
Wiring
Shielding
Terminal Block
CH 0
AI 0 +
AI 0 –
CH 7
AI 7 +
AI 7 –
Figure 2-9. Shielded Grounded Signal Source Connection Using a Terminal Block
Floating Signal Source Connection
Figures 2-2, 2-4, 2-6, and 2-8 illustrate floating signal source connections. In this configuration, the signal source being measured is a floating signal source, such as a battery. A floating signal source is not connected in any way to the building ground system.
To connect a floating signal source connection to the NI PXI-4224, the signal (V signal reference (V
© National Instruments Corporation 2-11 NI PXI-4224 User Manual
+) is connected to the NI PXI-4224 channel (AI X +). The
SIG
–) is connected to the channel reference (AI X –).
SIG
Chapter 2 Connecting Signals
Ground-Referenced Signal Connection
Figures 2-3, 2-5, 2-7, and 2-9 illustrate the ground-referenced signal connection. In this configuration, the voltage source being measured is referenced to its own ground reference that is connected through a conductive path to the instrument ground reference. For example, the path can be through a common earth ground or through the power line ground.
To connect a ground-reference signal source to the NI PXI-4224, the signal
+) is connected to the NI PXI-4224 channel (AI X +). The signal
(V
SIG
reference (V
–) is connected to the channel reference (AI X –).
SIG
Shielded Ground-Referenced Signal Connection (Recommended)
Figures 2-5 and 2-9 illustrate shielded ground-referenced signal connections. The connection to this signal source is identical to the ground-referenced signal connection with the addition of shielding around the field wiring. The shielding is grounded at the signal source ground (V
Ground Reference). Connect the signal (V
SIG
NI PXI-4224 channel (AI X +). Connect the signal reference (V channel reference (AI X –).
+) to the
SIG
–) to the
SIG
This shielding scheme is effective at reducing capacitive or electrically coupled noise. The same concerns regarding the difference in ground potentials, discussed in the Ground-Referenced Signal Connection section, also apply to this configuration.
For more information about the function of the NI PXI-4224 and other measurement considerations, refer to Chapter 4, Theory of Operation.
NI PXI-4224 User Manual 2-12 ni.com
3
Configuring and Testing
This chapter provides details about configuring and testing the NI PXI-4224 in MAX, including how to use device test panels and create and configure NI-DAQmx Tasks and NI-DAQmx Global Channels.

Verifying and Self-Testing the Signals Using Test Panels

After you have successfully installed the NI PXI-4224, verified the installation, and connected the signals, use the NI PXI-4224 device test panels to verify the device is measuring signals properly.
The test panels allow you to measure the signal connected to the NI PXI-4224 directly as well as configure some of the properties of your measurement. To open the NI PXI-4224 device test panels when in MAX, complete the following steps:
1. Expand Devices and Interfaces to display the list of devices and
interfaces.
2. Expand NI-DAQmx Devices to display the list of NI-DAQmx devices.
3. Click PXI-4224.
4. Click the Test Panels button in the device toolbar.
5. Configure the settings on the screen, and click Start to take a
measurement.
To measure scaled voltages, further configure channel properties, and configure timing settings, use an NI-DAQmx Task or NI-DAQmx Global Channel.
© National Instruments Corporation 3-1 NI PXI-4224 User Manual
Chapter 3 Configuring and Testing

Configuring the NI PXI-4224 in MAX

This section describes how to create NI-DAQmx Tasks and NI-DAQmx Global Channels in MAX that allow you to take measurements with the NI PXI-4224.

Creating a Voltage Task or Global Channel Using NI-DAQmx

An NI-DAQmx Global Channel gives a physical channel a name and provides scaling. An NI-DAQmx Task is a collection of channels with timing and triggering configured. To create a new NI-DAQmx Task or NI-DAQmx Global Channel, complete the following steps:
1. Double-click the Measurement & Automation Explorer icon on the desktop.
2. Right-click Data Neighborhood and select Create New.
3. Select NI-DAQmx Task or NI-DAQmx Global Channel and click Next.
4. Select Analog Input and select Voltage.
5. If you are creating a channel, you can select only one channel. If you are creating a task, select the channels to add to the task. You can select a range of channels by holding down the <Shift> key while selecting the channels. You can select multiple individual channels by holding down the <Ctrl> key while selecting channels. Click Next.
6. Enter the name of the task or channel, and click Finish.
7. Select the channel(s) you want to configure for input voltage range. While making the selections you can select blocks of channels by pressing the <Shift> key or individual channels by pressing the <Ctrl> key.
8. Under the Settings tab, set the input range by entering the Min and Max values.
9. Click the Device tab and select the Autozero mode.
10. Repeat steps 7 through 9 until you have configured all the channels.
Note For more information about how to further configure the NI PXI-4224, or how to use
LabVIEW to configure the device and take measurements, refer to Chapter 4, Theory of
Operation.
NI PXI-4224 User Manual 3-2 ni.com
Loading...
+ 59 hidden pages