This document contains step-by-step instructions for calibrating
National Instruments 6711/6713/6731/6733 for PCI/PXI/CompactPCI
analog output (AO) devices. Use this calibration procedure in conjunction
with the
for calibrating NI 6711/6713/6731/6733 devices.
ni671xCal.dll
file, which contains specific functions required
Refer to
Note
ni671xCal.dll
ni.com/support/calibrat/mancal.htm
file.
What Is Calibration?
Calibration consists of verifying the measurement accuracy of a device
and adjusting for any measurement error. Verification is measuring the
performance of the device and comparing these measurements to the
factory specifications. During calibration, you supply and read voltage
levels using external standards, then you adjust the module calibration
constants. The new calibration constants are stored in the EEPROM.
The calibration constants are loaded from memory as needed to adjust
for the error in the measurements taken by the device.
Why Should You Calibrate?
The accuracy of electronic components drifts with time and temperature,
which can affect measurement accuracy as the device ages. Calibration
restores these components to their specified accuracy and ensures that the
device still meets NI standards.
for a copy of the
CVI™, LabVIEW™, National Instruments™, NI™, ni.com™, and NI-DAQ™ are trademarks of National Instruments
Corporation. Product and company names mentioned herein are trademarks or trade names of their respective
companies. For patents covering National Instruments products, refer to the appropriate location: Help»Patents in your
software, the patents.txt file on your CD, or ni.com/patents.
The measurement requirements of your application determine how often
the NI 6711/6713/6731/6733 must be calibrated to maintain accuracy.
NI recommends that you perform a complete calibration at least once every
year. You can shorten this interval to 90 days or six months based on the
demands of your application.
Calibration Options: External Versus Internal
The NI 6711/6713/6731/6733 has two calibration options: an internal,
or self-calibration, and an external calibration.
Internal Calibration
Internal calibration is a much simpler calibration method that does not rely
on external standards. In this method, the device calibration constants are
adjusted with respect to a high-precision voltage source on the
NI 6711/6713/6731/6733. This type of calibration is used after the device
has been calibrated with respect to an external standard. However, external
variables such as temperature can still affect measurements. The new
calibration constants are defined with respect to the calibration constants
created during an external calibration, ensuring that the measurements can
be traced back to the external standards. In essence, internal calibration is
similar to the auto-zero function found on a digital multimeter (DMM).
External Calibration
External calibration requires using a high-precision DMM. During external
calibration, the DMM supplies and reads voltages from the device.
Adjustments are made to the device calibration constants to ensure that the
reported voltages fall within the device specifications. The new calibration
constants are then stored in the device EEPROM. After the onboard
calibration constants have been adjusted, the high-precision voltage source
on the device is adjusted. An external calibration provides a set of
calibration constants that you can use to compensate for the error in the
measurements taken by the NI 6711/6713/6731/6733.
NI 6711/6713/6731/6733 Calibration Procedure2ni.com
Equipment and Other Test Requirements
This section describes the equipment, test conditions, documentation,
and software you need to calibrate the NI 6711/6713/6731/6733.
Test Equipment
To calibrate the NI 6711/6713/6731/6733, you need a high-precision
DMM that is at least 10 ppm (0.001%) accurate. NI recommends that you
use the Agilent 3458A DMM for calibration.
If you do not have an Agilent 3458A DMM, use the accuracy specifications
to select a substitute calibration standard.
If you do not have custom connection hardware, you may need a connector
block such as the NI CB-68 and a cable such as the SH6868-D1. These
components give you easy access to the individual pins on the 68-pin
I/O connector.
Test Conditions
Follow these guidelines to optimize connections and test conditions during
calibration:
•Keep connections to the NI 6711/6713/6731/6733 short. Long cables
and wires act as antennae, picking up extra noise, which can affect
measurements.
•Use shielded copper wire for all cable connections to the device.
•Use twisted-pair wire to eliminate noise and thermal offsets.
•Maintain a temperature between 18 and 28 °C. To operate the module
at a specific temperature outside this range, calibrate the device at that
temperature.
•Keep relative humidity below 80%.
•Allow a warm-up time of at least 15 minutes to ensure that the
measurement circuitry is at a stable operating temperature.
Software
Because the NI 6711/6713/6731/6733 is a PC-based measurement device,
you must have the proper device driver installed in the calibration system
before attempting calibration. For this calibration procedure, you need
NI-DAQ version 6.9.2 or earlier installed on the calibration computer.
NI-DAQ, which configures and controls the NI 6711/6713/6731/6733,
is available at
NI-DAQ supports a number of programming languages, including
LabVIEW, LabWindows/CVI, Microsoft Visual C++, Microsoft Visual
Basic and Borland C++. When you install the driver, you only need to
install support for the programming language that you intend to use.
You also need copies of the
ni671xCal.h
The DLL provides calibration functionality that does not reside in
NI-DAQ, including the ability to protect the calibration constants, update
the calibration date, and write to the factory calibration area. You can
access the functions in this DLL through any 32-bit compiler. The factory
calibration area and the calibration date should only be modified by a
metrology laboratory or another facility that maintains traceable standards.
files.
ni671xCal.dll,ni671xCal.lib
,and
Configuring the NI 6711/6713/6731/6733
The NI 6711/6713/6731/6733 must be configured in NI-DAQ, which
automatically detects the device. The following steps briefly explain how
to configure the device in NI-DAQ. Refer to the NI 671X/673X UserManual for detailed installation instructions. You can install this manual
when you install NI-DAQ.
1.Power down the computer.
2.Install the NI 6711/6713/6731/6733 in an available slot.
3.Power on the computer.
4.Launch Measurement & Automation Explorer (MAX).
5.Configure the NI 6711/6713/6731/6733 device number.
6.Click Test Resources to ensure that the NI 6711/6713/6731/6733 is
properly working.
The NI 6711/6713/6731/6733 is now configured.
Note
After a device is configured in MAX, the device is assigned a device number, which
is used in each of the function calls to identify which DAQ device to calibrate.
Writing the Calibration Procedure
The calibration procedure in the Calibrating the NI 6711/6713/6731/6733
section provides step-by-step instructions on calling the appropriate
calibration functions. These calibration functions are C function calls from
NI-DAQ that are also valid for Microsoft Visual Basic and Microsoft
Visual C++ programs. Although LabVIEW VIs are not discussed in this
procedure, you can program in LabVIEW using the VIs that have similar
names to the NI-DAQ function calls in this procedure. Refer to the
NI 6711/6713/6731/6733 Calibration Procedure4ni.com
Loading...
+ 9 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.