The NI 5911 is warranted against defects in materials and workmanship for a period of one year from the date of shipment, as evidenced by
receipts or other documentation. National Instruments will, at its option, repair or replace equipment that proves to be defective during the
warranty period. This warranty includes parts and labor.
The media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects
in materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other documentation. National
Instruments will, at its option, repair or replace software media that do not execute programming instructions if National Instruments receives
notice of such defects during the warranty period. National Instruments does not warrant that the operation of the software shall be
uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before
any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are
covered by warranty.
National Instruments believes that the information in this document is accurate. The document has been carefully reviewed for technical
accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent
editions of this document without prior notice to holders of this edition. The reader should consult National Instruments if errors are suspected.
In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.
XCEPT AS SPECIFIED HEREIN,NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF
E
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE
NATIONAL INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE CUSTOMER.NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR
DAMAGES RESULTING FROM LOSS OF DATA
. This limitation of the liability of National Instruments will apply regardless ofthe form of action,whether in contract ortort, including
THEREOF
negligence. Any action against National Instruments must be brought within one year after the cause of action accrues. National Instruments
shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not cover
damages, defects, malfunctions, or service failures caused by owner’s failure to follow the National Instruments installation, operation, or
maintenance instructions; owner’s modification of the product; owner’s abuse, misuse, or negligent acts; and power failure or surges, fire,
flood, accident, actions of third parties, or other events outside reasonable control.
, PROFITS, USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY
Copyright
Under the copyright laws, this publication may not be reproduced ortransmitted in any form, electronic or mechanical, including photocopying,
recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National
Instruments Corporation.
Trademarks
CVI™,FLEXADC™,LabVIEW™, National Instruments™,NI™, and ni.com™are trademarks of National Instruments Corporation.
Product and company names mentioned herein are trademarks or trade names of their respective companies.
.CUSTOMER’S RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART OF
WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS
(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING FOR A LEVEL OF
RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL IMPLANTS OR AS CRITICAL COMPONENTS IN
ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE TO PERFORM CAN REASONABLY BE EXPECTED TO CAUSE SIGNIFICANT
INJURY TO A HUMAN.
(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE SOFTWARE PRODUCTS CAN BE
IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT LIMITED TO FLUCTUATIONS IN ELECTRICAL POWER SUPPLY,
COMPUTER HARDWARE MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE FITNESS, FITNESS OF COMPILERS
AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION, INSTALLATION ERRORS, SOFTWARE AND
HARDWARE COMPATIBILITY PROBLEMS, MALFUNCTIONS OR FAILURES OF ELECTRONIC MONITORING OR CONTROL
DEVICES, TRANSIENT FAILURES OF ELECTRONIC SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR
MISUSES, OR ERRORS ON THE PART OF THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS THESE ARE
HEREAFTER COLLECTIVELY TERMED “SYSTEM FAILURES”). ANY APPLICATION WHERE A SYSTEM FAILURE WOULD
CREATE A RISK OF HARM TO PROPERTY OR PERSONS (INCLUDING THE RISK OF BODILY INJURY AND DEATH) SHOULD
NOT BE RELIANT SOLELY UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM FAILURE. TO AVOID
DAMAGE, INJURY, OR DEATH, THE USER OR APPLICATION DESIGNER MUST TAKE REASONABLY PRUDENT STEPS TO
PROTECT AGAINST SYSTEM FAILURES, INCLUDING BUT NOT LIMITED TO BACK-UP OR SHUT DOWN MECHANISMS.
BECAUSE EACH END-USER SYSTEM IS CUSTOMIZED AND DIFFERS FROM NATIONAL INSTRUMENTS' TESTING
PLATFORMS AND BECAUSE A USER OR APPLICATION DESIGNER MAY USE NATIONAL INSTRUMENTS PRODUCTS IN
COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT EVALUATED OR CONTEMPLATED BY NATIONAL
INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS ULTIMATELY RESPONSIBLE FOR VERIFYING AND VALIDATING
THE SUITABILITY OF NATIONAL INSTRUMENTS PRODUCTS WHENEVER NATIONAL INSTRUMENTS PRODUCTS ARE
INCORPORATED IN A SYSTEM OR APPLICATION, INCLUDING, WITHOUT LIMITATION, THE APPROPRIATE DESIGN,
PROCESS AND SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.
Compliance
FCC/Canada Radio Frequency Interference Compliance*
Determining FCC Class
The Federal Communications Commission (FCC) has rules to protect wireless communications from interference. The FCC
places digital electronics into two classes. These classes are known as Class A (for use in industrial-commercial locations only)
or Class B (for use in residential or commercial locations). Depending on where it is operated, this product could be subject to
restrictions in the FCC rules. (In Canada, the Department of Communications (DOC), of Industry Canada, regulates wireless
interference in much the same way.)
Digital electronics emit weak signals during normal operation that can affect radio, television, or other wireless products. By
examining the product you purchased, you can determine the FCC Class and therefore which of the two FCC/DOC Warnings
apply in the following sections. (Some products may not be labeled at all for FCC; if so, the reader should then assume these are
Class A devices.)
FCC Class A products only display a simple warning statement of one paragraph in length regarding interference and undesired
operation. Most of our products are FCC Class A. The FCC rules have restrictions regarding the locations where FCC Class A
products can be operated.
FCC Class B products display either a FCC ID code, starting with the letters EXN,
or the FCC Class B compliance mark that appears as shown here on the right.
Consult the FCC web site
http://www.fcc.gov
FCC/DOC Warnings
This equipment generates and uses radio frequency energyand, if not installed and used in strict accordance with the instructions
in this manual and the CE Mark Declaration of Conformity**, may cause interference to radio and television reception.
Classification requirements are the same for the Federal Communications Commission (FCC) and the Canadian Department
of Communications (DOC).
Changes or modifications not expressly approved by National Instruments could void the user’s authority to operate the
equipment under the FCC Rules.
for more information.
Class A
Federal Communications Commission
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC
Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated
in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and
used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this
equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct
the interference at his own expense.
Canadian Department of Communications
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.
Class B
Federal Communications Commission
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the
FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not
occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can
be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of
the following measures:
• Reorient or relocate the receiving antenna.
• Increase the separation between the equipment and receiver.
• Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
Canadian Department of Communications
This Class B digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe B respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.
Compliance to EU Directives
Readers in the European Union (EU) must refer to the Manufacturer's Declaration of Conformity (DoC) for information**
pertaining to the CE Mark compliance scheme. The Manufacturer includes a DoC for most every hardware product except for
those bought for OEMs, if also available from an original manufacturer that also markets in the EU, or where compliance is not
required as for electrically benign apparatus or cables.
To obtain the DoC for this product, click Declaration of Conformity at
by product family. Select the appropriate product family, followed by your product, and a link to the DoC appears in Adobe
Acrobat format. Click the Acrobat icon to download or read the DoC.
* Certain exemptions may apply in the USA, see FCC Rules §15.103 Exempted devices,and§15.105(c). Also available in
sections of CFR 47.
** The CE Mark Declaration of Conformity will contain important supplementary information and instructions for the user or
installer.
ni.com/hardref.nsf/
. This website lists the DoCs
Conventions
The following conventions are used in this manual:
»The » symbol leads you through nested menu items and dialog box options
to a final action. The sequence File»Page Setup»Options directs you to
pull down the File menu, select the Page Setup item, and select Options
from the last dialog box.
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to
avoid injury, data loss, or a system crash.
boldBold text denotes items that you must select or click on in the software,
such as menu items and dialog box options. Bold text also denotes
parameter names.
italicItalic text denotes variables, emphasis, a cross reference, or an introduction
to a key concept. This font also denotes text that is a placeholder for a word
or value that you must supply.
monospace
Text in this font denotes text or characters that you should enter from the
keyboard, sections of code, programming examples, and syntax examples.
Text in this font is also used for proper names of functions or variables.
Contents
Chapter 1
Taking Measurements with the NI 5911
Installing the NI 5911 ....................................................................................................1-1
Thank you for buying a National Instruments (NI) 5911 digitizer, featuring
the FLEX ADC. This chapter provides information on installing,
connecting signals to, and acquiring data from your NI 5911.
Installing the NI 5911
There are two main steps involved in installation:
1.Install the NI-SCOPE driver software. You use this driver to write
programs to control your NI 5911 in different application development
environments (ADEs). Installing NI-SCOPE also allows you to
interactively control your NI 5911 with the Scope Soft Front Panel.
2.Install your NI 5911. For step-by-step instructions for installing
NI-SCOPE and the NI 5911, see Where to Start with Your NI Digitizer.
For multiple-board considerations, see the Operating Environment section
in Appendix A, Specifications, of this manual.
1
Connecting Signals
Figure 1-1 shows the front panel for the NI 5911. The front panel contains
three connectors—a BNC connector, an SMB connector, and a 9-pin mini
circular DIN connector (see Figure 1-2).
The BNC connector is for attaching the analog input signal you wish to
measure. The BNC connector is analog input channel 0. To minimize noise,
do not allow the shell of the BNC cable to touch or lie near the metal of the
computer chassis. The SMB connector is for external triggers and for
generating a probe compensation signal. The SMB connector is PFI1.
The DIN connector gives you access to an additional external trigger line.
The DIN connector can be used to access PFI2.
Note
The +5 V signal is fused at 1.1 A. However, NI recommends limiting the current
from this pin to 30 mA. The fuse is self-resetting.
You can acquire data either programmatically—by writing an application
foryourNI5911—or interactively with the Scope Soft Front Panel.
Programmatically Controlling Your NI 5911
To help you get started programming your NI 5911, NI-SCOPE comes
with examples that you can use or modify.
You can find examples for these different ADEs:
•LabVIEW—Go to
LabVIEW\Examples\Instr\niScopeExamples\
•LabWindows/CVI, C, and Visual Basic with Windows 98/95—Go to
vxipnp\win95\Niscope\Examples\c\
•LabWindows/CVI, C, and Visual Basic with Windows 2000/NT—Go
to
vxipnp\winnt\Niscope\Examples\
For information about using NI-SCOPE to programmatically control your
digitizer, refer to your NI-SCOPE Software User Manual. Other resources
include the NI-SCOPE Instrument Driver Quick Reference Guide.It
contains abbreviated information on the most commonly used functions
and LabVIEW VIs. For more detailed function reference help, see the
NI-SCOPE Function Reference Help file, located at Start»Programs»
National Instruments»NI-SCOPE. For more detailed VI help,use LabVIEW context-sensitive help (Help»Show Context Help)ortheNI-SCOPE VI Reference Help, located at Start»Programs»National
Instruments»NI-SCOPE.
Program Files\National Instruments\
Interactively Controlling Your NI 5911 with the Scope Soft Front Panel
The Scope Soft Front Panel allows you to interactively control your
NI 5911 as you would a desktop oscilloscope. To launch the Scope Soft
Front Panel select Start»Programs»National Instruments»NI-SCOPE»NI-SCOPE Soft Front Panel. Refer to the Scope Soft Front Panel Help
file for instructions on configuring the Scope Soft Front Panel for your
specific application.
Note
Press F1 with the Scope Soft Front Panel running to access the Scope Soft Front
This chapter includes an overview of the NI 5911, explains the operation of
each functional unit making up your NI 5911, and describes the signal
connections. Figure 2-1 shows a block diagram of the NI 5911.
2
Analog Input
Connector
Protect/
Calibration
Calibration
Generator
Digital IO
Connector
Mux
AC/DC Coupling
PGIA
Noise
Shaper
Timing IO/
Memory Control
Digital Signal
Processor
Figure 2-1. NI 5911 Block Diagram
A/D Converter
100 MHz, 8-Bit
Capture
Memory
Reference
Clock
Data
Differential Programmable Gain Input Amplifier (PGIA)
The analog input of the NI 5911 is equipped with a differential
programmable gain input amplifier. The PGIA accurately interfaces to and
scales the signal presented to the analog-to-digital converter (ADC)
regardless of source impedance, source amplitude, DC biasing, or
common-mode noise voltages.
When measuring high dynamic range signals, ground noise is often a
problem. The PGIA of the NI 5911 allows you to make noise-free signal
measurements. The PGIA differential amplifier efficiently rejects any
noise present on the ground signal. Internal to the PGIA, the signal
presented at the negative input is subtracted from the signal presented at the
positive input. As shown in Figure 2-2, this subtraction removes ground
noise from the signal. The inner conductor of the BNC is V+; the outer shell
is V–.
Input Signal
V+
V–
Ground Noise
Figure 2-2. Noise-Free Measurements of Signal
+
PGIA
–
V
out
Grounding Considerations
The path for the positive signal has been optimized for speed and linearity.
You should always apply signals to the positive input and ground to the
negative input. Reversing the inputs will result in higher distortion and
lower bandwidth.
The negative input of the amplifier is grounded to PC ground through a
10 kΩ resistor. The PGIA is therefore referenced to ground, so it is not
necessary to make any external ground connections. If the device you
connect to the NI 5911 is already connected to ground, ground-loop noise
voltages may be induced into your system. Notice that in most of these
situations, the 10 kΩ resistance to PC ground is normally much higher than
the cable impedances you use. As a result, most of the noise voltage occurs
at the negative input of the PGIA where it is rejected, rather than in the
positive input, where it would be amplified.
NI 5911 User Manual2-2ni.com
Input Ranges
Chapter 2Hardware Overview
To optimize the ADC resolution, you can select different gains for the
PGIA. In this way, you can scale your input signal to match the full input
range of the converter. The NI 5911 PGIA offers seven different input
ranges, from ±0.1 V to ±10 V, as shown in Table 2-1.
Table 2-1. Input Ranges for the NI 5911
RangeInput Protection Threshold
±10 V±10 V
±5 V±5 V
±2 V±5 V
±1 V±5 V
±0.5 V±5 V
±0.2 V±5 V
±0.1 V±5 V
Note
If you try to acquire a signal below the set input range the sensitive front-end
components of the NI 5911 may become unstable and begin returning invalid data. To
return the digitizer to a stable configuration, switch to the maximum input range setting and
acquire an AC-coupled or 0 V signal.
Input Impedance
The input impedance of the NI 5911 PGIA is 1 MΩ between the positive
and negative input, ±2% depending on input capacitance. The output
impedance of the device connected to the NI 5911 and the input impedance
of the NI 5911 form an impedance divider, which attenuates the input
signal according to the following formula:
VsR
--------------------=
RsRin+
in
where V
is the measured voltage, Vsis the source voltage, Rsis the external
m
source impedance, and R
V
m
is the input impedance.
in
If the device you are measuring has a very large output impedance, your
measurements will be affected by this impedance divider. For example,
if the device has 1 MΩ output impedance, your measured signal will be
one-half the actual signal value.
Input Bias
The inputs of the PGIA typically draw an input bias current of 1 nA at
25 °C. Attaching a device with a very high source impedance can cause
an offset voltage to be added to the signal you measure, according to
the formula R
example, if the device you have attached to the NI 5911 has an output
impedance of 10 kΩ, typically the offset voltage is 10 µV (10 kΩ ×1nA).
The NI 5911 features input-protection circuits that protect both the positive
and negative analog input from damage from AC and DC signals up to
±42 V.
×1nA,whereR
s
is the external source impedance. For
s
If the voltage at one of these inputs exceeds a threshold voltage, V
input clamps to V
minimize input currents to a nonharmful level.
The protection voltage, V
and a resistance of 100 kΩ is inserted in the path to
tr
, is input range dependent, as shown in Table 2-1.
tr
AC Coupling
When you need to measure a small AC signal on top of a large DC
component, you can use AC coupling. AC coupling rejects any
DC component in your signal before it enters into the PGIA. Activating
AC coupling inserts a capacitor in series with the input impedance. Input
coupling can be selected via software. See the Digitizer Basics appendix in
your NI-SCOPE Software User Manual for more information on input
coupling.
Oscilloscope and Flexible Resolution Modes
In oscilloscope mode, the NI 5911 works as a conventional desktop
oscilloscope, acquiring data at 100 MS/s with a vertical resolution of 8 bits.
This mode is useful for displaying waveforms and for deriving waveform
parameters such as slew rate, rise time, and settling time.
Flexible resolution differs from oscilloscope mode in two ways: it has
higher resolution (sampling rate dependent), and the signal bandwidth is
limited to provide antialiasing protection. This mode is useful for spectral
,the
tr
NI 5911 User Manual2-4ni.com
analysis, distortion analysis, and other measurements for which high
resolution is crucial.
Oscilloscope Mode
The ADC converts at a constant rate of 100 MS/s, but you can choose to
store only a fraction of these samples into memory at a lower rate. This
allows you to store waveforms using fewer data points and decreases the
burden of storing, analyzing, and displaying the waveforms. If you need
faster sampling rates, you can use Random Interleaved Sampling (RIS) to
effectively increase the sampling rate to 1 GS/s for repetitive waveforms.
In oscilloscope mode, all signals up to 100 MHz are passed to the ADC.
You need to ensure that your signal is band-limited to prevent aliasing.
Aliasing and other sampling terms are described more thoroughly in your
NI-SCOPE Software User Manual.
Sampling Methods—Real-Time and RIS
There are two sampling methods available in oscilloscope mode, real-time
and random interleaved sampling (RIS). Using real-time sampling, you can
acquire data at a rate of 100/n MS/s, where n is a number from 1 to 2
RIS sampling can be used on repetitive signals to effectively extend the
sampling rate above 100 MS/s. In RIS mode, you can sample at rates of
100 MS/s × n, where n is a number from 2 to 10.
Chapter 2Hardware Overview
32
.
Flexible Resolution Mode
Table 2-2 shows the relationship between the available sampling rates,
resolution, and the corresponding bandwidth for flexible resolution mode.
Table 2-2. Available Sampling Rates and Corresponding Bandwidth
in Flexible Resolution Mode
Sampling RateResolutionBandwidth
12.5 MS/s11 Bits3.75 MHz
5MS/s14 Bits2MHz
2.5 MS/s15.5 Bits1MHz
1MS/s17.5 Bits400 kHz
500 kS/s18 Bits200 kHz
200 kS/s18.5 Bits80 kHz
100 kS/s19 Bits40 kHz
50 kS/s19.5 Bits20 kHz
20 kS/s20.5 Bits8kHz
10 kS/s21 Bits4kHz
Like any other type of converter that uses noise shaping to enhance
resolution, the frequency response of the converter is only flat to its
maximum useful bandwidth. The NI 5911 has a bandwidth of 4 MHz.
Beyond this frequency, there is a span where the converter acts resonant
and where a signal is amplified before being converted. These signals are
attenuated in the subsequent digital filter to prevent aliasing. However,
if the applied signal contains major signal components in this frequency
range, such as harmonics or noise, the converter may overload and signal
data will be invalid. In this case, you will receive an overload warning. You
must then either select a higher input range or attenuate the signal.
How Flexible Resolution Works
The ADC can be sourced through a noise shaping circuit that moves
quantization noise on the output of the ADC from lower frequencies to
higher frequencies. A digital lowpass filter applied to the data removes all
but a fraction of the original shaped quantization noise. The signal is then
resampled to a lower sampling frequency and a higher resolution. Flexible
resolution provides antialiasing protection due to the digital lowpass filter.
NI 5911 User Manual2-6ni.com
Calibration
The NI 5911 can be calibrated for very high accuracy and resolution due
to an advanced calibration scheme. There are two different types of
calibration: internal, or self, calibration and external calibration. A third
option, internal restore, restores factory settings and should be used only in
the event of a self-calibration failure.
Internal calibration is performed via a software command that
compensates for drifts caused by environmental temperature changes. You
can internally calibrate your NI 5911 without any external equipment
connected. External calibration recalibrates the device when the specified
calibration interval has expired. See Appendix A, Specifications,forthe
calibration interval. External calibration requires you to connect an external
precision voltage reference to the device.
Internally Calibrating the NI 5911
Internally calibrate your NI 5911 with a software function or a
LabVIEW VI. See Chapter 3, Common Functions and Examples,ofyour
NI-SCOPE Software User Manual for step-by-step instructions for
calibrating your digitizer.
Chapter 2Hardware Overview
When Internal Calibration Is Needed
To provide the maximum accuracy independent of temperature changes,
the NI 5911 contains a heater that stabilizes the temperature of the most
sensitive circuitries on the board. However, the heater can accommodate
for temperature changes over a fixed range of ±5 °C. When temperatures
exceed this range, the heater no longer is able to stabilize the temperature,
and signal data becomes inaccurate. When the temperature range has been
exceeded, you receive a warning, and you need to perform an internal
calibration.
What Internal Calibration Does
Internal calibration performs the following operations:
•The heater is set to regulate over a range of temperatures centered at
the current environmental temperature. Thecircuit components require
a certain amount of time to stabilize at the new temperature. This
temperature stabilization accounts for the majority of the calibration
time.
•Gain and offset are calibrated for each individual input range.
•The linearity of the ADC is calibrated using an internal sinewave
generator as reference.
•The time-to-digital converter used for RIS measurements is calibrated.
Caution
internal calibration. For optimal calibration performance, disconnect the input signal from
the NI 5911.
Do not apply high-amplitude or high-frequency signals to the NI 5911 during
External Calibration
Why Errors Occur During Acquisition
The NI 5911 uses a heater circuit to maintain constant temperature on the
critical circuitry used in flexible resolution mode. If this circuit is unable to
maintain the temperature within specification, an error is generated. This
error indicates that the temperature of the ADC is out of range and should
be recalibrated by performing an internal calibration. During acquisition in
flexible resolution mode, an error will be generated if the input to the ADC
goes out of range for the converter. The fact that this condition has occurred
may not be obvious from inspecting the data due to the digital filtering that
takes place on the acquired data. Therefore, an error occurs to let you know
that the data includes some samples that were out of the range of the
converter and may be inaccurate.
External calibration calibrates the internal reference on the NI 5911.
The NI 5911 is already calibrated when it is shipped from the factory.
Periodically, the NI 5911 will need external calibration to remain within
the specified accuracy. For more information on calibration, contact NI, or
visit
ni.com/calibration
Appendix A, Specifications.
. For actual intervals and accuracy, refer to
Triggering and Arming
There are several triggering methods for the NI 5911. The trigger can be an
analog level that is compared to the input or any of several digital inputs.
You can also call a software function to trigger the board. Figure 2-3 shows
the different trigger sources. When you use a digital signal, that signal must
be at a high TTL level for at least 40 ns before any triggers will be accepted.
Note
The NI 5911 does not support delayed triggering.
NI 5911 User Manual2-8ni.com
Chapter 2Hardware Overview
Analog
Input
Gain
ATC_OUT
RTSI <0..6>
PFI1, PFI2
High
Level
Low
Level
a. Analog Trigger Circuit
Software
b. Trigger and Arm Sources
Figure 2-3.
+
COMP
COMP
–
7
2
Trigger Sources
Analog
Trigger
Circuit
Trigger
Arm
ATC_OUT
Analog Trigger Circuit
The analog trigger on the NI 5911 operates by comparing the current
analog input to an onboard threshold voltage. This threshold voltage is the
trigger value, and can be set within the current input range in 170 steps.
This means that for a ±10 V input range, the trigger can be set in increments
of 20 V/170 = 118 mV. There may also be a hysteresis value associated
with the trigger that can be set in the same size increments. The hysteresis
value creates a trigger window the signal must pass through before the
trigger is accepted. You can generate triggers on a rising or falling edge
condition. For a more complete discussion of triggering, see Chapter 3,
Common Functions and Examples, of your NI-SCOPE Software User
Manual.
Time to acquire posttrigger samples is (posttrigger samples)/(sample rate
(megahertz)).
Trigger hold-off is the minimum length of time (in seconds) from an
accepted trigger to the start of the next record. In other words, when a
trigger is accepted, the trigger counter is loaded with the desired hold-off
time. After completing its current record, the digitizer records no data and
accepts no triggers until the hold-off counter runs out. When the counter
runs out, the next record begins and a trigger may be accepted. Setting a
hold-off time shorter than posttrigger acquisition time has no effect, as
triggers are always rejected during an acquisition.
Trigger hold-off is provided in hardware using a 32-bit counter clocked
by a 25 MHz internal timebase. With this configuration, you can select
a hardware hold-off value of 40 ns to 171.8 s in increments of 40 ns. For
more information regarding trigger hold-off, see the Common TriggerParameters section in Chapter 3, Common Functions and Examples,of
your NI-SCOPE Software User Manual.
Memory
The NI 5911 allocates at least 4 kB of onboard memory for every
acquisition. Samples are stored in this buffer before transfer to the host
computer. Thus the minimum size for a buffer in the onboard memory is
approximately 4,000 8-bit oscilloscope mode samples or 1,000 32-bit
flexible resolution mode samples. Software allows you to specify buffers
of less than these minimum sizes. However, the minimum number of points
are still acquired into onboard memory, but only the specified number of
points are transferred into the memory of the host computer.
The total number of samples that can be stored depends on the size of the
acquisition memory module installed on the NI 5911 and the size of each
acquired sample.
Triggering and Memory Usage
During the acquisition, samples are stored in a circular buffer that is
continually rewritten until a trigger is received. After the trigger is
received, the NI 5911 continues to acquire posttrigger samples if you have
specified a posttrigger sample count. The acquired samples are placed into
onboard memory. The number of posttrigger or pretrigger samples is only
limited by the amount of onboard memory.
NI 5911 User Manual2-10ni.com
Multiple-Record Acquisitions
After the trigger has been received and the posttrigger samples have been
stored, the NI 5911 can be configured to begin another acquisition that is
stored in another onboard memory record. This is a multiple-record
acquisition. To perform multiple-record acquisitions, configure the
NI 5911 to the number of records you want to acquire before starting the
acquisition. The NI 5911 acquires an additional record each time a trigger
is accepted until all the requested records are stored in memory. You may
acquire up to 1024 records if your NI 5911 is equipped with 4 MB of
onboard memory, or 4096 records with 16 MB. Software intervention after
the initial setup is not required.
Multiple-record acquisitions can quickly acquire numerous triggered
waveforms because they allow hardware rearming of the digitizer before
the data is fetched. Therefore the dead time, or the time when the digitizer
is not ready for a trigger, is extremely small.
For more information on multiple-record acquisitions and dead time, see
the Making a Multiple-Record Acquisition section in Chapter 5, Tasks andExamples, of your NI-SCOPE Software User Manual.
Chapter 2Hardware Overview
RTSI Bus and Clock PFI
The RTSI bus allows NI digitizers to synchronize timing and triggering on
multiple devices. The RTSI bus has seven bidirectional trigger lines and
one bidirectional clock signal.
You can program any of the seven trigger lines to provide or accept a
synchronous trigger signal. You can also use any of the RTSI trigger lines
to provide a synchronization pulse from a master device if you are
synchronizing multiple NI 5911s.
You can use the RTSI bus clock line to provide or accept a 10 MHz
reference clock to synchronize multiple NI 5911 devices.
PFI Lines
The NI 5911 has two digital lines that can accept a trigger, accept or
generate a reference clock, or output a 1 kHz square wave. The function of
each PFI line is independent. However, only one trigger source can be
accepted during acquisition.
You can select PFI1 or PFI2 as inputs for a trigger or a reference clock.
Please see the Synchronization section below for more information about
the use of reference clocks in the NI 5911.
PFI Lines as Outputs
You can select PFI1 or PFI2 to output several digital signals.
Reference Clock is a 10 MHz clock that is synchronous to the 100 MHz
sample clock on the NI 5911. You can use the reference clock to
synchronize to another NI 5911 configured as a slave device or to other
equipment that can accept a 10 MHz reference.
Frequency Output is a 1 kHz digital pulse train signal with a 50% duty
cycle. The most common application of Frequency Output for the NI 5911
is to provide a signal for compensating a passive probe.
The NI 5911 uses a digital phase locked loop to synchronize the 100 MHz
sample clock to a 10 MHz reference. This reference frequency can be
supplied by an internal crystal oscillator or through an external frequency
input through the RTSI bus clock line or a PFI input.
The NI 5911 may also output its 10 MHz reference on the RTSI bus clock
line or a PFI line so that other NI 5911s or other equipment can be
synchronized to the same reference.
While the reference clock input is sufficient to synchronize the 100 MHz
sample clocks, it is also necessary to synchronize clock dividers on each
NI 5911 so that internal clock divisors are synchronized on each different
device. These lower frequencies are important because they are used to
determine trigger times and sample position.
To synchronize the NI 5911 clock dividers, you must connect the digitizers
with an NI RTSI bus cable. One of the RTSI bus triggers must be
designated as a synchronization line. This line will be an output from the
master device and an input on the slave device. To synchronize the
digitizers, a single pulse is sent from the master NI 5911 to the slaves. This
pulse supplies the slave devices with a reference time to clear their clock
dividers. Hardware arming cannot be used during an acquisition using
multiple devices. For more information about synchronization, refer to
your NI-SCOPE Software User Manual.
NI 5911 User Manual2-12ni.com
Specifications
This appendix lists the specifications of the NI 5911. These specifications
are typical at 25 °C unless otherwise stated.
Multiple NI 5911s in the same computer may raise operating temperatures beyond
specification and give rise to imprecise data. NI strongly recommends leaving an empty
PCI slot between multiple NI 5911s or adding a fan.
Ambient temperature ..............................5 to 40 °C
Relative humidity ...................................10% to 90%, noncondensing
Storage Environment
Ambient temperature ..............................–20 to 65 °C
EMC Compliance
CE2001, FCC
Calibration
Internal....................................................Internal calibration is done upon
software command. The
calibration involves gain, offset
and linearity correction for all
input ranges and input modes.
Interval.............................................1 week, or any time temperature
changes beyond ±5 °C. Hardware
detects temperature variations
beyond calibration limits, which
can also be queried by software.
National Instruments Web support is your first stop for help in solving
installation, configuration, and application problems and questions. Online
problem-solving and diagnostic resources include frequently asked
questions, knowledge bases, product-specific troubleshooting wizards,
manuals, drivers, software updates, and more. Web support is available
through the Technical Support section of
NI Developer Zone
ni.com
B
.
The NI Developer Zone at
building measurement and automation systems. At the NI Developer Zone,
you can easily access the latest example programs, system configurators,
tutorials, technical news, as well as a community of developers ready to
share their own techniques.
Customer Education
National Instruments provides a number of alternatives to satisfy your
training needs, from self-paced tutorials, videos, and interactive CDs to
instructor-led hands-on courses at locations around the world. Visit the
Customer Education section of
syllabi, training centers, and class registration.
System Integration
If you have time constraints, limited in-house technical resources, or other
dilemmas, you may prefer to employ consulting or system integration
services. You can rely on the expertise available through our worldwide
network of Alliance Program members. To find out more about our
Alliance system integration solutions, visit the System Integration section
of
National Instruments has offices located around the world to help address
your support needs. You can access our branch office Web sites from the
Worldwide Offices section of
up-to-date contact information, support phone numbers, e-mail addresses,
and current events.
If you have searched the technical support resources on our Web site and
still cannot find the answers you need, contact your local office or National
Instruments corporate. Phone numbers for our worldwide offices are listed
at the front of this manual.
noisean undesirable electrical signal—noise comes from external sources such
as the AC power line, motors, generators, transformers, fluorescent lights,
soldering irons, CRT displays, computers, electrical storms, welders, radio
transmitters, and internal sources such as semiconductors, resistors, and
capacitors. Noise corrupts signals you are trying to send or receive.
Nyquist frequencya frequency that is one-half the sampling rate. See also Nyquist Sampling
Theorem.
Nyquist Sampling
Theorem
the theorem states that if a continuous bandwidth-limited analog signal
contains no frequency components higher than half the frequency at which
it is sampled, then the original signal can be recovered without distortion.
O
Ohm’sLaw(R=V/I)—the relationship of voltage to current in a resistance
overrangea segment of the input range of an instrument outside of the normal
measuring range. Measurements can still be made, usually with a
degradation in specifications.
oversamplingsampling at a rate greater than the Nyquist frequency
P
passbandthe frequency range that a filter passes without attenuation
PCIPeripheral Component Interconnect—a high-performance expansion bus
architecture originally developed by Intel to replace ISA and EISA; it is
achieving widespread acceptance as a standard for PCs and workstations
and offers a theoretical maximum transfer rate of 132 Mbytes/s
peak valuethe absolute maximum or minimum amplitude of a signal (AC + DC)
posttriggeringthe technique to acquire a programmed number of samples after trigger
conditions are met
NI 5911 User ManualG-6ni.com
Glossary
pretriggeringthe technique used on a device to keep a buffer filled with data, so that when
the trigger conditions are met, the sample includes the data leading up
to the trigger condition
PXIPCI eXtensions for Instrumentation. PXI is an open specification that
builds off the CompactPCI specification by adding
instrumentation-specific features.
R
Rresistor
RAMrandom-access memory
real-time samplingsampling that occurs immediately
random interleaved
sampling
resolutionthe smallest signal increment that can be detected by a measurement
rmsroot mean square—a measure of signal amplitude; the square root of the
ROMread-only memory
RTSI busreal-time system integration bus—the National Instruments timing bus that
method of increasing the sample rate by repetitively sampling a repeated
waveform
system. Resolution can be expressed in bits or in digits. The number of bits
in a system is roughly equal to 3.3 times the number of digits.
average value of the square of the instantaneous signal amplitude
connects devices directly, by means of connectors on top of the boards, for
precise synchronization of functions
S
sseconds
Ssamples
S/ssamples per second—used to express the rate at which an instrument
samples an analog signal. 100 MS/s would equal 100 million samples each
second.
sensein four-wire resistance the sense measures the voltage across the resistor
being excited by the excitation current
settling timethe amount of time required for a voltage to reach its final value within
specified limits
source impedancea parameter of signal sources that reflects current-driving ability of voltage
sources (lower is better) and the voltage-driving ability of current sources
(higher is better)
system noisea measure of the amount of noise seen by an analog circuit or an ADC when
the analog inputs are grounded
T
temperature
coefficient
thermal driftmeasurements that change as the temperature varies
thermal EMFsthermal electromotive forces—voltages generated at the junctions of
thermoelectric
potentials
transfer ratethe rate, measured in bytes/s, at which data is moved from source to
triggerany event that causes or starts some form of data capture
the percentage that a measurement will vary according to temperature.
See also thermal drift
dissimilar metals that are functions of temperature. Also called
thermoelectric potentials.
see thermal EMFs
destination after software initialization and set up operations; the maximum
rate at which the hardware can operate
U
undersamplingsampling at a rate lower than the Nyquist frequency—can cause aliasing
update ratethe number of output updates per second
NI 5911 User ManualG-8ni.com
V
Vvolts
VACvolts alternating current
VDCvolts direct current
Glossary
V
error
voltage error
VIvirtual instrument—(1) a combination of hardware and/or software
elements, typically used with a PC, that has the functionality of a classic
stand-alone instrument (2) a LabVIEW software module (VI), which
consists of a front panel user interface and a block diagram program
V
rms
volts, root mean square value
W
waveform shapethe shape the magnitude of a signal creates over time
working voltagethe highest voltage that should be applied to a product in normal use,
normally well under the breakdown voltage for safety margin
filtering, A-4
acquisition modes specifications, A-7
acquisition system specifications, A-1
analog trigger circuit, 2-9
arming. See triggering and arming
B
bias, input, 2-4
block diagram of NI 5911, 2-1
BNC connector, 1-1