National Instruments NI 5911 User Manual

Computer-Based Instruments
NI 5911 User Manual
High-Speed Digitizer with FLEX ADC
NI 5911 User Manual
June 2001 Edition
Part Number 322150D-01

Worldwide Technical Support and Product Information

ni.com

National Instruments Corporate Headquarters

11500 North Mopac Expressway Austin, Texas 78759-3504 USA Tel: 512 794 0100

Worldwide Offices

Australia 03 9879 5166, Austria 0662 45 79 90 0, Belgium 02 757 00 20, Brazil 011 284 5011, Canada (Calgary) 403 274 9391, Canada (Montreal) 514 288 5722, Canada (Ottawa) 613 233 5949, Canada (Québec) 514 694 8521, Canada (Toronto) 905 785 0085, China (Shanghai) 021 6555 7838, China (ShenZhen) 0755 3904939, Denmark 45 76 26 00, Finland 09 725 725 11, France 01 48 14 24 24, Germany 089 741 31 30, Greece 30 1 42 96 427, Hong Kong 2645 3186, India 91805275406, Israel 03 6120092, Italy 02 413091, Japan 03 5472 2970, Korea 02 596 7456, Malaysia 603 9596711, Mexico 5 280 7625, Netherlands 0348 433466, New Zealand 09 914 0488, Norway 32 27 73 00, Poland 0 22 528 94 06, Portugal 351 1 726 9011, Singapore 2265886, Spain 91 640 0085, Sweden 08 587 895 00, Switzerland 056 200 51 51, Taiwan 02 2528 7227, United Kingdom 01635 523545
For further support information, see the Technical Support Resources appendix. To comment on the documentation, send e-mail to techpubs@ni.com.
Copyright © 1998, 2001 National Instruments Corporation. All rights reserved.

Important Information

Warranty

The NI 5911 is warranted against defects in materials and workmanship for a period of one year from the date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that proves to be defective during the warranty period. This warranty includes parts and labor.
The media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects in materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace software media that do not execute programming instructions if National Instruments receives notice of such defects during the warranty period. National Instruments does not warrant that the operation of the software shall be uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are covered by warranty.
National Instruments believes that the information in this document is accurate. The document has been carefully reviewed for technical accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent editions of this document without prior notice to holders of this edition. The reader should consult National Instruments if errors are suspected. In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.
XCEPT AS SPECIFIED HEREIN,NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS ANY WARRANTY OF
E
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE
NATIONAL INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE CUSTOMER.NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR
DAMAGES RESULTING FROM LOSS OF DATA
. This limitation of the liability of National Instruments will apply regardless ofthe form of action,whether in contract ortort, including
THEREOF
negligence. Any action against National Instruments must be brought within one year after the cause of action accrues. National Instruments shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not cover damages, defects, malfunctions, or service failures caused by owner’s failure to follow the National Instruments installation, operation, or maintenance instructions; owners modification of the product; owners abuse, misuse, or negligent acts; and power failure or surges, fire, flood, accident, actions of third parties, or other events outside reasonable control.
, PROFITS, USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSSIBILITY

Copyright

Under the copyright laws, this publication may not be reproduced ortransmitted in any form, electronic or mechanical, including photocopying, recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National Instruments Corporation.

Trademarks

CVI™,FLEXADC™,LabVIEW™, National Instruments™,NI™, and ni.com™are trademarks of National Instruments Corporation.
Product and company names mentioned herein are trademarks or trade names of their respective companies.
.CUSTOMERS RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART OF

WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS

(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING FOR A LEVEL OF RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL IMPLANTS OR AS CRITICAL COMPONENTS IN ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE TO PERFORM CAN REASONABLY BE EXPECTED TO CAUSE SIGNIFICANT INJURY TO A HUMAN.
(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE SOFTWARE PRODUCTS CAN BE IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT LIMITED TO FLUCTUATIONS IN ELECTRICAL POWER SUPPLY, COMPUTER HARDWARE MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE FITNESS, FITNESS OF COMPILERS AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION, INSTALLATION ERRORS, SOFTWARE AND HARDWARE COMPATIBILITY PROBLEMS, MALFUNCTIONS OR FAILURES OF ELECTRONIC MONITORING OR CONTROL DEVICES, TRANSIENT FAILURES OF ELECTRONIC SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR MISUSES, OR ERRORS ON THE PART OF THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS THESE ARE HEREAFTER COLLECTIVELY TERMED SYSTEM FAILURES). ANY APPLICATION WHERE A SYSTEM FAILURE WOULD CREATE A RISK OF HARM TO PROPERTY OR PERSONS (INCLUDING THE RISK OF BODILY INJURY AND DEATH) SHOULD NOT BE RELIANT SOLELY UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM FAILURE. TO AVOID DAMAGE, INJURY, OR DEATH, THE USER OR APPLICATION DESIGNER MUST TAKE REASONABLY PRUDENT STEPS TO PROTECT AGAINST SYSTEM FAILURES, INCLUDING BUT NOT LIMITED TO BACK-UP OR SHUT DOWN MECHANISMS. BECAUSE EACH END-USER SYSTEM IS CUSTOMIZED AND DIFFERS FROM NATIONAL INSTRUMENTS' TESTING PLATFORMS AND BECAUSE A USER OR APPLICATION DESIGNER MAY USE NATIONAL INSTRUMENTS PRODUCTS IN COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT EVALUATED OR CONTEMPLATED BY NATIONAL INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS ULTIMATELY RESPONSIBLE FOR VERIFYING AND VALIDATING THE SUITABILITY OF NATIONAL INSTRUMENTS PRODUCTS WHENEVER NATIONAL INSTRUMENTS PRODUCTS ARE INCORPORATED IN A SYSTEM OR APPLICATION, INCLUDING, WITHOUT LIMITATION, THE APPROPRIATE DESIGN, PROCESS AND SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.

Compliance

FCC/Canada Radio Frequency Interference Compliance*
Determining FCC Class
The Federal Communications Commission (FCC) has rules to protect wireless communications from interference. The FCC places digital electronics into two classes. These classes are known as Class A (for use in industrial-commercial locations only) or Class B (for use in residential or commercial locations). Depending on where it is operated, this product could be subject to restrictions in the FCC rules. (In Canada, the Department of Communications (DOC), of Industry Canada, regulates wireless interference in much the same way.)
Digital electronics emit weak signals during normal operation that can affect radio, television, or other wireless products. By examining the product you purchased, you can determine the FCC Class and therefore which of the two FCC/DOC Warnings apply in the following sections. (Some products may not be labeled at all for FCC; if so, the reader should then assume these are Class A devices.)
FCC Class A products only display a simple warning statement of one paragraph in length regarding interference and undesired operation. Most of our products are FCC Class A. The FCC rules have restrictions regarding the locations where FCC Class A products can be operated.
FCC Class B products display either a FCC ID code, starting with the letters EXN, or the FCC Class B compliance mark that appears as shown here on the right.
Consult the FCC web site
http://www.fcc.gov
FCC/DOC Warnings
This equipment generates and uses radio frequency energyand, if not installed and used in strict accordance with the instructions in this manual and the CE Mark Declaration of Conformity**, may cause interference to radio and television reception. Classification requirements are the same for the Federal Communications Commission (FCC) and the Canadian Department of Communications (DOC).
Changes or modifications not expressly approved by National Instruments could void the users authority to operate the equipment under the FCC Rules.
for more information.
Class A
Federal Communications Commission
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Canadian Department of Communications
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.
Class B
Federal Communications Commission
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications. However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
Canadian Department of Communications
This Class B digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations. Cet appareil numérique de la classe B respecte toutes les exigences du Règlement sur le matériel brouilleur du Canada.
Compliance to EU Directives
Readers in the European Union (EU) must refer to the Manufacturer's Declaration of Conformity (DoC) for information** pertaining to the CE Mark compliance scheme. The Manufacturer includes a DoC for most every hardware product except for those bought for OEMs, if also available from an original manufacturer that also markets in the EU, or where compliance is not required as for electrically benign apparatus or cables.
To obtain the DoC for this product, click Declaration of Conformity at by product family. Select the appropriate product family, followed by your product, and a link to the DoC appears in Adobe Acrobat format. Click the Acrobat icon to download or read the DoC.
* Certain exemptions may apply in the USA, see FCC Rules §15.103 Exempted devices,and§15.105(c). Also available in
sections of CFR 47.
** The CE Mark Declaration of Conformity will contain important supplementary information and instructions for the user or
installer.
ni.com/hardref.nsf/
. This website lists the DoCs

Conventions

The following conventions are used in this manual:
» The » symbol leads you through nested menu items and dialog box options
to a final action. The sequence File»Page Setup»Options directs you to pull down the File menu, select the Page Setup item, and select Options from the last dialog box.
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to avoid injury, data loss, or a system crash.
bold Bold text denotes items that you must select or click on in the software,
such as menu items and dialog box options. Bold text also denotes parameter names.
italic Italic text denotes variables, emphasis, a cross reference, or an introduction
to a key concept. This font also denotes text that is a placeholder for a word or value that you must supply.
monospace
Text in this font denotes text or characters that you should enter from the keyboard, sections of code, programming examples, and syntax examples. Text in this font is also used for proper names of functions or variables.

Contents

Chapter 1 Taking Measurements with the NI 5911
Installing the NI 5911 ....................................................................................................1-1
Connecting Signals ........................................................................................................1-1
Acquiring Data with Your NI 5911 ...............................................................................1-3
Programmatically Controlling Your NI 5911..................................................1-3
Interactively Controlling Your NI 5911 with the Scope Soft Front Panel ...... 1-3
Chapter 2 Hardware Overview
Differential Programmable Gain Input Amplifier (PGIA) ............................................2-1
Differential Input ............................................................................................. 2-2
Grounding Considerations ................................................................2-2
Input Ranges....................................................................................................2-3
Input Impedance ..............................................................................................2-3
Input Bias ..........................................................................................2-4
Input Protection ...............................................................................................2-4
AC Coupling....................................................................................................2-4
Oscilloscope and Flexible Resolution Modes................................................................2-4
Oscilloscope Mode ..........................................................................................2-5
Sampling MethodsReal-Time and RIS........................................................ 2-5
Flexible Resolution Mode ...............................................................................2-5
How Flexible Resolution Works.......................................................2-6
Calibration .....................................................................................................................2-7
Internally Calibrating the NI 5911 ..................................................................2-7
When Internal Calibration Is Needed..............................................................2-7
What Internal Calibration Does....................................................................... 2-7
Why Errors Occur During Acquisition ............................................. 2-8
External Calibration.........................................................................................2-8
Triggering and Arming ..................................................................................................2-8
Analog Trigger Circuit ....................................................................................2-9
Trigger Hold-Off .............................................................................................2-10
Memory..........................................................................................................................2-10
Triggering and Memory Usage .......................................................................2-10
Multiple-Record Acquisitions........................................................................................2-11
© National Instruments Corporation vii NI 5911 User Manual
Contents
RTSI Bus and Clock PFI ............................................................................................... 2-11
PFI Lines ......................................................................................................... 2-11
PFI Lines as Inputs ........................................................................... 2-12
PFI Lines as Outputs......................................................................... 2-12
Synchronization .............................................................................................. 2-12
Appendix A Specifications
Appendix B Technical Support Resources
Glossary
Index
NI 5911 User Manual viii ni.com
Taking Measurements with the NI 5911
Thank you for buying a National Instruments (NI) 5911 digitizer, featuring
the FLEX ADC. This chapter provides information on installing,
connecting signals to, and acquiring data from your NI 5911.

Installing the NI 5911

There are two main steps involved in installation:
1. Install the NI-SCOPE driver software. You use this driver to write programs to control your NI 5911 in different application development environments (ADEs). Installing NI-SCOPE also allows you to interactively control your NI 5911 with the Scope Soft Front Panel.
2. Install your NI 5911. For step-by-step instructions for installing NI-SCOPE and the NI 5911, see Where to Start with Your NI Digitizer.
For multiple-board considerations, see the Operating Environment section in Appendix A, Specifications, of this manual.
1

Connecting Signals

Figure 1-1 shows the front panel for the NI 5911. The front panel contains three connectorsa BNC connector, an SMB connector, and a 9-pin mini circular DIN connector (see Figure 1-2).
The BNC connector is for attaching the analog input signal you wish to measure. The BNC connector is analog input channel 0. To minimize noise, do not allow the shell of the BNC cable to touch or lie near the metal of the computer chassis. The SMB connector is for external triggers and for generating a probe compensation signal. The SMB connector is PFI1. The DIN connector gives you access to an additional external trigger line. The DIN connector can be used to access PFI2.
Note
The +5 V signal is fused at 1.1 A. However, NI recommends limiting the current
from this pin to 30 mA. The fuse is self-resetting.
© National Instruments Corporation 1-1 NI 5911 User Manual
Chapter 1 Taking Measurements with the NI 5911
CH0
PFI1
PFI2
(DIN)

Figure 1-1. NI 5911 Connectors

789
56
4
3
12
1 +5 Volts (Fused) 2GND 3 Reserved
4Reserved 5Reserved 6PFI2
7 Reserved 8 Reserved 9 Reserved

Figure 1-2. 9-Pin Mini Circular DIN Connector

NI 5911 User Manual 1-2 ni.com
Chapter 1 Taking Measurements with the NI 5911

Acquiring Data with Your NI 5911

You can acquire data either programmaticallyby writing an application foryourNI5911—or interactively with the Scope Soft Front Panel.

Programmatically Controlling Your NI 5911

To help you get started programming your NI 5911, NI-SCOPE comes with examples that you can use or modify.
You can find examples for these different ADEs:
LabVIEWGo to
LabVIEW\Examples\Instr\niScopeExamples\
LabWindows/CVI, C, and Visual Basic with Windows 98/95—Go to
vxipnp\win95\Niscope\Examples\c\
LabWindows/CVI, C, and Visual Basic with Windows 2000/NTGo to
vxipnp\winnt\Niscope\Examples\
For information about using NI-SCOPE to programmatically control your digitizer, refer to your NI-SCOPE Software User Manual. Other resources include the NI-SCOPE Instrument Driver Quick Reference Guide.It contains abbreviated information on the most commonly used functions and LabVIEW VIs. For more detailed function reference help, see the
NI-SCOPE Function Reference Help file, located at Start»Programs» National Instruments»NI-SCOPE. For more detailed VI help, use LabVIEW context-sensitive help (Help»Show Context Help)orthe NI-SCOPE VI Reference Help, located at Start»Programs»National Instruments»NI-SCOPE.
Program Files\National Instruments\

Interactively Controlling Your NI 5911 with the Scope Soft Front Panel

The Scope Soft Front Panel allows you to interactively control your NI 5911 as you would a desktop oscilloscope. To launch the Scope Soft Front Panel select Start»Programs»National Instruments»NI-SCOPE» NI-SCOPE Soft Front Panel. Refer to the Scope Soft Front Panel Help file for instructions on configuring the Scope Soft Front Panel for your specific application.
Note
Press F1 with the Scope Soft Front Panel running to access the Scope Soft Front
Panel Help.
© National Instruments Corporation 1-3 NI 5911 User Manual
Hardware Overview
This chapter includes an overview of the NI 5911, explains the operation of each functional unit making up your NI 5911, and describes the signal connections. Figure 2-1 shows a block diagram of the NI 5911.
2
Analog Input
Connector
Protect/
Calibration
Calibration
Generator
Digital IO
Connector
Mux
AC/DC Coupling
PGIA
Noise
Shaper
Timing IO/
Memory Control
Digital Signal
Processor

Figure 2-1. NI 5911 Block Diagram

A/D Converter
100 MHz, 8-Bit
Capture
Memory
Reference Clock
Data

Differential Programmable Gain Input Amplifier (PGIA)

The analog input of the NI 5911 is equipped with a differential programmable gain input amplifier. The PGIA accurately interfaces to and scales the signal presented to the analog-to-digital converter (ADC) regardless of source impedance, source amplitude, DC biasing, or common-mode noise voltages.
© National Instruments Corporation 2-1 NI 5911 User Manual
Chapter 2 Hardware Overview

Differential Input

When measuring high dynamic range signals, ground noise is often a problem. The PGIA of the NI 5911 allows you to make noise-free signal measurements. The PGIA differential amplifier efficiently rejects any noise present on the ground signal. Internal to the PGIA, the signal presented at the negative input is subtracted from the signal presented at the positive input. As shown in Figure 2-2, this subtraction removes ground noise from the signal. The inner conductor of the BNC is V+; the outer shell is V–.
Input Signal
V+
V–
Ground Noise
Figure 2-2. Noise-Free Measurements of Signal
+
PGIA
V
out
Grounding Considerations
The path for the positive signal has been optimized for speed and linearity. You should always apply signals to the positive input and ground to the negative input. Reversing the inputs will result in higher distortion and lower bandwidth.
The negative input of the amplifier is grounded to PC ground through a 10 kresistor. The PGIA is therefore referenced to ground, so it is not necessary to make any external ground connections. If the device you connect to the NI 5911 is already connected to ground, ground-loop noise voltages may be induced into your system. Notice that in most of these situations, the 10 kresistance to PC ground is normally much higher than the cable impedances you use. As a result, most of the noise voltage occurs at the negative input of the PGIA where it is rejected, rather than in the positive input, where it would be amplified.
NI 5911 User Manual 2-2 ni.com

Input Ranges

Chapter 2 Hardware Overview
To optimize the ADC resolution, you can select different gains for the PGIA. In this way, you can scale your input signal to match the full input range of the converter. The NI 5911 PGIA offers seven different input ranges, from ±0.1 V to ±10 V, as shown in Table 2-1.
Table 2-1. Input Ranges for the NI 5911
Range Input Protection Threshold
±10 V ±10 V
±5 V ±5 V
±2 V ±5 V
±1 V ±5 V
±0.5 V ±5 V
±0.2 V ±5 V
±0.1 V ±5 V
Note
If you try to acquire a signal below the set input range the sensitive front-end components of the NI 5911 may become unstable and begin returning invalid data. To return the digitizer to a stable configuration, switch to the maximum input range setting and acquire an AC-coupled or 0 V signal.

Input Impedance

The input impedance of the NI 5911 PGIA is 1 Mbetween the positive and negative input, ±2% depending on input capacitance. The output impedance of the device connected to the NI 5911 and the input impedance of the NI 5911 form an impedance divider, which attenuates the input signal according to the following formula:
VsR
--------------------= RsRin+
in
where V
is the measured voltage, Vsis the source voltage, Rsis the external
m
source impedance, and R
V
m
is the input impedance.
in
If the device you are measuring has a very large output impedance, your measurements will be affected by this impedance divider. For example,
© National Instruments Corporation 2-3 NI 5911 User Manual
Chapter 2 Hardware Overview

Input Protection

if the device has 1 Moutput impedance, your measured signal will be one-half the actual signal value.
Input Bias
The inputs of the PGIA typically draw an input bias current of 1 nA at 25 °C. Attaching a device with a very high source impedance can cause an offset voltage to be added to the signal you measure, according to the formula R example, if the device you have attached to the NI 5911 has an output impedance of 10 k, typically the offset voltage is 10 µV (10 kΩ ×1nA).
The NI 5911 features input-protection circuits that protect both the positive and negative analog input from damage from AC and DC signals up to ±42 V.
×1nA,whereR
s
is the external source impedance. For
s
If the voltage at one of these inputs exceeds a threshold voltage, V input clamps to V minimize input currents to a nonharmful level.
The protection voltage, V
and a resistance of 100 kis inserted in the path to
tr
, is input range dependent, as shown in Table 2-1.
tr

AC Coupling

When you need to measure a small AC signal on top of a large DC component, you can use AC coupling. AC coupling rejects any DC component in your signal before it enters into the PGIA. Activating AC coupling inserts a capacitor in series with the input impedance. Input coupling can be selected via software. See the Digitizer Basics appendix in your NI-SCOPE Software User Manual for more information on input coupling.

Oscilloscope and Flexible Resolution Modes

In oscilloscope mode, the NI 5911 works as a conventional desktop oscilloscope, acquiring data at 100 MS/s with a vertical resolution of 8 bits. This mode is useful for displaying waveforms and for deriving waveform parameters such as slew rate, rise time, and settling time.
Flexible resolution differs from oscilloscope mode in two ways: it has higher resolution (sampling rate dependent), and the signal bandwidth is limited to provide antialiasing protection. This mode is useful for spectral
,the
tr
NI 5911 User Manual 2-4 ni.com
analysis, distortion analysis, and other measurements for which high resolution is crucial.

Oscilloscope Mode

The ADC converts at a constant rate of 100 MS/s, but you can choose to store only a fraction of these samples into memory at a lower rate. This allows you to store waveforms using fewer data points and decreases the burden of storing, analyzing, and displaying the waveforms. If you need faster sampling rates, you can use Random Interleaved Sampling (RIS) to effectively increase the sampling rate to 1 GS/s for repetitive waveforms.
In oscilloscope mode, all signals up to 100 MHz are passed to the ADC. You need to ensure that your signal is band-limited to prevent aliasing. Aliasing and other sampling terms are described more thoroughly in your NI-SCOPE Software User Manual.

Sampling Methods—Real-Time and RIS

There are two sampling methods available in oscilloscope mode, real-time and random interleaved sampling (RIS). Using real-time sampling, you can acquire data at a rate of 100/n MS/s, where n is a number from 1 to 2 RIS sampling can be used on repetitive signals to effectively extend the sampling rate above 100 MS/s. In RIS mode, you can sample at rates of 100 MS/s × n, where n is a number from 2 to 10.
Chapter 2 Hardware Overview
32
.

Flexible Resolution Mode

Table 2-2 shows the relationship between the available sampling rates, resolution, and the corresponding bandwidth for flexible resolution mode.
© National Instruments Corporation 2-5 NI 5911 User Manual
Chapter 2 Hardware Overview
Table 2-2. Available Sampling Rates and Corresponding Bandwidth
in Flexible Resolution Mode
Sampling Rate Resolution Bandwidth
12.5 MS/s 11 Bits 3.75 MHz
5MS/s 14 Bits 2MHz
2.5 MS/s 15.5 Bits 1MHz
1MS/s 17.5 Bits 400 kHz
500 kS/s 18 Bits 200 kHz
200 kS/s 18.5 Bits 80 kHz
100 kS/s 19 Bits 40 kHz
50 kS/s 19.5 Bits 20 kHz
20 kS/s 20.5 Bits 8kHz
10 kS/s 21 Bits 4kHz
Like any other type of converter that uses noise shaping to enhance resolution, the frequency response of the converter is only flat to its maximum useful bandwidth. The NI 5911 has a bandwidth of 4 MHz. Beyond this frequency, there is a span where the converter acts resonant and where a signal is amplified before being converted. These signals are attenuated in the subsequent digital filter to prevent aliasing. However, if the applied signal contains major signal components in this frequency range, such as harmonics or noise, the converter may overload and signal data will be invalid. In this case, you will receive an overload warning. You must then either select a higher input range or attenuate the signal.
How Flexible Resolution Works
The ADC can be sourced through a noise shaping circuit that moves quantization noise on the output of the ADC from lower frequencies to higher frequencies. A digital lowpass filter applied to the data removes all but a fraction of the original shaped quantization noise. The signal is then resampled to a lower sampling frequency and a higher resolution. Flexible resolution provides antialiasing protection due to the digital lowpass filter.
NI 5911 User Manual 2-6 ni.com

Calibration

The NI 5911 can be calibrated for very high accuracy and resolution due to an advanced calibration scheme. There are two different types of calibration: internal, or self, calibration and external calibration. A third option, internal restore, restores factory settings and should be used only in the event of a self-calibration failure.
Internal calibration is performed via a software command that compensates for drifts caused by environmental temperature changes. You can internally calibrate your NI 5911 without any external equipment connected. External calibration recalibrates the device when the specified calibration interval has expired. See Appendix A, Specifications,forthe calibration interval. External calibration requires you to connect an external precision voltage reference to the device.
Internally Calibrating the NI 5911
Internally calibrate your NI 5911 with a software function or a LabVIEW VI. See Chapter 3, Common Functions and Examples,ofyour NI-SCOPE Software User Manual for step-by-step instructions for calibrating your digitizer.
Chapter 2 Hardware Overview

When Internal Calibration Is Needed

To provide the maximum accuracy independent of temperature changes, the NI 5911 contains a heater that stabilizes the temperature of the most sensitive circuitries on the board. However, the heater can accommodate for temperature changes over a fixed range of ±5 °C. When temperatures exceed this range, the heater no longer is able to stabilize the temperature, and signal data becomes inaccurate. When the temperature range has been exceeded, you receive a warning, and you need to perform an internal calibration.

What Internal Calibration Does

Internal calibration performs the following operations:
The heater is set to regulate over a range of temperatures centered at the current environmental temperature. Thecircuit components require a certain amount of time to stabilize at the new temperature. This temperature stabilization accounts for the majority of the calibration time.
Gain and offset are calibrated for each individual input range.
© National Instruments Corporation 2-7 NI 5911 User Manual
Chapter 2 Hardware Overview
The linearity of the ADC is calibrated using an internal sinewave generator as reference.
The time-to-digital converter used for RIS measurements is calibrated.
Caution
internal calibration. For optimal calibration performance, disconnect the input signal from the NI 5911.
Do not apply high-amplitude or high-frequency signals to the NI 5911 during

External Calibration

Why Errors Occur During Acquisition
The NI 5911 uses a heater circuit to maintain constant temperature on the critical circuitry used in flexible resolution mode. If this circuit is unable to maintain the temperature within specification, an error is generated. This error indicates that the temperature of the ADC is out of range and should be recalibrated by performing an internal calibration. During acquisition in flexible resolution mode, an error will be generated if the input to the ADC goes out of range for the converter. The fact that this condition has occurred may not be obvious from inspecting the data due to the digital filtering that takes place on the acquired data. Therefore, an error occurs to let you know that the data includes some samples that were out of the range of the converter and may be inaccurate.
External calibration calibrates the internal reference on the NI 5911. The NI 5911 is already calibrated when it is shipped from the factory. Periodically, the NI 5911 will need external calibration to remain within the specified accuracy. For more information on calibration, contact NI, or visit
ni.com/calibration
Appendix A, Specifications.
. For actual intervals and accuracy, refer to

Triggering and Arming

There are several triggering methods for the NI 5911. The trigger can be an analog level that is compared to the input or any of several digital inputs. You can also call a software function to trigger the board. Figure 2-3 shows the different trigger sources. When you use a digital signal, that signal must be at a high TTL level for at least 40 ns before any triggers will be accepted.
Note
The NI 5911 does not support delayed triggering.
NI 5911 User Manual 2-8 ni.com
Chapter 2 Hardware Overview
Analog
Input
Gain
ATC_OUT
RTSI <0..6>
PFI1, PFI2
High
Level
Low
Level
a. Analog Trigger Circuit
Software
b. Trigger and Arm Sources
Figure 2-3.
+
COMP
COMP
7
2
Trigger Sources
Analog
Trigger
Circuit
Trigger
Arm
ATC_OUT

Analog Trigger Circuit

The analog trigger on the NI 5911 operates by comparing the current analog input to an onboard threshold voltage. This threshold voltage is the trigger value, and can be set within the current input range in 170 steps. This means that for a ±10 V input range, the trigger can be set in increments of 20 V/170 = 118 mV. There may also be a hysteresis value associated with the trigger that can be set in the same size increments. The hysteresis value creates a trigger window the signal must pass through before the trigger is accepted. You can generate triggers on a rising or falling edge condition. For a more complete discussion of triggering, see Chapter 3,
Common Functions and Examples, of your NI-SCOPE Software User Manual.
© National Instruments Corporation 2-9 NI 5911 User Manual
Chapter 2 Hardware Overview

Trigger Hold-Off

Note
Time to acquire posttrigger samples is (posttrigger samples)/(sample rate
(megahertz)).
Trigger hold-off is the minimum length of time (in seconds) from an accepted trigger to the start of the next record. In other words, when a trigger is accepted, the trigger counter is loaded with the desired hold-off time. After completing its current record, the digitizer records no data and accepts no triggers until the hold-off counter runs out. When the counter runs out, the next record begins and a trigger may be accepted. Setting a hold-off time shorter than posttrigger acquisition time has no effect, as triggers are always rejected during an acquisition.
Trigger hold-off is provided in hardware using a 32-bit counter clocked by a 25 MHz internal timebase. With this configuration, you can select a hardware hold-off value of 40 ns to 171.8 s in increments of 40 ns. For more information regarding trigger hold-off, see the Common Trigger Parameters section in Chapter 3, Common Functions and Examples,of your NI-SCOPE Software User Manual.

Memory

The NI 5911 allocates at least 4 kB of onboard memory for every acquisition. Samples are stored in this buffer before transfer to the host computer. Thus the minimum size for a buffer in the onboard memory is approximately 4,000 8-bit oscilloscope mode samples or 1,000 32-bit flexible resolution mode samples. Software allows you to specify buffers of less than these minimum sizes. However, the minimum number of points are still acquired into onboard memory, but only the specified number of points are transferred into the memory of the host computer.
The total number of samples that can be stored depends on the size of the acquisition memory module installed on the NI 5911 and the size of each acquired sample.

Triggering and Memory Usage

During the acquisition, samples are stored in a circular buffer that is continually rewritten until a trigger is received. After the trigger is received, the NI 5911 continues to acquire posttrigger samples if you have specified a posttrigger sample count. The acquired samples are placed into onboard memory. The number of posttrigger or pretrigger samples is only limited by the amount of onboard memory.
NI 5911 User Manual 2-10 ni.com

Multiple-Record Acquisitions

After the trigger has been received and the posttrigger samples have been stored, the NI 5911 can be configured to begin another acquisition that is stored in another onboard memory record. This is a multiple-record acquisition. To perform multiple-record acquisitions, configure the NI 5911 to the number of records you want to acquire before starting the acquisition. The NI 5911 acquires an additional record each time a trigger is accepted until all the requested records are stored in memory. You may acquire up to 1024 records if your NI 5911 is equipped with 4 MB of onboard memory, or 4096 records with 16 MB. Software intervention after the initial setup is not required.
Multiple-record acquisitions can quickly acquire numerous triggered waveforms because they allow hardware rearming of the digitizer before the data is fetched. Therefore the dead time, or the time when the digitizer is not ready for a trigger, is extremely small.
For more information on multiple-record acquisitions and dead time, see the Making a Multiple-Record Acquisition section in Chapter 5, Tasks and Examples, of your NI-SCOPE Software User Manual.
Chapter 2 Hardware Overview

RTSI Bus and Clock PFI

The RTSI bus allows NI digitizers to synchronize timing and triggering on multiple devices. The RTSI bus has seven bidirectional trigger lines and one bidirectional clock signal.
You can program any of the seven trigger lines to provide or accept a synchronous trigger signal. You can also use any of the RTSI trigger lines to provide a synchronization pulse from a master device if you are synchronizing multiple NI 5911s.
You can use the RTSI bus clock line to provide or accept a 10 MHz reference clock to synchronize multiple NI 5911 devices.

PFI Lines

The NI 5911 has two digital lines that can accept a trigger, accept or generate a reference clock, or output a 1 kHz square wave. The function of each PFI line is independent. However, only one trigger source can be accepted during acquisition.
© National Instruments Corporation 2-11 NI 5911 User Manual
Chapter 2 Hardware Overview

Synchronization

PFI Lines as Inputs
You can select PFI1 or PFI2 as inputs for a trigger or a reference clock. Please see the Synchronization section below for more information about the use of reference clocks in the NI 5911.
PFI Lines as Outputs
You can select PFI1 or PFI2 to output several digital signals.
Reference Clock is a 10 MHz clock that is synchronous to the 100 MHz sample clock on the NI 5911. You can use the reference clock to synchronize to another NI 5911 configured as a slave device or to other equipment that can accept a 10 MHz reference.
Frequency Output is a 1 kHz digital pulse train signal with a 50% duty cycle. The most common application of Frequency Output for the NI 5911 is to provide a signal for compensating a passive probe.
The NI 5911 uses a digital phase locked loop to synchronize the 100 MHz sample clock to a 10 MHz reference. This reference frequency can be supplied by an internal crystal oscillator or through an external frequency input through the RTSI bus clock line or a PFI input.
The NI 5911 may also output its 10 MHz reference on the RTSI bus clock line or a PFI line so that other NI 5911s or other equipment can be synchronized to the same reference.
While the reference clock input is sufficient to synchronize the 100 MHz sample clocks, it is also necessary to synchronize clock dividers on each NI 5911 so that internal clock divisors are synchronized on each different device. These lower frequencies are important because they are used to determine trigger times and sample position.
To synchronize the NI 5911 clock dividers, you must connect the digitizers with an NI RTSI bus cable. One of the RTSI bus triggers must be designated as a synchronization line. This line will be an output from the master device and an input on the slave device. To synchronize the digitizers, a single pulse is sent from the master NI 5911 to the slaves. This pulse supplies the slave devices with a reference time to clear their clock dividers. Hardware arming cannot be used during an acquisition using multiple devices. For more information about synchronization, refer to your NI-SCOPE Software User Manual.
NI 5911 User Manual 2-12 ni.com
Specifications
This appendix lists the specifications of the NI 5911. These specifications are typical at 25 °C unless otherwise stated.
Acquisition System
Bandwidth ..............................................100 MHz maximum,
Number of channels ............................... 1
Number of flexible resolution ADC.......1
Max sample rate ..................................... 1 GS/s repetitive,
Resolution
A
see Table 2-2, Available
Sampling Rates and Corresponding Bandwidth in Flexible Resolution Mode
100 MS/s single shot
Sample Rate Mode Effective Resolution
100/n*MS/s Oscilloscope 8Bits
12.5 MS/s Flexible Resolution 11 Bits
5MS/s Flexible Resolution 14 Bits
2.5 MS/s Flexible Resolution 15.5 Bits
1MS/s Flexible Resolution 17.5 Bits
500 kS/s Flexible Resolution 18 Bits
200 kS/s Flexible Resolution 18.5 Bits
100 kS/s Flexible Resolution 19 Bits
50 kS/s Flexible Resolution 19.5 Bits
© National Instruments Corporation A-1 NI 5911 User Manual
Appendix A Specifications
Sample Rate Mode Effective Resolution
20 kS/s Flexible Resolution 20.5 Bits
10 kS/s Flexible Resolution 21 Bits
*1<n<232in oscilloscope mode
Sample onboard memory........................4 MB or 16 MB
Memory sample depth
Sampling
Frequency
Mode
Sample Depth
(4 MB)
Sample Depth
(16 MB)
100/n*MS/s Oscilloscope 4MS 16 MS
12.5 MS/s Flexible Resolution 1MS 4MS
5MS/s Flexible Resolution 1MS 4MS
2.5 MS/s Flexible Resolution 1MS 4MS
1MS/s Flexible Resolution 1MS 4MS
500 kS/s Flexible Resolution 1MS 4MS
200 kS/s Flexible Resolution 1MS 4MS
100 kS/s Flexible Resolution 1MS 4MS
50 kS/s Flexible Resolution 1MS 4MS
20 kS/s Flexible Resolution 1MS 4MS
10 kS/s Flexible Resolution 1MS 4MS
* 1<n<232in oscilloscope mode
Vertical sensitivity (input ranges)
Input Range Noise Referred to Input
±10 V 174 dBfs/
±5 V 168 dBfs/
±2 V 160 dBfs/
±1 V 154 dBfs/
NI 5911 User Manual A-2 ni.com
Hz
Hz
Hz
Hz
Appendix A Specifications
Input Range Noise Referred to Input
Acquisition Characteristics
Accuracy
DC gain accuracy ................................... ±0.05% signal ±0.0001% fs
DC offset accuracy................................. ±0.1 mV ±0.01% fs
Input coupling ........................................ DC and AC, software selectable
AC coupling cut-off frequency
(–3 dB) ................................................... 2.3 Hz ±13%
Input impedance..................................... 1 M±2%
Max measurable input voltage ............... ±10 V (DC + peak AC)
±0.5 V 148 dBfs/
±0.2 V 140 dBfs/
±0.1 V 134 dBfs/
for all input ranges at 1 MS/s in flexible resolution mode
for all input ranges at 1 MS/s in flexible resolution mode
Hz
Hz
Hz
Input protection ......................................±42 VDC (DC + peak AC)
Input bias current ................................... ±1 nA, typical at 25 °C
Common-Mode Characteristics
Impedance to chassis ground ................. 10 k
Common-mode rejection ratio ............... CMRR > –70 dB, (F
© National Instruments Corporation A-3 NI 5911 User Manual
<1kHz)
in
Appendix A Specifications
Filtering
Sampling
Frequency
100/n*MS/s Oscilloscope 100 MHz ±3 dB N/A
12.5 MS/s Flexible
5MS/s Flexible
2.5 MS/s Flexible
1MS/s Flexible
500 kS/s Flexible
200 kS/s Flexible
100 kS/s Flexible
50 kS/s Flexible
20 kS/s Flexible
Filter Mode Bandwidth Ripple
3.75 MHz ±0.2 dB –60 dB
Resolution
2MHz ±0.1 dB –70 dB
Resolution
1MHz ±0.05 dB –80 dB
Resolution
400 kHz ±0.005 dB –80 dB
Resolution
200 kHz ±0.005 dB –80 dB
Resolution
80 kHz ±0.005 dB –80 dB
Resolution
40 kHz ±0.005 dB –80 dB
Resolution
20 kHz ±0.005 dB –80 dB
Resolution
8kHz ±0.005 dB –80 dB
Resolution
Alias
Attenuation
10 kS/s Flexible
Resolution
*1<n<232in oscilloscope mode
4kHz ±0.005 dB –80 dB
Dynamic Range
Noise (excluding input-referred noise)
Sampling Frequency Bandwidth Noise Density Tot a l N o is e
100/n*MS/s
12.5 MS/s
5MS/s
2.5 MS/s
NI 5911 User Manual A-4 ni.com
100 MHz –120 dBfs/ –43 dBfs
3.75 MHz –135 dBfs/ –64 dBfs
2MHz –150 dBfs/ –83 dBfs
1MHz –155 dBfs/ –91 dBfs
Hz
Hz
Hz
Hz
Appendix A Specifications
Sampling Frequency Bandwidth Noise Density Tot a l N o is e
1MS/s
500 kS/s
200 kS/s
100 kS/s
50 kS/s
20 kS/s
10 kS/s
* 1<n<232in oscilloscope mode
400 kHz –160 dBfs/ –104 dBfs
200 kHz –160 dBfs/ –107 dBfs
80 kHz –160 dBfs/ –111 dBfs
40 kHz –160 dBfs/ –114 dBfs
20 kHz –160 dBfs/ –117 dBfs
8kHz –160 dBfs/ –121 dBfs
4kHz –160 dBfs/ –124 dBfs
Hz
Hz
Hz
Hz
Hz
Hz
Hz
Distortion
SFDR for input
Sampling Frequency
100 MS/s
12.5 MS/s 65 dB 85 dB 125 dB
5MS/s 70 dB 90 dB 130 dB
2.5 MS/s 75 dB 95 dB 135 dB
0dBfs
50 dB 50 dB N/A
SFDR for input
–20 dBfs
SFDR for input
–60 dBfs (typical)
1MS/s 85 dB 105 dB 145 dB
500 kS/s 90 dB 110 dB 150 dB
200 kS/s 100 dB 110 dB 160 dB
100 kS/s 100 dB 110 dB 160 dB
50 kS/s 100 dB 110 dB 160 dB
20 kS/s 100 dB 110 dB 160 dB
10 kS/s 100 dB 110 dB 160 dB
Timebase System
Reference clock...................................... 10 MHz
Clock accuracy (as master) .................... 10 MHz ±50 ppm
© National Instruments Corporation A-5 NI 5911 User Manual
Appendix A Specifications
Clock input tolerance (as slave)..............10 MHz ±100 ppm
Triggering Systems
Clock jitter ..............................................<75 pS
rms,
independent of
reference clock source
Clock compatibility ...............................TTL for both input and output
Interpolator resolution
(repetitive only) ......................................1 ns
Sampling clock frequencies
Oscilloscope mode...........................100 MHz/n,where1<n<2
32
Flexible resolution mode .................100 MHz/n,wheren = 8; 20; 50;
100; 200; 500; 1,000; 2,000; 5,000; 10,000
Reference clock sources .........................PFI lines, RTSI clock, or onboard
Phase difference between
multiple instruments ...............................<5 ns, at any input frequency
<100 MHz, from input connector to input connector
Modes .....................................................Above threshold, below
threshold, between thresholds, outside thresholds
Source .....................................................CH0, RTSI<0..6>, PFI 1,2
Slope .......................................................Rising/falling
Hysteresis................................................Full-scale voltage/n,wheren is
between 1 and 170; full-scale voltage on TRIG is fixed to ±5 V (without external attenuation)
Coupling .................................................AC/DC on CH0, TRIG
Pretrigger depth ......................................1 to 16 million samples
Posttrigger depth.....................................1 to 16 million samples
Holdoff time ...........................................5 µs – 171.85 s in increments
of 40 ns
NI 5911 User Manual A-6 ni.com
Acquisition Modes
Appendix A Specifications
Sensitivity............................................... 170 steps in full-scale voltage
range
TRIG input range ................................... ±5 V (without external
attenuation)
TRIG input impedance........................... 1 M± 1% in parallel
with 30 pF ± 15 pF
TRIG input protection............................±42 V [(DC + peak AC) < 10 kHz,
without external attenuation]
RIS ......................................................... 1 GS/s down to 200 MS/s
effective sample rate, repetitive signals only. Data is interleaved in software.
RIS accuracy ..........................................<0.5 ns
Single-shot ............................................. 100 MS/s down to 10 kS/s sample
rate for transient and repetitive signals
Power Requirements
+5 VDC ................................................. 4 A
+12 VDC................................................ 100 mA
–12 VDC ............................................... 100 mA
Physical
Dimensions............................................. 33.8 by 9.9 cm
(13.3by3.9in)
I/O connectors
Analog input CH0...........................BNC female
Digital triggers ................................ SMB female, 9-pin mini DIN
© National Instruments Corporation A-7 NI 5911 User Manual
Appendix A Specifications
Operating Environment
Note
Multiple NI 5911s in the same computer may raise operating temperatures beyond specification and give rise to imprecise data. NI strongly recommends leaving an empty PCI slot between multiple NI 5911s or adding a fan.
Ambient temperature ..............................5 to 40 °C
Relative humidity ...................................10% to 90%, noncondensing
Storage Environment
Ambient temperature ..............................–20 to 65 °C
EMC Compliance
CE2001, FCC
Calibration
Internal....................................................Internal calibration is done upon
software command. The calibration involves gain, offset and linearity correction for all input ranges and input modes.
Interval.............................................1 week, or any time temperature
changes beyond ±5 °C. Hardware detects temperature variations beyond calibration limits, which can also be queried by software.
External...................................................Internal reference requires
recalibration
Interval.............................................1 year
Warm-up time.........................................1 minute
NI 5911 User Manual A-8 ni.com
Technical Support Resources
Web Support
National Instruments Web support is your first stop for help in solving installation, configuration, and application problems and questions. Online problem-solving and diagnostic resources include frequently asked questions, knowledge bases, product-specific troubleshooting wizards, manuals, drivers, software updates, and more. Web support is available through the Technical Support section of
NI Developer Zone
ni.com
B
.
The NI Developer Zone at building measurement and automation systems. At the NI Developer Zone, you can easily access the latest example programs, system configurators, tutorials, technical news, as well as a community of developers ready to share their own techniques.
Customer Education
National Instruments provides a number of alternatives to satisfy your training needs, from self-paced tutorials, videos, and interactive CDs to instructor-led hands-on courses at locations around the world. Visit the Customer Education section of syllabi, training centers, and class registration.
System Integration
If you have time constraints, limited in-house technical resources, or other dilemmas, you may prefer to employ consulting or system integration services. You can rely on the expertise available through our worldwide network of Alliance Program members. To find out more about our Alliance system integration solutions, visit the System Integration section of
ni.com
ni.com/zone
ni.com
.
is the essential resource for
for online course schedules,
© National Instruments Corporation B-1 NI 5911 User Manual
Appendix B Technical Support Resources
Worldwide Support
National Instruments has offices located around the world to help address your support needs. You can access our branch office Web sites from the Worldwide Offices section of up-to-date contact information, support phone numbers, e-mail addresses, and current events.
If you have searched the technical support resources on our Web site and still cannot find the answers you need, contact your local office or National Instruments corporate. Phone numbers for our worldwide offices are listed at the front of this manual.
ni.com
. Branch office Web sites provide
NI 5911 User Manual B-2 ni.com

Glossary

Prefix Meanings Value
p- pico- 10
n- nano- 10
µ- micro- 10
m- milli- 10
k- kilo- 10
M- mega- 10
G- giga- 10

Symbols

% percent
+ positive of, or plus
12
9
6
3
3
6
9
negative of, or minus
/per
° degree
± plus or minus
ohm
A
A amperes
A/D analog to digital
AC alternating current
© National Instruments Corporation G-1 NI 5911 User Manual
Glossary
AC coupled the passing of a signal through a filter network that removes the
DC component of the signal
ADC analog-to-digital converteran electronic device, often an integrated
circuit, that converts an analog voltage to a digital number
ADC resolution the resolution of the ADC, which is measured in bits. An ADC with16 bits
has a higher resolution, and thus a higher degree of accuracy, than a 12-bit ADC.
alias a false lower frequency component that appears in sampled data acquired
at too low a sampling rate
amplification a type of signal conditioning that improves accuracy in the resulting
digitized signal and reduces noise
amplitude flatness a measure of how close to constant the gain of a circuit remains over a range
of frequencies
attenuate to reduce in magnitude
B
bbit—one binary digit, either 0 or 1
Bbyte—eight related bits of data, an eight-bit binary number. Also used
to denote the amount of memory required to store one byte of data.
bandwidth the range of frequencies present in a signal, or the range of frequencies to
which a measuring device can respond
buffer temporary storage for acquired or generated data (software)
bus the group of conductors that interconnect individual circuitry in a computer.
Typically, a bus is the expansion vehicle to which I/O or other devices are connected. Examples of PC buses are the PCI and ISA bus.
C
CCelsius
channel pin or wire lead to which you apply or from which you read the analog or
digital signal
NI 5911 User Manual G-2 ni.com
Glossary
clock hardware component that controls timing for reading from or writing to
groups
CMRR common-mode rejection ratioa measure of an instrument’s ability to
reject interference from a common-mode signal, usually expressed in decibels (dB)
counter/timer a circuit that counts external pulses or clock pulses (timing)
coupling the manner in which a signal is connected from one location to another
D
dB decibelthe unit for expressing a logarithmic measure of the ratio of
two signal levels: dB=20log10 V1/V2, for signals in volts
DC direct current
default setting a default parameter value recorded in the driver. In many cases, the default
input of a control is a certain value (often 0) that means use the current default setting.
device a plug-in data acquisition board, card, or pad. The NI 5911 is an example
of a device.
differential input an analog input consisting of two terminals, both of which are isolated from
computer ground, whose difference is measured
double insulated a device that contains the necessary insulating structures to provide electric
shock protection without the requirement of a safety ground connection
drivers software that controls a specific hardware instrument
E
EEPROM electrically erasable programmable read-only memoryROM that can be
erased with an electrical signal and reprogrammed
equivalent time sampling
event the condition or state of an analog or digital signal
© National Instruments Corporation G-3 NI 5911 User Manual
any method used to sample signals in such a way that the apparent sampling rate is higher than the real sampling rate
Glossary
F
filtering a type of signal conditioning that allows you to filter unwanted signals from
the signal you are trying to measure
fs full-scale—total voltage in the input range. A ±10 V input range is 20 V fs
G
gain the factor by which a signal is amplified, sometimes expressed in decibels
H
hardware the physical components of a computer system, such as the circuit boards,
plug-in boards, chassis, enclosures, peripherals, cables, and so on
harmonics multiples of the fundamental frequency of a signal
Hz hertzper second, as in cycles per second or samples per second
I
I/O input/outputthe transfer of data to/from a computer system involving
communications channels, operator interface devices, and/or data acquisition and control interfaces
in. inches
inductance the relationship of induced voltage to current
input bias current the current that flows into the inputs of a circuit
input impedance the measured resistance and capacitance between the input terminals of a
circuit
instrument driver a set of high-level software functions that controls a specific plug-in DAQ
board. Instrument drivers are available in several forms, ranging from a function callable language to a virtual instrument (VI) in LabVIEW.
interrupt a computer signal indicating that the CPU should suspend its current task
to service a designated activity
NI 5911 User Manual G-4 ni.com
interrupt level the relative priority at which a device can interrupt
ISA industry standard architecture
K
kkilo—the standard metric prefix for 1,000, or 103, used with units of
measure such as volts, hertz, and meters
kS 1,000 samples
L
LabVIEW laboratory virtual instrument engineering workbench—a graphical
programming ADE developed by National Instruments
LSB least significant bit
M
Glossary
m meters
MB megabytes of memory
memory buffer see buffer
MS million samples
MSB mostsignificantbit
© National Instruments Corporation G-5 NI 5911 User Manual
Glossary
N
noise an undesirable electrical signalnoise comes from external sources such
as the AC power line, motors, generators, transformers, fluorescent lights, soldering irons, CRT displays, computers, electrical storms, welders, radio transmitters, and internal sources such as semiconductors, resistors, and capacitors. Noise corrupts signals you are trying to send or receive.
Nyquist frequency a frequency that is one-half the sampling rate. See also Nyquist Sampling
Theorem.
Nyquist Sampling Theorem
the theorem states that if a continuous bandwidth-limited analog signal contains no frequency components higher than half the frequency at which it is sampled, then the original signal can be recovered without distortion.
O
OhmsLaw (R=V/I)—the relationship of voltage to current in a resistance
overrange a segment of the input range of an instrument outside of the normal
measuring range. Measurements can still be made, usually with a degradation in specifications.
oversampling sampling at a rate greater than the Nyquist frequency
P
passband the frequency range that a filter passes without attenuation
PCI Peripheral Component Interconnecta high-performance expansion bus
architecture originally developed by Intel to replace ISA and EISA; it is achieving widespread acceptance as a standard for PCs and workstations and offers a theoretical maximum transfer rate of 132 Mbytes/s
peak value the absolute maximum or minimum amplitude of a signal (AC + DC)
posttriggering the technique to acquire a programmed number of samples after trigger
conditions are met
NI 5911 User Manual G-6 ni.com
Glossary
pretriggering the technique used on a device to keep a buffer filled with data, so that when
the trigger conditions are met, the sample includes the data leading up to the trigger condition
PXI PCI eXtensions for Instrumentation. PXI is an open specification that
builds off the CompactPCI specification by adding instrumentation-specific features.
R
Rresistor
RAM random-access memory
real-time sampling sampling that occurs immediately
random interleaved sampling
resolution the smallest signal increment that can be detected by a measurement
rms root mean squarea measure of signal amplitude; the square root of the
ROM read-only memory
RTSI bus real-time system integration bus—the National Instruments timing bus that
method of increasing the sample rate by repetitively sampling a repeated waveform
system. Resolution can be expressed in bits or in digits. The number of bits in a system is roughly equal to 3.3 times the number of digits.
average value of the square of the instantaneous signal amplitude
connects devices directly, by means of connectors on top of the boards, for precise synchronization of functions
S
s seconds
S samples
S/s samples per secondused to express the rate at which an instrument
samples an analog signal. 100 MS/s would equal 100 million samples each second.
© National Instruments Corporation G-7 NI 5911 User Manual
Glossary
sense in four-wire resistance the sense measures the voltage across the resistor
being excited by the excitation current
settling time the amount of time required for a voltage to reach its final value within
specified limits
source impedance a parameter of signal sources that reflects current-driving ability of voltage
sources (lower is better) and the voltage-driving ability of current sources (higher is better)
system noise a measure of the amount of noise seen by an analog circuit or an ADC when
the analog inputs are grounded
T
temperature coefficient
thermal drift measurements that change as the temperature varies
thermal EMFs thermal electromotive forcesvoltages generated at the junctions of
thermoelectric potentials
transfer rate the rate, measured in bytes/s, at which data is moved from source to
trigger any event that causes or starts some form of data capture
the percentage that a measurement will vary according to temperature. See also thermal drift
dissimilar metals that are functions of temperature. Also called thermoelectric potentials.
see thermal EMFs
destination after software initialization and set up operations; the maximum rate at which the hardware can operate
U
undersampling sampling at a rate lower than the Nyquist frequency—can cause aliasing
update rate the number of output updates per second
NI 5911 User Manual G-8 ni.com
V
V volts
VAC volts alternating current
VDC volts direct current
Glossary
V
error
voltage error
VI virtual instrument(1) a combination of hardware and/or software
elements, typically used with a PC, that has the functionality of a classic stand-alone instrument (2) a LabVIEW software module (VI), which consists of a front panel user interface and a block diagram program
V
rms
volts, root mean square value
W
waveform shape the shape the magnitude of a signal creates over time
working voltage the highest voltage that should be applied to a product in normal use,
normally well under the breakdown voltage for safety margin
© National Instruments Corporation G-9 NI 5911 User Manual

Index

A
AC coupling, 2-4 accuracy characteristics, A-3 acquisition
multiple record, 2-11 Scope Soft Front Panel, 1-3
acquisition characteristics specifications
accuracy, A-3 common-mode characteristics, A-3 distortion, A-5 dynamic range, A-4
filtering, A-4 acquisition modes specifications, A-7 acquisition system specifications, A-1 analog trigger circuit, 2-9 arming. See triggering and arming
B
bias, input, 2-4 block diagram of NI 5911, 2-1 BNC connector, 1-1
C
calibration
errors occurring during acquisition, 2-8
external calibration, 2-8
internal calibration, 2-7
specifications, A-8 clock lines, 2-12 common-mode characteristics, A-3 connectors
BNC connector, 1-1
DIN connector, 1-1
location on front panel (figure), 1-2
SMB connector, 1-1
conventions used in the manual, vi customer education, B-1
D
dead time, in multiple record acquisition, 2-11 differential input
grounding considerations, 2-2 noise-free signal measurement (figure), 2-2
differential programmable gain input amplifier
(PGIA)
AC coupling, 2-4 differential input, 2-2 input bias, 2-4 input impedance, 2-3 input protection, 2-4 input ranges, 2-3
noise-free signal measurement (figure), 2-2 DIN connector, 1-1 distortion specifications, A-5 dynamic range specifications, A-4
E
EMC compliance, A-8 errors during acquisition, 2-8
F
filtering specifications, A-4 flexible resolution mode
available sampling rates (table), 2-6
definition, 2-5
memory sample depth (table), A-2
purpose and use, 2-6
sampling rate specifications (table), A-1
© National Instruments Corporation I-1 NI 5911 User Manual
Index
G
grounding considerations, 2-2
H
hardware overview
See also specifications acquisition system
PFI lines, 2-11
triggering and arming, 2-8 block diagram of NI 5911, 2-1 calibration, 2-7 differential programmable gain input
amplifier (PGIA)
AC coupling, 2-4
differential input, 2-2
grounding considerations, 2-2
input bias, 2-4
input impedance, 2-3
input protection, 2-4
input ranges, 2-3
noise-free signal measurement
(figure), 2-2 flexible resolution mode, 2-5 memory, 2-10 multiple record acquisition, 2-11 oscilloscope mode, 2-5 RTSI bus trigger and clock lines, 2-11 trigger hold-off, 2-10 triggering and arming
analog trigger circuit, 2-9 trigger hold-off, 2-10 trigger sources (figure), 2-9
I
impedance
formula for impedance divider, 2-3 input and output impedance, 2-3
input bias, 2-4 input impedance, 2-3
input protection circuits, 2-4 input ranges, 2-3 installing NI 5911, 1-1
M
memory
description, 2-10 triggering and memory usage, 2-10
multiple record acquisitions, 2-11
N
National Instruments Web support, B-1 NI 5911
See also hardware overview block diagram, 2-1 connectors
BNC connector, 1-1 DIN connector, 1-1 location on front panel (figure), 1-2
SMB connector, 1-1 front panel (figure), 1-2 installing, 1-1 Scope Soft Front Panel, 1-3 specifications
acquisition characteristics, A-3
acquisition modes, A-7
acquisition system, A-1
timebase system, A-5
triggering systems, A-6
NI Developer Zone, B-1 NI-SCOPE driver software
installing, 1-1 programmatically controlling
NI 5911, 1-3
noise-free measurements, 2-2
NI 5911 User Manual I-2 ni.com
Index
O
operating environment specifications, A-8 oscilloscope mode
definition, 2-4 purpose and use, 2-5 Real-Time and RIS sampling
methods, 2-5
output impedance, 2-3
P
PFI lines
as inputs, 2-12 as outputs, 2-12 overview, 2-11
PGIA. See differential programmable gain
input amplifier (PGIA) physical specifications, A-7 power requirement specifications, A-7 programmable gain input amplifier PGIA. See
differential programmable gain input
amplifier (PGIA) programmatically controlling NI 5911, 1-3
R
Random Interleaved Sampling (RIS), 2-5 real-time sampling, 2-5 RIS (Random Interleaved Sampling), 2-5 RTSI bus trigger and clock lines
PIF lines, 2-11 purpose and use, 2-11 synchronization, 2-12
specifications
acquisition characteristics
accuracy, A-3 common-mode characteristics, A-3 distortion, A-5 dynamic range, A-4
filtering, A-4 acquisition modes, A-7 acquisition system, A-1 calibration, A-8 EMC compliance, A-8 operating environment, A-8 physical, A-7 power requirements, A-7 storage environment, A-8 timebase system, A-5 triggering systems, A-6
storage environment specifications, A-8 synchronization, 2-12 system integration, by National Instruments,
B-1
T
technical support resources, B-1 timebase system specifications, A-5 triggering and arming
analog trigger circuit, 2-9 memory usage, 2-10 specifications, A-6 trigger hold-off, 2-10 trigger sources (figure), 2-9
V
S
sampling methodsreal-time and RIS, 2-5 sampling rate, flexible resolution mode
(table), 2-6 Scope Soft Front Panel, 1-3 SMB connector, 1-1
© National Instruments Corporation I-3 NI 5911 User Manual
vertical sensitivity specifications, A-2
W
Web support from National Instruments, B-1 worldwide technical support, B-2
Loading...