NI PXI-5422 16-Bit 200 MS/s Arbitrary Waveform Generator
Unless otherwise noted, the following conditions were used for each
specification:
•Analog Filter enabled.
•Signals terminated with 50 Ω.
•Direct Path set to 1 V
and High-Gain Amplifier Path set to 12 V
•Sample rate set to 200 MS/s and the Sample Clock Source set to
Divide-by-N.
Typical values are representative of an average unit operating at room
temperature (20 °C ±3 °C). Specifications are subject to change without
notice. For the most recent NI 5422 specifications, visit
To access all of the NI 5422 documentation, including the NI Signal Generators Getting Started Guide, which contains functional descriptions
of the NI 5422 signals, navigate to Start»Programs»National Instruments»NI-FGEN»Documentation.
Hot Surface If the NI 5422 has been in use, it may exceed safe handling temperatures and
cause burns. Allow the NI 5422 to cool before removing it from the chassis.
Where to Go for Support .........................................................................36
CH 0
(Channel 0 Analog Output, Front Panel Connector)
Table 1.
SpecificationValu eComments
Number of
1—
Channels
ConnectorSMB (jack)—
Output Voltage Characteristics
Output Paths1. The software-selectable Main Output Path setting
provides full-scale voltages from 12.00 V
5.64 mV
into a 50 Ω load. NI-FGEN uses either the
pk-pk
pk-pk
to
—
Low-Gain Amplifier or the High-Gain Amplifier when
the Main Output Path is selected, depending on the Gain
attribute.
2. The software-selectable Direct Path is optimized for
IF applications and provides full-scale voltages from
DAC
1.000 V
16 bits—
to 0.707 V
pk-pk
pk-pk
.
Resolution
NI 5422 Specifications2ni.com
Table 1. (Continued)
SpecificationValu eComments
Amplitude and Offset
Amplitude
Amplitude (V
)1. Amplitude
pk-pk
Range
PathLoad
Minimum ValueMaximum Value
Direct50 Ω0.7071.00
1 kΩ1.351.91
Open1.412.00
Low-
50 Ω0.005642.00
Gain
Amplifier
1 kΩ0.01073.81
Open0.01134.00
High-
50 Ω0.033812.0
Gain
Amplifier
1 kΩ0.064422.9
Open0.067624.0
Amplitude
3 digits
Resolution
Offset RangeSpan of ±50% of Amplitude Range with increments
<0.0028% of Amplitude Range.
values assume
the full scale
of the DAC is
utilized. If an
amplitude
smaller than
the minimum
value is
desired, then
waveforms less
than full scale
of the DAC can
be used.
2. NI-FGEN
compensates
for userspecified
resistive loads.
PathLoadMaximum Output Voltage (V)The combination
Output
Vo l t a g e
Direct50 Ω±0.500
1 kΩ±0.953
Open±1.000
LowGain
Amplifier
50 Ω±1.000
1 kΩ±1.905
Open±2.000
High-
Gain
Amplifier
50 Ω±6.000
1 kΩ±11.43
Open±12.00
Accuracy
DC AccuracyFor the Low-Gain or High-Gain Amplifier Path:
±0.2% of Amplitude ± 0.05% of Offset ± 500 µV
(within ±10 °C of self-calibration temperature)
±0.4% of Amplitude ± 0.05% of Offset ± 1 mV
(0 °C to 55 °C)
For the Direct Path:
Gain Accuracy: ±0.2% (within ±10 °C of self-calibration
temperature)
Gain Accuracy: ±0.4% (0 °C to 55 °C)
DC Offset Error: ±30 mV (0 °C to 55 °C)
of Amplitude and
Offset is limited
by the Maximum
Output Voltage.
All paths are
calibrated for
amplitude and
gain errors. The
Low-Gain and
High-Gain
Amplifier Paths
also are
calibrated for
offset errors.
Calibrated for
high impedance
load.
AC Amplitude
Accuracy
±1.0% of desired Amplitude ± 1 mV50 kHz sine
wave.
Output Characteristics
Output
50 Ω nominal or 75 Ω nominal, software-selectable.—
Impedance
Output
DC—
Coupling
NI 5422 Specifications4ni.com
Table 1. (Continued)
SpecificationValu eComments
Output Characteristics (Continued)
Output EnableSoftware-selectable. When the Output Path is disabled, the
CH 0 Output is terminated to ground with a 1 W resistor
equal to the selected output impedance.
Maximum
Output
Overload
The CH 0 output can be connected to a 50 Ω, ±12 V
(±8 V for the Direct Path) source without sustaining any
damage. No damage occurs if the CH 0 output is shorted to
ground indefinitely.
Waveform
Summing
The CH 0 output supports waveform summing among
similar paths—specifically, the outputs of multiple NI 5422
signal generators can be connected directly together.
Frequency and Transient Response
Analog
Filter
Pulse
Response
Software-selectable 7-pole elliptical filter for image
suppression.
PathValues are
DirectLow-Gain
Amplifier
High-Gain
Amplifier
Rise/Fall Time1.0 ns2.1 ns4.8 ns
Aberration16%6%8%
—
—
—
Available on
Low-Gain
Amplifier and
High-Gain
Amplifier Paths.
typical. Analog
Filter disabled.
Measured with a
1 m RG-223
cable.
Suggested Maximum Frequencies for Common Functions
FunctionPathDisable the
Analog Filter for
square, ramp,
and triangle
functions.
Direct
Low-Gain
Amplifier
High-Gain
Amplifier
Sine80 MHz80 MHz43 MHz
SquareNot Recommended50 MHz25 MHz
RampNot Recommended10 MHz10 MHz
TriangleNot Recommended10 MHz10 MHz
14
12
10
43 MHz
8
Not Recommended
6
4
Sine Amplitude (Vpp into 50 Ω)
2
0
0 20406080
NI 5422 Specifications8ni.com
Recommended Operation
Frequency (MHz)
Figure 5. Recommended Sine Wave Frequency Versus Amplitude
Table 1. (Continued)
SpecificationVa lu eComments
Spectral Characteristics
Spurious-Free
PathAmplitude
Dynamic
Range
(SFDR) with
Harmonics
Direct
Low-Gain
Amplifier
1 MHz–70 dBc–65 dBc–66 dBc
5 MHz–70 dBc–65 dBc–58 dBc
10 MHz–70 dBc–65 dBc–52 dBc
20 MHz–63 dBc–64 dBc–49 dBc
30 MHz–57 dBc–60 dBc–43 dBc
40 MHz–48 dBc–53 dBc–39 dBc
50 MHz–48 dBc–53 dBc—
60 MHz–47 dBc–52 dBc—
70 MHz–47 dBc–52 dBc—
80 MHz–41 dBc–52 dBc—
High-Gain
Amplifier
–1 dBFS.
Measured from
DC to 100 MHz.
Also called
harmonic
distortion.
SFDR with
harmonics at low
amplitudes is
limited by a
–148 dBm/Hz
noise floor. All
values are
typical and
include aliased
harmonics.
PathAmplitude
Dynamic
Range
(SFDR)
without
Harmonics
Direct
Low-Gain
Amplifier
1 MHz–85 dBFS–80 dBFS–77 dBFS
5 MHz–85 dBFS–80 dBFS–77 dBFS
10 MHz–80 dBFS–80 dBFS–77 dBFS
20 MHz–80 dBFS–80 dBFS–77 dBFS
30 MHz–73 dBFS–71 dBFS–68 dBFS
40 MHz–48 dBFS–58 dBFS–55 dBFS
50 MHz–48 dBFS–53 dBFS—
60 MHz–47 dBFS–52 dBFS—
70 MHz–47 dBFS–52 dBFS—
80 MHz–41 dBFS–52 dBFS—
Average Noise
Density
Path
Amplitude
Range
V
pk-pk
Average Noise Density
nV
-----------
dBmdBm/Hz
Hz
Direct1.004.019.9–141–145
High-Gain
Amplifier
dBFS/
Hz
–1 dBFS.
Measured from
DC to 100 MHz.
SFDR without
harmonics at low
amplitudes is
limited by a
–148 dBm/Hz
noise floor.
All values are
typical and
include aliased
harmonics.
Average Noise
Density at small
amplitudes is
limited by a
–168 dBm/Hz
noise floor.
NI Signal
Generators
Help»Devices»
NI 5422»
NI PXI-5422»
Waveform
Generation»
PXI_Trig<0..6>±2 Sample Clock
N/A
Marker Events.
Periods
Jitter40 ps rms (typical)—
NI 5422 Specifications26ni.com
Arbitrary Waveform Generation Mode
Table 10.
SpecificationVal ueComments
Memory
Usage
Onboard
Memory Size
Output ModesArbitrary Waveform mode and Arbitrary Sequence mode—
Arbitrary
Waveform
Mode
Arbitrary
Sequence
Mode
The NI 5422 uses the Synchronization and Memory Core
(SMC) technology in which waveforms and instructions
share onboard memory. Parameters, such as number of
segments in sequence list, maximum number of waveforms
in memory, and number of samples available for waveform
storage, are flexible and user defined.
8 MB standard:
8,388,608 bytes
32 MB option:
33,554,432 bytes
In Arbitrary Waveform mode, a single waveform is selected
from the set of waveforms stored in onboard memory and
generated.
In Arbitrary Sequence mode, a sequence directs the NI 5422
to generate a set of waveforms in a specific order. Elements
of the sequence are referred to as segments. Each segment is
associated with a set of instructions. The instructions
identify which waveform is selected from the set of
waveforms in memory, how many loops (iterations) of
the waveform are generated, and at which sample in the
waveform a marker output signal is sent.
The Minimum
Waveform Size
is sample rate
dependent in
Arbitrary
Sequence mode.
Table 10. (Continued)
SpecificationVal ueComments
Loop Count1 to 16,777,215.
—
Burst trigger: Unlimited
QuantumWaveform size must be an integer multiple of four samples—
Memory Limits
Arbitrary
Waveform
8 MB
Standard
4,194,176
Samples
32 MB
Option
16,777,088
Samples
256 MB
Option
134,217,600
Samples
512 MB
Option
268,435,328
Samples
All trigger modes
except where
noted.
Mode,
Maximum
Waveform
Memory
Arbitrary
Sequence
Mode,
4,194,048
Samples
16,776,960
Samples
134,217,472
Samples
268,435,200
Samples
Condition: One
or two segments
in a sequence.
Maximum
Waveform
Memory
Arbitrary
Sequence
Mode,
Maximum
65,000
Burst
trigger:
8,000
262,000
Burst
trigger:
32,000
2,097,000
Burst
trigger:
262,000
4,194,000
Burst
trigger:
524,000
Condition: One
or two segments
in a sequence.
Waveforms
Arbitrary
Sequence
Mode,
Maximum
104,000
Burst
trigger:
65,000
418,000
Burst
trigger:
262,000
3,354,000
Burst
trigger:
2,090,000
6,708,000
Burst
trigger:
4,180,000
Condition:
Wav eform
memory is
<4,000 samples.
Segments in a
Sequence
NI 5422 Specifications28ni.com
Calibration
Table 11.
SpecificationValueComments
Self-Calibration An onboard, 24-bit ADC and precision voltage reference are
used to calibrate the DC gain and offset. The self-calibration
is initiated by the user through the software and takes
approximately 90 seconds to complete.
External
Calibration
Calibration
Interval
Warm-up Time15 minutes—
The External Calibration calibrates the VCXO, voltage
reference, DC gain, and offset. Appropriate constants are
stored in nonvolatile memory.
Specifications valid within two years of External
Calibration.
Figure 13. 5 V Current Versus Frequency and Amplitude
NI 5422 Specifications30ni.com
Software
Table 13.
SpecificationValueComments
Driver
Software
Application
Software
Interactive
Control and
Configuration
software
NI-FGEN version 2.2.1 or later. NI-FGEN is an
IVI-compliant driver that allows you to configure, control,
and calibrate the NI 5422. NI-FGEN provides application
programming interfaces for many development
environments.
NI-FGEN provides programming interfaces for the
following application development environments:
•LabVIEW
™
• LabWindows
/CVI
™
• Measurement Studio
• Microsoft Visual C/C++
• Microsoft Visual Basic
• Borland C/C++
National Instruments provides several options for
interactively controlling and configuring your NI5422:
Note To ensure that the NI PXI-5422 cools effectively, follow the guidelines in the
Maintain Forced-Air Cooling Note to Users included in the NI 5422 kit. The NI PXI-5422
is intended for indoor use only.
Table 14.
SpecificationsValu eComments
Operating
Temperature
Storage
Temperature
Operating
Relative
Humidity
Storage
Relative
Humidity
Operating
Shock
Storage Shock50 g, half-sine, 11 ms pulse. Meets IEC-60068-2-27. Test
Operating
Vibration
0 ºC to +55 ºC in all NI PXI chassis except the following:
0 ºC to +45 ºC when installed in an NI PXI-101x or
NI PXI-1000B chassis. (Meets IEC-60068-2-1 and
IEC-60068-2-2.)
–25 ºC to +85 ºC. Meets IEC-60068-2-1 and
IEC-60068-2-2.
10% to 90%, noncondensing. Meets IEC-60068-2-56.—
5% to 95%, noncondensing. Meets IEC-60068-2-56.—
30 g, half-sine, 11 ms pulse. Meets IEC-60068-2-27. Test
profile developed in accordance with MIL-PRF-28800F.
profile developed in accordance with MIL-PRF-28800F.
5 Hz to 500 Hz, 0.31 g
. Meets IEC-60068-2-64.Spectral and jitter
rms
—
—
Spectral and jitter
specifications
could degrade.
—
specifications
could degrade.
Storage
Vibration
Altitude2,000 m maximum (at 25 °C ambient temperature)—
Pollution
Degree
NI 5422 Specifications32ni.com
5 Hz to 500 Hz, 2.46 g
profile exceeds requirements of MIL-PRF-28800F, Class B.
2—
. Meets IEC-60068-2-64. Test
rms
—
Safety, Electromagnetic Compatibility, and
CE Compliance
Table 15.
SpecificationVal ueComments
SafetyThe NI 5422 meets the requirements of the following
standards of safety for electrical equipment for
measurement, control, and laboratory use:
• IEC 61010-1, EN 61010-1
• UL 61010-1
• CAN/CSA-C22.2 No. 61010-1
EmissionsEN 55011 Class A at 10 m
FCC Part 15A above 1 GHz
ImmunityEN 61326:1997 + A2:2001, Table 1
Up to 4 mVpp noise (about –44 dBm) may be present
on the output during the conducted immunity test. Use
of the product at levels below –44 dBm will result in
self-recoverable errors.
Good screening (shielding) techniques must be
employed throughout the data acquisition system.
For UL and
other safety
certifications,
refer to the product
label or to
ni.com/
certification
search by model
number or product
line, and click the
appropriate link in
the Certification
column.
EMC/EMICE, C-Tick, and FCC Part 15 (Class A) Compliant
—
Notes:
1. This device is not intended for, and is restricted
from, use in residential areas.
2. For EMC compliance, operate this device with
shielded cabling.
3. When connected to other test objects, this product
may cause radio interference. If this occurs, you
may be required to take adequate measures to
reduce the interference.
This product meets the essential requirements of applicable European Directives as amended for
CE marking, as follows:
Low-Voltage
73/23/EEC—
Directive (safety)
Electromagnetic
89/336/EEC—
Compatibility
Directive (EMC)
Note: Refer to the Declaration of Conformity (DoC) for this product for any additional regulatory
compliance information. To obtain the DoC for this product, visit
ni.com/certification,
search by model number or product line, and click the appropriate link in the Certification column.
The National Instruments Web site is your complete resource for technical
support. At
troubleshooting and application development self-help resources to email
and phone assistance from NI Application Engineers.
A Declaration of Conformity (DoC) is our claim of compliance with the
Council of the European Communities using the manufacturer’s
declaration of conformity. This system affords the user protection for
electronic compatibility (EMC) and product safety. You can obtain the DoC
for your product by visiting
supports calibration, you can obtain the calibration certificate for your
product at
National Instruments corporate headquarters is located at
11500 North Mopac Expressway, Austin, Texas, 78759-3504.
National Instruments also has offices located around the world to help
address your support needs. For telephone support in the United States,
create your service request at
instructions or dial 512 795 8248. For telephone support outside the United
States, contact your local branch office:
National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation.
Refer to the Terms of Use section on ni.com/legal for more information about National
Instruments trademarks. Other product and company names mentioned herein are trademarks or trad e
names of their respective companies. For patents covering National Instruments products, refer to the
appropriate location: Help»Patents in your software, the patents.txt file on your CD, or
ni.com/patents.