Worldwide Technical Support and Product Information
ni.com
National Instruments Corporate Headquarters
11500 North Mopac Expressway Austin, Texas 78759-3504 USA Tel: 512 683 0100
Worldwide Offices
Australia 1800 300 800, Austria 43 662 457990-0, Belgium 32 (0) 2 757 0020, Brazil 55 11 3262 3599,
Canada 800 433 3488, China 86 21 5050 9800, Czech Republic 420 224 235 774, Denmark 45 45 76 26 00,
Finland 358 (0) 9 725 72511, France 01 57 66 24 24, Germany 49 89 7413130, India 91 80 41190000,
Israel 972 3 6393737, Italy 39 02 41309277, Japan 0120-527196, Korea 82 02 3451 3400,
Lebanon 961 (0) 1 33 28 28, Malaysia 1800 887710, Mexico 01 800 010 0793, Netherlands 31 (0) 348 433 466,
New Zealand 0800 553 322, Norway 47 (0) 66 90 76 60, Poland 48 22 3390150, Portugal 351 210 311 210,
Russia 7 495 783 6851, Singapore 1800 226 5886, Slovenia 386 3 425 42 00, South Africa 27 0 11 805 8197,
Spain 34 91 640 0085, Sweden 46 (0) 8 587 895 00, Switzerland 41 56 2005151, Taiwan 886 02 2377 2222,
Thailand 662 278 6777, Turkey 90 212 279 3031, United Kingdom 44 (0) 1635 523545
For further support information, see the Technical Support and Professional Servicesappendix. To comment on
National Instruments documentation, refer to the National Instruments Web site at ni.com/info and enter the
info code feedback.
The media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects
in materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other documentation. National
Instruments will, at its option, repair or replace software media that do not execute programming instructions if National Instruments receives
notice of such defects during the warranty period. National Instruments does not warrant that the operation of the software shall be
uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before any
equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are covered by
warranty.
National Instruments believes that the information in this document is accurate. The document has been carefully reviewed for technical accuracy. In
the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent editions of this document
without prior notice to holders of this edition. The reader should consult National Instruments if errors are suspected. In no event shall National
Instruments be liable for any damages arising out of or related to this document or the information contained in it.
E
XCEPTASSPECIFIEDHEREIN, NATIONAL INSTRUMENTSMAKESNOWARRANTIES, EXPRESSORIMPLIED, ANDSPECIFICALLYDISCLAIMSANYWARRANTYOF
MERCHANTABILITYORFITNESSFORAPARTICULARPURPOSE. CUSTOMER’SRIGHTTORECOVERDAMAGESCAUSEDBYFAULTORNEGLIGENCEONTHEPARTOF NATIONAL
I
NSTRUMENTSSHALLBELIMITEDTOTHEAMOUNTTHERETOFOREPAIDBYTHECUSTOMER. NATIONAL INSTRUMENTSWILLNOTBELIABLEFORDAMAGESRESULTING
FROMLOSSOFDATA, PROFITS, USEOFPRODUCTS, ORINCIDENTALORCONSEQUENTIALDAMAGES, EVENIFADVISEDOFTHEPOSSIBILITYTHEREOF. This limitation of
the liability of National Instruments will apply regardless of the form of action, whether in contract or tort, including negligence. Any action against
National Instruments must be brought within one year after the cause of action accrues. National Instruments shall not be liable for any delay in
performance due to causes beyond its reasonable control. The warranty provided herein does not cover damages, defects, malfunctions, or service
failures caused by owner’s failure to follow the National Instruments installation, operation, or maintenance instructions; owner’s modification of the
product; owner’s abuse, misuse, or negligent acts; and power failure or surges, fire, flood, accident, actions of third parties, or other events outside
reasonable control.
Copyright
Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical, including photocopying,
recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National
Instruments Corporation.
National Instruments respects the intellectual property of others, and we ask our users to do the same. NI software is protected by copyright and other
intellectual property laws. Where NI software may be used to reproduce software or other materials belonging to others, you may use NI software only
to reproduce materials that you may reproduce in accordance with the terms of any applicable license or other legal restriction.
Trademarks
National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation. Refer to the Terms of Use section
on ni.com/legal for more information about National Instruments trademarks.
Other product and company names mentioned herein are trademarks or trade names of their respective companies.
Members of the National Instruments Alliance Partner Program are business entities independent from National Instruments and have no agency,
partnership, or joint-venture relationship with National Instruments.
Patents
For patents covering National Instruments products/technology, refer to the appropriate location: Help»Patents in your software,
the
patents.txt file on your media, or the National Instruments Patent Notice at ni.com/patents.
WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS
(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING FOR A LEVEL OF
RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL IMPLANTS OR AS CRITICAL COMPONENTS IN
ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE TO PERFORM CAN REASONABLY BE EXPECTED TO CAUSE SIGNIFICANT
INJURY TO A HUMAN.
(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE SOFTWARE PRODUCTS CAN BE
IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT LIMITED TO FLUCTUATIONS IN ELECTRICAL POWER SUPPLY,
COMPUTER HARDWARE MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE FITNESS, FITNESS OF COMPILERS
AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION, INSTALLATION ERRORS, SOFTWARE AND HARDWARE
COMPATIBILITY PROBLEMS, MALFUNCTIONS OR FAILURES OF ELECTRONIC MONITORING OR CONTROL DEVICES,
TRANSIENT FAILURES OF ELECTRONIC SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR MISUSES, OR
ERRORS ON THE PART OF THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS THESE ARE HEREAFTER
COLLECTIVELY TERMED “SYSTEM FAILURES”). ANY APPLICATION WHERE A SYSTEM FAILURE WOULD CREATE A RISK OF
HARM TO PROPERTY OR PERSONS (INCLUDING THE RISK OF BODILY INJURY AND DEATH) SHOULD NOT BE RELIANT SOLELY
UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM FAILURE. TO AVOID DAMAGE, INJURY, OR DEATH,
THE USER OR APPLICATION DESIGNER MUST TAKE REASONABLY PRUDENT STEPS TO PROTECT AGAINST SYSTEM FAILURES,
INCLUDING BUT NOT LIMITED TO BACK-UP OR SHUT DOWN MECHANISMS. BECAUSE EACH END-USER SYSTEM IS
CUSTOMIZED AND DIFFERS FROM NATIONAL INSTRUMENTS' TESTING PLATFORMS AND BECAUSE A USER OR APPLICATION
DESIGNER MAY USE NATIONAL INSTRUMENTS PRODUCTS IN COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT
EVALUATED OR CONTEMPLATED BY NATIONAL INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS ULTIMATELY
RESPONSIBLE FOR VERIFYING AND VALIDATING THE SUITABILITY OF NATIONAL INSTRUMENTS PRODUCTS WHENEVER
NATIONAL INSTRUMENTS PRODUCTS ARE INCORPORATED IN A SYSTEM OR APPLICATION, INCLUDING, WITHOUT
LIMITATION, THE APPROPRIATE DESIGN, PROCESS AND SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.
Contents
About This Manual
Using the NI-488.2 Documentation...............................................................................xi
Accessing the NI-488.2 Help ..........................................................................xii
This manual describes the features and functions of the NI-488.2 software.
You can use the NI-488.2 software with Windows 2000/XP/Vista,
Mac OS X, and Linux.
Using the NI-488.2 Documentation
The following NI-488.2 documentation is available with your NI-488.2
software distribution CD:
•The Getting Started/Installation Guide briefly describes how to install
the NI-488.2 software and your GPIB hardware.
•This manual describes the features and functionality of the NI-488.2
software.
•The GPIB Hardware Guide contains detailed instructions on how to
install and configure your GPIB hardware. This guide also includes
hardware specifications and compliance information.
To view these documents, you need Adobe Acrobat Reader, which you can
download from
www.adobe.com.
Windows
To view these documents online, insert your NI-488.2 software distribution
CD and select the View Documentation option. The View Documentation
utility helps you find the documentation that you want to view. You can also
view these documents at
ni.com.
Mac OS X
To view these documents online, insert your NI-488.2 software distribution
CD and open the Documentation folder. You can also view these
documents at
ni.com.
Linux
To view these documents online, insert your NI-488.2 software distribution
CD and browse to the
documents at
Run GPIB Explorer by entering the following command:
<InstallDir>
/natinst/ni4882/bin/gpibexplorer
<InstallDir>
software. The default is
Select Help»Help Topics»NI-488.2.
is the directory where you chose to install the NI-488.2
/usr/local.
Conventions
The following conventions appear in this manual:
»The » symbol leads you through nested menu items and dialog box options
to a final action. The sequence File»Page Setup»Options directs you to
pull down the File menu, select the Page Setup item, and select Options
from the last dialog box.
This icon denotes a note, which alerts you to important information.
boldBold text denotes items that you must select or click in the software, such
as menu items and dialog box options. Bold text also denotes parameter
names.
IEEE 488 andIEEE 488 and IEEE 488.2 refer to the ANSI/IEEE Standard 488.1-2003
IEEE 488.2and the ANSI/IEEE Standard 488.2-1992, respectively, which define the
GPIB.
NI-488.2 User Manualxiini.com
About This Manual
italicItalic text denotes variables, emphasis, a cross-reference, or an introduction
to a key concept. Italic text also denotes text that is a placeholder for a word
or value that you must supply.
monospaceText in this font denotes text or characters that you should enter from the
keyboard, sections of code, programming examples, and syntax examples.
This font is also used for the proper names of disk drives, paths, directories,
programs, subprograms, subroutines, device names, functions, operations,
variables, filenames, and extensions.
monospace boldBold text in this font denotes the messages and responses that the computer
automatically prints to the screen. This font also emphasizes lines of code
that are different from the other examples.
monospace italic
PlatformText in this font denotes a specific platform and indicates that the text
Italic text in this font denotes text that is a placeholder for a word or value
that you must supply.
following it applies only to that platform.
Related Documentation
The following documents contain information that you may find helpful as
you read this manual:
•ANSI/IEEE Standard 488.1-2003, IEEE Standard Digital Interface
for Programmable Instrumentation
•ANSI/IEEE Standard 488.2-1992, IEEE Standard Codes, Formats,
Protocols, and Common Commands
This chapter describes how to set up your GPIB system.
Setting Up and Configuring Your System
Devices are usually connected with a cable assembly consisting of a
shielded 24-conductor cable with both a plug and receptacle connector at
each end. With this design, you can link devices in a linear configuration,
a star configuration, or a combination of the two configurations. Figure 1-1
shows the linear and star configurations.
Figure 1-2 shows an example of a multi-interface system configuration.
gpib0 is a PCI-GPIB and is the access interface for the voltmeter. gpib1
is a GPIB-ENET/100 and is the access interface for the plotter and printer.
One
GPIB
Another
GPIB
Digital
Voltometer
Plotter
gpib0
LINK
PWR/RDY
10/100
ENET
gpib1
TX RX TALK LISTEN
GPIB
GPIB-ENET/100
I
0
Printer
Figure 1-2. Example of Multiboard System Configuration
Configuration Requirements
To achieve the high data transfer rate that the GPIB was designed for,
you must limit the number of devices on the bus and the physical distance
between devices. The following restrictions are typical:
•A maximum separation of 4 m between any two devices and an
average separation of 2 m over the entire bus.
•A maximum total cable length of 20 m.
•A maximum of 15 devices or controllers connected to each bus,
with at least two-thirds powered on.
NI-488.2 User Manual1-2ni.com
Chapter 1Introduction
For high-speed (HS488) operation, the following restrictions apply:
•All devices in the system must be powered on.
•Cable lengths must be as short as possible with up to a maximum
of 15 m of cable for each system.
•There must be at least one equivalent device load per meter of cable.
If you want to exceed these limitations, you can use a bus extender to
increase the cable length or a bus expander to increase the number of
device loads. You can order bus extenders and expanders from
National Instruments.
Chapter 2Measurement & Automation Explorer (Windows)
Add a New GPIB Interface
For plug-and-play interfaces (such as PCI or USB), the system
automatically detects and installs the hardware.
To add a new GPIB-ENET/100 to your system, complete the following
steps:
1.Start Measurement & Automation Explorer as described in the
Starting Measurement & Automation Explorer section.
2.Right-click on the Devices and Interfaces folder and select
Create New.
3.In the Create New dialog window, select the GPIB-ENET/100 item
and click Finish.
The Add GPIB-ENET/100 Wizard appears.
4.Follow the prompts in the Add GPIB-ENET/100 Wizard to add your
interface.
5.Measurement & Automation Explorer automatically updates the list of
installed GPIB interfaces. You also can select View»Refresh to update
the list.
Remove a GPIB Interface
To remove a Plug and Play interface from your computer, disconnect it,
making sure to turn off the computer if the interface requires it.
To remove a GPIB-ENET/100 interface from your computer, you must
manually delete it from within Measurement & Automation Explorer by
completing the following steps:
1.Start Measurement & Automation Explorer as described in the
Starting Measurement & Automation Explorer section.
2.Expand Devices and Interfaces by clicking the + next to the folder.
3.Right-click on your GPIB-ENET/100 interface and select Delete from
the context menu.
4.When prompted, confirm your selection.
5.Select View»Refresh to update the list of interfaces in
Measurement & Automation Explorer.
NI-488.2 User Manual2-4ni.com
Scan for GPIB Instruments
To scan for instruments connected to your GPIB interface or to add a new
instrument to your system, complete the following steps:
1.Make sure that your instrument is powered on and connected to your
GPIB interface.
2.Start Measurement & Automation Explorer as described in the
Starting Measurement & Automation Explorer section.
3.Expand the Devices and Interfaces directory by clicking the + next
to the folder.
4.Right-click on your GPIB interface and select Scan for Instruments
from the drop-down menu that appears.
Measurement & Automation Explorer displays the connected
instruments in the right window pane.
Instruments Not Found
If Measurement & Automation Explorer reports that it did not find any
instruments, make sure that your GPIB instruments are powered on and
properly connected to the GPIB interface with a GPIB cable. Then, scan for
instruments again, as described in the Scan for GPIB Instruments section.
Chapter 2Measurement & Automation Explorer (Windows)
Too Many Listeners on the GPIB
If Measurement & Automation Explorer reports that it found too many
Listeners on the GPIB, refer to the following possible solutions:
•If you have a running GPIB Analyzer with the GPIB handshake option
enabled, disable the GPIB handshake option in the GPIB Analyzer.
•If you have a GPIB extender in your system, Measurement &
Automation Explorer cannot detect any instruments connected to
your GPIB interface. Instead, you can verify communication with
your instruments using the Interactive Control utility. To do so,
select Tools»NI-488.2»Interactive Control. For more
information about verifying instrument communication, type
at the Interactive Control command prompt.
Chapter 2Measurement & Automation Explorer (Windows)
Communicate with Your Instrument
To establish basic or advanced communication with your instruments,
refer to the following sections.
For more information about instrument communication and a list of the
commands that your instrument understands, refer to the documentation
that came with your GPIB instrument. Most instruments respond to the
*IDN? command by returning an identification string.
Basic Communication (Query/Write/Read)
To establish basic communication with your instrument, use the NI-488.2
Communicator, as follows:
1.Start Measurement & Automation Explorer as described in the
Starting Measurement & Automation Explorer section.
2.Expand the Devices and Interfaces directory by clicking the + next
to the folder.
3.Select your GPIB interface.
4.If you have not already done so, scan for connected instruments.
Right-click on your GPIB interface and select Scan for Instruments
from the drop-down menu that appears. Refer to the Scan for GPIB
Instruments section for more information.
Measurement & Automation Explorer displays the connected
instruments below your GPIB interface.
5.Right-click on your GPIB instrument in the left window pane and
select Communicate with Instrument from the drop-down menu that
appears.
NI-488.2 User Manual2-6ni.com
Chapter 2Measurement & Automation Explorer (Windows)
The NI-488.2 Communicator dialog box appears, as shown
in Figure 2-3.
Figure 2-3. NI-488.2 Communicator
6.Type a command in the Send String field and do one of the following:
•To write a command to the instrument then read a response back,
click the Query button.
•To write a command to the instrument, click the Write button.
•To read a response from the instrument, click the Read button.
To view sample C/C++ code that performs a simple query of a GPIB
instrument, click the Show Sample button.
Advanced Communication
For advanced interactive communication with GPIB instruments, use the
Interactive Control utility, as follows:
1.Start Measurement & Automation Explorer as described in the
Starting Measurement & Automation Explorer section.
2.Expand the Devices and Interfaces directory by clicking the + next
to the folder.
3.Right-click on your GPIB interface and select Interactive Control
from the drop-down menu that appears. Interactive Control
automatically opens a session to the selected GPIB interface.
4.At the command prompt, type NI-488.2 API calls to communicate
interactively with the your instrument. For example, you might use
Chapter 2Measurement & Automation Explorer (Windows)
To view the online help for Interactive Control, type help at the Interactive
Control command prompt. For more information on using this utility, refer
to Chapter 7, Interactive Control Utility.
View NI-488.2 Software Version
To view the NI-488.2 software version, complete the following steps:
1.Start Measurement & Automation Explorer as described in the
Starting Measurement & Automation Explorer section.
2.Expand the Software directory by clicking the + next to the folder.
3.Click NI-488.2.
Measurement & Automation Explorer displays the version number
of the NI-488.2 software in the right window pane.
Monitor, Record, and Display NI-488.2 Calls
To monitor NI-488.2 calls, use NI Spy, as follows:
1.Start Measurement & Automation Explorer as described in the
Starting Measurement & Automation Explorer section.
2.Expand the Devices and Interfaces directory by clicking the + next
to the folder.
3.Select your GPIB interface in the left window pane of Measurement &
Automation Explorer.
4.On the toolbar, click the NI Spy button to open the application.
5.On the NI Spy toolbar, click the blue arrow button to start a capture.
NI-488.2 User Manual2-8ni.com
Chapter 2Measurement & Automation Explorer (Windows)
6.Start the NI-488.2 application that you want to monitor.
NI Spy records and displays all NI-488.2 calls, as shown in Figure 2-4.
Figure 2-4. NI-488.2 Calls Recorded by NI Spy
For more information about using NI Spy, select Help»Help Topics
in NI Spy or refer to Chapter 6, NI Spy Utility.
View or Change GPIB Interface Settings
The default behavior of NI-488.2 software is to allow only administrators
to make changes to the GPIB interface settings. You can change this
behavior by following the instructions described in the Change GPIB
Permissions section.
To view or change your interface settings, complete the following steps:
1.Start Measurement & Automation Explorer as described in the
Starting Measurement & Automation Explorer section.
Note You must choose to run as Adminstrator to allow changes to GPIB Interface
Settings. To do this, right-click on the Measurement & Automation Explorer icon and
select Run as Administrator.
2.Expand the Devices and Interfaces directory by clicking the + next
to the folder.
Chapter 2Measurement & Automation Explorer (Windows)
3.Select your GPIB interface in the left window pane of Measurement &
Automation Explorer.
The interface properties appear in the right window pane of
Measurement & Automation Explorer, as shown in Figure 2-5.
Figure 2-5. GPIB Interface Properties in Measurement & Automation Explorer
4.(Optional) Change the settings for your interface and click Save to
apply the settings.
Change GPIB Permissions
NI-488.2 allows administrators to allow nonadministrators to make
changes to GPIB settings. This is controlled using the Measurement &
Automation Explorer menu item at Tools»NI-488.2»Require
Administrator Privileges. This menu is enabled only if you run MAX as
an administrator. In either case, a check next to the menu indicates that you
can change GPIB settings only when you run MAX as an administrator.
NI-488.2 User Manual2-10ni.com
Chapter 2Measurement & Automation Explorer (Windows)
To toggle between the different modes, complete the following steps:
1.Start MAX as an administrator as described in the Starting
Measurement & Automation Explorer section.
Note You must choose to run as Adminstrator to run in administrator mode. To do this,
right-click on the Measurement & Automation Explorer icon and select Run as
Administrator.
2.Select Tools»NI-488.2»Require Administrator Privileges from the
menu.
3.Enable or disable the requirement in the NI-488.2 Settings dialog box
and click OK.
View GPIB Instrumentation Information
To view information about your GPIB instruments, complete the
following steps:
1.Start Measurement & Automation Explorer as described in the
Starting Measurement & Automation Explorer section.
2.Expand the Devices and Interfaces directory by clicking the + next
to the folder.
3.Select your GPIB interface.
Measurement & Automation Explorer displays the connected
instruments in the right window pane.
4.If you have not already done so, scan for connected instruments.
Right-click on your GPIB interface and select Scan for Instruments
from the drop-down menu that appears. Refer to the Scan for GPIB
Instruments section earlier in this chapter for more information.
5.Double-click on the instrument displayed in the right window pane.
Measurement & Automation Explorer lists all the attributes for
the instrument, such as the primary address, the secondary address
(if applicable), the instrument’s response to the identification query
(
*IDN?), and the GPIB interface number to which the device is
Chapter 2Measurement & Automation Explorer (Windows)
Change GPIB Device Templates
For older NI-488.2 applications, you might need to modify one of the
device templates to find a given GPIB instrument by name, for example,
ibfind("fluke45"). Older applications still use ibfind instead of the
preferred
ibfind to obtain device handles and use ibdev instead. You can use
ibdev to dynamically configure your GPIB device handle. ibdev also
eliminates unneccessary device name requirements.
If you must modify a device template, run the GPIB Configuration utility.
1.Start Measurement & Automation Explorer as described in the
2.Select Help»Help Topics»NI-488.2 to view the NI-488.2 Help.
3.Search for the topic named How do I change a GPIB Device Template?
4.Double-click the device template you want to modify, such as DEV1.
5.Rename the template as described in your application documentation.
6.Click the OK button twice to save your changes and exit.
ibdev to obtain a device handle. In new applications, avoid using
Starting Measurement & Automation Explorer section.
and click the link to open the GPIB Configuration utility.
Enable/Disable NI-488.2 DOS Support
Note NI-488.2 DOS support is not available on Windows Vista or later operating systems.
To enable or disable NI-488.2 DOS support, complete the following steps:
1.Start Measurement & Automation Explorer as described in the
Starting Measurement & Automation Explorer section.
2.Click on your GPIB interface and select Tools»NI-488.2»
DOS Support... from the Explorer menu.
3.Enable or disable DOS support in the NI-488.2 Settings dialog box
and click OK.
4.If you are prompted to do so, restart your system.
NI-488.2 User Manual2-12ni.com
Chapter 2Measurement & Automation Explorer (Windows)
Access Additional Help and Resources
To access additional help and resources for the NI-488.2 software and your
GPIB hardware, refer to the following sections.
NI-488.2 Help
The NI-488.2 Help addresses questions you might have about NI-488.2 and
includes a function reference and troubleshooting information. You can
access the NI-488.2 Help as follows:
1.Start Measurement & Automation Explorer as described in the
Starting Measurement & Automation Explorer section.
2.Select Help»Help Topics»NI-488.2.
National Instruments GPIB Web Site
You can access the National Instruments GPIB Web site as follows:
1.Start Measurement & Automation Explorer as described in the
Starting Measurement & Automation Explorer section.
2.Select Help»National Instruments on the Web»GPIB Home Page.
View or Change GPIB-ENET/100 Network Settings
To view or change the network settings of your GPIB-ENET/100, refer to
the following sections. For more information about your GPIB-ENET/100
network settings, refer to the GPIB-ENET/100 information in the GPIB Hardware Guide.
Device Configuration
Use the NI Ethernet Device Configuration utility if you need to manually
configure the network parameters of the GPIB-ENET/100. If your network
uses DHCP, the network configuration is performed automatically at
startup and you do not need to run this utility unless you want to change the
hostname. Consult your network administrator if you do not know whether
your network uses DHCP.
1.Start Measurement & Automation Explorer as described in the
Starting Measurement & Automation Explorer section.
2.Expand the Devices and Interfaces directory by clicking the + next
to the folder.