This document contains step-by-step instructions for writing an external
calibration procedure for the National Instruments PXI/PCI-4070 and
NI PXI-4072 digital multimeters (DMMs). Each of these
National Instruments DMMs is a 6½-digit FlexDMM and 1.8 MS/s
Contents
isolated digitizer. For more information on calibration, visit
Where to Go for Support......................................................................... 68
.
ni.com/
™
Conventions
The following conventions are used in this document:
»The » symbol leads you through nested menu items and dialog box options
to a final action. The sequence File»Page Setup»Options directs you to
pull down the File menu, select the Page Setup item, and select Options
from the last dialog box.
♦The ♦ symbol indicates that the following text applies only to a specific
product, a specific operating system, or a specific software version.
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to
avoid injury, data loss, or a system crash. When this symbol is marked on a
product, refer to the Read Me First: Safety and Radio-Frequency Interference document included with the device for information about
precautions to take.
boldBold text denotes items that you must select or click in the software, such
as menu items and dialog box options. Bold text also denotes parameter
names.
italicItalic text denotes variables, emphasis, a cross-reference, hardware labels,
or an introduction to a key concept. Italic text also denotes text that is a
placeholder for a word or value that you must supply.
monospaceText in this font denotes text or characters that you should enter from the
keyboard, sections of code, programming examples, and syntax examples.
This font is also used for the proper names of disk drives, paths, directories,
programs, subprograms, subroutines, device names, functions, operations,
variables, filenames, and extensions.
Software Requirements
NI-DMM supports a number of programming languages including
LabVIEW, LabWindows
Visual Basic. When you install NI-DMM, you need to install support for
only the language you intend to use to write your calibration utility.
Note NI-DMM version 2.1 or later supports NI PXI-4070 calibration, NI-DMM
version 2.2 or later supports NI PCI-4070 calibration, and NI-DMM version 2.3 or later
supports NI 4072 calibration.
NI 4070/4072 Calibration Procedure2ni.com
™
/CVI™, Microsoft Visual C++, and Microsoft
The procedures in this document are described using C function calls.
You also can program in LabVIEW using the VIs that correspond to the
C function calls.
Documentation Requirements
In addition to this calibration document, you may find the following
references helpful in writing your calibration utility. All of these
documents are installed on your computer when you install NI-DMM.
To locate them, select Start»All Programs»National Instruments»NI-DMM»Documentation.
•NI Digital Multimeters Help
•NI Digital Multimeters Getting Started Guide
NI recommends referring to the following document online at
manuals
specifications:
•NI 4070/4072 Specifications
You may need the following documents, which are available at
manuals
•TB-2715 Terminal Block Installation Guide
•About Your NI 6608 Device
to ensure that you are using the latest NI 4070/4072
, to perform the optional frequency verification procedure:
Calibration Function Reference
For detailed information about the NI-DMM calibration functions used in
this procedure, refer to the LabVIEW Reference or the C/CVI/VB Reference
sections of the NI Digital Multimeters Help, located at Start»All Programs»National Instruments»NI-DMM»Documentation.
Password
The default calibration password in NI-DMM is "NI".
The accuracy requirements of your measurement application determine
how often you should calibrate the NI 4070/4072. NI recommends
performing a complete calibration at least once every two years. NI does
not guarantee the absolute accuracy of the NI 4070/4072 beyond this
two-year calibration interval. You can shorten the calibration interval based
on the demands of your application. Refer to Appendix A: Calibration
Options for more information.
Test Equipment
This section describes the required and optional equipment for calibration.
Required Test Equipment
Requirements for All NI 4070/4072 Devices
The following equipment is required for calibrating the NI 4070/4072:
•Fluke 5700A multifunction calibrator calibrated within the last
90 days, or a Fluke 5720A multifunction calibrator calibrated within
the last year
•Two sets of Fluke 5440 low thermal electromotive force (EMF) copper
cables
•Pomona 5145 insulated double banana plug shorting bar (or another
means of creating a short with low thermal EMF (≤150 nV) across the
HI and LO input banana plug connectors on the NI 4070/4072)
•Two Pomona B-4 banana-to-banana patch cords (cables) or similar
banana-to-banana cables with length not to exceed 4 in.
•National Instruments PXI chassis and controller, or a personal
computer (PC) with an available slot for the NI 4070/4072
Additional Requirements for the NI 4072
The following equipment is required for calibrating the capacitance and
inductance modes of the NI 4072:
•25Ω, 125 Ω, 5kΩ, and 100 kΩ resistors with thermal drift ≤5ppm/°C
and tolerance ≤1%. The distance between the resistor leads and the
NI 4072 terminals should be ≤1 in.
•Verification capacitors calibrated to at least four times the accuracy of
the NI 4072, with temperature coefficients ≤250 ppm/°C. The values
of the verification capacitors should cover the complete capacitance
range. NI suggests using traceable capacitor standards with values
≥10% of full range for all ranges, except the 300 pF range. For the
NI 4070/4072 Calibration Procedure4ni.com
300 pF range, a capacitor with values between 90–100% of full scale
should be used. NI suggests using the capacitance standards of the
SCA Series from IET Labs. This calibration procedure assumes the use
of 270pF, 1nF, 100nF, 10μF, and 1000 μF standards.
•If you are using cables to connect the verification capacitors to the
NI 4072 banana plug connectors, NI recommends using Pasternack
PE3005 banana-to-banana coaxial cables with length ≤4 inches and
total capacitance ≤40 pF. Before performing the verification
procedure, you should know the total capacitance up to the end of the
banana connectors that plug into the NI 4072.
Optional Test Equipment
The following equipment is optional for calibrating the NI 4070/4072 and
is only used for frequency verification:
•NI PXI-6608 timing and digital I/O module
•National Instruments SH68-68-D1 shielded cable
•National Instruments TB-2715 terminal block
•Pomona MDP 4892 double banana plug with strain relief
•Coaxial cable (for example, RG178)
Test Conditions
Follow these guidelines to optimize the connections and the environment
during calibration:
•Ensure that the PXI chassis fan speed is set to HI (if calibrating the
NI PXI-4070/4072) and that the fan filters are clean.
•Use PXI filler panels in all vacant slots to allow proper cooling.
•Plug the PXI chassis or PC and the calibrator into the same power strip
to avoid ground loops.
•Power on and warm up both the calibrator and the NI 4070/4072 for at
least 60 minutes before beginning this calibration procedure.
•Maintain an ambient temperature of 23 ±1 °C.
•Maintain an ambient relative humidity of less than 60%.
•Allow the calibrator to settle fully before taking any measurements.
Consult the Fluke 5700A/5720A user documentation for instructions.
•Allow the thermal EMF enough time to stabilize when you change
connections to the calibrator or the NI 4070/4072. The suggested time
periods are stated where necessary throughout this document.
•Keep a shorting bar connected between the VGUARD and
GROUND binding posts of the calibrator at all times.
•Clean any oxidation from the banana plugs on the Fluke 5440 cables
before plugging them into the binding posts of the calibrator or the
banana plug connectors of the NI 4070/4072. Oxidation tarnishes the
copper banana plugs so that they appear dull rather than shiny and
leads to greater thermal EMF.
•Keep the blue banana plugs on the Fluke 5440 cables connected to the
V GUARD binding post of the calibrator at all times.
•Prevent the cables from moving or vibrating by taping or strapping
them to a nonvibrating surface. Movement or vibration causes
triboelectric effects that can result in measurement errors.
Calibration Procedures
The calibration process includes the following steps:
1.Initial Setup—Set up the test equipment.
2.Verification Procedures—Verify the existing operation of the device.
This step confirms whether the device is operating within its specified
range prior to calibration. Figure 4 shows the procedural flow for
verification.
3.Adjustment Procedures—Submit the device to NI for a factory
calibration to adjust the calibration constants. Figure 5 shows the
procedural flow for adjustment.
4.Reverification—Repeat the verification procedure to ensure that the
device is operating within its specifications after adjustment.
These steps are described in more detail in the following sections.
Note In some cases, the complete calibration procedure may not be required. Refer to
Appendix A: Calibration Options for more information.
Initial Setup
Note This section is necessary for pre-adjustment verifications only. If you are performing
a post-adjustment verification, skip the setup and go directly to the Verifying DC Voltage
section.
To set up the test equipment, complete the following steps:
1.Remove all connections from the four input banana plug connectors on
the NI 4070/4072.
2.Verify that the calibrator has been calibrated within the time limits
specified in the Required Test Equipment section, and that DC zeros
calibration has been performed within the last 30 days. Consult the
NI 4070/4072 Calibration Procedure6ni.com
Fluke 5700A/5720A user documentation for instructions on
calibrating these devices.
Note Ensure that both the calibrator and the NI 4070/4072 (installed in a powered-on
PXI chassis or PC) are warmed up for at least 60 minutes before you begin this procedure.
3.Call
Note You use this session in all subsequent function calls throughout the verification
procedures.
4.Call
Verification Procedures
You can use the verification procedures described in this section for both
pre-adjustment and post-adjustment verification. The steps of each
verification procedure must be performed in the order listed; however, you
can omit entire sections (for example, the entire Verifying AC Current
section), if necessary.
The parameters Range, Resolution, and Sample Interval used in function
calls throughout this section have floating point values. For example, if
Range =
Sample Count, Array Size, and ParamValue have integer values. Refer
to the NI Digital Multimeters Help for more information about parameter
values.
niDMM_init with the resource name of the device to create a
session.
For more information on using
niDMM_init, refer to the NI Digital
Multimeters Help.
niDMM_SelfCal. This step is optional if you have adjusted the
NI 4070/4072 within the last 24 hours and the temperature has
remained constant to within ±1 °C of the calibration temperature (T
1, the floating point value is 1.0. The parameters Trigger Count,
cal
).
Note Many of the parameter values listed in this document are expressed in scientific
notation. Some programming languages do not support the direct entry of numbers in this
format. Be sure to properly enter these values with the appropriate number of zeros. For
example, enter the scientific notation number 10e–6 as
100000. If your programming language supports scientific notation, NI recommends that
0.00001 and the number 100e3 as
you use this feature to minimize possible data entry errors.
12. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
•Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
•Attribute_Value = NIDMM_VAL_10_MEGAOHM
13. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
14. Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_DC_VOLTS
•Range = 100
•Resolution = 100e–6
15. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
•Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
•Attribute_Value = NIDMM_VAL_10_MEGAOHM
16. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
17. Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_DC_VOLTS
•Range = 300
•Resolution = 300e–6
18. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
•Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
•Attribute_Value = NIDMM_VAL_10_MEGAOHM
19. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
20. Remove the shorting bar from the NI 4070/4072.
21. Reset the calibrator.
22. Fasten the connectors on one end of the Fluke 5440 cable to the
appropriate banana plug connectors of the NI 4070/4072, and fasten
the connectors on the other end of the cable to the appropriate
calibrator binding posts. Figure 1 shows the correct connections.
Table 1 lists the cable connections.
27. Call niDMM_ConfigureMultiPoint with the following parameters:
•Trigger Count =
1
•Sample Count = 10
•Sample Trigger = NIDMM_VAL_IMMEDIATE
•Sample Interval = –1
28. Call niDMM_ReadMultiPoint with the following parameters:
•Maximum Time =
NIDMM_VAL_TIME_LIMIT_AUTO
•Array Size = 10
Average the results by summing the returned reading array of the
function and dividing by the returned actual number of points. Store
the result as the 100 mV >10 GΩ mode offset.
29. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
•Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
•Attribute_Value = NIDMM_VAL_10_MEGAOHM
30. Call niDMM_ConfigureMultiPoint with the following parameters:
•Trigger Count =
1
•Sample Count = 10
•Sample Trigger = NIDMM_VAL_IMMEDIATE
•SampleInterval = –1
31. Call niDMM_ReadMultiPoint with the following parameters:
•Maximum Time =
NIDMM_VAL_TIME_LIMIT_AUTO
•Array Size = 10
Average the results by summing the returned reading array of the
function and dividing by the returned actual number of points. Store
the result as the 100 mV 10 MΩ mode offset.
32. Output 100 mV on the calibrator with the range locked to 2.2 V.
This range prevents a 50 Ω calibrator output resistance from creating
a voltage divider with the internal resistance of the NI 4070/4072.
niDMM_Read. Subtract the previously stored 100 mV >10 GΩ
mode offset from this measurement, and verify that the result falls
between the limits listed in Table 15.
36. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
•Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
•Attribute_Value = NIDMM_VAL_10_MEGAOHM
37. Call
niDMM_Read. Subtract the previously stored 100 mV 10 MΩ
mode offset from this measurement and verify that the result falls
between the limits listed in Table 15.
38. Output –100 mV on the calibrator with the range locked to 2.2 V.
This range prevents a 50 Ω calibrator output resistance from creating
a voltage divider with the internal resistance of the NI 4070/4072.
39. Set the input resistance of the NI 4070/4072 to >10 GΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
63. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
•Attribute_ID =
•Attribute_Value =
64. Call
niDMM_Read. Verify that this measurement falls between the
NIDMM_ATTR_INPUT_RESISTANCE
NIDMM_VAL_10_MEGAOHM
limits listed in Table 15.
65. Output 100 V on the calibrator.
Caution Avoid touching the connections when generating a high voltage from the
calibrator.
66. Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_DC_VOLTS
•Range = 100
•Resolution = 100e–6
67. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
•Attribute_ID =
•Attribute_Value =
NIDMM_ATTR_INPUT_RESISTANCE
NIDMM_VAL_10_MEGAOHM
68. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
69. Output –100 V on the calibrator.
70. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
NI 4070/4072 Calibration Procedure14ni.com
71. Call niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_DC_VOLTS
•Range = 300
•Resolution = 300e–6
72. Call niDMM_Read. Before you apply the voltage, the DMM must be in
the 300 V range.
73. Output 300 V on the calibrator.
74. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
75. Output –300 V on the calibrator.
76. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
77. Reset the calibrator for safety reasons.
You have completed verifying the DC voltage of the NI 4070/4072. Select
one of the following options:
•If you want to continue verifying other modes, go to the Verifying AC Voltage section.
•If you do not want to verify other modes and you are performing a
post-adjustment verification, go to the Completing the Adjustment
Procedures section.
•If you do not want to verify any additional modes and you are
performing a pre-adjustment verification, call
niDMM_close to close
the session.
Verifying AC Voltage
To verify AC voltage of the NI 4070/4072, complete the following steps:
1.Reset the calibrator.
2.Fasten the connectors on one end of the Fluke 5440 cable to the
appropriate banana plug connectors on the NI 4070/4072, and fasten
the connectors on the other end of the cable to the appropriate
calibrator binding posts. Figure 1 shows the correct connections.
Table 1 lists the cable connections.
You have completed verifying the AC voltage of the NI 4070/4072. Select
one of the following options:
•If you want to continue verifying other modes, go to the Verifying 4-Wire Resistance section.
•If you do not want to verify other modes and you are performing a
post-adjustment verification, go to the Completing the Adjustment
Procedures section.
•If you do not want to verify any additional modes and you are
performing a pre-adjustment verification, call
niDMM_close to close
the session.
Verifying 4-Wire Resistance
To verify the 4-wire resistance of the NI 4070/4072, complete the
following steps:
1.Reset the calibrator.
2.Fasten the connectors on one end of each Fluke 5440 cable to the
appropriate banana plug connectors on the NI 4070/4072. Fasten the
connectors on the other end of each Fluke 5440 cable to the appropriate
calibrator binding posts. Figure 2 shows the Fluke 5440 cables. Table 7
lists the cable connections.
1
VΩA
AUX
2
SENSE
VΩ
HI
LO
GUARDGROUND
6½-Digit FlexDMM
OUTPUT
HI
LO
HI
CURRENT
AUX
I/O
5V
MAX
300V
MAX
1A, 250V
MAX
300V
MAX
INPUT
V
AMPS
SENSE
W 4W
CAT II
HI
W
LO
HI
3
LO
1 NI 4070/40722 Fluke 5700A/5720A Calibrator3 Fluke 5440 Cables
Figure 2. Cable Connections for 4-Wire Resistance
NI 4070/4072 Calibration Procedure22ni.com
Table 7. Fluke 5440 Cable Connections
Fluke 5440
Cable Identification
First cableHIRedOUTPUT HI
Second cableHI SENSERedSENSE HI
Banana Plug Connector
(NI 4070/4072)
LOBlackOUTPUT LO
(No connection)BlueV GUARD
LO SENSEBlackSENSE LO
(No connection)BlueV GUARD
Banana Plug Color
(Fluke 5440 Cable)
(Fluke 5700A/5720A Calibrator)
3.Wait two minutes for the thermal EMF to stabilize if the Fluke 5440
cables were not previously connected in this configuration.
4.Call
niDMM_reset.
5.Refer to Table 8 for the appropriate calibrator output and function
parameter values as you complete the following steps:
a.On the calibrator, output the value listed in the Calibrator Output
column in Table 8 for the current iteration. Make sure that the
external sense is turned on but 2-wire compensation is turned off.
Note After setting the calibrator output to 0 Ω in the seventh iteration, you do not need to
continually set the calibrator to 0 Ω for iterations 8 through 12.
Binding Post
b.Call
niDMM_ConfigureMeasurement with the parameters set
as shown in Table 8 for the current iteration.
c.Call
niDMM_ConfigureOffsetCompOhms with
OffsetCompOhms set to either
NIDMM_VAL_OFFSET_COMP_OHMS_ON or
NIDMM_VAL_OFFSET_COMP_OHMS_OFF according to Table 8 for
the current iteration.
d.Call
niDMM_Read. Verify that this measurement falls between the
tolerances listed in Table 17. Tolerances are provided instead of
absolute limits because your calibrator will have different discrete
resistance values.
6.Repeat step 5 for each of the remaining iterations listed in Table 8.
Table 8. niDMM_ConfigureMeasurement Parameters
Calibrator
Iteration
110 MΩNIDMM_VAL_4_WIRE_RES10e610OFF
21MΩNIDMM_VAL_4_WIRE_RES1e61OFF
3100 kΩNIDMM_VAL_4_WIRE_RES100e30.1OFF
410 kΩNIDMM_VAL_4_WIRE_RES10e30.01ON
51kΩNIDMM_VAL_4_WIRE_RES1e31e–3ON
6100 ΩNIDMM_VAL_4_WIRE_RES100100e–6ON
70 ΩNIDMM_VAL_4_WIRE_RES10e610OFF
80 ΩNIDMM_VAL_4_WIRE_RES1e61OFF
90 ΩNIDMM_VAL_4_WIRE_RES100e30.1OFF
100 ΩNIDMM_VAL_4_WIRE_RES10e30.01ON
110 ΩNIDMM_VAL_4_WIRE_RES1e31e–3ON
120 ΩNIDMM_VAL_4_WIRE_RES100100e–6ON
Output
niDMM_ConfigureMeasurement Parameters
OffsetCompOhmsFunctionRangeResolution
You have completed verifying the 4-wire resistance of the NI 4070/4072.
Select one of the following options:
•If you want to continue verifying other modes, go to the Verifying 2-Wire Resistance section.
•If you do not want to verify other modes and you are performing a
post-adjustment verification, go to the Completing the Adjustment
Procedures section.
•If you do not want to verify any additional modes and you are
performing a pre-adjustment verification, call
niDMM_close to close
the session.
Verifying 2-Wire Resistance
To verify the 2-wire resistance of the NI 4070/4072, complete the
following steps:
1.Plug in the insulated banana plug shorting bar across the HI and LO
banana plug connectors on the NI 4070/4072.
2.Wait one minute for the thermal EMF to stabilize.
3.Call
NI 4070/4072 Calibration Procedure24ni.com
niDMM_reset.
4.Call niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_2_WIRE_RES
•Range = 10e3
•Resolution = 0.01
5.Call niDMM_ConfigureOffsetCompOhms with OffsetCompOhms
set to
NIDMM_VAL_OFFSET_COMP_OHMS_ON.
6.Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 18.
7.Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_2_WIRE_RES
•Range = 1e3
•Resolution = 1e–3
8.Call niDMM_ConfigureOffsetCompOhms with OffsetCompOhms
set to
NIDMM_VAL_OFFSET_COMP_OHMS_ON.
9.Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 18.
10. Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_2_WIRE_RES
•Range = 100
•Resolution = 100e–6
11. Call niDMM_ConfigureOffsetCompOhms with OffsetCompOhms
set to
NIDMM_VAL_OFFSET_COMP_OHMS_ON.
12. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 18.
13. Remove the shorting bar from the NI 4070/4072.
14. Reset the calibrator.
15. Fasten the connectors on one end of the Fluke 5440 cable to the
NI 4070/4072, and fasten the connectors on the other end of the cable
to the appropriate calibrator binding posts. Figure 1 shows the correct
connections. Table 1 lists the cable connections.
16. Wait two minutes for the thermal EMF to stabilize if the Fluke 5440
cable was not previously used in this configuration.
17. Output 0 Ω on the calibrator with 2-wire compensation turned on
but with external sense turned off.
18. Call
niDMM_reset to reset the NI 4070/4072 to a known state.
19. Call niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_2_WIRE_RES
•Range = 100e6
•Resolution = 100
20. Call niDMM_Read and store the result as the 100 MΩ range offset.
21. Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_2_WIRE_RES
•Range = 10e6
•Resolution = 10
22. Call niDMM_Read and store the result as the 10 MΩ range offset.
23. Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_2_WIRE_RES
•Range = 1e6
•Resolution = 1
24. Call niDMM_Read and store the result as the 1 MΩ range offset.
25. Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_2_WIRE_RES
•Range = 100e3
•Resolution = 0.1
26. Call niDMM_ConfigureMultiPoint with the following parameters:
•Trigger Count =
1
•Sample Count = 4
•Sample Trigger = NIDMM_VAL_IMMEDIATE
•Sample Interval = –1
27. Call niDMM_ReadMultiPoint with the following parameters:
•Maximum Time =
NIDMM_VAL_TIME_LIMIT_AUTO
•Array Size = 4
Average the results by summing the returned reading array of the
function and dividing by the returned actual number of points.
Store the result as the 100 kΩ range offset.
NI 4070/4072 Calibration Procedure26ni.com
28. Call niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_2_WIRE_RES
•Range = 10e3
•Resolution = 0.01
29. Call niDMM_ConfigureMultiPoint with the following parameters:
•Trigger Count =
1
•Sample Count = 4
•Sample Trigger = NIDMM_VAL_IMMEDIATE
•Sample Interval = –1
30. Call niDMM_ReadMultiPoint with the following parameters:
•Maximum Time =
NIDMM_VAL_TIME_LIMIT_AUTO
•Array Size = 4
Average the results by summing the returned reading array of the
function and dividing by the returned actual number of points.
Store the result as the 10 kΩ range offset.
31. Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_2_WIRE_RES
•Range = 1e3
•Resolution = 1e–3
32. Call niDMM_ConfigureMultiPoint with the following parameters:
•Trigger Count =
1
•Sample Count = 4
•Sample Trigger = NIDMM_VAL_IMMEDIATE
•Sample Interval = –1
33. Call niDMM_ReadMultiPoint with the following parameters:
•Maximum Time =
NIDMM_VAL_TIME_LIMIT_AUTO
•Array Size = 4
Average the results by summing the returned reading array of the
function and dividing by the returned actual number of points.
Store the result as the 1 kΩ range offset.
35. Call niDMM_ConfigureMultiPoint with the following parameters:
•Trigger Count =
1
•Sample Count = 10
•Sample Trigger = NIDMM_VAL_IMMEDIATE
•Sample Interval = –1
36. Call niDMM_ReadMultiPoint with the following parameters:
•Maximum Time =
NIDMM_VAL_TIME_LIMIT_AUTO
•Array Size = 10
Average the results by summing the returned reading array of the
function and dividing by the returned actual number of points.
Store the result as the 100 Ω range offset.
37. Output 100 MΩ on the calibrator without external sense or 2-wire
compensation.
38. Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_2_WIRE_RES
•Range = 100e6
•Resolution = 100
39. Call niDMM_Read. Subtract the previously stored 100 MΩ range offset
from this measurement. Verify that the result falls between the
tolerances listed in Table 18.
40. Output 10 MΩ on the calibrator without external sense or 2-wire
compensation.
41. Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_2_WIRE_RES
•Range = 10e6
•Resolution = 10
42. Call
niDMM_Read. Subtract the previously stored 10 MΩ range offset
from this measurement. Verify that the result falls between the
tolerances listed in Table 18.
43. Output 1 MΩ on the calibrator without external sense or 2-wire
compensation.
44. Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_2_WIRE_RES
•Range = 1e6
•Resolution = 1
NI 4070/4072 Calibration Procedure28ni.com
45. Call niDMM_Read. Subtract the previously stored 1 MΩ range offset
from this measurement. Verify that the result falls between the
tolerances listed in Table 18.
46. Output 100 kΩ on the calibrator without external sense or 2-wire
compensation.
47. Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_2_WIRE_RES
•Range = 100e3
•Resolution = 0.1
48. Call niDMM_Read. Subtract the previously stored 100 kΩ range offset
from this measurement. Verify that the result falls between the
tolerances listed in Table 18.
49. Output 10 kΩ on the calibrator with 2-wire compensation turned on
but with external sense turned off.
50. Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_2_WIRE_RES
•Range = 10e3
•Resolution = 0.01
51. Call niDMM_Read. Subtract the previously stored 10 kΩ range offset
from this measurement. Verify that the result falls between the
tolerances listed in Table 18.
52. Output 1 kΩ on the calibrator with 2-wire compensation turned on
but with external sense turned off.
53. Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_2_WIRE_RES
•Range = 1e3
•Resolution = 1e–3
54. Call niDMM_Read. Subtract the previously stored 1 kΩ range offset
from this measurement. Verify that the result falls between the
tolerances listed in Table 18.
55. Output 100 Ω on the calibrator with 2-wire compensation turned on
but with external sense turned off.
57. Call niDMM_Read. Subtract the previously calculated 100 Ω range
offset from this measurement. Verify that the result falls between the
tolerances listed in Table 18.
You have completed verifying the 2-wire resistance of the NI 4070/4072.
Select one of the following options:
•If you want to continue verifying other modes, go to the Verifying DC Current section.
•If you do not want to verify other modes and you are performing a
post-adjustment verification, go to the Completing the Adjustment
Procedures section.
•If you do not want to verify any additional modes and you are
performing a pre-adjustment verification, call
niDMM_close to close
the session.
Verifying DC Current
To verify the DC current of the NI 4070/4072, complete the following steps:
1.Reset the calibrator.
2.Fasten the connectors on one end of the Fluke 5440 cable to the
NI 4070/4072 HI SENSE and LO banana plug connectors, and connect
the connectors on the other end of the cable to the HI and LO calibrator
binding posts. Figure 3 shows the correct connections. Table 9 lists the
cable connections.
1
VΩA
AUX
2
SENSE
VΩ
HI
LO
GUARDGROUND
6½-Digit FlexDMM
OUTPUT
HI
LO
HI
CURRENT
AUX
I/O
5V
MAX
300V
MAX
1A, 250V
MAX
300V
MAX
INPUT
V
AMPS
SENSE
W 4W
CAT II
HI
W
LO
HI
LO
3
1 NI 4070/40722 Fluke 5700A/5720A Calibrator3 Fluke 5440 Cable
Figure 3. Cable Connections for Current
NI 4070/4072 Calibration Procedure30ni.com
Table 9. Fluke 5440 Cable Connections
Banana Plug Connector
(NI 4070/4072)
HI SENSERedOUTPUT HI
LOBlackOUTPUT LO
(No connection)BlueV GUARD
Banana Plug Color
(Fluke 5440 Cable)
(Fluke 5700A/5720A Calibrator)
3.Call niDMM_reset to reset the NI 4070/4072 to a known state.
4.Set the current output on the calibrator to NORM and output 0 A.
5.Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_DC_CURRENT
•Range = 0.02
•Resolution = 20e–9
6.Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table 19.
7.Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_DC_CURRENT
•Range = 0.2
•Resolution = 200e–9
8.Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table 19.
9.Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_DC_CURRENT
•Range = 1
•Resolution = 1e–6
10. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table 19.
11. Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_DC_CURRENT
•Range = 0.02
•Resolution = 20e–9
12. Call niDMM_Read to configure the NI 4070/4072 for a current mode
before applying current.
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 19.
15. Output –20 mA on the calibrator.
16. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 19.
17. Output 200 mA on the calibrator.
18. Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_DC_CURRENT
•Range = 0.2
•Resolution = 200e–9
19. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table 19.
20. Output –200 mA on the calibrator.
21. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 19.
22. Output 1 A on the calibrator.
23. Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_DC_CURRENT
•Range = 1
•Resolution =
1e–6
24. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table 19.
25. Output –1 A on the calibrator.
26. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 19.
You have completed verifying the DC current of the NI 4070/4072. Select
one of the following options:
•If you want to continue verifying other modes, go to the Verifying AC
Current section.
•If you do not want to verify other modes and you are performing a
post-adjustment verification, go to the Completing the Adjustment
Procedures section.
•If you do not want to verify any additional modes and you are
performing a pre-adjustment verification, call
niDMM_close to close
the session.
NI 4070/4072 Calibration Procedure32ni.com
Verifying AC Current
To verify the AC current of the NI 4070/4072, complete the following
steps:
1.Reset the calibrator.
2.Fasten the connectors on one end of the Fluke 5440 cable to the
NI 4070/4072 HI SENSE and LO banana plug connectors, and fasten
the connectors on the other end of the cable to the HI and LO calibrator
binding posts. Figure 3 shows the correct connections. Table 9 lists the
cable connections.
3.Call
4.Call
5.Call niDMM_Read to configure the NI 4070/4072 for a current mode
6.Output 1 mA at 1 kHz on the calibrator with the current output set
7.Call
8.Output 10 mA at 1 kHz on the calibrator.
9.Call
10. Call
11. Call niDMM_Read. Verify that this measurement falls between the
12. Output 100 mA at 1 kHz on the calibrator.
13. Call
niDMM_reset to reset the NI 4070/4072 to a known state.
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_AC_CURRENT
•Range = 0.01
•Resolution = 10e–9
before applying current.
to
NORM.
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 20.
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 20.
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_AC_CURRENT
•Range = 0.1
•Resolution = 100e–9
limits listed in Table 20.
niDMM_Read. Verify that this measurement falls between the
14. Call niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_AC_CURRENT
•Range = 1
•Resolution = 1e–6
15. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table 20.
16. Output 1 A at 1 kHz on the calibrator.
17. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 20.
You have completed verifying the AC current of the NI 4070/4072. Select
one of the following options:
•If you want to continue verifying other modes, go to the Verifying Frequency section.
•If you do not want to verify other modes and you are performing a
post-adjustment verification, go to the Completing the Adjustment
Procedures section.
•If you do not want to verify any additional modes and you are
performing a pre-adjustment verification, call
niDMM_close to close
the session.
Verifying Frequency
Notes The frequency of the NI 4070/4072 is not user adjustable. If this verification
procedure indicates that the frequency is out of specification, return the NI 4070/4072
to NI for repair.
This verification procedure is optional and requires additional test
equipment. If you do not want to verify frequency, select one of the
following options:
•If you are calibrating an NI 4072 and want to continue verifying other
modes, go to the Verifying Capacitance and Inductance (NI 4072
Only) section.
•If you do not want to verify other modes and are performing a
post-adjustment verification, go to the Completing the Adjustment
Procedures section.
•If you do not want to verify any additional modes and you are
performing a pre-adjustment verification, call
the session.
NI 4070/4072 Calibration Procedure34ni.com
niDMM_close to close
To verify the frequency of the NI 4070/4072, complete the following steps:
1.Remove all connections from the NI 4070/4072.
Note Polarity is not important in steps 2, 3, and 5.
2.Connect one end of the coaxial cable to the Pomona 4892 double
banana plug.
3.Tighten the other end of the coaxial cable in the screw terminal
channels 5 and 39 of the TB-2715 terminal block.
4.Connect the TB-2715 with the coaxial cable attached to the NI 6608.
5.Plug the Pomona 4892 into the HI and LO terminals of the
NI 4070/4072.
6.Call
7.Call
niDMM_reset to reset the NI 4070/4072 to a known state.
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_FREQ
•Range = 1
•Resolution = 0
8.Call niDMM_ConfigureFrequencyVoltageRange with
Voltage Range set to
9.Call
GPCTR_Control with the following parameters:
5.
•deviceNumber = the device number of the NI 6608, assigned by
Measurement & Automation Explorer (MAX)
•gpctrNum =
ND_COUNTER_0
•action = ND_RESET
10. Call GPCTR_Set_Application with the following parameters:
•deviceNumber = the device number of the NI 6608, assigned
by MAX
•gpctrNum =
ND_COUNTER_0
•application = ND_PULSE_TRAIN_GNR
11. Call GPCTR_Change_Parameter with the following parameters:
•deviceNumber = the device number of the NI 6608, assigned
12. Call GPCTR_Change_Parameter with the following parameters:
•deviceNumber = the device number of the NI 6608, assigned
by MAX
•gpctrNum =
ND_COUNTER_0
•paramID = ND_COUNT_2
•paramValue = 10e6
13. Call GPCTR_Control with the following parameters:
•deviceNumber = the device number of the NI 6608, assigned
by MAX
•gpctrNum =
ND_COUNTER_0
•action = ND_PROGRAM
14. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table 21.
15. Call
GPCTR_Control with the following parameters:
•deviceNumber = the device number of the NI 6608, assigned
by MAX
•gpctrNum =
ND_COUNTER_0
•action = ND_RESET
16. Repeat steps 10 through 15 with the following modification: in
steps 11 and 12, change paramValue to
function
GPCTR_Change_Parameter.
500 when you call the
17. Repeat steps 10 through 15 with the following modification: in
steps 11 and 12, change paramValue to
GPCTR_Change_Parameter.
20 when you call the function
You have completed verifying the frequency of the NI 4070/4072. Select
one of the following options:
•If you are calibrating an NI 4072 and want to continue verifying other
modes, go to the Verifying Capacitance and Inductance (NI 4072
Only) section.
•If you do not want to verify other modes and you are performing a
post-adjustment verification, go to the Completing the Adjustment
Procedures section.
•If you do not want to verify any additional modes and you are
performing a pre-adjustment verification, call
niDMM_close to close
the session.
NI 4070/4072 Calibration Procedure36ni.com
Verifying Capacitance and Inductance (NI 4072 Only)
This verification procedure only applies to the NI 4072 and requires
additional test equipment, as indicated in the Additional Requirements for
the NI 4072 section.
Note The NI 4072 inductance accuracy is theoretically verified if the capacitance
accuracy meets the specifications. If you have access to precision inductors, you can verify
the inductance measurements by comparing your results with the published accuracy
specifications.
NI suggests using traceable capacitor standards with low thermal drift. You
can use different verification capacitors to verify each capacitance range.
You can verify two ranges with the same verification capacitor as long as
its value is ≥10% of the higher capacitor range. For example, you can use a
1 nF verification capacitor to test both the 10 nF and 1 nF ranges.
After taking each measurement, verify that the measurement falls between
the tolerances listed in Table 22. Tolerances are provided instead of
absolute limits, because you can use capacitance verification values other
than the values suggested, or the calibrated value may differ slightly from
the nominal capacitance (for example, 272.43 pF instead of 270.00 pF).
The tolerances shown in Table 22 correspond to the NI 4072 accuracy
specifications.
The following verification procedure assumes the use of verification
capacitors with the following values: 270 pF, 1 nF, 100 nF, 10 μF, and
1000 μF.
The configuration of the cables and fixtures should be consistent
throughout each measurement. If you are using cables to connect the
verification capacitors to the NI 4072 banana plug connectors, minimize
noise by ensuring that the cables remain fixed and do not move during the
measurement.
Keep direct contact with the verification capacitors to a minimum so that
they are constantly kept at the ambient temperature. After connecting a
capacitor to the NI 4072 terminals, NI recommends waiting 30 seconds for
the capacitor temperature to stabilize.
Note You should know the total capacitance up to the banana connectors that plug into the
NI 4072 before performing the verification procedure.
To verify the capacitance measurements of the NI 4072, complete the
following steps:
1.Disconnect any fixtures or cables from the NI 4072.
2.Call
3.Call
niDMM_reset to reset the NI 4070/4072 to a known state.
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_CAPACITANCE
•Range = 300e-12
•Resolution = 50e-15
4.Set the number of averages of the NI 4072 to 20 by calling
niDMM_SetAttributeViInt32 with the following parameters:
•Attribute_ID =
NIDMM_ATTR_LC_NUMBER_MEAS_TO_AVERAGE
•Attribute_Value = 20
This measurement corresponds to a 0 pF capacitance.
Note
5.Call
niDMM_Read. Verify that this measurement falls between the
tolerances listed in Table 22.
6.Plug in the insulated banana plug shorting bar across the HI and LO
banana plug connectors of the NI 4072.
7.Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_INDUCTANCE
•Range = 10e-6
•Resolution = 1e-9
8.Set the number of averages of the NI 4072 to 40 by calling
niDMM_SetAttributeViInt32 with the following parameters:
•Attribute_ID =
NIDMM_ATTR_LC_NUMBER_MEAS_TO_AVERAGE
•Attribute_Value = 40
Note
This measurement corresponds to a 0 μH inductance.
9.Call
niDMM_Read. Verify that this measurement falls between the
tolerances listed in Table 22.
10. Remove the shorting bar and plug the 270 pF verification capacitor
into the HI and LO banana plug connectors of the NI 4072. Remember
to wait 30 seconds for the temperature to stabilize before performing
the next step.
NI 4070/4072 Calibration Procedure38ni.com
11. Call niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_CAPACITANCE
•Range = 300e-12
•Resolution = 50e-15
12. Set the number of averages of the NI 4072 to 20 by calling
niDMM_SetAttributeViInt32 with the following parameters:
•Attribute_ID =
NIDMM_ATTR_LC_NUMBER_MEAS_TO_AVERAGE
•Attribute_Value = 20
13. Call niDMM_Read. Verify that this measurement falls between the
tolerances listed in Table 22.
Note If you use capacitance verification values that differ from the values listed in
Table 10, verify that each measurement falls between the tolerances listed in Table 22.
The tolerances shown in Table 22 correspond to the NI 4072 accuracy specifications.
14. Remove the 270 pF verification capacitor, and plug the 1 nF
verification capacitor into the HI and LO banana plug connectors of the
NI 4072.
15. Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_CAPACITANCE
•Range = 1e-9
•Resolution = 100e-15
16. Set the number of averages of the NI 4072 to 20 by calling
niDMM_SetAttributeViInt32 with the following parameters:
•Attribute_ID =
NIDMM_ATTR_LC_NUMBER_MEAS_TO_AVERAGE
•Attribute_Value = 20
Note
If you use capacitance verification values that differ from the values listed in
Table 10, verify that each measurement falls between the tolerances listed in Table 22.
The tolerances shown in Table 22 correspond to the NI 4072 accuracy specifications.
17. Call
niDMM_Read. Verify that this measurement falls between the
19. Set the number of averages of the NI 4072 to 20 by calling
niDMM_SetAttributeViInt32 with the following parameters:
•Attribute_ID =
NIDMM_ATTR_LC_NUMBER_MEAS_TO_AVERAGE
•Attribute_Value = 20
Note
If you use capacitance verification values that differ from the values listed in
Table 10, verify that each measurement falls between the tolerances listed in Table 22.
The tolerances shown in Table 22 correspond to the NI 4072 accuracy specifications.
20. Call
niDMM_Read. Verify that this measurement falls between the
tolerances listed in Table 22.
21. Remove the verification capacitor, and plug into the HI and LO banana
plug connectors of the NI 4072 the next capacitor to be verified,
according to Table 10.
22. Repeat steps 18 through 21, using the parameters shown in
Table 10 for
NIDMM_ATTR_LC_NUMBER_MEAS_TO_AVERAGE for all verification
niDMM_ConfigureMeasurement and
capacitors listed.
Note If you use capacitance verification values that differ from the values listed in
Table 10, verify that each measurement falls between the tolerances listed in Table 22.
The tolerances shown in Table 22 correspond to the NI 4072 accuracy specifications.
Table 10. niDMM_ConfigureMeasurement Parameters
Valu e of
Verification Capacitor
100 nF100e-910e-12
10 uF10e-61e-920
niDMM Configure Measurement Parameters
1e-6100e-1220
100e-610e-93
Number of AveragesRangeResolution
20
1000 uF1e-3100e-93
10e-31e-63
You have completed verifying the capacitance and inductance of the
NI 4072. Select one of the following options:
•If you are performing a pre-adjustment verification, call
niDMM_close to close the session.
•If you are performing a post-adjustment verification, go to the
Completing the Adjustment Procedures section.
NI 4070/4072 Calibration Procedure40ni.com
Adjustment Procedures
This section explains how to adjust the NI 4070/4072. You can choose to
perform these adjustment procedures with or without performing the
verification procedures first.
The parameters Range, Resolution, Expected Measurement, and
Frequency used in function calls in this section have floating point values.
For example, if Range =
NI Digital Multimeters Help for more information about parameter values.
Note NI recommends repeating the verification procedures after you perform these
adjustment procedures. Reverification ensures that the device you have calibrated is
operating within specifications after adjustments.
Caution If you skip any of the steps within a section of the adjustment procedures,
NI-DMM does not allow you to store your new calibration coefficients. Instead, NI-DMM
restores the original coefficients to the EEPROM.
Setting Up the Test Equipment
If you have not already set up the test equipment, complete the following
steps:
1.Remove all connections from the four input banana plug connectors on
2.Verify that the calibrator has been calibrated within the time limits
1, the floating point value is 1.0. Refer to the
the NI 4070/4072.
specified in the Required Test Equipment section, and that DC zeros
calibration has been performed within the last 30 days. Consult the
Fluke 5700A/5720A user documentation for instructions on
calibrating these devices.
Note Ensure that the calibrator is warmed up for at least 60 minutes before you begin this
procedure.
3.Reset the calibrator.
4.If you have not already done so, allow the NI 4070/4072 to warm up
for 60 minutes within a powered-on PXI chassis or PC.
To adjust the DC voltage and resistance of the NI 4070/4072, complete the
following steps:
1.Fasten the connectors on one end of the Fluke 5440 cable to the
appropriate banana plug connectors on the NI 4070/4072, and fasten
the connectors on the other end of the cable to the appropriate
calibrator binding posts. Figure 1 shows the correct connections.
Table 1 lists the cable connections.
2.Wait two minutes for the thermal EMF to stabilize if the cable was not
previously connected in this configuration.
3.Call
Note You will use Cal Session in all subsequent function calls.
Note The default user password for adjusting the NI 4070/4072 is NI. Use
niDMM_SetCalPassword to change the password.
niDMM_InitExtCal with the resource descriptor of the
NI 4070/4072 and your valid user password to output a calibration
session (Cal Session) that you can use to perform NI-DMM
calibration or regular measurement functions.
4.Call
niDMM_ConfigurePowerLineFrequency with PowerLine
Frequency set to
50 or 60, depending on the power line frequency
(in hertz) that your instruments are powered from; select
50 for
400 Hz power line frequencies.
5.Output 100 mV on the calibrator with the range locked to 2.2 V.
6.Call
niDMM_CalAdjustGain with the following parameters:
•Mode =
NIDMM_VAL_DC_VOLTS
•Range = 0.1
•Input Resistance = NIDMM_VAL_10_MEGAOHM
•Expected Measurement = 0.1
7.Output –100 mV on the calibrator.
8.Call niDMM_CalAdjustGain with the following parameters:
•Mode =
NIDMM_VAL_DC_VOLTS
•Range = 0.1
•Input Resistance = NIDMM_VAL_10_MEGAOHM
•Expected Measurement = –0.1
9.Output 10 V on the calibrator.
NI 4070/4072 Calibration Procedure42ni.com
10. Call niDMM_CalAdjustGain with the following parameters:
13. Disconnect the Fluke 5440 cable from the NI 4070/4072 banana plug
connectors, leaving the other end of the cable connected to the
calibrator binding posts.
14. Plug in the insulated banana plug shorting bar across the HI and LO
banana plug connectors of the NI 4070/4072.
15. Wait two minutes for the thermal EMF to stabilize.
16. Call
niDMM_CalAdjustOffset with the following parameters:
23. Output 0 Ω from the calibrator without external sense or 2-wire
compensation.
24. Call
niDMM_CalAdjustGain with the following parameters:
•Mode =
NIDMM_VAL_2_WIRE_RES
•Range = 10e6
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
•Expected Value = the display on the calibrator for 0 Ω
25. Call
niDMM_CalAdjustOffset with the following parameters:
•Mode =
NIDMM_VAL_2_WIRE_RES
•Range = 10e6
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
26. Disconnect the Fluke 5440 cable from the NI 4070/4072.
27. Call
28. Call
niDMM_CalAdjustMisc with Type set to
NIDMM_EXTCAL_MISCCAL_ZINT.
niDMM_CalAdjustMisc with Type set to
NIDMM_EXTCAL_MISCCAL_2WIRELEAKAGE.
29. On the NI 4070/4072, plug a Pomona B-4 banana cable from the HI
input to the HI SENSE input. Plug another Pomona B-4 banana cable
from the LO input to the LO SENSE input.
Caution Make sure that the insulation of these cables does not touch.
30. Call
niDMM_CalAdjustMisc with Type set to
NIDMM_EXTCAL_MISCCAL_4WIRELEAKAGE.
31. Remove the banana cables, and plug the two sets of Fluke 5440 cables
into the appropriate banana plug connectors on the NI 4070/4072,
as shown in Figure 2 for 4-wire resistance.
32. Wait two minutes for the thermal EMF to stabilize.
33. Output 100 MΩ from the calibrator without external sense.
34. Call
niDMM_CalAdjustGain with the following parameters:
•Mode =
NIDMM_VAL_2_WIRE_RES
•Range = 100e6
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
•Expected Value = the display on the calibrator for 100 MΩ
35. Output 0 Ω from the calibrator without external sense or 2-wire
compensation.
NI 4070/4072 Calibration Procedure44ni.com
36. Call niDMM_CalAdjustGain with the following parameters:
•Mode =
NIDMM_VAL_2_WIRE_RES
•Range = 100e6
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
•Expected Value = the display on the calibrator for 0 Ω
37. Call
niDMM_CalAdjustOffset with the following parameters:
•Mode =
NIDMM_VAL_2_WIRE_RES
•Range = 100e6
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
38. Output 100 kΩ on the calibrator with external sense turned on
but without 2-wire compensation.
39. Call
niDMM_CalAdjustGain with the following parameters:
•Mode =
NIDMM_VAL_4_WIRE_RES
•Range = 100e3
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
•Expected Value = the display on the calibrator for 100 kΩ
40. Output 0 Ω on the calibrator with external sense turned on but without
2-wire compensation.
41. Call
niDMM_CalAdjustGain with the following parameters:
•Mode =
NIDMM_VAL_4_WIRE_RES
•Range = 100e3
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
•Expected Value = the display on the calibrator for 0 Ω
42. Output 10 kΩ on the calibrator with external sense turned on
but without 2-wire compensation.
43. Call
niDMM_CalAdjustGain with the following parameters:
•Mode =
NIDMM_VAL_4_WIRE_RES
•Range = 10e3
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
•Expected Value = the display on the calibrator for 10 kΩ
44. Output 0 Ω on the calibrator with external sense turned on but without
2-wire compensation.
45. Call
niDMM_CalAdjustOffset with the following parameters:
•If you are not performing additional adjustments, refer to one of the
following sections:
–Verification Procedures—to verify your new calibration
coefficients before saving them to the EEPROM
–Completing the Adjustment Procedures—if you do not want to
verify the adjustments you have just made
Adjusting AC Voltage (AC- and DC-Coupled) Modes
Note If you do not use the AC voltage modes for any measurements, or the accuracy of
these modes is irrelevant, you can skip this section in the calibration procedure and go
directly to the Adjusting Current Modes section.
To adjust the AC voltage of the NI 4070/4072, complete the following
steps:
1.Reset the calibrator.
2.Fasten the connectors on one end of the Fluke 5440 cable into the
appropriate banana plug connectors on the NI 4070/4072, and fasten
the connectors on the other end of the cable to the appropriate
calibrator binding posts. Figure 1 shows the correct connections.
Table 1 lists the cable connections.
3.Refer to Table 11 for the appropriate calibrator output and parameter
values as you complete the following steps:
a.On the calibrator, output the value listed in the Calibrator Output
column in Table 11 for the current iteration.
b.Call
c.Call
4.Repeat step 3 for each of the remaining iterations listed in Table 11.
niDMM_CalAdjustGain with Mode set to
NIDMM_VAL_AC_VOLTS. Set the remaining parameters as shown
You have completed adjusting the AC voltage modes of the NI 4070/4072.
Select one of the following options:
•If you are performing additional adjustments, refer to the following
sections, as applicable:
–Adjusting Current Modes
–Adjusting Capacitance and Inductance (NI 4072 Only)
(Hz)
Caution For the NI 4072, adjusting the capacitance and inductance is required. Skipping
this step causes an incorrect adjustment of the device.
•If you are not performing additional adjustments, refer to one of the
following sections:
–Verification Procedures—to verify your new calibration
coefficients before saving them to the EEPROM
–Completing the Adjustment Procedures—if you do not want to
verify the adjustments you have just made
Adjusting Current Modes
If you do not use the current modes (DC and AC), or the accuracy is
insignificant for your application, you can skip this section and select one
of the following options:
•If you skip this section and you are calibrating an NI 4072, go to the
Adjusting Capacitance and Inductance (NI 4072 Only) section.
•If you skip this section and you want to verify the new calibration
coefficients before saving them to the EEPROM, repeat the
Verification Procedures section (except for Initial Setup).
•If you skip this section and you do not want to verify the new
calibration coefficients, go to the Completing the Adjustment
Procedures section.
To adjust the current modes of the NI 4070/4072, complete the following
steps:
1.Reset the calibrator.
2.Fasten the connectors on one end of the Fluke 5440 cable to the
NI 4070/4072 HI SENSE and LO banana plug connectors, and fasten
the connectors on the other end of the cable to the HI and LO calibrator
binding posts. Figure 3 shows the correct connections. Table 9 lists the
cable connections.
3.Call
niDMM_ConfigureMeasurement with the following
parameters:
•Function =
NIDMM_VAL_DC_CURRENT
•Range = 0.02
4.Call niDMM_Read to configure the NI 4070/4072 for a current mode
before applying current.
5.Output 20 mA on the calibrator with the current output set to
6.Call
niDMM_CalAdjustGain with the following parameters:
•Mode =
NIDMM_VAL_DC_CURRENT
NORM.
•Range = 0.02
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
•Expected Value = 0.02
7.Output –20 mA on the calibrator with the current output set to NORM.
8.Call
niDMM_CalAdjustGain with the following parameters:
•Mode =
NIDMM_VAL_DC_CURRENT
•Range = 0.02
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
•Expected Value = –0.02
9.Output 0 A on the calibrator.
10. Call
niDMM_CalAdjustOffset with the following parameters:
•Mode =
NIDMM_VAL_DC_CURRENT
•Range = 0.02
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
NI 4070/4072 Calibration Procedure54ni.com
11. Call niDMM_CalAdjustOffset with the following parameters:
•Mode =
NIDMM_VAL_AC_CURRENT
•Range = 0.01
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
12. Output 200 mA on the calibrator.
13. Call
niDMM_CalAdjustGain with the following parameters:
•Mode =
NIDMM_VAL_DC_CURRENT
•Range = 0.2
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
•Expected Value = 0.2
14. Output –200 mA on the calibrator.
15. Call
niDMM_CalAdjustGain with the following parameters:
•Mode =
NIDMM_VAL_DC_CURRENT
•Range = 0.2
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
•Expected Value = –0.2
16. Output 0 A on the calibrator.
17. Call
niDMM_CalAdjustOffset with the following parameters:
•Mode =
NIDMM_VAL_DC_CURRENT
•Range = 0.2
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
18. Call niDMM_CalAdjustOffset with the following parameters:
•Mode =
NIDMM_VAL_AC_CURRENT
•Range = 0.1
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
19. Output 1 A on the calibrator.
20. Call
niDMM_CalAdjustGain with the following parameters:
22. Call niDMM_CalAdjustGain with the following parameters:
•Mode =
•Range = 1
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
•Expected Value = –1
23. Output 0 A on the calibrator with the current output set to NORM.
24. Call
25. Call niDMM_CalAdjustOffset with the following parameters:
26. Call niDMM_CalAdjustMisc with Type set to
You have completed adjusting the current modes of the NI 4070/4072.
Select one of the following options:
niDMM_CalAdjustOffset with the following parameters:
•Mode =
•Range = 1
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
•Mode =
•Range = 1
•Input Resistance = NIDMM_VAL_RESISTANCE_NA
NIDMM_EXTCAL_MISCCAL_SECTION.
NIDMM_VAL_DC_CURRENT
NIDMM_VAL_DC_CURRENT
NIDMM_VAL_AC_CURRENT
♦If you are calibrating an NI 4070, refer to one of the following sections:
•Verification Procedures—to verify your new calibration coefficients
before saving them to the EEPROM
•Completing the Adjustment Procedures—if you do not want to verify
the adjustments you have just made
♦If you are calibrating an NI 4072, refer to the Adjusting Capacitance and
Inductance (NI 4072 Only) section.
Adjusting Capacitance and Inductance
(NI 4072 Only)
♦If you are calibrating an NI 4070, skip this section and select one of the
following options:
•If you want to verify the new calibration coefficients before saving
them to the EEPROM, repeat the Verification Procedures section
(except for the Initial Setup section).
•If you do not want to verify the new calibration coefficients, go to the
Completing the Adjustment Procedures section.
NI 4070/4072 Calibration Procedure56ni.com
♦If you are calibrating an NI 4072, you must complete this section to attain
a valid calibration.
Caution It is necessary to adjust DC voltage and resistance before running these
adjustment steps. During this procedure, be sure to keep hands and any other moving
objects away from the fixture after calling every function.
To adjust the capacitance and inductance of the NI 4072, complete the
following steps:
1.Disconnect any fixtures or cables from the NI 4072.
2.Call
niDMM_CalAdjustLC with Type set to
NIDMM_EXTCAL_LC_OPEN.
3.Plug in the insulated banana plug shorting bar across the HI and LO
banana plug connectors of the NI 4072.
4.Call
niDMM_CalAdjustLC with Type set to
NIDMM_EXTCAL_LC_SHORT.
5.Remove the shorting bar and plug the 25 Ω resistor across the HI and
LO banana plug connectors of the NI 4072. The leads between the
resistor and the NI 4072 terminals should be ≤1 in.
6.Wait 30 seconds for the thermal EMF to stabilize.
7.Call
niDMM_CalAdjustLC with Type set to
NIDMM_EXTCAL_LC_25OHM.
8.Remove the resistor, and plug in across the HI and LO banana plug
connectors of the NI 4072 the next adjustment resistor, according to
Table 14.
9.Repeat steps 6 through 8 using the parameters shown in Table 14 for
niDMM_CalAdjustLC.
Table 14. niDMM_CalAdjustLC Parameters
niDMM_CalAdjustLC Parameters
Value of Resistor
125 ΩNIDMM_EXTCAL_LC_1KOHM
5kΩNIDMM_EXTCAL_LC_5KOHM
100 kΩNIDMM_EXTCAL_LC_100KOHM
Typ e
You have completed adjusting the capacitance and inductance modes of the
NI 4072. Select one of the following options:
•To verify that the NI 4072 is now operating within its specifications,
go to the Verification Proceduressection and complete the appropriate
procedures.
•To finish the calibration and close the session, go to the Completing the
Adjustment Procedures section.
Completing the Adjustment Procedures
To complete the adjustment procedure for the NI 4070/4072 and close the
session, call
•Action =
calibration were satisfactory and you want to save the new calibration
coefficients to the EEPROM.
Otherwise,
•Action =
calibration were unsatisfactory and you want to restore the original
calibration coefficients to the EEPROM.
Verification Limits
This section includes the verification limits for DC voltage, AC voltage,
4-wire resistance, 2-wire resistance, DC current, AC current, and frequency
for the NI 4070/4072, and the verification tolerances for capacitance on the
NI 4072. Compare these limits to the results you obtain in the Verification
Procedures section.
niDMM_CloseExtCal with the following parameter:
NIDMM_EXTCAL_ACTION_SAVE if the results of the
NIDMM_EXTCAL_ACTION_ABORT if the results of the
Note Use the values in the 24-Hour Limits column for a post-adjustment verification only.
Otherwise, use the values in the 2-Year Limits column.
Limits in the following tables are based upon the February 2007 edition of the
NI 4070/4072 Specifications. Refer to the most recent NI 4070/4072 specifications online
ni.com/manuals. If a more recent edition of the specifications is available, recalculate
at
the limits based upon the latest specifications.
DC Voltage
Table 15. NI 4070/4072 DC Voltage Verification Limits
Calibrator
Amplitude
0V1V>10 GΩ/10 MΩ–6 μV6 μV–2 μV2 μV
0V10 V>10 GΩ/10 MΩ–60 μV60 μV–20 μV20 μV
0V100 V10 MΩ–600 μV600 μV–200 μV200 μV
0V300 V10 MΩ–6 mV6mV–1.8 mV1.8 mV
100 mV100 mV>10 GΩ/10 MΩ0.099994 V0.100006 V0.099998 V0.100002 V
NI 4070/4072 Calibration Procedure58ni.com
RangeInput Resistance
2-Year Limits24-Hour Limits
LowerUpperLowerUpper
Table 15. NI 4070/4072 DC Voltage Verification Limits (Continued)
Calibrator
Amplitude
–100 mV100 mV>10 GΩ/10 MΩ–0.100006 V–0.099994 V–0.100002 V–0.099998 V
1V1V>10 GΩ/10 MΩ0.999969 V1.000031 V0.999992 V1.000008 V
–1 V1V>10 GΩ/10 MΩ–1.000031 V–0.999969 V–1.000008 V–0.999992 V
10 V10 V>10 GΩ/10 MΩ9.99969 V10.00031 V9.99994 V10.00006 V
–10 V10 V>10 GΩ/10 MΩ–10.00031 V–9.99969 V–10.00006 V–9.99994 V
100 V100 V10 MΩ99.9959 V100.0041 V99.9992 V100.0008 V
–100 V100 V10 MΩ–100.0041 V–99.9959 V–100.0008 V–99.9992 V
300 V300 V10 MΩ299.9835 V300.0165 V299.9964 V300.0036 V
–300 V300 V10 MΩ–300.0165 V–299.9835 V–300.0036 V–299.9964 V
RangeInput Resistance
2-Year Limits24-Hour Limits
LowerUpperLowerUpper
AC Voltage
Table 16. NI 4070/4072 AC Voltage Verification Limits
The complete calibration process for the NI 4070/4072 consists of
verifying, adjusting, and reverifying a device. During verification, you
compare the measured performance to an external standard of known
measurement uncertainty to confirm that the product meets or exceeds
specifications. Figure 4 shows the procedural flow for verification.
During adjustment, you correct the measurement error of the device by
adjusting the calibration constants and storing the new calibration constants
in the EEPROM. Frequency is the only mode that does not require
adjustment. Figure 5 shows the procedural flow for adjustment.
Reverifying all modes after adjustments ensures that the adjustment
procedures were performed correctly.
Normally, the calibration sequence is as follows:
1.Verify the operation of the NI 4070/4072 using the 2-year accuracy
limits (or the 90-day accuracy limits if it has been externally calibrated
within that time).
2.Adjust the NI 4070/4072.
3.Reverify the NI 4070/4072 using the 24-hour accuracy limits (or the
2-year accuracy limits when the 24-hour limits are not specified).
Depending on your measurement and accuracy requirements, a complete
calibration of the NI 4070/4072 may not be necessary. A number of options
are available that can shorten the calibration time. The following
adjustment options are available:
•Complete calibration—Performing the entire calibration procedure
from beginning to end; guarantees that the NI 4070/4072 performs at
or above the published specifications for all modes and ranges
•Complete calibration with exceptions:
–Omitting AC voltage mode steps if you do not use the AC voltage
modes or if the AC voltage accuracy is irrelevant
–Omitting DC/AC current mode steps if you do not use the current
modes or if the DC/AC current accuracy is irrelevant
–Omitting both AC voltage and DC/AC current mode steps if you
do not use those modes or if the accuracy of those measurements
is irrelevant
The National Instruments Web site is your complete resource for technical
support. At
troubleshooting and application development self-help resources to email
and phone assistance from NI Application Engineers.
A Declaration of Conformity (DoC) is our claim of compliance with the
Council of the European Communities using the manufacturer’s
declaration of conformity. This system affords the user protection for
electronic compatibility (EMC) and product safety. You can obtain the DoC
for your product by visiting
supports calibration, you can obtain the calibration certificate for your
product at
National Instruments corporate headquarters is located at
11500 North Mopac Expressway, Austin, Texas, 78759-3504.
National Instruments also has offices located around the world to help
address your support needs. For telephone support in the United States,
create your service request at
instructions or dial 512 795 8248. For telephone support outside the United
States, contact your local branch office:
Australia 1800 300 800, Austria 43 662 457990-0,
Belgium 32 (0) 2 757 0020, Brazil 55 11 3262 3599,
Canada 800 433 3488, China 86 21 5050 9800,
Czech Republic 420 224 235 774, Denmark 45 45 76 26 00,
Finland 385 (0) 9 725 72511, France 33 (0) 1 48 14 24 24,
Germany 49 89 7413130, India 91 80 41190000, Israel 972 3 6393737,
Italy 39 02 413091, Japan 81 3 5472 2970, Korea 82 02 3451 3400,
Lebanon 961 (0) 1 33 28 28, Malaysia 1800 887710,
Mexico 01 800 010 0793, Netherlands 31 (0) 348 433 466,
New Zealand 0800 553 322, Norway 47 (0) 66 90 76 60,
Poland 48 22 3390150, Portugal 351 210 311 210, Russia 7 495 783 6851,
Singapore 1800 226 5886, Slovenia 386 3 425 42 00,
South Africa 27 0 11 805 8197, Spain 34 91 640 0085,
Sweden 46 (0) 8 587 895 00, Switzerland 41 56 2005151,
Taiwan 886 02 2377 2222, Thailand 662 278 6777,
Turkey 90 212 279 3031, United Kingdom 44 (0) 1635 523545
ni.com/support you have access to everything from
ni.com/calibration.
ni.com/certification. If your product
ni.com/support and follow the calling
National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation.
Refer to the Terms of Use section on ni.com/legal for more information about National
Instruments trademarks. Other product and company names mentioned herein are trademarks or trad e
names of their respective companies. For patents covering National Instruments products, refer to the
appropriate location: Help»Patents in your software, the patents.txt file on your CD, or
ni.com/patents.