National Instruments NI 4070, NI4072 User Manual

CALIBRATION PROCEDURE
NI 4070/4072 6½-Digit FlexDMM
This document contains step-by-step instructions for writing an external calibration procedure for the National Instruments PXI/PCI-4070 and NI PXI-4072 digital multimeters (DMMs). Each of these National Instruments DMMs is a 6½-digit FlexDMM and 1.8 MS/s

Contents

isolated digitizer. For more information on calibration, visit
Conventions ............................................................................................2
Software Requirements ........................................................................... 2
Documentation Requirements................................................................. 3
Calibration Function Reference ....................................................... 3
Password ................................................................................................. 3
Calibration Interval ................................................................................. 4
Test Equipment ....................................................................................... 4
Required Test Equipment ................................................................ 4
Optional Test Equipment ................................................................. 5
Test Conditions ....................................................................................... 5
Calibration Procedures............................................................................ 6
Initial Setup......................................................................................6
Verification Procedures ................................................................... 7
Adjustment Procedures .................................................................... 41
Verification Limits.................................................................................. 58
DC Voltage ...................................................................................... 58
AC Voltage ...................................................................................... 59
4-Wire Resistance ............................................................................61
2-Wire Resistance ............................................................................61
DC Current.......................................................................................62
AC Current.......................................................................................62
Frequency......................................................................................... 63
Capacitance and Inductance............................................................. 63
Appendix A: Calibration Options ........................................................... 64
Where to Go for Support......................................................................... 68
.
ni.com/

Conventions

The following conventions are used in this document:
» The » symbol leads you through nested menu items and dialog box options
to a final action. The sequence File»Page Setup»Options directs you to pull down the File menu, select the Page Setup item, and select Options from the last dialog box.
The symbol indicates that the following text applies only to a specific
product, a specific operating system, or a specific software version.
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to avoid injury, data loss, or a system crash. When this symbol is marked on a product, refer to the Read Me First: Safety and Radio-Frequency Interference document included with the device for information about precautions to take.
bold Bold text denotes items that you must select or click in the software, such
as menu items and dialog box options. Bold text also denotes parameter names.
italic Italic text denotes variables, emphasis, a cross-reference, hardware labels,
or an introduction to a key concept. Italic text also denotes text that is a placeholder for a word or value that you must supply.
monospace Text in this font denotes text or characters that you should enter from the
keyboard, sections of code, programming examples, and syntax examples. This font is also used for the proper names of disk drives, paths, directories, programs, subprograms, subroutines, device names, functions, operations, variables, filenames, and extensions.

Software Requirements

NI-DMM supports a number of programming languages including LabVIEW, LabWindows Visual Basic. When you install NI-DMM, you need to install support for only the language you intend to use to write your calibration utility.
Note NI-DMM version 2.1 or later supports NI PXI-4070 calibration, NI-DMM
version 2.2 or later supports NI PCI-4070 calibration, and NI-DMM version 2.3 or later supports NI 4072 calibration.
NI 4070/4072 Calibration Procedure 2 ni.com
/CVI™, Microsoft Visual C++, and Microsoft
The procedures in this document are described using C function calls. You also can program in LabVIEW using the VIs that correspond to the C function calls.

Documentation Requirements

In addition to this calibration document, you may find the following references helpful in writing your calibration utility. All of these documents are installed on your computer when you install NI-DMM. To locate them, select Start»All Programs»National Instruments» NI-DMM»Documentation.
NI Digital Multimeters Help
NI Digital Multimeters Getting Started Guide
NI recommends referring to the following document online at
manuals
specifications:
NI 4070/4072 Specifications
You may need the following documents, which are available at
manuals
TB-2715 Terminal Block Installation Guide
About Your NI 6608 Device
to ensure that you are using the latest NI 4070/4072
, to perform the optional frequency verification procedure:

Calibration Function Reference

For detailed information about the NI-DMM calibration functions used in this procedure, refer to the LabVIEW Reference or the C/CVI/VB Reference sections of the NI Digital Multimeters Help, located at Start» All Programs»National Instruments»NI-DMM»Documentation.

Password

The default calibration password in NI-DMM is "NI".
ni.com/
ni.com/
© National Instruments Corporation 3 NI 4070/4072 Calibration Procedure

Calibration Interval

The accuracy requirements of your measurement application determine how often you should calibrate the NI 4070/4072. NI recommends performing a complete calibration at least once every two years. NI does not guarantee the absolute accuracy of the NI 4070/4072 beyond this two-year calibration interval. You can shorten the calibration interval based on the demands of your application. Refer to Appendix A: Calibration
Options for more information.

Test Equipment

This section describes the required and optional equipment for calibration.

Required Test Equipment

Requirements for All NI 4070/4072 Devices

The following equipment is required for calibrating the NI 4070/4072:
Fluke 5700A multifunction calibrator calibrated within the last 90 days, or a Fluke 5720A multifunction calibrator calibrated within the last year
Two sets of Fluke 5440 low thermal electromotive force (EMF) copper cables
Pomona 5145 insulated double banana plug shorting bar (or another means of creating a short with low thermal EMF (150 nV) across the HI and LO input banana plug connectors on the NI 4070/4072)
Two Pomona B-4 banana-to-banana patch cords (cables) or similar banana-to-banana cables with length not to exceed 4 in.
National Instruments PXI chassis and controller, or a personal computer (PC) with an available slot for the NI 4070/4072

Additional Requirements for the NI 4072

The following equipment is required for calibrating the capacitance and inductance modes of the NI 4072:
•25Ω, 125 Ω, 5kΩ, and 100 kΩ resistors with thermal drift ≤5ppm/°C and tolerance 1%. The distance between the resistor leads and the NI 4072 terminals should be 1 in.
Verification capacitors calibrated to at least four times the accuracy of the NI 4072, with temperature coefficients 250 ppm/°C. The values of the verification capacitors should cover the complete capacitance range. NI suggests using traceable capacitor standards with values 10% of full range for all ranges, except the 300 pF range. For the
NI 4070/4072 Calibration Procedure 4 ni.com
300 pF range, a capacitor with values between 90–100% of full scale should be used. NI suggests using the capacitance standards of the SCA Series from IET Labs. This calibration procedure assumes the use of 270pF, 1nF, 100nF, 10μF, and 1000 μF standards.
If you are using cables to connect the verification capacitors to the NI 4072 banana plug connectors, NI recommends using Pasternack PE3005 banana-to-banana coaxial cables with length 4 inches and total capacitance 40 pF. Before performing the verification procedure, you should know the total capacitance up to the end of the banana connectors that plug into the NI 4072.

Optional Test Equipment

The following equipment is optional for calibrating the NI 4070/4072 and is only used for frequency verification:
NI PXI-6608 timing and digital I/O module
National Instruments SH68-68-D1 shielded cable
National Instruments TB-2715 terminal block
Pomona MDP 4892 double banana plug with strain relief
Coaxial cable (for example, RG178)

Test Conditions

Follow these guidelines to optimize the connections and the environment during calibration:
Ensure that the PXI chassis fan speed is set to HI (if calibrating the NI PXI-4070/4072) and that the fan filters are clean.
Use PXI filler panels in all vacant slots to allow proper cooling.
Plug the PXI chassis or PC and the calibrator into the same power strip to avoid ground loops.
Power on and warm up both the calibrator and the NI 4070/4072 for at least 60 minutes before beginning this calibration procedure.
Maintain an ambient temperature of 23 ±1 °C.
Maintain an ambient relative humidity of less than 60%.
Allow the calibrator to settle fully before taking any measurements. Consult the Fluke 5700A/5720A user documentation for instructions.
Allow the thermal EMF enough time to stabilize when you change connections to the calibrator or the NI 4070/4072. The suggested time periods are stated where necessary throughout this document.
Keep a shorting bar connected between the VGUARD and GROUND binding posts of the calibrator at all times.
© National Instruments Corporation 5 NI 4070/4072 Calibration Procedure
Clean any oxidation from the banana plugs on the Fluke 5440 cables before plugging them into the binding posts of the calibrator or the banana plug connectors of the NI 4070/4072. Oxidation tarnishes the copper banana plugs so that they appear dull rather than shiny and leads to greater thermal EMF.
Keep the blue banana plugs on the Fluke 5440 cables connected to the V GUARD binding post of the calibrator at all times.
Prevent the cables from moving or vibrating by taping or strapping them to a nonvibrating surface. Movement or vibration causes triboelectric effects that can result in measurement errors.

Calibration Procedures

The calibration process includes the following steps:
1. Initial Setup—Set up the test equipment.
2. Verification Procedures—Verify the existing operation of the device. This step confirms whether the device is operating within its specified range prior to calibration. Figure 4 shows the procedural flow for verification.
3. Adjustment Procedures—Submit the device to NI for a factory calibration to adjust the calibration constants. Figure 5 shows the procedural flow for adjustment.
4. Reverification—Repeat the verification procedure to ensure that the device is operating within its specifications after adjustment.
These steps are described in more detail in the following sections.
Note In some cases, the complete calibration procedure may not be required. Refer to
Appendix A: Calibration Options for more information.

Initial Setup

Note This section is necessary for pre-adjustment verifications only. If you are performing
a post-adjustment verification, skip the setup and go directly to the Verifying DC Voltage section.
To set up the test equipment, complete the following steps:
1. Remove all connections from the four input banana plug connectors on the NI 4070/4072.
2. Verify that the calibrator has been calibrated within the time limits specified in the Required Test Equipment section, and that DC zeros calibration has been performed within the last 30 days. Consult the
NI 4070/4072 Calibration Procedure 6 ni.com
Fluke 5700A/5720A user documentation for instructions on calibrating these devices.
Note Ensure that both the calibrator and the NI 4070/4072 (installed in a powered-on
PXI chassis or PC) are warmed up for at least 60 minutes before you begin this procedure.
3. Call
Note You use this session in all subsequent function calls throughout the verification
procedures.
4. Call

Verification Procedures

You can use the verification procedures described in this section for both pre-adjustment and post-adjustment verification. The steps of each verification procedure must be performed in the order listed; however, you can omit entire sections (for example, the entire Verifying AC Current section), if necessary.
The parameters Range, Resolution, and Sample Interval used in function calls throughout this section have floating point values. For example, if
Range = Sample Count, Array Size, and ParamValue have integer values. Refer
to the NI Digital Multimeters Help for more information about parameter values.
niDMM_init with the resource name of the device to create a
session.
For more information on using
niDMM_init, refer to the NI Digital
Multimeters Help.
niDMM_SelfCal. This step is optional if you have adjusted the
NI 4070/4072 within the last 24 hours and the temperature has remained constant to within ±1 °C of the calibration temperature (T
1, the floating point value is 1.0. The parameters Trigger Count,
cal
).
Note Many of the parameter values listed in this document are expressed in scientific
notation. Some programming languages do not support the direct entry of numbers in this format. Be sure to properly enter these values with the appropriate number of zeros. For example, enter the scientific notation number 10e–6 as
100000. If your programming language supports scientific notation, NI recommends that
0.00001 and the number 100e3 as
you use this feature to minimize possible data entry errors.
© National Instruments Corporation 7 NI 4070/4072 Calibration Procedure

Verifying DC Voltage

To verify DC voltage of the NI 4070/4072, complete the following steps:
1. Plug in the insulated banana plug shorting bar across the HI and LO banana plug connectors on the NI 4070/4072.
2. Wait one minute for the thermal EMF to stabilize.
3. Call
4. Call
5. Set the input resistance of the NI 4070/4072 to >10 GΩ by calling
6. Call
7. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
8. Call
9. Call
10. Set the input resistance of the NI 4070/4072 to >10 GΩ by calling
11. Call niDMM_Read. Verify that this measurement falls between the
niDMM_reset.
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_DC_VOLTS
Range = 1
Resolution = 1e–6
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_GREATER_THAN_10_GIGAOHM
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_10_MEGAOHM
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_DC_VOLTS
Range = 10
Resolution = 10e–6
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_GREATER_THAN_10_GIGAOHM
limits listed in Table 15.
NI 4070/4072 Calibration Procedure 8 ni.com
12. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_10_MEGAOHM
13. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
14. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_DC_VOLTS
Range = 100
Resolution = 100e–6
15. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_10_MEGAOHM
16. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
17. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_DC_VOLTS
Range = 300
Resolution = 300e–6
18. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_10_MEGAOHM
19. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
20. Remove the shorting bar from the NI 4070/4072.
21. Reset the calibrator.
22. Fasten the connectors on one end of the Fluke 5440 cable to the appropriate banana plug connectors of the NI 4070/4072, and fasten the connectors on the other end of the cable to the appropriate calibrator binding posts. Figure 1 shows the correct connections. Table 1 lists the cable connections.
© National Instruments Corporation 9 NI 4070/4072 Calibration Procedure
1
6½-Digit FlexDMM
OUTPUT
VΩA
HI
LO
HI
AUX
CURRENT
AUX
I/O
5V MAX
300V MAX
1A, 250V
MAX
300V MAX
INPUT V
AMPS
SENSE
W 4W
CAT II
HI
W
LO
HI
LO
3
1 NI 4070/4072 2 Fluke 5700A/5720A Calibrator 3 Fluke 5440 Cable

Figure 1. Cable Connections for Voltage and 2-Wire Resistance

Table 1. Fluke 5440 Cable Connections

Banana Plug Connector
(NI 4070/4072)
HI Red OUTPUT HI
LO Black OUTPUT LO
(No connection) Blue V GUARD
Banana Plug Color (Fluke 5440 Cable)
Binding Post Label
(Fluke 5700A/5720A Calibrator)
2
SENSE
VΩ
HI
LO
GUARD GROUND
23. Wait two minutes for the thermal EMF to stabilize.
24. Generate 0 V on the calibrator.
25. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_DC_VOLTS
Range = 0.1
Resolution = 100e–9
26. Set the input resistance of the NI 4070/4072 to >10 GΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_GREATER_THAN_10_GIGAOHM
NI 4070/4072 Calibration Procedure 10 ni.com
27. Call niDMM_ConfigureMultiPoint with the following parameters:
Trigger Count =
1
Sample Count = 10
Sample Trigger = NIDMM_VAL_IMMEDIATE
Sample Interval = –1
28. Call niDMM_ReadMultiPoint with the following parameters:
Maximum Time =
NIDMM_VAL_TIME_LIMIT_AUTO
Array Size = 10
Average the results by summing the returned reading array of the function and dividing by the returned actual number of points. Store the result as the 100 mV >10 GΩ mode offset.
29. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_10_MEGAOHM
30. Call niDMM_ConfigureMultiPoint with the following parameters:
Trigger Count =
1
Sample Count = 10
Sample Trigger = NIDMM_VAL_IMMEDIATE
SampleInterval = –1
31. Call niDMM_ReadMultiPoint with the following parameters:
Maximum Time =
NIDMM_VAL_TIME_LIMIT_AUTO
Array Size = 10
Average the results by summing the returned reading array of the function and dividing by the returned actual number of points. Store the result as the 100 mV 10 MΩ mode offset.
32. Output 100 mV on the calibrator with the range locked to 2.2 V. This range prevents a 50 Ω calibrator output resistance from creating a voltage divider with the internal resistance of the NI 4070/4072.
33. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_DC_VOLTS
Range = 0.1
Resolution = 100e–9
© National Instruments Corporation 11 NI 4070/4072 Calibration Procedure
34. Set the input resistance of the NI 4070/4072 to >10 GΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_GREATER_THAN_10_GIGAOHM
35. Call
niDMM_Read. Subtract the previously stored 100 mV >10 GΩ
mode offset from this measurement, and verify that the result falls between the limits listed in Table 15.
36. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_10_MEGAOHM
37. Call
niDMM_Read. Subtract the previously stored 100 mV 10 MΩ
mode offset from this measurement and verify that the result falls between the limits listed in Table 15.
38. Output –100 mV on the calibrator with the range locked to 2.2 V. This range prevents a 50 Ω calibrator output resistance from creating a voltage divider with the internal resistance of the NI 4070/4072.
39. Set the input resistance of the NI 4070/4072 to >10 GΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_GREATER_THAN_10_GIGAOHM
40. Call
niDMM_Read. Subtract the previously stored 100 mV >10 GΩ
mode offset from this measurement, and verify that the result falls between the limits listed in Table 15.
41. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_10_MEGAOHM
42. Call
niDMM_Read. Subtract the previously stored 100 mV 10 MΩ
mode offset from this measurement and verify that the result falls between the limits listed in Table 15.
43. Output 1 V on the calibrator.
44. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_DC_VOLTS
Range = 1
Resolution = 1e–6
NI 4070/4072 Calibration Procedure 12 ni.com
45. Set the input resistance of the NI 4070/4072 to >10 GΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_GREATER_THAN_10_GIGAOHM
46. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
47. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_10_MEGAOHM
48. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
49. Output –1 V on the calibrator.
50. Set the input resistance of the NI 4070/4072 to >10 GΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_GREATER_THAN_10_GIGAOHM
51. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
52. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_10_MEGAOHM
53. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
54. Output 10 V on the calibrator.
55. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_DC_VOLTS
Range = 10
Resolution = 10e–6
56. Set the input resistance of the NI 4070/4072 to >10 GΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_GREATER_THAN_10_GIGAOHM
57. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
© National Instruments Corporation 13 NI 4070/4072 Calibration Procedure
58. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_10_MEGAOHM
59. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
60. Output –10 V on the calibrator.
61. Set the input resistance of the NI 4070/4072 to >10 GΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
NIDMM_ATTR_INPUT_RESISTANCE
Attribute_Value = NIDMM_VAL_GREATER_THAN_10_GIGAOHM
62. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
63. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
Attribute_Value =
64. Call
niDMM_Read. Verify that this measurement falls between the
NIDMM_ATTR_INPUT_RESISTANCE
NIDMM_VAL_10_MEGAOHM
limits listed in Table 15.
65. Output 100 V on the calibrator.
Caution Avoid touching the connections when generating a high voltage from the
calibrator.
66. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_DC_VOLTS
Range = 100
Resolution = 100e–6
67. Set the input resistance of the NI 4070/4072 to 10 MΩ by calling
niDMM_SetAttributeViReal64 with the following parameters:
Attribute_ID =
Attribute_Value =
NIDMM_ATTR_INPUT_RESISTANCE
NIDMM_VAL_10_MEGAOHM
68. Call niDMM_Read. Verify that this measurement falls between the limits listed in Table 15.
69. Output –100 V on the calibrator.
70. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
NI 4070/4072 Calibration Procedure 14 ni.com
71. Call niDMM_ConfigureMeasurement with the following parameters:
Function =
NIDMM_VAL_DC_VOLTS
Range = 300
Resolution = 300e–6
72. Call niDMM_Read. Before you apply the voltage, the DMM must be in the 300 V range.
73. Output 300 V on the calibrator.
74. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
75. Output –300 V on the calibrator.
76. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 15.
77. Reset the calibrator for safety reasons.
You have completed verifying the DC voltage of the NI 4070/4072. Select one of the following options:
If you want to continue verifying other modes, go to the Verifying AC Voltage section.
If you do not want to verify other modes and you are performing a
post-adjustment verification, go to the Completing the Adjustment
Procedures section.
If you do not want to verify any additional modes and you are performing a pre-adjustment verification, call
niDMM_close to close
the session.

Verifying AC Voltage

To verify AC voltage of the NI 4070/4072, complete the following steps:
1. Reset the calibrator.
2. Fasten the connectors on one end of the Fluke 5440 cable to the appropriate banana plug connectors on the NI 4070/4072, and fasten the connectors on the other end of the cable to the appropriate calibrator binding posts. Figure 1 shows the correct connections. Table 1 lists the cable connections.
3. Output 5 mV at 1 kHz on the calibrator.
4. Call
© National Instruments Corporation 15 NI 4070/4072 Calibration Procedure
niDMM_reset to reset the NI 4070/4072 to a known state.
5. Call niDMM_ConfigureMeasurement with the following parameters:
Function =
NIDMM_VAL_AC_VOLTS
Range = 0.05
Resolution = 50e–9
6. Call niDMM_Read. Verify that this measurement falls between the limits listed in Table 16.
7. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_AC_VOLTS_DCCOUPLED
Range = 0.05
Resolution = 50e–9
8. Call niDMM_Read. Verify that this measurement falls between the limits listed in Table 16.
9. Output 50 mV at 30 Hz on the calibrator.
10. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_AC_VOLTS_DCCOUPLED
Range = 0.05
Resolution = 50e–9
11. Call niDMM_Read. Verify that this measurement falls between the limits listed in Table 16.
12. Refer to Table 2 for the appropriate calibrator outputs and parameter values as you complete the following steps:
a. On the calibrator, output the value listed in the Calibrator Output
column in Table 2 for the current iteration.
b. Call
niDMM_ConfigureMeasurement with Mode set to
NIDMM_VAL_AC_VOLTS and the remaining parameters as shown
in Table 2 for the current iteration.
c. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 16.
d. Call
niDMM_ConfigureMeasurement again, changing Mode to
NIDMM_VAL_AC_VOLTS_DCCOUPLED.
e. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table 16.
NI 4070/4072 Calibration Procedure 16 ni.com
13. Repeat step 12 for each of the remaining iterations shown in Table 2.

Table 2. niDMM_ConfigureMeasurement Parameters

Calibrator Output niDMM_ConfigureMeasurement Parameters
Iteration
1 50 mV 50 Hz NIDMM_VAL_AC_VOLTS 0.05 50e–9
2 50 mV 1kHz NIDMM_VAL_AC_VOLTS 0.05 50e–9
3 50 mV 1kHz NIDMM_VAL_AC_VOLTS 0.5 500e–9
4 50 mV 20 kHz NIDMM_VAL_AC_VOLTS 0.05 50e–9
5 50 mV 50 kHz NIDMM_VAL_AC_VOLTS 0.05 50e–9
6 50 mV 100 kHz NIDMM_VAL_AC_VOLTS 0.05 50e–9
7 50 mV 300 kHz NIDMM_VAL_AC_VOLTS 0.05 50e–9
Amplitude Frequency Function Range Resolution
50 mV 50 Hz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.05 50e–9
50 mV 1kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.05 50e–9
50 mV 1kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.5 500e–9
50 mV 20 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.05 50e–9
50 mV 50 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.05 50e–9
50 mV 100 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.05 50e–9
50 mV 300 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.05 50e–9
14. Output 500 mV at 30 Hz on the calibrator.
15. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_AC_VOLTS_DCCOUPLED
Range = 0.5
Resolution = 500e–9
16. Call niDMM_Read. Verify that this measurement falls between the limits listed in Table 16.
17. Refer to Table 3 for the appropriate calibrator outputs and parameter values as you complete the following steps:
a. On the calibrator, output the value listed in the Calibrator Output
column in Table 3 for the current iteration.
b. Call
niDMM_ConfigureMeasurement with Mode set to
NIDMM_VAL_AC_VOLTS and the remaining parameters as shown
in Table 3 for the current iteration.
© National Instruments Corporation 17 NI 4070/4072 Calibration Procedure
c. Call niDMM_Read. Verify that this measurement falls between the
limits listed in Table 16.
d. Call
e. Call
niDMM_ConfigureMeasurement again, changing Mode to
NIDMM_VAL_AC_VOLTS_DCCOUPLED.
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 16.

Table 3. niDMM_ConfigureMeasurement Parameters

Calibrator Output niDMM_ConfigureMeasurement Parameters
Iteration
Amplitude Frequency Function Range Resolution
1 500 mV 50 Hz NIDMM_VAL_AC_VOLTS 0.5 500e–9
500 mV 50 Hz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.5 500e–9
2 500 mV 1kHz NIDMM_VAL_AC_VOLTS 0.5 500e–9
500 mV 1kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.5 500e–9
3 500 mV 1kHz NIDMM_VAL_AC_VOLTS 5 5e–6
500 mV 1kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 5 5e–6
4 500 mV 20 kHz NIDMM_VAL_AC_VOLTS 0.5 500e–9
500 mV 20 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.5 500e–9
5 500 mV 50 kHz NIDMM_VAL_AC_VOLTS 0.5 500e–9
500 mV 50 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.5 500e–9
6 500 mV 100 kHz NIDMM_VAL_AC_VOLTS 0.5 500e–9
500 mV 100 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.5 500e–9
7 500 mV 300 kHz NIDMM_VAL_AC_VOLTS 0.5 500e–9
500 mV 300 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 0.5 500e–9
18. Output 5 V at 30 Hz on the calibrator.
19. Call
niDMM_ConfigureMeasurement with the following
parameters:
Function =
NIDMM_VAL_AC_VOLTS_DCCOUPLED
Range = 5
Resolution = 5e–6
20. Call niDMM_Read. Verify that this measurement falls between the limits listed in Table 16.
NI 4070/4072 Calibration Procedure 18 ni.com
21. Refer to Table 4 for the appropriate calibrator outputs and parameter values as you complete the following steps:
a. On the calibrator, output the value listed in the Calibrator Output
column in Table 4 for the current iteration.
b. Call
niDMM_ConfigureMeasurement with Mode set to
NIDMM_VAL_AC_VOLTS and the remaining parameters as shown
in Table 4 for the current iteration.
c. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 16.
d. Call
e. Call
niDMM_ConfigureMeasurement again, changing Mode to
NIDMM_VAL_AC_VOLTS_DCCOUPLED.
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 16.

Table 4. niDMM_ConfigureMeasurement Parameters

Calibrator Output niDMM_ConfigureMeasurement Parameters
Iteration
1 5V 50 Hz NIDMM_VAL_AC_VOLTS 5 5e–6
2 5V 1kHz NIDMM_VAL_AC_VOLTS 5 5e–6
3 5V 1kHz NIDMM_VAL_AC_VOLTS 50 50e–6
4 5V 1kHz NIDMM_VAL_AC_VOLTS 300 300e–6
5 5V 20 kHz NIDMM_VAL_AC_VOLTS 5 5e–6
6 5V 50 kHz NIDMM_VAL_AC_VOLTS 5 5e–6
7 5V 100 kHz NIDMM_VAL_AC_VOLTS 5 5e–6
8 5V 300 kHz NIDMM_VAL_AC_VOLTS 5 5e–6
Amplitude Frequency Function Range Resolution
5V 50 Hz NIDMM_VAL_AC_VOLTS_DCCOUPLED 5 5e–6
5V 1kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 5 5e–6
5V 1kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 50 50e–6
5V 1kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 300 300e–6
5V 20 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 5 5e–6
5V 50 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 5 5e–6
5V 100 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 5 5e–6
5V 300 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 5 5e–6
22. Output 50 V at 30 Hz on the calibrator.
© National Instruments Corporation 19 NI 4070/4072 Calibration Procedure
23. Call niDMM_ConfigureMeasurement with the following parameters:
Function =
NIDMM_VAL_AC_VOLTS_DCCOUPLED
Range = 50
Resolution = 50e–6
24. Call niDMM_Read. Verify that this measurement falls between the limits listed in Table 16.
25. Refer to Table 5 for the appropriate calibrator outputs and parameter values as you complete the following steps:
a. On the calibrator, output the value listed in the Calibrator Output
column in Table 5 for the current iteration.
b. Call
niDMM_ConfigureMeasurement with Mode set to
NIDMM_VAL_AC_VOLTS and the remaining parameters as shown
in Table 5 for the current iteration.
c. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 16.
d. Call
e. Call
niDMM_ConfigureMeasurement again, changing Mode to
NIDMM_VAL_AC_VOLTS_DCCOUPLED.
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 16.

Table 5. niDMM_ConfigureMeasurement Parameters

Calibrator Output niDMM_ConfigureMeasurement Parameters
Iteration
1 50 V 50 Hz NIDMM_VAL_AC_VOLTS 50 50e–6
2 50 V 1kHz NIDMM_VAL_AC_VOLTS 50 50e–6
3 50 V 20 kHz NIDMM_VAL_AC_VOLTS 50 50e–6
4 50 V 50 kHz NIDMM_VAL_AC_VOLTS 50 50e–6
5 50 V 100 kHz NIDMM_VAL_AC_VOLTS 50 50e–6
6 50 V 300 kHz NIDMM_VAL_AC_VOLTS 50 50e–6
NI 4070/4072 Calibration Procedure 20 ni.com
Amplitude Frequency Function Range Resolution
50 V 50 Hz NIDMM_VAL_AC_VOLTS_DCCOUPLED 50 50e–6
50 V 1kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 50 50e–6
50 V 20 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 50 50e–6
50 V 50 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 50 50e–6
50 V 100 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 50 50e–6
50 V 300 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 50 50e–6
26. Call niDMM_ConfigureMeasurement with the following parameters:
Function =
NIDMM_VAL_AC_VOLTS_DCCOUPLED
Range = 300
Resolution = 300e–6
27. Call niDMM_Read. The DMM must be in the 300 V range before you apply the voltage.
28. Output 219 V at 30 Hz on the calibrator.
29. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 16.
30. Refer to Table 6 for the appropriate calibrator outputs and parameter values as you complete the following steps:
a. On the calibrator, output the value listed in the Calibrator Output
column in Table 6 for the current iteration.
b. Call
niDMM_ConfigureMeasurement with Mode set to
NIDMM_VAL_AC_VOLTS and the remaining parameters as shown
in Table 6 for the current iteration.
c. Call
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 16.
d. Call
e. Call
niDMM_ConfigureMeasurement again, changing Mode to
NIDMM_VAL_AC_VOLTS_DCCOUPLED.
niDMM_Read. Verify that this measurement falls between the
limits listed in Table 16.

Table 6. niDMM_ConfigureMeasurement Parameters

Calibrator Output niDMM_ConfigureMeasurement Parameters
Iteration
1 219 V 50 Hz NIDMM_VAL_AC_VOLTS 300 300e–6
2 219 V 1kHz NIDMM_VAL_AC_VOLTS 300 300e–6
3 219 V 20 kHz NIDMM_VAL_AC_VOLTS 300 300e–6
4 219 V 50 kHz NIDMM_VAL_AC_VOLTS 300 300e–6
5 70 V 300 kHz NIDMM_VAL_AC_VOLTS 300 300e–6
© National Instruments Corporation 21 NI 4070/4072 Calibration Procedure
Amplitude Frequency Function Range Resolution
219 V 50 Hz NIDMM_VAL_AC_VOLTS_DCCOUPLED 300 300e–6
219 V 1kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 300 300e–6
219 V 20 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 300 300e–6
219 V 50 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 300 300e–6
70 V 300 kHz NIDMM_VAL_AC_VOLTS_DCCOUPLED 300 300e–6
Loading...
+ 47 hidden pages