The GPIB-PCII/IIA is warranted against defects in materials and
workmanship for a period of two years from the date of shipment, as
evidenced by receipts or other documentation. National Instruments will, at
its option, repair or replace equipment that proves to be defective during the
warranty period. This warranty includes parts and labor.
The media on which you receive National Instruments software are
warranted not to fail to execute programming instructions, due to defects in
materials and workmanship, for a period of 90 days from date of shipment,
as evidenced by receipts or other documentation. National Instruments will,
at its option, repair or replace software media that do not execute
programming instructions if National Instruments receives notice of such
defects during the warranty period. National Instruments does not warrant
that the operation of the software shall be uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the
factory and clearly marked on the outside of the package before any
equipment will be accepted for warranty work. National Instruments will
pay the shipping costs of returning to the owner parts which are covered by
warranty.
National Instruments believes that the information in this manual is
accurate. The document has been carefully reviewed for technical accuracy.
In the event that technical or typographical errors exist, National
Instruments reserves the right to make changes to subsequent editions of
this document without prior notice to holders of this edition. The reader
should consult National Instruments if errors are suspected. In no event
shall National Instruments be liable for any damages arising out of or
related to this document or the information contained in it.
EXCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTS MAKES NO
WARRANTIES
ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A
PARTICULAR PURPOSE
CAUSED BY FAULT OR NEGLIGENCE ON THE PART OF
, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS
. CUSTOMER'S RIGHT TO RECOVER DAMAGES
NATIONAL
INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE
PAID BY THE CUSTOMER
LIABLE FOR DAMAGES RESULTING FROM LOSS OF DATA
USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES,
EVEN IF ADVISED OF THE POSSIBILITY THEREOF. This limitation of the
. NATIONAL INSTRUMENTS WILL NOT BE
, PROFITS,
liability of National Instruments will apply regardless of the form of action,
whether in contract or tort, including negligence. Any action against
National Instrument must be brought within one year after the cause of
action accrues. National Instruments shall not be liable for any delay in
performance due to causes beyond its reasonable control. The warranty
provided herein does not cover damages, defects, malfunctions, or service
failures caused by owner's failure to follow the National Instruments
installation, operation, or maintenance instructions; owner's modification of
the product; owner's abuse, misuse, or negligent acts; and power failure or
surges, fire, flood, accident, actions of third parties, or other events outside
reasonable control.
Copyright
Under the copyright laws, this book may not be copied, photocopied,
reproduced, or translated, in whole or in part, without the prior written
consent of National Instruments Corporation.
Trademarks
NAT4882® and NI-488.2™ are trademarks of National Instruments
Corporation.
Product and company names listed are trademarks or trade names of their
respective companies.
WARNING REGARDING MEDICAL AND
CLINICAL USE OF NATIONAL
INSTRUMENTS PRODUCTS
National Instruments products are not designed with components and testing
intended to ensure a level of reliability suitable for use in treatment and
diagnosis of humans. Applications of National Instruments products
involving medical or clinical treatment can create a potential for accidental
injury caused by product failure, or by errors on the part of the user or
application designer. Any use or application of National Instruments
products for or involving medical or clinical treatment must be performed by
properly trained and qualified medical personnel, and all traditional medical
safeguards, equipment, and procedures that are appropriate in the particular
situation to prevent serious injury or death should always continue to be
used when National Instruments products are being used. National
Instruments products are NOT intended to be a substitute for any form of
established process, procedure, or equipment used to monitor or safeguard
human health and safety in medical or clinical treatment.
FCC/DOC Radio Frequency
Interference Compliance
This equipment generates and uses radio frequency energy and, if not
installed and used in strict accordance with the instructions in this manual,
may cause interference to radio and television reception. This equipment
has been tested and found to comply with the following two regulatory
agencies:
Federal Communications Commission
This device complies with Part 15 of the Federal Communications
Commission (FCC) Rules for a Class B digital device. A Class B device is
distinguishable from a Class A device by the appearance of an FCC ID
number located on the Class B device.
Canadian Department of Communications
This device complies with the limits for radio noise emissions from digital
apparatus set out in the Radio Interference Regulations of the Canadian
Department of Communications (DOC).
Le présent appareil numérique n’émet pas de bruits radioélectriques
dépassant les limites applicables aux appareils numériques de classe B
prescrites dans le règlement sur le brouillage radioélectrique édicté par le
ministère des communications du Canada.
Instructions to Users
These regulations are designed to provide reasonable protection against
interference from the equipment to radio and television reception in
residential areas.
There is no guarantee that interference will not occur in a particular
installation. However, the chances of interference are much less if the
equipment is installed and used according to this instruction manual.
If the equipment does cause interference to radio or television reception,
which can be determined by turning the equipment on and off, one or more
of the following suggestions may reduce or eliminate the problem.
•Operate the equipment and the receiver on different branches of your
AC electrical system.
•Move the equipment away from the receiver with which it is interfering.
•Reorient or relocate the receiver’s antenna.
•Be sure that the equipment is plugged into a grounded outlet and that
the grounding has not been defeated with a cheater plug.
Notice to user: Changes or modifications not expressly approved by
National Instruments could void the user’s authority to
operate the equipment under the FCC Rules.
If necessary, consult National Instruments or an experienced radio/television
technician for additional suggestions. The following booklet prepared by
the FCC may also be helpful: How to Identify and Resolve Radio-TVInterference Problems. This booklet is available from the U.S. Government
Printing Office, Washington, DC 20402, Stock Number 004-000-00345-4.
Contents
About This Manual ..........................................................................xiii
Organization of This Manual......................................................xiii
Conventions Used in This Manual.............................................xiv
Related Documentation ..............................................................xv
This manual contains instructions for installing and configuring the
National Instruments GPIB-PCII/IIA interface board and the NI-488.2
software handler for MS-DOS/Windows graphics applications.
Organization of This Manual
This manual is organized as follows:
•Chapter 1, Introductionand Hardware Installation, lists what you need
to get started and optional equipment you can order, gives a brief
description of the hardware and the NI-488.2 software, and contains
instructions for installing your GPIB-PCII/IIA.
•Chapter 2, NI-488.2 Software Files, describes the software files
contained on your NI-488.2 distribution diskette.
•Chapter 3, Software Installation, contains instructions for installing the
NI-488.2 software files.
•Appendix A, Hardware Specifications, describes the electrical,
environmental, and physical characteristics of the GPIB-PCII/IIA and
the recommended operating conditions.
•Appendix B, Changing Board and Device Characteristics, contains
instructions for changing the default board and device characteristics of
the NI-488.2 handler for MS-DOS/Windows.
•Appendix C, Hardware and Software Configuration, contains
instructions for changing the configuration settings of your
GPIB-PCII/IIA.
•Appendix D, Interactive Software Installation, contains instructions for
running the interactive version of INSTALL.
•Appendix E, Customer Communication, contains forms you can use to
request help from National Instruments or to comment on our products
and manuals.
•The Glossary contains an alphabetical list and description of terms used
in this manual, including abbreviations, acronyms, metric prefixes,
mnemonics, and symbols.
Conventions Used in This Manual
Throughout this manual, the following conventions are used to distinguish
elements of text:
boldBold text denotes menus, menu items, dialog
buttons, or options.
italicItalic text denotes emphasis, a cross reference, or
an introduction to a key concept.
bold italicBold italic text denotes a note, caution, or
warning.
monospaceText in this font denotes text or characters that
are to be literally input from the keyboard,
sections of code, programming examples, and
syntax examples. This font is also used for the
proper names of disk drives, paths, directories,
programs, subprograms, subroutines, device
names, functions, variables, filenames, and
extensions, and for statements and comments
taken from program code.
Italic text in this font denotes that you must
supply the appropriate words or values in the
place of these items.
keyboard—for example, <Ctrl>.
enclosed in angle brackets denotes that you
should simultaneously press the named keys—
for example, <Ctrl-Alt-Del>.
About This Manual
EnterEnter is reserved to mean that the commands
immediately succeeding the word must be typed
into the computer, and then executed by pressing
the <Enter> key on the keyboard.
IEEE 488 andIEEE 488 and IEEE 488.2 refer to the
IEEE 488.2ANSI/IEEE Standard 488.1-1987 and the
ANSI/IEEE Standard 488.2-1987, respectively,
which define the GPIB.
Abbreviations, acronyms, metric prefixes, mnemonics, symbols, and terms
are listed in the Glossary.
Related Documentation
The following documents contain information that you may find helpful as
you read this manual.
•ANSI/IEEE Standard 488.1-1987, IEEE Standard Digital Interface for
Programmable Instrumentation
•ANSI/IEEE Standard 488.2-1987, IEEE Standard Codes, Formats,
Protocols, and Common Commands
•Microsoft MS-DOS User's Guide, Microsoft Corporation
•Microsoft Windows User's Guide, Microsoft Corporation
Customer Communication
National Instruments wants to receive your comments on our products and
manuals. We are interested in the applications you develop with our
products, and we want to help if you have problems with them. To make it
easy for you to contact us, this manual contains comment and configuration
forms for you to complete. These forms are in Appendix E, Customer
This chapter lists what you need to get started and optional equipment you
can order, gives a brief description of the hardware and the NI-488.2
software, and contains instructions for installing your GPIB-PCII/IIA.
What You Need to Get Started
GPIB-PCII/IIA interface board set for one of the following modes:
GPIB-PCIIA
or
GPIB-PCII
3.5 in. NI-488.2 Distribution Diskette for Graphics Applications
MS-DOS/Windows
One of the following operating systems installed on your computer:
MS-DOS version 3.0 or higher (or equivalent)
or
Windows version 3.1 or higher
You can contact National Instruments to order any of the following optional
equipment.
•GPIB Bus Extenders
Parallel – 100 to 120 VAC
Parallel – 220 to 240 VAC
Fiber-optic – 100 to 120 VAC
Fiber-optic – 220 to 240 VAC
Coaxial – 100 to 120 VAC
Coaxial – 220 to 240 VAC
•GPIB Connector Extender
•GPIB-SWITCH Box
•Shielded GPIB Cables*
Type X1 single-shielded cables (1 m, 2 m, or 4 m)
Type X2 double-shielded cables (1 m, 2 m, or 4 m)
*To meet FCC emission limits for this Class B device, you must use a
shielded (Type X1 or X2) GPIB cable. Operating this equipment with
a non-shielded cable may cause interference to radio and television
reception in residential areas.
Hardware Description
The GPIB-PCII/IIA interface board combines the functionality of the
National Instruments GPIB-PCII and GPIB-PCIIA interface boards. It can
be configured to function as either a GPIB-PCII or a GPIB-PCIIA,
depending on the setting of the configuration switches on the board. To
verify the GPIB-PC mode setting of your board, refer to Verifying the
GPIB-PC Mode Setting in Appendix C, Hardware and Software
Configuration, later in this manual.
The GPIB-PCII/IIA, equipped with the NAT4882 ASIC, transforms any
IBM PC, PC/XT, PC AT, or compatible computer into a full-functioning
IEEE 488.2 Talker/Listener/Controller. The NAT4882 controller chip is
fully compatible with the IEEE 488.2 standard.
You can use standard GPIB cables to connect the GPIB-PCII/IIA with up to
14 instruments. If you want to use more instruments, you can order a bus
extender or expander from National Instruments. Refer to Appendix A,
Hardware Specifications, for more information about the GPIB-PCII/IIA
hardware specifications and recommended operating conditions.
Software Description
Your kit includes NI-488.2 software, which National Instruments has
developed for use with the GPIB-PCII/IIA. The NI-488.2 software and
the GPIB hardware transform a general-purpose PC into a GPIB
Talker/Listener/Controller that has complete communications and bus
management capability. Your kit includes the NI-488.2 software for
MS-DOS and for Windows.
Installing the GPIB-PCII/IIA
Warning:Several components on your GPIB-PCII/IIA board can be
damaged by electrostatic discharge. To avoid such damage
in handling the board, touch the antistatic plastic package to
a metal part of your computer chassis before removing the
board from the package.
Perform the following steps to install the GPIB-PCII/IIA:
1.Turn off your computer and all external devices, such as monitors or
tape drives.
2.Unplug the power cord from the wall outlet.
3.Remove the top cover or access port of the I/O channel.
4.Remove the expansion slot cover on the back panel of the computer.
5.Insert the GPIB-PCII/IIA into any unused slot with the GPIB
connector sticking out of the opening on the back panel, as shown in
Figure 1-1. It might be a tight fit, but do not force the board into place.
1 GPIB-PCII/IIA board
2 Back panel
3 PC AT plug-in board
3
Figure 1-1. Installing the GPIB-PCII/IIA
6.Screw the mounting bracket of the GPIB-PCII/IIA to the back panel rail of
the computer.
7.Verify that the GPIB-PCII/IIA is securely installed.
8.Replace the retaining screw of the expansion slot cover if there is one.
9.Replace the cover on the computer.
10. Plug the power cord into the wall outlet.
11. Turn on your computer and external devices.
The GPIB-PCII/IIA interface board is now installed.
Note:Do not attach the GPIB cable to the extension receptacle on the
board until after you run the hardware and software diagnostic
programs, IBDIAG and IBTEST, which are explained in
Chapter 3, Software Installation. These programs may not
function properly if the GPIB cable is attached.
This chapter describes the software files contained on your NI-488.2
distribution diskette.
If you are using the NI-488.2 software for Windows, skip to NI-488.2Software Programs and Files for Windows, later in this chapter. If you are
using the NI-488.2 software for MS-DOS, proceed with the next section,
NI-488.2 Software Programs and Files for MS-DOS.
NI-488.2 Software Programs and Files for
MS-DOS
The NI-488.2 software for MS-DOS contains the following programs and
files.
•GPIB.COM is the NI-488.2 software handler file that is loaded at
system startup by DOS. Handler is a term used by National
Instruments to refer to a loadable device driver.
•IBDIAG.EXE is a program that tests the hardware installation. After
the handler is installed, IBTEST.BAT confirms that both the software
and hardware are installed and functioning properly.
•INSTALL.EXE is a multipurpose, menu-driven program that installs
the NI-488.2 software and tests the hardware and software
configuration. When INSTALL.EXE installs the NI-488.2 software, it
updates CONFIG.SYS, the MS-DOS system configuration file.
•IBTEST.EXE is a program that tests the installation of the NI-488.2
software.
•IBCONF.EXE is a software configuration program that changes the
configuration parameters of the NI-488.2 software handler.
•IBIC.EXE is an interactive control program that executes GPIB
functions you enter from the keyboard. It helps you establish
communication between your computer and your graphics device.
•README.DOC is a Readme file that discusses the NI-488.2 handler.
After you have reviewed all the programs and files on your distribution
diskette, skip to Chapter 3, Software Installation.
NI-488.2 Software Programs and Files for
Windows
The NI-488.2 software for Windows contains the following programs and
files.
•GPIB.DLL is a dynamic link library that is accessed by a Windows
GPIB application as it executes. It contains all of the GPIB functions.
•GPIB.INI is the private profile file that GPIB.DLL uses to determine
the software configuration parameters for each GPIB board and device
in the system. You can modify it by using either the WIBCONF.EXE
program or a text editor.
•WIBCONF.EXE, an MS-DOS application, is a software configuration
program that you can use to change the software parameters and other
data used by the handler. It operates in much the same way as the
GPIB MS-DOS configuration program, IBCONF.
•WIBCONF.PIF contains configuration information about the program
WIBCONF.EXE that is used by Windows when it is executed.
•IBDIAG.EXE, a DOS application, is a program that tests the hardware
settings on your GPIB board. It ensures that the board is properly
installed and that the hardware is accessible.
•IBDIAG.PIF contains configuration information about the program
IBDIAG.EXE that is used by Windows when it is executed.
•WIBTEST.EXE, a Windows application, can be used to test the
software handler. It verifies that the software configuration is
consistent with the GPIB hardware.
•WIBIC.EXE, a Windows application, is the Windows Interface Bus
Interactive Control program that executes GPIB functions that you
enter from the keyboard. It can be used to establish communication
between your computer and your graphics device.
After you have reviewed all the programs and files on your distribution
diskette, proceed to Chapter 3, Software Installation.
This chapter contains instructions for installing the NI-488.2 software files.
If you are using the NI-488.2 software for Windows, skip to the Installing
the NI-488.2 Software for Windows section later in this chapter. If you are
using the NI-488.2 software for MS-DOS, proceed with the following
section, Installing the NI-488.2 Software for MS-DOS.
Installing the NI-488.2 Software for MS-DOS
Follow these steps to install the NI-488.2 software for MS-DOS applications.
Note: You must have approximately 300 kilobytes of free disk space to
install the NI-488.2 software files.
Step 1. Run INSTALL
The quick version of the INSTALL program creates a directory,
C:\GPIB-PC, and copies the NI-488.2 software files to that directory.
Note:If you want to change the default file names and installation
settings of the INSTALL program, refer to Appendix C,
Interactive Software Installation, for information on running the
interactive version of INSTALL.
Complete the following steps to run the quick version of INSTALL.
1.Insert the NI-488.2 distribution diskette into an unused drive.
2.Install the software by entering the following command:
where x is the letter of the drive containing the distribution diskette.
INSTALL copies the NI-488.2 software files to the directory
C:\GPIB-PC. It also modifies the C:\CONFIG.SYS file to include
the following line:
device = \gpib-pc\gpib.com
INSTALL then runs the GPIB-PCII/IIA hardware diagnostic program,
IBDIAG. IBDIAG confirms that the hardware is functioning properly,
and verifies that the software configuration settings are set correctly.
If IBDIAG returns an error message, check the GPIB-PCII/IIA to
ensure that it is not connected to a GPIB cable. IBDIAG may not
function properly if a cable is connected to the GPIB-PCII/IIA.
If IBDIAG still returns an error message, you may need to check the
hardware configuration switch and jumper settings. Refer to Verifying
the Hardware Configuration Settings in Appendix C, Hardware and
Software Configuration, for information on checking and changing the
GPIB-PCII/IIA hardware configuration settings.
3.Restart your computer.
Step 2. Test the Software Installation
The file IBTEST.EXE on your NI-488.2 distribution diskette is a program
that tests the installation of the software. Run IBTEST from the
GPIB-PCII/IIA directory created by INSTALL by entering the following
commands:
cd \gpib-pc
ibtest
If IBTEST returns an error message, check the following:
•If you changed any of the hardware configuration settings on the
GPIB-PCII/IIA interface board, check the current software
configurations of the GPIB-PCII/IIA in IBCONF to make sure that they
match the hardware settings. If you make changes in IBCONF, be sure
to save them. Refer to Step 3. Configure the Software in Appendix C,
Hardware and Software Configuration, for information on changing
the software configuration.
•Confirm that the following line appears in your CONFIG.SYS file:
device = \gpib-pc\gpib.com
Is the file GPIB.COM located in the GPIB-PC directory? If it is not,
run INSTALL again and include it.
•Check the GPIB-PCII/IIA to ensure that it is not connected to a GPIB
cable. IBTEST may not function properly if a cable is connected to
the GPIB-PCII/IIA.
After you complete this checklist, run IBTEST again. If IBTEST still
fails, complete the GPIB-PCII/IIA Hardware and Software ConfigurationForm in Appendix E, Customer Communication, and contact National
Instruments for technical support.
If no errors occur, the NI-488.2 handler for MS-DOS is installed correctly
and you are ready to start using your GPIB-PCII/IIA with your graphics
application program.
If your graphics application program instructs you to change the default
settings of the software handler, refer to Appendix B, Changing Board andDevice Characteristics, for more information on how to run IBCONF and
on the configurable software parameters and their default values.
Installing the NI-488.2 Software for Windows
Complete the following steps to install the NI-488.2 software for Windows.
Note: You must have approximately 300 kilobytes of free disk space to
install the NI-488.2 software files.
Step 1. Run INSTALL
The quick version of the INSTALL program assumes that Windows is
installed in the default directory (C:\WINDOWS). It copies the NI-488.2
files for Windows to C:\WINDOWS and the GPIB destination directory,
GPIB-PCW.
Note:If you want to change the default file names and installation
settings of the INSTALL program, refer to Appendix C,
Interactive Software Installation, for information on running the
interactive version of INSTALL.
Complete the following steps to run the quick version of INSTALL.
1.Insert the NI-488.2 distribution diskette into an unused drive.
2.Install the software by entering the following command:
x
: install/qw
where x is the letter of the drive containing the distribution diskette
(this letter is usually A or B).
INSTALL copies the NI-488.2 software files to the directory
C:\GPIB-PCW and automatically runs the GPIB-PCII/IIA hardware
diagnostic program, IBDIAG. IBDIAG confirms that the hardware is
functioning properly, and verifies that the software configuration
settings are set correctly.
If IBDIAG returns an error message, check the GPIB-PCII/IIA to
ensure that it is not connected to a GPIB cable. IBDIAG may not
function properly if a cable is connected to the GPIB-PCII/IIA.
If IBDIAG still returns an error message, you may need to check the
hardware configuration switch and jumper settings. Refer to Verifying
the Hardware Configuration Settings in Appendix C, Hardware and
Software Configuration, for information on the GPIB-PCII/IIA
hardware configuration settings.
3.Restart your computer.
Step 2. Set Up the Windows Applications
To set up the NI-488.2 Windows applications, complete the following steps:
1.Run Windows Setup in the Main window.
2.Select Set Up Applications from the Options pull-down menu.
3.Add WIBIC and WIBTEST to the Windows Applications window.
If necessary, refer to the Microsoft Windows User's Manual for a more
detailed description of the Windows Setup procedure.
Step 3. Test the Software Installation
The file WIBTEST.EXE is a Windows application that tests the installation
of the software.
Run WIBTEST by selecting the WIBTEST icon in the Windows
Applications window. WIBTEST requires no interaction with the user and
takes about 10 seconds to complete.
If WIBTEST returns an error message, check the following:
•If you changed any of the hardware configuration settings on the
GPIB-PCII/IIA interface board, check the current software
configurations of the GPIB-PCII/IIA board in WIBCONF to make sure
that they match the hardware settings. If you make changes in
WIBCONF, be sure to save them. Refer to Step 3. Configure the
Software in Appendix C, Hardware and Software Configuration, for
information on changing the software configuration.
•Check the GPIB-PCII/IIA interface board to ensure that it is not
connected to a GPIB cable. WIBTEST may not function properly if a
cable is connected to the GPIB-PCII/IIA.
After you complete this checklist, run WIBTEST again. If WIBTEST still
fails, complete the GPIB-PCII/IIA Hardware and Software ConfigurationForm in Appendix E, Customer Communication, and contact National
Instruments for technical support.
If no errors occur, the NI-488.2 handler for Windows is installed correctly
and you are ready to start using your GPIB-PCII/IIA with your graphics
application program.
If your graphics application program instructs you to change the default
settings of the software handler, refer to Appendix B, Changing Board andDevice Characteristics, for more information on how to run the
configuration program and on the configurable software parameters and
their default values.
Appendix B
Changing Board and Device
Characteristics
This appendix contains instructions for changing the default board and device
characteristics of the NI-488.2 handler for MS-DOS/Windows.
Note:Most graphics application software is written to use the default
configuration of the NI-488.2 handler. Do not change the default
parameters of the NI-488.2 handler unless your graphics application
manual instructs you to do so.
Board and Device Characteristics
Two groups of features can be changed in the handler file or configuration file.
The first group consists of the characteristics of the devices attached to your GPIBPCII/IIA. The second group consists of the characteristics of each GPIB-PCII/IIA
installed in the computer. The following sections describe the characteristics of
each group.
Characteristics of the Graphics Devices
Each device used with the GPIB-PCII/IIA has the following characteristics:
•A symbolic name of each device on the GPIB (such as DEV5 or SCANNER).
•A primary and, if used, a secondary address for each device.
•A time limit imposed when executing certain functions. This time limit
ensures that transfers do not hang up the GPIB indefinitely.
•A way to terminate I/O transmissions to and from the device. Some devices
require or append an end of string (EOS) character, such as the ASCII line
feed character, to data strings. Others use the GPIB END message, which is
sent or received via the EOI signal line. Still others use both. Some terminate
I/O only when a predetermined number of bytes are sent or received.
Changing Board and Device CharacteristicsAppendix B
Characteristics of Each GPIB-PCII/IIA
Each GPIB-PCII/IIA board has the following characteristics:
•A symbolic name (either GPIB0 or GPIB1).
•A GPIB primary or secondary address.
•A computer I/O address.
•A GPIB-PC mode (either PCII or PCIIA).
•The capability to be designated as the System Controller of the devices on its
bus.
•A time limit imposed when executing certain functions.
•A way to terminate I/O transmissions to and from the board when executing
board calls—for example, by an EOS character, an END message, and/or a
byte count.
•An interrupt line that the board uses, if any.
•A DMA channel that the board uses, if any.
•High-speed, highest-speed, or normal timing when transmitting data to a
device. With normal timing, there is a delay of at least 2 µs after the data is
placed on the GPIB before the Data Valid (DAV) line is asserted. With highspeed timing, this delay is decreased to about 500 ns, and with highest-speed
timing, it is decreased to about 350 ns. The default setting of 2 µs is
recommended, as some devices cannot communicate at higher speeds.
Default Configurations
The default configurations of the software were set at the factory. For example, the
default device names of the 32 GPIB devices are DEV1 through DEV32, but your
application manual may instruct you to assign a more descriptive name to a
particular device, such as SCANNER.
Appendix BChanging Board and Device Characteristics
Note:Do not give a GPIB device the same name as a directory or file on your
disk drive.
You can also use the configuration programs, IBCONF or WIBCONF, to examine
the current default settings in the handler file.
If you do not make changes to the configuration, the default characteristics of the
software remain in effect.
Primary Default Characteristics
The following are the primary default characteristics of the handler:
•The GPIB-PC mode is assumed to be the default setting. See
Appendix C, Hardware and Software Configuration, for more information on
the GPIB-PC mode.
•There are 32 active devices with symbolic names DEV1 through DEV32.
•GPIB addresses of these devices are the same as the device number. For
example, DEV1 is at address 1.
•The first 16 devices are assigned to GPIB0 as their access board, and the
second 16 devices are assigned to GPIB1 as their access board. GPIB0 is the
symbolic name of the first GPIB-PCII/IIA board in your system. If you have
an additional GPIB-PCII/IIA board in your system, its symbolic name is
GPIB1.
•Each GPIB-PCII/IIA is System Controller of its independent bus and has a
GPIB address of 0.
•The END message is sent with the last byte of each data message to a device.
Each data message that is read from a device automatically terminates when
END is received. No EOS character is recognized.
•The time limit on I/O and wait function calls is approximately 10 seconds.
•GPIB0 is a GPIB-PCII board at base I/O address 02B8 hex using DMA
channel 1 and no interrupt line, or GPIB0 is a GPIB-PCIIA board at base I/O
address 02E1 hex using DMA channel 1 and no interrupt line.
Changing Board and Device CharacteristicsAppendix B
Repeat Addressing (Device Characteristic Only)
Normally, the GPIB address of a device is issued before each device function
pertaining to that device. If you enter yes for the Repeat Addressing field, this
addressing occurs even if the device is already properly addressed. The default
option for this characteristic is no.
Changing Board and Device Characteristics
If you are using the NI-488.2 software for Windows, skip to Changing
Characteristics for Windows later in this appendix. If you are using theNI-488.2 software for MS-DOS, proceed to the following section, Changing
Characteristics for MS-DOS.
Changing Characteristics for MS-DOS
Use the software configuration program, IBCONF, to change the board and device
characteristics.
Running IBCONF
Complete the following instructions to change a board or device characteristic
using IBCONF.
Note:Never run IBCONF from the distribution diskette without
write-protecting the diskette, as doing so will modify the
master copy.
1.Run IBCONF by entering the following commands:
cd \gpib-pc
ibconf
2.Press any key to display the configuration map.
3.When the configuration map appears, make sure that the board or device is
highlighted and press <F8>.
4.Using the arrow keys, move the highlight to the characteristic you want to
change.
Appendix BChanging Board and Device Characteristics
5.Use the arrow keys to change the value of the characteristic.
6.Press <F9> to stop editing.
7.Press <F9> once again to quit IBCONF.
8.IBCONF displays a prompt asking if you want to save your changes. Enter y
for yes.
9.IBCONF displays a prompt asking if you want to update the handler in
memory. Enter y for yes. In some cases, IBCONF instructs you to restart
your computer.
Changes to other configurable parameters can be made in the same manner.
Changing Characteristics for Windows
Use the software configuration program, WIBCONF, to change the board and
device characteristics.
Running WIBCONF
Complete the following instructions to change a board or device characteristic
using WIBCONF. WIBCONF edits the file GPIB.INI.
Note:Never run WIBCONF from the distribution diskette without
write-protecting the diskette, as doing so will modify the
master copy.
1.Select the WIBCONF icon by double-clicking on it.
2.WIBCONF displays a prompt asking for parameters. If you used the quick
version of INSTALL, press <Enter>.
If you used the interactive version of INSTALL and did not install the
software in the default directory, C:\WINDOWS, enter the name of your
Windows directory and the path to GPIB.INI—for example,
E:\WINDOWS\GPIB.INI.
The GPIB-PC mode of the GPIB-PCII/IIA is set using switch 9 in switch block
U2. Check to ensure that this switch is set to the mode indicated on the
identifying label on the mounting bracket of the interface board, beside the
GPIB connector.
Figure C-2 shows the GPIB-PC mode selection switch set for GPIB-PCII mode
and GPIB-PCIIA mode.
Key
= not used to select the GPIB-PC mode
1
PCII
a. GPIB-PCII
0
U2
OFF
Mode Selected
321
4
67895
PCIIA
1
U2
OFF
PCII
b. GPIB-PCIIA
Mode Selected
0
321
4
67895
PCIIA
Figure C-2. GPIB-PC Mode Selection Settings
If this switch is not set to the mode indicated on the identifying label, set the
switch to the appropriate mode. To select GPIB-PCII mode, push the switch
down on the side labeled PCII. To select GPIB-PCIIA mode, push the switch
down on the side labeled PCIIA.
If IBDIAG returns an error message and no cables are connected to the GPIBPCII/IIA, complete the instructions in this section to verify the hardware
configuration settings.
Table C-1 shows the factory settings and available configurations for the
switches and jumpers on the GPIB-PCII/IIA in GPIB-PCII mode.
Table C-2 shows the factory settings and available configurations for
the switches and jumpers in GPIB-PCIIA mode.
Table C-1. Factory Default Settings and Available Configurations for
GPIB-PCII Mode
GPIB-PCIIDefaultAvailable
Base I/O Address
(hex)
DMA Channel11, 2, 3, or Not Used
Interrupt Line
(IRQ)
Shield GroundConnectedConnected or disconnected
2B8100 to 3F8 in increments of 8
72, 3, 4, 5, 6, 7, or Not Used
Table C-2. Factory Default Settings and Available Configurations for
GPIB-PCIIA Mode
GPIB-PCIIDefaultAvailable
Base I/O Address
(hex)
DMA Channel11, 2, 3, or Not Used
Interrupt Line
(IRQ)
Shield GroundConnectedConnected or disconnected
Check Tables C-1 and C-2 to make sure that the factory default settings of your
GPIB-PCII/IIA board are the same as the values you entered when prompted by
the IBDIAG program. If they are not, run IBDIAG again, and enter the default
values when prompted. For instructions on running IBDIAG again, refer to
Appendix D, Interactive Software Installation.
If IBDIAG still returns an error message, one of the switches and jumpers on
your GPIB-PCII/IIA board may be configured to the same setting as another
board or device in your computer. Check the manuals of any other devices or
boards you are using to determine if they conflict with the configuration settings
of your GPIB-PCII/IIA board. If the manuals are not available, check Tables C3, C-4, and C-5 later in this appendix for possible conflicts.
If there is a conflict, you need to change the conflicting setting on either your
GPIB-PCII/IIA board or the other board or device and run IBDIAG again. To
change a hardware configuration setting on your GPIB-PCII/IIA board, follow
the instructions in the following section, Changing Hardware Configuration
Settings.
If you complete these instructions and IBDIAG still returns an error message,
complete the GPIB-PCII/IIA Hardware and Software Configuration Form in
Appendix E, Customer Communication, and contact National Instruments for
technical support.
Changing Hardware Configuration Settings
Complete the following steps to change the configuration settings of your GPIBPCII/IIA interface board.
Step 1. Configure the Hardware
Base I/O Address Selection
GPIB-PCII mode and GPIB-PCIIA mode use different regions of I/O address
space. By default, GPIB-PCII mode uses base I/O address 2B8 hex and GPIBPCIIA mode uses base I/O address 2E1 hex. To select a different base I/O
address for your GPIB-PCII/IIA, follow the instructions that pertain to the
GPIB-PC mode setting of your board.
Switch block U2 is used to set the address for address lines A3 through A9. The
addresses are in a consecutive block of eight beginning on any multiple of 8
between 100 and 3F8 hex. For example, for the default address, 2B8 hex, the
GPIB-PCII uses the address space 2B8 through 2BF hex.
Press the side marked 1 to select a binary value of 1 for the corresponding
address bit. Press the 0 side of the switch to select a value of 0 for the
corresponding address bit.
To change the base I/O address, press each switch to the desired position, then
check each switch to make sure it is pressed down all the way.
Figure C-3 shows two possible switch settings. Each of the address selections
shows how the base I/O address was calculated from the switch positions.
The GPIB-PCIIA base I/O address is set using switches 4 and 5 of the switch
block at U2. The four possible base I/O addresses are 2E1, 22E1, 42E1, and
62E1 hex.
Figure C-4 shows the switch settings for the four possible base I/O addresses
and the address space used for each setting. Figure C-4a shows how the base
I/O address was calculated from the switch positions.
Key
= not used to select the base I/O address
Push this side down (off) for logic 1
Push this side down (on) for logic 0
U2
0
1
OFF
321
4
67895
A14
A13
Binary Hex
0
0
0
0
0
0
0
2
1
0
Addresses Used
02E1
12E1
82E1
92E1
a. Switch Set to Base
I/O Address hex 2E1
06E1
16E1
86E1
96E1
0AE1
1AE1
8AE1
9AE1
0EE1
1EE1
8EE1
9EE1
1
1
E
1
0
0
0
1
0
1
Figure C-4. Base I/O Address Switch Settings for GPIB-PCIIA Mode
To change the base I/O address, locate the switches at U2, press each switch to
the desired position, and check each switch to make sure it is pressed down all
the way.
If you change the base I/O address setting from the default setting, record the
new setting on the GPIB-PCII/IIA Hardware and Software Configuration Form
in Appendix E, Customer Communication. Remember that you must run the
IBDIAG diagnostic program and configure the software, as explained later in
this appendix, after you change any hardware configuration settings.
Possible Conflicts
Table C-3 lists some of the I/O addresses used by other PC plug-in interface
boards and adapters. This is not a complete list, but it may help in determining
possible address conflicts. Symptoms of I/O address conflicts vary widely. At
one extreme, conflicts can prevent the computer from booting. At the other
extreme, they can cause problems that do not surface until a considerable
amount of time has elapsed. When conflicts do surface, the problems can
exhibit themselves simply as strange behavior.
National Instruments has made every effort to select a default base I/O address
that will work. However, because of the numerous different interface boards
available for use in the PC, it is not possible to select a base I/O address that is
guaranteed to work in all systems.
Note:In GPIB-PCII mode, eight consecutive addresses are used, while in
GPIB-PCIIA mode, sixteen addresses spread throughout the upper
address space are used.
218
219 to 21E
21F
220 to 23F
240 to 25F
248
258
260 to 27F
259 to 267
268
269 to 277
278 to 27F
280 to 29F
2A0 to 2A7
2A8
2A9 to 2AF
2B0 to 2DF
2B8
2B9 to 2BF
2C0 to 2DF
2E0 to 2FF
2E1
2E2 to 2E3
2E4 to 2E7
2E8
2E9 to 2F7
2F8 to 2FF
PC and PC AT Game Controller, reserved
LIM Expanded Memory Card
PC Expansion Unit
AT-DIO-24
Table C-3. I/O Addresses Used by Other Devices (Continued)
I/O Address
Range (Hex)Device
300 to 31F
300 to 30F
320 to 32F
330 to 347
348 to 357
358 to 35F
360 to 363
364 to 367
368 to 36B
36C to 36F
370 to 377
378 to 37F
380 to 38C
380 to 389
390 to 393
394 to 39F
3A0 to 3A9
3AA to 3AF
3B0 to 3BF
3C0 to 3CF
3D0 to 3DF
3E0 to 3EF
3F0 to 3F7
3F8 to 3FF
PC, AT Prototype card
3Com EtherLink (default)
IBM PC/XT Fixed Disk Controller
DCA 3278
PC Network (low address)
Reserved
PC Network (high address)
Reserved
PC, AT Parallel Printer Port 1
SDLC Communications
Bisynchronous (BSC) Communications (alternate)
Cluster Adapter 0
The GPIB-PCII/IIA interface board can use any of six interrupt lines available
on the PC, or no interrupts at all. The interrupt line is selected using the jumper
sets labeled IRQ2 through IRQ7 (see Figure C-1). The GPIB-PCII/IIA is set at
the factory to use line 7. To select another interrupt line, place the supplied
jumper across the two pins adjacent to the label designating that interrupt line.
Figure C-5 shows the selection of interrupt line 7.
Figure C-5. Default Interrupt Jumper Setting for GPIB-PCII Mode
If you do not want to use interrupts, you must logically disconnect the GPIBPCII/IIA from the IRQ lines by selecting NONE for the interrupt line when you
run the software configuration program, as explained in Step 3. Configure theSoftware, later in this appendix. The board can remain in the backplane and no
jumpers have to be moved or changed.
Shared Interrupts in GPIB-PCIIA Mode
Multiple GPIB-PCII/IIA boards can share the same interrupt level if they are all
configured for GPIB-PCIIA mode.
If you use the GPIB-PCII/IIA in GPIB-PCIIA mode and you want to change the
interrupt line, you must set switches I0, I1, and I2 in switch block U2 to the line
setting in addition to setting the interrupt jumpers.
Figure C-6 shows the switch and jumper settings for the default interrupt setting,
IRQ7, and shows how the interrupt setting was calculated from the switch
positions. Figure C-7 shows the switch and jumper settings for the five
remaining interrupt lines.
Key
= not used to select the interrupt line
OFF
U2
0
321
4
67895
I0
I1
I2
IRQ2
IRQ3
IRQ4
Binary Hex
IRQ6
IRQ5
1
1
1
IRQ7
7
1
Figure C-6. Default Interrupt Jumper Setting for GPIB-PCIIA Mode
If you change the interrupt jumper setting from the default setting, record the
new setting on the GPIB-PCII/IIA Hardware and Software Configuration Form
in Appendix E, Customer Communication. Remember that you must run the
IBDIAG diagnostic program and configure the software, as explained later in
this appendix, after you change any hardware configuration settings.
Possible Conflicts
Table C-4 lists some of the interrupt lines used by other PC plug-in interface
boards and adapters. This is by no means a complete list, but it may help in
determining possible interrupt conflicts. Symptoms of interrupt conflicts vary
widely. Conflicts can prevent the computer from booting. They may also cause
repeated time outs on GPIB function calls. When conflicts do surface, the
problems can exhibit themselves simply as strange behavior.
National Instruments has made every effort to select a default interrupt line that
will work. However, because of the numerous different interface boards
available for use in the PC, it is not possible to select an interrupt line that is
guaranteed to work in all systems. Therefore, be certain of your system’s
interrupt assignments before proceeding with installation.
Table C-4. Interrupt Lines Used by Other Devices
IRQDevice
7Parallel Port 1
Data Acquisition and Control (default)
GPIB-PCII/IIA
Table C-4. Interrupt Lines Used by Other Devices (Continued)
IRQDevice
4Serial Port 1
BSC
BSC Alt.
SDLC
3Serial Port 2
BSC
BSC Alt.
Cluster (Primary)
PC Network (default)
PC Network Alt. (default)
SDLC
WD EtherCard + (default)
3Com EtherLink (default)
2IRQ Chain for PC AT
1Keyboard Controller Output Buffer Full
0Timer Channel 0 Output
DMA Channel Selection
The GPIB-PCII/IIA can use DMA channels 1, 2, or 3, or no DMA at all. The
DMA channel is selected by the jumper sets labeled DRQ1 through DACK 3
(see Figure C-1).
Each DMA channel consists of two signal lines, as shown in Table C-5.
You must position two jumpers to select a DMA channel. One jumper selects
the DMA Request line, and the other selects the DMA Acknowledge line. You
must move these two jumpers as a pair, and the DMA Acknowledge and DMA
Request lines that you select must have the same numeric suffix for proper
operation.
Figure C-8 shows the jumper position for the default DMA channel setting,
DMA channel 1.
DMA
Acknowledge
DRQ1
DRQ2
DACK1
DRQ3
DACK2
DMA
Request
DACK3
Figure C-8. DMA Channel Jumper Setting for DMA Channel 1
If you do not want to use DMA for GPIB transfers (the GPIB-PCII/IIA
alternatively can use programmed I/O transfers), you must logically disconnect
the GPIB-PCII/IIA from the DMA lines by selecting NONE for the DMA line
when you run the software configuration program, as explained in Step 3.Configure the Software, later in this appendix. The board can remain in the back
panel and no jumpers have to be moved or changed.
Note:Disabling DMA will decrease performance substantially.
If you change the DMA jumper setting from the default setting, record the new
setting on the GPIB-PCII/IIA Hardware and Software Configuration Form in
Appendix E, Customer Communication. Remember that you must run the
IBDIAG diagnostic program and configure the software, as explained later in
this appendix, after you change any hardware configuration settings.
Possible Conflicts
There are only three DMA channels that can be used by the GPIB-PCII/IIA
interface board. If any device uses DMA channel 1, change the DMA channel
used by either the GPIB-PCII/IIA or the other device to DMA channel 2 or 3. If
no DMA channel is available, configure the software to run without DMA using
the software configuration program.
Shield Ground Configuration
The GPIB-PCII/IIA is set at the factory with the jumper in place to connect the
logic ground of the GPIB-PCII/IIA to its shield ground. This configuration
minimizes the EMI emissions.
Caution:The GPIB-PCII/IIA was tested for compliance with FCC
standards with the shield ground connected to logic ground.
Removing the jumper might cause EMI emissions to exceed any
or all of the applicable standards.
If your application requires that logic ground be disconnected from shield
ground, refer to Figure C-1 to locate the shield ground jumper W1 on the GPIBPCII/IIA. Remove the jumper and place it across only one of the jumper pins,
as shown in Figure C-9.
Be sure to record the jumper setting on the GPIB-PCII/IIA Hardware andSoftware Configuration Form in Appendix E, Customer Communication.
Step 2. Run the IBDIAG Diagnostic Program
If you have changed any hardware configuration settings, you must run the
diagnostic program, IBDIAG, again. For instructions on running IBDIAG,
refer to Appendix C, Interactive Software Installation.
Step 3. Configure the Software
If you have changed any hardware configuration settings, you must make
appropriate changes to the NI-488.2 handler. To do so, you must run the
configuration program, IBCONF or WIBCONF, and edit the board parameters
(such as the base I/O address or interrupt line) that you have changed.
If you are using the NI-488.2 software for Windows, skip to Configuring theSoftware for Windows, later in this appendix. If you are using the NI-488.2
software for MS-DOS, proceed to the next section, Configuring the Software for
Use the software configuration program, IBCONF, to change or examine the
configuration parameters of the NI-488.2 software handler file, GPIB.COM.
Perform the following steps to change a software parameter to match changes to
the hardware configuration.
1.Run IBCONF by entering the following commands:
cd \gpib-pc
ibconf
2.Press any key to display the configuration map.
3.When the configuration map appears, make sure that GPIB0 is highlighted
and press <F8> to edit the board configurations.
4.Using the arrow keys, move the highlight to the board parameter you want
to change.
5.Use the arrow keys to change the value of the parameter to match the
hardware configuration.
6.Press <F9> to stop editing.
7.Press <F9> once again to quit IBCONF.
8.IBCONF displays a prompt asking if you want to save your changes. Enter
y for yes.
9.IBCONF displays a prompt asking if you want to update the handler in
memory. Enter y for yes. In some cases, IBCONF instructs you to restart
your computer.
Changes to other configurable parameters can be made in the same manner.
Use the software configuration program, WIBCONF, to change or examine the
configuration parameters of the NI-488.2 software handler. WIBCONF is not a
Windows application. You can run it to configure the software directly from
Windows.
WIBCONF operates in much the same way as the NI-488.2 MS-DOS
configuration program, IBCONF. However, the following differences exist
between IBCONF for MS-DOS and WIBCONF for Windows:
•Instead of configuring the handler file, GPIB.COM, WIBCONF actually
modifies the configuration file, GPIB.INI.
•Auto-configuration for NI-488.2 GPIB DLLs is not supported.
•The option of configuring the loaded driver does not apply.
Perform the following steps to change a software parameter to match changes to
the hardware configuration.
1.Select the WIBCONF icon by double-clicking on it.
2.WIBCONF displays a prompt asking for parameters. If you used the quick
version of INSTALL, press <Enter>.
If you used the interactive version of INSTALL and did not install the
software in the default directory, C:\WINDOWS, enter the name of your
Windows directory, such as E:\WINDOWS\GPIB.INI, for example.
3.Press any key to display the configuration map.
4.When the configuration map appears, make sure that GPIB0 is highlighted
and press <F8> to edit the board configurations.
5.Using the arrow keys, move the highlight to the board parameter you want
to change.
6.Use the arrow keys to change the value of the parameter to match the
hardware configuration.
This appendix contains instructions for running the interactive version of
INSTALL.
By using he interactive version of INSTALL, you can change the default
filenames and directories used in installation. Enter the following command
to run INSTALL:
x
:install
where x is the letter of the drive containing the distribution diskette.
INSTALL displays a main menu with three options: DOS GPIB
Installation, Windows GPIB Installation, and Return to
DOS.
If you are using the NI-488.2 software for Windows, skip to INSTALL for
Windows, later in this appendix. If you are using the NI-488.2 software forMS-DOS, proceed to the following section, INSTALL for MS-DOS.
INSTALL for MS-DOS
Select DOS GPIB Installation from the main menu.
INSTALL displays a DOS installation menu with three options: GPIB
Installation, Diagnostics, and Return to DOS.
If you have already installed the NI-488.2 software files and need to run
IBDIAG again, skip to Running IBDIAG for MS-DOS, later in this section.
Select GPIB Installation to install the NI-488.2 software files.
INSTALL displays a screen asking you to confirm or change the default
installation settings, the destination directory and the boot drive.
•The destination directory setting is the drive and directory name of the
location to which INSTALL will copy the NI-488.2 software. The
default value of this setting is c:\gpib-pc. In most cases, the drive
name given in this field should be the drive off of which you start your
computer.
•The boot drive setting is the drive name of the disk (hard disk or
diskette) off of which you start your computer. If you start from a
diskette, this field typically contains A: or B:. If you start from a hard
disk, this field typically contains C:. The default value of this setting
is C:.
To change these settings, use the arrow keys to move the highlight to the
appropriate field and enter the new value. Press <Enter> to exit the menu.
INSTALL then checks the information you entered. If there is a problem
accessing any of the specified drives, it displays an error message and
prompts you to re-enter the offending setting. If there are no errors,
INSTALL installs the software files.
Note:Press <Esc> at any time to abort the installation.
When the installation is complete, INSTALL asks to modify your
CONFIG.SYS file. If you enter yes, INSTALL adds the following line toyour CONFIG.SYS file:
device=
where
the NI-488 software installed on your computer, INSTALL replaces the old
version's information in CONFIG.SYS with the new information.
If you enter no to the above prompt, INSTALL displays a message
informing you of the correct line that you should add to your CONFIG.SYS
file.
is the destination directory. If you have a previous version of
Appendix DInteractive Software Installation
Running IBDIAG for MS-DOS
Upon completion of the software installation, INSTALL returns to the main
menu. Select Diagnostics to run the hardware diagnostic program,
IBDIAG. IBDIAG confirms that the hardware is functioning properly, and
verifies that the software configuration settings are set correctly.
If IBDIAG returns an error message, check the GPIB-PCII/IIA to ensure
that it is not connected to a GPIB cable. IBDIAG may not function
properly if a cable is connected to the GPIB-PCII/IIA.
If IBDIAG still returns an error message, you may need to check the
hardware configuration switch and jumper settings. Refer to Appendix C,
Hardware and Software Configuration, for information on checking and
changing the GPIB-PCII/IIA hardware configuration settings.
If there are no errors, INSTALL displays the DOS installation menu.
Exiting INSTALL
You have completed the interactive version of INSTALL for
MS-DOS.
Select ReturntoDOS from the DOS installation menu to exit INSTALL.
Proceed to Step 2. Test the Software Installation in Chapter 3, SoftwareInstallation, to complete the installation procedure.
INSTALL for Windows
Select Windows GPIB Installation from the main menu.
INSTALL displays a Windows installation menu with three options:
Windows Installation, Diagnostics, and Return to
Windows.
If you have already installed the NI-488.2 software files and need to run
IBDIAG again, refer to Running IBDIAG for Windows, later in this section.
Select Windows Installation to install the NI-488.2 software files.
INSTALL displays a screen asking you to confirm or change the default
installation settings, the destination directory and the Windows directory.
•The destination directory setting is the drive and directory name of the
location to which INSTALL will copy the NI-488.2 software. The
default value of this setting is c:\gpib-pcw. In most cases, the
drive name given in this field should be the drive off of which you start
your computer.
•The Windows directory setting is the drive and directory name of the
location to which INSTALL will copy the Windows GPIB software.
The default value of this setting is c:\windows.
To change these settings, use the arrow keys to move the highlight to the
appropriate field and enter the new value. Press <Enter> to exit the menu.
INSTALL then checks the information you entered. If there is a problem
accessing any of the specified drives, it displays an error message and
prompts you to re-enter the offending setting. If there are no errors,
INSTALL installs the software files.
Note:Press <Esc> at any time to abort the installation.
Running IBDIAG for Windows
Upon completion of the software installation, INSTALL returns to the main
menu. Select Diagnostics to run the hardware diagnostic program,
IBDIAG. IBDIAG confirms that the hardware is functioning properly, and
verifies that the software configuration settings are set correctly.
If IBDIAG returns an error message, check the GPIB-PCII/IIA to ensure
that it is not connected to a GPIB cable. IBDIAG may not function
properly if a cable is connected to the GPIB-PCII/IIA.
If IBDIAG still returns an error message, you may need to check the
hardware configuration switch and jumper settings. Refer to Appendix C,
Hardware and Software Configuration, for information on checking and
changing the GPIB-PCII/IIA hardware configuration settings.
If there are no errors, INSTALL displays the Windows installation menu.
Exiting INSTALL
You have now completed the interactive version of INSTALL for
Windows.
Select ReturntoDOS from the Windows installation menu to exit
INSTALL. Proceed to Step 2. Set Up the Windows Applications in Chapter
3, Software Installation, to complete the installation procedure.
For your convenience, this appendix contains forms to help you gather the
information necessary to help us solve technical problems you might have
as well as a form you can use to comment on the product documentation.
Filling out a copy of the Technical Support Form before contacting
National Instruments helps us help you better and faster.
National Instruments provides comprehensive technical assistance around
the world. In the U.S. and Canada, applications engineers are available
Monday through Friday from 8:00 a.m. to 6:00 p.m. (central time). In other
countries, contact the nearest branch office. You may fax questions to us at
any time.
Corporate Headquarters
(512) 795-8248
Technical support fax:(800) 328-2203
Photocopy this form and update it each time you make changes to your
software or hardware, and use the completed copy of this form as a
reference for your current configuration. Completing this form accurately
before contacting National Instruments for technical support helps our
applications engineers answer your questions more efficiently.
If you are using any National Instruments hardware or software products
related to this problem, include the configuration forms from their user
manuals. Include additional pages if necessary.
Name
Company
Address
Fax ( ) Phone ( )
Computer brand
Model Processor
Operating system
Speed MHzRAM MB
Display adapter
Mouse yes no
Other adapters installed
Hard disk capacity MBBrand
Instruments used
National Instruments hardware product model
Revision
Configuration
(continues)
National Instruments software product
Version
Configuration
Application Program
Version
Graphics Devices Used
The problem is
List any error messages
The following steps will reproduce the problem
GPIB-PCII/IIA Hardware and
Software Configuration Form
Record the settings and revisions of your hardware and software on the line
located to the right of each item. Complete this form each time you revise
your software or hardware configuration, and use this form as a reference
for your current configuration. Completing this form accurately before
contacting National Instruments for technical support helps our applications
engineers answer your questions more efficiently.
National Instruments Products
•GPIB-PCII/IIA Hardware Revision
•GPIB-PC Mode of GPIB-PCII/IIA
•NI-488.2 Software Version Number on Disk
•Shield Ground Connected to Logic Ground (yes or no)
•Board Settings
Base I/O
Address
gpib0
gpib1
gpib2
gpib3
•GPIB-PC Handler Type set in IBCONF or WIBCONF
•Interrupt Level set in IBCONF or WIBCONF
•DMA Channel set in IBCONF or WIBCONF
•Base I/O Address set in IBCONF or WIBCONF
•Shield Ground set in IBCONF or WIBCONF
Interrupt
Level
DMA
Channel
Other Products
•Computer Make and Model
•Microprocessor
•Clock Frequency
•Type of Monitor Card Installed
•DOS or Windows Version
•Graphics Devices Used
•Application Program
Version
•Other Boards in System
•Base I/O Addresses of Other Boards
•DMA Channels of Other Boards
•Interrupt Levels of Other Boards
Documentation Comment Form
National Instruments encourages you to comment on the documentation
supplied with our products. This information helps us provide quality
products to meet your needs.
Title:Getting Started with Your GPIB-PCII/IIA and the
NI-488.2™ Software for MS-DOS/Windows Graphics
Applications
Edition Date:October 1994
Part Number:320269B-01
Please comment on the completeness, clarity, and organization of the
manual.
(continues)
If you find errors in the manual, please record the page numbers and
describe the errors.
Thank you for your help.
Name
Title
Company
Address
Phone ( )
Mail to:Technical Publications
National Instruments Corporation
6504 Bridge Point Parkway, MS 53-02
Austin, TX 78730-5039
Fax to:Technical Publications
National Instruments Corporation
MS 53-02
(512) 794-5678
Glossary
PrefixMeaningValue
nµ-
m-
c-
M-
°degrees
%percent
Aamperes
ACalternating current
ANSIAmerican National Standards Institute
ASCIIAmerican Standard Code for Information
ASICapplication-specific integrated circuit
CCelsius
DAVData Valid
DLLdynamic link library
DMAdirect memory access
EMIelectromagnetic interference
EOIend or identify
EOSend of string
FCCFederal Communications Commission
GPIBGeneral Purpose Interface Bus
hexhexadecimal
Hzhertz
IEEEInstitute of Electrical and Electronic Engineers
ininches
I/Oinput/output
mmeters
MBmegabytes of memory
PCpersonal computer
RAMrandom-access memory
sseconds
VDCvolts direct current