National Instruments™,NI™, and ni.com™are trademarks of National Instruments Corporation. Product and company names mentioned herein are
trademarks or trade names of their respective companies.
This document contains information and step-by-step instructions for
loading and running the calibration procedure for the NI 5102 using
Calibration Executive.
What Is Calibration?
Calibration consists of verifying the measurement accuracy of a device
and correcting for any measurement error. Verification is measuring the
performance of a device and comparing the results to the factory
specifications of the device. National Instruments calibrates every 5102
device at the factory. During the factory calibration process, the calibration
constants are stored on the EEPROM. These values are loaded from
memory and used as needed by the device.
Why Should You Calibrate?
The accuracy of electronic components drifts with time and temperature,
which can affect measurement accuracy as the device ages. Calibration
restores your device to its specified accuracy and ensures that it still meets
National Instruments standards.
How Often Should You Calibrate?
The measurement accuracy requirements of your application determine
how often you should externally calibrate your NI 5102 device.
National Instruments recommends that you perform a complete
calibration at least once every year. You can shorten this interval to
90 days or six months based on the demands of your application.
You can also use the verification procedure at a regular interval to
determine if your oscilloscope needs adjustment.
Calibrating the NI 5102 with Calibration Executive2ni.com
Equipment and Other Test Requirements
This section describes the equipment, documentation, software, and test
conditions needed for calibration.
Test Equipment
Tables 1 and 2 list specifications for equipment you can use to calibrate
your NI 5102 with Calibration Executive.
Recommended Equipment
Table 1 lists the equipment that National Instruments recommends for this
procedure.
Table 1. Required Equipment Specifications for NI 5102 Calibration
Required
Equipment
Scope CalibratorWav ete k 95 00
Active HeadWav ete k 95 10
Recommended
Equipment
Scope Calibrator
Active Head or
better
If you do not have the recommended instruments, use the accuracy
requirements listed above to select a substitute calibration standard.
Refer to Table 2 for a list of alternative instruments you can use for the
calibration procedure. Although these instruments are acceptable,
National Instruments recommends that you use the instruments from
Table 1.
Table 2 contains a list of alternative equipment you can use to calibrate
your NI 5102 with Calibration Executive.
Table 2. Alternative Equipment Specifications for NI 5102 Calibration
Required
Equipment
CalibratorFluke 5700
High-Speed Signal
Generator
DMMHP 34401Input Impedance2-wire resistance accuracy of
Recommended
Equipment
Calibrator
Tektronix SG503Bandwidth1.5% amplitude flatness
Parameter
Measured
Vertical Offset0 VDC ±0.1 mV
Ve r t i c a l G a i nDC ±40 mV to ±4.5 V, ±0.1%
AC Couplingsine wave 9–13 Hz ±100 ppm,
Necessary Specifications
into 1 MΩ
1.8 Vpp ±2%into1MΩ
for leveled sine wave
100 kHz–15 MHz ±50 ppm,
0.2–1.9 Vpp ± 2% into 1 MΩ
0.25% for 1 MΩ measurement
Connectors
If you are using the Wavetek 9500 and the Active Head, you do not need
special connectors.
If you are using the alternative instruments listed in Table 2, you need the
following connection equipment:
•50 Ω BNC cable
•50 Ω BNC terminator
•BNC T-connector
Documentation
This section describes the documentation you need to calibrate your 5102
device. In addition to this calibration procedure, you may need to refer to
the following documents:
•NI-SCOPE Instrument Driver Quick Reference Guide
•Where to Start with Your NI 5102 Oscilloscope
•NI 5102 User Manual
You can download these documents fromthe NationalInstruments Web site
at
ni.com/manuals
Calibrating the NI 5102 with Calibration Executive4ni.com
Software
Complete the following steps to install Calibration Executive:
1.Make sure that your computer and monitor are powered on and
that you have installed Windows 2000/NT/Me/9x.
2.Close all open applications.
3.Insert the installation CD into the CD-ROM drive.
4.Choose the Run option from the Start menu on the desktop task bar.
5.In the command line box, type
of the CD-ROM drive you are using), and click OK.
6.Follow the instructions that appear in the dialog boxes.
The setup program installs Calibration Executive as well as
the associated files listed in Table 3.
Table 3. Calibration Executive Directories and Supporting Files
Directory NameContents
x:\setup.exe
(wherexis the letter
CalExec\Procedures\NI 5102\Limits
CalExec\Procedures\NI 5102\Support Files
CalExec\Procedures\NI 5102\Support Files
Test Conditions
Follow these guidelines to optimize the connections and the environment
during calibration:
•Keep connections to the NI 5102 short. Long cables and wires act as
antennae, picking up extra noise that can affect measurements.
•Use a 50 Ω BNC coaxial cable for all connections to the device.
•Keep relative humidity between 10 and 90%, noncondensing,
or consult your device hardware manual for the optimum relative
humidity.
•Maintain the temperature at 25 °C.
•Allow a warm-up time of at least 15 minutes to ensure that the
measurement circuitry of the 5102 is at a stable operating temperature.
Microsoft Access database that
stores the calibration limits.
Directory structure that contains the
calibration procedures.
This section explains how to set up and run the Calibration Executive
procedure.
Using the Wavetek 9500 in automated mode, the calibration procedure
takes approximately ten minutes. In manual mode, the procedure takes
approximately 30 minutes.
Using the alternative instruments in automated mode, the calibration
procedure takes approximately 20 minutes. In manual mode, the procedure
takes approximately 40 minutes.
Connecting Your Calibrator, DMM, and NI 5102
The calibration procedure steps you through the connections between the
calibrator, function generator, DMM, and NI 5102.
Note
You do not need a function generator or a DMM if you are using the Wavetek 9500.
Running the Calibration Executive Procedure
To run the Calibration Executive calibration procedure, complete the
following steps:
1.Launch Calibration Executive and follow the instructions in the in the
Calibration Configuration Wizard to load the NI 5102 calibration
procedure. Refer to Chapter 1, Introduction to Calibration Executive,
in the Calibration Executive Software User Manual if you need more
information on configuring and loading a calibration procedure.
2.When prompted by Calibration Executive, enter information about the
installed hardware, such as the calibrator, function generator, and
DMM.
Note
If you are using the Wavetek 9500, you will not need a function generator or a
DMM, so you can leave
Information, Selected DMM Information,andAvailable Addresses. You must also
enter N/A in the appropriate fields for the tracking numbers and calibration due dates
before you can proceed. The calibration report does not include any of this extra
information if you are using the Wavetek 9500.
3.When the procedure is loaded, click Run Procedure to begin. The
procedure prompts you to enter the device number, which is the
number assigned by Measurement & Automation Explorer (MAX).
4.Follow any instructions you receive from Calibration Executive.
Calibrating the NI 5102 with Calibration Executive6ni.com
selected in the list boxes for Selected Generator
none
Viewing the Calibration Report
When the procedure is finished, complete the following steps to view the
calibration report:
1.Click View»Reports.
2.Select View to launch your browser and view your report. The
calibration report appears as an printable HTML file.
You have completed calibrating your NI 5102 with Calibration Executive.
Note
If your NI 5102 device fails after calibration, return the device to National
National Instruments Web support is your first stop for help in solving
installation, configuration, and application problems and questions. Online
problem-solving and diagnostic resources include frequently asked
questions, knowledge bases, product-specific troubleshooting wizards,
manuals, drivers, software updates, and more. Web support is available
through the Technical Support section of
Worldwide Support
National Instruments has offices located around the world to help address
your support needs. You can access our branch office Web sites from the
Worldwide Offices section of
up-to-date contact information, support phone numbers, e-mail addresses,
and current events.
If you have searched the technical support resources on our Web site
and still cannot find the answers you need, contact your local office or
National Instruments corporate. For telephone support in the United States,
dial 512 795 8248. For telephone support outside the United States, contact
your local branch office:
ni.com
ni.com
. Branch office Web sites provide
Australia 03 9879 5166, Austria 0662 45 79 90 0, Belgium 02 757 00 20,
Brazil 011 284 5011, Canada (Calgary) 403 274 9391,
Canada (Ottawa) 613 233 5949, Canada (Québec) 514 694 8521,
Canada (Toronto) 905 785 0085, China (Shanghai) 021 6555 7838,
China (ShenZhen) 0755 3904939, Denmark 45 76 26 00, Finland 09 725 725 11,
France 01 48 14 24 24, Germany 089 741 31 30, Greece 30 1 42 96 427,
Hong Kong 2645 3186, India 91805275406, Israel 03 6120092, Italy 02 413091,
Japan 03 5472 2970, Korea 02 596 7456, Malaysia 603 9596711,
Mexico 5 280 7625, Netherlands 0348 433466, New Zealand 09 914 0488,
Norway 32 27 73 00, Poland 0 22 528 94 06, Portugal 351 1 726 9011,
Singapore 2265886, Spain 91 640 0085, Sweden 08 587 895 00,
Switzerland 056 200 51 51, Taiwan 02 2528 7227,
United Kingdom 01635 523545
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.