National Instruments 373873C-01 User Manual

LabVIEWTM SignalExpress

Getting Started with LabVIEW SignalExpress

Getting Started with LabVIEW SignalExpress
TM
August 2007 373873C-01

Worldwide Technical Support and Product Information

ni.com

National Instruments Corporate Headquarters

11500 North Mopac Expressway Austin, Texas 78759-3504 USA Tel: 512 683 0100

Worldwide Offices

Australia 1800 300 800, Austria 43 662 457990-0, Belgium 32 (0) 2 757 0020, Brazil 55 11 3262 3599, Canada 800 433 3488, China 86 21 5050 9800, Czech Republic 420 224 235 774, Denmark 45 45 76 26 00, Finland 385 (0) 9 725 72511, France 01 57 66 24 24, Germany 49 89 7413130, India 91 80 41190000, Israel 972 3 6393737, Italy 39 02 413091, Japan 81 3 5472 2970, Korea 82 02 3451 3400, Lebanon 961 (0) 1 33 28 28, Malaysia 1800 887710, Mexico 01 800 010 0793, Netherlands 31 (0) 348 433 466, New Zealand 0800 553 322, Norway 47 (0) 66 90 76 60, Poland 48 22 3390150, Portugal 351 210 311 210, Russia 7 495 783 6851, Singapore 1800 226 5886, Slovenia 386 3 425 42 00, South Africa 27 0 11 805 8197, Spain 34 91 640 0085, Sweden 46 (0) 8 587 895 00, Switzerland 41 56 2005151, Taiwan 886 02 2377 2222, Thailand 662 278 6777, Turkey 90 212 279 3031, United Kingdom 44 (0) 1635 523545
For further support information, refer to the Technical Support and Professional Services appendix. To comment on National Instruments documentation, refer to the National Instruments Web site at ni.com/info and enter the info code feedback.
© 2004–2007 National Instruments Corporation. All rights reserved.

Important Information

Warranty

The media on which you receive National Instruments software are warranted not to fail to execute programming instructions, due to defects in materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace software media that do not execute programming instructions if National Instruments receives notice of such defects during the warranty period. National Instruments does not warrant that the operation of the software shall be uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of the package before any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of returning to the owner parts which are covered by warranty.
National Instruments believes that the information in this document is accurate. The document has been carefully reviewed for technical accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to make changes to subsequent editions of this document without prior notice to holders of this edition. The reader should consult National Instruments if errors are suspected. In no event shall National Instruments be liable for any damages arising out of or related to this document or the information contained in it.
XCEPT AS SPECIFIED HEREIN, NATIONAL INSTRUMENTS MAKES NO WARRANTIES, EXPRESS OR IMPLIED, AND SPECIFICALLY DISCLAIMS ANY WAR RANTY OF
E
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE . CUSTOMERS RIGHT TO RECOVER DAMAGES CAUSED BY FAULT OR NEGLIGENCE ON THE PART OF
N
ATIONAL INSTRUMENTS SHALL BE LIMITED TO THE AMOUNT THERETOFORE PAID BY THE CUSTOMER. NATIONAL INSTRUMENTS WILL NOT BE LIABLE FOR DAMAGES RESULTING FROM LOSS OF DATA, PROFITS, USE OF PRODUCTS, OR INCIDENTAL OR CONSEQUENTIAL DAMAGES, EVEN IF ADVISED OF THE POSS IBILITY THEREOF. This limitation of the liability of National Instruments will apply regardless of the form of action, whether in contract or tort, including
negligence. Any action against National Instruments must be brought within one year after the cause of action accrues. National Instruments shall not be liable for any delay in performance due to causes beyond its reasonable control. The warranty provided herein does not cover damages, defects, malfunctions, or service failures caused by owner’s failure to follow the National Instruments installation, operation, or maintenance instructions; owner’s modification of the product; owner’s abuse, misuse, or negligent acts; and power failure or surges, fire, flood, accident, actions of third parties, or other events outside reasonable control.

Copyright

Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical, including photocopying, recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written consent of National Instruments Corporation.
National Instruments respects the intellectual property of others, and we ask our users to do the same. NI software is protected by copyright and other intellectual property laws. Where NI software may be used to reproduce software or other materials belonging to others, you may use NI software only to reproduce materials that you may reproduce in accordance with the terms of any applicable license or other legal restriction.

Trademarks

National Instruments, NI, ni.com, and LabVIEW are trademarks of National Instruments Corporation. Refer to the Terms of Use section on ni.com/legal for more information about National Instruments trademarks.
Other product and company names mentioned herein are trademarks or trade names of their respective companies. Members of the National Instruments Alliance Partner Program are business entities independent from National Instruments and have no
agency, partnership, or joint-venture relationship with National Instruments.

Patents

For patents covering National Instruments products, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your CD, or
ni.com/patents.

WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS

(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING FOR A LEVEL OF RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL IMPLANTS OR AS CRITICAL COMPONENTS IN ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE TO PERFORM CAN REASONABLY BE EXPECTED TO CAUSE SIGNIFICANT INJURY TO A HUMAN.
(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE SOFTWARE PRODUCTS CAN BE IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT LIMITED TO FLUCTUATIONS IN ELECTRICAL POWER SUPPLY, COMPUTER HARDWARE MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE FITNESS, FITNESS OF COMPILERS AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION, INSTALLATION ERRORS, SOFTWARE AND HARDWARE COMPATIBILITY PROBLEMS, MALFUNCTIONS OR FAILURES OF ELECTRONIC MONITORING OR CONTROL DEVICES, TRANSIENT FAILURES OF ELECTRONIC SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR MISUSES, OR ERRORS ON THE PART OF THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS THESE ARE HEREAFTER COLLECTIVELY TERMED “SYSTEM FAILURES”). ANY APPLICATION WHERE A SYSTEM FAILURE WOULD CREATE A RISK OF HARM TO PROPERTY OR PERSONS (INCLUDING THE RISK OF BODILY INJURY AND DEATH) SHOULD NOT BE RELIANT SOLELY UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM FAILURE. TO AVOID DAMAGE, INJURY, OR DEATH, THE USER OR APPLICATION DESIGNER MUST TAKE REASONABLY PRUDENT STEPS TO PROTECT AGAINST SYSTEM FAILURES, INCLUDING BUT NOT LIMITED TO BACK-UP OR SHUT DOWN MECHANISMS. BECAUSE EACH END-USER SYSTEM IS CUSTOMIZED AND DIFFERS FROM NATIONAL INSTRUMENTS' TESTING PLATFORMS AND BECAUSE A USER OR APPLICATION DESIGNER MAY USE NATIONAL INSTRUMENTS PRODUCTS IN COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT EVALUATED OR CONTEMPLATED BY NATIONAL INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS ULTIMATELY RESPONSIBLE FOR VERIFYING AND VALIDATING
THE SUITABILITY OF NATIONAL INSTRUMENTS PRODUCTS WHENEVER NATIONAL INSTRUMENTS PRODUCTS ARE INCORPORATED IN A SYSTEM OR APPLICATION, INCLUDING, WITHOUT LIMITATION, THE APPROPRIATE DESIGN, PROCESS AND SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.

Contents

About This Manual
Conventions ...................................................................................................................vii
Related Documentation..................................................................................................vii
Chapter 1 Introduction to LabVIEW SignalExpress
Chapter 2 Installing LabVIEW SignalExpress
LabVIEW SignalExpress LE.........................................................................................2-1
Minimum System Requirements ...................................................................................2-1
Installing LabVIEW SignalExpress...............................................................................2-2
Chapter 3 Working with Projects
Opening a Project...........................................................................................................3-1
Running a Project and Displaying Signals ....................................................................3-2
Configuring a Step.........................................................................................................3-4
Moving and Deleting Steps............................................................................................3-8
Handling Errors and Warnings ......................................................................................3-8
Chapter 4 Working with Signals
Graphing Signals............................................................................................................4-1
Importing a Signal from a File.......................................................................................4-3
Aligning and Comparing Signals...................................................................................4-4
Signal Types in LabVIEW SignalExpress.....................................................................4-6
Exporting and Printing Signals ......................................................................................4-6
Saving Signals to File......................................................................................4-7
Exporting Signals to Microsoft Excel .............................................................4-7
Printing Signals ...............................................................................................4-7
Creating Reports in LabVIEW SignalExpress ................................................4-7
© National Instruments Corporation v Getting Started with LabVIEW SignalExpress
Contents
Chapter 5 Logging Data
Recording a Signal ........................................................................................................ 5-1
Viewing a Logged Signal .............................................................................................. 5-2
Logging Signals with Predefined Start and Stop Conditions ........................................ 5-3
Analyzing Logged Signals ............................................................................................5-5
Advanced Playback .......................................................................................................5-6
Chapter 6 Performing Sweep Measurements
Defining Sweep Ranges and Outputs............................................................................6-1
Plotting Sweep Results.................................................................................................. 6-4
Running Multidimensional Sweeps...............................................................................6-5
Chapter 7 Extending LabVIEW SignalExpress Projects with LabVIEW
Importing LabVIEW VIs into LabVIEW SignalExpress as Steps................................7-1
Converting LabVIEW SignalExpress Projects to LabVIEW Block Diagrams............. 7-4
Chapter 8 Where to Go from Here
LabVIEW SignalExpress Sample Projects.................................................................... 8-1
Using Hardware with LabVIEW SignalExpress...........................................................8-1
Web Resources ..............................................................................................................8-1
Appendix A Technical Support and Professional Services
Getting Started with LabVIEW SignalExpress vi ni.com

About This Manual

Use this manual to familiarize yourself with LabVIEW SignalExpress interactive measurements and the basic LabVIEW SignalExpress features that you use to acquire and analyze signals.
This manual contains exercises that help you begin working with LabVIEW SignalExpress. These exercises teach you how to run projects, configure steps, work with signals, configure sweep measurements, log data, and extend LabVIEW SignalExpress with LabVIEW graphical programming.

Conventions

The following conventions appear in this manual:
» The » symbol leads you through nested menu items and dialog box options
to a final action. The sequence File»Page Setup»Options directs you to pull down the File menu, select the Page Setup item, and select Options from the last dialog box.
This icon denotes a note, which alerts you to important information.
bold Bold text denotes items that you must select or click in the software, such
as menu items and dialog box options. Bold text also denotes input and output names, parameter names, dialog boxes, sections of dialog boxes, and menu names.
italic Italic text denotes variables, emphasis, a cross-reference, or an introduction
to a key concept. Italic text also denotes text that is a placeholder for a word or value that you must supply.
monospace Text in this font denotes text or characters that you should enter from the
keyboard. This font is also used for the proper names of disk drives, paths, directories, programs, subprograms, subroutines, device names, functions, operations, variables, filenames, and extensions.

Related Documentation

Refer to the LabVIEW SignalExpress Help, available by selecting Help» LabVIEW SignalExpress Help, for more information as you read this
manual.
© National Instruments Corporation vii Getting Started with LabVIEW SignalExpress
Introduction to LabVIEW SignalExpress
National Instruments provides innovative solutions for scientists and engineers to build automated measurement systems based on industry-standard computers and platforms. National Instruments develops robust, industry-leading programming environments for automating measurement systems, such as LabVIEW for graphical development, LabWindows for Microsoft Visual Studio programming. You can use these programming tools with National Instruments measurement hardware and interfaces to traditional instruments to build custom, advanced virtual instrumentation systems.
LabVIEW SignalExpress optimizes virtual instrumentation for design engineers by offering instant interactive measurements that require no programming. You can use LabVIEW SignalExpress interactively to acquire, generate, analyze, compare, import, and log signals. You can compare design data with measurement data in one step. LabVIEW SignalExpress extends the ease of use and performance of virtual instrumentation to those who must acquire or analyze signals without programming applications. You also can extend the functionality of LabVIEW SignalExpress by importing a custom virtual instrument (VI) created in the LabVIEW Development System or by converting a LabVIEW SignalExpress project to a LabVIEW block diagram so you can continue development in LabVIEW. Refer to Chapter 7, Extending
LabVIEW SignalExpress Projects with LabVIEW, for more information.
/CVI™ for ANSI C programming, and Measurement Studio
1
© National Instruments Corporation 1-1 Getting Started with LabVIEW SignalExpress
Installing LabVIEW SignalExpress
You can use the LabVIEW SignalExpress CD to install LabVIEW SignalExpress and complete the exercises in this manual.

LabVIEW SignalExpress LE

LabVIEW SignalExpress LE gives you a 30-day trial of the Full version of LabVIEW SignalExpress. After that period, you must activate your version of LabVIEW SignalExpress LE or purchase the Full version. If you activate your version of LabVIEW SignalExpress LE before your 30-day trial of the full version is complete, the full features are still available for the remainder of your 30-day trial. You can use LabVIEW SignalExpress LE to complete simple data acquisition and logging functions.
If you do not register your version of LabVIEW SignalExpress LE after 30 days, LabVIEW SignalExpress LE runs with the following limitations:
You can no longer save projects, so you cannot convert projects into LabVIEW block diagrams.
You can run projects for only 10 minutes per session.
2
Refer to the National Instruments Web site at purchase the full version of LabVIEW SignalExpress.
ni.com/signalexpress to

Minimum System Requirements

LabVIEW SignalExpress requires a minimum of 256 MB of RAM and a Pentium III or later or Celeron 866 MHz or equivalent processor, but National Instruments recommends 512 MB of RAM and a Pentium 4/M or equivalent processor.
© National Instruments Corporation 2-1 Getting Started with LabVIEW SignalExpress
Chapter 2 Installing LabVIEW SignalExpress

Installing LabVIEW SignalExpress

Complete the following steps to install LabVIEW SignalExpress on Windows Vista/XP/2000/NT.
Note If you want to convert LabVIEW SignalExpress projects to LabVIEW block
diagrams, you must install LabVIEW 7.1 or later before you install LabVIEW SignalExpress.
1. Disable any automatic virus detection programs before you install. Some virus detection programs interfere with the installation program.
2. Log on as an administrator or as a user with administrator privileges.
3. Insert the LabVIEW SignalExpress Development or Evaluation CD and follow the instructions that appear on the screen.
By default, LabVIEW SignalExpress installs in the
National Instruments\SignalExpress
4. After installation, check the hard disk for viruses and enable any virus detection programs you disabled.
The example projects you use with this manual are located in
SignalExpress\Examples\Tutorial. The solutions for each exercise
are located in
SignalExpress\Examples\Tutorial\Solutions.
Program Files\
directory.
If you use LabVIEW SignalExpress with National Instruments data acquisition or modular instrumentation hardware, you must install NI-DAQmx 8.5 or later, NI-SCOPE 2.7 or later, NI-FGEN 2.2 or later, or NI-DMM 2.4.2. LabVIEW SignalExpress includes a National Instruments Driver CD that includes these drivers.
Getting Started with LabVIEW SignalExpress 2-2 ni.com
Working with Projects
You can use LabVIEW SignalExpress to define measurement procedures by adding and configuring steps in an interactive measurement environment. A step is a configurable function that acquires, generates, analyzes, loads, or stores signals. Most steps process input signals and produce output signals. You can configure the operation of a step by specifying values in a configuration view. A saved sequence of configured steps is a LabVIEW SignalExpress project.
This chapter teaches you how to load and run existing projects and how to configure steps in these projects.

Opening a Project

Complete the following steps to load a sample project in LabVIEW SignalExpress.
1. Launch LabVIEW SignalExpress. Notice that LabVIEW SignalExpress is split into three main views: the
Project View on the left, the Data View in the middle, and the help on the right. Within the Data View, you should see the Data View tab, the Logging Options tab, and the Project Documentation tab.
2. If LabVIEW SignalExpress does not open in the factory-default configuration, select View»Data View to display the Data View.
3. Select File»Open Project, navigate to the
Examples\Tutorial Project.seproj LabVIEW SignalExpress project.
4. Examine the window that appears, as shown in Figure 3-1 with the help hidden, to learn about different components of LabVIEW SignalExpress.
3
SignalExpress\
directory, and double-click the First
© National Instruments Corporation 3-1 Getting Started with LabVIEW SignalExpress
Chapter 3 Working with Projects
1
2
3
4
5
6
5
7
1 Execution control buttons 2 Work Area pull-down menu 3Step
9
4 Input 5 Output 6Project View
8
7 Logged Data window 8 Data View 9 Record button
Figure 3-1. First Project.seproj
The left pane is the Project View, which presents the order of operations, or steps, for the project. The right pane is the Data View, which displays the signal that the project generates and analyzes.

Running a Project and Displaying Signals

LabVIEW SignalExpress has two execution modes—Run and Run Once. When you click the Run button, LabVIEW SignalExpress executes all steps in the project continuously until you click the Stop button. The Stop button appears in place of the Run button as a project runs. While the steps in the project execute, the Data View updates continuously. While the project runs, you can change the measurement configurations and view the
Getting Started with LabVIEW SignalExpress 3-2 ni.com
Chapter 3 Working with Projects
results immediately. If you modify the configuration of steps while a project runs, LabVIEW SignalExpress gives you direct, immediate feedback on the changes you make. When you click the Run Once button, LabVIEW SignalExpress executes all steps in the project one time.
Complete the following steps to run the sample project and display signals.
1. Click the Run button, shown at left, to execute all steps in the project continuously.
The project loads a signal from a text file and performs two operations on the signal—an amplitude and levels measurement and a distortion measurement. The Amplitude and Levels step and the Distortion step perform these measurements, respectively. When you run a project, steps analyze input signals and generate new output signals as a result of the analysis. In this project, the Load from ASCII step loads a distorted sine wave, the Amplitude and Levels step and the Distortion step analyze the sine wave, and both steps return new outputs. In the Project View, LabVIEW SignalExpress indicates inputs with red arrows and outputs with blue arrows.
The graph in the Data View still contains the loaded signal, which is a time-domain signal. Graphs display time-domain, frequency-domain, or x-y signals.
2. Click the exported spectrum output signal of the Distortion step in the Project View and drag it to the Data View to display the signal.
LabVIEW SignalExpress creates a new graph in the Data View. LabVIEW SignalExpress does not display the exported spectrum signal on the same graph as the time-domain signal because the exported spectrum signal is a frequency-domain signal. LabVIEW SignalExpress automatically recognizes different types of signals and renders them in the appropriate displays.
3. Select Help»LabVIEW SignalExpress Help, click the Search tab, and enter
"signal types" to refer to the LabVIEW SignalExpress
Help for more information about signal types. The help provides
information about using LabVIEW SignalExpress functionality such as projects, steps, and signals.
4. Click the dc output of the Amplitude and Levels step and drag it to the Data View to display the measurement.
LabVIEW SignalExpress creates a table to display the scalar measurement from the dc output.
© National Instruments Corporation 3-3 Getting Started with LabVIEW SignalExpress
Chapter 3 Working with Projects
5. Drag the rms output to the table to display the scalar RMS
measurement. LabVIEW SignalExpress creates a new row in the table to display the
second measurement. The project should appear as shown in Figure 3-2.
Figure 3-2. Outputs of First Project.seproj

Configuring a Step

A step is a configurable function that acquires, generates, analyzes, loads, or stores signals. Steps process input signals and produce output signals. You can configure the operation of a step in SignalExpress by specifying values in the Step Setup dialog box, or configuration view, for that step. While a project runs, you can modify the configuration of steps to view immediate feedback on the changes and adjust the measurements until you achieve the results you need.
Getting Started with LabVIEW SignalExpress 3-4 ni.com
Chapter 3 Working with Projects
Complete the following steps to configure the Distortion step and the Amplitude and Levels step.
1. Double-click the Distortion step in the Project View. The Step Setup dialog box for the Distortion step appears, as shown in Figure 3-3.
Figure 3-3. Distortion Step Setup Dialog Box
On the Configuration tab, the settings indicate that the Distortion step receives a time-domain waveform signal as an input, performs a power spectrum on the signal to convert it to the frequency domain, and computes the total harmonic distortion (THD) and fundamental frequency of the signal. The step generates three measurements as outputs—the spectrum, the THD, and the fundamental frequency of the original time-domain waveform input.
2. If the context help does not appear on the right side of the screen, click the Show Help button, shown at left, to display complete reference information about the step. The upper half of the context help displays information about the step, and the lower half of the context help displays information about an object when you move the cursor over the object.
© National Instruments Corporation 3-5 Getting Started with LabVIEW SignalExpress
Chapter 3 Working with Projects
3. On the Configuration tab, select Fundamental Tone from the Export signals (THD) pull-down menu.
The operation of the step changes from displaying the frequency-domain spectrum of the entire input signal to displaying only the frequency spectrum of the fundamental tone of the input signal. The graphed signals in the top half of the Step Setup dialog box update to reflect the change you made.
4. Select Harmonics Only from the Export signals (THD) pull-down menu.
The step changes to export only the spectrum of the harmonic signals from the input signal. Both the output signal of the Distortion step and the graph on the Data View update to reflect the change you made.
5. Click the Amplitude and Levels step in the Project View. The Step Setup dialog box changes from displaying the configuration
of the Distortion step to displaying the configuration of the Amplitude and Levels step.
Getting Started with LabVIEW SignalExpress 3-6 ni.com
Chapter 3 Working with Projects
6. Click the Input/Output tab to display the list of possible inputs and
outputs for this step, as shown in Figure 3-4.
Figure 3-4. Amplitude and Levels Step Setup Dialog Box
7. Place checkmarks in the Export +peak value, Export -peak value, and Export peak-peak value checkboxes to configure the Amplitude and Levels step to return three additional measurements.
Three additional outputs appear in the Project View.
8. Click the Data View tab to display the Data View.
9. Drag the three new outputs from the Project View to the table to display the scalar measurements.
10. Click the Stop button, shown at left, to stop the project. When you click the Stop button, the project stops running after
completing the current cycle of operations, or the current iteration. Click the down arrow on the Stop button and click the Abort button to completely stop the project without finishing the current iteration.
© National Instruments Corporation 3-7 Getting Started with LabVIEW SignalExpress
Chapter 3 Working with Projects
11. Select File»Save Project As and save the project as My First
Project.seproj in the SignalExpress\Examples\Tutorial
directory.
12. Select File»Close Project to close the project.

Moving and Deleting Steps

The steps in SignalExpress projects depend on input data, which means steps can operate only on signals exported from previous steps in the Project View. When you click the Input signal pull-down menu on the Input tab of the Step Setup dialog box of a step, the menu displays only compatible signals exported from a previous step. When the output of a step becomes the input of another step, the steps become dependent on each other, and the two steps execute sequentially at the same rate. The first step generates an output signal that the second step must receive as an input before the step can execute.
You can move a step within a project by dragging it up or down in the Project View. You can delete a step by right-clicking it in the Project View and selecting Delete from the shortcut menu. However, when you move or delete a step, the status of signals in the project changes. For example, if you delete a step that generates output signals, the operation of the project breaks if any of the deleted output signals are inputs for other steps, and an error indicator appears in the Project View. You also can cut, copy, and paste steps within a project by pressing the <Ctrl-X>, <Ctrl-C>, and <Ctrl-V> keys, respectively, or by right-clicking a step in the Project View and selecting Cut, Copy, Paste Before Selected Step, or Paste After Selected Step from the shortcut menu.

Handling Errors and Warnings

If an error occurs while a project runs, an error indicator, shown at left, appears in the Project View on the step that encountered the error. Double-click the step with the error to display an error description across the bottom of the Step Setup dialog box. Click the Details button to the right of the error description to display the full error description.
SignalExpress logs all errors and warnings in the Event Log while a project runs. To display the Event Log, select
Log
tab in the Data View area. Refer to the
more information about errors and warnings by selecting
SignalExpress Help
Getting Started with LabVIEW SignalExpress 3-8 ni.com
, clicking the
View»Event Log
LabVIEW SignalExpress Help
Search
tab, and entering
and click the
Help»LabVIEW
errors
Event
.
for
Working with Signals
You can use LabVIEW SignalExpress to generate and analyze signals to evaluate designs without programming. This chapter teaches you how to work with signals in LabVIEW SignalExpress, including how to plot signals on graphs, import signals from a file, interactively align and compare two signals, and save signals to a file.

Graphing Signals

Complete the following steps to plot signals in a sample project and examine the signals visually using cursors.
1. Select File»Open Project, navigate to the
Examples\Tutorial Signals.seproj. This project configures the Create Signal step
to create a square wave signal and the Filter step to perform a lowpass Butterworth filter.
2. Drag the step signal output of the Create Signal step to the Data View.
3. Drag the filtered step output of the Filter step to the Data View. You can plot signals of the same signal type on the same graph. Both
the step signal and the filtered step signals are time-domain signals, so they appear on the same graph. If you want to plot signals of different types, drag the signals to the same graph and LabVIEW SignalExpress creates a new graph.
4. Click the Add Display button, shown at left, on the toolbar to create a new graph.
5. Drag the filtered step output of the Filter step to the new graph.
4
SignalExpress\
directory, and double-click
© National Instruments Corporation 4-1 Getting Started with LabVIEW SignalExpress
Chapter 4 Working with Signals
6. Right-click the new graph and select Visible Items»Cursors from the
shortcut menu to display two interactive cursors, as shown in Figure 4-1.
Figure 4-1. Signals.seproj
As you drag the cursors, LabVIEW SignalExpress displays the x and y values of the cursors in the cursor table at the bottom of the Data View.
7. Select File»Save Project As and save the project as
My Signals.seproj in the SignalExpress\Examples\ Tutorial
Getting Started with LabVIEW SignalExpress 4-2 ni.com
directory.

Importing a Signal from a File

You can import signals from standard file formats such as ASCII comma­or tab-delimited files and LabVIEW measurement data files ( also can import signals from simulated results of EDA tools such as SPICE simulators.
Complete the following steps to import a signal from a file.
1. Right-click the Project View and select Load/Save Signals»Analog Signals»Load from ASCII from the shortcut menu.
2. Click the browse button, shown at left, in the Step Setup dialog box, navigate to the and double-click and displays the signals in the file.
In the File preview section, column 1 shows the time stamp data, and column 2 shows the actual voltage values of the signal.
3. Click the Import Signals tab to display the available signals in the file.
4. Place a checkmark in the Column 2 checkbox to import that signal, and remove the checkmark from the Column 1 checkbox.
The Step Setup dialog box displays a preview of the signal in the Imported Signal section.
5. Select Column 1 from the Input X values pull-down menu to set the x-axis data of the waveform to the appropriate values.
6. Open the Data View.
7. Right-click the Column 2 output in the Project View and select Rename from the shortcut menu.
8. Enter
9. Drag the step response output of the Load from ASCII step to the
step response and press the <Enter> key to rename the output.
lower graph in the Data View. The filtered step signal resembles the rising edge of the step response
output, as shown in Figure 4-2.
SignalExpress\Examples\Tutorial directory,
Step Response.txt. This step parses an ASCII file
Chapter 4 Working with Signals
.lvm). You
© National Instruments Corporation 4-3 Getting Started with LabVIEW SignalExpress
Chapter 4 Working with Signals
Figure 4-2. Signals of Signals.seproj
10. Select File»Save Project to save the project.

Aligning and Comparing Signals

Although the filtered step signal and the step response output both show an overshoot in the rising edge, assessing the similarity between the two is difficult because the signals come from different sources and vary in amplitude and timing. However, you can use the Interactive Alignment step to align and compare two signals, so you can choose which type of information you want to export from the operation to use in the project.
Complete the following steps to align two signals in the
My Signals.seproj project.
1. Right-click the step response output and select Send To» Processing»Analog Signals»Interactive Alignment from the
shortcut menu to pass the step response signal from the Load from ASCII step to the Interactive Alignment step.
Getting Started with LabVIEW SignalExpress 4-4 ni.com
Chapter 4 Working with Signals
The step selects the two most recent signals from the project to use as inputs and displays the signals on the graph of the Step Setup dialog box, as shown in Figure 4-3.
Figure 4-3. Interactive Alignment Step Setup Dialog Box
When you add a step to a project, LabVIEW SignalExpress selects input signals based on the signal types the step can accept. For example, the Interactive Alignment step can operate only on time-domain waveform signals. Therefore, the step selects as inputs the last two time-domain signals created in the project. To change the input signals for a step, select different signals from the pull-down menu of compatible signals in the Input/Output tab of the Step Setup dialog box.
2. Click the red signal in the Input Signals graph and drag it to another
point within the graph. You can drag, expand, and contract signals on the graph.
3. Try to align the rising edges of the two signals by dragging a signal within the graph. Click a signal to set an anchor point and press and hold the <Alt> key to stretch the signal around that anchor point in the x and y directions.
© National Instruments Corporation 4-5 Getting Started with LabVIEW SignalExpress
Chapter 4 Working with Signals
On the Alignment tab of the Step Setup dialog box, the step computes and displays the x and y gain and offset values you need to achieve alignment specifications as you drag the signals.
4. Select Auto - Step from the Mode pull-down menu to align the
signals. LabVIEW SignalExpress bases this alignment mode on built-in algorithms.
The lower graph in the Step Setup dialog box displays the difference between the two signals.
5. On the Input/Output tab, place a checkmark in the Export aligned signals checkbox to add the signals to the outputs of the step.
6. Select the Data View tab to open the Data View.
7. Click the Add Display button, shown at left, to add a third graph.
8. Drag the aligned reference and aligned test outputs of the Interactive Alignment step to the new graph to view the aligned signals.
9. Select File»Save Project to save the project.

Signal Types in LabVIEW SignalExpress

Some steps, such as the Arithmetic step, can operate on multiple signal types. For example, you can use the Arithmetic step to operate on time-domain or frequency-domain waveforms. The Arithmetic step changes behavior based on the type of input signals you select for the step. For example, if you add two time-domain signals, LabVIEW SignalExpress adds only their amplitudes. However, if you add two frequency-domain phase signals, LabVIEW SignalExpress adds the appropriate phase shift.
Refer to the LabVIEW SignalExpress Help for more information by selecting Help»LabVIEW SignalExpress Help, clicking the Search tab, and entering
"signal types".

Exporting and Printing Signals

You can use LabVIEW SignalExpress to document signals or continue analysis in another software application. This section teaches you how to export signals, including sending signals to an ASCII file, sending data to Microsoft Excel, printing signals, and using the built-in documentation feature to document the LabVIEW SignalExpress project.
Getting Started with LabVIEW SignalExpress 4-6 ni.com
Saving Signals to File
Complete the following steps to save a signal to a file.
1. Click the Add Step button and select Load/Save Signals»Analog
2. Click the Signals tab in the Step Setup dialog box and select filtered
3. On the File Settings tab, save the file to
4. Select Overwrite from the If file already exists pull-down menu.
5. Select Generic ASCII from the Export file type pull-down menu.
6. Click the Close button to close the Step Setup dialog box.
7. Click the Run button to run the project and save the resulting signal to
8. Select File»Save Project to save the project.
9. Select File»Close Project to close the project.
Chapter 4 Working with Signals
Signals»Save to ASCII/LVM.
step from the Input Data pull-down menu.
SignalExpress\
Examples\Tutorial\filtered signal.txt
path control.
You can use a Load/Save Signals step to save data to a file every time the project runs.
the specified ASCII file.
in the Export file
Exporting Signals to Microsoft Excel
To export signal data to Microsoft Excel, launch Excel and drag the output signal of a step in LabVIEW SignalExpress to an Excel spreadsheet.
Printing Signals
To print an image of a graph, open the Data View and select File»Print» Print Data View.
Creating Reports in LabVIEW SignalExpress
Select
View»Project Documentation
View. You can describe your project using text and drag and drop step outputs into your documentation. When you drag and drop a step output into the Project Documentation View, the value in the Project Documentation View automatically updates to match the current value of the step output.
To print your documentation, open the Project Documentation View and select File»Print»Print Documentation. To export your documentation to HTML, open the Project Documentation View and select File»Export» Export Documentation to HTML.
© National Instruments Corporation 4-7 Getting Started with LabVIEW SignalExpress
to display the Project Documentation
Logging Data
You can use LabVIEW SignalExpress to record and analyze measurements. You can record any time-domain, double, U32, or Boolean step output. You also can analyze and process logged data by playing it through analysis steps.
This chapter teaches you how to record data using the integrated data logging features in LabVIEW SignalExpress. You learn how to record a specified signal, play back that signal, and analyze the signal using analysis steps. You also learn how to use the Recording Options View to log signals based on specified start or stop conditions.

Recording a Signal

You can use the Record button to configure a data logging process.
Complete the following steps to specify a signal to record and to record the signal.
1. Select File»Open Project, navigate to the
Examples\Tutorial Logging.seproj.
This project uses the Create Signal step to generate a signal based on a formula.
2. Click the Record button, shown at left, to open the Logging Signals Selection dialog box.
The Logging Signals Selection dialog box displays the signals in the project available for recording. You can select one signal or multiple signals to record. You also can specify a name and description for the log.
3. Place a checkmark in the signal checkbox to record the formula signal generated in the Create Signal step.
4. Click the OK button to close the Logging Signals Selection dialog box and begin recording the signal. The logging operation continues until you click the Stop button.
5
SignalExpress\
directory, and double-click
© National Instruments Corporation 5-1 Getting Started with LabVIEW SignalExpress
Chapter 5 Logging Data
5. Click the Stop button, shown at left, to stop logging the signal. The logged data appears in the Logged Data window, as shown in
Figure 5-1.
Figure 5-1. Logged Data Window
By default, LabVIEW SignalExpress names the logged data according to the date and time you recorded the data. LabVIEW SignalExpress saves logged data in the in the Options dialog box.
6. Select Tools»Options and select the Logging option to specify the directory for LabVIEW SignalExpress to save the logged data and to customize various preferences for logged data.
7. Click the OK button to close the Options dialog box.
8. Select File»Save Project to save the project.
.tdms file format in the directory you specify

Viewing a Logged Signal

Complete the following steps to view the logged data.
1. If the Data View is not visible, select View»Data View to display the Data View.
2. The Logged Data window displays a list of all logged data in the current project. Select the data log you just recorded from the Logged Data window and drag it into the Data View. The Data View displays the logged data and a preview graph, as shown in Figure 5-2. The signal displayed in the Data View may differ from the signal displayed in Figure 5-2, depending on how long you record the signal.
Getting Started with LabVIEW SignalExpress 5-2 ni.com
Figure 5-2. Logging.seproj
Chapter 5 Logging Data
The preview graph provides a method for zooming and panning through data in the Data View. The preview graph appears by default when viewing logged data. When viewing live or non-logged data, right-click the Data View and select Visible Items»Preview to display the preview graph.
3. Click the Zoom In button next to the preview graph to zoom in on the
logged signal. The cursors on the preview graph show the subset of data currently displayed on the preview graph. Use the scroll bar beneath the preview graph to scroll through the data. Click and drag the cursors on the preview graph to increase or decrease the subset of data you are viewing.

Logging Signals with Predefined Start and Stop Conditions

You can configure start and stop conditions that the signals must meet before LabVIEW SignalExpress records or stops recording the signals. Complete the following steps to log data based on start and stop conditions.
1. If the Recording Options tab is not visible, select View»Recording Options to open the Recording Options View.
© National Instruments Corporation 5-3 Getting Started with LabVIEW SignalExpress
Chapter 5 Logging Data
2. Select Signal Selection in the Category list in the Recording Options
View.
3. Place a checkmark next to the signal in the Record column, as shown
in Figure 5-3.
Figure 5-3. Signal Selection
The Record button changes to the Record While Running button, shown at left. Ensure the Record While Running button is pressed. When the Record While Running button is pressed, LabVIEW SignalExpress records the selected signal when you click the Run or Run Once button.
4. Select Start Conditions in the Category list in the Recording Options
View.
5. Click the Add button in the Logging start conditions page to
customize a start condition for your logging task. a. Select the Signal option in the Condition source control to
specify for LabVIEW SignalExpress to begin recording when the input signal meets the specified condition.
b. Select signal in the Signal control.
Getting Started with LabVIEW SignalExpress 5-4 ni.com
Chapter 5 Logging Data
c. Select Rising slope in the Condition control to begin recording
the signal based on the value of the edge of the signal on a positive slope.
d. Enter
6. Select Stop Conditions in the Category list in the Recording Options View.
7. Click the Add button in the Logging stop conditions page to customize a stop condition for your logging task.
a. Select the Duration option in the Condition source control. b. Enter
c. Click the Run button, shown at left. LabVIEW SignalExpress
1 in the Value control to begin recording when the signal
crosses 1 on a rising slope.
5 in the Duration control to record the signal for 5 seconds
after the signal meets the start condition.
begins recording the signal when the signal crosses level 1 on a rising slope and continues recording the signal for 5 seconds.
The Recording indicator on the bottom of the Recording Options View will turn on when the signal meets the start condition and logging is in progress. The Disk information indicator displays the available hard disk space on the computer for the log.

Analyzing Logged Signals

After you log a signal, you can play back the logged data or run the logged signal through analysis steps, just as you can with live data. Complete the following steps to analyze a logged signal.
1. Navigate to the Work Area pull-down menu on the top left of the LabVIEW SignalExpress window, as shown in Figure 5-4. Click the down arrow and select Playback to switch to the Playback work area.
Figure 5-4. Work Area pull-down menu
Use work areas to perform multiple LabVIEW SignalExpress operations from within the same project. You can acquire data, process signals, record data, and perform measurements on logged data without opening a new project. When you save your project, LabVIEW SignalExpress saves every work area within the project in the same project file.
© National Instruments Corporation 5-5 Getting Started with LabVIEW SignalExpress
Loading...
+ 156 hidden pages