Nassetti Pendant Arm Installation manual

B
PPeennddaanntt AArrmm
B
RII
R
D
D
G
G
E
PPllaannnniinngg aanndd IInnssttaallllaattiioonn IInnssttrruuccttiioonnss
B
B
RII
R
D
D
PPeennddaanntt AArrmm
G
G
E
E
B
Contents Page
1. Notes On These Instructions 3
1.1. Explanation Of Symbols 3
2. Safety Instructions 3
2.1. Positioning Requirements 4
2.2. Intended Use 4
2.3. Instructions For Installation and Operation 4
3. Ceiling Fixations 4
3.1. Static Fixing Requirements 4
3.2. Definitions 5
3.2.1. Interface Plate 5
3.2.2. Ceiling Suspension System For Intermediate Ceiling 5
3.2.3. Counter Plate Set 5
3.3. Ceiling Height Adjustments 6
3.3.1. Solution 1 6
3.3.2. Solution 2 6
3.3.3. Solution 3 7
3.4. Minimum Ceiling Heights 8
3.5. Maximum Ceiling Heights With Basic Components 8
3.6. BRIDGE Ceiling Fixation System Components 8
3.6.1. Interface Plates 8
3.6.1.1. Interface Plates For Single Arm Type Pendant Arms 8
3.6.1.2. Interface Plates For Double Arm Type Pendant Arms 9
3.6.2. Suspension Systems For Intermediate Ceiling Structures 10
3.6.3. Counter-Plate Sets 11
3.6.4. Intermediate Ceiling Opening Covers
3.6.4.1. Intermediate Ceiling Opening Covers For Single Arm 11 Type Pendant Arms 11
3.6.4.2. Intermediate Ceiling Opening Covers For Double 12 Arm Type Pendant Arms
3.7. Options For Ceiling Fixations 12
3.7.1. Fixation On The Ceiling With Counter-Plate Sets 13
3.7.2. Fixation On The Ceiling With Heavy-Load Anchors 14
3.7.3. Fixation On The Ceiling Substructure With Bolts 15
3.8. Preparation For Installation Of The Ceiling Suspension Systems 16
3.9. Installation Of Interface Plate Onto The Ceiling Suspension System For Intermediate Ceiling 17
3.10. Preparation For Installation Of The Carrying Arms 17
4. Carrying Arms 18
4.1. Installation Of The Upper Joint Of The Single Arm Type Pendant Arms 18
4.2. Installation Of The Upper Joints Of The Double Arm Type Pendant Arms 19
4.3. Installation Of The Carrying Arms Onto The Upper Joints 21
4.4. Installation Of The Secondary Carrying Arms Onto The Middle Joints 21
4.5. Installation Of The Columns Onto The Lower Joints 22
B
RII
R
D
D
PPeennddaanntt AArrm
G
G
E
m
PLANLAMA VE MONTAJ TALİMATI BRIDGE-EN-R1 Page 1 / 39
B
Content Page
5. Consoles 23
5.1. Installation Of The Consoles Onto The Columns 23
6. Shelves 23
6.1. Installation Of The Shelves Onto The Consoles 23
7. I.V. / Infusion Pump Holders 24
7.1. Installation Of The I.V. / Infusion Pump Holders Onto The Consoles 24
8. Covers 25
8.1. Installation Of The Opening Covers Onto The Carrying Arms 25
8.2. Fixation Of The Covers Onto The Lower Joints 26
8.3. Fixation Of The Covers Onto The Middle Joints 26
8.4. Fixation Of The Covers Onto The Upper Joints 27
8.5. Intermediate Ceiling Opening Covers 27
8.5.1. Installation Of The Single Arm Type Ceiling Opening Covers 27
8.5.2. Installation Of The Double Arm Type Ceiling Opening Covers 28
9. Setting Operations 29
9.1. Settings The Mechanical Breaks 29
9.2. Settings The Rotation Stops 30
10. Electrical Installations 31
10.1. Specifications For The Mains Electrical Installation /
Equi-Potential Pin Laying With Strain Relief 31
10.2. Laying The Electrical Connections 32
10.2.1. Laying The Electrical Connections For Single Arm Type Pendant Arms 32
10.2.2. Laying The Electrical Connections For Double Arm Type Pendant Arms 32
10.3. Earthing Of The Apparatus 33
10.3.1. Earthing The Consoles 33
10.3.2. Earthing The Carrying Arms 33
10.3.3. Earthing The Upper Joints 34
10.4. Power Supply 34
10.5. Electrical Wiring Diagram Per Console 35
11. Gas Installations 36
12. Typical Gas Supply and Anaesthetic Gas Evacuation Ducts Diagram 37
13. Appendices 38
B
RII
R
D
D
PPeennddaanntt AArrm
G
G
E
m
PLANLAMA VE MONTAJ TALİMATI BRIDGE-EN-R1 Page 2 / 39
B
1. Notes On These Instructions
The preparatory measures described here ensure the safe planning of the main requirements for the installation and operation of BRIDGE pendant arms.
Our rights for making any technical modifications in order to improve the product are reserved.
Please read these instructions very carefully and observe the safety notes in these instructions and specifications
for installation.
The contents of these instructions can be altered at any time without warning.
Where there are translations into foreign languages, the English version of these instructions is binding.
Forms:
In Section 13 of these instructions, there are two forms, which must be used to document the following important data:
Form 1 Minimum specifications for fixing with heavy-load anchors
Form 2 Declaration of Acceptance
Trademarks:
BRIDGE is a registered trademark of Nassetti.
All other trademarks mentioned in these instructions are the sole property of the manufacturers concerned.
1.1. Explanation of Symbols
Symbols are used to alert you on possible operational or installation faults or give you useful tips on how to use the equipment.
B
RII
R
D
D
PPeennddaanntt AArrm
G
G
E
m
IMPORTANT!
The safety symbol alerts you on critical safety information and information necessary to obtain approval.
NOTE!
The note symbol gives you tips and useful information.
2. Safety Instructions
2.1. Positioning Requirements
The options and the conditions for fixing BRIDGE pendant arms are described in the following sections. The pendant arms must be fixed in consultation with the construction manager during installation procedures.
The methods described for the ceiling fixation, gas and power supply are general recommendations and their implementation is to be planned and designed for each individual case by qualified experts.
BRIDGE pendant arms are not suitable for use in potentially explosive areas.
An ambient temperature of 10°C to 40°C must be maintained during operation.
The relative humidity must not exceed 75%.
PLANLAMA VE MONTAJ TALİMATI BRIDGE-EN-R1 Page 3 / 39
B
2.2. Intended Use
BRIDGE pendant arms are intended to carry and position medical devices in operating theaters and intensive care units and to supply these devices with electricity and medical gases. In operating theaters, they also provide the option of evacuation of anaesthetic gases, compressed gases and provision of compressed driving gases.
BRIDGE pendant arms are suitable for continuous duty.
BRIDGE pendant arms can be customized with additional fittings
2.3. Instructions for Installation and Operation
Installation, testing and operation of pendant arms must be carried out in accordance with EN 60601-1, EN 737-2, and EN 737-3 and in accordance with the manufacturer’s instructions.
3. Ceiling Fixations
3.1. Static Fixing Requirements
The static certificate must be prepared prior to the installation of the ceiling fixation. Nassetti provides
a corresponding form with every ceiling fixation system supplied.
The static’s of the ceiling must be designed so that the loads of the pendant arms and the installed lighting systems or fittings, as well as any other loads present are safely carried and distributed.
The applicable regional building specifications must be observed.
Static calculations for the ceiling fixation and the load distribution in the ceiling must be provided.
The load-bearing capacity of the structure must be de signed, tested and confirmed by a structural engineer.
The interface plate or ceiling anchorage plate must lie flush with the raw ceiling.
If an existing interface plate or ceiling anchorage plate is used, care must be taken to ensure that:
The plate and the bolts are rust-free and in perfect condition.
The holes and bolts in the interface plate or ceiling anchorage plate fit the corresponding part of the upper joint of
the carrying arm.
A static certificate must also be provided for the existing interface plate or ceiling anchorage plate.
B
RII
R
D
D
PPeennddaanntt AArrm
G
G
E
m
The data for the calculation of the ceiling load can be taken from the following tables: table 1, maximum
masses and bending moments and table 2, load-bearing capacities. The specified regional safety factors should also be included in calculations.
Table 1:
Type of the pendant arm Vertical mass Bending moments Single arm 3000N 1900Nm Double arm 5300N 3500Nm
PLANLAMA VE MONTAJ TALİMATI BRIDGE-EN-R1 Page 4 / 39
B
Table 2:
Type of the pendant arm Maximum carrying capacities 500mm, 600mm, 650mm, 850mm and 1000mm single arms with single joint 500+500mm Single arm with double joint 150kg 600+600mm Single arm with double joint 125kg 650+650mm Single arm with double joint 115kg
3.2. Definitions:
Figure 1 show an interface plate (3), threaded rods, nuts and washers (4) on the interface plate, ceiling suspension system for intermediate ceilings (2) and a counter plate set (1) for single and double arm type pendant arms.
3.2.1. Interface Plate
BRIDGE pendant arms are installed using an interface plate. Interface plate is the component used for the installation of pendant arms and acts as an interface between the mains supply lines and the supply connections to the pendant arms.
3.2.2. Ceiling Suspension System For Intermediate Ceiling
Nassetti offers a ceiling suspension system for intermediate ceilings for the installation of BRIDGE pendant arms, where the distance between raw and intermediate ceilings necessitates the use of some intermediate ceiling structures.
3.2.3. Counter Plate Set
Nassetti offers a complete counter-plate set, which can be used for fixing with counter–plate.
Fig. 1:
B
RII
R
D
D
PPeennddaanntt AArrm
150kg
G
G
E
m
PLANLAMA VE MONTAJ TALİMATI BRIDGE-EN-R1 Page 5 / 39
B
3.3. Ceiling Height Adjustments
Depending on the ceiling structure, different components are used to fix the pendant arms to the ceiling.
The absolute mounting height is limited between the minimum ceiling height and the maximum ceiling
height. When fixing pendant arms, higher or lower values cannot be used. Precise details on this are listed in Tables 3 and 4.
3.3.1. Solution 1
If the distance between the raw ceiling and the intermediate ceiling is more than 140mm and less than or equal to 240mm for the single arm type pendant arms and more than 240mm and less than or equal to 280mm for the double arm type pendant arms; the interface plate is installed onto the raw ceiling. The threaded bolts on the interface plate can span this small intermediate distance.
Fig. 2:
3.3.2. Solution 2
If the distance between the raw ceiling and the intermediate ceiling is between 241mm and 960mm for the single arm type pendant arms and between 281mm and 960mm for the double arm type pendant arms, the ceiling height is adjusted as follows:
y In the area between the raw ceiling and the intermediate ceiling, using a ceiling suspension system for
intermediate ceiling.
y In the area between the intermediate ceiling and the finished floor, using columns of variable length.
B
1) Height of the raw ceiling from the finished
2) Height of the intermediate ceiling from the
3) Column (variable length)
4) Height of the upper edge of the console
5) Distance between intermediate ceiling and
6) Interface plate
RII
R
floor
finished floor
(1850mm) from the finished floor
raw ceiling
D
D
PPeennddaanntt AArrm
G
G
E
m
PLANLAMA VE MONTAJ TALİMATI BRIDGE-EN-R1 Page 6 / 39
B
Fig. 3:
3.3.3. Solution 3
If the distance between the raw ceiling and the intermediate ceiling is more than 960mm for either single or double arm type pendant arms, the ceiling height is adjusted as follows:
y In the area between the raw ceiling and the intermediate ceiling, using a mains ceiling substructure. y In the area between the intermediate ceiling and the finished floor, using columns of variable length.
Fig. 4:
B
1) Height of the raw ceiling from the finished floor
2) Height of the intermediate ceiling from the
3) Column (variable length)
4) Height of the upper edge of the console
5) Distance between intermediate ceiling and raw
6) Interface plate
7) Ceiling suspension system for Intermediate
8) Ceiling anchorage plate from ceiling suspension
1) Height of the raw ceiling from the finished floor
2) Height of the intermediate ceiling from the
3) Column (variable length)
4) Height of the upper edge of the console
5) Distance between intermediate ceiling and raw
6) Interface plate
7) Ceiling substructure
RII
R
finished floor
(1850mm) from the finished floor
ceiling (≤960mm)
ceiling
system for intermediate ceiling
finished floor
(1850mm) from the finished floor
ceiling
D
D
PPeennddaanntt AArrm
G
G
E
m
PLANLAMA VE MONTAJ TALİMATI BRIDGE-EN-R1 Page 7 / 39
B
3.4. Minimum Ceiling Heights
Table 3, shows the minimum possible ceiling heights for the installation of different pendant arm types.
The distance between the upper and lower edges of the pendant arm console and the finished floor must
remain between ergonomical limits. The values in Table 3 are determined, assuming that the distance between the upper edge of the console and the finished floor is at least 1850mm. In exceptional cases, the height may be different. In these cases, the installation should be made in accordance with the special working drawing.
Table 3:
Type of
pendant arm Single arm with single joint 2324mm 2464mm 140mm Single arm with double joint 2506mm 2646mm 140mm Double arm 2516mm 2756mm 240mm
For intermediate ceiling structures, the distance must not be less than the minimum distance between the raw ceiling and the intermediate ceiling.
3.5. Maximum Ceiling Heights With Basic Components
Table 4, shows the maximum possible ceiling heights for the installation of different pendant arm types when mounting them using a ceiling suspension system for intermediate ceiling.
Table 4:
Type of
pendant arm
Single arm with single joint 3024mm 3984mm 960mm Single arm with double joint 3256mm 4216mm 960mm Double arm 2516mm 3476mm 960mm
3.6. BRIDGE Ceiling Fixation System Components
3.6.1. Interface Plates
The interface plate consists of steel plate, threaded rods, and the appropriate nuts and washers.
3.6.4.2. Interface Plates For Single Arm Type Pendant Arms
Fixing holes:
For fixing to raw ceiling or ceiling substructure: Four long holes 22x50mm (1) at pitch circle diameter 470mm.
For fixing to ceiling suspension system for intermediate ceiling: Six holes ø17mm (2) at pitch circle diameter 470mm.
Material of plate: St 52-3 Dimensions of plate: ø530x15mm
Min. height
intermediate ceiling
Max. height
intermediate ceiling
B
Min. height raw
Max. height raw
ceiling
ceiling
RII
R
D
D
PPeennddaanntt AArrm
Min. distance intermediate
ceiling / raw ceiling
intermediate ceiling / raw
G
G
Max. distance
ceiling
E
m
PLANLAMA VE MONTAJ TALİMATI BRIDGE-EN-R1 Page 8 / 39
B
Fig. 5:
Supplied components:
6 pieces M16x315mm threaded rods DIN 976, material 8.8
18 pieces M16 hexagon nuts DIN 934
12 pieces spring washers
12 pieces plain washers
12 pieces plastic insulating washers
3.6.4.2. Interface Plates For Double Arm Type Pendant Arms
Fixing holes:
For fixing to raw ceiling or ceiling substructure: Four long holes 22x50mm (1) at pitch circle diameter 470mm.
For fixing to ceiling suspension system for intermediate ceiling: Six holes ø17mm (2) at pitch circle diameter 470mm.
Material of plate: St 52-3 Dimensions of plate: ø560x15mm
Fig. 6:
B
RII
R
D
D
PPeennddaanntt AArrm
G
G
E
m
PLANLAMA VE MONTAJ TALİMATI BRIDGE-EN-R1 Page 9 / 39
B
Supplied components:
6 pieces M16x315mm threaded rods DIN 976, material 8.8
6 pieces M16x205mm threaded rods DIN 976, material 8.8
36 pieces M16 hexagon nuts DIN 934
24 pieces spring washers
24 pieces plain washers
24 pieces plastic insulating washers
3.6.2. Suspension Systems For Intermediate Ceiling Structures
The ceiling suspension system for intermediate ceilings consists of steel ceiling anchorage plate, threaded rods, spacer tubes and the appropriate nuts and washers. It is used to span a maximum distance of 960mm between the raw ceiling and the intermediate ceiling.
Fixing holes:
For fixing to raw ceiling: Four long holes 22x50mm (1) at pitch circle diameter 470mm.
For fixing of the threaded rods and spacer tubes of the ceiling suspension system for intermediate ceiling: Six
threaded holes M16 (2) at pitch circle diameter 470mm. Material of plate: St 52-3 Dimensions of plate: 530x490x20mm
Fig. 7:
Supplied components:
y 6 pieces ø 48,4mm spacer tubes (max. 760mm) y 6 pieces M16 threaded rods DIN 976, material 8.8 (max. 830mm) y 12 pieces M16 hexagon nuts DIN 934 y 12 pieces spring washers y 6 pieces plain washers y 12 pieces plastic centering parts for the tubes
B
RII
R
D
D
PPeennddaanntt AArrm
G
G
E
m
PLANLAMA VE MONTAJ TALİMATI BRIDGE-EN-R1 Page 10 / 39
B
3.6.3. Counter-Plate Sets
The counter-plate set consists of four steel plates, threaded rods, and the appropriate nuts and washers.
Fixing holes: y For fixing to raw ceiling: one hole ø22mm (1). Material of plate: St 52-3 Dimensions of plate: ø150x15mm
Fig. 8:
Supplied components:
y 4 pieces M20x400mm threaded rods DIN 976, material 8.8 y 8 pieces M20 hexagon nuts DIN 934 y 8 pieces spring washers y 4 pieces plain washers
3.6.4. Intermediate Ceiling Opening Covers
The intermediate ceiling opening covers consists of two steel cover halves, threaded rods, and the appropriate nuts.
3.6.4.2. Intermediate Ceiling Opening Covers For Single Arm Type Pendant Arms
Fixing holes:
For fixing to interface plate: Four holes ø14,5mm (1) at pitch circle diameter 400mm. Dimensions of plate: 600x600x50mm
Fig. 9:
B
RII
R
D
D
PPeennddaanntt AArrm
G
G
E
m
PLANLAMA VE MONTAJ TALİMATI BRIDGE-EN-R1 Page 11 / 39