NARDA EFA-200, EFA-300 Operating Manual

EFA-200/-300 EM Field Analyzer
Operating Manual
2245/98.22
EFA-200/-300 EM Field Analyzer
2245 /20, 30, Series A ...
Operating Manual
If you have any questions about this product, please contact the Sales Company for your locality.
Narda Safety Test Solutions GmbH Sandwiesenstr. 7 72793 Pfullingen, Germany © 2006
Order no.: 2245/98.22 Edition : 05/06.03, A ...
Previous Edition: 04/05.03, A ... Subject to change without notice.
Our normal guarantee and delivery terms apply.
Printed in Germany
CONTENTS
1Introduction
2 Safety instructions
3 Preparation for use
1.1 About these instruments . . . . . . . . . . . . . . . . . . . 1-1
1.1.1 Applications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.1.2 Common features . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.1.3 Differing features . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
1.2 About this operating manual . . . . . . . . . . . . . . . . 1-3
2.1 Before connecting up . . . . . . . . . . . . . . . . . . . . . 2-1
2.2 Correct use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
2.3 AC Power Supply Unit. . . . . . . . . . . . . . . . . . . . . 2-2
2.4 Faults and excessive stresses . . . . . . . . . . . . . . 2-3
3.1 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.1.1 Packaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
3.1.2 Checking the package contents . . . . . . . . . . . . . 3-1
3.1.3 Checking for transport damage . . . . . . . . . . . . . . 3-1
3.1.4 Recovery after transport and storage . . . . . . . . . 3-1
3.2 Power supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
3.2.1 Operation from rechargeable batteries / dry cells 3-2
3.2.2 Charging the rechargeable batteries . . . . . . . . . . 3-3
3.3 Preparing for B-field measurements . . . . . . . . . . 3-4
3.3.1 Measurement setup. . . . . . . . . . . . . . . . . . . . . . . 3-4
3.3.2 Selecting the probe type . . . . . . . . . . . . . . . . . . . 3-4
3.3.3 Connecting / disconnecting the external
B-Field Probe . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
3.3.4 Switching on the Basic Unit. . . . . . . . . . . . . . . . . 3-5
3.4 Preparing for E-field measurements . . . . . . . . . . 3-8
3.4.1 Measurement setup . . . . . . . . . . . . . . . . . . . . . . . 3-8
3.4.2 Assembling the E-Field Unit measurement
setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
3.4.3 E-Field Unit function test . . . . . . . . . . . . . . . . . . 3-10
4 Control and display elements
4.1 Basic Unit keyboard . . . . . . . . . . . . . . . . . . . . . . . 4-1
4.2 Basic Unit display . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4.2.1 Liquid crystal display . . . . . . . . . . . . . . . . . . . . . . 4-5
4.2.2 Backlight. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4.2.3 Contrast . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
4.3 Basic Unit indicators. . . . . . . . . . . . . . . . . . . . . . . 4-6
4.4 E-Field Unit controls . . . . . . . . . . . . . . . . . . . . . . 4-7
5 Selecting an operating mode
5.1 Operating mode applications . . . . . . . . . . . . . . . . 5-1
5.2 Operating mode features . . . . . . . . . . . . . . . . . . . 5-2
5.2.1 Field Strength mode . . . . . . . . . . . . . . . . . . . . . . . 5-2
5.2.2 Exposure STD mode (Shaped Time Domain) . . . 5-3
5.2.3 Spectrum FFT mode
(Fast Fourier Transformation). . . . . . . . . . . . . . . . 5-4
5.2.4 Harmonic Analysis mode . . . . . . . . . . . . . . . . . . . 5-5
5.3 Checking the EFA operating modes. . . . . . . . . . . 5-6
6 Operating modes
6.1 Field Strength mode . . . . . . . . . . . . . . . . . . . . . . . 6-1
6.1.1 Activating Field Strength mode. . . . . . . . . . . . . . . 6-2
6.1.2 Selecting the measurement range . . . . . . . . . . . . 6-2
6.1.3 Selecting the detection mode . . . . . . . . . . . . . . . . 6-3
6.1.4 Selecting the display mode . . . . . . . . . . . . . . . . . 6-4
6.1.5 Selecting the frequency range . . . . . . . . . . . . . . . 6-4
6.1.6 Evaluating the results . . . . . . . . . . . . . . . . . . . . . . 6-6
6.1.7 Saving data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-8
6.2 Exposure STD mode (Shaped Time Domain) . . . 6-9
6.2.1 Activating STD mode. . . . . . . . . . . . . . . . . . . . . . 6-9
6.2.2 Selecting the measurement range . . . . . . . . . . . 6-9
6.2.3 Selecting the display mode . . . . . . . . . . . . . . . . 6-10
6.2.4 Frequency range . . . . . . . . . . . . . . . . . . . . . . . . 6-10
6.2.5 Evaluating the results . . . . . . . . . . . . . . . . . . . . 6-10
6.2.6 Saving data . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-12
6.3 Spectrum FFT mode . . . . . . . . . . . . . . . . . . . . . 6-13
6.3.1 Activating Spectrum FFT mode. . . . . . . . . . . . . 6-13
6.3.2 Selecting the measurement range . . . . . . . . . . 6-14
6.3.3 Selecting the measurement mode
(detection mode) . . . . . . . . . . . . . . . . . . . . . . . . 6-15
6.3.4 Selecting the display mode . . . . . . . . . . . . . . . . 6-15
6.3.5 Selecting the frequency range. . . . . . . . . . . . . . 6-16
6.3.6 Starting and stopping a measurement. . . . . . . . 6-17
6.3.7 Evaluating measurement results . . . . . . . . . . . . 6-17
6.3.7.1 Zooming in on the frequency axis . . . . . . . . . . . 6-20
6.3.7.2 Zooming in on the measured value axis . . . . . . 6-21
6.3.7.3 Marker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-22
6.3.7.4 Normalized display . . . . . . . . . . . . . . . . . . . . . . 6-24
6.3.8 Saving data . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-25
6.3.9 Possible problems . . . . . . . . . . . . . . . . . . . . . . . 6-25
6.4 Harmonic analysis mode . . . . . . . . . . . . . . . . . . 6-26
6.4.1 Activating Harmonic Analysis mode . . . . . . . . . 6-26
6.4.2 Selecting the measurement range . . . . . . . . . . 6-27
6.4.3 Selecting the measurement mode
(detection mode) . . . . . . . . . . . . . . . . . . . . . . . . 6-28
6.4.4 Selecting the display mode . . . . . . . . . . . . . . . . 6-28
6.4.5 Selecting the fundamental frequency . . . . . . . . 6-29
6.4.6 Starting and stopping a measurement. . . . . . . . 6-30
6.4.7 Evaluating measurement results . . . . . . . . . . . . 6-31
6.4.8 Saving data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-34
6.4.9 Possible problems . . . . . . . . . . . . . . . . . . . . . . . 6-34
7 Measurement data memory
7.1 Saving data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
7.1.1 Saving individual values . . . . . . . . . . . . . . . . . . . . 7-2
7.1.2 Recording timer-controlled measurement
sequences (AUTO). . . . . . . . . . . . . . . . . . . . . . . . 7-3
7.1.3 Recording spatially assigned measurement
sequences (MATRIX) . . . . . . . . . . . . . . . . . . . . . . 7-7
7.2 Displaying a data set . . . . . . . . . . . . . . . . . . . . . 7-14
7.3 Exporting data . . . . . . . . . . . . . . . . . . . . . . . . . . 7-16
7.4 Deleting data . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-18
8 Optional functions
8.1 Selecting an alarm . . . . . . . . . . . . . . . . . . . . . . . . 8-1
8.1.1 Activating the “ALARM” function . . . . . . . . . . . . . 8-1
8.1.2 Entering the alarm threshold . . . . . . . . . . . . . . . . 8-2
8.1.3 Activating alarm signals . . . . . . . . . . . . . . . . . . . . 8-2
8.2 Recording current values . . . . . . . . . . . . . . . . . . . 8-4
8.2.1 Activating the CURRENT function . . . . . . . . . . . . 8-4
8.2.2 Entering current values . . . . . . . . . . . . . . . . . . . . 8-4
8.2.3 Activating recording of current values . . . . . . . . . 8-5
8.3 Choosing a limit curve . . . . . . . . . . . . . . . . . . . . . 8-5
8.4 Selecting the measurement axis . . . . . . . . . . . . . 8-6
8.5 Selecting the measurement units . . . . . . . . . . . . . 8-7
8.6 Setting the averaging parameter NAVG . . . . . . . . 8-7
8.7 Editing the frequency selection list . . . . . . . . . . . . 8-8
8.7.1 Activating the “FREQ LISTS” function . . . . . . . . .8-8
8.7.2 Changing the frequency values . . . . . . . . . . . . . . 8-8
8.8 Setting the date and time . . . . . . . . . . . . . . . . . . . 8-9
8.8.1 Activating the “CLOCK” function . . . . . . . . . . . . . 8-9
8.8.2 Setting the current time . . . . . . . . . . . . . . . . . . . 8-10
8.8.3 Setting the date . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
9 Device data memory
9.1 Calling up a device setting (User Setup). . . . . . . 9-1
9.2 Scope and content of the default settings
(Factory Default Setup) . . . . . . . . . . . . . . . . . . . . 9-1
9.3 Saving device settings (User Setup) . . . . . . . . . . 9-4
10 Maintenance and repairs
10.1 Dry cells and rechargeable batteries . . . . . . . . . 10-1
10.1.1 Replacing the rechargeable batteries / dry cells 10-1
10.1.2 Handling rechargeable batteries . . . . . . . . . . . . 10-3
10.1.3 Lithium battery. . . . . . . . . . . . . . . . . . . . . . . . . . 10-3
10.1.4 Battery disposal. . . . . . . . . . . . . . . . . . . . . . . . . 10-4
10.2 Software update . . . . . . . . . . . . . . . . . . . . . . . . 10-4
10.2.1 Installing the software “EFA-200/-300 Tools”. . . 10-5
10.2.2 Setting up the EFA – PC connection. . . . . . . . . 10-6
10.2.3 Loading the device firmware . . . . . . . . . . . . . . . 10-6
10.3 Transferring STANDARDS parameters . . . . . . . 10-8
10.4 Enabling operating modes. . . . . . . . . . . . . . . . . 10-9
10.4.1 Entering the access code . . . . . . . . . . . . . . . . 10-10
10.5 Reading out the device characteristics . . . . . . 10-11
10.6 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-12
10.7 Power Supply Unit . . . . . . . . . . . . . . . . . . . . . 10-13
10.7.1 Opening the device . . . . . . . . . . . . . . . . . . . . . 10-13
10.7.2 Repairs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-13
10.7.3 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-13
11 Remote control
11.1 Abbreviations used. . . . . . . . . . . . . . . . . . . . . . . 11-1
11.2 Making the connection from EFA to PC . . . . . . . 11-1
11.3 Structure of device-specific commands . . . . . . . 11-3
11.4 Remote control commands . . . . . . . . . . . . . . . . 11-7
11.4.1 Measurement settings . . . . . . . . . . . . . . . . . . . . 11-8
11.4.2 Reading out current measurement values . . . . 11-12
11.4.3 Measurement data memory . . . . . . . . . . . . . . . 11-16
11.4.4 Optional functions. . . . . . . . . . . . . . . . . . . . . . . 11-22
11.4.5 General functions . . . . . . . . . . . . . . . . . . . . . . . 11-28
11.4.6 Test / Service . . . . . . . . . . . . . . . . . . . . . . . . . . 11-28
12 Specifications
12.1 Principle of operation . . . . . . . . . . . . . . . . . . . . . 12-1
12.2 Field Strength mode . . . . . . . . . . . . . . . . . . . . . . 12-1
12.2.1 Brief description . . . . . . . . . . . . . . . . . . . . . . . . . 12-1
12.2.2 Frequency range . . . . . . . . . . . . . . . . . . . . . . . . 12-2
12.2.3 Measurement range, dynamic range . . . . . . . . . 12-2
12.2.4 Measured value detection . . . . . . . . . . . . . . . . . 12-4
12.2.5 Averaging time / Refresh rate. . . . . . . . . . . . . . . 12-5
12.2.6 Measurement uncertainty . . . . . . . . . . . . . . . . . . 12-5
12.2.7 Result display . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-6
12.2.8 Frequency measurement
(Broadband setting only) . . . . . . . . . . . . . . . . . . 12-6
12.2.9 Auxiliary functions . . . . . . . . . . . . . . . . . . . . . . . 12-6
12.3 Exposure STD (Shaped Time Domain) mode . .12-7
12.3.1 Brief description . . . . . . . . . . . . . . . . . . . . . . . . . 12-7
12.3.2 Frequency range . . . . . . . . . . . . . . . . . . . . . . . . 12-7
12.3.3 Measurement range, dynamic range . . . . . . . . . 12-7
12.3.4 Limit value curves. . . . . . . . . . . . . . . . . . . . . . . . 12-8
12.3.5 Measurement value detection . . . . . . . . . . . . . . 12-9
12.3.6 Averaging time / Refresh rate . . . . . . . . . . . . . . 12-9
12.3.7 Measurement uncertainty . . . . . . . . . . . . . . . . . 12-9
12.3.8 Result display . . . . . . . . . . . . . . . . . . . . . . . . . 12-10
12.3.9 Additional functions . . . . . . . . . . . . . . . . . . . . . 12-10
12.4 Spectrum FFT . . . . . . . . . . . . . . . . . . . . . . . . . 12-10
12.4.1 Brief description. . . . . . . . . . . . . . . . . . . . . . . . 12-10
12.4.2 Frequency range . . . . . . . . . . . . . . . . . . . . . . . 12-11
12.4.3 Measurement range . . . . . . . . . . . . . . . . . . . . 12-11
12.4.4 Recording of measured values . . . . . . . . . . . . 12-14
12.4.5 Averaging / measured value rate . . . . . . . . . . 12-14
12.4.6 Result display . . . . . . . . . . . . . . . . . . . . . . . . . 12-15
12.4.7 Additional functions . . . . . . . . . . . . . . . . . . . . . 12-15
12.5 Harmonic Analysis mode. . . . . . . . . . . . . . . . . 12-15
12.5.1 Brief description. . . . . . . . . . . . . . . . . . . . . . . . 12-16
12.5.2 Frequency range . . . . . . . . . . . . . . . . . . . . . . . 12-16
12.5.3 Result display . . . . . . . . . . . . . . . . . . . . . . . . . 12-16
12.6 Miscellaneous . . . . . . . . . . . . . . . . . . . . . . . . . 12-17
12.6.1 Display type and warning devices . . . . . . . . . . 12-17
12.6.2 Data memory. . . . . . . . . . . . . . . . . . . . . . . . . . 12-17
12.6.2.1 Brief description. . . . . . . . . . . . . . . . . . . . . . . . 12-18
12.6.2.2 Memory capacity . . . . . . . . . . . . . . . . . . . . . . . 12-18
12.6.2.3 Clock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-18
12.6.3 Device settings memory . . . . . . . . . . . . . . . . . 12-19
12.6.4 Self test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-19
12.6.5 Remote control interface (serial) . . . . . . . . . . . 12-19
12.6.6 Power supply. . . . . . . . . . . . . . . . . . . . . . . . . . 12-20
12.6.6.1 Battery operation . . . . . . . . . . . . . . . . . . . . . . . 12-20
12.6.6.2 Charger unit. . . . . . . . . . . . . . . . . . . . . . . . . . . 12-21
12.7 General specifications . . . . . . . . . . . . . . . . . . . 12-21
12.7.1 Climatic environment conditions . . . . . . . . . . . 12-21
12.7.2 Conformance marks . . . . . . . . . . . . . . . . . . . . 12-22
12.7.3 Declaration of origin . . . . . . . . . . . . . . . . . . . . . 12-22
12.7.4 Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . 12-22
12.7.4.1 Calibration interval . . . . . . . . . . . . . . . . . . . . . . 12-22
12.7.4.2 Data memory battery (lithium battery) . . . . . . . 12-23
12.8 Dimensions and weight . . . . . . . . . . . . . . . . . . 12-23
12.9 Ordering information. . . . . . . . . . . . . . . . . . . . . 12-23
12.9.1 Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12-24
Annex A: Glossary
Annex B: Determining the matrix
Annex C: Fundamentals of FFT
Annex D: Data export format
Index
Operating concept
Chap. 3.3 Chap. 3.4
Chap. 6
Chap. 6
Chap. 7
Chap. 8
Chap. 9
Selecting an
operating mode
Chap. 5
Activating an
operating mode
Measurement
Measurement
data memory
Further
settings
Device
data memory

1 Introduction

1.1 About these instruments

1.1.1 Applications

Many people live and work in the vicinity of power lines and power distribution equipment. For safety’s sake, it must be ensured that these people are not exposed to any electromagnetic fields that could be injurious to health.
Global efforts to introduce effective protection in recent years have led to a range of national and international guidelines and standards. These set permitted limit values for electric and magnetic field strengths for various frequency ranges and signal shapes.
In practice, this means using simple equipment to measure the fields that occur at a particular location or place of work with sufficient accuracy to check the effectiveness of protective measures, ensure that the limit values stipulated by the standards and guidelines are not exceeded and to dispel any doubts in the minds of those affected. Lightweight, portable, battery­operated measuring instruments, which cover the dynamic range and frequencies of interest are needed for this. They must also be easy to use and deliver results that can be directly compared with the current limit values.

1.1.2 Common features

The EFA-200 and EFA-300 Field Analyzers are devices for monitoring low­frequency fields in the range 5 Hz to 32 kHz.
All instrument functions are designed for ease of use under practical conditions. Operation has been arranged so that the instruments can be set as simply as possible using function keys and menu keys. Menu keys are used to switch between the four main menus: MODE, CONFIGURATION, MEMORY and USER.
The MODE key thus gives access to up to 5 different operating modes. These modes can be activated individually at any time as needed, so you can always adapt your EFA to the specific task at hand.
1-1
1 Introduction EFA-200/-300
The MEM key allows you to record a timer-controlled set of measurements or to save the measured values in a spatial matrix, and to recall the stored results as required.
The CONF key enables access to additional functions such as activation of an alarm, entry of date and time information or the selection of stored limit value curves.
Function keys are used in all menus to enable access to further measurement functions.
By default, the parameters set the last time the instrument was used are set automatically when you switch on again. You can continue measuring straight away. You can also save your own settings under any name and recall these as required using the USER key.

1.1.3 Differing features

The main differences between the EFA-200 and EFA-300 result from the different measurement objectives. They are readily apparent from the different measurement setups that are used.
EFA-200 – magnetic field measurements
The EFA-200 is for measuring magnetic fields.
The measurement setup consists of the Basic Unit and a B-field probe. One B-field probe is already built in to the Basic Unit. Two external B-field probes of different designs are available for making more accurate measurements.
1-2
EFA-300 – electric and magnetic field measurements
The EFA-300 additionally measures electric fields.
The measurement setup also includes a Basic Unit, which is identical to the B-field measuring instrument as regards function and operation. An internal B-field probe is also fitted and the same external probes can also be used.
E-field measurements become possible when the cube-shaped E-Field Unit – a precisely aligned plate condenser – is connected. The connection between the Basic Unit and the E-Field Unit is a fiber optic cable. This allows the highly-sensitive E-Field Unit to be operated from a distance to ensure that the effects on the field caused by the operator do not influence the measurement.
EFA-200/-300 1 Introduction
In this measurement setup, the Basic Unit is only used to operate the E-Field Unit. The E-Field Unit performs all measurement functions and also stores the results independently of the Basic Unit. This unusual solution makes it possible to start a long-term measurement with the E-Field Unit and then to disconnect the Basic Unit and the fiber optic cable and make magnetic field measurements using the Basic Unit without the E-Field Unit.

1.2 About this operating manual

The two device types EFA-200 and EFA-300 differ in their measurement objectives (magnetic flux density or electric field strength). However, operation of the devices is to a large extent identical since the basic functions of the EFA-300 and EFA-200 are the same.
This is why it is possible to include both devices in a single operating manual and consider the two side by side. Important differences in handling and specifications are covered in the chapters “Preparation for use”, “Control and display elements” and “Specifications”.
Major differences are found, however, in chapters 5 to 9, which consider the operation of the devices.
In these chapters, indications that apply to both units refer to the common name of EFA.
Functions or restrictions that apply to only one of the instruments are preceded by an appropriate note, e.g. For EFA-200 only.
Restrictions in function are also preceded by appropriate notes. For example, some of the types of data set cannot be saved as individual values in the memory. The section “Saving individual values” is therefore prefixed with the note: For data set types Normal, List, Vector only.
1-3
1 Introduction EFA-200/-300
Notes:
1-4

2 Safety instructions

2.1 Before connecting up

This instrument left the factory in perfect condition. To ensure safe usage and to preserve it in good condition, please follow the safety instructions below.

2.2 Correct use

The instrument must only be used under the conditions and for the purposes for which it was designed.
Electric shock
The EFA range of instruments is not provided with any special insulation features.
Danger
Do not touch any parts carrying dangerous voltages with the measuring
instrument or the probe.
Strong fields
Very strong fields occur in the vicinity of powerful radiation sources. Persons fitted with electronic implants (pacemakers, etc.) must avoid
such areas.
Danger of explosion
Alkaline batteries can explode if you attempt to recharge them. Before starting to recharge the batteries, make sure that rechargeable
cells and not alkaline batteries are fitted in the battery compartment.
Check that the switch inside the battery compartment is set correctly.
2-1
2 Safety instructions EFA-200/-300
Malfunction
The accuracy and function of the instrument can be impaired by incorrect use, damage and unauthorized repairs.
Caution
Only use the instrument under the conditions and for the purposes for
which it was designed.
Regularly check the instrument for any signs of damage.Only allow repairs to be made by trained service engineers.
Unsuitable measuring range
The EFA series of instruments has a limited frequency range that extends
Hz up to 32 kHz. Fields outside this range will normally be assessed
from 5 incorrectly.
Select a suitable measuring range.Only measure within the valid measuring ranges.

2.3 AC Power Supply Unit

Electric shock
Parts carrying live voltages, which may cause injury can be exposed if the Power Supply Unit is damaged.
Danger
Never use a damaged Power Supply Unit.
Caution
2-2
Damage
Incorrect AC line voltages, condensation, extreme temperatures and poor ventilation can all result in damage to the Power Supply Unit.
Before connecting up the Power Supply Unit check that the operating
voltage of the Power Supply Unit matches the local AC line voltage.
Do not use a Power Supply Unit if condensation has formed on it. If
condensation is unavoidable, such as when the Power Supply Unit is cold and is brought into a warmer environment, it must be allowed to dry out completely before it is connected up.
Only use the Power Supply Unit indoors at temperatures between +5 °C
and +45 °C.
Ensure adequate ventilation for cooling during operation.
EFA-200/-300 2 Safety instructions

2.4 Faults and excessive stresses

The instrument must be taken out of service and secured against unauthorized use if safe operation can no longer be guaranteed. This is the case, for example:
if the instrument shows visible signs of damage
if the instrument no longer functions
if the instrument has been subjected to excessive stresses of any kind
that are beyond the permitted limits.
In such cases, please contact your local Service Center. The addresses are listed at the end of this operating manual.
2-3
2 Safety instructions EFA-200/-300
Notes:
2-4

3 Preparation for use

3.1 Unpacking

3.1.1 Packaging

Do not discard the packing materials. These items have been specially designed to be reused if they are not damaged. Always use the original packaging when transporting the instrument if possible.

3.1.2 Checking the package contents

Ordering information: see page 12-23

3.1.3 Checking for transport damage

After unpacking, check the instrument and all accessories for transport damage. Such damage is likely if the packaging itself has been severely damaged. Do not attempt to use an instrument which is visibly damaged.

3.1.4 Recovery after transport and storage

Condensation may form on an instrument which has been stored or transported at a low temperature when it is brought into a warm environment. To prevent any damage, wait until all signs of condensation on the instrument surfaces have disappeared before switching the instrument on. The instrument is only ready for use when it has reached the guaranteed operating temperature range of 0 to +50
°C.
3-1
3 Preparation for use EFA-200/-300

3.2 Power supply

The EFA-200 / -300 Basic Unit operates from dry cells or rechargeable
batteries.
The E-Field Unit for the EFA-300 contains built-in rechargeable
batteries.

3.2.1 Operation from rechargeable batteries / dry cells

Replacing the rechargeable batteries / dry cells: see page 10-1
Operation from rechargeable batteries
The Basic Unit is operated from five 1.2 V NiMH batteries (baby or size C,
R 14).
IEC
Fully charged batteries will provide power for about 10 hours of operation. The batteries are supplied pre-charged, but will need to be fully charged before the device is used for the first time.
Operation from dry cells
Basic Unit only
Five 1.5 V alkaline batteries (baby or size C, IEC LR 14) can be used in the Basic Unit instead of rechargeable batteries.
These batteries will provide power for about 20 hours of operation.
3-2
Low battery display
The message “LOW BATTERY” appears in the display when the rechargeable or dry batteries are almost discharged.
Note: You can continue measurements for about 15 minutes after the
“LOW BATTERY” warning appears.
The built-in lithium battery provides power for the result memory when the batteries are completely discharged or when they are being replaced.
Lithium battery: see page 10-3
EFA-200/-300 3 Preparation for use

3.2.2 Charging the rechargeable batteries

A complete recharge takes about 3 hours.
Danger of explosion
Alkaline batteries can explode if you attempt to recharge them. Before starting to recharge the batteries, make sure that rechargeable
Danger
cells and not alkaline batteries are fitted in the battery compartment.
Check that the switch inside the battery compartment is set correctly.
Starting the charging cycle
The local AC line voltage matches the operating voltage of the Power
Supply Unit.
The correct AC line adapter has been selected.
1. Connect the Power Supply Unit to the EFA charger socket.
2. Connect the Power Supply Unit to the AC line power supply.
The charge cycle starts.
The yellow CHARGE indicator LED lights up.
When the batteries are fully charged, the Power Supply Unit switches
to trickle charge mode automatically. The yellow CHARGE indicator LED flashes rapidly.
Possible malfunctions
The yellow CHARGE indicator LED flashes slowly.
Batteries are not ready to be charged, e.g. because they have not yet reached the charging temperature range (+10 °C to +40 °C).
Stop the charging cycle and start again later.
– or – Place the Power Supply Unit and the EFA in a warmer room.
3-3
3 Preparation for use EFA-200/-300

3.3 Preparing for B-field measurements

3.3.1 Measurement setup

The complete measurement setup for measuring a B-field consists of:
the Basic Unit
an external B-Field Probe (optional)
No person or object should be between the radiation source and the probe while the measurement is being made.

3.3.2 Selecting the probe type

B-Field Probes are used to measure magnetic fields. Three different types are available for use in different environments or for different measurement purposes.
Internal B-Field Probe
– The internal B-Field Probe is used when fields are to be detected and
measured with a minimum of work.
– Measurement accuracy: approx. 6 %
External B-Field Probe with d = 3 cm
– This very small B-Field Probe can be used for pinpoint
measurements in locations that are narrow or difficult to access.
– Measurement accuracy: approx. 6 %
External precision B-Field Probe with 100 cm2 cross-sectional area
– A probe with an effective cross-sectional area of 100 cm2 is stipulated
for measurements conforming to VDE standard 0848. The large surface eliminates the effects of small eddy fields and guarantees highly accurate measurements.
– Measurement accuracy: approx. 3%
Note: Physical constraints result in a higher background display level
when the built-in B-Field Probe and the external 3 cm diameter
field probe are used.
B It is possible that the sensitive measurement ranges are only completely usable when the external precision B-Field Probe with
2
100 cm
cross-sectional area is used.
3-4
EFA-200/-300 3 Preparation for use

3.3.3 Connecting / disconnecting the external B-Field Probe

The 12-way connector socket for the external B-field probe is on the head of the EFA.
Locking sleeve
Figure 3-1 External B-Field Probe
Connecting the B-Field Probe
1. Place the probe on the Basic Unit so that the red triangle on the probe points in the same direction as the red dot on the Basic Unit.
Push the plug into the Basic Unit with the locking sleeve until the plug
lock engages.
The probe is now connected.
The Basic Unit automatically detects the probe type that has been
connected. The display shows “B” for B-Field Probe.
Removing the B-Field Probe
Pull the locking sleeve on the plug towards the probe.
The plug lock disengages and the probe can be removed.

3.3.4 Switching on the Basic Unit

You have made sure that the batteries have enough power for the
measurement.
Briefly press the ON/OFF key.
The Basic Unit switches on. The green ON indicator LED lights up. The function test starts.
3-5
3 Preparation for use EFA-200/-300
Function test
The function test checks the internal memory, display, signal processor and the internal signal path.
While the function test is running:
– the test screen is displayed – the red Alarm LED flashes.
Figure 3-2 Test screen
If the function test is completed successfully:
– all the settings from the last measurement are retained – a beep is heard – the measurement screen is displayed. – Measurements can now be started.
3-6
Figure 3-3 B-field measurement screen
EFA-200/-300 3 Preparation for use
Possible malfunctions
The instrument does not operate.
The batteries are completely discharged. Recharge the rechargeable batteries or replace the dry cells.
Charging the rechargeable batteries: see page 3-3
Replacing the rechargeable batteries / dry cells: see page 10-1
– or –
A fatal error has occurred. Perform a reset, possibly also recall basic settings.
Basic Unit keyboard: see page 4-1
Display remains blank after switching on.
Display contrast is set too high / too low. Check the display contrast setting.
Contrast: see page 4-5
The default setting always appears after switching on.
The lithium battery is exhausted. Contact the Service Center.
Lithium battery: see page 10-3
Two beeps are heard close together.
The self test detected missing data in the device configuration. The instrument will be reset to the default setting automatically.
Measurements can be started.
Three beeps are heard close together and the error message “ERROR DETECTED” appears in the display.
The function test cannot be completed because of an internal fault in the instrument.
Contact the Service Center.
3-7
3 Preparation for use EFA-200/-300

3.4 Preparing for E-field measurements

3.4.1 Measurement setup

For EFA-300 only
The E-Field Unit is used to measure electric fields.
Unlike measurements of magnetic fields, electric field measurements are easily influenced by persons or objects in the vicinity of the sensor.
The complete EFA-300 measurement setup for electric fields includes:
The E-Field Unit fitted on a tripod
The Basic Unit
A fiber optic cable for connecting the E-Field Unit to the Basic Unit
3-8
Basic Unit
Figure 3-4 E-Field Unit measurement setup
The E-Field Unit is a stand-alone instrument that takes care of all processes, measurements and result storage. The EFA-300 Basic Unit serves only as a remote control unit in this setup, and can be used for other measurements once the settings have been made.
Table 3.1 shows the various functions of each part of the measurement setup.
E-Field Unit on tripod
Loading...