The instrument serial number and compliance information can be found on the rating plate on the back panel of each instrument. Please
refer to the rating plate and declaration of conformity (EU only) for compliance information.
Both instruments contain a touchscreen to enable operation of the instrument. The touchscreen is a touch-sensitive method of controlling
the instrument that enables the user to operate the system by touching a selection on the screen. Several buttons will appear in the
touchscreen user interface, such as:
• Next—Proceeds to the next screen.
• Back—Moves back to the previous screen.
• Cancel—Returns to beginning of the current workflow or the main menu.
|
10
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
B. Prep Station
The Prep Station is a multi-channel pipetting robot configured specifically for use with NanoString nCounter Test Kits. The instrument
performs liquid transfers, magnetic bead separations, and immobilization of molecular labels on the Sample Cartridge sur face (FIGURE 2.1).
FIGURE 2.1: The Prep Station
The Prep Station deck must be loaded with the appropriate consumables prior to use (FIGURE 2.2).
FIGURE 2.2: The Prep Station Deck
2017-07 MAN-C0034-06
|
11
USER MANUAL nCounter® Dx Analysis System
C. Digital Analyzer
The Dig ital Ana lyzer is a mult i-chann el epifluor escence sca nner config ured speci fically f or use with Na noString ’s nCounter Test Ki t Cartr idges.
Up to six Cartridges can be loaded into the instrument (FIGURE 2.3). When one or more scans have been completed, the instrument may be
paused to allow new Cartridges to be swapped in without disrupting the remaining Cartridges.
FIGURE 2.3: The Digital Analyzer
|
12
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
3
Selecting the Instrument Mode
Users who have the FLEX configuration are able to run their instruments in either Life Sciences or Diagnostics (Dx) application modes. The
FLEX configuration must be enabled by NanoString Support.
NOTE: If the FLEX configuration has not been enabled, the instructions provided in this chapter will not apply; only the
Diagnostics application mode will be available. Proceed to Chapter 4.
All NanoString IVD assays must be performed using the Diagnostics mode on these instruments. In Diagnostics mode, procedures and
controls are in place to restrict access to data and some features to authorized users only. For all other applications, individuals using
nCounter CodeSets or nCounter Elements should choose the Life Sciences mode and refer to the nCounter Analysis System User Manual for
further instruction.
017-07 MAN-C0034-06
|
13
USER MANUAL nCounter® Dx Analysis System
A. Prep Station
Select Instrument Mode
After turning on the Prep Station, the first screen asks the user to select either Diagnostics mode (blue, on the left) or Life Sciences mode
(green, on the right).
FIGURE 3.1: The Prep Station ‘Select Instrument Mode’ screen
Touch the blue icon labeled NanoString® Diagnostics to enter Diagnostics mode. The system will load the application and present the
Welcome screen (FIGURE 3.2). Before operating the Prep Station, a user must log on by selecting the Main Menu button.
FIGURE 3.2: The ‘Welcome’ screen
>>> The Sign In screen will appear (FIGURE 3.3).
|
14
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
Enter a valid user name and password and touch the Sign In button.
FIGURE 3.3: The sign in screen
>>> The Main Menu will appear (FIGURE 3.4).
FIGURE 3.4: The Prep Station Main Menu in Diagnostics mode
2017-07 MAN-C0034-06
|
15
USER MANUAL nCounter® Dx Analysis System
Switch Instrument Mode
The user may switch between Diagnostics mode and Life Sciences mode from the Main Menu. Press the Exit button located at the bottom
of the Main Menu (FIGURE 3.4).
>>> The ‘Exit Diagnostics Mode’ confirmation screen will appear.
FIGURE 3.5: The Prep Station ‘Exit Diagnostics Mode’ confirmation screen
Tou ch OK to exit Diagn ostics mode and ret urn to the ‘Select Instrument Mode’ screen (FIGURE 3.1). Touch Cancel to ret urn to the Main M enu.
|
16
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
B. Digital Analyzer
Select Instrument Mode
After turning on the Digital Analyzer, the first screen to appear asks the user to select either Diagnostics mode (blue, on the left) or Life
Sciences mode (green, on the right).
FIGURE 3.6: The Digital Analyzer ‘Select Instrument Mode’ screen
Touch the blue icon labeled NanoString® Diagnostics to enter Diagnostics mode. The system will load the application and present the
Welcome screen (FIGURE 3.7). Before operating the Digital Analyzer, a user must log on by selecting the Main Menu button.
FIGURE 3.7: The ‘Welcome’ screen
>>> The sign in screen will appear (FIGURE 3.8).
2017-07 MAN-C0034-06
|
17
USER MANUAL nCounter® Dx Analysis System
Enter a valid user name and password and press the Sign In button.
FIGURE 3.8: The sign in screen
>>> The Main Menu will appear (FIGURE 3.9).
18
FIGURE 3.9: The Digital Analyzer Main Menu in Diagnostics mode
|
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
Switch Instrument Mode
The user may switch between Diagnostics mode and Life Sciences mode from the Main Menu. Press the Exit button located at the bottom
of the Main Menu (FIGURE 3.9).
>>> The ‘Exit Diagnostics Mode’ confirmation screen will appear.
FIGURE 3.10: The Digital Analyzer ‘Exit Diagnostics Mode’ confirmation screen
Tou ch OK to exit D iagnostics mode and ret urn to the ‘Se lect Instr ument Mode’ screen (FIGURE 3.6). Touch Cancel to retur n to the Main Menu .
2017-07 MAN-C0034-06
|
19
USER MANUAL nCounter® Dx Analysis System
4
Operating the Web Application
This chapter provides directions for using the nCounter web application, which resides on a server contained within the nCounter Digital
Analyzer. When the system is connected to a net work, the web application can be used to communicate with both the Prep Station and the
Digital Analyzer. Its primary functions are as follows:
• Create and edit Run Sets
• View the statuses of Run Sets
• Download reports
• Perform administrative functions
A. Sign In and Profile Management
Sign In
The local nCounter system administrator must provide each user with the URL to access the nCounter web application and will also set up
the user account. Navigate to the URL provided from any computer connected to the organization’s local network. (It must be the same
network used by the Digital Analyzer.)
The Sign In page will appear. Enter the user ID and password provided by the administrator and click on the Sign In button (FIGURE 4.1).
FIGURE 4.1: The nCounter web application Sign In page
|
20
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
• User permissions for single assay: If the user has access to only one assay, the Dashboard page for that assay will immediately be
displayed (FIGURE 4.15).
• User permissions for multiple assays: If the user has access to more than one assay, the Ass ay Type control will be displayed. The
user will then need to select an assay, then click Continue (FIGURE 4.2).
FIGURE 4.2: The Assay Type control on the Sign In page
Forgot User ID/Password
• If you forget your user ID, contact your local nCounter system administrator to retrieve it.
• If the password is forgotten, the nCounter web application may be able to retrieve it.
• Click the link next to “Forgot your password?” to access the forgotten password submission page (FIGURE 4.3).
FIGURE 4.3: The submission page for a forgotten password
2017-07 MAN-C0034-06
|
21
USER MANUAL nCounter® Dx Analysis System
• Enter your user ID.
• If the nCounter system finds a matching profile, the password will be e-mailed to the address on record (FIGURE 4.4).
• If a matching profile is not found, you will be instructed to contact your nCounter system administrator in order to reset the
password.
FIGURE 4.4: The confirmation page after a password has been requested
|
22
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
Updating My Profile
User information can be updated by selecting the My Profile button from the menu bar at the top of the page (FIGURE 4.5).
FIGURE 4.5: Location of the My Profile button
The My Profile page enables the user to change the account password and/or e-mail address associated with the profile and view other
account information (FIGURE 4.6). See Section G: Manage Users for more information about user types and privileges.
FIGURE 4.6: The My Profile page
To accept any changes made to a profile, click the Save button. To discard any changes and return to the previous page, click the Cancel
button.
2017-07 MAN-C0034-06
|
23
USER MANUAL nCounter® Dx Analysis System
B. Application Layout and Navigation
Menu
The nCounter web application has a menu bar at the top that enables users to quickly navigate from one area of the application to another
(FIGURE 4.7). The menu items separate the application into five sections:
• Dashboard – Common tasks and statuses of recent activities
• Runs – Create a Run Set and view run statuses (a Run Set consists of 1–10 diagnostic samples and two mandatory reference
samples processed at the same time)
• Samples – View sample statuses
• Reports – Download repor ts
• Admin – Manage users and other system settings
FIGURE 4.7: The web application main menu.
Most of the menu options have sub-menus that are available when a user hovers over the item (FIGURE 4.8). If no sub-menus are available,
the user may simply click on the menu item.
FIGURE 4.8: Sub-menus are available for some menu options.
|
24
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
Table s
The nCounter web application displays tables to quickly view Run Set statuses, sample statuses, users, and reports.
Filter
Tables can become large after frequent use of the nCounter web application, making it difficult to find the data of interest. Filtering is
available in all tables to enable users to search for and view only data of interest.
By default, filtering is turned off for most status pages. (One exception: By default, reference samples are filtered out on the Samples
page.) To turn filtering on, click the + symbol next to the Filter Settings header. The header will expand and display available filter settings
(FIGURE 4.9).
FIGURE 4.9: Filter settings are available when viewing most data
NOTE: Filtering is not available for the Manage Users page.
Filter the data using any of the fields in the table. This will return any items that match the entered text anywhere in that field. Likewise,
users can choose to show all items or only items created recently by specifying the time frame of interest. To change the number of items
displayed per page, select the desired page size from the drop down menu (FIGURE 4.10).
2017-07 MAN-C0034-06
|
25
USER MANUAL nCounter® Dx Analysis System
FIGURE 4.10: An example of filter settings on the Run Sets page
Click Apply when finished. Only relevant rows will be displayed in the table (FIGURE 4.11).
NOTE: The presence or absence of a funnel icon next to the page title indicates whether or not filtering is applied.
26
FIGURE 4.11: An example of the funnel icon and filtered results on the Run Sets page
Different filters are available depending on the page displayed. For example, the Samples page provides the ability to hide the t wo reference
samples required for every run (FIGURE 4.12).
FIGURE 4.12: Check the Omit box to show or hide the reference samples
|
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
Sort
All table columns allow users to sort t he displayed rows in ascen ding or descending order. Click on the column heading for the data of interest
to sort the rows (FIGURE 4.13). To re-sort the data in the reverse order, click again on the same column heading.
FIGURE 4.13: An example of alphabetically sorting samples by clicking on the SAMPLE ID column heading
NOTE: Ta bles are typ ically displaye d by default wit h the most recently added items at the top and the oldest items at the bottom.
2017-07 MAN-C0034-06
|
27
USER MANUAL nCounter® Dx Analysis System
Forms
Forms are pages for manual data entry changes. Users with the ‘Create Run Set’ account privilege have access to the ‘Create Run Set’ and
‘Edit Ru n Set’ forms . Users who lac k this privil ege may view Run Se t and sample s tatuses , but may not creat e or edit Run Set s. Administ rators
may also access other forms, which include:
• Add New User
• Edit User
• Date and Time
• IP Address
• SSH Settings
• E-mail Configuration
All forms are designed with the expectation that users complete an action prior to moving to another area within the web application
(FIGURE 4.14). In order to navigate to another function after completing a form, click Save to retain the information entered or Cancel to exit
the form and discard any changes.
28
FIGURE 4.14: The menu bar is inaccessible while editing the contents of a form
|
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
C. Dashboard and Status Pages
Dashboard
The Dashboard is the landing page, the first screen displayed after logging into the nCounter web application (FIGURE 4.15). The Dashboard
can be accessed from other pages by clicking on the Dashboard menu item, but it will be unavailable when editing a form until that form has
been saved or discarded.
FIGURE 4.15: The Dashboard landing page and the location of the Dashboard menu option
The Dashboa rd provides quick access to the s tatuses of Run Sets, Samples, and Repor ts (assuming appropriate privileges have been applied
to the user profile). To view the complete status of any of these, click on the Show All button on the right hand side of the title bar.
The Dashboard also provides quick access to common tasks located on the left side of the page. Click on the appropriate link to navigate and
perform the desired action.
All navigations provided on the Dashboard can be accessed individually by using the menu bar at the top (FIGURE 4.15). The Dashboard
provides a consolidated view of these to make it easy to perform any function quickly from one place.
2017-07 MAN-C0034-06
|
29
USER MANUAL nCounter® Dx Analysis System
Run Set Status
The Run Sets page can be accessed from the Dashboard as described above, and is also available by clicking Runs on the menu bar.
The Run Sets page displays the statuses of all Run Sets that have been created (FIGURE 4.16)
FIGURE 4.16: Statuses of all Run Sets
|
30
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
The following information is available:
1. Run Set ID – The Run Set ID entered when the Run Set was created
2. Status – The current condition or state of the Run Set. Status possibilities include:
• BatchRegistered – Run Set Registered (Run Set is defined but has not started processing yet)
• PostHybProcessing – Prep Station Processing (Run Set is currently processing on the Prep Station)
• PostHybComplete – Prep Station Completed (Run Set has completed on the Prep Station)
• PostHybAbort – Prep Station Abort (manual abor t on the Prep Station)
• PostHybError – Prep Station Error (an error occurred during processing on the Prep Station)
• ScanProcessing – DA Scan Processing (currently processing on the Digital Analyzer)
• ScanError – DA Scan Error (an error occurred during processing on the Digital Analyzer)
• ScanAbort – DA Scan Abort (manual abort on the Digital Analyzer)
• BatchComplete – Batch Complete (Run Set has completed processing)
• ReportPending – Report Pending (scanning complete, but waiting for the algorithm to complete)
• ReportProcessing – Report Processing (algorithm is running but the report has not been produced yet)
• ReportComplete – Report Complete (Digital Analyzer has finished scanning and reports are available for download from the
Reports page)
• ReportError – Report Failed (all processing for the Run Set is complete but a report was not generated due to an error in the
algorithm)
• ReportCompleteWithError – Report Complete with Error (all processing for the Run Set is complete, and a report generated, but
the assay failed)
3. Created – The date the Run Set was initially created using the web application
4. Created By – The User ID that created the Run Set using the web application
5. Initiated By – The User ID that started processing the Run Set on the Prep Station
To view more details of the individual samples for a Run Set of interest, click on the hyperlink for that Run Set (FIGURE 4.16). The application
will navigate to the ‘Samples’ page and display sample information for the selected Run Set.
2017-07 MAN-C0034-06
|
31
USER MANUAL nCounter® Dx Analysis System
Sample Status
Similar to the Run Sets page, the Samples page can be accessed from the Dashboard and is also available from the Samples menu option
at the top of the page (FIGURE 4.17).
FIGURE 4.17: The location of the Samples menu option
The Samples page allows users to view the statuses of all samples for all Run Sets (FIGURE 4.18)
FIGURE 4.18: The statuses of all samples contained within all Run Sets
|
32
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
The following information is available:
1. Sample ID – The Sample ID entered manually or via a barcode scanner when creating the Run Set
2. Run Set ID – The Run Set ID entered when creating the Run Set in the web application
3. Status – The condition or state that the sample is currently in. The following are possible statuses for a sample:
• SampleRegistered – Sample Registered (Sample is defined; has not started processing yet)
• PostHybProcessing – Prep Station Processing (Run Set is currently processing on the Prep Station)
• PostHybComplete – Prep Station Completed (Run Set has completed on the Prep Station)
• PostHybAbort – Prep Station Abort (manual abor t on the Prep Station)
• PostHybError – Prep Station Error (an error occurred during processing on the Prep Station)
• ScanProcessing – DA Scan Processing (currently processing on the Digital Analyzer)
• ScanError – DA Scan Error (an error occurred during processing on the Digital Analyzer)
• ScanAbort – DA Scan Abort (manual abort on the Digital Analyzer)
• ReportPending – Report Pending (Scanning complete, but waiting for the algorithm to complete)
• ReportProcessing – Report Processing (Algorithm is running but the report has not been produced yet)
• ReportComplete – Report Complete (Digital Analyzer has finished scanning and reports are available for download from the
Reports page)
• ReportError – Report Failed (All processing for the run set is complete but a report was not generated due to an error in the
algorithm)
• ReportCompleteWithError – Report Complete with Error (All processing for the run set is complete, and a repor t generated, but
the assay failed)
4. Last Update – The date that the sample last changed status
5. Created By – The User ID that created the Run Set using the web application
6. Lane – The lane where the sample resides on the Car tridge
7. M e m o (optional) – Notes about the sample entered in the Memo field of the Run Set form
8. Prep Station – Name of the Prep Station on which the sample was processed, useful if more than one Prep Station is connected to
the Digital Analyzer
2017-07 MAN-C0034-06
|
33
USER MANUAL nCounter® Dx Analysis System
D. Run Sets
This section guides users through creating, editing, printing, and deleting Run Sets.
Create a Run Set
Users are required to create a Run Set associating the sample IDs with their strip tube well locations using the nCounter web application.
A user can choose to create a new Run Set from several places within the web application. The most common way is to select the option
Runs from the menu bar at the top of the page and choose Create New Run Set from the sub-menu that appears below (FIGURE 4.19).
FIGURE 4.19: The Create New Run Set option is available in the Runs sub-menu
>>> The Create New Run Set form will appear (FIGURE 4.20).
FIGURE 4.20: The Create New Run Set form
|
34
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
Enter the following information to create a Run Set:
1. nCounter Assay Type – If the user has access to more than one nCounter assay type, the assay that was previously selected during
sign in will be displayed (FIGURE 4.2). To change this selection, the user must log out, then sign in again and select a different assay
type.
2. Run Set ID – The Run Set ID must provide a unique name to identif y the Run Set.
3. Test Configuration Code – The Test Configuration Code is an alphanumeric barcode located inside the CodeSet box. It determines the
number of samples that can be processed.
4. CodeSet Kit Number – The CodeSet Kit Number is a numeric barcode located inside the CodeSet box and is also described as a
CodeSet Barcode Sticker. It determines the CodeSet expiration date. Because there may be a delay between creating a Run Set and
processing the samples, a warning will be provided if the CodeSet is within two weeks of its expiration date when the Run Set is
created.
5. Sample Data – The Reference Sample is always located in wells 1 and 2; they do not require any input and cannot be edited. Well
numbers 3-12 are used for patient RNA samples. Fields may vary for different assay types. As an example, Prosigna® requires the
following information:
a. Sample ID Label (required) – Sample IDs can be entered using barcoded sample tubes and a barcode scanner connected to the
computer. If a scanner is not available or if the barcodes are damaged, sample IDs can be manually entered using a keyboard.
NanoString recommends using unique sample IDs to track samples.
b. # of Positive Nodes (re quired) – Users can cho ose between zero positive n odes, 1-3 positive nodes or >= 4 positive no des (where
available).
c. Tumor Size (required) – Users can choose between <= 2 cm or > 2 cm.
d. Memo (optional) – Any notes about the sample can be entered here. There is a maximum limit of 32 characters.
NOTE: Leave the remaining fields blank if any st rip tube wells are not needed. If additional fields are required for more samples,
use a different test configuration that accommodates more samples.
IMPO RTANT: Some handheld scanners can misinterpret barcodes if they are not properly configured. It is essential that
the Test Configuration Code and CodeSet Kit Number be entered correctly. If you encounter errors, contact dxsupport@
nanostring.com for assistance.
6. Set E-mail Recipients – If desired, choose e-mail recipients by selecting users from the contact list on the lef t side and clicking the
Add>> button. Conversely, e-mail recipients can be removed by selecting the users from the list on the right side and clicking the
<<Remove button (FIGURE 4.21). Press the Ctrl key (or the Command key if using an Apple computer) on t he keyboard while selecting
multiple addresses to add or remove multiple recipients at once.
a. E-mail Status Updates to – Users assigned here will receive e-mail updates anytime the Run Set status has changed.
b. E-mail Report Notifications to – Users assigned here will receive an e-mail when the sample reports for the Run Set are
available for download. E-mails will include a link to log in and download the appropriate nCounter reports.
NOTE: A user must have proper permissions set in their user profile in order to be able to receive e-mail report notifications.
Attempts to assign a user without appropriate access to receive report notifications will result in an error message. To change
these permissions, an administrator must edit the user’s profile and add the privilege.
2017-07 MAN-C0034-06
|
35
USER MANUAL nCounter® Dx Analysis System
FIGURE 4.21: The location of the contact list and e-mail alerts on the Create New Run Set form
It is crucial to correctly enter sample information. Complete each well with all required information before entering sample information for
the next well. Click on Save Run Set when you have entered all information for the Run Set. After you save the Run Set, a prompt to print a
worksheet will appear (FIGURE 4.22).
FIGURE 4.22: The Print Worksheet prompt
Select OK to print the Run Set Worksheet.
NOTE: Clicking the Cancel button will navigate the user to the Run Sets page.
NOTE: The Run Set will still be saved in the application if a worksheet is not printed, but it is recommended that a printed
worksheet be used in the laborator y when preparing samples. The worksheet may also be printed at a later date.
|
36
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
>>>The worksheet will appear in a new window (FIGURE 4.23).
FIGURE 4.23: An example worksheet
Select Print to print the worksheet, and then close the worksheet. The application will return to the Run Sets page.
2017-07 MAN-C0034-06
|
37
USER MANUAL nCounter® Dx Analysis System
Edit a Run Set
Run Sets with the status ‘BatchRegistered’ can be edited. Users should edit a Run Set to match the final hybridization record if sample
information was altered. For example, the Sample ID Labels may be updated if patient samples are swapped out or if a new sample ID is
assigned.
To edit a Run Set, select the appropriate Run Set from the list on the Run Sets page and click the Edit button (FIGURE 4.24).
FIGURE 4.24: Location of the Edit button on the Run Sets page
>>> The Edit Run Set page will appear (FIGURE 4.25).
FIGURE 4.25: The Edit Run Set page
NOTE: Clicking on the Run Set ID hyperlink will NOT open the Run Set for editing. Clicking the hyperlink will open the Samples
status page, pre-filtered to view only that Run Set.
Verify that the correct Run Set has been opened, and then edit the Run Set as needed. Save changes by clicking the Save Run Set button at
the top. When prompted, print the worksheet as recommended earlier in this section.
|
38
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
Delete a Run Set
Run Sets with the status ‘BatchRegistered’ can be deleted if necessary. To delete a Run Set , select the appropriate Run Set from the list and
click the Delete button (FIGURE 4.26).
FIGURE 4.26: Location of the Delete button on the Run Sets page
>>> A prompt will appear to confirm deletion of the Run Set (FIGURE 4.27).
FIGURE 4.27: The ‘Delete runset?’ prompt
Click OK to delete the Run Set or Cancel to return to the Run Sets page.
Editing Sample Information
Occasionally, a report may need to be generated again if one or more sample parameters (e.g., for Prosigna®, # of Positive Nodes (Nodal
Status) or Tumor Size) were entered incorrectly. These parameters may be changed in the Create/Edit Run Set page before the sample is
started on the Prep Station, but after the sample has started on the Prep Station, only an administrator can edit these fields and generate
a new report. This can only be done once per sample. The new report will be marked as a revised report, and for reference, will include the
obsolete parameters and results. In addition, if the Prep Station or Digital Analyzer has been initiated prior to realizing the parameters were
entered incorrectly, do not abort the run, but allow it to fully complete, and then edit the sample parameters to revise the run.
IMPO RTANT: The user must have administrator privileges to edit an analyzed sample and re-run the report. The report can
only be re-run once.
See the Administrator section for details on Editing Sample Information.
2017-07 MAN-C0034-06
|
39
USER MANUAL nCounter® Dx Analysis System
E. Print Worksheets
It is highly recommended that wor ksheets be used when setting up hybridization reactions. As des cribed earlier, when a Run Set is saved t he
system will prompt the user to print the worksheet. However, it is also possible to print Run Set worksheets at any time from the nCounter
web application.
To print a Run Set worksheet, navigate to the Run Set’ page, select the Run Set row of interest, and click the Print Worksheet button
(FIGURE 4.28).
FIGURE 4.28: Location of the Print Worksheet button on the Run Sets page
>>> The worksheet (FIGURE 4.29) will appear in a new window.
FIGURE 4.29: An example worksheet
Click the Print button to begin the printing process for the worksheet or the Close but ton to close the window.
|
40
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
F. Reports
Users with the ‘Access Diagnostic Reports’ privilege can select and download diagnostic reports from the Reports page.
Users chosen to receive a report notification when setting up the Run Set should receive an e-mail stating that the sample processing is
complete and that the report is available for download. A link to download reports from the web application is included within the e-mail
notification.
NOTE: The download link requires the user to log in to the nCounter web application before diagnostic reports can
be accessed.
From the Reports page, select the row(s) of samples to be downloaded and click the Download button (FIGURE 4.30).
• If more than one language has been installed for an assay, the Report Languages menu will be displayed in the upper left corner of
the Report screen. The user will need to select desired language(s) for downloaded reports before clicking on the Download button.
• Language selections will be retained until the user modifies the selection.
(Note: Language options are assay-specific and are installed on a per-assay basis.)
FIGURE 4.30: The Reports page
Downloaded reports will be compressed into a ZIP file (*.zip). This file can be opened immediately or saved to a specified location on a
computer or network.
NOTE: Remember that reports contain confidential information. Be careful when saving to a network or location that may
be accessed by others.
2017-07 MAN-C0034-06
|
41
USER MANUAL nCounter® Dx Analysis System
G. Administration
For Administrators, the Admin button is available in the top menu bar of the web application. Administrators can choose these functions:
• Manage Users
• System Settings
• Add User
• Configure Report
In addition to the functions listed in the Admin menu, two additional areas are restricted to Administrators:
• LIS Integration
• Editing Sample Information
|
42
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
Add, Delete, and Manage Users
To add a new user, select Admin from the top menu bar and click on Add User (FIGURE 4.31).
FIGURE 4.31: The location of the Add User button in the Admin sub-menu
>>> The Add New User form will appear (FIGURE 4.32).
FIGURE 4.32: The Add New User form
2017-07 MAN-C0034-06
|
43
USER MANUAL nCounter® Dx Analysis System
Enter the following information and click the Save button.
• User ID (required)
• Password (required)
• Confirm Password (required)
• E-mail Address (required)
• User Type (required) – choose either:
• General – Access to standard functionality including creating a Run Set (access does not include processing samples)
• Administrator – General user access plus administrative access including user management and system settings
• Permissions (at least one permission is required) – choose all that apply:
• Run Post Hybridization – This permission is required to operate the Prep Station and initiate the post-hybridization process for
the assay(s) selected
• Access Reports – This permission is required to view the Repor ts tab on the Web Application and download diagnostic reports
for the assay(s) selected
• Create Run Set – This permission is required to create a new Run Set or edit an existing Run Set on the Web Application for the
assay(s) selected
NOTE: Administrators do not automatically have permission to access diagnostic reports and initiate post-hybridization
processing on the Prep Station; permission must still be assigned. Administrators are able to assign permissions to their own
accounts.
NOTE: Unique permission settings can be assigned for each assay type available on the system. For example, a user may be
able to create a Run Set for all assays on their system, initiate post-hybridization for a single assay, and be unable to access
any reports.
The Manage Users function allows an administrator to edit or delete users. To manage existing users, select the Admin option from the
menu bar at the top of the page and click on Manage Users.
>>> The Manage Users screen will appear (FIGURE 4.33).
FIGURE 4.33: The Manage Users page and the location of the Edit button
|
44
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
To edit a user, select the User ID link or select the row for the specified user and click the Edit button.
>>> The Edit User Data form will appear (FIGURE 4.34).
FIGURE 4.34: The Edit User Data form
Edit any of the following information and click the Save button to save any changes made.
• User ID
• Password
• Confirm Password
• E-mail Address
• User Type
• Assay Permissions
• Run Post Hybridization
• Access Reports
• Create Run Set
2017-07 MAN-C0034-06
|
45
USER MANUAL nCounter® Dx Analysis System
To delete a user, open the Manage Users page, select the row for the appropriate user, and click the Delete button (FIGURE 4.35).
FIGURE 4.35: Location of the Delete button on the Manage Users page
>>> A confirmation message will appear (FIGURE 4.36).
FIGURE 4.36: The ‘Delete User?’ prompt
Click OK to delete the user or click Cancel to return to the Manage Users page.
|
46
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
System Settings
System settings are available to users with the Administrative user type. The System Settings page supports four main functions, which
are displayed in separate tabs:
• Date and Time (default tab)
• E-mail Settings
• SSH Settings
• IP Address
To access the System Settings page, select the Admin option in the top menu bar and select the System Settings option (FIGURE 4.37).
FIGURE 4.37: Location of the System Settings option in the Admin sub-menu
>>> The System Settings page will appear, with the default Date and Time tab visible (FIGURE 4.38).
FIGURE 4.38: The Date and Time tab on the System Settings page
2017-07 MAN-C0034-06
|
47
USER MANUAL nCounter® Dx Analysis System
Date and Time
Date and time settings are initially set when NanoString personnel set up the nCounter. Change these settings only when necessary. Click
Save to save any changes or Cancel to discard changes and perform another function.
NOTE: Administrators cannot adjust the Time Zone or Daylight Saving Time options. Contact NanoString Support if these
values need to be adjusted.
NOTE: Changing the date and/or time in the nCounter web application adjusts these values on the Digital Analyzer and Prep
Station, as well. Take care when making changes to these settings.
|
48
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
E-mail Settings
E-mail settings (FIGURE 4.39) must be configured in order for the nCounter system to successfully send automated e-mails for status
updates such as Prep Station run completion or diagnostic report notifications.
FIGURE 4.39: The E-mail Settings tab
The following fields are available:
• SMTP Server (required) – (Simple Mail Transfer Protocol) – E-mail server name
• SMTP Port (required) – Port used for e-mail server
• SMTP Address (required) – Address from which automated e-mails are sent
• Enable SSL – Check if using a secure e-mail server
• SMTP Username – Account name for secure e-mail server (required if Enable SSL is checked)
• Password – Password for SMTP username (required if Enable SSL is checked)
• Confirm Password – Same password as above (required if Enable SSL is checked)
• Encryption Protocol – These options ef fectively encrypt e-mail data sent over the Internet between the client and e-mail server. The
following are possible values for encryption protocol (a selection is required if Enable SSL is checked):
• SSL3 (Secure Socket Layer) – Widely used encryption protocol
• TLS (Transport Layer Security) – More secure encryption protocol than SSL
• Test Address (optional) – This can be used to test if the e-mail set tings are configured correctly. Enter an e-mail address you have
access to and press the Test button. Check your e-mail account to confirm that the settings are working and the test e-mail was
received.
2017-07 MAN-C0034-06
|
49
USER MANUAL nCounter® Dx Analysis System
SSH Settings
The SSH Settings tab (FIGURE 4.40) allows an Administrator to configure the Secure Shell server (SSH server). The SSH server allows users
to programmatically pull nCounter report data or system backup data from the system in a secure manner. It also allows for unlimited
concurrent client connections. See the section “LIS Integration” below for more information.
FIGURE 4.40: The SSH Settings tab
The following fields are available:
Pre-configured fields:
• SSH Host Name – The name of the host used by the client for connecting to the SSH server
• SSH Port Number – The port used by the client for connection to the SSH server
• SSH User Name – SSH server account name used by the client for authenticating the connection
Configurable fields:
• Enable SSH Server – Check box for enabling or disabling the SSH server.
• Assay / Backup – Selection of assay report or system backup data to be accessed by the SSH server. A separate SSH user name
and password is required for each assay or system data.
• Password – SSH server account password used by the client for authenticating the connection (required when ‘Enable SSH
Server’ is checked)
• Confirm Password – Same password entered as above (required when enable SSH server is checked)
|
50
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
Once the Enable SSH Server checkbox is enabled, the Administrator can select which type of assay or backup data to configure for SSH
access (FIGURE 4.41).
FIGURE 4.41: Assay / Backup menu is active after the Enable SSH Server checkbox is selected
2017-07 MAN-C0034-06
|
51
USER MANUAL nCounter® Dx Analysis System
IP Address
IP address function (FIGURE 4.42) is an essential function to enable networking of the instruments. The Prep Station, Digital Analyzer, and
nCount er web applicat ion are all in fre quent communic ation. The IP a ddress is the ba sis for sett ing up this communicat ion. Avoid errors when
making changes to fields under the IP Address tab.
FIGURE 4.42: The IP Address tab
The following fields are available:
• Type – The following options are available to choose between:
• Static IP Address – This requires the network administrator to set an address that will never change.
• Dynamic IP Address (default) – Known as DHCP, this option allows the local network to automatically find and assign an IP
address. This address is subject to change.
• IP Address – IP address provided by the network administrator (for Static IP address)
• Subnet – Subnet provided by the network administrator (for Static IP address)
• Gateway – Gateway address provided by the network administrator (for Static IP address)
• DNS 1 – Domain Name Server provided by the net work administrator (for Static IP address)
• DNS 2 – A second DNS is optionally provided by the network administrator and used as a backup (for Static IP address)
NOTE: A Static IP Address is particularly useful because it ensures more reliable communication between the Digital Analyzer
and the Prep Station. If the Dynamic IP Address option is chosen, the Digital Analyzer may sometimes acquire a different IP
address, giving rise to the possibility of intermittent communication issues.
|
52
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
Configure Report
In many cases, organizations have common report formats standardized across all diagnostic reports produced by the organization. The
system allows administrators to customize the header and footer of the report, and as an option, allows a customized signature line on the
last page of the repor t. Administrators can define static text and include certain system-provided information about the sample. Sample
parameters which are on the NanoString standard diagnostic report header may be included in custom headers and footers. Additionally,
administrators may define fields which will be completed in the output PDF after download from the nCounter Dx Analysis System. These
fields may be used for patient name, date of birth, or any other Personally Identifiable Information (PII) desired. (Users will need to save the
completed forms outside the nCounter system as it is not currently engineered to suppor t storage of PII). The report body, including assay
results, logos, and explanator y graphs and text, may not be modified.
Each ins talled loca lized assay is custom ized independently. The administrator may only make changes to the a ssay selected at log in, thoug h
the administrator may customize any report locale installed for that assay. Changes are first made to a draft report; then, when the draft
meets the organization’s needs, t he administrator ca n move the draft to production. Administrators are also able to create a new draft from
the current production report, from a number of system-provided templates, or with blank header, footer, and signature area.
To access the Configure Repor t page, select the Admin option on the top menu bar and select Configure Report (FIGURE 4.43).
FIGURE 4.43: Location of the Configure Report option in the Admin menu
2017-07 MAN-C0034-06
|
53
USER MANUAL nCounter® Dx Analysis System
The Configure Report page will then appear, open to the default General tab (FIGURE 4.44).
FIGURE 4.44: The General tab of the Configure Report page (Content of report varies by regulatory clearances or approvals)
Within the Configure Report area, the following tabs are available:
• General
• Header
• Footer
• Signature
The Header, Footer, and Signature tabs allow the user to customize individual parts of the report.
The General tab has a Preview button (common across all tabs), and a series of buttons to choose and control the production and draft
reports:
• Discard Production Report – Selecting this button will remove all user-generated customization of the report and will revert to the
• Move to Production – Move the current draft report to production. This erases the current production report and cannot be undone.
• Revert Draft – Revert all parts of the draft report (header, footer, signature if configured) to their state at the start of this editing
session.
• Discard Draft – Remove all customization of the header, footer, and signature.
• Preview – Identical functionality to the “Preview” at the top of the page; displays a full-page PDF preview of the current Draft report.
• Edit – Opens the header editor to allow administrative users to customize their reports
|
54
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
If ther e is no working dr aft in progr ess (See FIGURE 4.45), then a dr opdown list ing of available t emplates is dis played inste ad of the thumb nail
of the draft report, along with a control to start a new draf t. Drafts may be started from the current production report, the NanoString
standard header and footer, blank header and footer, or one of a number of pre-supplied templates. These templates are intended to be
further modified by administrators to meet their organizations’ needs.
FIGURE 4.45: The General tab of the Configure Report page, displaying the menu for starting a new draft (Content of report varies
by regulatory clearances or approvals)
2017-07 MAN-C0034-06
|
55
USER MANUAL nCounter® Dx Analysis System
Header Tab
The Header tab allows administrators to modify the header for the report (see FIGURE 4.46).
FIGURE 4.46: The Header tab of the Configure Report page, displaying the NanoString standard header
The main body of the Header tab is a text entry area for specifying the content and layout of the customized report header. There is a
vertical ruler to assist in placement of content items. The red line represents the maximum size of the header—any content below this line
will be hidden by the body of the report and will not be visible. The body of the report will begin immediately following the header; unless
the customized header explicitly includes blank lines, there will be no visible gap between the user-generated header content and the
NanoString-supplied body. The Preview button may be used at any time to generate a PDF with the currently entered header fields.
Across the top of the text ent ry area, t he user will find s tandard text e diting and for matting tools (FIGURE 4.47) that allow te xt to be inser ted
and formatted at both character and paragraph level, import graphic images, use tables to more precisely position content, and insert
special fields
FIGURE 4.47: Text editing and formatting tools
|
56
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
When selected, the icon displays a menu (FIGURE 4.48) of special fields which can be inserted, formatted, and positioned as needed on
the report.
FIGURE 4.48: The Special Fields menu, displaying example assay-specific fields
These fields are:
• PDF Field – In the output PDF, these are blank fields which the user can complete using standard PDF viewers (Adobe Reader,
Apple OS X Preview, etc.). These fields can be used to enter any desired information about the sample, including HIPAA-protected
Personally Identifiable Information (PII) such as patient name or date of birth. The edited PDF will need to be downloaded and saved
to preserve such user-entered content.
• Run Set ID – The Run Set ID entered on the Create New Run Set page.
• Sample ID – The Sample ID entered on the Create New Run Set page.
• Run Date – The date that the sample was run on the DA. This date is automatically recorded by the nCounter Dx Analysis System.
• Comments – The Memo entered with the sample on the Create New Run Set page.
• Remaining fields are assay-specific parameters (e.g., Tumor Size and Node Count). Your specific assay may or may not have user-
entered parameters.
NOTE: After initial placement, images and tables may be further customized or refined by modifying cer tain properties. These
proper ties are accessed by right-clicking (control-clicking on systems without a right mouse button) on the object, then
selecting the appropriate items from the presented menu.
At the bot tom of the text entry area are a pair of tabs, Design and HTML, (FIGURE 4.49) that allow advanced users to switch between the
default interactive design mode and directly editing the underlying HTML. The HTML editor may be used by advanced users to fine-tune
layouts or apply affects not directly available through the interactive design mode.
FIGURE 4.49: Design and HTML tabs at the bottom of the text entry area
The Preview button at the bottom of the page functions identically to the Preview button at the top of the page. The Next button provides
an easy way to move from tab to tab when initially entering a customized report; additionally, users may directly select the tab they desire
to edit.
2017-07 MAN-C0034-06
|
57
USER MANUAL nCounter® Dx Analysis System
Foote r Tab
The Footer tab (FIGURE 4.50), allows administrators to modify the footer for the report. The footer is always placed at the bottom of each
page, leaving any empty space between the end of the body and the start of the footer. The formatting controls available for the header are
available for use in the footer area as well.
FIGURE 4.50: An example of the standard NanoString footer
Signature Tab
The Signature tab allows an administrator to optionally include a signature line on the last page of the report (FIGURE 4.51). When the
checkbox has been enabled, the signature line can be edited and will be included on generated PDF files. The editing tools available for the
header are also available for the signature.
FIGURE 4.51: The Signature tab with checkbox enabled
|
58
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
Header Configuration Example
To furthe r illustrat e how the heade r can be config ured, consid er the case o f a laborator y with a st andard hea der consist ing of the orga nization
logo; patient family name, given name, birth date, and gender; and assay-specific fields. This header example has been customized for
a specific assay (FIGURE 4.52), and demonstrates how to combine and nest multiple elements and adjust their individual configuration.
System-provided fields are used where available, and PDF fields are used for the PII in this example.
FIGURE 4.52: Example header
The specific steps to create this header are as follows:
1. Open the report configuration tool with Configure Report in the Admin menu.
2. If there is no draft in progress, skip to step 4.
3. If there is a current draft in progress, select the Discard Draft button at the bottom right of the pa ge. This will delete the current draft
and does not have an “undo” function. Select OK to confirm your intent to delete the draft.
4. Create a custom report. For this example, choose the Blank template, then select Go.
5. When the new draft is created, select the Header tab to edit the header.
6. Click in the content area to start editing.
7. Select the table tool to create a new table.
8. In the “Insert Table…” dialog, choose the following, then select OK when all parameters are entered:
a. 5 columns
b. 5 rows
c. Width: Custom, 90%
d. Border color: 4th entry in the 3rd row, “#339966”
e. Border size: 2
9. Make sure the cursor is in the upper left cell of the newly created table.
10. Select the image tool to import an image.
11. In the “Insert Image” dialog, choose the following, then select Insert when all parameters are entered:
a. For the source, choose “From your computer”
b. Select “Browse…”, then browse to the image you wish to use for your logo.
12. After the image has been inserted, right click and select Change Image…. In the “Change Image” dialog which is opened, check the
More options box to enable fine-tuning of the image display parameters.
a. Set Size: to “Custom Size”.
b. Set the width and height such that the width is less than 250 pixels and the height is less than 300 pixels. The exact values will
depend on the relative height and width of the image you have chosen.
c. Set the Position: to “Left-aligned”.
d. Select Change to apply these changes.
2017-07 MAN-C0034-06
|
59
USER MANUAL nCounter® Dx Analysis System
13. Right click on the logo, then select Merge Down from the presented menu. This will merge the top 2 cells of the leftmost column of
the table. Repeat until there is only 1 cell in the leftmost column of the table.
14. Enter the address or any additional tex t below the image.
15. Click in the topmost cell of the second column.
16. Enter the desired text label, “Family Name” and select icon to right-align the text.
17. Click in the second cell of the second column, then enter “DOB” and right-align.
18. Click in the third cell of the second column, then enter “Run Set ID” and right-align.
19. Click in the fourth cell of the second column, then enter “Run Date” and right-align.
20. Click the last cell of the second column, then enter “Comments” and right align.
21. Click in the first cell of the third column, then select Add PDF Field from the + menu.
22. Click in the second cell of the third column, then select Add PDF Field from the + menu.
23. Click in the third cell of the third column, then select Add Run Set ID Field from the + menu.
24. Click in the fourth cell of the third column, then select Add Run Date Field from the + menu.
25. Click in the last cell of the third column, then select Add Comments from the + menu.
26. Right click in the last cell of the third column, then select Merge Right. Repeat so that the cell spans the remainder of the table.
27. Click in the first cell of the fourth column, then enter “Given Name” and right-align.
28. Click in the second cell of the fourth column, then enter “Gender” and right-align.
29. Click in the third cell of the fourth column, then enter “Node Status” and right-align.
30. Click in the four th cell of the fourth column, then enter “Tumor Size” and right-align.
31. Click in the first cell of the last column, then select Add PDF Field from the + menu.
32. Click in the second cell of the last column, then select Add PDF Field from the + menu.
33. Click in the third cell of the last column, then select Add Number of Positive Nodes Field from the + menu.
34. Click in the fourth cell of the last column, then select Add Tumor Size Field from the + menu.
35. Select the Preview button at the top or bottom of the window to view a sample of the configured report.
|
60
2017-07 MAN-C0034-06
NanoString Technologies® USER MANUAL
Report Integration with External Systems
The nCo unter Dx Anal ysis System sup ports direct int egration w ith existing repor t generati on and management sy stems, such a s Laborat ory
Information Systems (LIS). Using SFTP for file transfers, external automated processes may retrieve the output report (either in whole or in
part), as well as the system backup archive.
The retrieved repor ts are in formats suitable for integration with existing processes and systems. For downstream integration, the system
makes the report available as a PDF file identical to that which can be downloaded through the web GUI, as a PDF of the NanoString default
repor t (if different from the current report), and as a modular set of graphic files which can be incorporated into downstream system output.
System Access
For LIS integration, the SSH Server must be enabled (see System Settings). The username and password specified in SSH Settings for the
assay can be used with any file transfer tool which supports SFTP (SSH File Transfer Protocol). This protocol is widely supported by such
tools as WinSCP and PuTT Y PSFTP on Windows and Transmit on Mac OS X. Access to the SSH Server is read-only: no files may be modified,
uploaded, or moved.
Assay-specific files: After gaining access to the SSH server, the assay-specific files are arranged as follows:
Assay directoryDescription and contents
/(login directory)
/pdf
/image
/image/<basename>
/deprecated
Top-level director y, contains XML export files and all subdirectories listed below
PDF files. For each completed sample, the NanoString standard report will always be present; if a
customized report has been defined, the PDF of that repor t will also be in this director y.
Organizing directory for modular images.
One director y for each sample, named as described below. Each directory contains a series of images
which compose the full assay report and an HTML file to assemble the images in the original order. See
below for details.
All file s which have been deprecated by using the Edit an Analyzed Sample function to generate a re-run
of the report algorithm or other update. These are the original, unchanged files.
Modular Images
The /image/<basename> directories contain a series of high-resolution (300 dpi+) images which may be used by advanced organizations.
Header, footer, and signature images may be modified to include organization-specific information, or may be completely replaced by
organization-provided equivalents, or may be omitted entirely. Body images may not be modified and should be used complete and as-is
only.
The supplied HTML file serves as an “inventory” and may be used to ensure that downstream systems have retrieved body images and/or
headers and footers for ever y page of the report. It can serve as a template for further modifications.
2017-07 MAN-C0034-06
|
61
USER MANUAL nCounter® Dx Analysis System
Filenames
All formats share a common basename format:
<YYMMDD>_<cartridgeID>_<scan>_<sampleName>_<Lane>
Where:
<YYMMDD>
<cartridgeID>
<scan>
<sampleName>
<Lane>
The filename extensions indicate the file format:
ExtensionContents
.xml
.pdf
.png
.html
.zip, .zXX
Image filenames have an additional component appended after the <basename> portion. This component serves to identify the precise
contents of the file. Elements listed in square brackets (“[ ]”) are optional and may not exist for all reports. The additional elements for image
files are:
ComponentContents
_head
_body#
_foot
[_sig]
[_h ead 2]
[_f oot 2]
Last 2 digits of the year, 2-digit (numeric) month, 2-digit numeric day
Barcode found on the Cartridge
Assigned by the system. Typically defaults to 1.
SampleID as entered in the Create New Run Set or Edit Run Set pages of the