Nanolive 3D Cell Explorer-fluo User Manual

3D Cell Explorer-uo
User Manual
1. Safety Notes 3
3D Cell Explorer-uo 3
CoolLED illumination system 5
2. Maintenance & Care 6
3D Cell Explorer-uo 6
CoolLED’s illumination system 7
3. Introduction 8
Instrument Description & Main Features 8
Typical Field of Applications 9
3D Cell Explorer-uo terminology 10
CoolLED illumination system terminology 11
5. Computer Technical Specifications 13
6. Start-Up 14
7. Handling 29
8. Sample Preparation 34
9. Cleaning procedure for optical elements 35
10. Software STEVE 37
11. Troubleshooting 58
12. Compliance 60
Contents
3
à go back to contents
1. Safety Notes for the 3D Cell Explorer-uo
• Verify that the computer is plugged and grounded earthed with its own power supply during usage with your 3D Cell Explorer-uo.
• Switch the 3D Cell Explorer-uo OFF before connecting it to a power supply.
• Disconnect and lockout the power supplies of the 3D Cell Explorer-uo before completing any maintenance work tasks or making adjustments.
• Do not use electrical equipment in wet conditions or damp locations.
Do not clean tools with ammable or toxic solvents.
2. Supply the 3D Cell Explorer-uo only with the provided power
supplies.
3. No special laser safety required when operating the 3D Cell
Explorer-uo (Class I laser).
For measurements, the 3D Cell Explorer-uo uses a Class I laser. A Class I laser is harmless for your eyes under any circumstances and thus does not require laser safety goggles when properly operated.
By no means should the user open the covers and thus have access
to the laser path within the microscope frame without a clear
authorization from Nanolive.
Operating the microscope in non-appropriate conditions may then be a risk. It is thus forbidden for a non-authorized user to open the
covers and manipulate the laser.
Thank you for purchasing the Nanolive 3D Cell Explorer-uo. Before using your 3D Cell Explorer-uo and the CoolLED illumination system, carefully read the safety notes to ensure safe handling and usage of the device at all times. Failure to do so may result in personal injury or damage to other
items.
1. Safety Notes
3D Cell Explorer-uo
4
à go back to contents
4. Contaminations
The user has the responsibility to avoid disastrous contaminations by biological samples. Nevertheless, contamination might occur
when working with biological samples. Please ensure appropriate cleaning of the 3D Cell Explorer-uo. Please refer to our Main­tenance and Care notes for proper cleaning instructions.
5. Cleaning
Please note that alcohol is highly ammable, do keep it away from re or potential sources of electrical sparks, and use it in a well- ventilated room
when using it for cleaning the 3D Cell Explorer-uo or other measures.
6. The 3D Cell Explorer-uo conforms to the requirements of
the Safety Standards as follows:
• EN/IEC 61010-1:2010 Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory use.
• EN62471:2008 Photo-biological Safety of Lamps and Lamp Systems/Guidance on manufacturing requirements relating to non-laser optical radiation safety. No risk
1. Safety Notes
3D Cell Explorer-uo
5
à go back to contents
7. Safety Notes for the CoolLED illumination system
• UV light may be emitted from this product depending on the
conguration selected. Avoid eye and skin exposure. Never look
directly into the light output beam from the Light Source or accessories. The emissions could damage the cornea and retina
of the eye if the light is observed directly.
• Always ensure that the Light Source is securely attached to the microscope with its light guide, prior to turning on the power. This will minimize the risk of injury and damage.
• If for any reason the Light Source is to be operated when not attached to a microscope, all personnel should wear eye shielding and clothing to protect the exposed skin.
Disconnecting the main supply is achieved by unplugging the power cord from the power supply block or the Light Source. Only plug in the power cable, once the Light Source is attached to the microscope.
There are no serviceable parts within the Light Source.
Removing any of the screws and covers will result in the safety
of the Light Source being impaired. The DC power supply unit should be inspected periodically throughout the lifetime of the system.
• To clean the exterior of the Light Source, use a slightly dampened
cloth with a simple water/detergent solution only. Avoid the
optical surfaces and lenses. Cleaning of optics should only be carried out using the provided optical swabs with grade isopro­panol or ethanol. Please note that the DC power supply unit should be isolated prior to cleaning.
• Safety precautions must always be exercised when using
electrical equipment to prevent operator shock, re hazard or
equipment damage.
FIRE HAZARD: Do not drape or cover the light source or the light
guide while it is in use. Do not place the light guide on a drape while it is in use.
8. Compliance:
The light source conforms to the requirements of the Safety S tandards as follows:
• EN/IEC 61010-1:2010 Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory use.
• EN62471:2008 Photo-biological Safety of Lamps and Lamp Systems/Guidance on manufacturing requirements relating to non-laser optical radiation safety. Risk Group 3.
1. Safety Notes
CoolLED illumination system
All warnings may not be applicable depending on the version/wavelength being used.
6
à go back to contents
Maintenance and Care for the 3D Cell Explorer-uo
1. Carefully, open the shipment box with the top face up to avoid damaging the 3D Cell Explorer-uo.
2. Handle the 3D Cell Explorer-uo from the side or the bottom plate; unplug all cables before moving it.
3. When moving the microscope, carefully carry it without lifting it from the camera and avoid any shocks.
4. Keep the instrument out of direct sunlight, high temperature or humidity, dusty and easy shaking environments.
Make sure the table surface is at, horizontal and rm.
5. All elements (e.g. microscope objective, mirrors, lenses) have been specially and carefully adjusted; please do not dismount, reposition, or modify them. In such case the warranty becomes automatically void and rework
shall be charged to the customer.
6. Do not block the rotating arm or prevent it to rotate freely.
7. Do not disassemble any parts of the microscope, as this affects the function or reduces the performance of the
microscope. Please be aware that the warranty expires if you remove the covers.
8. Keep the instrument clean, and do not contaminate the optical elements when wiping away the dust on the
instrument. It is highly recommended to leave the red cap in place on the microscope objective when the microscope is not in use in order to prevent contaminations to come on the optical surfaces.
9. Contaminations on the objective and the mirrors, like ngerprints and oil smudges, could be gently wiped with the provided cleaning swab and a bit of alcohol (Ethanol or Isopropanol). More information on the cleaning
procedure to be found in section “Cleaning Procedure for the optical elements”.
10. Do not attempt to use organic solvents to clean the microscope. To clean it, use a lint-free, soft cloth slightly moistened with clear water.
11. During use, if the microscope is splashed by liquid, cut off the power at once, and wipe away the splash.
12. Place the instrument in a cool, dry position. When not using the microscope or carrying it, keep it covered with
the provided dust cover. Furthermore, protect the microscope objective with the red vinyl cover.
2. Maintenance & Care
3D Cell Explorer-uo
7
à go back to contents
Maintenance & Care for the CoolLED illumination system
1. Carefully unpack the components from the shipping cartons.
2. The CoolLED illumination system requires little or no maintenance throughout its life. There are no eld service-
able parts so there is no need to remove the covers.
3. Cleaning of the external surfaces can be carried out with a mild soap and water solution used to lightly dampen a
lint-free cloth. Ensure that no liquid is allowed to enter the product through vents and panel edges. Avoid optical surfaces.
4. Cleaning of optical surfaces maybe necessary if debris or nger prints accidentally come into contact with the excitation lters during installation. In the rst instance remove any loose debris with an air duster (aerosol or rubber blower).
5. Finger prints or other liquid type contaminants should be removed using standard lens cleaning procedures. Do not ood the optical surfaces with uid as liquid could enter the product and cause damage.
Care of the light guide
The light guide is a ne, high quality optical instrument. The useful life can be prolonged by following these simple guidelines.
1. Avoid stretching the light guide, forming congurations, involving sharp angels or kinks, or contact with sharp or
pointed objects. The internal light bers are made of glass, a material that breaks under stress. Fiber breakage
will result in diminished light output.
2. Any inadvertent cut or puncture to the outer wall will render the light guide unsafe for use and must be taken
out of service immediately.
2. Maintenance & Care
CoolLED illumination system
8
à go back to contents
3. Introduction
Instrument description and main features
Plug & Play systems
• 2 imaging modalities:
- Physical structural imaging (based on refractive index, RI)
- Chemical uorescence imaging
3 Fluorescence Channels: 2 congurations available:
- DAPI (Excitation 392 nm / Emission 432 nm), FitC (Excitation 474 nm / Emission 515 nm), TritC (Excitation 554 nm / Emission 595 nm)
- FitC (Excitation 474 nm / Emission 515 nm), TritC (Excitation 554 nm / Emission 595 nm), Cy5 (Excitation 635 nm / Emission 730nm)
• 4 Fluorescence Channels:
- DAPI (Excitation 392 nm / Emission 432 nm), FitC (Excitation 474 nm / Emission 515 nm), TritC (Excitation 554 nm / Emission 595 nm), Cy5 ( Excitation 635 nm / Emission 730nm)
Note: TritC and Cy5 excitation lters are exchangeable on channel 3
Please refer to the following link for details on the Filter conguration:
https://www.semrock.com/SetDetails.aspx?id=3297
Likewise, refer to the following link for further data on the LED source:
http://www.coolled.com/product-detail/led-wavelengths/
• Operation modes:
- 2D epiuorescence snap-shot (every second) & time-lapse (up to 96 hours)
- 3D RI snap shot & time-lapse 3D RI & 2D epiuorescence snap-shot (every 3 seconds) & time-lapse (up to 96 hours)
Co-localisation of Fluorescence (3 channels) and Digital Stains (7 channels at single acquisition)
• Time-lapse possibility up to 96 hours
à go back to contents
9
à go back to contents à go back to contents
The typical fields of applications for the 3D Cell Explorer-uo
Merging of uorescence microscopy and 3D cell tomography
Calibration of the refractive index map with markers (Chemical dyes, uorescent antibodies, protein
labeling with GFP, etc.)
Combination of structural RI information with chemical information for specic protein/drug tracking, protein localization and mapping
long term (over a week) live cell imaging in 3D at high temporal resolution (1 image every 2 sec.) thank to our technology that generates no phototoxicity
Powerful partner for advanced image analysis strategies: high resolution and no phototoxicity allows to obtain more reliable cell measurements in subsequent image analysis
• Possibility to use 10 markers in parallel
3. Introduction
Typical Fields of Applications
10
à go back to contents à go back to contents
3. Introduction
3D Cell Explorer-uo terminology
Rotating arm
Scanning head
Peripheral mirror
Injection Head USB 2.0 slot Power switch
Power supply plug port
BNC cable plug port
Hi-grade stage for sample screening
Z knob for optical focusing
Cell container interface
XY knob for sample screening
Camera with USB
3.0 slot
Camera with USB
3.0 slot
3D Cell Explorer-uo – front view 3D Cell Explorer-uo – side view 3D Cell Explorer-uo – rear view
11
à go back to contents à go back to contents
3. Introduction
CoolLED illumination system terminology
CoolLED illumination system – front view
CoolLED illumination system – rear view
CoolLED illumination system – side view 1 Liquid Light Guide (LLG)
CoolLED illumination system – Top/Side View
Not needed BNC plugs
Excitation lters slots
Global Trigger
12
à go back to contents
4. TechnicalSpecicationSheet(TSS)
3D Cell Explorer-uo & CoolLED illumination system
à go back to contents
Specification 3D Cell Explorer-uo / Holotomography
Specifications CoolLED illumination system /
Epifluorescence
Dimensions (width x depth x height in mm) 380 x 170 x 445 77 x 186 x 162 Weight (in kg) 8 4 Environmental Operating Conditions Permissible ambient temperature, and can be
operated between 20° C and 30° C
5-35°C
Voltage ranges 100-240 VAC / 50-60 Hz / 1.0-0.5 A 110-240 VAC / 50-60 Hz / 2 A Camera USB3.0 CMOS Sony IMX174 sensor
Quantum Eciency (typical) 70 % (at 545 nm) Dark Noise (typical) 6,6 e¯ Dynamic Range (typical) 73,7 dB
Microscope Objective (MO) Dry objective / 60x magnication / NA 0.8 Illumination Source Class I laser low power (λ=520 nm,
sample exposure 0.2 mW/mm2)
(Class I. No eye protection needed)
High speed switchable <100 μs
Lifetime >20’000 hours each channel
Field-of-View (FoV) 90 x 90 x 30 μm
3
90 x 90 μm
Lateral resolution 200 nm 400 nm Axial resolution 400 nm - Tomography frame rate (acquisition) 0.5 fps 3 fps each channel Self-adjusting time < 90 seconds (Depending on sample mounting
medium thickness)
Channels Up to 7 simultaneously 3: DAPI (blue), FitC (green), TritC (orange)
3: FitC (green), TritC (orange), Cy5 (red)
4: DAPI (blue), FitC (green), TritC (orange) and
Cy5 (red) (Note: TritC and Cy5 excitation lters
are exchangeable on channel 3)
13
à go back to contents
5. ComputerTechnicalSpecications
3D Cell Explorer-uo
à go back to contents
The 3D Cell Explorer-uo is provided without a computer. Therefore, before you acquire a computer, please be aware of the minimum requirements listed below which are absolutely necessary for correct operation of the instrument.*
CPU: 1.5 GHz or faster multi-core CPU CPU Memory: minimum 8 GB DDR3 RAM Hard disk: 450 MB free space for STEVE and 1 TB for acquisition data GPU**: NVIDIA GeForce GTX 780 or later NVIDIA GTX or Quadro GPU of the Kepler or newer architectures GPU RAM: minimum 4 GB dedicated USB ports: 3 ports, minimum one USB 3.0 port
Operating System: 64-bit versions of Windows Vista, 7, 8 or 10
(Updated drivers: you need to have the latest Nvidia Graphics Card drivers, USB 3.0 drivers and fully updated Windows Operating System)
In addition: Internet access for STEVE installation and for the newest updates.
*We are happy to provide you with a list of computers and laptops that we have tested and validated. Please contact us.
**If you are unsure about the minimal requirements, please contact us for an updated list of tested GPU cards.
Please pay particular attention to the graphics card requirements. The full version of STEVE performs heavy calculations using a Nvidia GPU while the microscope is acquiring data. For this reason, please refer to the minimum required specications of the Nvidia GPU above. In addition, you
need at least one USB 3.0 port for camera image acquisition.
14
à go back to contents
6. Start-up
Components
à go back to contents
The 3D Cell Explorer-uo package contains the following:
(1) the 3D Cell Explorer-uo (2) 1 USB 2.0 cable (3) 1 USB 3.0 cable (4) 1 power supply (5) 1 power cable (6) 1 cleaning set
(7) WEEE certicate (8) protective cap for the MO and (9) protective cap for the LLG
connection to the microscope
(10) 1 BNC cable.
8
9
7
2
3
4
5
6
WEEE
Certicate
10
1
15
à go back to contents
The CoolLED Illumination System packaging contains the following:
(11) LED Light Source (12) DC Power Supply (13) IEC Power Cable (14) 1 USB 2.0 cable (15) liquid light guide (LLG) protected with red vinyl caps (use the red vinyl caps to protect both end of the LLG when not in use) (16) 3 holders equipped with excitation lters.
à go back to contents
6. Start-up
Components
11
16
12
13 14
15
16
à go back to contents
Important Information: It is critical to follow the installation and setup steps as described in the following pages to ensure the safety standards of the CoolLED illumination system. Please refer to chapter 1: Safety Notes.
1. Install the 3D Cell Explorer on a stable, at, horizontal and rm workbench.
• Lift the 3D Cell Explorer-uo carefully out of the box by holding it at the indicated spots in the image below.
Never lift the 3D Cell Explorer-uo by holding the camera.
à go back to contents
WARNING: Nanolive does not bear the damage produced by wrong handling of the 3D Cell Explorer-uo while it is removed out of its box.
In case of damages, Nanolive reserves the rights to examine the 3D Cell Explorer-uo and verify how the damage occurred. All warranty claims will be cancelled in case of
mishandling the 3D Cell Explorer-uo by ignoring the instructions. The warranty does not apply even if the 3D Cell Explorer-uo has been subject to accidental damage.
6. Start-up
Installation and setup
17
à go back to contents à go back to contents
2.   Carefully remove the protective accessories
• Properly dispose of the original packaging, or keep it for storage or return of the instrument to the manufacturer.
Remove the disposable plastic foil and red vinyl protective cap.
6. Start-up
Installation and setup
Disposable plastic foil Red vinyl cap
For safety reasons, please proceed to install the CoolLED illumination system rst.
18
à go back to contents à go back to contents
3. Install the CoolLED Illumination System on a stable, at, horizontal and rm workbench
• Lift the CoolLED Illumination System carefully out of the box together with its accessories.
6. Start-up
Installation and setup
16
12
13 14
15
19
à go back to contents à go back to contents
4. Install the excitation lters on the CoolLED Illumination System
Place the little drawers containing the excitation lters in the dedicated slots.
• Mind the order and the direction.
6. Start-up
Installation and setup
Channel 1
Channel 2
Channel 3
20
à go back to contents
6. Start-up
Installation and setup
5. Install the LLG on the CoolLED Illumination System
Remove the red cap from the LLG from one end (Figure 1)
Insert the LLG as far as it can go into the LED Light Source as shown below (Figure 2).
• Use an Allen key n°1.5 to secure the LLG in place. Once the screw is in mechanical contact with the LLG header, gently apply an extra force to make sure the LLG is maintained in place. There is a risk of damaging the LLG header and reducing the coupled power from the LED Light Source to the sample if too much force is applied by the screw.
LLG headerProtective cap
Pressure screw
Figure 1
Figure 2
21
à go back to contents
6. Start-up
Installation and setup
6. The CoolLED illumination system with the inserted LLG should look like the image below.
Place the light source close to the 3D Cell Explorer-uo in order to avoid any strain on the LLG or too short radius turns.
Ensure that the Light Source sits upright on a at surface and keep a clearance of 200mm on each side to ensure adequate airow for the
cooling system.
Clear space
200 mm
Clear space
200 mm
Figure 3
Figure 4
22
à go back to contents
6. Start-up
Installation and setup
7. Connect the LLG to the 3D Cell Explorer-uo
• Remove the red cap from the LLG from the other end (Figure 5)
• Insert the LLG as far as it can go into the injection head on the 3D Cell Explorer-uo as shown below (Figure 6).
Injection Head
LLG headerProtective cap
Figure 5
Figure 6
23
à go back to contents
6. Start-up
Installation and setup
• Use an Allen key n°1.5 to secure the LLG in place.
Bring the 3 screws rst in mechanical contact without effort one after the other. Gently apply some force to keep the LLG in place one after the other. Please be aware that any damage made to the LLG might lead to irreversible loss of coupled power brought to the sample (Figure 7).
1
2
3
Figure 7
24
à go back to contents
8. Handling note on the LLG (Figure 8)
• The LLG is carrying the light from the CoolLed illumination system to the 3D Cell Explorer-uo. It is fragile and should be handled with care.
Avoid scratches and damages to the LLG ends (especially the silica surfaces) when connecting them to the CoolLed illumination system and the 3D Cell Explorer-uo.
Do not stretch, squeeze or bend the LLG excessively.
6. Start-up
Installation and setup
Important Note
Short-term bending radius Rshort > 40 mm / Long-term bending radius Rpermanent > 75 mm (Figure 9)
To avoid permanent damage, it is highly recommended to leave a clear and safe path for the LLG from the CoolLed
illumination system to the microscope without any strain
(Figure 10).
Adequate bending radius: about 90 mm in the picture below
Metal
Teon
Fused Silica
R
Figure 8
Figure 9
Figure 10
25
à go back to contents
6. Start-up
Installation and setup
9. Diverse connections of your CoolLed illumination system
• Plug the power connector from the DC power supply to the CoolLED illumination system as shown below. Ensure that the DC power supply
is the one supplied with the product. Using a non-CoolLED power supply may damage the illumination system and will invalidate the warranty. At this stage do not connect the main power lead to the DC power supply (Figure 11).
• Plug the CoolLED illumination system USB 2.0 cable into the USB 2.0 port on the top of the CoolLED illumination system and connect it to
one of the USB slots on your computer (Figure 11).
• Then plug the BNC cable to the rear panel of the CoolLED illumination system and to the rear panel of the 3D Cell Explorer-uo (Figure 12).
CoolLED module Power plug Flat surface
USB 2.0 cable between module and computer
Power supply
BNC cable
Not Used
Not Used
Important Note
Orientation of the power plug: Plug the power cable with the at surface facing inward (Figure 13)
Figure 11
Figure 13
Figure 12
26
à go back to contents
6. Start-up
Installation and setup
• With the CoolLED illumination system now attached to the microscope it is safe to connect the power cable. Connect the power cable
supplied to a convenient socket, plug in the IEC connector into the DC power supply and switch the power on at the socket.
Computer
USB 2.0
USB 2.0
LLG
Power supply & power cable
BNC cable
Camera / data transfer USB 3.0
Power supply & power cable
3D Cell Explorer Nanolive SA
SN: YY MM DD 00 00 XX XX XX
27
à go back to contents
6. Start-up
Installation and setup
10. Diverse connections of your 3D Cell Explorer-uo (Figure 14)
• Make sure the power switch of your 3D Cell Explorer-uo is off.
Connect the Nanolive power cable to the Nanolive power supply. Then, plug the Nanolive power supply into the power port.
Plug the Nanolive USB 3.0 cable into the USB 3.0 port on the side of the 3D Cell Explorer (the camera), tightly screw the locks and connect it to the USB 3.0 slot on your computer.
Plug the Nanolive USB 2.0 cable into the USB 2.0 port on the back of the 3D Cell Explorer-uo and connect it to one of the USB slots on your computer.
LLG
BNC cable
USB 2.0 cable
Power supply cable
Figure 14
28
à go back to contents
6. Start-up
Installation and setup
11. Download and install STEVE
• Switch on your PC and make sure it is connected to a power source.
Updated drivers: you need to have the latest Nvidia Graphics Card drivers, USB 3.0 drivers and fully updated Windows.
• Go to www.nanolive.ch/register, ll out the form to register the 3D Cell Explorer-uo and download the software.
When requested, insert the serial number (you can nd your serial number on the bottom of your 3D Cell Explorer-uo, the delivery note or
on the back of the quick start guide).
• Download and install STEVE.
• Start STEVE while still connected to the internet.
• The user’s manual and important information are downloaded during installation.
• Switch on your 3D Cell Explorer-uo and make sure it is connected to a power source.
• Connect the power supply of the CoolLED source to the grid.
Please note that you must run Steve with Nvidia graphics processor
right click on the Steve exe le
• then click on "run with graphics processor"
select "Nvidia high-performance"
Before to start exploring your cells, check our Sample Preparation Manual on how to properly prepare your samples.
29
à go back to contents
7. Handling
Correct handling of the 3D Cell Explorer-uo
The following section’s purpose is to explain how to avoid any damages to the 3D Cell Explorer-uo during moving, carrying, transportation and
repacking.
If you want to move, carry, transport or repack the 3D Cell Explorer-uo, please make sure to:
1. Unplug all cables from both devices (3D Cell Explorer-uo and CoolLED illumination system): (a) power, (b) USB2.0, (c) USB3.0 (d) BNC and (e)
light guide.
2. Secure the sample stage of the 3D Cell Explorer-uo in top position.
3. Protect the Microscope Objective with its dedicated red vinyl cap.
4. Protect the injection head with its dedicated black vinyl cap
5. Respect the holding precautions of the 3D Cell Explorer-uo.
6. Protect both ends of the LLG with the red vinyl cap.
7. Avoid stretching the LLG, forming congurations, involving sharp angels or kinks, or contact with sharp or pointed objects.
8. Transport the 3D Cell Explorer-uo and the CoolLED illumination system in their original packaging.
30
à go back to contents
7. Handling
Correct handling of the 3D Cell Explorer-uo
Instructions on how to secure the sample stage
Before moving the 3D Cell Explorer-uo, please ensure that the sample stage is blocked in top position by using the z-screw to avoid damaging the Microscope Objective as shown in the picture below:
High-risk of damages during transport Secure top position for transport with z-screw in
top position
31
à go back to contents
Place the protective cap on the MO through the sample stage aperture
Push it until it leans on the knurled part of the MO
à go back to contents
7. Handling
Correct handling of the 3D Cell Explorer-uo
Instructions on how to protect the Microscope Objective (MO) with the red vinyl protective cap
32
à go back to contents
7. Handling
Correct handling of the 3D Cell Explorer-uo
Instructions on how to protect the LLG header and how to properly take care of the LLG
Use the protective cap for both ends when the LLG is not connected to the devices (Figure 15 & 16)
The light guide is a ne, high quality optical instrument. The useful life can be prolonged by following these simple guidelines
Avoid stretching the light guide, forming congurations, involving sharp angels or kinks, or contact with sharp or pointed
objects. Light guide breakage will result in diminished light output (Figure 16).
Any inadvertent cut or puncture to the outer wall will render the light guide unsafe for use and must be taken out of service
immediately.
Avoid scratch or damage to the LLG ends (especially the silica surfaces) when connecting them to the CoolLed illumination system and the 3D Cell Explorer-uo (Figure 17).
• Clean ends with optical tissue and grade alcohol
Storage of LLG: minimum bending radius of 75 mm
Metal
Teon
Fused Silica
LLG headerProtective cap
Figure 15Figure 17
Figure 16
33
à go back to contents
Do not lift the device by its sample
stage or by the rotating arm
Secure hand positions at the places provided for
lifting, carrying, and holding
à go back to contents
7. Handling
Correct handling of the 3D Cell Explorer-uo
Instructions on how to carry the 3D Cell Explorer-uo correctly
Once all cables are unplugged and the sample stage is secured and in top position, hold/carry your device as shown below.
34
à go back to contents
8. Sample Preparation
Sample Preparation Manual:
The 3D Cell Explorer-uo allows measuring the inside of a living cell offering the researcher the possibility to acquire high resolution 3D images of a cell within seconds:
cells xed on glass coverslips or FluoroDish
• cells grown on FluoroDish™
Thanks to extremely low light exposure and compatibility with cell culture accessories, the 3D Cell Explorer-uo is suitable for long-term
live cell imaging. We have designed a Sample Preparation Manual to guide researchers through the sample preparation methods for observation.
Please download here our Sample Preparation Manual.
Note: If you are using uorescent markers to label your cells, please follow the known concepts of Sample Preparation for Fluorescence
Microscopy!
à go back to contents
35
à go back to contents
9. Cleaning Procedure for optical elements
First, inspect the optical elements to determine the location of the con­taminants. This allows you to anticipate the cleaning (typically by a swiping
movement) so that the contaminant is removed from the surface of the optical element as soon as possible (avoid dragging it around).
Optical surfaces are sensitive to scratches and must thus be cleaned carefully with the appropriate equipment. We advise to use only new clean
swabs (Texwipe™ Microdenier or Alpha series) soaked with ethanol or iso­propanol to clean the optics.
Take a fresh and clean swab soaked in ethanol or isopropanol, and with­out pressure swipe the contaminant away from the center of the optical element. Use a fresh side of the swab after each contact with the optical
surface in order to avoid re-deposition of removed contaminants.
Please clean rst the most exposed optical surfaces, with thus higher
probability of contamination:
• The MO is quite exposed to contamination and it may thus be required
to clean it regularly. (Figure 18)
• The scanning mirror of the rotating arm is not much exposed, and its
cleaning should not be required too often. (Figure 19)
• Finally, the central mirror is not exposed and its cleaning should be
exceptional. (Figure 20)
High concentrations of contaminants may require repeated cleaning.
Figure 18
à go back to contents
36
à go back to contents
9. Cleaning Procedure for optical elements
Figure 19 Figure 20
à go back to contents
37
à go back to contents à go back to contents
10. Software STEVE
ImportantNote:Fluorescentlterconguration
Fluorescence filter configuration
Before you start your acquisitions, set up STEVE with the correct uorescence lter conguration as shown below. The 3D Cell Explorer-uo is delivered with the options FITC (green), TRITC (orange) and DAPI (blue) or Cy5 (red). The conguration may be changed from the Microscope —> Options menu. (Figure 21 & 22)
Figure 22 Fluo channels conguration window
Figure 21 Microscope Options
38
à go back to contents à go back to contents
Explanation of icons
Please also refer to our video on our homepage “Introduction and getting started video for STEVE”.
10. Software STEVE
Getting started with
STEVE
STEVE i s Nanolive’s softwa re for exploring the da ta acquired using th e 3D Cell Explorer. For each recorde d frame, the 3D Cell Exp lorer measures the Ref ractive Index of yo ur cell in three dimensions . Use STEVE to digitally stain your data and interactively explore your cell and its subcomponents
2D VIEW
The 2D view show s an X-Y slice of the sample’s Ref ractive Index over laid with the user- defined stains. I n this view you can: Move t o a different slice ( or via Sl ices slider) • Zoom in or out ( Move the view w hen zoomed in (SHIF T + ).
: Measure (or stop meas uring) a user-defined p ath length. Click t his button and click o n the image to dr aw a path : Show the Refrac tive Index and the I ndex Gradient in the Pa nel
Viewer of a sele cted voxel in the 2DView • : Take a screenshot of the 2D view • / : Switch between St andard mode and Vi ewer mode.
PANEL VIEWER
The panel view er represents a 2D space of Ref ractive Index (ho rizontal axis) and Index Gradient (vertical a xis). Stains are shown as r ectangles coverin g a region in the staini ng space. Regions on the right ha ve a high Refractive I ndex, regions on the t op have a high Index Gradi ent.
In this view yo u can: Change a stain’s pos ition ( on the stain’s cente r) or shape ( on the stain’s edge) • Chan ge a stain’s transparenc y ( on the stain’s center/edge o r via the Opacity and Edge Sof tness sliders).
3D VIEW
The 3D view show s the stained data in t hree dimensions. In t his view you can rotat e ( ) • Zoom ( ) • Move (SHIFT + ) the 3D dat a.
: Switch bet ween Voxel View and Surfa ce View • : Det ach the 3D view from S TEVE’s main window • : Take a screenshot of th e 3D view • : Start or st op a live video recordi ng
of the 3D view.
BUTTONS AREA
: Visuali ze your sample in white li ght. Move the micros cope’s knobs to search an d focus your cell of interest : S ingle shot acquisit ion: acquire a single 3D fr ame • : Time-lapse
acquisitio n: acquire multiple 3D fra mes in a range of time • : Load a measurement file (.vol) • : Save a 3D or 4D mea surement file (.vol). To save a 4D file, pleas e indicate the desi red range on the t ime line using marker s and • : Open the Ex port dialog.
SLIDER S AREA
Overlay : Fuzzy switch bet ween stained dat a and unaltered Refrac tive Index. Slices: Se lects the X-Y slice from t he 3D data stack. Opacity : Chooses the opa city for the curre ntly selected stain . Edge Softn ess: Use this to make a st ain’s edges soft and avoid s olarized images in 2D.
STAIN COLORI NG AREA (PANE L)
Load Panel: L oad a previously save d panel (.xml format ) Save Panel: S ave the current pan el (stains) Rescale: Mo dify the viewin g range of the panel vie wer to fit the current ti me frame.
STAIN PICKER
: New stain - choose a n ame (not required) and a c olor for the stain. Ne xt, begin painting your region of i nterest in the 2D view ( ). A rectangular st ain is computed based o n the
painted par t of the image and is show n in the panel viewer.
/ : Hide or show a st ain • : Delete a stain • : Show t he stain properti es and allow
background definition.
3D CONTROL A REA
: Change the ba ckground color for the 3D v iew. Hide Axes & Cu be: Show/hide the coo rdinate axes and the ou tline cube in the 3D vie w. Show Capti on: Show/hide the l abels in the 3D view.
Cropping sph ere: Cut away part of the 3D v iew. Move the plane thr ough the sphere to crop the 3D data. T he plane can be move d (SHIFT + ) or rotat ed ( ).
TRANSPORT CONTROL
Use this to exp lore your data in time. T he transport displ ay shows the frame n umber and acquisitio n time of the current f rame.
: Jump to t he beginning of a data s et : Set marker A to th e current cursor po sition • : Play/paus e playback • : S et marker B to the current c ursor position • : Jump to the en d
of a data set • : Set playb ack speed. (Note: Ma rkers A and B select a tim e range for saving / exporting purposes.)
TIME LINE AREA
Displays t he current time fram e ( ) and the select ed range ( ). Click on the t op part of the area to jump t o the specified fram e. Drag on the tick regi on to set
markers A and B .
August 2015
1x
Getting s tarted wi th Nanoliv es softwar e STEVE v 1.0
39
à go back to contents à go back to contents
10. Software STEVE
Interface Fluorescent
2D View
The 2D view shows a X-Y slice of the sample's Refractive Index overlaid with user-dened digital stains and uochromes resulting in a multi-channel image.
Transport Control
Buttons Area
Sliders Area
Digital Stain Colouring Area
Detailed Frame Dialogue Window
Digital Stain Picker
Time Line Area
3D Control Area
Frame Intervals
Fluorescence Sliders Area
Panel Viewer
The Panel Viewer represents a 2D space of Refractive Index and Index Gradient.
Multi-channel images are not represented here.
3D View
The 3D view shows the stained data in three dimensions.
40
à go back to contents à go back to contents
10. Software STEVE
Preparation for a measurement
Note!
In order to start your measurements, verify that all the cables are properly plugged in, the 3D Cell Explorer-uo and the CoolLED illumination system are turned on and connected to your computer and STEVE is running.
Operate and control the 3D Cell Explorer-uo and the CoolLED illumination system easily with STEVE if you wish to do the following:
• Fully automated microscope calibration
Single-frame image acquisition (single 3D reconstruction)
Time-lapse acquisition (multiple 3D reconstructions in a range of time)
• Localization of organelles at different cell depths
Co-localization of uorescence (3 channels) & Digital Stains (7 channels at single acquisition)
Sample screening using auxiliary bright eld mode (white light)
Data Storage
An acquisition generates a lot of data. It is thus necessary to store it on the control PC hard drive. The location of this acquisition buffer may be changed from the Microscope > Options menu. By default, it points to a location on your system drive (C:). You may choose another location with more available space to avoid acquisitions being interrupted due to insucient space.
Fully automated microscope calibration
With the 3D Cell Explorer-uo, there is no need for manual calibration. After positioning your sample, the microscope will instantly self-adjust so you can be sure you are getting the best possible images of your cells. You can always check the calibration progress as it is displayed on STEVE.
After calibration is completed the cell image is displayed on the left panel of STEVE (2D view).
Step-by-step guide on how to prepare for a measurement
Take the sample previously prepared (for details on how to prepare the sample, please refer to the “Sample Preparation Protocol”) and place it on
the 3D Cell Explorer-uo stage.
Open STEVE and click on the white light button. On the left panel of STEVE you can visualize your cells in the white light mode. Use the XY and Z
knobs of the 3D Cell Explorer-uo to search for the cell of interest and focus on it.
Now you can choose to either do a single-frame image acquisition or a time-lapse acquisition with your desired uorescent channel choice.
41
à go back to contents
10. Software STEVE
Selecting the mounting medium‘s refractive index
How to select the mounting medium’s refractive index
Before either a single-shot or a time-lapse acquisition is started, a dialog will appear in STEVE (Figure 23) prompting you to select the mounting medium refractive index (e.g. DMEM, Figure 24).
You may select the mounting medium from the drop-down list of commonly used media.
If yours is not listed you may add it to the list by clicking “New” and entering the new mounting medium refractive index and its name (Figure 25). In case you are using an aqueous mounting medium not found in the list for which you do not know the refractive index, it is safe to assume it is close to water. If using oil mounting media however, it is important to specify a correct refractive index as it can be signicantly larger than of
aqueous solutions.
Figure 23 Mounting medium Refractive Index
Selection Window
Figure 24 Example Mounting Medium Water
Figure 25 „New“
42
à go back to contents
10. Software STEVE
Single-frame image acquisition
How to do a single-frame image acquisition
A single frame image acquisition is an instantaneous measurement that results in a unique image of the object.
To perform a single shot, click on the 3D button or on the Single Shot from the Microscope menu. The mounting medium refractive index selection dialog will appear (see section "How to select the mounting medium’s refractive index" previously explained) and the acquisition will start once a refraction index is selected (Figure 26).
At the end of the acquisition, STEVE will enable to digitally stain the data and interactively explore the cell and its subcomponents. Further down, we explain the exact steps on how to achieve this. Please refer to “How to digitally stain cells with STEVE”.
Figure 26 Acquisition type dialogue window
43
à go back to contents
10. Software STEVE
Acquiringuorescenceimages
Multi-channel image
A multi-channel image combines a series of monochrome images into one image. The multi-channel image generally shows a sample that has
been stained with numerous different uorochromes. The multi-channel image is a combination of the single uorescence images. You can have the individual uorescence images displayed separately (showing, e.g., the plasma membrane or mitochondria) or also as a superimposition of all of the uorescence images (overview) (Figure 27).
Refractive Index (RI)
Nucleus Nucleoli Overview
Plasma membrane Mitochondria
Figure 27 Single uorescence images and overview uorescence image
44
à go back to contents
10. Software STEVE
Acquiringuorescenceimages
How to create multi-channel images
To perform a single shot, click on the 3D button or on the Single Shot from the Microscope menu.
A selection dialog will appear to choose the acquisition type which can be individually set up (Figure 28). Once a refractive index and the uorescence channels are selected, a new dialogue will appear to adjust the uorescence channel parameters (Figure 29). Adjust the uorescence parameters as per your needs. Then click on „ok“. Your acquisition will start.
Figure 29 Dialogue window uorescence parameters
Figure 28 Acquisition type dialogue window
45
à go back to contents
10. Software STEVE
Acquiringuorescenceimages
TIP
To be able to acquire good uorescence images, follow our guidelines hereafter.
1. At the time the dialogue appears you are able to adjust the uorescence channel parameters, in order to acquire the best image possible
(Figure 30).
Live-image of the selected channel
Tab for setting-up each of the selected channels
Tab for activating the white-light mode and perform bright-eld of the sample
Exposure setting of the camera in ms
Gain setting of the camera in percentage: it is recommended not to go beyond 40% to acquire a
high-quality image.
Intensity setting of the LED source in percentage: it is recommended to operate at about 10%
for the sample screening to limit the sample photobleaching
Figure 30 Dialogue window uorescence parameters
46
à go back to contents
10. Software STEVE
Acquiringuorescenceimages
TIP
For the best image quality, keep the intensity as low as possible in order to avoid bleaching while you search for a suitable uorescent cell.
You will need to maximize the gain to be able to see something. At the same time, the exposure time should be typically below 200 ms, to
have enough frame refreshing rate and screen comfortably the sample (Figure 31).
2. Now you are ready to focus on the uorescent image (use the Z knob for that).
Figure 31 Dialogue window uorescence parameters
47
à go back to contents
10. Software STEVE
Acquiringuorescenceimages
TIP
3. To acquire an image without background, adjust the intensity parameters. The intensity should stay low (<20-30) (again to minimize photobleaching). Now, reduce the gain until the background disappears. The maximum gain to be used can be up to 40%. Increase the exposure time if you need to see more details of the image but try to keep it as low as possible (100 ms or lower = very good; 200 ms = good) (Figure 32).
Figure 32 Dialogue window uorescence parameters
48
à go back to contents
10. Software STEVE
Displaying multi-channel images
When you visualize a multi-channel image, the overlay parameter must be set to 0 (Figure 33). The uorescence control area in the lower right part of your software window provides access to all of the uorescent channels (Figure 34).
You can have the individual uorescence image displayed separately or as a complete overview of all the uorescence images (refer to Figure
34); to do so use the navigation bars in the uorescence control area.
Move a color bar to make the channel appear or disappear. All of the color channels that are at the moment displayed on your monitor will be
recognized through the color bars (0 %to 100% transparency) (Figure 34).
Figure 34 Fluorescence control area
Figure 33 Overlay slider
49
à go back to contents
10. Software STEVE
Time-lapse acquisitions
How to do a time-lapse acquisition
In a time-lapse acquisition, multiple 3D reconstructions in a range of time are attained.
To perform a time lapse, click on the 4D button or on Time Lapse in the Microscope menu. The mounting medium refractive index selection dialog will appear (see section "How to select the mounting medium’s refractive index" previously explained). Once a refractive index is selected, a new dialog will open to specify the speed and the duration of the acquisition (Figure 35 green frame).
The time lapse can be performed either at the maximum speed or by setting the time interval between two consecutive frames.
The duration of the time lapse acquisition can be determined either by stopping it manually (by clicking on the Stop Acquiring button) or by setting
the ending time for it to stop automatically.
At the end of the acquisition, STEVE will enable to digitally stain the data and interactively explore the cell and its subcomponents. Further down, we explain the exact steps on how to achieve this. Please refer to “How to digitally stain cells with STEVE”.
Figure 35 time lapse setting window
50
à go back to contents
10. Software STEVE
Fluorescence time-lapse acquisition
Acquiring fluorescence time-lapses
To perform a time lapse, click on the 4D button or on Time Lapse in the Microscope menu. A selection dialog will appear to specify the speed and
the duration of the acquisition, which can be individually set up (Figure 36). Once the settings for the time lapse, the refractive index and the uo­rescence are selected, a new dialogue will appear to adjust the uorescence channel parameters (Figure 37). Adjust the uorescence parameters as per your needs (Figure xy9). Then click on „ok“. Your acquisition will start.
Please note: in case of maximal speed during a time-lapse acquisition, the fluorescence channels are disabled!
Figure 36 Time lapse setting window
Figure 37 Dialogue window uorescence
parameters
51
à go back to contents
10. Software STEVE
Time Line Area
While the 3D Cell Explorer-uo acquires multiple 3D reconstructions to create a time-lapse, you will be able to follow the exact chosen uorescent and refractive index parameters for each frame in the lower part of the STEVE interface (Figure 38).
Figure 38 Time Line Area
52
à go back to contents
10. Software STEVE
Digitally staining
How to digitally stain cells with STEVE
1. Choose a meaningful slice on the 2D visualization panel (left side).
Dragging the mouse up to down (left click pushed) on the 2D visualization panel or by moving the Slices slider.
2. Pick a new stain.
Click on the + button.
3. Choose a name (optional).
Click on the black-gray line edit form.
Write down the name in the line text input.
4. Choose a color.
Click on your desired color.
Option: drag the mouse to have a more precise vision of the color that you could select.
5. Go to the 2D visualization panel and draw.
Click and/or drag the mouse on your desired region of interest. The pixels under your cursor are glowing in the panel view. When you
release the mouse, we compute the stain from your selected pixels.
Option: You could zoom in on the 2D visualization panel by wheeling the mouse.
6. Now the stain is represented in the panel view and superimposed on the 2D visualization panel.
Option: you could change the weight of the colored image on the 2D visualization panel by moving the Overlay slider.
53
à go back to contents
10. Software STEVE
Digitally staining
7. How to manipulate your stain:
Change the opacity (Opacity slider or drag up-down with the right click pushed).
Change the edge softness (Edge softness slider or drag up-down with the right click pushed on a stain edge).
Move a stain in the panel view (click on the stain and drag it on the desired position in the refractive index – index gradient space).
Change the shape of the stain (select and edge or a corner and drag it).
8. To add other stains repeat the above steps
9. Stain representation: opacity, edge softness, stain edge /rectangle, panel view axes
10. Stains options:
Enable/Disable the visualization of the stain
Delete the stain
Change the color (by double clicking the colored button or by clicking on the palette icon)
Note! For more in-depth information about digital staining, please refer to our best staining practices on our support online page.
54
à go back to contents
10. Software STEVE
Data Saving
How to save your data
The data can be saved by clicking on the Save icon (Figure 39). The holograms are kept by default within the .vol le.
Figure 39 Saving Window
55
à go back to contents
10. Software STEVE
Data Saving
STEVE allows to save data as a 3D or a 4D .vol file (readable only with STEVE).
Alternatively, it is possible to export the data in other le formats by opening the export dialog shown in the Figure 40.
The data can be exported by choosing the following parameters:
Single acquisition or Time-lapse. For time-lapse, it is possible to export all frames or one every 2, 3, etc.
the type of data: RI or uorescence channel
the format of the le: Raw, Tiff, Obj, Png/Jpeg (available depending on the above selections).
Figure 40 Export data setting
56
à go back to contents
10. Software STEVE
Data Annotation
How to work with data annotation
For each measurement you can visualize and modify the measurement properties, called also data annotation. In the File Menu under “Properties” open the annotation dialog (Figure 41). All the visible properties, marked with “Eye icon” will be added to the current measurement after the button “Apply” is pushed. The invisible properties, marked with “Close Eye icon” will not be saved. The rst property entries, such as: “X size”, “X resolu­tion”, “Acquisition Time”, “Microscope”,etc. are not editable, you cannot modify them, but they will always be saved on the measurement.
You can add new entries by clicking on the “Add Entry” button or you can delete the selected entry by clicking on “Delete Entry”. By deleting an
entry, that entry will never be shown again on the annotation dialog, you will have deleted it permanently. Also, you can edit each entry by clicking
on the “Edit icon” button on the right side of the entry.
Figure 41 Export data setting
57
à go back to contents
10. Software STEVE
Software Updates
How to update the software
Steve will automatically update when a new version is available. The user will be alerted to close Steve and the Steve Maintenance Tool will start. In the Steve Maintenance Tool choose “Update components” and click the “Next” button. Follow the messages provided by the update dialog.
58
à go back to contents
11. Troubleshooting
Calibration
Calibration failed err.1.1
If applicable, reduce the quantity of liquid above your sample (-> Check our Sample Preparation Protocol in Section 5). Check that no opaque object
can intersect the laser beam while the arm is rotating. If using a dish, check that you are imaging close to its center such that the laser beam cannot intersect the wall of the dish instead of the top.
Calibration failed err.1.4
Sample may be too absorbent. Try to move in a different area and re-do the acquisition. If the error persists, please get in touch with our customer service.
Acquisition
Acquisition failure err.2.1
If applicable, try to move to another position within your sample. If the error persists try to clean the microscope objective and check that you com­ply with the sample preparation protocol ( -> Check our Sample Preparation Protocol).
Connection
Connection failure err.3.1
If STEVE cannot establish communication with the 3D Cell Explorer, please verify that the backside USB 2.0 is correctly plugged in to the 3D Cell
Explorer and to your computer, that the microscope has electrical power and that it is switched on. If the error persists, unplug all cables, reboot
your computer and plug the cables back (check Start-up - Section 4).
Connection failure err.3.2
If STEVE cannot establish communication with the 3D Cell Explorer, please verify that the USB 3.0 cable coming from the camera is correctly
screwed in and that it is properly connected to your computer’s USB 3.0 port.
59
à go back to contents
11. Troubleshooting
Connection failure err.3.3
STEVE cannot establish communication with the 3D Cell Explorer. The port to which the USB 3.0 cable (coming from the camera) was connected to was recognized as USB 2.0. This cable must be connected to a USB 3.0 port. If you have correctly connected the USB 2.0 and USB 3.0 cables and the error persists, please verify that your USB 3.0 adapter drivers are properly installed or contact your system administrator.
Connection failure err.3.4
STEVE cannot establish communication with the 3D Cell Explorer and the microscope cannot initialize itself. Please make sure the rotating arm is
not obstructed in any way. Try to switch it off and on again. If the error persists observe whether the arm rotates when you switch the microscope on. If this is not the case, please contact our customer service.
Connection failure err.3.6
The microscope needs maintenance. Please contact your distributor if applicable or our customer support.
Connection failure err.3.7
The connection to the uorescence module failed. Please make sure that the uorescence module is powered and that the USB cable from the
module is connected to a USB port on your computer.
Connection failure err.3.8
Microscope rmware update error. Your microscope likely needs maintenance. Please contact our customer support.
Processing
Processing error err.5.1
The STEVE processing tool chain encountered a fatal error. Please make sure your computer meets the STEVE hardware requirements. In particular,
make sure your graphical card is from the NVIDIA brand with at least 2GB of dedicated GPU memory. Please upgrade your graphical drivers using
the NVIDIA website at www.nvidia.com/Download/index.aspx. If the error persists, please contact our customer service including error detailed information displayed in the error popup.
60
à go back to contents
12. Compliance
WEEE
All qualifying products that are subject to the WEEE Directive and supplied by Nanolive are compliant with the WEEE marking requirements.
Such products are marked with the "crossed out wheelie bin" WEEE symbol and in accordance with European Standard EN 50419.
RoHS
Based on information obtained from our component suppliers, this
statement certies that ALL products manufactured and supplied by CoolLED Ltd are in compliance with Directive 2011/65/EU of the European Parliament and of the Council of 8 June 2011 on the restriction
of the use of certain hazardous substances in electrical and electronic
equipment (also known as “RoHS”).